S8 / S81 Ceiling Mounts

Planning Manual
M-30-1382-en
Issue 12.0
Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


2

Key to symbols

Warning!
The warning triangle indicates potential sources of danger which may
constitute a risk of injury for the user or a health hazard.

Caution:
The square indicates situations which may lead to malfunction, defects,
collision or damage of the instrument.

Note:
The hand indicates hints on the use of the instrument or other tips for the
user.

OPMI® is a registered trademark of Carl Zeiss Surgical GmbH.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Contents

New installation of S8 / S 81 ceiling mount 5

Technical data 7
On-site preparations 24
Constructional requirements 29
Planning the electrical installation 33
Planning the mechanical installation 42
Confirmation of the structural calculation and execution
of installation 47
Ordering data 48

Accessory interface 61

Accessory interface 62
Installation position 63
Mount flange and ceiling panel 64
Mount flange (flange plate) 66
Configuration example: mount flange with flange tube 67
Pre-installation set 68
Customer's responsibilities 69
Electrical connections for accessory modules 71
Confirmation of the structural calculation and execution
of installation 73

Discipline-specific installation positions 75

Installation positions for ophthalmology and ENT 77


Installation position for P&R, Hand 90
Installation position for Neuro, Spine, Uro and Gyn 93

Installation on an existing substructure 97

Installation options 99
Installation on an intermediate piece of S23 or EMD2 102
Installation on an existing ZDK intermediate piece 110
Installation on ceiling track mount 114

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


300 mm (11.8") ceiling panel extension 115
On-site preparations 116
Constructional requirements 118
Confirmation of the structural calculation and execution
of installation 119
Ordering data 121

Index 125

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 5

New installation of S8 / S 81 ceiling mount

Technical data 7
S8 ceiling mount, dimensions 7
S81 ceiling mount, dimensions 8
Pre-installation set 10
Installation position in the OR 12
Operating position 16
Operating range of the ceiling mount 20

On-site preparations 24
Customer's responsibilities 24
Building vibrations 26
Customer's preparatory responsibilities, overview 28

Constructional requirements 29
Forces and torques 29
Mounting a pre-installation set on ceilings with air-supply ceiling 30

Planning the electrical installation 33


Installing the conduits 34
Mounting the wall socket 36
On-site electrical installation 38
Checking the power connection 40

Planning the mechanical installation 42


300 mm (11.8") ceiling panel extension 46

Confirmation of the structural calculation and execution


of installation 47

Ordering data 48
S8 ceiling mount for ORs with false ceilings, ordering data 48
S8 ceiling mount for ORs without false ceilings, ordering data 51

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


6 New installation of S8 / S 81 ceiling mount

S81 ceiling mount for ORs with false ceilings, ordering data 54
S81 ceiling mount for ORs without false ceilings, ordering data 57

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 7

Technical data

S8 ceiling mount, dimensions

Key
1 False ceiling

620 mm / 24.4T S8
238 mm / 9.4T

1
228 mm / 9T

250 mm
9.8T

950 mm / 37.4T
950 mm / 37.4T
700 mm / 27.6T

350° ± 90° 270°

max. 1000 mm / 39.4T max. 930 mm / 36.6T

1 : 20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


8 New installation of S8 / S 81 ceiling mount

S81 ceiling mount, dimensions

Key
1 False ceiling
2 Floor
A Distance A is specifically defined for each project.
B Distance from floor
We recommend different distances from the floor, depending on the
application involved.
e ENT 1,650 mm (65")
e ophthalmology 1,800 mm (70.9")
e reconstructive and plastic surgery 1,800 mm (70.9")
e others 1,650 mm (65")

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 9

S81
238 mm / 9.4T
620 mm / 24.4T

228 mm / 9T
1
250 mm
9.8T

360 mm
14.2T
360 mm
14.2T
A max. 350° ± 90° 270°

B 600 mm / 23.6T max. 930 mm / 36.6T

2
1 : 20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


10 New installation of S8 / S 81 ceiling mount

Pre-installation set

Key
1 Stud bolt
2 Spacers
3 Ceiling anchor plate
4 Floor
5 Ceiling panel
6 Mount flange
7 False ceiling
8 Interface plate
9 Structural ceiling

A is the distance between the structural ceiling and false ceiling.


B is the distance between the floor and false ceiling.

Mount flange (6) and ceiling panel (5) are part of the ceiling mount and are
only shown here for information. They are not included in the pre-installa-
tion set.
A dummy cover is available for closing the opening in the false ceiling until
the ceiling mount is installed. Also see page 48 and page 52.
Caution:
When calculating the forces / torques applied to the structural ceiling, the
maximum load, i. e. the weight of the ceiling mount itself and the max-
imum weight of equipment attached to it, must be taken into account.
Note:
Dimension (A) must not exceed 800 mm (31.5"). Otherwise, a ceiling sub-
structure must be installed to bridge the excess distance. When designing
the substructure, make sure it is strong enough to withstand the bending
vibrations caused by the arms of the ceiling mount.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 11

1 9 8
2
3

45 mm / 1.8T
75mm / 3T

A
100 mm / 3.9T

150
75

7
620 mm / 24.4T
6
5

(1393 lbf)

(7007 lbf.ft)

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


12 New installation of S8 / S 81 ceiling mount

Installation position in the OR

The optimum installation position of the ceiling anchor plate is dependent


on the required position of the operating table in the OR. The following
overview shows the ceiling mounts in their optimum operating positions.
We recommend installing the ceiling anchor plate above the area of the
patients' feet.
Alternatively, the ceiling anchor plate can be installed laterally to the op-
erating table. Please see the illustration for the recommended distances
of the ceiling anchor plate from the surgical field.
In unfavorable on-site conditions, it may not be possible to implement the
installation in the recommended positions. Installation options for
cramped conditions are shown from page 16 onwards. OR illuminators
can only be used with restrictions or not at all in this case.

1 S8 installation above the operating table


2 S81 installation above the operating table
3 S8 installation laterally to the operating table
4 S81 installation laterally to the operating table
5 Area of possible installation positions for the ceiling anchor plate

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 13

S8 S81

1 2

1250 ± 100 mm
1500 ± 200 mm 49.2 ± 3.9T
59.1 ± 7.9T

1500 ± 200 mm 1250 ± 100 mm


59.1 ± 7.9T 49.2 ± 3.9T

3 4

5 5

1 : 50

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


14 New installation of S8 / S 81 ceiling mount

Distance from floor


1 Distance from floor
We recommend different distances from the floor, depending on the
discipline involved.
e ENT 1,650 mm (65")
e ophthalmology 1,800 mm (70.9")
e reconstructive and plastic surgery 1,800 mm (70.9")
e others 1,650 mm (65")

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 15

2800 - 3300 mm
110.2 - 129.9T

ca. 1800 mm
ca. 70.9T

S8
2650 - 3500 mm
104.3 - 137.8T

S81

1 : 50

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


16 New installation of S8 / S 81 ceiling mount

Operating position

The ceiling mounts are shown in a typical operating position. The angle
between the microscope arm and the carrier arm is 90°. At this angle, min-
imum effort is required for horizontal movement of the surgical micro-
scope.

Side view
1 S8 normal installation
2 S81 normal installation

If, due to other installations, the space available on the ceiling is not suf-
ficient for normal installation, you can use installation method 3 or 4 as an
alternative.
3 S8 installation in cramped conditions
The mount flange is installed in a position rotated through 180°. Also
see page 18. OR illuminators cannot be mounted in this case. The
same installation principle can also be used for the S81 ceiling mount
(not shown here).
4 S8 installation in cramped conditions
The brake arm is installed in a position rotated through 180°. Also see
page 18. OR illuminators cannot be mounted in this case. The same
installation principle can also be used for the S81 ceiling mount (not
shown here).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 17

S8 S81

1600 mm / 63T 1350 mm / 53.1T

1 2

S8
1100 mm / 43.3T (S8) 1400 mm / 55.1T (S8)
850 mm / 33.5T (S81) 1150 mm / 45.3T (S81)
250 mm / 9.8T

3 4

50 mm / 2T

1 : 50

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


18 New installation of S8 / S 81 ceiling mount

Top view
1 S8 normal installation
2 S81 normal installation

If, due to other installations, the space available on the ceiling is not suf-
ficient for normal installation, you can use installation method 3 or 4 as an
alternative.
3 S8 installation in cramped conditions
The mount flange is installed in a position rotated through 180°. Also
see page 16. OR illuminators cannot be mounted in this case. The
same installation principle can also be used for the S81 ceiling mount
(not shown here).
4 S8 installation in cramped conditions
The brake arm is installed in a position rotated through 180°. Also see
page 16. OR illuminators cannot be mounted in this case. The same
installation principle can also be used for the S81 ceiling mount (not
shown here).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 19

S8 S81
1 2
1600 mm / 63T 1350 mm / 53.1T

S8
1100 mm / 43.3T (S8) 1400 mm / 55.1T (S8)
850 mm / 33.5T (S81) 1150 mm / 45.3T (S81)

3 4

1 : 50

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


20 New installation of S8 / S 81 ceiling mount

Operating range of the ceiling mount

S8 ceiling mount
The operating range specified here applies to normal installation.
For installation in cramped conditions, the dimensions shown on page 16
and page 18 can be used to determine the operating range of the ceiling
mount.

1 Pivot axis 1
2 Pivot axis 2
3 Operating range for the surgical microscope

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 21

S8

1
2

1250 ± 500 mm / 49.2 ± 19.7T


250 mm / 9.8T

1 : 20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


22 New installation of S8 / S 81 ceiling mount

S81 ceiling mount


The operating range specified here applies to normal installation.
For installation in cramped conditions, the dimensions shown on page 16
and page 18 can be used to determine the operating range of the ceiling
mount.

