BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS
UNIT-4
                                            2 MARKS
1.   Mention the different types of electrodes used in biomedical field. (Sep 2020)
     There are three main types of electrodes used in biomedical field
         Microelectrodes.
         Depth and Needle electrodes
         Body Surface electrodes
2.   What are polarizable and non polarizable electrodes? (Sep 2020)
         A polarizable electrode is an electrode in an electrochemical cell that is
        characterized by charge separation at the electrode-electrolyte boundary.
         A non-polarizable electrode is an electrode in an electrochemical cell that can be
        characterized by no charge separation at the electrode-electrolyte boundary.
3.   What are the reasons for artifacts? (May 2019)
          Artifacts include tools, clothing, and decorations made by people. They provide
     essential clues for researchers studying ancient cultures material remains of a culture,
     such as tools, clothing, or food to expose by digging.
4.   Mention the need of an electrode gel. (May 2019, May 2015)
          Electrolyte gel is designed to provide carbohydrate for fast energy and
     key electrolytes for hydration. GO Energy + Electrolyte gels are particularly effective
     when exercising in hot conditions causing sweat rates to be high. Available in Lemon &
     Mint, Salted Caramel or Raspberry flavours, for a fresh or fruity taste.
5.   Define half cell potential. (Nov 2018, May 2018)
            Half-cell potential refers to the potential developed at the electrode of each half
     cell in an electrochemical cell. In an electrochemical cell, the overall potential is the
     total potential calculated from the potentials of two half cells.
6.   What are perfectly polarizable electrodes? Give an example. (Nov 2018, May 2018)
            Electrodes in which no actual charges cross the electrode-electrolyte interface
     when current is applied.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 1
     E.g. Platinum electrode
7.   What are motion artifacts. (Nov 2017, May 2016)
               Motion artifact is a patient-based artifact that occurs with voluntary or
     involuntary patient movement during image acquisition. Misregistration artifacts, which
     appear as blurring, streaking, or shading, are caused by patient movement during a CT
     scan.
8.   What are the types of surface electrodes. (Nov 2017, May 2015)
     There are five types of surface electrodes
        Metal Plate electrodes
        Suction cup electrodes
        Adhesive tape electrodes
        Multi point electrodes
        Floating electrodes.
9.   Define polarization. (May 2017)
             Polarization is the change of potential from a stabilized state, e.g. from the open-
     circuit electrode potential as the result of the passage of current. Polarization is
     sometimes also referred to as "over voltage" or "over potential".
10. List out the biomedical applications of silver chloride electrodes. (May 2017)
             Silver chloride electrodes are also used by many applications of biological
     electrode systems such as bio monitoring sensors as part of
        Electrocardiography (ECG)
        Electroencephalography (EEG)
        Transcutaneous electrical nerve stimulation (TENS) to deliver current.
11. Write down the method of insertion in needle electrode. (Nov 2016)
             The Needle Electrode insertion site is on the inner leg, approximately 3 finger
     widths (2”) away from the center of the ankle and one finger width towards the back of
     the leg. Place 3 fingers on the center of the ankle bone. Move the top finger one finger
     width towards the back of the leg.
12. Define microelectrodes. (Nov 2016)
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 2
               A microelectrode is    an     electrode   used    in electrophysiology either   for
      recording neural signals or for the electrical stimulation of nervous tissue.
13.    List the electrodes used in ECG and EEG. (May 2016)
         ECG
          Flat paper thin sticker electrode
          Self-adhesive circular pad electrode
         EEG
          Traditional Wet Ag/AgCl Electrodes
          Active Dry Single Gold Pin-Based Electrodes
          Hybrid Dry Multiple Spikes-Based Electrodes
          Passive Dry Solid-Gel Based Electrodes
14. Write a note on Einthoven triangle. (Nov 2015)
             Einthoven's triangle is an imaginary formation of three limb leads in
      a triangle used in electrocardiography, formed by the two shoulders and the pubis. The
      shape forms an inverted equilateral triangle with the heart at the center. It is named after
      Willem Einthoven, who theorized its existence.
15.    Define electrical double layer. (Nov 2015)
               A double layer (DL, also called an electrical double layer, EDL) is a structure
      that appears on the surface of an object when it is exposed to a fluid. The object might be
      a solid particle, a gas bubble, a liquid droplet, or a porous body.
                                            11 MARKS
1. Mention the electrodes used in biomedical application and explain the various
  electrodes with neat sketch. (Sep 2020, May 2019, Nov 2018, May 2018, Nov 2017,
  Nov 2016, May 2016, May 2015, Nov 2015)
  There are three main types of electrodes:
          Microelectrodes.
          Depth and Needle electrodes.
          Surface electrodes.
Microelectrodes:
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                              Page 3
A microelectrode is an electrode which is used to measure bio electric potential near or
within the cell.
There are two types of microelectrodes
        Metallic microelectrode
        Non-Metallic microelectrode
Metallic microelectrode:
        Metal Microelectrodes are made up of fine tungsten or stainless steel of 0.5 to 5
        microns cross section.
