DAT 400 Manual 1.4
DAT 400 Manual 1.4
Starting from the s/n 2007/12629 the DAT-400 units are equipped with a USB port.
                                        1
                                     Technical Specifications
Power Requirements
Environmental
Display
Performance
Serial Output RS-232, RS-422 or RS-485. USB port available from s/n 2007/12629
Maximum Cable Length 50 feet RS-232, 3200 feet for RS-422 and RS-485
Enclosure
                                                               2
                                             Installation
Install the instrument in a location where it will       Heavy electrical equipment should not be
not be subjected to excessive heat, humidity or          installed close to the weighing equipment.
vibration. For best results, avoid direct sunlight
on the front of the instrument. The unit should          Excessive vibration will affect the accuracy of
be installed at eye level so as to allow viewing         the weigh system and depending on the severity
the display and access to the front panel keys.          can cause damage to electrical and electronic
                                                         components.
Cable Types and sizes
                                                         The atmosphere should be dust free and not
Use a 6 x 0.5 mm2 shielded cable for the load            contain any corrosive gasses or materials which
cell/s connection. Use a 3 x 0.34 mm2 shielded           could adversely affect the equipment.
cable for the RS-232 connection, and a 2 x 0.34
mm2 shielded cable for the RS-485 connection.            Hazardous areas
                                                     3
Introduction                                            Operating Mode (cont’d)
This manual provides general information on             The “O” key is used to zero the unit in Gross
the installation, configuration, calibration, and       mode, tare the unit in Net mode, and reset the
operation of the Precise400 Series Digital /          value to zero in Peak Hold mode.
Analog Transmitters. There are two models
                                                        The “PRG” key is used to send the data to the
within this series, the Model 400 which features        RS-232 serial port for use with a computer.
a four-button keyboard, 6-digit Red LED display,
and four status LED’s. The Model 450 is a blind
unit which can only be configured and calibrated        Front panel key functions - Setup Mode
remotely via the serial port. Both models are
packaged in ABS Plastic DIN-Rail mounted                In setup mode three of the front panel keys are
enclosures.                                             used as directional keys to scroll through the
                                                        various menus. A label on the lower half of each
All units include the following features:               key identifies the direction provided by the key.
   Drives up to six 350 ohm load cells
   Supports 4 and 6 wire connections                  The “SET” and “FUN” keys are used to navigate
   RS-232 or RS-422/RS-485 outputs                    through the main menu and sub-menus. They
   USB port (from s/n 2007/12629)                     are also used to increment or decrement the
   (2) Logic inputs                                   numerical value of a selected digit.
   (2) Logic outputs (setpoints)
                                                        The “O” key is used to move through certain
   Isolated analog output 4/20mA, 0-5 Vdc
                                                        sub-menu parameters, or to select a specific
     or 0-10Vdc output utilizing a 16 bit DAC
                                                        digit when numerical values are displayed. It is
                                                        also used to return to the main menu from any
The Precise 400 Series Transmitters have               of the sub-menus.
three different modes of operation:
                                                        The “PRG” key switches the unit to the “Basic
Operating mode                                          Configuration” mode. It is also used to enter the
In this mode the unit displays gross weight, net        sub-menus, and store parameter changes.
weight, or peak force readings as required.
                                                        The “PRG” key when used in conjunction with
Setup mode                                              the “SET” key switches the unit to the “Complete
This mode is used for configuration and                 Configuration” mode. It is also used to enter the
calibration of the unit.                                sub-menus, and store parameter changes.
The test mode enables you to test the inputs,           There are four status LED’s on the front of the
outputs, and analog output.                             unit, one for Motion, one for Net mode, and one
                                                        for each of the outputs.
Front panel key functions - Operating Mode
(Model 400)
                                                        Enabling/Disabling front panel keys.
The “SET” key enables you to enter values for
Setpoints 1 and 2. Press the “SET” key, Set 1           In order to prevent any access to the instrument
appears on the display. Press the “PRG” key to          by non-authorized personnel, a procedure to
view the current value. To change the current           lock the front panel keys is also available.
value, press the “O” key to select a digit, then        Each key can be locked individually.
use the “UP” or “DOWN” keys to increment or             The 1/0 selection establish the lock/unlock
decrement the value.                                    condition for each single key.
Repeat this procedure for the remaining digits,         See page 6 for more details.
then press the “PRG” key to enter the new
value. To change the value for Setpoint 2, press        Enabling the “blind” function
the “SET” key again and repeat the above                This function allows to disable the visualization
procedure or press the “O” key to exit.                 of the display. A dash running counterclockwise
The “FUN” key is used to switch the unit from           on the display’s perimeter is displayed instead of
Gross mode to Net mode, or Peak Hold mode.              the weight value.
To switch to Peak Hold mode, press and hold             See page 6 for more details.
the “FUN” key until a ‘P’ appears on the display.
