Determination of Mtbe, Etbe, Tame, Dipe, Tertiary-Amyl Alcohol and C Toc Alcohols in Gasoline by Gas Chromatography
Determination of Mtbe, Etbe, Tame, Dipe, Tertiary-Amyl Alcohol and C Toc Alcohols in Gasoline by Gas Chromatography
for the
     Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1. Scope                                                                                     cally excluded from this test method. The methanol content of
   1.1 This test method covers the determination of ethers and                               M-85 fuel is considered beyond the operating range of the
alcohols in gasolines by gas chromatography. Specific com-                                   system.
pounds determined are methyl tert-butylether (MTBE), ethyl                                      1.5 Benzene, while detected, cannot be quantified using this
tert-butylether (ETBE), tert-amylmethylether (TAME), diiso-                                  test method and must be analyzed by alternate methodology
propylether (DIPE), methanol, ethanol, isopropanol,                                          (see Test Method D3606).
n-propanol, isobutanol, tert-butanol, sec-butanol, n-butanol,                                   1.6 The values stated in SI units are to be regarded as
and tert-pentanol (tert-amylalcohol).                                                        standard. Alternate units, in common usage, are also provided
  1.2 Individual ethers are determined from 0.20 % to 20.0 %                                 to increase clarity and aid the users of this test method.
by mass. Individual alcohols are determined from 0.20 % to                                      1.7 This standard does not purport to address all of the
12.0 % by mass. Equations used to convert to mass % oxygen                                   safety concerns, if any, associated with its use. It is the
and to volume % of individual compounds are provided. At                                     responsibility of the user of this standard to establish appro-
concentrations <0.20 % by mass, it is possible that hydrocar-                                priate safety, health, and environmental practices and deter-
bons may interfere with several ethers and alcohols. The                                     mine the applicability of regulatory limitations prior to use.
reporting limit of 0.20 % by mass was tested for gasolines                                      1.8 This international standard was developed in accor-
containing a maximum of 10 % by volume olefins. It may be                                    dance with internationally recognized principles on standard-
possible that for gasolines containing >10 % by volume                                       ization established in the Decision on Principles for the
olefins, the interference may be >0.20 % by mass. Annex A1                                   Development of International Standards, Guides and Recom-
gives a chromatogram showing the interference observed with                                  mendations issued by the World Trade Organization Technical
a gasoline containing 10 % by volume olefins.                                                Barriers to Trade (TBT) Committee.
   1.3 This test method includes a relative bias correlation for
ethanol in spark-ignition engine fuels for the U.S. EPA                                      2. Referenced Documents
regulations reporting based on Practice D6708 accuracy assess-                                  2.1 ASTM Standards:2
ment between Test Method D4815 and Test Method D5599 as                                         D1298 Test Method for Density, Relative Density, or API
a possible Test Method D4815 alternative to Test Method                                           Gravity of Crude Petroleum and Liquid Petroleum Prod-
D5599. The Practice D6708 derived correlation equation is                                         ucts by Hydrometer Method
only applicable for ethanol in fuels in the concentration range                                 D1744 Test Method for Water in Liquid Petroleum Products
from 2.28 % to 14.42 % by mass as measured by Test Method                                         by Karl Fischer Reagent3
D4815. The applicable Test Method D5599 range for ethanol is                                    D3606 Test Method for Determination of Benzene and
from 2.16 % to 14.39 % by mass as reported by Test Method                                         Toluene in Spark Ignition Fuels by Gas Chromatography
D5599.                                                                                          D4052 Test Method for Density, Relative Density, and API
  1.4 Alcohol-based fuels, such as M-85 and E-85, MTBE                                            Gravity of Liquids by Digital Density Meter
product, ethanol product, and denatured alcohol, are specifi-                                   D4057 Practice for Manual Sampling of Petroleum and
                                                                                                  Petroleum Products
   1
     This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
                                                                                                2
Subcommittee D02.04.0L on Gas Chromatography Methods.                                             For referenced ASTM standards, visit the ASTM website, www.astm.org, or
   Current edition approved Dec. 1, 2019. Published December 2019. Originally                contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1989. Last previous edition approved in 2015 as D4815 – 15b. DOI:                Standards volume information, refer to the standard’s Document Summary page on
10.1520/D4815-15BR19.                                                                        the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
                                                                                         1
                                                                  D4815 − 15b (2019)
  D4307 Practice for Preparation of Liquid Blends for Use as                           TABLE 1 Pertinent Physical Constants and Retention
   Analytical Standards                                                               Characteristics for TCEP/WCOT Column Set Conditions
                                                                                                            as in Table 2
  D4420 Test Method for Determination of Aromatics in
                                                                                                                    Relative Retention              Relative
   Finished Gasoline by Gas Chromatography (Withdrawn                                                                      Time
                                                                                                       Retention                         Molecular Density at
   2004)3                                                                            Component
                                                                                                       Time, Min.   (MTBE =   (DME =      Mass       15.56/
  D5599 Test Method for Determination of Oxygenates in                                                                1.00)    1.00)               15.5 6 °C
