Gima S.p.A.

Via Marconi, 1 – 20060 Gessate (MI) Italy


gima@gimaitaly.com – export@gimaitaly.com
www.gimaitaly.com

PROFESSIONAL MEDICAL PRODUCTS

SUPER VEGA SUCTION ASPIRATOR - 2 x 2 L - on trolley with footswitch

Instruction manual

28192

Gima S.p.A.
Via Marconi, 1 – 20060 Gessate (MI) Italy
Made in Italy
GB
SUPER VEGA SUCTION ASPIRATOR it’s an electrical powered surgical aspirator used for the nasal, oral and tracheal suction in
man and child of body liquids (mucus, catarrh or blood). The equipment is intended for continuous operation. .
The equipment is equipped with a trolley provided with 5 wheels (three of them are provided with loocking system in order to
avoid the equipment can overbalance) and an external plastic enclosure. Thanks to this characteristics and to the rating that it has,
this product is particularly suitable for hospital use, on the tracheotomized patients, minor surgical applications and post-
operative therapy at home. Made of highly heat resistant, electrically insulated plastic material in conformity with the latest
European safety standard, the product is supplied with a complete polycarbonate autoclavable jar with overflow valve and it is
equipped with aspiration regulator and vacuum indicator located on the front panel. Available under request with different
version for application and use (version with remote control, ecc..)

GENERAL WARNING
READ INSTRUCTION MANUAL CAREFULLY BEFORE USE

ONLY HIGHLY QUALIFIED STAFF USE RESERVED

THE INSTRUMENT MUST NOT BE DISASSEMBLED. FOR A TECHNICAL SERVICE ALWAYS CONTACT GIMA

KEEP OFF THE REACH OF CHILDREN OR NOT CAPABLE PEOPLE WITHOUT SUPERVISION

FULL CONTAINERS MUST BE HANDLED WITH GREAT CARE DURING TRANSFER TO THE DISPOSAL AREAS,
FOLLOWING THE LOCAL PROCEDURES AND REGULATIONS

IMPORTANT SAFETY RULES

1. Check the condition of the unit before each use. The surface of the unit should carefully inspected for visual damage. Check
the mains cable and do not connect to power if damage is apparent;
2. Before connecting the appliance always check that the electric data indicated on the data label and the type of plug used,
correspond to those of the mains electricity to witch it’s to be connected;
3. Respect the safety regulations indicated for electrical appliances and particularly:
 Use original components and accessories provided by the manufacturer to guarantee the highest efficiency and safety
of the device;
 The device can be used only with the bacteriological filter;
 Never immerge the appliance into water;
 Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;
 To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall and lead to a
malfunction and/or breakage. Should there be signs of damage to the plastic parts, which may expose inner parts of
the energised device, do not connect the plug to the electrical socket. Do not attempt to make the device work
before it has been thoroughly checked by qualified personnel and/or the GIMA technical service department.
 Don’t use in the presence of inflammable substances such as anaesthetic, oxygen or nitrous oxide;
 Don’t touch the device with wet hands and always prevent the appliance coming into contact with liquids;
 Don’t leave the appliance connected to the power supply socket when not in use;
 Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;
 Store and use the device in places protected against the weather and far from any sources of heat. After each use, it is
recommended to store the device in its own box away from dust and sunlight.
 Don’t use the device thoracic drainage.
 In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be necessary, you
must use ones that are in compliance with safety regulations, however, taking care not to exceed the maximum power
supply tolerated, which is indicated on the adapters and extensions.
4. For repairs, exclusively contact technical service and request the use of original spare parts.
Failure to comply with the above can jeopardise the safety of the device;
5. Use only for the purpose intended. Don’t use for anything other than the use defined by the manufacturer.
The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not
complying with current regulation.
6. The medical device requires special precautions regarding electromagnetic compatibility and must be installed and used in
accordance with the information provided with the accompanying documents: the SUPER VEGA SUCTION ASPIRATOR
device must be installed and used away from mobile and portable RF communication devices (mobile phones, transceivers,
etc.) that may interference with the said device.
7. Instrument and accessory discharging must be done according to current regulations in the country of use.
8. WARNING: Do not change this equipment without the permission of the manufacturer Gima S.p.A. None of electric or
mechanical parts have been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the
electric / mechanical parts. Always contact technical assistance
9. Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety
and the technical characteristics of the same.
10. The medical device is in contact with the patient by means of a disposable probe (supplied with the device) furnished with
the relative CE compliance certification according to the requirements of regulation ISO 10993-1: thus, no allergic reactions
and skin irritations may occur.
11. The product and its parts are biocompatible in accordance with the requirements of regulation EN 60601-1.

