Chemistry Analyzer User Guide

The document provides instructions for use and maintenance of the DR-7000D Semi-automatic Chemistry Analyzer. Key points include: 1. The analyzer uses colorimetric testing methods to test samples based on the principles of light absorption. 2. Users should carefully read the manual before operation and maintain the cuvettes by cleaning with alcohol regularly. 3. The manual covers installation, operation, maintenance, and specifications of the analyzer.

Uploaded by

sakata_abera4
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
2K views41 pages

Chemistry Analyzer User Guide

The document provides instructions for use and maintenance of the DR-7000D Semi-automatic Chemistry Analyzer. Key points include: 1. The analyzer uses colorimetric testing methods to test samples based on the principles of light absorption. 2. Users should carefully read the manual before operation and maintain the cuvettes by cleaning with alcohol regularly. 3. The manual covers installation, operation, maintenance, and specifications of the analyzer.

Uploaded by

sakata_abera4
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 41

DR-7000D Semi-automatic Chemistry Analyzer

Instruction:

Thanks for purchase DR-7000D Semi-automatic Chemistry Analyzer.


In order to use the product correctly, please read the user manual carefully before
operation.
Please keep the user manual safely for your anytime reference.

Caution:

Colorimetric cuvette should be marinated by the alcohol at least 60 minutes after self-
checking for the first time working, and then use distilled water to rinse it time by time
for 10 minutes. If not, the test result might be inaccurate.

I
DR-7000D Semi-automatic Chemistry Analyzer

Content

Chapter 1 Introduction.............................................................................................................................
1.1 About DR-7000D........................................................................................................................
1.2 Technical specification................................................................................................................
1.3 Warning Mark.............................................................................................................................
Chapter 2 The Basic Principle and Functional Feature............................................................................
2.1 Basic principle............................................................................................................................
2.2 Functions...................................................................................................................................
Chapter 3 Installation and Application....................................................................................................
3.1 Instrument Installation................................................................................................................
3.2 Main parts and Application.........................................................................................................
Chapter 4 Basic Operation.......................................................................................................................
4.1 Electrification..............................................................................................................................
4.2 Sample testing............................................................................................................................
4.3 Program measurement.............................................................................................................12
4.4 Quality control...........................................................................................................................20
4.5 Setup........................................................................................................................................22
4.6 Record search..........................................................................................................................24
4.7 Rinse.........................................................................................................................................25
Chapter 5 Instrument Maintenance.......................................................................................................26
5.1 Instrument surface cleaning....................................................................................................26
5.2 Colorimetric cuvette cleaning..................................................................................................26
5.3 Lamp-house replacement........................................................................................................26
5.4 Fuse replacement....................................................................................................................27
5.5 Pump pipette replacement.......................................................................................................27
5.6 Print paper replacement..........................................................................................................28
5.7 Printer crankshaft maintenance...............................................................................................28
Chapter 6 Transportation and Storage..................................................................................................29
6.1 Transportation requirements....................................................................................................29
6.2 Storage requirements..............................................................................................................29
Appendix A..............................................................................................................................................30
Appendix B.............................................................................................................................................31
Appendix C.............................................................................................................................................33
Appendix D.............................................................................................................................................37
Appendix E.............................................................................................................................................39

I
DR-7000D Semi-automatic Chemistry Analyzer

Chapter 1 Introduction

1.1 About DR-7000D


DR-7000D was designed in response to the needs of clinical tests. It is a newly intelligentized filter
chemistry analyzer.

The built-in dot matrix LCD display provided with apheliotropic board ensures the clear show of all
test results, functional status and operational instruction. The English indication system guides
users’ operation, and improves man-machine communication.

1.2 Technical specification


a) Lamp-house: 20W /12v Halogen tungsten lamp
b) Wavelength: 340、 405 、 450 、 492 、 505 、 546 、 578 、 630 nm optional
c) ABS: -0.3~3.0 Abs
d) Wavelength accuracy: ±1nm
e) Spectrum bandwidth : 6~12nm
f ) Resolution: 0.1m Abs
g) Linearity: 2Abs :≤ 2% 、 1Abs :≤ 0.5%
h) Stability : 20 minutes for warm-up and then ≤2mAbs/h
i ) Blank: Auto zero upon the colorimetric cuvette with full of water
j ) Interface : RS-232 serial port, parallel printer port
k) Memory :
──Quality control data (valid period: 1 month)
──200 testing programs, 3000 testing results
l) Test methods
──End-point ──Kinetic
──Two-point kinetic ──Absorbance
──Multipoint
m) Measurement time: 0~999 seconds programmable
n) Delay time: 0~999 seconds programmable
o) Flowcell
──Quartz flowcell, measuring volume is 32μl
p) Temperature control
──Semiconductor device control , 25℃ 、 30℃ 、 37℃ and room temperature optional
──Accuracy ±0.1℃
q) Sampling system
──Peristaltic pump, controlled by step motor
──Sampling volume: 100μl~9999μl programmable
──1μl/step sampling volume programmable
r) Power supply: 100~240VAC 50/60Hz

s) Environment condition:
1
DR-7000D Semi-automatic Chemistry Analyzer

──Ambient Temperature: 15℃~30℃, optimal temperature: 15℃~25℃


──Relative Humidity :≤ 85%RH
t) Dimension : 462mm×418mm×232mm
u) Weight: 6.5kg

1.3 Warning Mark

Biohazard mark Warning mark

There are two biohazard marks on the equipment, one is Biohazard mark, another is Force
electricity mark, they are all on the backboard of the equipment. As shown in Figure 1-1.
Biohazard mark

Warning mark
Figure 1-1 Warning Mark

There might be a potential infectivity for the sample & waste reagent, please make sure to take
protection action during the testing, cleaning process and instrument maintenance.

Samples and waste reagent should be disposed in a suitable package conforming to the local
regulation for laboratory.

2
DR-7000D Semi-automatic Chemistry Analyzer

Chapter 2 The Basic Principle and Functional Feature

2.1 Basic principle


The principle of the instrument is based on the phenomenon of different wave band absorbance
from substance, which is in line with Lambert-Bill Law.

A=log(I。 /I)=log(1/T)=∈bc
T= I /I。

T — Transparency c — Solution layer thickness(cm)


I — Transmitted light intensity b — Solution Moore concentration(mmol/ml)
I 。— Incident light intensity ∈ — Solution Moore absorbance coefficient(ml×mmol-1×cm-1)
A — Absorbance

2.1.1 Optical system


The prismatic part of DR-7000D Semi-automatic Chemistry analyzer adopts the filter in unique
research and manufacture techniques. It is featured by moist-resistant, heat-resistant, longevity of
service, little central wavelength excursion, high optical signal-noise ratio etc.. The principle
diagram is shown as Figure 2-1.

