Beacon 200 Gas Monitor Manual

The document is the operator's manual for the Beacon 200 Gas Monitor. It provides instructions on installation, operation, and maintenance of the device. Key points covered include a 1-year product warranty, mounting and wiring procedures, alarm indications, calibration instructions, and troubleshooting tips. Maintenance topics such as replacing fuses and preventative maintenance are also addressed.

Uploaded by

enriquepm06i
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
113 views53 pages

Beacon 200 Gas Monitor Manual

The document is the operator's manual for the Beacon 200 Gas Monitor. It provides instructions on installation, operation, and maintenance of the device. Key points covered include a 1-year product warranty, mounting and wiring procedures, alarm indications, calibration instructions, and troubleshooting tips. Maintenance topics such as replacing fuses and preventative maintenance are also addressed.

Uploaded by

enriquepm06i
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 53

Beacon 200 Gas Monitor

Operator’s Manual

Part Number: 71-0059RK


Revision: K
Released: 3/9/20

www.rkiinstruments.com
Product Warranty

RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects
in materials, workmanship, and performance for a period of one year from date of
shipment from RKI Instruments, Inc. Any parts found defective within that period will
be repaired or replaced, at our option, free of charge. This warranty does not apply to
those items, which by their nature, are subject to deterioration or consumption in
normal service, and which must be cleaned, repaired, or replaced on a routine basis.
Examples of such items are as follows:

a) Absorbent cartridges d) Batteries

b) Pump diaphragms and valves e) Filter elements

c) Fuses

Warranty is voided by abuse including mechanical damage, alteration, rough handling,


or repair procedures not in accordance with the operator’s manual. This warranty
indicates the full extent of our liability, and we are not responsible for removal or
replacement costs, local repair costs, transportation costs, or contingent expenses
incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation
of this gas monitor, and our warranty is limited to the replacement of parts or our
complete goods.

ii • Product Warranty
Table of Contents

Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
External Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Internal Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 3: Installation and Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mounting the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Wiring the Beacon 200 Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Starting Up the Beacon 200 Gas Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Chapter 4: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Recorder Output Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Viewing & Resetting Min/Max Readings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Chapter 5: Channel Control and Setup Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Enable/Disable Channel(s) Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Configure Channel Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
View System Information Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

Table of Contents • iii


Chapter 6: Input Setup Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Setting Up a New Channel or Changing an Existing Channel . . . . . . . . . . . . . . . . . . . . . 41
Chapter 7: Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Calibration Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Replacing the Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

iv • Table of Contents
Chapter 1: Introduction

Overview

This chapter briefly describes the Beacon 200 Gas Monitor. This chapter also describes the
Beacon 200 Gas Monitor Operator’s Manual (this document). Table 1 at the end of this
chapter lists the specifications for the Beacon 200.

About the Beacon 200 Gas Monitor

The Beacon 200 is a fixed-mounted, continuous-monitoring gas detection instrument. This


gas monitor is capable of detecting gas at up to two locations. The display screen
simultaneously displays the gas readings of the active channel or channels. Both direct
connect (internal amplifier type) detector heads and 4 - 20 mA transmitter (remote amplifier
type) detector heads may be used with the Beacon 200.

NOTE: If any of the Beacon 200’s channels have been set up for hydrogen-specific LEL
operation, that channel or channels can only be used with a hydrogen-specific
sensor. If you want to detect a different combustible gas or use a different detector,
the Beacon 200 must be sent back to RKI Instruments, Inc. for modification.

The Beacon 200 includes audible and visual alarms that warn you of hazardous gas
conditions. The alarm circuit includes two levels of gas alarms. The fail circuit alerts you to
failures in the gas detector heads or Beacon 200.
Three instrument programs allow you to display and change channel and calibration settings
and change channel types. They are the Channel Control & Setup Program, the Calibration
Program, and the Input Setup Program.

Overview • 5
About this Manual

The Beacon 200 Gas Monitor Operator’s Manual is organized as follows:


• Chapters 1 through 6 describe components of the Beacon 200 and procedures to install,
start up, operate, and maintain the Beacon 200.
The Beacon 200 Gas Monitor Operator’s Manual uses the following conventions for notes,
cautions, and warnings.

NOTE: Describes additional or critical information.

CAUTION: Describes potential damage to equipment.

WARNING: Describes potential danger that can result in injury or death.

! Caution: refer to accompanying documentation

~ Vac (AC voltage)


Vdc (DC voltage)

6 • About this Manual


Specifications

Table 1 lists specifications for the Beacon 200 Gas Monitor.


Table 1: Beacon 200 Specifications

Description Specification

Input Power 115/220V ~ ±10%, 50/60Hz, 0.5/0.3A


or
24 V +10% -8%, 0.6A
Construction (housing) Fiberglass/polyester with lexan window (NEMA 4X)
Dimensions 10.5 in. H x 8.5 in. W x 6.25 in. D
(267 mm H x 216 mm W x 158 mm D)
Weight 8 lbs.
Environmental • For indoor or outdoor locations (Type 4X)
Conditions • 2000m max. altitude
• -20°C to 50°C (-4°F to 122°F) max. ambient
• Maximum humidity of 80% relative
• Mains supply voltage fluctuations not exceeding ± 10% of nominal
• DC supply voltage fluctuations not exceeding +10% -8% of nominal
• Overvoltage Category II, Pollution Degree 2
Safety/Regulatory

C US
186718
Relays • Relay contacts rated for 10A @ 115/220V~ resistive or 10A @ 30V
resistive
• SPDT Form C
User Controls • Reset switch
• Program buttons: ESCAPE, UP/YES, DOWN/NO, and ENTER
Standard Accessory Operator’s manual (this document)

CAUTION: Cleaning is not to be performed.

CAUTION: The operator of this instrument is advised that if the equipment is used in a
manner not specified in this manual, the protection provided by the equipment
may be impaired.

Specifications • 7
Chapter 2: Description

Overview

This chapter describes external and internal components of the Beacon 200 Gas Monitor.

External Description

This section describes the housing and all external components of the Beacon 200. For the
purposes of this description, the housing door is considered the front of the monitor.

Housing
The Beacon 200’s fiberglass housing is weather- and corrosion-resistant. It is suitable for
installation where general purpose equipment is in use. The housing door is hinged on the left
side and is secured by two latches on the right side. The display screen and status lights are
visible through windows in the housing door. Four mounting feet are attached to the back of
the housing (one at each corner). The mounting feet allow you to install the housing to a
vertical surface. Three conduit hubs on the bottom of the housing are for external wiring
connections.

CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

CAUTION: To avoid electrical interference, do not route detector head and power wiring
through the same conduit hub.

Reset Switch
The reset switch is on the bottom of the housing. It is in front of the conduit hubs. The reset
switch serves three functions:
• Resets the alarm circuits for “latched” alarms after an alarm 1 or alarm 2 condition passes.
You can set each channel for latched or self-resetting alarms in the Channel Control &
Setup Program.
• Silences the buzzer during an alarm 1 or alarm 2 condition. You cannot silence failure
alarms.

8 • Overview
• Displays and resets the minimum and maximum gas concentration values.

Buzzer
The buzzer is on the bottom of the housing. It is on the far right. The buzzer sounds audible
alarms to warn you of gas alarms and instrument failures.

Internal Description

This section describes the internal components of the Beacon 200.

Main PCB

Display PCB
(see Figure 2)

Power
Supply
Power
Switch
DC Fuse

AC Fuses
Alarm Relays
(see Figure 3)
AC In
Terminal
Strip

Detector/Transm itter
Terminal Strip,
Channel 1

Detector/Transm itter
Terminal Strip,
Channel 2

Controller
Terminal
Strip

3/4" Conduit Alarm


Hub, 3X Terminal
Strip

Reset Switch Buzzer

Figure 1. Beacon 200 Gas Monitor Component Location

Internal Description • 9
Display Printed Circuit Board (PCB)
The display PCB is mounted to the power supply mounting plate which is in turn mounted to
the main PCB. The power supply mounting plate and main PCB are described below. The
display PCB includes the display, the status lights, and the program buttons.

Display
The display simultaneously indicates the channel number, current gas reading, measuring unit,
and target gas of all active channels.
The display also indicates messages, settings, and other data when you are operating the
instrument programs.

Display Cable Connector

ESCAPE
Button

UP/YES
Button

Fail LED
DOWN/NO
Button
Alarm 2
LED
ESCAPE
Alarm 1 Button
LED
Buzzer

Contrast LCD Pilot LED


Adjust Pot

Figure 2. Display Board Component Location

Status Lights
The Beacon 200 includes four status lights that indicate the current status of the monitor. The
status lights are to the left and right of the display (see Figure 2).
• Pilot Light. The pilot light is on when the Beacon 200 is receiving incoming power.
• Fail Light. The fail light turns on when the Beacon 200 is experiencing a fail condition. A
fail condition can be caused by a failure within the Beacon 200 or detector head(s) wired
to the Beacon 200. See Chapter 5, Maintenance, or the troubleshooting section in the
applicable detector head manual to respond to a fail condition.
• Alarm 1 Light. The alarm 1 light is on when the Beacon 200 is experiencing an alarm 1
gas condition.

10 • Internal Description
• Alarm 2 Light. The alarm 2 light is on when the Beacon 200 is experiencing an alarm 2
gas condition.

