Planning Manual
M-30-1382-en
Issue 12.0
Printed on 19. 11. 2007
Key to symbols
Warning!
The warning triangle indicates potential sources of danger which may
constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects,
collision or damage of the instrument.
Note:
The hand indicates hints on the use of the instrument or other tips for the
user.
Technical data 7
On-site preparations 24
Constructional requirements 29
Planning the electrical installation 33
Planning the mechanical installation 42
Confirmation of the structural calculation and execution
of installation 47
Ordering data 48
Accessory interface 61
Accessory interface 62
Installation position 63
Mount flange and ceiling panel 64
Mount flange (flange plate) 66
Configuration example: mount flange with flange tube 67
Pre-installation set 68
Customer's responsibilities 69
Electrical connections for accessory modules 71
Confirmation of the structural calculation and execution
of installation 73
Installation options 99
Installation on an intermediate piece of S23 or EMD2 102
Installation on an existing ZDK intermediate piece 110
Installation on ceiling track mount 114
Index 125
Technical data 7
S8 ceiling mount, dimensions 7
S81 ceiling mount, dimensions 8
Pre-installation set 10
Installation position in the OR 12
Operating position 16
Operating range of the ceiling mount 20
On-site preparations 24
Customer's responsibilities 24
Building vibrations 26
Customer's preparatory responsibilities, overview 28
Constructional requirements 29
Forces and torques 29
Mounting a pre-installation set on ceilings with air-supply ceiling 30
Ordering data 48
S8 ceiling mount for ORs with false ceilings, ordering data 48
S8 ceiling mount for ORs without false ceilings, ordering data 51
S81 ceiling mount for ORs with false ceilings, ordering data 54
S81 ceiling mount for ORs without false ceilings, ordering data 57
Technical data
Key
1 False ceiling
620 mm / 24.4T S8
238 mm / 9.4T
1
228 mm / 9T
250 mm
9.8T
950 mm / 37.4T
950 mm / 37.4T
700 mm / 27.6T
1 : 20
Key
1 False ceiling
2 Floor
A Distance A is specifically defined for each project.
B Distance from floor
We recommend different distances from the floor, depending on the
application involved.
e ENT 1,650 mm (65")
e ophthalmology 1,800 mm (70.9")
e reconstructive and plastic surgery 1,800 mm (70.9")
e others 1,650 mm (65")
S81
238 mm / 9.4T
620 mm / 24.4T
228 mm / 9T
1
250 mm
9.8T
360 mm
14.2T
360 mm
14.2T
A max. 350° ± 90° 270°
2
1 : 20
Pre-installation set
Key
1 Stud bolt
2 Spacers
3 Ceiling anchor plate
4 Floor
5 Ceiling panel
6 Mount flange
7 False ceiling
8 Interface plate
9 Structural ceiling
Mount flange (6) and ceiling panel (5) are part of the ceiling mount and are
only shown here for information. They are not included in the pre-installa-
tion set.
A dummy cover is available for closing the opening in the false ceiling until
the ceiling mount is installed. Also see page 48 and page 52.
Caution:
When calculating the forces / torques applied to the structural ceiling, the
maximum load, i. e. the weight of the ceiling mount itself and the max-
imum weight of equipment attached to it, must be taken into account.
Note:
Dimension (A) must not exceed 800 mm (31.5"). Otherwise, a ceiling sub-
structure must be installed to bridge the excess distance. When designing
the substructure, make sure it is strong enough to withstand the bending
vibrations caused by the arms of the ceiling mount.
1 9 8
2
3
45 mm / 1.8T
75mm / 3T
A
100 mm / 3.9T
150
75
7
620 mm / 24.4T
6
5
(1393 lbf)
(7007 lbf.ft)
S8 S81
1 2
1250 ± 100 mm
1500 ± 200 mm 49.2 ± 3.9T
59.1 ± 7.9T
3 4
5 5
1 : 50
2800 - 3300 mm
110.2 - 129.9T
ca. 1800 mm
ca. 70.9T
S8
2650 - 3500 mm
104.3 - 137.8T
S81
1 : 50
Operating position
The ceiling mounts are shown in a typical operating position. The angle
between the microscope arm and the carrier arm is 90°. At this angle, min-
imum effort is required for horizontal movement of the surgical micro-
scope.
Side view
1 S8 normal installation
2 S81 normal installation
If, due to other installations, the space available on the ceiling is not suf-
ficient for normal installation, you can use installation method 3 or 4 as an
alternative.
3 S8 installation in cramped conditions
The mount flange is installed in a position rotated through 180°. Also
see page 18. OR illuminators cannot be mounted in this case. The
same installation principle can also be used for the S81 ceiling mount
(not shown here).
4 S8 installation in cramped conditions
The brake arm is installed in a position rotated through 180°. Also see
page 18. OR illuminators cannot be mounted in this case. The same
installation principle can also be used for the S81 ceiling mount (not
shown here).
S8 S81
1 2
S8
1100 mm / 43.3T (S8) 1400 mm / 55.1T (S8)
850 mm / 33.5T (S81) 1150 mm / 45.3T (S81)
250 mm / 9.8T
3 4
50 mm / 2T
1 : 50
Top view
1 S8 normal installation
2 S81 normal installation
If, due to other installations, the space available on the ceiling is not suf-
ficient for normal installation, you can use installation method 3 or 4 as an
alternative.
3 S8 installation in cramped conditions
The mount flange is installed in a position rotated through 180°. Also
see page 16. OR illuminators cannot be mounted in this case. The
same installation principle can also be used for the S81 ceiling mount
(not shown here).
