An ST 2
An ST 2
Plating Systems
Operative Technique
Anchorage® 2 CP
Anchorage 2 CP
Precautions
The Stryker Anchorage 2 CP System has not been evaluated for safety and compatibility in the MR environment
and has not been evaluated for safety and compatibility in Magnetic Resonance (MR) environment and has not
been tested for heating or migration in the MR environment, unless specified otherwise in the product labeling.
Contraindications
The physician’s education, training and professional judgement must be relied upon to choose the most
appropriate device and treatment. Conditions presenting an increased risk of failure include:
• Any active or suspected latent infection or marked local inflammation in or about the affected area.
• Compromised vascularity that would inhibit adequate blood supply to the fracture or the operative site.
• Bone stock compromised by disease, infection or prior implantation that can not provide adequate support
and/or fixation of the devices.
• Material sensitivity, documented or suspected.
• Obesity. An overweight or obese patient can produce loads on the implant that can lead to failure of the
fixation of the device or to failure of the device itself.
• Patients having inadequate tissue coverage over the operative site.
• Implant utilization that would interfere with anatomical structures or physiological performance.
• Any mental or neuromuscular disorder which would create an unacceptable risk of fixation failure or
complications in postoperative care.
• Other medical or surgical conditions which would preclude the potential benefit of surgery.
This publication sets forth detailed recommended procedures for using Stryker devices and instruments. It offers guidance
that you should heed, but, as with any such technical guide, each surgeon must consider the particular needs of each patient
and make appropriate adjustments when and as required. A workshop training is recommended prior to performing your first
surgery. All non-sterile devices must be cleaned and sterilized before use. Follow the instructions provided in our reprocessing
guide (L24002000). Multi-component instruments must be disassembled for cleaning. Please refer to the corresponding
assembly / disassembly instructions. Please remember that the compatibility of different product systems have not been tested
unless specified otherwise in the product labeling. See package insert (Instruction for Use) [V15011, V15013] for a complete list
of potential adverse effects, contraindications, warnings and precautions. The surgeon must discuss all relevant risks including
the finite lifetime of the device with the patient when necessary.
Acknowledgments:
Stryker acknowledges the following surgeons for their support in the development of this technique guide: Keith L. Wapner, M.D.
2
VariAx 2 System Technology
VariAx 2 System
Screw Racks
All VariAx 2 screws can be stored in an easy-to-use storage rack. Our
screw rack assembly can be stored independently on the operative room
(OR) table and features a sloped design for easy access. Each screw
rack can either clip into the tray, or sit individually on a flat surface with
a rotating standing aid on the bottom of the rack for stabilization. The
storage rack’s transparent lid allows for identification of the content. All
four screw racks combined can hold up to 680 locking and non-locking
screws – 3.5mm T10, 2.7mm T8 and 2.4mm T8. Each rack features a
spot for up to 4 washers.
Polyaxial Locking
SMARTLock technology goes one step further. Every screw hole in
the VariAx 2 foot plate (except the compression slots) permits polyaxial
screw placement. Locking screws can be angled up to 15˚ in each
direction for a total range of 30˚. Simply place the polyaxial drill guide
in the screw hole to select the appropriate angle. Each plate allows a
surgeon to insert locking or non-locking screws at a variable angle of
their choice.
3
VariAx 2 System Technology
VariAx 2 System
Color Coding
The VariAx 2 screws and instruments follow a comprehensive color-coding
scheme whereby the screw anodization color matches the corresponding
instrument’s color. This allows for easy handling in the OR environment.
Overdrills and drill guides for interfragmentary lagging technique are
available for all screw sizes. The drill bits, lagging drill guides and screws
are also color coded to facilitate ease of use. Drill guides and screwdrivers
are also identified with a T10 or T8 symbol for quick reference.
One-Size-Fits-All Washer
A 7.0mm washer is available that works with all screw sizes (T8 or T10)
when non-locking screws are used independently from the plate.
4
VariAx 2 System Technology
VariAx 2 System
Plating Instrumentation
Drill bit, 2.6mm, scaled
Overdrills and drill guides for interfragmentary lagging technique are
available for all screw sizes. The drill bits, lagging drill guides and
screws are also color coded to facilitate ease of use.