        They are made by electrolytic itching.
        They are coated almost up to the tip by insulating material.
        To reduce impedance, micro tip is chloride.
Non-Metallic microelectrode:
        It consists of glass micropipette of diameter 1 micrometer.
        Micropipette filled with electrolyte solution that is compatible with cellular fluids
        is used.
        Stem of Micropipette has a thin flexible wire made out of chloride silver,
        stainless steel or tungsten.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                          Page 4
Depth electrode:
       Depth electrodes are placed through a small hole using stereotactic navigation to
      target locations identified through high-resolution imaging.
       Common targets include the mesial temporal lobe, inter hemispheric fissure, and
      per ventricular gray matter.
Needle electrode:
       A fine wire through which electrical current may flow when attached to a power
      source; used to carry high frequency electrical currents that create heat or destroy
      diseased tissue (called radiofrequency ablation) or seal blood vessels.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                      Page 5
Surface electrodes:
        Surface electrodes are those which are placed in contact with the skin of the subject
       in order to obtain bioelectric potentials from the surface.
        Electrode can be used to sense ECG, EEG, EMG etc.
There are five types of surface electrodes
        Metal Plate electrodes
        Suction cup electrodes
        Adhesive tape electrodes
        Multi point electrodes
        Floating electrodes
  1. Metal plate electrode:
        One of the most frequently used forms of biopotential sensing electrodes is
       the metal-plate electrode.
        In its basic form, it consists of a metallic conductor in contact with the skin.
        An electrolyte gel is used to establish and maintain the contact.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                             Page 6
 2. Suction cup electrode:
       To measure ECG from various positions on the chest, Suction cup electrodes are
      used.
       It suits well to attach electrodes on flat surface of the body and on soft tissue
      regions.
       They have a good contact surface. Physically they are large but the skin contacts
      only the electrode rim.
       It has high contact impedance. They have a plastic syringe barrel, suction tube and
      cables.
       Recently, due to infection and cleaning procedures, these electrodes are not used.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                        Page 7
 3. Adhesive tape electrode:
       In the surface electrode, the pressure of surface electrode against the skin squeezes
      out the electrode paste.
       To avoid this problem, adhesive electrodes are used. It has a lightweight metallic
      screen.
       They have a pad at behind for placing electrode paste. This adhesive backing hold
      the electrode on place and tight.
       It also helps to avoid evaporation of electrolyte present in the electrode paste.
 4. Multipoint electrode:
       Multipoint electrodes are very practical electrode setup for ECG measurements. It
      has more than 1000 active contact points.
       This helps to establish low resistance contact with the human. Under any
      environmental condition, doctors can use multipoint electrode.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 8
 5. Floating electrode:
       In metal plate or limb electrodes, the major disadvantage is the movement errors.
      Motion artefact occurs due to the motion at the interface between electrode and
      electrolyte.
       The interface gets stabilized using Floating electrodes. The floating electrodes do
      not contact the human subject directly.
       They contact the subject via electrolytic paste or jelly. The advantage of this type is
      the mechanical stability.
2. Brief on (May 2019)
  (a) Electrode tissue interface
  (b) Polarization
a) ELECTRODE TISSUE INTERFACE:
       The most commonly used electrodes in patient monitoring and related studies are
      surface electrodes.
       The notable examples are when they are used for recording ECG, EEG and
      respiratory activity by impedance pneumography.
       In order to avoid movement artefacts and to obtain clearly established contact (low
      contact impedance) an electrolyte or electrode paste is usually employed as an
      interface between the electrode and the surface of the source of the event.
      Metal - Electrolyte interface:
       The metal - Electrolyte interface appears to consist of a voltage source in series
      with a parallel combination of a capacitance and reaction resistance. The voltage
      developed is called the half-cell potential.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 9
              Metal                Ionic symbol                 Electrode potential
     Aluminium                     AI+++                        –1.66V
     Iron                          Fe++                         –0.44V
                                        ++
     Lead                          Pb                           –0.12V
     Hydrogen                      H+                               0
     Copper                        C++                          +0.34V
     Silver                         Ag+                        +0.80V
     Platinum                      Pt+                          +1.2V
     Gold                          Au+                          +1.69V
                Table: Electrode Potential for some Metals with Respect to Hydrogen
       This table shows that the electrode potentials are appreciable when dissimilar
      metals are used.
       They also exist, though of smaller magnitude, even if electrodes of similar
      materials are employed.
       The lowest potential has been observed to be in the silver-silver chloride
      electrodes.
       The values of the capacitance and the resistance depend upon many factors which
      include the current density, temperature, type and concentration of the electrolyte and
      the type of metal used.
       The difference in half-cell potentials that exists between two electrodes is also
      called ‘offset potential’.
       The differential amplifiers used to measure potentials between two electrodes are
      generally designed to cancel the electrode offset potential so that only the signals of
      interest are recorded.
