Apollo
Apollo
Apollo
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Working with these instructions for use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Use of terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Screen layouts and illustrations of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety information definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Definition of target groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Abbreviations and symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
General safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Product-specific safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Indications and contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Environment of use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Additional functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Accessory weight limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
System Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Machine Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Machine Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Gas supply block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Interface panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Vaporizers (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Vaporizer exclusion systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
O2 flush. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Auxiliary oxygen flow meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Writing table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
User Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Part Number: 9053586, 3rd edition
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
User controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
LED indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Screen colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Menu structure overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
System Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Preparation before first use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Charging the battery for emergency operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Installing the breathing system and flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Filling and installing the absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Connecting the gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Connecting the scavenger system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Connecting the endotracheal aspiration system (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Installing vaporizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Pre-use Checkout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Checking the Workstation according to the Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
System compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Leak tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Emergency start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Manual/Spontaneous ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Volume-Controlled Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Volume Mode AutoFlow - Volume AF (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Pressure-Controlled Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Pressure Support Ventilation (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Continuous Positive Airway Pressure CPAP - in Pressure Support Mode (Optional) . . . . . . . . . . . . . . . . . . . . . 145
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Standard screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Displayed parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Loops (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Mini trends (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Datalog. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Screen timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Data screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Trend screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
Monitoring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Alarm priorities and alarm signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Alarm displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Suppressing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Limit-based alarms activated in respective ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Enabling/disabling alarms globally during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Displaying and setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Configuring the default settings in Standby. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
System settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Interfaces datalog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Ventilator and gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214
System information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218
Configuration during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
Gas failure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260
Ventilator failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 262
Fresh-gas delivery failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263
Ventilator and fresh-gas delivery failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264
Gas measurement failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264
Display failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
User interface failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
System failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
Alarm - Cause - Remedy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283
Schematic diagram of alarm tones. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
EMC declaration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303
Device combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 306
Connections to IT-networks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 306
Relevant standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 308
Introduction
Contents
Working with these instructions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Use of terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Screen layouts and illustrations of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety information definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Definition of target groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Abbreviations and symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
General safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Strictly follow these Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Not for use in areas of explosion hazard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Safe connection with other electrical equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Information on electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Patient safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Patient monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Sterile accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installing accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Storing the instructions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Product-specific safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Part Number: 9053586, 3rd edition
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
List of abbreviations used in the software and on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
List of general abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
combines text and illustrations. The information is Configuring the default settings in
presented as sequential steps of action, giving the Standby
Default settings describe settings which the Figure 123. Location of Standby Config Key
user hands-on experience in learning how to use the anesthesia machine starts with when it is switched
on.
Apollo inhalation anesthesia machine. The default settings for ventilation, fresh-gas
delivery, and monitoring can be activated while in
Standby by pressing the >Restore Default
used both to refer to relevant details in the The four-digit password is assigned at the
factory.
illustrations and to specify the sequence of actions If desired, the function can be disabled by
DrägerService or a new password set.
where several actions are described. 3. Select and confirm the figures successively from
the line displayed using the rotary knob. The
password is represented by asterisks (****)
below the line of numbers
Right-Hand Column - the Illustrations The menu Standby Conf. for selecting the default
values is displayed when the password has been
e.g.:
>PEEP<
Screen messages and screen options are printed in
bold, e.g.:
Default Alarm Limits
Experts Maintenance
WARNING and CAUTION statements specific to Dräger has tested only the compatibility of
accessories listed in the current list of
subsystems or particular features of the medical device
accessories. If other, incompatible accessories
appear in the respective sections of these instructions
are used, there is a risk of patient injury due to
for use or in the Instructions for Use of another product medical device failure.
being used with this medical device.
Dräger recommends that the medical device is
Strictly follow these Instructions for Use only used together with accessories listed in the
current list of accessories.
WARNING!
Note: Strictly observe the instructions for use of all
Risk of incorrect operation and of incorrect use accessories such as:
Part Number: 9053586, 3rd edition
WARNING!
WARNING!
Risk of electric shock or device malfunction
Risk of fire
Any connected devices or device combinations
The flow sensor can ignite medications or other
not complying with the requirements mentioned substances based on highly flammable
in these instructions for use can compromise the substances.
correct functioning of the medical device and lead – Do not nebulize medications or other
to an electric shock. substances that are easily flammable or spray
– Electrical connections to equipment not listed them into the device.
in these instructions for use or these assembly – Do not use substances containing alcohol.
instructions must only be made when approved – Do not allow flammable or explosive
by each respective manufacturer. substances to enter the breathing system or the
– Before operating the medical device, strictly breathing circuit.
comply with the instructions for use of all – Do not use cyclopropane or ether.
connected devices or device combinations.
WARNING!
WARNING! Risk of device failure and/or danger to patient
Risk of explosion, fire Magnetic fields may negatively influence the
If an oxygen leak is suspected within or near the proper function of the medical device, thus
anesthesia machine, do not initiate operation. endangering the patient or user.
Disconnect all oxygen supplies and contact a The medical device must not be used in the
trained service technician. vicinity of magnetic resonance imagers (MRI,
NMR, NMI).
WARNING!
Patient monitoring
Risk of electric shock
The user of the medical device is responsible for
Connecting devices to the auxiliary choosing a suitable patient monitoring system that
outlets of the anesthesia machine can provides appropriate information on medical device
Part Number: 9053586, 3rd edition
Sterile accessories
CAUTION!
CAUTION!
Risk of patient injury
Risk of medical device failure and of patient injury
An incorrect diagnosis or misinterpretation of
Do not use sterile-packaged accessories if the measured values, or other parameters, may endanger
packaging has been opened, is damaged, or if there the patient.
are other signs of non-sterility.
Do not base therapy decisions on individual
Disposable articles must not be reprocessed and measured values or monitoring parameters only.
resterilized.
Installing accessories
WARNING!
Install accessories to the basic device in accordance If ventilation of the patient is no longer ensured
with the Instructions for Use of the basic device. Make due to an obvious fault in the equipment, the
patient must immediately be ventilated with a
sure that there is a safe connection to the basic
manual resuscitator.
device system.
Always keep a manual resucitator at hand.
Strictly observe Assembly Instructions and Instructions
for Use.
WARNING!
Storing the instructions for use
Risk of burns
CAUTION! Conductive breathing hoses or face masks may
cause burns during HF surgery.
Risk of incorrect use
Do not use these types of hoses and masks in
Instructions for use must be kept accessible to the combination with HF surgery.
user.
Training
CAUTION!
Training for users is available from the Dräger
Risk of malfunctions
Unapproved modifications to the medical device
can cause malfunctions.
No modifications must be made to this medical
device without the permission of Dräger. Dräger
does not accept responsibility for modifications
to the device made without the permission of
Dräger.
Functional safety
CAUTION! The essential performance consists of:
Risk of crushing – Supplying the anesthesia workstation with O2
Movable parts and attached parts can lead to crushing If the O2 supply (central supply or gas cylinder)
injuries. Pay special attention to edges, movable fails, an alarm is issued.
parts, and corners when working with the following – Supply of the patient with adequately oxygenated
parts: breathing gas
– Drawers
If the breathing gas contains insufficient levels of
– Ventilator module O2, an alarm is issued.
– Doors – Patients are not supplied with excessively high
– Writing table anesthetic gas concentrations
– Swivel arms for mounted devices If excessively high anesthetic gas concentrations
– Accessories such as gas cylinders, vaporizers, are delivered, an alarm is issued.
CLIC absorbers, and CLIC adapters
– Monitoring the airway pressure and the expiratory
minute volume
Alarms are issued depending on the set alarm
CAUTION! limits.
Risk of device failure
Compressed gas supply (pipeline supply or cylinder):
To avoid damaging the device(s) attached to a gas
supply, use only medical gases. Pay particular
attention to national and international standards
regulating the use of medical gases.
WARNING!
Risk due to barely audible alarms
The user must remain within the hearing range of
the acoustic alarm signal. This permits quick
Part Number: 9053586, 3rd edition
WARNING!
Risk due to a noisy environment
When operating in a noisy environment, the
volume of the alarm signals must be adjusted to
suit.
– O2 Uptake
– PEEP, patient compliance CPAT
Additional functions
MEDIBUS/MEDIBUS.X Protocol
MEDIBUS and MEDIBUS.X are software protocols for
use in transferring data between Apollo and an
external medical or non-medical device (e.g.
hemodynamic monitors, data management systems, or
a Windows-based computer) via the RS-232 interface
see:
– 9037426, 6th edition or higher
or
– 9052608, third edition or higher.
WARNING!
Risk of patient injury
Data transferred via MEDIBUS/MEDIBUS.X
interfaces are for information only and are not
intended as a basis for diagnosis or therapy
decisions. The data accessible via this interface
are not intended for a decentralized alarm system
in accordance with IEC60601-1-8:2012 (in the
sense of remote monitoring).
WARNING!
Risk of electric shock
Connecting devices to the auxiliary outlets of the
anesthesia machine can cause an increase in
leakage current beyond permissible values if the
protective conductor of a device fails.
Part Number: 9053586, 3rd edition
CAUTION! CAUTION!
CAUTION!
Risk of injury
If mounting accessories exceed the approved limits,
the anesthesia machine may tip over.
Maximum weight per arm = 33 lbs (15 kg).
Symbols
The following symbols appear on the Apollo and are
defined below. Symbol Explanation
XXXX
Date of manufacture
Exit menu, return to preceding menu
Automatic ventilation
Standby/operation switch
+ O2-Flush
Warning! Strictly follow these instructions
for use Operation without
Label on devices fitted with the "Operation
nitrous oxide! without nitrous oxide" option.
Rotary knob
SN Serial number
Temperature limit
Abbreviations
List of abbreviations used in the software Abbreviation Explanation
and on the device
insp. Inspiratory
Abbreviation Explanation inDes Inspiratory desflurane concentration
Agent/agent Anesthetic gas inEnf Inspiratory enflurane concentration
Air/AIR Compressed air for medical use inHal Inspiratory halothane concentration
APL Adjustable Pressure Limitation inIso Inspiratory isoflurane concentration
Aux CGO Auxiliary Common Gas Outlet inSev Inspiratory sevoflurane concentration
BW Body weight INOP Inoperable
CAL Calibration Iso. Isoflurane
CO2 Carbon dioxide LeakSYS System leakage
COM1 Interfaces used as MEDIBUS, MAC Minimum Alveolar Concentration
COM2 MEDIBUS.X interfaces Man.Spont., Manual/Spontaneous breathing
CPAP Continuous Positive Airway Pressure MAN/SPONT
CPAT Patient compliance MV Expiratory minute volume
CSYS System compliance MVLEAK Difference between inspiratory and
expiratory minute volume
ΔO2 Difference between inspiratory and
expiratory O2 MVMAND Mandatory breathed expiratory minute
volume
ΔPPS Difference in pressure to PEEP in
Pressure Support mode MVSPON Spontaneously breathed expiratory
minute volume
ΔVT Difference between inspiratory and
expiratory tidal volume MV*CO2 Expiratory minute volume CO2
Des. Desflurane N 2O Nitrous oxide
O2 Oxygen
Part Number: 9053586, 3rd edition
BTPS Body Temperature and Pressure, Vol.% Percentage gas rate in relation to total
Saturated gas volume
98.6 °F (37 °C), ambient pressure, V Volt
100 % relative humidity
xMAC Multiple of MAC
Units
Note: Throughout these instructions for use:
Ventilation pressures: cmH2O = mbar = hPa
Supply pressures: bar = kPa x 100
Part Number: 9053586, 3rd edition
System Components
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Machine Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Machine Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Gas supply block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Interface panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Vaporizers (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Vaporizer exclusion systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Dräger Vapor Interlock 2 System (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Selectatec (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Dräger Auto Exclusion 2-Vaporizer Mount (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Dräger Auto Exclusion 3-Vaporizer Mount (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
O2 flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Auxiliary oxygen flow meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Writing table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Part Number: 9053586, 3rd edition
Overview
This chapter identifies the major physical components
of the Apollo anesthesia machine and provides a brief
description of specific parts.
22
21 1
20
2
19 3
4
18
5
17 6
7
8
16
9
15 10
11
14
12
13
Part Number: 9053586, 3rd edition
11
10
2
9 3
4
7 5
6
1 2 3
6 5 4
1) This connection is not available with the "Operation without nitrous oxide" option.
Interface panel
Sensitivity
ESD
Accompanying
Documents
Caution!
!
Consult
IV System
4
SpO2
COM 2
2
COM 1
Vaporizers (Optional)
Note: Before operating the vaporizer, pay special
attention to the Instructions for Use of the
vaporizer being used. Note especially the
vaporizer flow limits.
The Dräger Vapor anesthetic agent vaporizers are
used to enrich the fresh gas with a precisely metered
quantity of vapor from the liquid anesthetic agent
being used, i.e. Isoflurane, Halothane, Enflurane,
Sevoflurane, or Desflurane.
When using a Desflurane vaporizer, it must be
connected to mains power. The auxiliary power outlet
(IEC/EN 60320-2-2/F) near the vaporizer exclusion
system is provided for that purpose.
The vaporizers being used must comply with
standard ISO 8835-4. If the internal gas
measurement system fails, an independent
measurement system complying with ISO 21647
must be used
WARNING!
Risk of patient injury
Risk of ambient air contamination
To prevent vaporizer leakage, which may lead to
low fresh-gas delivery, impaired manual
ventilation, or contamination of the ambient air,
the D-Vapor must be mounted very carefully.
Avoid catching the D-Vapor power cable behind/
underneath the housing. Make sure that the
D-Vapor is upright. Always perform a leak test
Part Number: 9053586, 3rd edition
CAUTION!
Risk of patient injury
If the vaporizer is not correctly mounted, the fresh-
gas flow will not be supplied with anesthetic agent
and the patient will not receive the correct
anesthesia.
Always double-check the position of the vaporizer,
make sure it is correctly mounted and do not mount
the vaporizer park holder close to the operable
vaporizer.
CAUTION!
Risk of ambient environment contamination and
patient injury
The parked vaporizer may be mistakenly opened if
the park holder is positioned right next to the
vaporizer mount on the anesthesia machine.
To avoid contaminating the ambient environment
and endangering the patient, always doublecheck
to make sure the correct vaporizer is being opened
before doing so.
Selectatec (Optional)
The interlock system for the Selectatec is built into
the vaporizers. When a vaporizer is selected for use,
the interlocking index pins will protrude from the
sides of the vaporizer thereby not allowing the
adjacent vaporizer to be opened. For more specific
information on the Selectatec, refer to the Selectatec
Vaporizer’s instruction manual.
