Proline CC
Proline CC
pan/ceph/Dimax3
TECHNICAL MANUAL
Publication part number 688402   En
Version 29
Published 2004-05
                                                                                                                                TABLE OF CONTENTS
      Chapter G    TROUBLESHOOTING
                   1   SERVICE MODE SIGNALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                         G-1
                   2   ERROR MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                     G-1
                   3   FUSE DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                                   G-1
                   4   EQUIPMENT COMPLETELY DEAD . . . . . . . . . . . . . . . . . . . . . . . . . .                                                G-5
                       4.1 Check the mains outlet voltage to be within limits. (±10% of nominal voltage). . . . . G-5
                   5   FLAT BATTERY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-7
                   6   DISTURBANCES IN ROTATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
                       6.1    Rotation part does not move at all . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
                       6.2    Stop in the middle of the rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
                       6.3    Rotation cycle gets stuck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-9
                       6.4    Trembles during rotating cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-9
                       6.5    Abnormal noise during rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-9
                   7   DISTURBANCES IN CASSETTE MOVEMENT . . . . . . . . . . . . . . . . G-10
                       7.1    Cassette carriage does not move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-10
                       7.2    Cassette carriage stops suddenly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-10
                       7.3    Trembles during movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-10
                       7.4    Abnormal noise from the mechanism . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-10
                   8   DISTURBANCES IN PRIMARY SLOT MECHANISM . . . . . . . . . . . G-11
                       8.1 Primary slot does not move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-11
                       8.2 Primary slot does not stop to the right position . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-11
                   9   TROUBLES IN ADJUSTMENT OF FOCAL TROUGH MOVEMENT G-12
                       9.1 Forwards/backwards movement does not function. . . . . . . . . . . . . . . . . . . . . . . . . G-12
                       9.2 Movement area abnormal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-12
                       9.3 Erroneous position indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-12
                   10 DISTURBANCES IN TEMPLE SUPPORT MECHANISM . . . . . . . . G-13
                       10.1 Temple supports do not move . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-13
                       10.2 Temple supports function erroneously . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-13
                   11 DISTURBANCES IN Z-CARRIAGE UP/DOWN MOVEMENT . . . . . G-14
                       11.1     Z-carriage does not move. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-14
                       11.2     Z-carriage moves only to one direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-14
                       11.3     Z-carriage gets stuck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-14
                       11.4     Mechanism does not observe limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-15
1 WARNINGS
                             WARNING
                             THE FOLLOWING WARNINGS, CAUTIONS AND NOTES MUST ALWAYS BE CONSIDERED
                             WHILE SERVICING THE UNIT, IN ORDER TO AVOID EITHER PERSONAL INJURY OR DAM-
                             AGE TO THE UNIT
     2    MANUAL VERSIONS
                     PLANMECA pursues a policy of continuous product development. Although every effort is made
                     to produce up-to-date product documentation this manual should not be regarded as an infallible
                     guide to current production specifications. Planmeca reserves the right to make changes to spec-
                     ifications without prior notice.
         NOTE      PLEASE NOTE THAT THIS TECHNICAL MANUAL IS VALID FOR SOFTWARE REVISIONS
                   PK 3.75 OR LATER AND PG 2.25 OR LATER (X-RAYS WITH DIMAX3 PK 5.08 OR LATER
                   AND PG 5.08 OR LATER). If you find that your X-ray unit doesn’t have some keyboard func-
                   tions, or its keyboard functions differs from procedures described in this manual, then
                   either the software must be replaced with at least the above mentioned revision, or you
                   have to relate to an older version of this manual. The former procedure is recommended.
         3         TECHNICAL SPECIFICATIONS
                               X-ray tube               Toshiba D-052SB
                               Focal spot size          0.5 x 0.5mm
                               Target angle             5°
                               Total filtration         2.5 mm Al
                               Generator                Constant potential, microprocessor controlled,
                                                        operating frequency 80 kHz
                               Anode voltage            60-80 kV ± 2.5 kV
                               Anode current            4-12 mA ± 1.0 mA
                               Exposure time            2.5 - 18 s as indicated ± 10%
                               Cooling period           Automatically controlled
                               Film size                15 x 30 cm (film-based x-ray)
                               Cassette                 Flat (film-based x-ray)
                               SID                      480 mm
                               Magnification            Constant 1.2
                               Line voltage             100, 117, 220, 230-240 V~ 50 or 60 Hz
                               Regulation               Automatic ± 10%
                               Line current             max. 8A at 230V~, 15A at 100V~
                   Electrical classification
                               Class                    I
                               Type                     B
                   Mechanical data
                               Weight                   108 kg
                               Dimensions               (HxDxW) 2200x930x890 mm, 86.6x36.6x35 inch
                               Color                    RAL 9002
                   Environmental requirements
                               Ambient temperature      Operating      +5°C to +40°C
                                                        Storage        -10°C to +50°C (film-based x-ray)
                                                                       0°C to +50°C (Dimax3 digital x-ray system)
                               Humidity                 25% - 75%
                   Manufacturer
                               PLANMECA Oy, Asentajankatu 6, 00810 Helsinki, FINLAND
                               phone: +358-0-7590 5500
          Minimum filtration
                     The Radiation port contains filtration of 1.5mm aluminum. The measured half-value layer is
                     0.52mmAl at 80kVp. The measured value corresponds to aluminum equivalent of 2.5mmAl.
          Exposure time
                     The beginning and end points of the exposure time are defined at 70% of the peak radiation
                     waveform measured with a calibrated x-ray monitor.
          Tube current
                     Is defined using the resistance and voltage over the feedback resistor measured with a calibrated
                     multimeter. The mA value is then the voltage divided by the resistance.
          The nominal x-ray voltage together with the highest x-ray tube current obtainable from the
          high-voltage generator when operated at its nominal x-ray tube voltage
                     80 kV 12mA
                   The highest x-ray tube current together with the highest x-ray tube voltage obtainable from
                   the high-voltage generator when operated at its highest x-ray tube current
                              12mA 80kV
                   The x-ray tube voltage and x-ray tube current which result in the highest electric output
                   power
                              80kV 12mA
                   The nominal electric power for a load time of 0.1s and at the nominal x-ray tube voltage
                              80 kV 12mA - 1500W
5 GENERAL FEATURES
                       In order to protect the x-ray tube high voltage generator circuitry as well as the tube head and the
                       x-ray tube itself against various kinds of overload conditions certain actions are taken, explained
                       in the following.
           Earthing check
                       The temperature measurement is done by using the tube head protective earth wire as signal
                       return to the generator processor board. This will create a less than 2 mA current to the wire and
                       allows to check that both the tube head and the electronics ground are connected to the protec-
                       tive earth. If not, a temperature sensor open error will be indicated and generator is disabled.
           Feedback check
                       Before the exposure is started both the kV- and mA-feedback inputs are measured. If the cables
                       are not connected properly or for some other reason the results are not normal the system does
                       not allow exposures and an error is indicated.
                   Starting exposure
                              If the pre-heat phase was passed without error the high voltage generator is enabled. In normal
                              conditions the kilovolts will rise within about 8 ms and the mA-value rises simultaneously.
                              Depending on the component tolerances the voltage rise will start from 5 to 15 ms after the ena-
                              ble first was active. After 40 ms from the enabling both the tube voltage and current are moni-
                              tored. If the voltage has not reached the desired value within 6 kV the generator will be stopped
                              with an error indication ‘kilovolts do not rise’. Correspondingly, if the current is not within 1 mA the
                              generator is stopped with indication ‘mA does not rise’.
                   Tube arching
                              If during the exposure a sudden drop of kV is detected the generator is stopped with an error indi-
                              cation ‘Sudden drop of kV detected’. If the x-ray tube is arching this is normally the procedure.
                              However, it might happen that the kV drops down and rises back up between the 2.5 ms micro-
                              processor measurement interval. In this case the generator is stopped by overvoltage and error
                              indication “overvoltage in tube head” will be displayed.
                   Overvoltage detection
                              The generator has a separate overvoltage detection circuitry that will stop the generator if the
                              tube head voltage exceeds about 92 kV. The microprocessor is monitoring and adjusting the volt-
                              age, but this circuit is a separate hardware comparator, working independently from the micro-
                              processor.
                     The maximum operating frequency is limited due to the heating of the X-ray tube. There are two
                     different limiting factors:
                     First is the time that has to be waited between exposures in order to allow the heat to spread out
                     from the tube anode into the oil of the tube head. If this was not done, the tube anode can start to
                     melt during the next exposure.
                     The second factor is the maximum allowed temperature of the tube head itself. It shall not
                     exceed 60 °C because of the thermal expansion of the oil and also because of the thermal limits
                     of the components inside.
                     In Planmeca Proline CC Panoramic X-ray unit, there is an automatic function implemented in the
                     microprocessor to calculate the cooling periods between the exposures. Depending on the type
                     of the X-ray tube and the exposure parameters used the wait time will be calculated and shown
                     in the clock display. This function protects the X-ray tube not allowing an exposure to be taken
                     while the wait time is on. The maximal wait time that can be generated by this function is about 1
                     min. 40 sec., so it does not remarkably reduce the actual exposure rate and still makes it impos-
                     sible to destroy the X-ray tube by misuse.
                     In order to control the tube head temperature the microprocessor is measuring it all the time. If it
                     raises above 40 °C the fan is started and if it raises above 60 °C the temperature is displayed in
                     the clock display and further exposures are prevented. The system is not able to calculate the
                     actual waiting time in this case because that depends mainly on the ambient parameters (room
                     temperature, air flow etc.), but the temperature is displayed to allow the user to estimate the wait-
                     ing time. (It will normally be about 10 min.) It is very important that the temperature measurement
                     is in order. Therefore it is checked all the time and an error message is generated, preventing fur-
                     ther exposures, if the measurement circuit is either short circuiting or an open circuit.
         1         KEYBOARD OVERVIEW
                                    This chapter describes keyboard functions in normal and service modes. There are some special
                                    control and display modes that are not normally available but which can be entered by certain
                                    control sequences. In the following it is explained what operations can be achieved with the keys
                                    and which indications appear on the displays.
20 19
                         1                                    C
                                                                                                           18                   17
          2      PAN         CEPH   3       B                                     D
           4     READY              5       A                 P                   E        16                                   15
                 PRET
21
          6                                              °                         kV                       mA
                                    7                                                                                    AUTO   14
                                                         mm
                                                  9                         10           11           12
                                    8                                                                                    CTL    13
                       The keyboard is always in the normal operating mode when the unit is turned on. Please refer to
                       the Planmeca Proline CC Panoramic X-ray User’s manual for descriptions for the normal key-
                       board operation and instructions for normal everyday use of the unit.
                       Press the control (CTL) key to enter the special function mode. The CTL-key’s indicator light
                       comes on. In the special function mode some of the keys have special functions, others do not
                       operate in this mode. When the CTL-key or a special function key is pressed, the indicator light
                       goes off and the keyboard is in the normal mode. In this section the special functions are
                       explained.
                                                                                                  FOR DETAILS,
   USER SETTING                                 KEY SEQUENCE
                                                                                                   REFER TO
TIME & DATE ADJUST-                                                                             “SETTING     THE
MENT                      CTL                                             READY
                                                                                                CLOCK” on page D-
                                                                          PRET
                                                                                                24
CTL
                                                                                                              FOR DETAILS,
      USER SETTING                                         KEY SEQUENCE
                                                                                                               REFER TO
    AUTOPRINT           SET-                                                                                “Autoprint settings”
                                  CTL          AUTO                                 READY
    TINGS                                                                           PRET                    on page B-5
                                              hold for
                                              3 seconds
AUTO
CTL
                               Press the CTL-key and then press and hold for 3 seconds the return- key to change the state of
                               the automatic return function. The indicator light of the return key indicates whether the function
                               is on or off. With the automatic return on the equipment will automatically return to cassette load-
                               ing position after panoramic exposures and when returning from the cephalostatic mode.
CTL
                                                hold for
                                              4 seconds
                               The temple rest motor can be inactivated. Press the CTL-key and press and hold down for 4 sec-
                               onds the temple support key. The text HLd 0 will appear on the display. Remove the supports,
                               the unit can be used without them.
                                AUTO
                                                                                        hLr0
                                                hold for
                                              4 seconds
                               The temple supports will automatically close when the ready key is pressed.
                               The automatic temple support closing can be switched off as follows: Press the auto key. The
                               indicator light will come on. Press and hold down the temple supports key for four seconds. The
                               text “hLr0” will appear briefly on the main display indicating that the automatic temple support
                               closing is inactivated.
                               To switch the automatic temple support closing on repeat the above keying sequence. The text
                               “hLr1” will appear briefly on the main display indicating that the automatic temple support closing
                               is activated.
kV mA
                            CTL
                                                            OR
                        The test mode enables the X-ray to be operated without radiation being generated. Press the
                        CTL-key and press any of the kV selection and mA selection keys. The kV and mA displays will
                        clear indicating that the X-ray tube has been switched off and that you have entered the test
                        mode.
                        Repeat the above procedure to exit the test mode and switch the radiation on again. The mA and
                        kV values will reappear on the display.
CTL
                        The parameters of the 25 previously taken exposures are stored into the memory of the equip-
                        ment and can be examined or taken into use if desired. Pressing CTL and then the UP or DOWN
                        key activates this parameter history display mode showing the last taken exposure parameters.
                        In this mode the keys ready, up, down and CTL only are operative. Operating the other keys
                        does nothing. The functions of these keys are:
                        UP            Find next parameters (newer one)
                        DOWN          Find previous parameters (elder one)
                        READY         Take the parameters shown into use and exit to normal
                                      mode
                        CTL           Exit to normal mode
                        All displayed parameters (but the time display) indicate the parameters selected during the spe-
                        cific exposure. If the exposure time display is blinking, it indicates that the taken exposure was a
                        cephalostatic one. Note that everything, including the special segment selections, sinus mode
                        etc. are stored into the memory and can be restored with ready key.
           NOTE         The cephalostatic exposures don’t store the layer position but the soft tissue wedge posi-
                        tion.
                        The time display indicates the used exposure mode, e.g. PAn, CEPH, P 30.
                        If less than 25 exposures have been taken the history does not go down to 25 but only to the
                        amount of exposures taken.
                        If the history parameters are not desired to be taken into use, there are two ways to exit this
                        mode without altering the currently selected parameters.
                        When the CTL- key is pressed the equipment returns to normal mode.
                        When ready is pressed while the running number 00 is displayed the currently selected parame-
                        ters are taken into use (no change).
                        Pressing ready changes the currently selected parameters to the ones that were displayed and
                        returns the equipment into normal mode.
AUTO
                               Pressing CTL and then the auto - key allows to change the setting of the Autoprint film marking
                               system. The film viewing side and the date format can be changed. A letter and a number will
                               appear on the display. The letter indicates the film viewing orientation and the date format, and
                               the number the contrast of the text that will appear on the film.
                               Pressing the auto key switches from one film viewing side and date format to the other.
                               U is American film orientation. The image appearing on the film is a view of the patient’s teeth
                               seen as if you were standing behind the patient. The date will also be American format - month/
                               day/year. The date will appear on the film with dashes between the day, month, and year (12-31-
                               92).
                               E is European film orientation. The image appearing on the film is a view of the patient’s teeth
                               when looking directly at the patient. The date will also be European format - day/month/year. The
                               date will appear on film with periods between the day, month, and year (31.12.92).
                               U1 is European orientation with the American date format.
                               Pressing the height adjusting keys changes the contrast of the text. There are eight contrast lev-
                               els, 1 will give you the lightest text and 8 the darkest.
CAUTION Operating the exposure switch will immediately start the X-Ray generator.
            NOTE          The operations of the rotation and cassette motors are disabled in this mode to allow gen-
                          erator operation without movements.
                          For further diagnosis of the equipment operation a separate service mode has been provided.
                          Various kinds of adjustments and calibrations can also be made in this operating mode. There
                          are also some special control and display modes that are not normally available but which can be
                          entered by certain control sequences. In this section it is explained what operations can be
                          achieved with the keys and which indications appear on the displays.
                               CTL
                                     +   CEPH
                          Press the CTL-key and simultaneously press and hold down the cephalometric mode selection
                          (CEPH) key until the temperature of the tube head appears on the display (about four seconds).
                          The ready key indicator light will start to flash. This indicates that you have entered the service
                          mode.
            CAUTION       Some of the functions in the service mode may be fatal to the unit’s proper operation.
                          Never use special service mode functions if you are not familiar with how they operate.
CEPH To exit the service mode press the CEPH key briefly.
                          Some of the keys, displays and indicator lights have different meanings in the service mode. In
                          the special function mode some of the indicator lights have third meaning. The special function
                          mode is entered by pressing the CTL-key.
                   °        Patient positioning keys are used to switch on the positioning “Patient position forward/
                   mm       lights and to move the patient.                                backward keys” on page
                                                                                           B-7
                            The rotation and cassette movements are not activated in the
                            service mode.
                                  mA setting keys are used to change the mA-value.                     “mA keys and display” on
                      mA                                                                               page B-8
                                  The display indicates the selected mA-value. During an exposure
                                  the display indicates the actual mA-value measured from the X-
                                  ray tube.
                                  Exposure time display shows the selected type of the X-ray tube.     “Exposure time display”
                                                                                                       on page B-9
                   Main display
                                The main display in the upper left corner normally displays the tube head temperature. However,
                                if the system has detected an error the main display indicates the error number occurred. The
                                error is cleared by pressing CTL-key and the temperature display reappears. Also, similarly to
                                the normal mode, if the tube head temperature exceeds 60 °C the temperature display blinks and
                                the exposures are not allowed and if it is not possible to make a new exposure due to the X - ray
                                tube cooling characteristics the blinking count down time is displayed.
                                When the CTL-key is pressed (the CTL key indicator light is on), the potentiometer signal of the
                                cephalostat soft tissue light is shown on the display.
                   CAUTION      To allow the seek of the limit sensor in the service mode there is no end stop before the
                                mechanical end. The motor shall not be driven against the ends to prevent it from being
                                damaged!
                   CAUTION      When operating the layer adjustment motor be sure that nothing gets between the moving
                                handles and the groove ends.
                   NOTE         During an exposure in the service mode this display does not indicate the selected kV-
                                value but the actual kV-value measured across the X-ray tube.
                     kV
                                With the minus (-) key the kV - value can be decreased in steps of 2 kV. When kept down for
                                more than 0.5 seconds an auto-repeat function activates.
                                With the plus (+) key the kV - value can be increased in steps of 2 kV. When kept down for more
                                than 0.5 seconds an auto-repeat function activates.
          NOTE         During an exposure in the service mode this display does not indicate the selected mA-
                       value but the actual mA-value measured from the X-ray tube.
              mA
                       With the minus (-) key the mA - value can be decreased in steps of 1 mA. When kept down for
                       more than 0.5 seconds an auto-repeat function activates.
                       With the plus (+) key the mA - value can be increased in steps of 1 mA. When kept down for
                       more than 0.5 seconds an auto-repeat function activates.
          NOTE         In transversal slicing system models the wide slot (number 7) substitutes the reduced
                       height panoramic slot (number 1).
                       The primary slot display indicates the number of the primary slot selected. While the slot is
                       changing the display is blinking. In the following table it is explained, which slots correspond to
                       which numbers.
                       Slot number           Slot
                       0 (or blank)          Normal panoramic slot
                       1                     Reduced height panoramic slot
                       2                     Lower jaw (not in ceph. models)
                       3                     Upper jaw (not in ceph. models)
                       Cephalostat models only:
                       4                     Vertical asymmetric
                       5                     Vertical symmetric
                       6                     Horizontal asymmetric
                       Transversal slicing system models only:
                       7                     Wide slot
                              If the letters Fd is shown on the display, it indicates that the system has not been adjusted for the
                 Fd           filament parameters of the current tube and must be allowed to do this by pressing the exposure
                              switch down.
                              During exposure in the service mode this display shows the measured tube filament voltage.
                   NOTE       Make sure that the unit is in the service mode before performing the functions described
                              in this section.
                              Press the control (CTL) key to enter the special function mode. The CTL-key’s indicator light
                              comes on. In the special function mode some of the keys have special functions, others do not
                              operate in this mode. When the CTL-key or a special function key is pressed, the indicator light
                              goes off and the keyboard is in the normal mode. In the following the special functions are
                              explained.
                                              hold for
                                            4 seconds
    PRIMARY SLOT CALI-                                                    °                             “Primary slot calibra-
                                 CTL                                                    CTL
    BRATION                                                               mm                            tion” on page D-7
                                                                                                        Dimax3:
                                                                                                        “Radiation    beam
                                             hold for                                                   adjustment” on page
                                           4 seconds                                                    E-14
    EXPOSURE         SWITCH                                                                             “Exposure    switch
                                  CTL           READY
    SELECTION                                   PRET                                                    selection” on page
                                                                                                        B-14
                                               hold for
                                             4 seconds
                                                                                                        Dimax3:
                                                    hold for                                            “Patient positioning
                                                  4 seconds                                             mechanism adjust-
                                                                                                        ment” on page E-31
    PANORAMIC/CEPHA-                                                                                    “Panoramic/cepha-
                                 CTL
    LOSTATIC SETUP                                                                                      lostatic setup” on
                                                                                                        page B-15
                                             hold for
                                            10 seconds
 MODEL SELECTION:                                                                      “Model selection” on
                        CTL                       mA
 CC/CC PROLINE                                                                         page B-16
                                     hold for
                                    10 seconds
 TUBE TYPE     PRO-                                                                    “Tube type program-
                        CTL                  C
 GRAMMING                                                              EXP             ming” on page B-15
                               B                            D
                                                                 hold until finished
                               A              P     hold for E
                                                  4 seconds
                                  CTL
                                                              Total exposure counter
                                   OR
                                  CTL
                                                              Panoramic exposure counter
                                  OR
                                  CTL                         Cephalometric exposure counter
                               In the service mode it is possible to verify the amount of exposures taken. There are three
                               counters in the system, one counting the normal panoramic exposures, one counting the cepha-
                               lostatic exposures and one counting every exposure (also the ones made in the service mode).
                               Exposures that caused an error and were interrupted are not counted. When one of the counters
                               is wanted to be seen, after the right key sequence the counter is displayed in the main display for
                               a few seconds after which the normal temperature display reappears. The key sequence always
                               needs to be started with the CTL-key to get the indicator light on and the second key determines,
                               which counter is displayed.
                               Because the main display has only four numbers on it and is therefore capable of showing counts
                               only up to 9999 a special method has been provided to be able to display counts up to 49 999.
                               The two dots between the numbers indicate the amount of 10 000 to be added to the displayed
                               number as follows:
                               Lower dot on                 Add 10 000
                               Upper dot also on            Add 20 000
                               Lower dot blinking           Add 30 000
                               Upper dot also blinking      Add 40 000
                               When 50 000 exposures has been taken the counter is cleared to zero and starts to count the
                               next 50 000.
                                  CTL                                            READY
                                                                                 PRET
                               In order to make the equipment failure analysis easier an error history display has been provided.
