Vivid Q N
Vivid Q N
2015
VQN_SVC_FRNT_CVR-
GE
Vivid™ q N
Service Manual
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Important Precautions
TRANSLATION POLICY
-i
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-ii
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-iii
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-iv
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-v
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-vi
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-vii
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-viii
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-ix
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-x
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-xi
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-xii
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
Service Documentation,
GE Vingmed Ultrasound AS
P.O.Box: 141
NO 3191 HORTEN
NORWAY
GE employees should use TrackWise to report service documentation issues. These issues will then
be in the internal problem reporting tool and communicated to the writer.
WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Safety Considerations section in Chapter 1.
-xiii
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE.
GE may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
COPYRIGHTS
© 2013 and 2015 by General Electric Company Inc. All Rights Reserved.
-xiv
3.SEP.2015
P R E L I M I N A R Y FRONTMATTER.FM
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Revision History
-xv
3.SEP.2015
FRONTMATTER.FM
P R E L I M I N A R Y
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
-xvi
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose of Chapter 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Important Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Conventions Used in this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Standard Hazard Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Human Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Vivid q N Battery Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 15
Patient Data Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 16
Customer Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 26
Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 26
System manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 27
-xviii
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 2
Site Preparations
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Purpose of Chapter 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
-xix
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 3
System Setup
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Purpose of Chapter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
-xx
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 69
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 69
Vivid q N Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 70
Service Screen Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 75
Configuring Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 81
Software Options Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 84
Connectivity Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 85
Connectivity Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 85
Physical Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 88
Connectivity Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 98
Set Up Connection to a DICOM Server in a Network . . . . . . . . . . . . . . . . . . . 3 - 110
Setup Connection to HL7 Server in a Network . . . . . . . . . . . . . . . . . . . . . . . . 3 - 117
Query/Retrieve (Q/R) Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 120
MPEGVue Export Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 127
eVue Dataflow Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 132
Using MPEGVue/eVue on a Remote PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 136
-xxi
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 4
General Procedures and Functional Checks
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Purpose of Chapter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Specific Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
General Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Power ON/Boot-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Power Shut Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 7
Log On to the System as ADM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
Using Removable Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 12
Labeling Removable Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Formatting Removable Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Verifying Removable Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
Archiving and Loading Presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 16
Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Basic Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 19
Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 24
SafeLock Cart Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
Back End Processor Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 29
Image Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 32
2D Mode (B Mode) Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 32
M Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 33
Color Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 34
PW/CW Doppler Mode Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 36
Tissue Velocity Imaging (TVI) Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 37
Probe/Connectors Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 38
Site Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 40
-xxii
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 5
Components and Function (Theory)
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Purpose of Chapter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
ECG Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 39
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 39
-xxiii
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
ECG/Respiratory Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
Internal ECG/Respiratory Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41
External ECG/Respiratory Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 43
Peripherals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 46
Cooling System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 48
-xxiv
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 6
Service Adjustments
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 - 1
Purpose of Chapter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
LCD Display Adjustments and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Keyboard Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Configuring Print Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Video System Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Beamformer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Battery Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 10
-xxv
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 7
Diagnostics/Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Purpose of Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Diagnostic Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Diagnostic Procedure Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Accessing the Diagnostic Test Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
Diagnostic Test Window Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
Monitoring Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 86
Accessing the Monitoring Diagnostic Test Options . . . . . . . . . . . . . . . . . . . . . 7 - 86
-xxvi
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 8
Replacement Procedures
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 - 1
Purpose of Chapter 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
-xxvii
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Software Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 99
Software Upgrade Procedure Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 99
Setting the BIOS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 114
Formatting and Partitioning the Hard Disk - Automatic Procedure (BT’12) . . 8 - 124
Peripherals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 133
Connecting and Removing Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 133
External USB Hard Disk Replacement Procedure . . . . . . . . . . . . . . . . . . . . . 8 - 135
-xxviii
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 9
Renewal Parts
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 1
Purpose of Chapter 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
List of Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Electronic Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 6
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 8
Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 10
Product Manuals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18
Product Manuals - BT’12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 18
-xxix
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CHAPTER 10
Care and Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Periodic Maintenance Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
System Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
Preliminary Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Input Power Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Physical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 13
Optional Diagnostic Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 14
Probe Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 15
Probe Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 15
Probe Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 15
Basic Probe Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 16
Probe Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 16
Returning and Shipping of Defective Probes . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 18
Chapter 1
Introduction
Section 1-1
Overview
1-1-1 Purpose of Chapter 1
This chapter describes important issues related to safely servicing the Vivid q N portable ultrasound
scanner. The service provider must read and understand all the information presented here before
installing or servicing a unit.
Section 1-2
Service Manual Overview
This manual provides installation and service information for the Vivid q N portable ultrasound scanner.
It contains 10 chapters as outlined in 1-2-1 "Contents in this Service Manual" on page 1-2 below.
In the beginning of the manual, before chapter 1, you will find the language policy for GE’s service
documentation, legal information, a revision overview and the Table of Contents (TOC).
• Chapter 1 - Introduction
Contains a content summary and warnings.
• Chapter 2 - Site Preparations
Contains pre-installation requirements for the Vivid q N portable ultrasound scanner.
• Chapter 3 - System Setup
Contains installation procedures and an installation checklist.
• Chapter 4 - General Procedures and Functional Checks
Contains functional checks that are recommended as part of the installation procedure, or as
required during servicing and periodic maintenance.
• Chapter 5 - Components and Function (Theory)
Contains block diagrams and functional explanations of the electronic circuits.
• Chapter 6 - Service Adjustments
Contains instructions on how to make any available service adjustments to the Vivid q N portable
ultrasound scanner.
• Chapter 7 - Diagnostics/Troubleshooting
Provides instructions for setting up and running diagnostic, troubleshooting and other related
routines for the Vivid q N portable ultrasound scanner.
• Chapter 8 - Replacement Procedures
Provides disassembly and reassembly procedures for all Field Replaceable Units (FRUs).
• Chapter 9 - Renewal Parts
Contains a complete list of field replaceable parts for the Vivid q N portable ultrasound scanner.
• Chapter 10 - Care and Maintenance
Provides periodic maintenance procedures for the Vivid q N portable ultrasound scanner.
The Vivid q N ultrasound imaging scanner is similar to the Vivid i n - however, designed to offer
additional features including the ability to support use of M4S-RS probes. Also weighing only 5 Kgs
(11 lbs), this laptop-style system is extremely versatile and, depending upon the installed software, can
be used for a variety of applications.
Section 1-3
Important Conventions
1-3-1 Conventions Used in this Manual
1-3-1-2 Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
• DANGER
• WARNING
• CAUTION
When a hazard is present that can cause property damage, but has absolutely no personal injury risk,
a NOTICE is used.
CAUTION CAUTION IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL OR CAN CAUSE
MINOR PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE IGNORED.
EQUIPMENT DAMAGE POSSIBLE.
NOTICE Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
NOTE: Notes are used to provide important information about an item or a procedure.
NOTE: Be sure to read the notes; the information contained in a note can often save you time or effort.
Important information will always be preceded by the exclamation point contained within a triangle,
as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be
used to make you aware of specific types of hazards that could possibly cause harm. Even if a symbol
isn’t used in this manual, it may be included for your reference.
LASER HEAT
LASER
LIGHT
Others icons make you aware of specific procedures that should be followed.
NOTE: The Vivid q N scanner and SafeLock Cart has no unintended or motorized moving parts that
could cause pinching; all moving parts are mechanically operated by the user. Pay attention to
move such parts carefully (e.g. when raising/lowering the SafeLock Cart main shelf).
Table 1-4 Standard Icons that indicates that a special procedure is to be used
AVOID STATIC ELECTRICITY TAG AND LOCK OUT WEAR EYE PROTECTION
TAG
&
LOCKOUT
Signed Date
OR
Section 1-4
Safety Considerations
1-4-1 Introduction
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID Q N, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
WARNING BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD,
PLACING PEOPLE IN AWKWARD POSITIONS, GE HAS LIMITED THE LIFTING WEIGHT
FOR ONE PERSON IN THE FIELD TO 16 KG (35 LBS). ANYTHING OVER 16 KG (35 LBS)
REQUIRES 2 PEOPLE.
WARNING HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE VIVID Q N.
ATTEMPTS TO MOVE THE UNIT WHEN MOUNTED ON THE SAFELOCK CART
CONSIDERABLE DISTANCES OR ON AN INCLINE BY ONE PERSON COULD RESULT
IN INJURY OR DAMAGE OR BOTH.
WARNING USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY.
CAUTION USE PROTECTIVE GLASSES DURING DRILLING, FILING AND DURING ALL OTHER WORK
WHERE EYES NEED PROTECTION.
CAUTION USE SAFETY SHOES WHEN DOING WORK WHERE THERE IS ANY CHANCE OF FOOT
DAMAGE.
DANGER WHEN THE VIVID Q N SCANNER IS USED WITH THE SAFELOCK CART, MAKE SURE
THE UNIT IS FIRMLY SECURED IN THE CORRECT POSITION ON THE CART SINCE IT
MAY BECOME UNSTABLE AND TIP OVER.
DANGER WHENEVER THE SAFELOCK CART IS TO BE MOVED ALONG ANY INCLINE, USE
EXTREME CAUTION. MAKE SURE THAT THE VIVID Q N SCANNER AND ALL
PERIPHERALS ARE SECURELY MOUNTED ON THE CART BEFORE ATTEMPTING TO
MOVE IT.
WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
DANGER NEVER USE A PROBE THAT HAS BEEN SUBJECTED TO MECHANICAL SHOCK OR
IMPACT. EVEN IF THE PROBE APPEARS TO BE UNBROKEN, IT MAY IN FACT BE
DAMAGED.
CAUTION THE Vivid q N portable ultrasound scanner WEIGHS 5KG (11 LBS.) OR MORE, DEPENDING ON
CARRY-ON PERIPHERALS WHEN READY FOR USE. IN ADDITION, THE SAFELOCK CART
WEIGHS APPROXIMATELY 40KG (88 LBS.), EXCLUDING PERIPHERALS. CARE MUST BE USED
WHEN MOVING THE UNIT OR REPLACING ITS PARTS. FAILURE TO FOLLOW THE
PRECAUTIONS LISTED COULD RESULT IN INJURY, UNCONTROLLED MOTION AND COSTLY
DAMAGE.
ALWAYS:
• Use the handle to move the system
• Be sure the pathway is clear.
• Use slow, careful motions.
• Do not let the system strike walls or door frames.
• Whenever the Vivid q N scanner is mounted on the SafeLock Cart and being moved on inclines, make sure that the scanner
and all peripherals are securely mounted on the SafeLock cart before attempting to move it.
WARNING ALWAYS LOCK THE SAFELOCK CART IN ITS PARKED (LOCKED) POSITION AFTER MOVING
THE SYSTEM. FAILURE TO DO SO COULD RESULT IN PERSONAL INJURY OR EQUIPMENT
DAMAGE.
WARNING EQUIPMENT DAMAGE COULD RESULT IF SPECIAL CARE IS NOT TAKEN WHEN
TRANSPORTING THE VIVID Q N SYSTEM (AND SAFELOCK CART IF APPLICABLE) IN
A VEHICLE.
ALWAYS:
• Eject any disks from the MOD (if installed).
• Ensure that the Vivid q N system is well prepared and packed in its original
packaging before transporting. Special care must be taken to correctly
position the packing material.
For further information, refer to Chapter 3 - System Setup.
• Place the probes in their carrying case.
• Secure the SafeLock cart in an upright position and lock the wheels (brake).
• Ensure that the Vivid q N system (and SafeLock cart if applicable) is firmly
secured while inside the vehicle.
• Secure the system with straps or as directed otherwise to prevent motion
during transport.
• Prevent vibration damage by driving cautiously. Avoid unpaved roads,
excessive speeds, and erratic stops or starts.
To ensure proper grounding, connect this equipment to a receptacle marked “HOSPITAL ONLY” OR
“HOSPITAL GRADE”.
The power outlet used for this equipment should not be shared with other types of equipment. Both the
system power cable and the power connector must meet international electrical standards.
WARNING CONNECTING A VIVID Q N SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY IT.
1-4-4-1 Probes
All the probes for the Vivid q N ultrasound unit are designed and manufactured to provide trouble-free,
reliable service. To ensure this, correct handling of probes is important and the following points should
be noted:
• Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
• Do not use a cracked or damaged probe. In this event, call your field service representative
immediately to obtain a replacement.
• Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
• To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the SafeLock Cart wheels.
• Inspect the probe prior to each use for damage or degradation to the:
- housing
- cable strain relief
- lens
- seal
- connector pins
- locking mechanism
• Do not use a damaged or defective probe.
• Never immerse the probe connector or adapter into any liquid.
• The system has more than one type of probe port. Use the appropriate probe port designed for the
probe you are connecting.
NOTE: For detailed information on handling endocavity probes and invasive intracardiac probes, refer
to the appropriate supplementary instructions for each probe. In addition, refer to the Vivid q N
User Manual for detailed probe handling instructions.
1-4-4-2 Peripherals
CAUTION When a non-medical peripheral powered from a non-isolated power source is connected to the
Vivid q N system using a network cable (RJ45), the cable shielding should NOT be soldered to
the LAN connector.
Refer to Figure 1-3, below.
1. Do NOT use a soldered LAN connector when connecting the scanner directly to a peripheral device.
Figure 1-3 Example of Soldered LAN Connector
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such a watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE: Regulations vary for different countries. Dispose of a used battery in accordance with local
regulations.
• Charge batteries only when the ambient temperature is between 0 °C and 65 °C (32 °F and
149 °F), and discharge the batteries between -10 °C and 55 °C (14 °F and 131 °F)
• Do not short-circuit the battery by directly connecting the negative terminals with metal
objects.
• Do not heat the battery or discard it in a fire.
• Do not expose the battery to temperatures over 60 °C (140 °F). Keep the battery away from
fire and other heat sources.
• Do not charge the battery near a heat source, such as, a fire or heater.
• Do not leave the battery in direct sunlight.
• Do not pierce the battery with a sharp object, hit it, or step on it.
• Do not use a damaged battery.
• Do not apply solder to a battery.
• Do not connect the battery to an electrical power outlet.
WARNING IN THE EVENT THAT THE VIVID Q N PORTABLE ULTRASOUND SCANNER WILL NOT
BE USED FOR A LONG PERIOD, IT IS NECESSARY TO REMOVE THE BATTERY FROM
THE SYSTEM WHILE NOT IN USE.
CAUTION TO PREVENT THE BATTERY BURSTING, IGNITING, OR FUMES FROM THE BATTERY CAUSING
EQUIPMENT DAMAGE, ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
• Do not immerse the battery in water or allow it to get wet.
• Do not place the battery into a microwave oven or pressurized container.
• If the battery leaks or emits an odor, remove it from all possible flammable sources.
• If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
• If you have any questions about the battery, consult your local GE representative.
• Storage of the battery pack:
Short term (less than 1 month): between 0o C (32o F) and 50o C (122o F).
Long-term (more than 3 months): between 10o C (50o F) and 35o C (95o F)
NOTE: When storing a battery pack for more than 6 months, it is necessary to charge the battery pack
at least once every 6 months in order to prevent leakage and deterioration in performance (as
a result of self-discharging).
NOTE: When charging the battery for the first time after long-term storage, recover the battery pack to
original performance through repeating several cycles of full charging and discharging.
NOTE: When shipped, the Vivid q N battery is in a state of being approximately 30% charged. Before use, it is
necessary to fully charge and discharge the battery pack up to 3 times, in order to utilize Li-lon smart
packs.
Important:
• For Vivid q N portable ultrasound scanners supplied from the factory with a dummy battery, when
unpacking the unit, do not discard the dummy battery.
• When charging the Vivid q N lithium ion battery on an external charger, since the battery in fact
forms one of the four legs of the scanner, removing it will leave the unit standing unbalanced. For
this reason, the dummy battery should be kept, so it may be inserted in position to provide stability
to the scanner while the lithium ion battery is being charged.
• In addition, use of the dummy battery is recommended during transportation or long-term storage
of the Vivid q N portable ultrasound scanner.
Section 1-5
Dangerous Procedure Warnings
Warnings, such as the warnings below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID Q N, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
Section 1-6
Lockout/Tagout (LOTO) requirements
Follow all Lockout/Tagout requirements by ensuring you are in total control of the AC power plug at all
times during the service process.
Signed Date
2. TURN OFF THE BREAKER.
3. UNPLUG THE Vivid q N.
4. MAINTAIN CONTROL OF THE POWER PLUG.
5. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE, AS THERE ARE NO TEST POINTS TO
VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON WILL TURN OFF.
BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER
IS TURNED OFF WHEN THE CORD IS STILL PLUGGED INTO THE AC OUTLET.
Section 1-7
Product Labels and Icons
The Vivid q N portable ultrasound scanner comes equipped with product labels and icons. These
represent pertinent information regarding the operation of the unit.
DESCRIPTION ILLUSTRATION
Type/Class Label Used to indicate the degree of safety or protection. Bottom panel of the adapter
Table 1-6 Label Icons and Symbols - Description and Location (cont’d) 2 of 3
“DANGER - Risk of explosion The system is not designed for use with flammable See "EXPLOSION WARNING" on
used in...” anesthetic gases. page 1-8.
Table 1-6 Label Icons and Symbols - Description and Location (cont’d) 3 of 3
Section 1-8
Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GE Healtcare policy states that body fluids must be properly removed from any part or equipment prior
to shipment. GE Healtcare employees, as well as customers, are responsible for ensuring that parts/
equipment have been properly decontaminated prior to shipment. Under no circumstance should a part
or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils
or an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
NOTE: The USER/SERVICE staff should dispose of all the waste properly, per federal, state, and local
waste disposal regulations.
The Ultrasound system is not meant to be used for long-term storage of patient data or images. The
user is responsible for the data on the system and a regular backup is highly recommended.
If the system is sent for repair, please ensure that any patient information is backed up and erased from
the system before shipping. It is always possible during system failure and repair to lose patient data.
GE is not responsible for the loss of this data.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes,
GE will ascertain agreement from the customer. Patient information shall only be transferred by
approved service processes, tools and devices restricting access, protecting or encrypting data where
required, and providing traceability in the form of paper or electronic documents at each stage of the
procedure while maintaining compliance with cross-border restrictions of patient information transfers.
Section 1-9
EMC, EMI, and ESD
1-9-1 Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings, including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due to interference from its environment,
or when the device produces unacceptable levels of emission. This interference is often referred to as
radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through space or
conducted over interconnecting power or signal cables. In addition to electromagnetic energy, EMC
also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-9-2 CE Compliance
The Vivid q N portable ultrasound scanner conforms to all applicable conducted and radiated emission
limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields
and power line transient requirements.
For applicable standards refer to the Safety Chapter in the Vivid q N User Manual.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps
are in good condition and installed tightly without skew or stress. Proper installation following
all comments noted in this service manual is required in order to achieve full EMC performance.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED
ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE
POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING RISK OF ELECTRICAL SHOCK, SYSTEM MUST BE TURNED OFF. AVOID ALL
CONTACT WITH ELECTRICAL CONTACTS, CONDUCTORS AND COMPONENTS.
ALWAYS USE NON-CONDUCTIVE HANDLES DESIGNED FOR THE REMOVAL AND
REPLACEMENT OF ESD SENSITIVE PARTS. ALL PARTS THAT HAVE THE POTENTIAL
FOR STORING ENERGY MUST BE DISCHARGED OR ISOLATED BEFORE MAKING
CONTACT.
CAUTION Any changes to accessories, peripheral units or any other part of the system must be approved
by the manufacturer. Ignoring this advice may compromise the regulatory approvals obtained
for the product.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps
are in good condition and installed tightly without skew or stress. Proper installation following
all comments noted in this service manual is required in order to achieve full EMC performance.
Section 1-10
Customer Assistance
1-10-1 Contact Information
If this equipment does not work as indicated in this service manual or in the Vivid q N User Manual, or
if you require additional assistance, please contact the local distributor or appropriate support resource,
as listed below.
Chapter 2
Site Preparations
Section 2-1
Overview
2-1-1 Purpose of Chapter 2
This chapter provides the information required to plan and prepare for the installation of a
Vivid q N ultrasound unit. Included are descriptions of the electrical and facility requirements that must be
met by the purchaser. A worksheet is provided at the end of this chapter (see Figure 2-2 on page 2-10)
to help ensure that all the required network information is available, prior to installation.
Section 2-2
Console Requirements
2-2-1 Unit Environmental Requirements
Relative Humidity
Requirement Temperature (non-condensing) Air Pressure
NOTE: The Vivid q N portable ultrasound scanner system may be operated at an altitude of up to 3000
meters (9842 ft)
CAUTION If the system has been in storage or has been transported, please see the acclimation
requirements before powering ON and/or using the system. See “Installation Warnings” on
page 3-2.
NOTE: Each person in the room places an additional 300 BTU/hr demand on the environmental
cooling.
CAUTION Only one person is required to unpack the Vivid q N ultrasound unit; at least two people must be
available to roll the system down the wheeling ramp. Attempts to move the system considerable
distances (or on an incline) by one person alone, could result in personal injury, and/or damage
to the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
Input Output
Adaptor Voltage Voltage Tolerances Op. Current Frequency
100V AC to
AC DC converter 20V ±10% 0.5 to 1A 50-60 Hz
240V AC
Inrush Current
Voltage
Console Only Console with all Peripherals
100 V 4.5 A 9A
• Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
NOTE: Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
• Medical lasers.
• Scanners.
• Cauterizing guns.
• Computers.
• Monitors.
• Fans.
• Gel warmers.
• Microwave ovens.
• Portable phones.
• Broadcast stations and mobile broadcasting machines.
Poor grounding is the most likely reason an ultrasound unit will have noisy images.
Ground the unit.
Check the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may
Be aware of RF sources. be required to eliminate interference problems caused by high frequency, high powered
radio or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all
Replace and/or reassemble
screws. Any cable with an external connection requires a magnet wrap at each end.
all screws, RF gaskets,
Install the shield over the front of the card cage. Loose or missing covers or RF gaskets
covers and cores.
allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Replace broken RF gaskets.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will
Do not place labels where
permit RF leakage. In case a label has been found in such a location, move the label to
RF gaskets touch metal.
a different, appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding.
Use GE-specified harnesses
Cable length, material, and routing are all important; do not make any changes that do
and peripherals.
not meet all specifications.
Properly address peripheral Do not allow cables to lie across the top of the system. Loop any peripheral cable excess
cables. length into one bundle.
Electronics
NOTE: System and electronic probes are designed for storage temperatures of -20o to +50o C. When
exposed to large temperature variations, the probes should be kept at room temperature for a
minimum of 10 hours before use.
Section 2-3
Facility Needs
2-3-1 Purchaser Responsibilities
The work and materials required to prepare the site are the responsibility of the purchaser. To avoid
delay, complete all pre-installation work before delivery. Use the Pre-installation Check List (provided
in Table 2-7 on page 2-12) to verify that all the required steps have been completed.
NOTE: All relevant preliminary electrical installations at the prepared site must be performed by
licensed electrical contractors. Other connections between electrical equipment, and
calibration and testing, must also be performed by qualified personnel. The products involved
(and the accompanying electrical installations) are highly sophisticated and special engineering
competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must utilize only
qualified personnel to perform electrical servicing of the equipment.
To avoid delays during installation, the individual or team who will perform the installation should be
notified at the earliest possible date (preferably prior to installation), of the existence of any of the
following variances:
• A dedicated single branch power outlet of adequate amperage (see Table 2-3 on page 2-3.) that
meets all local and national codes and is located less than 2.5 m (8.2 ft) from the unit’s proposed
location. See “Electrical Requirements” on page 2-3.
• A door opening of at least 76 cm (2.5 ft) in width (if using the SafeLock Cart).
• The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
• Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
• Power outlets for other medical equipment and gel warmer.
• Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
• Clean and protected space for storage of probes (either in their case or on a rack).
• Material to safely clean probes (performed using a plastic container, never metal).
• In the case of a network option:
• An active network outlet in the vicinity of the ultrasound unit.
• A network cable of appropriate length (regular Pin-to-Pin network cable).
• An IT administrator who will assist in configuring the unit to work with your local network. A fixed
IP address is required. Refer to the form provided in Figure 2-2 on page 2-10 for network details
that are required.
NOTE: All relevant preliminary network outlets installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
Hospital Network
GE Cabinet for
Software and Manuals
(optional)
36 IN.
(92 CM)
• Wired LAN
• Wireless LAN
CAUTION In order to achieve isolation between the Vivid q N scanner and the hospital network it is
mandatory to use a Category 5 LAN cable with non-grounded shielding.
(Cable to be supplied by the hospital).
• Vivid q N Details: DICOM network details for the Vivid q N unit, including the host
name, local port, IP address, AE title and net mask.
• Routing Information: IP addresses for the default gateway and other routers in use at
the site.
• DICOM Application Information: Details of the DICOM devices in use at the site, including the
DICOM host name, AE title and IP addresses.
Section 2-4
Connectivity Installation Worksheet
Dept: Room:
CONTACT INFORMATION
Name Title Phone E-Mail Address
Name - AE Title:
Scanner IP Settings IP Address:
Name - AE Title:
Subnet Mask:
IP Address:
Default Gateway:
Subnet Mask:
Server Name:
Default Gateway:
Remote DB User Name:
Vivid i n or Vivid q N
ROUTER1 . . . . . .
ROUTER2 . . . . . .
ROUTER3 . . . . . .
Store 1 . . .
Store 2 . . .
Store 3 . . .
Store 4 . . .
Store 5 . . .
Store 6 . . .
Work list . . .
Storage
. . .
Commit
MPPS . . .
Action Yes No
Notify installation team of the existence of any variances from the basic installation.
Make sure system and probes have been subject to acclimation period.
EMI precautions have been taken and all possible sources of interference have been
removed.
If a network is used, IP address has been set for the system and a dedicated network
outlet is available.
Chapter 3
System Setup
Section 3-1
Overview
3-1-1 Purpose of Chapter 3
This chapter provides instructions for installing the Vivid q N ultrasound unit. Before beginning the
installation process, an appropriate site must be prepared, as described in Chapter 2 - Site
Preparations. Once the site has been prepared, installation can proceed as described in this chapter.
Included in this chapter are guidelines for transporting the unit to a new site, as well as procedures that
describe how to receive and unpack the equipment, and (if necessary) how to file a damage or loss
claim. Instructions for checking and testing the unit, probes, and external peripherals for electrical safety
are also provided.
NOTE: As described in "Installation Acceptance Test Criteria" on page 3-180, a Vivid q N ultrasound
scanner is ready for use after the system has been configured successfully in accordance with
the information provided in - System Setup (this chapter).
NOTE: Depending on the customer’s specific requirements, the Vivid q N portable ultrasound scanner
may have been supplied with or without the SafeLock Cart (optional). Where applicable, the
installation procedures for each scenario are clearly identified in this chapter.
Section 3-2
Installation Reminders
3-2-1 Average Installation Time
The Vivid q N installation and functional checkout will take approximately one hour; Vivid q Nconsoles
with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
Table 3-2 Average Installation Time
Average
Description Installation Time Comments
Unpacking the scanner 20 minutes
Installing the scanner 30 minutes Time may vary, according to the required configuration
DICOM Option
30 minutes Time may vary, according to the required configuration
(connectivity)
CAUTION Turning the system ON after arrival at the site - without allowing time for acclimation - may cause
system damage!
Hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20
DANGER WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH
THE UNIT!
WARNING Two people are required to unpack the SafeLock Cart, as it is heavy. Two people are always
required whenever a part weighing 19kg (35 lb.) or more must be lifted.
CAUTION If the unit is very cold or hot, do NOT turn ON power to the unit until it has had sufficient time to
acclimate to its operating environment.
CAUTION To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
CAUTION To ensure proper grounding, connect this equipment to a receptacle marked “HOSPITAL ONLY”
OR “HOSPITAL GRADE”.
CAUTION Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
CAUTION Do NOT operate the unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling. (When covers are removed, EMI may be
present).
NOTICE When the SafeLock Cart is connected to the wall outlet and the main circuit breaker is in the ON
position, the AC Box Fan is operated, even if the system is turned OFF.
NOTE: The Vivid q N User Manual should be fully read and understood before operating the unit. Keep
the manual near the unit for reference.
3-2-4-1 Overview
Unproper handling during transportation may harm the equipment inside the package even if the
package itself is undamaged.
To make it easier to detection if the handling during transportation has been unproper, a set of Tilt &
Shock indicators have been attached to the transportation box.
Section 3-3
Receiving the Vivid q N
3-3-1 Examine All Packages
Examine all packages closely at time of delivery, as described in the procedure below.
1 Is damage apparent?
• If yes; continue with the instructions in subsection
3-3-2 - Damage in Transportation.
• If no; continue with step 2.
WARNING For Vivid q N portable ultrasound scanners supplied from the factory with a dummy
battery, when unpacking the unit, do not discard the dummy battery.
When charging the Vivid q N lithium ion battery on an external charger, since the
battery in fact forms one of the four legs of the scanner, removing it will leave the unit
standing unbalanced. For this reason, the dummy battery should be kept, so it may be
inserted in position to provide stability to the scanner while the lithium ion battery is
being charged.
In addition, use of the dummy battery is recommended during transportation or
long-term storage of the Vivid q N portable ultrasound scanner.
STEP TASK
1 Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent.
RELATIVE HUMIDITY
BETWEEN 30 and 95%
TOP.
UPRIGHT
KEEP AIR PRESSURE
TRANSPORTATION
BETWEEN
& STORAGE
70 kPa and 106 kPa
FRAGILE,
HANDLE WITH CARE
Section 3-4
Unpacking the Equipment
CAUTION Please read this section fully before unpacking the Vivid q N ultrasound unit.
1.) Remove sealers and unlock two locking levers at two sides of the Box as shown in Figure 3-5 on
page 3-8
2.) Remove the packing slip (shipping consignment note) detailing the contents of the Transportation
Box, keep close to hand and be ready to mark the check list.
3.) Remove the surface packing material (paper pad, silica gel, etc.). Continue to remove additional
packing material as applicable during the following procedure steps.
4.) Carefully remove the box containing the probes.
5.) Take out the package containing the following accessories: cables (AD/DC; ECG; Network),
Isolation Box, and software CD.
6.) Verify the content against the Delivery Note.
7.) Remove each of the boxes (one, two, or more, depending on options ordered) containing the
peripherals and Vivid q N.
8.) Carefully remove the Vivid q N ultrasound unit from the box and place at a flat and stable location-
refer to Figure 3-7 on page 3-10.
NOTICE It is recommended to keep and store the transportation box and all other packing materials in
case the unit has to be moved to a different location in the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
WARNING Setting the ultrasound components to the wrong voltage setting will most likely destroy the
equipment.
Section 3-5
Preparing for Installation
3-5-1 Confirming Customer Order
When preparing for installation of a Vivid q N system, it is important to verify that all items ordered by
the customer have been received. Compare all items listed on the packing slip (shipping consignment
note) with those received and report any items that are missing, back-ordered, or damaged, to your GE
sales representative.
NOTE: The transportation box contains an External Accessory Kit and Optional Peripherals Check
List. When checking the contents of the carton, make sure the Check List is completed. In the
event that any items are missing, contact your local GE representative.
3-5-2-1 Probes
The transportation box will contain the probes that have been ordered with the system.
For a list of probes available for use with the Vivid q N portable ultrasound scanner, refer to Section 9-7
"Probes" on page 9-8.
• A Damage Inspection Checklist for the Vivid q N portable ultrasound scanner is provided in
Table 3-7 below.
• A Damage Inspection Checklist for the SafeLock Cart (optional) is provided in Table 3-8 on
page 3-14.
1 Verify that the system is switched OFF and unplugged. Clean the console and
Console
control panel.
Physically inspect the control console for missing or damaged items. Verify the
2 Control Console
proper illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify
3 Probes
that the EMI fingers around the probe connector socket housing are intact. Check
the probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive
4 LCD Display
folded cloth. Inspect the monitor for scratches and raster burn.
5 Fans Verify that the system’s cooling fans and peripheral fans are operating.
Check the rear panel connectors for bent pins, loose connections and loose or
6 Rear Panel missing hardware. Screw all the cable connectors tightly to the connector sockets
on the panel. Verify that the labeling is in good condition.
Check that all screws are tightly secured in place, that there are no dents or
7 Covers
scratches and that no internal parts are exposed.
Check and clean the peripherals in accordance with the manufacturer’s directions.
8 Peripherals To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.
Check the AC DC unit is not damaged or cracked board and verify that the output
9 AC DC System
cable is properly secured.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or
10 Power Cord
any deterioration. Verify continuity. Replace the power cord, as required.
Peripheral Check the panel connectors for bent pins, loose connections and loose or missing
2 Power and USB hardware. Screw all the cable connectors tightly to the connector sockets on the
Audio Panels panel. Verify that the labeling is in good condition.
3 Covers Check that all screws are in place, all chassis and internal covers are installed.
Check and clean the peripherals in accordance with the manufacturer’s directions.
4 Peripherals To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.
5 Check the AC board connectors and the associated cabling for good connection
AC System
and proper insulation. Verify that the connections are secured.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or
any deterioration. Verify continuity.
Tighten the clamps that secure the power cord to the unit and the outlet plug to the
cord. Replace the power cord and/or clamp, as required.
6 Power Cord
Check that the front castors can swivel, and can be placed in the locked position
7 Front Castors by pressing the foot brake lever (lower) down on each castor. Check that the upper
lever releases the SafeLock Cart wheel lock on each castor wheel.
8 Rear Castors Check that the rear castors can roll and swivel.
1 Two interchangeable USB ports (digital printer, CD-RW and other peripherals)
2 Docking connector (currently not in use)
3 Port for DC IN (AC Adapter)
4 SVGA Output (VCR option or CRT monitor option)
5 LAN 10/100 Base-TX Ethernet network connector
6 PCMCIA port for PC card.
7 Ejection lever for PCMCIA device
2 12
8
9
6
7
5
10
11
3.) Make sure that the jumper is set to either 100 to 120 V AC or to 220 - 240 V AC, in accordance with
the voltage rating shown on the safety rating plate (step 1).
WARNING CONNECTING THE VIVID Q N TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY THE SCANNER.
Section 3-6
Ensuring Protection from EMI
The Vivid q N unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI).
Many of the covers, shields, and screws are provided primarily to protect the system from image
artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are
installed and secured before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
• Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
• Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
• Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
• Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
• Verify that all EMI rules listed in the following table are followed:
NOTE: The Vivid q N ultrasound unit is approved for use in hospitals, clinics and other environmentally
qualified facilities, in terms of the prevention of radio wave interference. Operation of the
ultrasound unit in an inappropriate environment can cause electronic interference to radios and
television sets situated near the medical equipment.
Poor grounding is the most likely reason an ultrasound unit will have noisy images. Check
Ground the unit.
the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may be
Be aware of RF sources. required to eliminate interference problems caused by high frequency, high powered radio
or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Replace and/or reassemble
Any cable with an external connection requires a magnet wrap at each end. Install the shield
all screws, RF gaskets,
over the front of the card cage. Loose or missing covers or RF gaskets allow radio
covers and cores.
frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Replace broken RF gaskets.
Do not turn on the unit until any loose metallic part is removed and replaced if needed.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
Do not place labels where
RF leakage. In case a label has been found in such a location, move the label to a different
RF gaskets touch metal.
appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding. Cable
Use GE specified harnesses
length, material, and routing are all important; do not make any changes that do not meet
and peripherals.
all specifications.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Properly address peripheral
Loop any peripheral cable excess length inside the peripheral bays or hang on the hooks
cables.
provided below the console. Attach the monitor cables to the frame.
Section 3-7
Completing the Hardware Installation
The hardware installation procedures described in this section provide instructions and information for
the possible scenarios, as follows:
NOTE: If required, it is possible to disable access to all Vivid q N scanner USB ports (including the
CD/DVD drive) in order to limit system vulnerability. This is implemented from the System
Configuration window, via the Service tab.
Activate the USB External Media disabled checkbox (as shown in XREF Figure 3-22 below), then
press the F2 or Esc key to save the changes.
NOTE: After disabling all Vivid q N scanner USB ports, all peripherals that are used by connecting to
a USB port (and also the CD/DVD drive) will not be accessible until such time as access to USB
External Media is re-enabled - i.e., the USB External Media disabled checkbox (shown above)
is unchecked.
