SWIRvs NIR
SWIRvs NIR
fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                                1
        Copyright (c) 2019 IEEE. Personal use of this material is permitted. However, permission to use this material for any other
        purposes must be obtained from the IEEE by sending a request to pubs-permissions@ieee.org.
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              2
        benefits of reduced photon scattering, negligible autofluores-                                                                 II. MATERIALS AND METHODS
        cence, and comparably low absorbance in SWIR window have                                              A. ICG based working standards calibrated with SI units
        been hypothesized to enable high-resolution imaging at greater                                        Previously we have used stable, working standards based upon
        penetration depths than achievable in NIR window [8-12].                                              the small fluorescent yield of QDots800 to assess performance
        Moreover, ICG has recently been found to have SWIR                                                    of various custom and commercial NIRF devices that operate
        fluorescence emission (>1064 nm) following excitation at 808                                          within the same or similar wavelength regimes. In this work,
        nm and its SWIR emission spectra can be found in [13]. ICG                                            detection of ICG at NIR and SWIR wavelengths defines the
        has been successfully used in small animals to demonstrate                                            task for which device performance is assessed. For this reason,
        SWIR intravital microscopy, noninvasive SWIR imaging in                                               we constructed working standards similar to that of our prior
        blood and lymph vessels and SWIR assessment of hepatobiliary                                          work [4, 5], using ICG as opposed to QDots800. The standard
        clearance using emerging commercial InGaAs array detectors                                            consists of 6 wells machined from a 96 well plate and filled with
        [14]. Despite the small fluorescent yield of ICG at 1064 nm                                           a mixture of polyurethane, TiO2 (1.5 mg/mL) and increasing
        following the NIR excitation, the large 200 nm Stokes shift may                                       concentrations of ICG diluted in water or in ethanol.
        be particularly advantageous for SWIR fluorescence imaging.                                           Polyurethane was chosen as the base material for the phantom
        Even with a comparatively larger fluorescent yield, the                                               due to its optical clarity, fast curing time, and affordability;
        incomplete rejection of excitation light due to the small (10 –                                       TiO2 was added to the phantom to exaggerate the optical
        60 nm) Stokes shift experienced in NIRF systems limits their                                          scattering properties of tissue that lead to incomplete excitation
        performance [15], a problem potentially avoided by the large                                          rejection; and 6 wells of the phantom had varying
        Stokes shift probed by SWIR imaging. As a result, SWIR has                                            concentrations of ICG. A low radiance phantom was
        thus been proposed as having advantage over NIRF imaging for                                          constructed to have similar fluorescent radiances to our prior
        clinical applications [13, 14], yet no quantitative comparison                                        work with ICG concentrations of 0 (acting as a non-fluorescent
        using a non-clinical or in vitro test assessment has been                                             background), 20, 40, 60, 80, and 100 nM and a high radiance
        presented. Non-clinical or in vitro testing provides an                                               phantom was constructed with ICG concentrations of 200, 400,
        inexpensive, repeatable, and when reported in SI units,                                               600, 800 and 1000 nM. To calibrate the radiance from the ICG
        traceable measurement that can hasten direct comparison and                                           based fluorescent solid phantoms with SI units in NIR window,
        enable evaluation of improvement in device advancements such                                          we used the same methodology described previously for
        as potentially offered by SWIR detectors. Figure 1 is a                                               calibration of QDots 800 based luminescent solid tissue
        schematic depicting the typical spectral responses of the                                             phantoms with SI units [5]. Briefly, a NIST calibrated power
        systems currently deployed in clinical imaging of ICG.                                                meter is used to assign a radiance scale to an 830 nm diffuse
                                                                                                              source (830 nm laser diode and integrating sphere, known as
                                                                                                              the “reference source”) with output matching emission from the
                                                                                                              fluorescent tissue phantoms. Both reference source and tissue
                                                                                                              phantoms are viewed by the camera system under the same
                                                                                                              radiometric conditions. Thus, the reference source is used to
                                                                                                              calibrate the arbitrary unit (AU) output of the camera system
                                                                                                              configured under specific radiometric conditions and
                                                                                                              transferring the radiance scale (mW ∙ cm−2 ∙ sr −1 ) to the tissue
                                                                                                              phantoms illuminated at a calibrated irradiance at 785 nm. The
                                                                                                              tissue phantoms can then be used as working standards at this
                                                                                                              wavelength to evaluate camera systems at any other radiometric
                                                                                                              configuration. To calibrate SI units in SWIR window, we
                                                                                                              replaced 785 nm excitation source with 808 nm excitation
                                                                                                              source, and the 830 nm reference source with 1064 nm
                                                                                                              reference source with all other configurations held constant.