4 S8 installation in cramped conditions
The brake arm is installed in a position rotated through 180°. Also see
page 16. OR illuminators cannot be mounted in this case. The same
installation principle can also be used for the S81 ceiling mount (not
shown here).
S8 S81
1 2
1600 mm / 63T 1350 mm / 53.1T
S8
1100 mm / 43.3T (S8) 1400 mm / 55.1T (S8)
850 mm / 33.5T (S81) 1150 mm / 45.3T (S81)
3 4
1 : 50
S8 ceiling mount
The operating range specified here applies to normal installation.
For installation in cramped conditions, the dimensions shown on page 16
and page 18 can be used to determine the operating range of the ceiling
mount.
1 Pivot axis 1
2 Pivot axis 2
3 Operating range for the surgical microscope
S8
1
2
1 : 20
1 Pivot axis 1
2 Pivot axis 2
3 Operating range for the surgical microscope
S81
1000 ± 400 mm
39.4 ± 15.7T
250 mm / 9.8T
1 : 20
On-site preparations
Customer's responsibilities
Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.
The actual load on the ceiling depends on a large number of different fac-
tors, which must be determined in detail by a structural engineer on a
case-to-case basis: see "Constructional requirements for ceiling mounts"
on page 29.
Caution:
g Make sure that a structural engineer checks the installation conditions
during the planning procedure.
Structural verification must be performed prior to the installation of the
mount.
We recommend filing the structural verification in the ceiling mount
documentation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g Please add a copy of the "Confirmation of structural calculation" to
your order (see page 47).
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the pre-installation set.
g The system should be installed on a hard, level surface, e.g. concrete.
Do not countersink the pre-installation set, but mount it directly on the
respective surface.
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
47).
Warning!
g If an existing ceiling anchor plate or intermediate piece must be ex-
changed, never re-use the old anchors. New anchor holes must be
drilled.
When calculating the effective strength of the new anchors, the struc-
tural engineer must take into account the weakening effect of the old
holes in the ceiling.
Building vibrations
Due to the large number of parameters involved and the variety of poten-
tial building/ceiling mount constellations, Carl Zeiss Surgical is unable to
give an absolute guarantee for the vibration-free suspension of the ceiling
mount, even if the building meets the applicable ISO standards.
However, it is highly improbable that constant vibrations are transferred
to the ceiling mount if all requirements regarding vibrations in the OR
ceiling are met.
e Max. vibration velocity (RMS) at the installation points for the ceiling
mount.
e Vmax < 0.1 mm / s or Vmax < 4 milli-inches / s or below the curves (di-
agram) for the specified frequency range.
Sources: Carl Zeiss Surgical (in-house study), ISO 10811, recommenda-
tions for ORs.
1
RMS of vibration velocity (mm / sec)
0,1
0,01
4 5 6,3 8 10 12,5 16 20 25 31,5 40 50 63 90
One-third octaveband center frequencles (Hz)
100
RMS of vibration velocity (mili-inch / sec)
10
1
4 5 6,3 8 10 12,5 16 20 25 31,5 40 50 63 90
One-third octaveband center frequencles (Hz)
Constructional requirements
The actual load on the ceiling depends on a large number of different fac-
tors. The requirements to be met by the ceiling or substructure result from
the addition of perpendicular forces and torques produced by the suspen-
sion system and accessories. These are the forces to be transmitted into
the structural ceiling via the ceiling anchors.
The column must be aligned in a vertical position (max. deviation ± 0.5°).
Warning!
The structural engineer must ensure in each individual case that the struc-
tural ceiling has a sufficient load capacity for the forces and torques listed
below. He must also take into account any additional loads on the ceiling
and add an appropriate safety margin, and must observe the applicable
national codes and regulations.
Caution:
e The perpendicular forces and torques specified below include an ad-
ditional load of 100 kg (220 lb), which is generated when a person
hangs on the end of the suspension arm (or the microscope). Further
safety margins have not been incorporated.
e The perpendicular forces and torques have been calculated on the ba-
sis of the maximum permissible load on the suspension arm.
Caution:
If the carrier arm systems are installed in air-supply ceilings, the method
of installation must be specifically adapted to the constructional features
of the type of air-supply ceiling involved.
A false ceiling set is mounted to bridge the distance between the struc-
tural ceiling and the non-structural false ceiling. The length of the stud
bolts and spacers used is adapted on site as required.
The necessary planning should be performed in consultation with the
manufacturer of the air-supply ceiling and Carl Zeiss.
When ordering, please specify distance (A), i.e. the distance between the
lower surface of the structural ceiling and the lower surface of the false
ceiling.
Note:
e Distance (A) is the height of the false ceiling and must not exceed
800 mm. Otherwise, a ceiling substructure must be installed to bridge
the excess distance. Distance (Z) is the height of the air-supply ceiling.
e It is essential that the false ceiling set (2,3) and the conduits for the
power lines be installed before the air-supply ceiling is attached. Sub-
sequent installation is not possible, for reasons of space.
e A manhole (5) which is free from compressed air must be available for
the installation work. The internal diameter (S) of the manhole must be
larger than that of the opening in the ceiling (M).
e The opening of the manhole to the OR can be closed with a separate
cover (7) by the manufacturer of the air-supply ceiling.