NOTE: Before connecting a printer to the Vivid q N scanner, refer to the "Safety and Environmental
Guidelines" on page 1-12.
This section provides information common to all printers approved for use with the Vivid q N ultrasound
scanner, as follows:
NOTE: The printers approved for use with the Vivid q N system are categorized in Table 3-11 below.
Refer to the appropriate instructions for connecting a specific printer, as applicable.
Report Printer (Color) with Network Capabilities • "HP Officejet Pro 8000 Color Printer" on page 3-28
Video Printer (B&W) • "Sony UP-D897 B/W Video Printer" on page 3-29
3-7-1-1-2 Configuring Hot Keys to Activate Printing Direct from the Control Panel
1.) Press Config (or F2).
2.) Select Connectivity (in the lower part of window).
3.) Select the Additional Outputs tab.
To configure printers for portrait/landscape printing, the change must be performed at the Windows
operating system level. The user first accesses the system at the Applications level from the Additional
Outputs screen. By selecting the Advanced option under Selected Devices, the user accesses the
Printer Properties dialog box and proceeds with the configuration procedure as explained below.
An additional procedure is required to configure the Sony UP-D897 Printer. This is described in
"Configuring the Sony UP-D897 Printer" on page 3-30.
Selected Devices
Select Advanced
3.) Click Configure in the Printer Properties dialog to enter Print Setup.
6.) Click Properties. The printer properties screen opens - Figure 3-17.
Orientation:
From scroll-down list,
select Portrait
NOTICE IMPORTANT
Do not connect the DeskJet Color Printer to the Vivid q N system while scanning is in progress!
NOTICE IMPORTANT
During Stand-by mode, it is NOT recommended to introduce or remove USB devices; this may
cause the system to lock-up during the boot-up procedure.
The DeskJet Color Printer connection in the different scenarios is explained below.
WARNING Do not attempt to use a different type of DeskJet Color Printer (brand or model) other than the
DeskJet Color Printer provided by GE. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
This only applies to HP 460 models which are supplied pre-installed with the appropriate software
drivers. The DeskJet Color Printer can be connected to either of the USB ports (upper or lower) on
the rear connectors panel.
When connecting directly to the system it is necessary to use an additional power source. As a
safety precaution, this must be done via an isolation transformer.
The DeskJet Color Printer can be connected once the system is powered ON, or after shutdown.
WARNING Do not attempt to use a different type of DeskJet Color Printer (brand or model) other than the
DeskJet Color Printer provided by GE. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
NOTE: The HP8000 Color Printer can be connected to the Vivid q N LAN only.
When using the USB connection, the printer should be connected through an isolated
transformer, since the Vivid q NUSB is not isolated.
NOTE: When powering the HP8000 Color Printer from the SafeLock Cart, the printer may be used via
a USB or LAN connection.
3-7-1-3-1 Connecting the HP Officejet Pro 8000 to the Vivid q N Scanner Mounted on the SafeLock Cart
NOTE: Follow the installation procedure for mounting the Vivid i/q system on the SafeLock cart, as
described in "Mounting the Vivid q N on the SafeLock Cart (optional)" on page 3-65.
NOTE: When using the Vivid i/q system mounted on the SafeLock cart, the HP Officejet Pro 8000
printer must be connected to the AC outlet at the rear of the SafeLock cart - see Figure 3-20.
1.) Connect the printer to the AC power outlet located at the rear of the SafeLock cart.
2.) Plug the SafeLock cart into the mains power supply.
3.) Using the USB communications cable, connect the HP Officejet Pro 8000 printer to the Vivid i/q
system.
USB Cable
AC Power Cable
Figure 3-20 AC Power Cable and USB Cable Connected to SafeLock Cart
4.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Color Printer Test" on page 4-24
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
NOTE: After physical connection of the printer to the Vivid q Nsystem, to configure the appropriate hot
keys to activate printing direct from the control panel, follow the instructions in "Configuring Hot
Keys to Activate Printing Direct from the Control Panel" on page 3-22.
WARNING Do not attempt to use a different type of B/W Printer (brand or model) other than the
B/W Printer provided by GE Medical Systems. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
NOTE: Once the printer is connected, perform the following functionality check:
1.) Click the Tcpip tab on the Connectivity screen (see Figure 3-14 on page 3-23).
2.) Select Advanced Settings - see Figure 3-22.
Click Advanced
Settings
Figure 3-23 Network Connections showing highlighted “Up One Level” button
4.) Navigate to Printers and Faxes and double-click Sony UP-D897 to open the Sony UP-D897
Printer Configuration dialog.
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
NOTE: After physical connection of the printer to the Vivid q Nsystem, to configure the appropriate hot
keys to activate printing direct from the control panel, follow the instructions in "Configuring Hot
Keys to Activate Printing Direct from the Control Panel" on page 3-22.
The Digital Color Printer connection in the different scenarios is explained below.
WARNING Do not attempt to use a different type of Color Printer (brand or model) other than the Color
Printer provided by GE. The ultrasound system is an extremely sensitive and complex medical
system. Any unauthorized peripherals may cause system failure or damage!
NOTE: Once the printer is connected, perform the following functionality check:
For instructions on mounting the Vivid q N scanner on the SafeLock Cart, refer to "Mounting the
Vivid q N on the SafeLock Cart (optional)" on page 3-65.
Following the connection and removal of peripherals, perform the following functionality tests:
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The USB Global Modem connection in the different scenarios is explained below.
When connecting the modem, there is no preference as to which cable to connect first. One end of the
telephone cable can be connected to the wall outlet and the other side to the modem, followed by
connecting one end of the USB cable to the modem and the other side to the lower USB port on the
rear connectors panel.
All software drivers for the modem are pre-installed for the specific modem type. One must follow the
i-Linq configuration procedure to complete the check-out process.
WARNING Do not attempt to use a different type of modem either via the PC MCIA or USB ports. The
ultrasound system is an extremely sensitive and complex medical system. Any unauthorized
peripherals may cause system failure or damage!
CAUTION For network security reasons, when configuring the wireless USB network adaptor on the
Vivid q N system, do not use the USB ports on the SafeLock cart.
CAUTION For network security reasons, both USB ports on the Vivid q N system must be configured
separately when installing the wireless USB network adaptor.
WARNING Do not use any other type of wireless adaptor other than the LinkSys WSUB54GC WLAN card.
The ultrasound system is an extremely sensitive and complex medical system. Any
unauthorized peripherals may cause system failure or damage!
Figure 3-28 Wireless USB Network Adaptor Connected to Rear of Vivid q N Scanner
NOTE: The procedures in Section 3-7-2-3-1 and 3-7-2-3-2 describe replacement of the Samsung
DVD/CD-RW drive (Model SE-TO84M/RSBN).
NOTE: .If replacing a new-type Samsung DVD/CD-RW drive (Model SE-TO84M/RSBN) with a new one
of the same model, prior to the procedure it is necessary to order the DVD Kit (P/N S2423870
- Table 3-1). Depending on the condition of the existing cables, it also may be necessary to
order the Samsung Cable Kit (S2423956) - see Table 3-2.
NOTE: If replacing only cables, it is only necessary to order the Samsung Cable Kit (S2423956) - see
Table 3-2.
The Samsung DVD SE-TO84M/RSBN Kit (P/N S2423870) includes the following:
Table 3-1 Samsung DVD SE-TO84M/RSBN Kit (P/N S2423870)
The Samsung cable kit for DVD SE-TO84M/RSBN (P/N S2423956) includes the following:
Table 3-2 Samsung Cable Kit (S2423956)
Part No Country
RS2423951 2 Port USB cable Vivid i n SafeLock Cart
NOTICE IMPORTANT - the following steps must be performed in the order stated.
(If the USB cable is connected first while power to the Vivid q N is ON, the drive will not be recognized.
In this event, it will be necessary to re-boot the system).
CAUTION Do not disrupt power to the Samsung DVD-R/RW drive from its AC/DC adapter while the USB
cable is connected to the Vivid q N system.
In this event, the system will stop recognizing the CD/DVD drive or any other device connected
to this USB port. A complete system re-boot will then be required to make the USB port
operational again.
NOTICE IMPORTANT: Use only CD-R, DVD-R or DVD+R media from a reputable vendor and only high-grade
media types.
NOTICE IMPORTANT - Always follow the safety instructions provided with the Samsung DVD-R/RW drive and
supplied by the manufacturer.
The Samsung DVD-R/RW drive, which is designed to operate with the Vivid q N system using the USB
port, combines the functions of a DVD or CD recorder, and a DVD or CD player.
This DVD device may be used on the Vivid q N system for the following purposes:
NOTE: For more details, refer to the latest version of the Vivid q N User Manual. Refer in particular to
the Archive chapter.
Removal Procedure
1.) To remove the Samsung DVD/CD-RW drive (Model SE-TO84M/RSBN):
a.) Disconnect the USB cable that connects the Vivid q N rear panel to the Samsung DVD/CD-RW
drive rear panel.
b.) Disconnect the 5V power supply output cable from the power port on the Samsung DVD-R/RW
drive rear panel.
2.) Remove the Samsung DVD/CD-RW drive from the system.
Installation Procedure
1.) Unpack the contents of the Samsung Cable Kit (P/N S2423956).
2.) Connect the mini USB connector (Figure 3-29) into the mini USB port located at the rear of the
replacement Samsung DVD/CD-RW Drive (Figure 3-30).
Samsung DVD/CD-RW
Drive Power Supply
Main Power
Connecting
Cable
3.) Connect the other end of the USB cable to the Vivid q N USB port at the rear of the Vivid q N - see
Figure 3-31.
USB Ports
NOTE: In the event that problems are experienced when exporting to DVD media, the error message
"Failed to finalize media" is displayed. It is therefore recommended to use the Medical Grade
5V Power Supply
for DVD Samsung Universal (P/N 2421169) and connect it to the peripheral outlet on the
SafeLock Cart, as shown in Figure 3-32.
Removal Procedure
1.) Turn OFF the power to the Vivid q N scanner and switch off the circuit breaker.
2.) Disconnect the power from the SafeLock Cart and from the Vivid q N scanner.
3.) Disconnect the USB cable from the USB hub located on the SafeLock Cart main column.
4.) Disconnect both ends of the 5V power outlet (one end from the SafeLock Cart main column power
and the other from the Samsung DVD-R/RW drive rear panel).
5.) Release the locking arm (retaining clamp lever) located beneath the specially-designed shelf
(DVD-R/RW drive compartment) beneath the upper shelf of the SafeLock Cart (see Figure 3-33).
6.) Slide the Samsung DVD-R/RW drive out of the compartment and remove it from the SafeLock Cart.
Installation Procedure
1.) Slide the replacement Samsung DVD-R/RW drive into the DVD-R/RW drive compartment located
beneath the upper shelf of the SafeLock Cart.
2.) Secure it in position by closing the retaining clamp lever - see Figure 3-33.
3.) Using the USB cable for the Samsung DVD-R/RW (see Figure 3-34), plug the mini USB connector
into the mini USB port located at the rear of the Samsung DVD/CD-RW Drive - see Figure 3-35.
USB connection
to DVD drive
Figure 3-35 USB Cable Plugged Into the DVD/CD Drive and the Cart’s USB Connectors
4.) Plug the USB connectors into the USB ports located below the upper shelf of the SafeLock Cart
- see Figure 3-35 and Figure 3-36.
Figure 3-36 USB ports Located Below the Support Tray of the Vivid q N Cart
5.) Connect the USB cable located on the SafeLock Cart main column, into the USB port located at the
rear of the Vivid q N system (Figure 3-31 on page 3-39).
6.) Re-connect power to the SafeLock Cart and to the Vivid q N system.
7.) Switch on the circuit breaker and turn ON power to the Vivid q N system.
8.) Perform the following test:
a.) Insert a CD-R disk in the DVD/CD-RW Drive.
b.) Eject and remove the CD-R disk to check the eject mechanism functions properly.
NOTICE IMPORTANT
Do not connect the MOD Drive to the Vivid q N system while scanning is in progress!
NOTICE IMPORTANT
During Stand-by mode, it is NOT recommended to introduce or remove USB devices; this may
cause the system to lock-up during the boot-up procedure.
WARNING Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
NOTICE IMPORTANT Do not connect the External USB Hard Disk to the Vivid q N system while scanning
is in progress! This will avoid the risk of causing image artifacts.
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The External USB Hard Disk can be connected to any of the USB ports on the SafeLock Cart.
One must use the additional power source (supplied with the External USB Hard Disk) to activate the
External USB Hard Disk.
WARNING Do not attempt to use a different type of mass storage device other than the
External USB Hard Disk supplied with the system. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
Refer to the External USB Hard Disk Installation Procedure section, page 8-136 for instructions on initial
configuration and connecting the External USB Hard Disk to the Vivid q N portable ultrasound scanner.
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The USB Memory Stick (disk-on-key) connection in the different scenarios is explained below.
3-7-2-7 Footswitch
NOTICE IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
NOTE: After connecting the Footswitch to the system, define the required functionality settings, as
follows:
3.) Using the Footswitch Functionality drop-down menus, select the required functionality for the
Left, Middle and/or Right Footswitch - see the example in Figure 3-42.
NOTE: When performing step 1, it may be necessary to use the VGA adaptor (P/N 2420728 - see
Figure 3-44) to connect the end of the VGA cable to the Vivid q N rear connector panel.
Since some VGA plugs have thicker housing, they do not fit properly into the
Vivid q N VGA connector socket on the rear panel (because the plug is not centrally aligned
within the frame). In this event, use of the VGA adaptor will eliminate the possibility of causing
damage to the BEP by the use of unnecessary force.
2) Connect the other end of the VGA cable to the VGA INPUT on the video converter - see Figure 3-45.
4.) Connect the other end of the S-Video cable to the VCR S-Video INPUT at the rear of the VCR.
NOTE: Dip switches are located beneath the video converter, as shown in Figure 3-46 below.
Dip Switches
5.) Follow the manufacturer’s guidelines to set the switches to the correct positions, as shown in
Figure 3-47.
6.) When all power and signal cables are connected to the system, VCR, and video converter, proceed
as follows (in the order stated):
a.) Press the power ON switch on the VCR.
b.) Press the power ON switch on the video converter.
c.) Turn ON power to the Vivid q N system.
7.) Press the Input Select button (see Figure 3-45 on page 3-48) on the video converter three times
in order to select VGA Input as your signal source.
The LED located on the video converter front panel will be illuminated, accordingly.
8.) Press on the three buttons <CTRL> <ALT> <Page Erase> simultaneously.
The following screen will appear.
9.) Select the Intel® Dual Display Clone icon and press OK - see Figure 3-48
NOTE: If the VGA cable is not properly connected to the rear panel, or in the event that the
scan-converter device is not switched ON, you may see a single icon displayed, as shown in
Figure 3-49.
NOTE: When the VGA cable is disconnected and reconnected while the system is ON, the VGA output
is maintained.
NOTE: If system is shut-down to standby or Full-shut-down while the VGA device is connected
continuously, the VGA output will appear after the system is turned ON
NOTE: If the system is shut-down to Standby or Full-shut-down and the VGA device is disconnected,
the VGA output will not appear after power is turned ON, and the above procedure will have to
be repeated.
DVI Cables
NOTE: Installation of Carto3 and the required VGA splitter (to connect to the Vivid q N) shall be done
by the BWI field engineer according to the BWI manuals and instructions.
WARNING To avoid possible electrical shock to user or patient, the Vivid q N should be connected to the
Carto3 only by using the Isolated VGA Splitter.
1.) Connect the VGA cable (Figure 3-52) to the VGA input port at the rear of the VGA splitter
(see Figure 3-53 and Figure 3-54).
Connect the other end of the cable into the Vivid q N system’s VGA output port.
Plug the other end of the power cable into the mains power wall outlet.
Interface Connections
• Do not expose the charger or power supply to water or conductive liquids (this is not a sealed case).
• Do not open the charger or power supply case. There are no user-serviceable parts inside.
• Do not cover the fan exhaust or obstruct the airflow, as this will cause overheating.
• Use only the manufacturer’s 24V 2.5A power supply.
• Place the charger in a cool location, away from external heat sources.
• During recalibration, the battery connector and base of the charger may become warm.
• Recharge Time:
The time for a full charge from 0% to 100% state of charge is typically 4.5 hours.
• Recalibration Time
The recalibration cycle begins by discharging away any residual capacity. Then a calibration
charge is delivered to the battery. This is followed by a calibration discharge. Finally the
battery is given a regular charge. A calibration cycle will be faster if the battery is fully
discharged to begin with.
The recalibration time is approximately 9 hours for an empty battery, and approximately
14 hours for a fully-charged battery.
NOTE: Calibration is initiated each time the button is pressed, so it is not recommended to press the
recalibration button part way through the recalibration cycle.
During the charging procedure, the green LED will be seen flashing.
At the end of this procedure, when the battery is fully charged, the solid green LED will be
illuminated in the status window, indicating a fully charged battery.
2.) Remove the battery from the charger unit battery bay and install it in the
Vivid q N portable ultrasound scanner.
NOTE: For more details on recalibration and an explanation of why it is necessary, refer to
Recalibration on page 5 - 82.
If the battery is in need of “fuel gauge” recalibration, the red LED will flash upon insertion of the battery.
This indicator provides feedback to the user on the accuracy of the fuel gauge and avoids unnecessary
battery calibration cycles.
The user has the option to calibrate the fuel gauge and charge the battery, or to only charge the battery.
This option is given because a recalibration cycle is longer than a charge cycle.
1) To recalibrate the battery, press the calibrate button on the front of the battery charger unit.
No action is required if only a recharge is required, as the charger will automatically begin to charge
the battery.
The blue calibration LED will flash to indicate that the battery is undergoing the recalibration cycle.
There may be a short delay before the calibration begins. During calibration the discharge resistors
will heat up and the fan will operate to maintain temperature within acceptable limits.
At the end of this procedure, when the battery is fully charged and the battery fuel gauge is
calibrated, the solid green and blue LEDs will be illuminated in the status window, indicating a fully
charged, fully calibrated battery.
2.) Remove the battery from the charger unit battery bay and install it in the Vivid q N portable
ultrasound scanner.
NOTE: The most common cause of calibration failure is overheating of the pack during discharge. For
this reason, keep the charger away from direct sunlight or heat sources.
• Two active probe connectors (one for a pencil probe) on the right side of the unit
Probes can be connected or changed any time, as described below, regardless of whether the system
is powered ON or OFF.
1) Place the probe’s carrying case on a stable surface and open the case.
2) Inspect the probe socket to verify that it is free of debris.
3) Carefully remove the probe and unwrap the probe cord.
DO NOT allow the probe head to hang freely. Impact to the probe head could result in
irreparable damage.
4) Press the probe connector locking lever downwards (refer to Figure 3-57).
5) Align the connector with the probe port and carefully push into place, as shown in Figure 3-58.
6) Press the connector locking lever upwards to the full vertical position to lock in place, as shown in
Figure 3-59.
7) Carefully position the probe cord so that it is free to move and is not resting on the floor.
When the probe is connected, it is automatically activated.
• To connect the internal ECG carefully push the end of the cable connector into the ECG connector
port (circular-shaped socket on the left side of the system beside the probe connectors), as shown
in Figure 3-60 below.
CAUTION When disconnecting the ECG Cable from the Vivid q N, do not pull on the cable directly. Carefully
remove the cable by holding the outer sleeve (closest to the system) while releasing the cable.
The same connector is used both for external and internal ECG. The appropriate ECG cable kit should
be ordered when external ECG is to be used.
See also information on Ext-ECG Module Connector provided in "ECG Module" on page 5-39.
NOTE: For optimal ECG operation, use only electrodes that meet universal standards - see Table 3-4.
Use of the External Respiratory option is enabled by removal of the ECG Board and installation of the
Internal ECG/Respiratory Board. For instructions, refer to the "ECG/Respiratory Module Assembly
Installation Procedure" on page 8-18.
NOTE: An External ECG (Respiratory) cable should be available. This is provided by GE (Part #
2413451 - see Section 9-5 "Cables" on page 9-6). Alternatively, the adaptor supplied with the
GE cable may be used to adapt an existing cable - refer to the optional connections for the
External Respiratory interface illustrated in Figure 5-34 on page 5-43.
Chapter 3 - System Setup 3-61
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CAUTION Verify compliance with all electrical and safety requirements and check the power cord to verify
that it is intact and of hospital-grade before connecting the unit to the power source. Products
equipped with a power source (wall outlet) plug should be connected to the fixed power socket
that has a protective grounding conductor. Never use an adapter or converter to connect with a
power source plug (for example, a three-prong to two-prong converter).
CAUTION Make sure the system is powered OFF or in standby mode before disconnecting the Vivid q N
unit from the electrical outlet and proceed as described in the steps below.
NOTICE When AC power is applied to the scanner, the On/Off button on the control console illuminates amber,
indicating the Back-end Processor is in Standby mode.
2) Hold down the On/Off button on the control panel for 3 seconds. The system automatically performs
an initialization sequence which includes the following:
• Loading the operating system.
• Running a quick diagnostic check of the system.
• Detecting connected probes.
3) When prompted, log in to the system as appropriate.
(Refer to the instructions for system login and user security setup as described in the
Vivid q N User Manual).
The system first enters 2D-Mode with the probe and application that were last used before the system
was shut down. If the probe has been removed since the system was last used, the currently-connected
probes and their available applications are displayed and selected by default.
NOTE: After turning OFF the system, wait at least 10 seconds before turning it on again. The system
may not be able to boot-up if power is recycled too quickly.
• Holding down the On/Off button for 3-5 seconds will evoke the display of the System Exit dialog
box (see Figure 3-61). From this, you can either select Full Shutdown (click the Shutdown button)
or choose to place the system in Standby Mode (click the Standby button).
NOTE: The system will remain in Standby Mode for as long as battery power will allow.
Section 3-8
Mounting the Vivid q N on the SafeLock Cart (optional)
NOTE: The Vivid q N portable ultrasound scanner may be mounted on the SafeLock Cart while power
to the system is either ON or OFF. When mounting the unit on the cart, it may be necessary to
adjust the locking mechanism, as described in 3-8-2 "Adjusting the SafeLock Cart Locking
Mechanism" on page 3-68.
NOTICE Only the SafeLock Cart supplied by GE and/or any medical-grade cart which has been approved as
suitable for use with the system by GE should be used with the Vivid q N portable ultrasound scanner.
NOTICE The Vivid q N system and the SafeLock Cart should be operated at a distance of more than 15 feet from
any magnetic field.
NOTICE When scanning with the Vivid q N portable ultrasound scanner mounted on the SafeLock Cart, the
Vivid q N should be connected to the Cart via the USB port.
WARNING Before mounting the Vivid q N portable ultrasound scanner on the SafeLock Cart make
sure there are no foreign objects on the main shelf surface or in front of the locking
tooth.
1.) Working from the front of the SafeLock Cart, place the Vivid q N scanner flat down on the main shelf
(towards the front) then slide it backwards between the sliding guides on either side of the shelf, as
shown below.
Sliding Guides
Figure 3-62 Placing the Vivid q N on the SafeLock Cart Main Shelf
3.) If necessary, perform the alignment procedure as described in 3-8-2 "Adjusting the SafeLock Cart
Locking Mechanism" on page 3-68.
4.) Connect the LAN Cable from the cart to the LAN connector on the Vivid q N rear panel.
Docking Connector
USB Port
LAN Connector
5.) Connect the Docking Connector to the rear of the scanner, as shown below.
Docking Connector
3-66 Section 3-8 - Mounting the Vivid q N on the SafeLock Cart (optional)
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
1.) Working from the left side of the SafeLock Cart, remove the DVD from the DVD/CD-RW device.
Working from underneath the main shelf, loosen and remove the four screws (two each side) from
the DVD/CD-RW device bracket, then remove the bracket and the device - see Figure 3-67 on page
3-68.
Screw
Screw
2.) Working from the underneath the main shelf, use a Phillips screwdriver to loosen the two screws
(half turn counter-clockwise) holding the middle locking tooth in position on both the left and right
sides of the main shelf, as shown below.
Middle Locking Teeth Viewed from Top of Shelf Screws Viewed from Beneath Shelf
3.) Check that the Vivid q N portable ultrasound scanner is locked correctly onto the two rear locking
teeth (if not, push the scanner fully back into position until an audible click is heard).
4.) Carefully tighten the four screws previously loosened in step 2.
5.) Reconfirm that the scanner is firmly locked into position and there is no excess play (movement)
between the scanner and the Cart before returning the DVD/CD-RW device and bracket.
3-68 Section 3-8 - Mounting the Vivid q N on the SafeLock Cart (optional)
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Section 3-9
Configuration
3-9-1 Overview
Table 3-5 below outlines the Vivid q N Configuration procedures described in the relevant sub-sections.
PAGE
SUB- SECTION DESCRIPTION NUMBER
3-9-2 Vivid q N Configuration 3-69
3-9-2-1 Preparations
Press CONFIG (F2) and log on as Adm (see "Log On to the System as ADM" on page 4-10).
3. Select the Department field, see (b) in Figure 3-68, and type the This name will be displayed on the image properties on saved images as
name of the department (max 64 characters). soon as the unit has been restarted.
NOTE: If the date and time displayed by the system are set to the BIOS default value, it may be
necessary to replace the BIOS battery - see "BIOS Battery Replacement" on page 8-47.
Language
2. Use the Language drop down dialog, see Figure 3-70, to select The selected language will be used as soon as the unit has been
your preferred language for the on-screen interface. restarted.
The following languages are available;
- ENG - English
- NOR - Norwegian
- DEU - German
- ITA - Italian
- FRA - French
- ESP - Spanish
- RUS - Russian (after selection, continue to step 3)
- GRE - Greek (after selection, continue to step 3
- DAN - Danish
- SWE - Swedish
4. Under Standards and formats, in the language field, scroll down The interface and keyboard input will be in Russian.
to the desired language (e.g. Russian) and click OK.
Reboot the system.
Units
2. Use the drop down Units dialog (see Figure 3-71) to select The selected units (Metric or US) will be used for measurements as soon
Metric or US Units. as the unit has been restarted.
3. Restart the scanner. All the changed settings will be used after the restart.
3-9-3-1 Overview
The Service Screen gives you access to:
NOTE: The Disable TCPIP Filtering option should only be used for Troubleshooting procedures, after
which it it is recommended to return immediately to the TCPIP Filtering enabled state.
For instructions, see "LCD Display Adjustments and Calibration" on page 6-1.
1) Press <F2>, select Service Tab. The Select Keyboard Setup dialog is displayed.
3-9-3-5-2 Languages
1.) Under the Languages tab, click Details to select appropriate Language.
2.) If necessary, activate applicable Supplemental Language Support check box, then click Apply.
3-9-3-5-3 Advanced
1.) Under the Advanced tab, select a language version of the non-unicode programs you want to use,
then click Apply.
NOTE: This option should not be used by un-authorized personnel. This option will prompt you to the
Windows Installation Wizard. Normally, all software drivers are pre-installed and there is no
need to use this functionality. In special cases where there is a need, follow the instructions
provided by GE Medical Systems or the instructions in the respective printer installation
manual.
1.) Select ADD PRINTER to start the Add Printer (Installation) Wizard.
NOTICE Vivid q N supports reading of DVD media, but not writing data to DVD.
- 100 V AC
Lexmark C762 Color Printer
- 110 V AC
Lexmark C762 Color Printer
- 220 V AC
Lexmark C762 Color Printer
Installation of the Lexmark C762 Color Printer is described in the Lexmark C752N/C762N Color
Printer Installation Manual, Direction Number FC194475.
Once the printer has been re-connected again, all the spooled print jobs will be printed out
automatically.
Sometimes this feature is device dependent. Therefore, in order to initiate the manual "print of all
spooled jobs" process, proceed as described below.
4.) Right-click and select Restart to resume print job/s for printing, as required
NOTICE When using Peripherals on the Vivid q N SafeLock Cart, always observe the general guidelines listed
below.
General Guidelines
1.) In Standby mode, no change should be made to USB-connected devices (regarding location,
introduction of new USB device, removal of a currently-connected USB device).
2.) USB devices should be connected and/or disconnected only in Full Shutdown or Normal Scanning
mode.
3.) USB devices should not be connected and/or disconnected before completion of tasks assigned to
them.
4.) Whenever the Vivid q N ultrasound scanner is mounted on the SafeLock Cart, always use the
lower USB port for the Hub.
5.) The Vivid q N ultrasound scanner and all connected devices working with scanner should be
powered and connected to the SafeLock Cart (only)!
For the SafeLock Cart, there are two basic peripheral configurations that are recommended:
Configuration - Recommendation 1
Configuration - Recommendation 2
• Memory Stick or HP460+Batt plugged directly into the upper USB port on the rear of the Vivid q N
ultrasound scanner.
New
Admin
Section 3-10
Connectivity Setup
3-10-1 Connectivity Introduction
The Vivid q N ultrasound unit can be connected to various connectivity devices, such as DICOM
devices and EchoPac servers. The following sections describe how to connect the system to a remote
archive/work station or a DICOM service, using a TCP/IP connection.
This section describes communication and connection options between the Vivid q N ultrasound unit
and other devices in the hospital information system.
Patient information can include demographic data and images, as well as reports and Measurement
and Analysis (M&A) data.
A dataflow is a set of pre-configured services. Selecting a dataflow will automatically customize the
ultrasound unit to work according to the services associated with this dataflow.
By utilizing dataflows, the user can configure the Vivid q N ultrasound unit to optimally meet the needs
of the facility, while keeping the user interface unchanged. Once the dataflow is selected, the actual
location of the database is entirely transparent.
A list of all the predefined dataflows is included in the Vivid q N User Manual.
Vivid q N
LocalArchive-Int.HD dataflow:
The local database is used for patient archiving.
Images are stored to internal hard drive.
Vivid q N
LocalArchive-MOD dataflow:
The local database is used for patient archiving.
Images are stored to a MOD
EchoPAC PC
RemoteArch-RemoteHD dataflow:
A remote database is used for patient archiving.
Images are also stored to a remote archive.
In this configuration, the scanner is just an intermediate acquisition unit which, after completion of a
study, will not contain any patient information, measurements, or images. The acquisition can be done
online or offline.
MDI Button
Controls the fifth
connector, making it
crossed or non-crossed.
X
To connect an end station or a Server to the Port
To PC
=
To connect a HUB or another Switch to the Port
To HUB
Wall connection
or Vivid7 Vivid3
EchoPAC PC
See the EchoPAC PC Service Manual (Part Number EP091298) for details about the EchoPAC PC,
and the required set-up procedures.
9.) Select "Install from a list or specific location (Advanced)" - see Figure 3-90 and click Next.
10.)Select Include this location in search (Figure 3-91) and browse to "D:\WIFI DRIVER\"
(Figure 3-92).
11.)Click Next.
The following screen opens (Figure 3-92).
3.) From the Wireless Network Connection screen navigate to Wireless Network Connection 8 and
double-click to open the Wireless Network Connection 8 screen - see Figure 3-95.
6.) Return to the instructions for "Installing the Compact Wireless-G Adaptor Software" on page 3-90
and repeat the entire procedure in order to configure the second USB port.
7.) When done, proceed to perform the following functionality tests:
- "Wireless USB Network Test" on page 4-25
1.) Connect one end of the Crossover Cable to the network connector on the Vivid q N.
2.) Connect the other end to the network connector to the EchoPAC PC Workstation, see Chapter 3 in
the EchoPAC PC Workstation Service Manual.
1.) Connect one end of the cable to the Ethernet connector on the Vivid q N.
2.) Connect the other end of the cable to the wall outlet.
If a Peer-to-Peer Network is connected to the hospital’s network, you may connect the
Vivid q N to the Peer-to-Peer Network.
3-10-3-1 Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
• Before you can set up the scanner, you need to collect some information. Refer to the "Worksheet
for DICOM Network Information in Figure 2-2 on page 2-10. Typically, the source for this
information is the network administrator.
• Follow the steps below to prepare the scanner for use on the network.
Computer Name:
For Vivid q N, this name is on the form:
Vivid_I-00nnnn, where “00nnnn” is a number
(nnnn is the scanner’s serial number).
IP Settings:
IP Settings area for the Vivid q N scanner.
AETitle
IP Address
Subnet Mask
Default Gateway
NOTE: Before starting this procedure, verify that the destination PC / Network has a Shared folder
(remote path) with user permissions. Refer to section 3-10-9-1-1 on page 3-137 for creating a
new user in the remote PC or network.
NOTE: The paths below (refer to Figure 3-103) are used as follows:
• Save As PDF Network Path - used for saving system Reports as PDFs.
• Remote Path - used for Save As, Export from Q-Analysis, and for
exporting error logs with "Alt + D".
1.) From the system keyboard press Config and log on as an ADM user with the appropriate password.
2.) Click Connectivity (from the bottom row of buttons) on the screen and select the Tools tab.
The Tools screen opens - see Figure 3-103.
Remote Path
3.) Under "Save As PDF Network Path", in the Remote Path field, type the remote path.
For example, in Figure 3-103 type: \\IP ADDRESS\DIRECTORY, where directory is the name of
the shared folder in the destination PC.
NOTE: When clicking the Check button, you might momentarily see a "failed" message that disappears
if you click Check a second time. The system will then confirm the setting.
5.) Under "Setting for Remote Path used for Save As...", type the remote path as follows:
\\IP ADDRESS\DIRECTORY.
NOTE: It is possible to configure an additional user name and password for all remote paths. These
settings are only used if the system fails to log on the default user and password.
• Under Configurable Remote Path User (refer to Figure 3-103), type the desired user name and
password, as configured on the destination PC.
(This is the password usually assigned by the local IT person).
Save settings
NOTE: The IP must be different from the IP address used for network computer.
NOTE: The Subnet mask should be the same for both the network computer and the
Vivid q N.
13.)Click OK
14.)Press the Config button on the control panel to close all windows and return to the regular scanning
mode.
NOTE: The IP address must be different from the one used for the Vivid q N system.
NOTE: The Subnet mask should be the same for both the network computer and the
Vivid q N.
9) Click OK, close all windows and approve all changes. Return to regular mode.
10) Connect the Vivid q N and network computer LAN connection using a cross over Network cable.
(Normally with RED connector covers on both sides, cable not longer than 100 meters).
The Vivid q N and the network computer are now ready for export, import of files.
NOTE: One can check the connection by pinging to the Vivid q N IP, as follows:
NOTE: The IP must be different from the IP address used for network computer.
12.)Type in the Subnet mask 255.255.255.0 or the one provided by the local network administrator.
NOTE: The Subnet mask should be the same for both the network computer and the
Vivid q N.
13.)Click OK
14.)Press the Config button on the control panel to close all windows and return to the regular scanning
mode.
NOTE: One can check the connection by pinging to the Vivid q N IP, as follows:
This scenario requires that the scanner is configured to be connected to the DICOM server as described
below.
3-10-4-1 Overview
To work against the DICOM server, the following information has to be entered in the scanner:
Arrow
The selected flow is shown, as seen in the example in Figure 3-116 below (Worklist/Local Archive
- DICOM Server/Int. HD).
6.) Select Worklist (so it is highlighted) and then select Properties to display the Properties dialog.
Select Worklist
and then select
Properties
9.) In the Select Tag drop down menu select "00080060 Modality".
10.)Type US in the Value text box and click on "Add to List".
The Search Criterias window is updated (see Figure 3-118).
NOTE: It is not possible to change the setting in the IP-Address field by editing it. See description
below, starting with step 13, to change the setting.
Select server
Modify
c.) Edit the name and/or the IP address of the server, see Figure 3-123.
Name
IP address
d.) Click OK to save the new settings and close the Edit Name dialog box.
e.) Click OK to close the IPs dialog to return to the Properties dialog box.
AE Title
Port No
Select device to
be verified.
Check
A green check-mark
indicates that the test
passed.
A red cross (as
illustrated) indicates
that the connection
failed.
3-10-5-1 Introduction
The procedures below describe how to set up the Vivid q N workstation so it can connect to the Hospital
Information System (HIS), via the Vivid HL7 Gateway. By connecting to the HIS, demographic
information (i.e. Patient ID, Name, Gender, etc.) can be retrieved from the HIS, reducing “double-work”
and typing errors. Patient Reports can be sent back to the HIS for storage, together with the rest of the
Patient records.