1 Pivot axis 1
2 Pivot axis 2
3 Operating range for the surgical microscope

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 23

S81

1000 ± 400 mm
39.4 ± 15.7T

250 mm / 9.8T

1 : 20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


24 New installation of S8 / S 81 ceiling mount

On-site preparations

Customer's responsibilities

Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.

The actual load on the ceiling depends on a large number of different fac-
tors, which must be determined in detail by a structural engineer on a
case-to-case basis: see "Constructional requirements for ceiling mounts"
on page 29.
Caution:
g Make sure that a structural engineer checks the installation conditions
during the planning procedure.
Structural verification must be performed prior to the installation of the
mount.
We recommend filing the structural verification in the ceiling mount
documentation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g Please add a copy of the "Confirmation of structural calculation" to
your order (see page 47).
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the pre-installation set.
g The system should be installed on a hard, level surface, e.g. concrete.
Do not countersink the pre-installation set, but mount it directly on the
respective surface.
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
47).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 25

Warning!
g If an existing ceiling anchor plate or intermediate piece must be ex-
changed, never re-use the old anchors. New anchor holes must be
drilled.
When calculating the effective strength of the new anchors, the struc-
tural engineer must take into account the weakening effect of the old
holes in the ceiling.

Planning the installation


g Inspect the mounting components supplied for completeness and
damage.
g Prior to the installation of a ceiling mount, you must always check that
the actual installation conditions - in particular the room height - corre-
spond to the specifications in the drawing.
g Make sure that the anchors calculated by your structural engineer are
properly installed, that the nuts and washers required to mount the
ceiling anchor plate are readily available at the installation site and
that the maximum tightening torque for the nuts is indicated.
g For new installation, the pre-installation set must be mounted before
the false ceiling is attached.
g At least 2 installers are required for mounting the pre-installation set.
g Installation of a ceiling mount in rooms with an air supply ceiling re-
quires proper preparation of the air supply ceiling (see page 30).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


26 New installation of S8 / S 81 ceiling mount

Building vibrations

The on-site requirements also include the low-vibration design of the


ceiling in the OR. This must be taken into account right during the plan-
ning phase for the ceiling mount.
The following information primarily refers to the ceiling mount, but it can
also be applied to a wall mount or floor stand.
Two types of excitation factors must be distinguished:
Single events which excite short-term vibration
Induced by inadvertent knocks against the suspension system or strong
impact against ceiling, wall or floor. This is the most frequent, but least
critical type of excitation. The ZEISS ceiling mount features excellent
damping against this type of vibration and displays a short recovery time.
In extreme cases, surgery has to be interrupted for a few moments.
Constant excitation causing sustained vibration
The excitation energy of factors such as elevators, air conditioning sys-
tems, construction work, traffic does not easily reach the ceiling mount via
the building. This type of excitation is extremely rare, but may lead to per-
manent vibration of the ceiling mount in extreme cases. This becomes
particularly visible at high magnifications as used in the surgical micro-
scope.
The following result of a study is intended to help you understand constant
vibration excitation occurring in rare cases in ceiling mounts, eliminate it
or prevent it from the outset in new installations.
Like any kinematic system, the Zeiss ceiling mount displays eigenfre-
quencies (=resonance frequencies) that range between 2 and 80 Hz de-
pending on the position of the system and the accessories attached to it.
The ZEISS ceiling mount provides very effective damping in particular
against higher frequencies above 10 Hz which are typical of buildings with
electrical excitation factors. Nevertheless it may happen in rare cases,
when the ceiling mount's suspension arm is in a specific position, that
even high frequencies which are critical for this particular constellation
lead to vibrations (e.g.: 17 Hz, ..., 19 Hz, ...).
Such excitation factors can usually be suppressed by the following meas-
ures:
e elimination of the excitation source (e.g. repair or damping of the air
conditioning system)
e constructional damping measures in the ceiling installation

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 27

Due to the large number of parameters involved and the variety of poten-
tial building/ceiling mount constellations, Carl Zeiss Surgical is unable to
give an absolute guarantee for the vibration-free suspension of the ceiling
mount, even if the building meets the applicable ISO standards.
However, it is highly improbable that constant vibrations are transferred
to the ceiling mount if all requirements regarding vibrations in the OR
ceiling are met.
e Max. vibration velocity (RMS) at the installation points for the ceiling
mount.
e Vmax < 0.1 mm / s or Vmax < 4 milli-inches / s or below the curves (di-
agram) for the specified frequency range.
Sources: Carl Zeiss Surgical (in-house study), ISO 10811, recommenda-
tions for ORs.

1
RMS of vibration velocity (mm / sec)

0,1

0,01
4 5 6,3 8 10 12,5 16 20 25 31,5 40 50 63 90
One-third octaveband center frequencles (Hz)

100
RMS of vibration velocity (mili-inch / sec)

10

1
4 5 6,3 8 10 12,5 16 20 25 31,5 40 50 63 90
One-third octaveband center frequencles (Hz)

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


28 New installation of S8 / S 81 ceiling mount

Customer's preparatory responsibilities, overview

e Mounting of the pre-installation set.


e Installation and connection of the systems's power cables on the pre-
installation set.
e The power line of the ceiling mount must be protected by a 1-pole au-
tomatic circuit breaker (230 V: 10 A, 115 V: 20 A) or a 1-pole slo-blo
fuse (230 V: 10 A, 115 V: 20 A). The fusing must be installed in the
phase wire.
e A 2-pole power switch with an admissible contact load of min. 230 V:
10 A (115 V: 20 A) must then be installed. Alternatively, a power
switch can be integrated in the system as part of the instrument con-
nector option. Also see page 48 and page 52.
e Installation of the wall socket (flush or surface socket).
e Installation of conduits and cables for
e power line, 3 x 1.5 mm2 (3 x AWG16)
e wall socket for foot control panel (as required)
e potential equalization (as required)
e video, internal diameter 50 mm (2") (as required)
e OR illuminators (24 V DC with backup power supply, customer's
responsibility) (as required)
e other OPMI accessories such as BIOM/SDI, additional illumina-
tion, AUX output port, etc. (as required)
e We recommend installing further conduits, e.g. to the central OR
control panel, for future upgrades.
e The maximum admissible cable length for SpeedFokus is 10 m
(394"). If this is not sufficient, a signal amplifier must be provided.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 29

Constructional requirements

The actual load on the ceiling depends on a large number of different fac-
tors. The requirements to be met by the ceiling or substructure result from
the addition of perpendicular forces and torques produced by the suspen-
sion system and accessories. These are the forces to be transmitted into
the structural ceiling via the ceiling anchors.
The column must be aligned in a vertical position (max. deviation ± 0.5°).

Forces and torques

Warning!
The structural engineer must ensure in each individual case that the struc-
tural ceiling has a sufficient load capacity for the forces and torques listed
below. He must also take into account any additional loads on the ceiling
and add an appropriate safety margin, and must observe the applicable
national codes and regulations.
Caution:
e The perpendicular forces and torques specified below include an ad-
ditional load of 100 kg (220 lb), which is generated when a person
hangs on the end of the suspension arm (or the microscope). Further
safety margins have not been incorporated.
e The perpendicular forces and torques have been calculated on the ba-
sis of the maximum permissible load on the suspension arm.

S8 or S81 ceiling mount


For an S8 or S81 ceiling mount alone (without accessories on the acces-
sory interface), the structural ceiling must have the following load ca-
pacity:

e Perpendicular force: minimum 2500 N (565 lbf)


e Torque: minimum 3000 Nm (2215 lbf.ft)

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


30 New installation of S8 / S 81 ceiling mount

Mounting a pre-installation set on ceilings with air-supply ceiling

Caution:
If the carrier arm systems are installed in air-supply ceilings, the method
of installation must be specifically adapted to the constructional features
of the type of air-supply ceiling involved.
A false ceiling set is mounted to bridge the distance between the struc-
tural ceiling and the non-structural false ceiling. The length of the stud
bolts and spacers used is adapted on site as required.
The necessary planning should be performed in consultation with the
manufacturer of the air-supply ceiling and Carl Zeiss.
When ordering, please specify distance (A), i.e. the distance between the
lower surface of the structural ceiling and the lower surface of the false
ceiling.
Note:
e Distance (A) is the height of the false ceiling and must not exceed
800 mm. Otherwise, a ceiling substructure must be installed to bridge
the excess distance. Distance (Z) is the height of the air-supply ceiling.
e It is essential that the false ceiling set (2,3) and the conduits for the
power lines be installed before the air-supply ceiling is attached. Sub-
sequent installation is not possible, for reasons of space.
e A manhole (5) which is free from compressed air must be available for
the installation work. The internal diameter (S) of the manhole must be
larger than that of the opening in the ceiling (M).
e The opening of the manhole to the OR can be closed with a separate
cover (7) by the manufacturer of the air-supply ceiling.
1 Structural ceiling
2 Ceiling anchor plate
3 Spacers
4 Upper surface of air-supply ceiling
5 Manhole
6 Interface plate
7 Cover of carrier arm system to OR (customer's responsibility)
8 Lower surface of air-supply ceiling
9 Ceiling panel

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 31

1
2

A 4
S
5
Z
6
7
8
9

2
3
4
6
5
7
8

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


32 New installation of S8 / S 81 ceiling mount

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 33

Planning the electrical installation

The power lines on the site of installation must be run up to the interface
plate. In addition, the potential equalization lines must be run to the termi-
nals on the interface plate.
Warning!
e The planning, execution and inspection of the on-site electrical instal-
lations must be performed by expert electrical technicians and official-
ly registered electrical firms.
e The electrical installations of the room concerned must comply with
the requirements of the applicable national regulations. In the Federal
Republic of Germany, VDE 0107 applies.
e The number of circuits is dependent on the customer's specific carrier
system configuration.
e If several circuits are required by the customer, appropriate allowance
must be made on the interface plate.
e Multiwire lines must be provided with wire end sleeves.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


34 New installation of S8 / S 81 ceiling mount

Installing the conduits

The wiring can be run either above the false ceiling or in conduits installed
in the structural ceiling. There are two options for the installation of con-
duits in the structural ceiling.
e The exit of the conduits is positioned beside ceiling anchor plate (8).
e The exit of the conduits is positioned at the center of ceiling anchor
plate (8).
The conduits have an internal diameter of 22 to 50 mm (0.9 - 2").
A conduit with a minimum internal diameter of 35 mm (1.4") is required for
the MediLive video line.

