Radiello Diffusive Sampler Guide
Radiello Diffusive Sampler Guide
Edition 01/2019
Index
Pag.
subject index ii
how does the diffusive sampler work? A1
why is radiello so special? A3
the components of radiello A5
the adsorbing cartridge A5
the diffusive body A5
the supporting plate A5
the label A5
how to use radiello A6
before sampling: assembling the supporting plate A6
on-field: to start the sampling A6
after the sampling A7
radiello maintenance A7
radiello-ready-to-use A8
accessories for radiello B1
vertical adapter B1
shelter B1
how to assemble the shelter B2
on-field temperature measurements B3
calibration solution for H2S B4
filtration kit B4
calibration solutions for aldehydes B4
calibration solutions for BTEX (CS2 desorption) B5
calibration solutions for VOCs in workplace environments B5
calibration solutions for BTEX (thermal desorption) B6
the spare parts of radiello B6
empty cartridge B6
barcode adhesive label B6
clip B6
tube B6
strip B6
aldheydes C1
volatile organic compounds (VOCs) chemically desorbed with CS2 D1
volatile organic compounds (VOCs) thermally desorbed E1
nitrogen and sulfur dioxides (NO2 and SO2) F1
ozone (O3) G1
hydrogen sulfide (H2S) H1
ammonia (NH3) I1
hydrochloric acid (HCl) J1
hydrofluoric acid (HF) K1
anaesthetic gases and vapours L1
phenol, methylphenol and dimethylphenol (thermally desorbed) M1
1,3-butadiene & isoprene N1
index by code Z1
Subject index
isoprene, N1 sterilization, L2
isopropanol, D2, D6 styrene, D2, D6, E3
isopropyl acetate, D2, D6, E3 sulphanilammide, F2
isopropylbenzene, D2 shelter, B1, B2
limonene, D2, E3 silica gel, G1, J1
maintenance of radiello, A7 strip for shelter B2, B6
MBTH, G1 sulfur dioxide, A8, F1
MBTH-azide, G1 supporting plate, A5
methanol, D2, D6 Tenax TA, M1
2-methoxyethanol, D2, E3 tetrachloroethylene, D2, D6, E3
2-methoxyethyl acetate, D2, E3 tetrahydrofuran, D2
1-methoxy-2-propanol, D2, D6, E3 thermal desorption, E1
1-methoxy-2-propyl acetate, D2, D6 thermal desorption, calibration, E5, M4
methyl acetate, D2, D6 thermal desorption, cartridge recovery, E6
3-methyl-2-benzothiazolinone hydrazone (v. MBTH) thermometer, B3
methyl-tert-buthylether (MTBE), D2 thermometer, reader, B3
methylcyclohexane, D2, D6 thermometer, software, B3
methylcyclopentane, D2 toluene, B5, B6, D2, D5, E3, E4
methylene blue, H1 1,1,1-trichloroethane, D2, D6, E3
methylethylketone, D2 trichloroethylene, D2, D6
methylisobuthylketone, D2, D6 triethanolamine, F1
methyl metacrylate, D2 1,2,4-trimethylbenzene, D2, D6, E3
2-methylpentane, D2, D6 n-undecane, D2, D6, E3
3-methylpentane, D2, D6 using radiello, A6
molecolar sieve, L1 vertical adapter, B1
molecolar sieve, duration and storage, L2 volatile organic compounds, thermal desorption, E1
naphtalene, D2 volatile organic compounds, thermal desorption, analyses, E4
NEDA, F2 volatile organic compounds, thermal desorption,
nitrogen dioxide, A8, F1 sampling rates, E1, E3
nitrous oxide, L1 volatile organic compounds, extraction with CS2, D1
n-nonane, D2, D6, E3 volatile organic compounds, extraction with CS2,
n-octane, D2, D6, E3 analyses, D4
ozone, A8, C4, G1 volatile organic compounds, extraction with CS2,
ozonide, G1 sampling rates, D1, D2
ozonolysis, G1 volatile organic compounds, extraction with CS2,
pentacyanonitrosylferrate (see cyanoferrate) retention times GC, D6
pentane, D2, D6 m-xylene, B5, B6, D2, D5, E3, E4
pentanal, B4, C1 o-xylene, B5, B6, D2, D5, E3, E4
permeative body, A5, L1 p-xylene, B5, B6, D2, D5, E3, E4
phenol, I1, M1 xylenol (see dimethylphenol)
a-pinene, D2, D6, E3
polycarbonate screw-thread cap for radiello-ready-to-use, A8
polypropylene tube, B6
propanal, B4, C1
propyl acetate, D2, D6
propylbenzene, D2
4-pyridylaldehyde, G1
radial diffusion, A1, A2
reader for on-field thermometer, B3
ready-to-use, radiello -, A8
sampling, ending, A7
sampling, preparing, A6
sampling, sampling rate, definition, A1
sampling, to start on-field, A6
sevorane, L1
snapping adapter, A8
sodium hypochlorite, I1
how does
t h e d i ff u s i v e s a m p l e r w o r k ?
The diffusive sampler is a closed box, usually cylindrical. Of its two opposite sides,
one is “transparent” to gaseous molecules which cross it, and are adsorbed onto
the second side. The former side is named diffusive surface, the latter is the adsor-
bing surface (marked with S and A in the figure). axis
Driven by the concentration gradient dC/dl, the gaseous molecules cross S and
diffuse towards A along the path l, parallel to the axis of the cylindrical box. The
molecules, which can be trapped by the adsorbing material, are eventually adsor-
bed onto A according to the equation: diffusive surface S l
dm dC
=DS [1]
adsorbing surface A
dt dl In the diffusive sampler, the adsor-
where dm is the adsorbed mass during time dt and D is the diffusion coefficient. bing and the diffusive surfaces are
Let C be the concentration at the diffusive surface and C0 the concentration at two opposing plane of a closed box.
Driven by the concentration gra-
the adsorbing surface, the integral of [1] becomes
dient, the gaseus molecules (colou-
m S (C-C ) red in the figure) pass through the
=D [2]
t l 0
diffusive surface and are trapped
from the adsorbing surface.
If the concentration at the adsorbing surface is negligible, the equation can be
approximated to
m S m
= D = Q and then C = [3]
tC l tQ
Q is the sampling rate and has the dimensions of a gaseous flow (if m is expressed in µg, t in minutes and C in
µg·l-1, Q is expressed in l·min-1).
Therefore, if Q is constant and measured, to calculate the ambient air concentration you need only to quantify the mass of
analyte trapped by the adsorbing material and to keep note of the time of exposure of the diffusive sampler.
To improve the analytical sensitivity the collected mass m should be increased by enlarging Q. As D is a
constant term, one can only try to improve the S/l ratio, namely the geometrical constant of the sampler.
Unfortunately, in the common axial simmetry sampler, if S is enlarged, the adsorbing surface A must be
enlarged too, in order to keep the two parallel surfaces at a fixed distance. Since the analytes can be reco-
vered from the axial sampler only by solvent extraction, any increase of A lead to a proportional increase
of the extraction solvent volume, thus the improvement of Q is canceled out by the effect of dilution.
The value of distance l could also be reduced, but under
the critical value of about 8 mm the diffusion law is no
longer valid in the case of low air velocity values, since diffusive
As S=2prh (where h is the height of the cylinder) and the diffusive path is as long as the radius r, we can then
express equation [1] as follows
dm dC
= D 2p h r [4]
dt dr
The integral of equation [4] from rd (radius of the diffusive cylindrical surface) to ra (radius of the adsorbing surface)
becomes
m 2p h
=D r =Q [5]
tC
ln d
ra
the ratio
2p h
r
ln d
ra
It is therefore very important to know the average temperature in order to ensure accuracy of experimental
data. See how you can perform on-field temperature measurements on page B3.
Even if some cartridges adsorb large quantities of water when exposed for a long time in wet atmosphere, gene-
rally this does not affect sampling by radiello.
radiello Some consequences, neverthless, can sometimes be felt on the
analysis. As an example, a very wet graphitised charcoal cartridge could generate ice plugs during cryogenic focu-
sing of thermally desorbed compounds or blow out a FID flame.
It is therefore important to protect radiello from bad weather. See page B1 how this can be easily done.
why is radiello so
sp ecial?
The diffusive sampling does not involve the use of heavy and encumbering pumping systems, does not have energy
power supply problems, does not require supervision, is noiseless, is not flammable and does not represent an
explosion hazard, can be performed by everybody everywhere and with very low costs.
Moreover, it is not subject to the breakthrough problem, which can be serious when active pumping is performed.
In pumped sampling the adsorbed compound beha-
ves as a chromatographic peak (top): air flow displa-
ces it along the adsorbent bed and its concentration
is distributed as a gaussian function. Eventually, the
compound comes out from the opposite end. When
its concentration in the outlet air is 10% of the con-
centration in the sampled air we say that the
breakthrough has been reached or, with a mislea-
ding expression, that the tube has been saturated.
Any further pumping leads to a loss of analyte and a
-3
100 µg.m
consequent underestimation of the environmental
100
pumpled
concentration. The extent of this phenomenon
µg.m-3 flowing out
samplers
depends weakly on the concentration of target com-
80
Moreover, radiello
4is able to work properly also with bad weather conditions
due to the water-repellent diffusive body
4has blank values lower than three times the instrumental
noise due to the complex conditioning procedures of the
bulk adsorbing (or chemiadsorbing) materials and to the
repeated quality controls along the whole production
4has low detection limits and high adsorbing capacities that
allow exposure time duration from 15 minutes to 30 days
and concentration measurements from 1 ppb to over 1000
ppm
4offers high precision and accuracy over a wide range of
exposure values
All the images in the manual concern the Environmental Research Center of the Istituti Clinici Scientifici Maugeri
t h e c o m p o n e n ts o f
radiello
The essential parts of radiello are the adsorbing cartridge, the diffusive body, the supporting plate and the adhesi-
ve label with the bar code indication. Apart from the adsorbing cartridge, if not differently stated, all of the other
components can be repeatedly used for several sampling
The ads or bing c ar t ri dge experiments.