Amplifier

Lamp-house Prismatic Flowcell Detector


System

Figure 2-1 Principle diagram of optical system

2.1.2 Electronic system

Flowcell Detector Amplifier


A/D
Monochromator

Driver control Computer system Temperature control

VG
Lamp-house VI Display/Keyboard Printer
Vn

Figure 2-2 Electronic principle diagram


3
DR-7000D Semi-automatic Chemistry Analyzer

The principle diagram of the electronic system is shown above in Figure 2-2. When passing
through the monochromator the ray of the lamp-house will be changed into monochromic light. The
result will be then received by the photoelectric detector when it passes through the flow
colorimetric cuvette, and further transformed into electric signal. The signal will be amplified and
filtered, and then transmitted into A/D transformer. When the information is transformed into
numbers, it will be sent into data collection system of the computer for kinds of calculation and
transaction. Users can either read the final result on the display or alternatively get it printed out.

2.2 Functions
1.Measurement methods :
a) End-point b) Two-point kinetic
c) Kinetic d) Absorbance
e) Multipoint
2. Internal and external printer is available for uses’ options.
3. Sampling volume, test method is programmable.
4. Users can determine whether to execute real time print out of the test results.
5. If any operational mistakes or other false results are found during the process of measurement,
users can choose the repeat key (“●”) to retest, the original serial number of the test remains.
6. Rinsing program is available between two sample tests to avoid cross contamination.
7. Up to 1 month quality control tracking, and the printing results provided.
8. Integrated test report printing is available.
9. Reaction curve display:
Reaction curve for sample and standard liquid will be displayed on the screen during the testing, as
shown in Figure 2-3.

ABS LIMIT
Absorbance Absorbance

ABS LIMIT
0 0
t1 time t1 time
(a)Positive reaction curve ( b ) Negative reaction curve

Figure 2-3 Activity reaction curve

Note: 1. Press “+/-” to check the reaction curve after testing the standard liquid or sample.
2. Press “Exit” to test next sample.

4
DR-7000D Semi-automatic Chemistry Analyzer

Chapter 3 Installation and Application

3.1 Instrument Installation


3.1.1 Installation
A. Move all the spare parts out after opening the package and then check the spare parts quantity
according to the packing list, if there are some spare parts broken please contact the delivery
forwarder or manufacturer.
B. Place the equipment on the selected work bench where is no vibration.
C. The equipment should not be placed close to the chemicals, caustic gas and electromagnet.
D. Prevent direct sunshine, humidity and high temperature.
E. Operating temperature range is 15℃ ~30℃ (optimal range: 15℃ ~25℃), relativity humidity is
less than 85%RH.
F. The power supply of the instrument is 220V/50Hz and the plug is on the back-board, see Figure
3-1.
The instrument is provided with special two cores electric wire, which should be connected with
well grounded receptacle after plugging in.

① ② ③

④ ⑤ ⑥ ⑦ ⑧

1. Power switch 2. Power socket 3. Fan 4. Fuse socket


5. Warning mark 6. Serial port 7. Parallel Port 8. Biohazard mark
9. Waste pipette
Figure 3-1 Back-board outline

3.1.2 Cautions for use


● Read the user manual carefully before working , especially for the caution & hazard info.
● Pre-heat for 20 minutes before working and use distilled water to wash the equipment before
any test.
● Wash over 3~5 times immediately after each test. After finish working, wash time after time
before turning off the instrument.

● Make sure to maintain some cleaning agant in the colorimetric cuvette when it stops working to
5
DR-7000D Semi-automatic Chemistry Analyzer

avoid the sample remain in the colorimetric cuvette.


● The colorimetric cuvette should be washed and dipped everyday, firstly use cleaning agent or
distilled water to wash it for several times, and then remain some cleaning agant in it.
● When doing sample feeding operation, please pay attention to the sucking pipette and waste
pipette, to avoid damaging and blocking.
● Prevent vibration and direct sunshine and keep way from the wall at least 20cm.

3.2 Main parts and Application


A. Display
The display of DR-7000D Chemistry analyzer is dot-matrix large angle of view LCD, which adopts
the latest STN technology. Its dimension is 108x58mm with dot-matrix 240x128. In addition, the
apheliotropic board ensures clear reading in a poor light available.

B . Keyboard
The keyboard comprises 19 key-presses, i.e. 12 number keys (including ± and radix point), 5
functional keys and “EXIT” and “ENTER” keys.
Number key — Entering into submenu or numeric parameter input.
ENTER — Press to confirm the executive command.
EXIT — Press to return to the former menu.
F1 — F5 (Functional key) — Press to execute the relevant functions indicated in the five small
squares at the bottom of the display.

C. Buzzer
1. As the key-press on the control board is made of pellicle without clear tactility, buzzer reproducer
is used to help inform operator as soon as the key-press is touched.
2. The buzzer will remind the user to move the sample away once the equipment finishes sampling.

D. Sample feeding switch and operation


The white sucking pipette is on the left of the instrument, and the switch of sample feeding is under it.
Sample feeding operation: During the measurement, user can follow the instructions on the screen
to do this operation, first put the sucking pipette into the bottom of prepared samples (including
distilled water, reagent blank, standard liquid, sample blank, sample, quality-control sample), press
the switch button and then the samples will be sucked automatically. Don’t move the sample until 2-
3 seconds later after the buzzer shows sample feeding is finished.

E. Printer
Both of internal printer and external printer are available for the instrument. The user can select one
of them for operation. There are 3 status optional: ON, OFF and AUTO. Real-time printing is
available once select AUTO status.

Refer to “5.6” for the replacement of thermal sensitive print paper.

6
DR-7000D Semi-automatic Chemistry Analyzer

Chapter 4 Basic Operation

4.1 Electrification
The power switch of the instrument is on the back-board (see Figure 3-1). Turn on the power switch
to get the instrument in the state of electrification, and then the info in Figure 4-1-1 and 4-1-2 will be
displayed on the screen orderly.

Semi-Automatic
Chemistry Analyzer
Version x.x x

Feb 6 2007

Figure 4-1-1

05—07—15
19:13:11
Self checking …
Please pump distilled water
Figure 4-1-2
Note: Under the Self checking interface(as shown in Figure 4-1-2), press the sucking button to
enter into the submenu as shown in Figure 4-1-3.

05—07—15
19:13:20
Self checking…
Please waiting
Figure 4-1-3

05—07—15
19:13:51
warm-up for 20 minutes should be done before testing
if not, test results might be inaccurate !!

Figure 4-1-4

Reminder info (see Figure 4-1-5) will be shown on the screen after self checking. The user also
could press “Exit” to omit it to enter into the test interface directly (To avoid inaccurate test result,
please don’t omit this process).

