SIREMOBIL Compact L
SP
Startup
System
Serial numbers 11000 and higher
© Siemens, 2009
10409105
© Siemens, 2009
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
Print No.: SPR2-135.815.10.03.02 English
Replaces: SPR2-135.815.10.02.02 Doc. Gen. Date: 11.11
CS PS XP WH&SG
n.a.
2009
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2009“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA
and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 2 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Table of Contents 3
0Table of Contents
1 _______ Startup ________________________________________________________ 4
Required Tools, Measuring Equipment and Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Notes and icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General safety information (in existing documents) . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General electrical safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Radiation safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
I.I. laser light localizer option or single tank laser light localizer option . . . . . . . . . . . . 7
Integrated I.I. laser light localizer option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
System binder information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Notes on adapting the power plug to local conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Power supply line information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Check of temperature indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Visual inspections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Country- and customer-specific programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Fluoroscopy time limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Audible signal during fluoroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Audible signal for direct exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Backup system parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
MEMOSKOP CX: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
MEMOSKOP CX: Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
MEMOSKOP CX: DICOM network configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
MEMOSKOP CX: Parameter backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Function checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Recorded tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Image quality (IQ) quick test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
C-arm movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Checks in accordance with IEC 62353 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Country-specific acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Customer documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Installation report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
2 _______ C-arm movements/report ________________________________________ 37
C-arm movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3 _______ Changes compared to previous version ____________________________ 38
© Siemens, 2009 SPR2-135.815.10.03.02 Page 3 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
4 Startup
1-
Required Tools, Measuring Equipment and Aids
1Startup
NOTE The items listed are detailed in the SPC (Spare Parts Catalog)
if there is no other reference. Exception: items marked with
an asterisk ( * ).
• Standard tool kit*
• Device for measuring protective conductor resistance and
leakage current (in accordance with IEC 62353), SECUTEST III+ or see SPC
equivalent
• Service PC *
• Serial interface cable for service PC – SIREMOBIL, e.g. 99 00 440
• Service software * (see system binder)
Also required for IQ quick test:
• Dynamic test case, e.g. 37 90 156
• Precision X-ray filter, 2.1 mm Cu, e.g. 99 00 598
• Radiation filter set, e.g. 97 98 596
• 25 mm AL measuring stand, type 26765 in accordance with DIN 6868
Part 50*
or
1.2 mm Cu (e.g., from radiation filter set 97 98 596) plus 11 67 662
17 micrometer Cu strips, e.g.
• Resolution test, Type 41 28 71 820
• Luminence measuring device, e.g., SMFit spotmeter 77 52 848
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 4 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 5
Notes and icons 1.1
Emphasized text in this documentation means the following:
Tab. 1 GEFAHR / DANGER
Bei einer unmittelbar drohenden Gefahr, die bei Nichtvermeidung
zum Tod oder zu einer schweren Körperverletzung führt.
Indicates when there is an immediate danger that leads to death
or serious physical injury.
Tab. 2 WARNUNG / WARNING
Bei einer Gefahr, die bei Nichtvermeidung zum Tod oder zu einer
schweren Körperverletzung führen kann.
Indicates a risk of danger that may lead to death or serious phys-
ical injury.
Tab. 3 VORSICHT / CAUTION
Bei einer Gefahr, die bei Nichtvermeidung zu einer leichten oder
mittleren Körperverletzung und/ oder zu einer Sachbeschädigung
führt oder führen kann.
Indicates a risk of danger that leads to slight or moderate physical
injury and/or damage to property.
Tab. 4 ACHTUNG / NOTICE
Bei einer Gefahr, die bei Nichtvermeidung zu einem unerwünscht-
en Ergebnis oder Zustand führt oder führen kann (nicht Tod,
Körperverletzung oder Sachbeschädigung).
Indicates a risk of danger that if disregarded leads or may lead to
a potential situation which may result in an undesirable result or
state (not death, physical injury or property damage).