Note: Only vaporizers labeled as “AUTO Table 1. Dräger Auto Exclusion Vaporizer Technical Data
EXCLUSION” vaporizers are compatible with Normal ≤ 10 L/min Dräger Vapor 2000
the Dräger Auto Exclusion 2-Vaporizer Operating Instruction for Use
Mount. See Table 1 for the Auto Exclusion Range Manual’s delivered
Vaporizer technical data. concentration accuracy
values apply.
When using a Desflurane vaporizer, it must be
plugged into the auxiliary power outlet located on the Extended >10 ≤ 15 L/min Dräger auto exclusion
side of the machine above the vaporizer mount. Operating vaporizer concentration
Range output accuracy may be
reduced.
Dräger Auto Exclusion 3-Vaporizer Figure 3. Dräger Auto Exclusion 3-Vaporizer Mount
Mount (Optional)
This system has an automatic interlock system that
ensures only one vaporizer can be used at a time.
When any one of the three vaporizers is selected for
use (opened), the interlock mechanism within that
Part Number: 9053586, 3rd edition
APL valve
The APL valve has two functions. It limits the
maximum pressure during manual ventilation and
exhausts excess gas into the scavenger system
during manual and spontaneous ventilation.
The APL valve is connected to the patient airway
through the ventilator. It functions only when the
ventilator is in Manual/Spontaneous mode.
The adjustment knob (1 in Figure 4) is used to select
between spontaneous and manual modes of
ventilation. It’s labeled to indicated approximate
pressure settings.
For spontaneous ventilation: Figure 4. APL Valve
Pressure is released for spontaneous ventilation
when the adjustment knob is rotated fully
counterclockwise, when the index mark on the knob
lines up with the index mark on the bottom of the APL
valve (2 in Figure 4). Spontaneous ventilation 1
eliminates both resistance to patient exhalation and
the need to readjust back pressure.
For manual ventilation:
In manual mode, the APL valve adjustment knob can
be rotated to change the approximate pressure at
which gas will flow through the valve and into the
scavenging system. Clockwise rotation of the
adjustment knob increases the pressure, and 2
counterclockwise rotation of the adjustment knob
decreases the pressure. Pulling up on the APL valve
head will temporarily relieve pressure.
Note: The APL valve is automatically excluded
WARNING!
Risk of patient injury
If the APL valve becomes blocked due to e.g.,
lines or cables being caught under the valve
head, the patient may be endangered.
Route all cables away from the APL valve; do
not hang lines, hoses or cables, e.g. the sample
line, on or near the APL valve.
CAUTION! 6
monitored.
Pressure monitoring must be ensured by the 1
connected device.
WARNING!
Part Number: 9053586, 3rd edition
Risk of fire
Cauterizing close to a source of oxygen can
lead to fire. Make sure that all connectors (e.g.,
Y-piece, breathing hoses including the
breathing bag, breathing system, external fresh-
gas outlet, oxygen therapy, anesthetic gas
receiving system) are leak-free so that oxygen
leakage cannot endanger the user or the patient.
WARNING!
Risk of patient injury
If the patient is connected to the auxiliary
oxygen outlet without a means of pressure
relief, high pressure will be applied and the
patient endangered.
Do not connect the patient directly to the
auxiliary oxygen outlet without ensuring a
means of pressure relief.
Writing table
The Apollo is equipped with a writing table Figure 7. Writing Table
(1 in Figure 7) which can be moved left or right or
folded down completely for convenient positioning.
To fold down the writing table, support the table with
one hand, then pull up on the release knob
(2 in Figure 7) and fold down. To bring the table up 2
again, swing it upward until it clicks into place.
User Interface
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Main screen display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
User controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Standard function keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Rotary knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Standby key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Flow control knobs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Ventilation control keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Monitoring/Configuration control keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
LED indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Screen colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Menu structure overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Part Number: 9053586, 3rd edition
Overview
This chapter provides a description of the Apollo user
interface, which enables you to view and change
monitoring, ventilation, and status information using
keys and the rotary knob.
12
11 10
User controls
Changes to system settings and screen displays are
made using the rotary knob, “keys” (keys with
permanently defined functions), and “buttons” (keys
with variable functions). All controls are described in
the following paragraphs.
Rotary knob
The rotary knob is located on the bottom right side Figure 10. Location of Rotary Knob & Standby Key
(1 in Figure 10). It is the main control used to select
Part Number: 9053586, 3rd edition
Vol. AF
Mode
1 2
Standby key
The standby key (2 in Figure 10) is used to switch
between operating modes and Standby.
1. to set the machine to Standby mode, press the
standby key > <. Then press the rotary knob
to confirm.
The standby key is also used to enter monitoring
mode while in Standby (see “Monitoring mode” on
page 177 for more information).
Vol. AF
Mode
N2O Air O2
_ + _ + _
+
LED indicators
Figure 15. LED Indicators
A number of LED indicators are located at the bottom
of the front panel. They can light up green or red, or
can remain extinguished, to indicate gas supply and
machine power status.
Vol. AF
Mode
Screen colors
Colors are used on the screen to indicate the status
of buttons and to highlight operating sequences.
Color Meaning
– can be operated, leads to another
Light green
menu or operating function
– not yet active, presettings
Part Number: 9053586, 3rd edition
Ventilation Buttons
The ventilation buttons appear dark green when
operable and turn yellow when selected. Once the
value is changed and confirmed, the button turns
back to dark green.
Monitoring/Configuration Buttons
The monitoring buttons along the right side of the
screen appear light green when operable. When
selected, their color changes to dark green. They also
change to dark green when another submenu or
function is displayed.
Colors of Parameter Settings/Values in Menus
When the user selects a menu the parameters and
values will appear on a dark green background.
Currently selected submenus are framed in an orange
border. Parameters in gray type are inactive and
cannot be selected.
Check List
Absorb. changed
Undo Change
Start Self Test
Accept
Cancel Test
Cancel Test
Standby
Alarm Limits
Self Test Results
Absorb. changed
Leak Test
Datalog
Page 1
Page 2
Delete Trend
Part Number: 9053586, 3rd edition
Cancel delete
Delete
Default Config
After entering the access code, the
menu Standby Config is opened with
the following submenus:
System Settings
Alarm Volume
Breathing Sound
(optional, only with breathing sound
module)
Pulse Volume
(optional)
Date/Time Language
Parameters
Scaling
Units
Gas Monitoring
Optional Parameters
Interfaces Datalog
Datalog entries triggered by
COM PORT 1 MEDIBUS
COM PORT 2 MEDIBUS
Select MEDIBUS
Screen Layout
Layout 1
Layout 2
Layout 3
Alarm Limits
Alarm Limits
Default Alarm Limits
Default Agent Limits
Alarms in Man/Spont.
Misc. alarm settings
Therapy related
Device related
Other
Ventilator and gas supply
Ventilator and gas supply
Parameter Default Values
Gas supply checks
Ventilator Default Settings
Weight related settings
Body Weight Related Ventilator
Settings
System Info
General Information
Activate Option
Trace 1
Trace 2
Trace 3
Remote Service
Monitor Mode
Part Number: 9053586, 3rd edition
Ventilation modes Man/Spont., Volume, Volume AF, Pressure, Press. Support, Aux CGO
Alarm Limits
Auto-set Limits 1)
CO2 Alrm ON -> off 2)
exit mode Bypass 3)
Param Settings
Scaling
Units
Agent Monitoring
Datalog Entries
Datalog entries triggered by
System Info
General Info
Trace 1
Trace 2
Trace 3
Exit Config
Start Timer
Part Number: 9053586, 3rd edition
System Setup
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Preparation before first use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Charging the battery for emergency operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Installing the breathing system and flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Filling and installing the absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Reusable absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Disposable CLIC absorber (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Connecting the gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Connecting pipeline supply of N2O, Air, and O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Connecting the backup gas cylinders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Caution when handling O2 cylinders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Connecting the scavenger system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Connecting the anesthetic gas receiving system AGS (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Connecting the passive scavenger system (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Connecting the endotracheal aspiration system (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Installing vaporizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Installing the flexible arm for the breathing bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Connecting the patient system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Connecting the patient circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Table with recommended hose configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Connecting AC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Part Number: 9053586, 3rd edition
Overview
This chapter provides information on how to set up
and install all system components needed to prepare
the Apollo for use. The setup procedure shall be
followed by the performance of the periodic
manufacturer’s procedure.
WARNING!
Risk of patient injury
Correct preparation of the anesthesia machine
is required to minimize the general risks
associated with the anesthesia machine.
Use only clean and disinfected parts and always
strictly follow the cleaning and assembly
instructions contained in these Instructions for
Use to prevent infection of patient or user.
WARNING!
Risk of device failure
If the batteries have not been sufficiently
charged and a power failure occurs, it may only
be possible to continue operation for a short
period of time.
Charge batteries for at least 10 hours before
first use or after storage.
The battery must be charged for 10 hours before Figure 16. Location of AC Power LED
using the anesthesia machine for the first time:
1. Plug the mains power plug of the Apollo
anesthesia machine into the mains outlet. Vol. AF
Mode
CAUTION!
Risk of device failure
In the event of a power failure, any devices
connected to auxiliary power outlets will not be
powered by the UPS.
Pay special attention to all power indicators of
connected devices.
WARNING!
Risk of battery failure
Allowing the battery to run low can damage it.
It must be charged at least every four weeks.
WARNING!
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits that
were not removed during reprocessing can
ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling such as residual
mucus, medication aerosols, and particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
1. Press the release button on the ventilator unit Figure 17. Installing the Flow Sensors
and pull it out. 2
2. Loosen the three sealing screws on the ventilator
(1 in Figure 17) a quarter turn counterclockwise
with the wrench supplied.
3. Pull the breathing system up and out by the 3
handle (2 in Figure 17). 1
4
4. Unscrew the inspiratory and expiratory ports (4 in
Figure 17) by turning them counterclockwise.
Part Number: 9053586, 3rd edition
6. Orient the inspiratory and expiratory ports Figure 18. Installing the Breathing System
(4 in Figure 17) so that the key on each port lines
up with the slot. Install the ports and tighten by
turning clockwise. Carefully seat the breathing
system onto the ventilator module, and tighten 1
the three sealing screws (1 in Figure 18) on the
ventilator cover.
1
WARNING!
Risk of scavenger becoming blocked
If objects such as packing foil get into the
device, e.g., the breathing system or the
ventilator drawer, the scavenger may become
blocked.
Make sure that there is no packing material left
inside the device.
Reusable absorber
1. Push the insert fully into the absorber canister Figure 19. Installing the Absorber Canister
(1 in Figure 19).
2. Fill the absorber canister with fresh soda lime up
to the MAX mark.
WARNING!
Risk of injury
Absorbent is caustic and is a strong eye, skin,
1
and respiratory tract irritant. 2
Use care when handling the absorbent to avoid
spills.
CAUTION!
Risk of device failure
It is recommended that Drägersorb 800 + or
Drägersorb FREE are used.
Do not use powdered soda lime, as a higher dust
load may impair functionality of the Apollo
anesthesia machine.
WARNING!
1
The disposable absorber must be clicked into
place before switching on the Apollo. This
ensures that the absorber is included in the leak
and compliance test for the anesthesia machine.
WARNING!
Risk of patient injury
The soda lime loses humidity. Generally, if the
CAUTION!
Risk of gas supply failure
If all gas supplies (pipeline or cylinder) are not
connected correctly, the reserve system will not be
available in the event of a gas supply failure.
Make sure that all supplies are connected according
to the engraving on the gas supply block and the
illustrations at the back of the machine. After
connecting the supplies, ensure proper functionality.
CAUTION!
Risk of device failure
Compressed gas supply (pipeline or cylinder): To
avoid damaging the device(s) attached to a gas
supply, use only medical gases. Pay particular
attention to national and international standards
regulating the use of medical gases.
Part Number: 9053586, 3rd edition
WARNING! Air
pipeline supply
Risk of patient injury connection
Pipeline delivery hoses used between wall
outlets and anesthesia machines have caused
accidents when, during assembly, an oxygen
fitting was placed on one end of the hose and a
nitrous oxide fitting on the other end.
Carefully check hoses each time you connect a
machine to a wall outlet to ensure that both
ends of the hose are indexed for the same gas.
CAUTION!
Risk of device failure
In order for the inhalation anesthesia machine to
operate as specified, the supply pressures at the N2O O2
machine inlet must be within a range of 2.7 and pipeline supply pipeline supply
6.9 kPa x 100. connection* connection
If the pipeline supply pressure LEDs remain Figure 22. Location of pipeline supply pressure LEDs
dark, it means that the pressure is below 39 psi
or that the hoses are not connected properly.
Vol. AF
Mode
1
Part Number: 9053586, 3rd edition
CAUTION!
Risk of gas supply failure
Should the pipeline gas supply fail, the backup gas
cylinders on the anesthesia machine will provide a
reserve gas supply.
To prevent a complete gas failure, the backup gas
cylinders should remain on the device, valves
closed, in reserve even if the anesthesia machine is
connected to pipeline gas supply.
WARNING!
Risk of gas supply failure
When attaching a cylinder, ensure that only one
washer is installed between the cylinder and the
yoke gas inlet. The use of multiple washers will
inhibit the pin-index safety system. Be sure to
verify the presence of the index pins each time a
cylinder is installed. Never attempt to override
the pin-index safety system.
WARNING!
If the cylinder pressure LEDs remain dark, it means Figure 26. Location of Cylinder Pressure LEDs
that the cylinder pressure is inadequate or that the
cylinders are not connected properly.
5. Close the cylinder valves. Vol. AF
Mode
WARNING!
Risk of gas supply failure
If the valves remain open when connected to the
pipeline gas supply, gas may be withdrawn from
the backup gas cylinders.
Close backup gas cylinder valves whenever
pipeline gas supply is sufficient.
Cylinders attached to the hanger yokes must contain Table 2. Recommended Cylinder Gas Pressures
gas at the recommended pressures shown in
PSI/bar - FULL
Table 2. (Indicated pressures are for E-size cylinders GAS PSI/bar - MIN
(typical full load)
at 70°F/21°C.) Cylinders measuring less than the
minimum recommended pressure (PSI - MIN) should Air 1900/131 1000/69
be replaced with new, full cylinders.
N2O 745/51 600/42
O2 1900/131 1000/69
WARNING!
Risk of explosion
CAUTION!
Risk of increased ambient gas concentration
Ambient air may become contaminated with
anesthetic agent if the scavenger hoses are
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
WARNING!
Risk of scavenger becoming blocked
If objects such as packing foil get into the
device, e.g., the breathing system or the
ventilator drawer, the scavenger may become
blocked.
Part Number: 9053586, 3rd edition
1. Install the receiving system on the machine by Figure 27. Anesthetic gas receiving system AGS
sliding its bracket onto the two shoulder screws
on the side of the machine.
2. Connect one end of the transfer hose to the
fitting on the receiving system (1 in Figure 27).
2
3. Connect the other end of the transfer hose to the
scavenger connection on the back of the
anesthesia machine (2 in Figure 27). 1
4
3
4. Connect the waste-gas vacuum hose to the
output connection on the receiving system
(3 in Figure 27).