Clear all sample data?

Yes- press Enter


No- press
7 Exit
DR-7000D Semi-automatic Chemistry Analyzer

图 4-1-3
Figure 4-1-5
Key in the final selection and then enter into the program main menu, as shown in Figure 4-1-6.
Note: If select Yes, all the test results will be deleted except for QC results and the sample serial
number will be reordered from 1 automatically.
If select No, the system will back to the main menu, as shown in Figure 4-1-6 :

1. Sample testing
2. Parameter setup
3. Quality control
4. Setup
5 .Record search
6. Rinse

Figure 4-1-6

There are 6 submenus under the main menu :


Press “1” to enter “Sample testing ” mode to do normal test ;
Press “2” to enter “Parameter setup” mode to edit and modify relevant test parameters.
Press “3” to enter “Quality control” mode to edit the control parameters or look over control
statistics.
Press “4” to enter “Setup” mode to setup the language, printer status, time, date, ID and lamp-
house luminosity.
Press “5” to enter “Record search” mode for the search of relevant record.
Press “6” to do the rinsing operation.

4.2 Sample testing


Under the main menu, press “1” to enter item display submenu. As shown in Figure 4-2-1.
There are 10 sorts for the test items: LIVER, KIDNEY, CARDIAC ENZYME, BLOOD, BLOOD
LIPID, HYDRONIUM , GROUP 1, GROUP 2, GROUP 3, GROUP 4; There are 20 items under
each sort and 200 items totally. Funtion key F4 and F5 are used for page up and page down.

The skip point at the first item after enter into the test item interface, there are 2 ways to select the
item.
Way 1: Input the item number (such as 001).
Way 2: Press the funtion key F1 and F2.

LIVER :
001)ALT 011)r-GT
002) ALP 012)CHE
003) T-BIL 013)D-BIL
004)ALB 014)TP
005)BUN 015)BUN
006) Cr 016)AST
007) CO2 017)Ca
008) CL 018)P
8
009) Fe 019)Mg
010) GLU 020)CHO

Pre. Next 002 Up Down


DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-2-1

Note : If inaccurate item number is input, it will be corrected by key the accurate one under the
same interface directly and the original one will be covered automatically, the item no. should be 3
digits.

Input the selected Item Number. (such as: 003 or press F2 to select 003) and then press “Enter” to
enter into the test status. As shown in Figure 4-2-2.

Test name:ALT 05—07—16


Mode:KINETIC
Sample No.:001
Medium: Sample
ABS:
Factor:-3376
Result:
Please Pump distilled water
Rinse Print Calibrate Zero QC
Figure 4-2-2

Press the sucking button to suck the distilled water and then the following info (as shown in Figure
4-2-3) will be shown on the screen. Please follow the screen reminder info to do the test operation
one by one.

Test name:TP 05—07—16


Mode: ENDPOINT
Sample No.:001 ID:
Medium: sample
ABS:
Factor :
Result:
Please pump sample

Rinse Print Calibrate Zero QC


Figure 4-2-3

Note:
a) Press sucking button to make the distilled water, reagent, sample, standard liquid and quality
control liquid are all sucked into the colorimetric pool by sucking pipette according to the
reminder info one by one, and then test results and ABS value will be shown on the screen
after finishing measurement. If printer status is ON, press F2 to print the test results out and
press 8 to run the print paper; If printer status is AUTO, the printer will execute real time test
9
DR-7000D Semi-automatic Chemistry Analyzer

result print-out. Test results will be saved automatically, if the test results exceed 3000 pieces,
the reminder info “THERE IS NO SPACE FOR DATA BASE, CLEAR ALL SAMPLE DATA?”,
press “EXIT” to out and press “ENTER” to clear it. The equipment will not be workable unless
do the clear operation.
b) One sample serial number is only correspond with one patient, it will be added 1 after each
testing automatically. Further more the operator also could press ENTER button to enter into
sample edit submenu, as shown in Figure 4-2-4, and then input a new sample serial number.
(within 001-999).

__ __ __ __

Input sample No.

Figure 4-2-4
Press ENTER after inputing the sample number and it will be the serial number for the next
testing sample.
c) If ID STATUS is ON, there are 2 ways to input the ID number, one is by barcode reader, another
is by press number 9 button to enter into the ID edit interface (as shown in figure 4-2-5), and
then input the ID number (0~9) by hand (the ID number should not more than 10 digits),
otherwise, there will be no ID number.

__ __ __ __ __ __ __

Input ID No.

Figure 4-2-5
Note:
(1) There is only one ID correspond with one daily sample number. One time ID number input is
avaiable for the samples which use the same sample serial number.
(2) There is no ID for new sample, repeat ID input is avaiable and the system will keep the final one
as the ID number.
d) Repeat test could be done after pressing the repeat key “●” and the original sample no. will not
be changed.
e) If quality control or re-calibration is required, press “F5” or “F3” and then “Pump standard liquid
again?” or “Pump control liquid again?” info will be shown on the screen
f )For the end point measurement, reminder info “Test the reagent blank again ?” will be shown
on the screen after setting parameters again, user could select the proper info, if select “1” the
reagent blank value will be tested again or it will be calculated according to the original value
by the equipment. As shown in Figure 4-2-6.
Test name:TP 05—07—16
Mode: ENDPOINT
Sample No.: ID:
Medium:
ABS:0.02
Factor : 1-YES
10 Result: 2-NO
Test reagent blank again?
DR-7000D Semi-automatic Chemistry Analyzer

Rinse Print Calibrate Zero QC


Figure 4-2-6
Enter into standard liquid test interface after finishing reagent blank test and the reminder info “Test
standard liquid again?” will be displayed on the screen, as shown in Figure 4-2-7 .

Test name:TP 05—07—16


Mode:ENDPOINT
Sample No.
Medium:
ABS:0.002
Factor: 1-YES
Result: 2-NO
Test standard liquid again?
Rinse Print Calibrate Zero QC
g) For the END point, two poin and Kinetic measurement,
Figure 4-2-7if the test time is not 0, the reaction curve
will be displayed on the screen in real time. After displaying the test results, press “±” button the
enlarge reaction curve will be displayed on the screen.

4.3 Program measurement


Under the main menu (as shown in Figure 4-1-6), press “2” to enter test program submenu, as
shown in Figure 4-3-1.

LIVER :
001)ALT 011)AMM
002) AST 012)CHE
003) TP 013)GLDH
004)ALB 014)ICDH
005)T-BIL 015)
006) D-BIL 016)
007) TBA 017)
008) ALP 018)
009) GGT 019)
010) LAP 020) 4-3-1
Figure
Pre Next 001 Up Down
4.3.1 Parameter setup:
For example, input “003” or press F2 (“↓”) 2 times to select item “TP” and then press “Enter” to
enter into its submenu. As shown in Figure 4-3-2.