3-10-5-2 Preparations
1.) Before you can set up the HL 7 Communication, you need to collect some information:
- The IP address of the Vivid HL7 Gateway
- Port Number used by the Vivid HL7 Gateway (Default port: 3320)
Typically, source for this information is the Mitra support personnel.
The "Connectivity Installation Worksheet" on page 2-10 can be used for gathering this information.
2.) Provide the computer name of each EchoPAC PC Workstation and Vivid q N being configured to
the Mitra support personnel.
NOTE: This configuration is optional. It should only be performed if the site is directing orders to the
Vivid q N that is scheduled to perform the study.
Please contact the Mitra support personnel to verify that this step is needed.
1.) In the Search Criterias dialog box, select 00400001 Scheduled Station AE Title from the Select
Tag pull-down menu.
2.) Enter the Vivid q N computer name for the Value.
3.) Click the Add to List button.
4.) Click the OK button.
NOTE: This configuration should be performed if the site will need to constrain the Worklist queries by
MRN (patient id)
1.) In the Search Criterias dialog box, select 00101000 Other Patient IDs from the Select Tag
pull-down menu.
2.) Enter a 9 digit number for the Value.
3.) Click the Add to List button.
4.) Click the OK button.
1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality
Worklist service.
2.) Expand the Gateway destination device.
3.) Select the Gateway Modality Worklist service.
4.) Click the Check button.
- If connectivity was achieved, then a green check should be displayed next to the name of the
Gateway's Modality Worklist service.
- If connectivity was not achieved, then there will be a red X next to the name of the Gateway's
Modality Worklist service.
b.) Verify that Text is select for the Export file format.
3-10-6-1 Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Vivid q N.
NAME
PULLDOWN
MENU
QUERY RETRIEVE
QUERY RETRIEVE
PROPERTIES DIALOG
SELECT
QUERY RETRIEVE AND THEN
SELECT PROPERTIES
IP-ADDRESS
DOWN-ARROW
AE TITLE
PORT NO
6.) Select the IP-address down-arrow to choose the DICOM Query/Retrieve server from the pull-down
menu. In some cases, the server to use is the same as used for DICOM Storage.
If the server to use is missing from the list, select <Modify> from the pull-down menu and edit the
setup for one of the predefined servers.
7.) Enter the correct AE Title and Port Number for the DICOM Query/Retrieve server in the respective
fields in the Query/Retrieve screen.
SEARCH CRITERIAS
SELECT TAG
ADD TO LIST
REMOVE
SMILEY
VERIFICATION
CHECK BOX
OK
2.) If the network connection to the server is OK, it will be illustrated by a smiling Smiley .
A sad Smiley indicates that the network connection is failing.
Typical cause: Network cable not connected.
3.) When ready, click OK to close the DICOM Query/Retrieve properties and save changes.
QUERYRETRIEVE
CHECK
- If the test fails, a pop-up warning is displayed, see left illustration in Figure 3-135.
Select OK to continue. A red “X” to the left of the Selected devices indicates that the test
failed.
3-10-7-1 Introduction
The procedures below describe how to set up the Vivid q N portable ultrasound scanner for MPEGVue
export. This provides the user with the ability to export an entire patient exam into standard supported
media (such as, CD, DVD, or USB Memory Stick [disk-on-key]) in standard MPEG or JPG format. The
exported exam can later be read on any PC.
Media containing one or several patient exams in MPEGVue format is easily generated on the
Vivid q N. When exporting data from the system for this purpose, the exported examination contains files
in HTML, MPEG and JPG formats for all of the images in the exam. It also contains measurements and
reports generated and saved during the exam; the reports are all saved as files in .CHM format.
Any media generated as MPEGvue also contains an embedded MPEGVue viewer, allowing the user to
view it on any PC without having to install any additional software.
The exported exam may be viewed on any PC operating on Windows 2000, 2003, Windows XP or
Windows Vista.
This can prove useful in many situations - for example, when sending patient data for a second opinion,
for academic lectures, conferences, presentations, etc. Another way of using this feature is to directly
provide the patient with an MPEGVue disk (CD or DVD) containing, for example, an OB exam.
3-10-7-2 Preparations
Obtain local IP addresses that will allow you to set up the system and the remote computer properly.
This information is available from the local system administrator.
5) From selected output devices, highlight evue and then click the Properties button - refer to
Figure 3-137 on page 3-129.
6) In the Remote Path field, type in the path to which you want to export MPEGVue (when this feature
is used).
NOTE: If the destination is a constant IP, you can type either the remote computer name or the remote
computer’s ID address in the Remote Path field.
After the path has been typed, it is available for selection from the Dataflow Name
drop-down menu (or you can select an existing path from this menu).
NOTE: Step 7 is recommended for the first transfer of MPEGVue files. This step is not needed for
Windows XP. After initial transfer of MPEGVue, it is recommended to disable this check box.
• CD
• MOD
• USB Memory stick (disk-on-key)
1.) Insert the required media into the appropriate drive (or USB port), when applicable.
2.) From the main screen, trackball to the Patient button and press SET.
3.) End the current exam (if still open), and select Patient-List.
The Patient List will be displayed with the following header:
4.) Trackball to the Export softkey in the bottom menu shown below and press SET.
7.) Trackball to the required Patient File and select it by pressing SET.
To select more than one Patient File from list, use the <Shift> key together with SET
8.) When ready, after the required Patients Files have been selected (highlighted,) press the COPY
button.
The selected files will be converted and copied into the selected media. When done, the following
message is displayed:
9.) If required, click OK then select any additional Patient Files for export.
10.)When export complete, click OK then remove the media containing the exported MPEGVue data
from the Vivid q N system.
NOTE: If any problems occurred with copying the Patient Files during the export process, an error
message such as the one below will be displayed, showing the status of each of the transferred
Patient Files.
3-10-8-1 Introduction
It is possible to configure the Vivid q N portable ultrasound scanner to allow the user to send images or
loops to a remote PC, in almost real time. When pressing Store, the loop will save not only to the local
archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”) on the network
(in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital in real-time is required.
3-10-8-2 Preparations
Before configuring the Vivid q N for eVue dataflow, it is necessary to decide upon a specific PC in the
hospital as being the e-Vue PC station. The service technician then configures the
Vivid q N to send images to that specific e-Vue PC, as described below.
10.)Trackball to the Remote Path field and type in the IP address and path of the remote computer:
\\10.0.0.2\EVUE_IMPORT, or type in the remote computer full name \\Name\EVUE_IMPORT
(refer to Figure 3-141).
11.)Trackball to the Destination field, open the drop-down menu and select Remote Path
(\\10.0.0.2\EVUE_IMPORT), or the computer name \\Name\EVUE_IMPORT
(refer to Figure 3-141).
12.)Activate the Copy Media Player Installation check box.
NOTE: Step 12 is recommended for the first transfer of eVue files. This step is not needed for Windows
XP. After initial transfer of eVue, it is recommended to disable this check box.
The Vivid q N dataflow is set and ready for export/import of eVue files into a network computer.
From now on, whenever the Vivid q N is used for scanning and the Store button is pressed, another
person will be able to view the stored loop on the remote e-Vue PC (after a short transmission time).
When new loops arrive, the following message will be displayed at the bottom of the screen:
“New images have arrived - To view please refresh the screen using F5".“
2.) Type the patient’s Last Name and ID in the appropriate fields (see Figure 3-142).
3.) Select the Dataflow you intend to use for this patient’s information (in this case, select Local
Archive – Int HD/eVue as shown in Figure 3-142).
4.) Using the soft key buttons, select CREATE PATIENT.
5.) Now start scanning. Each image you would like to store press the STORE button.
A few seconds after the Store button is pressed, images will be transferred into the remote
computer (which was previously set as the destination - refer to Figure 3-141 on page 3-133).
6.) To proceed, follow to the instructions as described for "Using MPEGVue Viewer on a Remote PC"
on page 3-152.
When a DVD/CD containing MPEGvue exams is placed into a DVD/CD drive on a remote PC, it should
self-boot within a few seconds and display the Patient List.
The first time this is done on the remote PC there will be a one-time operation where the MPEGvue
program will self-install. If the remote PC operating system is Windows XP the installation is automatic
and, when completed, the Patient List on the DVD/CD will be displayed.
In the event that the remote PC operating system is Windows 2000, the auto-installation will display one
of the messages shown below (see Figure 3-143 and Figure 3-144), requesting the user to download
and install some drivers.
The requested drivers may be downloaded from the Internet, or, if the DVD/CD was pre-configured,
those drivers will be located on the MPEGvue DVD/CD itself.
In addition to the above, the installation program will test the configuration of the remote PC and will
display a message if any component on the computer is not within the required specifications.
c.) Right-click on the New Folder icon and select Rename. Rename the folder as eVue _Import:
9.) Setting the Sharing properties for the new folder, as follows:
a.) Right-click on eVue_Import icon and select Properties;
The eVue_Import Properties dialog box opens, as shown in Figure 3-148 on page 3-141.
Note:
It is recommended to
remove this user.
g.) From the Select Users, Computers, or Groups dialog box (Figure 3-150), click on the
Locations button.
h.) In the Locations dialog box, select your computer from the list and press OK.
i.) In the Select Users or Groups dialog box, type the User Name in the empty window, as shown
in Figure 3-151.
Activate Allow
check boxes
for all
m.) Activate the Allow check boxes for all of the following: Full Control, Change, and Read.
n.) Click the OK button.
o.) Close all windows.
NOTE: The network computer is now ready to receive files from the Vivid q N system. At this stage, set
the appropriate path for export of files into the network computer. For example, for eVue, Data
Flow or MpegVue export into the network computer using the browsing option to select the
appropriate path. For an example of Vivid q N dataflow settings, refer to Figure 3-140 on page
3-133.
NOTE: In some cases, it is necessary to stop the firewall to allow functions such as Excel export and
eVue to access and load files into the Network Computer. The Firewall can be an external
service, or part of the Windows Basic Firewall tools.
NOTE: You may need to obtain these from the local IT person.
5.) From the C:\Program Files\GEULS\eStorePlayer\Version1_0\ folder, delete all files, as shown
in Figure 3-160 on page 3-151.
6.) Restart Windows, then confirm that the MPEGVue application is no longer installed.
NOTE: In the event that the MPEGvue DVD/CD is being run for the first time, the program will
self-install the MPEGvue viewer. For more details, refer to "Installing MPEGVue on a Remote
PC" on page 3-136.
1) Insert the media (MOD or Memory stick) containing the MPEGVue exams into the MOD drive (or
USB port) on the remote computer.
2) Double-click on the file START_MPEGVue.bat.
The MPEGVue Patient List is displayed (see Figure 3-161 below).
1) Click on the required examination Exam Date (labelled 1 in the example above) to select and
display images of that exam.
The MPEGVue screen is displayed as shown in the example in Figure 3-162 below.
1. Clipboard: select the image to display 8. Save image as MPEG or Bitmap file
There are several ways to select a thumbnail image from the clipboard:
• Use the mouse pointer. With the pointer on the required image, click the left mouse button to
select the image.
• Use the up or down keyboard arrow to view the next or previous loop in the exam.
• Click on the Next Image or Previous Image buttons (#11 in Figure 3-162), as required.
Chapter 3 - System Setup 3-153
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
• Clicking on the Freeze/Run icon (#5 in Figure 3-162) will freeze the loop. Clicking on same icon
once again will run the loop again.
• Clicking on the scrolling icons (#6 in Figure 3-162) will stop the loop, allowing the user to advance
forwards or backwards to the required frame.
• Using the left / right arrows - when the loop is in Run-Mode, pressing either the left or right arrows
on the PC keyboard will freeze the loop. Further pressing the left or right arrows will advance the
loop frame-by-frame.
In a 2D image:
NOTE: If the original image contains measurements that were taken and stored on the
Vivid q N scanner, those measurements will appear in the MPEGvue image. In this case,
additional measurements cannot be performed on that image.
NOTE: Currently, the measurement function is not available on DICOM images/loops, any secondary
captured screens 3D images, or images with discontinuities.
This feature allows users to archive some important measurements into the MPEGvue file.
When an exam contains Reports, an additional button is displayed on the bottom of the MPEGvue
Screen enabling the user to view the Reports of that exam.
NOTE: To use the MPEGVue over E-mail feature, Microsoft Outlook must be currently running. If not,
the following message is displayed prompting the user to first logon to Outlook:
It is possible to send images, loops, or full exams from any eVue PC. An eVue PC is a PC which
contains the MPEGvue player and an e-mail client application (e.g. Microsoft Outlook).
The currently-selected image, or the entire examination, can be sent by e-mail as an attachment to one
or several regular e-mail messages. Refer to Sending a Full Examination by E-mail below, or "Sending
an image by E-mail" on page 3-157, as applicable.
The recipient PC of those e-mail messages (where the eVue viewer is installed), will receive the exams
and will integrate them into its Patient List, for easy review.
NOTE: If this is the first time your recipient is receiving MPEGVue images, it is necessary to attach the
MPEGVue Player - refer to the instructions for "Sending the MPEGVue Player Software by
E-mail" on page 3-159.
1) From the MPEGVue Screen, open the examination to be sent and click the E-mail button (#10 in
Figure 3-162 on page 3-153).
The E-mail Dialog box is displayed (see Figure 3-163 on page 3-156).
2) Activate the Send Current Exam radio button (and if necessary, the Attach Player Software
check box), as shown in Figure 3-163 below.
5) Click Next.
The user is prompted to enter a password (for encryption of the study) in the window displayed:
NOTE: The password should be given to the recipient of the e-mail verbally by phone or by other
protected means.
(It is recommended to keep a separate record of passwords used).
7) Click Next.
The E-mail (addressed to the selected addressee) with the full examination attached is sent to the
Outbox; during this operation a series of progress windows are displayed, as seen in this example:
The extension of the attached files is .mpegvue and can only be opened with the MPEGvue
viewer.
NOTE: If this is the first time your recipient is receiving an MPEGVue image, it is necessary to attach
the MPEGVue Player - refer to the instructions for "Sending the MPEGVue Player Software by
E-mail" on page 3-159.
1) From the clipboard, select the image to be sent and click the E-mail button (#10 in Figure 3-162 on
page 3-153).
NOTE: If it is necessary to send the MPEGVue Player software to the recipient, first complete this
procedure, then proceed to "Sending the MPEGVue Player Software by E-mail" on page 3-159.
4) Click Next.
The E-mail (addressed to the selected addressee) with the image is sent to the Outbox; during this
operation a progress window is displayed:
NOTE: This procedure is required once - only the first time your recipient is receiving MPEGVue
image/s.
1) From the E-mail Dialog box, activate the Send Player Software check box:
NOTE: When the message has been sent from the Outbox and has arrived at the addressee, the
recipient should follow the instructions (included in the body of the email) in order to install the
MPEGVue Player.
1) From Outlook, open the received email message containing the installation of MPEGVue Player
files (three attachments: MPEGVuePlayer.cab; Start.wmv; Setup.html). Right-click on each
attachment and select Save As to save them to the Desktop.
3) When prompted by the Warning message shown below, make sure all Internet Explorer windows
are closed except the current window, then click Yes to proceed.
NOTE: At this point, the MPEGVue startup icon is now displayed on the desktop, ready for future use.
The user may choose to view the images immediately or later (as preferred). Follow the
instructions described in "Get New Exams by E-mail" on page 3-163.
The following message is displayed while the e-mail Inbox is being scanned for studies:
When complete, all studies found in the Inbox are listed in the Found Studies dialog box, as shown
in Figure 3-176 below.
NOTE: By default, the Open Exams for Review and Delete Received Mail from Inbox check boxes
are activated.
7) Click Next.
The message Receiving Selected Messages is displayed, followed by another message (see
Figure 3-178 below) showing progress of deletion of messages from the Inbox.
The E-mail message(s) containing the MPEG examination is deleted from “In-box” of Outlook and
is moved to the “Deleted-items” box of Outlook. When done, the following message is displayed:
NOTE: When viewing the examination which has been received from e-mail, a separate instance of
MPEGvue application is opened, dedicated only for one e-mail examination. If you were
previously viewing different MPEGvue Patient exams, you will need to switch over to the
previous instance of the MPEGvue.
In addition, if an e-mail message is lost or corrupted, the contents of the remainder of the received
messages is available for review.
After receiving even one e-mail message out of many that have yet to arrive, simply select the new
examination in the MPEGVue Screen. The screen display will show voids or missing thumbnail images
in all places relating to the e-mail messages that have not yet arrived.
Activate e-Vue
1) From the MPEGVue Startup Menu, click on the Activate e-Vue button.
The Patient List is displayed, listing all e-Vue exams received (refer to Figure 3-180, above).
2) Click on the required examination Exam Date (labelled 1 in the example above) to select and
display images of that exam.
The MPEGVue screen is displayed as shown in the example in Figure 3-162 on page 3-153.
Figure 3-181 MPEGVue Patient Manager Window - Manage Exams from Email Tab
The MPEGVue Patient Manager window has four tabs that provide access to the following utilities:
• Manage Exams from Email
• Manage Exams from e-Vue
• Manage Other Exams
• System Configuration
NOTE: By default, when the MPEGVue Patient Manager window is first opened, the Manage Exams
from Email tab is activated.
NOTE: When the selected studies have been copied to the destination folder, they will remain listed in
the MPEGVue Patient Manager Patient List. However, if they have been moved, they will no
longer be listed.
d.) To delete the selected exams, click the Delete Selected Patients button.
NOTE: When the selected studies have been deleted, they will no longer be listed in the MPEGVue
Patient Manager Patient List.
Figure 3-183 MPEGVue Patient Manager Window - Manage Exams from E-VueTab
The Patient List displays details - in table format - of all patient exams received via export from
E-Vue (Patient ID, Patient Name, Exam Date, etc.).
The current path information is displayed below the Patient List as well as number of studies,
cumulative size and amount of space available on the drive. By default, the destination path
Other Folder radio button is enabled.
b.) Select the required exam or exams - refer to step b on page 3-168.
c.) Copy or move the selected exam/s, as required - refer to step c on page 3-169.
d.) If required, delete the selected exam/s - refer to step d on page 3-169.
Figure 3-184 MPEGVue Patient Manager Window - Manage Other Exams Tab
NOTE: When the Manage Other Exams tab is first activated, the MPEGVue Patient Manager Window
opens (by default) with the most recently-selected source directory active (even if the user
selected this directory during Copy, Move, or Delete operations performed under another tab).
At this stage, no information will be displayed in the Patient List.
Once the user has specified any source directory that contains MPEGVue studies (these may be
located on CD/DVD-ROM, DOK or any local hard drive), exam details are displayed (in table format)
in the Patient List. The information shown includes Patient ID, Patient Name, Exam Date, etc. as
seen in the example in Figure 3-184, above.
The current path information is displayed below the Patient List together with details of the
removable media (e.g., CD/DVD-ROM). By default, the destination path Other Folder radio button
is enabled; additional information below shows the Path to destination folder.
b.) Select the appropriate Current Path Information - if necessary, click the adjacent Browse
button and select the correct source directory.
All patient exams found in the currently-selected source directory are listed in the Patient List.
c.) Select the appropriate Path to Destination Folder - if necessary, click the adjacent Browse
button and select the required directory.
d.) From the Patient List, select the required exam/s to be copied.
If the currently-selected source directory is CD/DVD-ROM (read-only media), the Copy
Selected Patients button is now enabled, as shown in Figure 3-185 below.
NOTE: If the currently-selected source directory is DOK or the local hard drive, the Delete Selected
Patients and the Move Selected Patients buttons are also now enabled.
When done, the currently-selected exam/s are available in the destination folder for review.
System Configuration read-only information is displayed (in table format) under the following
column headings: Item, Value, Minimum Requirements.
Below this information, the eVue Operation Utility buttons enable the user to enable or
disable the eVue mode, as required. An additional button (Configure eVue) is used to perform
the eVue configuration procedure.
NOTE: In order to enable eVue operation, it is necessary to first perform the eVue configuration
procedure as described in the following step.
If applicable, enable the peer-to-peer connection check box then type the correct current
IP address and Subnet mask details:
The Enable eVue button is now active, as shown in Figure 3-187 below.
NOTICE IMPORTANT To avoid system configuration inconsistencies, before using the Enable eVue or
Disable eVue functions, make sure that your PC is connected to the appropriate network and that all
network cables are properly connected. Otherwise, configuration of your PC may become incorrect and
the Configuration utility will be unable to execute the Enable eVue or Disable eVue operations.
In cases of a peer-to-peer connection, before using Enable eVue always make sure the two computers
(Vivid q N scanner and your PC) are properly configured and connected directly.
Before using Disable eVue, your PC should be connected back to the local network.
c.) Click the Enable eVue button to activate the eVue operation.
The following message is displayed:
d.) When required, click the Disable eVue button to de-activate the eVue operation.
The following message is displayed:
NOTE: On most computers, when inserting the DVD/CD media with MpegVue files, the MpegVue
Player will start automatically (due to the autoplay option of DVD/CD media).
In the event that any updates are required on the remote computer, the program will prompt the
user to install the appropriate components, as necessary. (Follow the on-screen instructions).
When done, the Patient List is displayed.
4.) From the Patient List, double-click on the required examination Exam Date (see the example
below) to select and display images of that exam.
The Image Review window opens, displaying thumbnail pictures of all the currently-selected
patient’s images - see Figure 3-188 on page 3-177.
NOTE: The MpegVue Player can be used as a regular player, using the Play, Stop, Forward, and
Rewind buttons. In addition, the user can forward images by e-mail, save images, and perform
measurements, as previously described.
The MpegVue Player displays two rows of command keys at the bottom of the screen. If the
remote computer is set up to view many Tool Bars, the lower row of the MpegVue command
keys might not be visible. Therefore, it is recommended to remove most Tool Bars (by selecting
<View> <Tool Bars> and un-checking all tools that are not required) until both lines of command
keys are visible, as shown below.
Command Keys
Section 3-11
Storing and Transporting the Unit
The Vivid q N ultrasound unit mounted on a SafeLock cart weighs up to 45 kg (99 lbs), depending on
the configuration. To avoid equipment damage and ensure maximum safety while transporting or
moving the unit, the following precautions are recommended:
CAUTION Whenever moving the Vivid q N ultrasound unit while mounted on the SafeLock Cart, always
observe the following safety precautions:
1) Proceed cautiously when crossing door or elevator thresholds. Grasp the front handle grips or the
back handle bar, and then push or pull. DO NOT attempt to move the unit by pulling its cables or
probe connectors.
2) Ensure that the unit does not strike walls or door frames.
3) Ensure that the pathway is clear.
4) Move the unit slowly and carefully.
5) Take extra care while moving the unit long distances and on inclines.
6) Avoid ramps with an incline of greater than ten degrees to prevent the unit from tipping over. Once
the destination is reached, lock the SafeLock Cart front wheel brakes.
For more details, refer to the appropriate procedure in Section 3-4 "Unpacking the Equipment" on page
3-8, and follow the instructions in reverse order, or see the Vivid q N Unpacking and Packing Procedure,
Direction FL094023. Use the latest revision of the manual.
Section 3-12
Completing the Installation Paperwork
NOTE: On completion of a Vivid q N ultrasound scanner installation, it is important to record the
relevant System Installation details (see Table 3-11 below) and Product Locator Installation
details (see Figure 3-190 on page 3-181), for future reference.
Location:
Date:
Customer Name:
Comments:
1.) Verify that all details have been entered in the System Installation Details form, as described in the
System Installation Details section, above. If this form has not been completed, enter the details
now.
2.) Locate the Product Locator Installation Card(s) that are affixed to the monitor and to each of the
peripherals, and fill in all details on both the front and back of the card(s) - see Figure 3-190
"Product Locator Installation Card (Example)" on page 3-181.
3.) Mail the Product Locator Installation Card(s) to the appropriate product center.
Section 3-13
Product Locator Installation Card
NOTE: The Product Locator Installation Card shown may not be the same as the provided Product
Locator card.
From the factory, a sheet with four Product Locator cards for transportation and one for Installation are
included.
Mail the Product Locator Installation Card(s) to the appropriate product center.
During and after installation, supplementary documentation (such as, User Manuals and Installation
Manuals for the peripherals) must be kept as part of the original system documentation. This will ensure
that all relevant safety and user information is available during the operation and service of the complete
system.
Chapter 4
General Procedures and Functional Checks
Section 4-1
Overview
4-1-1 Purpose of Chapter 4
This chapter describes the functional checks that must be performed either as a part of the installation
procedure, or as required during servicing and periodic maintenance of the
Vivid q N ultrasound scanner.
4-2 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Section 4-2
General Procedures
4-2-1 Overview
Since some general procedures are used more than others, for convenience, all the procedures
described in this section are listed below:
DANGER TO AVOID THE RISK OF FIRE, POWER TO THE SYSTEM MUST BE SUPPLIED
FROM A PROPERLY RATED OUTLET.
IT IS RECOMMENDED TO USE A DEDICATED POWER OUTLET. THE POWER
PLUG SHOULD NOT, UNDER ANY CIRCUMSTANCES, BE ALTERED TO A
CONFIGURATION RATED LESS THAN THAT SPECIFIED FOR THE CURRENT.
DO NOT USE AN EXTENSION CORD OR ADAPTOR PLUG. REFER TO 2-2-5
"Electrical Requirements" on page 2-3 FOR MORE DETAILS.
DANGER ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE
PROTECTIVE GROUNDING CONNECTOR.
DANGER NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE
SAFETY GROUND.
DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
NOTICE Use only power supply cords, cables and plugs provided by or designated by GE.
NOTE: When turning on a system from standby mode, it takes a few seconds before it responds. Do not push
the On/off button again during this period. A second push will initiate a full shutdown.
NOTE: If the unit has been in the OFF condition for an extended period of time, (3 to 5 days or more), the unit
may not boot, or may beep when turned on.
NOTE: If the system cannot boot up, or it boots up with a PS communication error, refer to the instructions for
8-3-4 "BIOS Battery Replacement" on page 8-47.
1) Ensure that the wall outlet is of appropriate type, and that the power switch is turned OFF.
2) Uncoil the power cable, allowing sufficient slack so that the unit can be moved slightly.
3) Verify that the power cable is without any visible scratches or any sign of damage.
4) Verify that the on-site mains voltage is within the limits indicated on the rating label on the base of
the Vivid q N portable ultrasound scanner.
5.) Connect the Mains Power Cable female plug via the AC Adapter to the Power Inlet (DC IN - refer
to Figure 4-1) at the rear of the unit.
DC IN
.
Figure 4-1 Mains Power Connector - Rear Panel
6.) Connect the other end (male plug) of the Mains Power Cable to a hospital-grade mains power outlet
with the proper rated voltage.
The unit is now ready for Power ON/Boot Up.
NOTICE When AC power is applied to the scanner, the On/Off button on the control console illuminates amber,
indicating the Back-end Processor is in Standby mode.
2) Hold down the On/Off button (Figure 4-2) on the control panel for 3 seconds. The system
automatically performs an initialization sequence which includes the following:
• Loading the operating system.
• Running a quick diagnostic check of the system.
• Detecting connected probes.
On/Off Button
a.) The unit’s ventilation fans starts on full speed, but slow down after a few seconds (listen to the
fan noise).
b.) Power is distributed to the peripherals, Operator Panel (control panel), Monitor, Front-End
Processor and Back-End Processor.
c.) The Back-End Processor and rest of the scanner starts with the sequence listed in the
following steps:
1.) Back-End Processor is turned ON and starts to load the software.
2.) The Start Screen (Vivid) is displayed on the monitor.
A start-up progress bar indicating the time used for software loading, is displayed on the monitor, as
shown in Figure 4-3.
3.) The software initiates and sets up the Front-End electronics and the rest of the scanner.
4.) The Keyboard backlight is lit.
5.) As soon as the software has been loaded, a 2D screen is displayed on the monitor, as
shown below.
NOTE: After turning OFF the system, wait at least 10 seconds before turning it on again. The system may not
be able to boot-up if power is recycled too quickly.
• Holding down the On/Off button for 3-5 seconds will evoke the display of the System Exit dialog
box (see Figure 4-5). From this, you can either select Full Shutdown (click the Shutdown button)
or choose to place the system in Standby Mode (click the Standby button). Both these procedures
are described in detail below.
Note: The system will remain in Standby Mode for as long as battery power will allow.
The System - Exit dialog box (Figure 4-5), used when switching OFF the unit, gives you the choices
described below.
• Standby
Use this button to select the Standby mode, allowing a shorter reboot time. Most of the system is
powered down, but a certain portion of the unit remains energized. The standby mode allows a
shorter re-boot time when the system is used on a daily basis or moved from one place to another.
• Logoff
Use this button to log off the current user.
The system remains ON and ready for a new user to log on.
(If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.)
• Shutdown
Use this button to Shut Down the system. The entire system is shut down. It is recommended to
perform a full shutdown at least once a week.
• Cancel
Use this button to exit from the System - Exit dialog box and return to the previous operation.
NOTE: Don’t press the On/Off button too long. If you press the button too long (more than 3-4 seconds), a
full shutdown is started.
1.) Hold down the On/Off button for 3-5 seconds to evoke the display of the System Exit dialog box
(see Figure 4-5 on page 4-7).
2.) Click the Standby button to activate Standby mode.
3.) Leave the mains power cable connected to keep the standby battery fully charged.
Follow steps below if you need to disconnect power cable when unit is in standby mode
1.) Switch OFF the Mains Power from the wall outlet.
2.) Disconnect the power cable.
The system will remain in Standby mode for approximately 15 minutes if the standby battery is fully
charged. If the system is unplugged for a longer period of time, a full shutdown is automatically
performed.
As described above, the battery is automatically charged when the system is plugged to the wall
outlet and the power is switched ON. It takes approximately 10 hours to charge a flat battery.
If the system is left in the OFF condition (system unplugged from the wall out let) for an extended
period of time (3 to 5 days or more), the system may not boot up or may beep when turned ON.
Should this occur, the system needs to recharge the batteries. This could take 15 minutes to as
long as 10 hours, depending on the battery age, system input voltage and system temperature. To
eliminate draining the batteries, the system should remain plugged into the wall outlet with the wall
outlet power switch in the ON position.
NOTE: Connect the power cable to the mains power outlet again as soon as possible and at least before
you switch on the mains power and press the On/Off button.
NOTE: When turning ON a system from standby mode, it takes a few seconds before it responds. Do not push
the On/Off button again during this period. A second push will initiate a full shutdown.
3.) Disconnect the mains power from the wall outlet only after the fans have stopped. This will cut
power distribution within the unit.
If the system is still running when you disconnect the mains power, the internal power supply will
switch over to the battery power and supply the system with power until either the battery is drained
to 30% capacity (invoking shut down), or shutdown has been invoked by the user.
NOTE: To completely remove power from the system, release and remove the battery.
4. Select Log on. - If this is the first time the unit is turned on, the “IMAGING AND
ANALYSIS - GLOBAL LEVEL” window is displayed.
- If another screen was displayed earlier, before logging out or turning
unit off, that screen will be displayed.
4-2-5-1 Introduction
Vivid q N is equipped with a Magneto Optical Disk Drive (MOD Drive) and a DVD/CD-RW Drive (as an
option from system version BT12).
The MOD Drive can be connected to either of the USB ports (upper or lower) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor being connected to
the Vivid q N USB port, and the other side to the rear of the MOD drive.
When connecting the MOD drive directly to the system, it is necessary to use an additional power
source. As a safety precaution, this must be done via an isolation transformer.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed designated to be used with the specific MOD
drive supplied by GE.
WARNING Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive
and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
NOTICE Never move the unit with a MO disk in the Magneto Optical Disk Drive because the drive actuator
will not be locked and the Magneto Optical Disk Drive could break.
NOTICE Do not eject the MO Disk using the button on front of the MO drive.
In some cases, the MO Disk is ejected automatically. If not automatically ejected, use the manual
ejection method as described in the following steps:
2.) Select 5 1/4 MO Disk in the Eject Device menu to eject the MO Disk.
3.) Remove the ejected MO Disk from the MO Drive and place it in its cover box for storage in a safe
place.
One must use the additional power source (supplied with the External USB Hard Disk) to activate the
External USB Hard Disk.
WARNING Do not attempt to use a different type of mass storage device other than the
External USB Hard Disk supplied with the system. The ultrasound system is an
extremely sensitive and complex medical system. Any unauthorized peripherals may
cause system failure or damage!
Refer to 8-7-2-5 "External USB Hard Disk Installation Procedure" on page 8-136 for instructions on
initial configuration and connecting the External USB Hard Disk to the Vivid q N portable ultrasound
scanner.
After connecting the External USB Hard Disk to the system, perform the functional checks described in
the 4-3-2-5 "External USB Hard Disk Test" on page 4-27.
NOTE: Use only CD-R, DVD-R or DVD+R media from a reputable vendor and only high-grade
media types.
1.) Press the Eject button on the right side of the DVD/CD Drive once to open the disk tray.
2.) Put a DVD/CD disk into the disk tray with the label side facing up.
3.) Press the Eject button once to close the disk tray.
After a few seconds, the disk is available for reading, formatting or writing, depending on the type
of disk you used.
NOTICE Do not eject the DVD/CD using the button on front of the DVD/CD Drive.
In some cases, the DVD/CD media is ejected automatically. If not automatically ejected, use the manual
ejection method as described in the following steps:
• Use a logical name and make sure to physically label each Removable Media.
• Verify that each Removable Media is given a unique name. Maximum eleven (11) characters may
be used for the label name.
• Remember that MO Disks have two sides, Side A and Side B. Each Side must have a unique name.
Example: MO Disks for storing images may be labeled “IMAGE_1A” (Side A on MO Disk #1), “IMAGE_1B (Side
B on MO Disk #2), “IMAGE_2A”...
Before you continue, be sure about what naming convention you will use for labelling the Removable Media.
NOTE: Use only CD-R, DVD-R or DVD+R media from a reputable vendor and only high-grade
media types.
NOTICE Do not eject the DVD/CD using the button on the DVD/CD drive.
6.) Press <Alt> + E on the alphanumeric keyboard to display the Eject Device menu (Figure 4-6 on
page 4-13)
7.) Select the relevant media.
The selected media is ejected.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an
MO disk for reloading on the system.
NOTICE Presets should not be saved on the same MO disk as images, because the Archive Menu lists the
images but not the presets stored on a MO disk.
2. Press Config (or F2); select Admin The Backup window is displayed.
then select Backup.
2. Press Config (or F2); select Admin The Restore window is displayed.
then select Restore.
Section 4-3
Functional Checks
4-3-1 Basic Controls
Move the handle from the maximum lowest position to Verify that the handle is able to reach the maximum
5
the maximum upper position. lowest and maximum highest positions smoothly.
Press all Alphanumeric keys one-by-one. As each key is pressed, the corresponding letter is
To quickly test every key (each letter in the alphabet), typed on the screen.
4 type the following sentence:
The quick brown fox jumps over the lazy dog
Repeat the test, while holding down the Shift key.
NOTE: The External Keyboard test also includes testing of the following components: TGC Sliders, Rotary,
Keyboard backlight, Button backlight, and Active Button.
3 Highlight the Keyboard Test and press <Start> The keyboard Test window opens
Press all extended KB keys one-by-one. The extended KB keys activate properly. Every hit on a
5
Press the Clear soft button to clear the Event window. key is followed by two rows (Depress/Release).
e.) Move all TGC sliders to the right position The position value is 63(-2) for all TGC Sliders
f.) Move all TGC sliders to the central position The position value is 31(±2) for all TGC Sliders
7 Perform the ROTARY TEST, as follows: The rotary is activated properly; pressing on the button
Rotate the rotary (in both directions) and press the will display a row of data for pressing and another row of
button at the center of the rotary. data for releasing (Depress/Release).
c.) Set the Keyboard Backlight to 255 All Alphanumeric Backlight LEDs are seen with same
intensity acceptable for the maximum level.
d.) Press the Test Software button All Backlight LEDs except the TGC sliders and the
Rotary are seen with same intensity acceptable for the
maximum level.
d.) Press the Test Software button All 10 Activity green LEDs and KB backlights are seen
with maximal intensity
b.) Place the test upper cover magnet close to the The status of the Lid switch is “Open” and changes to
Alphanumeric Keyboard Escape button. “Closed “after the magnet was placed.