Program Buttons
The Beacon 200 includes four program buttons that allow you to enter the instrument
programs, navigate through the programs, update instrument and channel settings, and save
changes to the program settings. When a program button is pressed, a buzzer located on the
display PCB beeps. The program buttons are near the right edge of the display PCB (see
Figure 2).
Table 2: Beacon 200 Program Button Functions

Button Function

ESCAPE • Moves backward through the program menus


• Cancels changes you make in the program menus
• Enters the Channel Control and Setup program (press with the ENTER button)
UP/YES • Accepts the displayed setting and proceeds to the next setting
• Changes the displayed setting
• Enters the Calibration program (press with the ENTER button)
DOWN/NO • Allows you to update the displayed setting
• Changes the displayed setting
ENTER • Saves changes you make in the programs
• Enters the Channel Control and Setup program (press with ESCAPE button)
• Enters the Calibration program (press with the UP/YES button)
• Enters the Input Setup Program (press while flipping power switch to ON
position)

Main PCB
The main PCB is mounted inside the housing. The power supply mounting plate is mounted to
the main PCB with four standoffs and the display PCB is mounted to the power supply
mounting plate with four standoffs. The main PCB includes the terminal strips, relays, fuses,
and power switch.

NOTE: If any of the Beacon 200’s channels have been set up for hydrogen-specific LEL
operation, the main PCB for that channel or channels has been modified and can
only be used with a hydrogen-specific sensor. If you want to detect a different
combustible gas or use a different detector, the Beacon 200 must be sent back to
RKI Instruments, Inc. for modification.

Internal Description • 11
Terminal Strips
The Beacon 200 includes four terminal strips for external wiring connections. See “Wiring the
Beacon 200 Gas Monitor” for detailed wiring procedures.
• Detector/Transmitter Terminal Strips. Two detector/transmitter terminal strips are
located near the bottom left corner of the main circuit board (see Figure 1). These two 9-
point terminal strips facilitate wiring connections to the detectors or transmitters.
Although each terminal strip can accommodate several different detector head models,
only one detector head at a time may be wired per channel. The top terminal strip is for
channel 1 and the bottom terminal strip is for channel 2.
• Alarm Terminal Strip. The alarm terminal strip is located along the bottom edge of the
main circuit board (see Figure 1). This 21-point terminal strips facilitates wiring
connections to external alarm devices (horn, strobe, etc.). Terminals are provided for
individual channel as well as common alarm relay contacts.
• Controller Terminal Strip. The 10-point controller terminal strip is near the lower right
edge of the main circuit board (see Figure 1). The controller terminal strip facilitates
various internal and external wiring connections. Table 3 lists the function of each
terminal.
Table 3: Terminal Assignments for the Controller Terminal Strip

Terminal Connects to:

BAT - - connection from 24 VDC power source1 (or 24 V backup battery)


BAT + + connection from 24 VDC power source1 (or 24 V backup battery)
+ CH1 OUT + connection of 4 - 20 mA output, channel 1
- CH 1 OUT - connection of 4 - 20 mA output, channel 1
+ CH2 OUT + connection of 4 - 20 mA output, channel 2
- CH2 OUT - connection of 4 - 20 mA output, channel 2
RESET (2) Reset switch (factory-wired)
BUZ-/BUZ+ Internal buzzer (factory-wired)
1 If DC power is used as the primary power source, do not make wiring connections to
the AC terminal strip.

12 • Internal Description
• AC In Terminal Strip. The AC in terminal strip is a 2-point terminal strip located above
the controller terminal strip (see Figure 1). It facilitates wiring from the AC power source.
Table 4 lists the function of each terminal.
Table 4: Terminal Assignments for the AC In Terminal Strip

Terminal Connects to:

LINE Hot wire from AC power source.


NEUT Neutral wire from AC power source.

NOTE: The AC power source’s ground wire must be connected to the ground stud. See
“Connecting the AC Power Source” on page 18 for instructions.

Relays
The Beacon 200 includes four channel relays (two per channel) and three common relays.
Both sets of relays are single-pole, double-throw (SPDT) and are rated for 10 amps at 115/220
VAC (resistive) or 10 amps at 30 VDC (resistive).

NOTE: You can select normally energized (NE) or normally de-energized (NDE) settings
for each channel in the Channel Control and Setup program. This section describes
the default setting: normally de-energized.
The alarm 1 and alarm 2 common relays are factory-set as NDE and the fail
common relay is factory-set as NE. The alarm 1, alarm 2, and fail common relays’
NE/NDE settings are not user-adjustable.

• Channel Relays. The four channel relays are above the alarm terminal strip (see
Figure 1). These relays are dedicated to specific channels and alarm levels.
For example, the channel 1, alarm 1 relay energizes if channel 1 recognizes an alarm 1
condition. Figure 3 below illustrates the allocation of the channel relays.
Channel 2, Alarm 2 Common Alarm 1

Channel 2, Alarm 1
Common Alarm 2
Channel 1, Alarm 2

Channel 1, Alarm 1 Common Fail

K1 K2 K3 K4 K5 K6 K7

Figure 3. Beacon 200 Channel Relay Allocation

Internal Description • 13
NOTE: The alarm 2 channel relays may be set to operate as individual channel fail relays.
See the Configure Channel Settings section of Chapter 5 for instructions.

• Common Relays. The three common relays, alarm 1, alarm 2, and fail, are to the left of
the controller terminal strip (see Figure 1). These relays are common for both channels.
For example, the alarm 1 common relay energizes if either channel 1 or channel 2
recognizes an alarm 1 condition.

Fuses
There are three fuses that are used in the Beacon 200. Two of them are AC fuses and one of
them is a DC fuse.
• AC Fuses. The two fuses located directly to the left of the controller terminal strip are the
AC fuses. They cut off the incoming AC power in the event of a short circuit or other
electrical fault which causes a high current draw in the Beacon 200. They are housed in
vertical fuse holders and are held in the holder by a quarter turn cover. They are labelled as
F2 (top fuse) and F3 (bottom fuse) on the PCB silk-screen and are rated to 3 A, 250 V, 1/4
x 1 1/4”, fast acting.
• DC Fuse. The fuse located to the left of the power switch is the DC fuse. It cuts off
incoming DC power in the event of a short circuit or other electrical fault which causes a
high current draw in the Beacon 200. It is also housed in a vertical fuse holder and is held
in the holder by a quarter turn cover. It is labelled as F1 on the PCB silk-screen and is
rated to 6A, 250 V, 1/4 x 1 1/4”, fast acting.

Power Switch
The power switch is located above the relays and in between the AC and DC fuses (see
Figure 1). The power switch turns the incoming AC power source on and off at the Beacon
200. When the switch is up, the power switch is on.

NOTE: The DC power input has no on/off switch and is not affected by the position of this
switch.

Power Supply
The power supply is mounted to the power supply mounting plate which is located behind the
display PCB. The power supply mounting plate is mounted to the main PCB with four
standoffs. The power supply receives the AC input from the external power source and
converts it to a DC voltage that is usable by the Beacon 200 circuitry.

14 • Internal Description
Chapter 3: Installation and Start Up

Overview

This chapter describes procedures to mount the Beacon 200 Gas Monitor, make wiring
connections to the monitor, and start up the monitor.

WARNING: Perform all installation and start-up procedures in a “fresh air” environment
(known to be free of combustible gas, toxic gas, and of normal oxygen
content). The Beacon 200 is not in operation as a gas monitoring system until
the start-up procedure is complete.

Mounting the Beacon 200 Gas


Monitor

WARNING: Only authorized and properly trained personnel should perform any
mounting procedures.

Perform the following procedure to install the Beacon 200 at the mounting site.
1. Select the mounting site. When you select the mounting site consider the following
factors:
• Is an AC or DC power source available?
• Is there enough room to open the housing door and make wiring connections through
the conduit hubs at the bottom of the housing?
• Are the display screen and status lights visible?
2. Close and latch the housing door.
3. Position the monitor on a vertical surface at eye level (4 1/2 to 5 feet from the floor).
4. The Beacon 200 is shipped with the mounting feet positioned behind the housing. Loosen
the screws that secure the feet to the housing, rotate the feet to their mounting position (as
shown in Figure 4), then tighten the screws.
5. Insert 1/4 in. or 5/16 in. screws through the slots in the mounting feet at each corner of the
housing to secure the housing to the mounting surface (see Figure 4).

Overview • 15
6. Each of the door clamps has a feature for locking device installation. A locking device that
requires a tool to unlock must be installed in each door clamp.

.41 8.50

Ø .38 x .50 slot, 4X 6.00 Mounting


Feet, 4X

ALARM 2 FAIL

10.94 10.50
ALARM 1 PILOT

BEACON 200
GAS MONITOR

RESET

.80 3/4 " Conduit Hub(3)

Figure 4. Beacon 200 Gas Monitor Outline and Mounting Dimensions

16 • Mounting the Beacon 200 Gas Monitor


AC In T erm inal S t rip W irin g

L in e (H ot)
1 15 V/2 2 0V ~
Cont rolle r Te rm ina l Ne utr al
S trip W iring
Ty pica l De te ct or/T rans mitt er
N OT E: A C g round
Te rmina l S trip W iring BAT c onnec ti on i s m ade
24 V
O nl y o ne d etec tor or t ransm i tter c a n be BAT + t o g round s tud i n
wi red to C h1 o r C h2 a t a tim e. S ee d ete ct or low er r ight c orne r
head w iring d iagr am for s peci fi c wi ring. C ha nn el 1 R e cord e r + CH 1
1 K M a x Im p e da nc e 4 - 2 0 m A O ut
O UT of c i rcui t b oard.