Program
1 、TEST NAME : TP P1
2 . TEMP : -- -- ℃
3 . MODE : END POINT
4 . UNITS : g/L
5 . L—LOW:0.0000
6 . L—HIGH: : 150

Figure 4-3-2
UP DOWN

The content in the bracket is the item instruction of program (Reminder info.) “P1”, “P2”, “P3” “P4”
11
DR-7000D Semi-automatic Chemistry Analyzer

on the top of the screen means the page number, Press F4, F5 to do page up and page down
operation, press 1-6 to edit and modify the content of relevant items.
There are totally 5 pages of program submenus, “P1” in Figure 4-3-2 means Page 1, press F5 to
view Page 2 (see Figure 4-3-3), Page 3 (See Figure 4-3-4) and Page 4 (See Figure 4-3-5).

Program P2

1 、R-BLANK : NO
2 . S-BLANK : NO
3 . STANDARD : 0.0
4 . FACTOR : 0.1
5 . VOLUME : 500
6. FLOAT: 002

Figure 4-3-3
UP DOWN

Program P3

1 、WL1 : 000
2 、WL2 : --------
3 、Delay ( S ) : 000
4 、MEAS ( S ) : 000
5 、REF-LOW: 0.1
6 、REF-HIGH: 0.1

Figure 4-3-4
UP DOWN

If select the Multipoint method, concentration parameter under P4 should be input, however, it is no
need for other measurement methods.

Program P4

STANDARD
P3 1: 0.000
STANDARD2: 0.000
STANDARD3: 0.000
STANDARD4: 0.000
STANDARD5: 0.000

Figure 4-3-5
UP DOWN

4.3.2 Parameters modification :


A) Modifying test name: Under the program submenu ( as shown in Figure 4-3-2), press “1”
and then press “ENTER” to enter into Program submenu, as shown in Figure 4-3-6:
Program ( Test name )

1 、Test name :
0 1 2 3 4 5 6 7 8 9
0 1 2 3 4 5 6 7 8 9
A B C D E F G H I J
12
K L M N O P Q R S T
U V W X Y Z — ()

↑ ↓ Del Page
DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-3-6

As known from above, ‘↑’ corresponds to F1, ‘↓’ corresponds to F3, ‘Delete’ corresponds to F4,

‘PA’( Page down) corresponds to F5.

For example, to change the name into GPT: Press F3 (or F1) to move the cursor to the line where
there is ‘G’, then press 6 (‘G’ is correspond to 6), ‘G’ will be displayed in the Name Column. Press
F3 (or F1) to move the cursor to the line where there is ‘P’, then press 5 (P is correspond to 5) ‘GP’
will be displayed in the Name Column, and so forth. When all finished, press ENTER to return to
the last submenu (if the input is more than 8 letters, it will return to last submenu automatically). If
any letter is entered incorrectly, press F4 to delete the letter. For lowercase entering, press F5 to
look up lowercase on the next page.

B ) Modifying Temperature : Press 2 under the submenu on the first page (as shown in
Figure 4-3-2), then press ENTER, the screen is shown as Figure 4-3-7 :

Program :( Temperature )

2 、TEMP : ℃
0 、──── ℃
1 、────25℃  
2 、────30℃
3 、────37℃
Figure 4-3-7

Key in the number in front of the selected temperature, then press ENTER to return to the
submenu.

C ) Modifying measure method : Press 3 under the submenu on the first page, then press
ENTER, the screen is shown as Figure 4-3-8.

Program ( Test mode )

3 . MODE :
0 ────ENDPOINT
1 ————TWOPOINT
2———— KINETIC
3———— MULTI-POINT
4———— ABS
Figure 4-3-8
Select the proper number and then press ENTER to back the upper menu.
13
DR-7000D Semi-automatic Chemistry Analyzer

D ) Modifying units : Press 4 under the submenu on the first page, then press ENTER. As
shown in Figure 4-3-9 :

Program ( Units )

4 . UNITS :
0 . ABS 4.μmol/L
1 . U/L 5.g/L
2 . μkat/L 6.g/dL
3 . mmol/L 7. mg/dL

Figure 4-3-9
Page
F5 is for page down ,as shown in Figure 4-3-10:

Program ( Units )

4 . UNITS :
0 . μg/dL 4.μg/L
1 . μU/mL 5.ng/L
2 . nmol/L 6. %
3 . mg/L

Figure 4-3-10 Page

Key in the number in front of the selected unit, then press ENTER to return to the submenu (as
shown in Figure 4-3-2).

E) Modifying linearity range: Press 5 under the submenu on page 1, then press “Enter” As shown
in Figure 4-3-11,4-3-12:

 ̄ ̄ ̄ ̄ ̄ ̄ ̄

Input L-LOW value: Figure 4-3-11


 ̄ ̄ ̄ ̄ ̄ ̄ ̄
 ̄   ̄  ̄  ̄  ̄ 


FigureL-HIGH
4-3-12 value:
Input
Input the proper parameter ̄   ̄  ̄  ̄  ̄ 
and then press “Enter” to back to the submenu (as shown in Figure 4-3-
2).
Note:
Input the linearity range according to the specification on the reagent kits, if the parameters exceed
14
DR-7000D Semi-automatic Chemistry Analyzer

the range during test, reminder mark “!” will be printed on the test report.
“!” means “Test the sample again after dilution and the result should be multiplied the dilution
multiple”.

F ) Modifying Reagent Blank : Press 1 under the program menu on the second page as
shown in Figure 4-3-3 , then press ENTER, as shown in 4-3-13:

Program ( Reagent Blank )

1.R-BLANK :
0─── NO
1─── YES
Figure 4-3-13

Press 0 or 1 for R-blank Yes or No, then press ENTER to return to the submenu ( as shown in
Figure 4-3-3).

G ) Modifying sample blank : Press 2 under the program menu on the second page (as
shown in Figure 4-3-3), then press ENTER, as shown in Figure 4-3-14:

Program ( Sample blank )

2.S-BLANK :
0─── NO
1─── YES
Figure 4-3-14

H ) Modifying standard liquid concentration: Press 3 under the submenu on the second
page (as shown in Figure 4-3-3), then press ENTER, as shown in Figure 4-3-15 :

 ̄ ̄ ̄ ̄ ̄ ̄ ̄

Input standard liquid


Figure 4-3-15
Input the proper standards liquid concentration, then press “Enter” to back to the submenu.

Note:
The factor will be turned into ‘----’ automatically after entering the standard liquid concentration.
The factor will be saved automatically and reminder info “Confirm the standard liquid again?” will be
shown on the screen once the user opens the submenu again, if select “yes”, the old coefficient will
15
DR-7000D Semi-automatic Chemistry Analyzer

be replaced by the new one.

