Calibration Technology
Hand-Held Pressure Calibrator with integrated pump
Model CPH6600
                                                                                                                     WIKA Data Sheet CT 16.01
 Applications
■■ Calibration service companies and service industry
■■ Instrument and control workshops
■■ Quality assurance
 Special Features
■■ Integrated pneumatic pump generates pressure to 20 bar
    and vacuum to -0.8 bar
■■ Accuracy to 0.025 % (incl. calibration certificate)
■■ Source/measure 4 ... 20 mA and 24 V loop power to power
    device under test
■■ Pt100 RTD input for temperature measurement, accurate
    to 0.1 °C (measurement only)
■■ Small, rugged compact design
                                                                                 Hand-Held Pressure Calibrator with integrated electric
                                                                                 pump Model CPH6600
Description
WIKA introduces a revolution in pressure calibration technol-                    An external pressure module connection supports all WIKA
ogy for the process industries. With the release of its new                      pressure modules, allowing even greater ranges.
CPH6600, technicians will have a small, lightweight hand-                        In combination with an external temperature probe, the
held calibrator that generates pressures up to 10 bar using a                    CPH6600 enables a simultaneous temperature measure-
high-performance integral electric pump. With manual pump                        ment. Additional functions, such as pressure switch testing
up to 20 bar is possible.                                                        and the indication of the test specimen deviation in percent,
                                                                                 increase the user-friendliness of the CPH6600.
The CPH6600 has a form factor similar to WIKA’s popular
CPH6400 Precision Hand-Held Pressure Indicator and
weighs less than 1 kg. The housing is contoured to easily fit a
technician’s hand with good balance top to bottom.
Performance doesn’t stop with the pump. The CPH6600
provides ± 0.025 % FS accuracy on its internal, isolated
pressure sensor. Temperature compensation on its internal
sensor ensures accuracy in field applications.
 WIKA Data Sheet CT 16.01 ∙ 03/2010                                                                                                 Page 1 of 6
Data sheets showing similar products and accessories:
ProcessCalibrator; Model CPH6000; see Data Sheet CT 15.01
Precision Hand-Held Pressure Indicator; Model CPH6400; see Data Sheet CT 14.01
Precision Digital Pressure Gauge; Model CPG1000; see Data Sheet CT 10.01
Dimensions in mm
Front view                  Right side view
Left side view
                                                          CPH6600
                                              F1   F2       F3
                   Pressure/vacuum                           ZERO
                   selector knob                   HOME
                                                           PUMP
                   Vent knob
                                                                       Electrical connec-
                   Pressure port                                       tions mA, volts,
                                                                       switch test
                                                                       Pressure, vernier
                                                                       adjustment
Page 2 of 6                                           WIKA Data Sheet CT 16.01 ∙ 03/2010
 Specifications                        Model CPH6600
 Pressure ranges                bar    -0.8 ... +2 bar*          -0.8 ... +10 bar*         -0.8 ... +20 bar**    0 … 20 bar abs.**
 Overpressure safety            bar    4 bar                     14 bar                    40 bar                40 bar
 Burst pressure                 bar    34 bar                    70 bar                    135 bar               135 bar
 Accuracy                       % FS   0.025 (6 month)
                                       0.035 (1 year)
 Pressure connection                   1/8” NPT female
 Permissible media                     clean, dry non-corrosive gases compatible with Silicon, Pyrex, RTV, Gold, Ceramic, Nickel and
                                       Aluminium
 Temperature compensation       °C     15 … 35
                                       Add 0.005 % FS/°C outside of 15 … 35 °C
 Units                                 psi, bar, mbar, kPa, MPa, kgcm², mmH2O @ 4°C, mmH2O @ 20 °C, cmH2O @ 4 °C,
                                       cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C
 Current
■■ Measuring range              mA     0 … 24 (max. load 1000 Ω)
■■ Resolution                   µA     1
■■ Accuracy                            0.015 % of reading ± 2 µA (simulation and measurement)
 Voltage
■■ Measuring range              DC     0 … 30 V
■■ Resolution                   mV     1
■■ Accuracy                            0.015 % of reading ± 2 mV (measurement)
 Output loop power              DC     24 V
 Temperature
■■ Measuring range              °C     -50 … +150
■■ Resolution                   °C     0.01
■■ Accuracy                            0.015 % of reading ± 20 mΩ, or 0.2 °C for complete measuring chain (Pt100 resistance thermo-
                                       meter and CPH6600)
 Permissible
■■ Operating temperature        °C     -10 … +50
■■ Storage temperature          °C     -20 … +60
 Power supply
■■ Battery                             DC 12 V, Eight (8) standard AA batteries
■■ Battery life                        300 pump cycles to 10 bar (minimum)
                                       1000 pump cycles to 2 bar
 Housing                               PolyCarbonate/ABS blend
 Ingress protection                    IP56