Key
1 Power line
3 x 1.5 mm2
2 Wall socket for foot control panel
As required.
3 Potential equalization
As required.
4 TV internal diameter 50 mm (2")
As required.
5 OR illuminators
24 V DC with backup power supply, customer's responsibility (as re-
quired)
6 Other OPMI accessories
e. g. BIOM/SDI, additional illumination, AUX output port, etc. (as re-
quired)
7 Further conduits
We recommend installing further conduits for future upgrades.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 35

Connection of SpeedFokus AF
Note:
g Determine the exact length between the Speedfokus AF connection
and the CCU+ monitor.
If the cable length exceeds 10 m (394"), a signal amplifier is required for
the SpeedFokus AF. The signal amplifier requires 110 - 230 V line power.

1-7 8 1-7 9

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


36 New installation of S8 / S 81 ceiling mount

Mounting the wall socket

Note:
e We recommend mounting wall socket (3) on the wall located behind
the surgeon. Height approx. 1.4 m (55.1").
e Please also note the chapter "Installing the conduits" on page 34.

Kit 1078-524 contains the following components for mounting the


wall socket:
e Wall socket for flush mounting
e Wall socket for surface mounting
e 10 m (394") cable
e Hanger for foot control panel
g Use the wall socket which is more suitable for your purposes.

Key
1 Wall socket for surface mounting
2 Wall socket for flush mounting
3 Wall socket
4 Central OR control panel (part of customer's installations)
5 Foot control panel
6 Remote control socket for an external signal of a maximum of 24 V /
0.5 A.

Remote control socket (6)


2
View on 1
connector side 2
3 1
3

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 37

6 6

1 2

3 4
1400 mm / 55.1T

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


38 New installation of S8 / S 81 ceiling mount

On-site electrical installation

The power lines on the site of installation must be run up to the interface
plate. In addition, the potential equalization lines must be run to the termi-
nals on the interface plate.
Warning!
e The planning, execution and inspection of the on-site electrical instal-
lations must be performed by expert electrical technicians and official-
ly registered electrical firms.
e The electrical installations of the room concerned must comply with
the requirements of the applicable national regulations. In the Federal
Republic of Germany, VDE 0107 applies.
e The number of circuits is dependent on the customer's specific carrier
system configuration.
e If several circuits are required by the customer, appropriate allowance
must be made on the interface plate.
e Multiwire lines must be provided with wire end sleeves.

Key
A Connectors for accessory modules (OR illuminators, monitor bracket)
For details of terminals B and C, see the section "Accessory interface",
page 71.
D Connectors for potential equalization
Maximum terminal cross sections
Terminals 1 - 4 = 16 mm2 (AWG 6)
Terminals 5 - 8 = 6 mm2 (AWG 10)
E On-site connector
F Ceiling mount connectors
G High-current connectors
Maximum terminal cross section = 4 mm2 (AWG 12)
Connector (part no. 6 001 215) with the contacts (H): L1, N1, PE1;
L2, N2, PE2 and L3, N3, PE3.
H Terminal numbers
I Circuit numbers

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 39

L1 L1
B L2 L2
PE PE

L1 L1
C L2 L2
PE PE

A D G

E
6 PE PE 7
5 PE PE 8
F
4 PE PE 1
D 3 PE PE 2

H I

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


40 New installation of S8 / S 81 ceiling mount

Checking the power connection

Caution:
The ceiling mount can be set at two points to the rated voltage used in the
country of destination:
e Transformer terminal on interface plate.
Settings: 100 - 115 - 230 V
e Power supply unit
Settings: 115 - 230 V
The correctness of the factory-adjusted setting is checked by our service
team during first-time installation.
e The power connection must conform to the regulations applicable in
the country of destination.
e A power switch (2-pin) has been provided.
e Line fuse
230 V: min. 10 A, slo-blo, class C
115 V: min. 20 A, slo-blo, class C

e All cables and plugs must be in perfect condition.


e Potential equalization: if required, the instrument can be incorporated
into protective "potential equalization" measures.
e A connector for OR illuminators is provided (option).
We recommend integrating the power switch in the central OR control
panel.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 41

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


42 New installation of S8 / S 81 ceiling mount

Planning the mechanical installation

Mounting the pre-installation set


Note:
g Check the installation conditions, in particular with respect to the room
height.
g Mount the pre-installation set, before the false ceiling is attached.
Two installers are required for mounting the pre-installation set. The
ceiling anchor plate alone weighs approx. 40 kg.
The pre-installation set comprises the following modules:
1 Ceiling anchor plate
2 Spacers (8 units)
3 Interface plate
4 Line power connection label (1167-817).
Line power connection label (4)
Line power connection label (1167-817)
identifies the line power terminals. The
installer attaches the label when
mounting the pre-installation set.

Auxiliaries required:
e Two sufficiently long ladders and a lifting device (lift table).
e Torque wrench: 0-120 Nm (minimum).
e Open-end wrenches and drive sockets for torque wrench:
24 mm and 55 mm (0.9" und 2.2").
e Anchors as specified by the structural engineer.
e Sheet specifying dimensions (supplied with the pre-installation set).

Mounting the ceiling anchor plate


The ceiling anchor plate weighs approx. 40 kg. Two installers are required
for mounting it.
g Measure the position in accordance with the plan and make sure that
one side of the ceiling anchor plate is parallel to the nearest wall.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 43

mm Inch
1
605 23.8
517 20.4
25 1.0
Ø400 Ø15.7
Ø560 Ø22.0
Ø26 Ø1.0

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


44 New installation of S8 / S 81 ceiling mount

Mounting the spacers


g Mount eight spacers (2) in the ceiling anchor plate (1).
g Firmly tighten the spacers (2) using a torque wrench with a 55 mm
(2.2") fork wrench bit. Apply a torque of 120 Nm.
g Adjust the length of the stud bolts (3) in such a way that the distance
between the lower surface of the false ceiling and the lower edge of
the stud bolt is 75 mm (3").
Note:
The mount flange (4) is only shown here for information. It is not included
in the pre-installation set. The mount flange (4) is part of the ceiling mount.

2
75 mm / 3"

3 4

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 45

Mounting the interface plate


Note:
The connector for potential equalization (1) and the terminals for power
line (2) are provided on interface plate (3).
g Check whether interface plate (3) corresponds to the drawing sup-
plied with the pre-installation set (reference no.).
g Remove nuts M16 (SW 24) (5) from all stud bolts.
g Slide interface plate (3) on the stud bolts in such a way that one side
is parallel to the nearest wall. The terminals for power line (2) must
point downward. Screw nuts M16 (SW 24) (5) onto the stud bolts.
g Tighten nuts M16 (SW 24) (5) with a torque of 70 Nm (51.6 lbf.ft).
Note:
The mount flange (4) is only shown here for information. It is not included
in the pre-installation set. The mount flange (4) is part of the ceiling mount.

1 2

3 5

75 mm
3"

3 4

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


46 New installation of S8 / S 81 ceiling mount

300 mm (11.8") ceiling panel extension

Ceiling panel extension (1) is required if the distance between the struc-
tural ceiling and the false ceiling (i.e. dimension "A" in the order forms) is
clearly shorter than 200 mm (7.9"), i.e. < 190 mm (<7.5"). In this case,
ceiling panel (2) alone will not suffice to completely cover pre-installation
set (3).
Ceiling panel extension (1) cannot enclose interface plate (5). If interface
plate (5) also protrudes from the false ceiling, you must order the "Cover
for pre-installation set" (Cat. No. 1139-023).
You can order a 300 mm (11.8") ceiling panel extension including
mounting instructions (1177-632) as a spare part from the Zeiss Service
Department.
The installer will cut the 300 mm (11.8") ceiling panel extension to the
length required on site and install it as per the mounting instructions
(1177-632).

Key
1 Ceiling panel extension 300 mm (11.8") (1177-632)
2 Ceiling panel for S8 / S81 ceiling mounts (1083-804)
3 Pre-installation set (1078-181)
4 False ceiling
5 Interface plate
6 Structural ceiling

1 2 3

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 47

Confirmation of the structural calculation and execution


of installation

Sales order no.: ...........................................................................................

Customer address / ...........................................................................................


Delivery address: ...........................................................................................
...........................................................................................

By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:

Name and address of ...........................................................................................


the structural engineer: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

The installer for


the proper mounting of the pre-installation set or the
ceiling or wall flange from Carl Zeiss:
Name and address of ...........................................................................................
the executing company ...........................................................................................
and installerrs name: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


48 New installation of S8 / S 81 ceiling mount

Ordering data

S8 ceiling mount for ORs with false ceilings, ordering data

Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.

