Infusion Pump HP-60 Service Manual: Medcaptain Medical Technology Co., LTD
Infusion Pump HP-60 Service Manual: Medcaptain Medical Technology Co., LTD
Service Manual
Before maintaining the HP-60 infusion pump, please read this Manual carefully and
follow the safety precautions and maintenance instructions contained herein.
Statement
Version Information
V1.0
Contents
STATEMENT ....................................................................................................................................... 2
4 HARDWARE ........................................................................................................................ 18
6 MAINTENANCE .................................................................................................................. 28
8 CALIBRATION .................................................................................................................... 34
     1.4     DISMANTLING THE PERISTALSIS PUMP MOTOR SPEED MEASURING PCBA AND PUMP FINGER POSITION
     DETECTION PCBA .................................................................................................................................. 55
     1.5     DISMANTLING THE SLIDING BLOCK AND HP-60 DOOR POTENTIOMETER ASSEMBLY ................................... 57
     1.6     DISMANTLING THE HP-60 TUBE DETECTION PCBA, ANTI-FREE-FLOW CLAMP PCBA, PATIENT SIDE OCCLUSION
     PRESSURE DETECTION PCBA AND ITS CONNECTION FFC ................................................................................ 59
     1.7     DISMANTLING THE PERISTALSIS PUMP ............................................................................................... 60
     1.8     DISMANTLING THE PERISTALSIS PUMP MOTOR ................................................................................... 61
     1.9     DISMANTLING THE DOOR DRIVE ASSEMBLY ........................................................................................ 63
     1.10         DISMANTLING THE HP-60 FLUID SIDE OCCLUSION PRESSURE DETECTION PCBA .................................. 65
     1.11         DISMANTLING THE SEPARATE AIR BUBBLE SENSOR MODULE A ......................................................... 65
     1.12         DISMANTLING THE HP-60 PUMP DOOR PCBA .............................................................................. 66
     1.13         DISMANTLING THE SEPARATE AIR BUBBLE SENSOR MODULE B.......................................................... 67
     1.14         DISMANTLING THE HP-60 VENEER .............................................................................................. 68
     1.15         DISMANTLING THE HP-60 PUMP DOOR BOARD’S CONNECTION FFC ................................................. 69
     1.16         DISMANTLING THE HP-60 MAIN PCBA ....................................................................................... 70
     1.17         DISMANTLING THE HP-60 DROP SENSOR INTERFACE PCBA ............................................................. 72
     1.18         DISMANTLING THE FASTENER ASSEMBLY ....................................................................................... 73
     1.19         ASSEMBLY............................................................................................................................... 74
1 Important Information
- The TSC requires that the instructions in the manuals are observed
- Depending on the unit type, the Service Program must be called which may lead to a dangerous
unit condition in case of inappropriate operation. Furthermore, a special service connector may be
necessary.
     Maintenance in violation of the requirements in this service manual may seriously damage
      the device.
     The service personnel must be trained and permitted in writing by MEDCAPTAIN to do so.
     The maintenance must be performed under electrostatic discharge (ESD) protective
      conditions. Do not touch PCBA or semiconductors by hand without any protective measure.
     Do not touch the display by using sharp objects. Otherwise, the display may be damaged.
     Do not autoclave the infusion pump.
     Before using the internal battery, check the battery to ensure that sufficient power is available.
      Recharge the battery if required.
     Liquid intrusion into the AC power socket or nurse call socket may cause short-circuit.
      Before connecting the power cable, check if the connecting parts are dry. If liquid has spilled
      onto the infusion pump, clean the pump with a dry cloth and then perform an inspection and
      maintenance.
     Use the maintenance parts and accessories provided by MEDCAPTAIN for replacement and
      maintenance.
     Do not maintain or use the infusion pump in a flammable environment.
     High-frequency surgical equipment, mobile phones, wireless devices, and defibrillators may
      cause interference on the infusion pump. Therefore, keep the infusion pump away from these
      devices when using the pump.
     After maintenance, perform a safety test and clean the pump according to this manual.
1.5    Quality Control
      MEDCAPTAIN meets the requirements for quality management system specified in
ISO9001 and ISO13485 standards. Complying with the requirements of the medical device
directive (MDD), the product mentioned in this manual has also passed CE certification.
1.6    Safety Warnings
           WARNING:
     The power supply of this device may reach up to 100-240V. Maintenance in violation of the
      requirements in this service manual may cause electric shock, serious injury, and even death.
     Maintenance in violation of the requirements in this service manual may seriously damage
      the device.
1.7    Acronyms and Abbreviations
ESD                            Electro-static discharge
PCBA                           Printed Circuit Board Assembly
ISO                            International Standardization Organization
CE                             Council of Europe
MDD                            Medical Device Directive
Important Information
                                                                     Non-ionizing
            Nurse caller
                                                                     electromagnetic radiation
                                                                     DISPOSAL:        Do     not
                                                                     dispose of this product as
            Refer to instruction                                     unsorted municipal waste.
            manual/booklet                                           Separate collection of such
                                                                     waste for special treatment
                                                                     is necessary.