Depending on the polluting compound to be
sampled, many different adsorbing or chemiad- T h e d i f f u s i v e b o d y
sorbing cartridges have been developed. Their Four kinds of diffusive bodies are available, with like outer dimen-
dimensions are neverthless the same for all: 60 sions: 60 mm height and 16 mm diameter.
mm length and 4.8 or 5.8 mm diameter. The white diffusive body, code RAD120, of general use, is made
They are contained in glass or plastic tubes of microporous polyethylene 1.7 mm thick and average porosity
wrapped up in a transparent polyethylene ther- 25 ± 5 µm. Diffusive path length is 18 mm.
mowelded bag. The blue diffusive body, code RAD1201, has the same proper-
The code number, printed onto the bag along ties of the white one but is opaque to light: it is suited to the sam-
with the lot number and expiry date indicates the pling of light-sensitive compounds.
kind of cartridge. The yellow diffusive body, code RAD1202, should be used whe-
Apart from the thermal desorption cartridges, all never the sampling rate must be reduced; it is made of micropo-
of the other kinds are for single use only. rous polyethylene 5 mm thick and average porosity 10 ± 2 µm.
Available in 5 or 20 pieces per package. Diffusive path length is 150 mm.
The cartridge has to be introduced into the The permeative diffusive body, code RAD1203, is a 50 µm thick
diffusive body. silicone membrane strengthened by a stainless steel net and a
microporous polyethylene cylinder. It is employed for anaesthetic
gases and vapours sampling.
Available in 20 pieces per package only.
The diffusive body has to be screwed onto the supporting plate.
T h e su p p o r ti n g p l a t e
It is identified by the code 121.
Made of polycarbonate, it acts
both as closure and support for RAD120 RAD1201 RAD1202 RAD1203
the diffusive body, which has
to be screwed onto the The lab el
thread. It comes along with a Self-adhesive, with printed bar-
clip and a transparent code number. Since each bar-
adhesive pocket to hold code number has been printed
the label. The three in only one copy, it allows an
parts are to be assem- unmistakable identification of
bled before use (see the sampling tube on field and
page A6). in the laboratory for the subse-
Available in 20 quent analysis.
pieces per pac- Each package of 20 adsorbing
kage only. cartridges contains also 21 labels.
code RAD190 If the labels are ordered separately,
code RAD121 they are shipped in 198 pieces per
package only.
user tip
assemble the supporting
plate in your laboratory
before the sampling cam-
paign: on the field they are
uselessly time-consuming.
and stick it onto the plate in a central position; 4
if you prefer, the pocket can be applied to the rear of the plate, but
BE CAREFUL, always with the label insertion slot on the side
(otherwise, if it starts raining the label can get wet)
o n-field
to start the sam pling
1
open the plastic bag, draw the cartridge out from the tube and put
it in the diffusive body. Keep the glass or the plastic tube and
stopper in the original plastic bag.
The lower part of the diffusive body holds a seat for the
central positioning of the cartridge. A correctly centered
cartridge should not stick out even by half a millime-
ter. If it is not so, the cartridge is not cor-
rectly positioned and is out of axis.
As a consequence, when the diffusive body is
screwed onto the supporting plate the cartridge is
bent, the geometry of the sampler is disturbed and 2 keeping the diffusive
the results obtained become unreliable. body in a vertical posi-
To place the cartridge centrally you need tion, screw it onto the
only to tap on the supporting plate.
diffusive body.
BE CAREFUL: do not hold
user tip the diffusive body hori-
Do not touch the zontally when you screw
cartridge with your it onto the plate, otherwise the cartridge could
fingers if possible, come out from its seat and stick out.
particularly if it is Insert a label in the pocket without peeling it
impregnated with 3 off. Keep note of the date and time and expo-
reactive se radiello. Sampling has started.
user tip
even if you can write date and time of the sampling start and end on the adhesive label, we suggest you to keep
note of these parameters also separately: after a week exposure with bad weather conditions, your writings could
become illegible!
DO NOT USE MARKING PENS to write on the label: they contain solvents that are sampled by radiello!
after th e sa mpling
Keep note of the date and time of the end of exposure.
4
Place the cartridge into the tube, peel off the label and stick it onto the
tube such that the barcode is parallel to the axis of the tube.
If you have performed the sampling of different polluting compounds at the same
time, BE CAREFUL NOT TO MIX UP THE TUBES: place the exposed cartridge in
its original tube, identified by the code printed on the plastic bag.
IMPORTANT
Always stick the label such that the barcode is parallel to the axis of the tube:
any other position will compromise the barcode automated reading by the optic
reading device.
maintenance
When exposed outdoors or in a workplace environment, the diffusive body may get dirty from airborne dust. Fine
particles (PM10) are especially harmful to yellow diffusive bodies since they can obstruct the pores. When the diffu-
sive bodies are dirty you can wash them as follows.
Immerse the diffusive bodies in a beaker with a soapy solution (e.g. dish detergent) and sonicate them for 20 minu-
tes. As the diffusive bodies float, you may make them sink by putting a smaller beaker on them, with water inside
enough to dip it a few centimeters.
Rinse the diffusive bodies with plenty of water and then deionized water; let them finally dry in the air.
After four or five washings, diffusive bodies need replacing: repeatedly adsorbed dust may have penetrated the
pores such deeply to be undisturbed by washing.
The following table shows the advised washing schedule:
radiello - r e a d y - t o - u s e
The ready-to-use version may be advantageous when you prefer not to assemble all of the components on field. It
can be purchased as it is or in separate parts to be assembled by the customer.
In the as-it-is version the adsorbing cartridge is already contained in a diffusive body closed with a poly-
carbonate screw-thread cap. The whole is closed in a polypropylene airtight container. Just before use
draw the diffusive body out of the container and fit it to the special snapping vertical adapter fixed to the
supporting plate. After the end of exposure, the diffusive body with its content is placed again in the poly-
propylene airtight container to be shipped to the laboratory for analysis. The ready-to-use as-
it-is radiello (polycarbonate cap, glass or plastic tube, special vertical adapter, barcode label
and polypropylene container comprised for each type) is available for the sampling of the fol-
lowing compounds:
ü diffusive bodies (of the required type, see following chapters) user tip
ü adsorbing cartridges (of the required type, see following chapters)
ü polycarbonate caps, code RAD1241 the ready-to-use version of radiello
ü special snapping adapters, code RAD1221 is very useful in the workplace sam-
ü polypropylene containers, code RAD1242 pling campaigns but is not advised if
ü supporting plates, code RAD121 very low concentrations in outdoor or
domestic environments are to be
measured
accessories
vertical adapter
code RAD122
Available in 20 pieces per package only
The diffusive body can be fitted to the supporting plate either in a vertical or
horizontal position, the vertical one being more comfortable when radiello is
used for personal sampling.
To assemble radiello in vertical position you have
to screw it to the vertical adapter code RAD122,
fitted to the supporting plate.
code RAD122
IMPORTANT
when mounting the diffusive body be careful to
keep it vertical with the thread upside (see
page A6).
shelter
code RAD196
For outdoor exposures a mountable polypropylene shelter is available
which can be hanged to lamp posts.
Available in 10 pieces per package only
It has been designed to be mounted easily and without any tool on
field, so that it is not cumbersome when you transport it from your labo-
ratory. Once assembled, it ensures the best compromise between pro-
tection against bad weather and ventilation.
It can house up to four radiello and is able to fit a wide range of pole
diameters.
Its colour is quite similar to that of the majority of lampposts: being less
visible, it is less subject to acts of vandalism.
two spacers
two strips
user tip
If the pole diameter is larger than the strip length, you can put two or more strips together to extend the fastening
system.
If the sampling site is very windy, do not introduce more than two radiello samplers in each shelter, otherwise rain
could dampen the outermost samplers.
O n - f i e l d t e m p e r a t u r e m e a s u r e m e n ts
codes RAD126 and RAD127
Since the uptake rate values of radiello depend on tempera-
ture, the concentration values obtained will be more accurate
if precise temperature measurements are performed during
the sampling.
To get reliable temperature data you may ask the local weather
station, if there is one, and if the measurements are performed
nearby your sampling sites. Bear in mind that you should take
into the account the urban heat island: did you know that there
can be a difference of even 4-5 °C between the center and the
suburbs of a big town?
With radiello you can create your own temperature
measurement station.
A thermometer with precision ± 0,5 °C between -20 and 80 °C thermometer code RAD126
and equipped with a data logger capable of recording 2048 data points Available in 3 pieces per package only
has been fixed to a vertical adapter (code RAD126). It is tiny enough (<
1 cm3) to go perfectly unobserved.
It has no battery to replace, needs no maintenance and works properly even with bad weather conditions.
Its memory allows you to record one temperature value every 15
minutes for 22 days, or every 30 minutes for 43 days, or every 60
minutes for 85 days, or... it lasts ten years or a million readings!
The thermometer is fitted to the supporting plate of radiello: use
the sampler normally and measure temperature and pollution at
the same time.
A very simple reader (code RAD127), connected to your PC by
a serial port, allows you to program the temperature sensor for
the measurements on field, to download the aquired data and to
perform data statistical and graphic processing by a very user-
friendly software.
One reader serves an unlimited number of thermometers. The reader code RAD127
SmartButton Reader Solution software needed to program the ther- single unit, serial port adapter included
mometers and download the data can be purchased from the
parent company's website at the link:
http://www.acrsystems.com
user tip
When performing urban monitoring install a thermometer every
ten sampling sites. IIf this may help you, contact us to discuss
sampling strategies.
c a l i b r a t i o n s o l u t i o n f o r H 2S
code RAD171
Code RAD171 relieves you from the task of preparing the sodium sulfide standard solution for the calibration curve
used for the determination of H2S by the cartridge code 170 (see page H1).
