Manual Gasgard XL
Manual Gasgard XL
In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-INST
To contact MSA International, dial 1-412-967-3354
Manufactured by
The warranties made by Mine Safety Appliances Company with respect to the product are
voided if the product is not used and serviced in accordance with the instructions in this
manual. Please protect yourself and others by following them. We encourage our cus-
tomers to write or call regarding this equipment prior to use or for any additional infor-
mation relative to use or repairs.
                                                                                    i
MSA Permanent Instrument Warranty
1. Warranty- Seller warrants that this product will be free from
   mechanical defect or faulty workmanship for a period of 18 months
   from date of shipment, or one year from installation, whichever
   occurs first, provided it is maintained and used in accordance with
   Seller's instructions and/or recommendations. This warranty does
   not apply to expendable or consumable parts whose normal life
   expectancy is less than one year such as, but not limited to, non-
   rechargeable batteries, filament units, filter, lamps, fuses etc. The
   Seller shall be released from all obligations under this warranty in
   the event repairs or modifications are made by persons other than
   its own or authorized service personnel or if the warranty claim
   results from physical abuse or misuse of the product. No agent,
   employee or representative of the Seller has any authority to bind
   the Seller to any affirmation, representation or warranty concerning
   the goods sold under this contract. Seller makes no warranty
   concerning components or accessories not manufactured by the
   Seller, but will pass on to the Purchaser all warranties of
   manufacturers of such components.
   THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
   EXPRESSED, IMPLIED OR STATUTORY, AND IS STRICTLY
   LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY
   DISCLAIMS ANY WARRANTY OF MERCHANT ABILITY OR OF
   FITNESS FOR A PARTICULAR PURPOSE.
2. Exclusive Remedy- It is expressly agreed that Purchaser's sole
   and exclusive remedy for breach of the above warranty, for any
   tortious conduct of Seller, or for any other cause of action, shall be
   the repair and/or replacement at Seller's option, of any equipment
   or parts thereof, which after examination by Seller is proven to be
   defective. Replacement equipment and/or parts will be provided at
   no cost to Purchaser, F.O.B. Seller's Plant. Failure of Seller to
   successfully repair any non-conforming product shall not cause the
   remedy established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser specifically
   understands and agrees that under no circumstances will seller be
   liable to purchaser for economic, special, incidental or
   consequential damages or losses of any kind whatsoever, including
   but not limited to, loss of anticipated profits and any other loss
   caused by reason of non-operation of the goods. This exclusion is
   applicable to claims for breach of warranty, tortious conduct or any
   other cause of action against seller.
 ii
Table of Contents
  Chapter 1,
  Product Introduction . . . . . . . . . . . . . . . . . . .1-1
   Correct Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
  Chapter 2,
  Description . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
   General Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
   Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
    Wall Mount Enclosure . . . . . . . . . . . . . . . . . . . . . . . . .2-2
    Figure 2-1. Enclosure . . . . . . . . . . . . . . . . . . . . . . . . .2-2
    Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
    Backup Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
    Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
    Display Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
    Sensor Extension Board . . . . . . . . . . . . . . . . . . . . . . .2-4
    Channel Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
    Channel Relay Board . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
   Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
    Graphic Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
    Figure 2-2. Front Panel . . . . . . . . . . . . . . . . . . . . . . . .2-6
    Figure 2-3.Graphic Display . . . . . . . . . . . . . . . . . . . . .2-6
    Control Push Buttons . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
    Channel Status Info LEDs . . . . . . . . . . . . . . . . . . . . . .2-7
    Figure 2-4. Channel Status Info LEDs . . . . . . . . . . . . .2-7
    Unit Status Info LEDs . . . . . . . . . . . . . . . . . . . . . . . . .2-9
    Figure 2-5. Unit Status Info LEDs . . . . . . . . . . . . . . . .2-9
                                                                                          iii
 Chapter 3,
 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
     Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
     Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
         " CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
      Figure 3-1. Mounting Drawing . . . . . . . . . . . . . . . . . .3-2
      Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
      Figure 3-2. Connecting the Main Supply . . . . . . . . . . .3-3
      Sensor Connection . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
         " CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
 Chapter 4,
 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
     Basic Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
      Figure 4-1. Basic Screen . . . . . . . . . . . . . . . . . . . . . . .4-1
     Status Info Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
     Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
      Using the Hotkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
     Viewing Channel Information . . . . . . . . . . . . . . . . . . . . .4-6
     Viewing Events Archive . . . . . . . . . . . . . . . . . . . . . . . . .4-7
          List of Assigned Events . . . . . . . . . . . . . . . . . . . . .4-8
 Chapter 5,
 Configuration . . . . . . . . . . . . . . . . . . . . . . . . .5-1
      (1) System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
      (2) Calibration Menu Direct Access . . . . . . . . . . . . . . .5-1
     System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
      Channel Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
      Relay Board Identification Mark . . . . . . . . . . . . . . . . .5-3
      Setting Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
      Description of Channel Parameters . . . . . . . . . . . . . . .5-4
      Modifying Channel Parameters . . . . . . . . . . . . . . . . . .5-6
     Relay Setting Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
      Description of Relay Parameters . . . . . . . . . . . . . . . . .5-8
      Example of Relay Configuration . . . . . . . . . . . . . . . .5-10
      Modifying Output Relays . . . . . . . . . . . . . . . . . . . . . .5-10
      Fixed Setting for Common Relays . . . . . . . . . . . . . . .5-12
     General Setting Menu . . . . . . . . . . . . . . . . . . . . . . . . .5-12
iv
   Manual Calibration (4 - 20 mA) . . . . . . . . . . . . . . . . .5-14
Chapter 6,
Configuration Software . . . . . . . . . . . . . . . . .6-1
 Installation and Start up . . . . . . . . . . . . . . . . . . . . . . . . .6-1
 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
   Menu Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
   Figure 6-1. Main Applications Screen . . . . . . . . . . . . .6-2
 Multilingual System . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
 Setting Up Channels . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
   Figure 6-2. Setting up Channels . . . . . . . . . . . . . . . . .6-4
 Setting Up Output Relays . . . . . . . . . . . . . . . . . . . . . . . .6-5
   Figure 6-3. Setting up Output Relays . . . . . . . . . . . . .6-5
Chapter 7,
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
        Table 7-1. Accessories and Spare Parts . . . . . . . .7-1
Chapter 8,
Technical Specifications and Certifications .8-1
     Table 8-1. Technical Specifications . . . . . . . . . . . .8-1
 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Appendix A,
Sensor List . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
        Table A-1.GasGard XL-compatible Sensors . . . . .A-1
Appendix B,
Wiring Diagrams . . . . . . . . . . . . . . . . . . . . . . .B-1
   Figure    B-1.Flamegard Flame Detector . . . . . . . . . . . .B-1
   Figure    B-2. SafEye Open Path Gas Detector . . . . . . .B-1
   Figure    B-3. Ultima X (Two-wire) . . . . . . . . . . . . . . . . .B-2
   Figure    B-4. Ultima X (Three-wire) . . . . . . . . . . . . . . . .B-2
   Figure    B-5. Ultima X IR Sensor . . . . . . . . . . . . . . . . . .B-3
                                                                                        v
 Appendix C,
 Individual Relay Connections . . . . . . . . . . .C-1
     Figure C-1. Individual Relay Connections . . . . . . . . . .C-1
 Appendix D,
 Terminal Connectors . . . . . . . . . . . . . . . . . . .D-1
     Figure D-1. Main Board . . . . . . . . . . . . . . . . . . . . . . . .D-1
     Figure D-2. Sensor Extension Board . . . . . . . . . . . . .D-1
     Figure D-3. Channel Relay Board . . . . . . . . . . . . . . . .D-2
vi
Chapter 1,
Product Introduction
Correct Use
The GasGard XL Wall Mount Controller is a compact control unit that:
 •   connects with up to eight active combustible, toxic and/or
     oxygen sensors
 •   is used to monitor indoor industrial locations for the presence of
     combustible, explosive, toxic gases or air/vapor mixtures and
     oxygen content
 •   provides power to the sensors, signal conditioning and display of
     measured gas concentrations, alarm thresholds, outputs for alarm
     devices and self monitoring diagnostics
 •   is suitable for many applications and industrial areas
 •   provides simple installation, easy editing and copying functions
 •   is ideal for smaller, independent gas warning systems.
The user must carefully read and observe:
 •   this operating manual, particularly, the safety, use and operating
     instructions
 •   the national regulations applicable in the user's country, including
     requirements specific to explosive risk approval.
Use or alterations outside these instructions is considered as
non-compliance.
