AXIOM Aristos Image System
VB21 and MX, VX, VX Plus, TX, FX, FX Plus System SW;
                       © Siemens, 2011
                                                       © Siemens, 2011
                                                       All documents may only be used for rendering
                                                       services on Siemens Healthcare Products. Any
                                                       document in electronic form may be printed
                                                       once. Copy and distribution of electronic docu-
                                                       ments and hardcopies is prohibited. Offenders
                                                       will be liable for damages. All other rights are re-
                                                       served.
                                                                                                         2011
2                                                                     Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
                                    Copyright
                                    “© Siemens, 2011“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
                                    sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
                                    Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
                                    Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
                                    ucts GmbH - Germany.
                                    Document Version
                                    Siemens reserves the right to change its products and services at any time.
                                    In addition, manuals are subject to change without notice. The hardcopy documents corre-
                                    spond to the version at the time of system delivery and/or printout. Versions to hardcopy
                                    documentation are not automatically distributed. Please contact your local Siemens office
                                    to order current version or refer to our website http://www.healthcare.siemens.com.
                                    Disclaimer
                                    Siemens provides this documentation “as is“ without the assumption of any liability under
                                    any theory of law.
                                    The installation and service of equipment described herein requires superior understand-
                                    ing of our equipment and may only be performed by qualified personnel who are specially
                                    trained for such installation and/or service.
1 _______ Requirements___________________________________________________ 6
             Manufacturer’s Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               How to use these Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
               Screen Shots & Graphics in this Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
             Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    7
               Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        7
               Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        7
               Protection of Electrostatically-sensitive Components . . . . . . . . . . . . . . . . . . . . . . . . .                            7
               Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        7
             General remarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
             Special Tools / Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
                For Collimator SW Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
                Collimator Software Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
             Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    12
                Log files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   12
                Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     15
                Uninstalling System Management (MNP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                           20
             Booting the COBR Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
             Installation with the Clone CD or DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
             Setting or Checking the IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
             Checking the Time Zone and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
             Resetting Regional and Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
             Installing the Aristos Service SW on the Radis PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
                 Generating the new XCU Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
             Installation of the VB21 Patch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
             Installation of DICOM Attrib. DAC Configurator (only if used) . . . . . . . . . . . . . . . . . . . . 33
             RADIS, Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
             Import Licenses for additional Functions (only if new functions are to be added) . . . . . 35
             RADIS, Final Setup, Setting Firewall, Security, and Radis Log . . . . . . . . . . . . . . . . . . . 36
             If needed: Deactivating the Serial Mouse at COM 1 (only for Aristos MX / VX /TX) . . . 37
             Updating the OT 16-bit Collimator N FD II (P/N 10092604 only) to VC00N . . . . . . . . . . 41
             Downloading the new XCU Floppy (only for Aristos MX / VX /TX) . . . . . . . . . . . . . . . . . 43
             Updating the Aristos FX System SW (only for FX/FX Plus) . . . . . . . . . . . . . . . . . . . . . .                                 44
               Downloading the FX VA04 SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                      44
               Downloading the new XCU Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                        46
               Checking and Configuring the new SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                           47
             Checking and Activating System Management (only if remote is configured) . . . . . . . . 49
               Activating System Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
             Restoring the Customer OGPs (for all systems) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
                Checking for RIS Match in the OGP Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
             Ortho (necessary only when installing for the first time). . . . . . . . . . . . . . . . . . . . . . . . . 54
             General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
             Creation of a repair plan (usually by USC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
                What to do before Reinstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
                On-site Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
             Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .       69
                Log files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     69
                Perform a virus scan if a virus is suspected (see the repair plan, if available). . . . . .                                         71
                Backup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .       71
             Booting the COBR Disk. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
             Reinstallation with a Recovery Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
             Checking the latest VB21 Patch according to the Repair Plan . . . . . . . . . . . . . . . . . . . . 78
             Installing the Aristos Service SW on the Radis PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
             Restore of the syngo and Radis Configuration, if the restored Image is older than the Back-
             ups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
             Check for the latest Licenses (according to the Repair Plan) . . . . . . . . . . . . . . . . . . . . . 81
             Checking and Activating System Management (only if remote is configured) . . . . . . . . 82
               Activating System Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
             Restoring the Customer OGPs (only if the OGPs are newer than in the Recovery Image)
              . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
                  Checking for RIS Match in the OGP Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
             Generating Backups if changes have been made. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
             Creating a current Recovery Image, if changes have been made . . . . . . . . . . . . . . . . . 88
                Labeling the Recovery Image(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
             Final check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
             Informing the Customer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
                 Reporting Errors by the Customer or CSE (for information only) . . . . . . . . . . . . . . . 92
             General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
             Creation of a repair plan (usually by USC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                       95
                What to do prior to Reinstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                    95
                Collimator Software Download (only if coming from VB10) . . . . . . . . . . . . . . . . . . . .                                     95
                On-site Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .              95
             Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
                Retrieve Log Files from a VB10x PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
                Screen shots and graphics used in this document should be viewed as examples only.
                They cannot describe all possible input combinations; they have an informative character
                and are used only for better orientation.
                The actual parameters that must be configured can be found in the text and the tables and
                depend on the requirements and configuration (items included in the shipment) of the indi-
                vidual system.
Requirements 1.1
Required Tools 0
Safety Information 0
             The “Safety Information“ contained in the system manual and in the particular unit instruc-
             tions must be observed.
             Integrated circuits and circuit boards on which MOS components are installed call for par-
             ticularly careful handling because of their highly sensitive electrical structures and because
             of their extremely high input resistance. The ESD regulations that apply for electrostati-
             cally-sensitive components must be observed.