            Fig. 1. The typical spectral response curves of detectors made of Si, GaAs,                       Radiance from the integrating sphere reference sources were
            and InGaAs materials.                                                                             measured by NIST-calibrated InGaAs- and Si-photodiode
                                                                                                              power meters at the wavelengths of 1064 nm and 830 nm
           The main objective of this work is to evaluate performance                                         (NIST15, Gaithersburg, MD; S2281-04, Hamamatsu, Japan).
        of Si-, InGaAs- and GaAs intensified- based camera systems
        for detecting ICG using an in vitro, fluorescent solid phantom                                        B. NIR and SWIR fluorescence imaging systems
        calibrated with SI-traceable units. A secondary objective is to                                       In NIRF imaging set-ups shown in Fig. 2(a), a 785 nm laser
        relate performance assessed from the quantitative in vitro test                                       diode (HPD1005-9mm-78503 model, High Power Devices
        to qualitative images from large and small animals. It should                                         Inc., NJ) that was typically employed in our investigational
        be noted that the comparisons made herein are specific to the                                         devices was used to illuminates the tissue/standard surfaces of
        devices compared, and that the methodology for assessing                                              interest through a diffuser. The collection of fluorescence
        performance in SI-traceable units could be used to drive                                              signals were implemented using two 830 nm band pass filters
        technological advancements that improve existing and future                                           (830FS10, optical density > 5 at 785 nm, Andover, Salem, NH)
        NIRF and SWIR fluorescent devices.                                                                    separated with the Nikon focus lens (AF NIKKOR 28 mm
                                                                                                              f/2.8D, Nikon, NY). The Nikon lens introduced a small amount
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              3
Fig. 2. Schematics of NIR (a) and SWIR (b) fluorescence imaging systems installed with Si, GaAs coupled Si, and InGaAs based cameras.
        of loss between two filters to efficiently cancel multiple-path                                       SWIR fluorescence imaging systems. To calculate the SNR and
        interference effects and increase the combined optical density.                                       contrast, the ICG-based fluorescent solid phantom with SI units
        This reduced the leakage of excitation light through the filters.                                     was first positioned under the FOV and the irradiance measured
        The collected fluorescence signals were then registered on a Si-                                      on the phantom surface was 3.7 mW/cm2 for 785 nm excitation
        based commercial electron-multiplying couple-charged device                                           and 4.2 mW/cm2 for 808 nm excitation. Fluorescence images
        (EMCCD) camera (Photon Max 512, Princeton Instruments,                                                were acquired at two different integrating times of 33 and 200
        Trenton, NJ) or a GaAs-based intensified scientific                                                   ms with the gain adjusted such that an individual pixel value
        complementary            metal–oxide–semiconductor (IsCMOS)                                           within the phantom well with the highest ICG concentration
        camera (a customized fiber optic coupled Zyla 5.5 sCMOS,                                              was just less than the full well capacity of the camera. From the
        ANDOR, Concord, MA) as developed in our past works [5]. In                                            acquired fluorescence images, the SNR corresponding to each
        the SWIR fluorescence imaging system shown in Fig. 2(b), a                                            well in the calibrated ICG-based fluorescent solid phantom was
        laser source operating at 808 nm (L808P1000MM model,                                                  calculated using the following equation:
        Thorlabs, NJ) was utilized to illuminate the tissue/standards.