       The electrode offset potential produced between electrodes may be unstable and
      unpredictable.
       The long-term change in this potential appears as baseline drift and short-term
      changes as noise on the recorded trace.
       If electrodes are used with ac-coupled amplifiers, the long term drift may be
      partially rejected by the low frequency characteristics of the amplifier.
       But it will depend upon the rate of change of electrode offset potential in relation to
      the ac-coupling time constant in the amplifier.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 10
       For example, if the electrode offset potential drift rate is 1 mV/s, satisfactory
      results can only be obtained if the low frequency response of the amplifier is1Hz.
      Electrolyte-Skin Interface:
       An approximation of the electrolyte-skin interface can be had by assuming that the
      skin acts as a diaphragm arranged between two solutions (electrolyte and body fluids)
      of different concentrations containing the same ions, which is bound to give potential
      differences.
       The simplest equivalent representation could then be described as a voltage source
      in series with a parallel combination of a capacitance and resistance.
       The capacitance represents the charge developed at the phase boundary whereas
      the resistance depends upon the conditions associated with ion-migration along the
      phase boundaries and inside the diaphragm.
       The above discussion shows that there is a possibility of the presence of voltages of
      non-physiological origin. These voltages are called contact potentials.
       The electrical equivalent circuit of the surface electrode suggests that the voltage
      presented to the measuring instrument from the electrode consists of two main
      components.
       One is the contact potential and the other is the biological signal of interest. The
      contact potential depends upon several factors and may produce an interference signal
      which exceeds several times the useful signal.
       The contact potential is found to be a function of the type of skin, skin preparation
      and composition of the electrolyte.
       When bioelectric events are recorded, interference signals are produced by the
      potential differences of metal-electrolyte and the electrolyte-skin interface.
       Normally, these potential differences are connected in opposition during the
      recording procedure, and in the case of a truly reversible and uniform electrode pair,
      their difference would be nil.
       However, in practice, a difference of potential may be extremely small and is found
      to exist between electrodes produced even under conditions of utmost care during
      manufacture.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 11
       Also, some of the elements in the equivalent circuit are time-dependent and are
      bound to show slow variations with time.
   b) POLARIZATION:
       If a low voltage is applied to two electrodes placed in a solution, the electrical
      double layers are disturbed.
       Depending on the metals constituting the electrodes, a steady flow of current may
      or may not take place.
       In some metal / liquid interfaces, the electrical double layer gets temporarily
      disturbed by the externally applied voltage, and therefore, very small current flows
      after the first surge, thus indicating a high resistance.
       This type of electrode will not permit the measurement of steady or slowly varying
      potentials in the tissues.
       They are said to, be polarized or non reversible. Thus, the phenomenon of
      polarization affects the electro-chemical double layer on the electrode surface and
      manifests itself in changing the value of the impedance and voltage source
      representing the transition layer.
       Parsons (1964) stated that electrodes in which no net transfer of charge takes place
      across the metal-electrolyte interface can be termed as perfectly polarized.
       Those in which unhindered exchange of charge is possible are called non-
      polarizable or reversible electrodes.
       The ionic double layer in metals of these electrodes is such that they allow
      considerable current to flow when a small voltage is applied, thus offering a low
      resistance.
       Non-polarizing electrodes on the other hand, are designed to rapidly dissipate any
      charge imbalance induced by powerful electrical discharges such as a defibrillation
      procedure.
       Depolarization enables the immediate reappearance of bioelectric signals on the
      monitor after defibrillation.
       For this reason, non-polarizing electrodes have become the electrodes of choice for
      monitoring in the intensive care units and stress testing procedures.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                          Page 12
       Although polarizable electrodes are becoming less common, they are still in use.
       They usually employ stainless steel and are used for resting ECGs or other
      situations where there is small likelihood that the electrodes would be exposed to a
      large pulse of energy (such as a defibrillation discharge) in which case they would
      retain a residual charge, become polarized, and will no longer transmit the relatively
      small bioelectric signals, thus becoming useless.
       The choice of metals for electrodes is not determined only by their susceptibility to
      polarization, but other factors such as mechanical properties, skin irritation or skin
      staining, etc. have also to be taken into consideration.
   3. How metal microelectrodes are formed and explain its electrical equivalent ciruit
      with neat diagram. (Sep 2020)
Metal Microelectrodes:
       Metal electrodes with very fine tips used for recording from single cells have the
      advantage over glass micropipetes of being relatively robust.
       Steel microelectrodes can be made from ordinary darning needles but preferably
      they should be of good stainless steel wire.
       They can be easily made up to 10 m diameter but great care has to be taken for
      diameters as small as 1 m.
       These very small tips are not very satisfactory as they are extremely brittle and
      have very high input impedance.
       Hubel (1957) described a method to make tungsten microelectrodes with a tip
      diameter of 0.4 m.
       He used electropointing technique which consists in etching a metal rod while the
      metal rod is slowly withdrawn from the etching solution, thus forming a tapered tip on
      the end of the rod.