1 Structural ceiling
2 Ceiling anchor plate
3 Spacers
4 Upper surface of air-supply ceiling
5 Manhole
6 Interface plate
7 Cover of carrier arm system to OR (customer's responsibility)
8 Lower surface of air-supply ceiling
9 Ceiling panel
1
2
A 4
S
5
Z
6
7
8
9
2
3
4
6
5
7
8
The power lines on the site of installation must be run up to the interface
plate. In addition, the potential equalization lines must be run to the termi-
nals on the interface plate.
Warning!
e The planning, execution and inspection of the on-site electrical instal-
lations must be performed by expert electrical technicians and official-
ly registered electrical firms.
e The electrical installations of the room concerned must comply with
the requirements of the applicable national regulations. In the Federal
Republic of Germany, VDE 0107 applies.
e The number of circuits is dependent on the customer's specific carrier
system configuration.
e If several circuits are required by the customer, appropriate allowance
must be made on the interface plate.
e Multiwire lines must be provided with wire end sleeves.
The wiring can be run either above the false ceiling or in conduits installed
in the structural ceiling. There are two options for the installation of con-
duits in the structural ceiling.
e The exit of the conduits is positioned beside ceiling anchor plate (8).
e The exit of the conduits is positioned at the center of ceiling anchor
plate (8).
The conduits have an internal diameter of 22 to 50 mm (0.9 - 2").
A conduit with a minimum internal diameter of 35 mm (1.4") is required for
the MediLive video line.
Key
1 Power line
3 x 1.5 mm2
2 Wall socket for foot control panel
As required.
3 Potential equalization
As required.
4 TV internal diameter 50 mm (2")
As required.
5 OR illuminators
24 V DC with backup power supply, customer's responsibility (as re-
quired)
6 Other OPMI accessories
e. g. BIOM/SDI, additional illumination, AUX output port, etc. (as re-
quired)
7 Further conduits
We recommend installing further conduits for future upgrades.
Connection of SpeedFokus AF
Note:
g Determine the exact length between the Speedfokus AF connection
and the CCU+ monitor.
If the cable length exceeds 10 m (394"), a signal amplifier is required for
the SpeedFokus AF. The signal amplifier requires 110 - 230 V line power.
1-7 8 1-7 9
Note:
e We recommend mounting wall socket (3) on the wall located behind
the surgeon. Height approx. 1.4 m (55.1").
e Please also note the chapter "Installing the conduits" on page 34.
Key
1 Wall socket for surface mounting
2 Wall socket for flush mounting
3 Wall socket
4 Central OR control panel (part of customer's installations)
5 Foot control panel
6 Remote control socket for an external signal of a maximum of 24 V /
0.5 A.
6 6
1 2
3 4
1400 mm / 55.1T
The power lines on the site of installation must be run up to the interface
plate. In addition, the potential equalization lines must be run to the termi-
nals on the interface plate.
Warning!
e The planning, execution and inspection of the on-site electrical instal-
lations must be performed by expert electrical technicians and official-
ly registered electrical firms.
e The electrical installations of the room concerned must comply with
the requirements of the applicable national regulations. In the Federal
Republic of Germany, VDE 0107 applies.
e The number of circuits is dependent on the customer's specific carrier
system configuration.
e If several circuits are required by the customer, appropriate allowance
must be made on the interface plate.
e Multiwire lines must be provided with wire end sleeves.
Key
A Connectors for accessory modules (OR illuminators, monitor bracket)
For details of terminals B and C, see the section "Accessory interface",
page 71.
D Connectors for potential equalization
Maximum terminal cross sections
Terminals 1 - 4 = 16 mm2 (AWG 6)
Terminals 5 - 8 = 6 mm2 (AWG 10)
E On-site connector
F Ceiling mount connectors
G High-current connectors
Maximum terminal cross section = 4 mm2 (AWG 12)
Connector (part no. 6 001 215) with the contacts (H): L1, N1, PE1;
L2, N2, PE2 and L3, N3, PE3.
H Terminal numbers
I Circuit numbers
L1 L1
B L2 L2
PE PE
L1 L1
C L2 L2
PE PE
A D G
E
6 PE PE 7
5 PE PE 8
F
4 PE PE 1
D 3 PE PE 2
H I
Caution:
The ceiling mount can be set at two points to the rated voltage used in the
country of destination:
e Transformer terminal on interface plate.
Settings: 100 - 115 - 230 V
e Power supply unit
Settings: 115 - 230 V
The correctness of the factory-adjusted setting is checked by our service
team during first-time installation.
e The power connection must conform to the regulations applicable in
the country of destination.
e A power switch (2-pin) has been provided.
e Line fuse
230 V: min. 10 A, slo-blo, class C
115 V: min. 20 A, slo-blo, class C
Auxiliaries required:
e Two sufficiently long ladders and a lifting device (lift table).
e Torque wrench: 0-120 Nm (minimum).
e Open-end wrenches and drive sockets for torque wrench:
24 mm and 55 mm (0.9" und 2.2").
e Anchors as specified by the structural engineer.
e Sheet specifying dimensions (supplied with the pre-installation set).
mm Inch
1
605 23.8
517 20.4
25 1.0
Ø400 Ø15.7
Ø560 Ø22.0
Ø26 Ø1.0
2
75 mm / 3"
3 4
1 2
3 5
75 mm
3"
3 4
Ceiling panel extension (1) is required if the distance between the struc-
tural ceiling and the false ceiling (i.e. dimension "A" in the order forms) is
clearly shorter than 200 mm (7.9"), i.e. < 190 mm (<7.5"). In this case,
ceiling panel (2) alone will not suffice to completely cover pre-installation
set (3).