   Gasoline by Gas Chromatography and Oxygen Selective                           Water                   2.90        0.58       0.43       18.0      1.000
   Flame Ionization Detection                                                    Methanol                3.15        0.63       0.46       32.0      0.7963
                                                                                 Ethanol                 3.48        0.69       0.51       46.1      0.7939
  D6708 Practice for Statistical Assessment and Improvement                      Isopropanol             3.83        0.76       0.56       60.1      0.7899
   of Expected Agreement Between Two Test Methods that                           tert-Butanol            4.15        0.82       0.61       74.1      0.7922
   Purport to Measure the Same Property of a Material                            n-Propanol              4.56        0.90       0.67       60.1      0.8080
                                                                                 MTBE                    5.04        1.00       0.74      88.2       0.7460
                                                                                 sec-Butanol             5.36        1.06       0.79       74.1      0.8114
3. Terminology                                                                   DIPE                    5.76        1.14       0.85      102.2      0.7282
   3.1 Definitions of Terms Specific to This Standard:                           Isobutanol              6.00        1.19       0.88       74.1      0.8058
                                                                                 ETBE                    6.20        1.23       0.91      102.2      0.7452
   3.1.1 low volume connector—a special union for connecting                     tert-Pentanol           6.43        1.28       0.95       88.1      0.8170
two lengths of tubing 1.6 mm inside diameter and smaller.                        1,2-Dimethoxyethane     6.80        1.35       1.00      90.1       0.8720
Sometimes this is referred to as zero dead volume union.                           (DME)
                                                                                 n-Butanol               7.04        1.40       1.04       74.1      0.8137
   3.1.2 oxygenate—any oxygen-containing organic compound                        TAME                    8.17        1.62       1.20      102.2      0.7758
that can be used as a fuel or fuel supplement, for example,
various alcohols and ethers.
   3.1.3 split ratio—in capillary gas chromatography, the ratio                    4.3 After benzene and TAME elute from the nonpolar
of the total flow of carrier gas to the sample inlet versus the                  column, the column switching valve is switched back to its
flow of the carrier gas to the capillary column, expressed by                    original position to backflush the heavy hydrocarbons.
                         split ratio 5 ~ S1C ! /C                     (1)           4.4 The eluted components are detected by a flame ioniza-
where:                                                                           tion or thermal conductivity detector. The detector response,
S = flow rate at the splitter vent, and                                          proportional to the component concentration, is recorded; the
C = flow rate at the column outlet.                                              peak areas are measured; and the concentration of each
                                                                                 component is calculated with reference to the internal standard.
  3.1.4 tert-amyl alcohol—tert-pentanol.
   3.2 Acronyms:                                                                 5. Significance and Use
   3.2.1 DIPE—diisopropylether.                                                    5.1 Ethers, alcohols, and other oxygenates can be added to
   3.2.2 ETBE—ethyl tert-butylether.                                             gasoline to increase octane number and to reduce emissions.
   3.2.3 MTBE—methyl tert-butylether.                                            Type and concentration of various oxygenates are specified and
                                                                                 regulated to ensure acceptable commercial gasoline quality.
   3.2.4 TAME—tert-amyl methylether.
                                                                                 Drivability, vapor pressure, phase separation, exhaust, and
   3.2.5 TCEP—1,2,3-tris-2-cyanoethoxypropane—a gas                              evaporative emissions are some of the concerns associated with
chromatographic liquid phase.                                                    oxygenated fuels.
   3.2.6 WCOT—a type of capillary gas chromatographic col-                         5.2 This test method is applicable to both quality control in
umn prepared by coating the inside of the capillary with a thin                  the production of gasoline and for the determination of
film of stationary phase.                                                        deliberate or extraneous oxygenate additions or contamination.
4. Summary of Test Method                                                        6. Apparatus
   4.1 An appropriate internal standard, such as 1,2-                               6.1 Chromatograph—While any gas chromatographic
dimethoxyethane (ethylene glycol dimethyl ether), is added to                    system, which is capable of adequately resolving the individual
the sample, which is then introduced into a gas chromatograph                    ethers and alcohols that are presented in Table 1, can be used
equipped with two columns and a column switching valve. The                      for these analyses, a gas chromatographic instrument, which
sample first passes onto a polar TCEP column, which elutes                       can be operated at the conditions given in Table 2 and has a
lighter hydrocarbons to vent and retains the oxygenated and                      column switching and backflushing system equivalent to Fig.
heavier hydrocarbons.                                                            1, has been found acceptable. Carrier gas flow controllers shall
  4.2 After methylcyclopentane, but before DIPE and MTBE                         be capable of precise control where the required flow rates are
elute from the polar column, the valve is switched to backflush                  low (see Table 2). Pressure control devices and gages shall be
the oxygenates onto a WCOT nonpolar column. The alcohols                         capable of precise control for the typical pressures required.
and ethers elute from the nonpolar column in boiling point                          6.1.1 Detector—A thermal conductivity detector or flame
order, before elution of any major hydrocarbon constituents.                     ionization detector can be used. The system shall have suffi-
                                                                                 cient sensitivity and stability to obtain a recorder deflection of
  3
    The last approved version of this historical standard is referenced on       at least 2 mm at a signal-to-noise ratio of at least 5 to 1 for
www.astm.org.                                                                    0.005 % by volume concentration of an oxygenate.
