Calibration Procedure

1. The document provides calibration instructions for a MA10-ME device. It involves entering calibration mode by pressing buttons, inputting a password, setting decimal points and minimum divisions, inputting maximum capacity, performing zero calibration with no load, and performing span calibration with a known standard load. 2. Steps include setting parameters like decimal place, minimum division, and maximum capacity, reading millivolt values during zero and span calibration, and inputting known weights to complete calibration. 3. Final steps confirm calibration is complete by inputting the password and displaying "CAL End".
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3K views2 pages

Calibration Procedure

1. The document provides calibration instructions for a MA10-ME device. It involves entering calibration mode by pressing buttons, inputting a password, setting decimal points and minimum divisions, inputting maximum capacity, performing zero calibration with no load, and performing span calibration with a known standard load. 2. Steps include setting parameters like decimal place, minimum division, and maximum capacity, reading millivolt values during zero and span calibration, and inputting known weights to complete calibration. 3. Final steps confirm calibration is complete by inputting the password and displaying "CAL End".
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

CALIBRATION INSTRUCTION OF MA10…ME

User can press to enter next or press to save setting to only change one
parameter, and then press to exit.
1. Under this status, press (twice), indicator will display CAL, then press to enter .

Normal Status

2. Display will show PASS, input password 000000 six times zero using then press
the display will show CALON for one second, then go to next step.

PASSWORD Input

3. Display will show Point, Press to select a desired value for decimal point among 0, 0.0,
0.00, 0.000 and 0.0000, and then press to set mini division and enter next step.

DECIMAL Point

4. Display will show Id, Press to select a desired value for min. division among 1, 2, 5, 10,
20 and 50 and then to save it and enter next step. If there’s no need to change the value,
then press .

Minimum Division
5. Display will show CP for few second, then Input max. Capacity (≤min. division×100000) by
using & key press to save it and enter next step.

Maximum Capacity

6. Display will show t, press to enter zero calibration. Display value near the output value
in millivolt between SIG+/SIG- of load cell.

Millivolt Value Display

7. Display will show o, when STAB lamp is on, press to finish zero calibration. If there’s no
need to calibrate zero, press . Make sure that t & o values should be nearly same.

Zero Calibration

8. Display will show Add Ld for few seconds, then it will show C 0.0000, At this state we need to
hook the standard known weight for calibration which should be 80% of maximum capacity,
display will show some millivolt values and then press to enter weight input.

9. Display will show 000000, At this status we have to enter standard weight value of loaded
weight by using & key. Press to finish calibration. Display will show S 10CAL,
Press then enter password six times 000000 to confirm calibration two times.
Display will show CAL End.