Ceiling panel extension (1) cannot enclose interface plate (5). If interface
plate (5) also protrudes from the false ceiling, you must order the "Cover
for pre-installation set" (Cat. No. 1139-023).
You can order a 300 mm (11.8") ceiling panel extension including
mounting instructions (1177-632) as a spare part from the Zeiss Service
Department.
The installer will cut the 300 mm (11.8") ceiling panel extension to the
length required on site and install it as per the mounting instructions
(1177-632).
Key
1 Ceiling panel extension 300 mm (11.8") (1177-632)
2 Ceiling panel for S8 / S81 ceiling mounts (1083-804)
3 Pre-installation set (1078-181)
4 False ceiling
5 Interface plate
6 Structural ceiling
1 2 3
By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:
.............................. ..................................................................
Date Signature
.............................. ..................................................................
Date Signature
Ordering data
Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
Description Cat.No.
S8 ceiling mount 1176-968
Pre-installation set for new installation 1078-181
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804
mm
max 800 mm
min 200 mm
A
Dimensions in inches:
mm
A= "
max 3300 mm
min 2800 mm
min 7.9", max 31.5"
B
B= "
min 110.2", max 129.9"
You must order the ceiling mount using the following order sheet only if
the OR does not provide a false ceiling.
If the OR does not provide a false ceiling, you will need a cover for the pre-
installation set (1078-181). You can order the cover for the pre-installation
set as a kit (1139-023).
Description Cat.No.
S8 ceiling mount 1176-968
Pre-installation set for new installation 1078-181
Cover for pre-installation set 1139-023
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804
Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
H Room height
Distance from the floor to the structural ceiling.
g You must exactly measure this height.
I Required installation height of the S8 ceiling mount
The installation height of the S8 ceiling mount is the distance from the
floor to the bottom surface of the mount flange.
g You must preselect the installation height.
Note:
The installation height can be between 2,800 mm (110.2") and 3,300
mm (129.9"). Ideally, you should select an installation height of
3,100 mm (122") which is the optimum compromise between working
height and head room.
A Ordered length of the pre-installation set
can be determined by subtracting the installation height (I) selected
and 45 mm (1.8") from the room height (H) measured.
F Length to which cover (1139-023) must be cut
can be determined by subtracting the installation height (I) selected
and 52 mm (2") from the room height (H) measured.
1
F
2
A = H - I - 45 mm
mm
A= mm
min 200 mm, max 800 mm
Dimensions in inches:
H
H= "
mm
F= " 5
min 7.9", max 31.5"
S81 ceiling mount for ORs with false ceilings, ordering data
Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
Order sheet for S81 ceiling mount in ORs with false ceilings
mm
max 800 mm
min 200 mm
A
D = B - C - 810 mm
D= mm
D min. 40 mm, max. 800 mm*)
*) Recommendation: max. 600 mm,
see note on page 54.
mm
Dimensions in inches:
C A= "
mm
min. 7.9", max. 31.5"
B= "
C= " B
Column length:
D = B - C - 31.9t
D= "
min. 1.6t, max. 31.5t *)
*) Recommendation:
max. 23.6",
see note on page 54.
S81 ceiling mount for ORs without false ceilings, ordering data
You must order the ceiling mount using the following order sheet only if
the OR does not provide a false ceiling.
If the OR does not provide a false ceiling, you will need a cover for the pre-
installation set (1078-181). You can order the cover for the pre-installation
set as a kit (1139-023).
Description Cat.No.
S81 ceiling mount 1176-969
Pre-installation set for new installation 1078-181
Cover for pre-installation set 1139-023
Wall socket for S8/S81 CM for foot control panel 1078-524
S8/S81 instrument socket for foot control panel 1141-820
Ceiling panel for S 8 / S81 ceiling mounts 1083-804
Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
Caution:
Use the following principle for planning the installation of the S81 ceiling
mount: Keep the ordered length (A) of the pre-installation set as short as
possible and select a longer column length (D) as required.
C Required distance from floor
We recommend defining different distances from the floor, depending
on the application involved: ENT 1650 mm (65"), ophthalmology
1800 mm (70.9"), reconstructive and plastic surgery 1800 mm (70.9"),
others 1650 mm (65").
H Room height (distance from the floor to the structural ceiling)
g You must exactly measure this height.
A Ordered height of pre-installation set
Start your planning using the shortest length of 200 mm (7.9").
D Column length (ordered length)
Calculate the column length: D = H - A - C - 810 mm (-31.9").
The column length may be a minimum of 40 mm (1.6") and a maxi-
mum of 800 mm (31.5").
Note: column lengths of >600 mm (>23.6T)
Depending on the type of construction involved, vibrations may have
a negative influence if the column length (D) exceeds 600 mm
(>23.6t).
g Use column lengths of >600 mm (>23.6t) only in exceptional cases. If
on-site conditions permit it, column lengths of >600 mm (>23.6t) can
be avoided by selecting an appropriately greater height (A) of the pre-
installation set.
Example: Use the maximum column length (D) of 600 mm (23.6") and
calculate the length (A) of the pre-installation set to be ordered using
the formula: A = H - D - C - 810 mm (-31.9").
F Length to which cover (F) (1139-023) must be cut
Calculate length (F) to which the cover must be cut:
F = A - 7 mm (-0.3").