5. Connect the other end of the vacuum hose to the
hospital waste-gas disposal system.
Note: Activate hospital vacuum system before
using the receiving system.
6. Make sure that the AGSS is ready for operation.
Check if the flow indicator at the AGS floats
between the two marks.
Note: During use, the float indicator in the flow
indicator should stay between the upper and
lower marks. If necessary, regulate flow
using the flow adjustment valve
(4 in Figure 27).
WARNING!
Risk of patient injury
If the side openings of the receiving system are
blocked, negative pressure may result in the
breathing system and the patient’s lungs.
CAUTION!
Risk of increased ambient gas concentration
Ambient air may become contaminated with
anesthetic agent if the scavenger hoses are
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
WARNING!
Risk of patient injury
If not used correctly, the suction unit may injure
the patient.
Prior to use, disconnect the patient from the
ventilator, and pay special attention to the
instructions for use of the suction unit.
Installing vaporizers
Install vaporizers as directed in the appropriate
Instructions for Use supplied with the vaporizers
available for use with the Apollo.
1
Part Number: 9053586, 3rd edition
WARNING!
Risk of burns
Conductive breathing hoses or face masks may
cause burns during HF surgery.
Do not use these types of hoses and masks in
combination with HF surgery.
WARNING!
Risk due to particles and dust
To protect the patient from particles and dust, a
filter must be used between the inspiratory limb
of the breathing system and the patient.
Use a Y-piece filter or filter on the inspiratory
port.
CAUTION!
WARNING!
Risk of patient injury
If the breathing hoses are wrongly connected,
the patient might be inadequately ventilated and 5
supplied with fresh gas.
Make sure that all breathing hoses are correctly
connected to the breathing system. 4
CAUTION!
Risk of gas measurement failure and device failure
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
Sample gas lines are single-use articles and must
not be disinfected.
WARNING!
Risk of gas measurement failure
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced at least every
four weeks.
WARNING!
Risk of gas measurement failure and device
failure
If alcohol or cleaning agents/disinfectants come
in contact with the inside of the water trap, they
can damage the diaphragm and the
measurement system may fail as a result.
Do not use these substances and do not wash,
flush, or sterilize the water trap.
CAUTION!
Risk of contamination of the device
Do not put the device into operation without a water
trap.
CAUTION!
Risk of incorrect measured values
Silicone can enter the measuring cuvette and distort
the gas measurement.
Do not spray the O-rings of the water trap holder
with silicone spray.
CAUTION!
Risk of incorrect measured values
Aerosols can damage the diaphragm and the
measurement system.
Do not use aerosols in the breathing system. The
water trap must not be used in combination with a
medical nebulizer.
Part Number: 9053586, 3rd edition
insp. insp.
exsp. exsp.
insp.
exsp.
WARNING!
Risk of negative lung pressure
If filters are blocked, the sample gas flow could
cause negative lung pressure.
When ventilating pediatric patients and
neonates, do not use HME filters or other filters
at the Y-piece if sample gas is being taken at the
tube adapter.
WARNING!
Risk due to additional components in breathing
circuit
When additional components are used or in the
case of hose configurations that differ from the
standard or recommended hose configurations,
the inspiratory and expiratory breathing
resistances may exceed the standard
requirements.
If such configurations are used, the user must
pay particular attention to the measured values.
Observe the instructions for use of the
additional components.
WARNING!
Risk of increased rebreathing
If coaxial hoses are used, leakages between the
inner and outer hose cannot be detected during
the leak test.
To avoid insufficient gas exchange and
rebreathing of CO2, monitor the measured gas
concentration extremely carefully.
Part Number: 9053586, 3rd edition
WARNING!
Risk due to incorrect measured values
Replacing the breathing hoses, filters,
vaporizers, or soda lime may change the
calculated leakage and compliance values of
the anesthesia machine and affect the therapy.
– Perform a leakage and compliance test after
replacing breathing hoses, particularly
extendable hoses, vaporizers, and soda
lime.
– Perform a leakage and compliance test after
adjusting the length of extendable hoses.
WARNING!
Risk when adjusting the hose length
When the hose length is changed, resistance
and compliance may change. This can result in
an increased or reduced ventilation volume for
neonates.
Do not use extendable hoses, particularly for
neonates.
Connecting AC power
Connecting auxiliary devices Figure 33. Location of Auxiliary Outlets on Back of
Machine
The Apollo has two auxiliary outlets on the back of
the machine (1 in Figure 33). Each outlet is rated 4
amps and is protected by circuit breakers.
1. Connect the external device to an outlet on the
back of the machine
1
1
Seneltivity
CAUTION!
ESD
Accompanying
Documents
Caution!
!
Consult
IV System
Risk of device failure
SpO2
COM 2
COM 1
In the event of a power failure, any devices
connected to auxiliary outlets will not be powered by
the UPS.
Pay special attention to all power indicators of
connected devices.
CAUTION! 22
!
WARNING!
Risk of device failure
If additional power extension sockets are
connected to the auxiliary outlets, device
internal electronics may be overloaded.
Do not connect additional power adapter
sockets to the auxiliary outlets.
WARNING!
Risk of electric shock
Connecting devices to the auxiliary outlets of
the anesthesia machine can cause an increase
in leakage current beyond permissible values if
the protective conductor of a device fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting a
device or several devices increases the leakage
current to a value which exceeds the
permissible value, do not use the auxiliary
outlets of the anesthesia machine. Instead, use
a mains power socket on the wall.
The system must fulfill the requirements for
medical electrical equipment in accordance with
the relevant standards, see “Relevant
standards” on page 308.
WARNING!
Risk of injury
If handled incorrectly, the anesthesia machine
may become top-heavy and tip over causing
injury to the patient and/or user.
Observe the following points to prevent this
hazard.
CAUTION!
Risk of physical injury
To avoid physical injury, e.g. pinching, pay special
attention to edges, moving parts and corners when
Part Number: 9053586, 3rd edition
working with
– drawers,
– the ventilator module,
– doors,
– the writing tray,
– swivel arms for mounted devices,
– gas cylinders,
– vaporizer units,
– CLIC absorbers and CLIC adapters,
as well as other accessories.
Pre-use Checkout
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Checking the Workstation according to the Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Power on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Check list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Self test results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
System compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Leak tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Leak system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Leak Man/Spont . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Locating and eliminating leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Additional suggestions to isolate components of the breathing system for leaks: . . . . . . . . . . . . . . . . . . . . . . . 106
Emergency start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Part Number: 9053586, 3rd edition
Overview
The pre-use checkout procedure must be performed
to ensure that the Apollo is ready for use. This is a
recommended procedure. Follow the institution’s
policies for specific procedures.
If the Apollo fails any checkout routine, do not use
the machine until corrective action is taken. If
indicated, contact an authorized representative of
DrägerService for inspection of the unit.
WARNING!
Risk of device failure and/or patient injury
Do not insert any additional components into or
modify the Apollo after the checkout procedure
has been started.
The anesthesia machine will not meet the
specified technical data.
WARNING!
Risk of patient injury
Inappropriate hose length affects compliance
and can result in incorrect tidal volume delivery
to the patient.
Patient hoses must be adjusted to the
appropriate lengths prior to performing the leak
and compliance tests.
Prerequisites
The device has been prepared (see “Cleaning and
Maintenance” on page 225) and assembled ready for
operation.
The pipeline supply and the power supply must be
connected.
WARNING!
Risk of explosion, fire
If an oxygen leak is suspected within or near the
inhalation anesthesia machine, do not initiate
operation.
Disconnect all oxygen supplies and contact a
trained service technician.
CAUTION!
Risk of inadvertent movement
If not properly secured, the device may move
inadvertently during operation.
Apply the brakes on the device to ensure it cannot
be moved accidentally.
WARNING!
Risk of electric shock
Connect the electrical power cable to a hospital-
grade live AC receptacle that accepts and
properly grounds the power cable. Do not use
“cheater plugs”. The term “cheater plug”
implies any and all electrical plugs or other
devices that can inhibit or prohibit the proper
grounding of the anesthesia machine.
Check list
After about 35 seconds, a check list for manual tests
to be performed by the user is displayed
(see Figure 37).
1. Check the components as instructed in the check Figure 37. Check List Screen
list on the screen and as described in this
procedure.
If the self test has to be interrupted, e.g. for a quick
start in an emergency:
2. Press the button >Cancel Test< (1 in Figure 37),
and proceed as specified in “Emergency start” on
page 107.
1
The self test can be canceled up to ten consecutive
times.
Part Number: 9053586, 3rd edition
WARNING!
Risk of device failure and/or patient injury
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency, carry out a complete self test as
soon as practicable.
WARNING!
Risk of gas supply failure
If the valves are open when connected to the
pipeline gas supply, gas may be withdrawn from
the backup gas cylinders.
Close cylinder valves whenever the pipeline gas
supply is sufficient.
1. Using the provided cylinder wrench, slowly open
the cylinder valves.
2. Verify that the cylinder pressure LEDs light up
green (2 in Figure 38).
The LEDs light up green when the cylinder pressure
for O2 and Air is over 290 psi and the pressure for
N2O is over 145 psi.
The cylinder pressures are shown in the Check List
WARNING!
Risk of device failure
The correct operation of the anesthesia
machine will be impaired if condensation enters
the breathing system and/or the ventilator
diaphragm.
If condensation is a frequent problem, install
water traps in the breathing hoses.
Part Number: 9053586, 3rd edition
WARNING!
Risk of strangulation
If not positioned with care, hoses, cables, and
similar machine components may endanger the
patient.
Take special care when connecting the patient.
WARNING!
Risk of patient injury
The soda lime loses humidity. Generally, if the
humidity falls below a minimum set point, the
following undesirable reactions can occur,
independent of the type of lime and the
inhalation anesthetic being used:
– reduced CO2 absorption,
– increased heat build-up in the absorber and
thus an increased breathing gas
temperature,
– formation of CO,
– absorption and/or decomposition of the
anesthetic agent.
These reactions could pose a danger to the
patient.
If using dry gases, only flush the anesthesia
system briefly, and only if necessary.
WARNING!
Risk of patient injury
If the flow controls are left open, the ensuing
flow of gas may dry out the soda lime,
endangering the patient.
The flow control valves should be closed when
the machine is in the standby mode or when it is
switched off.
Self test
Figure 43. Check List Screen
If all checks in the Check List are completed
successfully:
1. Press the rotary knob or the >Start Self Test<
key on the check list screen (1 in Figure 43) to
begin the Apollo automated self test. 1
WARNING!
Risk of device failure or patient injury
Functions coded yellow do not meet with the
specified technical data.
The error should be remedied as soon as
possible.
WARNING!
Risk of device failure or patient injury
Functions coded red must be remedied before
starting, e.g. if there is no O2 supply.
The device cannot be operated in this state.
WARNING!
Risk of inadequate monitoring
If the flow sensor, oxygen sensor, or gas sensor
is not operational, adequate substitute
monitoring must be ensured before starting the
anesthesia machine!
Part Number: 9053586, 3rd edition
Green FUNCTIONAL
2
Every component of the system is in satis-
factory operational order.
Yellow CONDITIONALLY FUNCTIONAL
A non-critical fault was detected. Apollo may
be used, but call DrägerService or your local
authorized service organization.
Empty The self test was canceled.
WARNING!
Risk of device failure and/or patient injury
Canceling the self test may lead to
malfunctions. Special attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, perform a complete
self test as soon as practicable.
System compliance
Apollo determines the current compliance of the
breathing circuit consisting of filters, hoses, and a
Y-piece. Typical values for the inspiratory system
compliance are between 0.5 and 2.6 mL/hPa
(mL/cmH2O).
In volume-controlled ventilation, system compliance
is compensated. For this purpose, Apollo increases
the applied tidal volume on the basis of the difference
between PEEP and plateau pressure in accordance
with the determined compliance value.
Leak tests
The Apollo tests for leaks in the mechanical
subsystem and in the Man/Spont system.
Part Number: 9053586, 3rd edition
Leak system
Figure 48. Mechanical Ventilation Leak Test
– Figure 48 shows the components tested in the
mechanical ventilation branch.
– This branch is tested with positive pressure.
– Leaks are indicated on the Self Test Results
screen by the yellow/green test result indicator
and by posting the leak value in mL/min (Leak
(system) test result in Figure 47).
Leak Man/Spont
Figure 49. Man/Spont Ventilation Leak Test
– Figure 49 shows the components tested in the
Man/Spont leak test.
– This test is performed with sub-atmospheric
pressure.
– Leaks are indicated on the Self Test Results
screen by the red/yellow/green test result
indicator and by posting the leak value if it is over
150 mL/min (Leak (Man/Spont) result in Figure
47).
The Apollo determines the current leakage of the
breathing system and breathing hoses. The system
tolerates leaks of up to 150 mL/min.
Note: For leaks of more than 150 mL/min:
Check the components of the breathing
system and the breathing hoses. Repair any
leaks and repeat the leak test.
Emergency start
WARNING!
Risk of incorrect delivery
The leak and compliance tests are not
performed if the self test is cancelled, and no
leak and compliance information is available.
The accuracy levels specified in the chapter
“Specifications” cannot be guaranteed.
appears.
6. Press the >Cancel Test< key on the Check List Figure 51. Check List Screen
screen (1 in Figure 51).
The machine runs through a minimal self test
that lasts about 10 seconds. Manual ventilation is
interrupted during this time, but spontaneous
breathing can continue.
WARNING!
Risk of device failure and/or patient injury
Canceling the self test may lead to
malfunctions; greater attention is required
during operation.
Always perform a complete self test, unless
acting in an emergency situation. If canceled for
an emergency situation, carry out a complete
self test as soon as practicable.
Operation Summary
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Typical operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Loading default settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Entering the patient’s age . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Entering the patient’s ideal body weight (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Setting the fresh-gas flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Setting vaporizer concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Setting ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Changing patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Switch to standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Changing soda lime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Reusable absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Disposable CLIC absorber (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Activating the CO2 bypass function (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
End of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
When Apollo is not in use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Part Number: 9053586, 3rd edition
Overview
This chapter of the instructions for use summarizes
basic operation of the Apollo, including starting
operation, changing patients, and ending operation.
Specific information on setting ventilation and
monitoring parameters is provided in later chapters of
the instructions for use.
Safety Information
WARNING!
Risk of electric shock
Touching the patient and electrical device
contacts could result in an electric shock.
Do not touch the patient and the electrical
device contacts at the same time.
Typical operation
Operation of the Apollo begins with the standby
screen which is displayed after the initial self tests.
This screen allows the user to restore default settings
and enter the patient parameters needed to begin a
case.
2 3 1
Entering the patient’s age
The set age influences the calculation of the MAC
value, the volumeter scale, the vertical axis of the
loops and ventilation monitoring graphs as well as
the alarm limits for (optional) SpO2 monitoring, and
CAUTION!