                               This error memory contains the 50 last errors occurred (not Er 00). The stack depth is not neces-
                               sarily 50, if less errors have occurred. In this case only the amount of errors that have happened
                               is displayed.
                               Press the CTL-key and the return key to enter this mode. The last error occurred (-01) is shown.
                               If no errors were memorized the displays appear blank. In this mode all the indicator lights are
                               undetermined and the numeric displays have a different function as is explained in the following:
                               PRIMARY SLOT DISPLAY                Blinking E to indicate Error display mode
                               kV-DISPLAY                          Day of month when error occurred
                               mA-DISPLAY                          Month when error occurred
                               MAIN DISPLAY                        Hours and minutes when error occurred
                               EXPOSURE TIME DISPLAY               Error code number that occurred
                               mm-DISPLAY                          Running number of this error in the stack (-01 - -50)
                               In this mode the keys READY, UP and DOWN only are operative. The functions of these keys
                               are:
                               UP                                  Find next error (newer one, towards -01)
                               DOWN                                Find previous error (elder one, towards -50)
                               READY                               Exit to service mode
                                                                      °
                        CTL                                                        READY
                                                                      mm           PRET
                        OR
                        CTL
                                    hold for
                                  4 seconds
                      The up/down motor first second speed can be adjusted individually to both directions to suit the
                      various needs of the equipment operators and to compensate for the individual frictions and
                      weight balances of the equipment.
                      Press the CTL and press and hold down the up key for 4 seconds to activate the adjustment
                      mode for up-direction (CTL and down key for down-direction). In this mode the keys ready, up,
                      down, mm- and mm+ only are operative. The functions of these keys are:
                      UP                      Move the equipment into direction up
                      DOWN                    Move the equipment into direction down
                      mm-                     Adjust the speed slower
                      mm+                     Adjust the speed faster
                      READY                   Exit this mode and program the new speed into the memory
                      The main display indicates the direction to be adjusted and the relative percentage of the current
                      speed with an arrowhead pointing up or down and the speed (00-99). The speed indication is rel-
                      ative, 00 meaning the minimum and 99 the maximum possible slow adjustment speed. 00 is not
                      stop and 99 is about half of the maximum speed of the motor. The factory default is 55.
                      The adjustment is altered with the mm- and mm+ keys after which the speed can be checked by
                      moving the equipment with UP or DOWN keys. Note that the direction under modification never
                      accelerates to the full speed but continues at the slow speed to allow the proper verification of it.
                      The opposite direction operates normally.
           NOTE       In this mode the end limit sensors are not checked to the direction under modification.
                      This allows to get the Z-carriage easily in it’s normal operating range if it was for some
                      reason jammed outside the limit sensor and had lost the right position information. See
                      chapter UP/DOWN motor for information of the operating philosophy.
                        CTL           READY
                                      PRET
                      The remote exposure switch type can be selected. The new type is equipped with a ready light
                      that will come on when the unit is correctly se up and ready to take an exposure.
                      Press the CTL-key and the READY key to select the desired exposure switch type. When the
                      switch type equipped with the ready light is selected, the ready light on the holder will come on. If
                      there is not a ready light on the holder, the exposure indicator light will come on to indicate that
                      the wrong type switch has been selected and you must select the other switch type. When the
                      switch type without the ready light is selected, the ready light on the holder is off and indicates
                      that the wrong type switch has been selected.
CTL
                               The equipment software has been readily equipped with the functions needed for cephalostatic
                               operation. In the service mode it is determined whether the equipment is equipped with a cepha-
                               lostat. Of course the cephalostatic locking coils have to be added, the special cassette holder
                               assembled, adjustments made etc. to gain the cephalostatic operation. This chapter only
                               describes the keyboard operations needed when changing the equipment to a cephalostatic one.
                               Press the CTL-key and then press either the sec- or sec+ key to change the mode of operation.
                               The new state of operation is indicated for a few seconds in the main display, PA for panoramic
                               operation and CE for panoramic/cephalostatic one.
CTL B D EXP
                                                             hold for
                                                           4 seconds
                               A possibility has been provided to be able to use two different types of X-ray tube in the equip-
                               ment. The possible types are Comet XL-90-5°/0.6B and Toshiba D-052SB. The main difference
                               is the size of the focal spot of these tubes: 0.6 x 0.6 mm and 0.5 x 0.5 mm correspondingly.
                               Because the filament characteristics and the loading curves of the tubes are not the same, the
                               type of the X-ray tube in the tube head must be entered into the system to ensure proper opera-
                               tion. Also, if the Toshiba type is used the jumper J3 (see picture below) on the generator board
                               must be placed in position “T” and with Comet types in position “C”. Change fuse F7 to 4 A.
                               Press the CTL-key and press and hold the segment key A, B or C for 4 seconds, the tube fila-
                               ment definition will start. The display will now show Fd.
                               To prevent from accidental programming the segment key has to be held down for 4 seconds
                               before it operates.
                   CAUTION     Always program the tube type into the system when the generator processor board or the
                               tube head is replaced. Even if the new tube type is the same than the previous one, the fil-
                               ament characteristics are not exactly the same and must be reprogrammed. Failing to do
                               this may damage the tube head.
                               Pressing down the exposure key after the tube type programming starts the generator with 70 kV
                               and 2 mA (the system adjusts for these values automatically). The system regulates the filament
                               voltage to get the 2 mA tube current accurate, changes to 3 mA and so on till all tube current val-
                               ues up to 12 mA have been adjusted. After the learning procedure the generator stops and the
                               tube type display stops blinking to indicate that the system is now ready for normal operation. If
                               the exposure switch is released before the procedure is complete (about 2 minutes), it will be
                               continued from that point on when the switch is operated again. If the power was switched off the
                               procedure starts from beginning. When the tube filament definition is done the exposure time dis-
                               play will indicate “C.6” for Comet 0.6 tube and “t.5” for Toshiba tube.
NOTE Always check the model type when the Keyboard processor board is replaced.
                      The equipment software has been equipped with the software for the PM 2002 CC, Planmeca
                      Proline CC, PM 2002 EC and Planmeca Proline EC X-ray units. In the service mode it is deter-
                      mined what the equipment model is.
CC / CC Proline
CTL mA
                                        hold for
                                       10 seconds
                      Press the CTL-key and press and hold the mA- -key for 10 seconds to change the model of the
                      X-ray unit.The X-ray unit will restart and the software version number is shown on the main dis-
                      play, 3.00 or larger for the Planmeca Proline CC and 2.xx for the PM 2002 CC X-ray unit.
EC Proline/ CC Proline
CTL kV
                                          hold for
                                         10 seconds
                      Press the CTL-key and press and hold the kV+ -key for 10 seconds to change the model of the
                      X-ray unit.The new model of the unit is indicated for a few seconds in the main display, EC for the
                      Planmeca Proline EC and CC for the Planmeca Proline CC X-ray unit.
AUTO
                                    hold for
                                  2 seconds
                      Press the auto key and press and hold the program key for 2 seconds to recall the factory default
                      quick exposure settings: A 60 kV/4 mA; B 62 kV/5 mA; C 64 kV/6 mA; D 68 kV/7 mA; E 70 kV/9
                      mA.
         1         HELP MESSAGES
                          The following is a list of the help messages. These messages appear on the main display if a key
                          is pressed that has no function, or if the function is not allowed for some reason. The help mes-
                          sage disappear automatically when you release the key.
HE 2 The error code must first be cleared from the display by pressing the control key.
                        HE 3            The unit must be driven to the ready position with the special program keys (in
                                        Transversal slicing system mode).
                        HE 4            The patient positioning mechanism can be moved only in the manual transversal
                                        slicing system mode.
                        HE 6            The temple support motor has been inactivated. The temple support movement
                                        does not operate.
                        HE 8            You can not change the horizontal segment, there is only one primary slot in the
                                        panoramic mode.
                                        In Transversal slicing system mode the primary slot is selected automatically.
                        HE 9            The soft tissue filter can not be moved when taking a posterior-anterior or ante-
                                        rior-posterior exposure.
HE 10 The X-ray unit is busy, wait until the previous function is completed.
                       HE 15            The x-ray unit is not communicating with the PC, e.g. the Dimaxis software is not
                                        running.
                       HE 18            In the cephalometric mode there is also a sensor head in the rotating unit in addi-
                                        tion to the sensor head attached to the cephalostat. Remove the sensor head
                                        from the rotating unit.
     2    ERROR MESSAGES
                      The purpose of the error messages is to make fault-finding easier. The list of the last 50 error
                      occurrences is of a great help. Please refer to the section “Error history display” on page B-13
                      how to display the error list.
Er 04 The cassette carriage not in ready position when the exposure switch was pressed.
Er 06 The rotating unit not in the end sensor (in transversal slicing system mode).
Er 08 The wrong type sensor head (i.e. panoramic type) attached to the cephalostat.
Er 20 Tube power generator time out. The exposure time is longer than a specified time.
Er 21 Up/down motor time out. The vertical carriage does not reach the limit sensor within a specified time.
   Er 23       Layer adjust motor time out. The layer mechanism does not reach the limit sensor within a specified
               time.
Er 24 Primary slot motor time out. The slot mechanism does not reach the sensor in specified time.
   Er 25       Cassette motor time out. The cassette holder mechanism does not reach the limit sensor within a
               specified time.
Er 26 Rotation motor time out. The rotating unit does not reach the limit sensor in the specified time.
   Er 28       Panoramic sensor head motor time out. The sensor head does not reach the panoramic position in the
               specified time. Or the safety limit switch is pressed and the movement of the sensor head is prevented.
Er 64 The radiation reaching the AEC sensor is very low or missing completely.
           Er 04 The cassette carriage not in ready position when the exposure switch was pressed.
                       The cassette carriage was not driven to the ready position before taking an exposure. Press the
                       ready key to move the cassette carriage and the rotating unit to the ready position and then take
                       an exposure.
           Er 06 Rotating unit not in the end sensor (in Transversal slicing system mode)
                       The rotating unit is not in the end sensor (Transversal slicing system mode). Drive the unit to the
                       end sensor with the special program keys and take an exposure.
                        - Friction drive slipping forward (centering spring drawing it to this forward direction)
                        - Inoperative sensor (giving pulses, not good signal)
                        - Check in service mode and adjust if necessary
NOTE If the unit is equipped with the STOP switch, check that it is not activated.
                   Er 64 The radiation reaching the AEC sensor is very low or missing completely
                              Check that the cassette is AEC compatible, i.e. marked with text “AEC COMPATIBLE”. If the
                              cassette type is correct, perform the AEC calibration, refer to section “AUTOMATIC EXPOSURE
                              CONTROL (AEC) CALIBRATION” on page D-26. If the calibration does not correct the situation,
                              change the AEC slide frame assembly.
           NOTE        If the generator processor is not working properly the error is indicated by Er 61. In some
                       cases it might happen that the improperly operating generator processor generates Er 74
                       by overloading the keyboard processor. Check by disconnecting generator processor.
                   ER 90 Keyboard processor
                               Keyboard processor EEPROM not activated.
                               - Change keyboard processor board
                   ER 91Keyboard processor
                               Keyboard processor stack overflow.
                               - Change the software on the keyboard processor board.
                   NOTE   This chapter contains the information required to set up and calibrate the Planmeca Pro-
                          line CC Panoramic x-ray. In case your unit is equipped with Dimax digital sensor head,
                          perform the adjustments according to the instructions given in chapter “DIMAX3 DIGITAL
                          SYSTEM ADJUSTMENT” on page E-1.
                          Protect yourself from radiation when you are checking the beam alignment. Since radiation
                          safety requirements vary from state to state and country to country it is the responsibility of the
                          installer to ensure that the correct precautions are observed.
                          The display values shown in this guide are only examples and should not be interpreted as rec-
                          ommended values unless otherwise stated.
                          WARNING
                          IT IS VERY IMPORTANT THAT THE ROOM IN WHICH THE X-RAY IS INSTALLED AND THE
                          POSITION FROM WHICH THE USER OPERATES THE EQUIPMENT ARE CORRECTLY
                          SHIELDED. SINCE RADIATION SAFETY REQUIREMENTS VARY FROM COUNTRY TO
                          COUNTRY AND STATE TO STATE IT IS THE RESPONSIBILITY OF THE INSTALLER TO
                          ENSURE THAT ALL LOCAL SAFETY REGULATIONS ARE MET.
     1    REQUIRED TOOLS
          Panoramic mode calibration
                     •   Alignment ruler (part number 50973). For checking the position of the patient positioning
                         mechanism.
                                    Alignrul.eps
                                                                                                   Alignment ruler
                     •   Ball phantom (part number 50971). For checking the position of the patient positioning mech-
                         anism and the positioning lights.
                                                                                             D
                                                                                          EN
                                                                                                          1
                                                                                                        97
                                                                                                     .50
                                                                                                   No
                                                                                                                        CA
                                                                                                                      ME
                                                                                                                    AN
                                                                                                                  PL
                                                                                          INE
                                                                                                                             T
                                                                                                                           AR
                                                                                       Y-L          INE                  ST
                                                                                                 X-L
Ball phantom
                     •   Frankfort plane alignment tool (part number 50977). Used with the ball phantom for checking
                         the position of the Frankfort plane light.
                                                                                                          Frankfort plane
                                                                                                          alignment tool
                                                                      D
                                                                     N
                                                                    E
                                                                                           71
                                                                                         09
                                                                                      o.5
                                                                                     N
                                                                                 7
                                                                                                        C
                                                                               97
                                                                                                       E
                                                                            .50
                                                                                                      M
                                                                                                     N
                                                                                                   LA
                                                                          NO
                                                                                                  P
                                                            OY
                                                                                                              T
                                                                      E
                                                                                                             R
                                                                    IN
                                                         CA
                                                                                                            A
                                                                                      E
                                                                  -L
                                                       ME
                                                                                    IN
                                                                                                          S
                                                                 Y
-L
                                                     AN
                                                                                 X
PL
                     •   Beam alignment tool (part number 50972). For checking the position of the panoramic x-ray
                         beam.
                                                                                                                                                         2
                                                                                                                                                      23
                                                                                                                                                    52
                                                                                                                                                  .6
                                                                                                                                                No
                                                                                      2                                            4
                                                                                   8x
                                                                                 t1
                                             6
                                                                              sta
                                                                           alo
                                                                         ph
                                                                      Ce
                                                                  e
                                                               lin
                                                             ro
                                                           2P
                                                        200
                                                   PM
                                              CA
                                            ME
                                          AN
                                        PL
                             •   Ceph head support alignment tools (part number 653126). The part number includes two
                                 pieces.
                           Before the x-ray is used the position of the radiation beam must be checked. Remove the inner
                           covers from tube head assembly and the cassette drive assembly. Four screws hold the tube
                           head cover in position, these are located at the rear of the tube head assembly. The screws that
                           hold the cassette drive cover in place are located at the front and lower side of the cover. The pri-
                           mary slot mechanism is located under the tube head cover.
                           Figure 1
                           Switch the unit on. Enter the test mode (see section “Test mode” on page B-4).
                           Press the ready key to drive the x-ray into the exposure position. The rotating unit and cassette
                  READY
                  PRET     carriage will move to their respective ready positions. When the x-ray is ready the indicator light
                           will come on.
                           The horizontal sector display will remain empty which indicates that the primary slot number 0
                           has been selected.
                           Figure 2
                           Start to take a test exposure by pressing and holding down the exposure button on the remote
                           control. Stop pressing the exposure button when the tube head has rotated to a convenient posi-
                           tion for viewing the secondary slot.
                               Place the beam alignment tool into the secondary slot. Note which way round it goes. It is advis-
                               able to leave the cassette in position as it will reduce the amount of scattered radiation.
                     Secondary slot
                                                           Beam alignment tool
                               Figure 3
                               Enter the service mode (see section “How to enter/exit the service mode” on page B-6). The radi-
                               ation beam from the first slot can now be checked without the rotating unit moving.
                               Darken the room sufficiently so that you will be able to see the image of the radiation beam on
                               the alignment tool (it is fluorescent and will glow when the radiation beam strikes it), but not so
                               dark that you cannot see the borders of the alignment rectangle.
                               Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in the
         4    mA
                    68    kV   darkened room. The actual values will depend on how dark the room is.
                               Protect yourself from radiation and press the exposure button. The beam image will appear on
                               the alignment tool. Observe the beam from behind the tube head.
                   CAUTION     Radiation is generated when the exposure button is pressed. Take adequate protection
                               measures. Keep the exposure time as short as possible.
                               The beam image should appear within the borders of the rectangle marked on the alignment tool
                               (if the beam is not correctly aligned see sections “Primary slot calibration” on page D-7 and
                               “Radiation beam too far to the right or left” on page D-8).
UP
                                                                Alignment rectangle
                                                                                                                                        no. 50972
                                                                                                                                        PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                           NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
                                                                                                                                                                                 Beam
                         Primary slot number 0
                                                                                                                       DOWN
                                                               Beam alignment tool
                               Figure 4
                               If the beam from the first slot is correctly aligned you should check the beam alignment of the
                               other slots. Note that the height of the other beams will vary as the slots are of different sizes.
                        Press the horizontal segment selection key to select the next primary slot. The number 1 will
                1       appear on the horizontal sector display. This is the second primary slot. Repeat the previously
                        described alignment checking procedure for this beam. If this beam is not aligned correctly refer
                        to sections “Primary slot calibration” on page D-7 and “Radiation beam too far to the right or left”
                        on page D-8 for information on how to align the beam. When it has been checked the number 2
                        and number 3 slots must be checked.
                                                                                                                                                                                                                                                UP
                                                                                                                                                                                                                                                                                         no. 50972
                                                                                                                                                                                                                                                                                         PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                                                                                                                        NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
                 Primary slot number 1
                                                                                                                                                                                                                                                                                                                                    Beam
                                                          Beam alignment tool                                                                                                                                                       DOWN
                        Figure 5
                        If the x-ray unit is equipped with the Transversal slicing system, the slot number 1 is replaced
                 7      with the slot number 7. The radiation beam from slot number 7 must be slightly wider than the
                        rectangle marked on the alignment tool and the same height as the rectangle.
                                                                                                                                                                                                        UP
                                                                                                                                                                                                                                                no. 50972
                                                                                                                                                                                                                                                PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                                                                                NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Beam
                        Figure 6
                        By repeatedly pressing this key you can select each slot in turn so that they can be checked and,
                        if necessary, adjusted.
                                                                                                                  UP                                                                                                                                                                                                                                                               UP
                                                                                                                       no. 50972
                                                                                                                                                                                                                                                                                                                                                                                        no. 50972
                                                                                                                       PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                      WARNING
                                                                      DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE X-RAYS
                                                                      ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION WITH
                                                                      THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM IS
                                                                      STILL MISALIGNED REPEAT THE OPERATION.
                                                                      If the beams are too high or low loosen the four screws that hold the primary slot plate in position
                                                                      and adjust as necessary. Tighten the screws and recheck the beam. Repeat the above proce-
                                                                      dure until the beam is correctly positioned.
               Beam too low                                                                  Beam too high                   Securing screws
                                                                      UP                                                                                                                                               UP
                                                                                                                                                                                                                                           no. 50972
                                                                           no. 50972
DOWN DOWN
Figure 8
                   NOTE                                               The four primary slots are machined in one plate. Any adjustments made to the primary
                                                                      slot plate or slot mechanism will automatically move all the slots. If any adjustments are
                                                                      necessary you will probably find that adjusting the first slot is all that is needed to cor-
                                                                      rectly align all the slots.
                                                                      If the beam is not vertical loosen the two screws that hold the primary slot mechanism to the col-
                                                                      limator tube and rotate the mechanism until the beam is correctly positioned. Tighten the screws
                                                                      and recheck the beam. Repeat the above procedure until the beam is correctly positioned.
                    Beam not vertical                                                                                                                       UP
                                                                                                                                                                                                                                                                                             Securing screws
                                                                                                                                                                                                  no. 50972
                                                                                                                                                                                                  PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                                    NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
DOWN
Figure 9
NOTE The ceph primary slots are also adjusted according to instructions given in this section.
                        WARNING
                        WARNING: DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE
                        X-RAYS ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION
                        WITH THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM IS
                        STILL MISALIGNED REPEAT THE OPERATION.
NOTE The ceph primary slots are also adjusted according to instructions given in this section.
                        If any of the beams are too far to the left or right of the alignment rectangle they must be cen-
                        tered.
                                                                                    UP                                                                                    UP
                                                                                         no. 50972
                                                                                                                                                                               no. 50972
                 Beam too far to the left                                                                                                                                                                                    Beam too far to the right
                                                                                         PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
DOWN DOWN
Figure 10
          NOTE          A microprocessor controlled motor determines the position of each slot. If the beam is
                        readjusted to the left or right the new position must be reprogrammed into the memory.
                        Press and hold down the primary slot key until the number of the slot that you are checking
               1        appears on the primary slot display and starts to flash. You are now in the primary slot calibration
                        mode.
                        For example, if you have just checked the position of the beam from the slot number 1, the
                        number 1 will appear on the display.
                        Adjust the position of the beam by pressing the patient positioning (mm) keys. The minus key will
                        move the slot mechanism, and beam to the left, and the plus key will move the slot mechanism,
                        and beam to the right.
                        Check the position of the beam again and, if necessary, repeat the above alignment procedure
                        until the beam is within the alignment rectangle on the fluorescent screen.
          CAUTION       Radiation is generated when the exposure button is pressed. Take adequate protection
                        measures. Keep the exposure time as short as possible.
                   NOTE   Always use the plus (+) key for the final adjustment. This makes sure that the slot mecha-
                          nism is correctly positioned by taking up the play between the draw nut and the position-
                          ing thread.
                          For example, if you are adjusting the beam to the left by pressing the minus (-) key you should
                          move the slot mechanism a little further to the left than necessary, and then press the plus (+)
                          key to move it back to the right to the correct position.
                                                               Final adjustment
                                                               in this direction
Position thread
Draw nut
                          Figure 11
                          If you cannot adjust the beam far enough to the left using the method described above you may
                          have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab slightly
                          and then tighten the screw. The slot position should then be rechecked.
Metal tab
Positioning thread
Draw nut
                          Figure 12
                          Press the horizontal segment selection key to store the slot position in the memory and move on
                      2   to the next slot. The number of the next slot will appear on the display and start to flash.
                          Repeat the alignment procedure described above for this slot and any other slots.
                          Exit the primary slot calibration mode and check the position of all the beams again. Realign if
       CTL                necessary.
                          Exit the service mode.
                            You should take an exposure of the ball phantom to check that the patient positioning mecha-
                            nism is correctly positioned.
                                                                  D
                                                                    N
                                                                      E
                                                                                   71
                                                                                 09
                                                                              o.5
                                                                                    N
                                                                                        A
                                                                                        C
                                                                                            E
                                                                                            M
                                                                                                N
                                                                                                LA
                                                                                                    P
                                                                     E
                                                                                                    T
                                                                   IN
                                                                                                     R
                                                                                                        A
                                                                Y-L
                                                                                                         T
                                                                            IN
                                                                                                            S
                                                                          -L
                                                                                X
                            Figure 13
                            Move the cassette carriage to the loading position. The patient positioning lights will come on
                            when the carriage stops moving.