Order sheet for S81 ceiling mount in ORs without false ceilings
1
A F
2
3
4
D
mm
Dimensions in inches:
Pre-installation set, ordered length (A)
A= "
H A=
min 7.9", max 31.5" mm
C= " min 200 mm, max 800 mm
H= "
Column length, ordered length (D):
Column length (D):
mm
D = H - A - C - 810 mm
D = H - A - C - 31.9"
C
D= " D= mm
min. 1.6", max. 31.5" *) min. 40 mm, max. 800 mm*)
*) Recommendation: *) Recommendation: max. 600 mm,
max. 23.6", see see note on page 58.
note on page 58.
Length (F) to which the Length (F) to which the cover must
cover must be cut: be cut:
F = A - 0.3" F = A - 7 mm.
5 F= "
F= mm
min 7.9", max 31.5"
min 193 mm, max 800 mm
Accessory interface
Accessory interface 62
Coupling of the Zeiss system with accessories from
other manufacturers 62
Installation position 63
Pre-installation set 68
Customer's responsibilities 69
Accessory interface
Next to the flange area for the carrier arm of the ceiling mount, the mount
flange also features a flange area for an accessory carrier. This is the ac-
cessory interface designed as a hole circle, see page 66.
This accessory interface permits you to mount an accessory carrier e.g.
for OR illuminators, supports, monitor brackets, trays for video systems or
similar devices.
Caution:
The coupling of a system from Carl Zeiss Surgical GmbH with accesso-
ries (OR illuminators, monitor brackets or similar) from other manufac-
turers results in a new medical system for which the creator of the new
system must meet the necessary requirements (approval, qualification,
structural requirements, etc.).
Please note the user manuals provided by the manufacturer of the acces-
sory modules. Further information is available from our service depart-
ment or from authorized representatives.
g Check the feasibility of the installation in detail together with the man-
ufacturer of the relevant accessory module (OR illuminator or monitor
bracket).
Installation position
Caution:
When defining the positions and ranges of movement of the accessory
equipment, please pay attention to the geometry of the respective ceiling
mount and to the risk of collisions. For the dimensions, please see:
e illustration on page 7 for S8 ceiling mount.
e illustration on page 8 for S81 ceiling mount.
Note:
The mount flange is usually installed 45 mm (1.8") below the false ceiling.
For the dimensions of the pre-installation set, please see the illustration
on page 68.
The S8 or S81 ceiling mount is secured with mount flange (1) on the pre-
installation set already installed in the ceiling. In addition to the flange
area for the carrier arm of the ceiling mount, the mount flange also fea-
tures a flange area for an accessory carrier e.g. for attaching an OR illu-
minator and/or monitor bracket. The mount flange is part of the ceiling
mount and is included in its delivery package.
Ceiling panel (2) is used to cover the mount flange and the entire opening
in the false ceiling. The opening in the false ceiling must not be larger than
600 mm x 600 mm (23.6" x 23.6"). Two openings are provided in the
ceiling panel, one for the ceiling mount and one for an accessory carrier.
Cover frame (3) closes the remaining space between the square tube of
the ceiling mount and the ceiling panel.
Blind cover (4) is supplied together with the ceiling panel, permitting the
opening for the accessory carrier to be completely closed. You can make
an opening in blind cover (4) which fits the accessory carrier used.
A dummy cover is available for closing the opening in the false ceiling until
the ceiling mount is installed. Also see page 48 and page 52.
Note:
The S8 or S81 ceiling mount is delivered with the following components
already mounted: mount flange (1), cover frame (2) and carrier arm of the
ceiling mount (5), as in the configuration example on page 67.
Key
1 Mount flange, part of the ceiling mount.
2 Ceiling panel consisting of two pieces which are fitted together during
installation.
3 Cover frame, closes the remaining space between the carrier arm of
the ceiling mount (square tube) and the ceiling panel.
4 Blind cover, completely closes the opening for the flange tube in the
ceiling panel. You can make an opening in the blind cover as required
for the accessory carrier used.
5 Upper part of the carrier arm of the S8 and S81 ceiling mounts, de-
signed as a square tube.
2 2
1 2 3 mm Inch
Ø17 Ø0.7
Ø270 Ø10.6
Ø160 Ø6.3
Ø165,5 Ø6.52
123 4.8
150 5.9
1 Mount flange
2 Ceiling panel
Only one piece of the ceiling panel is shown.
3 Flange tube (example)
A flange tube of this type can be used for mounting accessories (illu-
minators, monitor mount, etc.).
4 Upper part of the carrier arm
The upper part of the carrier arm of the S8 and S81 ceiling mounts
consists of a square tube.
3 4
Pre-installation set
1 Stud bolt
2 Spacers
3 Ceiling anchor plate
4 Floor
5 Ceiling panel
6 Mount flange
7 False ceiling
8 Interface plate
9 Structural ceiling
10 Dimension A
is the distance between the structural ceiling and false ceiling.
Mount flange (6) and ceiling panel (5) are part of the ceiling mount and are
only shown here for information. They are not included in the pre-installa-
tion set.
1 9 8
2
3
45 mm / 1.8T
75mm / 3T
A
100 mm / 3.9T
150
75
7 10
620 mm / 24.4T
6
5
Customer's responsibilities
Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.
The actual load on the ceiling depends on a large number of different fac-
tors, which must be determined in detail by a structural engineer on a
case-to-case basis: see "Constructional requirements for ceiling mounts"
on page 29.
Caution:
g Your structural engineer must separately take into account the addi-
tional perpendicular force and additional torque produced by accesso-
ries mounted on the accessory interface, i.e. he must add them to the
perpendicular force and torque caused on the ceiling by the ceiling
mount alone.
g Make sure that a structural engineer checks the installation conditions
during the planning procedure.
Structural verification must be performed prior to the installation of the
mount.