            Atmospheric             pressure
                                                                     Temperature limit
            limitation
             CE Mark: conforms to
             essential requirements of the
                                                  Class I      Class-I equipment
             Medical Device Directive
             93/42/EEC
             HOME button. Press this
             button to access the setting                      ON/OFF button. Press this
             interface or return to the                        button to power on/off the
             infusion            preparation                   pump.
             interface.
      If you have any question when using the infusion pump, please contact local
      distributor or directly contact us at any time.
      Address: 12th Floor, Baiwang Research Building, No.5158 Shahe West Road,
      Xili, Nanshan, 518055 Shenzhen, Guangdong, PEOPLE'S REPUBLIC OF
      CHINA
Fax: +86-755-26001651
Website: http://www.medcaptain.com/en/
E-mail: MC.Service@medcaptain.com
MEDCAPTAIN and its distributors all over the world have established after-sales service agencies
to effectively solve your problems in time.
System Overview
2 System Overview
      This product is intended to be used in conjunction with the IV set to control the dose
      of liquid infused into the patient’s body in clinical departments.
2.2    Composition
      Main unit + handle + pole clamp + power cable + nurse caller (optional) + drop sensor
(optional)
2.3    Appearance
    a) Front view
7 – Alarm indicator
System Overview
clamp block
pressure sensor
pressure sensor
c) Rear view
Accessories
Table 4-2 List of accessories
Accessory                     Description                Part Number
Pole clamp
                                                         1202-00003-01
Handle
                                                         1404-00105-01
AC power cable                European standard, 240V
                                                         1462-00004-01
                              AC 10A
3 Product Specifications
Model HP-60
3. IP23
Battery operating time is not smaller than 10 hours if the infusion rate
is 25 ml/h.
Key backlight
Interfaces Micro USB 2.0: enables connection to the external drop sensor.
USB 3.0: enables connection to a USB flash drive for software and
When infusion rate is not smaller than 1 ml/h: infusion accuracy ≤ ±3%
                       Dedicated IV set:
System Overview
Occlusion Level Patient side occlusion: 150 to 975 mmHg, 12 occlusion levels available
(default: level 6)
(default: level 5)
7 detection sensitivity levels: 25, 50, 100, 200, 300, 500, and 800 (ul)
Applicable IV Set Various brands of 20d/ml and 60d/ml IV sets conforming to relevant
standards
Infusion Modes Rate mode, time mode, weight mode, sequence mode, trapezia mode,
loading dose mode, drip mode, micro mode, and relay mode (used
Drug Library A maximum of 5000 drug types can be stored (default: 60 preset drug
types).
High-level alarms:
Infusion End, BAT Empty, Patient Side OCCL, Infusion End KVO
Start, KVO End, Relay Failed, Fluid Side OCCL, Air Bubble In Line,
No Drop, Too Many Drops, Too Few Drops, Drip Chamber Overflow,
Middle-level alarms:
Low-level alarms:
Door Failure
Special Functions      Repeated alarming: After the sound of an alarm is muted, this alarm is
System Overview
playback.
barcode scanning.
power failure.
Text size adjustment: Three text size levels are available for selection.
Wireless Networking The wireless networking function enables the infusion pump to be
4 Hardware
                      Interface
     Drop                                        J3                       J25                               J11
                      board for
    sensor
                     drop sensor
                                           J13
                                                   Motor speed
                                                                        Battery        Speaker
                                                 detection board
                                                                Figure 4
Hardware Description
4.2 Power Board
a. Circuit description:
    The AC/DC power board adopts a medical-grade switching power supply that features low leakage current
and high isolation voltage and meets the EMC requirements. The input AC voltage range of the AC/DC power
board is 100-240V, 50/60Hz. The output single DC power supply is 15.0V, and the maximum output current is
2A.
b. Main test point:
 No.      Test              Name                 Range              Unit          Remarks
          Point
1       TB2         DC output voltage      14.0-16.0             V            /
c. Circuit board socket numbering and definition
     The maximum voltage of the AC/DC power board may reach up to about 400 V. The workbench must be
      insulated to avoid touch of the high-voltage circuit part. High voltage may cause electric shock and
      personal injury.
     When performing a test with a multimeter or oscilloscope, note that the test probe must not short-circuit the
      live part on the circuit board to avoid damage of the circuit board.
                                                         19
Hardware Description
            hospital number. The USB3.0 interface is used to connect to the nurse call system for implementing
            the nurse call function.
    (4)     Audible alarm: The beeper and loudspeaker alarm circuit is used to implement the audible alarm
            function.