       The etched metal is then dipped into an insulating solution for placing insulation on
      all but the tip.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                         Page 13
       This arrangement offers lower impedance than the microcapillary electrode,
      infinite shelf life and reproduciable performance, with ease of cleaning and
      maintenance.
       The metal—electrolyte interface is between the metal film and the electrolyte of
      the cell.
       Skrzypek and Keller (1975) illustrated a new method of manufacturing tungsten
      microelectrode permitting close control of microelectrode para- meters.
       In this technique, the tips are dc electroetched to diameters below 500° A and
      completely covered with polymethyl methacrylate.
       An electron beam from a scanning electron microscope is then used to expose a
      precise area on the tip for later removal by chemical methods.
       Recording results with these electrodes suggested good desirable recording
      characteristics, i.e. ability to isolate and hold single cells.
       Tungsten is preferred for constructing micro-electrodes due to its mechanical
      strength and its apparent inertness.
       Although tungsten itself is reactive, a surface layer of tungsten oxide will, in most
      situations, protect the metal against corrosion.
       The electrical properties of tungsten microelectrodes made with a taper of the tip of
      about 1:10 and insulated with lacquer leaving a tip length of about 10–100 μm were
      studied by Zeuthen (1978).
       The resting potential in saline was found to be –0.3 V relative to a silver-silver
      chloride reference electrode for input currents less than 10–12 A.
       Metallic electrodes are formed from a fine needle of a suitable metal drawn to a
      fine tip.
       On the other hand, glass electrodes are drawn from Pyrex glass of special grade.
       The metal microelectrodes are used in direct contact with the biological tissue and,
      therefore, have a lower resistance. However, they polarize with smaller amplifier
      input currents.
       The glass microelectrode has a substantial current carrying capacity because of the
      large surface contact area between the metal and the electrolyte.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                         Page 14
Equivalent circuit:
       Equivalent circuit model and methods. (A) Equivalent circuit model of a metal
       microelectrode in the brain adapted from Robinson (1968). The entire circuit is
       comprised of the electrode in the brain and the amplifier with a filter.
       The effective impedance of the electrode (Z e ) is comprised of the resistance of the
       electrolyte solution (R s ), the resistance and capacitance at the double layer interface
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 15
      of the electrolyte and the un insulated electrode tip (R e and C e ) and the (negligible)
      resistance of the metal electrode (R m ).
       The effective input impedance of the amplifier (Z a ) is comprised of the input
      impedance of the head-stage amplifier (Z a ) and the shunt resistance and capacitance
      to ground from the tip of the electrode to the input of the amplifier (R sh and C sh ).
       Two aluminium plates were connected and separated from each by non-conducting
      plastic supports, shown here from a top and side view.
       The apparatus was immersed in dilute saline with voltage signals applied to the
      signal plate with an electrode suspended from above 3 mm away.
   4. Draw the equivalent diagram of an electrode tissue interface and explain the
      need of electrode jellies. (Nov 2018, May 2018)
       Warburg      (1899)   in   his   pioneering    studies   discovered    that   a   single
      electrode/electrolyte interface can be represented by a series capacitance C and
      resistance R as shown in Fig.
       However, C and R are unlike real capacitors and resistors because their values are
      frequency and current- density dependent. Often, these components are called the
      polarization capacitance and resistance.
       The consequence of this relationship is that the phase angle      is constant at /4 for
      all frequencies.
       However, only a limited number of studies have tested the accuracy of the
      Warburg model (Ragheb and Geddes, 1990).
       It has been observed that the Warburg series RC equivalent does not adequately
      represent the behaviour of an electrode/electrolyte interface as this equivalent does
      not truly account for the very low-frequency behaviour of the interface.
       It is well known that such an interface can pass direct current. Therefore, a
      resistance Rf placed in parallel with the Warburg equivalent is more appropriate.
       The below figure shows this equivalent circuit in which Rf represents the Faradic
      leakage resistance. The value of Rf is high in the low-frequency region and is
      dependent on current density, increasing with a increase in current density.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 16
       a) Equivalent circuit model of electrode–tissue interface, with interface reactance
      represented by a single capacitor CP, RP charge transfer resistance, RS
      Tissue/electrolyte resistance.
       b) Equivalent circuit model of electrode–tissue interface, with interface reactance
      represented by a constant phase element (ZCPA ).
       c) Illustration of experimental setup. Ag/AgCl reference electrode, Platinum core
      of the concentric bipolar electrode, and Stainless steel outer shell of the bipolar
      electrode were connected to the potentiostat as the reference, working, and counter
      electrode, respectively.
       d) Illustration of the geometry and dimension of the bipolar electrode tip.
ELECTRODE JELLIES:
       Conducting creams and jellies have for long been used to facilitate a more intimate
      contact between the subject’s skin and the recording electrodes.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 17
       The outer horny layer of the skin is responsible for the bulk of the skin contact
      impedance, and for this reason careful skin preparation is essential in order to obtain
      the best results.