We recommend filing the structural verification in the ceiling mount
documentation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g Please add a copy of the "Confirmation of structural calculation" to
your order (see page 73).
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the pre-installation set.
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
73).
The power supply for the electrical accessory components can be con-
nected to connectors "A" on the interface plate (connectors for OR illumi-
nators, terminal blocks B and C).
Key
A Connectors for OR illuminators
B Terminal 1 for OR illuminator 1
Maximum terminal cross section = 16 mm2 (AWG 6)
C Terminal 2 for OR illuminator 2
Maximum terminal cross section = 16 mm2 (AWG 6)
For details of the other terminals, see the section "On-site electrical instal-
lation", page 38.
L1 L1
B L2 L2
PE PE
L1 L1
C L2 L2
PE PE
By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:
.............................. ..................................................................
Date Signature
.............................. ..................................................................
Date Signature
The various surgical disciplines and on-site conditions require different in-
stallation positions of the S8 and S81 ceiling mounts.
This chapter describes a number of recommended installation positions
of the S8 and S81 ceiling mounts for different surgical disciplines.
Note:
In unfavorable on-site conditions, the recommended installation positions
must be changed appropriately. In cramped conditions, the use of OR il-
luminators is only possible with restrictions or indeed not at all.
Caution:
The recommended discipline-specific installation positions do not release
you from the responsibility of careful planning. Only careful planning will
ensure a smooth, collision-free workflow during surgery.
Scale 1:20
All recommended, discipline-specific installation positions are shown at a
scale of 1:20 and can usually be directly incorporated in the layout plan.
Transparent film
It may also be helpful for your planning work if you copy the scale 1:20
discipline-specific installation positions to transparent film. You can then
place the transparent film on the layout plan, and bring it into a suitable
position by shifting and turning it.
Note:
In the following illustrations, the representation of the ceiling mount in light
gray, dotted lines shows the alternative, mirrored (laterally reversed) in-
stallation option.
S8
.9 m
68 m
T
R 750
1
R
R 75
0
R 29 mm
.5T
M 1:20
S8
.9 m
68 m
T
R 7 50
1
R
R7
50
R 2 mm
9 .5 2
T
M 1:20
S81
m
m
0
0 .1T
14 5 5
R R
R 600
m
R 23.6 m
T
M 1:20
S81
.1 m
55 m
T
R 00
14
R
R 600 mm 2
R 23.6T
M 1:20
S8
1
0 mm
.5T
R 29
R 75
m
0m
R 1 . 9t
8
5
R6
7
.9T m
0m
75
68
R1
R
m
0m
.5t
R 29
R 75
M 1:20
S8
R
17
R 50
68 m
.9 m
T
1 2 2 1
m
5 0mt
17 . 9
R R 68
m
R
m
75
T
R
75
.5
29
29
R
m
.5
R
t
M 1:20
S81
R1
40
0m
R m
55
.1T
m
0m
R 60 3.6t
R2
R 14
00 m m
m 00 m
R 55 R6
. 1t 3 .6 t
R2
2 1 1 2
M 1:20
S81
1 2 2 1
R1
40
0m m
R5
5.1
m
4 00m
T R 1 55.1t
R
R6
00 mm
R 2 mm 00
3.6 R 6 23. 6T
t R
M 1:20
S8
R
75
1
0
R
m
29
m
.5
T
R1
R6
750
8 .9
mm
T
M 1:20
S8
2
R
7
R 50
29 m
.5 m
T
R
17 8. 9
R
50 T
6
mm
M 1:20
2
S81
R
60
R 0m
23
.6T m
R
14 5.1T
00
R
5
m
m
M 1:20
S81
1
R6
2 00
R 2 mm
3.6
T
R
14
00
R
55
m
m
.1
T
M 1:20
Note:
In P&R surgery, one or the other of the patient's hand may sometimes be
difficult to reach. In such cases, there is usually the possibility of turning
the operating table in such a way that the hand concerned is brought
within the optimum working range of the ceiling mount.
S8
2 1 1 2
0 mm
R 75 .5T R 75
0m
R 29 R 29 m
.5t
R1
m 7
5 0m R 50 m
7 t 68
R 1 6 8 .9 .9T m
R
M 1:20
S8
m
5 0mT
7 .9
R 1 68
R
R
75 9.5t
R
m
0m
2
m
0
m
.5
75
29
R
R
m
1 2 5 0mt 2 1
17 .9
R R 68
M 1:20
The S8 ceiling mount is used for the disciplines of neurology, spine, uro-
logy and gynecology.
The installation positions recommended in this chapter are also suitable
for ENT, with certain restrictions.
S8
T
mm
2
9.5
R2
7 50
R
mm
50 t
R 1 68.9
6 8 mm
.9T
R
7
R 50
17
R
29 mm
t
1
R 50
.5
7
R
M 1:20
S8
2
mm
7 50 T
R 29.5
R
mm
50 t
R 1 68.9
.9 m
68 m
T
R
R 750
7
1
R
m
m
0
75 9.5t
R 2
R
M 1:20
1
Installation options 99
Working ranges 99
Mounting options 100
Adapter plate (1071-924) 100
Installation options
If certain requirements are met, the S8 and S81 ceiling mounts can be in-
stalled on existing substructures provided for the S23, EMD2, S5, S51, S6
or S61 ceiling mount.
A special mounting kit is available for installing the S8 and S81 ceiling
mounts on existing substructures (pre-installation sets of earlier ceiling
mounts). A major part of the special mounting kit is the adapter plate. In
this application, the adapter plate replaces the mount flange included in
the standard delivery package of the ceiling mount.
