Chemistry Analyzer User Guide
Chemistry Analyzer User Guide
Instruction:
Caution:
Colorimetric cuvette should be marinated by the alcohol at least 60 minutes after self-
checking for the first time working, and then use distilled water to rinse it time by time
for 10 minutes. If not, the test result might be inaccurate.
I
DR-7000D Semi-automatic Chemistry Analyzer
Content
Chapter 1 Introduction.............................................................................................................................
1.1 About DR-7000D........................................................................................................................
1.2 Technical specification................................................................................................................
1.3 Warning Mark.............................................................................................................................
Chapter 2 The Basic Principle and Functional Feature............................................................................
2.1 Basic principle............................................................................................................................
2.2 Functions...................................................................................................................................
Chapter 3 Installation and Application....................................................................................................
3.1 Instrument Installation................................................................................................................
3.2 Main parts and Application.........................................................................................................
Chapter 4 Basic Operation.......................................................................................................................
4.1 Electrification..............................................................................................................................
4.2 Sample testing............................................................................................................................
4.3 Program measurement.............................................................................................................12
4.4 Quality control...........................................................................................................................20
4.5 Setup........................................................................................................................................22
4.6 Record search..........................................................................................................................24
4.7 Rinse.........................................................................................................................................25
Chapter 5 Instrument Maintenance.......................................................................................................26
5.1 Instrument surface cleaning....................................................................................................26
5.2 Colorimetric cuvette cleaning..................................................................................................26
5.3 Lamp-house replacement........................................................................................................26
5.4 Fuse replacement....................................................................................................................27
5.5 Pump pipette replacement.......................................................................................................27
5.6 Print paper replacement..........................................................................................................28
5.7 Printer crankshaft maintenance...............................................................................................28
Chapter 6 Transportation and Storage..................................................................................................29
6.1 Transportation requirements....................................................................................................29
6.2 Storage requirements..............................................................................................................29
Appendix A..............................................................................................................................................30
Appendix B.............................................................................................................................................31
Appendix C.............................................................................................................................................33
Appendix D.............................................................................................................................................37
Appendix E.............................................................................................................................................39
I
DR-7000D Semi-automatic Chemistry Analyzer
Chapter 1 Introduction
The built-in dot matrix LCD display provided with apheliotropic board ensures the clear show of all
test results, functional status and operational instruction. The English indication system guides
users’ operation, and improves man-machine communication.
s) Environment condition:
1
DR-7000D Semi-automatic Chemistry Analyzer
There are two biohazard marks on the equipment, one is Biohazard mark, another is Force
electricity mark, they are all on the backboard of the equipment. As shown in Figure 1-1.
Biohazard mark
Warning mark
Figure 1-1 Warning Mark
There might be a potential infectivity for the sample & waste reagent, please make sure to take
protection action during the testing, cleaning process and instrument maintenance.
Samples and waste reagent should be disposed in a suitable package conforming to the local
regulation for laboratory.
2
DR-7000D Semi-automatic Chemistry Analyzer
A=log(I。 /I)=log(1/T)=∈bc
T= I /I。
Amplifier
VG
Lamp-house VI Display/Keyboard Printer
Vn
The principle diagram of the electronic system is shown above in Figure 2-2. When passing
through the monochromator the ray of the lamp-house will be changed into monochromic light. The
result will be then received by the photoelectric detector when it passes through the flow
colorimetric cuvette, and further transformed into electric signal. The signal will be amplified and
filtered, and then transmitted into A/D transformer. When the information is transformed into
numbers, it will be sent into data collection system of the computer for kinds of calculation and
transaction. Users can either read the final result on the display or alternatively get it printed out.
2.2 Functions
1.Measurement methods :
a) End-point b) Two-point kinetic
c) Kinetic d) Absorbance
e) Multipoint
2. Internal and external printer is available for uses’ options.
3. Sampling volume, test method is programmable.
4. Users can determine whether to execute real time print out of the test results.
5. If any operational mistakes or other false results are found during the process of measurement,
users can choose the repeat key (“●”) to retest, the original serial number of the test remains.
6. Rinsing program is available between two sample tests to avoid cross contamination.
7. Up to 1 month quality control tracking, and the printing results provided.
8. Integrated test report printing is available.
9. Reaction curve display:
Reaction curve for sample and standard liquid will be displayed on the screen during the testing, as
shown in Figure 2-3.
ABS LIMIT
Absorbance Absorbance
ABS LIMIT
0 0
t1 time t1 time
(a)Positive reaction curve ( b ) Negative reaction curve
Note: 1. Press “+/-” to check the reaction curve after testing the standard liquid or sample.