2 Press Config
4 Click on View
From the Paint window, make sure that when moving the Movement is smooth (no jumping) and the trackball has
7 trackball you have good control and that the x and y no “dead” areas (i.e., when the trackball is moved there
coordinate values (displayed at the bottom of the are no areas in which no changes are seen).
screen) move digit-by-digit up and down.
2 Press Config
Make sure the System ID is indicated under the The computer description matches the system ID.
9
computer description.
4-3-1-6 CW Test
4-3-2 Peripherals
NOTE: When the Wireless Network USB Adaptor is installed in a new system, the functionality is verified as
described in "Configuring the Wireless Adaptor" on page 3-93.
NOTE: When the Wireless Network USB Adaptor is installed in a system that is already configured, proceed as
described below.
3.) From the Dataflow Name field scroll-down list (Figure 4-7), select a dataflow that the user works
with in day-to-day operations.
4.) Under Selected devices (Figure 4-7), select the appropriate device.
5.) Click Check.
The Information - Check Status box opens - Figure 4-8.
7.) Under Selected Devices (Figure 4-9), proceed to the next device (in this example “Remote
Database” that is under “Outputs”), click Check and proceed as before.
8.) Continue this procedure until all selected devices have been verified.
NOTE: If the External USB Hard Disk is not currently connected to the system, refer to 8-7-2-5 "External USB
Hard Disk Installation Procedure" on page 8-136 for configuration instructions.
Open the patient file and check that the images are The newly-imported images are displayed correctly on
6
available for review. the Vivid q N scanner.
Lift the release handle located on the left side of the The locking mechanism of the up and down movement
1
SafeLock Cart upwards. is released, allowing the main shelf to be raised.
While lifting the release handle, raise the shelf upwards Ensure that you do not apply too much force to move
2
and make sure it is able to reach its maximum height. the main shelf and that the shelf movement is smooth.
Table 4-17 Brakes Function Test (SafeLock Cart - Front Castor Wheels) - Recommended
Release the wheel lock (upper lever) on each front Ensure that the wheels move freely in all directions.
1 castor wheel by pulling it upwards. Push and pull the unit Check the wheels for wear and tear, and replace if
right, left, backwards and forwards. necessary.
Restart the system performing a full shutdown and During the boot-up process, verify that the messages
1 bootup, as described in "Power ON/Boot-up" on page 4- SCSI Adaptor Installation 2271149-2INS and MO are
3. displayed.
13 Press the Archive button on the keyboard. The Patient List window is displayed.
14 Press the Utilities button on the soft menu. The General Status window is displayed.
Type the new label (for example, “Side1”), then select The message Current Media Name is:SIDE1. Media
16 the Format button. was formatted successfully is displayed.
Confirm “YES” to the warning message. MO Label: SIDE1, MO STATUS: MOUNTED
Press <Patient> button and then select End Exam from The following message is displayed:
4
the archive menu. Unsaved image(s). Which do you want to save?
7 Select Test Patient and press Export. The Export pop-up is displayed.
10 Change the label to TEST-1 and press OK. A progress bar is displayed.
15 Delete the Test Patient that was added for the purpose
The Test Patient is removed from the patient list.
of this test.
16 Place the media into the drive and press the Import. The Export pop-up is displayed.
23 Press OK to confirm delete (must confirm three times). The Test Patient is deleted from patient list.
Turn ON the system’s circuit breaker. The system returns quickly to the Power-ON state -less
3
Press the On/Off Button. than 30 seconds.
Turn OFF the system’s circuit breaker. Turn ON the The system enters Standby mode and stays in this
4
system’s circuit breaker after entering to Standby mode. mode. The On/Off button LED is blinking.
Turn OFF the system’s circuit breaker. The system enters the Power-OFF state. The On/Off
5
Press the On/Off button. button LED is OFF.
Section 4-4
Image Testing
This section provides guidelines for the following image testing checks:
NOTE: For more details, refer to the information provided in the Vivid q N User Manual.
4-4-1-1 Introduction
The 2D Mode is the Vivid q N system’s default mode.
Status Window
Control Menu
4-4-1-2 Preparations
1.) Connect one of the probes.
See "Connecting Probes" on page 3-58 for information about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
WARNING Always use the minimum power required to obtain acceptable images in accordance
with applicable guidelines and policies.
1.) Press 2D on the Operator Panel to access 2D mode.
2.) The following Image Controls are used to optimize the 2D picture.
Verify that all the listed controls are working as intended:
- Use Gain and TGC controls to optimize the overall image together with the Power control.
- Use Depth to adjust the range to be imaged.
- Use Focus to center the focal point(s) around the region of interest.
- Use Frequency (move to higher frequencies) or Frame rate (move to lower frame rate) to
increase resolution in image.
- Use Frequency (move to lower frequency) to increase penetration.
- Use the Reject control to reduce noise in the image.
- Use the DDP control to optimize imaging in the blood flow regions and make a cleaner, less
noisy image.
- Use UD Clarity (Cardiac) or UD Speckle reduce (non-cardiac) to reduce image speckle.
Extra care must be taken to select the optimal Speckle reduction level, as too much filtering
of speckle can mask or obscure desired image detail.
- Use Adaptive reject (Cardiac) to reduce near field haze and blood pol artifact without diluting
tissue appearance of moving structures.
4-4-2-1 Preparations
1.) Connect one of the probes.
See "Connecting Probes" on page 3-58 for information about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
3.) Press MM on the Operator panel to bring up an M-Mode picture on the screen.
4.) Use the trackball to position the cursor over the required area of the image.
4-4-3-1 Introduction
Color Flow screens are 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode, or from M mode, or a combination of these.
4-4-3-2 Preparations
1.) Connect one of the probes to the scanner’s left-most probe connector.
See "Connecting Probes" on page 3-58 for information about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
NOTE: If the trackball control Pointer is selected, press trackball to be able to select between Position
and Size controls.
NOTE: The scale value may affect FPS, Low Velocity Reject, and Sample Volume.
• Adjust Low Velocity Reject to remove low velocity blood flow and tissue movement that reduces
image quality.
• Adjust Variance to detect flow disturbances.
• Adjust Sample volume (SV) to a low setting for better flow resolution, or a higher setting to more
easily locate disturbed flows
• Adjust Frequency to optimize the color flow display. Higher settings improve resolution. Lower
settings improve depth penetration and sensitivity. This does not affect the frequency used for 2D
and M-Mode.
NOTE: The Frequency setting may affect FPS, SV and Low Velocity Reject.
• Adjust Power to obtain an acceptable image using the lowest setting possible.
• Use Invert to reverse the color assignments in the color flow area of the display.
• Use Tissue priority to emphasize either the color flow overlay, or the underlying grey scale tissue
detail.
• Use Baseline to emphasize flow either toward or away from the probe.
• Use Radial and Lateral Averaging to reduce noise in the color flow area.
Radial and Lateral Averaging smooths the image by averaging collected data along the same
horizontal line. An increase of the lateral averaging will reduce noise, but this will also reduce the
lateral resolution.
NOTE: If the trackball control Pointer is selected, press trackball to be able to select between Position
and Size controls.
NOTE: The scale value may affect FPS, Low Velocity Reject, and Sample Volume.
• Adjust Low Velocity Reject to remove low velocity blood flow and tissue movement that reduces
image quality.
• Adjust Variance to detect flow disturbances.
• Adjust Sample volume (SV) to a low setting for better flow resolution, or a higher setting to more
easily locate disturbed flows.
• Adjust Frequency to optimize the color flow display. Higher settings improve resolution. Lower
settings improve depth penetration and sensitivity. This does not affect the frequency used for 2D
and M-Mode.
NOTE: NOTE: Frequency setting may affect FPS, SV and Low Velocity Reject.
Chapter 4 - General Procedures and Functional Checks 4-35
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
• Adjust Power to obtain an acceptable image using the lowest setting possible.
• Use Invert to reverse the color assignments in the color flow area of the display.
• Use Tissue priority to emphasize either the color flow overlay, or the underlying grey scale tissue
detail.
• Use Baseline to emphasize flow either toward or away from the probe.
• Use Radial and Lateral Averaging to reduce noise in the color flow area. Radial and Lateral
Averaging smooths the image by averaging collected data along the same horizontal line. An
increase of the lateral averaging will reduce noise, but this will also reduce the lateral resolution.
4-4-4-1 Introduction
PW and CW Doppler are used to measure velocity (most often in blood).
Doppler mode can be used with a special pencil probe or with an ordinary probe. By using an ordinary
probe, you can first bring up a 2D picture for navigation purposes and then add PW/CW Doppler.
4-4-4-2 Preparations
1.) Connect one of the probes to the scanner.
See "Connecting Probes" on page 3-58 for information about connecting the probes.
2.) Turn ON the scanner
The 2D Mode window is displayed (default mode).
• Adjust Low velocity reject to reduce unwanted low velocity blood flow and tissue movement.
• In PW mode, adjust Sample volume to low setting for better resolution, or higher setting to more
easily locate the disturbed flows.
• Adjust the Compress setting to balance the effect of stronger and weaker echoes and obtain the
desired intensity display.
• Adjust Frequency to optimize flow display. Higher setting will improve resolution and the lower
setting will increase the depth penetration.
• Adjust Frame rate to a higher setting to improve motion detection, or to a lower setting to improve
resolution.
NOTE: Frequency and Frame rate settings may affect the Low Velocity Reject.
• Adjust Power to obtain an acceptable image using the lowest setting possible. This is particularly
important in CW mode, as the energy duty cycle is 100% (constant).
NOTE: The Doppler Power setting affects only Doppler operating modes.
• Adjust the following settings to further optimize the display of the image.
• Use the Horizontal sweep to optimize the sweep speed.
• To view signal detail, adjust Scale to enlarge the vertical spectral Doppler trace.
• Use Invert to reverse the vertical component of the spectral Doppler area of the display.
• Use Angle correction to steer the ultrasound beam to the blood flow to be measured.
4-4-5-1 Introduction
TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by
sampling of tissue Doppler velocity values at discrete points. The information is stored in a combined
format with grey scale imaging during one or several cardiac cycles with high temporal resolution.
4-4-5-2 Preparations
1.) Connect one of the probes, to the scanner’s left-most probe connector.
See "Connecting Probes" on page 3-58 for information about connecting the probes.
2.) Turn ON the scanner
The 2D Mode window is displayed (default mode).
NOTE: PW will be optimized for Tissue Velocities when activated from inside TVI.
1. Press Probe on the Operator Panel. A list of the connected probes will pop up on the screen.
2. If not already selected. Use the trackball to select the desired An application menu for the desired probe is listed on
probe. the screen.
5. Verify there are no EMI/RFI or artifacts specific to the probe. No EMI/RFI or artifacts.
Section 4-5
System Turnover Checklist
Before returning the Vivid q N ultrasound scanner to regular operational use, the System Configuration
settings should be checked. In 2D-Mode, press Config on the keyboard and check the settings in
accordance with those listed or referenced in Table 4-22 below
1 Check the date and time settings. Refer to "Date and Time Adjustments" on page 3-71.
Ensure that all of the software options are configured correctly. Refer to "Software Options
3
Configuration" on page 3-84.
4 Ensure that the hospital details are configured correctly. Refer to Table 3-6 on page 3-70.
Ensure that the VCR is configured as required for the specific site and country. Refer to video
5
converter setup - see Figure 3-45 on page 3-48.
Ensure that the measurement units are configured as required for the specific site and country.Refer
6
to "Units of Measure" on page 3-74.
Ensure that the connectivity network settings have not been modified and are configured as required
7
for the specific site. Refer to "Set the Scanner’s Network Information" on page 3-99.
Section 4-6
Site Log
Chapter 5
Components and Function (Theory)
Section 5-1
Overview
5-1-1 Purpose of Chapter 5
This chapter explains concepts, component arrangement, and sub-system functions for both the
Vivid q N systems. It also describes the power distribution system and probes.
Section 5-2
General Information
The Vivid q N system is an ultra-light handheld ultrasound scanner that can be used with both phased,
linear and curved array ultrasound probes and Doppler (Pedof) probes.
Weighing only 5 kg (11 lb), each system is extremely versatile and - depending upon the installed
software - can be used for various imaging modes, such as:
5-2-1 Introduction
ECG Connector
Pencil Probe Connector
RS Probe Connector
Operator Panel
2 x USB Ports
(not visible in illustration - recommended for
connection of USB storage devices e.g. Memory Stick)
The Vivid q N systems are equipped with a highly efficient power supply, resulting in low power
consumption and minimal acoustical noise. Among other significant features are the following:
• Integrated, low-power Front End (uses advanced ASIC and FGPA technologies).
• Open connectivity using multiple USB ports
• High-performance 15" LCD monitor
• Low profile, backlit keyboard
Vivid q N systems have a digital beam forming system (incorporated in the Front End) that can handle
up to 192 element linear probes using multiplexing.
Signals flow from the Probe Connector Panel, to the Front End (FE) Electronics, to the Back End
Processor (BEP), and finally are displayed on the monitor and peripherals.
In addition, the Vivid q N systems enable acquisition of the following signals: ECG, and external ECG.
The ECG module, which supports acquisition of three bipolar ECG channels
(leads I, II, III), is incorporated in the cabinet cage assembly to provide ECG signals to synchronize
cardiac ultrasound image acquisition. Additionally, other analog inputs from devices such as treadmills
are processed.
Weak ultrasound echoes from blood cells and body structure are received by the probes and routed via
the Probe & MUX Board back to the TR32 Boards. The TR32 Boards amplify these signals, convert
them to a digital signal and do the beam forming.
The vector data streams (one per each MLA) generated by the Beam Former are routed to the
RFI Board for further processing by the Demodulator FPGA and DSP and then the signals are
transferred via the PCI bus to the Back-End Processor where they are converted to video. The video
signals are routed to the internal LCD monitor and to the VGA connector located on the back panel for
connection of external monitor or VCR (via special adaptor).
At power up, all necessary software is loaded from the hard disk.
5-2-2-3 Electronics
Vivid q N internal electronics are divided into three main components:
The main functional blocks in the Vivid q N systems are listed below, and described in detail in this
chapter, as follows:
• Front End (FE) Processor - see Section 5-3 "Front End Unit" on page 5-6
• Back End Processor (BEP) - see Section 5-4 "Back End Processor" on page 5-21
• Internal and external cable interconnections - see Section 5-5 "External Input/Output (I/O)" on page
5-29
• Monitor & Keyboard - see Section 5-7 "Monitor and Operator Panel" on page 5-34
• Power Supply System - see Section 5-6 "Power Supply System" on page 5-30
• Peripherals - see Section 5-10 "Peripherals" on page 5-46
For a detailed block diagram of the Vivid q N system, refer to Figure 5-2 "Vivid q N System Block
Diagram" on page 5-5.
$QDORJ %8))(56
5), %DVHERDUG
%HDPIRUPLQJ 'RSSOHU
'LDJQRVWLF $XGLR $XGLR
0L[HU &LUFXLW '$&
0L[HU
4
86%
&:73* 4'5 7((,) (;7'&
&RQWURO N[ [36
7HPS&RQWURO
&RQQHFWRU
(&*
'&'&3RZHU0RGXOH
[86%
60%XV
%DWWHU\
$&,QSXW $&'&36
Section 5-3
Front End Unit
5-3-1 General Information
The Vivid q N Front End system (see diagram in Figure 5-4 on page 5-8) is divided into two subsystems:
FRONT END
$QDORJ %8))(56
5), %DVHERDUG
%HDPIRUPLQJ 'RSSOHU
'LDJQRVWLF $XGLR $XGLR
0L[HU &LUFXLW '$&
0L[HU
4
86%
&:73* 4'5 7((,) (;7'&
&RQWURO N[ [36
7HPS&RQWURO
&RQQHFWRU
(&*
'&'&3RZHU0RGXOH
[86%
60%XV
%DWWHU\
$&,QSXW $&'&36
Figure 5-3 Vivid q N System Block Diagram showing Location of Front End
$QDORJ %8))(56
5),
%HDPIRUPLQJ 'RSSOHU
'LDJQRVWLF $XGLR
0L[HU &LUFXLW '$&
4
5)'DWD,Q '(02'
)3*$ &ORFN&LUFXLW
$'& 0!0
6\VWHP,&
$'&
(&*
BEP
Front End
Power Supply
RS Probe Connector
ECG Module
Keyboard
Assembly
WARNING When opening and closing the Keyboard Assembly, take care not to damage the cables
connected to the BEP (indicated in Figure 5-6, above).
RFI Board
TR32 Boards
(located below
the RFI Board)
P5 to P&M
J1 to TR32 Boards
J2 to TR32 Boards
J1 To TR32 Boards
J2 To TR32 Boards
5-3-3-7-1 Mechanism
The RFI Board performs the following probe management functions:
• Controls the selection of a probe connector
• Senses the Probe type
• Senses the Scan Plane Angle on TE probes
• Senses the Probe temperature
• If the temperature exceeds the maximum temperature, the RFI stops scanning and disables
Tx voltage.
The temperature references are set to 41.3 oC and 43.0 oC
• If the temperature reading falls below minimum temperature, the RFI stops scanning and disables
Tx voltage.
The minimum temperature references are set to 15 oC. A temperature reading as low as 15 oC
indicates a probe temperature-sensing error.
• 32 Transmitters
• 32 receiving channels, T/R switch, Pre-Amp, TGC
• 32 channel Beam Former
• First stage of Analog Beam-Former for CW mode, included in the receive input ASIC
• Clock distribution, starting with input at PECL level
P1
P3
P2
J1
J3
J2
P2 To RFI Board
P3 To P&M Board
Each TR32 Board has 32 channels and includes a 32-channel Digital Beam Former (consisting of one
FOC32 ASIC).
Inputs are 32 probe lines that have gone through aperture selection, performed in the P&M Board.
Outputs are MLA0 and MLA1 digital buses from the Beam Former and CW output lines to the
Analog CW BF.
The MLA output bus of TR32B is connected to the MLA input bus of TR32A, giving a 64-channel BF on
the MLA output.
All required control, memory and clocks for the two TR32 Boards reside in the RFI Board.
The two TR32 Boards are linked together in parallel. The way in which they operate depends upon the
probe currently in use, and also on the current type of Transmit/Receive mode.
Transmit/Receive modes can be divided into two types, as shown in the following examples:
• CW Mode - half the channels transmit signals, and half the channels receive
signals, simultaneously.
• 2D Mode - signals are propagated between the channels; there is a specific
time delay between transmit and receive - data is then interpolated
in accordance with the delay, and the image is created.
The TR32 Boards are connected to the RFI Board by stacked connectors, as shown Figure 5-13 below.
RFI
Connectors
TR32 1 & 2
Stacked
Connectors
J3
J5
J1
Pencil Probe
Connector
J2
RS Probe
Connector
P3
Not connected!
J2 RS Probe Connector
Section 5-4
Back End Processor
5-4-1 Introduction
The Back End Processor (BEP) supports the operation of the Vivid q N ultrasound unit and is the main
controller for the system.
A block diagram of the BEP is provided in Figure 5-17 "Vivid q N Back End Processor - Block Diagram"
on page 5-22.
The various components that comprise the Vivid q N BEP are described in the following sections:
%DFN(QG
(7;6%& /&'
´ /&'
0RQLWRU
´
%DVHERDUG
3&0&,$
3&,
(&* 86%
[86%3RUWV
For reference, a list of abbreviations used in this section is provided in Table 5-8 below.
Table 5-8 Abbreviations and Definitions
Abbreviation Definition
BGA Ball Grid Array
CS Component Side
PS Power Supply
The CPU utilized in the Vivid q N ultrasound unit is a Pentium -M, 1.4 GHz Processor.
• HDD
• Front End (RFI) Board
• USB ports
• Docking port, speaker phones and speaker jacks
• Keyboard
• Battery
• VGA
• Fans
• Power supply
The Power Management Controller (PMC), mounted on the ETX Base Board, ensures automatic
transition between the various power modes, stable DC condition, and provides power status LED display.
The Docking port, located on the rear panel of the ETX Base Board, provides the connections to the
Audio unit on the DIB, discrete control and status lines and 20V-DC power.
Refer to Figure 5-21 on page 5-27 for a view of the ETX Base Board showing the location of the various
connectors.
Figure 5-18 below shows a block diagram of the high level functions of the ETX Base Board.
• C: Drive: Execution of the operating system and the software application, and factory presets.
The partition size is 5 GB.
• D: Drive: User presets and data. The partition size is 5GB.
• E: Drive: Archiving of all images and reports. The partition size is the balance of available space
on the hard disk.
• F: Drive: Swap file partition. The partition size is 1GB.
The following additional partitions are hidden:
• Grub: 32MB
• Repository:10GB
NOTE: When connected to the system, a DVD device or disk-on-key will not necessarily be shown as
G drive.
Each Fan has 3 wires: two 12V wires that supply power to the fans; a 3rd wire for sensing the fans’
speed.
Fan Board
Internal Battery
KBD
LAN
Docking Connector
VGA
Upper View
USB Port
PCMCIA
(Not in use)
Lower View
J44 J13
J7
J19
J14
J18
J17
J16 J5
Table 5-9 BEP Connectors
Connector # Description
J7 VGA
J18 DOCKING
J20 Speakerphones
J16 RFI
P1:P4* ETX
J5 LCD PANEL
J6 KEYBOARD
J17 PS UNIT
J15 BATTERY
J13 CARDBUS
J8 Protection Key
J11:J12 FANS
J9 Speakers
Section 5-5
External Input/Output (I/O)
The Vivid q N system rear panel provides a few standard connectors for connecting the following
external devices:
1 2 3 4 5
6 7
Legend:
1. Two interchangeable USB ports (digital printer, DVD-RW/CD-RW and other peripherals).
2. Docking connector.
3. Port for DC IN (AC Adapter).
4. SVGA Output (VCR option or CRT monitor option).
5. LAN 10/100 Base-TX Ethernet network connector
6. PCMCIA port for PC card.
7. Ejection lever for PCMCIA device.
Section 5-6
Power Supply System
5-6-1 Electrical Power
The Vivid q N systems can operate at between 100 and 240 V at 50/60 Hz via an
AC/DC converter which supplies a 20 V DC output and 4.5 amps.
5-6-2 Introduction
The power system comprises two main building blocks:
• AC/DC Adaptor - an external unit connected to the system's rear panel (see Figure 5-22).
• DC/DC Power Supply Unit - located inside and mounted to the bottom assembly, provides the
system with both Low Voltage and High Voltage power, and also enables recharging of the battery.
- The Power Supply Unit controls the following main functions:
* Input switching stage - selects between the external DC source and the internal battery
* Back End Low Voltage power supply (LVPS)
* Standby power supply
* Front End Low Voltage power supply (LVPS)
* Front End High Voltage power supply (HVPS)
* Battery charger controlled by the Power Management controller
* Control circuitry (used for monitoring, charger control and auxiliary functions).
* The PS communicates with the host CPU via the SMBus
• Rechargeable Battery
• PS Controller and Monitor
• DC Source Selector
• Battery Charger
• High Voltage Power Supply (HVPS)
• Low Voltage Power Supply (LVPS)
J17
J14
J17 To BEP
The Vivid q N scanner has built-in charger functionality (see Figure 5-23 on page 5-31) and switches
automatically from battery operation to AC operation and vice versa.
When shutting down the system, leave the main power cable connected to keep the battery fully charged.
NOTE: The lithium ion technology used in the system's battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries.
CAUTION The battery is designed to work with Vivid q N systems only. Only use the batteries authorized
by GE.
WARNING Used batteries should not be placed with common household waste products. Contact
your local authority for the location of a chemical waste collection program nearest to
you.
5-32 Section 5-6 - Power Supply System
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
CAUTION If the Vivid q N portable ultrasound system is not being used for periods of one month or longer, the
battery should be removed during lengthy non-use period.
CAUTION To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,
observe the following precautions:
• Do not immerse the battery in water or allow it to get wet.
• Do not put the battery into a microwave oven or pressurized container. If the battery leaks or emits
an odor, remove it from all possible flammable sources.
• If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
• If you have any questions about the battery, consult GE or your local representative.
Short term (less than one month): 0° C (32 °F) to 50 °C (122 °F)
Long term (more than three months): 10° C (50 °F) to 35 °C (95 °F).
Section 5-7
Monitor and Operator Panel
5-7-1 Keyboard and Operator Panel
Figure 5-25 below illustrates the layout of the keyboard and operator panel. The buttons and controls
are grouped together for ease of use; the different areas of the keyboard and operator panel include the
following:
The Vivid q N keyboard unit comprise the main user-interactive devices: an alphanumeric keyboard, a
1.5" trackball, and an extended keyboard. In addition, it includes one or more printed circuit boards
(PCBs) designed to carry the various electronic components - these include the keyboards’ controller,
TGC potentiometers, USB Hub, and backlight illumination.
Figure 5-26 below shows a high-level functional block diagram of the keyboard unit. The individual
components are described in the Keyboard and Operator Panel Components section, page 5-35.
The alphanumeric keyboard has a PS/2 interface and supports all standard alpha-numeric functionality,
such as, auto-repeat, standard ASCII output codes, etc.
Specially designed, high-quality key caps - together with controllable backlighting - provide a keyboard
that is comfortable and easy to use.
The upper row of functional keys (F1 - F11), are clearly identified as being assigned to specific
ultrasound functions.
These keys, that slide horizontally, are used to compensate for depth-related attenuation in an image.
5-7-2-6 Trackball
The Vivid q N operator panel has a 1.5" trackball centrally located above the hand-rest area to provide
easy operation.
The trackball has a PS/2 interface and supports a variety of basic PC operations and specific ultrasound
controls.
• Keyboard controller
• Two general-purpose ports
(current up to 600 mA per port, over-current limit should be used).
1.) 3V standby for driving the power On/Off and status LEDs.
2.) 5V BE (Back End) is used for the remaining circuits.
The keyboard has current-limiting circuitry (implemented on the BEP), to protect it against excessive
current consumption (via the USB and/or short circuits).
5-7-2-10 Keyboard ID
The keyboard PCB contains an EEPROM which stores a variety of information about the keyboard such
as, GE part number, manufacturing date, and revision # of the PCB hardware.
The display monitor is a 15" LCD and is connected via a flex cable to the BEP, as shown in Figure 5-28.
Monitor
(LCD Display)
Flex Cable
BEP
Section 5-8
ECG Module
5-8-1 Overview
The Vivid q N ECG Patient I/O module is designed to enable acquisition of the following signals: ECG
and external ECG. The different pins on the ECG input connector are specifically assigned to support
either External ECG or Internal ECG:
• The Internal ECG module supports acquisition of two bipolar ECG channels: leads I, II.
• The External ECG module supports acquisition of a single bipolar ECG input channel ranging
from -1 to +1 volt allowing amplification of 1000. The minimal allowed impedance is 10 M ohm.
The ECG module is located under the keyboard assembly (see Figure 5-6 on page 5-9). All power and
communication is supplied to the ECG module via the USB hub located in the keyboard module as
shown in Figure 5-29 below.
Isolated Grounded
Part Part
Front Rear
A general block diagram of the ECG Patient I/O Module is shown in Figure 5-30 below.
Estimation: 80mA for ECG function and 50mA when working with EXT-ECG function.
5-8-2-3 Bandwidth
The ECG Bandwidth is 0.05 to 150Hz.
5-8-3 Isolation
The ECG signals are isolated from the Vivid q N system to comply with IEC601-1 type CF device
classification. All leakage and auxiliary currents apply the values listed in Table 5-12 below (taken from
table IV in IEC601-1, section 19.3):
Table 5-12 Leakage and Auxiliary Current Values
The ECG signals have double, reinforced isolation (B-a) from live parts. Hence, creepage distance and
air clearance in the module are 8 mm and 5 mm respectively, to withstand 4000 volts isolation between
applied part and grounded part.
The applied parts and enclosure have basic isolation (B-a).Therefore, creepage distance and air
clearance in the module are 4 mm and 2.5 mm respectively, to withstand 1500 volts isolation between
applied parts and enclosure.
Section 5-9
ECG/Respiratory Module
The Internal ECG/Respiratory Board enables ECG, Respiratory and External ECG functionality, as
explained in Section 5-9-1, below.
The installation of the Internal ECG/Respiratory Board is prerequisite for using the External ECG/
Respiratory Interface, as explained in "Internal ECG/Respiratory Board" on page 5-41.
5-9-1-1 Overview
The Vivid q N Internal ECG/Respiratory module is designed to enable acquisition of the following
signals: ECG, Respiratory and external ECG.
The Internal ECG/Respiratory Board is located in the cabinet cage assembly (refer to Figure 5-6 on
page 5-9). All power and communication is supplied to the ECG module via the USB hub located in the
keyboard module as shown in Figure 5-31, below.
Connections to and from the Internal ECG/Respiratory Board directly to the patient are illustrated in
Figure 5-32, below.
Internal
ECG/Respiratory P atient
Board
E C G & R es p
Ultras ound
S ys tem
Figure 5-32 Connection between the Internal ECG/Respiratory Board and the Patient
5-9-1-2 Specifications
Characteristic Specification
Characteristic Specification
Characteristic Specification
Characteristic Specification
5-9-2-1 Overview
Connections to the external monitor are illustrated in Figure 5-33, below.
There are three optional connections for the External Respiratory interface (refer to Figure 5-34),
depending on the Respiratory Monitor output type being used, as follows:
IN+ 1
2 6
7
3 5
4 IN-
GND
Figure 5-34 Pin Layout for External Respiratory Interface Connector
5-9-2-2 Specifications
Characteristic Specification
5-9-2-3 Connectors
6
3 4
1
2 5
J5
J3 J4
Section 5-10
Peripherals
5-10-1 Introduction
The following peripherals are used with the Vivid q N portable ultrasound scanner:
5-10-1-1 Printers
The Wireless Network Adapter provides a mobile network connection to the local area network. The
Wireless Network Adapter is supported via the USB port.
Section 5-11
Cooling System
5-11-1 General Information
The Vivid q N cooling system comprises five fans, which are directly controlled through the FAN
controller module, by the monitoring application. The fans withdraw air from the system through the side
vents, removing the warm air from around the internal components, as shown in Figure 5-35 below.
The structure of Vivid q N internal components has been designed to provide the best thermal
conduction, allowing the maximum amount of heat to be removed from the system. It is crucial that the
system ventilation access is not blocked; a distance of at least 20 cm from right and left sides of the
system should be kept clear. In addition, operating the system in an unclean environment could
introduce foreign objects into the unit and create blockage of the airflow, which may eventually cause
irreparable damage to the system. There are therefore some minimal environmental conditions that
should be provided for optimum operation of the system - for details and recommendations, refer to
Table 2-2 on page 2-2.
Chapter 6
Service Adjustments
Section 6-1
Overview
6-1-1 Purpose of Chapter 6
This section provides references to the procedures required for testing and adjusting the Vivid q N portable
ultrasound scanner. These tests are optional, and may be used to check the system for errors.
Important: LCD Display adjustments should be made in the correct order, as stated in the following
procedures.
NOTE: Incorrect calibration of the Vivid q N monitor may also result in less than optimal image quality
when reviewing images on a DICOM or EchoPAC station.
Access the Calibration Screen (by pressing <ALT>+<Q>), then follow the relevant procedure
instructions described below.
.
NOTE: The Hot keys used to exit the Calibration Screen are: <ALT>+<Q>.
NOTE: It is highly recommended to set the contrast to maximum. Press <ALT>+<Steer Right> to
increase the number of green bars to 7 (maximum).
To assist in making Brightness adjustments, an adjustment bar is displayed below the screen:
NOTE: The screen Brightness is calibrated when the dark blocks on the left (numbers 1 through 6) -
see Figure 6-1 - and the white bars on the right (numbers 1 through 12) are all just visible.
V Watermark
6-2 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
The Test Pattern in the Calibration Screen (refer to Figure 6-1 on page 6-1) may be used for Advanced
Settings. (Press Alt + Q to access the Test Pattern).
• To adjust the Blue Tint: hold down <ALT>, then use the < and > keys, as required.
Repeatedly pressing < will decrease the blue tint; repeatedly pressing > will increase the blue tint.
To assist in making blue tint adjustments, an adjustment bar is displayed below the screen:
NOTE: Systems with BT11 installed software have a Blue Tint “2” option that enables blue tint
enhancement. This option can be switched off if preferred by the user.
System: Vivid i n
Monitor: BOE HYDIS-1
NOTE: On older Vivid i/q systems where the IDTECH-1 monitor is supplied, the Blue Tint “2” option is
not applicable.
NOTE: The Blue Tint “2” option is incompatible with IDTECH-1/2 monitors, therefore, on these
systems, the Blue Tint “2” option must be turned off.
6-4 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
NOTE: The default blue tint test pattern is set at two bars - see Figure 6-7.
NOTE: To assist in making blue tint adjustments, an adjustment bar is displayed below the screen -
see Figure 6-8.
2.) Under LCD Setup, de-select Blue Tint “2” to turn off this option.
6-6 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
An external monitor can be connected to the Vivid q N system to serve as a secondary external monitor.
Connect the secondary monitor to the Vivid q N system via the VGA port and proceed as described in
the following scenarios:
A.) Scenario A
The user is able to make separate adjustments to the External Monitor, using the controls on the
External Monitor itself:
1.) Using the Vivid q N controls, press <Alt> + X and make sure that the Internal option is currently
selected.
2.) Calibrate the Vivid q N monitor as described in sections 6-1-2-1 and 6-1-2-2 above.
3.) After calibrating the Vivid q N monitor, calibrate the External Monitor using the External Monitor’s
controls.
B.) Scenario B
The user prefers to use the External Monitor as the Primary monitor, but does not have access
to it and therefore cannot make separate adjustments to it using the controls on the External
Monitor itself.
Moreover, the calibration preset currently in use on the Vivid q N system is not optimal for the
External Monitor.
In this scenario, adjustments (Brightness only) can be made to the External Monitor, using the
calibration controls on the Vivid q N system, as described below:
1.) Using the Vivid q N controls, choose the External Monitor to be the monitor that will serve as the
primary monitor to be calibrated, as follows:
Press <Alt> + X and select the External option.
2.) Using the Vivid q N controls, calibrate the External Monitor by making the required Brightness
adjustments:
To adjust the Brightness: hold down <ALT>:
NOTE: On completion of Brightness adjustment, the External Monitor calibration is now preset; only
this monitor (the primary monitor) will be optimized in this calibration process.
The VGA output of the Vivid q N is a non-isolated output, therefore it may only be connected directly to
a medical device.
1.) Purchase a standard VGA isolated splitter. Do not use the GE VGA isolated splitter
(P/N 5390022) as this has a special pin configuration (dedicated for use with CardioLab/MacLab/
Carto).
2.) Connect the device to an isolation transformer. Ensure that the device is not connected to another
non-medical device which can cause isolation failure through the third device.
3.) Power the second device via a medical grade power supply. Ensure that the device is not
connected to another non-medical device which can cause isolation failure through the third device.
6-8 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Vivid q N System
Next morning Battery calibrated
ready for use
NOTE: For step-by-step instructions, see "Performing the Battery Calibration Procedure" on page 6-11.
6-10 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
NOTE: It is recommended to commence this procedure towards the end of the working day (when the
battery is fully charged) then allow the procedure to be completed overnight, as outlined in
Figure 6-10.
3.) Wait for the confirmation window to be displayed showing that the battery is fully charged.
(See Battery Calibration Wizard Page 2 in Figure 6-12).
6-12 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
5.) Disconnect the AC power cable from the Vivid q N, then click Next.
The second phase of the process (Full Discharge) commences and the Battery Calibration Wizard
(Page 3) is displayed as shown in Figure 6-14 on page 6-14.
6.) Leave the system overnight to complete the Full Discharge process.
The Vivid q N system automatically shuts down at the end of this process; the battery calibration
procedure is now complete.
7.) The following morning, re-connect the AC power cable and turn ON power to the system.
NOTE: The system may now be used normally, without any dependence on battery capacity status.