C H A N N EL 1 C ha nn el 2 R e cord e r + CH 2
1 K M a x Im p e da nc e 4 - 2 0 m A O ut
L EL
O UT
A M P /P R E A M P O XYG EN

R ese t S wi t ch
RE S E T
+ S + R W G B
(F a ctor y W i re d)
RE S E T
Bu zze r
B UZ
(F a ctor y W i re d)
B UZ +

G re e n
O xyg e n
D e te cto r Ty pica l Ala rm Re lay T erm in al S trip W iring
W h it e
C H1 -A1 C H1 -A2 C H2 -A1 C H2 -A2 C O M -A1 C O M -A2 C O M -F AIL

NO NC C NO NC C NO NC C NO NC C NO NC C NO NC C NO NC C
- DC G RO UND
3 -W ire 4 -2 0 m a
FB ( 4-2 0 m A )
Tra nsm itt er
+ 24V

C H A N N EL 2
A M P /P R E A M P O XYG EN L EL

+ S + R W G B

Alarm D evice
Pow er

B lac k

G re e n

W h it e

Re d
L EL D e te ct or
A la rm D e vi ces

FB ( 4-2 0 m A ) Co nta ct R a ti ng o f 1 0 A m p s a t 1 1 5 /2 2 0 V ~
2 -W ire 4 -2 0 m a
Tra nsm itt er Re sistiv e o r 1 0A @ 3 0V R esi sti ve f or E ach
+ 24V S et o f A la rm R el ay C on tacts.

Figure 5. Beacon 200 Gas Monitor General External Wiring Diagram

Mounting the Beacon 200 Gas Monitor • 17


Wiring the Beacon 200 Gas Monitor

WARNING: Only authorized and properly trained personnel should perform any wiring
procedures.

This section describes procedures to connect the AC power source, DC power source, external
alarm(s), recorder, and detector head(s). See Figure 5 for a general wiring diagram of all
external wiring to the Beacon 200.

WARNING: Make all connections to the Beacon 200 before you plug in or connect mains
power. Before you make any wiring adjustments, always verify that mains
power to the Beacon 200 is turned off or disconnected.

Routing Wiring Into the Beacon 200 Housing


Wiring must be brought into the housing through one of the three factory-installed conduit
hubs on the bottom of the housing.
Do not drill into the Beacon 200 housing for any reason. Drilling the Beacon 200 housing and
routing wiring through holes not factory drilled will void the warranty and could result in:
• Damage to internal components from the drilling process.
• Moisture damage to internal components from poorly sealed holes.
• Unpredictable Beacon 200 behavior due to EMI/RFI interference caused by wires routed
across the PCBs.
• Possible shorting of Beacon 200 components due to wires routed across the PCBs.

Connecting the AC Power Source

NOTE: If you are using DC power as the primary power source, go to the next section,
“Connecting the DC Power Source.”

The AC In terminal strip will accept 24 - 14 AWG wire. For 115 VAC connection, select wire
that is rated to a minimum of 80°C, 0.75 A, and 150V. For 220 VAC connection, select wire
that is rated to a minimum of 80°C, 0.5 A, and 250 V. When selecting wire, be sure to meet the
local electrical code.
A certified switch or circuit breaker must also be installed in the AC line. For 115 VAC
connection, the switch or circuit breaker must be rated to 115 VAC, 50/60 Hz, and 0.5 A. For
220 VAC connection, the switch or circuit breaker must be rated to 220 VAC, 50/60 Hz, and
0.3 A.
Perform the following procedure to connect the AC power source to the Beacon 200.

18 • Wiring the Beacon 200 Gas Monitor


WARNING: Verify that mains power to the Beacon 200 is turned off or disconnected
before you continue with this procedure.

1. Turn off or disconnect mains power to the Beacon 200.


2. Open the housing door, then place the power switch in the OFF position.
3. Install an appropriately rated cable bushing or conduit to the right-most conduit hub on the
bottom of the Beacon 200 housing.

CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

4. Locate the 2-point AC In terminal strip (see Figure 1 on page 9). The terminals are
labelled LINE and NEUT.
5. Guide the AC power cord or wires through the right-most conduit hub on the bottom of
the Beacon 200 housing.

CAUTION: Do not route power and detector head wiring through the same conduit hub. The
power wiring may disrupt the transmission of the detector head signal to the
monitor.

6. Connect the line (hot) and neutral AC wires to the AC power terminal strip as shown in
Figure 6.

AC In Terminal Strip
Disconnect Device

Line (Hot) 115 V/220 V ~ ± 10%,


Neutral 50/60 Hz

Figure 6. AC Power Wiring


7. Connect the ground wire to the unused crimp terminal that is factory installed at the base
of the ground stud.
a. Remove all installed connections (from conduit hubs and power supply).
b. Remove the bottom lug.
c. Crimp the ground wire to the lug.
d. Reinstall the lug.

Wiring the Beacon 200 Gas Monitor • 19


e. Reinstall all of the other ground connections (from conduit hubs and power supply).

WARNING: Follow this grounding procedure to maintain the CSA classification of the
Beacon 200.

Connecting the DC Power Source

WARNING: Verify that mains power to the Beacon 200 is turned off or disconnected
before you continue with this procedure.

DC power may be used as a primary power source. If DC power (connected to the BAT
terminals on the controller terminal strip) is the primary power source, DO NOT connect AC
power to the AC In terminal strip.
A 24 VDC battery (connected to the BAT terminals on the controller terminal strip) may also
be used as a backup power source if AC power (connected to the AC In terminal strip) is the
primary power source.
The controller terminal strip will accept 24 - 14 AWG wire. Select wire that is rated to a
minimum of 80°C, 0.75 A, and 150V. When selecting wire, be sure to meet the local electrical
code.

WARNING: The BAT terminals on the controller terminal strip are only intended for
connecting a battery or a 24 VDC supply to the Beacon 200. Do not connect
any horns or strobes to the BAT terminals on the controller terminal strip.

1. Turn off or disconnect mains power to the Beacon 200.


2. Open the housing door, then place the power switch in the OFF position. Locate the DC
power terminals on the controller terminal strip (see Figure 1 on page 9). They are labelled
BAT - and BAT +
3. Install an appropriately rated cable bushing or conduit to one of the unused conduit hubs
on the bottom of the Beacon 200 housing.

CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

4. Guide a DC power cord or wires through the selected conduit hub.

CAUTION: Do not route power and detector head wiring through the same conduit hub. The
power wiring may disrupt the transmission of the detector head signal to the
monitor.

20 • Wiring the Beacon 200 Gas Monitor


CAUTION: When a battery is used as backup power, the Beacon 200 trickle charges the
battery. Do not use a non-rechargeable battery as backup power. Use RKI
backup battery 49-8102RK or an appropriately rated rechargeable lead acid
type battery.

5. Connect the DC wires to the controller terminal strip as shown in Figure 7.

Controller Terminal Strip

BAT
24 VDC ± 2.5 VDC
+ BAT +

Figure 7. DC Power Wiring

NOTE: If a 24 VDC battery is used as a backup power source, as long as AC power is on


the Beacon 200 will keep a trickle charge on the battery to maintain its charge. If AC
power is interrupted, the Beacon 200 will operate from the DC backup battery until
the battery voltage drops to 21.5 volts or less, or until AC power is restored. Battery
recharge time will vary depending on how much the battery was drained.

Connecting External Alarms


The relays are rated for 10 amps at 115/220 VAC (resistive) or 10 amps at 30 VDC (resistive).
The alarm terminals will accept 24 - 14 AWG wire. Select wire that is rated to a minimum of
80°C, up to 10 A depending on the device being powered, and up to 250V depending on the
device being powered.
Perform the following procedure to connect external alarm devices to the Beacon 200.

NOTE: The alarm terminal strip includes terminals for channel alarms and common
alarms. Channel alarms are activated by one particular channel. Common alarms are
activated by either of the two channels. The example used in this procedure
describes connecting an external alarm device to one of the channel alarm terminals:
the channel 1, alarm 1 terminals.

WARNING: Do not connect external alarms to the BAT terminals on the controller
terminal strip.

1. Turn off or disconnect mains power to the Beacon 200.


2. Open the housing door, then place the power switch in the OFF position. Locate the alarm
terminal strip (see Figure 1 on page 9.
3. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the
bottom of the Beacon 200 housing.

Wiring the Beacon 200 Gas Monitor • 21


CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

4. Guide the wiring of the external alarm device through the selected conduit hub on the
bottom of the Beacon 200 housing.

CAUTION: Do not route the external alarm wiring and detector head wiring through the
same conduit hub. The external alarm wiring may disrupt the transmission of the
detector signal to the Beacon 200.

5. Connect the leads from the external alarm device and power to the alarm terminals as
shown in Figure 8.
Beacon 200 Alarm
CH1-A1 Terminal Strip
NO NC C

+ (H)
External Alarm External Power
Device Source
- (N)

Figure 8. Typical External Alarm Wiring


6. Repeat steps 3 through 5 for additional external alarm devices.

Connecting a Recorder
The recorder output terminals will accept 24 - 14 AWG wire. Select wire that is rated to a
minimum of 80°C, 0.2 A, and 150V.
Perform the following procedure to connect an analog recording device to the Beacon 200.
The output at the recorder output terminals is a 4 - 20 mA signal that is proportional to the
detection range of the applicable detector head.
1. Turn off or disconnect mains power to the Beacon 200.
2. Open the housing door, then place the power switch in the OFF position. Locate the
recorder output terminals on the controller terminal strip (see Figure 1 on page 9).