                            Slide the ball phantom into the patient positioning mechanism.
                      °
            00        mm    Adjust the position of the patient positioning mechanism to 00.
                            In the darkroom open the cassette and place a film inside the cassette and then place a piece of
                            paper the same size on top, the paper will reduce the radiation that reaches the film. Close the
                            cassette.
Paper
Film
Cassette
                            Figure 14
                            Slide the cassette into the cassette holder. Make sure that you push it all the way in.
                            Enter the service mode.
                            Select a kilovolt value of 60 and a milliampere value of 2. Note that it is only possible to select
    2      mA
                 60    kV
                            these values in the service mode.
                            Drive the x-ray unit to the exposure position. The rotating unit will turn to the start position and the
                    READY
                    PRET    indicator light will come on.
Take an exposure. Hold the exposure down for the duration of the exposure.
                              Press the return key to drive the cassette to the loading position, remove the cassette and proc-
                              ess the film.
                              The images of 23 balls should appear on the film (Fig. 15). They should be round, all the same
                              size, and evenly spaced. Note that they will not necessarily be at the same vertical height. The
                              outermost balls on the left and right sides should appear symmetrically. If the distance from the
                              outermost ball to the edge of the exposed area on the right side differs more that 2mm from the
                              corresponding distance on the left side, the rotation start limit must be adjusted (see section
                              “Rotation limits” on page D-23). Note, that the adjustment of the patient positioning mechanism
                              must be performed first.
                              The distance from the center of the middle ball to the center of the rear middle ball should be less
                              than 2mm. If the distance is not correct you will have to adjust the position of the patient position-
                              ing mechanism (see section “Patient positioning mechanism adjustment” on page D-14).
120±2mm 120±2mm
Film
Middle ball
2mm
                              Figure 15
                              The distance from the center of the middle ball to the center of the tenth ball to the left and the
                              tenth ball to the right should be 120mm ± 2mm. If this distance is correct then the patient posi-
                              tioning mechanism need not be adjusted, and you should then check the patient positioning
                              lights (see section “PATIENT POSITIONING LIGHTS” on page D-19).
                              If these distances are not correct you will have to adjust the position of the patient positioning
                              mechanism. The following diagrams are a rough guide as to how to position it correctly. Note that
                              you may have to make more than one adjustment as the patient positioning mechanism could be
                              incorrectly positioned in more than one direction. For example it may be too far forward and too
                              near the column.
                     If the distance from the middle ball to the center of the tenth ball on the left-hand side is more
                     than 122mm and on the right-hand side less than 118mm then the patient positioning mechanism
                     must be moved away from the column (Fig. 16). Note, that the distance between the line on the
                     underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
                     section “Adjustment in ball phantom’s y-direction” on page D-14.
                                                                                  D
                                                                                 N
                                                                                E
                                                                                                      1
                                                                                                    97
                                                                                                  50
                                                                                                o.
                                                                                               N
                                                                                                                              A
                                                                                                                             C
                                                                                                                            E
                                                                                                                           M
                                                                                                                          N
                                                                                                                        LA
                                                                                                                       P      T
                                                                                  E
                                                                                                                             R
                                                                                IN
                                                                                                                                    A
                                                                                              E
                                                                              -L
                                                                                                                                   T
                                                                                            IN
                                                                                                                                  S
                                                                             Y
                                                                                          -L
                        Right-hand side
                                                                                         X
                                                  Film
                     Figure 16
                     If the distance from the middle ball to the center of the tenth ball on the left-hand side is less than
                     118mm and on the right-hand side more than 122mm then the patient positioning mechanism
                     must be moved towards the column (Fig. 17). Note, that the distance between the line on the
                     underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
                     section “Adjustment in ball phantom’s y-direction” on page D-14.
                                                                                           D
                                                                                          N
                                                                                         E
                                                                                                                   1
                                                                                                                 97
                                                                                                               50
                                                                                                             o.
                                                                                                            N
                                                                                                                                               A
                                                                                                                                              C
                                                                                                                                             E
                                                                                                                                            M
                                                                                                                                           N
                                                                                                                                         LA
                                                                                                                                        P         T
                                                                                           E
                                                                                                                                                 R
                                                                                         IN
                                                                                                                                                A
                                                                                                               E
                                                                                       -L
                                                                                                                                               T
                           Right-hand side                                                                   IN
                                                                                                                                              S
                                                                                      Y
                                                                                                           -L
                                                                                                          X
Film
                     Figure 17
                     If the distances from the middle ball to the center of the tenth ball on both the right and left are
                     more than 122mm then the patient positioning mechanism is too far backward and must be
                     moved forwards (Fig. 18).
                                                                                                 D
                                                                                                N
                                                                                               E
                                                                                                                     1
                                                                                                                   97
                                                                                                                 50
                                                                                                               o.
                                                                                                              N
                                                                                                                                                 A
                                                                                                                                                C
                                                                                                                                               E
                                                                                                                                              M
                                                                                                                                             N
                                                                                                                                           LA
                                                                                                                                          P
                           Right-hand side
                                                                                                                                                       T
                                                                                             E
                                                                                                                                                      R
                                                                                           IN
                                                                                                                                                     A
                                                                                                                 E
                                                                                         -L
                                                                                                                                                    T
                                                                                                               IN
                                                                                                                                                   S
                                                                                        Y
                                                                                                             -L
                                                                                                            X
Film
Figure 18
                   If the distances from the middle ball to the center of the tenth ball on both the right and left are
                   less than 118mm then the patient positioning mechanism is too far forward and must be moved
                   backwards (Fig. 19).
                                                                              D
                                                                             N
                                                                            E
                                                                                           1
                                                                                         97
                                                                                       50
                                                                                     o.
                                                                                    N
                                                                                                      A
                                                                                                     C
                                                                                                    E
                                                                                                   M
                                                                                                  N
                                                                                                LA
                                                                                               P
                     Right-hand side
                                                                                                         T
                                                                              E
                                                                                                        R
                                                                            IN
                                                                                                       A
                                                                                       E
                                                                          -L
                                                                                                      T
                                                                                     IN
                                                                                                     S
                                                                         Y
                                                                                   -L
                                                                                  X
                                             Film
Figure 19
               READY
                       Press the ready key to drive the x-ray unit to the exposure position. When the x-ray is ready the
               PRET
                       indicator light will come on.
                       Start to take a test exposure by pressing and holding down exposure button on the remote con-
                       trol. Stop pressing the exposure button about halfway through the exposure cycle so that the
                       rotating unit stops moving.
                       An error code will appear on the display. Clear the error code from the display by pressing the
           Er 00       CTL-key.
                       Manually position the rotating unit so that the center line that runs through the tube head and film
                       cassette mechanism is parallel to the wall to which the x-ray is secured. Slide the alignment pin
                       through the hole, nearest the column, in the top of the vertical carriage so that it goes into the
                       positioning hole in the rotating unit. (Fig. 20)
Alignment pin
Figure 20
           NOTE        Adjust the position of the rotating unit by holding the cassette carriage. Never use the
                       tube head to adjust the position.
                   Leave the ball phantom tool in position and place one end of the alignment ruler in the first pri-
                   mary slot (the largest one) and the other end in the secondary slot.
                                                                                             D
                                                                                                 N
                                                                                                 E
                                                                                                         1
                                                                                                             97
                                                                                                                  50
                                                                                                                   o.
                                                                                                                       N
                                                                                                                            A
                                                                                                                            C
                                                                                                                                E
                                                                                                                                M
                                                                                                                                    N
                                                                                                                                        LA
                                                                                                                                         P
                                                                                                                                         T
                                                                                             E
                                                                                                                                             R
                                                                                            IN
                                                                                                                                                 TA
                                                                                                     E
                                                                                        -L
IN
                                                                                                                                                  S
                                                                                        Y
                                                                                                         -L
                                                                                                             X
                                                                                                                                                           Primary slot
Alignment ruler
Secondary slot
                   Figure 21
                   Loosen the four screws that hold the patient position mechanism in place and adjust its position
                   to the left or right until the mechanism is centered.
Screws
                   Figure 22
                   Make sure that the y-line on the ball phantom is parallel to the line on the underside of the align-
                   ment ruler. Note that the line on the ball phantom and the line on the alignment ruler do not have
                   to coincide, but they must be parallel, the deviation may be ±1mm.
                                                                                                     Alignment ruler
                                               PLANMECA                No.50971
START END
                                                                                                                           Ball phantom
                                                     S
                   Figure 23
                   Tighten the screws, and remove the alignment pin.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
                      If the patient positioning mechanism is too far forward or too far backwards the fore/aft position
                      must be adjusted and the new position programmed into the memory.
              READY
                      Press the ready key to drive the x-ray unit to the exposure position. When the x-ray is ready the
              PRET
                      indicator light will come on.
                      Start to take a test exposure by pressing and holding down exposure button on the remote con-
                      trol. Stop pressing the exposure button after about two seconds so that the rotating unit stops
                      moving.
                      An error code will appear on the display. Clear the error code with the CTL-key.
        Er 00
                      Manually position the rotating unit so that the center line that runs through the tube head and film
                      cassette mechanism is at a right angle to the wall. Slide the alignment pin through the hole in the
                      top of the vertical carriage so that it goes into the positioning hole in the rotating unit. (Fig. 24)
Alignment pin
Figure 24
          NOTE        Adjust the position of the rotating unit by holding the cassette carriage. Never use the
                      tube head to adjust the position.
                                Leave the ball phantom tool in position and place the alignment ruler in position between the pri-
                                mary slot and the secondary slot as described earlier. The line on the alignment ruler must line
                                up with the x-line on the ball phantom (Fig. 25).
                     Secondary slot side                       PLANMECA                       No.50971                Tube head side
                                                       START                                             END
                                                                      X-LINE
                                 Alignment ruler
                                                                                     ENIL-Y
                                           Ball phantom
                                                                      X-LINECA.EPS
                                Figure 25          0° position
                                Enter the service mode.
                           °    Press either of the patient positioning keys once to switch the patient positioning lights on then a
             12            mm
                                second time to adjust the position of the patient positioning mechanism until the line on the ball
                                phantom lines up with the line on the alignment ruler. The position of the mechanism will appear
                                on the display.
                           °    Hold the patient positioning plus (+) key down (about five seconds) until you hear a signal and
               00          mm
                                the millimeter display zeroes. This indicates that the new position of the patient positioning mech-
                                anism has been programmed into the memory.
                                Remove the alignment pin and exit the service mode.
                                Take another ball phantom picture to check the alignment again. Readjust if necessary.
                       READY
                                Press the ready key to drive the x-ray unit to the exposure position. When the unit is ready the
                       PRET
                                indicator light will come on.
                                Start to take a test exposure by pressing and holding down exposure button on the remote con-
                                trol. Stop pressing the exposure button after about 16 seconds so that the rotating unit stops
                                moving (near the end position).
                                An error code will appear on the display. Clear the error code with the CTL-key.
             Er 00
                      Manually position the rotating unit so that the center line that runs through the tube head and film
                      cassette mechanism is at a right angle to the wall. Slide the alignment pin through the hole in the
                      top of the vertical carriage so that it goes into the positioning hole in the rotating unit.
          NOTE        Adjust the position of the rotating unit by holding the cassette carriage. Never use the
                      tube head to adjust the position.
                      Leave the ball phantom tool in position and place the alignment ruler in position between the pri-
                      mary slot and the secondary slot as described earlier. Measure the distance between the line on
                      the alignment ruler and the x-line on the ball phantom (Fig. 26). The distance must be less than
                      1mm.
                                                                              ENIL-Y
                                      Ball phantom             X-LINECA.EPS
                              The positions of the light beams from the three patient positioning lights should be checked.
                              Place the Frankfort plane alignment tool into position on top of the ball phantom.
                                                                                  D
                                                                                                                                  Frankfort plane alignment tool
                                                                                      N
                                                                                      E
                                                                                                               71
                                                                                                                    09
                                                                                                                        o.5
                                                                                                                          N
                                                                                                                                  A
                                                                                                          77
                                                                                                                                    C
                                                                                                                                      E
                                                                                                     09
                                                                                                                                         M
                                                                                                                                          N
                                                                                                .5
                                                                                                                                             LA
                                                                                             NO
                                                                                                                                              P
                                                                        OY
                                                                                                                                              T
                                                                                  E
                                                                                                                                                  R
                                                                                 IN
                                                                   CA
                                                                                                                                                  A
                                                                                                           E
                                                                             -L
                                                                                                                                                      T
                                                              ME
IN
                                                                                                                                                      S
                                                                             Y
                                                                                                                -L
                                                         AN
                                                                                                                    X
                                                    PL
Ball phantom
Figure 27
                         °    Press either of the patient positioning keys to switch the positioning lights on and then a second
               00        mm
                              time to drive the patient positioning mechanism to 00 if it is not already there.
                              The Frankfort plane light beam should be located between the two black lines on the alignment
                              tool. It should be horizontal.
                                                                                   D
                                                                                  N
                                                                                 E
                                                                                                                 71
                                                                                                               09
                                                                                                            o.5
                                                                                                           N
                                                                                                   7
                                                                                                                                    C
                                                                                                 97
                                                                                                                                   E
                                                                                              .50
                                                                                                                                  M
                                                                                                                                 N
                                                                                                                               LA
                                                                                            NO
                                                                                                                              P
                                                              OY
                                                                                                                                          T
                                                                              E
                                                                                                                                         R
                                                                            IN
                                                           CA
                                                                                                                                        A
                                                                                                           E
                                                                          -L
                                                         ME
                                                                                                         IN
                                                                                                                                      S
                                                                         Y
-L
                                                       AN
                                                                                                      X
PL
Figure 28
                              The plane-in-focus light beam should be positioned so that it is on the black reference line on the
                              side of the ball phantom.
                                                                                        D
                                                                                       N
                                                                                      E
                                                                                                                         71
                                                                                                                       09
                                                                                                                    o.5
                                                                                                                   N
                                                                                                       7
                                                                                                                                        C
                                                                                                     97
                                                                                                                                       E
                                                                                                 .50
                                                                                                                                      M
                                                                                                                                     N
                                                                                                                                   LA
                                                                                               NO
                                                                                                                                  P
                                                                  OY    INE
                                                                                                                                                  T
                                                                                                                                                 R
                                                               CA
                                                                                                                                                A
                                                                     Y-L
                                                                                                               E
                                                             ME
                                                                                                                   IN
                                                                                                                                              S
                                                                                                                -L
                                                           AN
                                                                                                                     X
PL
Figure 29
The midsagittal light beam should strike the black line on the front of the ball phantom.
                                                                                        D
                                                                                       N
                                                                                      E
                                                                                                               71
                                                                                                             09
                                                                                                          o.5
                                                                                                         N
                                                                                                                           A
                                                                                                     7
                                                                                                                          C
                                                                                                  97
                                                                                                                         E
                                                                                               .50
                                                                                                                        M
                                                                                                                       N
                                                                                                                     LA
                                                                                             NO
                                                                                                                    P
                                                                                Y
                                                                                       INE
                                                                                                                              T
                                                                             AO
                                                                                                                             R
                                                                                                                            A
                                                                          C         Y-L
                                                                                                                           T
                                                                        ME
IN
                                                                                                                          S
                                                                                                        -L
                                                                     AN
                                                                                                       X
                                                                   PL
                       Figure 30
                       If the light beams do not coincide with the black lines or they are not in focus, they should be
                       adjusted.
Bending the mirror bracket backwards or forwards will position the light beam.
                                                                                                                                                     Mirror bracket
                                          Adj.beam rightleft.eps
                       Figure 31
                       Rotating the mirror bracket will adjust the light beam position in vertical direction.
                                                                                                                               Adj.beam updown.eps
Figure 32
                              The clearance between the guidance profile and the counterprofile can be checked and adjusted
                              if necessary when the loading spring of the drive wheel is removed. Also the drive wheel with
                              transmission can be removed, but this is not necessary. Note that all major parts must be
                              installed so that the rotational part has about its final weight.
                              Provided that these requirements are fulfilled the clearance can be observed by moving the rotat-
                              ing unit sidewards from both ends (see figure below). Note that the rotating unit has to be half
                              way trough the rotating cycle. When moving the rotating unit parallelly sideways the longitudinal
                              clearance appears as clearance.
                              Figure 33
                              If there is any clearance it is eliminated by unscrewing screws (see figure below) and pushing the
                              counterprofile towards the cassette mechanism. Do not use any tools to move the counterprofile.
                              Tighten the screws by pressing the screw head with finger. Check once more that there is no
                              clearance left but also that the contact between the counterprofile and the guidance profile is not
                              tight. This can be done by carefully pushing the rotating unit from the cassette end using the
                              same speed at which it moves when it is driven by the stepper motor and turn it approximately
                              45Ž to both directions. If there is too much tension, the movement is stiff and continually sticks in
                              the middle area.
Figure 34
                     The guidance wheel can be adjusted if the driving wheel with gear is detached and clearance
                     between guidance and counter profile and is correctly adjusted. The guidance wheel is attached
                     to the rotating unit (1) through a flange (2). The flange is fixed to the rotating unit with screws (3)
                     and a bolt pin (4). By unscrewing the screws (3) the flange can be turned around the bolt pin (4).
                     Figure 35
                     The guidance is adjusted when the rotating unit is in a position corresponding to the middle point
                     of the exposure. In principle, the guidance wheel should be placed as near the inner edge of the
                     groove as possible. For proper adjustment turn the rotating unit with both hands. Always turn the
                     rotating unit from the cassette head end and move it throughout the rotating cycle calmly and at a
                     speed corresponding to the actual speed. If the guidance wheel has been adjusted too far from
                     the inner edge of the groove, a clack of the counter profile meeting the guidance profile can be
                     heard when moving the unit from the start position to the middle position. If the guidance wheel
                     has been adjusted too close to the inner edge of the driving groove the movement is sticky in an
                     area corresponding to a 45° angle in relation to the central area. The adjustment can be found
                     only by experimenting.
                   NOTE         The adjustment of the patient position mechanism must be performed before adjusting
                                the rotation limits.
                                The rotating movement is controlled by two Hall detectors: one determines the start position and
                                the other the end position. The latter has no importance for the normal function because the step
                                motor drives the rotating unit according to the programmed number of steps The end limit is of
                                great importance in case of abnormal function, because it informs the processor that the rotating
                                unit has not reached the end position.
                                The start limit is adjusted so that the movement of the guidance wheel in the drawing groove is
                                symmetric to the groove. The distance of the boss of the wheel from the end of the groove should
                                be the same both when the movement starts and when it ends. The symmetricity of the move-
                                ment has more importance to the quality of the radiograph than the absolute length of the move-
                                ment which varies a little because of production tolerances.
                                Remove the necessary covers. Slide the ball phantom into the patient positioning mechanism.
                                Check that the y-line on the ball phantom is parallel to the line on the underside of the alignment
                                ruler according to instructions given in section “Patient positioning mechanism adjustment” on
                                page D-14.
                                Remove the alignment pin. Enter the test mode if you are not already in it.
                                Press the ready key to drive the x-ray unit to the exposure position. When the x-ray is ready the
                        READY   indicator light will come on. Place the alignment ruler in position between the primary and sec-
                        PRET
                                ondary slot and write down the angle between the ball phantom “start”-line and the alignment
                                ruler.
                                Start to take a test exposure by pressing and holding down exposure button on the remote con-
                                trol until the rotation unit has reached the end position. Place the alignment ruler in position
                                between the primary and secondary slot and write down the angle between the ball phantom
                                “end”-line and the alignment ruler.
                                The positions corresponding to end and start positions should not differ more than ± 1 mm. The
                                position of the start limit (Hall detector) can be changed by unscrewing the screws. This should
                                be repeated until the tolerance area is reached.
                                The adjustment of the end position is done after the start position has been adjusted.
                                No special tools are needed. The end limit should activate about 5 mm before the end of the
                                movement and it should activate even when the rotating unit is at 5 mm past the end position.
                                The indicator light shows this when the equipment is in the service mode.
           CTL                                         READY
                                                       PRET           12:30            13.08              1997
                                                                         time              date              year
                     Press the CTL and then the CLOCK- key to enter into the calendar clock set mode. In this mode
                     the keys CLOCK, READY, UP and DOWN only are operative. The functions of these keys are:
                     CLOCK         Go to the next number to be adjusted.
                     UP            Increase the number to be adjusted.
                     DOWN          Decrease the number to be adjusted.
                     READY         Exit the set mode and start the clock.
                     When first entered into this mode the time display unit minutes start blinking indicating that they
                     can now be adjusted. Note that the clock is still running before the first actual adjustment is done
                     and exiting this mode now with the READY-key does not modify the clock at all. Using the UP
                     and DOWN keys the number can be modified, which also stops the clock.
                     Operating the CLOCK- key allows the setting of the next parameter in the following order:
                     Minutes (units)       0-9
                     Minutes (tens)        0-5
                     Hours                 00 - 23
                     Day                   01 - 31
                     Month                 01 - 12
                     Year (thousands)      0-9
                     Year (hundreds)       0-9
                     Year (tens)           0-9
                     Year (units)          0-9
                     From unit years the setting returns to the unit minutes. The clock set mode can be exit from any
                     point. To make it accurate a time signal should be waited and the clock set mode exit made at
                     the very moment. Exiting this mode clears the seconds (that cannot be seen) and starts the
                     clock.
                             When an x-ray exposure is taken you will hear a warning tone indicating that radiation is being
                             generated.
                             If the warning sound of the radiation cannot be heard or is considered to be too loud, the tone
                             (and loudness) can be adjusted. Pressing CTL and then the P-key activates this adjustment
                             mode turning the sound on continuously. In this mode the keys READY, UP and DOWN only are
                             operative. Operating the other keys does nothing. The functions of these keys are:
                             UP            Adjust the tone higher
                             DOWN          Adjust the tone lower
                             READY         Exit this mode and program the new tone into the memory
                             After the panoramic x-ray beam adjustment the AEC-signal for primary slots number 0 and 1
                             must be checked. The position of the slot must be adjusted so that the AEC-signal value is as
                             large as possible, and the value for the primary slot number 1 must be equal to the value for pri-
                             mary slot number 0.
                             Enter the service mode: press the CTL-key and simultaneously press and hold down the cepha-
                             lometric mode selection (CEPH) key until the temperature of the tube head appears on the dis-
         CTL
               +    CEPH
                             play (about four seconds). The ready key indicator light will start to flash. This indicates that you
                             have entered the service mode.
                   AUTO      Enter the AEC mode by pressing and holding down the auto key until the indicator light will come
                             on.
                             Place the AEC calibration filter to the front of the first primary slot.
                                                                                               Aectool.eps
Make sure that the primary slot number 0 is selected (the horizontal sector display is empty).
Load the cassette with a film and slide it into the cassette holder.
                   CAUTION       Radiation is generated when the exposure button is pressed. Take adequate protection
                                 measures. Keep the exposure time as short as possible.
                                 During the exposure the AEC-signal is shown on the main display. Adjust the slot position in hor-
               U nn              izontal direction until the signal reaches its maximum value as follows.