6-14 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 7
Diagnostics/Troubleshooting
Section 7-1
Overview
7-1-1 Purpose of Chapter
This chapter describes how to set up and run diagnostic tools to locate system problems and failures.
The Vivid q N ultrasound unit includes built-in diagnostic tools that can be accessed from the scanner
application.
Section 7-2
Diagnostics
7-2-1 Diagnostic Tools
The diagnostic tools are provided for checking the system which includes the power supply,
temperature, fan operation, board functions, Back-end signal processing performance, keyboard
operation, peripherals, and so on.
• When the system boots up, it loads all the required drivers and establishes communication with the
Front Board, via the PCI bus. Check the system presets to verify that good unit performance can
be achieved.
• When the Back End is operating, you can verify proper voltages and temperatures in the Front End
and in the Power Supply. To check the Front End parts, the tests must follow a logical sequence,
in order to identify the faulty component. For example, testing communication between the Front
End and the Back End begins by testing the RFI board and the PCI bus, before testing all the other
boards, in the following sequence (as applicable):
* RFI board, TR32 boards, Probe and MUX board
• Many error messages are displayed due to minor changes in calibration that may occur during normal
operation. Before performing diagnostic tests, perform calibration, as described in the DC Offset
Calibration section, page 7-25. If error messages are displayed after performing calibration, perform
the Front End diagnostics described in "Performing Front End (FE) Diagnostics" on page 7-24.
Diagnostic
Test
Tree
Data Flow
Map
Current Test
Controls &
Special
Instructions
Status Area
NOTICE IMPORTANT - The user-friendly Diagnostic Test Window displays dynamically-updated information.
The Data Flow Map area on the right is designed to provide an instant visual representation of the system
components; the color illumination of graphic buttons (light blue; light gray) and status indicators (green;
red; yellow) indicates which component, or set of components, is currently undergoing diagnostic testing
(see the examples in Figure 7-2 on page 7-4).
In this way, it is easy to visually follow the test sequence and progress of a specific test.
For a detailed explanation of the Diagnostic Test Window, refer to Table 7-2 on page 7-5.
Figure 7-2 Data Flow Map - Graphic Color Illumination of Test-Specific Components and Status Indicators
Full System Test (Front End and Back End) RX Test (TR32A and TR32B Boards)
NOTE: For a summary of the various Diagnostic Test Menu options, together with details of the
information displayed in different areas of the screen, see Table 7-2 on page 7-4 and Diagnostic
Test Window Menu Options section, on page 7-10.
NOTE: When performing a specific test for an intermittent problem, it is highly recommended to use the
Loop option (use the combo box or type the required number of loops) as appropriate. Looping
the test will assist in detecting the problem.
NOTE: Once the test has been completed, a log can be viewed, as described in the View Log section,
on page 7-6.
The Diagnostics Test Window (Figure 7-1) comprises the following areas and menu options:
Table 7-2 Vivid q N Diagnostics Test Window and Menu Options
Data Flow Displays a schematic view of the In the Data Flow map Status Window, status
indicators are highlighted by color, indicating the
Map Vivid q N data flow (Data Flow map).
following:
The functionality of each button in this area
• Light Blue: The selected test will be performed
corresponds to that of the test of the same name on this component, or group of components.
in the Diagnostic Test Tree.
• Light Gray: This component, or group of
You are able to select and perform the components will be excluded from the test/s.
Vivid q N diagnostic tests by trackballing to the • Green: All recommended tests for these
required test (listed in the Diagnostic Test Tree) components have been completed and no
and pressing Set. problems were identified.
• Red: At least one problem was identified in these
Note: Double-clicking a button in the Data Flow
components during the performance of the test.
Map expands the corresponding list of tests (in
• Yellow: The tests executed so far on these
the Diagnostic Test Tree) available for that
components have passed, but not all
component (or group of components). recommended tests have been performed.
Controls and Displays any special instructions for checks The area also includes these controls:
Special required before performing the test. • a Start/Stop button used to commence a test.
Instructions For example, Make sure system is calibrated. After pressing Start, the button may be used to
stop a test before completion.
Disconnect probe and press Start.
• a Pause button used to pause a test before
resuming.
• a Clear History button used to clear the
highlighted status indication in the Data Flow
map (schematic view).
Note: this button is not used to clear the test
sequence log files (refer to Clear Log button in
Table 7-3 on page 7-6).
• an On Fail combo box - used to select the
action required in the event that the test fails
(Stop or Continue).
• a Loops checkbox. If activated, the test will be
repeated for the selected number of loops.
Status Area Displays instructions during the test, as well as The following messages are displayed:
test status and result messages. • Test status: For example, Ready for test,
Executing, or Finished.
• Instructions: During the course of a diagnostic
test, any instructions to the user are displayed
in the Status area. For example, Wait.
• Test results: For example, Passed or Failed.
The area also includes two progress bars:
• Progress bar: Shows progress of an individual
test (within a sequence of tests).
The area also includes:
• a View Log button used to review the test
results. (Enabled only after test execution).
Refer to View Log section, on page 7-6.
Loop Lists name of loop and the sub-tests within the loop
Reports Displays the specific report (or list of reports) for the Double-clicking on a currently-selected
currently-selected test sequence. report displays that report in the Report
(shows all OK and failed tests) window (lower portion of the View Log
Certain tests will have more than one report (see the dialog box) - refer to Figure 7-4 below.
example in Figure 7-3 below).
Exit Button Closes the View Log dialog box Returns to the Diagnostic Test window.
Lower portion of the View Log dialog box where the Visible only after test execution.
currently-selected report is displayed in a separate window
Report Window (see the example in Figure 7-3).
The displayed Average, Deviation, Max and Min values
are dynamically updated.
In Figure 7-3 above, the selected report (Test signal amplitude, first harmonic) is displayed in the Report
window. When there is more than one report listed, the first report is displayed (by default).
Double-clicking on another report (for example, Test signal amplitude second harmonic), will evoke the
display of that report in the Report window as shown in the example in Figure 7-4 below.
Figure 7-4 ADC Test - Viewing Report (Test Signal Amplitude 2nd Harmonic)
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also
evokes the display of the following information relative to that specific point: Average, Deviation
(standard deviation from the Average), Max, and Min.
The signal should be within the limits marked in the white area (the dotted line represents the
Average), as shown in Figure 7-4.
NOTE: The Full System Test requires approximately 10 minutes for completion.
For test procedure instructions, see the Full System Test (FE and BEP) section, on page 7-20.
NOTE: The Front End Tests may be performed in any order. However, due to the functional
dependencies between each of the Front End boards, it is recommended to perform the tests
on the boards in the following order (as described in the Performing Front End (FE) Diagnostics
section, on page 7-24):
7-2-4-2-1 Calibration
• DC Offset Calibration
This is used to test proper operation of the Beamformer, and TR32A and TR32B boards. The test
generates no signal at the input to the Beamformer and is used to calibrate the system, or to test
system calibration in order to check that the current calibration values are valid for the current
environment.
For test procedure instructions, see the DC Offset Calibration section, page 7-25.
• DC Offset Verification
This test verifies that the calibration saved on the Hard Disk is within acceptable limits.
For test procedure instructions, see the DC Offset Verification section, page 7-28.
NOTE: To conduct a complete audio test of the system, it is also necessary to perform an audio test
which is part of the Back End - for details, refer to the PC Doctor section, page 7-15.
Therefore, in the first phase of the test, the Beamformer on the TR32A board generates a test
pattern signal (while TR32B generates pattern zero).
In the second phase of the test, the Beamformer on the TR32A board generates a test pattern
zero - while the BF on the TR32B board generates a test pattern signal.
Both the test patterns are summed and focused on the BF 32A and generated to the RFI
demodulator.
For test procedure instructions, see the TR32 BF Digital Test section, page 7-46.
For test procedure instructions, see the System BF Digital Test section, page 7-49
- ADC Test
The ADC Test is used to check that 1st harmonic, 2nd harmonic, 3rd harmonic, and
background noise are within pre-defined limits.
In this test, a signal is generated from the RFI into the TR switches. For all 64 channels of the
system, each signal goes through the following: pre-amplification, filtering, analog-to-digital
conversion, and focusing. The test signal is at a frequency of 2 MHz.
For test procedure instructions, see the ADC Test section, page 7-52.
- RX Test
The RX Test is performed with no probe connected. This is used to check that the signal
values at the input to the TR32 board are as expected.
When performing the RX Test, test signals are generated from the RFI, and the results are
expected to be of fairly high amplitudes with no distortion of the frequency.
For test procedure instructions, see the RX Test section, page 7-57.
7-2-4-2-4 PM Tests
• ACW Test
For test procedure instructions, see the ACW IQ Symmetry Test section, page 7-61.
For test procedure instructions, see the ACW Matrix Test section, page 7-64.
For test procedure instructions, see the ACW Delay Range Test section, page 7-68.
For test procedure instructions, see the ACW Pre-Mixer Gain Test section, page 7-72.
For test procedure instructions, see the ACW BPF Test section, page 7-75.
For test procedure instructions, see the ACW HPF Test section, page 7-80.
• Pencil Test
- Pencil Rx Test
When performing the Pencil Rx Test, no probes or jig are connected to the system.
This test is performed to check the receive functions. During this test, signals are transmitted
through the circuitry. The received frequency should match the transmitted test frequency, as
shown in the Test report.
For test procedure instructions, see the Pencil Rx Test section, page 7-83.
For test procedure instructions, see the Full System Monitoring Test section, page 7-86.
For test procedure instructions, see the Fixed Voltages Test section, page 7-93.
For test procedure instructions, see the Temperatures Test section, page 7-96.
For test procedure instructions, see the Fans Speed Test section, page 7-99.
• Battery capacity
• Charging voltage being supplied to the battery (or the voltage being supplied by the battery)
• Source of the power being provided to the system (i.e. external power via the AC/DC or via the
battery [indicating the battery is either being charged or discharged]).
For test procedure instructions, see the Battery Parameters Test section, page 7-102.
For test procedure instructions, see the Tx Voltages Test section, page 7-105.
For test procedure instructions, see the Extended Keyboard Test section, page 7-108.
NOTE: For instructions on performing the alphanumeric keyboard test, refer to "Alphanumeric
Keyboard Test" on page 4-19.
For test procedure instructions, see the ECG Test section, page 7-112.
7-2-4-4-3 PC Doctor
The PC Doctor Tests are divided into two main groups: Automatic and Interactive. The automatic tests
do not require user intervention. The procedure instructions for both groups are provided in the PC
Doctor section, page 7-115.
• Automatic
For test procedure instructions, see the Full Back End Test w/o Hard Disk Scan section, page
7-115.
For test procedure instructions, see the Full PC Doctor Test (50 min) section, page 7-117.
- CPU Tests
The CPU Tests are used to perform a variety of CPU checks, as described below.
For test procedure instructions, see the CPU Tests section, page 7-119
• CPU Full Test: Performs a battery of tests on the CPU to ensure that it is working
properly. All the remaining tests are performed as part of the CPU Test.
• CPU Registers: Tests the CPU Registers using the default test patterns. On 32 bit CPUs,
the extended registers are also checked. If errors are detected, the failed registers are
listed.
• CPU Arithmetics: Tests the proper function of the arithmetic commands ADC, ADD,
DEC, DIV, IDIV, IMUL, INC, MUL, SBB and SUB with 16 and 32 bit operands. If errors
are detected, the failed commands and operands are listed.
• CPU String Operations: Tests the proper operation of the string commands LODS,
MOVS, SCAS and STOS with 8, 16 and 32 bit operands. If errors are detected, the failed
commands and operands are listed.
• CPU Logical Operations: Tests the proper operation of the logical operations AND,
NOT, OR and XOR with 16 and 32 bit operands. If errors are detected, the failed
operands are listed.
• CPU Math Registers: Tests the coprocessor register stack, pointers and commands
FLDLZ, FLDPI, FLDLN2, FLDLG2M FLDL2T FLDL2E and FLD1. If errors are detected,
the failed steps and commands are listed.
• CPU Math Commands: Tests the coprocessor commands FBLD/FBSTP, FLD/FST,
FXCH, FCOM, FICOM, FTST, FXAM and FUCOM. If errors are detected, the failed
instructions are listed.
• CPU Transcendental: Tests the coprocessor transcendental commands F2XM1,
FPATAN, FPTAN, FYL2XP1, FCOS, FSIN and FSINCOS. If errors are detected, the
failed commands are listed.
• CPU MMX Arithmetic: Tests the MXX commands PADD, PSUB, PMUL and PMADD. If
errors are detected, the failed commands are listed.
• CPU MMX Logical: Tests the MXX commands PAND, PANDN, POR and PXOR. If errors
are detected, the failed commands are listed.
• CPU MMX Shift: Tests the MXX commands PSLL and PSRL. If errors are detected, the
failed commands are listed.
• CPU MMX Data Transfer: Tests the MXX commands MOVD and MOVQ. If errors are
detected, the failed commands are listed.
• CPU MMX Misc.: Tests the MXX commands PCMPEQ, PCMPGT, PACKSS, PACKUS
and PUNPCK. If errors are detected, the failed commands are listed.
- HD Test
The HD Tests are used to perform a variety of Hard Drive checks, as described below.
For test procedure instructions, see the HD Test section, page 7-122.
• Hard Drive Full Test (w/o Scan): Performs a battery of tests to examine the hard drive
and ensure that it can perform all its functions. All the remaining tests are performed on
the hard drive to ensure that the hard disk controller and the drive mechanism are working
correctly. The disk surface itself is also checked. This test may take 3.5 minutes.
• Hard Drive Linear Seek: Moves the hard disk drive heads from track 0 to the maximum
track one track at a time.
• Hard Drive Funnel Seek: Moves the hard disk drive heads in a funnel fashion (from the
first track to the last, then to the second track, the second last track, the third track and so
on).
• Hard Drive Surface Scan: Scans the surface for defects on the hard disk drive.
- Memory Test
The Memory Tests are used to perform a variety of memory checks, as described below.
For test procedure instructions, see the Memory Test section, page 7-124.
• Full Memory Test: Uses 18 test patterns to test memory locations. If errors are detected,
the address of the block in which the error was detected is shown. During testing, these
diagnostics will also look for parity errors and other exceptions. If encountered, these are
added to the test log as errors.
• Memory Pattern Test: Uses several test patterns to test as much memory as possible.
• Memory Parity Test: Checks for parity errors on the memory bus during intensive string
transfer operation. If an error is detected, the address of the failed block is displayed.
- Video Test
The Video Tests are used to perform a variety of video diagnostic tests, as described below.
These tests use 18 test patterns to fill the video buffer, testing your graphics acceleration and
text output
For test procedure instructions, see the Video Test section, page 7-126.
• Full Video Test: Test the system’s video capabilities. This involves testing the memory,
testing the graphics acceleration and testing the text output. You can follow the tests
being performed on the monitor. This test can be cancelled at any time by pressing <Esc>
on the keyboard.
• Video Memory: Tests the video memory by filling the video buffer with 18 test patterns,
one pattern at a time. These tests fill the entire screen with a single color.
• Video Data Transfer: Tests the graphics acceleration part of the video controller. This
test fills the screen with black and white concentric squares, and rectangles of various
sizes and colors. If errors are detected, the locations of the problems are displayed.
• Video Text Output: Prints a text string in random sizes and colors to test the video device
driver and video controller.
- PCI Test
A Peripheral Component Interconnect (PCI) bus is a fast standard bus that is common in
Pentium systems. The PCI Test is performed to ensure that the PCI bus is working up to
speed. The motherboard is scanned to verify the configuration space and to ensure that the
diagnostics can communicate with the board.
For test procedure instructions, see the PCI Test section, page 7-128.
- Network Test
A Network Interface Card (NIC) is a computer circuit board or card that is installed in a
computer so that it can be connected to a network. Networks are used by businesses and
some home users to connect PCs together so that they can share files and resources. Data
travels over a network connection much faster than over standard internet connections.
The PC Doctor Network Test will reveal that a network card is present without a driver
installed. These cards are likely to not show up on a PC's list of network cards available for
test. This may cause the test to return a result of “not applicable”, meaning no network cards
were found. The test may detect other issues such as electrical interference (noise), circuitry
damage, loose connectors, and crimped or bad network cables.
Chapter 7 - Diagnostics/Troubleshooting 7-17
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
The Network Test will test DSL or cable internet connections, which operate within a network
environment. However, this test does not test the DSL or cable modem used for accessing
the Internet. Currently, no PC-Doctor test exists for testing cable or DSL modems.
The Network Communications Test checks if your PC's network card works and if your PC
properly communicates over a network. It makes a list of all the remote PCs on the network that
the NIC uses and sends information in the form of several “data packets” from the tested PC to
the “host” PCs. These “data packets” are then returned by the “host” PCs to the tested PC where
the contents of the “data packet” are checked. If too many of the “data packets” fail to get
returned or are thrown away because of data corruption, the Network test logs as FAILED.
For test procedure instructions, see the Network Test section, page 7-130.
• Interactive
- Audio Test
The Audio Test is performed to check for proper operation of the audio system; the following
selections are available: Left Channel, Right Channel, Both Channels, and MIDI.
For test procedure instructions, see the Audio Test section, page 7-132.
- Monitor Test
The Monitor Test is used to perform a variety of monitor diagnostic tests, as described below.
• Combination Test: Checks the monitor alignment, color depth and monitor resolution.
Four cross hairs are placed in each corner of the screen, as well as RGB spectrum arrays
and horizontal and vertical alignment bars.
• Solid Color Test: Tests the panel display on laptops and LCD screens. This test display
a solid color on the screen to ensure that all the color pixels are operating correctly.
• VESA (Video Electronics Standards Association) Test Patterns: Checks luminance,
geometry and focus. These tests are designed for professional measurement use.
For test procedure instructions, see the Monitor Test section, page 7-134.
Section 7-3
Performing Full System Diagnostics
When performing the automatic Full System Diagnostic tests, it is strongly recommended to start with
the Full System Test (FE and BEP) - especially if the Vivid q N system shows signs of an intermittent
problem. As described in this section, the testing sequence can be paused and restarted at any stage,
providing a quick and easy method of reviewing any detected faults.
The different types of Full System Diagnostic Tests are described in detail in the Diagnostic Test
Window Menu Options section, on page 7-10. For instructions on performing each of these tests, refer
to the following information:
• "Full System Test (FE and BEP)" on page 7-20
NOTE: For a description of this test, refer to the "Full System Test (FE and BEP)" on page 7-10.
NOTE: The Full System Test (FE and BEP) performs a full sequence of complete system tests
(automatically) and requires approximately 10 minutes for completion.
1.) In the Diagnostic Test window, trackball to Full System Test (FE and BEP) in the list of tests
available for selection from the Diagnostic Test Tree (see Figure 7-5 below) and press Set. The Full
System Test (FE and BEP) name is displayed under Execute, and all system components are
highlighted blue in the Schematic View, as shown below:
2.) As indicated in the Special Instructions area, make sure the system is calibrated and disconnect
any probe currently connected to the system.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Automatic Test Sequence commences. While the sequence of automatic testing or the group
of tests proceeds, the name of the test currently in progress (Full System Test FE and BEP) is
displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bars in the Status area:
7-20 Section 7-3 - Performing Full System Diagnostics
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full System Test (FE and BEP) sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-6.
Figure 7-6 View Test Log - Full System Test (BEP and FE)
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
If the problem still persists, contact the GE Medical Systems On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
Section 7-4
Performing Front End (FE) Diagnostics
NOTE: When performing Front End Diagnostic tests, it is strongly recommended to start with the Full
System Test (FE and BEP) especially if the system shows signs of an intermittent problem.
(For details, see "Full System Test (FE and BEP)" on page 7-20).
The Front End diagnostic tests can be initiated manually and may be performed in any order. However,
due to the functional dependencies between each of the Front End boards, it is recommended to
perform the tests on the boards in the following order:
1.) RFI board
2.) TR32 boards
3.) PM board
Performing the tests in the proper sequence will help to derive the correct conclusions regarding the
source of the problem.
As described in this section, the testing sequence can be paused and restarted at any stage, providing
a quick and easy method of reviewing any detected faults. For instructions on performing a specific test,
refer to the relevant test in the appropriate sub-sections:
NOTE: If the system must be re-booted, do so. Continue testing from the point at which you were
required to reboot. If a failure occurs, re-boot the system and test again to verify the outcome
of the test.
NOTE: For a description of this test, refer to the information provided in "DC Offset Calibration" on page
7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to DC Offset Calibration in the list of Front End tests
available for selection from the Diagnostic Test Tree (see Figure 7-7 below) and press Set. The DC
Offset Calibration name is displayed under Execute; the relevant system components to be tested
(in this case, all components) are highlighted light blue in the Schematic View as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The DC Offset Calibration Test commences. While the test proceeds, the name of the test currently
in progress (DC Offset Calibration) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the DC Offset Calibration Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing the test performed and showing
whether it passed or failed, as shown in the example in Figure 7-8.
NOTE: There is one test report (ADC Offset values) which is listed in the Reports window on the right.
By default, this report is displayed in the Reports window below.
8.) Use the scroll arrows at the bottom of the Reports window (see Figure 7-8) to scroll through the
channels and view the corresponding results.
9) When done, trackball to the Close button and press Set to close the View Report window.
10) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, repeat the
DC Offset Calibration test again and review the newly-acquired results.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE Medical Systems On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information provided in "DC Offset Verification" on page
7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to DC Offset Verification in the list of tests available for
selection from the Diagnostic Test Tree (see Figure 7-9 below) and press Set. The DC Offset
Verification name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (PM, TR32 and RFI boards), as shown below:
Figure 7-9 DC Offset Verification Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The DC Offset Verification Test commences. While the test proceeds, the name of the test currently
in progress (DC Offset Verification) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the DC Offset Verification Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-10.
Figure 7-10 View Test Log - DC Offset Verification Test
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (ADC Offset verification values) which is listed in the Reports
window on the right. By default, this report is displayed in the Reports window below.
9.) Use the scroll arrows at the bottom of the Reports window (see Figure 7-10) to scroll through the
channels and view the corresponding results.
10) When done, trackball to the Close button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
NOTE: In the event that the surrounding temperature is above 38°C, the DC Offset Verification Test
may fail. Upper/lower test limits of ±10 (instead of ±5) are acceptable.
12.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
13.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information provided in "Full RFI Test" on page 7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to Full RFI Test in the list of RFI tests available for selection
from the Diagnostic Test Tree (see Figure 7-11 below) and press Set. The Full RFI Test name is
displayed under Execute; the relevant system component to be tested is highlighted light blue in the
Schematic View (RFI), as shown below:
Figure 7-11 Full RFI Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Full RFI Test commences. While the test proceeds, the name of the test currently in progress
(Full RFI Test) is displayed in the Execute field and the current Loop # and specific test are shown
Chapter 7 - Diagnostics/Troubleshooting 7-31
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full RFI Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in the example in Figure 7-12.
Figure 7-12 View Test Log - Full RFI Test
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For the Full RFI test, four tests are performed in the test sequence; these are listed in the Tests
window as shown in Figure 7-12. For each currently-selected test (for example, RFI FE Cache
Test) the corresponding report is listed in Reports on the right.
For details of each report, refer to the following explanations:
If the problem still persists, contact the GE On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information provided in "RFI FE Cache Test" on page
7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to RFI FE Cache Test in the list of RFI tests available for
selection from the Diagnostic Test Tree (see Figure 7-9 below) and press Set. The RFI FE Cache
Test name is displayed under Execute; the relevant system component to be tested remain
highlighted light blue in the Schematic View (RFI), as shown below:
Figure 7-13 RFI FE Cache Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The RFI FE Cache Test commences. While the test proceeds, the name of the test currently in
progress (RFI FE Cache Test) is displayed in the Execute field and the current Loop # and specific
test are shown below. The messages Init Done and Please Wait are displayed beside the progress
bar in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
7-34 Section 7-4 - Performing Front End (FE) Diagnostics
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI FE Cache Test sequence is complete
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-14.
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (FE Cache Test report) which is listed in the Reports window
on the right. By default, this report is displayed in the Reports window below.
9) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information provided in "RFI Analog Circuits Test" on
page 7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to RFI Analog Circuits Test in the list of RFI tests available
for selection from the Diagnostic Test Tree (see Figure 7-15 below) and press Set. The RFI Analog
Circuits Test name is displayed under Execute; the relevant system component to be tested are
highlighted light blue in the Schematic View (RFI), as shown below:
Figure 7-15 RFI Analog Circuits Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The RFI Analog Circuits Test commences. While the test proceeds, the name of the test currently
in progress is displayed in the Execute field and the current Loop # and specific test are shown
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI Analog Circuits Test sequence is complete,
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-16.
Figure 7-16 View Test Log - RFI Analog Circuits Test
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (Analog Circuits Test report) which is listed in Reports on the
right. By default, this report is displayed in the Reports window below (Figure 7-16). Use the
Report window scroll bars and/or arrows to scroll through complete report if necessary.
9) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "RFI Demodulator Short Test" on page
7-11.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to RFI Demodulator Short Test in the list of RFI tests
available for selection from the Diagnostic Test Tree (see Figure 7-17 below) and press Set. The
RFI Demodulator Short Test name is displayed under Execute; the relevant system component to
be tested are highlighted light blue in the Schematic View (RFI), as shown below:
Figure 7-17 RFI Demodulator Short Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The RFI Demodulator Short Test commences. While the test proceeds, the name of the test
currently in progress is displayed in the Execute field and the current Loop # and specific test are
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in
the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI Demodulator Short Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-18.
Figure 7-18 View Test Log - RFI Demodulator Short Test
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there is one report (Demodulator Short Test report) which is listed in the Reports
window on the right. By default, this report is displayed in the Reports window below (Figure 7-
18). Use the Report window scroll bars and/or arrows to scroll through complete report if
necessary.
9) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information provided in "RFI Audio Test" on page 7-12.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to RFI Audio Test in the list of RFI tests available for
selection from the Diagnostic Test Tree (see Figure 7-9 below) and press Set. The RFI Audio Test
name is displayed under Execute; the relevant system component to be tested is highlighted light
blue in the Schematic View (RFI), as shown below:
Figure 7-19 RFI Audio Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The RFI Audio Test commences. While the test proceeds, the name of the test currently in progress
is displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
When the upper progress bar reaches approximately midway in the Status area, a loud,
high-pitched sound will be heard coming from the speakers.
NOTE: If no sound is heard, it is necessary to perform the BEP Audio Test as described in "Audio Test"
on page 7-132. In the event this test also fails, the problem is probably in the BEP - refer to
"Performing Back End Diagnostics on the System" on page 7-108.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the RFI Audio Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-20.
Figure 7-20 View Test Log - RFI Audio Test
NOTE: For this test, this one report (Demodulator Short Test report) which is listed in the Reports
window on the right. By default, this report is displayed in the Reports window below, as shown
in the example in Figure 7-20.
7-44 Section 7-4 - Performing Front End (FE) Diagnostics
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
NOTE: For a description of this test, refer to the information in "TR32 BF Digital Test" on page 7-12.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to TR32 BF Digital Test in the list of TR32, BF tests
available for selection from the Diagnostic Test Tree (see Figure 7-21 below) and press Set. The
TR32 BF Digital Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (TR32A and TR32B), as shown below:
Figure 7-21 TR32 BF Digital Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
5.) Trackball to the Start button and press Set.
The TR32 BF Digital Test commences. While the test proceeds, the name of the test currently in
progress is displayed in the Execute field and the current Loop # and specific test are shown below.
The messages Init Done and Please Wait are displayed beside the progress bar in the Status area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the TR32 BF Digital Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-22.
Figure 7-22 View Test Log - TR32 BF Digital Test (MLA0)
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there are two reports (MLA0 and MLA1) which is listed in Reports on the right. By
default, the MLA0 report is displayed in the Reports window below, showing the results for both
the TR32A and TR32B boards (Figure 7-22).
9.) To view the MLA1 test results in the Report window, select MLA1 test results under Reports.
The MLA1 report is displayed in the Reports window below:
Figure 7-23 View Test Log - TR32 BF Digital Test (MLA1)
10) When done, trackball to the Exit button and press Set to close the View Report window.
11) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
12.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
13.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "System BF Digital Test" on page 7-12.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to System BF Digital Test in the list of TR32, BF tests
available for selection from the Diagnostic Test Tree (see Figure 7-9 below) and press Set. The
System BF Digital Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (TR32A. TR32B), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The System BF Digital Test commences. While the test proceeds, the name of the test currently in
progress (System BF Digital Test) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the System BF Digital Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-25.
Figure 7-25 View Test Log - System BF Digital Test (MLA0)
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of Test results.
NOTE: For this test, there are two reports (MLA0 and MLA1) which is listed in Reports on the right. By
default, the MLA0 report is displayed in the Reports window below (Figure 7-25).
9.) To view the MLA1 test results, select MLA1 test results under Reports.
The MLA1 test results are displayed in the Report Window, as shown below.
10) When done, trackball to the Exit button and press Set to close the View Report window.
11) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
12.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
13.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ADC Test" on page 7-12.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to ADC Test in the list of TR32, BF tests available for
selection from the Diagnostic Test Tree (see Figure 7-9 below) and press Set. The ADC Test name
is displayed under Execute; the relevant system components to be tested are highlighted light blue
in the Schematic View (TR32A. TR32B), as shown below:
Figure 7-27 ADC Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The ADC Test commences. While the test proceeds, the name of the test currently in progress (ADC
Test) is displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ADC Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-28.
Figure 7-28 View Test Log - First Harmonic Report
Expected Average
Selected channel
Average Result
Scroll between
channels here
NOTE: The ADC test has four test reports which are listed in Reports on the right. By default, the First
Harmonic report is displayed in the Reports window below (Figure 7-28).
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels. Left-clicking on a currently-selected channel marks a red square on the report graph,
and also evokes the display of the following information relative to that specific point: Average,
Deviation (standard deviation from the Average), Max, and Min.
The signal should be within the limits marked in the white area (the dotted line represents the
Average), as shown in Figure 7-28.
8) When done, trackball to the Close button and press Set to close the View Report window.
9.) Select the second report (the report is highlighted in the Reports list, and the corresponding results
are shown in the Report window below, as shown in Figure 7-29).
Figure 7-29 View Test Log - Second Harmonic
10.)Use the tools provided to scroll through the channels and view the results, as previously described.
11) When done, trackball to the Close button and press Set to close the View Report window.
12.)Select the third report (the report is highlighted in the Reports list, and the corresponding results are
shown in the Report window below, as shown in Figure 7-30).
13.)Use the tools provided to scroll through the channels and view the results, as previously described.
14) When done, trackball to the Close button and press Set to close the View Report window.
15.)Select the fourth report (the report is highlighted in the Reports list, and the corresponding results
are shown in the Report window below, as shown in Figure 7-31).
16.)Use the tools provided to scroll through the channels and view the results, as previously described.
17) When done, trackball to the Close button and press Set to close the View Report window.
18) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
19.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
20.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-4-4-3-1 RX Test
NOTE: For a description of this test, refer to the information in "RX Test" on page 7-12.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to Rx Test in the list of RX tests available for selection from
the Diagnostic Test Tree (see Figure 7-32 below) and press Set. The RX Test name is displayed
under Execute; the relevant system components to be tested are highlighted light blue in the
Schematic View (TR32A. TR32B), as shown below:
Figure 7-32 Rx Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Rx Test commences. While the test proceeds, the name of the test currently in progress
(Rx Test) is displayed in the Execute field and the current Loop # and specific test are shown below.
The messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Rx Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-33.
Figure 7-33 View Test Log - Rx Test Signal Amplitude
Expected Average
Selected channel
Average Result
Scroll between
channels here
NOTE: The Rx test has two test reports (Test Signal Amplitude and Frequency) which are listed in the
Reports window on the right. By default, the Test Signal Amplitude report is displayed in the
Reports window below (Figure 7-33).
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also
evokes the display of the following information relative to that specific point: Average, Deviation
8) When done, trackball to the Close button and press Set to close the View Report window.
9) Select the second report (Frequency); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-29.
Figure 7-34 View Test Log - Rx Test Signal Frequency
10.)Use the tools provided to scroll through the channels and view the results, as previously described.
11) When done, trackball to the Close button and press Set to close the View Report window.
12) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
13.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
14.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW IQ Symmetry Test" on page 7-13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to ACW IQ Symmetry Test in the list of ACW tests
available for selection from the Diagnostic Test Tree (see Figure 7-32 below) and press Set. The
ACW IQ Symmetry Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (P&M), as shown below:
Figure 7-35 ACW IQ Symmetry Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW IQ Symmetry Test commences. While the test proceeds, the name of the test currently
in progress (ACW IQ Symmetry Test) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW IQ Symmetry Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-36.
NOTE: The ACW IQ Symmetry Test test has one test report. By default, this is displayed in the Reports
window below (Figure 7-36).
7) When done, trackball to the Exit button and press Set to close the View Report window.
8) After reviewing the test results, proceed as follows:
• Re-boot the system (this is obligatory).
• Perform calibration, as described in "DC Offset Calibration" on page 7-25.
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
9.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
10.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW Matrix Test" on page 7-13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to ACW Matrix Test in the list of ACW tests available for
selection from the Diagnostic Test Tree (see Figure 7-37 below) and press Set. The ACW Matrix
Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (P&M), as shown below:
Figure 7-37 ACW Matrix Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW Matrix Test commences. While the test proceeds, the name of the test currently in
progress (ACW IQ Symmetry Test) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW Matrix Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-38.
Figure 7-38 View Test Log - ACW Matrix Test - Signal Frequency
Expected Average
Selected channel
Average Result
Scroll between
channels here
NOTE: The ACW Matrix Test test has three test reports (Signal Frequency, Signal RMS, and Signal
SNR) which are listed in the Reports window on the right. By default, the Signal Frequency
report is displayed in the Reports window below (Figure 7-38).
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also
evokes the display of the following information relative to that specific point: Average, Deviation
7) When done, trackball to the Close button and press Set to close the View Report window.
8) Select the second report (Signal RMS); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-39.
Figure 7-39 View Test Log - ACW Matrix Test - Signal RMS
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the Close button and press Set to close the View Report window.
11) Select the third report (Signal SNR); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-40.
Figure 7-40 View Test Log - ACW Matrix Test - Signal SNR
12.)Use the tools provided to scroll through the channels and view the results, as previously described.
13) When done, trackball to the Close button and press Set to close the View Report window.
14) After reviewing the test results, proceed as follows:
• Re-boot the system (this is obligatory).
• Perform calibration, as described in "DC Offset Calibration" on page 7-25
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
15.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
16.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW Delay Range Test" on page 7-13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to ACW Delay Range Test in the list of ACW tests
available for selection from the Diagnostic Test Tree (see Figure 7-41 below) and press Set. The
ACW Delay Range Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (P&M), as shown below:
Figure 7-41 ACW Delay Range Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW Delay Range Test commences. While the test proceeds, the name of the test currently
in progress (ACW IQ Delay Range Test) is displayed in the Execute field and the current Loop #
and specific test are shown below. The messages Init Done and Please Wait are displayed beside
the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW Delay Range Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-42.
Figure 7-42 View Test Log - ACW Delay Range Test - Signal Frequency
Expected Average
Selected channel
Average Result
Scroll between
channels here
NOTE: The ACW Delay Range Test test has three test reports (Signal Frequency, Signal RMS, and
Signal SNR) which are listed in the Reports window on the right. By default, the Signal
Frequency report is displayed in the Reports window below (Figure 7-42).
NOTE: Left and right arrows provided in the View Report window enable scrolling between the different
channels.
Left-clicking on a currently-selected channel marks a red square on the report graph, and also
evokes the display of the following information relative to that specific point: Average, Deviation
7) When done, trackball to the Close button and press Set to close the View Report window.
8) Select the second report (Signal RMS); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-43.
Figure 7-43 View Test Log - ACW Delay Range Test - Signal RMS
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the Close button and press Set to close the View Report window.
11) Select the third report (Signal SNR); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-44.
Figure 7-44 View Test Log - ACW Delay Range Test - Signal SNR
12.)Use the tools provided to scroll through the channels and view the results, as previously described.
13) When done, trackball to the Close button and press Set to close the View Report window.
14) After reviewing the test results, proceed as follows:
• Re-boot the system (this is obligatory).
• Perform calibration, as described in "DC Offset Calibration" on page 7-25.