Risk of patient injury
The use of minimum flow or low flow settings may
lead to the accumulation of metabolic products in
the breathing system.
To avoid this risk, use appropriate soda lime, or set
higher fresh-gas flows, and always use the gas
measurement module provided by the anesthesia
machine. Vol. AF
Mode
N2O Air O2
_ + _ + _
+
WARNING!
Risk of patient awareness
If a complete gas supply failure occurs, the
anesthesia machine will continue to function
with ambient air. However, anesthestic agents
will no longer be delivered and the inspiratory
gas composition will be diluted.
Carefully monitor the gas mixture and, if
necessary, use IV anesthetics.
Changing patients
Follow the steps below for successive patient cases.
WARNING!
Risk of patient injury
Restored default settings may contain settings
inappropriate for a new patient.
After default settings have been restored, make
sure the ventilation monitoring settings are
appropriate to the patient connected.
Reusable absorber
WARNING!
Risk of increased inspiratory CO2
concentrations
When the absorber is swung out, no CO2 is
absorbed.
Always make sure the absorber is clicked into
place after installing or replacing.
Leak test
WARNING!
Risk of patient injury
The system will be pressurized during the leak
test.
To prevent patient injury, do not perform the
leak test with a patient connected to the
anesthesia machine.
WARNING!
Risk of misleading data
Changing the breathing hoses, vaporizers, or
soda lime can modify the calculated leak and
compliance values of the anesthesia machine
and influence the therapy settings.
Perform a leak test after the breathing hoses,
vaporizers, or soda lime have been replaced.
WARNING!
Risk of increased inspiratory carbon dioxide
concentrations
CO2 is not absorbed in the breathing system
when the absorber is swung out.
Always make sure the absorber is clicked into
place after installing or replacing.
To deactivate the CO2 bypass function, swing the Figure 59. Pushing CLIC Adapter Back into Place
disposable absorber (2 in Figure 59) back into the
system until it engages.
End of operation
To set the Apollo to Standby mode:
WARNING!
Risk of fire
In order to avoid the accumulation of potentially
hazardous oxygen concentrations in the
anesthesia machine or the operating room, all
sources of oxygen must be closed and the
anesthesia machine disconnected from them
when the anesthesia machine is not in use.
WARNING!
Risk of gas supply failure
If the valves remain open when connected to the
pipeline gas supply, gas may be withdrawn from
the backup gas cylinders.
Close cylinder valves whenever the pipeline gas
supply is sufficient.
WARNING!
Risk of gas supply contamination
When the central gas supply is connected, the
smallest internal leakage can cause
contamination of the supply gases.
Always disconnect the medical gas hoses from
the terminal unit when the device is not in use.
WARNING!
Risk of device failure and/or patient injury
The self test checks sensitive internal device
processes the functionality of which, if not
regularly tested, may fail or not be available.
It is strongly recommend that the Apollo be
Part Number: 9053586, 3rd edition
CAUTION!
Risk of device failure
Larger quantities of condensation may impair
operation of the anesthesia machine and/or lead to
failure of the equipment.
Remove any water which may have accumulated in
the ventilator diaphragm.
Ventilation
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Manual/Spontaneous ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Setting the APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
For manual ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
For spontaneous breathing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Starting Manual/Spontaneous Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Presetting the Manual/Spontaneous mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Starting the Manual/Spontaneous mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
O2 flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Volume-Controlled Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Compliance compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Starting volume-controlled ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Presetting the volume-controlled ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Starting the volume-controlled ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Synchronized volume-controlled ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Synchronized volume-controlled ventilation with Pressure Support (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . 132
Volume Mode AutoFlow - Volume AF (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Starting Volume Mode AutoFlow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Presetting the Volume AutoFlow ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Starting the Volume AutoFlow ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Synchronized volume-guaranteed ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Synchronized volume-guaranteed ventilation with Pressure Support (Optional) . . . . . . . . . . . . . . . . . . . . . . . . 137
Part Number: 9053586, 3rd edition
Overview
The Apollo supports the following ventilation modes:
– Manual/Spontaneous ventilation Man/Spont
– Volume-controlled ventilation Volume Mode.
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional)
– Pressure-controlled ventilation Pressure Mode.
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional)
– Pressure-assisted spontaneous breathing
Pressure Support CPAP (optional)
– Volume AF (Volume Mode AutoFlow) (optional).
With activation of:
– Sync. (Synchronization)
– Press. Support (Pressure Support) (optional).
In addition, the optional auxiliary common gas outlet
(Aux CGO) is available for use of non-rebreathing
systems.
CAUTION!
Risk of inadequate alarm monitoring
Some Alarm limits may be automatically modified
when the ventilation is changed or the settings are
modified.
Check or adapt alarm limits each time the
ventilation mode is changed or when the settings
are modified while a patient is being ventilated.
Part Number: 9053586, 3rd edition
Manual/Spontaneous ventilation
To use the Manual/Spontaneous ventilation mode,
the user must first set the APL valve to the
appropriate mode and then select the mode using
the button >Man/Spont< on the front display panel.
WARNING!
Risk of patient injury
If the APL valve becomes blocked due to for
example, lines or cables being caught under the
knob, the patient may be endangered.
Route all cables away from the APL valve. Do
not hang lines, hoses or cables, for example,
the sample line, on or near the APL valve.
The parameter buttons turn dark green and parameters ranges settings1)
display the preset parameter values. The “Man/
Spont” indication in the status field stops flashing Age <1 to 120 40
and is displayed continuously. Manual/ >Age<
Spontaneous ventilation begins. [years]
Ideal body weight 1 to 120 kg, --
5. Set an appropriate fresh-gas flow. Verify
>Weight<2) 1 to 240 lbs.
adequate flow by checking the total flow meter.
[kg/lbs.]
If a Man/Spont parameter has to be changed
during ventilation: 1) Site defaults can be set instead.
2) Optional
– Press the button for the parameter to be
changed, turn the rotary knob to the desired
value, and press the rotary knob to confirm.
O2 flush
For flushing and rapidly filling the breathing system Figure 63. Location of the O2 flush button
and breathing bag with O2 while bypassing the
vaporizer.
Press the >O2+< button (1 in Figure 63). O2 flows
into the breathing system without anesthetic gas as
long as the button is pressed.
Volume-Controlled Ventilation
The Apollo has a volume-controlled ventilation mode Figure 64. Respiratory Cycle - Volume Mode
with fixed mandatory tidal volume (VT) and frequency PMAX
Pressure
Trigger indicator Trigger indicator
(Freq.). Synchronization can be activated, as well as PPEAK
variable Pressure Support for spontaneous breathing PPLAT Trigger
efforts (optional). PPS
on
Compliance compensation
Ventilator compliance compensation is continuously
applied during volume-controlled ventilation so that
the tidal volume delivered to the patient corresponds
to the VT setting. Ventilator compliance is determined
during the leak test performed in Standby mode. To
have compliance compensation work accurately, it is
important that the patient hoses used during the leak
test match the type of hoses used during the proce-
dure.
Note: When the ventilator settings for Volume Figure 65. Volume Mode Screen
Mode cause the ventilator to operate at its 2
limits of performance, it is not possible for
the Apollo to apply compliance
compensation. If the ventilator's performance
limit is reached, it is not possible to
Part Number: 9053586, 3rd edition
2. Press the button for the parameter to be Table 4. Adjustment ranges and factory settings for
changed; its color changes to yellow to indicate Volume Mode
that it is selected. Ventilation Adjustment Factory
3. Turn the rotary knob to adjust the parameter to parameters range settings 1)
the desired value, and press the rotary knob to Pressure limitation 10 to 70 40
confirm (4 in Figure 65). >PMAX< min. PEEP
[cmH2O] +10
Continue to set the values for the other
Tidal volume 20 to 14002) 600
parameters.
>VT<
The parameters that can be set for Volume [mL]
Mode are shown in Table 4, along with their Frequency 3 to 100 12
adjustment ranges and factory default values. >Freq<3),4)
[bpm]
Starting the volume-controlled ventilation mode Inspiratory time 0.2 to 6.7 1.7
>TINSP<4)
4. When all the parameters have been preset,
[sec.]
press the rotary knob to start volume ventilation
Insp. pause time : 0 to 60 10
(4 in Figure 65).
insp. time
The parameter buttons turn dark green and >TIP : TINSP<
display the preset parameter values. The [%]
“Volume” indication in the status field stops >PEEP< 0 to 20 0
flashing and is displayed continuously. Volume [cmH2O] max. PMAX –10
ventilation begins. Trigger sensitivity OFF, 3.0
>Trigger< 0.3 to 15 (Press. Supp.)
If a volume parameter has to be changed during [L/min] OFF
ventilation: (Vol./Press.
– Press the button for the parameter to be Mode)
changed, turn the rotary knob to the desired Pressure Support OFF,
value, and press the rotary knob to confirm. >ΔPPS<5) 3 to 50 5
[cmH2O] (Press. Supp.)
Note: Volume Mode can also be started without max. PMAX–PEEP OFF
presetting values: (Vol./Press.
Mode)
– Press the >Vol Mode< key located at the
bottom of the display panel, and press the Rise time 0.0 to 2.0 0.0
>TSLOPE<
rotary knob. Default parameter values are
1. When all the parameters have been preset, Table 5. Adjustment ranges and factory settings for
press the rotary knob to start Volume AF Volume AF mode
ventilation (4 in Figure 69). Ventilation Adjustment Factory
The parameter buttons turn dark green and display parameters range settings 1)
the preset parameter values. The “Vol. AF” indication Pressure limitation 10 to 70 40
in the status field stops flashing and is displayed >PMAX< min. PEEP
continuously. Volume AutoFlow ventilation begins. [cmH2O] +10
Tidal volume 20 to 14002) 600
If a volume parameter has to be changed during >VT<
ventilation: [mL]
1. Press the button for the parameter to be Frequency 3 to 100 12
changed, turn the rotary knob to the desired >Freq<3),4)
value, and press the rotary knob to confirm. [bpm]
Inspiratory time 0.2 to 6.7 1.7
Note: Volume AF can also be started without >TINSP<4)
presetting values: [sec.]
– Press the >Vol. AF Mode< key located at >PEEP< 0 to 20 0
the bottom of the display panel, and press [cmH2O]
the rotary knob. Default parameter values Trigger sensitivity OFF, 3.0
are used. >Trigger< 0.3 to 15 (PressSupp.)
[L/min] OFF
(Volume AF)
Pressure Support OFF,
>ΔPPS<5) 0 to 50 5
[cmH2O] (PressSupp.)
max. PMAX–PEEP OFF
(Volume AF)
Rise time 0.0 to 2.0 0.0
>TSLOPE<
[sec.]
Age <1 to 120 40
>Age<
[years]
Part Number: 9053586, 3rd edition
Pressure-Controlled Ventilation
The Apollo has a pressure-controlled ventilation Figure 72. Respiratory Cycle - Pressure Mode
mode with fixed pressure limitation PINSP and Pressure Trigger indicator Trigger indicator
frequency Freq. as well as with optional
synchronization and variable Pressure Support for PINSP
spontaneous breathing efforts (optional).
Trigger
A continuous pressure is applied to the patient during TSLOPE Time [s]
on
the inspiratory time TINSP (refer to Figure 72). The TINSP TEXP
1/Freq.
rate at which the pressure curve rises is preset via
the rise time TSLOPE. Synchronization and Pressure
Flow
Support are controlled by the sensitivity of the flow on
trigger and the level of ΔPPS. The maximum time
interval for controlled ventilation is set via the
frequency. To maintain a constant frequency, a time 25%
1. When all the parameters have been preset, Table 6. Adjustment ranges and factory settings for
press the rotary knob to start pressure ventilation Pressure Mode
(4 in Figure 73). Ventilation Adjustment Factory
The parameter buttons turn dark green and parameters range setting1)
display the preset parameter values. The
Pressure limitation 5 to 70 15
“Pressure” indication in the status field stops >PINSP< min. PEEP
flashing and is displayed continuously. Pressure [cmH2O] +5
ventilation begins.
Frequency 3 to 100 12
If a pressure parameter has to be changed >Freq<2),3)
during ventilation: [bpm]
– Press the button for the parameter to be Inspiratory time 0.2 to 6.7 1.7
changed, turn the rotary knob to the desired >TINSP<3)
[sec.]
value, and press the rotary knob to confirm.
>PEEP<4) 0 to 20 0
Note: Pressure Mode can also be started without [cmH2O] max. PINSP –5
presetting values:
Trigger sensitivity OFF, 3.0
– Press the >Press Mode< key located at >Trigger< 0.3 to 15 (Press. Supp.)
the bottom of the display panel, and press [L/min]
the rotary knob. Default parameter values OFF
are used. (Vol./Press.
Mode)
Note: There is 15-second timeout period for
making ventilation mode changes, with a Pressure Support OFF,
5-second audible tone sequence after the >ΔPPS<5) 3 to 50 5
[cmH2O] (Press. Supp.)
first 10 seconds. If the new setting is not
confirmed within the timeout period, the max. PMAX–PEEP OFF
current ventilation setting remains in effect. (Vol./Press.
Mode)
Rise time 0.0 to 2.0 0.0
>TSLOPE<
Part Number: 9053586, 3rd edition
[sec.]
Age <1 to 120 40
>Age<
[years]
Ideal body weight 1 kg to 120 kg, --
>Weight<5) 1 lbs. to 240 lbs.
[kg/lbs.]
1) Site defaults can be configured instead.
2) Depending on the configuration, the inspiratory time (TINSP) can be
automatically changed together with adjustment of the frequency so that the
resultant ratio of inspiration to expiration (I : E) remains constant. Only
applies if trigger = OFF. See the chapter “Configuration”.
3) The resultant ratio of inspiration to expiration (I : E) is also displayed in
parallel.
4) Depending on the configuration, the pressure limit (PINSP) can be changed
automatically together with adjustment of the PEEP value.
See the chapter “Configuration”.
5) Optional.
peak flow. Any leakage is compensated Flow trigger Flow trigger No trigger
TINSP changes
Part Number: 9053586, 3rd edition
Frequency changes
Depending on the configuration, the inspiratory time Figure 78. Automatic Freq/TINSP Change
TINSP can be automatically changed together with
adjustment of the frequency Freq. in volume-
controlled or pressure-controlled ventilation modes
without synchronization, so that the resultant ratio of
inspiration to expiration I:E remains constant.
To make a combined Freq./TINSP parameter change: 1 1
PEEP changes
Depending on the configuration, the inspiratory Figure 79. Automatic PEEP/PINSP Change
pressure PINSP can be automatically changed when
the PEEP value is changed in the pressure-
controlled ventilation mode.
To make a combined PEEP/PINSP parameter
change:
1 1
1. Press the >PEEP< button on the Pressure Mode
screen. The key lights up yellow, along with the
value for PINSP, to indicate that both values will
change (1 in Figure 79).
