جامعة االنبار
كلية العلوم التطبيقية – هيت
                   قسم الفيزياء الحياتية
    االجهزة الطبية
Blood Gas Analyzer
 Mohammed Qasim Taha
                           Definition
• Blood gas analysis
  • Also called Arterial Blood Gas (ABG) analyzer
  • It uses 3 electrodes to test arterial blood
• Used in
   1. Respiratory (breathing) therapy departments
   2. Clinical and cardiopulmonary (heart and the lungs) labs
   3. Critical care units, surgical suites
   4. Physician offices, and hospital nurseries to monitor
      patients’ acid-base balance and oxygen (O2)- carbon
      dioxide (CO2) exchange, providing the clinician with
      information to use in patient diagnosis and regulation
      of therapy.                                          2
                   Purpose
Blood gas/pH analyzers measure:
1. The partial pressure of oxygen (O2) → po2
2. Carbon dioxide (CO2) gases → pco2
3. The (hydrogen ion concentration) p(h+)
4. Sodium, potassium, and calcium concentrations
• Values for PO2, PCO2 and PH reflect the
  concentrations of these gases in arterial blood as well
  as the concentration of hydrogenions (H+)
• hence, the state of respiration, metabolism and
  body’s acid production of a patient can be diagnosed
                                                      3
          Components of Blood Gas
                       Blood Gas
Oxygenation            Ventilation   Acid-Base
     PaO2
     SaO2
                         PCO2           pH
  Sa: saturation
Pa: partial pressure
                                                 4
ABG Analyzer example
                       5
                 Calibration
• Calibration: with known concentrations of standard
  buffers and calibrated solutions.
 1. Gas mixtures with high and low concentrations of
    O2 and CO2 are alternately admitted to the sample
    chamber,
 2. O2 and CO2 electrode responses are used to set
    high and low points of the PO2 and PCO2 curves.
 3. Calibrations are done by adjusting the electrode
    response and are usually referred to:
   • One level (either high or low)
   • Two levels (both high and low)
• The electrode systems and the sample chamber are
  located inside a temperature-controlled block
  maintained at 37°C
                                                     6
        Reference Electrode
•The reference electrode is
 used in the measurement
 of pH and electrolyte
 parameters, located in the
 pH/Blood Gas module.
Components
1. Pole: AgCl
2. Electrolyte: KCl
3. Permeable seal
                              7
              pH electrode
•The pH measurement is
 performed using two
 separate electrodes:
  1. Ph-measuring Electrode
  2. Reference Electrode
•The pH-sensitive glass
 membrane is located at
 the tip and seals the inner
 buffer solution with a
 constant and known pH.
                               8
                   pH electrode
• A saturated electrolyte
  solution (Potassium
  Chloride) in the reference
  electrode and a leaky
  membrane permit current
  flow from the reference
  electrode through the
  sample in the
  measurement chamber to
  the measuring electrode.
• The potential difference is
  displayed on a voltmeter
  calibrated in pH units.         9
                      pO2 electrode
• Oxygen electrode measures the
  oxygen partial pressure in a
  blood or gas sample.
1. Cathode: platinum
2. Anode: a silver/silver chloride
3. Electrolyte Solution: sodium
   chloride, Cathode and anode
   are placed in the electrolyte
4. Applied voltage: 700 mV
5. Permeable membrane:
   (Plastic) designed to allow
   only O2 to leak
                                      10
pO2 electrode
                11
                   Pco2 Electrode
• The pCO2 electrode is a combined
  pH and Ag/AgCl reference
  electrode mounted in a plastic
  jacket, which is filled with a
  bicarbonate electrolyte.
• The PCO2 electrode also contains
  a spacer (usually a porous
  membrane of nylon) that acts as a
  support.
→ As CO2 diffuses through the
 membrane and into the support,
 The pH of the electrolyte changes
→ The output of this modified pH
 electrode is proportional to the
 PCO2 present in the sample.          12
Electrode Cartridges
                       13
           Other Electrodes
•The K+ electrode is an ion-selective electrode
  • Sensing Element is a PVC membrane containing a
   potassium-neutral ion carrier.
• The Na+ electrode is an ion-selective electrode
  • Sensing Element is a Na+-sensitive ceramic pin
    contained in the tip of the jacket.
•The Ca++ electrode is an ion-selective electrode
  • Sensing Element is a PVC membrane containing a
    calcium-neutral ion carrier.
                                                     14
15
16
17
    In-line (extracorporeal) monitors
   • Connected to the patient’s existing arterial line
   • These devices consist of a sensor, an arterial blood gas (ABG)
     module, and a monitor.
Operation:
Blood is drawn into the sensor
     → Fluorescent dyes in the sensor detect the pH, PO2, and PCO2
    levels and transmit light of a specific wavelength
The ABG module emits excitation energy
   → dyes transmit light of a specific wavelength (λ1) through a
   fiber-optic cable to the instrument.
   → The dyes then emit light of a longer wavelength (λ2) and
   transmit it to the instrument through the cable.
    →The parameters are then determined by the difference in the
   two wavelengths
                                                                  18
 In-line (extracorporeal) monitors
SensicathTM system consists of patented fiber optic technology in a sensor which
                                      T.Elsarnagawy, MASH210, 311,2010             19
attaches directly to the patient’s arterial     line
•Continuous intravascular blood gas monitoring
 • This technique uses electrochemical and optical
   detection methods; however, the measurements
   are taken in vivo
                                                     20
Modern blood gas analyzers features
•Measure/time:
  • pH (15 Seconds)
  • Po2 (50 Seconds)
  • Pco2 (30 Seconds)
•Have Monitor
•Analog Amplifier and digital control and storage
 circuits
•Easy calibration
                                                21
   Basic Safety Consideration
 When analysis is complete, the blood
 specimen is disposed of in one of two ways:-
1. Most analyzers pump the specimen into a waste
   container, and the system is flushed with a rinse
   or wash solution.
2. Some newer units retain the specimen in the
   disposable sealed reaction cartridge which is
   then discarded.
                                                   22
Important Terminologies
1. Analysis time, sec: The time from sample
   insertion to a displayed or printed result.
2. Calibration: The operator can initiate random
   calibrations and calibrations during standby
   mode.
3. Standby mode: Unit calibrates only when the
   unit is in standby mode, otherwise it will
   calibrate according to programmed intervals
                                                 23
    Basic Work Specifications
1. The equipment should possess electrodes
   with long life at least 2 years
2. Assessment of the instrument should be
   provided by the company.
3. All results should be available within 3 min.
4. The results should be microprocessor
   controlled and of latest technology version
                                             24
   Basic Work Specifications, cont.
5. The instrument should have facilities like
   monitor screen, external keyboard, mouse,
   and barcode reader
6. The instrument should have the capability to
   interface a computer and a computer should
   be supplied for data acquisition and patient
   record with recommended software. The
   system should have RS232 serial port.
7. Display language should have English
                                              25