2. Press “Exit” to test next sample.
4
DR-7000D Semi-automatic Chemistry Analyzer
① ② ③
④ ⑤ ⑥ ⑦ ⑧
● Make sure to maintain some cleaning agant in the colorimetric cuvette when it stops working to
5
DR-7000D Semi-automatic Chemistry Analyzer
B . Keyboard
The keyboard comprises 19 key-presses, i.e. 12 number keys (including ± and radix point), 5
functional keys and “EXIT” and “ENTER” keys.
Number key — Entering into submenu or numeric parameter input.
ENTER — Press to confirm the executive command.
EXIT — Press to return to the former menu.
F1 — F5 (Functional key) — Press to execute the relevant functions indicated in the five small
squares at the bottom of the display.
C. Buzzer
1. As the key-press on the control board is made of pellicle without clear tactility, buzzer reproducer
is used to help inform operator as soon as the key-press is touched.
2. The buzzer will remind the user to move the sample away once the equipment finishes sampling.
E. Printer
Both of internal printer and external printer are available for the instrument. The user can select one
of them for operation. There are 3 status optional: ON, OFF and AUTO. Real-time printing is
available once select AUTO status.
6
DR-7000D Semi-automatic Chemistry Analyzer
4.1 Electrification
The power switch of the instrument is on the back-board (see Figure 3-1). Turn on the power switch
to get the instrument in the state of electrification, and then the info in Figure 4-1-1 and 4-1-2 will be
displayed on the screen orderly.
Semi-Automatic
Chemistry Analyzer
Version x.x x
Feb 6 2007
Figure 4-1-1
05—07—15
19:13:11
Self checking …
Please pump distilled water
Figure 4-1-2
Note: Under the Self checking interface(as shown in Figure 4-1-2), press the sucking button to
enter into the submenu as shown in Figure 4-1-3.
05—07—15
19:13:20
Self checking…
Please waiting
Figure 4-1-3
05—07—15
19:13:51
warm-up for 20 minutes should be done before testing
if not, test results might be inaccurate !!
Figure 4-1-4
Reminder info (see Figure 4-1-5) will be shown on the screen after self checking. The user also
could press “Exit” to omit it to enter into the test interface directly (To avoid inaccurate test result,
please don’t omit this process).
图 4-1-3
Figure 4-1-5
Key in the final selection and then enter into the program main menu, as shown in Figure 4-1-6.
Note: If select Yes, all the test results will be deleted except for QC results and the sample serial
number will be reordered from 1 automatically.
If select No, the system will back to the main menu, as shown in Figure 4-1-6 :
1. Sample testing
2. Parameter setup
3. Quality control
4. Setup
5 .Record search
6. Rinse
Figure 4-1-6
The skip point at the first item after enter into the test item interface, there are 2 ways to select the
item.
Way 1: Input the item number (such as 001).
Way 2: Press the funtion key F1 and F2.
LIVER :
001)ALT 011)r-GT
002) ALP 012)CHE
003) T-BIL 013)D-BIL
004)ALB 014)TP
005)BUN 015)BUN
006) Cr 016)AST
007) CO2 017)Ca
008) CL 018)P
8
009) Fe 019)Mg
010) GLU 020)CHO
Figure 4-2-1
Note : If inaccurate item number is input, it will be corrected by key the accurate one under the
same interface directly and the original one will be covered automatically, the item no. should be 3
digits.
Input the selected Item Number. (such as: 003 or press F2 to select 003) and then press “Enter” to
enter into the test status. As shown in Figure 4-2-2.
Press the sucking button to suck the distilled water and then the following info (as shown in Figure
4-2-3) will be shown on the screen. Please follow the screen reminder info to do the test operation
one by one.
Note:
a) Press sucking button to make the distilled water, reagent, sample, standard liquid and quality
control liquid are all sucked into the colorimetric pool by sucking pipette according to the
reminder info one by one, and then test results and ABS value will be shown on the screen
after finishing measurement. If printer status is ON, press F2 to print the test results out and
press 8 to run the print paper; If printer status is AUTO, the printer will execute real time test
9
DR-7000D Semi-automatic Chemistry Analyzer
result print-out. Test results will be saved automatically, if the test results exceed 3000 pieces,
the reminder info “THERE IS NO SPACE FOR DATA BASE, CLEAR ALL SAMPLE DATA?”,
press “EXIT” to out and press “ENTER” to clear it. The equipment will not be workable unless
do the clear operation.
b) One sample serial number is only correspond with one patient, it will be added 1 after each
testing automatically. Further more the operator also could press ENTER button to enter into
sample edit submenu, as shown in Figure 4-2-4, and then input a new sample serial number.