2. Turn the rotary knob to adjust the PEEP value.
The PINSP value is adjusted at the same time.
3. When the desired PEEP value is displayed,
press the rotary knob to set the value. Both the
>PEEP< and the >PINSP< keys turn green.
If so configured, the lower alarm limit for the
airway pressure PAW will be automatically
changed when the PEEP value is changed.
Example: Bain system Figure 80. Auxiliary CGO - Bain Circuit shown
1. Prepare the system according to the
corresponding Instructions for Use.
2
To monitor O2, CO2, and anesthetic gases: 1
3
2. Connect the sample line to the Luer lock 1
connection on the mask manifold and to the
water trap connection on the front of the Apollo
(1 in Figure 80).
For mask manifolds without a sample line connector:
– Place a T-piece with filter between the mask pipe
and fresh-gas connection port.
or:
– where applicable, use a Luer lock filter
connection.
3. Connect the fresh-gas hose of the Bain system
to the auxiliary common gas outlet
(2 in Figure 80).
4. To remove exhaled gas, connect the non-
rebreathing system connector (3 in Figure 80) to
the Y-piece.
WARNING!
Risk of patient injury
If the bag does not inflate, the patient will not
receive adequate ventilation. Switch to the
Apollo internal breathing system and ventilate
the patient using an automatic ventilation mode.
CAUTION!
Risk of increased ambient gas concentration
Ambient air may become contaminated with
anesthetic agent when using non-rebreathing
systems.
Ensure sufficient ambient air circulation.
Monitoring
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Standard screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Selecting a default layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Modifying current layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Adjusting display brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
Displayed parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
CO2 concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157
O2 concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Anesthetic gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Airway pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Respiratory flow and volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
SpO2 concentration (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Loops (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Mini trends (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Virtual flow tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Gas supply module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Ventilation source module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
VT/PAW module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
ΔVT module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
The low flow wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Volumeter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Part Number: 9053586, 3rd edition
Overview
The Apollo has three basic screens for the display of
monitoring information: standard, data, and trend.
The gas measurement and gas delivery windows
remain displayed in all three screens, but the
information presented in the graphical/numerical
window will change, depending on the selected
screen and user configuration.
This chapter defines all monitoring parameters and
provides instructions for selecting and configuring
monitoring screens and parameters.
WARNING!
Risk of patient injury
If the display loses patient data, it is possible
that active monitoring is not being performed.
Close patient observation or alternate
monitoring de-vices should be used until
monitor function is restored.
Standard screen
The standard screen is automatically displayed
whenever a ventilation mode is selected.
Figure 82. Standard Screen
This screen can always be selected during operation: Gas Measurement Graphical/Numerical
Display Display
1. By pressing the > < key to display the
standard screen directly (1 in Figure 82).
or
Part Number: 9053586, 3rd edition
Screen layout
Selecting a default layout
The user can select a default screen layout for the Figure 83. Standard Screen - Screen Layout Window
standard screen. The selection determines the three
curves that are displayed, as well as the three 2
modules that are shown below the curves.
Three default layouts are available. They can only be
configured in the standby Configuration screen. See
the chapter “Configuration” for complete information. 1
1. Press the >Screen Layout< key on the standard
screen (1 in Figure 83). The screen layout
window appears (2 in Figure 83), with the
currently selected layout highlighted.
2. Turn the rotary knob to select a different layout,
and press the rotary knob to confirm.
The screen layout window is removed and the
standard screen is displayed with the selected
layout.
Displayed parameters
The following parameters are displayed on the
Apollo:
Note: The specific parameters that are displayed
Part Number: 9053586, 3rd edition
WARNING!
Risk of insufficient ventilation
The displayed spontaneous minute volume
(MVSPON) indicates the volumes of the patient's
breathing and the mechanical breathing
support. If the mechanical breathing support is
triggered by small tidal volumes of the patient, a
major portion of the spontaneous minute
volume is not achieved by the patient's
breathing but the mechanical breathing
support. In this case, MVSPON indicates a high
value although the actual spontaneous minute
volume is very low.
Do not make therapy decisions based solely on
the displayed value of MVSPON.
– Numerical display:
– SpO2 (functional O2 saturation level of blood)
– ♥ (pulse rate)
– Trend curve for SpO2 and pulse
Loops (Optional)
Graphical display for the following measured and
calculated values:
– Pressure/Volume
– Flow/Volume
See “Loops (Optional)” on page 168 for more
detailed information.
Automatic ventilation
VT/PAW module
Shows the tidal volume VT and airway pressure PAW
as bar graphs, as well as a numerical value for tidal
volume. See “Volumeter Module” on page 162 for
more information on the tidal volume graph.
ΔVT module
Shows the tidal volume VT and airway pressure PAW
as bar graphs, as well as a numerical value for
inspiratory tidal volume VTINSP, expiratory tidal
volume VT, and the difference between inspiratory
and expiratory tidal volume ΔVT. See “Volumeter
Module” on page 162 for more information on the
tidal volume graph.
WARNING!
Risk of patient injury
If used incorrectly, the reaction time of fresh-
gas concentration changes could increase,
increasing also the risk of undesirable soda
lime compounds.
This tool should not be used when higher flows
are required such as during induction,
emergence, or other times when rapid changes
to the concentration of gases in the circuit are
desired, or when the chemical pharmacology of
the agent being used indicates otherwise.
Bar graph
Indication Meaning
color
Gas measurement
The concentration of O2, CO2, and of the anesthetic
agents N2O, halothane, enflurane, isoflurane,
desflurane, and sevoflurane is measured.
Part Number: 9053586, 3rd edition
WARNING!
Risk of gas measurement failure
The presence of aerosols in the breathing
circuit should be avoided, as the displayed
agent concentrations and/or the water trap
membrane may be affected.
The presence of organic cleaning agents or
gases containing haloalkanes (e.g. CFC) will
impair the accuracy of the integrated gas
analyzer.
Calibration
The patient-gas measuring module is automatically Figure 86. Example of Calibration Messages
calibrated every time the device is started and then 1 2 3
regularly with ambient air, as long as the device is
switched on (see “Zeroing interval” on page 298).
During calibration, messages informing the user
about the calibration appear on the screen
(1 and 2 in Figure 86), and the measurement values
are replaced with the word CAL (3 in Figure 86).
CO2 and O2
The CO2 and O2 concentrations are side-stream
measured, thus delaying an indication of the real-
time values by approximately four seconds. The
displayed CO2 and O2 curves are not synchronized
with the pressure and flow curves.
If apnea occurs, the display for etCO2 is replaced by
the message apnea CO2. The apnea time [min:sec]
Anesthetic agents
The anesthetic agents are side-stream measured in
the same way as CO2 and O2.
Mixture detection
Apollo automatically detects the anesthetic gas used
and switches the measurement and monitoring of
anesthetic gas concentration to the gas detected.
If there is a mixture of two volatile anesthetic agents,
the concentration of the secondary anesthetic agent
is displayed if the xMAC value is 0.1 MAC or greater.
The gas with the higher expiratory xMAC value is
displayed above the secondary anesthetic agent.
A secondary anesthetic agent becomes the main
anesthetic agent if its xMAC value exceeds the MAC
value of the main anesthetic agent by 0.2 MAC.
A mixture of more than two volatile anesthetic agents
cannot be reliably detected. In this case, the
anesthetic gas values are no longer displayed and an
alarm is generated.
Note: A mixture of more than two agents may lead
to a temporary time-out of the measured O2
value.
WARNING!
Risk of patient injury
If the anesthetic agent displayed does not
match the label of the anesthetic vaporizer
being applied (open), make sure the vaporizer is
filled correctly.
Part Number: 9053586, 3rd edition
MAC definition
1 MAC is the anesthetic gas concentration in the
blood at 760 mmHg (1013 hPa), at which 50% of
patients no longer respond to a skin incision with
movement.
The integrated MAC algorithm is based on the MAC
values as indicated on the list. These values are
merely guideline values. It is the information on the
slip accompanying the anesthetic agents which is
binding.
Age (years)
Loops (Optional)
Display loops on the screen
1. Press the >Loops< button on the standard Figure 89. Loops Key on Standard Screen
screen (1 in Figure 89).
The label of the button changes to >Exit Loops< and
the P/V Loop and the Flow/Vol loop are displayed
with a value table instead of the two lower curves
(refer to Figure 90). Each loop remains on display for
three breathing cycles; the color intensity of the loop
decreases with each cycle.
1
The scale of the Pressure and Flow axes depends on
the scale selected for the real-time curves. The scale
of the volume axis depends on the scale in the VT/
PAW module.
In addition to the factory setting for the axis
orientation complying with ISO 80601-2-13, inverted
display is also available. To change the display
setting, contact service personnel.
See the chapter “Configuration” for more information
on scale configuration.
VT MV*CO2
ml 0
50 50
150 150
500 500
1000 1000
VT O2 uptake
ml
50 50 400
150 150
500 500
0
1000 1000
VT CPAT
ml
50 10
0
150 50
500 100
1000 100
Part Number: 9053586, 3rd edition
Datalog
The Datalog is a chronological record of measured Figure 95. Datalog Button - Standard Screen
values and events that occur during the case. The
Apollo automatically records events such as
performed or canceled tests, changes of agent, and
changes of ventilation mode (followed by the date).
At the end of each case, the case duration, the use of
each fresh gas, patient uptake, and total use of 1
anesthetic agent are recorded. The device also
records the time when the device was switched off.
The datalog can also be configured by the user to
record each WARNING or CAUTION or to record
patient parameter data at specified intervals of 1, 2,
5, or 10 minutes. Each WARNING and CAUTION
entry is followed by the measured values recorded at
the time of the alarm’s occurence. See the chapter
“Configuration” for more information.
The Datalog can be accessed during operation as
well as in Standby. It consists of two pages:
>page 1< lists standard patient parameters, and
>page 2< lists more standard parameters as well as
optional parameters, such as SpO2 and pulse.
To display the Datalog:
– Press the >Datalog< button on the standby
screen, data screen, or standard screen
(1 in Figure 95).
Page 1 of the Datalog is displayed (see Figure 96).
To display the second page:
– Press the >Page 2< button on the Datalog Figure 96. Page 1 of the Datalog
screen (1 in Figure 96)
Screen timer
The timer function allows the user to time an event Figure 97. Location of Start Timer Key - Standard Screen
using a button available in any operating mode.
To start the timer:
– Press the >Start timer< button (1 in Figure 97)
on the standard, data, or trend screen.
Data screen
1. Press the > < key repeatedly until the data Figure 100. Location of Next Key
screen appears (1 in Figure 100).
All numerical values are displayed on the data Figure 101. The Data Screen
screen with their units of measurement (see Figure
101).
It shows patient data for all monitored parameters,
including data for optional SpO2 and pulse (if
available). System compliance CSYS and leakage
LeakSYS, along with the date and time of the last leak
test, are displayed in the middle left part of the data
screen. The modules displayed below and the
numerical values are the same as those configured
for the standard screen (see “Screen layout” on
page 156).
Trend screen
Displays the measured values over an interval Figure 102. Location of Next Key
beginning with the start of the measurement.
The maximum storage time is eight hours, data older
than eight hours is erased.
It displays up to four graphical trends on the screen
at one time (see Figure 103). 1
Selecting other display combinations Figure 103. Trend Screen - Agent Trends Shown
data is available.
The trend for inspiratory and expiratory values is
represented by bar graphs. The expiratory value is
always indicated by a black line.
The trends for agents, N2O, and O2 are displayed
with the relevant color coding.
Zoom function
The zoom function allows the user to magnify a
portion of the trend display. It becomes available
after 30 minutes of trend data is collected.
The zoom window appears as a rectangle on the
trend. The rectangle can be moved by the user to
select the area to magnify.
Monitoring mode
The monitoring mode is a special mode that can be Figure 106. Standby screen
activated only in Standby and is used for specific
types of monitoring, e.g., sidestream CO2 monitoring
during supplemental O2 via nasal cannula with SpO2
monitoring. There is no fresh-gas delivery in
monitoring mode and the machine is not in any
ventilation mode.
The alarms that are active in monitoring mode are
comparable to those in Man/Spont mode (see the 1
chapter “Alarms” for more information.)
If fresh-gas flow is activated while in monitoring
mode, the system switches to Man/Spont mode.
2
standby screen (1 in Figure 106).
or
Vol. AF
Mode
2 1
Sensitivity
ESD
Accompanying
Documents
SpO2 sensor
Caution!
!
Consult
socket
IV System
SpO2
COM 2
COM 1
Part Number: 9053586, 3rd edition
Safety-relevant information
WARNING!
WARNING!
Risk of patient injury
Risk of electric shock
If the SpO2 sensor is used in the presence of
If the SpO2 sensor becomes damaged during intravascular dyes, such as methylene blue,
use, discontinue use, especially if there are measurement accuracy may be inaccurate.
uncovered electrical contacts.
Do not rely on measurement data if the SpO2
sensor is used under these conditions.
WARNING!
Risk of patient injury CAUTION!
Incorrectly positioned sensors may result in Risk of misleading data
incorrect measurements which may lead to
patient injury. Immersing the SpO2 sensor in liquid may lead to a
malfunction and thus misleading data.
Only use Nellcor sensors in the recommended
positions. Do not immerse the SpO2 sensor in liquid.
CAUTION! CAUTION!
Risk of patient injury Risk of failure or inaccurate data
The Apollo anesthesia machine has been verified to If positioned close to a bright light source, the pulse
function with Nellcor pulse oximeter probes. The signal may fail or the results may be inaccurate.
use of other probes may result in patient injury.
The sensor must be protected from exposure to
Use only verified probes with the Apollo. bright light (e.g. surgical lamps and direct sunlight).
CAUTION!
WARNING!
Risk of failure or inaccurate data
Risk of patient injury
High intrathoracic pressure, pressure on the If the sensor is positioned on limbs together with an
WARNING! CAUTION!
Risk of patient injury Risk of failure or inaccurate data
If the SpO2 sensor is used in the presence of Electrocautery can influence the measuring
significant concentrations of dyshemoglobins, accuracy.
such as carboxyhemoglobin or methemoglobin,
measurement accuracy may be reduced. Leads and the SpO2 sensor should be positioned as
far away from the electrocautery and its neutral
Do not rely on measurement data if the SpO2
sensor is used under these conditions. electrode as possible.
CAUTION!
Risk of inaccurate data
Sensor performance may be impaired and lead to
inaccurate results if the patient moves violently.
The sensor should be positioned at a quite/stable
site in order to reduce the risk of artifacts due to
movements.
CAUTION!
Risk of inaccurate data
If incorrectly used, the adhesive strips may cause
the pulse signal to fail. The adhesive strips must not
be stretched unduly.
Never use two adhesive strips as this can lead to
venous pulsation, causing the pulse signal to fail.
CAUTION!