                                                                                                                                                              S t C   S b 0
        The choice of 808 nm as the excitation wavelength was based                                                                                SNR                                                   (1)
        upon prior literature reports of SWIR fluorescence detection of                                                                                                
        ICG. The generated fluorescence signals were collected with a                                         where S t C  and S b 0 represent the averaged arbitrary units
        1064 nm long-pass filter (BLP01-1064R, optical density >5 at
        808 nm, Semrock, Rochester, NY) and then focused through a                                            (AU) values for the pixels corresponding to the entire well
        SWIR lens (SR2343-A01 model, StingRay Optics, Keene, NH)                                              containing C concentration of ICG and to the well without ICG
        onto an InGaAs-based PIN Photodiode camera (Ninox SWIR                                                respectively, and σ represents the standard deviation of AU
        640, Raptor Photonics, kindly loaned by Phoenix Engineering,                                          values in the well without ICG. The calculated SNR was then
        Carson, CA). The working distances and hence radiometric                                              plotted as a function of radiance (mW ∙ cm−2 ∙ sr −1 ) of each
        conditions were identical for all systems. All cameras were                                           well in ICG based fluorescent solid phantom. The contrast was
        cooled to their corresponding temperatures suggested by the                                           computed by the following relationship:
        manufacturers and the field of views (FOVs) of both NIR and                                                                                                 S t C 
                                                                                                                                                 Contrast                   1                           (2)
                                                                                                                                                                    S b 0
        SWIR fluorescence imaging systems were adjusted to 10 cm in
        diameter.
                                                                                                              The computed contrast was then plotted as a function of
        C. Figures of merit for fluorescence imaging                                                          radiance (mW ∙ cm−2 ∙ sr −1 ) of each well of the working
        Signal-to-noise ratio (SNR) and contrast were used to compare                                         standard.
        and quantify the measurement sensitivity of both NIR and
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              4
        D. Small and large animal imaging with NIRF and SWIR                                                  emission radiance to excitation irradiance for both NIR and
        imaging systems                                                                                       SWIR fluorescence channels is proportional to the ICG
        Animal studies were conducted under protocols approved by                                             concentration, although at the lower and upper ranges of
        the University of Texas Health Science Center Animal Welfare                                          radiance, the radiometric-specific function curves slightly
        Committee. Dynamic imaging was performed using the                                                    deviate from linearity, demonstrating the dark noise of the
        imaging systems described above on anesthetized athymic mice                                          detector and maximum measurable radiance prior to camera
        after 100 µL of 323 µM ICG was injected into the tail vein.                                           saturation. The slope of the lines refer to the fluorescent
        Images were acquired dynamically with acquisition times of 50                                         characteristic of the standard, which for our past work with
        ms using the EMCCD system and 200 ms using the GaAs-based                                             inorganic QDots800 is stable for years, but for ICG deteriorates
        IsCMOS and InGaAs systems. The irradiance measured on the                                             with light exposure and time slowly [5]. As a result, these
        animal surfaces has a spatial variation of 3.0~4.4 mW/cm2 for                                         standards were prepared and used in a darkroom and exposure
        785 nm excitation and 3.1-5.1 mW/cm2 for 808 nm excitation.                                           to illumination for no more than a few seconds over which time
        In ancillary studies of swine, dynamic imaging was conducted                                          their output appeared to be stable. It is noteworthy, that despite
        using the IsCMOS and InGaAs systems following                                                         the small fluorescence yield at SWIR over NIR wavelengths,
        intramuscular injection of 1-2 mL of 62.5 µM using an 18-                                             the radiance in SWIR fluorescence channel is higher than that
        gauge needle.                                                                                         in NIRF channel. We believe that the narrow bandwidth of 830
                                                                                                              nm bandpass filter limited the amount of NIR fluorescence
                                                                                                              collected as compared to the SWIR fluorescence collected
                                              III.    RESULTS                                                 using the 1064 nm long pass filter. We did not evaluate whether
        A. Radiance of ICG based fluorescent solid phantom in SI                                              enhancement of optical effects that occur due to the low
        units                                                                                                 absorption and high refractive index of TiO2 scattering particles
        When reconstituted in water prior to incorporation with                                               in the presence of a fluorophore could be responsible the large
        polyurethane and TiO2, or when diluted in ethanol at                                                  differences between the NIR and SWIR measurement of
        concentrations lower than 200 nM, ICG emission was too weak                                           fluorescent characteristics of the working standard.