Description Cat.No.
S8 ceiling mount 1176-968
Pre-installation set for new installation 1078-181
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804

A Distance between structural ceiling and false ceiling


(ordered length of pre-installation set)
B Distance between false ceiling and floor

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 49

Order sheet for S8 ceiling mount in ORs with false ceilings

mm

max 800 mm
min 200 mm
A

Dimensions in inches:

mm
A= "

max 3300 mm
min 2800 mm
min 7.9", max 31.5"
B
B= "
min 110.2", max 129.9"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


50 New installation of S8 / S 81 ceiling mount

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 51

S8 ceiling mount for ORs without false ceilings, ordering data

You must order the ceiling mount using the following order sheet only if
the OR does not provide a false ceiling.
If the OR does not provide a false ceiling, you will need a cover for the pre-
installation set (1078-181). You can order the cover for the pre-installation
set as a kit (1139-023).

Description Cat.No.
S8 ceiling mount 1176-968
Pre-installation set for new installation 1078-181
Cover for pre-installation set 1139-023
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


52 New installation of S8 / S 81 ceiling mount

Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.

H Room height
Distance from the floor to the structural ceiling.
g You must exactly measure this height.
I Required installation height of the S8 ceiling mount
The installation height of the S8 ceiling mount is the distance from the
floor to the bottom surface of the mount flange.
g You must preselect the installation height.
Note:
The installation height can be between 2,800 mm (110.2") and 3,300
mm (129.9"). Ideally, you should select an installation height of
3,100 mm (122") which is the optimum compromise between working
height and head room.
A Ordered length of the pre-installation set
can be determined by subtracting the installation height (I) selected
and 45 mm (1.8") from the room height (H) measured.
F Length to which cover (1139-023) must be cut
can be determined by subtracting the installation height (I) selected
and 52 mm (2") from the room height (H) measured.

1 Structural ceiling, bottom surface


2 Cover of the pre-installation set (1139-023)
3 Ceiling panel (1083-804)
4 Mount flange, bottom surface
5 Floor, top surface

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 53

Order sheet for S8 ceiling mount in ORs without false ceilings

1
F
2

Length of pre-installation set


to be ordered:

A = H - I - 45 mm
mm

A= mm
min 200 mm, max 800 mm
Dimensions in inches:
H
H= "
mm

Length to which the cover


I = "
must be cut:
Length of pre-installa-
I tion set to be ordered: F = H - I - 52 mm
A = H - I - 1.8"
F= mm
A= "
min 197 mm, max 800 mm
min 7.9", max 31.5"
Length to which the
cover must be cut:
F = H - I - 2"

F= " 5
min 7.9", max 31.5"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


54 New installation of S8 / S 81 ceiling mount

S81 ceiling mount for ORs with false ceilings, ordering data

Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.

Description Cat. No.


S81 ceiling mount 1176-969
Pre-installation set for new installation 1078-181
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804
A Distance between structural ceiling and false ceiling
(ordered length of pre-installation set)
B Distance between false ceiling and floor
C Distance from floor
We recommend different distances from the floor, depending on the
application involved.
e ENT 1650 mm (65")
e ophthalmology 1800 mm (70.9")
e reconstructive and plastic surgery 1800 mm (70.9")
e others 1650 mm (65")
D Column length (ordered length)
The following formula is used for calculation:
D = B - C - 810 mm (31.9").
Note: column length (D) >600 mm (>23.6T) Depending on the type
of construction involved, vibrations may have a negative influence if
the column length (D) exceeds 600 mm (>23.6t).

g Use column lengths of >600 mm (>23.6t) only in exceptional cases. If


on-site conditions permit it, column lengths of >600 mm (>23.6t) can
be avoided by selecting an appropriately greater height (A) of the pre-
installation set. Should the pre-installation set protrude from the false
ceiling, you can cover it using ceiling panel extension (1177-632).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 55

Order sheet for S81 ceiling mount in ORs with false ceilings

mm

max 800 mm
min 200 mm
A

Column length (ordered length):

D = B - C - 810 mm

D= mm
D min. 40 mm, max. 800 mm*)
*) Recommendation: max. 600 mm,
see note on page 54.
mm

Dimensions in inches:
C A= "

mm
min. 7.9", max. 31.5"
B= "
C= " B
Column length:
D = B - C - 31.9t
D= "
min. 1.6t, max. 31.5t *)
*) Recommendation:
max. 23.6",
see note on page 54.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


56 New installation of S8 / S 81 ceiling mount

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 57

S81 ceiling mount for ORs without false ceilings, ordering data

You must order the ceiling mount using the following order sheet only if
the OR does not provide a false ceiling.
If the OR does not provide a false ceiling, you will need a cover for the pre-
installation set (1078-181). You can order the cover for the pre-installation
set as a kit (1139-023).

Description Cat.No.
S81 ceiling mount 1176-969
Pre-installation set for new installation 1078-181
Cover for pre-installation set 1139-023
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


58 New installation of S8 / S 81 ceiling mount

Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
Caution:
Use the following principle for planning the installation of the S81 ceiling
mount: Keep the ordered length (A) of the pre-installation set as short as
possible and select a longer column length (D) as required.
C Required distance from floor
We recommend defining different distances from the floor, depending
on the application involved: ENT 1650 mm (65"), ophthalmology
1800 mm (70.9"), reconstructive and plastic surgery 1800 mm (70.9"),
others 1650 mm (65").
H Room height (distance from the floor to the structural ceiling)
g You must exactly measure this height.
A Ordered height of pre-installation set
Start your planning using the shortest length of 200 mm (7.9").
D Column length (ordered length)
Calculate the column length: D = H - A - C - 810 mm (-31.9").
The column length may be a minimum of 40 mm (1.6") and a maxi-
mum of 800 mm (31.5").
Note: column lengths of >600 mm (>23.6T)
Depending on the type of construction involved, vibrations may have
a negative influence if the column length (D) exceeds 600 mm
(>23.6t).
g Use column lengths of >600 mm (>23.6t) only in exceptional cases. If
on-site conditions permit it, column lengths of >600 mm (>23.6t) can
be avoided by selecting an appropriately greater height (A) of the pre-
installation set.
Example: Use the maximum column length (D) of 600 mm (23.6") and
calculate the length (A) of the pre-installation set to be ordered using
the formula: A = H - D - C - 810 mm (-31.9").
F Length to which cover (F) (1139-023) must be cut
Calculate length (F) to which the cover must be cut:
F = A - 7 mm (-0.3").

1 Structural ceiling, bottom surface


2 Cover of the pre-installation set (1139-023)
3 Ceiling panel (1083-804)
4 Mount flange, bottom surface
5 Floor, top surface

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


New installation of S8 / S 81 ceiling mount 59

Order sheet for S81 ceiling mount in ORs without false ceilings

1
A F
2

3
4

D
mm

Dimensions in inches:
Pre-installation set, ordered length (A)
A= "
H A=
min 7.9", max 31.5" mm
C= " min 200 mm, max 800 mm

H= "
Column length, ordered length (D):
Column length (D):
mm

D = H - A - C - 810 mm
D = H - A - C - 31.9"

C
D= " D= mm
min. 1.6", max. 31.5" *) min. 40 mm, max. 800 mm*)
*) Recommendation: *) Recommendation: max. 600 mm,
max. 23.6", see see note on page 58.
note on page 58.
Length (F) to which the Length (F) to which the cover must
cover must be cut: be cut:

F = A - 0.3" F = A - 7 mm.
5 F= "
F= mm
min 7.9", max 31.5"
min 193 mm, max 800 mm

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


60 New installation of S8 / S 81 ceiling mount

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 61

Accessory interface

Accessory interface 62
Coupling of the Zeiss system with accessories from
other manufacturers 62

Installation position 63

Mount flange and ceiling panel 64

Mount flange (flange plate) 66

Configuration example: mount flange with flange tube 67

Pre-installation set 68

Customer's responsibilities 69

Electrical connections for accessory modules 71

Confirmation of the structural calculation and execution


of installation 73

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


62 Accessory interface

Accessory interface

Next to the flange area for the carrier arm of the ceiling mount, the mount
flange also features a flange area for an accessory carrier. This is the ac-
cessory interface designed as a hole circle, see page 66.
This accessory interface permits you to mount an accessory carrier e.g.
for OR illuminators, supports, monitor brackets, trays for video systems or
similar devices.

Coupling of the Zeiss system with accessories from other manufacturers

Caution:
The coupling of a system from Carl Zeiss Surgical GmbH with accesso-
ries (OR illuminators, monitor brackets or similar) from other manufac-
turers results in a new medical system for which the creator of the new
system must meet the necessary requirements (approval, qualification,
structural requirements, etc.).
Please note the user manuals provided by the manufacturer of the acces-
sory modules. Further information is available from our service depart-
ment or from authorized representatives.
g Check the feasibility of the installation in detail together with the man-
ufacturer of the relevant accessory module (OR illuminator or monitor
bracket).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 63

Installation position

Caution:
When defining the positions and ranges of movement of the accessory
equipment, please pay attention to the geometry of the respective ceiling
mount and to the risk of collisions. For the dimensions, please see:
e illustration on page 7 for S8 ceiling mount.
e illustration on page 8 for S81 ceiling mount.
Note:
The mount flange is usually installed 45 mm (1.8") below the false ceiling.
For the dimensions of the pre-installation set, please see the illustration
on page 68.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


64 Accessory interface

Mount flange and ceiling panel

The S8 or S81 ceiling mount is secured with mount flange (1) on the pre-
installation set already installed in the ceiling. In addition to the flange
area for the carrier arm of the ceiling mount, the mount flange also fea-
tures a flange area for an accessory carrier e.g. for attaching an OR illu-
minator and/or monitor bracket. The mount flange is part of the ceiling
mount and is included in its delivery package.
Ceiling panel (2) is used to cover the mount flange and the entire opening
in the false ceiling. The opening in the false ceiling must not be larger than
600 mm x 600 mm (23.6" x 23.6"). Two openings are provided in the
ceiling panel, one for the ceiling mount and one for an accessory carrier.
Cover frame (3) closes the remaining space between the square tube of
the ceiling mount and the ceiling panel.
Blind cover (4) is supplied together with the ceiling panel, permitting the
opening for the accessory carrier to be completely closed. You can make
an opening in blind cover (4) which fits the accessory carrier used.
A dummy cover is available for closing the opening in the false ceiling until
the ceiling mount is installed. Also see page 48 and page 52.
Note:
The S8 or S81 ceiling mount is delivered with the following components
already mounted: mount flange (1), cover frame (2) and carrier arm of the
ceiling mount (5), as in the configuration example on page 67.