The disciplines concerned are ophthalmic surgery, ENT surgery and P&R
surgery.
Note:
Use this special mounting kit only if the pre-installation set for new instal-
lations (Cat. No. 1078-181) cannot be installed in the OR.
Only the pre-installation set for new installations ensures the typical
working distances (microscope axis - pivot axis 1) of 1,250 mm (49.2") for
the S8 ceiling mount or 1,000 mm (39.4") for the S81 ceiling mount, see
page 21 and page 23.
Working ranges
The EMD2 and S23 ceiling mounts have been designed for an optimum
working distance (center of mount flange to microscope axis) of 650 mm
(25.6"). With a stationary operating table, working ranges of 900 mm
(35.4") can be achieved if the adapter plate is rotated through 180°. If, in
addition, the adapter plate is mounted with an offset of 90 mm (3.5"), the
working range can be increased to 990 mm (39").
Note:
The recommended typical working ranges (= distance of pivot axis 1 to
microscope axis) for the S8 S81 ceiling mounts are:
e 1,250 mm ± 500 mm (49.2 ± 19.7") for the S8 ceiling mount.
e 1,000 mm ± 400 mm (39.4" ± 15.7") for the S81 ceiling mount.
This means that an S8 ceiling mount will be at its rear mechanical limit po-
sition in this type of installation, see illustration on page 102.
Mounting options
Using the adapter plate, the S8 or S81 ceiling mount can be mounted
on:
e existing intermediate piece (305658-0000.000) of an S23 or EMD2
ceiling mount, with the adapter plate in a laterally offset position and
the electrical components mounted on the mounting bracket.
e existing intermediate piece (305658-0000.000) of an S23 or EMD2
ceiling mount, with the adapter plate in a centered position and option-
ally:
e electrical components mounted on the mounting bracket or
e electrical components mounted on the adapter plate.
e existing ZDK intermediate piece (305658-0000.000) of an S5, S51, S6
or S61 ceiling mount, and optionally:
e electrical components mounted on the mounting bracket or
e electrical components mounted on the adapter plate.
e existing ceiling track mount (305620-0000.000) of the S23 ceiling
mount, with electrical components mounted on the adapter plate.
The different mounting options are described in more detail in the fol-
lowing.
1 Hole circle
The hole circle with dia. 440 mm (17.3") has been provided for the in-
stallation on an existing ZDK intermediate piece (305942-9103.000) of
the S5, S51, S6 or S61 ceiling mount.
2 Rows of bores
The rows of bores have been provided for the installation on an ex-
isting intermediate piece (305658-0000.000) of the S23 or EMD2
ceiling mount and for the installation on an existing ceiling track mount
of the S23 ceiling mount.
The adapter plate can also be mounted with a 90 mm (3.5") lateral
offset on the intermediate piece of the S23 or EMD2 ceiling mount.
t = 20 mm (0.8"), thickness of the adapter plate
t = 20 mm (0.8T)
mm Inch mm Inch
Ø14 Ø0.55 250 9.8
Ø15 Ø0.6 500 19.7
Ø270 Ø10.6 90 3.5
Ø160 Ø6.3 123 4.8
Ø440 Ø17.3 150 5.9
120 4.7 275 10.8
140 5.5 550 21.7
990 mm / 39T
750
mm S8
m R 1
0 m 5T 8.9T
7 5 . R6
1 R 29
R
M 1:20
1 Pivot axis
The radii of the working range
R = 600 mm (23.6") and R = 1,400 mm (55.1") refer to this pivot axis.
990 mm / 39T
S81
mm
400
.6 m
1
23 m
R
5.1T
T
R 00
1 R5
6
R
M 1:20
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 50 mm (1.97"), if the electrical components are attached to
the side of the mounting bracket.
4 5
2
min. 50 mm
min. 2.0"
100 mm
3.9"
45 mm
1.8"
100 (50) mm
3.9 (2.0)"
240 (200) mm
3 9.4 (7.9)"
4 5
2
min. 50 mm
min. 2.0"
100 mm
3.9"
45 mm
1.8"
150 mm
5.9"
3
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 108 mm (4.3"), if the electrical components are mounted
on the adapter plate.
2
min. 108 mm
min. 4.3"
100 mm
3.9"
45 mm
1.8"
150 mm
5.9"
Note:
The mounting bracket can only be used to mount the electrical compo-
nents on intermediate piece (1) if intermediate piece (1) has a clearance
of at least 310 mm (12.2").
Otherwise, the electrical components can also be mounted on adapter
plate (3).
.
1 ZDK intermediate piece (305942-9103.000)
of S5, S51, S6 and S61 ceiling mounts
2 False ceiling
3 Adapter plate
4 Electrical components
mounted on mounting bracket.
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 50 mm (1.97"), if the electrical components are attached to
the side of the mounting bracket.
4
min. 310 mm / 12.2"
2
min. 50 mm
min. 2.0"
100 mm
3.9"
45 mm
1.8"
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
The probable distance of the adapter plate from the intermediate piece
will be at least 108 mm (4.3"), if the electrical components are mounted
on the adapter plate.
min. 108 mm
min. 4.3"
45 mm / 1.8"
100 mm
3.9"
You can install the S81 ceiling mount on an existing ceiling track mount.
The S8 ceiling mount cannot be installed on a ceiling track mount.
The following is required for installing an S81 ceiling mount on an existing
ceiling track mount:
e Kit for mounting S8/S81 CM on existing ceiling anchor plate (1120-
925)
e and, in addition, the "Kit for mounting S81 on ceiling track mount"
(1144-462).