                                                                             2
                                                                       D4815 − 15b (2019)
          TABLE 2 Chromatographic Operation Conditions                                 or flame ionization detection are used. Split injection is
        Temperatures          Flows, mL/min            Carrier Gas: Helium             necessary to maintain the actual chromatographed sample size
Column Oven              60   to injector   75   Sample size, µLA        1.0–3.0       within the limits of column and detector optimum efficiency
Injector, °C            200   Column         5   Split ratio                15:1       and linearity.
Detector—TCD, °C        200   Auxiliary      3   Backflush, min          0.2–0.3
          —FID, °C      250   Makeup        18   Valve reset time       8–10 min          6.1.4.1 Some gas chromatographs are equipped with on-
Valve °C                 60                      Total Analysis time   18–20 min       column injectors and autosamplers, which can inject small
A
  Sample size must be adjusted so that alcohols in the range of 0.1 % to 12.0 %        samples sizes. Such injection systems can be used provided
by mass and ethers in the range of 0.1 % to 20.0 % by mass are eluted from the         that sample size is within the limit of the column and detectors
column and measured linearly at the detector. A sample size of 1.0 µL has been
introduced in most cases.                                                              optimum efficiency and linearity.
                                                                                          6.1.4.2 Microlitre syringes, automatic syringe injectors, and
                                                                                       liquid sampling valves have been used successfully for intro-
                                                                                       ducing representative samples into the gas chromatographic
   6.1.2 Switching and Backflushing Valve—A valve, to be                               inlet.
located within the gas chromatographic column oven, capable                              6.2 Data Presentation or Calculation, or Both:
of performing the functions described in Section 11 and
                                                                                         6.2.1 Recorder—A recording potentiometer or equivalent
illustrated in Fig. 1. The valve shall be of low volume design
                                                                                       with a full-scale deflection of 5 mV or less can be used to
and not contribute significantly to chromatographic deteriora-
                                                                                       monitor detector signal. Full-scale response time should be 1 s
tion.
                                                                                       or less with sufficient sensitivity and stability to meet the
   6.1.2.1 Valco Model No. A 4C10WP, 1.6 mm (1⁄16 in.) fit-
                                                                                       requirements of 6.1.1.
tings. This particular valve was used in the majority of the
                                                                                         6.2.2 Integrator or Computer—Means shall be provided for
analyses used for the development of Section 15.
                                                                                       determining the detector response. Peak heights or areas can be
   6.1.2.2 Valco Model No. C10W, 0.8 mm (1⁄32 in.) fittings.
                                                                                       measured by computer, electronic integration, or manual tech-
This valve is recommended for use with columns of 0.32 mm
                                                                                       niques.
inside diameter and smaller.
   6.1.2.3 Some gas chromatographs are equipped with an                                  6.3 Columns, Two as Follows:
auxiliary oven, which can be used to contain the valve and                               6.3.1 Polar Column—This column performs a preseparation
polar column. In such a configuration, the nonpolar column is                          of the oxygenates from volatile hydrocarbons in the same
located in the main oven and the temperature can be adjusted                           boiling point range. The oxygenates and remaining hydrocar-
for optimum oxygenates resolution.                                                     bons are backflushed onto the nonpolar column in 6.3.2. Any
   6.1.3 An automatic valve switching device must be used to                           column with equivalent or better chromatographic efficiency
ensure repeatable switching times. Such a device should be                             and selectivity to that described in 6.3.1.1 can be used. The
synchronized with injection and data collection times.                                 column shall perform at the same temperature as required for
   6.1.4 Injection System—The chromatograph should be                                  the column in 6.3.2, except if located in a separate auxiliary
equipped with a splitting-type inlet device if capillary columns                       oven as in 6.1.2.3.
                                                                                   3
                                                            D4815 − 15b (2019)
   6.3.1.1 TCEP Micro-Packed Column, 560 mm (22 in.) by                   nents of the same boiling point range in a gasoline sample. The
1.6 mm (1⁄16 in.) outside diameter by 0.76 mm (0.030 in.)                 following procedure has been used successfully.