Risk of inaccurate data
In the presence of shock, low blood pressure,
severe vasoconstriction, major anemia,
hypothermia, arterial occlusion proximal to
thesensor, and asystolia, the pulse signal may fail.
CAUTION!
Risk of inaccurate data
If simulators are used as calibrators, the SpO2
module may produce incorrect data.
Simulators must not be used as calibrators.
Alarms
Contents
Alarm priorities and alarm signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Advisory/technical message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Downgrading alarm priorities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
Setting the alarm volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
Alarms in Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
Alarm displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Suppressing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Silencing audible alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Disabling alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Alarm behaviour when changing ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
Limit-based alarms activated in respective ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Enabling/disabling alarms globally during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Enabling/disabling CO2 alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Enabling/disabling SpO2 alarms (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Bypass mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194
Displaying and setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195
To set an alarm limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
Auto-Set of alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
Part Number: 9053586, 3rd edition
Warning
– Message with highest priority.
– A warning message requires immediate action.
– Text flashes on red background.
– Red LED (2 in Figure 111) flashes, accompanied
by a repetitive tone sequence.
Caution
– Message with medium priority.
– A caution message requires immediate action.
– Text flashes on yellow background.
– Yellow LED (3 in Figure 111) flashes,
accompanied by a repetitive 3-tone sequence.
Advisory/technical message
– Message with lowest priority.
Part Number: 9053586, 3rd edition
Technical message
– Yellow LED (3 in Figure 111) illuminates
continuously without any acoustic tone.
– These messages must be noted and action
taken if necessary.
Dräger recommends the user to remain close to the
anesthesia machine, i.e. within a range of up to
12 feet (4 meters), to allow for quick recognition and
action in the event of an alarm.
Whenever an alarm message is displayed, the alarm
LED flashes or lights up continuously and an
acoustic tone sequence indicates the alarm priority
class.
In addition, a flashing help text is displayed in the
prompt field.
In the case of limit-based alarms, the corresponding
measured values will be highlighted by a colored
background (1 in Figure 111) and will flash.
The color of the background reflects the color-coding
of the alarm priority (red, yellow, cyan).
For a complete list of Apollo alarm messages, see
“Alarm - Cause - Remedy” on page 267.
Alarm displays
Alarm messages are displayed in the alarm message Figure 112. Multiple Alarms Display
field (1 in Figure 112) in order of priority. 1 2
All displayed alarms are sorted according to the three
classes defined on page 185. Within these classes,
the alarms are sorted and displayed according to an
internal priority system. A priority of 31 indicates the 3
highest, a priority of 1 the lowest priority. The priority
numbers are given in the table in the chapter “Alarm -
Cause - Remedy” on page 267. The internal priorities
are not displayed.
Up to three messages can be displayed
simultaneously. In some cases, the corresponding
measured values are highlighted on the screen by a
flashing background in addition to the alarm
message.
If more than three alarms occur simultaneously, the Figure 113. Additional Alarm Fields
more
symbol > < appears to the right of the alarm
message field (2 in Figure 112), and the button
>Show All Alarms< is activated on the right side of
the screen (3 in Figure 112).
When this button is pressed, the upper curve display
is replaced by up to six additional alarm message
fields for 15 seconds (see Figure 113).
If more than nine alarm messages are
simultaneously active, the lowest priority alarms will
not be displayed until the total number of active
alarm messages falls below nine.
Part Number: 9053586, 3rd edition
Suppressing alarms
Silencing audible alarms
The audible alarms can be silenced for 2 minutes. Figure 114. Alarm Indications
To silence all audible alarms: 2 1
Disabling alarms
Figure 115. Disabling alarms
CAUTION!
1 2
Risk of inadequate monitoring
National standards require a minimum monitoring
with some alarm functions. These standards may
not be met if the alarm function of the etCO2
monitoring parameter is disabled.
Only disable this monitoring parameter after
CAUTION!
Risk of inadequate monitoring
National standards require a minimum monitoring
with some alarm functions. These standards may
not be met if the alarm function of the SpO2
monitoring parameter is disabled.
Only disable this monitoring parameter after
consulting national standards.
Audio and visual alarms can be disabled by adjusting
the default alarm limits in the standby configuration
screen. Certain alarms can also be disabled
automatically, based on ventilation mode. See the
chapter “Configuration” for complete information.
WARNING!
Risk of patient injury
As anesthesia machines within one care area
might have different alarm limit configurations,
make sure that the preset alarm limits are
appropriate for the new patient. Also make sure
that the alarm system has not been rendered
useless by setting the alarm limits to extreme
values or by their alarm tone being disabled.
See “Configuring the default settings in
Standby” on page 201.
Mode Volume, Volume AF, Press. Support Aux CGO Monitoring, Factory setting
Alarm Pressure, Press. Support CPAP Man/Spont
SpO2 > ON ON ON ON ––
[%] < ON ON ON ON 92
Pulse > ON ON ON ON 120
[bpm] < ON ON ON
1)
ON 50
etCO2 > ON ON
1)
1) 50
[mmHg] < ON ON 1) ––
FiCO2 > ON ON 1) 1) 5
[mmHg]
MV > ON ON OFF 1) 12
[L/min] < ON ON OFF 1) 3.0
FiO2 > ON ON 1) 1) ––
[Vol.%] < ON ON ON ON 20
Mode Volume, Volume AF, Press. Support Aux CGO Monitoring, Factory setting
Alarm Pressure, Press. Support CPAP Man/Spont
inSev. > ON ON ON ON 4.2
[Vol.%] < ON ON 1) 1) ––
PAW > ON ON ON ON 40
[cmH2O] < ON ON OFF ON 8
APNEA PRESSURE ON ON OFF OFF 8
APNEA FLOW ON ON OFF OFF –––
Apnea CO2 ON ON ON2) ON2) –––
CAUTION!
Risk of inadequate monitoring
National standards require a minimum monitoring
with some alarm functions. These standards may
not be met if the alarm function of the etCO2
CAUTION!
Risk of inadequate monitoring
National standards require a minimum monitoring
with some alarm functions. These standards may
not be met if the alarm function of the SpO2
monitoring parameter is disabled.
Only disable this monitoring parameter after
consulting national standards.
Part Number: 9053586, 3rd edition
Bypass mode
The bypass mode permits patient monitoring without Figure 119. Bypass Mode Indication
unnecessary alarms during extra-corporal Bypass Mode Indication
oxygenation of the patient by a heart lung machine.
In the bypass mode:
– All gas concentrations are measured
independently of the breathing phase.
– CO2 apnea and pressure apnea alarms are
inactive.
– SpO2 monitoring alarms are inactive.
– The “MAC LOW?” alarm is inactive.
– MV alarms and flow apnea monitoring can be
configured (see page 209).
The bypass mode can be used in all active
ventilation modes.
Enabling/disabling bypass mode
1. Press the >Config< key (1 in Figure 117) on the
standard or data screen.
The submenu volumes/Alarms is opened (refer to
Figure 119).
2. Select and confirm the column “Alarms On/Off”
via the rotary knob.
3. Select and confirm the line “Bypass Mode” via
the rotary knob.
4. Select and confirm “On” or “Off” via the rotary
knob.
5. The bypass mode can also be deactivated by
pressing the button “Exit mode Bypass”.
CAUTION!
Risk of insufficient monitoring
If the setting MV alarms in bypass mode? is set to
no, MV alarms and flow apnea monitoring are
deactivated.
Special attention is required.
concerned.
For this purpose, the alarm limits menu can be
selected in Standby via the button >Alarm Limits<.
Configuration
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Configuring the default settings in Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
System settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
Breathing sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203
Pulse volume (for SpO2, optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Date/Time/Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
Gas monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
Optional parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
Interfaces datalog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
Datalog entries triggered by . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
COM PORT 1 MEDIBUS, COM PORT 2 MEDIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
MEDIBUS default configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207
Setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Default alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Default agent limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
Alarms in Man/Spont . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
Therapy related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 211
Part Number: 9053586, 3rd edition
Volumes/Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Pulse volume (for SpO2, optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Alarms on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Param Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Agent monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Datalog entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Datalog entries triggered by . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
System information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Trace 1, Trace 2, Trace 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Exit SysInfo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Exiting configuration during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Setting the patient’s age and weight during operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
Overview
The user can configure settings on the Apollo in
Standby mode as well as during operation. Standby
configuration allows the user to save a complete set
of defaults that are invoked automatically when the
machine is switched on (see “Configuring the default
settings in Standby” below). The configuration
settings that can be made during operation are more
limited and are valid only until the machine is
switched off (see “Configuration during operation” on
page 220).
System settings
The following settings can be selected in the System
Settings standby configuration screen:
Alarm volume
See 1 in Figure 125.
The minimum alarm volume can be set to a value
between 1 to 9. The standard alarm volume cannot
be set below this limit.
WARNING!
Risk of use error
If using features like the ’breathing sound’ or
when operating under loud ambient conditions,
the auditory alarm signals may not be heard.
Always set the alarm tone to a sufficiently loud
volume.
Breathing sound
See 2 in Figure 125.
0 = off
Part Number: 9053586, 3rd edition
to
9 = maximum volume
Factory setting: 0
The breathing sound is produced by the Breathing
Sound Emulator (BSE) module. This module
converts the measured inspiratory and expiratory
flow values into audible sounds similar to a breathing
sound.
The volume depends on the set patient age. This
dependency allows optimal volume at all flow levels.
All alarms are still audible with the breathing sound
volume set to the maximum.
Date/Time/Language
See 4 in Figure 125.
Date/Time
– day, month, year
– hh:mm
Time format
– 12 hrs.
– 24 hrs.
Factory setting: 12 hrs.
Language
Selects the language of the display text.
– English (US)
Factory setting: English (US)
Parameters
The following settings can be selected in the Figure 126. Parameters Standby Configuration Screen
Parameters standby configuration screen 1 2 3 4
(refer to Figure 126):
Scaling
– Tidal Volume
auto: A suitable scale is selected automatically in
accordance with the set age:
– <1 year: 50 mL
– 1 to 2 years: 150 mL
– >2 years to 10 years: 500 mL
– >10 years: 1000 mL
Units
See 2 in Figure 126.
Gas monitoring
See 3 in Figure 126.
Optional parameters
Part Number: 9053586, 3rd edition
Interfaces datalog
The following settings can be selected in the
interfaces logbook standby configuration screen.
Datalog entries triggered by Figure 127. Interfaces Logbook Standby Configuration
Screen
See 1 in Figure 127.
These settings determine when automatic entries will 1 2 3 4
be made in the Datalog.
Screen layout
The menu screen layout contains three default Figure 128. Screen Layout Standby Configuration Screen
layouts of the home screen: Layout 1, Layout 2, and
Layout 3.
The layouts comprising the following elements which
can be freely configured:
– Three curves with the associated numerical 1
Each curve/module can only be displayed once. If a Figure 129. Available Options
curve/module is selected twice, the preceding
selection automatically becomes “blank”.
To configure Layout 1, Layout 2, or Layout 3:
1. Select and confirm a layout via the rotary knob.
2. Select and confirm a curve or a module.
3. Change and confirm the selection via the rotary
knob.
1
CAUTION!
Risk of inadequate monitoring
Certain monitoring options are mandatory
depending on the applicable national requirements.
Some monitoring options may not be covered by
certain screen layout configurations.
Always take national standards into account when
configuring the screen layout.
Adjustment
Alarm Factory setting
range
SpO2 > 81 to 99, -- --
[%] < 80 to 98 92
Pulse > 21 to 250 120
[bpm] < 20 to 249 50
etCO2 > 1 to 75, -- 50
[mmHg] < 0 to 74, -- --
inCO2 < 1 to 10 5
[mmHg]
MV > 0.1 to 20.0, -- 12
[L/min] < 0 to 19.9, -- 3.0
FiO2 > 19 to 99, -- --
[Vol.%] < 18 to 98 20
PAW > 5 to 99 40
[cmH2O] < 0 to 35 8
Part Number: 9053586, 3rd edition
Adjustment
Alarm Factory setting
range
inHal. > 0.1 to 8.4 1.5
[Vol.%] < 0 to 8.3, -- --
inIso. > 0.1 to 8.4 2.3
[Vol.%] < 0 to 8.3, -- --
inEnf. > 0.1 to 9.9 3.4
[Vol.%] < 0 to 9.8, -- --
inDes. > 0.1 to 21.9 12.0
[Vol.%] < 0 to 21.8, -- --
inSev. > 0.1 to 9.9 4.2
[Vol.%] < 0 to 9.8, -- --
Two dashes > < indicate that the corresponding
alarm is disabled.
Note: The new default alarm limits are effective
whenever the anesthesia machine is
switched on and after selecting >Restore
default settings< in Standby.
Alarms in Man/Spont
Therapy related
See 1 in Figure 131.
CAUTION!
Risk of insufficient monitoring
If the setting MV alarms in Bypass mode? is set to
no, MV alarms and flow apnea monitoring are
deactivated.
Special attention is required.
Device related
See 2 in Figure 131.
Other
See 3 in Figure 131.
– APNEA PRESSURE
– APNEA FLOW
In this case, the flow apnea alarm cannot be
configured to be a low-priority alarm (see page 222).
If set to no, these alarms will be raised with normal
alarm cascade behavior.
WARNING!
Risk of device failure
The anesthesia machine does not operate
without at least one oxygen supply.
Either the O2 pipeline supply or the O2 cylinder
Part Number: 9053586, 3rd edition
– Preset configuration
Select, edit, and confirm the VT to be changed
via the rotary knob.
The settings for VT are interpolated for weights
between the four predetermined classes.
System information
The system information standby configuration Figure 134. System Settings Standby Configuration
screens display useful information and allow the Screen: General Information Page
activation of software options by an authorized
DrägerService representative. It also allows access
to the Remote Service Box (see Figure 134).
General Information
– Software Versions of the individual components
– Enabled Software Options
– Gas Consumption and sampling rate of the
patient gas module
– Operating Hours of individual components
Activate Option
Software options can be activated by entering a Figure 135. Activate Option Screen
multi-digit code.
Options and the associated activation codes are
available from the respective Dräger sales
organization.
To activate an option:
1. Select and confirm the figures successively from
the line displayed via the rotary knob (refer to
Figure 135).
2. When the complete code is entered, activate,
select, and confirm the menu item via the rotary
knob.
Remote Service
An inspection for the technical status of the device
can take place by using Remote Service.
Figure 136. Remote Service key
Before activating the Remote Service:
1. Carry out a self test.
WARNING!
Risk of patient injury
The patient may be injured if connected to the
1
device when the remote service function is
active.
Only use the Remote Service Link on medical
devices which are not otherwise in use.
Volumes/Alarms
The following settings can be selected in the Figure 140. Volumes/Alarms Configuration Screen
Volumes/Alarms configuration screen
(refer to Figure 140).
Alarm volume
4 = minimum volume (>45 dB(A))
to
9 = maximum volume (<75 dB(A))
WARNING!