VariAx 2 offers three standard types of drill guides: a 2.0mm for use
with all T8 screws, a 2.0mm for use with optional T10 2.7mm screws
and a 2.6mm for use with T10 3.5mm screws.
Drill guide for 3.5mm screws
The 2.0mm drill guide for T8 screws facilitates drilling a 2.0mm pilot
hole for a 2.4 or 2.7mm T8 screw centrically for locking or non-locking
screws. Additionally, the opposite side of the guide facilitates eccentric
drilling for use in a compression hole when compression is desired.
The 2.0mm drill guide for T10 screws facilitates drilling a 2.0mm pilot
hole for a 2.7mm T10 screw centrically for locking or non-locking
Overdrill, AO, 3.5mm x 122mm screws. Additionally, the opposite side of the guide facilitates eccentric
drilling for use in a compression hole when compression is desired.
The 2.6mm drill guide facilitates drilling a 2.6mm pilot hole for a 3.5mm
screw centrically for locking or non-locking screws. Additionally, the
opposite side of the guide facilitates eccentric drilling for use in a
compression hole when compression is desired.
Lag overdrill guide for 3.5mm screws Note: When drilling eccentrically, the arrow marked on the
compression side of the drill guide should be pointing toward
the fracture line/osteotomy site. The compression drill guide must
be inserted perpendicularly into the compression hole and can
not be angulated. When angulating the drill guide or using the
wrong instrument, there is a risk that the screw will not properly
sit in the plate or slip through the hole.
5
VariAx 2 System Technology
VariAx 2 System
6
VariAx 2 System Technology
VariAx 2 System
Taps and Countersink
Tap for 2.7mm screws
2.4, 2.7 and 3.5mm taps are available in the system. Although all
screws are self-tapping, it is recommended to use a tap if excessive
Tap for 3.5mm screws
resistance is felt during insertion or if the bone is dense.
A countersink is also available for reducing the screw head
prominence when the screw is used independently of a plate.
Countersink
Modular Handle
VariAx 2 offers a fully modular handle system. This is composed of
two handle grip sizes (medium and large) that can be interchanged
with either a bi-directional ratcheting AO-coupling insert or a
Large handle standard AO-coupling insert.
Both handle sizes are equipped with a spin-cap to allow insertion
using a two-finger technique. In order to disengage the insert from
the handle, push down on the button on the distal part of the handle
and pull the insert away from the handle.
Note: The inserts must be removed from the handles before
cleaning.
The ratcheting insert can work in three modes: clockwise ratcheting,
Ratcheting insert counter-clockwise ratcheting or neutral. To switch between the
different modes, simply twist the distal part of the insert to the desired
driving direction.
Note: To ensure appropriate ratcheting function, perform
appropriate maintenance on the insert by applying medical-
Medium handle grade lubricant oil through the marked cut-outs.
7
VariAx 2 System Technology
VariAx 2 System
Level 4 (plates)
8
Anchorage 2 CP System Technology
Anchorage 2 CP
reconstruction cases. The VariAx SMARTLock allows
the option of T10 3.5/2.7mm or T8 2.7/2.4mm locking
screws capable of locking at up to 15˚ off axis to the plate.
Additionally, the indication-specific Anchorage 2 CP
plates merge the advantages of rigid plate fixation and
incorporates compression generated through the plate with
either a T10 4.1mm or T8 3.6mm partially threaded screw
across the fusion site at variable angle (varies on CP plate).
Level 1 Level 2
• 4 Plate Modules: MTP/Universal Module, • F&A Reduction Instruments 3: K-wire
Midfoot Module, Lapidus, Left Module, Cutter (1), Pointed Reduction Clamp (1)
Lapidus, Right Module (2 CP plates each)
• Reamer Inlay: Cup & Cone Reamers
NOTE: Templates do not have a module on their 14mm-22mm (1 of each)
own. They are found within the corresponding
plate module, and are to be used with the
VariAx 2 T10 or T8 cannulated joysticks.