inside diameter stainless steel tube packed with 0.14 g to 0.15 g            8.1.2 Completely dissolve 10 g of TCEP in 100 mL of
of 20 % (mass/mass) TCEP on 80/100 mesh Chromosorb                        methylene chloride. Next add 40 g of 80/100 mesh Chromo-
P(AW). This column was used in the cooperative study to                   sorb P(AW) to the TCEP solution. Quickly transfer this mixture
provide the precision and bias data referred to in Section 15.            to a drying dish, in a fume hood, without scraping any of the
   6.3.2 Nonpolar (Analytical) Column—Any column with                     residual packing from the sides of the container. Constantly,
equivalent or better chromatographic efficiency and selectivity           but gently, stir the packing until all of the solvent has
to that described in 6.3.2.1 and illustrated in Fig. 2 can be used.       evaporated. This column packing can be used immediately to
   6.3.2.1 WCOT Methyl Silicone Column, 30 m (1181 in.)                   prepare the TCEP column.
long by 0.53 mm (0.021 in.) inside diameter fused silica
WCOT column with a 2.6 µm film thickness of cross-linked                  9. Sampling
methyl siloxane. This column was used in the cooperative                    9.1 Every effort should be made to ensure that the sample is
study to provide the precision and bias data referred to in               representative of the fuel source from which it is taken. Follow
Section 15.                                                               the recommendations of Practice D4057, or its equivalent,
                                                                          when obtaining samples from bulk storage or pipelines.
7. Reagents and Materials
                                                                            9.2 Upon receipt in the laboratory, chill the sample in its
  7.1 Carrier Gas—Carrier gas appropriate to the type of
                                                                          original container to 0 °C to 5 °C (32 °F to 40 °F) before any
detector used. Helium has been used successfully. The mini-
                                                                          subsampling is performed.
mum purity of the carrier gas used must be 99.95 mol %.
                                                                            9.3 If necessary, transfer the chilled sample to a vapor tight
  7.2 Standards for Calibration and Identification—Standards
                                                                          container and store at 0 °C to 5 °C (32 °F to 40 °F) until needed
of all components to be analyzed and the internal standard are
                                                                          for analysis.
required for establishing identification by retention time as well
as calibration for quantitative measurements. These materials
                                                                          10. Preparation of Micro-Packed TCEP Column
shall be of known purity and free of the other components to be
analyzed. (Warning—These materials are flammable and can                    10.1 Wash a straight 560 mm length of 1.6 mm outside
be harmful or fatal if ingested or inhaled.)                              diameter (0.76 mm inside diameter) stainless steel tubing with
                                                                          methanol and dry with compressed nitrogen.
   7.3 Methylene Chloride, used for column preparation, re-
agent grade, free of nonvolatile residue. (Warning—Harmful                  10.2 Insert six to twelve strands of silvered wire, a small
if inhaled. High concentrations may cause unconsciousness or              mesh screen, or stainless steel frit inside one end of the tube.
death.)                                                                   Slowly add 0.14 g to 0.15 g of packing material to the column
                                                                          and gently vibrate to settle the packing inside the column.
8. Preparation of Column Packings                                         When strands of wire are used to retain the packing material
   8.1 TCEP Column Packing:                                               inside the column, leave 6.0 mm (0.25 in.) of space at the top
   8.1.1 Any satisfactory method used in the practice of the art          of the column.
that will produce a column capable of retaining the C1 to C4                10.3 Column Conditioning—Both the TCEP and WCOT
alcohols and MTBE, ETBE, DIPE, and TAME from compo-                       columns are to be briefly conditioned before use. Connect the
                                                                      4
                                                           D4815 − 15b (2019)
columns to the valve (see 11.1) in the chromatographic oven.             baseline and the system is ready for another analysis. The
Adjust the carrier gas flows as in 11.3 and place the valve in the       chromatogram should appear similar to the one illustrated in
RESET position. After several minutes, increase the column               Fig. 2.
oven temperature to 120 °C and maintain these conditions for                11.5.1.1 Ensure that the BACKFLUSH time is sufficient to
5 min to 10 min. Cool the columns below 60 °C before                     quantitatively transfer the higher concentrations of the ethers,
shutting off the carrier flow.                                           specifically MTBE, into the nonpolar column.
                                                                            11.5.2 It is necessary to optimize the valve BACKFLUSH
11. Preparation of Apparatus and Establishment of                        time by analyzing a standard blend containing oxygenates. The
Conditions                                                               correct BACKFLUSH time is determined experimentally by
                                                                         using valve switching times between 0.20 min and 0.35 min.
   11.1 Assembly—Connect the WCOT column to the valve
                                                                         When the valve is switched too soon, C5 and lighter hydro-
system using low volume connectors and narrow bore tubing.
                                                                         carbons are backflushed and are co-eluted in the C4 alcohol
It is important to minimize the volume of the chromatographic
                                                                         section of the chromatogram. When the valve BACKFLUSH is
system that comes in contact with the sample; otherwise, peak
                                                                         switched too late, part or all of the ether component (MTBE,
broadening will occur.
                                                                         ETBE, or TAME) is vented, resulting in an incorrect ether
   11.2 Adjust the operating conditions to those listed in Table         measurement.
2, but do not turn on the detector circuits. Check the system for           11.5.2.1 DIPE may require a BACKFLUSH time slightly
leaks before proceeding further.                                         shorter than the other ethers. The system may require reopti-
   11.2.1 If different polar and nonpolar columns or capillary           mization if the analysis of DIPE is required.
columns of smaller ID, or both, are used it can be necessary to             11.5.3 To facilitate setting BACKFLUSH time, the column
use different optimum flows and temperatures.                            vent in Fig. 1 can be connected to a second detector (TCD or
   11.3 Flow Rate Adjustment:                                            FID), as described in Test Method D4420, and used to set
                                                                         BACKFLUSH TIME based on the oxygenates standard con-
   11.3.1 Attach a flow measuring device to the column vent
                                                                         taining the ethers of interest.
with the valve in the RESET position and adjust the pressure to
the injection port to give 5.0 mL ⁄min flow (14 psig). Soap
                                                                         12. Calibration and Standardization
bubble flow meters are suitable.