(within 001-999).
__ __ __ __
Figure 4-2-4
Press ENTER after inputing the sample number and it will be the serial number for the next
testing sample.
c) If ID STATUS is ON, there are 2 ways to input the ID number, one is by barcode reader, another
is by press number 9 button to enter into the ID edit interface (as shown in figure 4-2-5), and
then input the ID number (0~9) by hand (the ID number should not more than 10 digits),
otherwise, there will be no ID number.
__ __ __ __ __ __ __
Input ID No.
Figure 4-2-5
Note:
(1) There is only one ID correspond with one daily sample number. One time ID number input is
avaiable for the samples which use the same sample serial number.
(2) There is no ID for new sample, repeat ID input is avaiable and the system will keep the final one
as the ID number.
d) Repeat test could be done after pressing the repeat key “●” and the original sample no. will not
be changed.
e) If quality control or re-calibration is required, press “F5” or “F3” and then “Pump standard liquid
again?” or “Pump control liquid again?” info will be shown on the screen
f )For the end point measurement, reminder info “Test the reagent blank again ?” will be shown
on the screen after setting parameters again, user could select the proper info, if select “1” the
reagent blank value will be tested again or it will be calculated according to the original value
by the equipment. As shown in Figure 4-2-6.
Test name:TP 05—07—16
Mode: ENDPOINT
Sample No.: ID:
Medium:
ABS:0.02
Factor : 1-YES
10 Result: 2-NO
Test reagent blank again?
DR-7000D Semi-automatic Chemistry Analyzer
LIVER :
001)ALT 011)AMM
002) AST 012)CHE
003) TP 013)GLDH
004)ALB 014)ICDH
005)T-BIL 015)
006) D-BIL 016)
007) TBA 017)
008) ALP 018)
009) GGT 019)
010) LAP 020) 4-3-1
Figure
Pre Next 001 Up Down
4.3.1 Parameter setup:
For example, input “003” or press F2 (“↓”) 2 times to select item “TP” and then press “Enter” to
enter into its submenu. As shown in Figure 4-3-2.
Program
1 、TEST NAME : TP P1
2 . TEMP : -- -- ℃
3 . MODE : END POINT
4 . UNITS : g/L
5 . L—LOW:0.0000
6 . L—HIGH: : 150
Figure 4-3-2
UP DOWN
The content in the bracket is the item instruction of program (Reminder info.) “P1”, “P2”, “P3” “P4”
11
DR-7000D Semi-automatic Chemistry Analyzer
on the top of the screen means the page number, Press F4, F5 to do page up and page down
operation, press 1-6 to edit and modify the content of relevant items.
There are totally 5 pages of program submenus, “P1” in Figure 4-3-2 means Page 1, press F5 to
view Page 2 (see Figure 4-3-3), Page 3 (See Figure 4-3-4) and Page 4 (See Figure 4-3-5).
Program P2
1 、R-BLANK : NO
2 . S-BLANK : NO
3 . STANDARD : 0.0
4 . FACTOR : 0.1
5 . VOLUME : 500
6. FLOAT: 002
Figure 4-3-3
UP DOWN
Program P3
1 、WL1 : 000
2 、WL2 : --------
3 、Delay ( S ) : 000
4 、MEAS ( S ) : 000
5 、REF-LOW: 0.1
6 、REF-HIGH: 0.1
Figure 4-3-4
UP DOWN
If select the Multipoint method, concentration parameter under P4 should be input, however, it is no
need for other measurement methods.
Program P4
STANDARD
P3 1: 0.000
STANDARD2: 0.000
STANDARD3: 0.000
STANDARD4: 0.000
STANDARD5: 0.000
Figure 4-3-5
UP DOWN
1 、Test name :
0 1 2 3 4 5 6 7 8 9
0 1 2 3 4 5 6 7 8 9
A B C D E F G H I J
12
K L M N O P Q R S T
U V W X Y Z — ()
↑ ↓ Del Page
DR-7000D Semi-automatic Chemistry Analyzer
Figure 4-3-6
As known from above, ‘↑’ corresponds to F1, ‘↓’ corresponds to F3, ‘Delete’ corresponds to F4,
For example, to change the name into GPT: Press F3 (or F1) to move the cursor to the line where
there is ‘G’, then press 6 (‘G’ is correspond to 6), ‘G’ will be displayed in the Name Column. Press
F3 (or F1) to move the cursor to the line where there is ‘P’, then press 5 (P is correspond to 5) ‘GP’
will be displayed in the Name Column, and so forth. When all finished, press ENTER to return to
the last submenu (if the input is more than 8 letters, it will return to last submenu automatically). If
any letter is entered incorrectly, press F4 to delete the letter. For lowercase entering, press F5 to
look up lowercase on the next page.