Risk of use error
If using features like the ’breathing sound’ or
when operating under loud ambient conditions,
the auditory alarm signals may not be heard.
Always set the alarm tone to a sufficiently loud
volume.
Breathing sound
0 = off
Part Number: 9053586, 3rd edition
to
9 = maximum volume
The breathing sound is produced by the Breathing
Sound Emulator (BSE) module. This module
converts the measured inspiratory and expiratory
flow values into audible sounds similar to a breathing
sound.
The sound volume depends on the set patient age.
This dependency allows optimal volume at all flow
levels.
All alarms are still audible with the breathing sound
volume set to the maximum.
Alarms on/off
This setting is used to enable/disable CO2 alarms,
optional SpO2 alarms, and bypass mode (for further
information see page 191 and page 193).
The flow apnea alarm can be configured to be a low-
priority alarm (technical alarm). This setting remains
active until switching to standby and starting a new
case.
This configuration is only possible when the setting
Immediate alarm upon apnea? (patients < 6
years) is set no (see page 213).
Param Settings
The following settings can be selected in the Param Figure 141. Param Settings Configuration Screen
Settings configuration screen (refer to Figure 141).
Scaling
– CO2
– PAW
– Flow
– O2
The setting is made automatically or by selecting
a pre-set scale.
auto: Automatic adjustment to the next higher or
lower scale after two passes if the scaling frame
Agent monitoring
– MAC display: Yes or No
This setting determines whether the MAC factor
is displayed or not.
– Age corrected: Yes or No
This setting determines whether the MAC factor
is corrected for patient age or not. See page 166
for a detailed explanation of the MAC definition
and calculation.
Datalog entries
The following settings can be selected in the Datalog Figure 142. Datalog Entries Configuration Screen
Entries configuration screen (refer to Figure 142).
System information
Part Number: 9053586, 3rd edition
The system info configuration screens display the Figure 143. System Info Configuration Screen
following information:
General information
– Software Versions of the individual components
– Software Options
– Gas Consumption and sampling rate of the
patient gas module
– Operating Hours of individual components
Trace 1, Trace 2, Trace 3
Description of internal equipment states and
parameters.
Exit SysInfo
1. Press the >Exit SysInfo< to exit the System
Information menu and return to the configuration
screen.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Disassembling components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Observe before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Removing the sample line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227
Removing the water trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
Removing the patient system hoses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
Removing the absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
Reusable absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
Disposable CLIC absorber (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
Removing the breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
Removing the ventilator diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Removing the flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Opening the breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232
Removing the anesthetic gas receiving system AGS (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
Removing the passive scavenger system (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234
Removing the endotracheal aspiration system (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234
Cleaning and Disinfection Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
Proper Cleaning/Disinfection Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
Cleaning/Disinfection Objective and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
Disinfecting/Cleaning/Sterilizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
Choice of disinfectant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
Part Number: 9053586, 3rd edition
Overview
This chapter provides complete instructions for the
disassembly and cleaning of the Apollo anesthesia
machine.
Note: Set the Apollo to Standby before
disassembly.
Disassembling components
Observe before disassembly
Disconnecting from the mains
Switch off the device and accessory devices and
remove their mains plugs.
CAUTION!
Risk of gas measurement failure and device failure
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
Sample gas lines are single-use articles and must
be replaced, not disinfected.
WARNING!
Risk of measurement failure and device failure
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced at least every
four weeks.
CAUTION!
Risk of component damage and patient injury
If mishandled, the spiral ribbing on the breathing
hoses can become detached from the sleeve.
Breathing hoses with damaged spiral ribbing can
easily be kinked and interrupt the flow of gas!
When attaching or removing the breathing hoses,
always hold them by the connection sleeve and not
by the spiral ribbing! Always check the breathing
hoses for damage prior to use. Damaged breathing
hoses must be replaced.
CAUTION!
Risk of flow measurement failure 2
Disinfecting or cleaning the flow sensors by 1
machine will damage them and cause the flow
measurement to fail.
Disinfect and clean the flow sensors as described in 2
Part Number: 9053586, 3rd edition
CAUTION!
Risk of flow measurement failure
Sterilizing the Spirolog flow sensors in high-
temperature steam will damage them and cause the
flow measurement to fail.
Disinfect and clean the Spirolog flow sensor as
described in the Instructions for Use of the Spirolog
and SpiroLife flow sensors.
1 1
1
1 1
3. Lift off the metal valve plate (1 in Figure 155). Figure 155. Removing the Valve Plate
4. Prepare the housing parts for conditioning in a
cleaning and disinfection machine.
5. Place the metal valve plate in the cleaning and
disinfection machine.
1
7
Note the Instructions for Use of the anesthetic gas
receiving system AGS.
WARNING!
Risk of infection
Fluids gathered in the suction bottle may be
infectious and therefore dangerous for hospital
personnel.
Always wear gloves when emptying the suction
bottle.
Proper Cleaning/Disinfection
Sequence
If equipment parts are to be cleaned by hand, they
must always be disinfected first for personal
protection.
When equipment parts are to be cleaned and
disinfected by machine, they should always be
cleaned first and then disinfected.
Cleaning/Disinfection Objective
and Methods
The purpose of the described cleaning/disinfection
measures is to provide each patient with a reliably
disinfected anesthetic machine, i.e. a machine free
from all unhealthy micro-organisms. Sterility is
required only for the intubation tube and bronchial
aspiration catheter.
The following disinfection methods may be used:
– Disinfection by wiping the surface of the device
(see “Choice of disinfectant” on page 237)
– Mechanical cleaning with high-temperature
disinfection (≥199 °F (93 °C) , ≥10 min). This is
the preferred method for many Apollo
components; a suitable cleaning agent must be
added.
– Manual disinfection in a bath. Extensive personal
protection is required due to inhalation of vapors.
Equipment parts should be cleaned and disinfected
by machine for hygienic conditioning purposes
(EN ISO 15883, in preparation).
Complex, thermally stable functional components,
such as the breathing system, can be easily cleaned
and disinfected by machine, but are not always dried
sufficiently. Subsequent vacuum disinfection in hot
steam or steam sterilization is recommended to dry
off the remaining water.
– Vacuum disinfection in hot steam at 167 °F
(75 °C) for 20 minutes or 221 °F (105 °C) for
1 minute, for example.
– Steam sterilization, for example at 250 °F
Disinfecting/Cleaning/Sterilizing
Choice of disinfectant
Only products suitable for surface disinfection may
be used for disinfection. Testing was performed with
Incidin Extra N and Incidur (wiping), and Gigasept FF
and Korsolex Extra (disinfection by immersion).
Surfaces
Surfaces of the Apollo, compressed gas hoses, and
cables:
WARNING!
Risk of electric shock
Part Number: 9053586, 3rd edition
CAUTION!
Risk of failure of flow measurement
Improper reprocessing and soiling, such as
deposits or particles, can damage the flow sensor.
– No machine cleaning or disinfection
– No plasma sterilization or radiation sterilization
– No water jets, compressed air, brushes or the
like
– No ultrasonic bath
– No hot-steam sterilization with Spirolog and
Infinity ID flow sensors
– Clean and disinfect the flow sensor in
accordance with the corresponding instructions
for use.
– For disinfecting the flow sensor use only clean
disinfectant solutions.
CAUTION!
Risk of device failure
If inappropriate substances are used for hygienic
preparation, the device and its components may be
damaged (corrosion, condensation).
Apollo and its components must not be treated with
formaldehyde vapors or ethylene oxide.
Reassembling components
WARNING!
Risk of scavenger becoming blocked
If objects such as packing foil get into the
device, e.g., the breathing system or the
ventilator drawer, the scavenger may become
blocked.
Make sure that there is no packing material left
inside the device.
Visual inspection
CAUTION!
Risk of faulty components
Even accessories designed to be reused and
WARNING!
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits that
were not removed during reprocessing can
ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
Part Number: 9053586, 3rd edition
1. Insert the flow sensors (1 in Figure 161) into the Figure 161. Installing the Flow Sensors
two port connections on the breathing system,
with the electric connection on each sensor
facing down in the slot.
2. Orient the inspiratory and expiratory ports
(2 in Figure 161) so that the key on each port
lines up with the slot. Install the ports and tighten 1
them by turning clockwise. 2
Note: Flow sensors must be recalibrated after
replacement by performing the power-on self
test (see chapter “Pre-use Checkout”). 1
2
Installing the ventilator diaphragm
1. Insert the ventilator diaphragm so that the Dräger Figure 162. Inserting the Ventilator Diaphragm
legend is visible (1 in Figure 162).
1
WARNING! 1
Risk of injury
Absorbent is caustic and is a strong eye, skin,
and respiratory tract irritant.
3
Use care when handling the absorbent to avoid
spills.
CAUTION! 2
Risk of device failure
It is recommended that Drägersorb 800 + or
Drägersorb Free is used. Do not use powdered
soda lime, as a higher dust load may impair
functionality of the Apollo anesthesia machine.
2
4. Slide the new disposable absorber onto the
mounting (2 in Figure 165).
5. Push the absorber into the anesthesia machine
until it engages.
6. Reset the soda lime change log to current date
by pressing the >soda lime changed< button,
see page 97.
WARNING!
Risk of patient injury
The soda lime loses humidity. Generally, if the
humidity falls below a minimum set point, the
following undesirable reactions can occur,
independent of the type of lime and the
inhalation anesthetic being used:
– reduced CO2 absorption;
– increased heat build-up in the absorber and
thus, an increased breathing gas
temperature;
– formation of CO;
– absorption and/or decomposition of the
inhalation anesthetic.
WARNING!
Risk of burns
Conductive breathing hoses or face masks may
cause burns during HF surgery. 3 2
Do not use these types of hoses and masks in
combination with HF surgery.
WARNING! 1
Risk of patient injury
If the breathing hoses are incorrectly
connected, the patient might be inadequately
ventilated and supplied with fresh gas.
Make sure that all breathing hoses are correctly
connected to the breathing system.
1
Part Number: 9053586, 3rd edition
CAUTION!
2
Risk of increased ambient gas concentration
Ambient air may become contaminated with 1
anesthetic agent if the scavenger hoses are 4
3
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
WARNING!
Risk of patient injury
If the side openings of the receiving system are
blocked, negative pressure may result in the
breathing system and the patient’s lungs.
Always make sure the side openings of the
receiving system are not blocked.
CAUTION!
Risk of increased ambient gas concentration
Ambient air may become contaminated with
anesthetic agent if the scavenger hoses are
functionally inhibited.
The scavenger hoses must not be pinched, kinked,
or blocked in any manner.
Maintenance
Overview
This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
WARNING!
Risk of infection
Users and service personnel can become
infected with pathogenic germs.
Disinfect and clean device or device parts
before any maintenance measures and also
before returning the medical device for repair.
WARNING!
Risk of electric shock
Current-carrying components are located under
the cover.
Do not remove the cover.
Maintenance measures must be performed by
the personnel responsible. Dräger recommends
DrägerService to perform these measures.
Inspection
Perform inspections at regular intervals and observe the fol-
lowing specifications.
Preventive maintenance
WARNING!
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the proper operation of all
components, this device must undergo
inspection and preventive maintenance at
specified intervals.
WARNING!
Risk of electric shock
Before performing any maintenance work,
disconnect all electrical connectors and gas
connectors from power supply and gas supply.
Repair
Part Number: 9053586, 3rd edition
WARNING!
Risk of gas measurement failure and device
failure
If alcohol or cleaning agents/disinfectants come
in contact with the inside of the water trap, they
can damage the diaphragm and the
measurement system.
Do not use these substances and do not wash,
flush, or sterilize the water trap.
CAUTION!
Risk of misleading data
Aerosols can damage the diaphragm and the
measurement system may fail as a result.
Do not use aerosols in the breathing system. The
water trap must not be used in combination with a
medical nebulizer.
WARNING!
Risk of gas measurement failure and device
failure
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced at least every
four weeks.
WARNING!
Risk of infection
The device and its components must be
disinfected and cleaned before disposal!
WARNING!
Risk of explosion and of chemical burns
Improper handling of batteries can result in
explosions and chemical burns.
Do not throw batteries into fire. Do not force
batteries open.
Troubleshooting
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259
Gas failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260
Ventilator failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 262
Fresh-gas delivery failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263
Ventilator and fresh-gas delivery failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264
Gas measurement failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264
Display failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
User interface failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
System failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
Alarm - Cause - Remedy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267
Part Number: 9053586, 3rd edition
Overview
This chapter discusses several types of failure that
may occur on the Apollo and provides courses of
action following the failure. An alphabetical list of all
Apollo alarms and their causes and remedies is
provided on page 267.
Note: If the remedies suggested in this chapter do
not resolve a fault that may impair the proper
functioning of the Apollo, use another device.
WARNING!
Risk of patient injury
If all power supplies fail, the screen display will
be dark and automatic ventilation will cease.
The patient must be ventilated manually.
Gas failure
If the gas supply fails, the Apollo displays a Figure 175. NO O2 SUPPLY Alarm Message
corresponding message in the status field at the top 1
of the screen (1 in Figure 175):
NO AIR SUPPLY, NO N2O SUPPLY, or
NO O2 SUPPLY
1. Open the valve on the corresponding backup
cylinder at the back of the machine
(see Figure 176).
2. Restore the pipeline supply.
CAUTION!
Risk of patient injury
In case of a gas supply failure, adjust the gas
concentration manually.
Note the contra-indications for 100 % O2.
WARNING!
Risk of patient awareness
If a complete gas supply failure occurs, further
operation may continue by supplying the
anesthesia machine with ambient air.
Anesthetic agents will no longer be delivered
and the inspiratory gas composition will be
diluted thereby raising the issue of patient
awareness.
Therefore carefully monitor the gas mixture
and, if necessary, use IV anesthetics.
WARNING!
Risk of gas supply contamination
When the central gas supply is connected, the
Part Number: 9053586, 3rd edition
CAUTION!
Risk of increased ambient gas concentrations
If the breathing bag is not attached, expiratory
anesthetic agents can escape from the breathing
system.
Make sure the breathing bag is attached and
ensure sufficient ambient air circulation.
WARNING!
Risk of gas supply failure
If the valves remain open when connected to
the pipeline gas supply, gas may be withdrawn
from the backup gas cylinders.
Close cylinder valves whenever the pipeline gas
supply is sufficient.
Ventilator failure
If the ventilator fails, the following message is Figure 177. Ventilator Failure Messages
displayed in the status field on the screen 3 1
(1 in Figure 177):
VENTILATOR FAIL
The ventilation buttons are removed from the screen
and a prompt appears advising the user how to
proceed (2 in Figure 177):
“Ventilator failure!
Only manual ventilation available.”
The machine automatically switches to Man/Spont
mode (3 in Figure 177).
WARNING!
Risk of patient injury 2
WARNING!