       The recording site should first be cleaned with an ether-meth mixture. In addition
      to having good electrical conductivity, the electrode jelly must have a particular
      chloride ion concentration (about 1%) close to the physiological chloride
      concentration.
       This is primarily important for long-term monitoring because it should not produce
      a harmful diffusion between the jelly and the body.
       It is to be particularly ensured that the jelly chosen is of a bland nature and does not
      contain soap or phenol which can produce a marked irritation of the skin after a few
      hours.
       The electrical conductivity of different makes of electrode cream can be measured
      (Hill and Khandpur, 1969) by means of the Schering ac bridge circuit.
       The cream is placed in a perspex conductivity cell of known dimensions and the
      resistive component of the cell impedance is measured at 10 Hz, the conductivity
      being calculated from the cell dimensions.
       The contact impedance of the skin depends upon the type of electrolyte used and
      the time (Trimby, 1976)
       A low concentration sodium chloride electrolyte has 0.5% NaCI and a high
      concentration electrolyte has a concentration in the range of 5 to 10% NaCI.
       The impedance is found to fall rapidly to 40% between 7 to 30 min. Stabilization
      occurs at about 30 to 45 min.
       An interesting observation from this figure is that while pre-rubbing the skin will
      lower the initial impedance value, the final value after using a high concentration
      electrolyte becomes nearly the same.
       Electrode jelly can be replaced in certain cases by using a conducting plastic as an
      interface between the electrode and the surface of the body.
   5. Explain the construction and working of Ag/AgCl electrode. (May 2017, Nov
      2016, May 2016)
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 18
       The silver/silver chloride or Ag/AgCl reference electrode is many electrochemists'
       reference electrode of choice. It is easily and cheaply prepared. It is stable, and quite
       robust.
       It is sometimes referred to as "SSCE" (Silver/Silver Chloride Electrode) but that
       abbreviation has been used for Sodium Saturated Calomel Electrode also.
Construction:
       The figure to the right shows an easily constructed Ag/AgCl reference
       electrode. The body of the electrode is made from 4 mm glass tube.
       A porous glass is available in 4 mm diameter rod and serves as the ionicly
       conducting electrical pathway between the inside of the reference electrode and the
       bulk of your cell.
       It has low electrical resistance (under 10 kohm for the common filling solutions)
       and a modest leak rate.
       The electrical resistance of the reference electrode 'frit' is an important factor in
       determining the stability and speed of your potentiostat in actual use.
       The leak rate may be important because of possible contamination of your solution
       by the reference electrode filling solution and vice versa.
Production of Silver-Silver Chloride Electrodes:
       Silver-silver chloride electrodes are normally prepared by electrolysis. Two silver
       discs are suspended in a saline solution.
       The positive pole of a dc supply is connected to the disc to be chlorided and the
       negative pole goes to the other disc.
       A current at the rate of 1 mA/cm2 of surface area is passed through the electrode
       for several minutes.
       A layer of silver chloride is thus deposited on the surface of the anode. The
       chemical changes that take place at the anode and cathode respectively are:
       To prepare silver-silver chloride electrodes of good quality, only pure silver should
       be used and the saline solution should be made from analar grade sodium chloride.
       Before chloriding, silver must be cleaned—preferably by the electrolytic method.
       They demonstrated that the impedance was different for different layers of chloride
       and that there is an optimum chloriding, which gives the lowest impedance.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 19
       A layer of silver chloride is thus deposited on the surface of the anode. The
      chemical changes that take place at the anode and cathode respectively are:
              NaCl     = Na+ + Cl–
              Cl– + Ag+ = AgCl
       The positively charged sodium ions generate hydrogen when they reach the
      cathode surface.
              2Na+ + 2H2O + 2 electrons = 2NaOH + H2
       They concluded that the lowest electrode-electrolyte impedance in the frequency
      range of 10 Hz to 10 kHz was found to occur with a chloride deposit ranging between
      100 and 500 mAs/cm2 of electrode area.
       Higher values may be used with a corresponding reduction in time to achieve the
      100–500 mAs/cm2 chloride deposit.
       The use of a chloride deposit in excess of this range did not alter the resistive
      nature of the electrode-electrolyte impedance although it increased its magnitude.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                         Page 20
       Geddes (1972) confirmed that an optimal coating of silver chloride applied to a
      silver electrode minimizes the electrical impedance.
       This is supported by Getzel and Webster (1976) who concluded that silver chloride
      may be applied to cleaned silver electrodes in the amount of 1050–1350 mA s/cm2 in
      order to reduce the impedance of the electrodes.
       However, to further reduce the impedance of the electrodes, they should be coated
      with at least 2000 mAs/cm2 of silver chloride followed by immersion in a
      photographic developer for 3 minutes.
       Grubbs and Worley (1983) obtained a lower and more stable impedance electrode
      by placing a heavier initial chloride coat on an etched silver electrode, and then
      electrolytically removing a portion of that coat.