    (5)     Motor drive control: The motor drive chip manufactured by TRINAMIC is used, which supports
            24V/2A drive output. In addition, the stepper motor supports a subdivision of up to 256, ensuring
            infusion accuracy and stability at low infusion rate. The SPI communication interface is used for
            connection to M0. Motor start/stop, motor rotation speed, and motor acceleration/deceleration are
            all controlled by SPI signals.
    (6)     Air bubble detection circuit: Air bubble drive and monitoring circuit is provided to monitor whether
            air bubbles exist in the infusion tube.
    (7)     Fluid side/patient side occlusion pressure detection circuit: The fluid side/patient side occlusion
            pressure detection board is connected to test the fluid side/patient side occlusion pressure.
    (8)     WiFi function: WiFi circuit and protocol are used for communication with the nurse station,
            implementing output of infusion data.
    (9)     Wired network function: Wired network is used for communication with the nurse station,
            implementing output of infusion data.
b. Main test point:
 No.      Test              Name                    Range                Unit           Remarks
         Point
1       TP15       24V       DC      output 22-26                   V                /
                   voltage
2       TP25       5V DC output voltage 4.9-5.1                     V                /
3       TP31       3.3V DC voltage            3.2-3.4               V                /
a. Circuit description: Two optical coupling circuits are used to detect the motor rotation speed. The rotation
    speed and rotation direction can be determined based on signal existence/absence and sequence of optical
    couplers A and B.
b. Main test point: None
c. Circuit board socket numbering and definition
 No.    Socket       PIN                    Name                           Description
                  Quantity
1      J13        6             Socket for connection to J12 /
                                of HP-60 main board
2      J2         3             Socket for connection to J2 /
                                of HP-60 pump finger
                                position detection board
Note: The sliding resistor is a very sensitive and precise device. When disassembling or assembling the resistor,
be careful not to touch peripheral components and not to pull the connection cable.
                                                        24
Sensor Information
5     Sensor Information
     This chapter only describes the software UIs related to servicing and test. For the description of other
software functions, see the corresponding operation manual.
      c) Pump finger signal: When the pump finger board optocoupler is shielded, the status displayed is 1.
           Otherwise, the status displayed is 0. If the display status is incorrect, check the pump finger
           optocoupler and related connection cable.
      d)   Air bubble detection: When no tube is installed or the tube is empty, a value in the range of 0-100 is
           displayed. When a full tube is installed, a value larger than 1000 is displayed. Otherwise, check
           whether the tube is installed in position and whether the air bubble sensor and detection board are
           faulty.
      e)   Dedicated consumable: Check whether the consumable used is a dedicated consumable. If yes, the
           status displayed is 1. Otherwise, the status displayed is 0.
      f)   AC test: Check whether a value in the range of 14V-16V is displayed when the pump is connected to
           AC power supply and a value smaller than 2V is displayed when the pump is not connected to AC
           power supply. If the displayed value is incorrect, the device generates the ERR007 alarm, which may
           be caused by a power module failure.
                                                        25
Sensor Information
  g) DC test: This value varies with the input DC input voltages. Connect the infusion pump to the DC
       power supply and check whether the test value falls within the range of 14-16V. Then, disconnect the
       pump from the DC power supply and check whether the test value is about 6V. If the voltage is
       incorrect, the power module may be faulty.
  h)   Battery test: Install the battery in the infusion pump and check whether the test value is stable and falls
       within the range of 9-13V. If the test value keeps changing, the battery may be damaged or the main
       board is faulty.
  i)   VDD test: The normal value is 3.3±0.2V. If the test value is abnormal, the main board may be faulty.
  j)   VAA test: The normal value is 24±2V. If the test value is abnormal, the main board may be faulty.
k) VCC test: The normal value is 3.3±0.2V. If the test value is abnormal, the main board may be faulty.
                                                      26
Sensor Information
  l) Fluid/patient side pressure test: Press the pressure sensor to check whether the pressure value changes.
     If the pressure value does not change, the pressure sensor may be faulty.
  m) Pump door Hall signal: Check whether 1 is displayed when the door is open and 0 is displayed when
     the door is closed. If the displayed value is incorrect, check whether the electromagnet is properly
     installed and whether the Hall switch functions properly.
  n) Number of drops: Check whether the actual number of drops is consistent with that detected by the
     drop sensor with naked eyes. If there is a large difference between them, check whether the drop
     sensor is properly installed and whether the drop sensor is faulty.
                                                    27
Maintenance
6       Maintenance
6.1      Overview
      To ensure long-term steady operation of the infusion pump, service personnel must check, maintain, and
test the pump at regular intervals. This chapter describes the basic test method of the infusion pump and
recommends appropriate test frequency and test tools. Service personnel must select appropriate test tools to
maintain and test the infusion pump based on actual requirements.
      The test and inspection methods provided in this chapter are used to verify that the performance of the
infusion pump meets the specification requirements. During the test, if the test results do not meet requirements,
a certain functional module of the infusion pump is faulty. In this case, immediate maintenance or replacement
must be performed. In case of any other doubts, please contact the after-sales service department of Medcaptain
in time.