   11.3.2 Attach a flow measuring device to the split injector               12.1 Identification—Determine the retention time of each
vent and adjust the flow from the split vent using the A flow            component either by injecting small amounts separately, in
controller to give a flow of 70 mL ⁄min. Recheck the column              known mixtures, or by comparing the relative retention times
vent flow set in 11.3.1 and adjust if necessary.                         with those in Table 1.
   11.3.3 Switch the valve to the BACKFLUSH position and                     12.1.1 To ensure minimum interference from hydrocarbons,
adjust the variable restrictor to give the same column vent flow         it is strongly recommended that a fuel devoid of oxygenates be
set in 11.3.1. This is necessary to minimize flow changes when           chromatographed to determine the level of any hydrocarbon
the valve is switched.                                                   interference.
   11.3.4 Switch the valve to the inject position RESET and                 12.2 Preparation of Calibration Samples—Prepare multi-
adjust the B flow controller to give a flow of 3.0 mL ⁄min to            component calibration standards of the oxygenates and con-
3.2 mL ⁄min at the detector exit. When required for the particu-         centration ranges of interest, by mass, in accordance with
lar instrumentation used, add makeup flow or TCD switching               Practice D4307.
flow to give a total of 21 mL ⁄min at the detector exit.                    12.2.1 For each oxygenate, prepare a minimum of five
                                                                         calibration standards spanning the range of the oxygenate in
   11.4 When a thermal conductivity detector is used, turn on
                                                                         the samples. As an example, for full range calibration, 0.1 %,
the filament current and allow the detector to equilibrate. When
                                                                         0.5 %, 2 %, 5 %, 10 %, 15 %, and 20 % by mass of each
a flame ionization detector is used, set the hydrogen and air
                                                                         oxygenate may be used.
flows and ignite the flame.
                                                                            12.2.2 Before preparing the standards, determine the purity
   11.5 Determine the Time to Backflush—The time to back-                of the oxygenate stocks and make corrections for the impurities
flush will vary slightly for each column system and must be              found. Whenever possible, use stocks of at least 99.9 % purity.
determined experimentally as follows. The start time of the              Correct the purity of the components for water content,
integrator and valve timer must be synchronized with the                 determined by Test Method D1744.
injection to accurately reproduce the backflush time.                       12.2.3 To minimize evaporation of light components, chill
   11.5.1 Initially assume a valve BACKFLUSH time of                     all chemicals and gasoline used to prepare standards.
0.23 min. With the valve RESET, inject 1 µL to 3 µL of a blend              12.2.4 Prepare standards by transferring a fixed volume of
containing at least 0.5 % or greater oxygenates (see 7.3), and           oxygenates, using pipettes or eye droppers (for volumes below
simultaneously begin timing the analysis. At 0.23 min, rotate            1 % by volume), to 100 mL volumetric flasks or septum capped
the valve to the BACKFLUSH position and leave it there until             vials as follows. Cap and record the tare weight of the
the complete elution of TAME is realized. Record this time as            volumetric flask or vial to 0.1 mg. Remove the cap and
the RESET time, which is the time at which the valve is                  carefully add the oxygenate to the flask or vial. Do not
returned to the RESET position. When all of the remaining                contaminate with sample the part within the flask or vial that
hydrocarbons are backflushed, the signal will return to a stable         contacts the cap. Cap and record the net mass (Wi) to 0.1 mg of
                                                                     5
                                                                                        D4815 − 15b (2019)
        TABLE 3 Example Calculation of Correlation Coefficient
   Xi          Yi       x = Xi − x̄        y = Yi − ȳ         xy            x2           y2
    1.0         0.5        −2.0               −1.0             2.0         4.0       1.0
    2.0         1.0        −1.0               −0.5             0.5         1.0       0.25
    3.0         1.5        0.0                0.0              0.0         0.0       0.0
    4.0         2.0        +1.0               0.5              0.5         1.0       0.25
    5.0         2.5        +2.0               1.0              2.0         4.0       1.0
 x̄ = 3.0    ȳ = 1.5                                        (^xy)      ^x2 = 10.0 ^y2 = 2.5
                                                            2 = 25.0
                                                               s o xyd 2           25.0
                                                     r2 5                     5              5 1.0
                                                            s o x 2 ds o y 2 d s 10.0ds 2.5d
                                                                                                     6
                                                               D4815 − 15b (2019)
sample, cap, and record the net mass (Wg) to the nearest 0.1 mg              where:
of the sample added. Mix thoroughly and inject into the gas                  wi = mass % of each oxygenate, as determined using Eq 19,
chromatograph. If using an automatic sampler, then transfer an               Vi = volume % of each oxygenate to be determined,
aliquot of the solution into a glass gas chromatographic (GC)                Di = relative density at 15.56 °C (60 °F) of the individual
vial. Seal the GC vial with a TFE-fluorocarbon-lined septum. If                     oxygenate, as found in Table 1, and
the sample is not immediately analyzed, store below 5 °C                     Df = relative density of the fuel under study, as determined
(40 °F).                                                                            by Test Method D1298 or D4052.