Required Parts 0
             The table provides an overview of the required CDs/DVDs for different software installation
             situations. These are:
             • New installation (from scratch), e.g. upgrade (retrofit) from VB10... to VB21C and
                  higher
             • Re-installation, e.g. because the existing software crashed
             • Latest patch beginning with VB21C
                                    VA20C COBR                              x       x
                                    Aristos RADIS VB21 Patch CD             x       x        x
                                    Aristos Recovery CDs (blank)            x
                                    Aristos I.S. VB21A 460 Clone CD         x
                                    from initial installation: recorded             x
                                    recovery CDs
                                    Stand system SW, and backups            x       x
Tab. 2 PC M410
                                    VA20C COBR                              x       x
                                    Aristos RADIS VB21 Patch CD             x       x        x
                                    Aristos Recovery CDs (blank)            x
                                    from initial installation: recorded             x
                                    recovery CDs
                                    Aristos I.S. VB21A M410 Clone CD x
                                    Stand system SW, and backups            x       x
Tab. 3 PC M430
                                    VA20C COBR                              x       x
                                    Aristos RADIS VB21 Patch CD             x       x        x
                                    Aristos Recovery DVD (blank)            x
                                    from initial installation: recorded             x
                                    recovery DVDs
                                    Aristos I.S. M430 VB21A Clone           x
                                    DVD
                                    Stand system SW, and backups            x       x
                                 VA20C COBR                              x         x
                                 Aristos RADIS VB21 Patch CD             x         x          x
                                 Aristos Recovery DVDs (blank)           x
                                 from initial installation: recorded               x
                                 recovery DVDs
                                 Aristos I.S. M440 VB21A Clone           x
                                 DVD
                                 Stand system SW, and backups            x         x
Tab. 5 PC M460
                                                            This procedure describes the steps necessary to install the Radis VB21 SW on the AXIOM
                                                            Aristos Image System for the 1st time, e.g. only if VB10 or VB20 was on the system
                                                            before.
                                                            • The following is an outline of some of the work steps that will be covered in this chapter:
                                                            Tab. 6   Work Steps
             Check the Med CS Intranet/Extranet for the latest version of this document (For Service ->
             Product Information) or contact your local Siemens office if you do not have access to the
             Siemens Intranet/Extranet.
             To avoid extra work, check the XCS SSW to determine whether the collimator SW is lower
             than VC00N. If so, follow the following work steps. If not, skip this paragraph.
             • Go to the download location on the CS Home page: For Service > Product Informa-
                  tion > AX Systems > Radiography > Direct Technique > Software.
             • The required download file is the 10092710_OT_II_VC00N.exe and is located under
                  the description Collimator G052G.
             • Double-click on it and unzip it to a folder of your choice on the C or D drive of your ser-
                  vice PC.
             • You will find two files in the folder: CollimatorSWLoader.exe and
                  SW_D21_TBN_OT_II_VC00N.H86. Keep them there for later use.
Preparations 2.2
Log files 0
             1. Unlock the access to Windows® with Tab - Del - Transfer image (“+”).
             2. Press the Windows® key, and then Run; enter radissysinfo > OK.
3. Press Start.
5. The log file will be created and saved to the selected directory.
               8. Copy the compressed log file (2/Fig. 5 / p. 15) to a USB stick or similar medium, and
                  keep it ready in case it is requested by RSC/HSC/QIB/DCU.
                       ¹ The log files look like this: “Log<date>_<time>.zip“
               9. If a customer complaint has to be filed, send the file along with the complaint to Health-
                  care AX Complaint.
Backup 0
               The following configurations have to be saved manually because they are related to Win-
               dows® and therefore are not backed up with the syngo backup.
               • Make a note of the service key:
                    - Options > Service > Local Service...
Service Key
             • Make a note of the host name (= computer name) and the IP addresses:
                 - Options > Service > Local Service... > Configuration > TCP/IP LAN
                 - Home > close the window.
                 IP address
                 Subnet mask
                 Default gateway
                 Customer Configuration:
                 - > Options > Configuration.. > Regional and Language...
                 - “Regional Options”: Make a note of the date and time format.
                 - “ Languages > Details...”: Make a note of the language and keyboard.
                 Date Format
                 Time Format
                 Language
                 Keyboard (input locales)
Network Speed
                 Select the Advanced tab card, read the value for Speed & Duplex or Link
                 Speed & Duplex and make a note of it.
             The following work steps describe how to create backups of “SW Settings 02” on CD, “SW
             Settings, Calibration” on CD, “Customer OGPs” on floppy (or USB stick) (the other pack-
             ages do not have to be saved because there are newer versions with the new VB21 SW).
             • To do this, go to:
                 > Options > Service > Local Service > Backup & Restore
                  Server name
                  TCP/IP address
             • Options > Service > local Service > Configuration > System Mgmt.
                 Deselect all check boxes in the Agent Control window and start with Go in the Action
                 Bar.
                 Uninstalling System Management (MNP) in VB10x:
             • The following screen displays the restore progress (Fig. 16 / p. 25). This will take a
                  while ...
                  - If the medium set consists of more than one CD/DVD, you will be prompted to insert
                    the next medium.
                    ¹ Follow the instructions on the screen.
                  - When the restore has ended, a success message is shown in a new screen
                    (Fig. 17 / p. 25).
                  - Close the success window with <OK>.
                  - Remove the Ghost image from the drive and close the tray.
                    ¹ The window will close and the PC starts to reboot.
             • After restart, enter only the host name (= computer name) that was noted before.
                  Do NOT change the password.
             • Go to Start > Settings > Control Panel > Switch to Classic View > Network Con-
                  nections.
             • Highlight “Local Area Connection” and right mouse-click on “Properties”.
             • Highlight “Internet Protocol (TCP/IP)” and click on “Properties”.
             • Check or set “Use the following IP address”. (There must be a fixed IP address and
                  NOT DHCP.) Insert the correct IP address (noted down before).
             • Confirm by clicking on “OK”.
             • Set the network speed to the value noted down before:
                  Without selecting anything, click on Configure.
                 Select the Advanced tab card and set the value for Speed & Duplex or Link Speed
                 & Duplex to the value previously noted.
             •    Go to Start > Settings > Control Panel > Date and Time.
             •    Go to the “Time Zone” tab card and set the correct time zone.