Tab. 5 HINWEIS / NOTE
Ist als Tipp zu verstehen. Der Anwender muss diese Anweisung
nicht unbedingt beachten. Er erfährt jedoch Vorteile, wenn er dies
tut.
Should be understood as a tip. The user does not absolutely have
to observe these instructions. However, there will be advantages
if he does.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 5 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
6 Startup
Safety notes 1.2
General safety information (in existing documents) 0
WARNING Danger of injury, death, or material damage.
Non-compliance can lead to death, injury, or material dam-
age.
Please note:
¹ The product-specific safety notes in these instruc-
tions,
¹ The general safety information in TD00-000.860.01...,
and
¹ the respective safety-relevant guidelines of your ser-
vice organization.
General electrical safety information 0
WARNING Electrical safety!
Non-compliance can lead to severe injury or even death, as
well as material damage.
¹ Parts under an electrical charge are accessible when
the covers are open. To avoid danger, disconnect the
system from the line voltage before opening the cov-
ers. Disconnect the power plug.
¹ If an uninterruptible power supply (UPS) unit is
installed in the system, the voltage output of the UPS
must also be deenergized or the voltage output plug
must be disconnected.
¹ If work steps must be performed using electrical volt-
age, the general safety information according to
TD00-000.860.01... must be observed.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 6 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 7
Radiation safety notes 0
WARNING X-ray radiation!
Non-compliance can lead to illness, irreversible damage to
body cells and genetic material, severe injury, and even
death.
When working on systems where radiation must be released,
follow the respective safety-relevant guidelines of your ser-
vice organization.
Please note:
¹ Use available radiation protection devices.
¹ Wear radiation protection clothing (lead apron).
¹ Stay as far away as possible from the radiation source.
¹ Release radiation only if necessary.
¹ Set the radiation activity as low as possible. (low kV
and mA values, short radiation time)
¹ Release radiation for as short a time as possible.
X ¹ Checks requiring the release of radiation are identified
by the radiation warning symbol shown on the left.
I.I. laser light localizer option or single tank laser light localizer option 0
CAUTION Laser emissions!
This product contains class 2 lasers. (USA: Laser class 2)
Disregarding safety precautions can lead to bodily injury,
especially to the retina of the eye, resulting in irreversible
damage to vision.
¹ Please follow the respective safety-relevant guidelines
of your service organization.
¹ When working with the laser light localizer, do not look
directly into the laser beam.
NOTE Laser emissions!
There is no direct hazard to the eye (blinking reflex). Never-
theless, do not look directly into the laser beam.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 7 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
8 Startup
Integrated I.I. laser light localizer option 0
CAUTION Laser emissions!
This product contains lasers of the class 1M.
Disregarding safety precautions can lead to bodily injury,
especially to the retina of the eye, resulting in irreversible
damage to vision.
¹ Please follow the respective safety-relevant guidelines
of your service organization.
¹ When working with the laser light localizer, do not look
into the laser beams directly with optical instruments.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 8 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 9
System binder information 1.3
The system binder is located in the monitor cart behind the keyboard. It is accessible when
the cover on the back of the monitor cart is removed.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 9 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
10 Startup
Notes on adapting the power plug to local conditions. 1.4
The customer can have the power plug on the supplied power cable replaced with an
appropriate local plug by an electrician, provided that:
- The power plug to be used can transmit the necessary power for operating the sys-
tem (see the current and voltage values on the line voltage label or in the operator
manual) and
- The required line resistance is reached (see the system planning guide) and
- The selective protection of the fusing is in accordance with the relevant national stan-
dards.
The customer must also ensure that connection to unsuitable power supply systems is pre-
vented (e.g., using labeling or constructive measures).
Protective conductor measurements should be performed and documented as a final step.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 10 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 11
Power supply line information 1.5
The delivery state of the system with respect to the power line connection values is indi-
cated by the line voltage label or in the operator manual.
