Picture ‘What to do / Protocol Screenshot bellavista calibration menu Pressure gain calibration Calibrates the most important pressure sensors at 30 mbar in one calibration step 1. Connections according to picture 2. Press Prepare ‘Wait until the Calibrate button becomes ac- tive, but don't press it yet 4. Press button Blower ... % (see bellavista screenshot on the lef) 5. Adjust the blower % until you read a pres- sure of = 30 mbar on your FlowAnalyser PF- 300 (measurement reading: P (HF) / CIT- REX (measurement reading: Channel) us- ing the appeared slider (go as close as pos sible) Press Close Press button Pines ... mbar (see bellavista screenshot on the left) 8. Type in the exact value you read on your xo 9. 10. W. FlowAnalyser PF-300 / CITREX using the appeared numeric keypad (e. g. 30.08) Press Apply Press Calibrate Wait until the Reset button becomes active, then press it 12. Check the calibration date and time ‘Setup with CITREX: Connection to green connector with / second T-piece™ only needed fo G5 * tno second piece is available, 3 proximal flow sensor can be used alternatively (instead of the two T-pieces) with one ne connected to the blue port and the other one connected tothe green port or

You might also like