Caution
3. Switch on unit.
5. Remove battery.
                                                          28
Maintenance
When the battery always gives a low battery indication or the battery is not installed correctly, a
     battery check is needed.
 Use a screwdriver to loosen the battery cover fixing screws at the bottom of the pump.
        CAUTION:
Remove the battery if the infusion pump is not likely to be used for some time
WARNING:
     l The battery’s replacement must be done by specialist who has been trained to finish such operation.
     Otherwise there will be a risk of danger.
     l Please strictly follow the instruction to replace the battery, and the battery should be provided by the
     manufacturer. Otherwise there will be a risk of danger.
      Do not disassemble, short circuit or throw the battery into fire in case of the danger caused by linkage or
     explosion.
                                                         29
Maintenance
                                                      30
Cleaning and Disinfection
7     Cleaning and Disinfection
      It is highly recommended that the materials and methods listed in this chapter be used for
cleaning and disinfection of the device. If other materials or methods are used, the device may be
damaged or its service life may be shortened.
           CAUTION:
 In case of any doubts about the use of the detergent or disinfectant, please consult the local
      distributor.
 Please dispose of the wastes generated after the cleaning and disinfection according to the
      relevant regulations of the local hospital.
    7.1     Preparations
      1. Before the cleaning and disinfection, disconnect the device from the patient.
      2. Power off the device and disconnect the device from the AC or DC power supply.
      3. Remove the infusion consumable and accessories (for example, drop sensor and barcode
scanner) connected to the pump.
      4. Wear a pair of rubber gloves and a gauze mask to prevent contaminants from splashing
onto your skin during the cleaning and disinfection.
      5. You are not allowed to disassemble this device for cleaning and disinfection. To
disassemble this device for further cleaning and disinfection, please contact the local distributor.
      6. Prepare several pieces of soft medical gauze, a detergent container, and a disinfectant
container.
7.2 Cleaning
           WARNING:
          Do not immerse the device in the detergent solution.
          Prevent the solution from seeping into the device.
          Do not use organic solvent, halogenated solvent, petroleum-based solvent, glass detergent,
              acetone, or other irritant detergents.
          Only manual cleaning is allowed to be adopted for this device. Do not adopt the automatic
              cleaning mode for this device.
Cleaning procedure:
                                                       31
Cleaning and Disinfection
       1. Completely immerse a piece of soft medical gauze in neutral or slightly alkaline detergent
        solution, wring out the gauze, and then use the gauze to wipe the device surface.
       2. Wipe all the surfaces of the device in sequence until all the contaminants are removed
        from the device surface.
       3. Ensure that all the edges and corners of the device are completely cleaned.
       4. After the cleaning, use a piece of dry medical gauze to remove the residual detergent
        solution.
    The following table lists the detergents recommended for the device.
Table 6-1 recommended detergents
7.3 Disinfection
        WARNING:
       Do not immerse the device in the disinfectant solution.
       Prevent the solution from seeping into the device.
       Use the disinfectant according to its operation manual.
       Do not autoclave the device.
       Only manual disinfection is allowed to be adopted for this device. Do not adopt the automatic
           disinfection mode for this device.
    Disinfection procedure:
       1. Before the disinfection, clean the device according to the method provided in section 9.2.
       2. Completely immerse a piece of soft medical gauze in the intermediate-efficiency or
        high-efficiency disinfectant solution, wring out the gauze, and then use the gauze to wipe
        the device surface.
       3. Wipe all the surfaces of the device in sequence. For the contact time of the disinfectant, see
        the operation manual of the disinfectant.
       4. Ensure that all the edges and corners of the device are completely disinfected.
                                                    32
Cleaning and Disinfection
       5. After the disinfection, immerse another piece of soft medical gauze in clean water, wring
        out the gauze, and then use the gauze to wipe the device surface for removing the residual
        disinfectant solution.
       The following table lists the disinfectants recommended for the device and the required
contact time for the disinfection.
        WARNING:
       Do not dry the device by using a drying machine or similar products.
       Connect the device to the power supply again after the device is completely dry.
       1. After cleaning and disinfection, place the device in a shady, cool, and ventilated
environment for air drying.
       2. If you are not going to use the device soon after air drying, place the device in its original
package for storage and transportation.
                                                     33
Calibration
8 Calibration
                                                        34
Calibration
(Schematic diagram)
                    35
Calibration
                                                 (Picture of real product)
                                                          36
Calibration
                                                        37
Calibration
150 225 300 375 450 525 600 675 750 825 900 975
     3      Manually input the rates of the 18 points of a known IV set, fluid side occlusion pressure, and level 11
            patient side occlusion pressure.
                                                          39
Calibration
    2    Select a brand of consumable to be calibrated (if a desired consumable does not exist, execute method
         1 to add this consumable).