Any necessary adjustments to local conditions must be performed in accordance with the
service instructions/the circuit diagram of the system.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 11 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
12 Startup
Check of temperature indicators 1.6
• Check the exterior temperature indicators for 37°C and 40°C on the I.I. housing.
- If the inner, square fields of both indicators show white, the temperature has not been
exceeded. Then remove the temperature indicators.
- If an indicator is discolored (inner field black), proceed according to IQ document
AX51-030.812.01...
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 12 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 13
Visual inspections 1.7
• No damage to the system is apparent.
• All pieces are properly attached.
• All moving parts can move freely.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 13 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
14 Startup
Options 1.8
• Options that were not already installed at the factory must be attached or installed and
put into operation according to their accompanying documentation.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 14 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 15
Country- and customer-specific programming 1.9
Fluoroscopy time limitation 0
X Setting made at the factory in the scope of DHHS guidelines:
• The buzzer sounds after 5 minutes of totaled fluoroscopy time.
Compulsory radiation shut-down occurs after 5 minutes of uninterrupted fluoroscopy
time (the buzzer sounds 30 seconds before compulsory shut-down).
Setting made at the factory in the rest of the world:
• The buzzer sounds after 5 minutes of totaled fluoroscopy time.
Compulsory radiation shut-down occurs after 10 minutes of uninterrupted fluoroscopy
time (the buzzer sounds 30 seconds before compulsory shut-down).
Using the Service PC the fluoroscopy time limitation can be reprogrammed to:
• The buzzer sounds after 5 minutes of totaled fluoroscopy time.
No compulsory radiation shut-down.
• The buzzer sounds after 10 minutes of totaled fluoroscopy time.
Compulsory radiation shut-down occurs after 10 minutes of uninterrupted fluoroscopy
time (the buzzer sounds 30 seconds before compulsory shut-down).
• The buzzer sounds for 2 seconds after 5 minutes of totaled fluoroscopy time.
No compulsory radiation shut-down.
• The buzzer sounds for 2 seconds after 5 minutes of totaled fluoroscopy time.
Compulsory radiation shut-down occurs after 10 minutes of uninterrupted fluoroscopy
time (the buzzer sounds 30 seconds before compulsory shut-down).
Country-specific regulations must be observed.
Applicable parameters: “Buzzer and block time settings” - “Buzzer warning/ X-ray blocking”
Audible signal during fluoroscopy 0
X • The audible signal during fluoroscopy is inactive at delivery. No audible signal will
sound during fluoroscopy. If national regulations require that an audible signal sound
during fluoroscopy, the service PC can be used to program the system. Applicable
parameters: “Buzzer and block time settings” - “Buzzer mode (FL / IFL / DR).”
Audible signal for direct exposure 0
X • When using direct exposure, an audible signal sounds when radiation ends. If national
regulations require that an audible signal sound during radiation, the service PC can be
used to program the system. Applicable parameters: “Buzzer and block time settings” -
“Buzzer mode (CEX).”
© Siemens, 2009 SPR2-135.815.10.03.02 Page 15 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
16 Startup
Backup system parameters 0
• After any customer-specific programming has been completed, this programming must
be saved to diskette using the host service software and be stored in the system binder.
MEMOSKOP CX: Operation 0
NOTE Note for systems without operating keyboard:
The SIREMOBIL Compact L with MEMOSKOP CX and license
for 25 images (CX 25) can be delivered without the operating
keyboard if the customer so wishes. However, a keyboard is
necessary for configuring and backing up the MEMOSKOP
CX.
In this case, connect the service keyboard included with
delivery (before switching on the system) to the MEMOSKOP
CX. (The service keyboard is kept in the compartment for the
system binder at the back of the monitor cart.)
The service keyboard has all the keys from the operating key-
board (except for the "Home" key) necessary for performing
the settings in User Setup and Technical Setup.
For the "Home" key ( ->I<- ), use the "Escape" key on the ser-
vice keyboard.