   13.2 Chromatographic Analysis—Introduce a representative                    14.4 Report the volume % of each oxygenate to the nearest
aliquot of the sample, containing internal standard, into the gas            0.01 % by volume.
chromatograph, using the same technique and sample size as
                                                                                14.5 Mass % Oxygen—To determine the oxygen content of
used for the calibration analysis. An injection volume of 1.0 µL
                                                                             the fuel, convert and sum the oxygen contents of all oxygen-
to 3.0 µL with a 15:1 split ratio has been used successfully.
                                                                             ated components determined above in accordance with the
Start recording and integrating devices in synchronization
                                                                             following equation:
with sample introduction. Obtain a chromatogram or inte-
grated peak report, or both, which displays the retention times                                                   w i 3 16.0 3 N i
and integrated area of each detected component.
                                                                                                    W tot 5   (          Mi
                                                                                                                                            (21)
                                                                         7
                                                                       D4815 − 15b (2019)
                                       TABLE 4 Precision Interval as Determined from Cooperative Study Data
                                                                                  Repeatability
Component                                                                                                                                                      Total
             MEOH        EtOH        iPA        tBA        nPA       MTBE        sBA         DIPE        iBA           ETBE    tAA        nBA       TAME
                                                                                                                                                              Oxygen
  Wt. %
   0.20       0.04       0.02       0.02       0.02       0.01       0.02        0.01     0.03           0.03          0.01    0.02       0.02       0.02
   0.50       0.06       0.04       0.03       0.03       0.02       0.03        0.02     0.05           0.05          0.03    0.03       0.04       0.03
   1.00       0.09       0.06       0.04       0.04       0.03       0.05        0.03     0.08           0.08          0.05    0.04       0.06       0.05      0.02
   2.00       0.14       0.09       0.06       0.06       0.05       0.07        0.05     0.12           0.12          0.09    0.06       0.09       0.08      0.05
   3.00       0.17       0.12       0.07       0.07       0.06       0.09        0.06     0.15           0.15          0.12    0.08       0.12       0.11      0.08
   4.00       0.20       0.14       0.09       0.09       0.07       0.11        0.07     0.17           0.17          0.16    0.09       0.14       0.13      0.12
   5.00       0.23       0.16       0.10       0.10       0.08       0.12        0.08     0.20           0.20          0.19    0.11       0.16       0.15      0.15
   6.00       0.26       0.18       0.11       0.11       0.08       0.14        0.09     0.22           0.22          0.22    0.12       0.18       0.17
  10.00       0.35       0.24       0.15       0.15       0.11       0.18        0.12     0.29           0.29          0.33    0.16       0.24       0.25
  12.00       0.39       0.27       0.16       0.16       0.12       0.20        0.14     0.32           0.32          0.38    0.18       0.27       0.29
  14.00                                                              0.22                 0.35                         0.44                          0.32
  16.00                                                              0.24                 0.38                         0.49                          0.35
  20.00                                                              0.27                 0.43                         0.58                          0.41
                                                                                 Reproducibility
Component                                                                                                                                                      Total
             MEOH        EtOH        iPA        tBA        nPA       MTBE        sBA         DIPE        iBA           ETBE    tAA        nBA       TAME
                                                                                                                                                              Oxygen
  Wt. %
   0.20       0.14       0.09       0.14       0.07       0.04       0.04        0.15         0.14       0.14          0.11    0.06       0.09       0.14
   0.50       0.24       0.16       0.26       0.12       0.07       0.08        0.28         0.26       0.26          0.21    0.10       0.15       0.22
   1.00       0.37       0.23       0.42       0.19       0.11       0.12        0.44         0.42       0.42          0.46    0.15       0.22       0.31      0.09
   2.00       0.57       0.34       0.67       0.30       0.16       0.19        0.70         0.67       0.67          0.61    0.22       0.33       0.44      0.22
   3.00       0.72       0.43       0.80       0.40       0.21       0.25        0.92         0.88       0.88          0.83    0.28       0.41       0.54      0.36
   4.00       0.86       0.51       1.06       0.48       0.24       0.30        1.11         1.06       1.06          1.03    0.33       0.49       0.63      0.52
   5.00       0.99       0.58       1.23       0.56       0.28       0.35        1.29         1.23       1.23          1.22    0.38       0.55       0.70      0.70
   6.00       1.10       0.64       1.40       0.63       0.31       0.40        1.46         1.40       1.40          1.41    0.42       0.61       0.77
  10.00       1.51       0.86       1.97       0.89       0.41       0.56        2.06         1.97       1.97          2.07    0.56       0.82       1.00
  12.00       1.68       0.95       2.22       1.00       0.45       0.63        2.33         2.22       2.22          2.38    0.62       0.91       1.10
  14.00                                                              0.70                     2.46                     2.68                          1.19
  16.00                                                              0.77                     2.69                     2.96                          1.28
  20.00                                                              0.89                     3.13                     3.51                          1.43
  14.6 Report the total mass % of oxygen in the fuel to the                               where X is the mean mass % of the component.