                                                      40
Calibration
3 Tap Purge to remove the air bubbles and fill the entire infusion tube with liquid.
4 Tap to access the next page, and tap IV Set Cali. to access the IV set calibration page.
                                                     41
Calibration
                                                    42
Alarms and Troubleshooting
9      Alarms and Troubleshooting
Warnings:
 During troubleshooting by service personnel, there is a 100-240V AC high voltage in the HP-60 and
       the switching power supply circuit may generate a 400V DC high voltage. An accidental touch may
       cause serious personal injury. Please follow the requirements in related regulations and ensure that
       the maintenance is performed by professional qualified personnel.
      When a device is used to measure a circuit board or other parts, short circuit or incorrect
       measurement method may cause serious personal injury or serious device damage.
   The device safety and functions must be verified after being disassembled for maintenance and then
    assembled each time. For details, see chapter 7.
Basic tools for maintenance and inspection:
 PC (used for downloading software and filling table)
 Software downloading connector
 Digital multimeter
 20-50 ml measuring cylinder (accuracy: 0.2 ml) or electronic balance
 Time keeping device like a stopwatch
Notes:
 Among the aforementioned tools, the measuring cylinder or electronic balance is used to check the
    infusion accuracy.
 For the tools required for disassembling and assembling the device, see chapter 8.
 Dedicated safety test devices must be used to test safety items after disassembling and then
    assembling the device.
                                                           43
Alarms and Troubleshooting
                                                 46
 Alarms and Troubleshooting
 9.4    System Faults and Troubleshooting
     System Faults and Troubleshooting
Alarm                                Alarm
               Alarm Content                  Alarm Cause                    Troubleshooting
Message                              Level
                                                                             1. Check whether the infusion tube is
                                                                             twisted or extruded and whether the
                                                                             retaining needle leaks.
                                                                             2. Check whether the occlusion level
                                                                             setting is reasonable (an older patient
                                                                             indicates a higher drug liquid viscosity
                                                                             and higher infusion rate, and the
                                                                             occlusion level also needs to be
                                                                             higher).
                                                                             3. Calibrate the pressure value of the
                                              Real-time pressure in the      consumable.
Patient Side
                Patient Side OCCL      High   tube ≥ occlusion alarm         4. Calibrate the pressure value of
OCCL
                                              value configured               the pressure sensor when it does
                                                                             not touch the infusion tube, and
                                                                             check that the KB value is correct.
                                                                             (Generally, the K value is
                                                                             404mV/N±50 and B value is about
                                                                             500mV±200.)
                                                                             5. Check the pressure sensor.
                                                                             6. Check the drive head board, nut
                                                                             displacement detection board, main
                                                                             board, and related connection cable.
                                                                             1. Check the optocoupler and related
                                              1. The motor is locked.        connection cable.
                System error: motor
                                              2. The motor rotation          2. Check the motor and related
                drive locked
ERR 001                                High   speed is too low.              connection cable.
                (motor rotation
                                              3. The motor rotation          3. Check the speed measurement board
                speed is abnormal)
                                              speed is too high.             and related connection cable.
                                                                             4. Check the main board.
                System error:
                reversed rotation of                                         1. Check the optocoupler and related
                                              The motor rotation
                motor drive                                                  connection cable.
ERR 002                                High   direction is opposite to the
                (motor rotation                                              2. Check the speed measurement board
                                              expected direction.
                direction is                                                 and related connection cable.
                abnormal)
                System error:                                                1. Check the pressure sensor and
                pressure sensor                                              related connection cable.
                                              The pressure sensor status
ERR 003         failure                High                                  2. Check the drive head board and
                                              is incorrect.
                (pressure sensor is                                          related connection cable.
                faulty)                                                      3. Check the main board.
                                                                             1. Check the main board and related
                Main chip voltage
                                              The monitored voltage is       connection cable (generally, check
                3.3V
ERR 005                                High   smaller than 1.67V or          whether the C71 voltage on the main
                (CPU voltage is
                                              greater than 2.25V.            board falls within the range of
                abnormal)
                                                                             1.67-2.25V).
                DC input voltage              The monitored voltage is       1. Check the power board and related
ERR 006         (DC voltage is         High   smaller than 2.15V or          connection cable.
                abnormal)                     greater than 3.3V.             2. Check the main board.
                AC input voltage              The monitored voltage is       1. Check the power board and related
ERR 007                                High
                (AC power supply              smaller than 2.12V or          connection cable.
                                                  47
 Alarms and Troubleshooting
                   is abnormal)                      greater than 2.87V.
                                                         48
 Alarms and Troubleshooting
                                                                               1. Check the fluid/patient side
                Drive
                                                                               occlusion pressure board and related
                communication                   A communication error
                                                                               connection cable.
ERR 013         error: M4 fails to     High     occurs between M0 and
                                                                               2. Check the connection cable between
                detect the data of              M4.
                                                                               the main board and pump door board.
                M0.