B ) Modifying Temperature : Press 2 under the submenu on the first page (as shown in
Figure 4-3-2), then press ENTER, the screen is shown as Figure 4-3-7 :
Program :( Temperature )
2 、TEMP : ℃
0 、──── ℃
1 、────25℃
2 、────30℃
3 、────37℃
Figure 4-3-7
Key in the number in front of the selected temperature, then press ENTER to return to the
submenu.
C ) Modifying measure method : Press 3 under the submenu on the first page, then press
ENTER, the screen is shown as Figure 4-3-8.
3 . MODE :
0 ────ENDPOINT
1 ————TWOPOINT
2———— KINETIC
3———— MULTI-POINT
4———— ABS
Figure 4-3-8
Select the proper number and then press ENTER to back the upper menu.
13
DR-7000D Semi-automatic Chemistry Analyzer
D ) Modifying units : Press 4 under the submenu on the first page, then press ENTER. As
shown in Figure 4-3-9 :
Program ( Units )
4 . UNITS :
0 . ABS 4.μmol/L
1 . U/L 5.g/L
2 . μkat/L 6.g/dL
3 . mmol/L 7. mg/dL
Figure 4-3-9
Page
F5 is for page down ,as shown in Figure 4-3-10:
Program ( Units )
4 . UNITS :
0 . μg/dL 4.μg/L
1 . μU/mL 5.ng/L
2 . nmol/L 6. %
3 . mg/L
Key in the number in front of the selected unit, then press ENTER to return to the submenu (as
shown in Figure 4-3-2).
E) Modifying linearity range: Press 5 under the submenu on page 1, then press “Enter” As shown
in Figure 4-3-11,4-3-12:
 ̄ ̄ ̄ ̄ ̄ ̄ ̄
 ̄
 ̄ ̄ ̄ ̄ ̄ ̄ ̄
 ̄  ̄  ̄  ̄  ̄
 ̄
FigureL-HIGH
4-3-12 value:
Input
Input the proper parameter ̄  ̄  ̄  ̄  ̄
and then press “Enter” to back to the submenu (as shown in Figure 4-3-
2).
Note:
Input the linearity range according to the specification on the reagent kits, if the parameters exceed
14
DR-7000D Semi-automatic Chemistry Analyzer
the range during test, reminder mark “!” will be printed on the test report.
“!” means “Test the sample again after dilution and the result should be multiplied the dilution
multiple”.
F ) Modifying Reagent Blank : Press 1 under the program menu on the second page as
shown in Figure 4-3-3 , then press ENTER, as shown in 4-3-13:
1.R-BLANK :
0─── NO
1─── YES
Figure 4-3-13
Press 0 or 1 for R-blank Yes or No, then press ENTER to return to the submenu ( as shown in
Figure 4-3-3).
G ) Modifying sample blank : Press 2 under the program menu on the second page (as
shown in Figure 4-3-3), then press ENTER, as shown in Figure 4-3-14:
2.S-BLANK :
0─── NO
1─── YES
Figure 4-3-14
H ) Modifying standard liquid concentration: Press 3 under the submenu on the second
page (as shown in Figure 4-3-3), then press ENTER, as shown in Figure 4-3-15 :
 ̄ ̄ ̄ ̄ ̄ ̄ ̄
Note:
The factor will be turned into ‘----’ automatically after entering the standard liquid concentration.
The factor will be saved automatically and reminder info “Confirm the standard liquid again?” will be
shown on the screen once the user opens the submenu again, if select “yes”, the old coefficient will
15
DR-7000D Semi-automatic Chemistry Analyzer
I) Modifying Factor : Press 4 under the program menu on page 2 (as shown in Figure 4-3-3),
then press “Enter”. The factor is “00000.” input proper value, then press “Enter” to return to the
submenu. As shown in Figure 4-3-16.