                                 Press the CTL and press and hold down the primary slot key for 4 seconds to enter the beam
       CTL                       alignment mode.
                   hold for 4
                   seconds
                            °    Adjust the position of the beam by pressing the patient positioning (mm) keys. The minus key will
                            mm
                                 move the slot mechanism, and beam to the left, and the plus key will move the slot mechanism,
                                 and beam to the right. The primary slot adjustment in horizontal direction is described in section
                                 “Radiation beam too far to the right or left” on page D-8.
                                 Check the x-ray beam position with the beam alignment tool after the adjustment. If the beam
                                 image is not within the borders of the rectangle marked on the alignment tool, adjust the beam
                                 position slightly.
                                 After the primary slot number 0 has been adjusted, select the primary slot number 1 and adjust it
                        1        in the same way. Make sure that the value for the primary slot number 1 equals to the value for
                                 the primary slot number 0.
                                 Check the x-ray beam position with the beam alignment tool after the adjustment. Adjust the
                                 beam position slightly, if needed.
                   NOTE          Remove the AEC calibration filter from the primary slot before selecting the primary slot
                                 number 1. Position the AEC calibration filter to the front of the second primary slot before
                                 adjusting the AEC-signal.
                            The default kilovolt value, i.e. the value that is used when the AEC exposure starts, is 68 kV. The
                            film density is adjusted by changing the default milliampere value. The adjustment range is from
                            -3 to +3. The density value 0 corresponds the mA value 8. Changing the density value will
                            change the mA value as follows: one density step equals to one mA step. The negative value
                            decreases the mA value from the value 8mA, whereas the positive value increases it.
                                                                                   Density
                                                       kV             mA
                                                                                    value
                                                       68              5              -3
                                                       68              6              -2
                                                       68              7              -1
                                                       68              8               0
                                                       68              9               1
                                                       68             10               2
                                                       68             11               3
                            To find the exposure values that give the film density closest to value 1.0 ±0.1 OD the test expo-
                            sures are taken. The films are either compared to the reference film which density is approx. 1.0
                            OD (supplied with the system), or the film density is measured using a film densitometer.
                            It is recommended to start with exposure values 68kV and 7mA and check the film density. In
                            case other exposures are needed, decrease/increase the mA value until the correct film density
                            is achieved.
Place the AEC calibration filter to the front of the first primary slot (primary slot number 0).
                            In the darkroom open the cassette and place a film inside the cassette. Close the cassette. Slide
                            the cassette into the cassette holder.
                NOTE        The film density adjustment must be performed in normal panoramic mode. Make sure
                            that the unit is NOT in the AEC mode, i.e. make sure that the auto key indicator light is
                            NOT on.
                            Drive the x-ray unit to the exposure position. The rotating unit will turn to the start position and the
                    READY
                    PRET    indicator light will come on.
Take an exposure.
                               Press the return key to drive the cassette to the loading position, remove the cassette and proc-
                               ess the film.
                               Compare the film to the reference film: the film must be as dark as the reference film. Repeat the
                               procedure with different mA values until the correct film density is achieved, and then select the
                               density value that gives the right mA value as follows.
                               Enter the service mode: press the CTL-key and simultaneously press and hold down the cepha-
                               lometric mode selection (CEPH) key until the temperature of the tube head appears on the dis-
             CTL
                   +   CEPH
                               play (about four seconds). The ready key indicator light will start to flash. This indicates that you
                               have entered the service mode.
       CTL
                               To change the density value enter the film density adjustment mode: press the control key and
                       PAN
                               then press the pan key until the density value, indicated with “d”, is shown on the main display.
                               Change the value with the height adjusting keys.
                               Select the density value that corresponds the mA value, e.g. if the mA value is 7mA, select the
                   d -1        density value -1.
                       READY
                               Exit the film density adjustment mode by pressing the ready key.
                       PRET
           NOTE              Take five 10 seconds exposures in the service mode with a kilovolt value of 68 and a milli-
                             ampere value of 7 before adjusting the AEC-signals.
                             Make sure that the unit is in the service mode and that the AEC mode is selected. If needed,
                             enter the AEC mode by pressing and holding down the auto key until the indicator light will come
                             on.
                             In the service mode the AEC detector offset value (base signal) is shown on the main display
           U             8   when the unit is in the AEC mode and no exposure is taken. Set the offset to value 8 ±2 with the
                             offset adjustment trimmer located on the Cassette limit PCB. Turning the trimmer clockwise
                             increases the value and turning it counterclockwise decreases it.
            AdjAEC.eps
                             After the offset value is set the AEC-signal during the exposure is adjusted.
                             Make sure that the primary slot number 0 is selected (the horizontal sector display is empty).
                             Check that the aluminium block is in front of the slot.
                             In the darkroom open the cassette and place a film inside the cassette. Close the cassette. Slide
                             the cassette into the cassette holder.
                             Protect yourself from radiation and press the exposure button.
           CAUTION           Radiation is generated when the exposure button is pressed. Take adequate protection
                             measures. Keep the exposure time as short as possible.
                                         During the exposure the AEC-signal value is shown on the main display. Set this value to approx.
               U 60                      60 with the gain trimmer located on the Cassette limit PCB. Turning the trimmer clockwise
                                         decreases the value and turning it counterclockwise increases it. Repeat the adjustment until the
                    AdjAEC.eps
                                         correct value has been reached.
                   NOTE                  The final adjustment of the AEC-signal during the exposure must be done in the normal
                                         mode.
                   NOTE                  In the normal mode the AEC-signal is only shown on the main display if you previously
                                         were in the film density adjustment mode. In case the AEC-signal is not shown on the
                                         main display, enter the service mode and enter the film density adjustment mode. Then
                                         exit the film density adjustment and the service modes and the signal can be seen on the
                                         main display.
CEPH To exit the service mode press the CEPH key briefly.
                                         Drive the x-ray unit to the exposure position by pressing the ready key.
                                 READY
                                 PRET
                                         Take an exposure. Note that you can stop radiation in the middle of the exposure (approx. 4 sec-
                                         onds after starting the exposure). An error code 00 will appear on the display and start to flash.
                                         Press the CTL-key to clear the error code.
                     After the exposure the difference between the offset value and the AEC-signal during the expo-
         U 51        sure is shown on the main display. Set this value to 51 ±1 with the gain trimmer located on the
                     Cassette limit PCB. Repeat the adjustment until the correct value has been reached and check
                     the value by taking at least two exposures.
                     Check the value for the primary slot number 1 and adjust the slot position in horizontal direction,
                     if needed. Do not adjust the value by turning the trimmer.
          NOTE       Remove the AEC calibration tool from the primary slot before selecting the primary slot
                     number 1. Position the AEC calibration tool to the front of the second primary slot
                     (number 1) before adjusting the AEC-signal.
          NOTE       In the service mode the difference between the offset value and the AEC-signal value of
                     the latest exposure (taken in the normal mode) can be seen by pressing the CTL key. To
                     exit this mode press the CTL key a second time.
9 CEPHALOSTAT ADJUSTMENT
                              Loosen the two M4x8 BN 5950 bolts on the top of the cover with a 5.5 mm ring spanner. Loosen
                              the two M4x4 DIN 7985 screws on the bottom of the cover. Pull the cover away from the cassette
                              holder.
                                                                                                     RemCephCover2.eps
             M4x8 BN 5950
                        Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
  4              CEPH   ing unit and cassette carriage will move automatically to the correct position for taking cephalom-
                        etric exposures. Collimator number 4 will be automatically selected.
                        Remove the tube head covers and cassette carriage covers (see sections “Tube head covers” on
                        page H-1, “Secondary slot cover” on page H-1 and “Cassette carriage back cover” on page H-1).
                        Place the ceph head support alignment tools in the ear post holders as far as they go (Fig. 36, 1).
                        Rotate the head support to the 90° position.
                        Look through the opening on the cephalostat cassette holder and the opening on the head sup-
                        port alignment tools. The fourth primary slot is seen symmetrically if the adjustment is correct
                        (Fig. 36, 2).
                        Figure 36
                        If the fourth primary slot is not seen symmetrically, the head support requires readjustment. See
                        instruction given in section “Head support adjustment” on page D-35.
                               Switch the unit off. Remove the cephalostat arm upper cover, attachment casting cover and
                               attachment mechanism cover.
                               Make sure that the locking screw of the adjustment flange is loosened (Fig. 37, 1). Adjust the
                               head support position according to the Fig. 37. Rotate the head support around its vertical axis to
                               adjust the ear posts in horizontal direction (Fig. 37, 2) and around its horizontal axis to adjust the
                               ear posts in vertical direction (Fig. 37, 3).
                               If the cephalostat cassette holder does not run parallel with the primary slot, adjust the cassette
                               holder in vertical direction (Fig. 37, 4).
4 5
                                                                                                                                         2
                                                                                   6
Figure 37
Figure 38
9.4 Checking the soft tissue filter and the filter positioning light position
                                Clip the soft tissue filter positioning tool to the soft tissue filter which is at the end of the cassette
                                carriage (Fig. 39). This will enable you to see the soft tissue filter when you check its position.
Cassette carriage
Figure 39
                                Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
      4                  CEPH   ing unit and cassette carriage will move automatically to the correct position for taking cephalom-
                                etric exposures. Collimator number 4 will be automatically selected.
                                Orientate the Cephalometric beam alignment tool so that it is vertical and then place it in the
                                cephalometric cassette holder. Position it so that it is on the left of the cassette holder (lateral
                                exposure position).
                                        Cassette
                                        holder
4 5
                                        Beam
                                        alignment
                                                                                                                                        Positioning light 25mm
                                                                                       Soft tissue filter 0mm
                                        tool
                                                                         6
Figure 40
                                                                                                              Slide the soft tissue filter positioning light to left as far as possible. The light will come on and the
                                                                                                              soft tissue filter, on the cassette carriage, will automatically move to the correct position.
                                                                                                                           Patient positioning light
                                                                                                              Figure 41
                                                                                                              The beam from the soft tissue filter positioning light must strike the line on the beam alignment
                                                                                                              tool marked “Positioning light 0mm” and the number on patient positioning display must be 00. If
                                                                                                              the light is in the wrong position it must be repositioned and if the display number is wrong it must
                                                                                                              be recalibrated. If the light beam itself is crooked or out of focus the lens or mirror must be repo-
                                                                                                              sitioned. Refer to the section “Soft tissue filter positioning light calibration and adjustment” on
                                                                                                              page D-40.
4 5
                                                                                                                                                                             °
                                                                                                                                                                    00       mm
                     6
                                                                                                              Figure 42
                                                                                                              Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in a
                                                                                                              darkened room for example 62kV and 6mA. The actual values will depend on how dark the room
                                                                                                              is.
                                                                                                              Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                                                                                                              sure button. The image of the radiation beam from slot number 4 will appear on the alignment
                                                                                                              tool.
                                                                                                                       The radiation beam must appear within borders of the rectangle marked on the beam alignment
                                                                                                                       tool and to the right of the line marked “Soft tissue filter 0mm”.
                                                                                                                                              4         5
                                                                                                                                                                                                                                                       Radiation beam
                                                                                                                       Figure 43
                                                                                                                       If the left-hand side of the radiation beam does not reach the line marked “Soft tissue filter 0mm”
                                                                                                                       or extends passed the line the soft tissue filter is in the wrong position it must be repositioned.
                                                                                                                       Refer to the section “Soft tissue filter adjustment” on page D-43.
                                                                                                                       Move the soft tissue filter positioning light until it strikes the line on the beam alignment tool
                                                                                                                       marked “Positioning light 25mm”. The number 25 must appear on the on patient positioning dis-
                                                                                                                       play. If the display number is wrong it must be recalibrated. Refer to the section “Soft tissue filter
                                                                                                                       positioning light calibration and adjustment” on page D-40.
4 5
                                                                                                                                                                                                                                                                        °
                                                                                                                                                                                                                                                              25        mm
                           6
Figure 44
                               Cassette
                               holder
4 5
                               Beam
                               alignment
                        Figure 45
                        Press the CTL-key, the light beam position number will appear on the main display.
                        Slide the light to the right as far as possible and check that the position number on the display is
                        >10. If it is not, adjust the light assembly position with the light assembly lever (Fig. 46, 1) so that
                        the number will be >10.
2 3
                                  4        5
                                                                                                                                                                                                                                                         4
                                                                                                                                                                                                                                              0 - 48mm
                                                                                                                  Posiotioning light 25mm
                                                                         Posiotioning light 0mm
                                                Soft tissue filter 0mm
                                                                                                                                                                                                                                                             1
                                      6
                        Figure 46
                        Press the Vertical sector selection D key. The text “CAL” will appear on the main display and the
                        rotating unit and cassette carriage will move automatically to the cephalometric mode position.
                        You are now in the soft tissue filter positioning light calibration mode.
                        Slide the patient positioning light to left as far as possible. The beam from the soft tissue filter
                        positioning light should strike the line on the beam alignment tool marked “Positioning light 0mm”
                        (Fig. 46, 2).
                                 Move the soft tissue filter positioning light until it strikes the “Positioning light 25mm” line on the
                                 beam alignment tool (Fig. 46, 3).
                                 Press the Ready key. The soft tissue filter and the soft tissue filter positioning light are now cali-
                         READY
                         PRET
                                 brated and the text “CEPH” appears on the main display.
                                 Make sure that the soft tissue filter positioning light movement range is 0-48mm (Fig. 46, 4).
                                 Adjust the light assembly position with the light assembly lever, if necessary.
                                                4        5
                                                                 Posiotioning light 0mm
                                                                                          1
                                                    6
Figure 47
Turn the mirror bracket (Fig. 48, 1) until the light beam is in right position.
4 5
                     Figure 48
                     Turn the bulb housing (Fig. 49, 1) until the light beam is vertical. Fasten the screws of the bulb
                     fixing plate (Fig. 49, 2).
4 5
                                                                                                                      2
                                                    Posiotioning light 0mm
                                                                                                      1
                                   6
Figure 49
                                Check the radiation beam position according to instructions given in section “Checking the soft
                                tissue filter and the filter positioning light position” on page D-37. If the left side of the radiation
                                beam is too far left or right the radiation beam must be repositioned.
                                                                                                                                                                                                         2
                                                                         4
3 3
3 6 9 12 15 mm
                                                                                                                                                                                                             1
                                                                                            9   6                                            3 0                    3            6    9   12 mm
                                                                                   4        5
                                                                                                                                                       Posiotioning light 25mm
                                                                                                                          Posiotioning light 0mm
                                                                                                 Soft tissue filter 0mm
                                                                                       6
Figure 50
                   NOTE         Place the ceph head support alignment tools in the ear post holders and check the cepha-
                                lostat arm adjustment according to instruction given in section “Checking the head sup-
                                port position” on page D-34.
                         Press the ceph mode selection key to enter the ceph mode. The rotating unit and cassette car-
                         riage will move automatically to the correct position for taking cephalometric exposures.
              CEPH
Enter the service mode, see section “How to enter/exit the service mode” on page B-6.
                         Orientate the cephalometric beam alignment tool so that it is vertical and then place it in the
                         cephalometric cassette holder. Position it so that it is on the left of the cassette holder.
                                                                                                Cassette
                                                                                                holder
4 5
                                                                                                Beam
                                                                                                alignment                                                                                                        Positioning light 25mm
                                                                                                                                                                Soft tissue filter 0mm
                                                                                                tool
                                                                                                                                                  6
                         Figure 51
                         Press the collimator selection key twice to skip passed the two panoramic collimator slots and
               4         select the third collimator slot (the first cephalometric slot). The number 4 will appear on the dis-
                         play.
                         Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                         sure button. The image of the radiation beam from slot number 4 will appear on the alignment
                         tool.
NOTE The soft tissue filter will filter a narrow area of the left side of the radiation beam.
                         The radiation beam must appear within the borders of the rectangle marked on the alignment
                         tool. If it does not refer to section“Primary slot calibration” on page D-7. The adjustment is per-
                         formed in the same way as when adjusting the standard panoramic slot.
                             4         5
                                                                                                Positioning light 25mm
                                               Soft tissue filter 0mm
Beam
Figure 52
Slide the beam alignment tool to the right of the cassette holder.
                                                                                                                           Cassette
                                                                                                                           holder
                                                                                                                                                  4         5
                                                                                                                                                                                                                                                 6
                                                                                                                           Beam
                                               Figure 53
                                               Press the collimator selection key to reposition the third collimator slot (the first cephalometric
                   5                           slot). The number 5 will appear on the display. Note that the third collimator slot (the first ceph
                                               slot) is used to take both vertical symmetric and vertical asymmetric exposures. The slot is
                                               moved slightly to accommodate the two exposure views.
                                               Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                                               sure button. The image of the radiation beam from slot number 5 will appear on the alignment
                                               tool.
                                               The radiation beam must appear within the borders of the rectangle marked on the alignment
                                               tool. If it does not refer to the section “Primary slot calibration” on page D-7. The adjustment is
                                               performed in the same way as when adjusting the standard panoramic slot.
Beam
                               4         5
                                                                                                                            6
                                                                                                  Positioning light 25mm
                                                 Soft tissue filter 0mm
Figure 54
                                    Cassette
                                    holder
                                                                                                                                   No. 652232
                                                                                                   6
                                     Beam
                                     alignment
                                                                                                   5
                                                                                                   4
                       Figure 55
                       Press the collimator selection key to select the fourth collimator slot (the second cephalometric
              6        slot). The number 6 will appear on the display.
                       Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                       sure button. The image of the radiation beam from slot number 6 will appear on the alignment
                       tool.
NOTE The soft tissue filter will filter a narrow area of the left side of the radiation beam.
                       The radiation beam must appear within the borders of the rectangle marked on the alignment
                       tool. If it does not refer to “Primary slot calibration” on page D-7. The adjustment is performed in
                       the same way as when adjusting the standard panoramic slot.
                                                                                                   Beam
                                                      No. 652232
                   6
                                                      PLANMECA PM 2002 Proline Cephalostat 18x24
Figure 56
                       If you need to check any of the slots again you can scroll through the slots by repeatedly pressing
                       collimator selection key.
                       When all the slots have been checked, and if necessary adjusted, exit the service mode by
                       pressing the ceph key.
                   NOTE   This chapter contains the information required to set up and calibrate the Planmeca Pro-
                          line CC Panoramic x-ray with Dimax3 sensor head. In case your unit is a film-based pano-
                          ramic x-ray, perform the adjustments according to the instructions given in chapter
                          “ADJUSTMENT & CALIBRATION” on page D-1.
                          Protect yourself from radiation when you are checking the beam alignment. Since radiation
                          safety requirements vary from state to state and country to country it is the responsibility of the
                          installer to ensure that the correct precautions are observed.
                          The display values shown in this guide are only examples and should not be interpreted as rec-
                          ommended values unless otherwise stated.
                          WARNING
                          IT IS VERY IMPORTANT THAT THE ROOM IN WHICH THE X-RAY IS INSTALLED AND THE
                          POSITION FROM WHICH THE USER OPERATES THE EQUIPMENT ARE CORRECTLY
                          SHIELDED. SINCE RADIATION SAFETY REQUIREMENTS VARY FROM COUNTRY TO
                          COUNTRY AND STATE TO STATE IT IS THE RESPONSIBILITY OF THE INSTALLER TO
                          ENSURE THAT ALL LOCAL SAFETY REGULATIONS ARE MET.
     1    REQUIRED TOOLS
                     •   Alignment ruler (part number 50973). For checking the position of the patient positioning
                         mechanism.
                                 Alignrul.eps
                                                                                                              Alignment ruler
                     •   Ball phantom (part number 50971). For checking the position of the patient positioning mech-
                         anism and the positioning lights.
                                                                                                                                Ball phantom
                                                                                    D
                                                                                  EN
                                                                                                               1
                                                                                                             97
                                                                                                          .50
                                                                                                        No
                                                                                                                               CA
                                                                                                                             ME
                                                                                                                           AN
                                                                                                                         PL
                                                                                  INE
                                                                                                                                     T
                                                                                                                                   AR
                                                                            Y-L                         INE                      ST
                                                                                                     X-L
                     •   Frankfort plane alignment tool (part number 50977). Used with the ball phantom for checking
                         the position of the Frankfort plane light.
                                                                                                                   Frankfort plane
                                                                                                                   alignment tool
                                                                    D
                                                                     N
                                                                        E
                                                                                                 71
                                                                                              09        5
                                                                                                     o.
                                                                                                        N
                                                                                          7
                                                                                                                     C
                                                                                        97
                                                                                                                    E
                                                                                     .50
                                                                                                                       M
                                                                                                                         N
                                                                                                                           LA
                                                                                   NO
                                                                                                                             P
                                                         OY
                                                                                                                             T
                                                                    E
                                                                                                                                 R
                                                                IN
                                                      CA
                                                                                                                                A
                                                                                          E
                                                               -L
                                                    ME
                                                                                                IN
                                                                                                                                     S
                                                              Y
-L
                                                  AN
                                                                                                 X
PL
• Sensor alignment tool (part number 10002699). For attaching the beam alignment tool.
                                                                         calibtool.eps
                                                       Sensor alignment tool
• Beam alignment tool (part number 50972). For checking the position of the x-ray beam.
• Cephalostat calibration tool (part number 10002935). For calibrating the cephalostat.
                   •   Panoramic mode calibration block (part number 665413). For calibrating the panoramic sen-
                       sor head.
Figure 1
                              Unscrew the two attachment screws of the quick connector mechanism cover with the 4mm allen
                              screw and slide the cover from its position.
                   NOTE       In case the x-ray is a pan/ceph unit with two fixed sensor heads, the cover is attached to
                              two attachment pins. Unscrew the two attachment screws with the 2.5mm allen key and
                              remove the cover by sliding it horizontally from its position.
Cover
                              Figure 2
                              Loosen the four screws located on the lock disc.
Lock disc
Figure 3
                      Loosen the lock disc attachment screw with the 3mm allen key (Fig. 4, 1) and remove the lock
                      disc from the adapter (Fig. 4, 2).
1 2
                      Figure 4
                      Loosen the button axle attachment screw and push the button axle inwards and remove the sen-
                      sor head from its adapter.
                                                                                    Sensor head
             Button axle
                      Figure 5
                      The sensor alignment tool can now be attached to the adapter pins.
Sensor head
                              Figure 6
                              Place the lock disc to the adapter (Fig. 7, 1) and tighten the attachment screw (Fig. 7, 2) with the
                              3mm allen key.
1 2
Figure 7
                     Push the button axle towards the lock disk. Make sure that the head of the axle goes into the hole
                     on the lock disc and tighten the attachment screw with 3mm allen key.
Button axle
                     Figure 8
                     To remove the sensor head clearance tighten the four screws located on the lock disc slightly
                     with 2.5mm allen key.
                     Figure 9
                     Attach the covers to the adapter.
2.2 Sensor head with quick connector mechanism - removing the covers
                   NOTE      In certain market areas the movable sensor head is not available, in which case all cepha-
                             lostat-equipped units have two fixed sensor heads.
                             Unscrew the bottom cover plate attachment screw with the 4mm allen key and remove the cover
                             plate.