Since sodium sulfide is deliquescent, its weight is not a primary standard and sodium sulfide solution need titration
once prepared. Moreover, titration must be repeated often due to the instability of diluted solution (one hour time is
sufficient to decrease sulfide content by 10%).
Code RAD171 is a methylene blue concentrated solution
that, once diluted 1:50, provides the same absorbance equivalent to
Solution ml of ml of water
value at 665 nm of a sodium sulfide solution of with con- µg.ml-1 of S=
centration 1.145 µg·ml-1 sulfide ions. A 2 di codice 171 98 1.145
This concentration value has been chosen to obtain the
B 25 di A 25 0.572
highest absorbance value within the linearity range of the
spectrophotometer. C 10 di A 40 0.229
filtration kit
code RAD174
Code RAD174 filtration kit is composed by 20 single use plastic syrin-
ges and 20 single use micropore hydrophilic polypropylene filters with
diameter 13 mm and 0.45 µm porosity.
Both filter and syringe are suited to filtration of aqueous solutions with
pH in the range of 0 to 12 with commonplace eluents for ion chromato-
graphy and reverse phase HPLC.
which are blanks and nine, divided into three groups of m-xylene 12 24 48
three, preloaded with VOCs to simulate 8 hours exposures
(480 minutes) to the concentrations listed in the table. The p-xylene 12 24 48
values shown are indicative, actual ones are certified for o-xylene 12 24 48
each lot.
The composition of the mixture is simple but it includes butanol 15 30 60
compounds with different polarity. The loaded mass is cal- 2-etoxyiethyl acetate 2.5 5 10
culated in order to represent exposures to 0.5, 1 and 2
times the TLV value for the mixture.
VOCs loading is performed by injection of precisely known amounts of calibrated mixtures of the eight compounds
under nitrogen flow.
Kept at 4 °C, the cartridges are stable for at least four months.
codice RAD175
stainless steel net, Clips
100 mesh, 5.9 mm Code RAD195
Available in 20 pieces
per package only.
codice RAD176
stainless steel net,
100 mesh,
4.9 mm diameter
codice RAD177 Tubes
stainless steel net, Available in 20 pieces per package only.
3x8 µm,
4.9 mm diameter
code RAD1991
glass tube,
working volume 2.8
Strip ml
Code RAD198
Useful for repositioning of
radiello shelter. code RAD1992
Lenght 75 cm. polypropylene tube,
Available in 100 pieces working volume
per package only. 12 ml
Aldehydes
what you need
blue diffusive body code RAD1201
supporting plate code RAD121
vertical adapter code RAD122 (optional)
chemiadsorbing cartridge code RAD165
filtration kit code RAD174 (only for analysis)
Principle
Code 165 is a stainless steel net cartridge filled with 2,4-dinitrophenylhydrazine (2,4-DNPH) coated Florisil®.
Aldehydes react with 2,4-DNPH to give the corresponding 2,4-dinitrophenylhydrazones
O2N R O2N R
F-NH-NH2 + O=C F-NH-NH=C + H 2O
O2N H O2N H
2,4-DNPH aldehyde 2,4-DNPhydrazone
The 2,4-dinitrophenylhydrazones are then extracted with acetronitrile and analyzed by reverse phase HPLC and UV
detection.
Sampling rates
Sampling rates values at 298 K (25 °C) and 1013 hPa are listed below:
QK = Q298 ( 298
K
) 0.35
where QK is the sampling rate at the temperature K and Q298 is the reference value at 298 K. This produces a varia-
tion of ± 1% for 10 °C variation (upwards or downwards) from 25 °C.
Sampling rate is invariant with humidity in the range 15-90% and with wind speed between 0.1 and 10 m·s-1.
Calculations
The average concentration C over the whole sampling time (in µg·m-3) is calculated according to the expression:
m [µg]
C [µg·m-3] = 1,000,000
Q [ml·min-1] · t [min]
where:
m = mass of aldehyde in µg
t = exposure time in minutes
Exposure
The optimum exposure duration varies with the expected concentration. Taking formaldehyde as an example, con-
centration values of 5-30 µg·m-3 are usually found in outdoor urban measurements while 20-200 µg·m-3 are expec-
ted in workplace environments. In workplace environments concentrations may be as high as 2,000-3,000 µg·m-3
for short time intervals: it can therefore be interesting to evaluate the peak value (usually referred to by STEL). The
corresponding advised exposure time is shown in the table below:
Advised exposure times
minimum 8h 8h 2h 15 minutes
30,0
µg found onto the cartridge
20,0
Do not expose all of the cartridges belonging to the same lot: keep at least two cartridges as blanks.
15,0
Storage 10,0
The cartridges need to be kept, properly sealed, in a dark place at 4 °C for ensuring a shelf life (according to EN
0 1 3 5 7 14 21 28 35 42 60
13528-2) of six months. If stored at -18 °C, the shelf life will be twelve months. Each lot is approved for use when
days after exposure
20.0
18.0
16.0
they are stable for 60 days. After solvent desorption (see Analysis) discard the cartridge. The resulting solution, well
capped and stored at 4 °C, is stable for at least 42 days.
Analysis
Desorption
Materials user tip
- HPLC grade acetonitrile
For a reliable and rapid filtra-
- class A volumetric pipette, capacity 2 ml
tion employ the filtration kit
- micropore filter membranes, porosity 0.45 µm, solvent resistant
code RAD174.
Procedure
To obtain an accurate calibra-
Introduce 2 ml acetonitrile directly in the cartridge tube, recap and stir from time
tion curve we offer you the cali-
to time for 30 minutes. Discard the cartridge. Filter the resulting solution and keep
bration solution code RAD302.
it well capped until analysis time. If analysis has to be delayed, store the solution
at 4 °C.
Instrumental analysis
The method suggested below is only indicative; the analyst can choo- IMPORTANT: verify the presence
se an alternative method, on the basis of its personal experience. and the abundance of the 2,4-DNPH
chromatographic peak: otherwise, the
cartridge could be saturated.
Materials
- reverse phase C18 HPLC column, length
formaldehyde
acetaldehyde
benzaldehyde
isopentanal
Procedure
Set the detector at the wavelength of 365 pentanall
nm. Inject between 10 and 50 µl of solu-
hexanal
acrolein
propanal
IMPORTANT
Acrolein gives place to three chromatographic peaks, two of them are unresolved. Calculate the concentration
basing onto this most abundant peak and ignore the others.
Isopentanal appears as two unresolved peaks: its concentration should be obtained by integration of both peaks
as a sum.
user tip
If you perform several analyses, a barcode reader will greatly improve productivity in your laboratory and will also
minimize the possibility of errors in the copying of sample labels.
Please contact us to help you in the implementation of the reader.
Interferences
Other carbonyl compounds
All carbonyl compounds, ketones included, react with 2,4-DNPH but do not interfere in the analysis if proper
chromatographic parameters are selected.
In the described chromatographic conditions acetone-2,4-DNPH peak is well resolved from acrolein-2,4-DNPH.
Neverthless, if acetone concentration is higher than 50,000 µg·m-3, acrolein-2,4-DNPH peak intensity is depressed
by 25%.
Ozone
Examples of ozonolysis of dinitrophenylhydrazones on Effetto dell'ozono sulla portata
active supporting materials as silica gel are found in the
literature.
105
95
acetaldeide
val, is higher than 100 ppb. Since this is not usually the
isopentanale
75
pentanale
Principle
Code RAD130 cartridge is a stainless steel net cylinder, with 100 mesh grid opening and 5.8 mm diameter, packed
with 530 ± 30 mg of activated charcoal, particle size is 35-50 mesh. Volatile organic compounds are trapped by
adsorption and recovered by carbon disulfide displacementI, analysis is performed by FID gas chromatography.
Sampling rates
The table on pages D2 and D3 lists sampling rate values at 298 K (25 °C) and 1013 hPa, experimentally measured
in a standard atmosphere chamber. For other compounds whose diffusion coefficient1 is known sampling rate can
be calculated according to equation [5] on page working principle, taking into account that white diffusive body and
code 130 cartridge give the geometric constant of radiello the value of 14.145 ± 0.110 cm. Several experiments per-
formed in the standard atmosphere chamber demonstrate that the calculated sampling rates seldom deviate by
more than ±10% from the experimentally measured values.
QK = Q298 ( 298
K
) 1,5
where QK is the sampling rate at the temperature K and Q298 is the reference value at 298 K. This produces a varia-
tion of ±5% for 10 °C variation (upwards or downwards) from 25 °C.
Sampling rate is invariant with humidity in the range 15 ÷ 90% and with wind speed between 0.1 and 10 m·s-1.
1
Lugg G.A.: Diffusion Coefficients of Some Organic and Other Vapors in Air. Anal. Chem. 40-7:1072-1077 (1968).
Calculations
The listed sampling rate values already take into account for the desorption efficiency with carbon disulfide. The ave-
rage concentration over the exposure time interval is therefore calculated from the mass of analyte found onto the
cartridge and exposure time without introducing any corrective factor, apart from corrections due to average tem-
perature different from 25 °C.