22 • Wiring the Beacon 200 Gas Monitor


3. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the
bottom of the Beacon 200 housing.

CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

4. Guide the wiring from the recording device through the selected conduit hub on the
Beacon 200.
5. Connect the wires from the recording device to the recorder output terminals as shown in
Figure 9.

Recording Device
#1, 1Kohm Max
+
Impedance
+ CH1
OUT
+ CH2
Recording Device
#2, 1Kohm Max
+ OUT
Impedance

Figure 9. Recorder Output Wiring

Connecting RKI Detector Heads


The detector/transmitter terminals will accept 24 - 14 AWG wire. Select wire that is rated to a
minimum of 80°C, 0.75 A, and 150V.
Perform the following procedure to connect an RKI detector head to the Beacon 200.

NOTE: If any of the Beacon 200’s channels have been set up for hydrogen-specific LEL
operation, that channel or channels can only be used with a hydrogen-specific
sensor. If you want to detect a different combustible gas or use a different detector,
the Beacon 200 must be sent back to RKI Instruments, Inc. for modification.

1. Turn off or disconnect mains power to the Beacon 200.


2. Open the Beacon 200 door and place the power switch in the off position.
3. See the detector head instruction manual for instructions on how to connect wires to the
detector head.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the
bottom of the Beacon 200 housing.

Wiring the Beacon 200 Gas Monitor • 23


CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

5. Route the wires in conduit or shielded cable from the detector head through the selected
conduit hub into the Beacon 200. See Table 5 below for wire size and distance guidelines.
Unshielded twisted pair cable in conduit or shielded twisted pair cable is recommended
for all the direct connect detector heads. For the LEL detector, pair the R & B wires and
the W & G wires. Shielded cable or wires in conduit are recommended for the 2-wire and
3-wire 4 - 20 mA transmitters.
6. Connect the wires from the detector head to the appropriate detector/transmitter terminals.
The top detector terminal strip is for channel 1 and the bottom one is for channel 2. See the
detector head instruction manual for controller terminal connections.

CAUTION: Do not route power and detector head wiring through the same conduit hub. The
power wiring may disrupt the transmission of the detector head’s to the Beacon
200.

Table 5: Wire Size Guidelines for RKI Detector Head Wiring

Number of Max Distance to Max Distance to Max Distance to


Detector Head Type Wires to Controller w/18 Controller w/16 Controller w/14
Controller Gauge Wire Gauge Wire Gauge Wire

Direct Connect LEL 4 500 ft. 1,000 ft. 2,000 ft.

Direct Connect Oxygen 2 500 ft. 1,000 ft. 2,000 ft.

Direct Connect H2S 2 500 ft. 1,000 ft. 2,000 ft.

Direct Connect CO 2 500 ft. 1,000 ft. 2,000 ft.

Direct Connect Cl2 2 500 ft. 1,000 ft. 2,000 ft.

Direct Connect SO2 2 500 ft. 1,000 ft. 2,000 ft.

2-Wire 4 - 20 mA Transmitter 2 2,500 ft. 5,000 ft. 8,000 ft.

3-Wire 4 - 20 mA Transmitter 3 2,500 ft. 5,000 ft. 8,000 ft.

Connecting User-Supplied 4 - 20 mA Transmitters


The Beacon 200 may be used with a user supplied 2-wire or 3-wire 4 - 20 mA transmitter
which runs on 24 VDC. When this is done, the following Beacon 200 parameters are normally
setup at RKI Instruments: unit of measure, item name, and full scale. For example, “PSI AIR”
with a full scale of 10 PSI.
The detector/transmitter terminals will accept 24 - 14 AWG wire. Select wire that is rated to a
minimum of 80°C, 0.75 A, and 150V.

24 • Wiring the Beacon 200 Gas Monitor


Perform the following procedure to connect a 4 - 20 mA transmitter, which you supply, to the
Beacon 200.
1. Turn off or disconnect mains power to the Beacon 200.
2. Open the Beacon 200 door and turn off the power switch.
3. See the transmitter’s instruction manual for instructions on how to connect wires to the
transmitter.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on the
bottom of the Beacon 200 housing.

CAUTION: Only use the three factory installed conduit hubs on the bottom of the housing
for wire entry into the housing. Do not drill the housing for any reason. See
“Routing Wiring Into the Beacon 200 Housing” on page 18 for more
information.

5. Route the wires from the transmitter through the selected conduit hub into the Beacon 200.
6. Connect the wires from the transmitter to the appropriate detector/transmitter terminals.
The top detector terminal strip is for channel 1 and the bottom one is for channel 2. See the
transmitter instruction manual for controller terminal connections. Figure 10 below
illustrates typical transmitter wiring connections.

CAUTION: Do not route power and transmitter wiring through the same conduit hub. The
power wiring may disrupt the transmission of the transmitter’s signal to the
Beacon 200.

3-Wire Connection 2-Wire Connection


AMP/PREAMP OXYGEN LEL AMP/PREAMP OXYGEN LEL

+ S + R W G B + S + R W G B

- DC GROUND
3-Wire 4-20 mA
FB (4-20 mA) Transmitter FB (4-20 mA) 2-Wire 4-20 mA
Transmitter
+ 24 V + 24 V

Figure 10. Generic 4 to 20 mA Transmitter Output Wiring

Wiring the Beacon 200 Gas Monitor • 25


Starting Up the Beacon 200 Gas Monitor

WARNING: Only authorized and properly trained personnel should perform any startup
procedures.

Perform the following procedure to place the Beacon 200 into normal operation.
1. Complete the mounting and wiring procedures described earlier in this chapter.
2. Complete all installation procedures described in the detector head or user supplied 4 - 20
mA transmitter instruction manual.
3. Verify that all wiring connections are correct and secure and that the Beacon 200’s power
switch is in the OFF position.
4. Turn on or connect mains power to the Beacon 200.
5. Place the Beacon 200’s power switch in the ON position. RKI INSTRUMENTS
BEACON 200 appears on the display for a few seconds, then WARMING UP appears
for each active channel. The warm-up period will last for one minute.

NOTE: To prevent unwanted alarms during warm up, the alarm circuits are not active while
the WARMING UP message is displayed.

6. Verify that the PILOT light is on. If the PILOT light is not on, see the troubleshooting
guide in Chapter 5, Maintenance.
7. Perform the start-up procedure for each detector head or user supplied 4 - 20 mA
transmitter as described in the detector head or transmitter instruction manual

26 • Starting Up the Beacon 200 Gas Monitor


Chapter 4: Operation

Overview

This chapter describes the Beacon 200 Gas Monitor in normal operation. This chapter also
describes the Beacon 200 in alarm 1, alarm 2, and fail conditions and suggests response to
these conditions.

Normal Operation

Normal operation is defined as follows:


• The start-up procedure is complete.
• The Beacon 200 is not indicating an alarm 1, alarm 2, or fail condition.
• The Beacon 200 is not running the Channel Control & Setup or Calibration Programs.
During normal operation, the Beacon 200 simultaneously displays the current gas reading,
unit of measure, and target gas for each active channel.
1 : 0 % L E L C H 4
2 : 2 0 . 9 % O x y g e n

The PILOT light is on during normal operation indicating that the Beacon 200 is receiving
incoming power.

Overview • 27
Recorder Output Operation

The output at the recorder output terminals is a 4 - 20 mA signal for each active channel that is
proportional to the detection range of the channel. A channel that is set as CHANNEL NOT
USED or CHANNEL DISABLED in the Channel Control & Setup Program (see Chapter 5)
has an output of 0 mA.
There are several special circumstances where the recorder output will behave as follows:
• When a channel is in WARMUP after the Beacon 200 is turned on, the recorder output
will be at 4 mA for all channel types except oxygen. For oxygen channels, the output will
be 17.4 mA while in WARMUP.
• If the Beacon 200 is being powered by a battery and is in low battery alarm, the recorder
output for each channel will be 0 mA.
• When a channel is added or a channel type changed, the display will indicate NEEDS
CALIBRATION for that channel when the Beacon 200 is first turned on and will continue
to indicate this until the channel is calibrated. In this situation, the recorder output will be
3.2 mA until the channel is calibrated.
• If a channel goes into a fail condition, the recorder output will be 0 mA.
• If you enter any of the instrument programs, such as the Calibration Program, the recorder
output will hold at the value it was at when you entered the program was entered until you
return to normal operation.

28 • Recorder Output Operation


Alarm Indications

This section describes the Beacon 200 in alarm 1, alarm 2, and fail conditions and suggests
response to these conditions. Table 6 below lists the alarm indications for each condition.

NOTE: The Beacon 200 includes alarm on and alarm off delay settings for each channel and
level of gas alarm. The alarm indications described in this section operate according
to the factory set delay settings. See the Configure Channel Settings Menu section of
Chapter 5 for all the factory settings.

Table 6: Visual and Audible Alarm Indications

Audible
Condition Cause Visual Indication(s)
Indication

Alarm 11 Increasing (decreasing for O2) gas • ALARM 1 light is on Pulsing tone
reading at or above the alarm 1 • Gas reading flashes and
setpoint alternates with ALARM-1
message

Alarm 21 Increasing gas reading at or above • ALARM 2 light is on Pulsing tone


the alarm 2 setpoint • Gas reading flashes and
alternates with ALARM-2
message
Fail • Disconnected or misconnected • FAIL light is on Steady tone
detector wiring • FAIL message flashes in
• Display reading below -10% place of gas reading
of full scale or lower
• Defective components
Low No AC power and DC power • FAIL light is on None
Battery2 source (primary or backup) less • SUPPLY VOLTAGE IS
than 21.5 volts. TOO LOW LOW
POWER STANDBY
message and actual voltage
of incoming DC power

* 1If the Beacon 200 is in both an alarm 1 and an alarm 2 condition, both alarm lights are on
and the display alternates between the gas reading and the ALARM-1 ALARM-2 message.
2
* If DC power is used as primary or backup power source.