After completing all work steps for settings and backup on
the MEMOSKOP CX, unplug the service keyboard and put it
back in the compartment on the back of the monitor cart.
NOTE A service keyboard (either PS2, or USB with PS2 adapter) will
be delivered ex factory. To connect via PS2, connect the key-
board to the Memoskop "PC keyboard" port on the back of
the Memoskop. If a USB keyboard was delivered, it can be
connected to the USB port on the front of the Memoskop CX.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 16 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 17
Fig. 1: Operation Example 1
Fig. 2: Operation Example 2
Calling User Set-up Press the "CTRL" and "U" keys at the same
time.
Calling Technical Set-up Press the "CTRL" and "T" keys at the same
time.
Enter the password and press the "Return"
key.
Selecting a submenu Press the key (letter or number) displayed
above or to the right of the menu description.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 17 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
18 Startup
Example (1/Fig. 1 / p. 17), Main menu User
Set-up: Pressing the "C" key opens the sub-
menu "System Configuration."
Closing a submenu Press the "Home" key.
If data has been changed, a confirmation
window is displayed.
To save the changes, press the "Return" key.
To delete the changes, press the "Home" key.
Switching from one input field to another Cursor keys UP DOWN LEFT RIGHT, also
in a submenu refer to (2/Fig. 2 / p. 17).
Selecting/deselecting checkboxes in a Select the checkbox using the cursor keys.
submenu (setting/deleting a check) Press the "Return" key.
Selecting/deselecting list boxes in a sub- Select the list box using the cursor keys.
menu Press the "Return" key.
Selecting an entry in a list box Select the list box using the cursor keys.
(1/Fig. 2 / p. 17) Open the list box by pressing the "Return"
key.
Select the desired entry using the UP and
DOWN cursor keys.
Confirm the selection by pressing the
"Return" key.
Transferring data sets to text fields Press the "CTRL" and "A" keys at the same
(3/Fig. 2 / p. 17) time.
Example (3/Fig. 2 / p. 17), DICOM Printer
Configuration: After entering the DICOM
node data and setting the list box entries,
press the “CTRL” and “A” keys to transfer the
entry to the text field (3/Fig. 2 / p. 17) as a
data row.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 18 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 19
Modifying data sets displayed in text fields Select the data row to be modified in the text
(3/Fig. 2 / p. 17) field (3/Fig. 2 / p. 17).
Press the "Return" key to enter the data in
the following text fields as well.
Select the text to be modified or the list box
using the cursor keys.
Modify the text or open the list box by press-
ing the "Return" key. Use the cursor keys to
select another entry and press the "Return"
key.
Press the "CTRL" and "U" keys at the same
time.
The modified data row is displayed in the text
field (3/Fig. 2 / p. 17).
Deleting data sets displayed in text fields Select the data row to be deleted in the text
(3/Fig. 2 / p. 17) field (3/Fig. 2 / p. 17).
Press the "CTRL" and "D" keys at the same
time.
The data row is deleted.
MEMOSKOP CX: Configuration 0
• The following parameters may need to be changed on site:
Fig. 3: User Setup_System Configuration_1
© Siemens, 2009 SPR2-135.815.10.03.02 Page 19 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
20 Startup
User Setup, "Set Date/Time"
System Configuration: "Hospital name"
Programming:
Refer to SIREMOBIL Compact L operator manual
Fig. 4: Technical Setup - Display Setup (SW version VE00C and higher)
Fig. 5: Technical Setup - System Parameters Setup (SW version VE00C and higher)
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 20 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 21
Technical Setup, Check “Date/Time format”
Display Setup Check “UI Language”
Technical Setup, Check “Product Name”: SIREMOBIL Compact L
System Parameters Setup Check "Serial Number": The SIREMOBIL Compact L
serial number must be entered.
You can find an overview of all the Technical Setup parameters for the MEMOSKOP CX
in Troubleshooting Guide SIREMOBIL Compact L, section “Overview of MEMOSKOP CX
parameters”.