Average concentration (in µg·m-3) over the whole exposure time is calculated according to the following expression:
m [µg]
C [µg·m ] = -3
1,000,000
QK [ml·min-1] · t [min]
where:
m = mass of analyte in µg
t = exposure time in minutes
Notes:
a = weakly adsorbed compound. If its concentration is higher than the TLV for the workplace environments it may be partially
displaced by other compounds that are more strongly trapped if their concentration is also high. If this is the case, it is advi-
sable to reduce sampling time under 8 hours.
b = prolonged exposure of charcoal cartridges at relative average humidity higher than 80% causes adsorption of up to 100 mg
of water. Water does not interfere with adsorption mechanisms but is displaced by carbon disulfide and gives raise to a sepa-
rate layer. Some very water soluble polar compounds will distribute between the two solvents, thus provoking an undere-
stimation of the actual air concentration since only the carbon disulfide is injected in the gas chromatograph. When the con-
centration of polar compounds has to be determined, the calibration curve should be prepared by spiking 50 µl of water in
each tube containing the cartridge and the 2 ml of carbon disulfide standard solution (see Analysis).
c = better reproducibility obtained by use of methanol as extraction solvent instead of carbon disulfide.
Limit of quantitation
The limit of quantitation depends on the instrumentation and on the analytical conditions. The minimum revealable
environmental concentration can be estimated on the basis of the equation on chapter Calculations, where m is the
minimum revealable mass, experimentally measured for each compound. Under the analytical conditions described
on the following chapter Analysis, the limit of quantitation for 7 days exposure usually ranges from 0.05 and 1 µg·m-3,
depending on the compound.
In any case, the limit of quantitation can never be lower than the inferior limit of the linearity range indicated in the
previous table.
Exposure
Code RAD130 cartridge has a very large loading capacity: about 80 mg, corresponding to an overall VOCs con-
centration of 3,000 - 3,500 mg·m-3 sampled for 8 hours or 70,000 - 80,000 µg·m-3 sampled for 14 days. Neverthless,
if the quantified overall adsorbed mass should be near 80 mg, sampling rate could have deviated from linearity. If
this is the case, it is advisable to repeat the sampling experiment reducing exposure time.
Workplace environment
In workplace environments complex mixtures of airborne solvent vapours are often found at concentrations 2,000-
3,000 mg·m-3. The outstanding adsorbing capacity of code RAD130 cartridges allows you to sample them for the
whole working shift of 8 hours. On the other hand, the very high values of sampling rates for a variety of compounds
allow you to perform accurate concentration measurements even after very short exposures. For example, 15 minu-
tes are enough to measure 0.1 mg·m-3 of benzene.
radiello can therefore be employed to evaluate both TWA and STEL concentrations.
Storage
The activated charcoal cartridges have undergone a complex conditioning process that ensures an outstanding
chromatographic blank level, never exceeding three times the instrumental noise of a FID detector at the lowest atte-
nuation.
Kept in a cool place and away from volatile organic compounds, the cartridges mantain unchanging blank level and
adsorbing capacity for at least two years. Expiry day and lot number are printed onto the plastic bag wrapping each
cartridge: its integrity stands as warranty seal.
After exposure the cartridges, well capped and kept in a cool and solvent-free place, maintain their content unalte-
rated for at least six months
Analysis
Extraction
Introduce 2 ml of CS2 and 100 µl of internal standard solution (see next) directly in the radiello glass tube without
drawing out the cartridge. Always use class A volumetric pipettes or dispensers. Stir from time to time for 30 minu-
tes. If analysis is not performed soon after, draw out the cartridge and
discard it. IMPORTANT
always use high purity grade CS2, for
example Honeywell cod. 342270.
Calibration
outdoor environment sampling BE CAREFUL
If benzene, toluene, ethylbenzene and xylenes (BTEX) have to be even refrigerated, CS2 permeates the
analyzed, prepare three or four standard solutions in CS2 having tube plastic cap: its volume decreases by
decreasing concentrations of the analytes in the following ranges (in 4-5% a day. If the internal standard has
mg·l-1): been added, it is only matter of unplea-
benzene 0.04 ÷ 17.6 etilbenzene 0.04 ÷ 17.7 sant odour...
toluene 0.09 ÷ 34.8 m-xilene 0.04 ÷ 17.2
o-xilene 0.04 ÷ 17.6 p-xilene 0.04 ÷ 17.2
user tip
Analysis of unknown samples For a very accurate calibration
Identify the sample that has been exposed for the longest time or at the highest we offer the preloaded cartrid-
expected concentration. Introduce 2 ml of CS2 but do not add the internal stan- ges code 405 (outdoor envi-
dard, stir and let the sample stand for 30 minutes. Without discarding the cartrid- ronment) and code 406 (work-
ge, inject the CS2 solution in the gas chromatograph with FID detector (see place environment).
below), identify the compounds appearing in the chromatogram and make an esti-
mation of the order of magnitude of their concentrations.
Prepare a CS2 solution of the identified compounds with doubled concentration with respect to the sample. Dilute this
solution in order to obtain standard solutions of concentration respectively about 0.1, 0.5 and 1 times the concentra-
tion estimated in the sample. Introduce 2 ml of each standard solution onto a blank code RAD130 cartridge in its glass
tube, along with the chosen internal standard solution.
The chosen internal standard should have a retention time that does not cause interference with other compounds in
the chromatogram. Compatibly with this requirements, we suggest to employ a solution of 2-fluorotoluene (e.g. Aldrich
F 1,532-3) in CS2 with concentration of 100 µl·l-1 for outdoor samples and 2 ml·l-1 for workplace samples.
Add 2 ml of CS2 and the internal standard to all of the samples, stir, let the samples stand for 30 minutes and discard
the cartridges prior to the analysis.
USER TIP
If you perform several analyses, a barcode reader will
greatly improve productivity in your laboratory and will
also minimize the possibility of errors in the copying of
sample labels.
Please contact us to help you in the implementation of the
reader.
the production
propyl acetate 11.868
n-eptane 12.068
1-ethoxy-2-propanol 12.775 The cartridge is filled up with charcoal by a very complex automated appara-
methylcyclohexane 12.912
methylisobutylketone 13.258
tus that was designed and realised in our laboratory. It avoids any contamina-
isobutyl acetate 14.005 tion of the adsorbing material during the delicate process of cartridge produc-
toluene 14.055 tion and ensures a very accurate dosing of the material itself, providing a varia-
butyl acetate 15.279 bility of less than 2% of the weight of the activated charcoal among the car-
n-octane 15.435
tetrachloroethylene 15.601
tridges.
diaceton alcohol 15.915
Principle
Code RAD145 is a stainless steel net cylinder, with 3x8 µm mesh opening and 4.8 mm diameter, packed with 350
± 10 mg of graphitised charcoal (Carbograph 4), particle size is 35-50 mesh.
Volatile organic compounds are trapped by adsorption and recovered by thermal desorption, analysis is performed
by capillary gas chromatography and FID or MS detection.
General considerations
Thermal desorption is an easy-to-use technique,
but it implies some precautions and is of less
k3>k2
25 °C
sampling rates are reduced to less than one third compared to those obtained with white diffusive bodies.
Some compounds, moreover, are thermally unstable. Thermal degradation of such com-
pounds will cause an underestimation of their concentration or the appearance of ghost
peaks.
Thermal desorption is neverthless an outstanding analytical technique because it is easy to
perform, it does not require the use of toxic solvents as carbon disulfide, it ensures very low
limits of detection, is suited to mass spectrometric detection and allows the recovery of the
adsorbing cartridges. Basing on our experience, we have chosen Carbograph 4 as the best
compromise between sampling efficiency and recovery yields for a wide range of organic
compounds.
Sampling rates
Sampling rate values at 298 K (25 °C) and 1013 hPa are listed in table on page E3. All of the values shown have
been experimentally measured. Exposure tests have been performed up to the levels shown (in µg·m-3·min) and
sampling rates are guaranteed to be linear up to the limit values and for overall concentration of volatile organic
compounds in air not exceeding 2,000 µg·m-3.
( )
K 1,5
QK = Q298
298
where QK is the sampling rate at the temperature K and Q298 is the reference value at 298 K. This produces a varia-
tion of ± 5% for 10 °C variation (upwards or downwards) from 25 °C.
Sampling rate is invariant with humidity in the range 15-90% and with wind speed between 0.1 and 10 m·s-1.
Do not expose directly radiello to rain: even if small amounts of water are adsorbed by Carbograph 4, they can neverthless
interfere with analysis.
Calculations
The listed sampling rate values take already into account the recovery yields of adsorbed compounds. The avera-
ge concentration over the sampling period is therefore calculated from sampled mass of analyte and exposure time
without introducing any other corrective factor, apart from temperature variations of Q.
Average concentration C in µg·m-3 over the whole exposure time is calculated according to the following expression:
m [µg]
C [µg·m-3] = 1,000,000
QK [ml·min-1] · t [min]
where:
m = mass of analyte in µg
t = exposure time in minutes
Exposure
Workplace environment
The use of ligth adsorbing media is not recommended in the workplace environment.
1
after 7 days exposure and with MS detection; analytical conditions as described in the Analysis paragraph
2
for overall VOCs concentrations not exceeding 500 µg·m-3
Storage
The cartridges are thermally conditioned in a high temperature stove with an inert atmosphere, with an oxygen con-
tent lower than 10 ppm. The duration of the adsorbent capacity of graphitized carbon is virtually unlimited and has
been tested six months after production. Cartridges should be stored in a clean and solvent-free environment, in the
refrigerator or at room temperature. The expiry date and the lot number are printed on the transparent plastic casing,
whose integrity acts as a guarantee seal
Analysis
The methods proposed here have been elaborated with the Perkin-Elmer Turbomatrix thermal desorber coupled to
the Agilent 6890 gas chromatograph and Agilent 5973 mass spectrometer. They can naturally be transferred to other
instruments.
A method for BTEX and one for VOC are proposed here. The first refers to samples from urban air monitoring where
research is usually limited to benzene, toluene, ethylbenzene and xylene isomers. The second is more suitable for
indoor investigations, allowing the quantification of all the compounds listed in the table above and the more gene-
ral qualitative research, which also includes analytes with medium polarity.