       The electrode functions as a redox electrode and the equilibrium is between
      the silver metal (Ag) and its salt—silver chloride (AgCl, also called silver(I)
      chloride).
      The corresponding half-reactions can be presented as follows:
                      Ag+ + e- ----- Ag(s)
                      AgCl(s) + e- ------ Ag(s) + Cl-
      or can be written together:
                       AgCl(s) + Ag(s) + e- --------- Ag(s) + e- + Ag+ + Cl-
      which can be simplified:
                      AgCl(s) ------- Ag+ + Cl-
       This reaction is characterized by fast electrode kinetics, meaning that a sufficiently
      high current can be passed through the electrode with 100% efficiency of
      the redox reaction.
       The reaction has been proven to obey these equations in solutions of pH values
         between 0 and 13.5.
       The Nernst equation below shows the dependence of the potential of the silver-
      silver chloride electrode on the activity or effective concentration of chloride-ions:
                             E = E0 – RT/F ln(Cl-)
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 21
       The standard electrode potential E0 against standard hydrogen electrode (SHE) is
       0.230 V ± 10 mV. The potential is however very sensitive to traces of bromide ions
       which make it more negative.
Application of Ag/AgCl
       An Ag/AgCl electrode used as a corrosion sensor in a reinforced concrete structure
       is considered as having good application prospect.
6. Describe about the placement of electrodes for ECG and EEG. (May 2017, May 2015)
Placement of electrodes for ECG:
3-electrode system
        Uses 3 electrodes (RA, LA and LL)
       Monitor displays the bipolar leads (I, II and III)
       To get best results – Place electrodes on the chest wall equidistant from the heart
       (rather than the specific limbs)
5-electrode system
        Uses 5 electrodes (RA, RL, LA, LL and Chest)
       Monitor displays the bipolar leads (I, II and III)
       AND a single unipolar lead (depending on position of the brown chest lead
       (positions V1–6))
12-lead ECG
        10 electrodes required to produce 12-lead ECG
       4 Electrodes on all 4 limbs (RA, LL, LA, RL)6 Electrodes on precordium (V1–6)
       Monitors 12 leads (V1–6), (I, II, III) and (aVR, aVF, aVL)
       Allows interpretation of specific areas of the heart
       Inferior (II, III, aVF)Lateral (I, aVL, V5, V6)Anterior (V1–4)
12-lead Precordial lead placement
        V1: 4th intercostal space (ICS), RIGHT margin of the sternum
       V2: 4th ICS along the LEFT margin of the sternum
       V4: 5th ICS, mid-clavicular line
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                       Page 22
        V3: midway between V2 and V4
        V5: 5th ICS, anterior axillary line (same level as V4)
        V6: 5th ICS, mid-axillary line (same level as V4)
Additional Lead placements
Right sided ECG electrode placement
There are several approaches to recording a right-sided ECG:
        A complete set of right-sided leads is obtained by placing leads V1-6 in a mirror-
       image position on the right side of the chest (see diagram, below).
        It can be simpler to leave V1 and V2 in their usual positions and just transfer leads
       V3-6 to the right side of the chest (i.e. V3R to V6R).
        The most useful lead is V4R, which is obtained by placing the V4 electrode in the
       5th right intercostal space in the mid-clavicular line.
        ST elevation in V4R has a sensitivity of 88%, specificity of 78% and diagnostic
       accuracy of 83% in the diagnosis of RV MI.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                          Page 23
V4R ECG lead placement
      Erhardt et al first described the use of a right sided precordial lead (CR4R or V4R)
       in the diagnosis of right ventricular infarction which had previously been thought to
       be electrocardiographically silent.
        Single right-sided precordial lead in the diagnosis of right ventricular involvement
       in inferior myocardial infarction.
Posterior leads
Leads V7-9 are placed on the posterior chest wall in the following positions:
        V7 – Left posterior axillary line, in the same horizontal plane as V6.
        V8 – Tip of the left scapula, in the same horizontal plane as V6.
        V9 – Left paraspinal region, in the same horizontal plane as V6.
Lewis lead(S5-lead)
Lewis lead placement
        Right Arm (RA) electrode on manubrium
        Left Arm (LA) electrode over 5th ICS, right sternal border.
        Left Leg (LL) electrode over right lower costal margin.
Fontaine leads:
Fontaine bipolar precordial leads (F-ECG) are used to increase the sensitivity of epsilon
wave detection. Named after French cardiologist and electrophysiolgist Guy Hugues
Fontaine (1936-2018). Leads are placed as shown:
        Right Arm (RA) over the manubrium;
        Left Arm (LA) over the xiphoid process;
        and Left Leg (LL) in the standard V4 position (5th ICS MCL).