Alarm Indications • 29
NOTE: You can select normally energized (NE) or normally de-energized (NDE) channel
relay settings in the Channel Control & Setup menu. The following sections
describe the default setting for the channel relays which is NDE.
Common alarm 1 and alarm 2 relays are factory-set as NDE, and the common fail
relay is factory set as NE. The common relays’ NE/NDE settings are not user-
adjustable.

Alarm 1 Condition
This section describes the audible and visual indications for an alarm 1 condition and suggests
response to an alarm 1 condition.

Alarm 1 Condition Indications


When the gas reading of an active channel reaches the alarm 1 setpoint, the Beacon 200 senses
an alarm 1 condition. The Beacon 200 alerts you to an alarm 1 condition as follows:
• The ALARM 1 light turns on.
• The gas reading in alarm 1 condition flashes and alternates with the ALARM-1 message.
• The buzzer sounds a pulsing tone.
• The common alarm 1 relay energizes.
• The applicable alarm 1 channel relay energizes.

Responding to an Alarm 1 Condition


This section suggests response to an alarm 1 condition.
1. Follow your established procedure for a low level combustible or toxic gas condition or a
decreasing oxygen content condition.
2. Oxygen alarms are self-resetting and will automatically clear when the oxygen rises above
the alarm 1 setpoint.
3. Alarms for all other gas types are latching. After the gas reading falls below the alarm 1
setpoint, press the reset switch to reset the alarm 1 circuit. Resetting the alarm 1 circuit
silences the buzzer, turns off the ALARM 1 light, resets the channel display, and de-
energizes the common and channel alarm 1 relays.

NOTE: To silence the buzzer while in an alarm 1 condition, press the reset switch.
You cannot de-energize the alarm 1 relays until the gas reading falls below (above
for oxygen) the alarm 1 setpoint.

30 • Alarm Indications
Alarm 2 Condition
This section describes the audible and visual indications for an alarm 2 condition and suggests
response to an alarm 2 condition.

Alarm 2 Condition Indications


When the gas reading of an active channel reaches the alarm 2 setpoint, the Beacon 200 senses
an alarm 2 condition. The Beacon 200 alerts you to an alarm 2 condition as follows:
• The ALARM 2 light turns on.
• The gas reading in alarm 2 condition continues to flash and alternates with the ALARM-2
messages.
• The buzzer sounds a pulsing tone.
• The common alarm 2 relay energizes.
• The applicable alarm 2 channel relay energizes.

Responding to an Alarm 2 Condition


This section suggests response to an alarm 2 condition.
1. Follow your established procedure for a high level combustible or toxic gas condition or
an increasing oxygen content condition.
2. Oxygen alarms are self-resetting and will automatically clear when the oxygen rises above
the alarm 2 setpoint.
3. Alarms for all other gas types are latching. After the gas reading falls below the alarm 2
setpoint, press the reset switch to reset the alarm circuit. Resetting the alarm circuit turns
off the ALARM 2 light, and de-energizes the common and channel alarm 2 relays.

NOTE: To silence the buzzer while in an alarm 2 condition, press the reset switch.
You cannot de-energize the alarm 2 relays until the gas reading falls below the alarm
2 setpoint.

Fail Condition
This section describes the audible and visual indications for a fail condition and suggests
response to a fail condition.

Fail Condition Indications


The Beacon 200 senses a fail condition for any of the following:
• The detector head wiring to the Beacon 200 is disconnected or incorrectly connected.
• The detector head’s detector is disconnected or incorrectly connected.
• The display reading is -10% of full scale or lower.

Alarm Indications • 31
• The Beacon 200 or detector head is malfunctioning.
When the Beacon 200 senses a fail condition, it alerts you as follows:
• The FAIL light turns on.
• The gas reading for the failing channel is replaced by the FAIL message.
• The buzzer sounds a steady tone.
• The common fail relay de-energizes.

NOTE: If you elected to use the channel’s alarm 2 relay as an individual fail relay in the
Channel Control & Setup menu, the relay de-energizes in a fail condition. See the
Channel Control & Setup section of Chapter 5 for a description of this setting.

Responding to a Fail Condition


This section suggests response to a fail condition.
1. Verify that the detector head wiring to the Beacon 200 is correctly and securely connected.
2. Verify that the detector head’s detector is correctly and securely connected.
3. See the troubleshooting guide in the detector head instruction manual.

Low Battery Condition


This section describes the audible and visual indications for a low battery condition and
suggests response to a low battery condition. This condition only applies when DC power is
used as a primary or backup power source.

NOTE: When a 24 VDC battery is used as a backup power source, the Beacon 200 keeps the
battery charged by providing a trickle charge from the AC power source. If AC
power is interrupted, the Beacon 200 will operate from the DC backup battery until
the battery voltage drops to 21.5 volts or less, or until AC power is restored.

Low Battery Condition Indications


The Beacon 200 senses a low battery condition when:
• AC power is disconnected, misconnected, or interrupted
AND
• the DC power source is 21.5 volts or less
When the Beacon 200 senses a low battery condition, it alerts you as follows:
• The FAIL light turns on.
• The top display screen displays the SUPPLY VOLTAGE IS TOO LOW, LOW
POWER STANDBY message and the actual voltage of incoming DC power.

32 • Alarm Indications
Responding to a Low Battery Condition
This section suggests response to a low battery condition.
If DC power is the primary power source:
1. For a temporary DC power source, disconnect primary DC power at the Beacon 200, then
connect a 24 VDC backup battery.
2. Determine and correct the cause of primary DC power loss.
When the DC power source rises above 22.0 volts, the Beacon 200 begins the warm up
process.
If DC power is the backup power source:
1. Replace or recharge the 24 VDC backup battery to resume backup power capability.
2. Determine and correct the cause of primary AC power loss.
When backup DC or primary AC power is restored, the Beacon 200 begins the warm up
process. When AC power is restored, the Beacon 200 charges the backup battery until it is
fully recharged. Charge time varies depending on the battery size and how much the
battery was depleted. Once the battery is fully charged, the Beacon 200 reverts to a trickle
charge to maintain the battery charge.

Viewing & Resetting Min/Max Readings

The reset switch may be used to view and reset the minimum and maximum gas readings for
the active channel(s).
1. While the Beacon 200 is in normal operation, press and hold the reset switch button for 3
seconds.
2. The display will indicate MIN / MAX Display Press RESET when done viewing . . . for
5 seconds before displaying the minimum and maximum readings for the active
channel(s). The minimum reading is on the left and the maximum is on the right side of
the display for each channel.
3. Press and release the reset switch button to exit the min/max screen. The display will
indicate To RESET Min/MAX values Press and HOLD RESET Button for 10 seconds
and then return to normal operation.
• To return to normal operation without resetting the minimum and maximum readings,
do not press the reset switch button and allow the unit to return to normal operation.
• To reset the minimum and maximum readings, before the unit returns to normal
operation press and hold the reset switch button until the display indicates Min/Max
Values Have Been Reset. Release the reset switch button.The unit will then return to
normal operation.

Viewing & Resetting Min/Max Readings • 33


Chapter 5: Channel Control and Setup Program

Overview

The Channel Control & Setup Program allows viewing of and changes to instrument setup
parameters. It is accessed using the program buttons. The Channel Control & Setup Program
includes three menus as described in Table 7.
Table 7: Channel Control & Setup Program Menus

Menu Function

Enable/Disable Channel(s) Configures channels as enabled, disabled, or not used


Configure Channel Settings Configures alarm settings, noise filter setting, and
zero suppression setting for each channel
View System Information Displays the firmware version number and the
instrument operating voltage

To enter the Channel Control & Setup Program, simultaneously press and hold the ESCAPE
and ENTER buttons for approximately 5 seconds.
The Channel Control & Setup Program menu includes a 5-minute time-out feature. If you do
not press a button for 5 minutes, the Beacon 200 automatically returns to normal operation.

NOTE: If the Beacon 200 returns to normal operation because of a program time-out, the
active channels enter a warm-up period just as they do when the unit is first turned
on.
If you are installing a new system, the channels have been setup at the factory for
the ordered detector heads. Use the Channel Control & Setup Program only if you
want to disable or enable a channel, delete a channel, or change channel settings. If a
channel is being added or a channel is being changed from one type to another,
contact RKI Instruments, Inc. for additional documentation required to define the
channel type.

34 • Overview
Enable/Disable Channel(s) Menu

1. From normal operation, simultaneously press and hold the ESCAPE and ENTER buttons
for approximately 5 seconds to enter the Channel Control & Setup Program. Release the
buttons when the Control & Setup Program Proceed? [YES] or [NO] message appears
on the display screen.
2. Press the UP/YES button to continue.
3. Press the UP/YES or DOWN/NO button until the 1) Enable/Disable Channel(s) message
appears on the display screen, then press the ENTER button.
4. Use the UP/YES and DOWN/NO buttons to select the channel you want to enable or
disable, then press the ENTER button.
5. Press the DOWN/NO button. The CHANNEL USAGE setting displays on the display
screen.
6. Use the UP/YES and DOWN/NO buttons to display the setting you want, then press the
ENTER button to select the setting. The table below describes the three available settings.
Table 8: Beacon 200 Channel Usage Settings

Setting Description

CHANNEL The Beacon 200 displays gas readings and initiates gas and channel failure
ENABLED alarms when appropriate.
Use this setting for normal operation when the channel has a detector head
wired to it.
CHANNEL The Beacon 200 displays DISABLED for the channel and the channel’s
DISABLED alarm circuit is not active.
Use this setting when the channel has a detector head wired to it, but gas
readings and alarms are not required for the channel (for example if the
detector head requires maintenance or is malfunctioning).
CHANNEL The Beacon 200 leaves the channel blank on the display screen.
NOT USED Use this setting when the channel does not have a detector head wired to it.