• CP Instrument Inlay: K-wire 1.4mm x 100mm (5),
T10 and T8 CP Drill Guides (1 of each),
CP Reamers (2)
• CP Screw Rack (1)
9
Anchorage 2 CP System Technology
Anchorage 2 CP
Anchorage 2 CP Screws
CP Lag Screw Lengths Hole Size Anatomical Indications
VariAx 2 Screws
Screw Type Lengths Hole Size Anatomical Indications
Anchorage 2 CP was created in order to provide a complete, modular system which contains various plates
– each with their own specific indications – for reconstructive and trauma procedures for the foot and ankle.
The addition of the four new plates is intended to fill procedure gaps in the VariAx and VariAx 2 Foot Systems.
The Anchorage 2 CP System features a two-level tray that is compatible with the VariAx 2 four-level tray.
Anchorage 2 CP plates feature a CP hole that allows for a variable screw angle. All other holes feature VariAx
SMARTLock1 technology, which allows for +/- 15˚ screw insertion.
The Anchorage 2 CP Plating System contains pre-contoured, anatomic locking plates. Screw locations have
been designed to maximize bone purchase for each respective fusion without violating joint lines.
• Plate thickness: 1.5mm
• Plates are either T10 or T8 plates. T10 plates accept 4.1mm CP screws. T8 plates accept 3.6mm CP screws.
• Screws and instruments are color coded for ease of identification
• Screw material: Ti6Al4V ELI
• Plate material: Titanium grade 2
Anchorage 2 CP foot plates are designed to have a low profile. Additionally the VariAx 2 screws have
minimal screw head prominence and therefore may reduce the risk of soft tissue irritation. The Anchorage 2
CP screws have no screw head prominence when inserted through CP hole.
10
Anchorage 2 CP System Technology
T8 Cross Plates
Only VariAx 2 screws (locking/non-locking, T8 2.7/2.4mm) must
be used in circular holes, and only T8 3.6mm partially threaded
CP screws must be used in the MTP and Universal CP holes.
Anchorage 2 CP
VariAx 2 T8 drill bit and screwdriver blades are compatible with
both VariAx 2 and Anchorage 2 CP screws. The Anchorage 2
CP specific drill guide (T8) is needed for the CP screw holes of
the MTP and Universal CP plates.
11
Anchorage 2 CP System Technology
Raised edge
on plantar side
Cannulated CP Reamer
The cannulated CP reamer is advanced over the 1.4mm x
100mm K-wire to the appropriate depth limit. This reamer is
sharp with new aggressive cutting flutes. Please apply light
pressure on the bone and always use caution when operating.
12
Anchorage 2 CP System Technology
Introduction to Cross Plate (CP) Technology
CP Screws
The Anchorage 2 CP System uses an inset diagonal,
partially threaded screw to develop mechanical
compression across the fusion site. This approach
Anchorage 2 CP
minimizes the need for separate, independent stabilization
and is designed to create compression. The Anchorage 2
CP screws are self-tapping and feature an advanced new
conical pitch thread and cutting flute that catches the far
cortex faster and allows for maximum compression across
the joint, resulting in successful unions.
Quick Tip
35° angle
Midfoot CP Plate
+/- 15°
angle
35° angle
Lapidus CP Plate
25°
angle
50° angle
Universal CP Plate
+/- 15°
angle
35° angle
MTP CP Plate
+/- 7.5°
angle
14
Anchorage 2 CP System Technology
MTP Joint Preparation Options
Anchorage 2 CP
to the shaft. Hold the phalanx in position (15˚ Valgus;
10-15˚ Dorsiflexion). Make a parallel cut on proximal
phalanx and proceed to the Medial Eminence Removal.
Place a guidewire into the middle of the base of the
proximal Phalanx and drive it out distal medial. Reduce
the joint and then drive the pin back in a retrograde
fashion into the metatarsal shaft for provisional fixation.