14
GB
12. Operation of the device is very simple and therefore no further explanations are required other than those indicated in the
following user manual.

The manufacturer cannot be held liable for accidental or indirect damages should the device be modified,
repaired without authorization or should any of its component be damaged due to accident or misuse.

Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance
with the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its
normatives.

TECHNICAL CHARACTERISTICS

Model SUPER VEGA SUCTION ASPIRATOR


Typology (MDD 93/42/EEC) Medical Device Class IIa
Classification UNI EN ISO 10079-1 HIGH VACUUM / HIGH FLOW
Main Voltage 230 V ~ / 50 Hz
Power consuption 110 VA
Fuse F 1 x 1.6A L 250 V
Maximum suction aspiration (without jar) -80kPa (- 0.80bar)
Minimum suction aspiration (without jar) Less than -40kPa (-0.40 bar)
Maximum flow (without jar) 40 l/min
Weight 6.2 Kg
Dimension 380 x 920 (h) x 360mm
Functioning NON-STOP OPERATED
Working condition Room temperature: 5 ÷ 35 °C
Room humidity percentage: 30 ÷75 % RH
Atmospheric pressure: 800 ÷ 1060 hPa
Conservation and transport condition Room temperature: - 40 ÷70 °C
Room humidity percentage: 10 ÷ 100% RH
Atmospheric pressure: 500 ÷ 1060 hPa

IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC


DIRECTIVE 2012/19/UE-WEEE:
The symbol on the device indicates the separated collection of electric and electronic equipment. At the end of
life of the device, don’t dispose it as mixed solid municipal waste, but dispose it referring to a specific
collection centre located in your area or returning it to the distributor, when buying a new device of the
sample type to be used with the same functions.
This procedure of separated collection of electric and electronic devices is carried out forecasting a European
environmental policy aiming at safeguarding, protecting and improving environment quality, as well as avoiding potential effects
on human health due to the presence of hazardous substances in such equipment or to an improper use of the same or of parts of
the same Caution: The wrong disposal of electric and electronic equipment may involve sanctions.

15
GB

SYMBOLS

Class 2 isolation equipment

CE marking in conformity with EC directive 93/42/CEE and subsequent


changes

General warnings and/or specifications

Consult the instruction manual

Manufacturer: Gima S.p.A.


Via Marconi, 1 - 20060 Gessate (MI) Italy

Applied Part type BF (suction probe)

Keep in a cool, dry place

Fuse

Conservation temperature: - 25 ÷ 70° C

Atmospheric Pressure

~ Alternate current
Hz Mains frequency
ON / OFF
0 / I / II Remote Control
Batch production

Serial number

Model / Ref number


Degree of protection an electrical device provides in the case of
accidental or intentional contact with the human body or with objects,
and protection in the case of contact with water.
IPX1
1st DIGIT 2nd DIGIT
(on the footswitch control label)
PENETRATION OF SOLIDS PENETRATION OF LIQUIDS
Protected against the vertical
No protection
flow of drops of water

CLEANING DEVICE

Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always use a cotton cloth
dampened with detergent. Don’t use abrasive or solvent detergents.

PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO
NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.

During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in
contact with contaminating substances (after each utilization cycle of the machine).

16
GB
ACCESSORIES SUPPLIES

DESCRIPTION CODE
N°2 COMPLETE ASPIRATION JAR 2000cc RE 210351/01
CONICAL FITTING RE 210420
TUBES SET 8 mm x 14 mm RE 210355/03
ANTIBACTERIAL AND HYDROFOBIC FILTER SP 0121
FOOTSWITCH CONTROL Cod. 52130
(for versions equipped with footswitch control)
EUROPEAN POWER SUPPLY CORD SP 0020

The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit. It should be changed
immediately if it becomes wet or if there is any sign of contamination or discolouration. If should also be changed if the unit is used
with a patient whose risk of contamination is unknown. Don’t use the suction unit without the protection filter. If the suction
unit is used in an emergency or in a patient where the risk of contamination is not know the filter must be changed after each use.
On request, are available versions with disposable collection system (FLOVAC® 2000ml) composed by a reusable polycarbonate
vessel and polyethylene disposable bag (collection liner for fluids suction canisters).

Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-
infections. Don’t use after lapse of the sell-by date

Check the expiry date on the original packaging of the suction catheter and check the integrity of the sterile
packaging. GIMA declines any liability for injury to the patient correlated to the deterioration of the above-
mentioned sterile packaging due to handling of the original packaging by third parties.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO
10993-1 standards on material biocompatibility.

Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this
limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you to
change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube.
Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must
be replaced if there are visible signs of decay of the material constituting the said component.

Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the
said component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said component.
Service life of the device: More than 850 hours of operation (or 3 years) in accordance with the standard conditions of testing and
operation. Shelf life: maximum 5 years from the date of manufacture.

CLEANING OF ACCESSORIES
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:

 Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating
substances;
 Disconnect the tank from the device and remove the said container from the support of the device.
 Separate all the parts of the cover (overflow device, washer).
 Disconnect all tubes from the jar and the protection filter
 Wash each part of the container from secretions under cold running water and then clean every single part in hot
water (temperature not exceeding 60°C)
 Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any deposits. Rinse
with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with commercial
disinfectants by carefully following the instructions and dilution values supplied by the manufacturer. After cleaning,
leave the parts to dry in an open, clean environment.
 Dispose of the aspiration catheter according to that provided by local laws and regulations.

The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed
60°C). After cleaning, leave the parts to dry in an open, clean environment.

When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:

 Place the overflow valve into its seat in the cover (under VACUUM connector)
 Insert floating valve keeping the o-ring towards the opening of the cage
 Place the o-ring into its seat around the cover

17
GB
 After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or
liquid exit

After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to
remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).

The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is
guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit the
physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.

After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.

The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).

DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER

Instruction for disposal Liner Flovac®:

If the device is equipped with disposable collection systems FLOVAC ® carry out the disposal of the bag as follows:
Turn off the Vacuum and remove all the tubes connected to the Liner, giving particular attention to avoid accidental contamination.
Fit the appropriate plugs to the “PATIENT” and “TANDEM” ports, pressing the home firmly, taking care to avoid accidental
contamination. Turn the butterfly connector to OFF. Remove the liner bag from the rigid container and transfer it to the waste
disposal area, ensuring that all the openings are sealed, keeping in mind the product is potentially infectious.
This product must be disposed of in accordance with the current hospital regulations.

PERIODICAL MAINTENANCE CHECKS


The SUPER VEGA SUCTION ASPIRATOR does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. With regard to training, given the information contained in the user
manual and since it is easy to understand the said device, it doesn’t appear to be necessary.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been damaged during
previous use. Connect the cable to electrical network and turn the switch on.
Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum indicators reaches at
least -80kPa (-0.80 bar). Rotate the knob from right to left. The vacuum indicator should go down -40kPa (-0.40 bar).
Check that no loud noises are present. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situated in the plug
protects the instrument. For fuses replacing, always the type and the range.
Before changing the fuse, disconnect the plug from the power supply socket.
Internally, the device (only for devices fitted with a circuit board) is protected by a fuse (T 50mA L 250V) that cannot be reached
from the outside, so please contact a technician authorised by the manufacturer for its replacement.