                                                                               3. Check the main board.
                System error: fluid
                                                                               1. Calibrate the pressure sensor.
                side pressure sensor            The AD value of the fluid
                                                                               2. Check the pressure sensor and
ERR 009         failure                High     side pressure sensor
                                                                               related connection cable.
                (fluid side pressure            exceeds the normal range.
                                                                               3. Check the main control board.
                sensor is faulty)
                                                                               1. Check whether the drop sensor is
                                                                               properly installed.
                                                1. No drop sensor is           2. If no drop sensor is installed, check
                Fluid side occlusion            connected or the drop          whether the drop detection function is
Fluid Side      (mutual exclusion               detection is disabled.         disabled.
                                       High
OCCL            with the drop                   2. Real-time occlusion         3. Calibrate the pressure sensor.
                sensor)                         pressure ≥ preset occlusion    4. Check the pressure sensor and
                                                alarm value                    related connection cable.
                                                                               5. Check the main control board and
                                                                               related connection cable.
                                                                               1. Check whether air bubbles exist in
                                                                               the tube.
                                                                               2. Check the air bubble sensor and
                Air bubbles/air                 The detected air bubble
Air Bubble In                                                                  related connection cable.
                bubble                 High     value is greater than the
Line                                                                           3. Check the pump door board and
                accumulation                    preset air bubble threshold.
                                                                               related connection cable.
                                                                               4. Check the main control board and
                                                                               related connection cable.
                                                                               1. Check whether dust or dirt exists on
                                                When the second pump           the pump finger optocoupler.
                Pump finger failure             finger reaches the highest     2. Check the pump finger detection
                (peristalsis pump               position, the pump finger      board and related connection cable.
ERR 011                                High
                finger drive                    encoding disk does not         3. Check the motor speed measurement
                failure)                        reach the specified            board and related connection cable.
                                                position.                      4. Check the main control board and
                                                                               related connection cable.
                                                                               1. Check whether the drop detection
                                                                               function is enabled while no drop
                                                                               sensor is connected.
                                                The drop detection             2. Check whether the drop sensor
No Drop
                No Drop Sensor         Middle   function is enabled but no     works properly.
Sensor
                                                drop sensor is connected.      3. Check the drop sensor interface
                                                                               board and related connection cable.
                                                                               4. Check the main control board and
                                                                               related connection cable.
                                                                               1. Check whether the drop detection
                                                1. The infusion tube is
                                                                               function is enabled while no drop
                                                inserted into the drop
                                                                               sensor is connected.
                                                sensor.
                                                                               2. Check whether the drip chamber is
                                                2. Drop detection is
No Drop         No Drop                High                                    dripping liquid.
                                                enabled.
                                                                               3. Check whether the drop sensor is
                                                3. The drop signal does not
                                                                               properly installed.
                                                generate any signal all the
                                                                               4. Check whether the drop sensor is
                                                time.
                                                                               faulty.
                                                    49
 Alarms and Troubleshooting
                                                                                   5. Check the drop sensor interface
                                                                                   board and related connection cable.
                                                                                   6. Check the main control board and
                                                                                   related connection cable.
 Notes:
      For the alarm message ERR 00X, X indicates a character. The specific alarm message may be ERR 001,
       ERR 002, and etc.
      For all problems related to the potentiometer, judge whether the hardware is faulty. Access the AD value on
       the Maintenance screen, move the potentiometer, and observe whether the AD value changes to judge
       whether the sensor and related wires are damaged.
      Access the AD value displayed on the screen, press the pressure sensor, and observe whether the AD value
       displayed changes to judge whether the pressure sensor and related wires are damaged.
                                                         50
Disassembly and Assembly
4 Battery 1
         1.       Use tweezers to remove the five lower shell hole plugs (2).
         2.       Use a Philips screwdriver to remove the four M3 screws (3).
         3.       Take down the face shell (1).
                   Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 51 / 93
Disassembly and Assembly
        4.       Press the fastener on the battery connector to separate the plug and socket, and take out the
  battery (4).
        1.       Remove the AC power supply connection cable, ground wire, and loudspeaker connection
  cable.
        2.       Use a Phillips screwdriver to remove the two PT3*8 cross recessed pan head drilling PT thread
  tapping screws (1).
        3.       Take down the rear panel socket assembly.
  1.2      Dismantling the Loudspeaker and AC Socket
2 Loudspeaker 1
3 AC socket 1
8 Interface socket 1
             1. Remove the five cross recessed pan head drilling PT thread tapping screws (4), and take
                 down the AC socket and loudspeaker.
             2. Screw out the two cross recessed countersunk drilling tapping screws (7), and take down
                 the AC socket.
2 HP-60 AC shield 1
4 AC-DC module 1
             4. Use a Phillips screwdriver to remove the two PT3*8 cross recessed pan head drilling PT
                 thread tapping screws (3).