                             " WARNING
      This product is supporting life and health. Incorrect use,
      maintenance or servicing can affect its function and seri-
      ously compromise the user’s life. When turned ON for the
      first time, the controller performs a self-check. Do Not use
      the monitor if it fails the self-check, which is indicated by a
      failure light. Do Not use the monitor if it is damaged, improp-
      er servicing/maintenance has been performed, or if genuine
      MSA spare parts have not been used.
                                                                        1-1
Chapter 2,
Description
General Overview
The general-purpose control system enables:
 •   Combustible gas monitoring
     •   for protecting operational plants and workers by monitoring
         atmospheres for potentially explosive gases/vapors in air
         before they reach the lower explosion limit and by activating
         alarms and initiating risk aversion measures.
 •   Oxygen monitoring
     •   for protecting individuals by monitoring the atmosphere for
         oxygen deficiency or oxygen enrichment.
 •   Toxic gas monitoring
     •   for protecting individuals by continuously monitoring the
         atmosphere for toxic gas concentrations.
NOTE: An alarm activates if set limit values are reached
      (e.g. minimum/ maximum concentration in the atmosphere).
                                                                     2-1
Control Unit
        2-2
Power Supply
 •   The 100 W power supply accepts main AC and DC power.
 •   In case of loss of main power, the power supply automatically
     switches to backup battery.
 •   Power source for detector is 24 VDC.
Backup Battery
Two batteries in series (12 VDC/2.2 Ah lead acid) can be used in the
Control Unit as an option. The batteries supply 20 minutes of backup for
eight catalytic combustible sensors.
Main Board
The Main board microcontroller includes:
 •   watchdog
 •   internal buzzer (85 dB)
 •   standard common outputs
 •   battery charger circuit.
All necessary information is stored on the Main board.
                                                                2-3
 •    For SCADA visualization software, all three interfaces can be used.
 •    For service access, as a safety precaution, only the two ModBus
      RTU communications can be used.
Standard common outputs are incorporated:
 •    two SPDT (Single Pole Double Throw) relays common alarm
      (ALARM 1 and ALARM 2)
 •    one DPDT (Double Pole Double Throw) relay common fault
 •    two addressable common horn relays.
Display Board
The Display board, located on the back side of the lid, includes:
 •    microcontroller
 •    watchdog
 •    push buttons
 •    graphic display (128x64).
Channel Board
The channel boards are set to communicate with 4-20 mA transmitters.
Every board:
 •    includes four LEDs (A1, A2, Status, Power)
 •    attaches to the Channel connectors
      (four on the Main board, four on the Sensor Extension board).
Sensor power and connection is designed as "fail-safe"
(wrong connection/configuration does not damage the remote sensor).
     2-4
Channel Relay Board
Every Channel Relay board:
 •   provides eight output relays
 •   is connected to the Main board or Sensor Extension board (one by
     one) via connectors.
Relays are fully user-configurable from the front panel or connected PC
via configuration software.
Two optional SPDT (Single Pole Double Throw) relays per channel
(eight relays per four channels) are configurable for individual alarms:
 •   Normally open/closed
 •   Normally energized/de-energized
 •   Latching/non latching
 •   Increasing/decreasing
 •   N out of m voting
 •   Grouping
     NOTE: Contacts are resistive load.
                                                                   2-5
Front Panel
The Front Panel enables communication with the Control Unit
and is used to:
 • monitor the status of all connected field sensors
 • determine system settings
 • configure all settings.
Graphic Display
The Control Unit has a large, 128 x 64 pixel graphic display which
shows information for all channels. See Chapter 4 for symbol
explanation.
    2-6
Control Push Buttons
Using the control push buttons, the user can operate the menu prompt
as shown below. For more information, see Chapter 4.
Right button [ ↵ ]
Up Button [ ]
Down button [ ]
Alarm Acknowledgement
Alarm Reset
                                                                2-7
1. The green Power LED:
   Channel is powered and enabled by the system.
2. The yellow Status LED:
   Channel is in a fault condition.
    •   Blinking LED (0.5 Hz):
        Channel is disabled.
3. The red Alarm 1 LED:
   First alarm level is reached. It further indicates the specific status:
    •   4 Hz blinking:
        First alarm level is reached, but not yet acknowledged
    •   Steady:
        First alarm level is acknowledged, but still inside alarm range
    •   0.5 Hz blinking:
        First alarm is configured as latching. Measured value is out of
        alarm level and not yet reset.