             •    Place a checkmark in “Automatically adjust clock for daylight saving changes”.
             •    Confirm by clicking on “Apply”.
             •    Go to the “Date & Time” tab card and set the correct date and time.
             •    Confirm by clicking on “OK”.
             • Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
                  MX/VX, TX or FX/FX Plus into the CD drive.
             • Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
                  “start.exe” file. The Installer window opens.
             • Click on the link under "Service PC (+ Download Software) ...".
             • If queried, choose Full Install.
             • Choose Continue > OK.
             • For Aristos FX/FX Plus: Insert a blank, formatted floppy into the floppy drive of the
                  RADIS PC and click on "VA04B" (or higher) under "XCU floppy Software of Aristos FX".
                  A copy routine is started and the floppy is created. When doing this, observe the sys-
                  tem messages and, where applicable, acknowledge them.
             • For Aristos MX/VX: Insert a blank, formatted floppy into the floppy drive of the RADIS
                  PC and click on "VC01B" (or higher) under "XCU floppy Software of Aristos MX".
                  A copy routine is started and the floppy is created. When doing this, observe the sys-
                  tem messages and, where applicable, acknowledge them.
             • For Aristos VX Plus: Insert a blank, formatted floppy into the floppy drive of the RADIS
                  PC and click on "VA00C" (or higher) under "XCU floppy Software of Aristos VX Plus".
                  A copy routine is started and the floppy is created. When doing this, observe the sys-
                  tem messages and, where applicable, acknowledge them.
             • For Aristos TX: Insert a blank, formatted floppy into the floppy drive of the RADIS PC
                  and click on "VA00M" (or higher) under "XCU floppy Software of Aristos TX".
                  A copy routine is started and the floppy is created. When doing this, observe the sys-
                  tem messages and, where applicable, acknowledge them.
             • Label the created floppy appropriately. Keep it for later use.
             • Close the Service SW windows and remove the CD and floppy.
             To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
             •   Insert the Aristos RADIS VB21C Patch or higher CD into the CD drive.
             •   Navigate to the CD/DVD drive with the Windows® Explorer.
             •   Double-click on ParamUpdate.exe.
             •   The patch installation will begin; installation may take 15 ...20 sec.
             •   Click on OK.
             • If the system is using a DAC, install the DAC program from the "DICOM Attribute Con-
                  fig. VA20A", P/N 7264435 or higher.
             • Follow the "DICOM Attrib. Configurator" Instructions, AX00-000.816.01.02… or higher.
             • To do this:
                  -   Go to Options > Service > Local Service... > Configuration
                  -   > Next
                  -   > Licensing
                  -   Insert the data medium with the new License.dat (it is possible that the new license
                      file still needs to be renamed to “license.dat”).
                  -   Click on “Browse..” and go to the new “license.dat”.
                  -   Click on “Open” > “Import License file” and confirm with “OK”.
                  -   Check in the Licenses Window (above) if the added license is shown and has in X in
                      the “Avail” column.
                  -   Select “Home” and close the window.
                  -   Reboot the Radis PC: Options > End Session... > Restart System... > Yes
RADIS, Final Setup, Setting Firewall, Security, and Radis Log 2.13
             • Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Win-
                 dows® Explorer.
             • Navigate to C:\post_clone_tools\Postclone.bat.
             • Double-click on C:\post_clone_tools\Postclone.bat.
                 > OK.
             • Connect all cables.
             In some cases, the mouse cursor sticks in the top right corner when the RADIS PC is con-
             nected to the XCU; this is because the RADIS PC thinks that an RS232 mouse is con-
             nected to its RS232 interface and interprets the signals from the XCU as a "mouse
             command". To prevent this, the RS232 mouse must be disabled on the RADS PC.
             • Select Start > Run > "devmgmt.msc" using the "Windows® key", the arrow keys and
                  the Enter key.
             Fig. 20:
             • After starting the Device Manager, select the "Mice and other pointing devices"
                  hardware components and select the context menu with the right mouse button.
             • Connect the serial XCU service cable to the Radis PC.
             • Select "Scan for hardware changes".
                  - If the serial cable is connected, the serial mouse driver will be found and the mouse
                    cursor will begin to move by itself.
                  - If the “mice and other pointing devices” was not found, repeat “scan for hard-
                    ware changes” several times.
             Fig. 21:
             • Press Alt + Tab to activate the Device Manager window.
             • Use the up/down arrow keys to navigate to "Mice and other pointing devices" (use
                 left/right arrow keys to open the sub-menus).
             Fig. 22:
             •   Select "Microsoft Serial Ball Point".
             •   Press Enter.
             •   Press Tab until "Device Usage" is selected.
             •   Use the up/down arrow keys to select "Do not use this device (disable)".
             Fig. 23:
             • The mouse for this hardware profile will now be deactivated for COM 1 and the Device
                  Manager will display a red cross in front of the deactivated component.
Fig. 24:
             • Loosen the two No. 2 Allen screws (1/Fig. 25 / p. 41). Swing out the front panel of the
                  collimator.
             If there is no FX, skip this chapter and go to (Checking and Activating System Management
             (only if remote is configured) / p. 49).
                     NOTICE           Before starting a download, disable the screen saver and any
                                      power-saving features which might shut down the PC unex-
                                      pectedly! This update takes approx. 30...45 min. and should
                                      be done parallel to the Radis update using the Service PC.
                                      Screen shots do not always show the correct software ver-
                                      sions.
                                          ¹ Follow chapter headlines, notes and hints closely!
             • Connect the Service PC to the gateway master (use the short adapter cable) or use the
                 Radis PC.
                 If using the Service PC: Make sure that the latest FX Service SW is installed.
             • If using the Radis PC, go to C:\SSW\Aristos\Aristos.exe to start the Aristos FX SSW:
                 - Log in with your name. The password is "_Siemens" (case-sensitive!).
             • Perform a backup with SSW Aristos FX > Backup/Restore > Backup ...by pressing
                 the backup button.