             5. Take out the AC-DC module (4).
  1.4   Dismantling the Peristalsis Pump Motor Speed Measuring PCBA and Pump Finger Position
        Detection PCBA
        1. Remove the connection cable (7008000021) of the pump finger position detection board.
        2. Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
              GB 9074.4 (1), and take down the HP-60 pump finger position detection PCBA (2).
        3. Remove the connection cable (7000000025) of the HP-60 motor speed measuring board.
        4. Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
              GB 9074.4 (4), and take down the HP-60 motor speed measuring PCBA (3).
  1.5    Dismantling the Sliding Block and HP-60 Door Potentiometer Assembly
      1. Turn the HP-60 door drive gear G3 (1) to make the HP-60 sliding block (2) slide to the
             rightmost shown in the figure.
      2. Screw out the HP-60 door drive push rod (3) and the two HP-60 sliding block set pin (4).
      3. Take down the HP-60 sliding block.
      4. Remove the connection cable plug on the HP-60 door potentiometer assembly.
      5. Screw out the two PM2*3 cross recessed small pan head machine thread screws (6), and take
            Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 58 / 93
Disassembly and Assembly
     1.6    Dismantling the HP-60 Tube Detection PCBA, Anti-Free-Flow Clamp PCBA, Patient Side
            Occlusion Pressure Detection PCBA and its Connection FFC
        1. Take down the HP-60 rear panel socket assembly and HP-60 AC shield.
        2. Remove the HP-60 patient side occlusion signal connection cable (7008000019).
        3. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2), and
            take down the HP-60 tube detection PCBA (1).
        4. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (4), and
            take down the HP-60 anti-free-flow clamp PCBA (3).
        5. Remove the HP-60 patient side occlusion pressure detection board’s connection FFC (7).
        6. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (6), and
            take down the HP-60 patient side occlusion pressure detection PCBA (5).
  1.7    Dismantling the Peristalsis Pump
2 PM3*6 screw 4
        1. Take down the HP-60 sliding block according to the method in section 8.6.
        2. Take down the HP-60 common consumable middle plate (1).
        3. Screw down the four PM3*6 screws (2).
        4. Remove the HP-60 door potentiometer assembly (9121000008), HP-60 motor speed measuring
              board’s connection cable (7000000025), and plug of the peristalsis pump motor’s connection
              cable.
        5. Take down the peristalsis pump assembly (3).
  1.8    Dismantling the Peristalsis Pump Motor
9 Stepper motor 1
        1. Take down the peristalsis pump assembly according to the method in section 8.8.
        2. Remove the three cross recessed pan head, spring washer, and flat washer combination screws
              M3x6 GB 9074.4 (1 and 5).
        3. Remove the two HP-60 peristalsis pump springs (6), HP-60 pump adjustment nut (7), and HP-60
              pump spring support (8).
        4. Screw out the only PM2.5*6 cross recessed small pan head machine thread screw+spring
              washer+flat washer assembly (4).
        5. Take down the HP-60 speed measuring encoding disk (3).
        6. Screw out the four PM2.5*10 cross recessed small pan head machine thread screw+spring
              washer+flat washer assembly (2).
        7. Take down the stepper motor (9).
  1.9    Dismantling the Door Drive Assembly
     1. Remove the two cross recessed pan head, spring washer, and flat washer combination screws
         M2x6 GB 9074.4 (2) and the PT2.3*6 cross recessed pan head drilling PT thread tapping screw
         (3).
     2. Remove the motor connection cable, and take down the door drive assembly.
3. Remove the two PM2.5*6 cross recessed small pan head machine thread screw+spring
      1. Remove the three PT2.3*6 cross recessed pan head drilling PT thread tapping screws (1).
      2. Take down the HP-60 fluid side occlusion pressure detection PCBA (2).
  1.11 Dismantling the Separate Air Bubble Sensor Module A
      1. Disconnect the separate air bubble sensor module A and the main board connection plug.
      2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
      3. Take down the separate air bubble sensor module A (1).
  1.12 Dismantling the HP-60 Pump Door PCBA
      1. Use tweezers to remove the seven pump door hole plugs (1).
      2. Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
      3. Take down the HP-60 cover plate (3).
      4. Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (5).
      5. Take out the HP-60 pump door PCBA (4).
  1.13 Dismantling the Separate Air Bubble Sensor Module B
      1. Screw out the four cross recessed pan head screws M2x4 GB 818 (1).
      2. Take down the HP-60 patient side occlusion pressure detection board (2).
      3. Take down the separate air bubble sensor module B (3).
  1.14 Dismantling the HP-60 Veneer
1 HP-60 veneer 1
1. Use tweezers to strip off the veneer (1) from the seam between the veneer and shell.
      1. Use tweezers to remove the HP-60 pump door hole plug (1).
      2. Remove the PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
      3. Take down the HP-60 bottom shell FPC cover (4).
      4. Disconnect one end of the HP-60 pump door board’s connection FFC (3).
          No.                        Part Name                       Quantity
      5. Take down the HP-60 cover plate (60Z0000026) according to the method in section 8.13.
      6. Take down the HP-60 pump door board connection FFC (5).
  1.16 Dismantling the HP-60 Main PCBA
2 PM3*6 screw 8
     1. Remove the eight HP-60 lower shell hole plugs (1) and eight PM3*6 screws (2).
     2. Take down the HP-60 face shell (3).
5 HP-60 AC shield 1
      3. Take out the HP-60 middle plate assembly (10) and HP-60 pump door assembly (11).
      4. Remove the two PT3*8 cross recessed pan head drilling PT thread tapping screws (6), and take
          down the HP-60 AC shield (5).