MEMOSKOP CX: DICOM network configuration 0
NOTE If DICOM functionality is not installed, skip this paragraph.
Continue with (MEMOSKOP CX: Parameter backup / p. 30).
NOTE A PING command to the MEMOSKOP CX is not answered by
the MEMOSKOP CX.
• If the DICOM option is available, the following network configurations must be per-
formed to adapt to the customer-specific network.
DICOM system-related entries
Fig. 6: Technical Setup DICOM Settings_6
• Select Technical Setup (CTRL + T).
• Input the password and confirm by pressing the Return key.
• Select "System DICOM Setup."
© Siemens, 2009 SPR2-135.815.10.03.02 Page 21 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
22 Startup
• Input the appropriate data and confirm by pressing the Return key.
• Close the "DICOM Config Setup" window.
Technical Setup, Local AE Title AE Title of SIREMOBIL Compact L.
DICOM Config Setup: Default: SM_COMPACT
Local Server Port Default: 104
Only change if instructed to do so by
the administrator.
Connection Timeout default 300 (in seconds).
Only change if network problems
with exceeding time limits occur.
Local Station Name Department designation or other ID.
Used as a search criterion to deter-
mine worklists for this department.
Field can also remain empty.
Modality Modality designation of the trans-
ferred DICOM images.
Default: RF
Possible: RF, XA, SC
Change as necessary.
Network system-related entries
Fig. 7: Technical Setup - System Parameters Setup (SW version VE00C and higher)
Technical Setup is still open.
• Disconnect the network cable (LAN cable) on the SIREMOBIL.
• Select "System Parameters Setup."
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 22 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 23
• Input the appropriate data and confirm by pressing the Return key.
• Close the "System Parameters Setup" window.
• Close "Technical Setup."
• Reconnect the network cable (LAN cable) on the SIREMOBIL.
• Switch the system off and back on.
NOTE The saved system IP address is only correctly displayed in
"System Parameters Setup" if the network cable (LAN cable)
is connected and after the system has been booted. (Display
if network cable is disconnected: 0.0.0.0).
Technical Setup, System Host Name TCP/IP network host name.
System Parameters The default name entered can be
Setup changed.
System IP Address IP address (e.g., 192.168.20.1)
System Net Mask Subnet Mask (e.g., 255.255.255.0)
Default Gateway IP address (e.g., 194.202.12.1)
DHCP Enable Default: Activated (checked).
Do not use DHCP if possible
(unchecked)!
Only use DHCP if instructed to do so
by the administrator.
DICOM storage node entries
Fig. 8: Technical Setup DICOM Settings_7
• Select Technical Setup (CTRL + T).
© Siemens, 2009 SPR2-135.815.10.03.02 Page 23 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
24 Startup
• Select "Storage Setup."
• Input the appropriate data and confirm each by pressing the Return key.
• After entering the required data for a DICOM storage node (including the data for a
DICOM storage commitment node if desired) press the "CTRL" and "A" keys at the
same time.
¹ This data is transferred to the text field. (1/Fig. 8 / p. 23).
• The "DICOM Verify" function is available to check the DICOM storage node. To do this,
select the DICOM storage node in the text field (1/Fig. 8 / p. 23) and press the "CTRL"
and "V” keys at the same time.
¹ A corresponding message is displayed.
The "SC Verify" function is also available to check the DICOM storage commitment
node. To do this, likewise select the DICOM storage node in the text field and press the
"CTRL" and "C” keys at the same time.
¹ A corresponding message is displayed.
• After the first DICOM storage node has been entered, additional DICOM storage nodes
can also be entered.
• Close the "DICOM Storage Setup" window.
• Confirm the request to save data by pressing the Return key.
Technical Setup, Provider Name/IP IP address of the first DICOM stor-
DICOM Storage Setup Address age node (e.g., 192.168.20.2). The
host name of the DICOM storage
node must be entered when using
DHCP.
Service Provider AE AE title of the first DICOM storage
Title node (e.g., DICOM_Archive).
