Xylene C H(CH ) 6 4 3 2 No.

123

n=1
ppm
Scale

123
150
200
250
100
20
50
Performance The minimum scale value (10ppm) is not printed on the tube, but only the scale line is printed.
Measuring range 5 to 10 ppm (10)to 250 ppm 250 to 625 ppm
Number of pump strokes 2(200 mL) 1(100 mL) 1/2(50 mL)
Correction factor 1/2 1 2.5
Sampling time 3 min 1.5 min 1 min
Detecting limit : 1 ppm(2 pump strokes)
Colour change : White → Brown
Operating conditions : Temperature 0 to 40 ℃ (32 to 104 ℉) correction not used
Relative humidity 0 to 90 % correction not used
Relative standard deviation : 10 %(for 10 to 50 ppm), 5 %(for 50 to 250 ppm)
Tube quantity and number of tests per box : 10 tubes for 10 tests
Shelf life : 36 months

Reaction principle
C6H4(CH3)2 + I2O5 + H2SO4 → I2

Possible coexisting substances and their interferences


Substance Concentration Interference Changes colour by itself to
Toluene ≧ 1/5 + Brown
Acetylene, Hexane ≧ 2000 ppm +(Two layers)
Pale brown(Whole layer)
Carbon monoxide ≧ 1000 ppm +(Two layers)
Benzene ≧ 1/5 + Pale yellow

Other substance measurable with this detector tube


Substance Correction No. of pump strokes Measuring range
Trimethylbenzene by scale 2 10 to 300 ppm

Calibration gas generation


Diffusion tube method