4. The red Alarm 2 LED:
   Second alarm level is reached. It further indicates specific status:
    •   4 Hz blinking:
        Second alarm level is reached, but not yet acknowledged
    •   Steady:
        Second alarm is acknowledged, but still inside alarm range
    •   0.5 Hz blinking:
        Second alarm is configured as latching. Measured value is out
        of alarm level and not yet reset.
 2-8
Unit Status Info LEDs
                                                     2-9
Chapter 3,
Installation
NOTE: Reference shipping documents and carton label to check that
      delivered components are correct before installation.
 •   The installation location for the Control Unit must be outside the
     potentially explosive area and free of corrosive gases.
 •   Sensors for use in the potentially explosive area must have the
     appropriate approval and be installed in accordance with all
     relevant local and national regulations.
 •   The gas warning system must be installed by qualified personnel
     after reading the supplied documentation.
 •   All relevant local and national regulations and instructions must be
     observed.
                                                                   3-1
Mechanical Installation
Electrical Installation
                             " CAUTION
     3-2
         If an external earth connection is permitted or demanded by the
         local authorities, it serves merely as additional earthing.
 •   Select an installation location that complies with the environmental
     conditions indicated in the technical data.
 •   When installing the Control Unit, the following conditions must be
     met to comply with the European EMC Directive
     •   A fault-free ground or fault-free potential bonding conductor
         must be provided when connecting devices to the main
         power supply.
     •   Ensure an appropriate supply voltage in accordance with
         EMC directives.
     •   If the devices are supplied from a direct voltage (DC) source,
         the supply cable must be shielded.
     •   All sensor and control cables must be shielded.
          •   Shielded cables must have at least 80% coverage.
     •   Control and sensor cables must be installed physically
         separate from power supply cables.
     •   Shielded cables must be laid in a group. If cable lengths are to
         be extended using terminal boxes, the boxes must be shielded
         and internal connections must be kept as short as possible.
     •   The control system can be simultaneously connected to
         AC voltage and 24 VDC (to enable an automatic power switch
         to 24 VDC voltage if the AC main supply fails).
Power Supply
 •   Standard power supply 110-230 VAC/24 VDC (100 W) is used to
     power internal circuits and attached sensors.
 •   A switch or circuit-breaker shall be included in the building
     installation in close proximity to the equipment and within easy
     reach of the operator. It shall be marked as the disconnecting
     device for the equipment.
 •   Main voltage is connected to terminal L (Line), N (Neutral), and
     ground.
                                                                  3-3
                   Figure 3-2. Connecting the Main Supply
Sensor Connection
 •    The sensor must be connected to the terminals on the Main board
      or Sensor Extension board (See Appendices C and D for details).
                                " CAUTION
           Follow the instructions for components subject to damage
           from static electricity!bb
 •    Incorrect connection of the sensors will not damage the Control
      Unit or the sensor.
 •    The wiring diagrams for different sensors are shown in Appendix B.
      The sensors must be connected to the terminals using shielded
      cables.
 •    For the number of wires and the maximum cable length for each
      sensor type, refer to the specific sensor operating manual.
     3-4
Chapter 4,
Operation
The integrated operation/display unit:
    •   is the control system user interface
    •   displays alarms, warnings, and system parameters.
         •   Connecting the operating unit to a PC enables a user-friendly
             operator interface.
         •   Input fields are set up as selection fields as much as possible,
             with all known inputs displayed. Selection is cursor-controlled
             for easy display unit use.
Basic Screen
The basic screen displays information about active channel status
(active, inhibit, and in service) including information about the system
(system date and time).
                                                                      4-1
 •    System Date and Time:
      Actual preset values for events archive (default is DD/MM)
 •    Channel Number:
      Position of the channel board [counted from left to the right on the
      Main board (1 … 4) and the Sensor Extension board (5 … 8)].
 •    Status Information:
      Information about channel status (alarm, fault, in service, etc.;
      see symbols)
 •    Actual Value:
      Measured value of gas concentration.
 •    Events Archive:
      Last 700 events (alarms, alarm acknowledgement and reset, faults,
      calibrations, etc.) are stored in memory and can be user-restored.
           Alarm 2:
              Displays if alarm level is reached
           Overrange:
             Displays if channel signal is above the range
             (more than 105 % of full scale).