Note:
Use the order sheet on page 123 to order the S81 ceiling mount for instal-
lation on an existing ceiling track mount.
1 2 3
On-site preparations
Customer's responsibilities
Note:
Zeiss service staff can install the ceiling mount only if all points of the fol-
lowing checklist applicable to the relevant installation conditions have
been fulfilled.
Caution:
g Obtain the structural verification for an existing substructure from a
structural engineer, i.e. make sure that the engineer confirms the ef-
fective strength of the existing anchors.
Structural verification must be performed prior to the installation of the
mount.
Please add a copy of the "Confirmation of structural calculation" to
your order (see page 119).
g On-site conditions also include building vibrations, which the structural
engineer responsible must take into account right during the planning
phase (see page 26).
Obtain a written confirmation from your structural engineer stating that
possible building vibrations have been taken into account (see page
119).
g If an existing substructure is to be used, make sure that the maximum
inclination of the ceiling anchor plate under load does not exceed 0.2°.
g The structural engineer must check that no modifications have been
made to the substructure since the original installation.
g Obtain a written confirmation from a structural engineer stating that
the applicable national codes and regulations have been complied
with.
g If any differences exist between the planning documents and the ac-
tual on-site situation, please inform your contact at Carl Zeiss or the
planning expert prior to the installation of the ceiling flange or pre-in-
stallation set.
g Please note the information provided on page 104 for ordering an S81
ceiling mount for installation on an existing substructure (existing in-
termediate piece).
Warning!
g If an existing flange plate or intermediate piece must be exchanged,
never re-use the old anchors. New anchor holes must be drilled.
When calculating the effective strength of the new anchors, the struc-
tural engineer must take into account the weakening effect of the old
holes in the ceiling.
Constructional requirements
The actual load on the ceiling depends on a large number of different fac-
tors. The requirements to be met by the ceiling or substructure result from
the addition of perpendicular forces and torques produced by the suspen-
sion system and accessories. These are the forces to be transmitted into
the structural ceiling via the ceiling anchors.
The column must be aligned in a vertical position (max. deviation ± 0.5°).
Warning!
Your structural engineer must ensure in each individual case that the ex-
isting substructure and structural ceiling have a sufficient load capacity for
the forces and torques listed below. He must also take into account any
additional loads on the ceiling and add an appropriate safety margin, and
must observe the applicable national codes and regulations.
Caution:
e The perpendicular forces and torques specified below include an ad-
ditional load of 100 kg (220 lb), which is generated when a person
hangs on the end of the suspension arm (or the microscope). Further
safety margins have not been incorporated.
e The perpendicular forces and torques have been calculated on the ba-
sis of the maximum permissible load on the suspension arm.
By signing below, the following persons confirm that they have per-
formed their work in a proper and orderly way:
The structural engineer for
e the selection and layout of the installation site, taking possible build-
ing vibrations into account
e the structural calculation, taking into account the applicable national
regulations and the planning manual
e the structural checking of an existing substructure
e the structural calculation of a substructure built on site
e the final checking and release of the structural calculations:
.............................. ..................................................................
Date Signature
.............................. ..................................................................
Date Signature
Ordering data
Description Cat.No.
Caution:
When ordering an S81 ceiling mount for installation on an existing inter-
mediate piece, you must determine and specify dimension B in the "Order
sheet for S81 ceiling mounts in ORs with false ceilingst (see page 55) as
follows:
g Measure the distance between the bottom surface of the intermediate
piece to the floor.
g Subtract the probable distance between the adapter plate and the in-
termediate piece and enter the value obtained as dimension B in the
order sheet.
Note:
Please note that the relevant dimensions have to be entered in the fol-
lowing drawing for each purchase order.
g Enter the data and enclose a copy of the drawing with your purchase
order.
Description Cat.No.
Column length:
D = B - C - 910 mm
D D= mm
min 40 mm, max 800 mm
mm
mm
C
B
Dimensions (inches):
B= "
C= "
Column length:
D = B - C - 35.8t
D= "
min 1.6t, max 31.5t
Index
Numerics
300 mm ceiling panel extension (1177-632) 115
A
Accessories from other manufacturers, coupling with Zeiss system 62
Accessories on OPMI, cable installation 34
Accessory components, power supply 71
Accessory interface 62
Adapter plate 99, 100
Adapter plate (1071-924) 115, 121, 122
Adapter plate mounted in centered position 106
Adapter plate, offset 102
B
Blind cover 64
Building vibrations 26
C
Ceiling anchor plate 10
Ceiling anchor plate, installation positions 12
Ceiling anchor plate, mounting of 42
Ceiling panel 10, 46, 64, 115
Ceiling panel extension 300 mm (11.8") (1177-632) 46
Ceiling track mount (305620-0000.000) 114
Column length 54, 58, 122
Column, vertical alignment 29, 118
Conduits for future upgrades 34
Conduits, installation 34
Confirmation of execution of installation 47, 73, 119
Confirmation of structural calculation 47, 73, 119
Constructional requirements 29, 118
Coupling of Zeiss system with accessories from other manufacturers 62
Cover frame 64
Customer's responsibilities 24, 69, 116
D
Dimensions, S8 ceiling mount 7
Dimensions, S81 ceiling mount 8
Discipline-specific installation positions 76
Distance from floor 14, 54, 58, 122
E
Electrical components on adapter plate 108, 112
Electrical components on mounting bracket 106, 110