7. Press the ESCAPE button, then press the DOWN/NO button to return to normal operation.

Enable/Disable Channel(s) Menu • 35


Configure Channel Settings Menu

This section describes how to view and change channel parameters for the installed gas
channels.
1. Simultaneously press and hold the ESCAPE and ENTER buttons for approximately 5
seconds to enter the Channel Control & Setup Program. Release the buttons when the
Control & Setup Program Proceed? [YES] or [NO] message appears on the display
screen.
2. Press the UP/YES button to continue.
3. Press the UP/YES or DOWN/NO button until the 2) Configure Channel Setting(s)
message appears on the display screen, then press the ENTER button.
4. Use the UP/YES and DOWN/NO buttons to select the channel for which you want to set
parameters, then press the ENTER button.
5. Press the UP/YES button until the parameter you want to set appears on the display
screen. The screen will display the current setting and ask if it is OK.
Table 9 lists the parameters you can set for a channel. Table 9 also lists the factory set
value for each parameter.

NOTE: Use the ESCAPE button to go back to a previously displayed parameter.

6. If the setting is not OK and you want to change the it, press the DOWN/NO button. The
parameter is now adjustable.
7. Use the UP/YES or DOWN/NO button to update the parameter, then press the ENTER
button to continue.
8. Repeat steps 5 through 7 to set any other channel parameters.
9. Press the UP/YES button until the following message appears on the display screen.

C o n f i g u r a t i o n f o r
C H A N NE L 1
h a s b e e n c o m p l e t e d
S a v e S e t t i n g s ? [ Y / N ]

36 • Configure Channel Settings Menu


10.Press the UP/YES button to save the configuration. The screen will then return to the
Channel Control & Setup menu.
11.Press ESCAPE to return to the screen which asks Control & Setup Program Proceed?
[YES] or [NO].
12.{Press the DOWN/NO button to return to normal operation.

Table 9: Channel Setting Parameters

Parameter
Description
(Factory-Set Value)

ALARM-1 Level The gas reading at which the Beacon 200 initiates an alarm 1
See the Beacon 200 Detector Head condition for this channel.
Specification Sheet for the
detector head installed on this
channel
ALARM-1 ON DELAY The amount of time the Beacon 200 delays activation of the alarm
(1 sec) 1 circuit once an alarm 1 condition is initiated.
ALARM-1 OFF DELAY The amount of time the Beacon 200 delays turning off the alarm 1
(0 sec) circuit once an alarm 1 condition passes.
ALARM-1 (activation) Indicates if the alarm 1 circuit is activated by gas readings
(DECREASING for oxygen INCREASING or DECREASING to the ALARM-1 Level.
channels, INCREASING for all
other channel types)
ALARM-1 Relay (action) If set as NORMALLY DE-ENERGIZED, the channel’s alarm 1
(NORMALLY DE-ENERGIZED) relay is de-energized in normal operation and energizes when an
alarm 1 condition is initiated.
If set as NORMALLY ENERGIZED, the channel’s alarm 1
relay is energized in normal operation and de-energizes when an
alarm 1 condition is initiated.
ALARM-1 Relay (reset) If set as LATCHING, you must press the RESET button to reset
(SELF RESETTING for oxygen, the alarm 1 circuit after the alarm 1 condition passes.
LATCHING for all other channel If set as SELF RESETTING, the Beacon 200 automatically
types) resets the alarm 1 circuit after the alarm 1 condition passes.
ALARM-2 Relay (used for) If set as ALARM-2 Condition, the channel’s alarm 2 relay
(ALARM-2 Condition) activates when an alarm 2 condition is initiated for the channel.
If set as FAIL Condition, the channel’s alarm 2 relay activates
when a fail condition is initiated for the channel.
ALARM-2 Level The gas reading at which the Beacon 200 initiates an alarm 2
See the Beacon 200 Detector Head condition for this channel.
Specification Sheet for the
detector head installed on this
channel

Configure Channel Settings Menu • 37


Table 9: Channel Setting Parameters (Continued)

Parameter
Description
(Factory-Set Value)

ALARM-2 ON DELAY The amount of time the Beacon 200 delays activation of the alarm
(2 min for oxygen, 1 sec for all 2 circuit once an alarm 2 condition is initiated.
other channel types)
ALARM-2 OFF DELAY The amount of time the Beacon 200 delays turning off the alarm 2
(0 sec) circuit once an alarm 2 condition passes.
ALARM-2 (activation) Indicates if the alarm 2 circuit is activated by gas readings
(INCREASING) INCREASING or DECREASING to the ALARM-2 Level.
ALARM-2 Relay (action) If set as NORMALLY DE-ENERGIZED, the channel’s alarm 2
(NORMALLY DE-ENERGIZED) relay is de-energized in normal operation and energizes when an
alarm 2 condition is initiated.
If set as NORMALLY ENERGIZED, the channel’s alarm 2
relay is energized in normal operation and de-energizes when an
alarm 2 condition is initiated.
ALARM-2 Relay (reset) If set as LATCHING, you must press the RESET button to reset
(SELF RESETTING for oxygen, the alarm 2 circuit after the alarm 2 condition passes.
LATCHING for all other channel If set as SELF RESETTING, the Beacon 200 automatically
types) resets the alarm 2 circuit after the alarm 2 condition passes.
NOISE FILTER The noise filter feature helps “smooth out” jumpy or noisy signals
(3) from the detector head. You can set the noise filter from 1 to 8.
A setting of 8 produces the greatest amount of smoothing but also
responds slowest to changes in the response reading.
A setting of 1 responds fastest to changes in the response reading
but produces the least amount of smoothing.
ZERO SUPPRESSION The zero suppression feature helps prevent “jumpy” readings near
(0.0% of the detection range in the fresh air reading.
terms of the detection units for For example, if the zero suppression setting on a %LEL channel
carbon dioxide channel types, is 2.0% and the full scale is 100% LEL, the Beacon 200 will
0.5% oxygen for oxygen channels, display a reading of 0% LEL for gas readings from -2% LEL to
2.0% of the detection range in 2% LEL.
terms of the detection units for all
other channel types)

View System Information Menu


The View System Information Menu consists of only one display screen which indicates the
version number of the firmware that is running the instrument and the system voltage. The
system voltage is the voltage that is directly running the instrument’s circuitry. When the unit
is running from AC power, this voltage is normally 24.0 volts ± 0.2 volts. When the unit is
running from DC power, this voltage reading can be used as an indication of the DC voltage
powering the unit.

38 • Configure Channel Settings Menu


Chapter 6: Input Setup Program

Overview

This chapter describes how to use the Input Setup Program to add a channel or change the
channel type of an installed channel on the Beacon 200. The Input Setup Program allows you
to define the type of detector head, the units and gas type, and the full scale for that channel.
To enter the Input Setup Program, the Beacon 200 must first be off. While the Beacon 200 is
off, press and hold the ENTER button, then turn on the Beacon 200 with the ON/OFF switch.
The Input Setup Program menu includes a 5-minute time-out feature. If you do not press a
button for 5 minutes, the Beacon 200 automatically begins normal operation.

NOTE: If the Beacon 200 enters normal operation because of a program time-out, the active
channels enter a warm-up period just as they do when the unit is first turned on.

Setting Up a New Channel or


Changing an Existing Channel

1. While the Beacon 200 is off, press and hold the ENTER button, then turn on the Beacon
200 with the ON/OFF switch.
2. The Beacon 200 will beep repeatedly while you are holding down the ENTER button and
then the screen will show INPUT SETUP PROGRAM on the top line.
3. Press the ENTER button to continue.
If you press the ESCAPE button, the unit will start-up and enter its warm-up period.
4. Press the UP/YES or DOWN/NO button until the channel you wish to add or change
appears on the display screen, then press the ENTER button. The detector head type
screen appears.

Overview • 39
5. The display screen indicates what type of detector head is currently selected for that
channel and asks if it is OK. Table 1 below briefly describes each type.
Table 10: Beacon 200 Detector Head Types

Detector Head Type Description

4 - 20 mA Amp [+S(-)] A 4 - 20 mA detector head is connected to the Beacon


200 with 2 or 3 wires, depending on the detector head
model, using the AMP/PREAMP +, S, and - terminals
on a detector head terminal strip. All calibration
adjustments are made at the detector head.
PreAmp [+S] A PreAmp detector head is connected to the Beacon
200 with 2 wires using the AMP/PREAMP + and S
terminals on a detector head terminal strip. All
calibration adjustments are made at the Beacon 200.
O2 Direct [+-] An O2 Direct detector head is an oxygen detector head
in which the oxygen detector is wired to the Beacon
200 with 2 wires using the OXYGEN + and - terminals
on a detector head terminal strip. All calibration
adjustments are made at the Beacon 200.
LEL Direct [RWGB] An LEL Direct detector head is a combustible gas
detector head in which the combustible gas detector is
wired to the Beacon 200 with 4 wires using the LEL R,
W, G, and B terminals on a detector head terminal strip.
All calibration adjustments are made at the Beacon
200.