           Underrange:
             Displays if channel signal is below the range
             (less than -5 % of full scale)
           Sensor Fault:
             Displays if a sensor connection is broken or shorted or
             if signal falls below -10% of full scale
           In Service:
              Displays if channel is in service mode (calibration)
     4-2
        Inhibit:
           Displays if channel is inhibited
        STEL Alarm:
          Displays if STEL alarm is reached
        TWA Alarm:
          Displays if TWA alarm is reached
                                                                 4-3
Controls
Located on the front panel are four control push buttons for Control Unit
operation and configuration.
   4-4
Using the Hotkeys
Pressing some of the front panel control push buttons for a longer
period accesses some additional basic screen features.
         UP:
           Pressing button for five seconds changes the
           language from English to one of 10 local languages
           and back
         ESCAPE:
           Pressing button for five seconds starts the GasGard
           XL Selftest Process, testing the:
          •   LCD display
          •   LEDs
          •   internal buzzer.
After the selftest, the firmware version of all assigned boards displays:
          •   C1 …C8 - Channel boards
          •   R1 … R2 - Relay boards
          •   D- Display board
          •   M – Main board.
                                                                   4-5
Viewing Channel Information
NOTE: To change gas type, unit of measure or alarm levels, see the
      Quick Startup Guide (P/N 10095757) or Chapter 6.
   4-6
Viewing Events Archive
          1. Viewing the Basic Screen, press [ENTER].
                                                                4-7
List of Assigned Events
00   Device Turned On             27 Master Calibration Saved
01   Communication OFF            28 New Calibration Saved
02   Communication ON             29 Coil Relay 1 FAULT
03   Board Reset                  30 Coil Relay 2 FAULT
04   Alarm1 ON                    31 Coil Relay 3 FAULT
05   Alarm1 OFF                   32 Coil Relay 4 FAULT
06   Alarm1 Acknowledgement       33 Coil Relay 5 FAULT
07   Alarm1 Reset                 34 Coil Relay 6 FAULT
08   Alarm2 ON                    35 Coil Relay 7 FAULT
09   Alarm2 OFF                   36 Coil Relay 8 FAULT
10   Alarm2 Acknowledgement       37 Main Power Supply OFF
11   Alarm2 Reset                 38 Main Power Supply N
12   UnderRange                   39 Low Battery
13   OverRange                    40 Battery FAULT
14   STEL alarm ON                41 Common Coil Relay Horn2 FAULT
15   STEL alarm OFF               42 Common Coil Relay Horn1 FAULT
16   STEL alarm Acknowledgement   43 Common Coil Relay Alarm2 FAULT
17   TWA alarm ON                 44 Common Coil Relay Alarm1 FAULT
18   TWA alarm OFF                45 Common Coil Relay Fault FAULT
19   TWA alarm Acknowledgement    46 RAM Check FAULT
20   Sensor FAULT                 47 ROM Check FAULT
21   System FAULT                 48 System Password entered
22   Start of Calibration         49 Calibration Password entered
23   ZG accepted                  50 Read FAULT
24   SG accepted                  51 Write FAULT
25   FAULT of ZG result
26   FAULT of SG result
     4-8
Chapter 5,
Configuration
All GasGard XL Control Unit parameters can be configured by using:
 •   front panel control push buttons
 •   configuration software tool (see Chapter 6).
Two menus allow the user to change unit parameters:
NOTE: There are different passwords for the System menu and
      Calibration menu. Passwords can be set in the General Setting
      menu.
                                                                  5-1
System Menu
The System menu allows full unit configuration:
 •     Modifying channel parameters
 •     Setting output relays (Relays Setting menu)
 •     Calibration (Calibration menu)
 •     General settings (General setting menu).
     5-2
1. Channel Number
If selected, the Channel Number allows user to set channel parameters.
3. Setting Icons
Marking icons enable the user to enter an additional submenu:
     Relays setting menu (individual relays modification menu;
     see Chapter 5, "Modifying Channel Parameters").
ENTER:
UP:
DOWN
ESCAPE:
  5-4
    NOTE: Use [DOWN] to toggle between the two screens.
    •    The second screen displays:
                                                                5-5
Modifying Channel Parameters
 1. Enter into System menu (see Chapter 5, "System Menu").
 2. Use control push buttons to select a channel number to modify and
    press [ENTER] to confirm.