 ̄ ̄ ̄ ̄ ̄ ̄ ̄
J ) Modifying the sampling volume : Press 5 under the program menu on Page2 (as shown
in Figure 4-3-3), then press ENTER. Input the proper number, then press ENTER to return to the
submenu. As shown in Figure 4-3-17.
uL
K ) Modifying the decimal digits of test results: Press 6 under the program menu on page 2
(as shown in Figure 4-3-3), then press Enter to input the required decimal digits, then press Enter
to turn to the submenu. As shown in figure 4-3-18. (Decimal digits is the numbers after the radix
point of the test result, the max. one is 4 digits such as 1.2334).
Decimal digi
Figure 4-3-18
L) Modifying wavelength : Press 1 under the program menu on Page 3 (as shown in Figure 4-
3-4), then press ENTER. Input the selected wavelength number, if correct, press ENTER to back to
the submenu (sub. wavelength press 2), as shown in Figure 4-3-19.
WAVE :
1 . 340 5 . 505
2 . 405 6 . 546
3 . 450 7 . 578
4 . 492 8 . 630
16 Input wavelength 1:
 ̄
 ̄ ̄  ̄ ̄  ̄ ̄  ̄ ̄
n m ̄
DR-7000D Semi-automatic Chemistry Analyzer
Figure 4-3-19
Note: There will be a reminder info “Input wavelength again” on the screen, once main wavelength
value input is unavailable and the operator should input the wavelength again.
M) Modifying delay time : Press 3 under the main menu on page 3 (as shown in Figure 4-3-
4), then press ENTER. Input the proper time, if correct, press ENTER to back to submenu. (Delay
time setup should be according to the specification on the reagent kits), as shown in Figure 4-3-20.
N) Modifying test time : Press 4 under the main program menu on page 3 (as shown in figure
4-3-4), then press “ENTER, input proper number, if it is correct, press ENTER to back to the
submenu. As shown in Figure 4-3-21.
Note: Measure time is only applied when using two-point kinetic and kinetic mode.
O) Modifying reference high : Under the main program press 6 on page 3 (as shown in figure
4-3-4), then press ENTER. Input the proper number , if it is correct, press ENTER to back to the
submenu ,as shown in Figure 4-3-22.
________________________
P) Modifying reference low : Under the main program press 5 on page 3 (as shown in figure 4-
3-4, then press ENTER. Input the proper number , if it is correct, press ENTER to back to the
submenu , as shown in Figure 4-3-23 :
17
DR-7000D Semi-automatic Chemistry Analyzer
__________________________
Quality Control :
1 . Program :
2 . Control statistic :
3 . Clear data :
LIVER :
001)ALT 011)AMM
002) AST 012)CHE
003) TP 013)GLDH
004)ALB 014)ICDH
005)T-BIL 015)
006) D-BIL 016)
007) TBA 图 1-1017)
008) ALP 018)
009) GGT 019)
010) LAP 020)
Figure 4-4-2
Pre Next 001 UP DOWN
Input the selected Item number then press ENTER to enter into the submenu, as shown in Figure
4-4-3, then input the proper number, if it is correct, press ENTER, as shown in Figure 4-4-3.
Figure 4-4-3
Input the proper number of control lot, target value and standard deviation one by one. If they are
correct, press ENTER to back to main menu. Two types of control liquid are available for the
program.
Input No.
Figure 4-4-4
001)ALB 3SD
Tested 2SD
X=
SD = X
CV =
DAY = 3 -2SD
C03 = 0.0 -3SD
Standard:
X=
SD=1 5 15 25
Figure 4-4-5
QC 1 Print ← →
Press “←” or “→” to view the control data within 1 month.
Figure 4-4-7
Press “ENTER” to clear all the statistic data and then back to upper menu, as shown in Figure 4-4-
6.
Figure 4-4-8
Press “Enter” to enter into the submenu, key in the selected number and then press ENTER to
clear it.
4.5 Setup
Under the program main menu, press 4 to enter into the submenu, as shown in Figure 4-5-1.
1 . Language-English
2 . Printer
3 . Date & time
4 . Lamp-house replacement
5. ID: ON
6. Resume:
4.5.1 Printer
Under the setup menu (as shown in figure 4-5-1), press 2 to enter into the printer submenu, as
shown in Figure 4-5-2:
1 . Internal printer : ON
2 . External printer : OFF
3. Printer space: 2
Press 2 to select the status of external printer, then press ENTER to back to the submenu.
Press 3 to select the run paper distance (per time) and the printed test result line-distance, there
are 5 values optional: 2, 4, 6, 8, 10 line.
2005-05-19 16:30:25
MM-DD-YY
Lamp-house
Luminosity:
Figure 4-5-4
Program indicates pump, and set wavelength 340nm automatically, then read the number of the
luminosity continually. Press EXIT to back to last menu.