        for detection or quantification using the SWIR system. Figure
        3(a) shows that ICG in ethanol exhibited higher emission at                                           B. Measurement sensitivity of both NIR and SWIR
        >1064 nm than ICG in water, which has been attributed to its                                          fluorescence imaging systems
        higher absorption cross section in ethanol [13], but could also                                       Figure 5 shows that the SNR of NIRF imaging systems are
        be impacted by water absorption at SWIR wavelengths.                                                  much higher than that of SWIR system despite that there is an
        Because SWIR fluorescence imaging systems can detect                                                  2-3 order of magnitude greater radiance in SWIR fluorescence
        emission of the lowest ethanol-diluted ICG concentration (200                                         channel. At longer integrating times of 200 ms, the SNR is
        nM) as shown in Fig. 3(b), working standards prepared with                                            improved as compared to that at 33 ms for both the Si-based
        ethanol dilution of ICG at concentrations of 200 to 1000 nM                                           EMCCD and InGaAs systems as depicted in Fig. 5(a) and (c).
        were adopted. Figure 4 illustrates the ratio of emission radiance                                     For the GaAs-based IsCMOS imager (shown in Fig. 5(b))
        at 830 nm to the incident 785 nm excitation irradiance (ratio                                         shorter integrating time is compensated by the increased gain of
        uncertainty = 0.10) as well as the ratio of emission radiance at                                      the intensifier. Figure 6 shows that the Si-based EMCCD NIRF
        1064 nm to the incident 808 nm excitation irradiance (ratio                                           imaging system provides approximately twice the contrast as
        uncertainty = 0.11) for each of the wells containing different                                        that of SWIR system. In comparison to both, GaAs-based-
        concentrations of ICG. It can be observed that the ratio of                                           IsCMOS NIRF imaging system gives greater contrast as shown
           Fig. 3. (a) The acquired SWIR fluorescence image of Eppedorf tubes filled                          Fig. 4. Emission radiance normalized to excitation irradiance of ICG
           with ICG diluted in water and ethanol, demonstrating higher fluorescence                           based working standards in NIR (Left axis) and SWIR (Right axis)
           signals in ethanol. (b) Three acquired fluorescence images from ICG based                          fluorescence imaging channels, where arrows to show which y-axis
           fluorescent solid phantom using NIR and SWIR fluorescence imaging                                  corresponds to each dataset.
           systems.
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              5
           Fig. 5. The plots of signal-to-noise ratio as a function of radiance at 33 ms and 200 ms exposure times using EMCCD (Si) (a) or IsCMOS (GaAs intensified)
           (b) camera based NIRF imaging system and InGaAs (c) camera based SWIR fluorescence imaging system.
          Fig. 6. The plots of contrast as a function of radiance at 33 ms and 200 ms exposure times using EMCCD (Si) (a) or IsCMOS (GaAs Intensified) (b) camera
          based NIRF imaging system and InGaAs (c) camera based SWIR fluorescence imaging system.
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              6
            Fig. 8. (a) Fluorescence images in NIR and SWIR fluorescence imaging channels of a rod consisting ICG (322 μM) at depths of 0, 2, and 4 mm in 1.0%
            Liposyn. (b) Full-width-half-maximum (FWHM) of fluorescent rod as a function of depth in 1.0% Liposyn, demonstrating improved spatial-resolution at
            SWIR as opposed to NIRF wavelengths.
        images, from which the full-width-half-maximum (FWHM) of                                              fluorescence imaging of ICG or any other potential fluorescent
        feature width was calculated, as shown in Fig. 8(b). At this                                          contrast agent [16] that could be translated clinically for
        high concentration of ICG, the FWHM of the SWIR ICG signal                                            molecularly targeted detection of diseased tissues. With more
        deteriorates less with increasing depth than does the NIRF ICG                                        SWIR sensitive array technologies made of other materials such
        signal.                                                                                               as InSb and HgCdTe, or perhaps more impactful, the advent of
           To relate performance characterized by the working                                                 SWIR sensitive intensifiers that couple to conventional Si-
        standards and in vitro tests to actual imaging, we also evaluate                                      based detectors, it may be possible that sensitivity could be
        the devices with small and large animal tests. Dynamic in vivo                                        improved beyond that possible with NIRF imaging systems.