Key
1 Mount flange, part of the ceiling mount.
2 Ceiling panel consisting of two pieces which are fitted together during
installation.
3 Cover frame, closes the remaining space between the carrier arm of
the ceiling mount (square tube) and the ceiling panel.
4 Blind cover, completely closes the opening for the flange tube in the
ceiling panel. You can make an opening in the blind cover as required
for the accessory carrier used.
5 Upper part of the carrier arm of the S8 and S81 ceiling mounts, de-
signed as a square tube.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 65

2 2

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


66 Accessory interface

Mount flange (flange plate)

1 Accessory interface (flange area for an accessory carrier)


with 6 tapped holes M12 in a hole circle of 270 mm (10.6") dia.
Accessories (illuminators, monitor bracket etc.) can be mounted here
using an accessory carrier, e.g. a flange tube.
2 Hole circle with 8 bores of 17 mm (0.7") dia.
This is the interface for the pre-installation set. The hole circle is used
to mount the flange on the stud bolts of the pre-installation set.
3 Flange area for the carrier arm of the ceiling mount
The flange area contains four bores arranged in a square. On delivery,
the carrier arm of the ceiling mount (square tube) is attached to this
flange surface.

Mount flange, bottom view

1 2 3 mm Inch
Ø17 Ø0.7
Ø270 Ø10.6
Ø160 Ø6.3
Ø165,5 Ø6.52
123 4.8
150 5.9

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 67

Configuration example: mount flange with flange tube

1 Mount flange
2 Ceiling panel
Only one piece of the ceiling panel is shown.
3 Flange tube (example)
A flange tube of this type can be used for mounting accessories (illu-
minators, monitor mount, etc.).
4 Upper part of the carrier arm
The upper part of the carrier arm of the S8 and S81 ceiling mounts
consists of a square tube.

Mount flange with flange tube, bottom view

3 4

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


68 Accessory interface

Pre-installation set

1 Stud bolt
2 Spacers
3 Ceiling anchor plate
4 Floor
5 Ceiling panel
6 Mount flange
7 False ceiling
8 Interface plate
9 Structural ceiling
10 Dimension A
is the distance between the structural ceiling and false ceiling.

Mount flange (6) and ceiling panel (5) are part of the ceiling mount and are
only shown here for information. They are not included in the pre-installa-
tion set.

1 9 8
2
3
45 mm / 1.8T
75mm / 3T

A
100 mm / 3.9T

150
75

7 10
620 mm / 24.4T
6
5

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 69

Customer's responsibilities

Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.

The actual load on the ceiling depends on a large number of different fac-
tors, which must be determined in detail by a structural engineer on a
case-to-case basis: see "Constructional requirements for ceiling mounts"
on page 29.
Caution:
g Your structural engineer must separately take into account the addi-
tional perpendicular force and additional torque produced by accesso-
ries mounted on the accessory interface, i.e. he must add them to the
perpendicular force and torque caused on the ceiling by the ceiling
mount alone.
g Make sure that a structural engineer checks the installation conditions
during the planning procedure.
Structural verification must be performed prior to the installation of the
mount.
We recommend filing the structural verification in the ceiling mount
documentation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g Please add a copy of the "Confirmation of structural calculation" to
your order (see page 73).
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the pre-installation set.
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
73).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


70 Accessory interface

Forces and torques


Warning!
The structural engineer must ensure in each individual case that the struc-
tural ceiling has a sufficient load capacity for the forces and torques listed
below. He must also take into account any additional loads on the ceiling
and add an appropriate safety margin, and must observe the applicable
national codes and regulations.
Note:
The maximum permissible perpendicular force and the maximum permis-
sible torque for the accessory interface result from the difference between
the maximum permissible values for the pre-installation set and ceiling
mount, i.e. pre-installation set - ceiling mount = accessory interface.
Pre-installation set
The pre-installation set can accept the following maximum perpendicular
force and maximum torque:

e Perpendicular force: max. 6200 N (min. 1393 lbf)


e Torque: max. 9500 Nm (max. 7007 lbf.ft)

Since the pre-installation set is subjected to the following maximum per-


pendicular force and maximum torque by the ceiling mount:

e perpendicular force: min. 2500 N (min. 565 lbf)


e torque: min. 3000 Nm (min. 2215 lbf.ft)

the following maximum perpendicular force and maximum torque remain


for the accessories:
Accessory interface (maximum)

h Perpendicular force: max. 3700 N (max. 832 lbf)


h Torque: max. 6500 Nm (max. 4794 lbf.ft)

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 71

Electrical connections for accessory modules

The power supply for the electrical accessory components can be con-
nected to connectors "A" on the interface plate (connectors for OR illumi-
nators, terminal blocks B and C).

On-site requirements for the connection of accessories


g Route the power lines in the building up to the interface plate.
g Route any required control lines and/or monitor connections in the
building up to the interface plate.
g Make sure (for in-wall installation) that a sufficient number of conduits
with sufficient cross section are provided.
g Route additional lines for potential equalization to the terminals on the
interface plate.
Warning!
e The planning, execution and inspection of the on-site electrical instal-
lations must be performed by expert electrical technicians and official-
ly registered electrical firms.
e The electrical installations of the room concerned must comply with
the requirements of the applicable national regulations. In the Federal
Republic of Germany, VDE 0107 applies.
e The number of circuits is dependent on the customer's specific carrier
system configuration.
e If several circuits are required by the customer, appropriate allowance
must be made on the interface plate.
e Multiwire lines must be provided with wire end sleeves.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


72 Accessory interface

Key
A Connectors for OR illuminators
B Terminal 1 for OR illuminator 1
Maximum terminal cross section = 16 mm2 (AWG 6)
C Terminal 2 for OR illuminator 2
Maximum terminal cross section = 16 mm2 (AWG 6)
For details of the other terminals, see the section "On-site electrical instal-
lation", page 38.

L1 L1
B L2 L2
PE PE

L1 L1
C L2 L2
PE PE

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Accessory interface 73

Confirmation of the structural calculation and execution


of installation

Sales order no.: ...........................................................................................

Customer address / ...........................................................................................


Delivery address: ...........................................................................................
...........................................................................................

By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:

Name and address of ...........................................................................................


the structural engineer: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

The installer for


the proper mounting of the pre-installation set or the
ceiling or wall flange from Carl Zeiss:
Name and address of ...........................................................................................
the executing company ...........................................................................................
and installerrs name: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


74 Accessory interface

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 75

Discipline-specific installation positions

Installation positions for ophthalmology and ENT 77

Installation position for P&R, Hand 90

Installation position for Neuro, Spine, Uro and Gyn 93

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


76 Discipline-specific installation positions

The various surgical disciplines and on-site conditions require different in-
stallation positions of the S8 and S81 ceiling mounts.
This chapter describes a number of recommended installation positions
of the S8 and S81 ceiling mounts for different surgical disciplines.
Note:
In unfavorable on-site conditions, the recommended installation positions
must be changed appropriately. In cramped conditions, the use of OR il-
luminators is only possible with restrictions or indeed not at all.
Caution:
The recommended discipline-specific installation positions do not release
you from the responsibility of careful planning. Only careful planning will
ensure a smooth, collision-free workflow during surgery.

Scale 1:20
All recommended, discipline-specific installation positions are shown at a
scale of 1:20 and can usually be directly incorporated in the layout plan.

Transparent film
It may also be helpful for your planning work if you copy the scale 1:20
discipline-specific installation positions to transparent film. You can then
place the transparent film on the layout plan, and bring it into a suitable
position by shifting and turning it.
Note:
In the following illustrations, the representation of the ceiling mount in light
gray, dotted lines shows the alternative, mirrored (laterally reversed) in-
stallation option.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 77

Installation positions for ophthalmology and ENT

The S8 or S81 ceiling mount is used for ophthalmology and ENT.

Recommendations for ophthalmology


Preferred installation position for ophthalmology
The S8 or S81 ceiling mount is positioned above the patient's foot area
(centered relative to the operating table).
Alternative installation position for ophthalmology
The S8 or S81 ceiling mount is positioned laterally to the patient (centered
relative to the patient's head).

Recommendation for ENT


General
Two OR illuminators are normally used for ENT. Do not mount these OR
illuminators (ceiling lights) on the mount flange of the ceiling mount, but
on separate ceiling flanges. Otherwise, collisions are liable to occur if both
ceiling lights are pointed at the surgical field at an angle from behind the
surgeon.
Preferred installation position for ENT
The S8 or S81 ceiling mount is centered above the operating table.
The OR illuminators are not mounted on the flange of the ceiling mount,
but on separate ceiling flanges.
Alternative installation position for ENT
The S8 or S81 ceiling mount is installed in a centered position behind the
patient's head.
Again, the OR illuminators are not mounted on the flange of the ceiling
mount, but on separate ceiling flanges.