If the detector head type is not correct, press the DOWN/NO button and continue with step
6.
If the detector head type is correct, press the UP/YES button to proceed to the units and
gas type screen and skip to step 7.

NOTE: See the detector head operator’s manual and the Beacon 200 Detector Head
Specification sheet for the detector head to determine the detector head type.

6. Use the UP/YES and DOWN/NO buttons to scroll through the list of detector head types
until the correct one is displayed. Press the ENTER button to accept the type. The units
and gas type screen appears.
7. If the units and gas type are not correct, press the DOWN/NO button and continue with
step 8.
If the units and gas type are correct, press the UP/YES button to proceed to the full scale
screen and skip to step 9.

40 • Setting Up a New Channel or Changing an Existing Channel


8. Use the UP/YES and DOWN/NO buttons to scroll through a list of units and gas type
choices.

NOTE: See the Beacon 200 Detector Head Specification Sheet for the detector head to
determine the correct units and gas type.

When the desired units and gas type appears, press the ENTER button to accept the
choice. The full scale screen appears.
One of the choices is User Will Specify. If the desired units and gas type setting is not in
the list, this setting will allow you to enter a 10 character units and gas type setting. With
this choice displayed, press the ENTER button and a screen will appear which allows you
to input the characters. Use the UP/YES and DOWN/NO buttons to select a character and
then press ENTER to accept it and continue with the next character. When all characters
have been entered, the full scale screen will appear.
9. If the full scale setting is not correct, press the DOWN/NO button and continue with step
10.
If the full scale setting is correct, press the UP/YES button to proceed to the save screen
and skip to step 11.
10.Use the UP/YES and DOWN/NO buttons to scroll through a list of full scale settings.

NOTE: See the Beacon 200 Detector Head Specification Sheet for the detector head to
determine the correct full scale setting.

When the desired full scale setting appears, press the ENTER button to accept the setting.
The save screen appears.
One of the choices is User Will Specify. If the desired full scale setting is not in the list,
this setting will allow you to enter a full scale setting. With this choice displayed, press the
ENTER button and a screen will appear which prompts you to choose how many decimal
places you want in the full scale setting. You can select up to 3 decimal places, then press
the ENTER button to proceed to the select full scale screen. Use the UP/YES and DOWN/
NO buttons to select a full scale setting, then press ENTER to accept it. The save screen
appears.
11.To save the settings, press the UP/YES button and the settings will be saved. The display
will return to the first screen of the Input Setup Program.
If you do not want to save the settings, press the DOWN/NO button. The display will
return to the first screen of the Input Setup Program.

Setting Up a New Channel or Changing an Existing Channel • 41


12.Press the ENTER button to perform additional setup or the ESCAPE button to exit the
Input Setup Program and begin the Beacon 200’s warm-up sequence.

NOTE: Once the Beacon 200 channels are configured correctly, see the detector head
operator’s manual(s) for a complete description of detector head installation
procedures.

42 • Setting Up a New Channel or Changing an Existing Channel


Chapter 7: Maintenance

WARNING: Only authorized and properly trained personnel should perform any
maintenance procedures.

Overview

This chapter describes use of the Calibration Program and corrective maintenance procedures
for the Beacon 200. It includes a troubleshooting guide for problems you may encounter with
the Beacon 200. Procedures to replace components of the Beacon 200 are at the end of this
chapter.

Calibration Program

The Calibration Program is used to calibrate the Beacon 200’s active channel(s). Since the
Beacon 200 can support both direct connect (internal amp) and 4-20 mA transmitter (remote
amp) detector heads, when calibrating the active channel(s) there are three possible detector
head combinations:
• Direct connect detector head(s) only.
If one or two direct connect detector heads are active, then all calibration adjustments are
made at the Beacon 200 after calibration gas is applied at the detector(s).
• 4-20 mA transmitter detector head(s) only.
If one or two 4-20 mA transmitter detector heads are active, then all calibration
adjustments are made at the detector head(s) after calibration gas is applied to the
detector(s).
• A direct connect and a 4-20 mA transmitter detector head.
If one direct connect and one 4-20 mA transmitter detector head are installed, then
calibration adjustments must be made at the Beacon 200 for the direct connect detector
head after applying gas to its detector, and at the detector head for the 4-20 mA transmitter
detector head.

Overview • 43
Calibration Program Flow
Figures 11, 12, and 13 below illustrate the general flow of the Calibration Program for the
three possible detector head combinations. See the next section, Entering the Calibration
Program, for instructions to enter the Calibration Program. In general, the program screens
provide instructions to guide you through the program. At any point in the calibration
program, the ESCAPE button may be used to either return to the previous screen or abort a
process.

ESCAPE

Calibration Calibration
Normal ENTER Calibrate
Program Enter Timeout
Operation Channel 1 Y/N?
ENTER/ESCAPE Selection

Calibrate Air Adjust


Air Adjust Span
Channel 2 Y/N? Channel 2
Channel 1 Channel 1 Y/N?
(If Installed) (If Installed)

Span Select Cal. Gas Select Cal. Gas Apply Gas to


Channel 2 Y/N? Concentration Concentration Ch. 1 & Ch. 2
(If Installed) for Channel 1 for Channel 2 Detectors

Press Enter to
Adjust Span

Figure 11. Direct Connect Detector Head(s) Only

44 • Calibration Program
ESCAPE

Calibration Calibration
Normal ENTER Calibrate
Program Enter Timeout
Operation Channel 1 Y/N?
ENTER/ESCAPE Selection

Calibrate Calibrate at
Press Enter
Channel 2 Y/N? Detector
When Done
(If Installed) Head(s)

Figure 12. 4 - 20 mA Transmitter Detector Head(s) Only

Calibration Program • 45
Normal
Operation

ESCAPE Calibration
Program Enter
ENTER/ESCAPE

ENTER

Calibration
Timeout
Selection

No No
Calibrate Calibrate
Channel 1 Y/N? Channel 2 Y/N?

Yes Yes

Transmitter Direct Connect Transmitter Direct Connect

Calibrate Calibrate
Channel 1 at Air Adjust Air Adjust
Channel 1 Channel 2 at Channel 2
Detector Head Detector Head

Press Enter Select Cal. Gas Press Enter Select Cal. Gas
When Done Concentration When Done Concentration
for Channel 1 for Channel 2

Apply Gas to Apply Gas to


Ch. 1 Detector Ch. 2 Detector

Press Enter to Press Enter to


Adjust Span Adjust Span

Figure 13. 4 - 20 mA & Direct Connect Detector Head(s)

46 • Calibration Program
Entering the Calibration Program
1. Assemble the calibration kit(s). See the instruction manual for each detector head for
procedures specific to that detector head.
2. Simultaneously press and hold the ENTER and UP/YES buttons for approximately 5
seconds to enter the Calibration Program. Release the buttons when the CALIBRATION
PROGRAM... message displays and asks if you want to continue or return to normal
operation.

NOTE: While in the Calibration Program, the alarm status of the Beacon 200 will be locked
in the state it was in when the Calibration Program was entered.

3. Press the ENTER button to continue and display the Calibration Time-out setting.
The Beacon 200 will remain in the Calibration Program for the amount of time indicated
by the Calibration Time-out setting or until you exit the program. If necessary, adjust the
setting using the UP/YES and DOWN/NO buttons. Make sure you have allotted enough
time to perform the calibration procedure. Consider the type of detector head(s) installed
and the distance from the Beacon 200 when determining the time required.

NOTE: If you do not accept a Time-out setting and stay at this screen or press ESCAPE and
do not press control buttons again, then the Beacon 200 will exit the Calibration
Program and begin its warm-up period 5 minutes after the last button was pushed.
This warm-up period functions as if the unit were just powered up.

4. Press the ENTER button to accept the Time-out setting, start the Time-out period and
continue in the Calibration Program.

NOTE: The following describes calibration with a typical channel allocation which may be
different from yours. This example allocation has two active channels with both
types of detector heads installed. Channel 1 is a 4-20 mA detector head and channel
2 is a direct connect detector head. See the detector head instruction manuals for
calibration information and procedures specific to the detector heads.

Calibrating a 4-20 mA Detector Head


1. The display asks if you want to calibrate channel 1 (in this example a 4-20 mA detector
head). Press the UP/YES button to continue with calibrating channel 1.
If you press the DOWN/NO button, the display will skip channel 1 and ask if you want to
calibrate channel 2.
2. If you pressed the UP/YES button, since channel 1 is a 4 - 20 mA detector head in this
example, the unit will display the following message for a few seconds before continuing:
Reminder. Calibration Must Be Done at the Detector Head. Then the display will
alternate between the current gas reading for channel 1 and the message CALIBRATE

Calibration Program • 47
AT HEAD above the time remaining in the calibration Time-out.
3. Verify that the detector head is in a fresh-air environment. (If necessary, use a zero-
emission air cylinder, also known as zero air, to introduce a fresh-air sample when
adjusting the zero reading below.)
4. Adjust the detector head’s zero reading (fresh air reading for oxygen). See the detector
head’s instruction manual for instructions on how to adjust the zero reading (fresh air
reading for oxygen).
5. Apply calibration gas to the detector head’s detector and adjust the detector head’s span
reading (zero reading for oxygen). See the detector head’s instruction manual for
instructions on how to adjust the span reading (zero reading for oxygen).
6. Press the ENTER button to indicate that you are done with calibrating channel 1 and are
ready to continue.