                             Figure 10
                             Turn the lock disc from above the locking plate and push in the quick connector mechanism lock-
                             ing knob. Unscrew the two attachment screws of the quick connector mechanism cover with the
                             4mm allen screw and slide the cover from its position.
Locking knob
Cover
Figure 11
                      Before the x-ray is used the position of the radiation beam must be checked. Remove the inner
                      cover from tube head assembly. Four screws hold the tube head cover in position, these are
                      located at the rear of the tube head assembly.
                                                                            Tube head
                                                                            assembly
                      Figure 12
                      Remove the sensor head from the quick connector mechanism (see instructions given in Plan-
                      meca Proline CC panoramic x-rays user’s manual), or remove the fixed sensor head from its
                      adapter (refer to section “Removing and attaching the fixed sensor head” on page E-4). Attach
                      the sensor alignment tool to the adapter. Place the beam alignment tool to the sensor alignment
                      tool as shown on the Fig. 13. Note which way round it goes. Slide the beam alignment tool as up
                      as it goes in this sensor alignment tool position.
                      Figure 13
                      Switch the unit on. Enter the test mode (see section “Test mode” on page B-4).
                      Press the ready key to drive the x-ray into the exposure position. The rotating unit will move to
              READY
              PRET    the ready position. When the x-ray is ready the indicator light will come on.
                              The horizontal sector display will remain empty which indicates that the primary slot number 0
                              has been selected.
Figure 14
                              Start to take a test exposure by pressing and holding down the exposure button on the remote
                              control. Stop pressing the exposure button when the tube head has rotated to a convenient posi-
                              tion for viewing the sensor head.
                              Enter the service mode (see section “How to enter/exit the service mode” on page B-6). The radi-
                              ation beam from the first slot can now be checked without the rotating unit moving.
                              Darken the room sufficiently so that you will be able to see the image of the radiation beam on
                              the alignment tool (it is fluorescent and will glow when the radiation beam strikes it), but not so
                              dark that you cannot see the borders of the alignment rectangle.
                              Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in the
         4    mA
                    68   kV   darkened room. The actual values will depend on how dark the room is.
                              Protect yourself from radiation and press the exposure button. The beam image will appear on
                              the alignment tool. Observe the beam from behind the tube head.
                   CAUTION    Radiation is generated when the exposure button is pressed. Take adequate protection
                              measures. Keep the exposure time as short as possible.
                      The beam image should appear within the borders of the rectangle marked on the alignment tool
                      (if the beam is not correctly aligned see section “Radiation beam adjustment” on page E-14).
UP
Alignment rectangle
                                                                                                                                                                                                              no. 50972
                                                                                                                                                                                                              PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                      NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
                                                                                                                                                                                                                                                                                                 Beam
                Primary slot number 0
                                                                                                                                  DOWN
                                                      Beam alignment tool
                      Figure 15
                      If the beam from the first slot is correctly aligned you should check the beam alignment of the
                      other slots. Note that the height of the other beams will vary as the slots are of different sizes.
                      Press the horizontal segment selection key to select the next primary slot. The number 1 will
               1      appear on the horizontal sector display. This is the second primary slot. Repeat the previously
                      described alignment checking procedure for this beam. If this beam is not aligned correctly refer
                      to section “Radiation beam adjustment” on page E-14 for information on how to align the beam.
                      When it has been checked the number 2 and number 3 slots must be checked.
                                                                                                                                                                                      UP
                                                                                                                                                                                                                                                       no. 50972
                                                                                                                                                                                                                                                       PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                                                              NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
                      Figure 16
                      If the x-ray unit is equipped with the Transversal slicing system, the slot number 1 is replaced
               7      with the slot number 7. The radiation beam from slot number 7 must be slightly wider than the
                      NORMAL BEAM and the same height as the rectangle marked on the alignment tool.
                                                                                                                                       UP
                                                                                                                                                                                      no. 50972
                                                                                                                                                                                      PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                  NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Beam
Figure 17
                             By repeatedly pressing this key you can select each slot in turn so that they can be checked and,
                             if necessary, adjusted.
                                                                                                                  UP                                                                                            UP
no. 50972
                                                                                                                                                                                                                     no. 50972
                                                                                                                       PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                             Figure 18
                             When all the slots have been checked, and if necessary adjusted, exit the service mode and take
                             a ball phantom exposure to check the position of the patient positioning mechanism, see section
                             “PATIENT POSITIONING MECHANISM” on page E-27.
           NOTE            In case the radiation beam is not correctly aligned first check the sensor head position by
                           using a spirit level. The sensor head (i.e. the sensor alignment tool) must be perpendicu-
                           lar to the rotating unit (see Fig. 20 on page E-15), otherwise the angle of the quick connec-
                           tor mechanism must be adjusted according to the instructions given in section “Adjusting
                           the sensor head position” below.
                           After adjusting the sensor head position adjust the primary slot according to the instruc-
                           tions given in sections “Primary slot adjustment - radiation beam too low, high or not ver-
                           tical” on page E-16 and “Primary slot adjustment - radiation beam too far to the right or
                           left” on page E-17, if needed.
Figure 19
                   Adjust the head support angle with the two screws (use 2.5mm allen key) located on the left side
                   of the mechanism (see Fig. 19 on page E-14 and Fig. 20 below).
90°
Spirit level
Figure 20
Primary slot adjustment - radiation beam too low, high or not vertical
                                                               WARNING
                                                               DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE X-RAYS
                                                               ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION WITH
                                                               THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM IS
                                                               STILL MISALIGNED REPEAT THE OPERATION.
                                                               If the beams are too high or low loosen the four screws that hold the primary slot plate in position
                                                               and adjust as necessary. Tighten the screws and recheck the beam. Repeat the above proce-
                                                               dure until the beam is correctly positioned.
        Beam too low                                                                  Beam too high                   Securing screws
                                                               UP                                                                                                                                               UP
                                                                                                                                                                                                                                    no. 50972
                                                                    no. 50972
DOWN DOWN
Figure 21
          NOTE                                                 The four primary slots are machined in one plate. Any adjustments made to the primary
                                                               slot plate or slot mechanism will automatically move all the slots. If any adjustments are
                                                               necessary you will probably find that adjusting the first slot is all that is needed to cor-
                                                               rectly align all the slots.
                                                               If the beam is not vertical loosen the two screws that hold the primary slot mechanism to the col-
                                                               limator tube and rotate the mechanism until the beam is correctly positioned. Tighten the screws
                                                               and recheck the beam. Repeat the above procedure until the beam is correctly positioned.
            Beam not vertical                                                                                                                        UP
                                                                                                                                                                                                                                                                             Securing screws
                                                                                                                                                                                           no. 50972
                                                                                                                                                                                           PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
                                                                                                             NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
DOWN
Figure 22
NOTE The ceph primary slots are also adjusted according to instructions given in this section.
                   Primary slot adjustment - radiation beam too far to the right or left
                                                                                                                                             °
                                                                                                                                             mm
                           CTL                                                                                                                                                                                                          CTL
                                  WARNING
                                  WARNING: DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE
                                  X-RAYS ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION
                                  WITH THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM
                                  IS STILL MISALIGNED REPEAT THE OPERATION.
                                  If any of the beams are too far to the left or right of the alignment rectangle they must be cen-
                                  tered.
                                                                                               UP                                                                                         UP
                                                                                                    no. 50972
                                                                                                                                                                                               no. 50972
                           Beam too far to the left                                                                                                                                                                                     Beam too far to the right
                                                                                                    PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
DOWN DOWN
Figure 23
                   NOTE           A microprocessor controlled motor determines the position of each slot. If the beam is
                                  readjusted to the left or right the new position must be reprogrammed into the memory.
                                  Press and hold down the primary slot key until the number of the slot that you are checking
                        1         appears on the primary slot display and starts to flash. You are now in the primary slot calibration
                                  mode.
                                  For example, if you have just checked the position of the beam from the slot the number 1 will
                                  appear on the display.
                                  Adjust the position of the beam by pressing the patient positioning (mm) keys. The minus key will
                                  move the slot mechanism, and beam to the left, and the plus key will move the slot mechanism,
                                  and beam to the right.
                                  Check the position of the beam again and, if necessary, repeat the above alignment procedure
                                  until the beam is within the alignment rectangle on the fluorescent screen.
                   CAUTION        Radiation is generated when the exposure button is pressed. Take adequate protection
                                  measures. Keep the exposure time as short as possible.
          NOTE       Always use the plus (+) key for the final adjustment. This makes sure that the slot mecha-
                     nism is correctly positioned by taking up the play between the draw nut and the position-
                     ing thread.
                     For example, if you are adjusting the beam to the left by pressing the minus (-) key you should
                     move the slot mechanism a little further to the left than necessary, and then press the plus (+)
                     key to move it back to the right to the correct position.
                                                          Final adjustment
                                                          in this direction
Position thread
Draw nut
                     Figure 24
                     If you cannot adjust the beam far enough to the left using the method described above you may
                     have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab slightly
                     and then tighten the screw. The slot position should then be rechecked.
Metal tab
Positioning thread
Draw nut
                     Figure 25
                     Press the horizontal segment selection key to store the slot position in the memory and move on
              2      to the next slot. The number of the next slot will appear on the display and start to flash.
                     Repeat the alignment procedure described above for this slot and any other slots.
                     Exit the primary slot calibration mode and check the position of all the beams again. Realign if
   CTL               necessary.
                     Exit the service mode.
                              Make sure that the unit is in the panoramic mode and that the primary slot number 0 is selected
                              (the horizontal sector display will remain empty).
                              The calibration must be performed with 2x2 (high resolution), 3x3 (enhanced resolution) and 4x4
      10     mA     60   kV
                              (normal resolution) binning. With high resolution select a kilovolt value of 60 and a milliampere
                              value of 10. With enhanced resolution select 60 kV and 7 mA, and with normal resolution select
                              60 kV and 5 mA. Note that these values must be selected to produce a clear exposure.
                              Remove the inner cover from tube head assembly, if it is in position. Four screws hold the tube
                              head cover in position, these are located at the rear of the tube head assembly.
                                                                                        Tube head
                                                                                        assembly
                              Figure 26
                              Place the calibration tool to the front of the primary slot.
Primary slot
Calibration tool
Figure 27
                      Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon.
                      The Dimax2Tool window appears.
Figure 28
                      Figure 29
                      The calibration must be performed with 2x2 (high resolution), 3x3 (enhanced resolution) and 4x4
                      (normal resolution) binning. Select the binning from the Setting / Binning pulldown menu.
                      Figure 30
                      Click the Dimax3 panoramic exposure button. A dark current exposure (i.e. an exposure without
                      radiation) must be taken first and a message window shown below appears.
   Dimax3 panoramic
   exposure
Figure 31
Press all the five sector keys to SWITCH OFF the radiation. The indicator lights will COME ON.
B D
A P E
                           Figure 32
                           The pmpcal2 window with text Waiting appears.
                           Figure 33
                           Press the ready key to drive the x-ray to the exposure position.
                   READY
                   PRET
                           Figure 34
                           Take an exposure. After the exposure the message window shown below appears.
                           Figure 35
                           Press all the five sector keys to SWITCH ON the radiation. The indicator lights will GO OUT.
B D
A P E
Figure 36
Figure 37
                      Press the ready key to drive the x-ray to the exposure position.
              READY
              PRET
                      Figure 38
                      Take an exposure.
                      The sensor head is now calibrated. After the exposure has been taken the image is shown in the
                      Dimax2Tool window.
Figure 39
                              Figure 40
                              The pmpcal2 window with text Waiting appears.
                              Figure 41
                              Touch the ready field to drive the x-ray to the exposure position.
                              The text Exposure appears.
                              Figure 42
                              Take an exposure. The calibration block image should be evenly grey. If it is not, recalibrate the
                              sensor head. If recalibration does not help, check that the Dimaxis, DIDAPI and drivers you are
                              using are all from the same CD rom. Check also the x-ray beam and patient positioning mecha-
                              nism adjustments by taking a ball phantom exposure.
Zoom in
Zoom out
                                                                         Hand
                                                                          Marking tool
                     Figure 43
                     The message window shown below appears. Confirm the selection by clicking Yes, or cancel the
                     selection by clicking No. While marking the row bad the sensor is simultaneously recalibrated.
                     Figure 44
                     The panoramic sensor consists of four sensor chips, and normally the stripes are located in the
                     chip boundaries. You can find the chip boundaries by checking the row numbers: the row num-
                     bering is not consecutive. The amount of sensor chip rows that can be marked bad is limited to
                     six. In case you have already marked six rows, more rows cannot be marked and the following
                     message window appears.
                     Figure 45
                     The row can be returned back to normal by clicking it with the right mouse button.
                   NOTE      If the temple supports are in position, remove them before taking the ball phantom expo-
                             sure.
Figure 46
Figure 47
Press the ready key to drive the x-ray to the exposure position.
Figure 48
                     Take an exposure. The images of 23 balls should appear on the image. They should be round,
                     all the same size, and evenly spaced.
                     Also the Results of the ball phantom test window appears. In this window the distances from the
                     center of the middle ball to the center of the tenth ball on the right and left side, and the distance
                     from the center of the middle ball to the center of the rear middle ball are given.
                     Figure 49
                     The outermost balls on the left and right sides should appear symmetrically. If the distance from
                     the center of the middle ball to the center of the tenth ball on the right side differs more than 2mm
                     from the corresponding distance on the left side, the rotation start limit must be adjusted (in the
                     same way as in the film-based panoramic x-ray, see section “Rotation limits” on page D-23).
                     Note, that the adjustment of the patient positioning mechanism must be performed first.
                     The distance from the center of the middle ball to the center of the rear middle ball should be less
                     than 2mm. If the distance is not correct you will have to adjust the position of the patient position-
                     ing mechanism (see section “Patient positioning mechanism adjustment” on page E-31).
                                 You should take an exposure of the ball phantom to check that the patient positioning mecha-
                                 nism is correctly positioned.
                                                                             D
                                                                           EN
                                                                                           1
                                                                                         97
                                                                                       50
                                                                                    No.
                                                                                                       A
                                                                                                     EC
                                                                                                   NM
                                                                                                PLA
                                                                        INE
                                                                                                           T
                                                                                                         AR
                                                                     Y-L             INE               ST
                                                                                  X-L
Ball phantom
                                 Figure 50
                                 Remove the sensor alignment tool from the adapter mechanism and attach the sensor head to its
                                 position (for fixed sensor head refer to section “Removing and attaching the fixed sensor head”
                                 on page E-4).
                                 Slide the ball phantom into the patient positioning mechanism.
                           °
                00         mm    Adjust the position of the patient positioning mechanism to 00.
                                 Drive the x-ray unit to the exposure position. The rotating unit will turn to the start position and the
                        READY
                        PRET     indicator light will come on.
                    NOTE         The ready indicator light will be flashing if the PC is not in the ready state. Refer to the
                                 Dimaxis user’s manual on how to prepare the system ready for the exposure.
                                 Take an exposure using the normal resolution (refer to the Dimaxis user’s manual). Hold the
                                 exposure down for the duration of the exposure.
                     The images of 23 balls should appear on the image (Fig. 51). They should be round, all the same
                     size, and evenly spaced. Note that they will not necessarily be at the same vertical height. The
                     outermost balls on the left and right sides should appear symmetrically. If the distance from the
                     center of the middle ball to the center of the tenth ball on the right side differs more than 2mm
                     from the corresponding distance on the left side,, the rotation start limit must be adjusted (in the
                     same way as in the film-based panoramic x-ray, see section “Rotation limits” on page D-23).
                     Note, that the adjustment of the patient positioning mechanism must be performed first. The dis-
                     tance from the center of the middle ball to the center of the rear middle ball should be less than
                     2mm. If the distance is not correct you will have to adjust the position of the patient positioning
                     mechanism (see section “Patient positioning mechanism adjustment” on page E-31).
R L
Figure 51
          NOTE       Before you can measure the true distance between the balls, you have to define a refer-
                     ence distance as follows:
                     Click the CAL button and then draw a long horizontal line to the image by clicking the
                     start point, releasing the mouse button and clicking the end point. The Reference distance
                     window appears.
                     Calculate the line length using the formula A x 0.132mm = B, where A is the distance in
                     pixels and B is distance in millimetres. Enter the value B (line length in millimetres) to the
                     Reference distance window and click OK.
                     The distance from the center of the middle ball to the center of the tenth ball to the left and
                     the tenth ball to the right can now be measured.
                     The distance from the center of the middle ball to the center of the tenth ball to the left and the
                     tenth ball to the right should be 120mm ± 2mm. If this distance is correct then the patient posi-
                     tioning mechanism need not be adjusted, and you should then check the patient positioning
                     lights (see section “PATIENT POSITIONING LIGHTS” on page D-19).
                     If these distances are not correct you will have to adjust the position of the patient positioning
                     mechanism. The following diagrams are a rough guide as to how to position it correctly. Note that
                     you may have to make more than one adjustment as the patient positioning mechanism could be
                     incorrectly positioned in more than one direction. For example it may be too far forward and too
                     near the column.
                               If the distance from the middle ball to the center of the tenth ball on the left-hand side is more
                               than 122mm and on the right-hand side less than 118mm then the patient positioning mechanism
                               must be moved away from the column (Fig. 52). Note, that the distance between the line on the
                               underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
                               section “Adjustment in ball phantom’s y-direction” on page E-31.
                                                    D
                                                         N
                                                             E
                                                                                              71
                                                                                                   09
                                                                                                       o.5
                                                                                                             N
                                                                                                                                A
                                                                                                                                C
                                                                                                                                    E
                                                                                                                                    M
                                                                                                                                        N
                                                                                                                                        LA
                                                                                                                                            P
                                                             E
                                                                                                                                            T
                                                        IN
                                                                                                                                             R
                                                                                                                                                A
                                                  Y-L
                                                                                                                                                 T
                                                                                              IN
                                                                                                                                                    S
                                                                                               -L
                                                                                                   X
           Right-hand side
                               Figure 52
                               If the distance from the middle ball to the center of the tenth ball on the left-hand side is less than
                               118mm and on the right-hand side more than 122mm then the patient positioning mechanism
                               must be moved towards the column (Fig. 53). Note, that the distance between the line on the
                               underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
                               section “Adjustment in ball phantom’s y-direction” on page E-31.
                                                                         D
                                                                       EN
                                                                                                                      1
                                                                                                                    97
                                                                                                                  50
                                                                                                                          No.
                                                                                                                                                    A
                                                                                                                                                  EC
                                                                                                                                                NM
                                                                                                                                             PLA
                                                                            E                                                                              RT
                                                                       IN
                                                                 Y-L                                             IN
                                                                                                                   E                                    STA
                                                                                                              X-L
Right-hand side
Figure 53
                           If the distances from the middle ball to the center of the tenth ball on both the right and left are
                           more than 122mm then the patient positioning mechanism is too far backward and must be
                           moved forwards (Fig. 54).
                                                               D
                                                                    N
                                                                        E
                                                                                                         71
                                                                                                              09
                                                                                                                  o.5
                                                                                                                    N
                                                                                                                                    A
                                                                                                                                    C
                                                                                                                                        E
                                                                                                                                        M
                                                                                                                                            N
                                                                                                                                            LA
                                                                                                                                                P
                                                                        E
                                                                                                                                                T
                                                                   IN
                                                                                                                                                 R
                                                                                                                                                    A
                                                             Y-L
                                                                                                                                                     T
                                                                                                         IN
                                                                                                                                                        S
                                                                                                          -L
                                                                                                              X
         Right-hand side
                           Figure 54
                           If the distances from the middle ball to the center of the tenth ball on both the right and left are
                           less than 118mm then the patient positioning mechanism is too far forward and must be moved
                           backwards (Fig. 55).
                                                D
                                                     N
                                                         E
                                                                                          71
                                                                                               09
                                                                                                   o.5
                                                                                                         N
                                                                                                                        A
                                                                                                                        C
                                                                                                                            E
                                                                                                                            M
                                                                                                                                N
                                                                                                                                LA
                                                                                                                                    P
                                                         E
                                                                                                                                    T
                                                    IN
                                                                                                                                        R
                                                                                                                                        A
                                              Y-L
                                                                                      E
                                                                                                                                            T
                                                                                          IN
                                                                                                                                            S
                                                                                           -L
                                                                                               X
Right-hand side
Figure 55
                       READY
                                         Press the ready key to drive the x-ray unit to the exposure position. When the x-ray is ready the
                       PRET
                                         indicator light will come on.
                                         Start to take a test exposure by pressing and holding down exposure button on the remote con-
                                         trol. Stop pressing the exposure button about halfway through the exposure cycle so that the
                                         rotating unit stops moving.
                                         An error code will appear on the display. Clear the error code from the display by pressing the
               Er 00                     CTL-key.
                                         Manually position the rotating unit so that the center line that runs through the tube head and
                                         sensor head is parallel to the wall to which the x-ray is secured. Slide the alignment pin through
                                         the hole, nearest the column, in the top of the vertical carriage so that it goes into the positioning
                                         hole in the rotating unit. (Fig. 56)
Alignment pin
Figure 56
                   NOTE                  Adjust the position of the rotating unit by holding the sensor head end of the rotating unit.
                                         Never hold the tube head when adjusting the position.
                     Leave the ball phantom tool in position and place alignment tool adapter to the sensor head.
                     Place one end of the alignment ruler in the first primary slot (the largest one) and the other end to
                     the sensor alignment tool.
                                                                                                 D
                                                                                                N
                                                                                               E
                                                                                                              1
                                                                                                            97
                                                                                                          50
                                                                                                        o.
                                                                                                       N
                                                                                                                          A
                                                                                                                         C
                                                                                                                        E
                                                                                                                       M
                                                                                                                      N
                                                                                                                    LA
                                                                                                                   P         T
                                                                                                E
                                                                                                                            R
                                                                                              IN
                                                                                                                           A
                                                                                                          E
                                                                                            -L
                                                                                                                          T
                                                                                                        IN
                                                                                                                         S
                                                                                           Y
                                                                                                      -L
                                                                                                     X
                                                                                                                                 Alignment ruler
                      Dimax2_adj1.eps
                     Figure 57
                     Loosen the four screws that hold the patient position mechanism in place and adjust its position
                     to the left or right until the mechanism is centered.
Screws
                     Figure 58
                     Make sure that the y-line on the ball phantom is parallel to the line on the underside of the align-
                     ment ruler. Note that the line on the ball phantom and the line on the alignment ruler do not have
                     to coincide, but they must be parallel, the deviation may be ±1mm.
                                                                                                                  Alignment ruler
                                                PLANMECA                      No.50971
                                                      X-LINE
                                                                     ENIL-Y
                                                                                                                   Ball phantom
                                                      Y-lineca.EPS
                     Figure 59
                     Tighten the screws, and remove the alignment pin.
                     Take another ball phantom picture to check the alignment again. Readjust if necessary.
                       READY
                                                   Press the ready key to drive the x-ray unit to the exposure position. When the x-ray is ready the
                       PRET
                                                   indicator light will come on.