Desorption
The thermal desorber is equipped with 1/4” s.s. sample tubes, they have to be hollow and free: discard the stain-
less steel gauze disk which is fitted to the groove and discard also the springs if present.
Code RAD145 cartridge has been dimensioned to fit the diameter of Turbomatrix thermal desorption tubes. Its
length is such that, when the cartridge is introduced into the tube and is stopped by the groove, it is positioned
exactly centrally with respect to the tube length. The same considerations apply to the Markes Unity thermal desorber.
Once capped, the Turbomatrix steel tube has to be positioned in the carousel with the grooves on the bottom.
The described conditions have been optimized for seven days exposures to typical concentrations of urban atmo-
spheres and indoor environments. Shorter exposure times or considerably higher concentrations would require dif-
ferent settings of split flows.
Instrumental analysis
Column
J&W PONA, length 50 m, d.i. 0.2 mm, film thickness 0.5 µm; the column head can be connected directly to the
Turbomatrix six-way valve.
Temperatures
GC oven: 36 °C for 1 minute, 6 °C·min-1 up to 110 °C, mantain for 1 minute, 20 °C·min-1 up to 250 °C, final isotherm
5 minutes.
Flows
Carrier gas: nitrogen at 0.8 ml·min-1
Here, below, we display two total ion current chromatograms from an outdoor urban site and an indoor sampling
respectively.
In the first case, the benzene peak corresponds to an average concentration of 2.2 µg·m-3 ; in the second the con-
centration of 1,4-dichlorobenzene was 14 µg·m-3.
Calibration
Calibration curves are obtained by gas-phase injection of
methanol solutions of the target compounds onto blank
cartridges. Injections are performed through a GC injec-
tor, where a short piece (10 cm) of wide-bore (0.25 mm
i.d.) deactivated uncoated column is installed. The other
end bears a Swagelock reducing connection (1/16”-1/4”).
The 1/4” Swagelock nut has to be equipped with a PTFE
ferrule instead of the original steel one (use PTFE ferru-
les that come along with the Turbomatrix caps).
Introduce a blank cartridge in a Turbomatrix tube and fit
the tube to the Swagelock nut. Mantain the injector at 170
°C but do not heat the oven. Inject slowly 1 µl of each cali-
bration solution under nitrogen flow (40 ml·min-1) and let
the system purge for 2 minutes. Analyze the cartridge as
you would do with a sample. We suggest you to prepare
a complete set of calibration solutions by subsequent
dilutions such as they contain, for example, 8, 4, 2, 1, To prepare the calibration standards fit a 1/16”-1/4”
Swagelock reducing connection to the GC injector by a
0.04, 0.02 and 0.01 µg·µl-1 of each compound.
short piece (10 cm) of wide-bore deactivated uncoated
column.
TIC chromatograms of an
outdoor urban sampling (left)
and of indoor air (bottom).
Mass spectra of benzene and
of 1,4-dichlorobenzene are
shown on bottom of each pic-
ture, at concentrations of 2.2
and 14 µg·m-3 respectively.
Despite the low concentration
values, the signal-to-noise
ratio is very high in both
cases. As a consequence,
very reliable mass spectral
identification is possible by
comparison with mass spec-
tral data libreries with no
need of further processing.
Cartridge recovery
The cartridges can be recon-
ditioned using the thermal
desorber in "tube conditio-
ning" mode, heating them to
350 °C for at least 20 minutes
in an inert gas flow (helium or
nitrogen at a flow of 50 ÷ 100
ml·min-1).
Principle
The cartridge code RAD166 is made of mycroporous polyethylene coated with triethanolamine (TEA). Nitrogen
(NO2) and sulfur (SO2) dioxide is chemiadsorbed onto TEA as nitrite and sulphite or sulphate ions respectively.
Nitrite is quantified by visible spectrophotometry while sulphite and sulphate are analysed by ion chromatography
(NO2 and SO2 can be analysed together by ion chromatography).
Sampling is selective for gaseous molecules: any airborne nitrite, sulphite or suplhate will not cross the diffusive
membrane.
Sampling rates
NO2
The sampling rate value Q298 at 298 K (25°C) and 1013 hPa is 0.141 ± 0.007 ng·ppb-1·min-1.
SO2
The sampling rate value Q298 at 298 K (25°C) and 1013 hPa is 0.466 ± 0.022 ng·ppb-1·min-1.
Sampling rate of NO2 varies from the value at 298 K on the effect of temperature (in Kelvin) following the equation:
QK=Q298· K ( )
7,0
298
where QK is the sampling rate at the temperature K ranging from 263 to 313 K (from -10 to 40 °C) and Q298 is the
reference value at 298 K.
Sampling rate for SO2 does not vary with temperature between 263 and 313 K (from -10 to 40 °C).
Sampling rate is invariant with humidity in the range 15 - 90% and with wind speed between 0.1 and 10 m·s-1 for both
gases.
SO2
Convert the sulphite found onto the cartridge into sulphate by multiplying its mass by 1.2, then sum the obtained value to
the sulphate found in the cartridge. The concentration in ppb is calculated according to the equation:
mSO4
CSO2 =
0.466· t
where mSO is the overall sulphate mass in ng found in the cartridge (sulphate itself and sulphite converted into sulphate)
and t is exposure time in minutes.
4
Exposure
Exposure up to 15 days is feasible but if relative humidity is higher than 70% for the entire sampling duration it is not
advisable to sample for more than 7 days. Due to the fact that TEA is very hygroscopic in fact, even if water does not
actually interfere with sampling or analysis, the excess water adsorbed by the cartridge could cause some loss of adsor-
bing medium by percolation.
WARNING: NO2 results may differ from those produced by automatic chemiluminescent instrumentation due to expo-
nential variation of the sampling rate of radiello with temperature. This phenomenon is characteristic of all NO2 samplers
that use TEA as an absorbent medium. The reason is not yet completely clear, but it is assumed that it depends in part
on the balance in the air between the species NO2 and N2O4, whose ratio is strongly linked to temperature: the TEA cap-
tures only the species NO2.
Storage
The cartridges are stable for at least 12 months before and 4 months after the sampling, if kept in the dark at 4 °C.
Expiry date is printed on the plastic bag.
Do not expose all of the cartridges belonging to the same lot, keep at least two of them as blanks.
Analysis
Add 5 ml of water in the plastic tube with the cartridge and stir vigorously by a vortexer for 1 minute. Do the same
with two-three unexposed cartridges.
Colorimetric determination of nitrite ion
Nitrogen dioxide is quantitatively converted to nitrite ion. Prepare the following reactives:
• sulphanilamide: dissolve 10 g of sulphanilamide in 100 ml concentrated HCl and dilute to 1,000 ml with water
• NEDA: dissolve 250 mg of N-(1-naphthyl)ethylendiamine dihydrochloride in 250 ml of water (discard the solu-
tion when it turns brown).
Transfer 0.5 ml (or a different volume, see the table below) of the cartridge extraction solution to a plastic or glass
10 ml tube along with 5 ml of sulphanilamide reactive. Cap tigthly, stir and wait for 5 minutes. Add 1 ml of NEDA
reactive, stir and wait for 10 minutes. Do the same with unexposed cartridges.
Measure the absorbance of samples at 537 nm using water to zero the spectrophotometer, then subtract the blank
value from unexposed cartridges. Prepare the calibration standards in the same way from sodium nitrite solutions
of concentration ranging from 0.1 to 20 mg·l-1 expressed as NO2-.
When nitrite ion concentration is higher than 20 µg·ml-1 (corresponding to 7 days of exposure to 70 ppb) the absor-
bance value is no longer comprised in the average expected sample volume water volume
calibration curve. To analyse the samples, concentration for 7 ml to be added
draw smaller amounts of the extraction solu- days exposure in ppb ml
tion as shown in the table. In order to mantain
the overall volume unaltered, add the listed up to 70 0.5 0
volume of water. from 70 to 150 0.25 0.25
higher than 150 0.1 0.4
Ozone (O3)
What you need
blue diffusive body code RAD1201
supporting plate code RAD121
vertical adapter code RAD122 (optional)
chemiadsorbing cartridge code RAD172
Principle
The adsorbing cartridge is formed by a micropore polyethylene tube filled with silica gel coated with 4,4’-dipyridy-
lethylene and closed, at one end, by a PTFE cap. Upon exposure, acid-catalysed ozonolysis of 4,4’-dipyridylethy-
lene leads to 4-pyridylaldehyde. Silica gel ensures the presence of water, necessary to complete ozonolysis reactions.
+ O3
1,2-di(4-pyridyl)ethylene ozonure
+ H2O
2 + H2O2
In the laboratory, 4-pyridylaldheyde is condensed with 3-methyl-2-benzothiazolinone hydrazone (MTBH) to yield the
ozonure 4-pyridylaldehyde
The absorbance of the solution is measured at 430 nm. Production of 4-pyridylaldehyde is a specific reaction of
ozone; neither nitrogen oxides nor organic compounds, if present, do interfere.
Sampling rate
The sampling rate value Q298 at 298 K (25°C) and 1013 hPa is 24.6 ml·min-1.
Sampling is linear in the exposure range from 10,000 to 4,000,000 µg·m-3·min-1.
QK= Q298 ( K
298
) 1.5
where QK is the sampling rate at the temperature K and Q298 is the reference value at 298 K.
Sampling rate is not influenced by humidity or wind speed.
Calculations
The average concentration over the whole exposure time is calculated according to the equation
m [µg]
C [µg·m-3] = 1,000,000
24.6 t [min]
where m is ozone mass in µg sampled by radiello and t is exposure time in minutes.
Exposure
Introduce the cartridge in the diffusive body and make sure that the PTFE cap is positioned at the same end of
the screw.
In outdoor environments, where typical ozone concentrations range from 2 to 400 µg·m-3, we suggest exposure time
from 24 hours to 14 days. The ideal range is from 3 to 7 days.