Description Name of the storage node, displayed
in the User Menu of the MEMOSKOP
CX.
Provider Port Number 104 (default) Only change if
instructed to do so by the administra-
tor.
Overlay plane enable Check this box to display comments
in the image. (You can only select
“Small Image Size” or “Overlay plane
enable,” not both.)
1024 Lines If the programmed DICOM storage
node can only display square
images, check this box. (Transferred
image size is 1024 x 1024 pixels.)
Small Image Size If the programmed DICOM storage
node can only display reduced
images, check this box. (Transferred
image size is 512 x 512 pixels.)
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 24 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 25
Storage Commitment Default: Deactivated (unchecked).
Check the box if the DICOM "Storage
Commitment" service is to be used.
SC Provider Name/IP Enter the IP address of the DICOM
storage commitment node when
using storage commitment.
The host name of the DICOM stor-
age commitment node must be
entered when using DHCP.
SC AE Title Enter the AE title of the DICOM stor-
age commitment node when using
storage commitment.
SC Port Number When using storage commitment:
104 (default) Only change if
instructed to do so by the administra-
tor.
SC Connection Time- When using storage commitment:
out Default: 300 (in seconds).
Only increase if network problems
with exceeding time limits occur.
DICOM worklist node entries
Fig. 9: Technical Setup DICOM Settings_3
Technical Setup is still open.
• Select "Worklist Setup."
• Input the appropriate data and confirm each by pressing the Return key.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 25 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
26 Startup
• After entering the data necessary for a DICOM worklist node, press the “CTRL" and "A”
keys at the same time.
¹ The data are transferred to the text field.
• The "DICOM Verify" function is available to check the DICOM worklist node. To do this,
select the DICOM worklist node in the text field and press the "CTRL" and "V” keys at
the same time.
¹ A corresponding message is displayed.
• After the first DICOM worklist node has been entered, additional DICOM worklist nodes
can also be entered.
• Close the "DICOM Worklist Setup" window.
• Confirm the request to save data by pressing the Return key.
Technical Setup, Provider Name/IP IP address of the first DICOM
DICOM Worklist Setup Address worklist node (e.g., 192.168.20.3).
The host name of the DICOM
worklist node must be entered when
using DHCP.
Service Provider AE AE title of the first DICOM worklist
Title node (e.g., DICOM_Worklist).
Description Name of the worklist node, displayed
in the User Menu of the MEMOSKOP
CX.
Provider Port Number 104 (default) Only change if
instructed to do so by the administra-
tor.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 26 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 27
DICOM Query/Retrieve Node entries
Fig. 10: Technical Setup DICOM Settings_8
Technical Setup is still open.
• Select "Query/Retrieve Setup."
• Input the appropriate data and confirm each by pressing the Return key.
• After entering the data necessary for a DICOM query/retrieve node, press the “CTRL"
and "A” keys at the same time.
¹ The data are transferred to the text field.
• The "DICOM Verify" function is available to check the DICOM query/retrieve node. To
do this, select the DICOM query/retrieve node in the text field and press the "CTRL" and
"V” keys at the same time.
¹ A corresponding message is displayed.
• After the first DICOM query/retrieve node has been entered, additional DICOM
query/retrieve nodes can also be entered.
• Close the "DICOM Query/Retrieve Setup" window.
• Confirm the request to save data by pressing the Return key.
Technical Setup, Provider Name/IP IP address of the first DICOM
DICOM Query/Retrieve Address query/retrieve node.
Setup The host name of the DICOM
query/retrieve node must be entered
when using DHCP.
Service Provider AE AE title of the first DICOM
Title query/retrieve node (e.g.,
DICOM_QUERY).
© Siemens, 2009 SPR2-135.815.10.03.02 Page 27 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
28 Startup
Description Name of the query/retrieve node, dis-
played in the User Menu of the
MEMOSKOP CX.
Provider Port Number 104 (default) Only change if
instructed to do so by the administra-
tor.