                                                   Start to take a test exposure by pressing and holding down exposure button on the remote con-
                                                   trol. Stop pressing the exposure button after about two seconds so that the rotating unit stops
                                                   moving.
                                                   An error code will appear on the display. Clear the error code with the CTL-key.
              Er 00
                                                   Manually position the rotating unit so that the center line that runs through the tube head and
                                                   sensor head is at right angles to the wall. Slide the alignment pin through the hole in the top of
                                                   the vertical carriage so that it goes into the positioning hole in the rotating unit. (Fig. 60)
Alignment pin
Figure 60
                   NOTE                            Adjust the position of the rotating unit by holding the sensor head end of the rotating unit.
                                                   Never hold the tube head when adjusting the position.
                          Leave the ball phantom tool in position and place alignment tool adapter to the sensor head.
                          Place the alignment ruler in position between the primary slot and the sensor alignment tool as
                          described earlier.
                                                                                                              D
                                                                                                                  N
                                                                                                                  E
                                                                                                                          71
                                                                                                                           09
                                                                                                                               o.5
                                                                                                                                 N
                                                                                                                                     A
                                                                                                                                     C
                                                                                                                                         E
                                                                                                                                         M
                                                                                                                                             N
                                                                                                                                             LA
                                                                                                                                                 P
                                                                                                                                                 T
                                                                                                              E
                                                                                                                                                  R
                                                                                                             IN
                                                                                                                                                     A
                                                                                                                      E
                                                                                                         -L
                                                                                                                                                      T
                                                                                                                      IN
                                                                                                                                                         S
                                                                                                         Y
                                                                                                                          -L
                                                                                                                           X
                           Dimax2_adj2.eps
                          Figure 61
                          The line on the alignment ruler must line up with the x-line on the ball phantom (Fig. 62).
                    Sensor head side                           PLANMECA                       No.50971                                                       Tube head side
                                                       START                                                          END
                                             Ball phantom
                                                                      X-LINECA.EPS
                          Figure 62                0° position
                          Enter the service mode.
                    °     Press either of the patient positioning keys once to switch the patient positioning lights on then a
        12          mm
                          second time to adjust the position of the patient positioning mechanism until the line on the ball
                          phantom lines up with the line on the alignment ruler. The position of the mechanism will appear
                          on the display.
                    °     Hold the patient positioning plus (+) key down (about five seconds) until you hear a signal and
         00         mm
                          the millimeter display zeroes. This indicates that the new position of the patient positioning mech-
                          anism has been programmed into the memory.
                          Remove the alignment pin and exit the service mode.
                          Take another ball phantom picture to check the alignment again. Readjust if necessary.
                       READY
                                     Press the ready key to drive the x-ray unit to the exposure position. When the unit is ready the
                       PRET
                                     indicator light will come on.
                                     Start to take a test exposure by pressing and holding down exposure button on the remote con-
                                     trol. Stop pressing the exposure button after about 16 seconds so that the rotating unit stops
                                     moving (near the end position).
                                     An error code will appear on the display. Clear the error code with the CTL-key.
              Er 00
                                     Manually position the rotating unit so that the center line that runs through the tube head and
                                     sensor head is at a right angle to the wall. Slide the alignment pin through the hole in the top of
                                     the vertical carriage so that it goes into the positioning hole in the rotating unit.
                   NOTE              Adjust the position of the rotating unit by holding the sensor head end of the rotating unit.
                                     Never hold the tube head when adjusting the position.
                                     Leave the ball phantom tool in position. Place the alignment ruler in position between the primary
                                     slot and the sensor alignment tool as described earlier.
                                                                                        D
                                                                                         N
                                                                                             E
                                                                                                     71
                                                                                                      09
                                                                                                          o.5
                                                                                                            N
                                                                                                                A
                                                                                                                C
                                                                                                                    E
                                                                                                                    M
                                                                                                                        N
                                                                                                                        LA
                                                                                                                            P
                                                                                             E
                                                                                                                            T
                                                                                                                             R
                                                                                       LIN
                                                                                                                                A
                                                                                                 E
                                                                                  Y-
                                                                                                 IN
                                                                                                                                    S
                                                                                                     -L
                                                                                                      X
                               Dimax2_adj3.eps
Figure 63
                                               Measure the distance between the line on the alignment ruler and the x-line on the ball phantom
                                               (Fig. 64). The distance must be less than 1mm.
                                                                                                      ENIL-Y
                                                               Ball phantom
                                                                                       X-LINECA.EPS
                                               Figure 64         180° position
                                               If the distance between the line on the alignment ruler and the x-line on the ball phantom is more
                                               than 1mm, the sensor head position must be adjusted. Adjust the sensor head towards the ball
                                               phantom’s x-line until the distance is less than 1mm according to the instructions given in section
                                               “Adjusting the position of the sensor head” on page E-36. Correct only half of the deviation, so
                                               that the deviation will be the same in both the 180° and 0° positions.
              NOTE                             This section applies to panoramic units without cephalostat and pan/ceph units with one
                                               sensor head (i.e. sensor head with quick connector mechanism). In case your x-ray unit is
                                               a pan/ceph model with two fixed sensor heads, the position of the entire panoramic sen-
                                               sor head can not be adjusted. Refer to section “Adjusting the position of the optical sen-
                                               sor of the sensor head” on page E-37 for instructions on adjusting the optical sensor
                                               position.
                                               The position of the sensor head can be slightly adjusted. Loosen the two screws that hold the
                                               sensor head in rotating unit and move the sensor head to required position.
                        D2_adjsenshead_1.eps
Figure 65
4.4 Adjusting the position of the optical sensor of the sensor head
NOTE This section applies to pan/ceph units with two fixed sensor heads.
                               The panoramic position of the sensor head can be slightly adjusted by adjusting the optical sen-
                               sor position. Loosen the two sensor assembly attachment screws with the 2.5mm allen key and
                               move the assembly to desired position.
Figure 66
           6       CEPHALOSTAT ADJUSTMENT
                   NOTE       Switch the unit off after all the checks and adjustments have been performed, before tak-
                              ing any patient exposures.
                              Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
       4               CEPH
                              ing unit will move automatically to the correct position for taking cephalometric exposures. The
                              latest used image area will be automatically selected.
                              Remove the inner cover from tube head assembly. Four screws hold the tube head cover in posi-
                              tion, these are located at the rear of the tube head assembly.
                                                                                 Tube head
                                                                                 assembly
                              Figure 67
                              Rotate the cephalostat head support to the lateral ceph position. Remove the ear posts.
                     Figure 68
                     Unscrew the two slot attachment screws with the 4mm allen key and detach the slot plate.
Figure 69
                   Cephalostat sensor head with quick connector mechanism - removing the covers
                             Unscrew the bottom cover plate attachment screw with the 4mm allen key and remove the cover
                             plate.
                             Figure 70
                             Turn the lock disc from above the locking plate and push in the quick connector mechanism lock-
                             ing knob. Unscrew the attachment screw of the quick connector mechanism cover with the 4mm
                             allen key.
Cover
Locking plate
Figure 71
Figure 72
Figure 73
Unscrew the attachment screw of the adapter cover and slide the cover from its position.
                         Figure 74
                         Unscrew the upper cover attachment screw with the 4mm allen key and remove the cover.
Upper cover
Figure 75
Lock disk
                      Figure 76
                      Loosen the lock disc attachment screw with the 3mm allen key and remove the lock disc.
Figure 77
                   Loosen the button axle attachment screw with the 3mm allen key and push the button axle
                   inwards.
Button axle
                   Figure 78
                   Remove the sensor head from its adapter.
                   Figure 79
                   The sensor alignment tool can now be attached to the adapter pins.
                      The sensor head must be perpendicular to the head support. Use a sprit level to check the sen-
                      sor head position, i.e. the sensor alignment tool position (see Fig. 81 on page E-47). In case it is
                      not perpendicular to the head support, the angle of the quick connector mechanism must be
                      adjusted.
                      In case the sensor head is attached to the cephalostat, remove the sensor head from the quick
                      connector mechanism (see instructions given in Planmeca Proline CC panoramic x-rays user’s
                      manual). Remove the covers according to the instructions given in section “Cephalostat sensor
                      head with quick connector mechanism - removing the covers” on page E-41.
                      In case the cephalostat is equipped with fixed sensor head, remove the covers and the sensor
                      head according to the instructions given in section “Removing and attaching the fixed sensor
                      head” on page E-42.
                      Attach the sensor alignment tool to the quick connector. Loosen the two screws located on the
                      outer side of the mechanism with the 3mm allen key.
Figure 80
                   Adjust the head support angle with the two screws located on the left side of the mechanism (use
                   2.5mm allen key) (see Fig. 81) so that the sensor alignment tool is perpendicular to the cephalo-
                   stat head support.
90°
Spirit level
Figure 81
                     Enter the service mode (see section “How to enter/exit the service mode” on page B-6). The radi-
                     ation beam can now be checked without the sensor head moving.
                     Press the CTL-key.
               CTL
                     Press and hold down the primary slot key until the number on the primary slot display starts to
               0     flash. You are now in the primary slot calibration mode.
              C      Press the collimator selection key twice to skip passed the two panoramic collimator slots. The
                     letter “C” will appear on the display. This selection is used to check the position of the left side of
                     the radiation.
                     Place the beam alignment tool to the sensor alignment tool to horizontal position (see figure
                     below).
Figure 82
                     Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in a
                     darkened room for example 80kV and 12mA.
                     Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                     sure button. The image of the radiation beam will appear on the beam alignment tool.
                       The radiation beam must appear on the beam alignment tool as shown on the figure below. It
                       must reach, but not overlap, the middle line of the tool. If it does not, the cephalostat arm posi-
                       tion must be adjusted. Note that the beam does not have to reach the right side line. Refer to the
                       section “Adjusting the cephalostat arm position” on page E-51.
                                                                          D2alinmhoriz.eps
                           Middle line
                                                                                             Right side line
Figure 83
                   d   Press the primary selection key once. The letter “d” will appear on the display.
                       Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                       sure button. The image of the radiation beam will appear on the beam alignment tool.
                       The radiation beam must appear symmetrically on the beam alignment tool as shown on the fig-
                       ure below. If it does not refer to the section “Radiation beam adjustment” on page E-14. The
                       adjustment is performed in the same way as when adjusting the standard panoramic slot.
                                                    A         A
                                                                  D2adj_2prim.eps
Figure 84
                     To check the beam position in vertical direction place the beam alignment tool to the sensor
                     alignment tool to vertical position as shown on the Fig. 85 below.
                     Figure 85
                     Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                     sure button. The image of the radiation beam will appear on the beam alignment tool.
                     The radiation beam must reach the lower edge of the rectangle marked on the beam alignment
                     tool. If it does not refer to “Radiation beam adjustment” on page E-14. The adjustment is per-
                     formed in the same way as when adjusting the standard panoramic slot.
                                                                         D2adj_1prim.eps
                     Beam alignment
                     tool in upper
                     position
                                                                                           Beam alignment
                                                                                           tool in lower
                                                                                           position
                     Figure 86
                     Move the beam alignment tool to the upper position on the sensor alignment tool as shown on
                     the Fig. 86.
                     The radiation beam must reach the upper edge of the rectangle marked on the beam alignment
                     tool (see figure above). If it does not refer to “Radiation beam adjustment” on page E-14. The
                     adjustment is performed in the same way as when adjusting the standard panoramic slot.
                     When all the slots have been checked, and if necessary adjusted, exit the service mode by
                     pressing the ceph key.
                              The left side of the radiation beam is checked as described in section “Checking the cephalomet-
                              ric beam alignment” on page E-48. If the left side of the radiation beam is too far left or right the
                              cephalostat arm position must be adjusted.
                              Adjust the arm position by turning the two cephalostat arm adjusting screws. Turning the other
                              screw clockwise with 4mm allen key will move the beam to the right, and turning the other screw
                              clockwise with the 6mm allen key to the left.
                              Figure 87
                              If the cephalostat arm position adjustment does not enough correct the beam alignment, the
                              position of the latch of the rotating unit locking solenoid must be adjusted. The adjustment is pos-
                              sible only when the rotating unit is in the end position.
                              Enter the test mode (see section “Test mode” on page B-4). Press the exposure button until the
                              rotating unit has reached the end position. When the rotating unit is in this position you will be
                              able to see a small recess in one side of the guide block which is located under the upper shelf.
                              Switch the unit off. Loosen the latch attachment screws with the 3mm allen key and move the
                              latch to required direction.
Figure 88
                             Remove the sensor alignment tool from the cephalostat quick connector mechanism and attach
                             the sensor head to the quick connector.
                             Remove the nasial positioner and the ear posts from their holders. Attach the cephalostat calibra-
                             tion tool to the right-hand ear post holder (near the sensor head).
Nasial positioner
                                                                                           Cephalostat
                                                                                           calibration tool
                             Figure 89
                             Make sure that the second primary slot is in its position.
                   CEPH
                             Press the ceph mode selection key to enter the ceph mode if you are not already in it.
6 Select the image area number 6 with the primary slot selection key.
                             The calibration must be performed with 3x3 (enhanced resolution) and 4x4 (normal resolution)
    7      mA     60    kV
                             binning. With enhanced resolution select 60 kV and 7 mA, and with normal resolution select 60
                             kV and 5 mA. Note that these values must be selected to produce a clear exposure.
                               Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon
                               located in the Dimaxis program folder. The DimaxTool window appears.
Figure 90
                               Figure 91
                               The calibration must be performed with both 3x3 (enhanced resolution) and 4x4 (normal resolu-
                               tion) binning. Select the binning from the Settings / Binning pulldown menu.
                               Figure 92
                               Click the Dimax3 cephalometric exposure button. The pmpcal2 window with text Waiting
                               appears.
       Dimax3 cephalometric
       exposure
                               Figure 93
                               Press the ready key to drive the x-ray to the exposure position.
                       READY
                       PRET
                     Figure 94
                     Take an exposure. The sensor head is now calibrated. After the exposure has been taken the
                     image is shown in the Dimax2Tool window. The triangle on the image must be even, without any
                     irregularities.
Figure 95
                              In case the triangle is not evenly shaped, it can be slightly edited. Enlarge the image if needed.
                              The sensor joint is marked with the red lines when the keyboard’s L-key is pressed. Move the
                              lower part of the triangle with the keyboard’s arrow keys as follows: the right edge of the triangle
                              (vertical) can be adjusted with the right/left arrow keys and the left edge with the up/down keys.
Red lines
                              Figure 96
                              While editing the shape of the triangle the sensor is simultaneously recalibrated.
                              If the evenly shaped triangle can not be reached, the second primary slot position must be
                              adjusted in vertical direction. Refer to the section “Second primary slot adjustment” on page E-64
                              for information on how to align it.
Figure 97
                     Figure 98
                     Press the ready key to drive the x-ray to the exposure position.
                     The text Exposure appears.
                     Figure 99
                     The calibration tool image should be evenly grey. If it is not, recalibrate the sensor head. If recal-
                     ibration does not help, check that the Dimaxis, DIDAPI and drivers you are using are all from the
                     same CD rom. Check also the cephalostat adjustments by taking an exposure from the ear posts
                     (see section “Checking the head support position” on page E-59).
Zoom in
Zoom out
Hand
Marking tool
                     Figure 100
                     The message window shown below appears. Confirm the selection by clicking Yes, or cancel the
                     selection by clicking No. While marking the row bad the sensor is simultaneously recalibrated.
Figure 101
                              The cephalostat sensor consists of eight sensor chips, and normally the stripes are located in the
                              chip boundaries. You can find the chip boundaries by checking the row numbers: the row num-
                              bering is not consecutive. The amount of sensor chip rows that can be marked bad is limited to
                              six. In case you have already marked six rows, more rows cannot be marked and the following
                              message window appears.
                              Figure 102
                              The row can be returned back to normal by clicking it with the right mouse button.
Figure 103
Figure 104
Touch the ready field to drive the x-ray to the exposure position.
                     Figure 105
                     Take an exposure. On the image you will see the images of a metal ball and a metal ring. The
                     must be concentric, or if not exactly concentric there must be a small gap between the edge of
                     the ball and the inner edge of the circle. If the ball and ring are correctly positioned, the test
                     image is accepted.
                     If the ball touches the ring the ear posts must be adjusted, refer to section “Head support adjust-
                     ment” on page E-60.
                              Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
       6               CEPH
                              ing unit will move automatically to the correct position for taking cephalometric exposures. Image
                              area 4 will be automatically selected. Press the collimator selection key twice to select the image
                              area 6.
                              Remove the sensor alignment tool from the cephalostat quick connector mechanism, if needed.
                              Attach the sensor head to the quick connector.
                              Rotate the head support to the 0° position and place the two ear posts into their holders if they
                              are not already installed.
                                 Dim
                                    219
                                       _2.e
                                           ps
Head support
                              Figure 106
                              Embedded in the end of the left (V or L) ear post there is a metal ball and embedded in the end of
                              the right ear post (O or R) is a metal ring. The images of these that appear on an exposed image
                              are used to check the alignment of the ear posts.
Stand behind the tube head, protect yourself from radiation and take an exposure.
                              On the image you will see the images of a metal ball and a metal ring. The must be concentric, or
                              if not exactly concentric there must be a small gap between the edge of the ball and the inner
                              edge of the circle.
Ball
           Ring
                       Correct                           Correct                             Wrong
                              If the ball touches the ring the ear posts must be adjusted. Refer to section “Head support adjust-
                              ment” on page E-60.
                         Switch the unit off. Remove the cover of cephalostat head support.
                         Make sure that the locking screw of the adjustment flange is loosened (Fig. 108, 1).
                         Adjust the head support position according to the Fig. 108. Rotate the head support around its
                         vertical axis to adjust the ear posts in horizontal direction (Fig. 108, 2) and around its horizontal
                         axis to adjust the ear posts in vertical direction (Fig. 108, 3).
1 2
Figure 107
                          Adjust the cephalostat head support position to horizontal position, if needed (see Fig. 107, 4 and
                          Fig. 108 below)
                          Figure 108
                          Fasten the loosened screws and replace the covers.
                   NOTE   CHECK THE RADIATION BEAM ALIGNMENT ACCORDING TO THE INSTRUCTIONS GIVEN
                          IN SECTION “Checking the cephalometric beam alignment” on page E-48 AFTER
                          ADJUSTING THE HEAD SUPPORT POSITION. ADJUST THE BEAM, IF NEEDED. REPEAT
                          THE BEAM AND HEAD SUPPORT ADJUSTMENTS UNTIL THE RADIATION BEAM POSI-
                          TION AND HEAD SUPPORT POSITION ARE ACCEPTABLE.
           NOTE       You must now check the beam alignment with the second primary slot in its position.
                      Attach the second primary slot to its position.
                      Remove the sensor head and attach the sensor alignment tool to the cephalostat head support.
                      Enter the service mode (see section “How to enter/exit the service mode” on page B-6). The radi-
                      ation beam can now be checked without the sensor head moving.
                      Press the CTL-key.
                      Press and hold down the primary slot key until the number on the primary slot display starts to
                  0   flash. You are now in the primary slot calibration mode.
d Press the collimator selection key until the letter “d” will appear on the display.
                      Place the beam alignment tool to the sensor alignment tool to vertical position as shown on the
                      figure below.
                      Figure 109
                      Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                      sure button. The image of the radiation beam will appear on the beam alignment tool.
                          The beam image should appear within the borders of the rectangle marked on the beam align-
                          ment tool. If it does not adjust the second primary slot position. Refer to section “Second primary
                          slot adjustment” on page E-64.
                                                                      D2filt_D.eps
                           Beam alignment
                           tool in upper
                           position
                                                                                         Beam alignment
                                                                                         tool in lower
                                                                                         position
                          Figure 110
                          Move the beam alignment tool to the upper position on the sensor alignment tool as shown on
                          the figure above.
                          Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
                          sure button. The image of the radiation beam will appear on the alignment tool.
                          The beam image should appear within the borders of the rectangle marked on the beam align-
                          ment tool as shown on the figure above. If it does not adjust the second primary slot position.
                          Refer to section “Second primary slot adjustment” on page E-64.
                   NOTE   After checking and adjusting the second primary slot position the cephalostat sensor
                          head must be calibrated according to the instructions given in section “Calibrating the
                          cephalostat sensor head” on page E-52.
                        Figure 111
                        Move the second primary slot in vertical direction until the slot is correctly positioned. Tighten the
                        loosened screws.
Figure 112
                               Rotate the second primary slot with the two screws located on the right side of the slot until the
                               slot is correctly positioned (use the 2.5mm allen key).
Figure 113
Figure 114
                   NOTE        After checking and adjusting the second primary slot position the cephalostat sensor
                               head must be calibrated according to the instructions given in section “Calibrating the
                               cephalostat sensor head” on page E-52.
           NOTE        The Image area shifting is possible from software versions PK 5.12 and PG 5.12. Upgrade
                       the software, if needed.
7.1 Adjusting the position of the frame cover fastening flat spring
                       Remove the frame cover as follows. Unscrew the two attachment screws using the 2.5mm Allen
                       key. Remove the attachment screws and the washers (figure below, 1). Pull the frame cover
                       carefully downwards as shown on the figure below (2). Slide the frame cover away from the
                       cephalostat head support (3).
Figure 115
                   The frame cover fastening flat spring must be on the same level as the trail holder (see figure
                   below, 1). If needed, loosen the attachment screws (2) and move the fastening flat spring.
                   Tighten the attachment screws.
                   Figure 116
                   Attach the frame cover.
NOTE The Image area sifting is possible from software versions PK 5.12 and PG 5.12.
                        In the lateral view (image area number 4 or 6) the image area on the patient’s nasal side can be
                        shifted backwards and forwards as shown on the figure below.
CS=1 CS=1
  Ear post
                                                       Image area
                                                       (grey area)
CS=2 CS=2
CTL + CEPH
Press the CTL-key and the PAN key until the text CS1 or CS2 appears on the display.
CTL PAN
                          The parameter value CS1 indicates the “old” image area (moves the image area backwards).
                          The parameter value CS2 moves the image area forwards approx. 14mm (0.55 in.) regarding to
                          the image area with value CS1. Change the parameter value with the up/down keys.
                            READY
                            PRET
CEPH
                          You have to adjust the cephalometric beam after changing the parameter value. Refer to the
                          panoramic unit’s technical manual.
                   NOTE   If you do not want to move the image area when upgrading the software from the revi-
                          sions smaller than 5.12, select the parameter value CS=1. In that case you do not have to
                          adjust the cephalometric beam.
F PREVENTIVE MAINTENANCE
1 SYSTEM MAINTENANCE
1.1 Cleaning
                   CAUTION    When disinfecting the unit surfaces, always disconnect the unit from the mains. The unit
                              must not be exposed to gaseous disinfectants or explosive anesthetics. Never spill any
                              liquids into the unit.
                              The chin support and the Ceph cassette holder can be cleaned with NON-ALCOHOLIC disinfec-
                              tion solutions.
                              The bite piece of the chin rest can be removed and cleaned with alcohol-based solutions. The
                              patient support handles can be cleaned with alcohol-based solutions.
                              Other unit surfaces can be cleaned with a soft cloth damped in a mild cleaning solution.