 Dimensions                     mm     see technical drawing on Page 2
 Weight                         g      950
 CE-conformity
■■ EMC directive                       2004/108/EC, EN 61 326 Emission (Group 1, Class B) and Immunity (portable measurement
                                       equipment)
 Calibration                           Factory calibration certificate (optional: DKD calibration certificate)
* electric pump
** manual pump
WIKA Data Sheet CT 16.01 ∙ 03/2010                                                                                         Page 3 of 6
Features
Temperature, Current and Voltage                                  Functions
A Pt100 resistance thermometer (RTD) input is provided for        The new CPH6600 Pressure Calibrator provides a number of
temperature measurements, accurate to ± 0.1 °C. In addition,      convenient functions. Switch tests can be performed on the
the CPH6600 measures and simulates 4 ... 20 mA loop               internal or external pressure input. Percentage (%) Error and
current signals. It can measure up to 30 V DC. An internal 24     damping functions are provided. Up to five frequently-used
V DC instrument power supply supplies power to a transmit-        setups can be stored and retrieved with one-button recall.
ter under test.
                                                                  Compact and rugged
Display                                                           The compact, rugged design operates on eight (8) standard
The CPH6600 displays up to 3 calibration values at one time       AA Alkaline batteries. Due to the power saving designs incor-
from the choice of internal pressure sensor, external pressure    porated in the CPH6600, it can perform approximately 300
sensor, temperature from the optional RTD probe or electri-       calibration cycles to 10 bar on one set of batteries.
cal (mA or V DC). The display is a large backlit graphics-style
LCD that is easy to read under any lighting conditions.
Pressure ranges
The CPH6600 can be delivered in 4 different ranges of 2 bar,
10 bar and 20 bar. Also an absolute range of 0 … 20 bar abs
is available. The 20 bar versions use a manually-operated
hand pump.
               [P1]                             psi           3   1) Primary Parameters
                                                                     Indicate what is being measured
1                         0.000
               RTD                                                2) Span Indicator
                                                              4      Indicates the percentage of the 4 to 20 mA span (For mA
                       72.50    °F                                   and mA Loop functions only)
               mA                      -25.00%
                                                              2
                               0.000 mA                           3) Pressure Units
                CONF UPPER MORE                                      Indicates one of 15 pressure units available for display
                                                                  4) Units
                                                                     Indicates the unit of measure for the display
Keypad (Electric pump versions)
                                                      1           1) Function Keys
                                                                     These are soft keys used to configure the calibrator
                                                                  2) ON/OFF Key
                                                                     This key is used to turn the calibrator on and off
                          F1     F2     F3
                                                          3
                                                                  3) ZERO Key
                                         ZERO
                                HOME                                 This key is used to zero pressure measurements
         2                                                4
                                                                  4) Arrow Keys
                   5                   PUMP
                                                                     Used to control mA source/simulation and to set pump
                                                      6
                                                                     and % error limits
                                                                  5) Home Key
Keypad (Manual pump version)                                         Return to main menu screen
                                                                  6) Pump Key
                                                      1
                                                                     Push to run pump (electric pump version)
                                                                  7) Backlight
                         F1      F2     F3                           Turn on backlight (manual pump version)
                                                          7
                                HOME
        2                                                 4
                   5                   ZERO
Page 4 of 6                                                                                  WIKA Data Sheet CT 16.01 ∙ 03/2010
Special operating modes
Operating mode: Pressure switch test
                                                                  Pressure switch test:
                    CPH6600
                                                                  The pressure switch function allows the CPH6600 to display
    [P1] ST CLOSE     psi
                                                                  the pressures when the switch closes and opens. In addition
           0.000                                                  the hysteresis can be displayed.