DICOM printer node entries
Fig. 11: Technical Setup DICOM Settings_9
Technical Setup is still open.
• Select "Print Setup."
• Input the appropriate data or select options and confirm by pressing the Return key.
• After entering the data necessary for a DICOM printer node, press the “CTRL" and "A”
keys at the same time.
¹ The data are transferred to the text field.
• The "DICOM Verify" function is available to check the DICOM printer node. To do this,
select the DICOM printer node in the text field and press the "CTRL" and "V” keys at the
same time.
¹ A corresponding message is displayed.
• After the first DICOM printer node has been entered, additional DICOM printer nodes
can also be entered.
• Close the "DICOM Print Setup" window.
• Confirm the request to save data by pressing the Return key.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 28 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 29
Technical Setup, Provider Name/IP IP address of the first DICOM printer
DICOM Print Setup Address node (e.g., 192.168.20.4). The host
name of the DICOM printer node
must be entered when using DHCP.
Service Provider AE AE title of the first DICOM printer
Title node (e.g., AGFA_Drystar).
Description Name of the printer node, displayed
in the User Menu of the MEMOSKOP
CX.
Provider Port Number 104 (default) Only change if
instructed to do so by the administra-
tor.
Film Label Preferred standard labeling of film for
printouts from this SIREMOBIL Com-
pact L.
Destination Select the (Film) tray to be used on
the DICOM printer.
Priority LOW (default). Do not change.
Media Select film medium.
Polarity NORMAL (default) corresponds to
image on monitor. REVERSE corre-
sponds to X-ray film.
Border Select film border (black or unex-
posed).
Trim NO (default). Do not change.
Blank Area without image information
(black or unexposed).
NOTE To select "Film Settings," press the "CTRL" and "P" keys at
the same time.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 29 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
30 Startup
Fig. 12: Technical Setup DICOM Settings_5
Technical Setup, Film Formats Preloaded with four standard film for-
DICOM Print Setup - mats.
Film Settings Adjustable to suit customer's needs.
Film Sizes Preloaded with three standard film
sizes.
Adjustable to suit customer's needs.
Minimum Density Default: 0
Adjustable to suit customer's needs.
Maximum Density Default: 100
Adjustable to suit customer's needs.
MEMOSKOP CX: Parameter backup 0
After any customer-specific programming has been completed, this programming must be
saved on the backup CD-ROM (see system binder) and be stored in the system binder.
Technical Setup is still open.
• Remove the backup CD-R from the system binder and insert it in the CD drive of the
MEMOSKOP CX.
• In the Technical Setup main menu select "Save Settings" ("S" key).
• Take note of menu navigation on the MEMOSKOP.
• After saving the parameters, remove the CD-R from the drive.
• Close "Technical Setup."
• Switch the system off and back on.
• Store the CD-R in the system binder.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 30 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 31
Function checks 1.10
X Operating the SIREMOBIL Compact L, refer to operator manual
• Stand and C-arm system movements
• Mechanical functions of the monitor cart
• Semitransparent slot diaphragm and iris diaphragm
• Collimation of the semitransparent slot and iris diaphragm
• Image intensifier format conversion
• Image reversal
• Image position and rotation
• Image memory functions
• Fluoroscopy/Pulsed Fluoroscopy
• Image memory (ATB and LIH functions)
• Digital radiography DR
• Subtraction option
• Roadmap option
• Direct exposure option
• CD burning option
• Dicom storage option
• Dicom worklist option
• Dicom query/retrieve option
• Dicom print option
• Laser light localizer option
• Dose are product option: display cGy * cm2
• Air kerma option: display mGy (without radiation); mGy/min (during radiation with fluo-
roscopy).
• Video printer option
© Siemens, 2009 SPR2-135.815.10.03.02 Page 31 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
32 Startup
Recorded tests 1.11
• The required protocols and instructions are stored in the system binder.
Image quality (IQ) quick test 0
X
P • Perform an image quality quick test.