I) Modifying Factor : Press 4 under the program menu on page 2 (as shown in Figure 4-3-3),
then press “Enter”. The factor is “00000.” input proper value, then press “Enter” to return to the
submenu. As shown in Figure 4-3-16.

 ̄ ̄ ̄ ̄ ̄ ̄ ̄

Input factor Figure 4-3-16


Note : The standard concentration will turn into ‘-----’ after entering the factor value automatically.

J ) Modifying the sampling volume : Press 5 under the program menu on Page2 (as shown
in Figure 4-3-3), then press ENTER. Input the proper number, then press ENTER to return to the
submenu. As shown in Figure 4-3-17.

uL

Input sample volume Figure 4-3-17


Note: the min. sample volume is 100μl, however, in order to reduce the cross-contamination of
samples with high concentration, it is better to adjust it to 500μl at least.

K ) Modifying the decimal digits of test results: Press 6 under the program menu on page 2
(as shown in Figure 4-3-3), then press Enter to input the required decimal digits, then press Enter
to turn to the submenu. As shown in figure 4-3-18. (Decimal digits is the numbers after the radix
point of the test result, the max. one is 4 digits such as 1.2334).

Input float 0-4


__

Decimal digi
Figure 4-3-18
L) Modifying wavelength : Press 1 under the program menu on Page 3 (as shown in Figure 4-
3-4), then press ENTER. Input the selected wavelength number, if correct, press ENTER to back to
the submenu (sub. wavelength press 2), as shown in Figure 4-3-19.

WAVE :
1 . 340 5 . 505
2 . 405 6 . 546
3 . 450 7 . 578
4 . 492 8 . 630

16 Input wavelength 1:

 ̄     
 ̄ ̄  ̄ ̄  ̄ ̄  ̄ ̄
n m ̄
DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-3-19
Note: There will be a reminder info “Input wavelength again” on the screen, once main wavelength
value input is unavailable and the operator should input the wavelength again.

M) Modifying delay time : Press 3 under the main menu on page 3 (as shown in Figure 4-3-
4), then press ENTER. Input the proper time, if correct, press ENTER to back to submenu. (Delay
time setup should be according to the specification on the reagent kits), as shown in Figure 4-3-20.

___ ___ ___ S

Input delay time Figure 4-3-20

N) Modifying test time : Press 4 under the main program menu on page 3 (as shown in figure
4-3-4), then press “ENTER, input proper number, if it is correct, press ENTER to back to the
submenu. As shown in Figure 4-3-21.

___ ___ ___ S

Input measuring timeFigure 4-3-21

Note: Measure time is only applied when using two-point kinetic and kinetic mode.

O) Modifying reference high : Under the main program press 6 on page 3 (as shown in figure
4-3-4), then press ENTER. Input the proper number , if it is correct, press ENTER to back to the
submenu ,as shown in Figure 4-3-22.

________________________

Input REF-High Figure 4-3-22

P) Modifying reference low : Under the main program press 5 on page 3 (as shown in figure 4-
3-4, then press ENTER. Input the proper number , if it is correct, press ENTER to back to the
submenu , as shown in Figure 4-3-23 :

17
DR-7000D Semi-automatic Chemistry Analyzer

__________________________

Input REF-Low Figure 4-3-23


Note: All parameters above should be selected according to the specification on the reagent kit. (If
test result is not within the ref value range, “↑” & “↓” reminder info will be marked behind the result.)

4.4 Quality control


Under the program main menu (as shown in figure 4-1-6), press 3 to enter into the Quality Control
submenu , as shown in Figure 4-4-1:

Quality Control :

1 . Program :
2 . Control statistic :
3 . Clear data :

Input No. : Figure 4-4-1

4.4.1 Program quality control


Press 1 under the quality control submenu, then press ENTER to enter into the submenu, as
shown in Figure 4-4-2:

LIVER :
001)ALT 011)AMM
002) AST 012)CHE
003) TP 013)GLDH
004)ALB 014)ICDH
005)T-BIL 015)
006) D-BIL 016)
007) TBA 图 1-1017)
008) ALP 018)
009) GGT 019)
010) LAP 020)
Figure 4-4-2
Pre Next 001 UP DOWN

Input the selected Item number then press ENTER to enter into the submenu, as shown in Figure
4-4-3, then input the proper number, if it is correct, press ENTER, as shown in Figure 4-4-3.

Control name : ALT


Control lot 1
Target value 1
SD 1:
Control lot 2:
Target value 2:
SD 2:
18
Input No.:
DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-4-3

Input the proper number of control lot, target value and standard deviation one by one. If they are
correct, press ENTER to back to main menu. Two types of control liquid are available for the
program.

4.4.2 Modifying control statistics


Under the program main menu, press 2 to enter into the Control Statistic submenu, as shown in
Figure 4-4-4.

1. Present control data


2. Individual control data

Input No.
Figure 4-4-4

Press 1 ----- Present control data will be shown on the screen.


Press 2 -----Individual control data will be shown on the screen.
Press F2 or “↓” to select the edited QC item and then press ENTER to enter into the submenu, as
shown in Figure 4-4-5.

001)ALB 3SD
Tested 2SD
X=
SD = X
CV =
DAY = 3 -2SD
C03 = 0.0 -3SD
Standard:
X=
SD=1 5 15 25

Figure 4-4-5
QC 1 Print ← →
Press “←” or “→” to view the control data within 1 month.

4.4.3 Statistic data clearance


Press 3 under the program main menu (as shown in Figure 4-4-1), then press ENTER to enter into
the submenu, as shown in Figure 4-4-6:

1. All control data


2. Individual control data

Input No. Figure 4-4-6

4.4.3.1 All control data clearance


Press 1 under the main menue (as shown in Figure 4-4-6) to enter into the submenu, as shown in
Figure 4-4-7.

Clear all statistic data?


Yes-Press Enter
19
No-Press Exit
DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-4-7

Press “ENTER” to clear all the statistic data and then back to upper menu, as shown in Figure 4-4-
6.

4.4.3.2 Individual control data clearance


Press 2 under the main menue (as shown in Figure 4-4-6) to enter into the individual control data
submenu as shown in Figure 4-4-8.

Clear the control data ?


Yes-Press Enter
No-Press Exit

Figure 4-4-8
Press “Enter” to enter into the submenu, key in the selected number and then press ENTER to
clear it.