Instead of regular leads I, II, and III there are now three bipolar chest leads that are termed FI,
FII, and FIII which record the potentials developed in the right ventricle, from the
infundibulum to the diaphragm.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                              Page 24
The vertical bipolar lead FI, (similar to aVF) magnifies the atrial potentials and can be used
to record:
        Epsilon waves;
        Search for AV dissociation in ventricular tachycardia;
        And to study abnormal atrial rhythms when the P waves are too small on regular
       leads.
Placement of electrodes for EEG:
       The Fp2, F8, T4, T6, and O2 electrodes are placed at intervals of 5%, 10%, 10%,
       10%, 10%, and 5%, respectively, measured above the right ear, from front (Fpz) to
       back (Oz).
        The same is done for the odd-numbered electrodes on the left side, to complete the
       full circumference.
10–20 system (EEG)
        The 10–20 system or International 10–20 system is an internationally recognized
       method to describe and apply the location of scalp electrodes in the context of
       an EEG exam, polysomnograph sleep study, or voluntary lab research.
        This method was developed to maintain standardized testing methods ensuring that
       a subject's study outcomes (clinical or research) could be compiled, reproduced, and
       effectively analyzed and compared using the scientific method.
        The system is based on the relationship between the location of an electrode and
       the underlying area of the brain, specifically the cerebral cortex.
        During sleep and wake cycles, the brain produces different, objectively recognized
       and distinguishable electrical patterns, this can be detected by electrodes on the skin.
        The "10" and "20" refer to the fact that the actual distances between adjacent
       electrodes are either 10% or 20% of the total front–back or right–left distance of the
       skull. For example, a measurement is taken across the top of the head, from
       the nasion to inion.
        Most other common measurements ('landmarking methods') start at one ear and
       end at the other, normally over the top of the head. Specific anatomical locations of
       the ear used include the tragus, the auricle and the mastoid.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 25
Electrode labeling:
       Each electrode placement site has a letter to identify the lobe, or area of the brain it
      is reading from: pre-frontal (Fp), frontal (F), temporal (T), parietal (P), occipital (O),
      and central (C).
       Note that there is no "central lobe"; due to their placement, and depending on the
      individual, the "C" electrodes can exhibit/represent EEG activity more typical of
      frontal, temporal, and some parietal-occipital activity, and are always utilized in
      polysomnography sleep studies for the purpose of determining stages of sleep.
       There are also (Z) sites: A "Z" (zero) refers to an electrode placed on
      the midline sagittal plane of the skull, (FpZ, Fz, Cz, Oz) and is present mostly for
      reference/measurement points.
       These electrodes will not necessarily reflect or amplify lateral hemispheric cortical
      activity as they are placed over the corpus callosum, and do not represent either
      hemisphere adequately.
       Even-numbered electrodes (2,4,6,8) refer to electrode placement on the right side
      of the head, whereas odd numbers (1,3,5,7) refer to those on the left; this applies to
      both EEG and EOG (electrooculogram measurements of eyes) electrodes, as well as
      ECG (electrocardiography measurements of the heart) electrode placement.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                           Page 26
       Chin, or EMG (electromyogram) electrodes are more commonly just referred to
      with "right," "left," and "reference," or "common," as there are usually only three
      placed, and they can be differentially referenced from the EEG and EOG reference
      sites.
       The "A" (sometimes referred to as "M" for mastoid process) refers to the
      prominent bone process usually found just behind the outer ear (less prominent in
      children and some adults). In basic polysomnography, F3, F4, Fz, Cz, C3, C4, O1,
      O2, A1, A2 (M1, M2), are used.
       Cz and Fz are 'ground' or 'common' reference points for all EEG and EOG
      electrodes, and A1-A2 are used for contralateral referencing of all EEG electrodes.
       This EEG montage may be extended to utilize T3-T4, P3-P4, as well as others, if
      an extended or "seizure montage" is called for.
Measurement:
       Specific anatomical landmarks are used for the essential measuring and positioning
      of the EEG electrodes. These are found with a tape measure, and often marked with a
      grease pencil, or "China marker."
       Nasion to Inion: the nasion is the distinctly depressed area between the eyes, just
      above the bridge of the nose, and the inion, is the crest point of back of the skull,
      often indicated by a bump (the prominent occipital ridge, can usually be located with
      mild palpation). Marks for the Z electrodes are made between these points along the
      midline, at intervals of 10%, 20%, 20%, 20%, 20% and 10%.
       Preauricular to preauricular (or tragus to tragus: the tragus refers to the small
      portion of cartilage projecting anteriorly to the pinna).
       The preauricural point is in front of each ear, and can be more easily located with
      mild palpation, and if necessary, requesting patient to open mouth slightly.
       The T3, C3, Cz, C4, and T4 electrodes are placed at marks made at intervals of
      10%, 20%, 20%, 20%, 20% and 10%, respectively, measured across the top of the
      head.