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
15.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
16.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW Pre-Mixer Gain Test" on page 7-
13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to ACW Pre-Mixer Gain Test in the list of ACW tests
available for selection from the Diagnostic Test Tree (see Figure 7-45 below) and press Set. The
ACW Pre-Mixer Gain Test name is displayed under Execute; the relevant system components to
be tested are highlighted light blue in the Schematic View (P&M), as shown below:
Figure 7-45 ACW Pre-Mixer Gain Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW Pre-Mixer Gain Test commences. While the test proceeds, the name of the test currently
in progress (ACW Pre-Mixer Gain Test) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW Pre-Mixer Gain Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-46.
Figure 7-46 View Test Log - ACW Pre-Mixer Gain Test
NOTE: The ACW Pre-Mixer Gain Test has one test report. By default, this is displayed in the Reports
window below (Figure 7-46).
7) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
9.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
10.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW BPF Test" on page 7-13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
NOTICE It is necessary to disable the Loop option where it should not be used.
After performing this test, you must re-boot the system and perform the DC Offset Calibration procedure
as described in "DC Offset Calibration" on page 7-25.
1.) In the Diagnostic Test window, trackball to ACW BPF Test in the list of ACW tests available for
selection from the Diagnostic Test Tree (see Figure 7-47 below) and press Set. The ACW BPF Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (P&M), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW BPF Test commences. While the test proceeds, the name of the test currently in progress
(ACW BPF Test) is displayed in the Execute field and the current Loop # and specific test are shown
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW BPF Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-48.
NOTE: The ACW BPF Test has four test reports (BPF0, BPF1, BPF2, and BPF3) which are listed in
the Reports window on the right. By default, the BPF0 report is displayed in the Reports window
below (Figure 7-48).
7) When finished reviewing the first report, trackball to the Exit button and press Set to close the View
Report window.
8) Select the second report (BPF1); the report is highlighted in the Reports list, and the corresponding
results are shown in the Report window below, as shown in Figure 7-49.
Figure 7-49 View Test Log - ACW BPF Test - BPF1
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the Close button and press Set to close the View Report window.
11) Select the third report (BPF2); the report is highlighted in the Reports list, and the corresponding
results are shown in the Report window below, as shown in Figure 7-50.
12.)Use the tools provided to scroll through the channels and view the results, as previously described.
13) When done, trackball to the Close button and press Set to close the View Report window.
14) Select the fourth report (Signal SNR); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-51.
15.)Use the tools provided to scroll through the channels and view the results, as previously described.
16) When done, trackball to the Close button and press Set to close the View Report window.
17) After reviewing the test results, proceed as follows:
• Re-boot the system (this is obligatory).
• Perform calibration, as described in "DC Offset Calibration" on page 7-25.
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
18.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
19.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ACW HPF Test" on page 7-13.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
NOTICE It is necessary to disable the Loop option where it should not be used.
After performing this test, you must re-boot the system and perform the DC Offset Calibration procedure
as described in "DC Offset Calibration" on page 7-25.
1.) In the Diagnostic Test window, trackball to ACW HPF Test in the list of ACW tests available for
selection from the Diagnostic Test Tree (see Figure 7-52 below) and press Set. The ACW HPF Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (P&M), as shown below:
Figure 7-52 ACW HPF Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
4.) Trackball to the Start button and press Set.
The ACW HPF Test commences. While the test proceeds, the name of the test currently in progress
(ACW HPF Test) is displayed in the Execute field and the current Loop # and specific test are shown
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW HPF Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-53.
NOTE: The ACW HPF Test has two test reports (HPF300 and HPF600) which are listed in the Reports
window on the right. By default, the HPF300 report is displayed in the Reports window below
(Figure 7-53).
7) When finished reviewing the first report, trackball to the Exit button and press Set to close the View
Report window.
8) Select the second report (HPF600); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-54.
9) When done, trackball to the Exit button and press Set to close the View Report window.
10) After reviewing the test results, proceed as follows:
• Re-boot the system (this is obligatory).
• Perform calibration, as described in "DC Offset Calibration" on page 7-25.
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance
11.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
12.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Pencil Rx Test" on page 7-14.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) In the Diagnostic Test window, trackball to Pencil RX Test in the list of Pencil tests available for
selection from the Diagnostic Test Tree (see Figure 7-55 below) and press Set. The Pencil RX Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (P&M), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Pencil RX Test commences. While the test proceeds, the name of the test currently in progress
(Pencil RX Test) is displayed in the Execute field and the current Loop # and specific test are shown
below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Pencil RX Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-56.
NOTE: The Pencil RX Test has one test report which is listed in Reports on the right and displayed in
the Reports window below (Figure 7-56).
8) When finished reviewing the first report, trackball to the Exit button and press Set to close the View
Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
Section 7-5
Monitoring Tests
7-5-1 Accessing the Monitoring Diagnostic Test Options
• Access the Diagnostic Test window, as described in the Accessing the Front End Diagnostic
Options section, on page 7-25.
NOTE: For a description of this test, refer to the information in "Full System Monitoring Test" on page
7-14.
1.) In the Diagnostic Test window, trackball to Full System Monitoring Test in the list of Monitoring
tests available for selection from the Diagnostic Test Tree (see Figure 7-57 below) and press Set.
The Full System Monitoring Test name is displayed under Execute; all system components to be
tested are highlighted light blue in the Schematic View, as shown below:
Figure 7-57 Full System Monitoring Test
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
The Full System Monitoring Test commences. While the test proceeds, the name of the test currently
in progress (ADC Test) is displayed in the Execute field and the current Loop # and specific test are
7-86 Section 7-5 - Monitoring Tests
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in the
Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the Pause button and press Set, if required.
7.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full System Monitoring Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-58.
Figure 7-58 View Test Log - Full System Monitoring Test
NOTE: The Full System Monitoring Test has five sub-tests (Fixed Voltages, Temperatures, Fans
Speed, Battery Parameters, and Tx Voltages). These are listed in the Test column, as shown
in Figure 7-58.
8.) Select the first report (Fixed Voltages); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-59.
Figure 7-59 View Test Log - Full System Monitoring Test - Fixed Voltages
9.) Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in Figure 7-59.
10.)Select the second report (Temperatures); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-60.
Figure 7-60 View Test Log - Full System Monitoring Test - Temperatures
11.)Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in Figure 7-60.
12.)Select the third report (Fans Speed); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-61.
Figure 7-61 View Test Log - Full System Monitoring Test - Fans Speed
13.)Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in Figure 7-59.
14.)Select the fourth report (Battery Parameters); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-62.
Figure 7-62 View Test Log - Full System Monitoring Test - Battery Parameters
15.)Check that the sub-test passed and that result values are within the pre-defined Upper and Lower
limits, as shown in Figure 7-62.
16.)Select the fifth report (Tx Voltages); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-63.
Figure 7-63 View Test Log - Full System Monitoring Test - Tx Voltages
17.)Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in Figure 7-63.
18) When done, trackball to the Exit button and press Set to close the View Report window.
19) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
20.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
21.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Fixed Voltages Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Fixed Voltages in the list of Monitoring tests available
for selection from the Diagnostic Test Tree (see Figure 7-64 below) and press Set. The Fixed
Voltages Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
Figure 7-64 Voltages Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Fixed Voltages Test commences. While the test proceeds, the name of the test currently in
progress (Fixed Voltages) is displayed in the Execute field and the current Loop # and specific test
are shown below. The messages Init Done and Please Wait are displayed beside the progress bar
in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
Chapter 7 - Diagnostics/Troubleshooting 7-93
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Fixed Voltages Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-65.
Figure 7-65 View Test Log - Fixed Voltages Test
NOTE: For this test, there is one report (Voltages) which is listed in the Reports window on the right.
By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values are within the pre-defined Max and Min limits, as
shown in Figure 7-65.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Temperatures Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Temperatures in the list of Monitoring tests available for
selection from the Diagnostic Test Tree (see Figure 7-66 below) and press Set. The Temperatures
Test name is displayed under Execute; all system components (excluding the P&M) will be tested
are highlighted light blue in the Schematic View (all), as shown below:
Figure 7-66 Temperatures Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Temperatures Test commences. While the test proceeds, the name of the test currently in
progress (Temperatures) is displayed in the Execute field and the current Loop # and specific test
are shown below. The messages Init Done and Please Wait are displayed beside the progress bar
in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
7-96 Section 7-5 - Monitoring Tests
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Temperatures Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-67.
Figure 7-67 View Test Log - Temperatures Test
NOTE: For this test, there is one report (Temperatures) which is listed in the Reports window on the
right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values are within the pre-defined Max and Min limits, as
shown in Figure 7-67.
8) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Fans Speed Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Fans Speed in the list of Monitoring tests available for
selection from the Diagnostic Test Tree (see Figure 7-68 below) and press Set. The Fans Speed
Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
Figure 7-68 Fans Speed Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Fans Speed Test commences. While the test proceeds, the name of the test currently in
progress (Fans Speed) is displayed in the Execute field and the current Loop # and specific test are
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in
the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
Chapter 7 - Diagnostics/Troubleshooting 7-99
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Fans Speed Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-69.
Figure 7-69 View Test Log - Fans Speed Test
NOTE: For this test, there is one report (Fans Status) which is listed in the Reports window on the right.
By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each fan are within the pre-defined Max and
Min limits, as shown in Figure 7-69.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If the test fails, repeat the Fan Speed Test.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Battery Parameters Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Battery Parameters in the list of Monitoring tests
available for selection from the Diagnostic Test Tree (see Figure 7-70 below) and press Set. The
Battery Parameters Test name is displayed under Execute; the relevant system components to be
tested are highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
Figure 7-70 Battery Parameters Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Battery Parameters Test commences. While the test proceeds, the name of the test currently
in progress (Battery Parameters) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Battery Parameters Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-71.
Figure 7-71 View Test Log - Battery Parameters Test
NOTE: For this test, there is one report (Battery Status) which is listed in the Reports window on the
right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each sensor are within the pre-defined Lower
and Upper limits, as shown in Figure 7-71.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
Chapter 7 - Diagnostics/Troubleshooting 7-103
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
• If the test fails again, repeat the Fan Speed Test.
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Tx Voltages Test" on page 7-14.
1.) In the Diagnostic Test window, trackball to Tx Voltages in the list of Monitoring tests available for
selection from the Diagnostic Test Tree (see Figure 7-72 below) and press Set. The Tx Voltages
Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
Figure 7-72 Tx Voltages Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Tx Voltages Test commences. While the test proceeds, the name of the test currently in
progress (Tx Voltages) is displayed in the Execute field and the current Loop # and specific test are
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in
the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Tx Voltages Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in Figure 7-73.
Figure 7-73 View Test Log - Tx Voltages Test
NOTE: For this test, there is one report (Tx Voltages) which is listed in the Reports window on the right.
By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values for each fan are within the pre-defined Max and
Min voltage limits, as shown in Figure 7-73.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
7-106 Section 7-5 - Monitoring Tests
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
assistance.
• If errors were identified but the results were very close to the upper or lower limits, perform
calibration, as described in "DC Offset Calibration" on page 7-25, and then repeat the test.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
Section 7-6
Performing Back End Diagnostics on the System
The Back End Diagnostics Tests enable you to test the functionality of various Back End hardware
components.
NOTE: The Back End Diagnostic Tests may be performed independently, regardless of sequence.
NOTE: For a description of this test, refer to the information in the "Extended Keyboard Test" on page
7-15.
1.) In the Diagnostic Test window, trackball to Extended Keyboard Test in the list of BEP tests available
for selection from the Diagnostic Test Tree (see Figure 7-74 below) and press Set. The Extended
Keyboard Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-74 Extended Keyboard Test
The interactive Keyboard Test screen is displayed as shown below, enabling testing of the various
external keyboard components.
Figure 7-75 Extended Keyboard Test Screen
NOTE: When performing this step, make sure that a minimum of 32 clicks are heard when turning the
Gain rotary in both the clockwise and counter-clockwise directions to ensure that a complete
check is performed.
NOTE: when the trackball cursor is positioned at the extreme top left of the screen:
X = 0 and Y = 0.
b.) As you move the trackball cursor, make sure the coordinate values increase/decrease in small
steps, without jumping more than two or three digits. If for example, the values increment by
steps of 10, this indicates that the trackball operation is faulty (possibly the trackball requires
cleaning, or replacement).
16) When done with all the above tests, trackball to the OK button and press Set.
The View Test Log dialog box is displayed, as shown in Figure 7-76 below.
Figure 7-76 View Test Log - Extended Keyboard Test
17.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
18) Proceed as follows:
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
19.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "ECG Test" on page 7-15.
NOTE: When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) Connect the ECG cable to the input on the ECG port on the right side of the Vivid q N scanner.
2.) Connect the ECG leads to yourself (as the simulated patient).
3.) In the Diagnostic Test window, trackball to ECG Test in the list of BEP tests available for selection
from the Diagnostic Test Tree (see Figure 7-77 below) and press Set. The ECG Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-77 ECG Test
4.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
5.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
6.) Trackball to the Start button and press Set.
The interactive ECG Device Test screen is displayed as shown in Figure 7-78.
7.) Make sure that the resolution scale is set to 1 Sec. (as shown in Figure 7-78, above).
8.) Trackball to the Open Device button and press Set.
The ECG signal is displayed in the Data area, as shown in the example below:
Figure 7-79 Test ECG - Data
9.) Verify that a single clear QRS signal is displayed on the screen, as shown in Figure 7-79.
10.)When done, trackball to the Exit button and press Set.
11.)When prompted by the following message, select Yes or No as applicable:
7-6-1-3 PC Doctor
NOTE: For a complete description of all PC Doctor tests, refer to the information in "PC Doctor" on page
7-15.
This procedures in this section are divided into two groups. Refer to the relevant procedure, as follows:
• Automatic Tests
- Full Back End Test w/o Hard Disk Scan - see below
- "Full PC Doctor Test (50 min)" on page 7-117
- "CPU Tests" on page 7-119
- "HD Test" on page 7-122
- "Memory Test" on page 7-124
- "Video Test" on page 7-126
- "PCI Test" on page 7-128
- "Network Test" on page 7-130
NOTE: For a description of this test, refer to "Full Back End Test without Hard Disk" on page 7-15.
1.) In the Diagnostic Test window, trackball to Full Back End Test w/o Hard Disk Scan in the list of PC
Doctor tests available for selection from the Diagnostic Test Tree (see Figure 7-80 below) and press
Set. The Test name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-80 Full Back End Test w/o Hard Disk Scan
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Full Back End Test w/o Hard Disk Scan commences. While the test proceeds, the name of the
test currently in progress is displayed in the Execute field and the current Loop # and specific test
are shown below. The messages Init Done and Please Wait are displayed beside the progress bar
in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full Back End Test w/o Hard Disk Scan sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-81.
Figure 7-81 View Test Log - Full Back End Test w/o Hard Disk Scan
NOTE: For the Full Back End Test w/o Hard Disk Scan, many tests are performed in the test sequence;
these are listed in the Tests window as shown in Figure 7-81. For each currently-selected test
(for example, CPU Full Test) the corresponding report is listed in Reports on the right.
For details of each report, refer to the explanations provided in "CPU Tests" on page 7-119:
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required.
8) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Full PC Doctor Test (50 min)" on page
7-15.
1.) In the Diagnostic Test window, trackball to Full PC Doctor Test in the list of PC Doctor tests available
for selection from the Diagnostic Test Tree (see Figure 7-82 below) and press Set. The Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-82 Full PC Doctor Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Full PC Doctor Test commences. While the test proceeds, the name of the test currently in
progress is displayed in the Execute field and the current Loop # and specific test are shown below.
The messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full PC Doctor Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-81.
Figure 7-83 View Test Log - Full PC Doctor Test
NOTE: For the Full PC Doctor Test, many tests are performed in the test sequence; these are listed in
the Tests window as shown in Figure 7-83. For each currently-selected test (for example, CPU
Full Test) the corresponding report is listed in Reports on the right.
For details of each report, refer to the explanations provided in "CPU Tests" on page 7-119:
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required.
8) When done, trackball to the Exit button and press Set to close the View Report window.
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of these tests, refer to the information in "CPU Tests" on page 7-15.
1.) In the Diagnostic Test window, trackball to CPU Full Test in the list of CPU tests available for selection
from the Diagnostic Test Tree (see Figure 7-84 below) and press Set. The Test name is displayed
under Execute; the relevant system components to be tested are highlighted light blue in the
Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-84 CPU Full Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The CPU Full Test commences. While the test proceeds, the name of the test currently in progress
is displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the CPU Full Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-81.
Figure 7-85 View Test Log - CPU Full Test
NOTE: For the CPU Full Test, many tests are performed in the test sequence; these are listed in the
Tests window as shown in Figure 7-83. For each currently-selected test (for example,
Registers) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
- CPU Registers
- CPU Arithmetics
- CPU String Operations
- CPU Logical Operations
- CPU Math Registers
- CPU Math Commands
7-120 Section 7-6 - Performing Back End Diagnostics on the System
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
- CPU Transcendental
- CPU MMX Arithmetic
- CPU MMX Logical
- CPU MMX Shift
- CPU MMX Data Transfer
- CPU MMX Misc.
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-6-1-3-4 HD Test
NOTE: For a description of these tests, refer to the information in "HD Test" on page 7-16.
1.) In the Diagnostic Test window, trackball to Hard Drive Full Test w/o Scan in the list of HD tests
available for selection from the Diagnostic Test Tree (see Figure 7-86 below) and press Set. The Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-86 Hard Drive Full Test w/o Scan
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Hard Drive Full Test w/o Scan commences. While the test proceeds, the name of the test
currently in progress is displayed in the Execute field and the current Loop # and specific test are
shown below. The messages Init Done and Please Wait are displayed beside the progress bar in
the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Hard Drive Full Test w/o Scan sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-87.
Figure 7-87 View Test Log - Hard Drive Full Test w/o Scan
NOTE: For the Hard Drive Full Test w/o Scan, several tests are performed in the test sequence; these
are listed in the Tests window as shown in Figure 7-87. For each currently-selected test (for
example, Linear Seek) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
- Linear Seek
- Random Seek
- Funnel Scan
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of these tests, refer to the information in "Memory Test" on page 7-17.
1.) In the Diagnostic Test window, trackball to Full Memory Test in the list of Memory tests available for
selection from the Diagnostic Test Tree (see Figure 7-88 below) and press Set. The Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Full Memory Test commences. While the test proceeds, the name of the test currently in
progress is displayed in the Execute field and the current Loop # and specific test are shown below.
The messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full Memory Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-89.
Figure 7-89 View Test Log - Full Memory Test
NOTE: For the Full Memory Test, several tests are performed in the test sequence; these are listed in
the Tests window as shown in Figure 7-89. For each currently-selected test (for example,
Memory Pattern Test) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
- Memory Pattern Test
- Memory Parity Test
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of these tests, refer to the information in "Video Test" on page 7-17.
1.) In the Diagnostic Test window, trackball to Full Video Test in the list of Video tests available for
selection from the Diagnostic Test Tree (see Figure 7-88 below) and press Set. The Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Full Video Test commences. While the test proceeds, the name of the test currently in progress
is displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Full Video Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-91.
Figure 7-91 View Test Log - Full Video Test
NOTE: For the Full Video Test, several tests are performed in the test sequence; these are listed in the
Tests window as shown in Figure 7-91. For each currently-selected test (for example, Video
Memory) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
- Video Memory
- Video Data Transfer
- Video Text Output
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "PCI Test" on page 7-17.
1.) In the Diagnostic Test window, trackball to PCI Test in the list of PC Doctor tests available for selection
from the Diagnostic Test Tree (see Figure 7-92 below) and press Set. The Test name is displayed
under Execute; the relevant system components to be tested are highlighted light blue in the
Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The PCI Test commences. While the test proceeds, the name of the test currently in progress is
displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the PCI Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-89.
Figure 7-93 View Test Log - PCI Test
NOTE: For the PCI Test, there is one report which is listed in the Tests window as shown in
Figure 7-93.
7) When done, trackball to the Exit button and press Set to close the View Report window.
8) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
9.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
10.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to the information in "Network Test" on page 7-17.
1.) In the Diagnostic Test window, trackball to Network Test in the list of PC Doctor tests available for
selection from the Diagnostic Test Tree (see Figure 7-94 below) and press Set. The Test name is
displayed under Execute; the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-94 Network Test
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The Network Test commences. While the test proceeds, the name of the test currently in progress
is displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
• Green: All recommended tests for these components have been completed and no problems
were identified.
• Red: Problems were identified in these components during the performance of the test.
• Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Network Test sequence is complete:
- the Diagnostic Test window displays the Finished message
- the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in Figure 7-95.
Figure 7-95 View Test Log - Network Test
NOTE: For the Network Test, there is one report which is listed in the Tests window as shown in
Figure 7-95.
7) When done, trackball to the Exit button and press Set to close the View Report window.
8) After reviewing the test results, proceed as follows:
• If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
• If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
• If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in Chapter 5 - Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE On-Line Center for assistance.
9.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
10.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of this test, refer to "Audio Test" on page 7-18.
1.) In the Diagnostic Test window, trackball to Audio Test in the list of Interactive PC Doctor tests
available for selection from the Diagnostic Test Tree (see Figure 7-96 below) and press Set. The Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-96 Audio Test
3.) Trackball to the Left Channel, Right Channel, Both Channels or Midi button and press Set to select
the speaker from which to play sound.
4.) Check that the sound produced from the speaker is OK and audible.
5.) Repeat step 3 to test the other channels.
NOTE: If no sound is heard, trackball to the More Info button and press Set. Follow the on-screen
suggestions provided.
6.) When done, trackball to the Close button and press Set to exit the test.
NOTE: For the Audio Test, there is one report which is listed in the Tests window as shown in
Figure 7-98.
If the problem still persists, contact the GE On-Line Center for assistance.
8.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
9.) Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
NOTE: For a description of the monitor tests, refer to "Monitor Test" on page 7-18.
1.) In the Diagnostic Test window, trackball to Monitor Test in the list of Interactive PC Doctor tests
available for selection from the Diagnostic Test Tree (see Figure 7-99 below) and press Set. The Test
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
Figure 7-99 Monitor Test
3.) Select and drag the Test screen to another region so that the display screen is not obscured.
4.) Trackball to the required button (Solid Color Test: Red, Green or Blue) and press Set to perform
the corresponding test.
5.) When in the displayed screen, press any key to return to PC Dr for Windows Monitor Test screen.
6.) When done, trackball to the Close button and press Set to exit the test.
NOTE: For the Monitor Test, there is one report which is listed in the Tests window as shown in
Figure 7-101.
If the problem still persists, contact the GE On-Line Center for assistance.
8.) Trackball to the Exit button and press Set to close the View Test Log dialog box.
9.) Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
Section 7-7
Automatic Error Log
The Vivid q N automatically logs the software activity, sequences and error messages that occur during
normal operation. Each day’s activity is compressed into a single daily logger report, with a time stamp
appearing in each line of the report. Additional user comments and screen captures, can also be
included in the report. Each day’s report is saved in the system for 90 days, during which time they can
be recalled and saved to disk.
The bookmark will allow users to place a time-stamp where a particular problem has occurred, while
continuing with their current activity with minimum interruption.
NOTE: The log file bookmark will later make it easy for service engineers to locate the location of the
failure logging more precisely.
1.) While the system is turned-ON in any mode, press <ALT + D> on the alphanumeric keyboard.
The System Problem Reporting dialog opens - see Figure 7-103 on page 7-138.
2.) In the Description of Issue window, type a description of the problem.
Notes should be made regarding the selected probe, the imaging mode and the application that was
being used at the time of the malfunction.
If applicable, try to describe the button or key-pushing sequence that immediately preceded the
problem.
3.) If use of the Advanced features is required, proceed to the following instructions, as applicable:
• "Enabling Extensive Logging" on page 7-141
• "Advanced Options" on page 7-143
• "DB Scan" on page 7-146
After the required advanced features have been selected, continue with step 4 and step 5, below.
4.) From the Destination drop-down menu, select the required export destination (DVD/CD or Memory
Stick) then click Save and Export.
A Zip file (named “logfile_<date>_<time>.zip”) is copied onto the selected removable media.
NOTE: Previously, users were required to save log files before exporting them. With version BT06,
these two operations are performed as one step.
When selecting DVD/CD as the export destination, first make sure that the DVD/CD has been
formatted. For details, see "Formatting a DVD/CD" on page 7-139.
To Format a DVD/CD:
1.) Connect the DVD/CD-RW media drive to the USB port and make sure the DVD/CD-RW power
switch on the rear of the drive is set in the External position, as shown in Figure 7-104, below.
Figure 7-104 Power Switch in External Position
2.) Insert the blank DVD/CD media into the DVD/CD-RW drive tray.
NOTICE IMPORTANT: Use only CD-R, DVD-R or DVD+R media from a reputable vendor and only
high-grade media types.
NOTE: In some cases, the user may need to activate the Extensive Logging feature. This will allow the
system to record logs in a more detailed format, specifically addressing an issue of suspected
failure.
As shown in Figure 7-106 on page 7-142, checkboxes provided in the Extensive Logging dialog
enable the user to activate the appropriate failure category (any one or more; any combination)
to collect more information regarding particular functionality.
Table 7-4 provides recommendations for activating a specific Extensive Log category.
Extensive Log
Type of Failure
Category
Media Handling System Lock-up when using removable media (MOD, DVD/CD, Disk-on-Key)
NOTE: The impact of activating a checkbox would be to generate larger, detailed log files.
After usage, make sure to return to the default settings (unflagged) to conserve the size of the
log files. After 7 days, the system will automatically return to the default settings.
NOTE: By default, the log files may be very large as they are not limited by time/date. The user is able
to limit the size/timeframe of the log files by choosing the required settings, as described below.
1.) Activate the Export Logs Using checkbox, as shown in Figure 7-107, above.
2.) Proceed to set the required Log Range, as follows:
Either:
Activate the Time Range radio button, then use the From:/To: drop-down lists to define the required
time range.
Or:
Activate the Bookmark radio button, then select the required Bookmark from the drop-down list of
bookmarks.
This will export log files which are limited to the range of 1 hour before through 1 hour after the
selected bookmark.
NOTE: In the Advanced Export Options dialog, from the lower area (labelled Attach - see
Figure 7-107 on page 7-143) users may flag the different items they may wish to attach to the
log files (or alternatively disable the checkboxes for those not required).
Table 7-5 outlines the different categories of information that may be attached to a log file.
Diagnostic Reports Should be used whenever a diagnostic test has failed and it is necessary to forward the
information to GE for analysis.
Crash and Memory Dumps Useful for any type of memory crash or system lock-up problem.
Keyboard Strokes Log Shows the keyboard strokes that were recorded by the system - may be useful in detection of
a specific error.
Windows Logs Contains the Windows operating system logs - supports the application operation.
Sniffer Logs Applicable if the Network Sniffer Log has been used.
Data Base Folder Provides more Archive information. Useful for loss of patient information and/or analysis of
database, archive, or back-up problems.
D: and E: File Listing Lists all file names and locations in the hard disk drives: D and E.
Useful in the event of loss of patient information, missing images, lock-up problems, etc.
"Upload to GE" Tool Enables the user to export the log directly to GE for analysis.
For details, see "Sending Log Files to GE" on page 7-145.
DBScan Log files created when performing a scan of the database contain useful information regarding
the integrity of the database.
1.) Make sure the required Log File Ranges are set, as previously described.
2.) Under Attach, use the checkboxes provided to flag the required attachments, as appropriate.
NOTE: Clicking the Restore Defaults button returns to the default setting whereby all checkboxes in the
left column are flagged and all on the right are disabled, as shown Figure 7-107.
3.) When done, from the Destination drop-down menu, select the required export destination (CD or
Memory Stick) then click Save and Export.
A Zip file (named "logfile_<date>_<time>.zip") is copied onto the selected removable media.
NOTE: Previously, users were required to save log files before exporting them. With version BT06,
these two operations are performed as one step.
When selecting DVD/CD as the export destination, first make sure that the DVD/CD has been
formatted. For details, see "Formatting a DVD/CD" on page 7-139.
When export is completed, a message is displayed saying (number of) reports successfully exported.
4.) Click OK to close the window.
5.) When export completed, press <Ctrl>+E to eject the media.
NOTE: Successful upload of logs to the GE ftp site can only be achieved when the designated PC is
connected to the Internet and when all security measures (such as, firewalls) taken at the local
network do not prevent the utility from performing the upload.
1.) Place the media in any Windows-based PC and run the UploadLog.exe.
The Upload Log dialog is displayed:
Figure 7-108 Uploading Log Files
NOTE: In the event that the upload process fails, check your local network security settings to make
sure that they did not prevent the upload. Otherwise, contact GE personnel.
7-7-2-4 DB Scan
To Perform the Database Scan Procedure:
During this procedure, a log file which contains information about the integrity of the database will
automatically be created and stored in the Logs directory (in the folder name Log).
If required, this log file may be sent as an attachment when exporting log files. For more details, see
"Setting the Log File Attachments" on page 7-144.
Chapter 8
Replacement Procedures
Section 8-1
Overview
8-1-1 Purpose of Chapter 8
This chapter provides replacement procedures for the Vivid q N system parts, as outlined below.
Section 8-2
External Component Replacement Procedures
8-2-1 Bearing Handle Replacement Procedure
NOTE: The term “Bearing Handle” relates to the Vivid q N portable carrying handle.
8-2-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Bearing Handle replacement
procedure.
8-2-1-3 Preparations
None
Latch Latch
2.) Release the captive screw on either side of the Bearing Handle (there are two screws that secure
the handle in position), by turning 1/4 turn in a counterclockwise direction (Figure 8-2).
Figure 8-2 Releasing the Bearing Handle Captive Screw
Captive Screw
3.) Using both hands, pull the Bearing Handle slightly apart to release it from the system.
4.) Remove the Bearing Handle.
Figure 8-3 Bearing Handle Removed
NOTE: When tightening the captive screws, do not use excessive force do not apply more than
1/4 turn; overtightening will damage the screw thread.
8-2-2-1 Tools
None required.
8-2-2-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Latch Latch
NOTE: When removing A defective batterY, ensure that it IS disposed of in accordance with local
regulations. Alternatively, forward IT to GE for proper disposal.
CAUTION Use only batteries approved by GE as suitable for use with the Vivid q N portable ultrasound
scanner
1) With the Vivid q N lying face down (refer to Figure 8-5 on page 8-4), hold the unit steady with one
hand while sliding the battery sideways into the battery compartment.
2.) Make sure the battery is properly inserted all the way, before pushing the locking latch to the closed
position.
3.) Turn the Vivid q N over, release the front latches and open the LCD cover.
4.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
5.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Battery Parameters Test" on page 7-102
8-2-3-1 Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Hard Disk replacement procedure.
8-2-3-3 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
3.) With both hands, tilt the rear end of the scanner upwards then, holding it steady with one hand,
remove the Hard Disk from the system by holding the grip and pulling it straight out of the hard disk
module holder, as shown in Figure 8-7 below.
NOTE: When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access
codes already registered in the Hard Disk. Only qualified personnel should perform these tasks.
1) Place the new hard disk in position in the hard disk module holder - see Figure 8-7 on page 8-7.
2.) Return and fasten the two screws (previously removed) to secure the Hard Disk in position - refer
to Figure 8-6 on page 8-6.
NOTICE If the supplied SATA hard disk is not formatted, a Formatting Utility CD will be supplied as part of the
kit - perform the Hard Disk Formatting procedure as described in "Formatting the Hard Disk" on page
8-8.
NOTE: The formatting process limits the size of the partition “E:\” on the hard disk to approximately
120GB.
NOTICE For BT12 systems, use BT12 software media to perform the Hard Disk Formatting procedure.
To identify which hard disk is installed on your system, boot up the system and enter the BIOS
configuration as described in "Setting the BIOS" on page 8-114.
The hard disk is identified by the Primary Master: ST9500325AS-(PM) - see Figure 8-8.
NOTE: If you need to abort the process at this stage, prior to formatting, do so by pressing the On/Off
switch and holding down for a few seconds until the system shuts down.
6.) Select the appropriate formatting action corresponding with your system’s software installation.
7.) Select “Y” to continue with the formatting process or “N” to restart the procedure.
The system will start the formatting procedure and reboot automatically.
After rebooting, the system will boot from the software installation CD.
8.) Perform a “Full Clean Install” software procedure corresponding to the BT level of your system.
- Refer to the "Software Upgrade Procedure Overview" on page 8-99
Note: IMPORTANT - if both the Hard Disk and Back End are being replaced simultaneously, it is
necessary to re-configure the LCD type.
10.)After you have completed the formatting procedure, store the Formatting Utility Disk at the
customer’s site where it can be accessed for future use.
8-2-4-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the following Control Panel and
Keyboard (and sub-component) replacement procedures.
8-2-4-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
This screw
not captive
(do not
lose!)
4.) Turn the Vivid q N over, release the front latches and open the LCD Display cover.
5.) Carefully lift the Control Panel and Keyboard upwards from the front, then release it from the rear;
remove it from the Vivid q N and place it next to the system, as shown in Figure 8-15.
Figure 8-15 Removing the Control Panel and Keyboard
NOTE: When removing the Control Panel and Keyboard during other replacement procedures, Step 6
will only be required if indicated.
6.) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the control panel/keyboard cable from the system as shown in Figure 8-16.
Figure 8-16 Disconnecting the Control Panel/Keyboard Cable
Disconnect here
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the Control Panel and Keyboard Removal
Procedure.
NOTE: When installing the Control Panel and Keyboard during other replacement procedures, Step 1
will only be required if indicated.
1) Stand the replacement Control Panel and Keyboard beside the Vivid q N system and tilt it on its
side to gain access to the cable connector underneath. Re-connect the keyboard/control panel
cable from the system to the cable connector (refer to Figure 8-16, above).
2.) Carefully lift the Control Panel and Keyboard into position on the Vivid q N (refer to Figure 8-15,
above). Make sure it is properly seated and take care to align the screw holes with those in the
Bottom Assembly.
3.) Close the Vivid q N LCD Display cover, making sure both front latches are securely fastened.
4.) Carefully turn the Vivid q Nover, placing it face down.
5.) Return and fasten each of the 9 screws that secure the Control Panel and Keyboard onto the
Bottom Assembly (refer to Figure 8-14, above).
6.) Re-install the battery as described in the "Battery Installation Procedure" on page 8-5.
7.) Return the Handle as described in the "Bearing Handle Installation Procedure" on page 8-3.
8.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
9.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
- "Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
- "Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
- "Alphanumeric Keyboard Test" on page 4-19
- "External Keyboard Test" on page 4-20
- "Trackball Test" on page 4-22
- "Calibration Tests" on page 7-25
NOTE: Replacement Keycaps for the Vivid q N system are supplied as part of the Keycap Kit - Part No.
S2423359. (For Scandinavian countries, Part No. 2420372-1).
1) Carefully place a small flat screw driver between the keycap you wish to remove and its neighboring
keycap.
2.) Use the screw driver to gently lever off the defective keycap, until it is completely loosened from its
base (see Figure 8-17 below).
3.) Lift off and remove the defective keycap.
Figure 8-17 Keycap Replacement
NOTE: This section describes the replacement procedure for Trackball Assembly - Part No. S2423360.
1) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the keyboard/control panel cable from the system - refer to Figure 8-16 on page 8-11.
2.) Tilt the Control Panel and Keyboard on its side and loosen and remove the four screws (two on
either side of the bracket) that fasten the trackball assembly to the base of the Control Panel, as
shown in Figure 8-18 below.
Figure 8-18 Trackball Removal
3.) Disconnect the three trackball cable connectors from the trackball assembly (carefully pull them out)
and remove the trackball assembly from the Control Panel.
NOTE: This section describes the replacement procedure for ATGC Slider keycaps (supplied as part
of the Keyboard Assembly) - Vivid i n systems: Part No. S423357; Vivid q N systems: Part No.
S423358.
1) Carefully place a small flat screw driver between the ATGC slider keycap you wish to remove and
the surface of the keyboard.
2.) Use the screw driver to gently lever off the defective keycap, until it is completely loosened from its
base.
3.) Lift off and remove the defective keycap - see Figure 8-19 below.
Figure 8-19 ATGC Slider Keycap Replacement
NOTE: This section describes the replacement procedure for ATGC Rotary knob (supplied as part of
the Keyboard Assembly) - Vivid i n systems: Part No. S423357; Vivid q N systems: Part No.