4.5 Setup
Under the program main menu, press 4 to enter into the submenu, as shown in Figure 4-5-1.

1 . Language-English
2 . Printer
3 . Date & time
4 . Lamp-house replacement
5. ID: ON
6. Resume:

Setup Figure 4-5-1

4.5.1 Printer
Under the setup menu (as shown in figure 4-5-1), press 2 to enter into the printer submenu, as
shown in Figure 4-5-2:

1 . Internal printer : ON
2 . External printer : OFF
3. Printer space: 2

Printer state Figure 4-5-2


Press 1 to select the status of internal printer, then press ENTER to back to the submenu.
20
DR-7000D Semi-automatic Chemistry Analyzer

Press 2 to select the status of external printer, then press ENTER to back to the submenu.
Press 3 to select the run paper distance (per time) and the printed test result line-distance, there
are 5 values optional: 2, 4, 6, 8, 10 line.

4.5.2 Date & Time


Press 3 under the setup menu (as shown in Figure 4-5-1) and then press ENTER to enter into the
submenu as shown in Figure 4-5-3:

2005-05-19 16:30:25
MM-DD-YY

Date & time setup Figure 4-5-3


Input the proper number, then press ENTER to back to the submenu.

4.5.3 Lamp- house replacement


Press 4 under the setup menu (as shown in Figure 4-5-1) to enter into the lamp-house submenu,
as shown in Figure 4-5-4:

Lamp-house

Luminosity:

Please pump distilled water

Figure 4-5-4
Program indicates pump, and set wavelength 340nm automatically, then read the number of the
luminosity continually. Press EXIT to back to last menu.

4.5.4 ID setup
Press 5 under the Setup menu (As shown in Figure 4-5-1) to set up the ID and then press ENTER
TO back the upper menu.

4.6 Record search


Under the program main menu (as shown in Figure 4-1-6), press 5 to enter into the record search
submenu, as shown in Figure 4-6-1:

1. Daily test results


2. Sample No.
Test date
ID
5. Record clearance

Search mode: Figure 4-6-1

4.6.1 Record search


21
DR-7000D Semi-automatic Chemistry Analyzer

Under the RECORD SEARCH submenu( as shown in Figure 4-6-1), search function is avaiable
once press any key within 1 to 4, such as press 1 ,the following interface will be shown on the
screen.

05-07-16 15:55:08
Test name:
Sample No.: 001
ID:
REF-HIGH:83.00
REF-LOW:60:00
UNITS:
RESULT:0.0
Single Print Del Prior Next
Figure 4-6-2
Press F1 to select the printing report format (single or integrate).
Press F2 to print the test report.
Press F3 to delete the current note.
Press F4 to display the previous record.
Press F5 to display the next record.
Press EXIT to back to the upper menu (as shown in Figure 4-6-1).

4.6.2 Record clearance


Under the record clearance menu (see Figure 4-6-1), press 5 to enter into the record clearance
submenu, as shown in Figure 4-6-3:

Clear all records

Figure 4-5-6

Record clearance
Figure 4-6-3

Press 1 to clear all the records and press EXIT to back to upper menu as shown in figure 4-6-1.

4.7 Rinse
Under the program main menu (as shown in Figure4-1-6), press 6 to enter into RINSE submenu,
as shown in Figure 4-7-1.

___ ___

Input wicking counts

22
DR-7000D Semi-automatic Chemistry Analyzer

Figure 4-7-1

Input the rinse time by hand and then press ENTER. The equiment will do the rinse operation
according to the selected time automatically. (Rinse time scope 1~99)

Chapter 5 Instrument Maintenance

5.1 Instrument surface cleaning


Use the wet cloth and cleaning agent to clean the instrument. To avoid damage the DOS shell and
the liquid flow into the equipment, prohibit using any corrosive impregnants.

5.2 Colorimetric cuvette cleaning


(1) The colorimetric cuvette should be washed by the cleaning agent or distilled water between
each test.
(2) After finish working everyday, wash several times by the cleaning agent to rinse the remains in
the colorimetric cuvette.
(3) To avoid serious contamination, adjust the temperature to 37 ℃ to accelerate the cleaning
speed, first in cleaning agent, then in distilled water to wash it several times.
(4) Make sure to maintain some distilled water in the colorimetric cuvette when the equipment stops
working everyday.

5.3 Lamp-house replacement


(1)
Turn off power firstly when replacing Lamp-house.
(2)
Push the sucking pipette out of the instrument smoothly, and then take down the screw from the
bottom of the instrument. Lift the DOS shell smoothly from the back shell (do not take it off
rapidly), leaving the front on the table. Pull out the sucking pipette from inside completely
before DOS shell is lifted up.
23
DR-7000D Semi-automatic Chemistry Analyzer

(3)
Take off the lampshade and pressure pad first ,and then take down the illuminant. (Warning: light
bulb and ambient temperature may be very high, to avoid dangers, please take gloves to do
this operation).
(4)
Insert the new bulb (12V, 20W) in the lamp stand without touching the top of the light bulb, and then
screw the pressure pad. To avoid effecting the test results, please do not touch the bulb by
hand directly. Use alcohol cotton to clean it.
(5)
Under such circumstance, switch on the power to enter into setup menu ,then press 4 to enter into
the lamp house replacement submenu, according to the reminder info to pump the distilled
water and then the Luminosity value will be shown on the screen.
(6)
Adjust the position of lamp-house until the figure on the bottom of the display turned into maximum
(Reference range 18000~50000).
(7)
Ensure the figure on the bottom of the display unchanged, and tighten all screws.
(8)
Turn off the power, and put on the lampshade. Thus the illuminant replacement is finished.
(9)
Insert the pipette properly. Put DOS shell on and tighten all screws.

Figure 5-3-1

5.4 Fuse replacement


The fuse socket is on the back board of the instrument, below the power switch and on the top of
the entrance of electric wire. As shown in Figure 5-4-1. Pull out the electric wire, pick off the fuse
socket to replace the fuse offered by manufacturer.

24
DR-7000D Semi-automatic Chemistry Analyzer

Fuse socket Fuse

Figure 5-4-1

5.5 Pump pipette replacement


Open the DOS shell, the method is same to the lamp-house replacement (see step1&2). Press the
stay bar to open the block board under the pump, move the pump pipette out and then replace a
new one.
To avoid leaking, tighten the screw to tense the pipette properly.

Pump pipette:
Diameter : 2.06mm ( inner )
Length : 90mm

Stay bar Block board pump bearing

Figure 5-5-1

5.6 Print paper replacement


SPRM printer is provided with paper scroll before leaving factory, but it is not been inserted. This is
for the safety of the printer during the extended transportation and storage. Insert the paper scroll
25
DR-7000D Semi-automatic Chemistry Analyzer

before use, refer to the following steps:


(1)
Press the cover below the paper exit gently, the cover will flick automatically and be open.
(2)
Cut the paper as Figure 5-6-1.
(3)
Switch on the printer power, insert the paper into the paper entrance, it will enter into the crankshaft
automatically.