Key for the following drawings


1 Pivot point of the ceiling mount
2 Flange area for an accessory carrier

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


78 Discipline-specific installation positions

Preferred installation position for ophthalmology and ENT


S8 ceiling mount positioned above the patient's foot area (centered rela-
tive to the operating table). Flange area for accessory carrier (2) in out-
ward position.

S8

.9 m
68 m
T
R 750
1
R

R 75
0
R 29 mm
.5T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 79

Preferred installation position for ophthalmology and ENT


S8 ceiling mount positioned above the patient's foot area (centered rela-
tive to the operating table). Flange area for accessory carrier (2) in inward
position.

S8

.9 m
68 m
T
R 7 50
1
R

R7
50
R 2 mm
9 .5 2
T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


80 Discipline-specific installation positions

Preferred installation position for ophthalmology and ENT


S81 ceiling mount positioned above the patient's foot area (centered rel-
ative to the operating table). Flange area for accessory carrier (2) in out-
ward position.

S81

m
m
0
0 .1T
14 5 5
R R

R 600
m
R 23.6 m
T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 81

Preferred installation position for ophthalmology and ENT


S81 ceiling mount positioned above the patient's foot area (centered rel-
ative to the operating table). Flange area for accessory carrier (2) in in-
ward position.

S81

.1 m
55 m
T
R 00
14
R

R 600 mm 2
R 23.6T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


82 Discipline-specific installation positions

Alternative installation position for ophthalmology


S8 ceiling mount positioned laterally to the patient (centered relative to the
patient's head). Flange area for accessory carrier (2) in outward position.

S8

1
0 mm
.5T
R 29
R 75
m
0m
R 1 . 9t
8
5

R6
7
.9T m
0m
75
68
R1
R

m
0m
.5t
R 29
R 75

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 83

Alternative installation position for ophthalmology


S8 ceiling mount positioned laterally to the patient (centered relative to the
patient's head).Flange area for accessory carrier (2) in inward position.

S8

R
17
R 50
68 m
.9 m
T
1 2 2 1
m
5 0mt
17 . 9
R R 68

m
R

m
75

T
R

75

.5
29

29
R
m
.5

R
t

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


84 Discipline-specific installation positions

Alternative installation position for ophthalmology


S81 ceiling mount positioned laterally to the patient (centered relative to
the patient's head). Flange area for accessory carrier (2) in outward posi-
tion.

S81

R1
40
0m
R m
55
.1T
m
0m
R 60 3.6t
R2

R 14
00 m m
m 00 m
R 55 R6
. 1t 3 .6 t
R2

2 1 1 2

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 85

Alternative installation position for ophthalmology


S81 ceiling mount positioned laterally to the patient (centered relative to
the patient's head). Flange area for accessory carrier (2) in inward posi-
tion.

S81

1 2 2 1
R1
40
0m m
R5
5.1
m
4 00m
T R 1 55.1t
R

R6
00 mm
R 2 mm 00
3.6 R 6 23. 6T
t R

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


86 Discipline-specific installation positions

Alternative installation position for ENT


S8 ceiling mount positioned behind the patient's head. Flange area for ac-
cessory carrier (2) in outward position.

S8

R
75
1

0
R

m
29

m
.5
T
R1
R6
750
8 .9
mm
T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 87

Alternative installation position for ENT


S8 ceiling mount positioned behind the patient's head. Flange area for ac-
cessory carrier (2) in inward position.

S8
2
R
7
R 50
29 m
.5 m
T

R
17 8. 9
R
50 T
6
mm

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


88 Discipline-specific installation positions

Alternative installation position for ENT


S81 ceiling mount positioned behind the patient's head. Flange area for
accessory carrier (2) in outward position.

2
S81

R
60
R 0m
23
.6T m

R
14 5.1T
00
R
5

m
m

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 89

Alternative installation position for ENT


S81 ceiling mount positioned behind the patient's head. Flange area for
accessory carrier (2) in inward position.

S81
1

R6
2 00
R 2 mm
3.6
T

R
14
00
R
55

m
m
.1
T

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


90 Discipline-specific installation positions

Installation position for P&R, Hand

The S8 ceiling mount is used for P&R and hand surgery.


The installation positions recommended in this chapter are also suitable
for the following applications:
e ENT
e Neurology
e Urology
e Oral and maxillofacial
e Spine (limited suitability).

Note:
In P&R surgery, one or the other of the patient's hand may sometimes be
difficult to reach. In such cases, there is usually the possibility of turning
the operating table in such a way that the hand concerned is brought
within the optimum working range of the ceiling mount.

Key for the following drawings


1 Pivot point of the ceiling mount
2 Flange area for an accessory carrier

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 91

Recommended installation position for P&R, Hand


S8 ceiling mount positioned laterally to the operating table, approximately
midway along the table. Flange area for accessory carrier (2) in outward
position.
Preferable if OR illuminators are mounted separately.

S8

2 1 1 2

0 mm
R 75 .5T R 75
0m
R 29 R 29 m
.5t

R1
m 7
5 0m R 50 m
7 t 68
R 1 6 8 .9 .9T m
R

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


92 Discipline-specific installation positions

Recommended installation position for P&R, Hand


S8 ceiling mount positioned laterally to the operating table, approximately
midway along the table. Flange area for accessory carrier (2) in inward
position.
Preferable if OR illuminators are mounted on the flange tube.

S8

m
5 0mT
7 .9
R 1 68
R

R
75 9.5t
R
m

0m
2
m
0

m
.5
75
29
R
R

m
1 2 5 0mt 2 1
17 .9
R R 68

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 93

Installation position for Neuro, Spine, Uro and Gyn

The S8 ceiling mount is used for the disciplines of neurology, spine, uro-
logy and gynecology.
The installation positions recommended in this chapter are also suitable
for ENT, with certain restrictions.

Key for the following drawings


1 Pivot point of the ceiling mount
2 Flange area for an accessory carrier

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


94 Discipline-specific installation positions

Recommended installation position for Neuro, Spine, Uro and Gyn


S8 ceiling mount positioned laterally to the operating table, approximately
midway along the table. Flange area for accessory carrier (2) in outward
position.

S8

T
mm
2
9.5
R2
7 50
R
mm
50 t
R 1 68.9
6 8 mm
.9T

R
7
R 50
17
R

29 mm
t

1
R 50
.5
7
R

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Discipline-specific installation positions 95

Recommended installation position for Neuro, Spine, Uro and Gyn


S8 ceiling mount positioned laterally to the operating table, approximately
midway along the table. Flange area for accessory carrier (2) in inward
position.
Preferable if OR illuminators are mounted on the flange tube.

S8

2
mm
7 50 T
R 29.5
R
mm
50 t
R 1 68.9
.9 m
68 m
T

R
R 750

7
1
R

m
m
0
75 9.5t
R 2
R

M 1:20
1

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


96 Discipline-specific installation positions

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 97

Installation on an existing substructure

Installation options 99
Working ranges 99
Mounting options 100
Adapter plate (1071-924) 100

Installation on an intermediate piece of S23 or EMD2 102


Mounting the adapter plate in an offset position 102
Mounting the adapter plate in a centered position 106

Installation on an existing ZDK intermediate piece 110


Mounting electrical components on the mounting bracket 110

Installation on ceiling track mount 114

300 mm (11.8") ceiling panel extension 115

On-site preparations 116


Customer's responsibilities 116

Constructional requirements 118


Forces and torques 118

Confirmation of the structural calculation and execution


of installation 119

Ordering data 121


Mounting an S8/S81 ceiling mount on an existing intermediate piece 121
Mounting an S81 ceiling mount on an existing ceiling track mount 122

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


98 Installation on an existing substructure

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 99

Installation options

If certain requirements are met, the S8 and S81 ceiling mounts can be in-
stalled on existing substructures provided for the S23, EMD2, S5, S51, S6
or S61 ceiling mount.
A special mounting kit is available for installing the S8 and S81 ceiling
mounts on existing substructures (pre-installation sets of earlier ceiling
mounts). A major part of the special mounting kit is the adapter plate. In
this application, the adapter plate replaces the mount flange included in
the standard delivery package of the ceiling mount.
The disciplines concerned are ophthalmic surgery, ENT surgery and P&R
surgery.
Note:
Use this special mounting kit only if the pre-installation set for new instal-
lations (Cat. No. 1078-181) cannot be installed in the OR.
Only the pre-installation set for new installations ensures the typical
working distances (microscope axis - pivot axis 1) of 1,250 mm (49.2") for
the S8 ceiling mount or 1,000 mm (39.4") for the S81 ceiling mount, see
page 21 and page 23.