In workplace environments it is advisable to sample over the entire 8 hours shift.
Storage
The cartridges need only protection from direct sunlight: keep them in a drawer or a cupboard at room temperatu-
re. In these conditions, the blank level does not exceed 0.015 absorbance units for up to six months.
Expiry date is printed onto the plastic bag wrapping each cartridge.
Generally, an increase of blank level does not imply that the cartridge must be discarded. The only consequence is
a corresponding increase of the analytical limit of quantification.
After exposure the samples have to be stored in the dark as before, along with three unused cartridges to be analy-
sed as blanks. Analyse them within a week.
Analysis
Reactives and materials user tip
• 3-methyl-2-benzothiazolinone hydrazone hydrocloride (MBTH): dissolve 5 g For a simple and accurate fil-
per litre in water and add 5 ml of concentrated sulphuric acid; this solution is to tration make use of the filtra-
be freshly prepared. tion kit code RAD174.
• 4-pyridylaldehyde
• micropore filter membrane 0.45 µm
IMPORTANT
Procedure
If the absorbance value is higher than the cali-
Draw the cartridge out from the plastic tube, discard the PTFE bration curve upper limit dilute the sample with
cap and pour the silica gel into the tube. Add 5 ml of MBTH the MBTH solution: never use water to dilute!
solution, recap the tube and stir vigorously. Let the tube stand Water alters the pH of the solution with unpre-
for at least one hour to react, stirring from time to time. Filter dictable variations in the linearity of absorbance
through the micropore filter (if you make use of the code 174, values vs concentration.
act as follows: fit the filter to the syringe, transfer the solution
from the tube to the syringe and filter it into a second tube or directly into the spectrophotometer measure cell).
Measure absorbance at 430 nm using water to zero the spectrophotometer. The yellow colour is stable for several
days if the solution is kept well capped in its tube.
Treat in the same manner three unused cartridges of the same lot and subtract the average blank value from the
absorbance values of the samples.
Calibration
Dissolve 100 µl (112.2 mg at 20° C) of 4-pyridylaldehyde in 1 litre of water and dilute this solution (e.g. 1:2, 1:5, 1:10)
to obtain calibration solutions. Transfer 0.5 ml of each calibration solution in a plastic tube together with 4.5 ml of MTBH
solution. Stir and let stand for one hour, then read the absorbance at 430 nm (filtration is not needed). Plot the cali-
bration curve for ozone mass vs measured absorbance, taking into account that:
1 µg of 4-pyridylaldehyde = 0.224 µg of ozone.
Principle
The cartridge code RAD170 is made of microporous polyethylene and impregnated with zinc acetate. Hydrogen sul-
phide is chemiadsorbed by zinc acetate and transformed into stable zinc sulfide.
The sulfide is recovered by extraction with water. In contact with an oxidizing agent as ferric chloride in a strongly
acid solution it reacts with the N,N-dimethyl-p-phenylendiammonium ion to yield methylene blue.
NH2 H2N N
+ + H2S +
H3C N N CH3 H3C N S N CH3
H3C CH3 H3C CH3
H H FeCl3
Sampling rate
Sampling rate Q298 at 298 K (25°C) and 1013 hPa is 0.096 ± 0.005 ng·ppb-1·min-1.
where QK is the sampling rate at the temperature K ranging from 268 to 313 K (from -5 to 40 °C).
Sampling rate is invariant with humidity in the range 10 - 90% and with wind speed between 0.1 and 10 m·s-1.
Calculations
Once QK at the sampling temperature has been calculated, the concentration C is obtained according to the equation:
m
C= 1,000
Q K· t
where m is the mass of sulphide ion in µg found onto the cartridge and t is exposure time in minutes.
Exposure
Exposure duration may vary from 1 hour to 15 days. Sampling is linear from 2,000 to 50,000,000 ppb·min of H2S.
Storage
The cartridges are stable at least for 12 months before and 6 months after exposure. Do not expose all of the car-
tridges of the same lot: keep at least two of them as blanks.
Analysis
Reactives and materials
• sulphuric acid: slowly add 25 ml of concentrated sulphuric acid to 10 ml water and let the solution cool;
• amine: dissolve 6.75 g of N,N-dimethyl-p-phenylendiammonium oxalate in the sulphuric acid solution. Dilute
this solution to 1 litre with sulphuric acid - water 1:1 v/v. Kept in a dark bottle and well capped, this solution is
stable for at least four weeks. CAUTION: this solution is very poisonous.
• ferric chloride: dissolve 100 g of ferric chloride hexahydrate (FeCl3·6H2O) in 40 ml of water.
• ferric chloride-amine: mix 10 ml of ferric chloride solution with 50 ml of amine solution. This solution has to be
freshly prepared;
• sulphuric acid for dilution: slowly dissolve 40 ml of concentrated sulphuric acid in 900 ml of water, let the solu-
tion cool and make up to 1,000 ml.
Procedure
Add 10 ml of water to the plastic tube containing the cartridge, recap and stir vigorously, preferably by a VORTEX
stirrer.
Add 0.5 ml of ferric chloride - amine solution, recap immediately and stir. The tube must be capped immediately in
order to avoid that the developed hydrogen sulfide can escape from the tube before reacting.
Wait for 30 minutes and measure absorbance at 665 nm using water to zero the spectrophotometer. The colour is
stable for several weeks.
Do the same with two or three unexposed cartridges of the
same lot and obtain the average blank value, then subtract it to IMPORTANT
the samples. Absorbance is linear up to 1,200 absorbance
units, corresponding to an exposure value of
about 80,000 ppb·min. If higher absorbance
Calibration values are obtained, dilute the samples with the
sulphuric acid for dilution.
Calibration curves may be prepared by sodium sulfide standard
solutions, which have to be titrated just before use. As diluted Be careful to apply the same dilution ratio to the
sodium sulfide solutions are very unstable (the sulfide content samples and the blanks.
can diminish as much as the 10% in an hour) it is strongly NEVER USE WATER TO DILUTE.
recommended to make use of the calibration solution code
RAD171, following the instructions included.
user tip
Code RAD171 calibration
solution relieves you from the
task of preparation and titration
of the sodium sulfide solutions.
Ammonia (NH3)
What you need
blue diffusive body code RAD1201 or white code RAD120
supporting plate code RAD121
vertical adapter code RAD122 (optional)
chemiadsorbing cartridge code RAD168
Principle
The cartridge code RAD168 is made of microporous polyethylene and impregnated with phosphoric acid. Ammonia
is adsorbed as ammonium ion. Airborne ammonium salts dispersed as particulate matter do not cross the diffusive
membrane of radiello.
Ammonium ion is quantified by visible spectrometry as indophenol: at basic buffered pH ammonium ion reacts with
phenol and sodium hypochlorite, with pentacyanonitrosylferrate catalysis (in the following cyanoferrate), to form
indophenol. The reaction product is intensely coloured in blue, and its absorbance measured at 635 nm.
NaClO
HO + NH3 + OH O N ONa
Na2Fe(CN)5NO.2H2O
(cyanoferrate)
indophenol
Sampling rate
Sampling rate Q298 at 298 K (25°C) and 1013 hPa is 235 ml·min-1.
Calculations
The concentration C in µg·m-3 is obtained according to the equation:
m
C = 0.944 1,000,000
235· t
where m is the mass of ammonium ion in µg found onto the cartridge and t is exposure time in minutes.
0.944 is the numerical factor necessary to convert ammonium ion into ammonia (see Analysis)
Exposure
IMPORTANT
Introduce the cartridge in the diffusive body and make sure that the
PTFE cap is positioned at the same end of the screw. Do not touch the microporous portion
of the cartridge with your fingers:
Ammonia is sampled linearly in the range from 2,000 - 20,000,000 µg·m-3·min. sweat contains ammonium ions.
Exposure time is allowed to range from 1 hour to 14 days.
Storage
The cartridges are stable at least for 12 months before and after exposure if kept at room temperature in an ammo-
nia-free environment. Do not expose all of the cartridges of the same lot: keep at least two of them as blanks.
Analysis
Materials
• plastic or glass tube, volume 12 ml, with cap
• micropipet with variable volume from 0.1 to 1.0 ml
• 5 ml glass pipet
Reactives
• buffer solution (pH 10.6): dissolve 1.1 g of NaOH and 3.04 g of NaHCO3 in one litre of water
• phenol: dissolve 10 g of phenol in 100 ml of ethanol
• cyanoferrate: dissolve 0.5 g of sodium pentacyanonitrosylferrate dihydrate (Na2Fe(CN)5NO·2H2O) in 100 ml of
water and add a few drops of 10% NaOH. Keep this solution in a dark bottle and prepare it freshly.
• oxidising solution: sodium hypochlorite with 1% of active chlorine in 0.2 M NaOH. Keep cool in a dark bottle.
Principle
Code RAD169 cartridge is made of stainless steel net loaded with silica gel (0.1 to 0.4 mm particle size). Gaseous
hydrochloric acid is adsorbed by silica gel and subsequently extracted with water to be quantified by ion chromato-
graphy as chloride ion.
Sampling is selective for the gaseous molecules: any airborne chloride salt will not cross the diffusive membrane of
radiello.
Sampling rate
Sampling rate (Q298) at 25 °C (298 K) and 1013 hPa is 103 cm3·min-1.
QK = 103 ( 298
K
) 1,5
where QK is the sampling rate at temperature K and Q298 is the sampling rate value at the reference temperature of
298 K. This yields a ± 5% variation of Q for a 10 °C variation (upwards or downwards) from 25 °C.
Sampling rate is invariant with humidity in the range 15 - 90% for short exposure time (see Exposure) and with wind speed
between 0.1 and 10 m·s-1.