    •   First of configuration screens appear (left figure).
    NOTE: Use [DOWN] to toggle between the two screens:
Up Button [ ]
Down button [ ]
  5-6
 5. Use the control push buttons to change parameter value.
 6. Press [ENTER] to accept the changes.
     NOTE: Pressing [ESCAPE] toggles to the upper menu.
 •   If FlameGard is selected from the sensor list, the Setting menu
     changes to the following single screen:
                                                                 5-7
Description of Relay Parameters
 1. Enter into the System menu.
 2. In the System menu, select the item for “Relay Setting Menu” and
    press [ENTER] to confirm (see Chapter 5, "System Menu").
Up Button [ ]
Down button [ ]
  5-8
(3) Relays Voting (V):
              Allows user to set the voting of the selected relays
(4) Relay Number (R01-R16):
              1-8      (Channel Relay board No. 1)
              9 - 16 (Channel Relay board No. 2)
(5) ATEX Status:
             Defines the selected relay is related to the channel
             being configured according to ATEX regulation. ATEX
             channels are displayed in bold numbers on the screen.
             Some relay parameters are limited [e.g., user can only
             set the normally energized (S: NE) selection]
(6) Channel Number:
              Relates the relay to the selected channel
              (bold number indicates channel is configured according
              to ATEX regulation)
(7) Configuration Grid:
              Setting the following items to the grid defines the
              status on the selected channel that activates the
              selected relay:
       - Alarm 1 is set
- Alarm 2 is set
- Fault is set
- Selected
                                                                5-9
Example of Relay Configuration
     5-10
 3. Use control push buttons to select and change required parameter.
Up Button [ ]
Down button [ ]
Up Button [ ]
Down button [ ]
     5-12
 1 - Modbus     RTU: Defines the RS 485 / USB communication parameters
     ADR –       unit address
     Speed –     communication speed
     Parity –    setting the parity (Even/Odd)
 2 - Modbus     TCP / IP: Defines the Ethernet communications parameters
     ADR –       unit IP address
     MASK -      subnetwork mask
     GATE -      gateway
                                                                           5-13
            Right (ENTER) button [ ↵ ]
Up Button [ ]
Down button [ ]
Calibration
Calibration (4 - 20 mA)
Channel boards are configured to accept 4 - 20 mA input signals from
two-wire or three-wire remote transmitters. The controller does not
require any calibration.
     5-14
Chapter 6,
Configuration Software
                                                                    6-1
Application
Menu Bar
The menu bar has four pull down menus for application functions:
 •     File
 •     View
 •     Settings
 •     Help.
     6-2
File
From the File menu, the user can:
   • send or receive a configuration to the device unit
   • load or save a configuration to or from a file
   • print
   • exit the application
View
From the View menu, the user can switch between four basic screens and show or hide
the toolbar and status bar.
NOTE: For easy access click on the toolbar icon to show a particular screen.
This application allows communication with devices via serial or USB port. The user must
select one port to use for communication.
Multilingual System
The Configuration software is multilingual; for individual languages, use
the "lang.properties" configuration file.
                                                                                6-3
Setting Up Channels
  6-4
Setting Up Output Relays
From the View menu, user can select the [OUTPUTS] option.
For each channel and each output relay, the user can select:
 •   Alarm 1
 •   Alarm 2
 •   STEL
 •   TWA or
 •   Fault.
For each selected channel, the drop-down menu enables voting
selection. Voting can be set from 1:1 up to 42:42.
                                                               6-5
Chapter 7,
Maintenance
  •   The Control Unit requires no special maintenance or cleaning,
      apart from verifying that it responds appropriately when performing
      sensor calibration checks.
  •   For SIL 2 (Safety Integrity Level) sensor applications, the
      calibration intervals must be reduced appropriately.
  •   Check the calibration of sensor(s) connected to the Control Unit
      according to your company’s safety manager.
                                                                      7-1
Chapter 8,
Technical Specifications and
Certifications
 8-2
Appendix A,
Sensor List
                                          A-1
Appendix B,
Wiring Diagrams
                                                  B-1
      Figure B-3. Ultima® X (Two-wire)
B-2
Figure B-5. Ultima® X IR Sensor
                                  B-3
Appendix C,
Individual Relay Connections
                                                  C-1
Appendix D,
Terminal Connectors
                                             D-1
      Figure D-3. Channel Relay Board
D-2