    CONFIG LOWER MORE
                                                                  In the switch test mode the display update rate is increased
                                                                  to help capture changing pressure inputs.
     F1      F2       F3
                       ZERO
             HOME
                                                                  If necessary, it is possible to measure the ambient or media
                                                                  temperature with an external Pt100 resistance thermometer
                     PUMP
                                                Pressure switch
Use low volume
tubing when pos-
sible
Operating mode: Calibrating Transmitters & Percentage error function
                                                                  Calibrating Transmitters:
                    CPH6600
                                                                  The mA input function allows the 4 ... 20 mA output from the
                                                                  device being calibrated to be read back. This can be done in
    [P1]             psi
            0.00
    mA              0.00%
          4.000 mA
    % ERROR
          0.00 %
                                                                  one of two ways.
    CONFIG LOWER MORE
     F1       F2      F3
                                                                  1) Passively – Where the device under test generates
                       ZERO       SIGNAL
                                                                     4 ... 20 mA directly and can be read by the calibrator.
             HOME
                              +             –                     2) Actively – Where the calibrator supplies 24 V DC loop
                                                                     power to the device under test to power the device while
                     PUMP
                                                                     reading the resulting 4 ... 20 mA signal.
                                     TEST
                                                                  The calibrator features a unique function which can calculate
                                                                  pressure versus milliamp error as a percentage of the 4 ...
                                                                  20 mA loop span. The percent error mode uses all 3 screens
                                                                  and has a unique menu structure. It simultaneously displays
                                                                  pressure, mA and percentage error.
                                                                  Example:
                                                                  Suppose a pressure transmitter under test has a measuring
                                                                  range of 2 bar and outputs a corresponding 4 ... 20 mA signal.
                                                                  The user can programme a 0 to 2 bar pressure span into the
                                                                  calibrator and the calibrator will calculate and dis-play the
                                                                  deviation or % Error from the expected 4 ... 20 mA output.
                                                                  This eliminates the need for manual calculations and also
                                                                  helps if it becomes difficult to set an exact pressure with an
                                                                  external pump.
WIKA Data Sheet CT 16.01 ∙ 03/2010                                                                                    Page 5 of 6
Scope of Supply
■■ CPH6600 Hand-Held Pressure Calibrator with integrated
    pump
■■ Operating instructions
■■ Test leads
■■ Calibration hose kit with fittings ⅛ NPT male
■■ Adapter set ⅛ NPT female on ¼ NPT female, ⅛ NPT
    female on ¼ BSP female, ⅛ NPT female on G ½ female
■■ PTFE thread seal tape
■■ 8 batteries type AA
■■ 3.1 Calibration Certificate per DIN EN 10 204
Accessories                                                                                         Hand-Held Pressure Calibrator with integrated manual
                                                                                                    pump
■■ Pt100 resistance thermometer
■■ Battery set, 4 NiMH batteries
■■ Battery charger set, incl. 4 batteries, quick charger, power
    lead, adapters
■■ Service case
Option
■■ DKD calibration
                                                                                                    Hand-Held Pressure Calibrator with service case
Products and Services within our Calibration Technology Program
■■ DKD calibration services for pressure                                                           ■■ DKD calibration services for temperature
■■ Repair of calibration units of all makes                                                        ■■ Portable measuring devices and calibrators
■■ Portable pressure measuring devices for test and calibra-                                       ■■ Dry well temperature calibrators
   tion tasks                                                                                      ■■ Calibration baths and furnaces
■■ Precision pressure measuring units and pressure control-                                        ■■ Precision thermometers
   lers                                                                                            ■■ Primary standards for temperature
■■ Primary standards for pressure                                                                  ■■ Consulting and seminars
■■ Testing technology system solutions
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.
 Page 6 of 6                                                                                                                WIKA Data Sheet CT 16.01 ∙ 03/2010
                                                                                                                                                                 03/2010 GB
                                                                                                                      WIKA Alexander Wiegand SE & Co. KG
                                                                                                                      Alexander-Wiegand-Straße 30
                                                                                                                      63911 Klingenberg/Germany
                                                                                                                      Tel.	   (+49) 9372/132-0
                                                                                                                      Fax	    (+49) 9372/132-406
                                                                                                                      E-mail	 info@wika.de
                                                                                                                      www.wika.de