Type of fault Cause Remedy


1. The suction unit doesn’t Cable is damaged Replace the cable
work External power source failure Check the external power source
2. No aspiration Jar Cap not screwed on correctly Unscrew the cap, and re-screw it
3. No aspiration Lid seal not in its seat Unscrew the cap and insert the seal properly in
its seat
4. The Vacuum power on the  Vacuum regulator set to  Turn the vacuum regulator clockwise and
patient side is either very low minimum check the value of the vacuum on the gauge
or absent  Protection filter blocked or  Replace the filter
damaged  Replace or reconnect the tubes, check the jar
 Connection tubes blocked, kinked connections
or disconnected  Empty the jar, or disconnect the tube from the
 Shut-off valve blocked or jar and unblock the shut-off valve. The unit
damaged twill only work in the upright position
 Pump motor damaged  Refer to authorised service personnel
5. The float doesn’t close If the cap has been washed, ensure Fit the float into it’s place
that the float is not partially
detached
6. The float doesn’t close The float it’s covered by dirty Unscrewed the cap, leave the and put in on
material autoclave
7. Low suction Foam inside the jar Fill the jar to 1/3 full of ordinary water
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 None of the procedures have Contact GIMA customer service
achieved the desired results
18
GB

If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the
device to GIMA technical service.

Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.

BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.

INSTRUCTION FOR USE

Assembly of the device SUPER VEGA SUCTION ASPIRATOR:


Take the 5 arm base and set up the 5 wheels that come with the above device. The wheels provided with braking device must be
placed one next to the order. Take the support bar that comes with the device SUPER VEGA SUCTION ASPIRATOR and place it in
the hole on the 5-arm base. From under the base, lock the two parts by means of the supplied screw.
Eventually, place the device on the trolley.

 The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
 The working position must be such as to allow one to reach the control panel and to have a good view of the empty
indicator, the jar and the antibacterial filter.
 If the device is to be transported from one place to another, to prevent the liquid collection jar from falling and consequently
the liquid from spilling, removing the jar from the device is recommended.

WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device. The vacuum jar, during use, must be used in vertical mode, to prevent the action of the
backflow valve. If this protection is triggered, turn the device off and disconnect the pipe connected to the vacuum jar (indicated
with the word VACUUM) on its cover.

 Place the unit on a flat, horizontal surface


 Connect one end of the short silicon tube, with antibacterial filter, to the suction connector on the lid of the jar.
 The other tube already connected to the filter has to be connected to the “VACUUM” jar outlet, where has been fixed the red
float (security float). When the 90% of the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the device.

WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
“VACUUM”. A wrong connection causes immediate destruction in case of contact with sucked liquids.

Filter assembling
FLOW DIRECTION

IN / Fluid Inside

IN
Suction Pump Inlet Jar Air Tube

 Connect the long silicone tube to the “PATIENT” jar outlet


 Connect the other end of the long silicon tube to the probe plastic connector, then connect the suction probe to it.
 Connect the power cord to the device then connect the plug to the electrical mains supply.
 Push switch on position I to start suction.
 Unscrew the lid of the jar and fill the jar 1/3 full or ordinary water (this assists the unit to reach peak vacuum performance
and makes clean-up easier) then re-screw the lid on the jar correctly.
 During operation the jar has to be in vertical position to avoid overflow valve to cut off aspiration. Should this happen,
switch off the device and disconnect the tube from the jar cover (from “VACUUM” outlet).

19
GB
 Once finished push switch on O position and unplug.
 Remove the accessories and clean.
 At the end of each use, place the device in its box away from dust.
WARNING: The power supply cable plug is the element of separation from the electrical mains system: even if the units
equipped with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to
allow a further method of disconnection from the mains supply system.

Antibacterial Filter
Autoclavable Silicon 8x14 mm

Vacuum (Kpa / Bar) Vacuum Regulator Knob

“PATIENT” Port ON / OFF Switch


Conical Fittings
“VACUUM” Port

Collection Jar in Autoclave PC

WARNING: NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER

Footswitch control device: The equipment, on request, is provided with a footswich control device. It allows the continuous
use of the surgical aspirator. In this case, the plug of the footswitch device shall be inserted into the appropriate socket outlet
placed on the back side of the equipment.

Near the footswitch control plug, there is a commutator that operates the device
directly using the mains or by using the footswitch control.

When the switch is turned on (green light) and the commutator is in position I, the
device works directly using the mains.

When the switch is turned on (green light) and the commutator is in position II, the
device works via remote control (footswitch control)

When the switch is turned on (green light) and the commutator is in position 0, the
device is powered but not in operation

Using the footswitch control and the flow deviator: If using equipment fitted with a flow deviator, users may direct suctioned
liquids in any of the two collection tanks provided. Flow deviator comes with two complete suction kits (2 sets of tubes, 2
antibacterial and hydrophobic filters and two conical junctions).

Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the blu ring fitted
on the tank holder for a more comfortable insertion of the same container.
- After opening the package, fully stretch the bag and then flatten it concentrically to eliminate as much air as possible.
- Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing firmly around the entire
perimeter. Make sure that the system is completely sealed.
- close the connector marked as “TANDEM” with the lid provided.
- Connect the power source of the vacuum to the VACUUM port equipped with specific reusable conical fitting with "male"
connection.
- Connect the patient tube to the PATIENT port of the cover
20
GB
- Before use, check all closures and make sure there are no leaks, starting the aspiration source. If the bag expands to fully
adhere to the walls of the rigid container and the cover bends towards the inside of the glass, the system is not leaking.
- Start the aspiration and periodically check the filling level of the container. The overflow valve will cause the interruption
of aspiration if the aspirated fluids have reached the maximum filling level of the device.
- When the float valve intervenes signalling the device is too full, the suction source must be disconnected within no more
than 5 minutes.

Warning: The accidental inversion of connections may cause contamination for the operator and/or for the vacuum
generation equipment.

RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES

This section contains information regarding the conformity of the compliance with the EN 60601-1-2 Standard.
The SUPER VEGA SUCTION ASPIRATOR is an electro-medical device that requires particular precautions regarding electro-
magnetic compatibility and which must be installed and commissioned according to the electro-magnetic compatibility
information supplied. Portable and mobile radio communication devices (mobile phones, transceivers, etc.) may interfere with
the medical device and should not be used in close proximity with, adjacent to or on top of the medical device. If such use is
necessary and unavoidable, special precautions should be taken so that the electro-medical device functions properly in its
intended operating configuration (for example, constantly and visually checking for the absence of anomalies or malfunctions).
The use of accessories, transducers and cables different to those specified, with the exception of transducers and cables sold by
the appliance and system manufacturer as spare parts, can lead to an increase in emissions or in a decrease of the immunity of
the device or system. The following tables supply information regarding the EMC (Electromagnetic Compatibility)
characteristics of the electro-medical device.

Guidance and manufacturer’s declaration – electromagnetic Emissions


The SUPER VEGA SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPER VEGA SUCTION ASPIRATOR should assure that it’s used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
Irradiated / Conducted emissions Group 1 The SUPER VEGA SUCTION ASPIRATOR only used RF
CISPR11 energy only for its internal functioning. Therefore, its RF
emissions are very low and are not cause interference in
proximity of any Electronic appliances.
Irradiated / Conducted emissions CISPR11 Class [B] The SUPER VEGA SUCTION ASPIRATOR can be used in all
Harmonic emissions EN 61000-3-2 Class [A] environments, including domestic and those connected
Voltage fluctuations / flicker emissions Complies directly to the public mains distribution that supplies power
EN 61000-3-3 to environments used for domestic scopes.

Guidance and manufacturer’s declaration – Immunity Emissions


The SUPER VEGA SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below.
The customers or the user of the sSUPER VEGA SUCTION ASPIRATORshould assure that it’s used in such an environment.
Immunity Test Level indicated by the Compliance Level Electromagnetic environment - guidance
EN 60601-1-2
Electrostatic ± 6kV on contact The device doesn’t Floors should be wood, conceret or ceramic tile.
discharge (ESD) ± 8kV in air change its state If floors are covered with synthetic material, the
IEC/EN 61000-4-2 relative humidity should be at least 30%.
Electrical fast ± 2kV power supply The device doesn’t Mains power quality should be that of a typical
transient / burst lines change its state commercial environment or hospital.
EN 61000-4-4
± 1kV for input /
output lines
Surge ± 1kV differential mode The device doesn’t Mains power quality should be that of a typical
EN 61000-4-5 change its state commercial environment or hospital.
Loss of voltage, brief 5%UT (>95% dip UT) - Mains power quality should be that of a typical
voltage interruptions for 0.5 cycle commercial environment or hospital If the user of
and variations 40%UT (>60% dip UT) the SUPER VEGA SUCTION ASPIRATOR request
EN 61000-4-11 for 5 cycle that the appliance operates continuously, the use
70%UT (>30% dip UT) of a continuity unit is recommended.
for 25 cycle
<5%UT (>95% dip UT)
for 5 sec
Magnetic field 3A/m The device doesn’t The power frequency magnetic field should be
EN 61000-4-8 change its state measured in the intended installation location to
assure that it’s sufficiently low.
Nota UT is the value of the power supply voltage