      5. Remove the two PT3*8 cross recessed pan head drilling PT thread tapping screws (8), and take
          down the HP-60 rear panel socket assembly (7).
      6. Remove the five cross recessed pan head, spring washer, and flat washer combination screws
          M3x6 GB 9074.4 (9), and take down the HP-60 main PCBA (4).
  1.17 Dismantling the HP-60 Drop Sensor Interface PCBA
      1. Take down the HP-60 middle plate assembly (3) and HP-60 pump door assembly (4) according
          to the method in section 8.17.
      2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (1).
      3. Take down the HP-60 drop sensor interface PCBA (2).
  1.18 Dismantling the Fastener Assembly
4 HP-60 fastener 1
1. Take down the shifting block plate (1). The shifting block plate is connected through the fastener.
Be careful not to break off the fastener when taking down the shifting block plate.
2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws.
3. Take down the fastener seat (3) together with the fastener (4) and shifting block plate pressure
           spring (5).
     1.19 Assembly
     Perform assembly in the reversed order of the disassembly procedure. When performing assembly, note
the following points:
3. For the parts where lubricating grease has been applied but already erased, the operator must
4. When installing the main PCBA, do not forget to install the main board shielding case and main
5. When installing the main motor, pay attention to the direction of the outgoing line.
6. After installing the main motor and reduction gear assembly, manually make the gear rotate for
over one circle to ensure that the gear assembly rotates steadily and smoothly. Ensure that the
speed measuring encoding disk and speed measuring optocoupler do not rub each other.
7. When installing the pump door board, ensure that the surface of the display screen and transparent
      8. All cables and wires must be connected correctly and firmly. The wire arrangement must be tidy,
           and the wires must be bundled and fastened as requested
Note: The list of spare parts is updated as requested based on product changes and maintenance at regular
intervals.
11 Software Upgrade
                                             83
Appendix
1)    Connect the white end of the USB-to-RS232 connection cable to the USB port of the process
      equipment PC.
     Note: A drive needs to be installed at the first use. After drive installation, determine a
fixed USB port for connection of the USB-to-RS232 connection cable in the future. If you
change the port for connection of this line, you need to reinstall the drive.
2)    Connect the black end of the USB-to-RS232 connection cable to the external USB port of the
      HP-30/HP-60 rear panel assembly.
4) Restart the integrated machine or power on the board. The system displays a message
      indicating that the connection is successful.
                                                  84
Appendix
7) Click start download. A download progress bar is displayed. It takes about 20 minutes for
     programming of the “main program resource”.
                                                  85
Appendix
                                              86
Appendix
2. Standard screwdriver with Philips head (Diameter Φ 3mm, All length is ok)
4. Tweezers
                                              27
Appendix
                                     88
      Appendix
            months
Hospital:                       Department:                                       Contact person:
                                                       89
    Appendix
□ Check that no crack or   □Shell Leakage Current Test                   □Check air bubble sensor
damage exists              1Connect the 601 safety analyzer to the       Check method: make a 3mm
                           AC power supply.                              air bubble in the tube, then set
                           Connect the applied part of the tested        a speed to infusion; it would
                           device to the SUM end of the process          be alarm when air bubble
                           equipment and the RA end of the safety        passing the pump’s sensor.
                           analyzer.
                           Connect the tested device to the auxiliary
                           power output socket of the 601 safety
                           analyzer through a power cable.
                           Connect one end of the red test lead to the
                           red input terminal of the safety analyzer
                           and clamp the other end of the red test
                           lead onto the metal foil attached to the
                           shell of the tested device.
                           Plug in the 601 safety analyzer, and press
                           5-Enclosure leakage on the panel of the
                           safety analyzer to access the shell leakage
                           current test page.
                           Normally, the shell leakage current is not
                           greater than 100 μA. In case of a single
                           fault, the shell leakage current is not
                           greater than 500 μA.
                                                27
     Appendix
                                                       27
     Appendix
                                                     92
MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD.
After-sales service address: 12th Floor, Baiwang Research Building, No.5158 Shahe West
Road, Xili, Nanshan, 518055 Shenzhen, Guangdong, PEOPLE'S REPUBLIC OF CHINA
Fax: +86-755-26001651
Postal code: 518055
Website: http://www.medcaptain.com
Email: MC.service@medcaptain.com
©2017 MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. All Rights Reserved.
27