        imaging of mice injected i.v. with 100 L of 323 M ICG was                                           Interestingly, a recent work by Carr et al [17] demonstrates that
        performed using NIRF (EMCCD and GaAs-based IsCMOS)                                                    the SWIR system has highest image contrast with a higher
        and SWIR fluorescence demonstrate comparable performance                                              imaging penetration depth at the peak absorbance of water
        with potentially less scatter and greater clarity with the SWIR                                       (~1450 nm). The SWIR system evaluated herein was less
        system, as shown in supplemental videos 1- 3 and Fig. 9. This                                         sensitive in detecting ICG than NIRF systems as depicted in
        result may be expected in the context of the results in Figure 8.                                     Figures 5-7. Future comparisons of imaging systems need to
        However, when using the larger swine animal, i.m. injection                                           be based on quantitative characterization using traceable
        with 2 mL of 62.5 µM can be seen using the GaAs-based                                                 working standards that are calibrated in SI units and that are
        IsCMOS NIRF system, but not with the InGaAs-based SWIR                                                appropriately used to objectively address performance in
        system, as shown in Fig. 10. This latter result is likely due to                                      clearly defined tasks through receiver operator curve analysis.
        the relatively strong absorption of water at SWIR wavelengths                                         Before task-based assessments are performed within expensive
        that limits penetration depth and is expected with the results                                        in vivo or clinical tests, an in vitro or non-clinical test can
        obtained in the ethanol and water-based working standards.                                            provide a simple means of assessing and comparing
                                                                                                              performance of different devices, their operation, and
                                                                                                              radiometric geometries in which they are used. Subsequent
                                            IV.    DISCUSSION                                                 evaluation of task-based performance within in vivo or clinical
                                                                                                              testing will require calibration with SI units to be effective.
        Herein we used an ICG-based working standard calibrated with
                                                                                                                 Previously, we developed a QDots 800 based working
        SI-traceable units to report the quantitative performance of two
                                                                                                              standard that, unlike the ICG working standard described
        NIR and a single SWIR fluorescence imaging system and then
                                                                                                              herein, possessed stable emission over a period of a year or
        followed with a qualitative comparison in large and small
                                                                                                              greater, but did not possess emission into the SWIR wavelength
        animal studies. The advantages of SWIR, namely reduced
                                                                                                              regimes. The broad emission spectra of ICG ranging from NIR
        tissue scattering, a large Stokes shift that minimizes
                                                                                                              to SWIR spectra region allows performance comparisons of
        contributions from autofluorescence and excitation light
                                                                                                              both NIR and SWIR fluorescence imaging systems possible.
        leakage, and the larger bandwidth for fluorescent collection
                                                                                                              On the downside, these ICG based working standards
        bode well for its deployment in biomedical optical imaging.
                                                                                                              photodecay with exposure to light and therefore are not as
        However increased water absorption and the smaller
                                                                                                              desirable as inorganic ones. Future work to develop stable
        fluorescent yield at SWIR wavelengths may offset these
                                                                                                              working standards with durable stability and spanning NIR and
        advantages, especially for clinical applications that use ICG.
                                                                                                              SWIR wavelength ranges could help spur the development of
        Because this study was confined to the three separate systems
                                                                                                              successful fluorescent imaging devices and the imaging agents
        studied, we would like to emphasize that we do not broadly
                                                                                                              they are designed to detect.
        conclude that NIRF imaging is advantageous over SWIR
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              7
            Fig. 9. Example of ICG imaging of small animals using EMCCD (Si) (a) or IsCMOS (GaAs) (b) camera based NIRF imaging system and InGaAs (c) camera
            based SWIR fluorescence imaging system.
            Fig. 10. Example ICG imaging in large animals using IsCMOS (GaAs) (a) camera based NIRF imaging system and InGaAs (b) camera based SWIR
            fluorescence imaging system.
        *
         Note: References are made to certain commercially available                                                                                 REFERENCES
        products in this paper to adequately specify the experimental                                         [1]           E. Sevick-Muraca, "Translation of near-infrared fluorescence
        procedures involved. Such identification does not imply                                                             imaging technologies: emerging clinical applications," Annual
        recommendation or endorsement by the National Institute of                                                          Review of Medicine, vol. 63, pp. 217-231, 2012.
                                                                                                              [2]           J. T. Alander et al., "A review of indocyanine green fluorescent
        Standards and Technology, nor does it imply that these products                                                     imaging in surgery," Journal of Biomedical Imaging, vol. 2012, p.
        are the best for the purpose specified.                                                                             7, 2012.