15
Anchorage 2 CP
MTP CP – Surgical Technique
Incision/Exposure
A dorsal longitudinal incision is commonly used for
MTP joint exposure. A medial approach may also be
considered according to surgeon experience and
Anchorage 2 CP
Preparation of
Metatarsophalangeal Joint
The first MTP joint is prepared using flat cuts technique
or the provided cup and cone reamers within the
Anchorage 2 CP set (Refer to page 15 for detailed
MTP exposure). Once prepared, the cup and cone
surfaces can be aligned into desired position and
temporarily stabilized using either a 1.4 x 100mm
guidewire or a 1.6 x 200mm fully threaded wire.
Stabilization
Proper positioning can be evaluated by placing a
support against the plantar surface of the foot. Once
the desired position has been achieved, drive a K-wire
1.6mm x 200mm from the mid portion of the base
of the proximal phalanx distally out the medial side.
Then reduce the joint and drive it back in a retrograde
fashion into the metatarsal for provisional fixation.
MTP CP Preparation
The MTP CP template is oriented either R (705178) or
L (705177). Loosen and insert VariAx joystick for T8
screw holes (703927) into the large hole of the MTP
template. After inserting the joystick tip in the circular
hole, turn the knob on the upper part of the joystick
clockwise to fix it. Proper template position is achieved
when the raised edge on the MTP CP template
is directly over the metatarsophalangeal joint line. A
1.4mm x 100mm K-wire (45-80200) is inserted through
the cannulated VariAx joystick. Once the K-wire is
positioned, the template and joystick are removed.
16
MTP CP Preparation (continued)
The CP reamer (705172) is advanced over the same
K-wire until the upper edge of the reamer is flush with
the bone surface. The plate is positioned and temporarily
Anchorage 2 CP
fixated on the distal aspect of plate with a 1.4mm x 100mm
K-wire . Once fixated, the proximal holes are drilled with
T8 drill bits and filled with appropriate T8 screws.
Note: All circular holes accommodate both locking
and non-locking T8 VariAx 2 screws (2.7mm and
2.4mm) with the exception of the CP hole.
CP Screw
After proximal fixation is achieved, the T8 CP drill guide
(705173) is positioned in the CP plate hole. A drill bit
(703896) for T8 screws is used to drill to the plantar
cortex of the proximal phalanx. Use the VariAx 2/CP
depth gauge (705170) to measure the appropriate length
T8 CP screw to be inserted. Once appropriate screw
is measured, place through the CP hole. As CP screw
seats in plate, all temporary fixation is removed. Final
tightening of T8 CP screw develops compression across
the joint with lag effect.
Drill Bit, AO, Dia Convex/Concave VariAx 2 Joystick Template for MTP
2.0mm x 135mm, Surface Reamers for T8 Screw Holes Cross-Plate, Left
Scaled 703896 703927 705177
17
Anchorage 2 CP
Midfoot TN – Surgical Technique
Anchorage 2 CP
Incision/Exposure
Incision for the talonavicular joint is made dorsomedially,
beginning at the tip of the medial malleolus and in line
with the medial border of the foot, extending 6-7cm
toward the great toe longitudinally and ending between
the anterior and posterior tibial tendons. Once exposed,
all of the remaining cartilage from the joints is removed
and prepared for fusion.
18
Midfoot CP Preparation
After inserting the joystick tip in the circular hole, turn
the knob on the upper part of the joystick clockwise
to fix it. Proper position is achieved when the raised
Anchorage 2 CP
edge on the Midfoot CP template is directly over the
talonavicular joint line. A 1.4mm x 100mm K-wire
(45-80200) is inserted through the cannulated VariAx
joystick. Once the K-wire is positioned, the template
and joystick are removed. The CP reamer (705172) is
advanced over the same K-wire until the upper edge
of the reamer is flush with the bone surface.
Note: All circular holes accommodate both locking
and non-locking T10 VariAx 2 screws (3.5mm and
2.7mm) with the exception of the CP hole.