             • In the SSW, select Aristos FX > Backup/Restore > Download (1/Fig. 28 / p. 45).
             • Check whether the clutch power settings are correct according to Speed Info 21929:
                 Set the clutch power to 0%....
                 To do so, enter the Aristos FX SSW in developer mode (a developer password for 2010
                 can be found in the Knowledge Base).
                 Click on SSW Aristos FX > Development > GW Config > Mechanical offset.
                 Click on Read. Change all values to 0 in the Clutch Power column.
                 Click on Save.
                 Switch the entire system off and on and check the function.
             • Wait until the SW has downloaded. After the program has downloaded, an audible sig-
                  nal is sounded (TOOT-TA-TOOT) (three beeps).
             • Remove the floppy.
             • Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.
             • Switch the system off completely (hold the switch at the control cabinet for more than 3
                  sec.) and switch on again.
             • Click on the buttons Update Cmp Ver and Update IF Ver. Check the software version
                  of all components (Fig. 30 / p. 47) against the provided the Compatibility List (the
                  screen shot is just an example!).
             • To test the Connectivity, click on the Test button. Since this is a very time-critical action,
                 it is possible that a connection will not be immediately established. Because of this, sev-
                 eral attempts should be made.
                     ¹ The server must send back a message, e.g.:
                       NOTE            If outside NAT is configured, the test function will not work
                                       and the message “...not alive” will be displayed.
                                       The new server name and IP address will then have to be reg-
                                       istered with the local administrator and the SRS Help Desk.
                                       Despite this, continue to the next step of the installation.
                      NOTE           If the text under Status is underlined, a link to the error mes-
                                     sage is overlaid. See (1/Fig. 36 / p. 51).
             • Activate a test patient in RIS with several studies and check whether the studies match
                  after opening the patient on the image system.
Ortho (necessary only when installing for the first time) 2.20
             Enter the correct distance between the Ortho support and the detector:
             • Options>Service>Local Service>Radis Tools>Image Proc. Params.
             • Enter the correct value in “Ortho Support Detector Distance” and “Save”.
                      NOTE           A table with FX Ortho OGPs can be found in the last chapter
                                     of these instructions, “Useful Information and Error Correc-
                                     tion”.
             • Configure a network printer (Fig. 40 / p. 55) only if you want to print to a print server
                  (Fig. 59 / p. 60) and (Fig. 60 / p. 60).
• Select Create a new port and Standard TCP/IP Port (Fig. 41 / p. 55).
Fig. 44:
Fig. 43:
• Select the printer manufacturer and type or have the printer’s floppy to load the driver.
Fig. 52:
Fig. 51:
             • Click on Finish.
             • For printer setup, click on Properties (properties depend on the type of printer; images
                 and configuration may differ from this description)(Fig. 52 / p. 58).
• Click on Next.
             • In the menu that comes up, enter the printer’s name and path (Fig. 60 / p. 60) (the
                 screen shot is in German, please apply accordingly).
             • Continue with (Fig. 41 / p. 55).
             • The following work steps describe how to create backups. This is necessary only if
                  some changes were made since the installation. To save time, save packages only
                  where changes have been made.
             • To do this, go to:
                  > Options > Service > Local Service > Backup & Restore > Backup
                  - Insert the backup CD / diskette / USB stick
                  - Select Drives: [-R-]CD-R; select the following packages one after the other.
                  - In each case, press GO and wait for the Ready messages.
                  - SW Settings02
                  - SW Settings Calibration
                  - IQAP_Database
                  - Customer_OGP_DBs
                  - Clinical Assurance
                  - Insert a blank diskette, labeled "Customer OGP Diskette (Date)". Go back to the
                    Local Service main menu. Select Radis Tools > Organ Programs > Backup. Wait
                    for the "Successfully..." message.
                  - Insert a floppy (blank or the site structure backup floppy) into the floppy drive. Go
                    back to Radis Tools. Click on Service Software. Click on OK in the first message.
                    Observe the second message. Log in to the XCS SSW with User name and OK. Pro-
                    duce a backup of the site structure with Data > Backup to disk
                  - Go to the Aristos FX / FX Plus SSW: SSW Aristos FX > Backup/Restore > Backup
                    and wait for the completion message.
             • Generate a recovery CD / DVD for the C and D drives by following the work steps
                  below. Destroy the clone medium. Do not leave it at the customer site.
                  l)   A new window comes up (Fig. 62 / p. 63) displaying the backup process of the sys-
                       tem partition. This will take a while ...
                  m) At the end of the backup routine, the COBR Disk Backup window appears
                     (Fig. 63 / p. 63) indicating backup success.
                  n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
                     paragraph.
                  o) Confirm the COBR Disk Backup window with <OK>(Fig. 63 / p. 63).
                  p) Go back to step c, select (D:) Med_Data (3/Fig. 61 / p. 62) and repeat the steps.
                     Then continue with step q.
                  q) In the COBR Disk Backup mask, press <Quit> (Fig. 61 / p. 62).
                       ¹ The window will close and a command window is displayed (Fig. 64 / p. 64).
             Perform a function check of the system in accordance with the modifications that have been
             made.
             • The customer or the CSE can easily report problems with the system online.
                 - Under Help, select Report Error.
                 -
                                                       This procedure describes the steps necessary to reinstall the Radis VB21 SW on the
                                                       AXIOM Aristos Image System when a VB21 was installed before, e.g. after a system
                                                       crash.
                                                       • Here is a guideline of what will be covered (among other things) in this chapter:
                                                       Tab. 7   Work Steps
             Creating a repair plan is the best way to ensure that all necessary steps are performed in
             an efficient way and that multiple visits after setup of the PC are avoided.
             • Identify the SW version before the crash (possibly consider open patches or new ver-
                  sions - > to save time) .
                  List all performed updates and remote updates available in the update handling tool /
                  country SAP (notifications)] and the sequence of updates that need to be installed.
             • Check whether a virus protection contract exists -> if yes, download the latest released
                  scan engine + latest virus pattern from the Intranet.