                   Exposure indicators
                              Confirm that the exposure indicator lights turn on in the control panel and in the remote control
                              for the length of the exposure.
                   Exposure switch
                              Confirm that the exposure switch requires continuous activation to maintain the exposure.
                              Releasing the exposure switch during the radiation should produce the ER00 message on the
                              displays (clear it by pressing the CTL-key once). Make a visual check that the exposure switch
                              spiral cable is not damaged.
                   Labels
                              Check that no labels are detached or worn and that they are all legible.
                      Annually or after every 10 000 exposures whichever appears first the electrical exposure param-
                      eters of the Planmeca Proline CC Panoramic X-ray should be checked in order to maintain the
                      initially accurate operation of the equipment despite of the possible long term component value
                      alterations.
                      There are no adjustments in the equipment because of its’ self-calibrating microprocessor based
                      operation and therefore no actual parameter trimming of the kV, mA and exposure time can be
                      done.If one of the parameters to be checked by this periodical preventive maintenance proce-
                      dure is found to be beyond the limit values allowed, the corresponding electrical circuit board is to
                      be changed in whole and sent back to the factory for recalibration that is done by replacing the
                      faulty component.
                      The parameters to be checked are the following:
                      Before exposure:
                      •   kV measurement circuit impedance
                          mA measurement circuit impedance
                      During exposure:
                      •   kV value
                      •   mA value
                      •   exposure time
                      In order to perform the calibration measurements the x-ray equipment must be powered off and
                      the rotating arm bottom cover is to be removed. The generator processor board connectors P2
                      and P13 are disconnected from their connectors and the PM 2002 CC Panoramic X-ray Calibra-
                      tion unit (order No. 50980) is connected in between these connectors. Please, be certain that the
                      calibration unit is connected to both the cable end connectors and to the generator processor
                      board connectors.
                      Before powering on the x-ray equipment the circuit impedance measurements should be per-
                      formed in order to verify the condition of the measurement resistors in the tube head and the
                      operational amplifier input resistors on the generator processor board.
                      Because the calibration unit is battery operated first put it in the BATT mode and check that the
                      battery voltage is above 8 V. Note that the battery check loads the battery and should therefore
                      be done shortly to avoid unnecessary battery discharge.
                   CAUTION   Put the lead radiation protection plate in front of the collimator to avoid radiation during
                             measurements.
                   kV - measurement
                             Select for kV value measurement and select x-ray equipment for kV values from 60 kV to 80 kV
                             (selected mA value has no effect but 2 mA is recommended to avoid unnecessary radiation),
                             press the exposure switch at each kV value and read the calibration unit display when the gener-
                             ator is on (warning sound of higher pitch can be heard), allow few seconds to the calibrator to
                             settle. The results must be the values selected ± 2.5 kV. If not, the generator processor board is
                             to be replaced.
                   CAUTION   DO NOT OPERATE THE RESISTOR/BATT BUTTON WHILE THE X-RAY IS OPERATING.
                             DOING THIS MAY CAUSE PERMANENT DAMAGE TO THE TUBE HEAD.
                   mA - measurement
                             Select for mA value measurement and select x-ray equipment mA values from 2 mA to 12 mA
                             (selected kV value has no effect but 60 kV is recommended to avoid unnecessary radiation),
                             press the exposure switch at each mA value and read the calibration unit display when the gen-
                             erator is on (warning sound of higher pitch can be heard), allow few seconds to the calibrator to
                             settle. The results must be the values selected ± 1.0 mA. If not, the generator processor board is
                             to be replaced.
                   CAUTION   DO NOT OPERATE THE RESISTOR/BATT BUTTON WHILE THE X-RAY IS OPERATING.
                             DOING THIS MAY CAUSE PERMANENT DAMAGE TO THE TUBE HEAD.
                   CAUTION   DO NOT OPERATE THE RESISTOR/BATT BUTTON WHILE THE X-RAY IS OPERATING.
                             DOING THIS MAY CAUSE PERMANENT DAMAGE TO THE TUBE HEAD.
                     When using various focal trough forms or when in cephalostatic mode the exposure time varies.
                     The selected time can be seen from the x-ray equipment exposure time display and the meas-
                     ured time must correspond to that within ± 10%. If not, the generator processor must be replaced
                     and most likely also the keyboard processor. That is because the processor clocks are synchro-
                     nized to each other and variation in exposure times can only be due to the processor clock fre-
                     quency alteration that must appear in both processors within 1%. If the time differs from the
                     selected one because of some other reason, an error indication will always be present.
     3    MECHANICAL CHECKS
                     Refer to the chapter “ADJUSTMENT & CALIBRATION” on page D-1.
G TROUBLESHOOTING
         2         ERROR MESSAGES
                                Please refer to sections “ERROR MESSAGES” on page C-2 and “ERROR MESSAGE EXPLA-
                                NATIONS” on page C-4 for details. These error messages make fault-finding easier. Also the list
                                of the last 50 error occurrences is of great help, refer to section “Error history display” on page B-
                                13 how to display the error list.
         3         FUSE DESCRIPTION
                   Main fuses
                                Location:      Two pieces located near the main switch.
                                Description:   6 x 32 mm glass tube fuse. Rated 8 A for 220/240 V; 15 A for 100/110 V
                                               SLOW BLOW.
                                Function:      To protect the whole equipment. All power is delivered via these fuses.
                                Reasons for blow:     - Mains wiring short circuit after fuses
                                                      - Filter defective
                                                      - Bad fuse or wrong rating
                                Indication:    No power or action in the unit.
                                Analysis:      Mains voltage found at wall outlet but not at the low voltage power supply
                                               board.
                            - Connect the wires back one by one and try if the fuse still disconnects. If it
                            disconnects, follow the last connected wire and disconnect its other end
                            from the processor board and try if the fuse still blows. If it still disconnects
                            the wiring is defective
                            - Disconnect all other connectors from the processor board, connect the
                            power connector and try if the fuse still disconnects. If it disconnects, the
                            processor board motor control circuitry is defective.
                            - Connect the processor board connectors back one by one and try if the
                            fuse disconnects. If it disconnects, follow the last connected wire and find
                            the short circuiting part.
4.1 Check the mains outlet voltage to be within limits. (±10% of nominal voltage)
                          a)   Check the main fuses near the mains switch below the lower fixed part. If the fuses were OK take
                               step 4, if not take step 3.
                          b)   Replace the fuses only with fuses of the same rating. With power off open the lid on top of the
                               upper fixed part and detach connector P3 from the low voltage power supply board. If fuses hold
                               take step 4, if not there is a short circuit in the mains wiring from mains switch to the filter in upper
                               fixed part, in the mains wiring from the filter to the loose connector P3 or the mains filter is
                               defective.
                          c)   With power off open the lid on top of the upper fixed part if it was not already open, connect
                               connector P3 if it was taken off. If the fuse F3 was blown, it could be a result from an occasional
                               overvoltage in the mains. To check this replace the fuse and power on. If everything operates
                               normally this was the case, if the fuse blows again detach connectors P1, P2, P4 and P5. Check
                               all the fuses on the low voltage power supply board and replace if necessary with fuses of the
                               same rating. If the fuses hold take step 5, if not replace the low voltage power supply board.
                               Check that the primary voltage setting is the same.
                          d)   Connect connector P2 while powered off. If the fuses hold when powered on take step 6. If not,
                               there is a short circuit to the protective earth in the mains wiring from this board to the tube power
                               generator board or on the tube power generator board itself. If the relay K2 on the low voltage
                               power supply board is inoperative and is connecting all the time the board must be replaced but
                               there is also a short circuit between the mains live wires from this board to the tube power
                               generator board or on the tube power generator board itself. Open the bottom cover of the
                               rotating part and detach the tube power generator board connector P1. If fault remains, replace
                               the cable harness. If not, replace the tube power generator board.
                          e)   With power on measure the voltage across GND and +VCC labeled connector pins on either
                               connector P1 or P4. If it is within 30 - 39 VDC, take step 7, otherwise replace the low voltage
                               power supply board. Check that the primary voltage setting is the same.
                f)   Power off and wait for 1 minute to allow the capacitors to discharge. Connect connector P1. If the
                     fuses hold when powered on take step 8.
                     If not, there is a short circuit in the cable harness from this board to the generator processor or in
                     the generator processor and the circuitry connected to it. Open the bottom cover of the rotating
                     part and detach the generator processor board connector P5. If a fuse blows, replace the cable
                     harness. If not, power off and connect P5 to the generator processor board and detach all the
                     other connectors from it. Check the fuse on the board and replace if necessary. If a fuse blows
                     when powered on, replace the generator processor board (see chapter Circuit boards/generator
                     processor). If not, power off and connect the connectors back to the generator processor board
                     one at a time checking every time for the fault to reappear. Check the cable that caused the fault
                     and all the circuitry connected to it. Replace the defective parts.(see appropriate chapters for
                     adjustments that should be made).
                g)   Power off and connect connector P4 to the low voltage power supply board. Open the lower fixed
                     part cover and detach connector P1 from the keyboard processor board. If the fuses hold when
                     powered on take step 9.
                     If not, there is a short circuit in the cable harness from the low voltage power supply board to the
                     keyboard processor board. Replace the cable harness.
                h)   Power off and connect P1 to the keyboard processor board and detach all the other connectors
                     from it. Check the fuse on the board and replace if necessary. If a fuse blows when powered on,
                     replace the keyboard processor board (see chapter Circuit boards/keyboard processor). If not,
                     power off and connect the connectors back to the keyboard processor board one at a time (leave
                     the connector P7 out) checking every time for the fault to reappear. Check the cable that caused
                     the fault and all the circuitry connected to it. Replace the defective parts.(see appropriate
                     chapters for adjustments that should be made). If the fuses didn’t blow take step 10.
                i)   Power off and connect P7 to the keyboard processor board. Open the keyboard housing and
                     detach connector P2 from the keyboard/display board. If the fuses hold when powered on take
                     step 11. If not, there is a short circuit in the cable harness from the keyboard processor board to
                     the keyboard/display board. Replace the cable harness.
                j)   Power off and connect P2 to the keyboard/display board and detach connector P1. Check the
                     fuse on the board and replace if necessary. If a fuse blows when powered on, replace the
                     keyboard/display board. If not, the system should now start normally and you can take the step
                     12. If the display is still blank, look at the keyboard processor board for a red LED. If it is off or
                     blinking, replace the board. If it is continuously illuminating, replace the keyboard/display board.
                     If neither did help, power off and detach connector P1 from the low voltage power supply board.
                     If the system now starts, the fault should be located on the generator processor board. Replace
                     (see chapter Circuit boards/generator processor).
                k)   Power off and connect connector P1 to the keyboard/display board. Turn the rotating part
                     manually to the starting point (on the sensor) if it was not there. Shift the cassette on the sensor
                     at the patient positioning end (backwards when looking from the keyboard). connect the low
                     voltage power supply connector P5. Power on. Push either +mm or -mm button to switch on the
                     positioning lights. If a fuse blows, check the lamps and their wirings. If not take step 13.
                l)   Using the keyboard drive one motor at a time. (HOLD, LAYER, CASSETTE, PRIMARY
                     APERTURE). If a fuse blows, check the last motor operated and its wirings. Replace if
                     necessary. The rotation motor cannot be driven separately and it must be checked last. Push
                     READY and wait till cassette stops. Keeping the power on move manually the rotating part off
                     from it’s start sensor, after which the system automatically tries to return the rotating arm onto the
                     sensor, driving the rotation motor only. Check. If everything is normal take step 14.
                m)   Enter the service mode. Select for 60 kV and 2 mA and try one exposure. If still OK, select for
                     maximum values and try one exposure. If still OK enter normal mode and try one normal
                     exposure with maximum values. If still OK the problem actually was a bad fuse that had blown for
                     no reason or a bad connection on one end of the connectors. Check. If a fuse was blown during
                     the exposures the tube power generator takes too much current for some reason and must be
                     replaced. In this case fuses should not blow but an error indication should appear. Check the
                     ratings of the fuses.
         5         FLAT BATTERY
                             The equipment has two different types of parameter memories. The most critical parameters that
                             are seldom changed are stored into an EEPROM-type memory that is non volatile but the
                             amount of changes made is restricted to about 10’000 times per single data.
                             The other parameters that are not fatal if lost and which are altered more often are stored into a
                             battery backup RAM-type memory on the keyboard processor board. This information is lost and
                             substituted with minimum or less harmful ones if the battery goes flat.
                             When the equipment is powered on the battery is charged. A completely flat battery will be fully
                             charged within about two hours. If the equipment is never switched on or just shortly after long
                             intervals as may be the case when stock keeping it, the battery will last at least two months.
                             When the equipment is turned on with a flat battery the clock display and mm display are blinking
                             all zeroes. In the following is explained, which information is lost and with which values it is sub-
                             stituted.
                             After detecting this kind of a situation the substituted information must be corrected by resetting
                             the above listed parameters to actual ones.
CAUTION There are two parameters that must be corrected in order to gain the normal operation!
                             If the equipment was a cephalostatic one it must be once again programmed to be cephalostatic
                             as it is explained in the chapter “Panoramic/cephalostatic setup” on page B-15.
                             The correct place of the layer positioning motor must be learned into the system by driving the
                             motor in service mode twice past its reference sensor till the mm display ceases blinking. Take
                             care not to drive the motor into the mechanical ends that may cause damage.
6 DISTURBANCES IN ROTATION
                          The position indication is correct if the display shows 00 when the light meets the bite piece.
                          Erroneous position indication may depend on the fact that the display has not been adjusted to
                          zero after installing the reference Hall sensor or keyboard processor PCB or that the accumulator
                          is empty.
12 PROBLEMS IN RADIOGRAPHS
                                 Most often troubles are caused by other reasons, not by the equipment. Check that the fault on
                                 the radiograph does not result from defective patient positioning or fault in the film.
12.2 No radiograph
Film has certain tolerance in the holder which causes considerable variations.
         1         COVERS
                   CAUTION    The unit contains hazardous voltages. Always disconnect the unit from mains by remov-
                              ing the plug from the wall outlet, and wait for 2 minutes before starting to remove covers.
                              Remove the front cover of the tube head by unscrewing the 4 fixing screws at the back side.
                              Remove the back cover of the tube head by unscrewing the 2 fixing screws that can be seen
                              after removing the front cover. (Fig. 1, 1)
Remove the secondary slot cover by unscrewing the 4 screws at the cassette head (Fig. 1, 2).
                                                                                                                                    lA    l
                                                                                                                                       .uq A mm
                                                                                                                                          em 5
                                                                                                                                             m ,2 N
                                                                                                                                               5,2 O
                                                                                                                                                  ELA ITAR
                                                                                                                                                     TO TLI
                                                                                                                                                        TN FL
                                                                                                                                                           OI ATO
                                                                                                                                                             TA
                                                                                                                                                                RT T
                                                                                                                                                                  LIF
                                                                        TN
                                                                         SREITA
                                                                            O P
                                                                            DETCAOT S      GN
                                                                               V FE U
                                                                           ER RESBRUSORE INR
                                                                             TE      O OP GN AW
                                                                             RU TU     ER XE A
                                                                                ET EP    A    E D
                                                                                  AR X S SNOFASEB Y
                                                                                     EP N      IT SS AM
                                                                                                                                                                        2
                                                                                 TNEDOO'LOYA CURELN TI
                                                                                    EM M TE R A      T U N
                                                                                       ES EL TN LI SNI ROU YA
                                                                                         UE TE EI ER       G T
                                                                                            LU ET TAP AP NITARER-X
                                                                                              PU IRU E PA      A PO SI
                                                                                                 RC CEL R TE REPO DH    N
                                                                                                                          T
                                                                                                   S S S E UO C N
                                                                                                      AP D P X O DNA
                                                                                                         TNSRU UE ITN A
                                                                                                            OS ET RE ET
                                                                                                               EN CA GN TA
                                                                                                                  IOF SEAD
                                                                                                                  SE LPM L
                                                                                                                    TC E IS
                                                                                                                       EP 'D
                                                                                                                         SE
                                                                                                                            R
Figure 1
2 MOTORS
2.1 Z-motor
                          a)   Lock the rotating unit and the Z-carriage with transportation pins
                          b)   Place a support on the floor where you can tilt the machine.
                          c)   Detach the unit from the wall and overturn it carefully on the support.
                          d)   Pull the top trim cover out.
                          e)   Remove the steel cables and connectors.
                          f)   Slack the stop screws on the drawing side of the motor axle.
                          g)   Slack the stop screw on the sliding side and pull the steel cable off.
                          h)   Cut the panduit and pull the motor away.
                          i)   Install a new motor.
                          a)   Remove the outer secondary slot and back cover of the cassette head.
                          b)   Remove the loading spring of the cassette motor.
                          c)   Remove the connector.
3 CIRCUIRT BOARDS
                         a)   Remove the lower shelf cover. If the unit has old version of covers the holder of the chin rest has
                              to be removed before removing the cover.
                         b)   Remove the old board and replace it with a new one.
                   NOTE        If the first slot is adjusted mechanically by loosening the screws and turning the tube
                               head so that the beam is aligned in the middle, normally the other slots need no or only a
                               small adjustment.
                               Make the ball phantom test picture. Adjust if necessary (see section “PATIENT POSITIONING
                               MECHANISM” on page D-10 for film-based panoramic x-ray and section “PATIENT POSITION-
                               ING MECHANISM” on page E-27).
                               Install the covers.
I FUNCTIONAL DESCRIPTION
                     The rotating mechanism is designed so that the x-ray beam passes the patient’s teeth and tem-
                     poro-mandibular joints in an optimal way throughout the excursion.
                     The geometry of the driving groove of the block (Fig. 3, 4) is as follows. The groove has the
                     shape of a curve of a circle symmetrically to both directions from the center the driving groove to
                     ± 25° turning angle (point A being the center of the circle). From ± 25° turning angle to ± 85° turn-
                     ing angle, the radius of the shape of the groove varies and reminds of an evolvent curve. There-
                     after the groove follows the tangent of the curve.
                                                  25 °
                                     °
                                    80
Figure 1
                   The rotating unit (Fig. 2, 7) is attached with bearings to the upper horizontal part by means of a
                   cam (Fig. 2, 1). For both axles of the cam there are two angular contact ball bearings (Fig. 2, 2)
                   which are preloaded without clearance by means of nuts (Fig. 2, 3).
                   The cam has two main tasks: Firstly, it transfers the weight of the rotating unit to the main block
                   (Fig. 2, 4) through bearings. Secondly, it determines the path of the bearing point B during the
                   rotating cycle. Apart from point B, the path of the rotating movement is also determined by drive /
                   guide point C. On the main block (Fig. 2, 4) there is a U-groove and a guide wheel (Fig. 2, 5) rolls
                   along the inner edge of the groove. The bearing point of the wheel is fixed in relation to the rotat-
                   ing unit (Fig. 2, 7). There is also a driving wheel (Fig. 2, 6) which is concentrically located in rela-
                   tion to the bearing point. The fixing point of the driving wheel (transmission) is located so that the
                   driving wheel can move unhindered crosswise in relation to the groove.
                   The mechanism has a loading spring (Fig. 2, 8) which forces the driving wheel against the outer
                   edge with great pressure.
8 4
                                                 2
                                                                                          6
                                                  3
Figure 2
                     The primary function of the cassette mechanism is to enable controlled linear movement of the
                     film cassette across the radiation beam synchronically in relation to the rotation movement.
                     Besides, the cassette mechanism has certain secondary functions such as secondary slot which
                     prevents the radiation scattering from patient from exposing the film and Frankfort positioning
                     light reflector which produces a horizontal light beam to facilitate patient positioning.
                     The assembly (Fig. 3, 1) is the frame of the cassette mechanism. A carriage (Fig. 3, 2) is
                     attached with bearings and axles (Fig. 3, 3 and 4) to the frame. The panoramic carriage is locked
                     to the frame.
                     The cassette holder (Fig. 3, 5) is attached with linear bearings to the carriage through axle (Fig.
                     3, 6) and support wheel (Fig. 3, 7). The cassette carriage is driven by an friction wheel (Fig. 3, 8)
                     which is in contact with the top surface of the carriage.
                     The lateral force caused by the cassette holder, the weight of the film cassette and the loading
                     spring of the driving wheel is compensated by a support wheel in the carriage (Fig. 3, 2). The
                     cassette holder rolls against the support wheel. The secondary slot (Fig. 3, 9) is attached to the
                     carriage (Fig. 3, 2). Commanded by the microprocessor a stepper motor (Fig. 3, 10) rotates the
                     friction wheel analogically with the rotation mechanism through worm gear transmission.
                     The cassette mechanism is shown below
                                                                            8
                                                                     10
                                                                                                     1
                                     2
                                9
                                         4
                                                        3
Figure 3
                               Tube head tank itself has got no serviceable parts and must always be returned to the factory for
                               repair.
                               For changing tube head see section“CHANGING THE TUBE HEAD” on page H-7.
                               The collimator tube (Fig. 4, 1) is attached with screws to the radiation window frame. Under a nut
                               there is an inner primary slot sheet which consists of an aluminium filter plate and lead plates
                               with holes. The lead plate reduces the radiation beam for the actual primary slot. The aluminium
                               plate filters the harmful soft radiation. In the outer part of the collimator tube there is a second
                               inner primary slot (Fig. 4, 2) which reduces secondary radiation scattering from the inner parts of
                               the collimator tube.
                               Part (Fig. 4, 3) is the frame of the primary slot mechanism. The primary slot (Fig. 4, 4) is attached
                               to a slot carriage (Fig. 4, 5) which is attached with bearings to the frame. The stepper motor
                               draws the slot carriage through a screw (Fig. 4, 6) and a nut (Fig. 4, 7). The drive nut has a plate
                               which together with an optic sensor (Fig. 4, 2) composes the place reference for the primary slot.
                               The microprocessor controls the stepper motor. Motor commandment is described in section
                               “Motors” on page I-10.
                               The function of the PCB 105-10-08 is described in section “Circuit boards” on page I-8.
                                           6
                                                7   1
                                                        3    2
Figure 4
                                                   9
                                                                                     2
                                       10                                       6
                                12
                                                                                              1
                                  11
Figure 5
28 27
10
22
27 44
                                                                15              1
                                     4
11
12
11
20
                                      7
                                                                  32
                                      5
Figure 6
                       The tube head is located at the other end of the rotating part.The function of it is to convert the
                       high frequency power, supplied by the generator circuit board, into X-rays. The high voltage
                       transformer and cascade circuit board in the tube head rectify and multiply this power to a maxi-
                       mum of 80 kV DC to the X-ray tube which in turn generates the radiation. The metal case gives a
                       radiation-proof hermetically closed housing to the X-ray tube. Because of the high voltage
                       needed by the tube, the tube head is filled with transformer oil to give successive insulation. The
                       tube head inner parts cannot be serviced in the field, because the opening of it needs extra clean
                       environment and after opening the oil must be dried in vacuum, but the tube head temperature
                       sensor is located in the center of the connector P1 and can be replaced without removing the
                       tube head.