       Skull circumference is measured just above the ears (T3 and T4), just above the
      bridge of the nose (at Fpz), and just above the occipital point (at Oz). The Fp2, F8,
      T4, T6, and O2 electrodes are placed at intervals of 5%, 10%, 10%, 10%, 10%, and
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                          Page 27
       5%, respectively, measured above the right ear, from front (Fpz) to back (Oz). The
       same is done for the odd-numbered electrodes on the left side, to complete the full
       circumference.
        Measurement methods for placement of the F3, F4, P3, and P4 points differ. If
       measured front-to-back (Fp1-F3-C3-P3-O1 and Fp2-F4-C4-P4-O2 montages), they
       can be 25% "up" from the front and back points (Fp1, Fp2, O1, and O2).
        If measured side-to-side (F7-F3-Fz-F4-F8 and T5-P3-Pz-P4-T6 montages), they
       can be 25% "up" from the side points (F7, F8, T5, and T6).
        If measured diagonally, from Nasion to Inion through the C3 and C4 points, they
       will be 20% in front of and behind the C3 and C4 points. Each of these measurement
       methods results in different nominal electrode placements.
7. Write in detail notes on electrical conductivity of jellies and its types.(May 2017)
ELECTRICAL CONDUCTIVITY OF ELECTRODE JELLIES:
        Conducting creams and jellies have for long been used to facilitate a more intimate
       contact between the subject’s skin and the recording electrodes.
        The outer horny layer of the skin is responsible for the bulk of the skin contact
       impedance, and for this reason careful skin preparation is essential in order to obtain
       the best results.
        The recording site should first be cleaned with an ether-meth mixture. In addition
       to having good electrical conductivity, the electrode jelly must have a particular
       chloride ion concentration (about 1%) close to the physiological chloride
       concentration.
        This is primarily important for long-term monitoring because it should not produce
       a harmful diffusion between the jelly and the body.
        It is to be particularly ensured that the jelly chosen is of a bland nature and does not
       contain soap or phenol which can produce a marked irritation of the skin after a few
       hours.
        The electrical conductivity of different makes of electrode cream can be measured
       (Hill and Khandpur, 1969) by means of the Schering ac bridge circuit.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                            Page 28
       The cream is placed in a perspex conductivity cell of known dimensions and the
      resistive component of the cell impedance is measured at 10 Hz, the conductivity
      being calculated from the cell dimensions.
       The contact impedance of the skin depends upon the type of electrolyte used and
      the time (Trimby, 1976)
       A low concentration sodium chloride electrolyte has 0.5% NaCI and a high
      concentration electrolyte has a concentration in the range of 5 to 10% NaCI.
       The impedance is found to fall rapidly to 40% between 7 to 30 min. Stabilization
      occurs at about 30 to 45 min.
       An interesting observation from this figure is that while pre-rubbing the skin will
      lower the initial impedance value, the final value after using a high concentration
      electrolyte becomes nearly the same.
          Electrode jelly can be replaced in certain cases by using a conducting plastic as
      an interface between the electrode and the surface of the body.
       The impedance is found to fall rapidly to 40% between 7 to 30 min.
       Stabilization occurs at about 30 to 45 min.
       An interesting observation from this figure is that while pre-rubbing the skin will
      lower the initial impedance value, the final value after using a high concentration
      electrolyte becomes nearly the same.
       Electrode jelly can be replaced in certain cases by using a conducting plastic as an
      interface between the electrode and the surface of the body.
       Jenkner (1967) used silastic S-2086 by Dow Corning with EEG electrodes and
      showed that contact resistance was almost the same as with a conventional electrode
      which make use of electrode jelly.
TYPES OF GEL:
PEO - BASED POLYMER GEL ELECTROLYTE:
       For preparation of this electrolyte EC and/or PC are taken as plasticizers. LiClO4,
      LiCFSO3, LiN(SO2CF3)2 etc are used as the salt.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                          Page 29
       This electrolyte has a conductivity of 10-3 Scm-1 . The polymer used (i.e. PEO) is
      not soluble to a greater extent with the solvent resulting in poor mechanical strength
      of the gels.
PAN - BASED POLYMER GEL ELECTROLYTE:
       The polymer used is PAN (Polyacrylonitrile). Lithium salts are the principle salt
      used here.
       The Li+ ion transference number were found to be very high in case of PAN- based
      gel electrolyte.
PMMA - BASED POLYMER GEL ELECTROLYTE:
       The polymer, solvent and salt are PMMA, EC and/or PC and Lithium and sodium
      salts respectively.
       An ionic conductivity of 10-3 S cm-1 has been achieved in this gel electrolyte.
PVDF - BASED POLYMER GEL ELECTROLYTE:
       PVDF is the polymer, EC/PC or DMF is the solvent and lithium salts such as
      LiCF3SO3, LiPF6 or LiN(SO3CF3)2 or NaI is mainly taken as the salt.
       PVDF based gel has good electrochemical properties but they are reactive to
      lithium and lithium salts.
DEPARTMENT OF BIOMEDICAL ENGINEERING
BM T46 – BIOMEDICAL SENSORS AND TRANSDUCERS                                        Page 30