Risk of patient injury
If pressure or volume monitoring fails, the
patient cannot be adequately monitored.
Ensure adequate substitute monitoring.
WARNING!
Risk of patient injury
If the fresh-gas delivery fails, the anesthesia
machine automatically discontinues the fresh-
gas flow. An O2 flow must be delivered to the
patient.
Check vaporizer setting and set the O2 flow to a
sufficient level.
WARNING!
Display failure
If the screen display fails:
1. Switch off the machine.
2. Set O2 flow to the desired level using the total
flow meter.
3. Check the vaporizer setting.
4. Ventilate the patient manually.
5. Ensure adequate substitute monitoring.
System failure
If the system no longer responds to an action:
1. Ventilate the patient by hand.
2. Switch the machine off and on again.
3. Cancel the self test.
WARNING!
Risk of patient injury
If the breathing bag does not fill with fresh gas,
the patient cannot be adequately ventilated.
Check the oxygen supply, open cylinder valves
if necessary.
Advisory AGENT SENSOR FAIL Anesthetic gas measurement system has Use external gas measuring system.
(1) (in Standby only) failed. Call DrägerService.
Advisory/ AIR CYLIND. CONNECT?* Pressure sensor of backup cylinder not Check pressure sensor connection.
– connected.
(8/–)
Caution/ AIR CYLIND. EMPTY* Backup Air cylinder empty and central Air Use a new backup Air cylinder.
Advisory supply not available or not connected. Use the pipeline supply.
(24/7)
Caution AIR FLOW MEAS. FAIL Fresh-gas flow measurement for Air has Use only oxygen as fresh gas and
(11) failed. observe total flow meter.
Check fresh-gas flow settings using total
flow meter and set a fresh-gas flow
greater than or equal to measured or set
minute volume.
Call DrägerService.
Caution
(24)
0 to 30 sec.
Leak or blockage in tube or hose system. Check hose system, tube, and microbial
filter.
Patient not connected. Connect patient correctly.
Caution/ APNEA VENTILATION No spontaneous breathing efforts by the Check the patient's trigger capability.
Part Number: 9053586, 3rd edition
Caution (24) =
insp. gas concentration > upper alarm
limit for 31 to 180 seconds.
Warning (31) =
Caution (24) =
insp. gas concentration > upper alarm
limit for 31 to 180 seconds.
Warning (31) =
insp. gas concentration > upper alarm
limit for >180 seconds.
Optional Air cylinder is empty or closed. Connect a full Air cylinder or open the
cylinder valve.
Compressed Air compressor has failed. Check compressor.
Warning/ NO FRESH GAS* This alarm can be deactivated by trained
Advisory service personnel at customer request.
(26/9)
No fresh-gas flow set. Open flow control valves and adjust flow.
Warning/ NO N2O SUPPLY* N2O supply has failed. Open N2O backup cylinder.
Advisory Check pipeline supply.
(25/10)
Pipeline supply hose not connected or Check connection to pipeline supply.
kinked.
N2O cylinder empty or closed. Connect a full N2O cylinder or open the
cylinder valve.
“ “ displayed beside The apnea alarm for the measured value For more details see page 190.
measured values concerned has been disabled.
Specifications
Contents
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283
Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284
Monitor screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284
Life span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284
Operating data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285
Protection class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
UMDNS Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
GMDN code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
Fresh-gas delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
Auxiliary O2 flow meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288
Breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288
Auxiliary common gas outlet (CGO) (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291
Measuring systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 292
Frequency measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295
Gas measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 296
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300
Latex use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 301
Schematic diagram of alarm tones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
Part Number: 9053586, 3rd edition
Specifications
General information
Ambient conditions
Depending on the type of anesthetic agent delivery
unit used, this data may vary.
During operation
Temperature1) 59 °F to 104 °F
(max. 95 °F for charging the battery)
During storage/transportation
Temperature1) –4 °F to 140 °F
Battery: 5 °F to 104 °F2)
1) Depending on the type of anesthetic agent delivery unit used, this data may vary.
2) The longer-term storage at a temperature outside this range may shorten the life of the batteries.
Dimensions
(Variations may occur depending on the configuration)
Machine dimensions (W x H x D) approx. 31.5 x 59.1 x 31.5 in. (85 x 150 x 80 cm)
Breathing system dimensions (W x H x D) approx. 14.8 x 15.9 x 13.6 in. (37.5 x 40.5 x 34.5 cm)
Weight
Various attached parts (e.g., baskets, flexible approx. 22 lbs (10 kg)
breathing bag holder, park holder for vaporizer,
cylinder pressure reducers, gas cylinder holder,
halogen light)
Monitor screen
Flat screen, color, TFT, 12.1” diagonal,
800 x 600 pixels
Life span
10 years
Operating data
Power input 200 W typically, max. 1.5 kW with power drawn from auxiliary outlets
Standby 180 W
Typical 200 W
Internal battery
Supply pressure for O2, N2O, and Air 39 psi to 100 psi
(2.7 to 6.9 kPa x 100)
Noise emissions from device Free field measurements complying with ISO 3744
Sound pressure level L(A) of the alarm The alarm volume is adjustable, refer to “Alarm volume” on page 221.
tones at the workstation, measured
according to IEC 60601-1-8 Edition 2.1:
Protection class
Classification
UMDNS Code
GMDN code
Fresh-gas delivery
Settings:
Fresh-gas flow, delivery 0 to min. 12 L/min per gas (O2, N2O, Air)
Fresh-gas flow, electrical measuring 0 to 12 L/min volumetric flow per gas (O2, N2O, Air)1)
Standardized to ATPD conditions.
1) The maximal permissible fresh-gas flow may be limited depending on the anesthetic agent vaporizer used.
Accuracy ±10% of the max. displayed value for 50% O2 and 50% N2O
calibrated at 5 L/min
Breathing system
Absorber volume
Volume 0.13 L
Total system leakage (as per ISO 8835-2) <150 mL/min at 30 cmH2O
Standardized to BTPS conditions.
Accuracy between 5 and 15 L/min ±15% of set value or ±3 cmH2O, whichever is greater
Pressure drop at 30 L/min 2.8 cmH2O (in Spont position; wet and dry)
As per ISO 8835-2, dry, max. ±6 cmH2O, -4.4 cmH2O 4.2 cmH2O -4.6 cmH2O 4.2 cmH2O
with hose set for adults M30146
As per ISO 8835-2, dry, sole breathing -3.0 cmH2O 3.0 cmH2O -3.3 cmH2O 3.3 cmH2O
system without patient hoses
Ventilator
(electronically controlled, electrically driven piston ventilator, fresh gas decoupled)
Ventilation modes Volume Mode, Pressure Mode, Volume AF (optional), CPAP (optional),
and Pressure Support mode (optional), synchronized volume and
pressure modes
Settings:
Accuracy (5 mL to 150 mL) ±10% of set value or ±10 mL, whichever is greater
Accuracy (over 150 mL) ±5% of set value or ±15 mL, whichever is greater
FrequencyMIN
Part Number: 9053586, 3rd edition
1) Due to gas measurement sampling, leakage (both at the patient and in the device), and resistance/compliance of the patient and the
breathing circuit, the maximum delivered tidal volume may be limited.
PEEP
ΔPPS
TSLOPE
1) Due to gas measurement sampling and leaks (both at the patient and in the device), the end-expiratory PEEP value may be lower than specified at the end of
long expiratory phases.
Measuring systems
Respiratory pressure
Cylinder pressure
Part Number: 9053586, 3rd edition
Pressure Measurement
(pressure indicator, dial type)
Flow measurement
(hot wire anemometry)
All values apply under ATPD conditions:
Flow
Tidal volume VT
Range 0 mL to 9999 mL
Delta VT
Range 0 mL to 9999 mL
Volume VTINSP
Range 0 mL to 9999 mL
Compliance CPAT
MVLEAK
MVMAND
MVSPON
O2 Uptake
MVSPON
Frequency measurement
Frequency (Freq.)
Gas measurement
Time after switch-on until the specified accuracy Less than 500 s
is attained
Maximum time until emptying of the water trap is 41 h (sample gas under BTPS
necessary conditions, ambient air 23 °C
(73.4 °F))
Sampling rate1) 150 mL/min ±20 mL/min 200 mL/min ±20 mL/min
Delay for sampling (typical value, depends on less than 4 seconds less than 4 seconds
sample line)
gas measurement module with consumption-free not applicable less than 500 ms
O2 measuring
Response time t10..90 CO2 less than 500 ms less than 350 ms
Response time t10..90 N2O less than 500 ms less than 500 ms
1) The respective value depends on the patient-gas measurement module (PGM) used which is displayed on the System Information page.
Enflurane 0 to 10 Vol.%
Sevoflurane 0 to 10 Vol.%
Desflurane 0 to 20 Vol.%
MAC (xMAC)
Range 0 to 9.9
Secondary agent At no later than 0.4 Vol.%1); becomes primary agent if expiratory xMAC is
more than 0.2 MAC above former primary agent.
Cross-sensitivity None referred to alcohol (<3000 ppm), acetone (<1000 ppm), methane,
water vapor, NO, and CO
Drift of measurement accuracy Compensated by cyclic zeroing. Zeroing is performed automatically and
with ambient air. This means there is minimal change to the gas
concentrations in the breathing circuit.
Zeroing interval
Maximum time until emptying of the water trap is 41 h (sample gas under BTPS conditions, at 73 °F (23 °C) ambient
necessary temperature)
Fresh-gas consumption per case 0 L to 9999 L per gas (O2, N2O, Air) (O2 value not including gas used for
the O2 flush and the auxiliary O2 flow meter)
Resolution 1L
Total agent consumption per case (liquid agent) 0 mL to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane)
Resolution 1 mL
Agent consumption due to patient uptake per 0 mL to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
case (liquid agent) Sevoflurane, Desflurane)
Resolution 1 mL
1) Exception: If a measured desflurane concentration of at least 4Vol.% is present, a mixed agent identification is available as soon as the measured concentration
of the secondary agent reaches at least 10% of the desflurane concentration.
2) The respective value depends on the patient-gas measurement module (PGM) used which is displayed on System Information page.
Accuracy ±3 bpm
Sensors
Acoustic pulse signal A tone is generated for each pulse detected. The pitch of the tone
proportional to the oxygen saturation. Increasing saturation increases the
pitch.
The displayed plethysmogram is a relative indicator of the pulse amplitude. Its scale is not absolute and it is only used to
judge the quality of the SpO2 measurement.
Interfaces
Only connect devices that meet the requirements of IEC 60950-1 for ungrounded SELV circuits and the requirements of
IEC 60601-1 (as of the 2nd edition) for touchable secondary circuits with a maximum nominal voltage of 24 VDC.
Pin allocation
2 TX transmit
3 RX receive
5 GND ground
9 NC not connected
Shields DTR and DSR, as well as RTS and CTS are internally connected.
Hardware handshake is not supported.
Dräger Base IV system (not sold in the U.S.) Power supply for IV systems
Latex use
Apollo is not made with natural rubber latex.
Part Number: 9053586, 3rd edition
Warning Yes
Depending on the overall alarm situation, this tone sequence may
be played as a 5-tone sequence due to the timing of the individual
alarms.
Caution Yes
Note No
Backup speaker
The alarm volume for the backup speaker cannot be
adjusted. The backup speaker issues simplified
acoustic alarm signals. The pitch of the tone
sequence is the same, but the intervals are different:
EMC declaration
General information
The EMC compliance of the product has been evaluated
with the external cables, transducers, and accessories
specified in the list of accessories. Other accessories which
do not affect EMC compliance may be used if no other rea-
sons forbid their use (see other sections of the instructions
for use). The use of noncompliant accessories can result in
increased emissions or decreased immunity of the medical
device.
The medical device may only be used adjacent to or
stacked with other devices when the configuration is ap-
proved by Dräger. When use adjacent to or stacked with
other devices is absolutely necessary without the configura-
tion being approved by Dräger, the correct operation of the
device in this configuration must be tested before the prod-
uct is used. In any case, strictly observe the instructions for
use of the other devices.
Electromagnetic emissions
Electromagnetic environment
The medical device is intended for use in an electromag-
netic environment as specified below. The user must ensure
that the medical device is used in such an environment.
Electromagnetic immunity
The medical device is intended for use in an electromag-
netic environment as specified below. The user must ensure
that the medical device is used in such an environment.
Electrical fast transients Power supply lines: ±2 kV ±2 kV Mains voltage quality should be that of a
/ bursts Longer input / output lines: ±1 kV typical commercial or hospital
(IEC 61000-4-4) ±1 kV environment.
Device combinations
This device can be operated in combination with
other Dräger devices or with devices from other
manufacturers. Observe the accompanying
documents for the individual devices.
If a device combination is not approved by Dräger,
the safety and the functional state of the individual
devices can be compromised. The operating
organization must ensure that the device
combination complies with the applicable editions of
the relevant standards for medical devices.
Device combinations approved by Dräger meet the
requirements of the following standards:
– IEC 60601-1, 3rd edition (general requirements
for safety, device combinations, software-
controlled functions)
– IEC 60601-1-2 (electromagnetic compatibility)
– IEC 60601-1-8 (alarm systems)
Or:
– IEC 60601-1, 2nd edition (general requirements
for safety)
– IEC 60601-1-1 (device combinations)
– IEC 60601-1-2 (electromagnetic compatibility)
– IEC 60601-1-4 (software-controlled functions)
– IEC 60601-1-8 (alarm systems)
Connections to IT-networks
Data can be exchanged across an IT-network by
using hard-wired and wireless technologies. An
Serial ports
The following interfaces are supported:
– RS232 interfaces conforming to EIA RS-232
(CCITT V.24/V.28) for the following applications:
– MEDIBUS, MEDIBUS.X
– Connections with third party medical devices
Consequences of using an inappropriate network
Part Number: 9053586, 3rd edition
Relevant standards
In addition to the standards listed here, the medical device also complies with various other standards, e.g.,
standards concerning special national requirements.
IEC 60601-1 2nd ed. Part 1:
Medical electrical equipment Requirements for safety
IEC 60601-1-2 Part 1-2:
Medical electrical equipment General requirements for safety,
Collateral standard: Electromagnetic compatibility; Requirements and tests
ISO 9919 Particular requirements for basic safety and essential performance of pulse
Medical electrical equipment oximeter equipment for medical use
ISO 21647 Particular requirements for basic safety and essential performance of respi-
Medical electrical equipment ratory gas monitors
For devices from production date July 2015 the following also applies:
IEC 60601-1 3rd ed. Part 1:
Medical electrical equipment General requirements for basic safety and essential performance
IEC 60601-1-2 Part 1-2:
Medical electrical equipment General requirements for safety - Collateral standard: Electromagnetic
compatibility - Requirements and tests
As of 2015-08:
Dräger Medical GmbH
changes to
Drägerwerk AG & Co. KGaA
As of 2016-05:
Draeger Medical, Inc.
changes to
Draeger, Inc.