Calibrating a Direct Connect Detector Head


1. The display asks if you want to calibrate channel 2 (in this example a direct connect
detector head). Press the UP/YES button to continue with calibrating channel 2.
If you press the DOWN/NO button, the display will skip channel 2 and return to the first
calibration program screen which asks if you want to continue or escape from the
program.
2. If you pressed the UP/YES button, since channel 2 is a direct connect detector head in this
example, the unit will display the following message for a few seconds before continuing:
Expose Detector(s) To Fresh Air. . . When Done Press ENTER. Then the display will
alternate between the gas reading and the message FRESH AIR ENTER to
ACCEPT ESCAPE to ABORT above the time remaining in the calibration Time-out.
3. If the detector is in a fresh air environment, press the ENTER button. The unit will adjust
the zero reading and display the message Fresh Air Adjust Passed for: Channel 2
before continuing.
If you suspect the detector area is not a fresh air environment, apply zero air to the detector
before pressing the ENTER button. See the detector head instruction manual for
instructions to apply zero air to the detector. The Beacon 200 will freeze the display
reading for channel 2 at the lowest level reached while applying zero air (highest level for
an oxygen channel). After applying zero air for the required amount of time, usually two
minutes, remove the zero air source from the detector. Press the ENTER button at the
Beacon 200. The unit will adjust the zero reading and display the message Fresh Air
Adjust Passed for: Channel 2 before continuing.
4. The display asks if you want to perform a span (zero for an oxygen channel) operation on
channel 2 by applying gas. Press the UP/YES button to continue with adjusting the span
on channel 2.
If you press the DOWN/NO button, the unit will return to the first calibration program
screen which asks if you want to continue or escape from the Calibration Program.

48 • Calibration Program
5. If you pressed the UP/YES button, the display will prompt you for the span gas (zero gas
for an oxygen channel) concentration that will be used. Adjust the displayed concentration
up or down as needed using the UP/YES and DOWN/NO buttons so that it matches the
concentration in the calibration cylinder.
6. Press the ENTER button to accept the calibration gas concentration and continue. The unit
will display the message Expose Detector(s) To Gas. . . When Done Press ENTER
for a few seconds. It will then alternate between the gas reading for channel 2 and the
message APPLYING GAS ENTER to ACCEPT ESCAPE to ABORT above the
time remaining in the calibration time-out.
7. Apply the calibration gas to the detector. See the detector head instruction manual for
instructions on how to apply gas to the detector.
When the calibration gas is applied, the Beacon 200 will freeze the display gas reading at
the highest level reached while the gas was applied (lowest for an oxygen channel).
8. After applying calibration gas for the required amount of time, usually two minutes,
remove the gas from the detector.
9. Press the ENTER button at the Beacon 200 to proceed with the calibration adjustment.
If the Beacon 200 is able to successfully make the calibration adjustment, it will display
the message Cal Passed for: Channel 2. It will then return to the first calibration
program screen which asks if you want to continue or escape from the program. Press
ESCAPE to return to normal operation.
If the Beacon 200 is not able to make the calibration adjustment, it will display the
message Not Enough Response! CAL Failed for: Channel 2, then return to the first
calibration program screen which asks if you want to continue or escape from the
program. Press ENTER to start the calibration process again or ESCAPE to return to
normal operation.

NOTE: Be sure to allow enough time between the application of calibration gas and
returning to normal operation for the gas reading to return to normal levels to avoid
false alarms. The alarm circuits are active when the Beacon 200 returns to normal
operation.

Calibration Program • 49
Replacing the Fuses

The Beacon 200 includes three fuses: one DC and two AC fuses. The replacement DC fuse
must be rated to 6 A, 250 V, and 1/4 x 1 1/4”, fast acting. The replacement AC fuses must be
rated to 3 A, 250 V, 1/4 x 1 1/4”, fast acting.
To replace other components of the Beacon 200, contact RKI Instruments, Inc. for further
information.
1. Turn off or disconnect mains power to the Beacon 200.
2. Open the housing door of the Beacon 200, then place the power switch in the OFF
position.
3. Locate the vertical fuse holders on either side of the power switch.
The DC fuse which is labelled F1 on the PCB silk-screen is to the left of the power switch.
The AC fuses are labelled F2 and F3 and are to the right of the power switch.
4. Use a flat-blade screwdriver to rotate the applicable fuse holder 1/4 turn counterclockwise.
The fuse holder releases from the socket.
5. Remove the fuse holder from the socket, then remove the fuse from the fuse holder.

CAUTION: Verify that the replacement fuse is the same type and rating as the fuse you are
replacing.

6. Install the appropriate replacement fuse in the fuse holder, then place the fuse holder in the
socket.
7. Push the fuse holder into the socket, then turn the holder 1/4 turn clockwise to secure it in
the socket.
8. Turn on or connect mains power to the Beacon 200.
9. Place the Beacon 200’s power switch in the ON position, then verify that the PILOT light
is on.
10.Close and secure the housing door.

Preventive Maintenance

Preventive maintenance of the Beacon 200 consists of daily, monthly, and quarterly
procedures to ensure that the detector head is operating properly. See the detector head
instruction manual(s) for preventive maintenance procedures that apply to the detector head(s)
supplied with your Beacon 200.

50 • Replacing the Fuses


Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended action for
problems you may encounter with the Beacon 200.

NOTE: This troubleshooting guide describes controller problems only. See the detector head
instruction manual(s) for troubleshooting procedures that apply to the detector
head(s) supplied with your Beacon 200.

Table 11: Troubleshooting the Beacon 200

Condition Symptom(s) Probable Causes Recommended Action

No Power • The PILOT light • The power wiring is 1. Verify that the wiring to the power
is off. disconnected or source is correct and secure.
• The display misconnected. 2. At the Beacon 200, verify that the
screens are blank. • One or both AC wiring to the AC terminals is
fuses or the DC fuse correct and secure.
is blown. 3. Check the continuity of the
• The display cable is applicable (AC or DC) fuse.
disconnected or 4. Verify that the display cable is
misconnected. connected.
The display (ribbon) cable plugs
into connectors on the top edge of
the display board (labeled J1) and
on the top edge of the main circuit
board (also labelled J1).
5. If the power difficulties continue,
contact RKI for further
instruction.
Frequent or • The Beacon 200 • The Beacon 200 is 1. Verify that the detector head
Suspect alerts you to experiencing false wiring is properly shielded (see
Alarms frequent or readings due to RFI the detector head instruction
suspect alarms. or EMI. manual).
• The detector • The detector head 2. Verify that the detector head
head’s fresh air wiring is wiring is correct and secure.
readings remain disconnected, 3. Verify that power and detector
on zero misconnected, or head wiring is routed through
(20.9 for oxygen). intermittent. separate conduit hubs on the
bottom of the Beacon 200
housing.
4. Increase the alarm on delay
setting in the Channel Control &
Setup Program.
5. If the frequent or suspect alarm
difficulties continue, contact RKI
for further instruction.

Troubleshooting • 51
Table 11: Troubleshooting the Beacon 200 (Continued)

Condition Symptom(s) Probable Causes Recommended Action

Flickering • The display • The Beacon 200 is 1. Verify that the detector head
Display readings flicker experiencing false wiring is properly shielded (see
often. readings due to RFI the detector head instruction
or EMI. manual).
• The noise filter 2. Verify that the detector head
setting is too low. wiring is correct and secure.
• The zero 3. Verify that power and detector
suppression setting head wiring is routed through
is too low. separate conduit hubs on the
• The display screen bottom of the Beacon 200
is malfunctioning. housing.
4. Increase the noise filter setting in
the Channel Control & Setup
Program.
5. Increase the zero suppression
setting in the Channel Control &
Setup Program
6. If the display difficulties continue,
contact RKI for further
instruction.
Buzzer not • The buzzer does • The buzzer is 1. Verify that the wiring to the BUZ-
Working not sound an disconnected or and BUZ+ terminals of the
audible alarm misconnected. controller terminal strip is correct
during alarm • The buzzer is and secure.
conditions. malfunctioning. 2. If the buzzer difficulties continue,
• The buzzer contact RKI for further
sounds weak or instruction.
broken.
Reset • The buzzer does • The reset switch is 1. Verify that the wiring to the
Switch not not silence when disconnected or RESET (2) terminals of the
Working you press the reset misconnected. controller terminal strip is correct
switch. • The reset switch is and secure.
• The applicable malfunctioning. 2. If the reset switch difficulties
alarm circuit does continue, contact RKI for further
not reset when instruction.
you press the reset
switch after an
alarm condition
passes.

52 • Troubleshooting
Parts List

The table below lists the part numbers and descriptions for replacement parts and accessories
offered for the Beacon 200 Gas Monitor.
Table 12: Parts List, Beacon 200 Gas Monitor

Part No. Description

18-0107RK Conduit Hub (3/4 in.)


43-0440RK Reset Switch
43-4165RK AC Fuse, 1/4” x 1 1/4”, fast acting, 3A, 250 V
43-4170RK DC Fuse, 1/4” x 1 1/4”, fast acting, 6A, 250 V
52-1016RK Buzzer
57-0050RK Main PCB, standard, for use with standard LEL sensors
57-0050-H2-01 Main PCB, for hydrogen-specific LEL Channel 1
57-0050-H2-02 Main PCB, for hydrogen-specific LEL Channel 1 and Channel 2
71-0059RK Beacon 200 Gas Monitor Operator’s Manual (this document)

Parts List • 53