21
GB
Guidance and manufacturer’s declaration – Immunity Emissions
The SUPER VEGA SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPER VEGA SUCTION ASPIRATOR should assure that it’s used in such an environment.
Immunity Test Level indicated by Compliance Electromagnetic environment - guidance
the EN 60601-1-2 Level
Conducted Immunity 3Vrms 150kHz to V1 = 3 V rms The portable and mobile RF communication devices,
EN 61000-4-6 80Mhz including cables, must not be used closer to the ASKIR C30
(for non life-supporting device, than the separation distance calculated by the
devices) equation applicable to the transmitter frequency.
Recommended separation distance
Radiated Immunity 3V/m 80MHz to E1 = 3 V / m
EN 61000-4-3 2.5GHz d = [3.5 / V1] P
(for non life-supporting
devices) d = [12 / E1] P from 80 MHz to 800MHz
d = [23 / E1] P from 800 MHz to 2.5 GHz

Where P is the maximum nominal output voltage of the


transmitter in Watt (W) depending on the manufacturer of
the transmitter and the recommended separation distance
in metres (m). The intensity of the field from the fixed RF
transmitters, as determined by an electro-magnetic study
of the sitea), could be lower than the level of conformity of
each frequency interval b). It is possible to check for
interference in proximity to devices identified by the
following symbol:

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.

a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial
mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and
accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic
study of the site should be considered. If the field intensity measured in the place where the device will be used surpasses the
above mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnormal
performance arises, additional measures such as changing the device's direction or positioning may be necessary.

b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.

Recommended separation distance between portable and mobile radio-communication devices and the monitor
The SUPER VEGA SUCTION ASPIRATOR is intended to operate in an electro-magnetic environment where RF irradiated interferences are
under control. The client or operator of the SUPER VEGA SUCTION ASPIRATOR device can help prevent electro-magnetic interference by
keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and the SUPER VEGA SUCTION
ASPIRATOR device, as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal Separation distance from the frequency transmitter (m)
output power of the 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Transmitter W d = [12/E1] P
d = [3.5 / V1] P d = [23/E1] P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied

Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people.

22
GB
WARRANTY CONDITIONS

This product is guaranteed for a period of 24 months from the date of purchase. The warranty includes the repair or replacement
of defect spare parts free of charge, if the defect has been clearly described by the customer and determined by technical service.
Inspections on the part of the seller, performed at the request of the customer and intended to determined wjether the device is
fully functional, are not covered by the free-of-charge warranty service. This service will be charged to the customer depending on
the effort required. The consumables components are not subject to warranty. Consumable components are silicon tubes, filters,
seals, conical adaptor and suction catheter. Also exluded from warranty is all damage resulting from improper handling, wilful
damage or improper care of the device.
The warranty shall expire if repairs and servicing are not carried out by technical service.

RULES FOR RETURNING AND RAPAIRING

UNDER NEW EUROPEAN RULES, GIMA REQUIRES THE FOLLOWING PROCEDURES TO BE CARRIED OUT TO PROTECT THE
INSTRUMENT AND THE SAFETY OF ALL WHO COME IN CONTACT WITH IT.

Before returning an instrument for repair, the external surfaces and all accessories MUST be carefully disinfected with a cloth
soaked in methylated spirits or hypochlorite-based solution. The instrument and accessories should then be placed in a bag with a
note outlining the disinfection undertaken.
Failure to follow this procedure will result in the instrument being returned to the purchaser unrepaired.
Instruments returned for repair MUST be accompanied by a description of the problem. Gima will not be responsible for damage
caused through improper use. To avoid such damage, please read the instruction carefully.
Where Gima determines that an instrument is faulty, a replacement will be provided only if a SALES RECEIPT and STAMPED
GUARANTEE are provided. Gima will not be responsible for damage accessories. These may be replaced at the customer’s expence.

23