                                                                                                              [3]           B. Zhu and E. Sevick-Muraca, "A review of performance of near-
        Acknowledgement: The authors thank Jeanne Houston at NIST                                                           infrared fluorescence imaging devices used in clinical studies," The
                                                                                                                            British Journal of Radiology, vol. 88, no. 1045, p. 20140547, 2014.
        for the loan of the calibrated InGaAs photodiode used in the                                          [4]           B. Zhu, J. C. Rasmussen, and E. M. Sevick‐Muraca, "A matter of
        SWIR measurements.                                                                                                  collection and detection for intraoperative and noninvasive near‐
                                                                                                                            infrared fluorescence molecular imaging: To see or not to see?,"
                                                                                                                            Medical Physics, vol. 41, no. 2, 2014.
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/TMI.2019.2937760, IEEE
                                                                                               Transactions on Medical Imaging
        IEEE Transactions on Medical Imaging                                                                                                                                                              8
        [5]           B. Zhu, J. C. Rasmussen, M. Litorja, and E. M. Sevick-Muraca,                           [12]          H. Wan et al., "A bright organic NIR-II nanofluorophore for three-
                      "Determining the performance of fluorescence molecular imaging                                        dimensional imaging into biological tissues," Nature
                      devices using traceable working standards with SI units of                                            Communications, vol. 9, no. 1, p. 1171, 2018.
                      radiance," IEEE Transactions on Medical Imaging, vol. 35, no. 3,                        [13]          Z. Starosolski, R. Bhavane, K. B. Ghaghada, S. A. Vasudevan, A.
                      pp. 802-811, 2016.                                                                                    Kaay, and A. Annapragada, "Indocyanine green fluorescence in
        [6]           G. Hong, A. L. Antaris, and H. Dai, "Near-infrared fluorophores for                                   second near-infrared (NIR-II) window," Plos One, vol. 12, no. 11,
                      biomedical imaging," Nature Biomedical Engineering, vol. 1, no. 1,                                    p. e0187563, 2017.
                      p. 0010, 2017.                                                                          [14]          J. A. Carr et al., "Shortwave infrared fluorescence imaging with the
        [7]           S. L. Jacques, "Optical properties of biological tissues: a review,"                                  clinically approved near-infrared dye indocyanine green,"
                      Physics in Medicine & Biology, vol. 58, no. 11, p. R37, 2013.                                         Proceedings of the National Academy of Sciences, vol. 115, no. 17,
        [8]           K. Welsher, S. P. Sherlock, and H. Dai, "Deep-tissue anatomical                                       pp. 4465-4470, 2018.
                      imaging of mice using carbon nanotube fluorophores in the second                        [15]          E. M. Sevick-Muraca and J. C. Rasmussen, "Molecular imaging
                      near-infrared window," Proceedings of the National Academy of                                         with optics: primer and case for near-infrared fluorescence
                      Sciences, vol. 108, no. 22, pp. 8943-8948, 2011.                                                      techniques in personalized medicine," Journal of Biomedical
        [9]           G. Hong et al., "Multifunctional in vivo vascular imaging using                                       Optics, vol. 13, no. 4, p. 041303, 2008.
                      near-infrared II fluorescence," Nature Medicine, vol. 18, no. 12, p.                    [16]          B. K. Byrd et al., "Characterizing short-wave infrared fluorescence
                      1841, 2012.                                                                                           of conventional near-infrared fluorophores," Journal of Biomedical
        [10]          D. Naczynski et al., "Rare-earth-doped biological composites as in                                    Optics, vol. 24, no. 3, p. 035004, 2019.
                      vivo shortwave infrared reporters," Nature Communications, vol. 4,                      [17]          J. A. Carr, M. Aellen, D. Franke, P. T. So, O. T. Bruns, and M. G.
                      p. 2199, 2013.                                                                                        Bawendi, "Absorption by water increases fluorescence image
        [11]          H. Kantamneni et al., "Surveillance nanotechnology for multi-organ                                    contrast of biological tissue in the shortwave infrared," Proceedings
                      cancer metastases," Nature Biomedical Engineering, vol. 1, no. 12,                                    of the National Academy of Sciences, vol. 115, no. 37, pp. 9080-
                      p. 993, 2017.                                                                                         9085, 2018.
0278-0062 (c) 2019 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.