4.5.4 ID setup
Press 5 under the Setup menu (As shown in Figure 4-5-1) to set up the ID and then press ENTER
TO back the upper menu.
Under the RECORD SEARCH submenu( as shown in Figure 4-6-1), search function is avaiable
once press any key within 1 to 4, such as press 1 ,the following interface will be shown on the
screen.
05-07-16 15:55:08
Test name:
Sample No.: 001
ID:
REF-HIGH:83.00
REF-LOW:60:00
UNITS:
RESULT:0.0
Single Print Del Prior Next
Figure 4-6-2
Press F1 to select the printing report format (single or integrate).
Press F2 to print the test report.
Press F3 to delete the current note.
Press F4 to display the previous record.
Press F5 to display the next record.
Press EXIT to back to the upper menu (as shown in Figure 4-6-1).
Figure 4-5-6
Record clearance
Figure 4-6-3
Press 1 to clear all the records and press EXIT to back to upper menu as shown in figure 4-6-1.
4.7 Rinse
Under the program main menu (as shown in Figure4-1-6), press 6 to enter into RINSE submenu,
as shown in Figure 4-7-1.
___ ___
22
DR-7000D Semi-automatic Chemistry Analyzer
Figure 4-7-1
Input the rinse time by hand and then press ENTER. The equiment will do the rinse operation
according to the selected time automatically. (Rinse time scope 1~99)
(3)
Take off the lampshade and pressure pad first ,and then take down the illuminant. (Warning: light
bulb and ambient temperature may be very high, to avoid dangers, please take gloves to do
this operation).
(4)
Insert the new bulb (12V, 20W) in the lamp stand without touching the top of the light bulb, and then
screw the pressure pad. To avoid effecting the test results, please do not touch the bulb by
hand directly. Use alcohol cotton to clean it.
(5)
Under such circumstance, switch on the power to enter into setup menu ,then press 4 to enter into
the lamp house replacement submenu, according to the reminder info to pump the distilled
water and then the Luminosity value will be shown on the screen.
(6)
Adjust the position of lamp-house until the figure on the bottom of the display turned into maximum
(Reference range 18000~50000).
(7)
Ensure the figure on the bottom of the display unchanged, and tighten all screws.
(8)
Turn off the power, and put on the lampshade. Thus the illuminant replacement is finished.
(9)
Insert the pipette properly. Put DOS shell on and tighten all screws.
Figure 5-3-1
24
DR-7000D Semi-automatic Chemistry Analyzer
Figure 5-4-1
Pump pipette:
Diameter : 2.06mm ( inner )
Length : 90mm
Figure 5-5-1
Paper Entry
Stay bar
Figure 5-6-1
Cautions:
The instrument should be rinsed several times and the sucking pipette should be kept empty & dry
26
DR-7000D Semi-automatic Chemistry Analyzer
before transportation or storage. If the transportation ambient temperature is lower than 0℃, it is
better to use 95% alcohol to rinse the pipette in order to avoid frost crack.
Appendix A
Manufacturer’s warranty
Dear Customer :
Thanks for purchasing DR-7000D semi-automatic chemistry analyzer, our company will provide the
following services for you:
4. We will provide updated service for free according to the technology improvement.
Any technological service requests, please contact us according to the following address:
Appendix B
Trouble shooting
3. Tested result is not good. Sample feeding generates After several rinsing or sample
air bubble, pump delivery feeding, if it still doesn’t work properly,
and speed is instable. wash it with specific cleaning agent for
2-3 times, then with distilled water for
5-10 times. Re-input the pump
parameters.
28
DR-7000D Semi-automatic Chemistry Analyzer
4. The printer doesn’t work Printer plug falls out or is Open the DOS shell, reinsert plug and
and display E1, E2, E3, E4 loose connection. check if the light for printer board
etc info. flashes.
5. The pump does not run Switch damaged or loose Open the DOS shell, reinsert plug or
after switching on the connection. replace the switch.
sample feeding power.
6. Sampling abnormity after Interface from the outside Turn off the power and restart the
system self-checking. electric net. equipment after several seconds.
7. Temperature abnormity Temperature sensor Turn off the power in normal room
after system self-checking. socket is not connected temperature, restart it after 5
well or the temperature minutes.If it is not workable, open the
exceed the scope of equipment’s shell and refix the sensor
5℃~40℃. socket.
8. The buzzer doesn’t work. Computer shuts down, Turn off the power, and restart a few
which is usually caused by seconds later.
external interference.