CP Screw
The T10 CP drill guide (705171) is positioned inside the
CP hole. Use the 2.6mm drill bit to accommodate the
4.1mm CP screw. Once drilled, use the VariAx 2/CP
depth gauge to measure the appropriate length for the
T10 CP screw to be inserted. When the appropriate
screw length is measured, place the 4.1mm CP screw
through the CP hole. As the CP screw is seated in
the plate, all the temporary fixation is removed. Final
tightening of the T10 CP screw develops compression
across the joint with lag effect. Insert remaining proximal
screws on the talus side of the joint and position under
fluoroscopy.
19
Anchorage 2 CP
Midfoot NC – Surgical Technique
Incision/Exposure
The incision can be extended distally for access to the cuneiforms,
first metatarsal base and naviculocuneiform and intertarsal joints.
Remove cartilage in the naviculocuneiform joint. If there is significant
Anchorage 2 CP
Midfoot CP Preparation
The Midfoot CP template comes in either short (705176) or long
(705183) depending on the anatomy. The correct template size
is chosen based on the length of plate that spans the fusion site
and leaves an adequate bony bridge between the plate screws.
Loosen and insert the VariAx joystick for T10 screw holes (703928)
into the appropriate Midfoot CP template hole. After inserting the
joystick tip in the circular hole, turn the knob on the upper part of
the joystick clockwise to fix it. Proper position is achieved when
the raised edge on the Midfoot CP template is directly over the
naviculocuneiform joint line. A 1.4mm x 100mm K-wire (45-80200)
is inserted through the cannulated VariAx joystick. Once the K-wire
is positioned, the template and joystick are removed. The CP
reamer (705172) is advanced over the same K-wire until the
upper edge of the reamer is flush with the bone surface.
Note: All circular holes accommodate both locking and
non-locking T10 VariAx 2 screws (3.5mm and 2.7mm) with
the exception of the CP hole.
CP Screw
The T10 CP drill guide (705171) is positioned inside the CP hole.
Use the 2.6mm drill bit to accommodate the 4.1mm CP screw.
Once drilled, use the VariAx 2/CP depth gauge to measure the
appropriate length for the T10 CP screw to be inserted. When
the appropriate screw length is measured, place 4.1mm CP
screw through the CP hole. As the CP screw sits in the plate, all
temporary fixation is removed. Final tightening of T10 CP screw
develops compression across the joint with lag effect. Insert
remaining proximal screws on the navicular side of the joint and
position under fluoroscopy.
20
Anchorage 2 CP
Midfoot CC – Surgical Technique
Incision/Exposure
Expose the calcaneocuboid joint through a lateral (Ollier) incision.
Distract the joint with a lamina spreader, and sharply debride
the articular cartilage to expose the bleeding subchondral bone.
Anchorage 2 CP
Completely remove the cartilage from the calcaneocuboid joint.
Midfoot CP Preparation
The Midfoot CP template comes in either short (705176) or long
(705183) depending on the anatomy. The correct template size is
chosen based on the length of plate that spans the fusion site and
leaves an adequate bony bridge between the plate screws. Loosen
and insert the VariAx joystick for T10 screw holes (703928) into the
appropriate Midfoot CP template hole. After inserting the joystick tip
in the circular hole, turn the knob on the upper part of the joystick
clockwise to fix it. Proper position is achieved when the raised edge
on the Midfoot CP template is directly over the calcaneocuboid
joint line. A 1.4mm x 100mm K-wire (45-80200) is inserted through
the cannulated VariAx joystick. Once the K-wire is positioned, the
template and joystick are removed. The CP reamer (705172) is
advanced over the same K-wire until the upper edge of the reamer
is flush with the bone surface.
Note: All circular holes accommodate both locking and
non-locking T10 VariAx 2 screws (3.5mm and 2.7mm) with
the exception of the CP hole.
CP Screw
The T10 CP drill guide (705171) is positioned inside the CP hole.
Use the 2.6mm drill bit to accommodate the 4.1mm CP screw.
Once drilled, use VariAx 2/CP depth gauge to measure the
appropriate length for the T10 CP screw to be inserted. When
appropriate screw length is measured, place the 4.1mm CP
screw through the CP hole. As the CP screw sits in the plate,
all temporary fixation is removed. Final tightening of the T10 CP
screw develops compression across the joint with lag effect. Insert
remaining proximal screws on calcaneus side of the joint and
position under fluoroscopy.