             • Options & Licenses:
                  Check whether there is a distribution list (ROD Remote Options Distribution) [current
                  on-site license file + ROD] using the remote connection.
             • Download all needed remote updates, incl. UI (by CSE (on laptop HD) or by USC (pre-
                  pare DVD etc. if applicable))
             • If a virus is suspected, check the system using the bootable virus scanner.
             • Check the external removable devices (CD, DVD, USB stick, USB disk) if a virus is sus-
                  pected.
             •    Unplug the network cable (virus protection).
             •    Install the latest Ghost image.
             •    Install all required updates and remote updates.
             •    If a virus protection contract exists, install the latest virus pattern.
             •    Install the remote option (ROD), if required.
             •    Plug in the network cable only after all remote updates and virus patterns (VP) have
                  been installed.
Preparations 3.2
                      NOTE            Please consider the steps created in the repair plan. Prior to
                                      a new installation, all patients must be archived from the
                                      local database, because the database is over-written during
                                      the new installation.
Log files 0
              4. Copy the compressed log file (2/Fig. 69 / p. 71) to a USB stick or similar medium, and
                 keep it ready in case it is requested by RSC/HSC/QIB/DCU.
                         ¹ The log files look like this: “Log<date>_<time>.zip“
              5. If a customer complaint has to be filed, send the file along with the complaint to Health-
                 care AX Complaint.
Perform a virus scan if a virus is suspected (see the repair plan, if available). 0
Backup 0
              The following configurations have to be saved manually because these are not part of the
              syngo backup.
Service Key
             The following work steps describe how to create backups of all syngo settings on CD, “Cus-
             tomer OGPs” on floppy (or USB stick).
             • To do this, go to:
                 > Options > Service > Local Service > Backup & Restore > Command > Backup
                 -   Insert the backup CD or USB stick.
                 -   Under Command > Backup > Drives: [-R-]CD-R, select the following packages :
                 -   SW Settings02
                 -   SW Settings Calibration
                 -   IQAP_Database
                 -   Customer_OGP_DBs
                 -   Clinical Assurance
                 -   Press GO for each one and wait for the Ready messages.
                 -   Go back to the Local Service > Configuration menu. Click on Next until you come
                     to the actual Configuration menu.
             • If System Management is configured, manually save the settings:
                 Go back to the Local Service > Configuration menu.
                 Make a note of the settings for System Mgmt.:
                  Server name
                  TCP/IP address
                  Local Host NAT
                  Network adapter
             • The following screen displays the restore progress (Fig. 75 / p. 77). This will take a
                  while ...
                  - If the medium set consists of more than one CD/DVD, you will be prompted to insert
                    the next medium.
                    ¹ Follow the instructions on the screen.
                  - When the restore has ended, a success message is shown in a new screen
                    (Fig. 76 / p. 77).
                  - Close the success window with <OK>.
                  - Remove the FD recovery image from the drive and close the tray.
                    ¹ The window will close and the PC will start to reboot.
Checking the latest VB21 Patch according to the Repair Plan 3.5
             To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
             •    Insert the Aristos RADIS Patch VB21C CD into the CD drive.
             •    Navigate to the CD/DVD drive with the Windows® Explorer.
             •    Double-click on ParamUpdate.exe.
             •    The patch installation will begin; installation may take 15 ...20 sec.
             •    Click on OK.
             • Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
                  MX/VX, TX or FX/FX Plus into the CD drive.
             • Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
                  “start.exe” file. The Installer window opens.
             • Click on the link under "Service PC (+ Download Software) ...".
             • If queried, choose Full Install.
             • Choose Continue > OK.
Check for the latest Licenses (according to the Repair Plan) 3.8
             • To do this:
                  -   Go to Options > Service > Local Service... > Configuration
                  -   > Next
                  -   > Licensing
                  -   Insert the data medium with the new License.dat (it is possible that the new license
                      file still needs to be renamed to “license.dat”).
                  -   Click on “Browse..” and go to the new “license.dat”.
                  -   Click on “Open” > “Import License file” and confirm with “OK”.
                  -   Select “Home” and close the window.
                  -   Reboot the Radis PC: Options > End Session... > Restart System... > Yes
             • To test the Connectivity, click on the Test button. Since this is a very time-critical action,
                  it is possible that a connection will not be immediately established. Because of this, sev-
                  eral attempts should be made.
                     ¹ The server must send back a message, e.g.:
                       NOTE             If outside NAT is configured, the test function will not work
                                        and the message “...not alive” will be displayed.
                                        The new server name and IP address also have to be regis-
                                        tered with the local administrator and SRS Help Desk.
                                        Despite this, continue to the next step of the installation.
                     NOTE            If the text under Status is underlined, a link to the error mes-
                                     sage is overlaid. See (1/Fig. 81 / p. 84).
Restoring the Customer OGPs (only if the OGPs are newer than in
the Recovery Image)                                           3.10
             • Activate a test patient in RIS with several studies and check whether the studies match
                  after opening the patient on the image system.
                  l)   A new window comes up (Fig. 85 / p. 89) displaying the backup process of the sys-
                       tem partition. This will take a while ...
                  m) At the end of the backup routine, the COBR Disk Backup window appears
                     (Fig. 86 / p. 89) indicating backup success.
                  n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
                     paragraph.
                  o) Confirm the COBR Disk Backup window with <OK>(Fig. 86 / p. 89).
                  p) Go back to step c, select (D:) Med_Data (3/Fig. 84 / p. 88) and repeat the steps.
                     Then continue with step q.
                  q) In the COBR Disk Backup mask, press <Quit> (Fig. 84 / p. 88).
                       ¹ The window will close and a command window is displayed (Fig. 87 / p. 90).
             Perform a function check of the system in accordance with the modifications that have been
             made.
             • The customer or the CSE can easily report online any system problems that may occur.
                 - Under Help, select Report Error.