Calculations
Let m be the mass of chloride ion in µg found onto the cartridge and t the exposure time in minutes, the environmental
concentration C of hydrochloric acid in µg·m-3 is obtained according to the equation:
1.028 m 1,000,000
C=
QK t
where QK is the sampling rate at temperature K (in Kelvin) and 1.028 is the ratio between molecular masses of HCl
and Cl-(see Analysis).
Exposure
Hydrochloric acid is sampled linearly in the range from 20,000 ÷ 20,000,000 µg·m-3·min.
Workplace environment
In workplace environment we recommend exposure time from 15 minutes to 8 hours: the ceiling values can be mea-
sured.
Outdoor environment
We recommend exposure time from 2 hours to 2 days. Exposure time as long as 7 days is allowed if average rela-
tive humidity does not exceed 50%, taking into account the water absorbing properties of silica gel.
We also recommend to protect radiello from rain by the mountable shelter code RAD196.
Interferences
Gaseous chlorine is adsorbed by silica gel and is revealed as 0.02 ng of chloride ion for 1 µg·m-3·min of chlorine.
Storage
Kept in a clean environment free from gaseous hydrochloric acid, the cartridges code 169 are stable for at least 24
months before and after sampling.
If more than six months have passed since you received the cartridges, before environmental sampling campaigns,
it is advisable to analyse some cartridges to check for contamination from the background. Discard the cartridges if
they contain more than 5 µg of chloride ion.
Analysis
Add 2 ml of deionised water to the cartridge in its tube (make sure that no trace of chloride ion is found in the water
you use). Recap the tube and stir vigorously by a VORTEX stirrer for 1-2 minutes. Analyse the solution by ion chro-
matography. Subtract the blank value obtained from two unexposed cartridges.
Prepare the calibration solutions with sodium or potassium chloride concentrations ranging from 0.5 to 25 mg/litre
as Cl-.
Principle
The cartridge code RAD166 is made of microporous polyethylene coated with triethanolamine (TEA). Gaseous
hydrofluoric acid is adsorbed by TEA and subsequently extracted with water to be quantified by ion chromatography
or by ion selective electrode as fluoride ion.
Sampling is selective for the gaseous molecules: any airborne fluoride salt will not cross the diffusive membrane of
radiello.
Sampling rate
Sampling rate at 25 °C and 1013 hPa is 187 cm3·min-1.
Calculations
Let m be the mass of fluoride ion in µg found onto the cartridge and t the exposure time in minutes, the environmental
concentration C of HF in µg·m-3 is obtained according to the equation:
1.053 m 1,000,000
C=
187 t
where 1.053 is the ratio between molecular masses of HF and F-(see Analysis).
Exposure
Hydrofluoric acid is sampled linearly in the range from 10,000 to 50,000,000 µg·m-3·min.
Workplace environment
In workplace environments we recommend exposure time from 15 minutes to 8 hours: the ceiling values can be
measured.
Outdoor environment
We recommend exposure time from 2 hours to 14 days.
Protect radiello from rain by the mountable shelter code RAD196.
Storage
Kept in a dark place at 4 °C, the cartridges stay unaltered for at least 12 months before exposure and 4 months after
sampling. Expiry date is printed on the plastic bag wrapping each cartridge.
If more than six months have passed since you received the cartridges, before environmental sampling campaigns,
it is advisable to analyse some cartridges to measure any contamination from the background. Discard the cartrid-
ges if they contain more than 2 µg of fluoride ion.
Keep at least two unexposed cartridges for each lot and analyse them as blanks.
Analysis
Ion chromatography
Add 5 ml of the eluent solution to the radiello tube. Stir vigorously by a VORTEX stirrer for 1-2 minutes. Let the tube
stand for 10 minutes, then stir manually and inject the solution in the ion chromatographic apparatus without further
treatment.
Analyse 1-2 unexposed cartridges and subtract the average blank value to the samples.
Principle
Code RAD132 cartridge is made of stainless steel net loaded with a mixture of molecular sieve and activated char-
coal 35-50 mesh.
Nitrous oxide and halogenated anaesthetic gases permeate the silicone membrane and are sampled by the molecu-
lar sieve and by activated charcoal respectively.
The sampled compounds are displaced by a water-methanol mixture and are quantified by capillary gas chromato-
graphy and a headspace sampler.
N2O, isoflurane, ethrane and halothane are detected by the Electron Capture Detector (ECD) with very good sensiti-
vity; sevorane can not be quantified by ECD detection and has to be analyzed by mass spectrometry.
Sampling rate
Sampling rate values at 25 °C (298 K) and 1013 hPa are listed in the table on the right.
sampling rate
Effect of temperature, humidity and wind speed (ml·min-1)
Sampling rate varies from the values at 298 K on the effect of tem-
perature (in Kelvin) as expressed by the following equation: N 2O 1.01
isoflurane 2.25
QK = Q298 ( 298
K
) 1.5
ethrane 3.39
where QK is the sampling rate at temperature K and Q298 is the sam-
pling rate value at reference temperature of 298 K. This yields a ± 5% halothane 4.93
variation of Q for a 10 °C variation (upwards or downwards) from 25
°C. sevorane 0.92
Sampling rate is invariant with humidity in the range 10 ÷ 90% for expo-
sure time not exceeding 8 hours and with wind speed between 0.1 and
10 m·s-1.
Calculations
Concentration in air is obtained by the following equation:
C = m· 1,000
QK·t
where:
C = concentration in mg·m-3
m = mass of analyte found on the cartridge in µg
QK = sampling rate in ml·min-1
t = exposure time in minutes
Exposure
Sampling rate is constant for exposure time up to 8 hours at relative humidity up to 80% with N2O concentration up
to 500 ppm and overall halogenated anaesthetic compounds concentration up to 100 ppm.
Exposure time longer than 8 hours in presence of relative humidity higher than 80% leads to the loss of the nitrous
oxide already sampled by the effect of competing water vapour adsorption on the molecular sieve sites.
Storage IMPORTANT
The sampling kit code RAD125 is sterilized by g-rays. Use of DO NOT STERILIZE THE SAMPLER BY
the sampler makes it no longer sterile. With the exception of the AUTOCLAVING. Autoclaving treatment perma-
adsorbing cartridge, the sampler is indefinitely re-usable. After nently damages the silicone permeative mem-
the first sampling, if you can arrange for sterilization by your- brane.
selves you only need to re-order code RAD132 cartridges to
perform other sampling campaigns. Adsorbing cartridges need not to be sterile.
If kept in a dry place free from chemical contamination, the cartridges are stable for at least 12 months.
After the sampling, the cartridges are stable for 30 days if stored with the same precautions.
Analysis
Materials needed for the analysis
• 20 ml headspace glass vials with open-top aluminum crimp caps and rubber/PTFE septa
• water/methanol mixture 60/40 v/v
• usual laboratory glassware
water impurity
time was 4 hours at the concentration values indicated and with methanol
relative humidity of 70%.
water
MS detection
The instrumental conditions are as described above, with the exception of the carrier gas (helium has to be used
instead) and the make-up gas, which is not employed. Acquire by SIM (Single Ion Monitoring) focussing the detec-
tor on the following signals (the base peak is underlined):
N2O: 44; isoflurane and ethrane: 51, 67, 117; halothane: 117, 198, 179; sevorane: 33, 131, 181
If high concentrations of CO2 interfere (it gives a strong signal at m/z 44), N2O can be quantified basing on the signal
at m/z 30. On page L4 a typical GC-MS chromatogram (as total ion current) is displayed. It can be observed that,
as an effect of the vacuum applied on the detector end of the column, retention times are shorter with respect to
those obtained with ECD detection.
Calibration
Calibration curves for N2O and halogenated anaesthetics can be prepared simultaneously.
Draw pure N2O in a gas sampling bulb. Transfer 20 ml of pure N2O in the 1 litre bottle through the septum by a
gastight syringe. Switch on the magnetic stirrer and let the mixture equilibrate for 30 minutes.
Standard solutions of the halogenated compounds must be prepared in water/methanol 60/40 v/v in order to con-
MSD Analysis
ethrane
isoflurane
air
N 2O
halotane
sevorane
time (min)
tain from 0.05 to 3.0 mg/l of each compound; five calibration levels are recommended.
For each level pipet 10 ml of calibration solution in an empty vial, add a blank code 132 cartridge and cap imme-
diately.
Add also a precisely measured volume of diluted N2O drawn from the bottle by a gastight syringe (usually added
volume ranges from 50 to 1,000 µl), stir and let equilibrate at 45 °C for 1 hour.
The values above generally comprise the usual conditions of operating theatres. The analyst may choose different
values if needed, but equivalent exposure values should not exceed 400,000 mg·m-3·min for nitrous oxide and 50,000
mg·m-3·min for each of the halogenated compounds.
Pay attention: the ECD and/or MSD response may not be linear. If this should be the case, use a second order cali-
bration curve.
Useful data
name chemical formula molecular weight 1 mg·m-3 at 25°C = ppm
nitrous oxide N 2O 44 0.556
forane CHF2-O-CHCl-CF3 184.5 0.133
ethrane CHF2-O-CF2-CHClF 184.5 0.133
halothane CF3-CHBrCl 197.4 0.124
sevorane CH2F-O-CH(CF3)2 200 0.123
Principle
Code RAD147 cartridge is a stainless steel net cylinder with 100 mesh opening and 4.8 mm diameter, packed with
250 ± 10 mg of Tenax-TA, particle size 20-35 mesh. Phenols are trapped by adsorption and recovered by thermal
desorption, analysis is performed by capillary gas chromatography and MS detection.
The method has been optimized for the following compounds:
OH OH OH OH OH
CH3 CH3
CH3 CH3
phenol o-cresol m-cresol 2,3-dimethylphenol
CH3
p-cresol
OH OH OH OH
CH3 H3C CH3
3,4-dimethylphenol
Calculations
The listed sampling rate values take already into account the recovery yields of adsorbed compounds. The avera-
ge concentration over the sampling period is therefore calculated from sampled mass of analyte and exposure time
without introducing any other corrective factor, apart from temperature variations of Q.