Paper Entry

Stay bar

Figure 5-6-1

5.7 Printer crankshaft maintenance


In order to ensure the normal work of the printer, please do not take the crankshaft down
discretionarily.
(1)
When replacing the paper, check whether any wastepaper maintains on the crankshaft, blow off it if
there is any.
(2)
Ensure the clean of external board.
(3)
When thermal sensitive printer is dirty (print results is not clear), clean it with alcohol cotton.
(4) Do not pull the paper emphatically before the crankshaft is put up.

Chapter 6 Transportation and Storage

6.1 Transportation requirements


1. Anti-moisture and anti-water works should be done well before transportation.
2. Avoid strong shake and extrusion.
3. Take care to move, pack or unpack the instrument.

6.2 Storage requirements


1. Indoor
2. Temperature: -40℃~ +50℃
3. In a well-ventilated, clean and dry area, away from chemicals and caustic gas.

Cautions:
The instrument should be rinsed several times and the sucking pipette should be kept empty & dry
26
DR-7000D Semi-automatic Chemistry Analyzer

before transportation or storage. If the transportation ambient temperature is lower than 0℃, it is
better to use 95% alcohol to rinse the pipette in order to avoid frost crack.

Appendix A

Manufacturer’s warranty

Dear Customer :

Thanks for purchasing DR-7000D semi-automatic chemistry analyzer, our company will provide the
following services for you:

1. Technical consultations are provided at any time.


2. Maintenance free of charge within one year from the day you purchase the instrument.
3. Maintenance will be charged in the following conditions.

(1) Product which has expired the warrantee date.


(2) Damage caused by accidental factors or improper use.
(3) Damage caused by the operation that is not according to the user manual.
(4) Damage caused by your own repair without our company’s permission.
27
DR-7000D Semi-automatic Chemistry Analyzer

4. We will provide updated service for free according to the technology improvement.

Any technological service requests, please contact us according to the following address:

DIRUI INDUSTRIAL CO., LTD.


95 Yunhe street, New & High Tech. Development Zone
Changchun, JiLin 130012 P.R. China
Tel: 86-431-85100409
Fax: 86-431-85172581
E-mail:dirui@dirui.com.cn
http://www.dirui.com.cn

Appendix B

Trouble shooting

Fault Reason Trouble shooting


1. The fan and W lamp do The fuse is broken. Replace the fuse (2A).
not work after switching on
the power.
2. After switching on the W lamp is broken or First check whether the voltage for
power, the fan works but W without voltage. lamp house is 11.0V. If the voltage is
lamp doesn’t work. as normal, then W lamp is broken,
replace a new one.

3. Tested result is not good. Sample feeding generates After several rinsing or sample
air bubble, pump delivery feeding, if it still doesn’t work properly,
and speed is instable. wash it with specific cleaning agent for
2-3 times, then with distilled water for
5-10 times. Re-input the pump
parameters.

28
DR-7000D Semi-automatic Chemistry Analyzer

4. The printer doesn’t work Printer plug falls out or is Open the DOS shell, reinsert plug and
and display E1, E2, E3, E4 loose connection. check if the light for printer board
etc info. flashes.

5. The pump does not run Switch damaged or loose Open the DOS shell, reinsert plug or
after switching on the connection. replace the switch.
sample feeding power.
6. Sampling abnormity after Interface from the outside Turn off the power and restart the
system self-checking. electric net. equipment after several seconds.

7. Temperature abnormity Temperature sensor Turn off the power in normal room
after system self-checking. socket is not connected temperature, restart it after 5
well or the temperature minutes.If it is not workable, open the
exceed the scope of equipment’s shell and refix the sensor
5℃~40℃. socket.
8. The buzzer doesn’t work. Computer shuts down, Turn off the power, and restart a few
which is usually caused by seconds later.
external interference.

9. some a key does not Key cover is used too Replace the key cover.
work. much.
10. Display with variegation It is usually caused by Turn off the power, and restart a few
after switching on. external interference. seconds later. Or open case, reinsert
all cables connection.

11. The pump works Flow system is blocked or Check that if the input and output
without sample feeding. pipette aging or Blocked. pipette are twisted or extruded, and
check if pump pipette is tightened or
aging. To replace with a new one or
cut it shorter. And clean and
impregnate flowcell.

12. Instrumet alarm. Re-operate based on manual. If it


happens often, please contact the
supplier.

29
DR-7000D Semi-automatic Chemistry Analyzer

Appendix C

Communication Interface with Computer

Baud rate: 9600

Data bit: 8 bits

Stop bit: 1 bit

Verification: none

30
DR-7000D Semi-automatic Chemistry Analyzer

Hardware hold hands: none

Start character: 02H

2 bytes space: 20H

Line feed character: 0DH0AH

Ending character: 03H

Connection between chemistry analyzer and computer :

computer Chemistry analzyer port


port ( RS232 )

31
DR-7000D Semi-Automatic Chemistry Analyzer

Test report data output format :

1 、 Data output format :( individual data )

1 2 3 4 5 6 7 8 9 10 1 12 1 1 15 16 1 18 1 20 21 2 23 2 2 26 2 28 2 30
1 3 4 7 9 2 4 5 7 9
STX C LF                                                
R
SP S SP T e s t i n g S R e p o r t CR L                
P P F
CR L                                                  
F
SP D a t e : × × × × - × × - × × S × × : × × : × × CR L
P F
SP S a m p l e S N O . : × × × CR L                    
P F
SP I D : × × × × × × × × × × CR LF                      
SP T e s t S N a m e : × × × × × × × × CR LF            
P
SP R E F : S × × × × × × × × × × × × × × × C LF        
P R
SP S e q . S S R e s u l t S SP U n i t SP SP H i n t CR L
P P P F
SP × × × S × × × × × × × × × SP × × × × × × × SP × × CR L
P F
CR L                                                  
F
ETX                                                    

34
DR-7000D Semi-Automatic Chemistry Analyzer

Note :
SP------- space character (20h hex)
X--------- virtual value bit
STX----- start character (02h hex)
CR------- return character (0dh hex)
LF-------- line feed character (0ah hex)
ETX------ ending character (03h hex)

2 、 Data input format :( Integrated data )

1 2 3 4 5 6 7 8 9 1 1 1 13 1 1 16 1 1 1 20 2 22 23 2 2 26 27 2 2 30
0 1 2 4 5 7 8 9 1 4 5 8 9
STX CR L                                              
F
SP SP Q u a l i t y S C o n t r o l S R e p o r t C L
P P R F
CR LF                                                
SP D a t e : × × × × - × × - × × S × × : × × C L    
P R F
SP T e s t S N a m e : × × × × × × × × CR L          
P F
SP I D : × × × × × × × × × × C LF                    
R
SP S t a n d a r d S X : SP × × × × × × × × CR L      
P F