nearest 0.01 % by mass.                                                                   15.1.2 Reproducibility—The difference between two single
                                                                                        and independent results obtained by different operators work-
15. Precision and Bias4
                                                                                        ing in different laboratories on identical material would, in the
   15.1 Precision—The precision of this test method as deter-                           long run, exceed the following values in Table 4 only in one
mined by a statistical examination of interlaboratory test results                      case in twenty.
is as follows:                                                                                         Reproducibility Estimates in Oxygenates in Gasolines
   15.1.1 Repeatability—The difference between successive                               Component                              Reproducibility
results obtained by the same operator with the same apparatus                           Methanol (MeOH)                        0.37 (X0.61)
                                                                                        Ethanol (EtOH)                         0.23 (X0.57)
under constant operating conditions on identical test materials                         Isopropanol (iPA)                      0.42 (X0.67)
would, in the long run, in the normal and the correct operation                         tert-Butanol (tBA)                     0.19 (X0.67)
of the test method, exceed the following values in Table 4 only                         n-Propanol (nPA)                       0.11 (X0.57)
                                                                                        MTBE                                   0.12 (X0.67)
in one case in twenty.                                                                  sec-Butanol (sBA)                      0.44 (X0.67)
                Repeatability Estimates for Oxygenates in Gasoline                      DIPE                                   0.42 (X0.67)
Component                              Repeatability                                    Isobutanol (iBA)                       0.42 (X0.67)
Methanol (MeOH)                        0.09 (X0.59)                                     ETBE                                   0.36 (X0.76)
Ethanol (EtOH)                         0.06 (X0.61)                                     tert-Pentanol (tAA)                    0.15 (X0.57)
Isopropanol (iPA)                      0.04 (X0.56)                                     n-Butanol (nBA)                        0.22 (X0.57)
tert-Butanol (tBA)                     0.04 (X0.56)                                     TAME                                   0.31 (X0.51)
n-Propanol (nPA)                       0.003 (X0.57)                                    Total Oxygen                           0.09 (X1.27)
MTBE                                   0.05 (X0.56)
sec-Butanol (sBA)                      0.003 (X0.61)                                       where: X is the mean mass % of the component.
DIPE                                   0.08 (X0.56)
Isobutanol (iBA)                       0.08 (X0.56)                                        15.2 Bias—The National Institute of Standards and Tech-
ETBE                                   0.05 (X0.82)                                     nology (NIST) provides selected alcohols in reference fuels. As
tert-Pentanol (tAA)                    0.04 (X0.61)                                     an example, the following standard reference materials (SRM)
n-Butanol (nBA)                        0.06 (X0.61)
TAME                                   0.05 (X0.70)                                     in reference fuels are available from NIST (www.nist.gov).
Total Oxygen                           0.02 (X1.26)                                                                                Nominal Concentration, Mass % of
                                                                                         SRM                    Type
                                                                                                                                  MeOH     EtOH     MeOH + tBuOH
                                                                                         1829   Alcohols in Reference Fuel        0.335    11.39       10.33 + 6.63
   4
    Supporting data have been filed at ASTM International Headquarters and may           1837   Methanol and tert-butanol                              10.33 + 6.63
be obtained by requesting Research Report RR:D02-1296. Contact ASTM Customer             1838   Ethanol                                    11.39
Service at service@astm.org.                                                             1839   Methanol                          0.335
                                                                                  8
                                                                      D4815 − 15b (2019)
   15.3 Relative Bias—A relative bias assessment of Test                              observed for some samples after applying the bias-correction
Method D4815 versus Test Method D5599 for the determina-                              for the material types and considered random.
tion of ethanol in spark-ignition engine fuel was conducted                             15.3.2 Correlation Equation:
using data from the ASTM D02 Interlaboratory Crosscheck                                         Predicted Test Method D5599=
Program. The assessment was performed in accordance with
                                                                                                   bias-corrected Test Method D4815 =C D481510.03          (23)
the requirements of Practice D6708 with a successful outcome.
It was based on measurements of 82 spark-ignition engine fuels                        where:
supplied to the ASTM Proficiency Test Program by participat-                          CD4815 = Test Method D4815 reported mass percent of
ing laboratories between February 2007 and October 2014 and                                    ethanol.
is documented in Research Report RR:D02-1819.5
                                                                                         15.3.2.1 The correlation equation is only applicable for
   NOTE 3—In the United States, the EPA requires the measurement of                   fuels in the concentration range of ethanol from 2.28 % to
ethanol and other oxygenates in spark ignition engine fuels by Test                   14.42 % by mass as reported by Test Method D4815.