C-arm movements 0
P • Check and confirm the movements of the C-arm system in the report located here:
(C-arm movements/report / p. 37).
Refer to the operator manual, "Basic System" tab, "C-arm System Movements" section.
Checks in accordance with IEC 62353 0
NOTE Follow the safety guidelines given in "Performing the checks
in accordance with IEC 62353." (print no. XP00-000.860.01...).
The document is filed in the "Safety" register of the system
binder when the system is delivered.
NOTE The protocol "Record of checks in accordance with IEC
62353" (print no. XP00-000.835.02...) is filed in the "Startup"
register of the system binder when the system is delivered.
Fill out this protocol completely and then save it with the
existing protocols in the system binder.
P • Perform the checks listed in "Performing the checks in accordance with IEC 62353" and
document them in the protocol "Record of checks in accordance with IEC 62353."
Tab. 6 Measuring points for the protective earth resistance test
Measuring points for protective earth resistance test
Basic unit, a handle attachment screw for I.I. cover
Basic unit, an attachment screw for I.I. grid (not with integrated I.I. laser light local-
izer)
Basic unit, an attachment screw for I.I. housing (only with integrated I.I. laser light
localizer)
Basic unit, an attachment screw for SIREPHOS cover
Basic unit, equipotential connection at base of stand
Basic unit, a screw in the lift column
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 32 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 33
Measuring points for protective earth resistance test
Base unit, a screw on the underside of horizontal carriage
Base unit, a lateral screw at brake pedal
Basic unit, a screw on rear cover (electronic box)
Basic unit, base plate of electronic box (at unpainted location)
Monitor cart, a screw at the back of each monitor
Monitor cart, a screw at monitor support
Monitor cart, a screw at the rear panel (bottom)
Monitor cart, a screw at the rear panel (top)
Monitor cart, at a metal part of a wheel
Monitor cart, a screw at the floor plate (from below)
Monitor cart, a screw at lateral cable outlet
P • Completely fill out the protocol "Record of checks in accordance with IEC 62353" and
file it with the other reports already filed in the system binder.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 33 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
34 Startup
Country-specific acceptance 1.12
• Perform country-specific acceptance (DHHS/RöV, etc.).
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 34 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
Startup 35
Customer documentation 1.13
• Customer documentation must be given to the customer.
© Siemens, 2009 SPR2-135.815.10.03.02 Page 35 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
36 Startup
Installation report 1.14
NOTE The manufacturer of this product requires information and
the legislature demands proof that a product delivered free of
defects from the factory continues to possess the required
and certified quality properties on installation and startup.
It is therefore absolutely necessary that the installation
report with the installation and start-up data be sent without
delay after completion of the work to the address stated on
the installation report.
You will find the installation report in the product system
binder.
P • Fill out the installation report and immediately send it to the address given on the report.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 36 of 38 © Siemens, 2009
11.11 CS PS XP WH&SG
C-arm movements/report 37
2-
C-arm movements
2C-arm movements/report
System: SIREMOBIL Compact L
Material no.:
Serial no.:
Customer-specific ID no.:
Tab. 7
Movement OK Name Date Signa-
Yes/No ture
Releasing/locking the brakes
Downward movement, posi-
tion 1
Downward movement, posi-
tion 2
The warning signals sound as
described.
Function of
EMERGENCY_STOP_button
Comments:
Tab. 8
Date Comments Name Signature
© Siemens, 2009 SPR2-135.815.10.03.02 Page 37 of 38 SIREMOBIL Compact L
11.11 CS PS XP WH&SG
38 Changes compared to previous version
3Changes compared to previous version 3-
• Section "Information on the protective conductor resistance test" deleted.
• Section "Information on the leakage current measurement" deleted.
• Section “recorded tests” revised.
• Chapter "C-arm movements / report" added.
SIREMOBIL Compact L SPR2-135.815.10.03.02 Page 38 of 38 © Siemens, 2009