35
DR-7000D Semi-Automatic Chemistry Analyzer
SP S t a n d a r d S S D : S × × × × × × × × C L    
P P R F
SP R e s u l t : × × × × × × × × × C L              
R F
ETX                                                  

34
DR-7000D Semi-Automatic Chemistry Analyzer

3 、 Data output format :( QC data )

1 2 3 4 5 6 7 8 9 1 1 1 13 1 1 16 1 1 1 20 2 22 23 2 2 26 27 2 2 30
0 1 2 4 5 7 8 9 1 4 5 8 9
STX CR L                                              
F
SP SP Q u a l i t y S C o n t r o l S R e p o r t C L
P P R F
CR LF                                                
SP D a t e : × × × × - × × - × × S × × : × × C L    
P R F
SP T e s t S N a m e : × × × × × × × × CR L          
P F
SP I D : × × × × × × × × × × C LF                    
R
SP S t a n d a r d S X : SP × × × × × × × × CR L      
P F
SP S t a n d a r d S S D : S × × × × × × × × C L    
P P R F
SP R e s u l t : × × × × × × × × × C L              
R F
ETX                                                  

36
DR-7000D Semi-Automatic Chemistry Analyzer

Appendix D

Regular Chemistry Reagent List

Parameter Cod Product Methodology Wavelength Temperature Standard Factor Unit Measure Lag Reagent
e concentrati ment time
No. on time(S) (S)
Liver 01 ALT Kinetic 340nm 37℃ ── -3376 U/L 60 60 double reagent
Function 02 GGT Kinetic 405nm 37℃ ── 1158 U/L 60 60 double reagent
03 ALP Kinetic 405 nm 37℃ ── 3378 U/L 60 60 double reagent
04 CHE Kinetic 405 nm 37℃ ── 8369 U/L 60 60 double reagent
05 T-BIL End point 546 nm 37℃ 89.6 ── umol/L ── ── double reagent
06 D-BIL End point 546 nm 37℃ 26 ── umol/L ── ── double reagent
07 ALB End pint 630 nm 37℃ 40 ── g/L ── ── simple reagent
08 TP End pint 546 nm 37℃ 80 ── g/L ── ── simple reagent
09 NH3 2-point kinetic 340 nm 37℃ 200 ── umol/L ── ── simple reagent
10 AST Kinetic 340 nm 37℃ ── -3376 U/L 60 60 double reagent
11 LAP Kinetic 405 nm 37℃ ── 2121 U/L 60 60 double reagent
12 ICDH Kinetic 340 nm 37℃ ── 3376 U/L 60 120 double reagent

37
DR-7000D Semi-Automatic Chemistry Analyzer
13 ADA Kinetic 340 nm 37℃ ── 4921 U/L 60 180 double reagent
Kidney 14 BUN 2-point kinetic 340 nm 37℃ 7.14 ── mmol/L 60 30 double reagent
Function 15 Cr 2-point kinetic 505 nm 37℃ 400 ── umol/L 60 30 double reagent
16 UA End point 505 nm room temperature 7.2 ── mmol/L ── ── double reagent
Hydronium 17 Ca End pint 578 nm room temperature 2.5 ── mmol/L ── ── double reagent
18 Ca End pint 630 nm room temperature 2.5 ── mmol/L ── ── simple reagent
19 CI End point 505 nm room temperature 100 ── mmol/L ── ── simple reagent
20 Mg End point 546 nm room temperature 1 ── mmol/L ── ── simple reagent
21 Fe End point 546 nm room temperature 90 ── umol/L ── ── double reagent
22 P End point 340 nm room temperature 1.29 ── mmol/L ── ── simple reagent

Glucose 23 GLU End point 505 nm 37℃ 5.55 ── mmol/L ── ── double reagent
metabolism 24 GLU End point 340 nm 37℃ 5.55 ── mmol/L ── ── double reagent
Cod Product Methodology Wavelength Temperature Standard Factor Unit Measure Lag Reagent
e concentrati ment time
No. on time(S) (S)
25 FTS End point 546 nm 37℃ 1.95 ── mmol/L ── ── simple reagent
Blood 26 CHO End point 505 nm 37℃ 7.0 ── mmol/L ── ── double reagent
lipoid 27 HDL-C End point 546 nm 37℃ 53 ── mmol/L ── ── double reagent
28 LDL-C End point 546 nm 37℃ 77 ── mmol/L ── ── double reagent
29 TG End point 546 nm 37℃ 2.2 mmol/L ── ── double reagent
30 LDH-L Kinetic 340 nm 37℃ ── 9683 U/L 60 60 double reagent
31 CK Kinetic 340 nm 37℃ ── 8095 U/L 60 120 double reagent
32 CK-MB Kinetic 340 nm 37℃ ── 8254 U/L 60 120 double reagent
33 HBDH Kinetic 340 nm 37℃ ── 4920 U/L 60 60 double reagent
Acute 34 AMY Kinetic 405 nm 37℃ ── 5138 U/L 60 60 double reagent
abdomen 35 AMY Kinetic 405 nm 37℃ ── 5138 U/L 60 60 double reagent

38
DR-7000D Semi-Automatic Chemistry Analyzer
Note:

1. The above info just for regular items, for the details, please check the specification on the reagent kits.

2. User also could program the new parameters according to the specification on the reagent kits.

39
DR-7000D Semi-Automatic Chemistry Analyzer

Appendix E

Bar code reader installation and operation

1. Installation
The bar code reader is the optional spare parts for DR-7000 series semi-automatic chemistry
analyzer, there is a serial port on analyzer back board for it installation. The operation is very easy,
operator only need to insert the bar code reader connector plug into the proper serial port.

Notice: The equipment should be powered off before bar code reader installation.

2. Operation
(1) The bar code reader could read Code-39, Code-128, EAN-8, UPC-E and Inter leaved 2 of 5 bar
code automatically and the total digit for the bar code number should not more than 10 letters,
otherwise, it will not workable.
(2) The operator could do the scan operation before each test (ID should be setup ON) and then the
bar code number will be saved into ID number. There is only one bar code number for each
sample, therefore, it is only need to scan one time for the same sample or the equipment will save
the last scan one as the final one.
(3)The bar code reader belongs to far-distance Reader, the operator only need to aim the scan point
at the barcode number and then press the trigger, so that, the real-time bar code info will be
shown on the screen.
(4)The bar code reader reads the barcode by the light line center on the top of it, to avoid mistakes,
the light line should cover the whole barcode and keep proper bevel with the trigger before
reading.

Connection of stylus printer

Instrument could connect EPSON LQ300K+ stylus printer. (Other type stylus printer can be used as
well, but it may cause some change in print speed and character position.)