Working ranges

The EMD2 and S23 ceiling mounts have been designed for an optimum
working distance (center of mount flange to microscope axis) of 650 mm
(25.6"). With a stationary operating table, working ranges of 900 mm
(35.4") can be achieved if the adapter plate is rotated through 180°. If, in
addition, the adapter plate is mounted with an offset of 90 mm (3.5"), the
working range can be increased to 990 mm (39").
Note:
The recommended typical working ranges (= distance of pivot axis 1 to
microscope axis) for the S8 S81 ceiling mounts are:
e 1,250 mm ± 500 mm (49.2 ± 19.7") for the S8 ceiling mount.
e 1,000 mm ± 400 mm (39.4" ± 15.7") for the S81 ceiling mount.
This means that an S8 ceiling mount will be at its rear mechanical limit po-
sition in this type of installation, see illustration on page 102.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


100 Installation on an existing substructure

Mounting options

Using the adapter plate, the S8 or S81 ceiling mount can be mounted
on:
e existing intermediate piece (305658-0000.000) of an S23 or EMD2
ceiling mount, with the adapter plate in a laterally offset position and
the electrical components mounted on the mounting bracket.
e existing intermediate piece (305658-0000.000) of an S23 or EMD2
ceiling mount, with the adapter plate in a centered position and option-
ally:
e electrical components mounted on the mounting bracket or
e electrical components mounted on the adapter plate.
e existing ZDK intermediate piece (305658-0000.000) of an S5, S51, S6
or S61 ceiling mount, and optionally:
e electrical components mounted on the mounting bracket or
e electrical components mounted on the adapter plate.
e existing ceiling track mount (305620-0000.000) of the S23 ceiling
mount, with electrical components mounted on the adapter plate.
The different mounting options are described in more detail in the fol-
lowing.

Adapter plate (1071-924)

1 Hole circle
The hole circle with dia. 440 mm (17.3") has been provided for the in-
stallation on an existing ZDK intermediate piece (305942-9103.000) of
the S5, S51, S6 or S61 ceiling mount.
2 Rows of bores
The rows of bores have been provided for the installation on an ex-
isting intermediate piece (305658-0000.000) of the S23 or EMD2
ceiling mount and for the installation on an existing ceiling track mount
of the S23 ceiling mount.
The adapter plate can also be mounted with a 90 mm (3.5") lateral
offset on the intermediate piece of the S23 or EMD2 ceiling mount.
t = 20 mm (0.8"), thickness of the adapter plate

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 101

t = 20 mm (0.8T)

mm Inch mm Inch
Ø14 Ø0.55 250 9.8
Ø15 Ø0.6 500 19.7
Ø270 Ø10.6 90 3.5
Ø160 Ø6.3 123 4.8
Ø440 Ø17.3 150 5.9
120 4.7 275 10.8
140 5.5 550 21.7

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


102 Installation on an existing substructure

Installation on an intermediate piece of S23 or EMD2

Mounting the adapter plate in an offset position

Adapter plate mounted with a 90 mm (3.5") lateral offset


The following two illustrations show an adapter plate mounted with a 90
mm (3.5") lateral offset. The working range for the surgical microscope is
located between the circles shown in dashed lines. Therefore, you should
check for the installation of an S8 or S81 ceiling mount whether the op-
timum working range can be obtained by repositioning the operating
table: 1,250 mm (49.2") with S8, and 1,000 mm (39.4") with S81.

Working range of the S8 ceiling mount with a 90 mm (3.5") lateral off-


set of the adapter plate
1 Pivot axis
The radii of the working range
R = 750 mm (29.5") and R = 1,750 mm (68.9") refer to this pivot axis.

990 mm / 39T

750
mm S8
m R 1
0 m 5T 8.9T
7 5 . R6
1 R 29
R

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 103

Working range of the S81 ceiling mount with a 90 mm (3.5") lateral


offset of the adapter plate

1 Pivot axis
The radii of the working range
R = 600 mm (23.6") and R = 1,400 mm (55.1") refer to this pivot axis.

990 mm / 39T
S81

mm
400
.6 m

1
23 m

R
5.1T
T
R 00

1 R5
6
R

M 1:20

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


104 Installation on an existing substructure

Adapter plate mounted with a lateral offset and


electrical components mounted on the mounting bracket
Note:
The mounting bracket can only be used to mount the electrical compo-
nents on intermediate piece (1) if intermediate piece (1) has a clearance
of at least 310 mm (12.2").

1 Intermediate piece of S23 or EMD2 ceiling mount


(305658-0000.000)
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on mounting bracket.
5 Mounting bracket
The mounting bracket can also be mounted in a position rotated
through 180°.

Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 50 mm (1.97"), if the electrical components are attached to
the side of the mounting bracket.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 105

4 5

min. 310 mm / 12.2"

2
min. 50 mm
min. 2.0"

100 mm
3.9"
45 mm
1.8"

100 (50) mm
3.9 (2.0)"

240 (200) mm
3 9.4 (7.9)"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


106 Installation on an existing substructure

Mounting the adapter plate in a centered position

Electrical components mounted on the mounting bracket


Note:
The mounting bracket can only be used to mount the electrical compo-
nents on intermediate piece (1) if intermediate piece (1) has a clearance
of at least 310 mm (12.2").
If necessary, the electrical components can also be mounted on adapter
plate (3).

1 Intermediate piece of S23 or EMD2 ceiling mount


(305658-0000.000)
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on mounting bracket
5 Mounting bracket
The mounting bracket can also be mounted in a position rotated
through 180°.
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 50 mm (1.97"), if the electrical components are attached to
the side of the mounting bracket.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 107

4 5

min. 310 mm / 12.2"

2
min. 50 mm
min. 2.0"

100 mm
3.9"
45 mm
1.8"

150 mm
5.9"
3

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


108 Installation on an existing substructure

Electrical components mounted on the adapter plate


Note:
The electrical components can only be mounted on adapter plate (3) if the
clearance between intermediate piece (1) and adapter plate (3) is at least
108 mm (4.3").
If necessary, the electrical components can also be mounted at the side
of intermediate piece (1) using the mounting bracket.

1 Intermediate piece of S23 or EMD2 ceiling mount


(305658-0000.000)
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on adapter plate.

Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 108 mm (4.3"), if the electrical components are mounted
on the adapter plate.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 109

2
min. 108 mm
min. 4.3"

100 mm
3.9"
45 mm
1.8"

150 mm
5.9"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


110 Installation on an existing substructure

Installation on an existing ZDK intermediate piece

The ZDK intermediate piece (305942-9103.000) was used to install the


S5, S51, S6 and S61 ceiling mounts.

Mounting electrical components on the mounting bracket

Note:
The mounting bracket can only be used to mount the electrical compo-
nents on intermediate piece (1) if intermediate piece (1) has a clearance
of at least 310 mm (12.2").
Otherwise, the electrical components can also be mounted on adapter
plate (3).
.
1 ZDK intermediate piece (305942-9103.000)
of S5, S51, S6 and S61 ceiling mounts
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on mounting bracket.

Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 50 mm (1.97"), if the electrical components are attached to
the side of the mounting bracket.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 111

4
min. 310 mm / 12.2"

2
min. 50 mm
min. 2.0"

100 mm
3.9"
45 mm
1.8"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


112 Installation on an existing substructure

Electrical components mounted on the adapter plate


Note:
The electrical components can only be mounted on adapter plate (3) if the
clearance between intermediate piece (1) and adapter plate (3) is at least
108 mm (4.3").
If necessary, the electrical components can also be mounted on the side
of intermediate piece (1) using the mounting bracket (1124-211).

1 ZDK intermediate piece (305942-9103.000)


of S5, S51, S6 and S61 ceiling mounts
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on adapter plate.

Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 108 mm (4.3"), if the electrical components are mounted
on the adapter plate.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 113

min. 108 mm
min. 4.3"
45 mm / 1.8"

100 mm
3.9"

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


114 Installation on an existing substructure

Installation on ceiling track mount

You can install the S81 ceiling mount on an existing ceiling track mount.
The S8 ceiling mount cannot be installed on a ceiling track mount.
The following is required for installing an S81 ceiling mount on an existing
ceiling track mount:
e Kit for mounting S8/S81 CM on existing ceiling anchor plate (1120-
925)
e and, in addition, the "Kit for mounting S81 on ceiling track mount"
(1144-462).
Note:
Use the order sheet on page 123 to order the S81 ceiling mount for instal-
lation on an existing ceiling track mount.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 115

300 mm (11.8") ceiling panel extension

Ceiling panel extension (1) is required to completely cover any interme-


diate pieces of the S23, EMD2, S5, S51, S6 or S61 ceiling mount protru-
ding from the false ceiling.
To allow this, a 300 mm (11.8") ceiling panel extension and the relevant
mounting instructions (1177-632) are included in each "Kit for mounting
S8/S81 CM on existing ceiling anchor plate" (1120-925).
An installer will cut the 300 mm (11.8") ceiling panel extension to the
length required on site and install it as per the mounting instructions
(1177-632).

1 300 mm (11.8") ceiling panel extension (1177-632)


2 Ceiling panel for S8 / S81 ceiling mounts (1083-804)
3 Adapter plate (1071-924)
4 False ceiling
5 Intermediate piece of S23 or EMD2 ceiling mount
The intermediate piece of the S23 or EMD2 is shown here as an ex-
ample. Ceiling panel extension (1) can also be used with the ZDK in-
termediate piece of the S5, S51, S6 or S61 ceiling mount.
6 Structural ceiling

1 2 3

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


116 Installation on an existing substructure

On-site preparations

Customer's responsibilities

Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.

Caution:
g Obtain the structural verification for an existing substructure from a
structural engineer, i.e. make sure that the engineer confirms the ef-
fective strength of the existing anchors.
Structural verification must be performed prior to the installation of the
mount.
Please add a copy of the "Confirmation of structural calculation" to
your order (see page 119).
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
119).
g If an existing substructure is to be used, make sure that the maximum
inclination of the ceiling anchor plate under load does not exceed 0.2°.
g The structural engineer must check that no modifications have been
made to the substructure since the original installation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the ceiling flange or pre-in-
stallation set.
g Please note the information provided on page 104 for ordering an S81
ceiling mount for installation on an existing substructure (existing in-
termediate piece).