S423358.
1) Carefully place a small flat screw driver between the ATGC Rotary knob and the surface of the
keyboard - see Figure 8-17 below.
2.) Use the screw driver to gently lever off the defective rotary knob, until it is completely loosened from
its base.
3.) Lift off and remove the defective knob - see Figure 8-20 below.
Figure 8-20 ATGC Rotary Knob Replacement
NOTE: For Vivid q N systems that previously did not have the Respiratory Option, it is necessary
remove the ECG Board from the ECG Module Assembly and replace it with an
ECG/Respiratory Board. The preparation and removal procedures below describe the required
steps.
Preparation
1) Access the D:\ directory and delete the file named traces.res, as shown below:
D:\Marlin\Test\resources\Nemo\acquis\HWConfig\traces.res
NOTE: In the event that the traces.res file is not deleted prior to installing the new ECG/Respiratory
board, the following error message will later be displayed:
2.) Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Removal Procedure
1) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the keyboard/control panel cable from the system - refer to Figure 8-16 on page 8-11.
2.) Loosen and remove the four screws (two long and two short) on either side of the bracket that fasten
the ECG Module assembly to the base of the Control Panel, as shown in Figure 8-21.
Figure 8-21 ECG Module Assembly Removal
3.) Lift and tilt the ECG Module on its side, then disconnect the ECG Module cable - see Figure 8-22.
Figure 8-22 Disconnecting the ECG Module Cable
Release
cable
here
4.) Carefully remove the ECG module assembly (board and bracket) from the base of the Control Panel
and place it on a table.
5.) Remove the tape from the Flex Cable on the ECG Board and disconnect the Flex Cable.
6.) Unfasten the two screws that secure the bracket to the ECG Board and remove the bracket.
7.) Fit the bracket to the new ECG/Respiratory Board, ready for the installation procedure.
NOTE: In the event that an error message is displayed prompting for deletion of the traces file, this
indicates failure to perform the Preparation procedure prior to installation of the new
ECG/Respiratory board. Refer to "Preparation" on page 8-17.
8.) Press Config and log on as Adm - see "Log On to the System as ADM" on page 4-10.
9.) Select About, then click on the HwVersion tab.
10.)Verify that the correct ECG/Respiratory module Part # is displayed, as shown in Figure 8-23.
Figure 8-23 Hardware Version Verification
8-2-5-1 Tools
None required.
8-2-5-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
7.) Lift the Rear Cover and Latch Assembly away from the Bottom Assembly and remove completely
from the system.
8-2-6-1 Tools
Phillips screwdriver; Loctite 242 glue
8-2-6-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in - System Setup.
5.) Detach the Rear Cover and Latch Assembly from the Bottom Assembly and remove.
6.) Loosen and remove the 2 screws that secure the left hinge to the LCD Display cover, then carefully
remove the hinge.
7.) Loosen and remove the 2 screws that secure the right hinge to the LCD Display cover, then
carefully remove the hinge.
NOTE: When installing the replacement hinges, apply a coating of Loctite 242 glue to each of the
8 screws, as and when these are returned to their respective locations.
NOTICE It is important to properly adjust the alignment of the LCD cover with the Bottom Assembly,
before returning the previously-removed parts, as described below.
1.) While supporting the LCD Display cover with one hand, carefully place a replacement hinge in the
correct position on the Left side of the LCD Display cover.
2.) Carefully place a replacement hinge in the correct position on the Right side of the LCD Display
cover.
NOTE: The cover has a degree of sideways free play i.e. it can slide from left to right on the hinges’
axle. The hinges on the right-hand side must be aligned with the holes in the Bottom Assembly
as explained in the next step.
3.) Gently maneuver the LCD cover by raising and lowering it while exerting pressure to slide the
LCD Display Cover until the hinge holes are aligned with the Bottom Assembly screw holes - see
right hinge in Figure 8-27 on page 8-23.
4.) On the Left hinge: return and fasten the 2 screws (previously removed) to the Bottom Assembly,
making sure the hinge is properly secured. Do not over-tighten the screws. Refer to Figure 8-27.
5.) On the Right hinge: return and fasten the 2 screws (previously removed) to the Bottom Assembly,
making sure the hinge is properly secured. Do not over-tighten the screws. Refer to Figure 8-27.
Figure 8-27 LCD Display Hinges Secured to Base Assembly - Location of Screws
6.) Check that when closing the LCD Display cover it aligns properly with the Bottom Assembly. If
necessary, make further adjustments.
7.) Return and fasten the 2 screws (previously removed) to each of the LCD hinges, making sure each
hinge is securely fastened to the Base Assembly. Do not over-tighten the screws.
8.) Reconnect the LCD flex cable to the BEP (refer to Figure 8-26 on page 8-22).
9.) Refit the Control Panel and Keyboard, as described in the "Control Panel and Keyboard Installation
Procedure" on page 8-12.
10.)Close and open the lid a few times to ensure proper closure of the LCD Display cover.
11) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
12.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
8-2-7-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-7-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in - System Setup.
8-2-8-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-8-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Figure 8-28 Removing the Screw Caps from the LCD Display Frame
3.) With the LCD Display in the fully open position, carefully release the LCD display frame from the
securing clips on all four sides, using both hands to release and lift it away from the LCD display.
Figure 8-30 Removing the LCD Display Frame
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the LCD Display Frame Removal Procedure.
1) With the LCD Display in the fully open position, return the LCD display frame to the correct position
and using both hands, press the frame under the securing clips on all four sides to fasten it against
the LCD display. Refer to Figure 8-30 on page 8-28.
2.) Return and tighten the four screws (previously removed).
3.) With the LCD Display in the open (upright) position, return the screw cap to each of the
four screws that fasten the LCD frame in position, as shown in Figure 8-28 on page 8-27.
4.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Section 8-3
Internal Component Replacement Procedures
NOTE: When performing Internal Component Replacement Procedures, generally it is not necessary
to disconnect and completely remove the Control Panel and Keyboard from the system. Simply
remove it from the Bottom Assembly and lay it next to the Vivid q N portable ultrasound
scanner, as shown in Figure 8-15 on page 8-11.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED
ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE
POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
8-3-1-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-3-1-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
4.) Loosen the two screws that fasten the P&M Board to the Bottom Assembly, as shown in
Figure 8-32 below.
Figure 8-32 Loosening the Screws from the P&M Board
5.) Lift the P&M Board upwards away from the Bottom Assembly and remove it completely, as shown
in Figure 8-33 below.
Figure 8-33 Removing the Probe and MUX Board from the Bottom Assembly
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the P&M (Probe and MUX) Board Removal
Procedure.
CAUTION When installing boards, ESD may cause damage to a board. Always have the ESD hand strip
connected to the machine chassis and to your hand.
1) Return a new P&M Board to the system, pushing it downwards into position on the Bottom
Assembly. Make sure it is properly seated before returning and fastening the two screws (previously
removed). Refer to Figure 8-33 and Figure 8-32, above.
2.) Return and fasten the four screws to secure the P&M Board to the RFI Board, as shown in
Figure 8-31 on page 8-30.
3.) Return the Control Panel and Keyboard, as described in the "Control Panel and Keyboard
Installation Procedure" on page 8-12
4.) Return the bearing handle, as described in the "Bearing Handle Installation Procedure" on page
8-3.
5.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
NOTE: When replacing the TR32 Boards, both boards must be of the same version. (A TR32 Board
V12 cannot be used with a TR32 Board V7). Always verify this by examining the version number
stated on the board, as shown in Figure 8-34, below.
TR32 Board
Version (V12)
Verification
NOTE: Before installing the TR32-V12 board, check that the DCDC V4 board (P/N 2403248-4) is
already installed. Always verify this by examining the Power Supply DCDC Board version
number stated in the About Screen (HwVersion tab), as shown in Figure 8-35, below. If the
DCDC V2 is installed, replace it with the DCDC V4 board.
DCDC Board
Version (V4)
Verification
CAUTION When removing the boards, ESD may cause damage to a board. Always have the ESD hand strip
connected to the machine chassis and to your hand.
8-3-2-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-3-2-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
5.) Disconnect the PCI cable flex (BEP-to-RFI) from the cable connector on RFI Board as shown in
6.) Disconnect each of the two HVPS cables from their sockets on DC Power Supply, as shown in
Figure 8-39 below.
Figure 8-39 Disconnecting the DC Power Supply
7.) Gently lift one side of the P&M Board upwards and hinge it open to gain access to the RFI and TR32
Board assembly beneath, as shown in Figure 8-41 below.
(The RFI and TR32 Boards are seated together in the system as one unit).
Figure 8-40 Accessing the RFI and TR32 Boards
8.) Lift the RFI and TR32 Board assembly upwards and remove it from the Bottom Assembly, as shown
9.) Remove the RFI Board from the TR32 Boards (separate by gently pulling them apart), as shown in
Figure 8-42 below.
Figure 8-42 Removing the RFI Board from the TR32 Boards
10.)Separate the two TR32 Boards from each other, as shown in Figure 8-43 below.
Figure 8-43 Separating the Two TR32 Boards from Each Other
NOTE: Replace the defective board, as described in the appropriate steps of the following procedure.
NOTE: This procedure applies to replacement of one or more of the defective boards (TR32A and/or,
TR32B and/or RFI).
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the TR32 and RFI Boards Removal Procedure.
CAUTION WHEN INSTALLING BOARDS, ESD MAY CAUSE DAMAGE TO A BOARD. ALWAYS
HAVE THE ESD HAND STRIP CONNECTED TO THE MACHINE CHASSIS AND TO
YOUR HAND.
a.) Apply a small quantity of heat sink paste (Dow Corning® 340 Heat Sink Compound, or similar)
to each of the eleven areas shown in Figure 8-44 (marked with arrows).
Figure 8-44 TR32 Board Showing Areas for Application of Heat Sink Paste
b.) Attach the TR32 Board to its respective heat sink pad by placing the heat sink on the
TR32 Board (the obverse side of the heat sink is shown in Figure 8-45).
Figure 8-45 TR32 Board Shown with Respective Heat Sink Pad
8-38 Section 8-3 - Internal Component Replacement Procedures
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
2) Re-connect the two TR32 Boards to each other (gently press together as shown in Figure 8-43 on
page 8-37).
3.) Connect the TR32 Boards to the RFI Board (gently press together as shown in Figure 8-42 on page
8-36).
4.) Return the TR32 and RFI Board assembly to the system, placing it in position on the Bottom
Assembly. (While performing this step, gently lift one side of the P&M Board upwards and hinge it
open in order to insert the TR32 and RFI Board assembly underneath). Refer to Figure 8-40 and
Figure 8-41 on page 8-36.
5.) Reconnect each of the two HVPS cables to the DC Power Supply (see Figure 8-39 on page 8-35).
6.) Reconnect the PCI cable flex (BEP-to-RFI) to the cable connector on RFI Board (see Figure 8-38
on page 8-35).
7.) Return the eight screws and fasten the TR32 and RFI Board assembly firmly in position on the
Bottom Assembly (refer to Figure 8-36 and Figure 8-37 on page 8-34).
8.) Return the Control Panel and Keyboard to its original position, as described in the "Control Panel
and Keyboard Installation Procedure" on page 8-12.
9.) Return the bearing handle, as described in the "Bearing Handle Installation Procedure" on page
8-3.
10.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
8-3-3-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-3-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
6.) Remove the Control Panel and Keyboard, as described in the "Control Panel and Keyboard
Removal Procedure" on page 8-10.
7.) Disconnect all the cables from the BEP.
NOTE: When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access
codes already registered in the Hard Disk. Only qualified personnel should perform these tasks.
Proceed as follows:
a.) Disconnect the LCD Flex ribbon cable (carefully pull back the two connector latches to
release).
b.) Disconnect the BEP-to-RFI Flex cable.
Fan/Speaker-to-BEP cable
8.) Unscrew and remove the four screws that secure the BEP to the Bottom Assembly as shown in
2
1
4
3
NOTE: Beneath one of the four screws removed in the previous step, is a small plastic securing tab
(illustrated in Figure 8-49) that serves to seat the BEP firmly in position. It is necessary to loosen
and release this tab (lift it up using a screw driver if necessary) in order to facilitate complete
removal of the BEP.
Figure 8-49 Releasing the Plastic Securing Tab from the BEP
Plastic
Securing Tab
9.) Using both hands, lift the BEP out of the Bottom Assembly and remove it from the system, as shown
NOTE: When removing the BEP for replacement with the SATA BEP, disconnect the control
panel/keyboard cable (Figure 8-51) from the parallel BEP you have just removed and keep this
cable for installing on the SATA BEP.
Control Panel
Keyboard Cable
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the BEP Removal Procedure.
NOTE: The SATA BEP has an adaptor bracket, which is required for its installation. The bracket is
shown on the left in Figure 8-52; in the picture on the right, the bracket is shown in position in
the Vivid q N scanner. Notice that the bracket sits on the PCMCIA slot-flap, which is pressed
open while the adaptor bracket is positioned on the flap itself. To install the adaptor bracket, the
flap is pressed open (Figure 8-53) and the adaptor bracket is placed in position. The procedure
is described below.
NOTE: Figure 8-54 shows a view from the rear of the Vivid q N showing the adaptor bracket closing off
the PCMCIA slot.
Figure 8-54 Rear View of the Vivid q N Scanner after SATA BEP Installation
VGA Connector
Hard Disk shown without Adaptor Bracket
its retaining screws (external view)
3.) Connect the control panel/keyboard cable (if previously removed from an ATA BEP; see Figure
8-51) to the SATA BEP.
Proceed to install the SATA BEP as described in the next procedure.
NOTE: When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access
codes already registered in the Hard Disk. Only qualified personnel should perform these tasks.
2.) Check that the BEP is seated correctly then return and fasten the four screws (previously removed)
to secure the BEP to the Bottom Assembly - refer to Figure 8-48 on page 8-42.
3.) Re-connect all the cables to the BEP:
Proceed as follows (refer to Figure 8-47 on page 8-41):
7.) Re-install the battery as described in the "Battery Installation Procedure" on page 8-5.
8.) Return the bearing handle as described in the "Bearing Handle Installation Procedure" on page 8-3.
9.) Turn ON the system and boot-up.
10.)Perform a complete functional check of the system, as described in Chapter 4 - General Procedures
and Functional Checks.
11.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the
Vivid q N system is currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
8-3-4-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
BIOS Battery
1) Fit a new BIOS battery in position on the BEP - refer to Figure 8-55.
2.) Re-install the BEP, as described in the "BEP Installation Procedure" on page 8-44.
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see "Date and Time Adjustments" on page 3-71.
5.) Proceed to perform the following functionality test:
- "Battery Parameters Test" on page 7-102
8-3-5-1 Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the HVPS replacement
procedure.
8-3-5-3 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
4.) Loosen the five screws that secure the DC Power Supply to the Bottom Assembly, as shown in
5.) Carefully lift the DC power supply upwards and remove it from the Bottom Assembly, as shown in
Figure 8-58 below.
Figure 8-58 Removing the DC Power Supply
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the DC-to-DC Power Supply Removal
Procedure.
1) Return the HVPS to its original position - refer to Figure 8-58, above.
2.) Return the five screws and fasten the HVPS in position on the Bottom Assembly, as shown in Figure
8-57 on page 8-49.
3.) Reconnect each of the two Power Supply cables to the DC Power Supply (HVPS-to-BEP and
HVPS-to-RFI), as shown in Figure 8-56 on page 8-48.
4.) Return the bearing handle, as described in the "Bearing Handle Installation Procedure" on page
8-3.
5.) Make sure the Vivid q N LCD Display cover is securely closed and place the scanner face down.
Re-install the battery as described in the "Battery Installation Procedure" on page 8-5.
6.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
8-3-6-1 Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the fan replacement procedure.
8-3-6-3 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
4.) Disconnect the speaker cable from the fan assembly, as shown in Figure 8-60 below
Figure 8-60 Disconnecting the Speaker Cable
5.) Using both hands, lift the Fan Assembly upwards from the front and then the rear to remove it from
the Bottom Assembly; lay it down adjacent to the system, ready for the next step as shown in
Figure 8-61 below.
6.) Disconnect the Fan cable from the BEP and completely remove the Fan Assembly from the system,
as shown in Figure 8-62 below.
NOTE: When performing the following installation procedure steps, it may be helpful to review the
appropriate movie clips previously referenced in the Fan Removal Procedure.
1) Connect the replacement Fan cable to the BEP- refer to Figure 8-62, above.
2.) Using both hands, carefully mount the replacement Fan Assembly into position on the Bottom
Assembly - refer to Figure 8-61 on page 8-52.
3.) Make sure the Fan Assembly is seated in the correct position. Return and fasten the two screws
(previously removed) to secure the Fan Assembly to the Bottom Assembly, as shown in
Figure 8-59 on page 8-51.
4.) Reconnect the speaker cable to the fan assembly, as shown in Figure 8-60.
5.) Return the Control Panel and Keyboard, as described in the "Control Panel and Keyboard
Installation Procedure" on page 8-12.
6.) Return the bearing handle, as described in the "Bearing Handle Installation Procedure" on page
8-3.
7.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
8-3-7-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrench
keys as indicated in the speaker replacement procedure.
8-3-7-3 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
NOTE: The AOU LCD PN5461952 kit is delivered with the LCD monitor assembled in the display frame
as an integrated unit.
NOTE: If replacing an old-type LCD Monitor with an AUO LCD Monitor, it is necessary to uninstall the old-type
hinges and install a pair of new-type hinges (P/N KTZ280195), as shown in Figure 8-64.
NOTE: Prior to installing the AUO LCD monitor, a replacement DC power supply, P/N 5439596 should be
available if this is required. See Note on page 8-58.
8-3-8-1 Tools
Use the appropriate Phillips and flat screw drivers, and Loctite 402 glue, as indicated in the AUO LCD
Display replacement procedure.
8-3-8-3 Preparation
1) Make sure all required tools and parts are available.
2.) Adhere the appropriate Vivid label onto the new display frame.
3.) Remove the display frame from the new AOU LCD Display (to provide access to the hinges).
4.) Shut down the Vivid q N ultrasound unit, as described in - System Setup
3.) Slide to open the two connector latches and disconnect the LCD Flex ribbon cable from BEP - refer
to Figure 8-65.
5.) Detach the Rear Cover and Latch Assembly from the Bottom Assembly and remove.
NOTE: When performing this procedure, care must be taken to correctly attach the Display Frame on the
Bottom Assembly hinges.
NOTE: Prior to inserting the hinge securing screws in their respective locations, apply a coating of
Loctite 402 glue to each of the hinge screws (supplied).
1.) Protect the Bottom Assembly with a protective cloth, as shown in Figure 8-67.
Note: Check that there is a coating of Loctite 402 on the screw threads, but do not tighten the
screws completely - see Figure 8-68.
Flex Cable
Figure 8-68 Mounting the AUO LCD Display Assembly onto the Bottom Assembly
4.) Gently pull the LCD display upwards and fully tighten the hinge screws on both the left and right
sides of the display.
5.) Re-connect the flex cable (indicated in Figure 8-68).
6.) Fit the display frame onto the AOU LCD Display and cover the securing screws with the screw-caps
provided.
7.) Verify that the Vivid q N lid opens and closes freely and check that there is a “paper-thin” gap
between the top rear cover and the keyboard surface as shown in Figure 8-69.
Figure 8-69 Checking the Gap Between the Vivid q NLid and the Bottom Assembly
8.) Fully tighten the hinge screws.
9.) Return the Keyboard Assembly to its location on the Bottom Assembly, as described in "Control
Panel and Keyboard Installation Procedure" on page 8-12.
10.)Refit the bearing handle, as described in the "Bearing Handle Installation Procedure" on page 8-3.
11) For the final stage of the LCD Display installation procedure, proceed to calibrate the
newly-installed LCD Display by making any adjustments required to the Brightness, Contrast
and/or Blue Tint, as described in the "LCD Display Adjustments and Calibration" on page 6-1.
12.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Full System Test" on page 7-10
NOTE: In the event that the installed AUO LCD monitor flickers after installation, the DC Power Supply should
be replaced with a new DC Power Supply, P/N 5439596. Follow the instructions provided in the
"DC-to-DC Power Supply Replacement Procedure" on page 8-48.
Table 8-2 AUO LCD Cable and Flex RSPL Kit (P/N 5443867) Contents
Flex Board 1
Ferrite core 1
NOTE: For reference, the relevant parts are illustrated in Table 8-3, below:
Table 8-3 AUO LCD Cable and Flex RSPL Kit - Illustrated Parts
Item # as Labeled in
Part Illustration Figures and Referenced
in Procedures
8-3-9-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the AUO LCD Flex Cable replacement
procedure.
8-3-9-3 Preparation
Shut down the Vivid q N ultrasound unit, as described in - System Setup.
NOTE: The numbers shown in square brackets correspond to the numbers listed in Table 8-3.
1.) Remove the AUO LCD display - refer to the "AUO LCD Display Removal Procedure" on page 8-55.
2.) Disconnect the AUO LCD cable [3] from the rear of the LCD panel.
3.) Disconnect and remove the flex board [4] by unscrewing the two retaining screws encircled in
Figure 8-70.
Disconnect
here
1
3
2
4 5
Flex Board LCD Backlight
Inverter
Figure 8-70 Flex Board and LCD Back Light Inverter Board
4.) Disconnect the Inverter Flex Board cable [2] from the LCD Backlight Inverter board [5], then
disconnect the LCD Back Light Inverter board from the AUO LCD cable [3] (arrow at right in
Figure 8-70).
5.) Disconnect the Flex Board from the LCD flex [1] - Figure 8-70.
6.) Gently remove the Flex Board and the LCD Back Light Inverter Board.
7.) Disconnect the LCD flex [1] from the rear panel of the LCD Display and remove - Figure 8-71.
NOTE: The numbers shown in square brackets correspond to the numbers listed in Table 8-3 on page 8-59.
1.) Connect the LCD Flex to the LCD as shown in Figure 8-71.
2.) Thread the Flex Board cable through the Display Flex Cover [6] making sure that the Display Flex
Cover is correctly orientated as shown in Figure 8-72.
Figure 8-72 Flex Board Cable Threaded through Flex Board Cover and Ferrite
3.) Thread the Flex Board cable through the ferrite core [7] and secure the ferrite with a strip of Kapton
tape - Figure 8-72.
4.) Connect the Inverter Flex Board cable [2] to the Flex Board [4], then connect its other end to the
LCD Backlight Inverter board [5] - Figure 8-70.
5.) Connect the LCD Backlight Inverter board [5] to the AUO LCD cable [3] - Figure 8-70.
NOTE: To install the AUO LCD display in the rear cover (the system lid), the flex cable must be prepared by
folding as shown in Figure 8-73. Folding the flex cable in this manner enables the LCD panel to lie flat
in the lid housing.
20 cm
Tip: Fold here at +/- 20 cm from
right end
8.) Place the AUO LCD Display in position in the system’s top rear cover (the system lid) - Figure 8-75. .
Figure 8-75 AUO LCD Display located in the Vivid q N Top Rear Cover
9.) Position the LCD Backlight Inverter board on the retaining pins located in the system’s rear cover.
10.)Secure the Flex Board with the two retaining screws as show in Figure 8-75.
11.)Refit the AUO LCD Display in the system’s top rear cover as described in the "AUO LCD Display
Installation Procedure" on page 8-57.
12) Check that the LCD Display can be tilted upwards and downwards, and that it holds its set position.
13.)Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Full System Test" on page 7-10
Section 8-4
SafeLock Cart Components Replacement
NOTE: This section describes replacement procedures for a new-type SafeLock Cart only.
WARNING THIS EQUIPMENT MUST NOT BE OPERATED WITHIN 15 FEET OF ANY MAGNETIC EQUIPMENT.
DANGER Electrical hazards exist at several points in the system. Familiarize yourself with
all hazardous voltages and high current levels before removing any of the
covers.
DANGER DO NOT wear the ESD wrist band strap when removing parts from the power
supply unit. Before removing any part of the power unit, turn the power OFF and
disconnect the power cord.
WARNING BEFORE REMOVING CIRCUIT BOARDS, TURN THE POWER OFF AND WEAR THE ESD WRIST
BAND STRAP.
2
1
12 8
9
10
11
13 14
8-4-2-1 Tools
Appropriate Phillips screwdriver.
8-4-2-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Retaining Screw
8-4-3-1 Tools
Appropriate Phillips screwdriver.
8-4-4-1 Tools
Appropriate Phillips screwdriver and Loctite glue.
NOTE: If this is a first-time installation of the bumper protector, it is necessary to remove the two
existing Phillips screws located on the rear of the SafeLock Cart (AC Cable connection area),
as shown in Figure 8-79.
1.) Apply a little Loctite glue on the threads of each of the two shoulder screws (Figure 8-80).
Figure 8-80 Shoulder Screws
2.) Insert a shoulder screw into each screw hole (refer to Figure 8-79).
8-68 Section 8-4 - SafeLock Cart Components Replacement
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3.) Mount the AC Power Cable Bumper Protector onto the shoulder screws by aligning the mounting
tabs (ears) with the screws, then pushing the bumper downwards until it is seated securely.
See Figure 8-81.
Figure 8-81 AC Power Cable Bumper Protector Mounted on SafeLock Cart
8-4-5-1 Tools
Appropriate Phillips screwdriver.
1.) Thoroughly clean the surface of the Top Shelf Assembly, in particular the recessed area from where
the ECG Cable Protector (or plastic guide holders) were removed.
2.) Position the ECG Cable Protector on the Top Shelf Assembly surface, ensuring that the recessed
groove on the protector aligns with the groove on the SafeLock Cart, as shown in Figure 8-83.
CAUTION When placing the ECG Cable Protector in position, take care not to damage the plastic.
IMPORTANT: Perform the next steps in the correct order. When securing the ECG Cable Protector, first fasten the rear
securing screw, then the front securing screw.
3.) Return and fasten the rear securing screw (Figure 8-84), then the front securing screw.
Figure 8-84 Fastening the Rear Securing Screw
Figure 8-85 illustrates the ECG Cable Protector secured in position on the SafeLock Cart.
ECG Cable
Connected to Scanner
8-4-6-1 Tools
Appropriate Phillips screwdriver.
8-4-6-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Retaining Screw
8-4-7-1 Tools
Appropriate flat and Phillips screwdrivers and a wire cutter
8-4-7-2 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Retaining Screw
8) Loosen and remove the ground screw fastened to the AC Distribution Assembly - see Figure 8-88.
Figure 8-88 AC Distribution Assembly and Cable Sockets
Peripheral power
cable Ground Screw
Jumper
AC Assembly mounting
screw
AC Assembly mounting
screw
9) Loosen and remove the two AC Assembly mounting screws that secure the AC Distribution
Assembly to the mounting bracket on the SafeLock Cart chassis - see Figure 8-88.
10) Carefully pull the AC Distribution Assembly out towards you and remove it remove from the
SafeLock Cart.
8-4-8-1 Tools
Appropriate Phillips screwdriver and wire cutters.
8-4-8-2 Preparation
1) Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
2.) Remove the Upper Cover (rear) and Lower Cover (rear) of the SafeLock Cart as described in
"Upper Cover (Rear) Removal Procedure" on page 8-66 and "Lower Cover (Rear) Removal
Procedure" on page 8-73.
NOTE: Before starting the routing procedure adhere two plastic cable holders to the metal linear guide
bracket as shown in Figure 8-90 (middle picture).
1.) Insert the replacement Peripheral Power cable and power socket in position in the SafeLock’s
Upper Cover (front) (see item 4 in Figure 8-76 page 8-65).
2.) With the AC power socket plugged into the cart (Figure 8-90, left picture), connect the ground cable
to the body of the cart.
3.) Route the Peripheral AC Outlet cable as shown in the middle picture of Figure 8-90.
4.) Secure the cable with the cable clamp in the direction shown in Figure 8-91.
Figure 8-91 Correct Installation of Cable Clamp
5.) Route the cable and the ground cable over the plastic cable holders (Figure 8-90 middle picture)
and secure the cables with the tie-wraps.
6.) Continue routing the Peripheral AC Outlet cable towards the right hand side of the Rear Upper
Cover and downwards - Figure 8-92.
7.) Fasten the Peripheral AC Outlet cable to the Main Harness cable as shown in Figure 8-93.
8.) Plug the Peripheral AC Outlet cable into the power supply and connect the Ground lead to the body
of the cart - Figure 8-94.
10.)Shorten the USB power cable by folding it into a “u” shape and fasten together with tie wraps as
shown in Figure 8-95.
Figure 8-95 USB Power Cable Folded and Fastened with Tie-wraps
11.)Check that the routing procedure has been implemented correctly: lower the cart’s top-shelf and
make sure that the cables do not interfere with the movement of the cart’s top shelf.
NOTE: The movement should be smooth and there must be no pinching of the cables.
Slide the top-shelf up and down a number of times and check that the cables do not disturb the
up-down movement of the cart’s top-shelf - see Figure 8-96.
12.)Return the Upper Rear Cover as described in Service Manual Direction R2423163-100, Revision
3, Chapter 8, section 8-4-3-4.
13.)Return the Upper Cover (rear) and Lower Cover (rear) of the SafeLock Cart as described in "Upper
Cover (Rear) Installation Procedure" on page 8-66 and "Lower Cover (Rear) Installation Procedure"
on page 8-73.
14.)Once more, slide the top-shelf up and down a number of times and check that the movement is
smooth and that there is no interference of the cables with the up-down movement of the cart’s
top-shelf.
15) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
8-4-9-1 Tools
Appropriate Phillips screwdriver and wire cutters.
8-4-9-2 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Remove the Upper Cover (rear) and Lower Cover (rear) of the SafeLock Cart as described in "Upper
Cover (Rear) Removal Procedure" on page 8-66 and "Lower Cover (Rear) Removal Procedure" on
page 8-73.
8-4-10-1 Tools
Appropriate socket wrench
8-4-10-2 Preparation
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
4.) Remove the upper and lower bolts that attach the USB board to the upper frame, as shown in
Figure 8-97.
5) Remove the USB board from the SafeLock Cart.
8-4-11-1 Tools
Appropriate Phillips screwdrivers, Allen key and open wrench.
8-4-11-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8) Remove the Gas Spring by turning it counter-clockwise until it is released from the Gas Spring
Handle assembly.
NOTE: To remove the Gas Spring handle, unscrew the upper Gas Spring securing screw.
NOTE: When packed, the Gas Spring replacement part is fully extended. The spring must be
compressed prior to installation.
1.) Compress the Gas Spring by pressing downwards on the Gas Spring with its pressure release
button pressed against a solid surface until the Gas Spring is full compressed.
Solid Surface
2) Turn the Gas Spring locking nut counter-clockwise until it rises about a quarter-way upwards on the
thread.
3) Insert the Gas Spring handle into the Gas Spring Housing Assembly until it is flush with the outer
cylindrical surface of the housing assembly.
Figure 8-100 Gas Spring Holder Inserted Flush with Housing Assembly
Lock Nut
4) Attach the Gas Spring to the housing assembly by turning it clockwise until the locking nut makes
contact with the housing assembly - see Figure 8-100.
5) Align the lower Gas Spring connecting bracket with the SafeLock chassis until it is aligned with the
screw aperture on the chassis.
Figure 8-101 Lower Gas Spring Securing Bracket
6) Insert the lower Gas Spring connecting screw and tighten it securely.
7) Replace the Main Harness cable by coiling it once around the Gas Spring and then reconnect it to
the AC unit.
8) Secure all cables in position with new cable ties.
9) Return the Lower Cover (Rear), as described in the "Lower Cover (Rear) Installation Procedure" on
page 8-73.
10) Return the Upper Cover (Rear), as described in the "Upper Cover (Rear) Installation Procedure" on
page 8-66.
11) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
8-4-12-1 Tools
Appropriate Phillips screwdriver.
2.) Remove the DVD Holder from the base of the Top Shelf.
8-4-13-1 Tools
Appropriate Phillips screw driver.
8-4-14-1 Tools
Appropriate Phillips screwdrivers, Allen key and open wrench.
8-4-14-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8-4-15-1 Tools
Appropriate Phillips screwdrivers.
8-4-15-2 Preparations
Shut down the Vivid q N ultrasound unit.
Handrest
Securing screws
Main Shelf
8-4-16-1 Tools
Appropriate Phillips screw driver.
8-4-16-2 Preparations
Shut down the Vivid q N ultrasound unit.
3.) While supporting the shelf with one hand, loosen and remove the two upper screws (one on
either side - see Figure 8-108); remove the shelf.
Figure 8-108 Probe Shelf Assembly - Upper Securing Screws
Upper Screws
8-4-17-1 Tools
Appropriate Phillips screwdriver.
8-4-18-1 Tools
Appropriate flat screw driver and open wrench.
8-4-18-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Brake
8-4-19-1 Tools
Appropriate flat screw driver and open wrench.
8-4-19-2 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
Section 8-5
Software Loading
8-5-1 Software Upgrade Procedure Overview
NOTE: The software installation/upgrade procedure to software version BT12 can be performed on any
Vivid q N system currently running software BT11.
NOTE: For Vivid q N systems with software version BT12, a DVD Drive is an option. BT12 software installation
is available only through the USB Flash Drive.
NOTE: For systems running a software version earlier than BT11, it is necessary to first perform the
required software upgrade procedures (in the correct sequential order) up to software version
BT11. Then, follow the instructions for "Vivid q N Software Upgrade Overview" on page 8-100.
NOTE: For all Vivid q N systems, before upgrading the system software, it is necessary to contact your local
OTR department to receive an appropriate software password. Make sure you specify the system serial
number located on the base of the Vivid q N unit.
NOTICE The Software Installation/Upgrade procedure cannot be performed when the Vivid q N portable
ultrasound scanner is mounted on/connected to the SafeLock Cart.
NOTICE During software installation, using CD or DVD, the DVD/CD-RW must be connected directly to
the Vivid q N portable ultrasound scanner and not through the SafeLock Cart or USB outlets.
WARNING DO NOT ATTEMPT TO INSTALL SOFTWARE THAT WAS NOT DESIGNATED FOR YOUR
VIVID Q N UNIT. ATTEMPTING TO INSTALL UN-APPROVED SOFTWARE WILL CAUSE
IRREVERSIBLE DAMAGE TO HARDWARE AND SOFTWARE!
Normally, the Vivid q N system is supplied from the manufacturer with the software already installed. In some
cases, it is necessary to re-install the software.
NOTICE Before performing a Vivid q N Software Upgrade from Version BT11 to BT12, it is mandatory to backup
the Archive, then format and partition the Hard Disk (due to partitioning changes). Follow the
instructions as described in 8-5-1-1-1 "Preparation" on page 8-101.
NOTE: The PMC software is automatically updated with the system upgrade.
NOTE: The Vivid q N BT11 to BT12 upgrade software can only be purchased by the customer through
the Sales Dept, using the appropriate ordering number (HCat #), as listed in Table 8-5.
HCAT # Description
NOTE: After performing an upgrade, all logs are deleted. It is recommended that these should be
recorded should they be needed in the future.
8-5-1-1-1 Preparation
NOTE: Make sure that the required Software Option Keys are available prior to commencing the
upgrade process. Do not use the Software Option Keys from the previous BT version.
WARNING REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
BEFORE STARTING THE UPGRADE PROCEDURE.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the “Copy” option
(for instructions, refer to the Vivid q N User Manual on your document media).
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Vivid q N User Manual on your document media).
3.) Write down the following Connectivity settings as it will be necessary to restore them at the end of
the process:
- TCP/IP address
- Network printer (if present)
- Wireless network settings (if present)
- Trackball speed
- InSite ExC settings
4.) Modify the BIOS settings, as described below in Modifying BIOS Settings prior to Software
Installation.
5.) Proceed to perform the "Formatting and Partitioning the Hard Disk - Automatic Procedure (BT’12)"
on page 8-124.
6.) When done, continue to Table 8-6 on page 8-103 and follow the Vivid q N Software
Upgrade/Installation/Reinstallation Procedure - BT11 to BT12 step-by-step instructions.
NOTE: If USB HDD is in the Excluded menu (as shown below), move it to the Boot priority order list by selecting
USB HDD and pressing the x keyboard key.