Method D5599. Effective Jan. 1, 2016, there is an allowance in the
regulation to use other test methods if they are formally correlated with the            15.3.2.2 The correlation equation is applicable for fuels that
specified test method by a consensus organization, for example, ASTM.                 when determined by Test Method D5599 are in the concentra-
This relative bias statement is intended to satisfy the requirement and               tion range of range of 2.16 % to 14.39 % by mass.
allow use of Test Method D4815 bias-corrected results in the stated
concentration ranges in place of Test Method D5599 for ethanol content.                   NOTE 4—The Test Method D5599 concentration range used to develop
                                                                                      the Practice D6708 assessment may not cover the entire scope indicated in
   15.3.1 The degree of agreement between results from Test                           the scope of Test Method D5599 for blended ethanol.
Method D4815 and Test Method D5599 can be further im-                                     NOTE 5—The correlation equation was developed from a variety of fuel
proved by applying a correlation equation (Eq 23) (Research                           samples from the ASTM Interlaboratory Crosscheck programs; however,
Report RR:D02-1819),5 and this correlation equation shall be                          it is recommended that the correlation equation be verified for samples of
                                                                                      interest to ensure applicability.
utilized when reporting compliance with EPA fuels program.
Sample-specific bias, as defined in Practice D6708, was                               16. Keywords
                                                                                        16.1 alcohols; DIPE (disopropylether) ; ETBE (ethyl tert-
   5
     Supporting data have been filed at ASTM International Headquarters and may
                                                                                      butylether); ethers; gas chromatography; gasoline; MTBE
be obtained by requesting Research Report RR:D02-1819. Contact ASTM Customer          (methyl tert-butylether); oxygenates; TAME (tert-
Service at service@astm.org.                                                          amylmethylether)
ANNEX
(Mandatory Information)
  A1.1 Fig. A1.1 shows the interference from hydrocarbons                             olefin gasoline, and the resulting chromatogram was compared
for a gasoline containing 10 % by volume olefins. Each alcohol                        with that obtained with no ethers or alcohols added.
and ether was added at 0.1 % by mass into the 10 % by volume
                                                                                  9
                                                                    D4815 − 15b (2019)
APPENDIX
(Nonmandatory Information)
   NOTE X1.1—This appendix contains instrument conditions and results               is much easier to obtain at lower cost. This appendix contains instrument
obtained using nitrogen as an alternative carrier gas. At this time, because        conditions and results obtained using nitrogen carrier gas for the analysis
the test method precision and bias performance information using the                of ethanol and MtBE in pump gasoline samples.
alternative carrier gases and conditions listed in this appendix have not
been studied in accordance with the proper ASTM ILS process, this                      X1.1 This section lists the GC operating conditions for
appendix is included only for information purposes. Results obtained
under the conditions described in this appendix are not considered to be            D4815 utilizing nitrogen carrier gas. These conditions are same
valid D4815 results, and shall not be represented as such.                          as those for helium carrier gas described in Table 2. The back
   NOTE X1.2—Helium is a widely used carrier gas for most capillary gas             flush time and valve reset time shown in Table X1.1 were
chromatographic applications. Recent disruptions in helium supplies                 experimentally determined for the GC system used in this
combined with higher prices have prompted the search for alternative
carrier gases where helium’s chromatographic properties are not critical to         study. These values will vary from system to system depending
the method’s performance. Nitrogen is a suitable helium replacement and             on differences found in the columns.
                                                                               10
                                            D4815 − 15b (2019)
                         TABLE X1.1 GC Conditions Utilizing Nitrogen Carrier Gas
                     Carrier gas                            Nitrogen (99.9995 %)
                     Inlet                                       Split/splitless
                     Inlet temperature                              200 °C
                     Inlet flow                                   75 mL/min
                     Split ration                                     15:1
                     Column flow                                  5 mL/min
                     Auxiliary flow                               3 mL/min
                     Valve temperature                               60 °C
                     Oven temperature                                60 °C
                     Detector                                  Flame ionization
                     Detector temperature                           250 °C
                     Back flush time                               0.24 min
                     Valve reset time                               14 min
                     Total analysis time                            16 min
FIG. X1.1 Analysis of Ethanol in Gasoline Utilizing Nitrogen Carrier Gas Conditions Described in Table X1.1
FIG. X1.2 Analysis of MtBE in Gasoline Utilizing the Nitrogen Carrier Gas Conditions Described in Table X1.1
                                                    11
                                                         D4815 − 15b (2019)
                          TABLE X1.2 Analysis of Four Gasoline Samples using D4815 and
                                              Nitrogen Carrier Gas
                            NOTE 1—Observed repeatability (r) was determined from replicate
                          analysis of each sample.
                           Compound          Mass %                Observed r                   D4815r
                          Ethanol             0.99                   0.01                        0.06
                          Ethanol             6.63                   0.03                        0.19
                          MtBE                2.10                   0.01                        0.08
                          MtBE                11.29                  0.05                        0.19
    ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
    This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
  This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
12