                 -
                                                • The following is a guideline of what will be covered (among other things) in this chapter:
                                                Tab. 8   Work Steps
             Creation of a repair plan is the best way to ensure that all necessary steps are performed
             efficiently and that multiple visits after setup of the PC are avoided.
             • Identify the SW version before the crash (possibly consider open patches or new ver-
                  sions - > to save time).
                  If going from VB10 to the new PC (latest version of VB21), additional collimator SW has
                  to be downloaded from Intranet.
                  List all performed updates and remote updates available in the update handling tool /
                  country SAP (notifications). List the sequence of updates needed to be installed.
             • Check whether a virus protection contract exists -> if yes, download the latest released
                  scan engine + latest virus pattern from the Intranet.
             • Options & Licenses:
                  Check whether there is a distribution list via remote (current license file on-site + ROD
                  Remote Options Distribution)
             • Download all required remote updates incl. UIs (by CSE (on laptop HD) or by USC (pre-
                  pare DVD etc., if applicable))
             To avoid extra work, check whether the collimator SW is lower than VC00N using the XCS
             SSW. If so, follow the work steps below. If not, skip this paragraph.
             • Go to the download location on the CS Home page: For Service > Product Informa-
                  tion > AX Systems > Radiography > Direct Technique > Software.
             • The required download file is the 10092710_OT_II_VC00N.exe and is located under
                  the description Collimator G052G.
             • Double-click on it and unzip it to a folder of your choice on the C or D drive of your ser-
                  vice PC.
             • You will find two files in the folder: CollimatorSWLoader.exe and
                  SW_D21_TBN_OT_II_VC00N.H86. Keep them there for later use.
             • If a virus is suspected, check the system using the bootable virus scanner.
             • Check the external removable devices (CD, DVD, USB Stick, USB disk) if a virus is sus-
                  pected.
             • Consider using the escalation routine.
                  What to do in case of a Virus Infection This should be implemented in CB Doc with a link
                  to the Knowledge Base.
             • Unplug the Net cable (virus protection).
Preparations 4.2
             1. Unlock the access to Windows® with Tab - Del - Transfer image (“+”).
             2. Press the Windows® key, and then Run; enter radissysinfo > OK.
3. Press Start.
5. The log file will be created and saved to the selected directory.
               8. Copy the compressed log file (2/Fig. 94 / p. 100) to a USB stick or similar medium, and
                  keep it ready in case it is requested by RSC/HSC/QIB/DCU.
                       ¹ The log files look like this: “Log<date>_<time>.zip“
               9. If a customer complaint has to be filed, send the file along with the complaint to Health-
                  care AX Complaint.
AXIOM Aristos Image System                AXB4-320.816.08.03.02   Page 100 of 134               © Siemens, 2011
                                          05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                      101
© Siemens, 2011                 AXB4-320.816.08.03.02   Page 101 of 134        AXIOM Aristos Image System
                                05.11                   CS PS XP RFU
102                        PC Replacement: SW Installation and Settings
              4. Copy the compressed log file (2/Fig. 97 / p. 102) to a USB stick or similar medium, and
                 keep it ready in case it is requested by RSC/HSC/QIB/DCU.
                         ¹ The log files look like this: “Log<date>_<time>.zip“
              5. If a customer complaint has to be filed, send the file along with the complaint to Health-
                 care AX Complaint.
Backup 0
              The following configurations have to be saved manually because these are Win-
              dows®-related configurations and are not part of the syngo backup.
              • Make a note of the service key:
                    - Options > Service > Local Service...
Service Key
AXIOM Aristos Image System           AXB4-320.816.08.03.02   Page 102 of 134                 © Siemens, 2011
                                     05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                     103
             • Make a note of the host name (= computer name) and the IP addresses:
                  - Options > Service > Local Service... > Configuration > TCP/IP LAN
                  - Home > close the window.
                  IP address
                  Subnet mask
                  Default gateway
                  Customer Configuration:
                  - > Options > Configuration.. > Regional and Language...
                  - “Regional Options”: Make a note of the date and time format.
                  - “ Languages > Details...”: Make a note of the language and keyboard.
                  Date Format
                  Time Format
                  Language
                  Keyboard (input locales)
Network Speed
© Siemens, 2011                 AXB4-320.816.08.03.02   Page 103 of 134       AXIOM Aristos Image System
                                05.11                   CS PS XP RFU
104                   PC Replacement: SW Installation and Settings
AXIOM Aristos Image System      AXB4-320.816.08.03.02     Page 104 of 134          © Siemens, 2011
                                05.11                     CS PS XP RFU
PC Replacement: SW Installation and Settings                                                    105
                  Select the Advanced tab card, read the value for Speed & Duplex or Link
                  Speed & Duplex and make a note of it.
             The following work steps describe how to create backups of “SW Settings02”,“ SW Settings
             Calibration” and “Customer OGPs” on floppy (or USB stick).
             The other packages do not have to be saved because there are newer versions with the
             new VB21 SW.
             • To do this, go to:
                  > Options > Service > Local Service > Backup & Restore
© Siemens, 2011                AXB4-320.816.08.03.02   Page 105 of 134       AXIOM Aristos Image System
                               05.11                   CS PS XP RFU
106                   PC Replacement: SW Installation and Settings
                 Server name
                 TCP/IP address
                 Local Host NAT
                 Network adapter
AXIOM Aristos Image System     AXB4-320.816.08.03.02   Page 106 of 134                 © Siemens, 2011
                               05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                      107
© Siemens, 2011                 AXB4-320.816.08.03.02   Page 107 of 134        AXIOM Aristos Image System
                                05.11                   CS PS XP RFU
108                   PC Replacement: SW Installation and Settings
AXIOM Aristos Image System      AXB4-320.816.08.03.02    Page 108 of 134                   © Siemens, 2011
                                05.11                    CS PS XP RFU
PC Replacement: SW Installation and Settings                                          109
© Siemens, 2011          AXB4-320.816.08.03.02   Page 109 of 134   AXIOM Aristos Image System
                         05.11                   CS PS XP RFU
110                   PC Replacement: SW Installation and Settings
                      NOTE            Before switching on the PC, make sure that only power, the
                                      mouse keyboard and the monitor are connected. Do not yet
                                      connect the network cable, fiber optic cables or the XCS ser-
                                      vice cable.