9. some a key does not Key cover is used too Replace the key cover.
work. much.
10. Display with variegation It is usually caused by Turn off the power, and restart a few
after switching on. external interference. seconds later. Or open case, reinsert
all cables connection.
11. The pump works Flow system is blocked or Check that if the input and output
without sample feeding. pipette aging or Blocked. pipette are twisted or extruded, and
check if pump pipette is tightened or
aging. To replace with a new one or
cut it shorter. And clean and
impregnate flowcell.
29
DR-7000D Semi-automatic Chemistry Analyzer
Appendix C
Verification: none
30
DR-7000D Semi-automatic Chemistry Analyzer
31
DR-7000D Semi-Automatic Chemistry Analyzer
1 2 3 4 5 6 7 8 9 10 1 12 1 1 15 16 1 18 1 20 21 2 23 2 2 26 2 28 2 30
1 3 4 7 9 2 4 5 7 9
STX C LF
R
SP S SP T e s t i n g S R e p o r t CR L
P P F
CR L
F
SP D a t e : × × × × - × × - × × S × × : × × : × × CR L
P F
SP S a m p l e S N O . : × × × CR L
P F
SP I D : × × × × × × × × × × CR LF
SP T e s t S N a m e : × × × × × × × × CR LF
P
SP R E F : S × × × × × × × × × × × × × × × C LF
P R
SP S e q . S S R e s u l t S SP U n i t SP SP H i n t CR L
P P P F
SP × × × S × × × × × × × × × SP × × × × × × × SP × × CR L
P F
CR L
F
ETX
34
DR-7000D Semi-Automatic Chemistry Analyzer
Note :
SP------- space character (20h hex)
X--------- virtual value bit
STX----- start character (02h hex)
CR------- return character (0dh hex)
LF-------- line feed character (0ah hex)
ETX------ ending character (03h hex)
1 2 3 4 5 6 7 8 9 1 1 1 13 1 1 16 1 1 1 20 2 22 23 2 2 26 27 2 2 30
0 1 2 4 5 7 8 9 1 4 5 8 9
STX CR L
F
SP SP Q u a l i t y S C o n t r o l S R e p o r t C L
P P R F
CR LF
SP D a t e : × × × × - × × - × × S × × : × × C L
P R F
SP T e s t S N a m e : × × × × × × × × CR L
P F
SP I D : × × × × × × × × × × C LF
R
SP S t a n d a r d S X : SP × × × × × × × × CR L
P F
35
DR-7000D Semi-Automatic Chemistry Analyzer
SP S t a n d a r d S S D : S × × × × × × × × C L
P P R F
SP R e s u l t : × × × × × × × × × C L
R F
ETX
34
DR-7000D Semi-Automatic Chemistry Analyzer
1 2 3 4 5 6 7 8 9 1 1 1 13 1 1 16 1 1 1 20 2 22 23 2 2 26 27 2 2 30
0 1 2 4 5 7 8 9 1 4 5 8 9
STX CR L
F
SP SP Q u a l i t y S C o n t r o l S R e p o r t C L
P P R F
CR LF
SP D a t e : × × × × - × × - × × S × × : × × C L
P R F
SP T e s t S N a m e : × × × × × × × × CR L
P F
SP I D : × × × × × × × × × × C LF
R
SP S t a n d a r d S X : SP × × × × × × × × CR L
P F
SP S t a n d a r d S S D : S × × × × × × × × C L
P P R F
SP R e s u l t : × × × × × × × × × C L
R F
ETX
36
DR-7000D Semi-Automatic Chemistry Analyzer
Appendix D
Parameter Cod Product Methodology Wavelength Temperature Standard Factor Unit Measure Lag Reagent
e concentrati ment time
No. on time(S) (S)
Liver 01 ALT Kinetic 340nm 37℃ ── -3376 U/L 60 60 double reagent
Function 02 GGT Kinetic 405nm 37℃ ── 1158 U/L 60 60 double reagent
03 ALP Kinetic 405 nm 37℃ ── 3378 U/L 60 60 double reagent
04 CHE Kinetic 405 nm 37℃ ── 8369 U/L 60 60 double reagent
05 T-BIL End point 546 nm 37℃ 89.6 ── umol/L ── ── double reagent
06 D-BIL End point 546 nm 37℃ 26 ── umol/L ── ── double reagent
07 ALB End pint 630 nm 37℃ 40 ── g/L ── ── simple reagent
08 TP End pint 546 nm 37℃ 80 ── g/L ── ── simple reagent
09 NH3 2-point kinetic 340 nm 37℃ 200 ── umol/L ── ── simple reagent
10 AST Kinetic 340 nm 37℃ ── -3376 U/L 60 60 double reagent
11 LAP Kinetic 405 nm 37℃ ── 2121 U/L 60 60 double reagent
12 ICDH Kinetic 340 nm 37℃ ── 3376 U/L 60 120 double reagent
37
DR-7000D Semi-Automatic Chemistry Analyzer