Lapidus CP Preparation
The Lapidus CP template is oriented either R or L (705180,
705179). Loosen and insert the VariAx joystick for the T10 screw
holes (703928) into the appropriate Lapidus CP template hole.
After inserting the joystick tip in the circular hole, turn the knob on
the upper part of the joystick clockwise to fix it. Proper position
is achieved when the raised edge on the Lapidus CP template is
directly over the tarsometatarsal joint line. A 1.4mm x 100mm K-wire
(45-80200) is inserted through the cannulated VariAx joystick. Once
the K-wire is positioned, the template and joystick are removed. The
CP reamer (705172) is advanced over the same K-wire until the
upper edge of the reamer is flush with the bone surface.
Note: All circular holes accommodate both locking and
non-locking T10 VariAx 2 screws (3.5mm and 2.7mm) with
the exception of the CP hole.
CP Screw
The T10 CP drill guide (705171) is positioned inside the CP hole.
Note: The Anchorage 2 CP Lapidus CP plate can be angulated
in one direction only.
Use the 2.6mm drill bit to accommodate the 4.1mm CP screw.
Once drilled, use the VariAx 2/CP depth gauge to measure the
appropriate length for the T10 CP screw to be inserted. When the
appropriate screw length is measured, place the 4.1 CP screw
through the CP hole. As the CP screw sits in the plate, all temporary
fixation is removed. Final tightening of the T10 CP screw develops
compression across the joint with lag effect. Insert remaining
proximal screws and position under fluoroscopy.
22
Anchorage 2 CP
Universal CP – Surgical Technique
Universal CP Preparation
Insert the VariAx joystick for the T8 screw holes (703927) into
the Universal CP template hole (705182). After inserting the
joystick tip in the circular hole, turn the knob on the upper part
Anchorage 2 CP
of the joystick clockwise to fix it. Proper position is achieved
when the raised edge on the Universal CP template is directly
over the joint line. A 1.4mm x 100mm K-wire (45-80200) is
inserted through the cannulated VariAx joystick. Once the
K-wire is positioned, the template and joystick are removed.
The CP reamer (705172) is advanced over the same K-wire
until the upper edge of the reamer is flush with the bone surface.
Note: All circular holes accommodate both locking and
non-locking T8 VariAx 2 screws (2.7mm and 2.4mm) with
the exception of the CP hole.
CP Screw
After the T8 distal screws are fixated, the T8 CP drill guide
(705173) is positioned in the CP plate hole. A 2.0mm drill
bit (703896) for the T8 screws is used to drill to the proximal
cortex. Use the VariAx 2/CP depth gauge (705170) to measure
the appropriate length T8 CP screw to be inserted. When
appropriate screw length is measured, place the 3.6mm
CP screw through the CP hole. As the CP screw sits in the
plate, all temporary fixation is removed. Final tightening of the
T8 CP screw develops compression across the joint with lag
effect. Insert remaining proximal screws and position under
fluoroscopy.
23
References Manufacturer:
1. Licensed from Prof. Wolter Stryker GmbH
This document is intended solely for the use of healthcare professionals. A surgeon must always rely on his or her own professional Bohnackerweg 1
clinical judgment when deciding whether to use a particular product when treating a particular patient. Stryker does not dispense 2545 Selzach
medical advice and recommends that surgeons be trained in the use of any particular product before using it in surgery. Switzerland
The information presented is intended to demonstrate a Stryker product. A surgeon must always refer to the package insert, www.stryker.com
product label and/or instructions for use, including the instructions for Cleaning and Sterilization (if applicable), before using any
Stryker product. Products may not be available in all markets because product availability is subject to the regulatory and/or
medical practices in individual markets. Please contact your Stryker representative if you have questions about the availability of
Stryker products in your area.
Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following trademarks or service
marks: Anchorage, SMARTLock, Stryker, VariAx. All other trademarks are trademarks of their respective owners or holders.
The products listed above are CE marked.
Content ID: AN-ST-2, Revision 1, 4-2016
Copyright © 2016 Stryker