AXIOM Aristos Image System     AXB4-320.816.08.03.02   Page 110 of 134                 © Siemens, 2011
                               05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                       111
© Siemens, 2011                  AXB4-320.816.08.03.02     Page 111 of 134      AXIOM Aristos Image System
                                 05.11                     CS PS XP RFU
112                   PC Replacement: SW Installation and Settings
             • Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
                 MX/VX, TX or FX/FX Plus into the CD drive.
             • Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
                 “start.exe” file. The Installer window opens.
             • Click on the link under "Service PC (+ Download Software) ...".
             • If queried, choose Full Install.
             • Choose Continue > OK.
AXIOM Aristos Image System      AXB4-320.816.08.03.02   Page 112 of 134               © Siemens, 2011
                                05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                           113
Checking the latest VB21 Patch according to the Repair Plan 4.7
             To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
             •    Insert the Aristos RADIS Patch VB21C CD into the CD drive.
             •    Navigate to the CD/DVD drive with the Windows® Explorer.
             •    Double-click on ParamUpdate.exe.
             •    The patch installation will begin; installation may take 15 ...20 sec.
             •    Click on OK.
© Siemens, 2011                  AXB4-320.816.08.03.02     Page 113 of 134          AXIOM Aristos Image System
                                 05.11                     CS PS XP RFU
114                   PC Replacement: SW Installation and Settings
             • If the system is using a DAC, install the DAC program from the "DICOM Attribute Con-
                 fig. VA20A", P/N 7264435 or higher.
             • Follow the "DICOM Attrib. Configurator" Instructions, AX00-000.816.01.02… or higher.
AXIOM Aristos Image System     AXB4-320.816.08.03.02   Page 114 of 134                © Siemens, 2011
                               05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                        115
             • Go to “Options > Service > Local Service” and log in with the current service key.
             • Insert the backup CD / diskette / USB stick
             • Select Backup & Restore > Command > Backup > Drives: [-R-]CD-R; select the fol-
                  lowing packages :
             •    SW Settings02
             •    SW Settings Calibration
             •    IQAP_Database
             •    Only if coming from VB21: Customer_OGP_DBs
             •    Only if coming from VB21: Clinical Assurance
             •    Press GO for each one and wait for the Ready messages.
             •    Go back to the Local Service > Configuration menu. Click on Next until you come to
                  the actual Configuration menu.
             • If there is an newer “defect map” on site, it must be restored.
                  For this, go to RADIS Tools > Import FD Calib. Data.
                  Follow the description on the screen.
             • Select Home.
             • Close the window.
© Siemens, 2011                 AXB4-320.816.08.03.02     Page 115 of 134        AXIOM Aristos Image System
                                05.11                     CS PS XP RFU
116                    PC Replacement: SW Installation and Settings
Check for the latest Licenses (according to the Repair Plan) 4.10
             • To do this:
                 -   Go to Options > Service > Local Service... > Configuration
                 -   > Next
                 -   > Licensing
                 -   Insert the data medium with the new License.dat (it is possible that the new license
                     file still needs to be renamed to “license.dat”).
                 -   Click on “Browse..” and go to the new “license.dat”.
                 -   Click on “Open” > “Import License file” and confirm with “OK”.
                 -   Select “Home” and close the window.
                 -   Reboot the Radis PC: Options > End Session... > Restart System... > Yes
AXIOM Aristos Image System       AXB4-320.816.08.03.02    Page 116 of 134                    © Siemens, 2011
                                 05.11                    CS PS XP RFU
PC Replacement: SW Installation and Settings                                                       117
© Siemens, 2011                 AXB4-320.816.08.03.02    Page 117 of 134        AXIOM Aristos Image System
                                05.11                    CS PS XP RFU
118                    PC Replacement: SW Installation and Settings
             • To test the Connectivity, click on the Test button. Since this is a very time-critical action,
                 it is possible that a connection will not be immediately established. Because of this, sev-
                 eral attempts should be made.
                     ¹ The server must send back a message, e.g.:
                       NOTE             If outside NAT is configured, the test function will not work
                                        and the message “...not alive” will be displayed.
                                        The new server name and IP address also have to be regis-
                                        tered with the local administrator and SRS Help Desk.
                                        Despite this, continue to the next step of the installation.
AXIOM Aristos Image System        AXB4-320.816.08.03.02   Page 118 of 134                     © Siemens, 2011
                                  05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                       119
© Siemens, 2011                 AXB4-320.816.08.03.02      Page 119 of 134      AXIOM Aristos Image System
                                05.11                      CS PS XP RFU
120                   PC Replacement: SW Installation and Settings
                      NOTE            If the text under Status is underlined, a link to the error mes-
                                      sage is overlaid. See (1/Fig. 108 / p. 119).
AXIOM Aristos Image System     AXB4-320.816.08.03.02   Page 120 of 134                 © Siemens, 2011
                               05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                     121
Restoring the Customer OGPs (only if the OGPs are newer than in
the Recovery Image)                                           4.12
             • Activate a test patient in RIS with several studies and check whether the studies match
                  after opening the patient on the image system.
© Siemens, 2011                 AXB4-320.816.08.03.02    Page 121 of 134      AXIOM Aristos Image System
                                05.11                    CS PS XP RFU
122                   PC Replacement: SW Installation and Settings
             • The following work steps describe how to create backups for all “Customer OGPs” set-
                 tings on floppy (or USB stick) and Configuration Data on CD (or USB stick)
             • To do this, go to:
                 > Options > Service > Local Service > Backup & Restore
                 - Insert the backup CD / USB stick
                 - Under Command > Backup > Drives: [-R-]CD-R, select the following packages:
                 - SW Settings02
                 - Customer_OGP_DBs
                 - IQAP_Database
                 - Clinical Assurance
                 - SW Settings Calibration
                 - Press GO for each one and wait for the Ready messages.