13 ADA Kinetic 340 nm 37℃ ── 4921 U/L 60 180 double reagent
Kidney 14 BUN 2-point kinetic 340 nm 37℃ 7.14 ── mmol/L 60 30 double reagent
Function 15 Cr 2-point kinetic 505 nm 37℃ 400 ── umol/L 60 30 double reagent
16 UA End point 505 nm room temperature 7.2 ── mmol/L ── ── double reagent
Hydronium 17 Ca End pint 578 nm room temperature 2.5 ── mmol/L ── ── double reagent
18 Ca End pint 630 nm room temperature 2.5 ── mmol/L ── ── simple reagent
19 CI End point 505 nm room temperature 100 ── mmol/L ── ── simple reagent
20 Mg End point 546 nm room temperature 1 ── mmol/L ── ── simple reagent
21 Fe End point 546 nm room temperature 90 ── umol/L ── ── double reagent
22 P End point 340 nm room temperature 1.29 ── mmol/L ── ── simple reagent
Glucose 23 GLU End point 505 nm 37℃ 5.55 ── mmol/L ── ── double reagent
metabolism 24 GLU End point 340 nm 37℃ 5.55 ── mmol/L ── ── double reagent
Cod Product Methodology Wavelength Temperature Standard Factor Unit Measure Lag Reagent
e concentrati ment time
No. on time(S) (S)
25 FTS End point 546 nm 37℃ 1.95 ── mmol/L ── ── simple reagent
Blood 26 CHO End point 505 nm 37℃ 7.0 ── mmol/L ── ── double reagent
lipoid 27 HDL-C End point 546 nm 37℃ 53 ── mmol/L ── ── double reagent
28 LDL-C End point 546 nm 37℃ 77 ── mmol/L ── ── double reagent
29 TG End point 546 nm 37℃ 2.2 mmol/L ── ── double reagent
30 LDH-L Kinetic 340 nm 37℃ ── 9683 U/L 60 60 double reagent
31 CK Kinetic 340 nm 37℃ ── 8095 U/L 60 120 double reagent
32 CK-MB Kinetic 340 nm 37℃ ── 8254 U/L 60 120 double reagent
33 HBDH Kinetic 340 nm 37℃ ── 4920 U/L 60 60 double reagent
Acute 34 AMY Kinetic 405 nm 37℃ ── 5138 U/L 60 60 double reagent
abdomen 35 AMY Kinetic 405 nm 37℃ ── 5138 U/L 60 60 double reagent
38
DR-7000D Semi-Automatic Chemistry Analyzer
Note:
1. The above info just for regular items, for the details, please check the specification on the reagent kits.
2. User also could program the new parameters according to the specification on the reagent kits.
39
DR-7000D Semi-Automatic Chemistry Analyzer
Appendix E
1. Installation
The bar code reader is the optional spare parts for DR-7000 series semi-automatic chemistry
analyzer, there is a serial port on analyzer back board for it installation. The operation is very easy,
operator only need to insert the bar code reader connector plug into the proper serial port.
Notice: The equipment should be powered off before bar code reader installation.
2. Operation
(1) The bar code reader could read Code-39, Code-128, EAN-8, UPC-E and Inter leaved 2 of 5 bar
code automatically and the total digit for the bar code number should not more than 10 letters,
otherwise, it will not workable.
(2) The operator could do the scan operation before each test (ID should be setup ON) and then the
bar code number will be saved into ID number. There is only one bar code number for each
sample, therefore, it is only need to scan one time for the same sample or the equipment will save
the last scan one as the final one.
(3)The bar code reader belongs to far-distance Reader, the operator only need to aim the scan point
at the barcode number and then press the trigger, so that, the real-time bar code info will be
shown on the screen.
(4)The bar code reader reads the barcode by the light line center on the top of it, to avoid mistakes,
the light line should cover the whole barcode and keep proper bevel with the trigger before
reading.
Instrument could connect EPSON LQ300K+ stylus printer. (Other type stylus printer can be used as
well, but it may cause some change in print speed and character position.)
Rev: 06/2009