Average concentration C in µg·m-3 over the whole exposure time is calculated according to the following expression:
m [µg]
C [µg·m-3] = 1.000.000
QK [ml·min-1] · t [min]
where:
m = mass of analyte in µg
t = exposure time in minutes
Exposure
Workplace environment
Exposure time can range from 2 to 8 hours.
Storage
The duration of Tenax adsorbent capacity is virtually unlimited. If kept in a cool place not contaminated by phenols,
white and adsorbent capacity remain unchanged for at least twenty-four months. The expiry date and the lot num-
ber are printed on the transparent plastic casing, whose integrity acts as a guarantee seal.
After exposure the cartridges, well capped and kept in a cool and solvent-free place, maintain their content unalte-
red for at least three months.
Analysis
The analytical method hereafter described have been set up by the Perkin-Elmer Turbomatrix thermal desorber and
Agilent 5973 MSD mass spectrometer detector. They may be implemented on other instruments by introducing
minor adjustements.
Desorption
The thermal desorber is equipped with 1/4” s.s. sample tubes, they have to be hollow and free: discard the stain-
less steel gauze disk which is fitted to the groove and discard also the springs if present.
Code 147 cartridge has been dimensioned to fit the diameter of Turbomatrix thermal desorption tubes. Its length is
such that, when the cartridge is introduced into the tube and is stopped by the groove, it is positioned exactly cen-
trally with respect to the tube length. The same considerations apply to the Markes Unity thermal desorber.
Once capped, the Turbomatrix steel tube has to be positioned in the carousel with the grooves on the bottom.
The desorption conditions described below have been developed to obtain the best results from cartridges expo-
sed for seven days to the usual concentrations of urban and indoor pollution. Shorter exposure times or much higher
concentrations than usual may make it necessary to readjust the splits.
Temperatures and timing
• Desorption: 280°C for 10 minutes
• Cryofocusing trap (Tenax TA): during primary desorption maintain at 2 °C, secondary desorption at 99 °C·sec-1
up to 290 °C, maintain at 290 °C for 1 minute
• Six port valve: 150 °C
• Transfer line: 200 °C
Flows
• Carrier gas: helium, 24 psi
• Desorption flow: 100 ml·min-1
• Flow from tube to cryofocusing trap: 20 ml·min-1
• Outlet split: 25 ml·min-1
Instrumental analysis
Column
J&W HP-5MS Ultra Inert or equivalent, length 60 m, internal diameter 0.25 mm, film thickness 0.25 µm; the column
is directly fitted to the six-port valve of Turbomatrix apparatus
Temperatures
• GC oven: 40 °C for 5 minutes, 5 °C·min-1 up to 115°C, 10 °C·min-1 up to 165 °C, 30 °C·min-1 up to 285 °C, final
isotherm 3 minutes
• GC-MS interface: 260 °C
•
Flows
• helium carrier gas: 1.6 ml·min-1
•
Calibration
Calibration curves are obtained
by gas-phase injection of metha-
nol solutions of the target com-
pounds onto blank cartridges.
Injections are performed through
a GC injector, where a short piece
of wide-bore (0.53 mm i.d.) deac-
tivated uncoated column is instal-
led. The other end bears a
Swagelock reducing connection
(1/16”-1/4”).
The 1/4” Swagelock nut has to be
equipped with a PTFE ferrule
instead of the original steel one
(use PTFE ferrules that come along with the Turbomatrix caps).
Introduce a blank code RAD147 cartridge in a Turbomatrix tube and fit the tube to the Swagelock nut. Keep the
injector at 170 °C but do not heat the oven. Inject slowly 1 µl of each calibration solution under nitrogen flow (40 ml·min-1)
and let the system purge for 2 minutes. Analyze the cartridge as you would do with a sample. We suggest you to
prepare a complete set of calibration solutions by subsequent dilutions such as they contain, for example, 4, 2, 1,
0.05, 0.025 and 0.01 µg·µl-1 of each compound.
Principle
Cartridge code RAD141 is a 4.8 mm diameter stainless steel mesh tube with a mesh size of 3x8 µm, filled with
approximately 480 mg of graphite carbon (Carbopack X) 40/60 mesh.
1,3-butadiene and isoprene are trapped by adsorption, recovered by thermal desorption and analysed by capillary
gas chromatography with MS detector.
Sampling rates
Sampling rate values were measured experimentally at 20°C (273 K) and 1013 hPa in a dynamic controlled atmo-
sphere chamber.
The sampling rate for 1,3-butadiene in the workplace is 30.5 ± 0.3 ml·min-1 (nominal value at a concentration bet-
ween 114 and 226 µg·m-3 for 8-hour exposures). For the longer term (7 days) sampling the value is 4.7 ml·min-1
[Strandberg et al. (1), (2)].
For isoprene the sampling rate is 41.2 ± 4.9 ml·min-1 (in the range 2 ÷ 6,680 µg·m-3 for exposures of 30 to 480 min).
Calculations
The average concentration C over the exposure time interval is calculated from the mass of the analyte found on
the cartridge (corrected for the white, if any) and from the exposure time, using the absorption values above, as fol-
lows:
m [µg]
C [µg·m-3] = 1,000,000
QK [ml·min-1] · t [min]
where:
m = mass of analyte in µg
t = exposure time in minutes
Limit of detection
The blank response, the limit of detection (LOD) and the limit of quantification (LOQ) depend on the instrumenta-
tion and the analytical conditions.
Under the analytical conditions specified below, the blank value is not detectable, i.e. it is less than 0.5 ng for both
compounds.
The LOQ for 8-hour workplace exposure is 0.1 µg·m-3. For a 7-day exposure to ambient air, the LOQ is 0.03 µg·m-3;
see also Strandberg et al. (2).
Measurements uncertainty
The following table shows the values of uncertainty in 1,3-butadiene measurements in the workplace, evaluated with
two different approaches. Uncertainties were first determined under laboratory conditions, following the methods of
the ISO GUM (Guide to Expression of Uncertainty in Measurement, International Organization for Standardization)
and ISO 5725
(Accuracy (trueness Uncertainty of measurement for an 8-hour sampling of 1,3-butadiene in working environment
and precision) of Relative combined 200 µg·m-3 442 µg·m 2210 µg·m-3 4420 µg·m-3
-3
Storage
After the exposure, the samples, well capped in their glass tubes, have to be stored in the freezer, because 1,3-
butadiene and isoprene are reactive compounds. Laboratory tests according to EN 838 showed for both compounds
a loss of analyte of 7-8% after 14-day storage. The samples shall therefore be analysed within 14 days from the end
of exposure, in order to ensure the maximum loss of analyte remain within 10%.
Analysis
The method proposed here was developed with the Perkin-Elmer Turbomatrix thermal desorber coupled to the
Agilent 6890 gas chromatograph and Agilent 5973 mass spectrometer. It can of course be transferred to other instru-
ments.
Desorption
The 1/4 "pipe supplied with the Turbomatrix must be empty and free: remove the stainless steel disk placed inside
it in correspondence with the circular incision and, if present, also the springs.
The code 141 cartridge has been sized so that its outer diameter coincides with the inner diameter of the
Turbomatrix tube. Moreover, its length is such that, when the cartridge is introduced into the tube and is stopped by
the groove, it is positioned exactly centrally with respect to the tube length. The same considerations apply to the
Markes Unity thermal desorber.
Once capped, the Turbomatrix steel tube has to be positioned in the carousel with the grooves on the bottom.
The described conditions have been optimized for seven days exposures to typical concentrations of urban atmo-
spheres and indoor environments. Shorter exposure times or considerably higher concentrations would require dif-
ferent settings of split flows.
Instrumental analysis
Column
J&W GS-GASPRO, length 60 m, i.d. 0.32 mm; the column is directly fitted to the six-port valve of Turbomatrix appa-
ratus.
Temperatures
• GC oven: 80 °C for 1 minute, 25 °C·min-1 up to 175 °C, mantain for 8 minutes, 25 °C·min-1 up to 250 °C, final
isotherm 11.2 minutes
• Interface GC-MS: 290 °C
• Ionic source: 230 °C, quadrupole 150 °C
Flows
• Carrier gas: helium at 1.8 ml·min-1
Calibration
The calibration curve is performed by injecting known aliquots of a gaseous mix-
ture certified as 1,3-butadiene in nitrogen onto virgin cartridges. The operation is
carried out with the injector of a gas chromatograph whose output is grafted with
a short piece (10 cm) of a deactivated capillary column (0.25 mm id) connected
to a Swagelock reducer 1/16 "-1/4" .
Instead of the 1/4 ”steel ferrule of the reducer, one of those in PTFE used for clo-
sing the Turbomatrix tubes is used
Introduced a virgin cartridge in the Turbomatrix tube and inserted the tube in the Swagelock reducer, keeping the
injector at 50 °C and the oven cold, injecting different volumes of the gas mixture under a flow of nitrogen of 30
ml·min-1, leaving flow the gas for 2 minutes.
It is recommended to use a mixture of 1,000 ppm of 1,3-butadiene in nitrogen and to inject aliquots of 20, 40, 60,
80 and 100 µl of mixture (with a 100 µl gas-tight syringe) or 100 aliquots , 200, 300, 400 and 500 µl of mixture (with
a 500 µl gas-tight syringe) according to the desired calibration range.
Cartridges conditioning
The cartridges can be reconditioned using the thermal desorber in "tube conditioning" mode, heating them to 350
°C for at least 20 minutes in an inert gas flow (helium or nitrogen at a flow of 50 ÷ 100 ml·min-1).
Index by code