                 - Press Home > Radis Tools > Organ Programs.
                 - Select A:\ for diskette (recommended) or the drive for USB stick under “Drive
                   Select”.
                 - Press “Backup” to save the OGPs.
                 - Close the window and press “Home”.
AXIOM Aristos Image System     AXB4-320.816.08.03.02    Page 122 of 134                  © Siemens, 2011
                               05.11                    CS PS XP RFU
PC Replacement: SW Installation and Settings                                                     123
© Siemens, 2011                 AXB4-320.816.08.03.02   Page 123 of 134       AXIOM Aristos Image System
                                05.11                   CS PS XP RFU
124                   PC Replacement: SW Installation and Settings
                 l) A new window comes up (Fig. 112 / p. 124) displaying the backup process of the
                    system partition. This will take a while ...
                 m) At the end of the backup routine, the COBR Disk Backup window appears
                    (Fig. 113 / p. 124) indicating backup success.
                 n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
                    paragraph.
                 o) Confirm the COBR Disk Backup window with <OK>(Fig. 113 / p. 124).
                 p) Go back to step c, select (D:) Med_Data (3/Fig. 111 / p. 123) and repeat the steps.
                    Then continue with step q.
                 q) In the COBR Disk Backup mask, press <Quit> (Fig. 111 / p. 123).
                      ¹ The window will close and a command window is displayed (Fig. 114 / p. 125).
AXIOM Aristos Image System      AXB4-320.816.08.03.02   Page 124 of 134                  © Siemens, 2011
                                05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                      125
© Siemens, 2011                  AXB4-320.816.08.03.02   Page 125 of 134       AXIOM Aristos Image System
                                 05.11                   CS PS XP RFU
126                   PC Replacement: SW Installation and Settings
             Perform a function check of the system in accordance with the modifications that have been
             made.
AXIOM Aristos Image System     AXB4-320.816.08.03.02   Page 126 of 134                   © Siemens, 2011
                               05.11                   CS PS XP RFU
PC Replacement: SW Installation and Settings                                                         127
             • The customer or the CSE can easily report problems with the system online.
                  - Under Help, select Report Error.
                  -
© Siemens, 2011                  AXB4-320.816.08.03.02    Page 127 of 134         AXIOM Aristos Image System
                                 05.11                    CS PS XP RFU
128                                                           Useful Information and Correction of Errors
                                               5-
For Information Only
5Useful Information and Correction of Errors
                                               • The customer or the CSE can easily report problems with the system online.
                                                    - Under Help, select Report Error.
                                                    -
                                               If there are display problems with an EIZO monitor connected to Radis via the Digital Video
                                               Interface (DVI) in the Celsius M410 (gray housing), carry out these settings:
AXIOM Aristos Image System                                         AXB4-320.816.08.03.02   Page 128 of 134                 © Siemens, 2011
                                                                   05.11                   CS PS XP RFU
Useful Information and Correction of Errors                                                         129
             • Unlock Windows® by pressing “Tab + Del. + Send (Num field)”, open the Explorer and
                  go to Control Panel; press GO; if necessary, select "classical view" in the window
                  that comes up; select: the Display > Settings tab card.
                  -   Color quality: Highest (32-bit)
                  -   Screen resolution: 1280 x 1024 pixels
                  -   Click on "Advanced" > Monitor > Set Screen refresh rate: 60 Hz
                  -   Close all windows
Using XP Explorer 0
Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Explorer.
Configuring the CDs/DVDs as Offline Devices for Export of 12-bit Images (only if
needed)                                                                         0
             This feature can be used only if the CD/DVD burner is configured in the Local Service con-
             figuration.
             • Go to Options > Local Service... > Configuration
             • Make sure that "DICOM Offline Devices" is enabled.
© Siemens, 2011                   AXB4-320.816.08.03.02   Page 129 of 134        AXIOM Aristos Image System
                                  05.11                   CS PS XP RFU
130                          Useful Information and Correction of Errors
AXIOM Aristos Image System      AXB4-320.816.08.03.02      Page 130 of 134          © Siemens, 2011
                                05.11                      CS PS XP RFU
Useful Information and Correction of Errors                                                       131
• After reboot, go to Options > Local Service... "RADIS Tools" > "CD Recording".
© Siemens, 2011                  AXB4-320.816.08.03.02     Page 131 of 134     AXIOM Aristos Image System
                                 05.11                     CS PS XP RFU
132                          Useful Information and Correction of Errors
             •   Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Explorer.
             •   Go to "D:\Images\Exported14bitImages”
             •   Delete all files in this directory.
             •   Close all windows.
             The reason for a failed installation could be that fragments from the old installation were not
             completely deleted. Following a failed installation, a low level formatting of both hard drives
             should be performed prior to making a second attempt. To do this:
             •   During the reboot, press the "CTRL + A" keys.
             •   Confirm the selection.
             •   Under "Options", select "SCSI Disk Utilities" with the arrow keys.
             •   Confirm the selection with Enter.
             •   Select the 1st hard drive (SCSI ID #0) in the table.
             • Confirm the "Format Disk" selection with Enter.
             • Confirm the safety inquiry with "Yes".
             • Repeat the procedure for the 2nd hard drive (SCSI ID #1) and then exit the Adaptec
                 tool.
AXIOM Aristos Image System        AXB4-320.816.08.03.02   Page 132 of 134                    © Siemens, 2011
                                  05.11                   CS PS XP RFU
Changes to Previous Version                                                                                     133
6Changes to Previous Version   6-
                               Chapters 2, 3 and 4: included backup and restore of the “IQAP Database”.
© Siemens, 2011                                 AXB4-320.816.08.03.02   Page 133 of 134      AXIOM Aristos Image System
                                                05.11                   CS PS XP RFU
134                                              Changes to Previous Version