With USB HDD now listed in Boot priority order (as shown below), promote it to the second position in
the list by selecting it, then while pressing and holding <Shift>, repeatedly pressing the + button until
USB HDD is in the number 2 position (Figure 8-112).
4.) Click the Exit menu and select Exit Saving Changes.
5.) When prompted, click Yes to confirm and exit.
NOTE: The following upgrade procedure is based on the Vivid i n system. The procedure is the same
for the Vivid q N system (displayed screens will of course reflect Vivid q N information).
NOTE: During the software installation procedure, it is important to observe the USB Flash Drive connected to
the rear panel and make sure the LED is blinking.
Note:
For a Software Reinstallation procedure, Hard Disk
Formatting is optional.
8. Press [Yes] to continue. The following message is displayed, providing the option to reboot the system
manually (if preferred) when software installation complete:
User documentation files (User Manual in all languages) are then copied:
10. Select the following option: The following Warning message is displayed:
Reboot Machine
A series of progress bars, blank and additional screens are displayed, shown
in these examples:
NOTE: This procedure is valid for Vivid q N systems running BT12 software later than Build 31.
As described in the procedure below, it is not necessary to reinstall the software since the option is
provided to load only the latest version of the User Manuals (in all languages).
Table 8-7 Loading Latest Version Vivid q N User Manuals
5. Select the following option: File loading starts. While in progress, the following screen is displayed:
Load Installation
Note:
Since the existing User documentation is recognized by
the system, an additional button Update User Manual
on the right provides the option to load that latest version
User Manual (without performing software installation)
6.) Set correct System Time and System Date. Set Date using next format MM/DD/YYYY
7.) Check that IDE Primary Master is detected.
8.) Proceed to "Setting the BIOS Advanced Parameters" on page 8-115.
1.) From the BIOS main menu (see Figure 8-114), select the <Advanced> menu and press <Enter>.
3.) Set the parameters for Enable memory gap: as shown above in Figure 8-116.
4.) Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
5.) Scroll to PCI/PNP Configuration and press <Enter> to open the PCI/PNP Configuration screen.
Figure 8-117 BIOS Advanced PCI/PNP Parameters
6.) Set the parameters in the PCI/PNP Configuration screen as shown above in Figure 8-117.
7.) Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
8-116 Section 8-5 - Software Loading
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8.) Scroll to Memory Cache and press <Enter> to open the Memory Cache screen.
Figure 8-118 BIOS Advanced Memory Cache Parameters
9.) Set the parameters in the Memory Cache screen as shown above in Figure 8-118.
10.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
11.)Scroll to I/O Device Configuration and press <Enter> to open the I/O Device Configuration
screen.
Figure 8-119 BIOS Advanced I/O Device Configuration Parameters
12.)Set the parameters in the I/O Device Configuration screen as shown above in Figure 8-119.
13.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
14.)Scroll to Keyboard Features and press <Enter> to open the Keyboard Features screen.
Figure 8-120 BIOS Advanced Keyboard Features Parameters
15.)Set the parameters in the Keyboard Features screen as shown above in Figure 8-120.
16.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
17.)Scroll to Hardware Monitor and press <Enter> to open the Hardware Monitor screen.
Figure 8-121 BIOS Advanced Hardware Monitor Parameters
18.)Set the parameters in the Hardware Monitor screen as shown above in Figure 8-121.
19.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
20.)Scroll to Watchdog Settings and press <Enter> to open the Watchdog Settings screen.
Figure 8-122 BIOS Advanced Watchdog Settings
21.)Set the parameters in the Watchdog Settings as shown above in Figure 8-122.
22.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
23.)Scroll to Display Control and press <Enter> to open the Display Control screen.
Figure 8-123 BIOS Advanced Display Control Parameters
24.)Set the parameters in the Display Control screen as shown above in Figure 8-123.
Chapter 8 - Replacement Procedures 8-119
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
25.)Press <Esc> to return to the Advanced menu home screen (Figure 8-115).
26.)Scroll to Miscellaneous and press <Enter> to open the Miscellaneous screen.
Figure 8-124 BIOS Advanced Miscellaneous Parameters
27.)Set the parameters in the Miscellaneous screen as shown above in Figure 8-124.
28.)Proceed to "Additional BIOS Settings" on page 8-121.
1.) From the BIOS main menu (see Figure 8-114), select the <Advanced> menu and press <Enter>.
2.) Select the <Security> menu and press <Enter>.
Figure 8-125 BIOS Security Parameters
3.) Set the parameters in the Security screen as shown above in Figure 8-125.
4.) Press <Esc> to return to the main BIOS screen (Figure 8-114 on page 8-114).
6.) Set the parameters in the Power screen as shown above in Figure 8-126.
7.) Press <Esc> to return to the main BIOS screen (Figure 8-114 on page 8-114).
8.) Select the <Boot> menu and press <Enter>.
Figure 8-127 BIOS Boot Parameters
9.) Set the parameters in the Boot parameters screen as shown above in Figure 8-127.
10.)Press <Esc> to return to the main BIOS screen (Figure 8-114 on page 8-114).
12.)Reboot the system and check that the Supervisor PASSWORD was installed properly.
13.)Proceed to perform the following functionality tests:
- "Software Configuration Checks" on page 4-39
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
8-5-3 Formatting and Partitioning the Hard Disk - Automatic Procedure (BT’12)
NOTE: This procedure is applicable to Vivid q N systems running software version BT12.
Table 8-8 Formatting and Partitioning the Hard Disk - Automatic Procedure 1 of 2
5. Select the following option: A short series of screens displays information while information is being read
Format Hard Disk from the Flash Drive.
When complete, the following message is displayed:
then press <Enter>.
6. Press [Y] to confirm the Hard Disk formatting command. A short series of screens are displayed. Thereafter, the screen appears blank.
After a few minutes, the following message is displayed:
Note: Pressing [N] would evoke display of the message
Press any key to Reboot. After a system reboot, the
message show in Step 4 above would be redisplayed.
Table 8-8 Formatting and Partitioning the Hard Disk - Automatic Procedure (cont’d) 2 of 2
Section 8-6
Performing a Complete System Back-up
NOTICE IMPORTANT: Contact the customer and request that a full incremental back-up be performed prior to
upgrade. If necessary, refer them to the Vivid q N User Manual. Ask for removal (delete) of un-necessary
patient information, so that the back-up process will require minimal time.
Remind customers that GE strongly recommends that periodic back-ups of their Archive be performed.
This routine back-up process will assure the retention of all the information that is temporarily saved
into the local Archive (hard-disk). The permanent Archive is the backed-up Archive on the variety of
medias.
Do not proceed with the upgrade process if a complete incremental back-up process has not
been performed. Contact the On-Line Center to resolve any back-up process issues, before
proceeding with the upgrade. Ignore this note if back-up of patient information is not required, as per
the customer’s instructions.
8-6-1-1 Preparation
1.) If not already connected, connect the DVD\CD-RW to the USB port at the rear of the system and
connect power to the DVD\CD-RW drive.
2.) Press the <Eject> button of the DVD\CD-RW drive to verify that no disc is present in the disk drive.
NOTE: See Disk Management in the Vivid i n or Vivid q N User’s Manual on your document CD.
NOTE: The Operator Logon screen only appears after the first boot up of the system.
8.) Under Destination Device (see Figure 8-130), select the required media device.
9.) Click Start and follow the on-screen wizard instructions.
The system stores the images.
NOTE: If the Export button is not displayed, click More consecutively until the Export button is
displayed:
The following warning (Figure 8-135) is displayed to indicate that only the database is exported to
the target media.
6.) From the Export screen, select the required destination media (Figure 8-136) and click OK.
The Export Patient List screen opens - Figure 8-137.
7.) Click the soft-keys at the bottom of the screen consecutively in the following order as the system
processes the actions: Select All, Copy, and Done - see Figure 8-138.
NOTE: This procedure is applicable when importing files from various media devices.
NOTE: If the Import button is not displayed, click More consecutively until the Import button is
displayed.
NOTE: A warning is displayed (Figure 8-141) - to indicate that only the database is imported to the
target device.
6.) Under “From” in the Import screen, select the required media source from the drop-down list (see
Figure 8-141).
NOTE: If an Import / Export Conflict screen displays (this could occur if two separate examinations (this
could occur if two separate examinations were saved under the same name), click OK and
continue.
7.) Click the soft keys Select All, Copy, and Done, consecutively in that order as the system processes
the actions - see Figure 8-142.
NOTE: Wait while the system processes each action before pressing the next soft-key.
Section 8-7
Peripherals
NOTE: This section provides references to the installation procedures for the various peripherals used
with the Vivid q N portable ultrasound scanner. Connection to the system is via one of the USB
ports (unless otherwise stated) on the rear panel, as shown in Figure 8-143 below.
Following the connection and removal of peripherals, perform the following functionality tests:
• "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is currently
mounted on the SafeLock cart
• "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
NOTE: For details on connecting and configuring printers, refer to "Connecting Printers" on page 3-21.
NOTE: This section describes removal and replacement procedures, including initial configuration.
8-7-2-1 Tools
Appropriate flat screwdriver and open-ended wrench.
8-7-2-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
NOTE: Ensure that there is no activity on the drive before you turn it off.
If power is turned off while files are being written to the drive, the files will be corrupted.
NOTE: If the External USB Hard Disk is filled with data, store it in a safe place and label it with the date,
and a note that the drive is full.
Figure 8-145 External USB Hard Disk - Rear View and Cables
Power IN Cable
Power OUT Cable
WARNING IN THE EVENT THAT THE AC/DC POWER CONVERTER AND CABLES HAVE BEEN SUPPLIED
WITH THE POWER IN CABLE PART # 066E0661 (COLOR WHITE), DISCARD THIS CABLE AND
DO NOT USE.
ORDER THE CORRECT POWER IN CABLE PART # 5400905 (COLOR BLACK).
4.) Verify that both MODE switches are set to the ON position as illustrated
NOTE: The total RAID 1 capacity is half the physical capacity of the two drives.
CAUTION CHANGING THE VOLUME CONFIGURATION WILL ERASE ALL EXISTING DATA!
IF THE DRIVE CONTAINS FILES YOU WANT TO SAVE, MOVE THEM TO ANOTHER DRIVE
BEFORE CHANGING THE SWITCH SETTINGS.
5.) Connect the AC power cable from the power supply to the AC power outlet on the SafeLock Cart.
6.) Switch ON the power.
7.) Study the status light indicators and verify that they function as described in Figure 8-146.
NOTE: It is important that the External USB Hard Disk has a unique volume name. The site should
choose a naming convention to use for marking this Hard Drive and any future hard drives, in
case more than one drive will be needed in the future.
NOTE: Only connect one USB storage device (like this External USB Hard Disk) to the Vivid q N
scanner at a time.
1) Connect the USB cable from the External USB Hard Disk to a laptop or PC.
2.) Power up the External USB Hard Disk.
3.) Under My Computer, after a while, the disk will show as a new drive, named Removable Disk
- refer to Figure 8-147 on page 8-138. (It will be allocated the next available drive letter.)
4.) Right-click the drive name (Removable Disk) and select Rename from the menu.
5.) Enter the name you have decided to use for the drive.
6.) When done, click somewhere outside the name. The name will then be changed.
7.) Record the Volume’s name on a sticker or a piece of paper and attach it to the drive so that it will
be easier to locate the correct drive in the future.
8.) Use the “Safely Remove Hardware” tool in the Task Bar before continuing with the next step.
9.) Disconnect the USB cable from the laptop or PC.
8-7-2-7 Connecting the External USB Hard Disk to the Vivid q N Scanner
1) Connect the USB cable to a USB port on the Vivid q N scanner.
2.) Turn ON power to the Vivid q N ultrasound scanner.
3.) Under My Computer (refer to Figure 8-147 on page 8-138), verify that the drive appears for
selection.
4) Proceed to perform the following functionality tests:
- "External USB Hard Disk Test" on page 4-27
Chapter 9
Renewal Parts
Section 9-1
Overview
9-1-1 Purpose of Chapter 9
This chapter gives you an overview of replacement parts for the Vivid q N.
Section 9-2
List of Abbreviations
• Assy - Assembly
• BEP - Back End Processor
• Ctrl - Control
• FEP - Front End Processor
• FRU 1 - Replacement part available in parts hub
• FRU 2 - Replacement part available from the manufacturer (lead time involved)
• LCD - Liquid Crystal Display
• Int - Internal
• I/O - Input/Output
• PWA - Printed Wire Assembly
• Recv - Receive
• XFRMR - Transformer
Section 9-3
Mechanical Hardware Parts
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1. KEYCAP KIT VIVID q N BT11 S2423463 BT’11 and higher. 1 Y
2. Internal Battery Pack 5423275 Replaces: 2378963-2, S2423267, FL200027 and FL200068 1 Y
For Vivid q N
For Vivid q N
Section 9-4
Electronic Parts
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1. Vivid q IR KB RSPL Kit 5503081 1 Y
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
For Vivid q N.
11. ECG Module Assy N5448727 1 Y
Replaces NR2421861.
For Vivid q N
12. TR32-v12 Board N5444671 1 Y
Replaces NS2423369
For Vivid q N
13. Probe and MUX Assy N5452376 1 Y
Replaces N5420885
Section 9-5
Cables
Section 9-6
Software
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1. SW and Doc disk on key (BT12) N5444462 Software for Vivid q N - BT’12 1 Y
Section 9-7
Probes
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
2290776
1. Probe: 3C - RS 1 Y
H40402LL
2. Probe: 4C - RS 5131629 1 Y
3. Probe: 8C - RS 2354971 1 Y
2376127
5. Probe: 8L RS 1 Y
H40402LT
2355686
7. Probe: 3S- RS 1 Y
H4000PD
8. Probe: 5S - RS 5135960 1 Y
5198571
9. Probe: 6S- RS 2.7MHz to 8.0MHz BT’09 1 Y
H45021RP
2377059
10. Probe: 7S - RS 1 Y
H4000PE
2377040
11. Probe: 10S - RS 1 Y
H4000PF
KN100093
14. Probe: 6T - RS Must have TEE Probe Interface Module (H45021D) 1 Y
H45531MZ
KN100104
15. Probe: 6Tc - RS 2.9 MHz to 8.0 MHz 1 Y
H45551ZE
KN100073
16. Probe: 9T - RS Must have TEE Probe Interface Module (H45021D) 1 Y
H45531YM
2411689
17. Probe:P2D - RS 1 Y
H45021C
2411690
18. Probe: P6D - RS 1 Y
H45021CA
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
26. 12L-RS Biopsy Kit H40432LC 1 Y
Section 9-8
Peripherals
H45531XE
7. Video Scan Converter 1 Y
2412280
H45531XG
10. USB Flash Card (512MB) 1 Y
2411544
H45021GB
15. MO Drive 9.1GB 5.25" Ext 1 Y
066E0677
Section 9-9
SafeLock Cart Parts
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
2. Handrest S2423250 1 Y
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
Section 9-10
Optional Modo Cart
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
Section 9-11
Spare Part Kits
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
BT’09
This kit includes:
• Vi_q SATA BEP SERVICE KIT
• TR32 Board
BT’09
This kit includes:
• Vi_q SATA BEP SERVICE KIT
• RFI Board
INCLUDES:
• RFI BOARD V2-NATURAL,
7. FRONTEND ASSY.-Vivid-i,q NATURAL N2375187-2 1 Y
• PROBE AND MUX ASSY.-NATURAL,
• TR32 BOARD V12-NATURAL-ROHS
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
8. SPEAKERS MODULE-VIVID-I R2378953 For Vivid q N 1 Y
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
BT’09
This kit includes:
• Vi_q SATA BEP SERVICE KIT
BT’09
This kit includes:
15. Spare Part Kit 5 FL200095-6 1 Y
• Doc CD - FL200117
• SW DVD - FL200112
BT’06
This kit includes:
• Doc CD - FL200028
16. Spare Part Kit 5 FL200095-5 • SW CD - FL200103 1 Y
SW ver. 1.36.19,Appl. SW 6.1.0. build 122
• FL200107
Application SW Patch for Vivid q N
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
Section 9-12
Product Manuals
9-12-1 Product Manuals - BT’12
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1. Vivid q N Service Manual - BT12 FQ091019 This Manual 1 N
Chapter 10
Care and Maintenance
Section 10-1
Overview
10-1-1 Periodic Maintenance Inspections
Having been determined by GE engineers that your Vivid q N system has no high-wearing components
likely to fail due to frequent use, no Periodic Maintenance Inspections are mandatory. However, some
Customer Quality Assurance Programs may require additional tasks
and/or inspections to be performed at periods of frequency different from those listed in this manual.
After delivery and installation of the Vivid q N ultrasound unit, your service representative will contact
you to schedule the first of several routine service visits. In addition to routine maintenance by trained
Field Service Engineers (FSEs), it is recommended that the procedures described in this chapter be
observed as part of your own internal routine maintenance program.
Table 10-1 Contents in Chapter 10
10-1-3 Warnings
CAUTION Do not pull out or insert circuit boards while power is ON.
CAUTION Do not operate this unit unless all board covers and frame panels are securely in place. System
performance and cooling require this.
10-2 -
GE
DIRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Section 10-2
Why Perform Maintenance Procedures?
10-2-1 Keeping Records
It is good business practice that ultrasound facilities maintain records of quality checks and corrective
maintenance. The records provide the customer with documented confirmation that the Vivid q N
portable ultrasound scanner is maintained on a periodic basis.
Routine Quality Control testing of the system must be conducted regularly. The same tests are
performed regularly during each period, so that changes can be monitored over time and effective
corrective action taken, if required.
Testing results, corrective action, and the effects of corrective action, must be documented and
maintained on site.
Your GE Service Representative can help you with establishing, performing and maintaining records for
a Quality Assurance program. Please contact us for details of your nearest
GE Service Representative, who will provide a quotation for this service, as required.
Section 10-3
Maintenance Task Schedule
10-3-1 How Often Should Care & Maintenance Procedures be Performed?
The Customer Care & Maintenance Task Schedule (provided in Table 10-2 on page 10-5) specifies how
often the Vivid q N ultrasound scanner should be serviced, and outlines items requiring special
attention.
NOTE: It is the customer’s responsibility to ensure the Care and Maintenance procedures are
performed on the Vivid q N ultrasound scanner as scheduled in order to retain the high levels
of safety, dependability, and system performance.
The service procedures and recommended intervals shown in the Customer Care & Maintenance Task
Schedule assumes that you use your Vivid q N scanner for an average patient load (10-12 patients per
day) and that you do not use it as a primary mobile unit which is transported between diagnostic
facilities.
NOTE: If conditions exist which exceed typical usage and patient load, it is strongly recommended to
increase the maintenance frequencies.
Surgical Probe
As prescribed in Probe Manual.
Current Leakage Checks
Section 10-4
System Maintenance
10-4-1 Preliminary Checks
The preliminary checks take approximately 15 minutes to perform. Refer to the Vivid q N User Manual
whenever necessary.
St
ep Item Description
1 Ask & Listen Ask the customer if they have any problems or questions about the equipment.
3 Turn the system power ON and verify that all fans and peripherals turn on. Watch the displays
Power-up
during power up to verify that no warning or error messages are displayed.
NOTE: Refer also to Chapter 4 - General Procedures and Functional Checks, for additional details
about the functional checks described in this section.
Applicable Software Verify the basic operation of all optional modes such as Multi-Image, 3D, Contrast,
5 Harmonics, Cine, Stress Echo,... etc.
Optionsa Check the basic system controls that affect each the operation of each option.
Perform an Element Test on each probe to verify that all the probe elements and system
9 Probe Elements channels are functional.
Perform the Automatic Test to verify that all boards are functioning according to
10 System Diagnostics
specifications.
Use the Visual Channel Utility on the 10L probe to verify that all system xmit/recv
11 Xmit/Recv Elements
channels are functional.
12 Keyboard Test Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
Refer to the “Measurement and Analysis” chapter in the Vivid q N User Manual, for
14 Measurements
measurement accuracy specifications.
a. Some software may be considered as standard, depending upon system model configuration
2 B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 DICOM Verify that DICOM is functioning properly. Send an image to a DICOM device.
2 Mains Cable Inspect the mains cable and its connectors for any damage.
3 Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals,
Terminals
and that no strands may cause a short circuit.
10-4-4 Cleaning
CAUTION When performing Cleaning Procedures, to prevent the risk of system damage, always observe
the following precautions:
• Use only cleaning materials and solutions (such as Isopropyl Alcohol [IPA])
as recommended in the procedures described in Table 10-7 on page 10-11.
• Do not use any solutions or products not listed in the Vivid q N User Manual.
• Do not spray any liquid directly onto the Vivid q N covers, LCD Display or keyboard!
• Do not allow any liquid to drip or seep into the system.
• Prior to cleaning, turn OFF power to the system.
2 LCD Display On a weekly basis, gently wipe the LCD Display with a dry, soft, lint-free non-abrasive folded
cloth.
Note: In the event that you see a scratch-like mark on the LCD Display, this may be a stain
transferred from the Keyboard or Trackball when the LCD Display Cover was pressed from the
outside. Proceed as follows:
Wipe or dust the stain gently with a soft, lint-free cloth moistened with water or a 70-30 mixture
of isopropyl alcohol and water that does not contain impurities. Wring out as much of the liquid
as possible then wipe the LCD Display again. Do not let any liquid drip into the computer!
Be sure to dry the LCD Display before closing the cover.
4 Probe Holders Clean the probe holders with warm water and a damp cloth to remove all traces of gel
(soaking may be required to remove excess gel).
MOD Clean the drive head and media with the vendor-supplied cleaning kit. Advise the user to
5 repeat this often, to prevent future problems. MOD disks must be stored away from dust and
cigarette smoke. Do not use alcohol or benzene to clean the MOD cartridge.
DVD - CDRW Clean the drive head and media with the vendor-supplied cleaning kit. Advise the user to
6 Drive repeat this often, to prevent future problems. CDs must be stored away from dust and cigarette
smoke. Do not use alcohol or benzene to clean the CD drive.
7 Peripherals Clean the peripherals in accordance with the respective manufacturer’s directions.
10-4-4-2 Disinfection
NOTICE IMPORTANT - To avoid contamination when the system is in use during operations, always
cover the keyboard and the monitor with the OR covers provided for this purpose.
The disinfection procedures described in this section should be routinely performed by medical
personnel, as shown in Table 10-8. The information is provided in this manual for reference purposes.
Monitor display-surface Weekly. Clean with dry soft, Soft, Staff users/doctors
In event of lint-free cloth. lint-free cloth.
contamination, When contaminated Disinfect with a
clean immediately with blood or other 70:30 mixture of
with a 70:30 stains secretions, use a isopropyl
mixture of isopropyl 70:30 mixture of alcohol and
alcohol and water. isopropyl alcohol and water.
water.
Devices' external surfaces Weekly. Clean the washable Use a soft, Trained cleaning staff
(incl. monitor housing, handle In event of outer surfaces with a non-abrasive
and SafeLock Cart wheels) contamination, soft, non-abrasive cloth folded cloth or
clean immediately including the SafeLock sponge with a
with a 70:30 Cart wheels. Do not use mild, general
mixture of isopropyl abrasive powder or purpose,
alcohol and water. strong non-abrasive
chemicals! soap and water
solution.
2 Scratches and Dents Inspect the system for dents, scratches or cracks.
Clean the LCD Display Cover as described in Table 10-7 on page 10-11
3 LCD Display Cover Verify ease of up/down movement and make sure that the LCD Display cover holds
its set position when open and that there is no play on the mechanism.
Check that the cover latches open and lock correctly.
Clean the Control Panel and Keyboard as described in Table 10-7 on page 10-11
Inspect the Control Panel and Keyboard. Note any damaged or missing items.
5 Control Panel and Keyboard
(Replace faulty components, as required).
Check all four SafeLock Cart wheels and castors for wear.
Check the brakes on each of the two front wheels. Make sure that pressing the
brake locks the wheel, preventing it from swivelling or rolling and therefore stopping
7 SafeLock Cart movement of the SafeLock Cart. Verify that releasing the brakes allows free
movement of the SafeLock Cart.
Check the SafeLock Cart raise/lower motion and verify proper operation of the
release mechanism.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
MOD user to repeat this often, to prevent future problems. MOD disks must be stored
8
(if option installed) away from dust and cigarette smoke. Do not use alcohol or benzene to clean the
MOD cartridge.
Clean the drive head and media with the vendor-supplied cleaning kit. Advise the
9 DVD- CDRW Drive user to repeat this often, to prevent future problems. CDs must be stored away from
dust and cigarette smoke. Do not use alcohol or benzene to clean the CD drive.
Check all internal cable harnesses and connectors for wear, and secure connector
10 Cables and Connectors
seating. Pay special attention to probe strain or bend reliefs.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation or
other deterioration, and verify continuity. Tighten the clamps that secure the power
11 Power Cord
cord to the unit and the outlet plug to the cord.
Replace the power cord and clamp, as required.
14 External I/O Check all connectors for damage and verify that the labeling is good.
Check that the battery is not damaged, does not leak, does not emit an odor, and
is not deformed or discolored. Perform the Battery Parameters Test as described
16 Battery in "Battery Parameters Test" on page 7-102.
Observe all warnings and cautions for battery handling, recharging, storing, and/or
disposal, as listed in "Vivid q N Battery Safety" on page 1-15.
Section 10-5
Probe Maintenance
WARNING Any evidence of wear on a probe indicates that it must not be used. Improper handling may
easily damage ultrasound probes. See the Vivid q N User Manual and also refer to the probe
manufacturer’s handling instructions, for more details.
Failure to follow these precautions can result in serious injury and equipment damage.
Failure to properly handle or maintain a probe may also void its warranty.
Check there are no traces of gel on the probe holders. If any gel present, clean the probe
1 Probe Holder holders with warm water and a damp cloth to remove all traces of gel (soaking may be required
to remove excess gel).
Check all the probes for wear and tear on the lens, cable and connector. Look for bent pins on
the connector and in the connector socket on the unit. Remove any dust from inside the
connector sockets.
2 Probes Verify that the EMI fingers around the probe connector socket housing are intact.
Check the probe locking mechanism and probe switch.
Perform a visual check of transducer pins and connector sockets before connecting a probe.
Verify that the system properly recognizes all probes.
NOTE: For detailed information on handling endocavity probes, refer to the appropriate supplementary
instructions for each probe.
NOTE: The most recent, up-to-date information on probes and probe care is available at:
http://www.gehealthcare.com/usen/ultrasound/products/probe_care.html
WARNING Any evidence of wear on a probe indicates that it must not be used. Ultrasound probes can be
easily damaged by improper handling. See the Vivid q N User Manual and also refer to the probe
manufacturer’s care cards, for more details.
NOTE: Failure to follow these precautions can result in serious injury and equipment damage. Failure
to properly handle or maintain a probe may also void its warranty.
Always perform a visual check of the probe pins and system sockets before plugging in a probe.
When handling probes, always observe the precautions listed in "Probe Handling" on page 10-15.
The TEE and Interoperative probes often have special usage considerations; always refer to the
individual probe manufacturers’ handling instructions/user manual.
NOTE: For details on general probe cleaning, refer to the information provided in the Vivid q N User
Manual.
NOTE: For specific probe cleaning instructions, refer to the individual probe Users Manual (or care card
supplied with the probe).
CAUTION Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s
warranty. DO NOT soak the probe or wipe the probe lens with any product not listed in the
Vivid q N User Manual. Doing so could result in irreparable damage to the probe and/or Vivid q N
system. Follow the care instructions supplied with the probe.
CAUTION To help protect yourself from blood-borne diseases when cleaning and handling probes, wear
approved, non-allergic disposable gloves.
WARNING Always disinfect a defective probe before returning it to the manufacturer. Be sure to
tag the probe as being disinfected.
WARNING Neurological and intraoperative procedures must be performed with the use of legally
marketed, sterile, pyrogen-free probe sheaths. Probes used during neurological
surgery must NOT be sterilized with liquid chemical sterilants because of the possibility
of neurotoxic residue remaining on the probe.
10-5-4-2-1 Cleaning and Sterilizing Probes Used Externally
Table 10-11 Cleaning & Sterilizing Procedure for Probes Used Externally
Step Description
1 To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2 Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80°F). Use only mild
non-abrasive soap.
3 Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4 Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5 When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6 Check the lens for cracking and the cable insulation for separation.
Table 10-12 Cleaning & Sterilizing Procedure for Probes (TEE) Used Internally
Step Description
1 To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2 Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80°F). Use only mild
non-abrasive soap.
3 Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4 Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5 When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6 Check the lens for cracking and the cable insulation for separation.
WARNING Always disinfect a defective probe before returning it to the manufacturer. Be sure to
tag the probe as being disinfected.
CAUTION To help protect yourself from blood-borne diseases when cleaning and handling probes, wear
approved, non-allergic disposable gloves.
Equipment being returned must be properly clean and free of blood and other potentially infectious
contaminants.
GE policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have
been properly decontaminated prior to shipment. Under no circumstances should a part or equipment
be shipped before being visibly clean and properly disinfected.
The purpose of the regulation is to protect employees in the transportation industry, as well as the
persons who will receive and/or open the package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
Section 10-6
Electrical Safety Tests
10-6-1 Safety Test Overview
The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC
60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of
cord-connected, electrically operated, patient care equipment. If additional information is needed, refer
to the NFPA 99 (for USA) and IEC 60601-1 documents.
WARNING THE USER MUST ENSURE THAT THE SAFETY INSPECTIONS ARE PERFORMED AT
LEAST EVERY 12 MONTHS ACCORDING TO THE REQUIREMENTS OF THE PATIENT
SAFETY STANDARD IEC-EN 60601-1. ONLY TRAINED PERSONS ARE ALLOWED TO
PERFORM THE ABOVE-MENTIONED SAFETY INSPECTIONS.
CAUTION Compare all safety-test results with safety-test results of previously performed safety tests e.g.
last year etc. in event of unexplainable abrupt changes of safety-test results, consult
experienced authorized service personnel or ge for further analysis.
CAUTION To avoid the risk of electrical shock, the unit under test must not be connected to other electrical
equipment. Remove all interconnecting cables and wires. The operator and patient must not
come into contact with the unit while the tests are being performed.
CAUTION Possible risk of infection. Do not handle soiled or contaminated probes and other components
that have been in patient contact. Follow appropriate cleaning and disinfecting procedures
before handling the equipment.
NOTICE When the SafeLock Cartis connected to the wall outlet and the main circuit breaker is in the ON position,
the AC Box Fan is operated - even if the system power is turned OFF.
To minimize the risk of a probe causing electrical shock, the customer should observe the following
recommendations:
• Do not use a probe that is cracked or damaged in any way
NOTE: Because the Vivid q N scanner is operated from low voltage DC (considered extremely safe
extra low voltage) and because the transformer is enclosed in plastic parts only (and therefore
does not have any accessible metal parts), when the scanner is not mounted on the
SafeLock Cart, it is not necessary to perform the following safety tests (i.e., they apply only to
the SafeLock Cart itself):
Table 10-15 Type CF Applied Part Current Leakage Limits - Surgical Probes and ECG Connections
NOTE: No outlet tester can detect the condition in which the Neutral (grounded supply) conductor and
the Grounding (protective earth) conductor have been reversed. If later tests indicate high
current leakages, this should be suspected as a possible cause and the outlet wiring should be
visually inspected.
NOTE: This test is applicable to the SafeLock Cart only. It is not necessary to perform this test on the
Vivid q N scanner (for details, refer to the note on page 10 - 20).
Measure the resistance between the third pin of the attachment plug and any exposed metal parts on
the SafeLock Cart. The Ground wire resistance should be less than 0.2 Ohms. Refer to the procedure
in the IEC 601-1.1.
GROUND PIN
OHMMETER
NOTE: Follow the steps below to test the Ground wire resistance. If the test passes, there is no safety
risk. However, the input power cable should be tested with a substitute Safety Tester.
1.) Connect one of the Ohmmeter probes to the tested AC wall outlet cable side (Ground pin).
2.) Using the other Ohmmeter probe, touch the exposed metal part of the SafeLock Cart.
3) Set the meter Function switch to the Resistance position.
4) Measure and record the Ground wire resistance.
This should be less than 0.2 Ohms.
NOTE: Since very low resistance is measured, the Ohmmeter probe resistance should be subtracted
from the reading.
NOTE: Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
10-6-5-1 Definition
This test measures the current that would flow in a grounded person who touched accessible metal
parts of the bedside station, if the Ground wire should break. The test verifies the isolation of the power
line from the chassis.
The meter is connected from accessible metal parts of the case to Ground. Measurements should be
made with the unit ON and OFF, with the power line polarity Normal and Reversed. Record the highest
reading.
Figure 10-3 Set Up for Chassis Source Current Leakage, IEC 601-1 Clause 19
Continuous Current Leakage and Patient, Auxiliary Currents
When using the Microguard or a similar test instrument, its power plug may be inserted into the wall
outlet and the equipment under test be plugged into the receptacle on the panel of the meter. This
places the meter in the grounding conductor and the current flowing from the case to Ground will be
indicated in any of the current ranges. The maximum allowable limit for chassis source leakage is shown
in Table 10-13 on page 10-21.
NOTE: Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
Table 10-16 Typical Data Sheet for Chassis Source Current Leakage Test
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
10-6-6-1 Definition
This test measures the current which would flow to Ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface.
Measurements should be made with the Ground open and closed, with power line polarity normal and
reversed, and with the ultrasound console OFF and ON. For each combination the operating controls,
such as the lead switch, should be operated to find the worst case condition.
CAUTION Equipment damage possibility. To avoid damaging the unit, never switch the polarity and the
neutral status when the unit is powered ON. Be sure to turn the unit power OFF before switching
them, using the Polarity switch and/or the Neutral switch.
NOTE: Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
NOTE: Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
NOTE: Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer
tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.
CAUTION Line voltage is applied to the ECG leads during this test. To avoid the risk of possible electric
shock, the system being tested must not be touched by anyone (patient or operator), while the
ISO TEST switch is depressed.
NOTE: It is not necessary to test each lead individually, or power condition combinations, as required
in previous tests.
Table 10-17 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
Maximum
Allowance Limit
Table 10-18 Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power Maximum
Source Allowance Limit
115V 20uA
Patient Lead Isolation Current Test
220/240V 5mA
ON REVERSE CLOSED
ON NORM OPEN
ON REVERSE OPEN
10-6-9-1 Definition
This test measures the current that would flow to Ground (from any of the probes) through a patient who
is being scanned, and who becomes grounded by touching some other grounded surface.
CONSOLE
POWER N (WHITE)
OUTLET
G (GREEN)
MOMENTARY
SWITCH LEAKAGE TEST
METER
NOTE: Each probe will have a certain amount of current leakage, depending on its design. Small
variations in probe current leakages are normal from probe to probe. Other variations will result
from differences in line voltage and test lead placement.
CAUTION When power to the unit is ON, never switch the Polarity and the status of Neutral. Be sure to turn
OFF power to the unit before switching them, using the POLARITY switch and/or the NEUTRAL
switch. Failure to comply with this warning may cause damage to the unit!
Table 10-20 Typical Data Sheet for Transducer Source Current Leakage Test
Transducer Tested:
Tester Tester
Power GROUND
POLARITY or NEUTRAL
Unit Power Switch Switch Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
Section 10-7
Excessive Current Leakage
10-7-1 Possible Causes of Excessive Current Leakage
Tighten all Grounds. Ensure star washers are under all Ground studs.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
NOTE: No outlet tester can detect the condition where the white neutral wire and the green grounding
wire are reversed. If later tests indicate high current leakages, this should be suspected as a
possible cause and the outlet wiring should be visually inspected.
NOTE: Each probe will have some amount of leakage, depending on its design. Small variations in
probe current leakages are normal from probe to probe. Other variations will result from
differences in line voltage and test lead placement. The maximum allowable current leakage for
body surface contact probes differs from that of inter-cavity probes. Be sure to enter the correct
probe type in the appropriate space on the check list.
If excessive current leakage is slot dependent, inspect the system connector for bent pins, poor
connections, and Ground continuity.
GE Vingmed Ultrasound AS
P.O.Box: 141
N-3191 HORTEN
NORWAY
www.gehealthcare.com