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 117

Warning!
g If an existing flange plate or intermediate piece must be exchanged,
never re-use the old anchors. New anchor holes must be drilled.
When calculating the effective strength of the new anchors, the struc-
tural engineer must take into account the weakening effect of the old
holes in the ceiling.

Planning the installation


g Inspect the mounting components supplied for completeness and
damage.
g Prior to the installation of a ceiling mount, you must always check that
the actual installation conditions - in particular the room height - corre-
spond to the specifications in the drawing.
g Two installers are required for mounting the mounting kit required for
installation on existing ceiling anchors.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


118 Installation on an existing substructure

Constructional requirements

The actual load on the ceiling depends on a large number of different fac-
tors. The requirements to be met by the ceiling or substructure result from
the addition of perpendicular forces and torques produced by the suspen-
sion system and accessories. These are the forces to be transmitted into
the structural ceiling via the ceiling anchors.
The column must be aligned in a vertical position (max. deviation ± 0.5°).

Forces and torques

Warning!
Your structural engineer must ensure in each individual case that the ex-
isting substructure and structural ceiling have a sufficient load capacity for
the forces and torques listed below. He must also take into account any
additional loads on the ceiling and add an appropriate safety margin, and
must observe the applicable national codes and regulations.
Caution:
e The perpendicular forces and torques specified below include an ad-
ditional load of 100 kg (220 lb), which is generated when a person
hangs on the end of the suspension arm (or the microscope). Further
safety margins have not been incorporated.
e The perpendicular forces and torques have been calculated on the ba-
sis of the maximum permissible load on the suspension arm.

S8 or S81 ceiling mount


For an S8 or S81 ceiling mount alone (without accessories on the acces-
sory interface), the existing substructure and the structural ceiling must
have the following load capacity:

e Perpendicular force: minimum 2500 N (565 lbf)


e Torque: minimum 3000 Nm (2215 lbf.ft)

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 119

Confirmation of the structural calculation and execution


of installation

Sales order no.: ...........................................................................................

Customer address / ...........................................................................................


Delivery address: ...........................................................................................
...........................................................................................

By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:

Name and address of ...........................................................................................


the structural engineer: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

The installer for


the proper mounting of the pre-installation set or the
ceiling or wall flange from Carl Zeiss:
Name and address of ...........................................................................................
the executing company ...........................................................................................
and installerrs name: ...........................................................................................
...........................................................................................

.............................. ..................................................................

Date Signature

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


120 Installation on an existing substructure

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 121

Ordering data

Mounting an S8/S81 ceiling mount on an existing intermediate piece

Description Cat.No.

Kit for mounting an S8/S81 CM on existing ceiling an- 1120-925


chor plate
The mounting kit (1120-925) includes:
e adapter plate (1071-924)
e mounting bracket (1124-211).
e 300 mm (11.8") ceiling panel extension (1177-632)

Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


122 Installation on an existing substructure

Mounting an S81 ceiling mount on an existing ceiling track mount

Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.

Description Cat.No.

Kit for mounting an S8/S81 CM on existing ceiling an- 1120-925


chor plate
Kit for installing an S81 on a ceiling track mount 1144-462

Controlling the ceiling track mount using a hand switch


The "Kit for installing an S81 on a ceiling track mount (1144-462)" in-
cludes a hand switch.
g If you wish to operate the ceiling track mount using this hand switch,
please also order the instrument socket (1141-820) for the connection
of this hand switch.

B Distance from the bottom surface of the ceiling track mount


(bottom surface of cover) from the floor
C Distance from floor
Depending on the discipline, we recommend using different
distances from the floor:
e ENT 1,650 mm (65")
e ophthalmology 1,800 mm (70.9")
e reconstructive and plastic surgery 1,800 mm (70.9")
e others 1,650 mm (65")
D Column length
The dimension is calculated using the formula:
D = B - C - 910 mm (35.8").

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Installation on an existing substructure 123

Order sheet for S81 ceiling mount on ceiling track mount

Column length:

D = B - C - 910 mm

D D= mm
min 40 mm, max 800 mm
mm

mm
C
B

Dimensions (inches):
B= "
C= "
Column length:
D = B - C - 35.8t
D= "
min 1.6t, max 31.5t

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


124 Installation on an existing substructure

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Index 125

Index

Numerics
300 mm ceiling panel extension (1177-632) 115
A
Accessories from other manufacturers, coupling with Zeiss system 62
Accessories on OPMI, cable installation 34
Accessory components, power supply 71
Accessory interface 62
Adapter plate 99, 100
Adapter plate (1071-924) 115, 121, 122
Adapter plate mounted in centered position 106
Adapter plate, offset 102
B
Blind cover 64
Building vibrations 26
C
Ceiling anchor plate 10
Ceiling anchor plate, installation positions 12
Ceiling anchor plate, mounting of 42
Ceiling panel 10, 46, 64, 115
Ceiling panel extension 300 mm (11.8") (1177-632) 46
Ceiling track mount (305620-0000.000) 114
Column length 54, 58, 122
Column, vertical alignment 29, 118
Conduits for future upgrades 34
Conduits, installation 34
Confirmation of execution of installation 47, 73, 119
Confirmation of structural calculation 47, 73, 119
Constructional requirements 29, 118
Coupling of Zeiss system with accessories from other manufacturers 62
Cover frame 64
Customer's responsibilities 24, 69, 116
D
Dimensions, S8 ceiling mount 7
Dimensions, S81 ceiling mount 8
Discipline-specific installation positions 76
Distance from floor 14, 54, 58, 122
E
Electrical components on adapter plate 108, 112
Electrical components on mounting bracket 106, 110

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


126 Index

Electrical connections for accessory modules 71


Electrical installation, on site 38
EMD2 and S2 ceiling mounts 99
ENT, alternative installation position 86, 87, 88, 89
Execution of installation, confirmation of 47, 73, 119
F
Flange area for accessory carrier 66
Flange area for ceiling mount 66
Forces 10
Forces and torques 29, 70, 118
G
Gynecology, recommended installation positions 93
H
High-current connectors 38, 72
Hole circle 100
Hole circle with 8 bores of 17 mm dia. 66
I
Installation above operating table 12
Installation in cramped conditions 16, 18
Installation laterally to operating table 12
Installation position 63
Installation position in the OR 12
Installation positions for Ophthalmology and ENT 77
Installation positions of ceiling anchor plate 12
Installation positions, discipline-specific 76
Installation site, layout 47, 73, 119
Installation, proper 47, 73, 119
Interface plate 10, 46
Interface plate, mounting 45
Intermediate piece (305658-0000.000) 104, 106, 108
Intermediate piece ZDK (305658-0000.000) 110
Intermediate piece ZDK (305942-9103.000) 112
Intermediate piece, existing 100
K
Key to symbols 2
L
Layout plan 76
Load on the ceiling, actual 29, 118
M
Mechanical installation 42
Mount flange 10, 64, 66
Mount flange and ceiling panel 64
Mount flange with flange tube 67

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Index 127

Mounting bracket (1124-211) 121, 122


Mounting kit (1120-925) 121, 122
Mounting kit, special 99
Mounting options 100
Mounting the spacers 44
N
Neurology, recommended installation positions 93
O
On-site conditions 76
On-site electrical installation 38
On-site preparations 24, 116
Operating position 16
Operating range for surgical microscope 20, 22
Operating range of ceiling mount 20
Ophthalmology and ENT, preferred installation position 78, 79, 80, 81
Ophthalmology, alternative installation position 82, 83, 84, 85
Optimum installation position 12
OR illuminators, 24 V DC with backup power supply 34
OR illuminators, connectors for 38, 72
Ordering data 48, 51, 121
P
P&R, Hand, installation position 90
P&R, Hand, recommended installation position 91, 92
Perpendicular force 29
Pivot axis 1 20, 22
Pivot axis 2 20, 22
Planning the electrical installation 33
Planning, careful 76
Potential equalization 34
Potential equalization, connectors for 38, 72
Power connection, checking 40
Power line, installation 34
Pre-installation set 10
Pre-installation set, completely covered 46
Pre-installation set, mounting on ceilings with air-supply ceiling 30
Pre-installation sets of earlier ceiling mounts 99
R
Review and release 47, 73, 119
Rows of bores 100
S
S8 ceiling mount 20
S8 ceiling mount, dimensions 7
S81 ceiling mount 22
S81 ceiling mount, dimensions 8

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


128 Index

Scale, same as layout plan 76


Signature, confirm by 47, 73, 119
Spacers 10
Structural calculation 47, 73, 119
Structural calculation, confirmation of 47, 73, 119
Structural ceiling 10
Structural engineer 47, 73, 119
Stud bolt 10
Substructure, produced on-site 47, 73, 119
Substructures, existing 99
T
Technical data 7
Torque 29
Torques 10
Transparent film 76
Turning the operating table 90
TV, cable installation 34
U
Upper part of carrier arm 64
Urology, recommended installation positions 93
V
Vibration, short-term 26
Vibration, sustained 26
W
Wall socket for foot control panel 34
Wall socket, mounting 36
Working ranges 99

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


129

M-30-1382-en S8 / S81 Ceiling Mounts Issue 12.0


Printed on 19. 11. 2007

Downloaded from www.Manualslib.com manuals search engine


Carl Zeiss Surgical GmbH
A Carl Zeiss Meditec Company Fax: ++49 (0) 7364-20 48 23
Email: surgical@meditec.zeiss.com
73446 Oberkochen Internet: www.meditec.zeiss.com
Germany

M-30-1382-en Printed in Germany CG-ST