                       When the tube head is replaced it must be correctly aligned and the tube type must be given to
                       the system, after which the system must be allowed to learn the filament characteristics.
                       There are nine circuit boards in the equipment. In the following their locations and functions are
                       described as well as the actions that should be taken when replacing one. Also the on board fuse
                       ratings are marked.
           CAUTION     When connecting or disconnecting cables and/or circuit boards the power must be
                       switched off and a 10 second time must be allowed to the capacitors to discharge. Other-
                       wise damage can be caused to the circuit boards.
           CAUTION     To avoid damage caused by static discharge always discharge yourself by touching the
                       earthed metal frame before touching the electrical parts.
           CAUTION     Always turn off the mains switch before handling the board because the mains voltage
                       comes onto this board. Always allow at least two minutes of time to the capacitors to dis-
                       charge after the power has been on.
                   CAUTION     Make sure that the power is switched off before handling the board to avoid electrical
                               shock and unnecessarily blown fuses if some live parts touch the frame.
           Display/keyboard (105-10-09)
                       This circuit board is located in the keyboard housing below the lower fixed part. It’s function is
                       rather obvious, to drive the displays according to the information it receives from the keyboard
                       processor and to pass the information of the keyboard states to the keyboard processor via a 6-
                       pole cable.
                       This circuit board can be replaced without any further actions.
                       F1 protects the circuits on this board and is rated 1AT
2.3 Motors
                       There are six motors in the equipment that perform all the movements needed. In the following is
                       explained where the motors are, what they do, where they get their driving power from and what
                       is the system how they are operated.
                       Five of the motors are stepper motors the operating method of which is different from the DC-
                       motor. They have two coils with two wires to each and the current is flowing through the coils in
                       both directions in turn with a phase difference that causes the rotation. Now, if the stepper motor
                       is wrongly connected or one or two wires are broken the motor might run not at all, run with low
                       torque, run to wrong direction, change the rotation direction by itself or run at double or half
                       speeds. To check this the coil resistances should be about 5 ohms each and one coil connected
                       to connector pins 1 and 2 and the other to pins 3 and 4. If it runs to wrong direction just reverse
                       one coil polarity. Reversing both causes nothing. Measuring the voltages needs an oscilloscope
                       because the motors are driven by about 40 kHz switch mode current regulated power supplies.
                       Best to have a reference motor with or to cross connect two motors to check the driver circuit
                       operations. If cross connecting, remember that the limit sensors operate wrongly.
           Up/down motor
                       The up/down motor is located on the top of the column. It’s function can clearly be seen, it is
                       moving the actual X-ray part in vertical direction along the column. It is a DC brush motor with a
                       worm gear of 1:72 and it gets it’s drive from the keyboard processor first downwards along the
                       spiral cable (on the left side seen from the front) and then upwards along a cable located inside
                       the column.
                       When this motor is activated, it is first driven one second slowly and then accelerated softly to full
                       speed. When the button is released the motor stops immediately if it was still in the slow mode,
                       but slows down softly if it was running faster. If the motor is driven to the end of it’s traverse path
                       (limit sensor) the system performs a quick stop by using the motor as a brake (short circuit). The
                       motor speed is controlled with pulse width control at 200 Hz frequency, that can be heard as a
                       humming sound.
                       The slow speed can be adjusted independently to both directions to fulfil all needs.
                              When the motor is driven to the limit sensor and stopped, it will not start to that direction but only
                              to the other. When started to the other direction it will run slowly as long as it gets back to the
                              sensor and over it. To avoid hazards the system will speed up the motor if it doesn’t reach the
                              sensor within 5 seconds when the button is kept pressed, that normally is long enough, and
                              ignore the sensor pulse. Therefore it should be noted that if for some reason the up/down move-
                              ment is driven over the limit sensor and cannot be driven back within this time (jammed) the but-
                              ton should not be operated more than 4 seconds at a time while helping the motor by hand. If this
                              is not taken care of, the system detects the sensor as if it was the sensor of the other end and
                              never comes to the normal operating range.
                              In this case enter the slow speed adjustment mode to the direction you want to go and press the
                              button. This mode ignores the sensor and enables you to reach the normal range between sen-
                              sors. While in the normal range exit the adjust mode and then drive the motor to the other end
                              sensor to get it operative again.
                   Rotation motor
                              This motor is located in the rotating part beside the generator processor board. It is rotating the
                              rotating part during the exposure and return. The drive to this stepper motor is delivered by the
                              generator processor.
                              ÷This motor cannot be operated independently but is activated by the system whenever the rota-
                              tion is needed. The operating range of the rotation is controlled by two sensors fixed to the rotat-
                              ing part frame.
                      The rotation start sensor defines the starting point (home position) of the rotating arm. When
                      pressed READY the system will continuously seek for the sensor and rotate the arm to it if it was
                      pushed aside. Note that if the arm was pushed over to wrong side of the sensor (not towards
                      center) the system tries to find the sensor by rotating the arm but in wrong direction. In this case
                      the arm must be pushed over the sensor to the actual operating range and thus enable the sys-
                      tem to find the sensor.
                      The rotation end sensor is polled in the end of the rotation. The rotation must stop on the sensor,
                      otherwise an error is generated. If the sensor was reached and gone over the rotation is stopped
                      immediately with an error.
           Cassette motor
                      The cassette motor is located on the top of the cassette holder in the cassette end of the rotating
                      part. It drives the cassette holder (and the film cassette) linearly across the X-ray beam during
                      the exposure and back during return. In cephalostatic models it also takes care of opening the
                      secondary slot wide enough and of moving the soft tissue wedge. It is a stepper motor and it gets
                      it’s drive from the generator processor.
                      The operating range of this motor is controlled by four sensors on the CASSETTE LIMITS circuit
                      board. Two of them control the position of the secondary aperture (cassette holder carriage) and
                      two the end limits of the cassette holder movements.
                      When driven to READY position the motor is operated till both the cassette holder and the cas-
                      sette holder carriage start sensors are active. If the cassette is pushed aside the system will seek
                      back for the sensors. When driven to the end direction normally a certain length is driven and no
                      sensors are used. The exception is when the equipment is powered on with the cassette not on
                      the start sensors. Then the cassette is driven all the way till it finds the end sensor of the cassette
                      holder.
                      In panoramic models the end sensor of the cassette holder carriage is never used. It is used
                      together with the end sensor of the cassette holder to detect the right position of the secondary
                      aperture when in cephalostatic mode.
2.4 Lamps
                      To make it as easy as possible to position the patient rightly, three positioning lights are provided.
                      All the lights are similar in construction and each of them has an individual fuse on the LOW
                      VOLTAGE POWER SUPPLY circuit board. The lamps are 6 VAC 10 W halogen bulbs with spe-
                      cial filament form. The light beam adjustment is done by bending the mirror rather than moving
                      the bulb and therefore the bulb replacement does not necessarily need the adjustment proce-
                      dure. It is good practice though to check the adjustments.
2.5 Wiring
                               The whole electronics of the equipment is constructed in order to keep the amount of wires at
                               minimum and especially minimize the wires at such points where they are subject to wear due to
                               movement etc.
                               The equipment wiring diagram (105-02-01-1/1-B) illustrates the whole wiring concept. It can be
                               seen that through the most critical rotating joint goes only seven wires that carry all the power
                               and information that is needed. The accurate information about each wire or cable harness can
                               be found from this wiring diagram or from the separate cable harness drawings that the reference
                               numbers on the diagram point to.
                               In the following few aspects are covered to give an overall view to the wiring concept.
                   Mains cables
                               The mains cables are tried to keep separate from all the other wires as good as possible. In the
                               220 V and 240 V equipments, where the currents are smaller the mains comes in along a spiral
                               cable on the right side of the column (front view). It ends up to the mains fuses and mains switch
                               that are located on the lower fixed part on the other side of the metal wall than the keyboard proc-
                               essor.
                               In 100 V and 110V equipments the thicker mains cable comes straight to this lower fixed part and
                               fuses.
                               From the mains switch the mains cable goes up to the upper fixed part and a mains filter. After
                               the filter it goes to the low voltage power supply that generates the operating voltages to the elec-
                               tronics.
                               Through a relay on the low voltage power supply the mains continues through the rotating joint to
                               the tube power generator circuit board that is in the rotating arm. Note that the relay only discon-
                               nects one of the mains wires and therefore even when the power is not connected to the genera-
                               tor by this relay the electrical shock hazard remains, depending on the polarity of the mains plug.
                               Always use the mains switch to disconnect both wires.
                               The mains filter in the upper fixed part takes care that no disturbances generated by the genera-
                               tor are allowed to the mains. Still there is a small filter on the generator circuit board, the purpose
                               of which is to cut off the very high frequency components that might be coupled to the low voltage
                               wires that are squeezed near to the mains wires when going through the rotating joint.
                               From the main switch on the mains cables are thick enough to be used also in the 100 V equip-
                               ments.
          Protective earthing
                     In the 200 V equipments, where the mains comes in at the bottom of the column, the first earth-
                     ing is done there. From there it continues along the mains spiral cable to the lower fixed part,
                     where it is connected to the lower fixed part. In 100 V equipments the first earthing is done here.
                     From this point the vertical part is earthed by a separate wire. The earth wire continues also to
                     the upper fixed part, where it is connected to it and also to the mains filter. From this point on the
                     earthing wire proceeds through the rotating joint and is connected to the rotating part near the
                     bearing. This is the very point, where the ground of the electronics is connected to the protective
                     earthing. From this point goes one more earthing wire to the tube head.
                     When servicing the equipment always take care that all the earthing wires are properly con-
                     nected. If the earthing is not in order the equipment can be lethal to the user.
                     The microprocessor detects if the ground of the electronics is not connected to the protective
                     earthing (tube head not earthed) and will not start the generator, but it cannot tell if the earthing is
                     not connected to the mains plug.
          Signals
                     To be able to minimize the wires two microprocessors were installed, both as near as possible to
                     the generator, motors and sensors. To pass all the information needed to operate the equipment
                     there is only one wire that connects the microprocessors together providing a serial communica-
                     tions channel. The keyboard/display is connected to the keyboard processor with a three wire
                     serial link and needs thus only five wires (actually six with double ground). The keyboard alone
                     should otherwise have needed 29 wires.
                     From the keyboard processor leaves a seven pole spiral cable down the left side of the column.
                     Two of these are needed for the up/down motor that couldn’t possibly be placed near to the proc-
                     essors. The other five wires go to the remote control box. Three of these are needed for the nor-
                     mal operation of the equipment and the two others provide one extra serial communications
                     channel to allow various kinds of new functions and equipment to be adapted to the equipment.
                     In the following are shortly described the signals in various interconnecting cables that can easily
                     be checked. The rightmost columns define the voltage range and type of the signal:
                          TUBE HEAD
                                                   SHIELD
                                                     WHT
                                                                                                                                                                                                                                                             3 SHIELD
                                                     BRN
                                                                                                                                                                                                                                                                2 WHT
                                                                                                                                                                                                                                                                1 BRN         P13
                                                                             1 WHT   T+                                                                                                                        1                                                                                           1
                                                                             2 RED   mA-                                                                                                                                                                                                                            P1                  1    P14
                                                                             3 GRN         P6
                               PI                                            4 BLK
                                                                                     mA+
                                                                                                                                                                                                         P2
                                                                             5 BRN
                                                                                     VH-                                                                                                                                                                                      P2
                                                                                     VH+
                                                                                                                                                                                                                                                                              1
                                                                               BRN    1
                                                                               BRN
                                                                                      2    P4                                                                                                                                                                                                                                                                            CASSETTE MICRO SWITCH
                                                                                      3                                                                                                                                                   CEPH. LOCK
                                                                                                                                                                                                                                                                                                                                                        1 RED
                                                                                                                                    105-10-02                                                                                                                                                                                               P12         2 BLU
                                                                                                                                    GENERATOR                                                                                                                                                                                                            1 BLU
                               P2                                                                                                                                                                                                                                                                                                                        2 GREEN
                                                                                                                                                                                                                                                                                                                                            P11          3 RED
                                                                                                                                                                                                                                                                                                                                                                                                       P1
                                                                               BRN
                                                                                      1                                                                                                                                                                                                  105-10-03                                                       1 SHIELD
                                                                                                                                                                                                                                                                                                                                                                                                      105-10-08
                                                                                      2    P3                                                                                                                                                                                                                                                            2 BRN
                                                                               BRN
                                                                                      3
                                                                                                                                                                                                                                                                                         GENERATOR                                          P10          3 WHT                                         PRIMARY
                                                                                                                                                                                                                                                                                                                                                                                              END
                                                                                                                                                                                                                                                                                         PROSESSOR                                                       1 SHIELD                             LIMIT
                                                                                                                                                                                                                                                                                                                                                                                                       POSITION
                                                                                                                                                                                                                                                                                                                                                         2 BRN                                         DETECTOR
                                                                                                                                                                                                                                                                                                                                            P9           3 WHT
                                                                                                                                                                                                                                                                                                                                                                                    START
                                                                   COIL                                                                                                                                             3                                                                                                                                                               LIMIT
                                                                                                                                                                                                                           N
                                                                                                                                                                                                                    2
                                                                                                                                                                                                                           L                                                                                                                           BLU
                                                                                                                                                                                                                    1                                         BLK
                                                                                                                                                                                                                                                                                                                                                       RED
                                                                                                                                                                                                                                                              RED             P3                                                            P8         YELL
                                                                                                                                                                                                                                                                                                                                                                                              ROTATION
                                                                                                                                                                                                                                                                                                                                                                                              MOTOR
                                                                                                                                                                                                                    P1                                        BLU                                                                                      GRN
                                                                                                                                                                                                                                                              RED             P4                                                                       BLU
                                                                                                                                                                                                                                                              1 GRN                                                                                    RED                                                  PRIMARY
                                                                                                                                                                                                                                                                        GND
                                                                                                                                                                                                                                                              2 GRY                                                                         P7         YELL                                                 SLOT
                                                                                                                                                                                                                                                              3 WHT
                                                                                                                                                                                                                                                                        SCL   P5                                                                       GRN                                                  MOTOR
                                                                                                                                                                                                                                                                        VCC
                                                                                                                                                                                                                                                                                                                                                       BLU
                                                                                                                                                                                                                                                                                                                                                       RED
                                                                                                                                                                                                                                                                                                                                            P6         YELL                                    CASSETTE
                                                                                                                                                                                                                                                                                                                                                       GRN                                     MOTOR
                                                                                                                                                                                                                                                                                                  P18
                                                                                                                                                                                                                                                                                                                                                              4 BLU
                                                                                                                                                                                                                                                                                                                                                              3 RED
                                                                                                                                                                                                                                                                                  P17                                                                         2 YELL                                        CEPH. LOCK
                                                                                                                                                                                                                                                                                                               P18                                            1 WHT
                                                                                                                                                                                                                                                                                                                                            P16 P15
                                                                                                                                                                                                                                                                                                     BLU                                                                                    POSITIONING
                                                                                                                                                                                                                                                                                                                                                                                            LIGHT
                                                                                                                                                                                                                                                                                                     BRN                  GRN
                                                                                                                                                                                                                                                                                                                                                                                            FRANKFORT
VCC 4 WHT
                                                                                                                                                                                                                                                               GND 2 GRN
                                                                                                                                                                                                                                                               SCL 3 GRY
                                                                                                                                                                                                                                                               LAC 1 BRN
                                                                                                                                                                                                 BLU
                                                                                                                                                                                                        1 N
                                                                                                                                                                                                        2 P2                              105-10-05     P1
                                                                                                                                                                                                 BRN
                                                                                                                                                                                                        3 L                               LOW VOLTAGE POWER
                                                                                                                                                                                                 BRN    1     L                                        BLU
                                                                                                                                                                                                        2
                                                                                                                                                                                                                                                                                                                                                                        UPPER FIXED PART
                                                                                                                                                                                                 BLU    3                                              BRN
                                                                                                                                                                                                              N
                                                                                                                                                                                                       P3
                                                                                                                         GRN/YELL
GRN/YELL
                                                                                                                                    GRN/YELL
                                                                                                                                                                                                                               LAMP
                                                                                                                                                                                                                               GND
                                                                                                                                                                                                                               GND
                                                                                                                                                                                                                               VCC
                                                                                                                                                                                                                               PEN
                                                                                                                                                                                                                               EXP
                                                                                                                                                                                                                               SCL
                                                                                                                                                                                                                               LAC
                                                                                                                                                                                                                   P5 P4
                                                                                                                                                                                                        GRY
                                                                                                                                                                                                        GRY
                                                                                                                                                                                                                           4 YELL
                                                                                                                                                                                                                           3 WHT
                                                                                                                                                                                                                           7 VIOL
                                                                                                                                                                                                                           5 GRN
                                                                                                                                                                                                                           8 BRN
                                                                                                                                                                                                                           1 RED
                                                                                                                                                                                                                           6 BLU
2 BLK
                                                                                                                                                                                                                                                                                                                                                                                             POSITIONING
                                                                                                                                                                                                                                                                                                                                                                                             LIGHT
                                                                                                                                                                                                                                                                                                                                                                                             SAGITTAL
                                                                                                                                                                                                                                                                                                                                                                                             POSITION
                                                                                                                                                                                                                                                                                                                                                                                             LIGHT
                                                                                                                                                                                                                                                                                                                                                                                             LAYER
                          105-10-24
                          CEPH. CONTROL
                                                                                                                                                                                                                            4 YELL
                                                                                                                                                                                                                                                             5 YELL
                                                                                                                                                                                                                                                             3 PINK
                                                                                                                                                                                                                            3 WHT
                                                                                                                                                                                                                                                             6 WHT
                                                                                                                                                                                                                            7 VIOL
                                                                                                                                                                                                                            5 GRN
                                                                                                                                                                                                                                                             2 GRN
                                                                                                                                                                                                                            1 RED
8 BRN
                                                                                                                                                                                                                                                             4 GRY
                                                                                                                                                                                                                                                 1 GRY
                                                                                                                                                                                                                                                 2 GRY
                                                                                                                                                                                                                                                             8 BRN
                                                                                                                                                                                                                            6 BLU
                                                                                                                                                                                                                                                             7 BLU
                                                                                                                                                                                                                            2 BLK
                                                            6
                                                            5
                                                                                                                                                                                                                               LAMP
                                                                                                                                                                                                                                                             ELMP
                                                                                                                                                                                                                               GND
GND
                                                            4             ASSYM.
                                                                                                                                                                                                                               VCC
                                                                                                                                                                                                                                                             VCC
                                                                                                                                                                                                                               PEN
EXP
                                                                                                                                                                                                                                                             EXP
                                                                                                                                                                                                                               SCL
                                                                                                                                                                                                                                                             SRL
                                                                                                                                                                                                                               LAC
                                                                                                                                                                                                                               LAC
                                                                                                                                                                                                                                                  LAC
                                                                                                                                                                                                                                                  LAC
                                              P2
                                                                                                                                                                                                                                                             Z+
                                                            3             LIMIT
                                                                                                                                                                                                                                                             Z-
                                                            2
                                                            1             SYMM.                                                                                                                                                  P1               P2                    P3
                                                                          LIMIT
                                                                1 YELL                                                                                                                                        1 YELL
                                                                2 RED                                                                                                                                         2 RED                                                                      BLU
                                                                3 BRN                                                                                                                                         3 BRN                                                                      RED
                YELL                                            4 BLK                                                                                                                                         4 BLK                                                          P4                                                                                 HOLD
                                                                                                                                                                                                                                                                                         WHT
                YELL                P3 P1                       5 WHT                                                                                                                                         5 WHT              P9       105-10-06                                      YELL                                                                   MOTOR
                                                                6 PURP                                                                                                                                        6 PURP
                                                                7 GRN                                                                                                                                         7 GRN
                                                                                                                                                                                                                                          KEYBOARD
                                                                                                                                                                                                                                                                                         BLU
                                                                8 BLU                                                                                                                                         8 BLU                       PROCESSOR                                      RED                                                                    LAYER
                                                                                                                                                                                                                                                                             P5          YELL                                                                   MOTOR
                                                                                                                                                                                                                                                                                         WHT
                                                                                                                                                                                                                                                                                         RED
                                                                                                                                                                                                              3 WHT            P8                                                        BLK
                                                                                                                                                                                   UP/DOWN                    2 BRN
                                                                                                                                                  GRN / YELL
                                                                                                                                                                                                                                                                                         BLU
                                                                                                                                                                                   MOTOR LIMIT                1 SHIELD                                                            P6
                                                                                                                                                                   BRN
BLU
                                                                                                                                                                                                                                                                                         GREEN
                                                                                                                                                                                                                                                                                         YEL
                                                                                                                                                                                                                                                                                                                                                          105-10-07
                                                                                                                                                                                                                                                                                         5 YELL
                                                                                                                                                                                                                                                                                         6 WHT
                                                                                                                                                               F               F
                                                                                                                    L
                                                                                                                 EL
                                                                                                                                                                                                                                    P10
                                                                                                             /Y
                                                                                                             N
                                                                                                         R
                                                                                                         G
                                                                                                                                                                                                                                                                                                YELL
                                                                                                                                                                                                                                                                                                WHT
GRN
                                                                                                                                                                                                                                                                                                RED
                                                                                                                                                                                                                                                                                                BLU
                                                                                                                                                                                                                                                                                                BLK
                                                                                            GRN / YELL
GRN / YELL
  COLUMN
                                                                                                                 GRN / YELL
                                                                                                                                                                                                                                                                                         105-10-09
                                                                                                                                                                                                                                                                                         DISPLAY/KEYBOARD
                                                                                                                                                 N
                                                                                                                                                 L
YELL
                                                                                                                                                                                                                                                                                                                                                                                      PINK
                                                                                                                                                                                                                                                                                                                                                                                      WHT
                                                                                                                                                                                                                                                                                                                                                                                      GRN
                                                                                                                                                                                                                                                                                                                                                                                      GRY
                                                                                                                                                                                                                                                                                                                                                                                      BRN
                                                                                                                                                                                                                                                                                                                                                                                      BLU
                                                                                                                                      GRN / YELL
                                                                                                                                      BRN
                                                                                                                                      BLU
                                                                                                                                                                                                                                                                                                                          + BRN
                                                                                                                                                                                                                                                                                                  UP/DOWN
                                                                                                                                                                                                                                                                                                  MOTOR
                                                                                                                                                                                                                                                                                                                           - BLU                                                                                            OUTSIDE VIEW
                                                                                                                                                                                                                                                                                                                                                                                                                              (FEMALE)
                                                                                                                                                                                                                                                                                                                                                                                                                                WHT
                                                                                                                                                                                                                                                                                                                                                                                                                      GRY                 PINK
                                                                                                                                                                                                                                                                                                                                                                                                                                4 2 5
                                                                                                                                                                                                                                                                                                                                                                                                                     YELL      1      3    GRN
                                                                                                                                                                                                                                                                                                                     4 YELL
                                                                                                                                                                                                                                                                                                                     2 WHT
                                                                                                                                                                                                                                                                                                                     3 RED
                                                                                                                                                                                                                                                                                                                     5 BLU
                                                                                                                                                                                                                                                                                                                     1 BLK
6 n.c