                                                    4
        Main Power Connections                                                                             Logic Input Connections (cont’d)
        Use terminal 8 for the (+) input, and terminal 9                                                   Input (2) functions as a “send data” command,
        for the (-) input. The supply voltage must be                                                      closing the input causes the weight data to be
        24 Vdc + 15%. The external power supply must                                                       sent to the RS-232 serial port.
        have a minimum rating of 5 watts.
                                                                                                           Logic Output Connections
        Transducer/s Connections                                                                           Use terminal 1 for Output (1) and terminal 2 for
        Attach cable from transducer/s to transmitter                                                      Output (2). Terminal 3 is the common for both
        (terminals 13 through 18). If using a 4-conductor                                                  outputs. The outputs utilize solid-state relays
        cable, (no remote sensing) jumpers must be                                                         with a maximum load rating of 24 Vdc @ 100mA
        installed between the + Excitation and + Sense
        leads (terminals 14 to 15) and between                                                             Analog Output Connections
        the - Excitation and - Sense leads (terminals 13                                                   Use terminal 10 for the 4/20mA output (300 Ohm
        to 16) as shown below in Figure 1.                                                                 load max), or terminal 11 for a 0 to 5 Vdc or 0 to
                                                                                                           10 Vdc output (10K Ohm load max). Terminal
        Logic Input Connections                                                                            12 is used as the common for either output.
        Use terminal 4 for Input (1), and terminal 5 for
        Input (2). Terminal 6 is the common for both                                                       Serial Output Connections
        inputs. These are powered inputs, which require                                                    The transmitters serial port supports RS-232 or
        an input voltage of 24 Vdc to activate them. See                                                   RS-422/RS-485 or USB communications. Serial
        Figure 1 for additional information.                                                               parameter selections can be found on page 19.
        Input (1) is used to zero the unit in Gross mode,                                                  A detailed description of the serial data formats
        tare the unit in Net mode, and reset the peak                                                      is provided on pages 21 through 25.
        value when operating in Peak Hold mode.                                                            Wiring connections are shown on pages 26
                                                                                                           through 27.
                                       Common
Output 2
Output 1
                                                                                                           Figure 1
                                                                         1
                                                                                                                DAT
                                                                                                                400
     (24 Vdc)
                                                            Input 1
                                                                                                                                1
                                                                                                                                2
                                                                                                                                3
                                                                         4
                                                            Input 2
                                                                         5
SET
                                                      Common
                                                                         6
                                                             Not used
                                                                         7
                    SUPPLY
                    POWER
                                                            +24 Vdc
                                                                         8
                     24 Vdc
                                                              0 Vdc
                                                                         9
FUN
                                                                                                                                27 26 25 24 23 22 21 20 19
                              Analog
                              Output
RS232
                                                            Common                                                                                              Gnd
                                                                - Exc
                                                                                                          O
                                                                                                                                                             Not used
                                                              + Exc                                                                                              Gnd
                                                            + Sense                                                                                             + Tx
                                                                                                                                                                                           (1000m max.)
                                                                                                                                                                                            32 unit max.
                                                                                                                                                                                            RS422/485
                                                             - Sense                                                                                            - Tx
                                                                                                          PRG
                                                                 - Sig                                                                                          + Rx
                                                               + Sig                                                                                            - Rx
                    LOAD
                    CELL
                                                                                                      5
When no programming procedures are in progress, the display shows the weight value.
In some particular conditions the following MESSAGES are displayed:
                                         The weight displayed by the instrument is more than 9 divisions over the
                                         “Live” weight value (see “NET” parameter, page 13)
                                         The load cells input signal is lower than - 0,5 (-3,9) mV/V.
               O-L                       The load cells input signal is higher than +3,5 (+3,9) mV/V.
                                         The load cell input signal is missing
The values in brackets refer to the instruments having load cells input signal range from - 3,9 to + 3,9 mV/V (optional)
Locking the front panel keys and enabling the “blind” function
In order to prevent any access to the instrument by non-authorized personnel, a procedure to lock the front
panel keys is available. Each key can be locked individually.
The 1/0 selection establish the lock/unlock condition for each single key.
The “blind” function allows to disable the visualization of the display. A dash running counterclockwise on the
display’s perimeter is displayed instead of the weight value once the programmed time (01 to 99 sec.) expires.
When the instrument is in “blind” mode all its functions are active.
Press and hold the “PRG” and “O” keys. The “O” key must be pressed after the “PRG” key.
Release both keys when the message LOCK appears on the display.
                                 (
( PRG ) +           (       O         (press and hold)                                        Front panel keys
O PRG
                        (
          (     O                    LOCK             ( PRG )
                                                                                                                           Perform the selection
                                                                                                0000                       0/1 of the blinking digit
                                                                                     ( PRG )
                                                                                                                             Move the blinking
                                                                                                                    O
                                                                                                                             digit to the right
                        (
          (     O                    BLIND            ( PRG )
                                                                      ( PRG )                      OFF
                                                                                                                        Default = OFF
                                                                      ( PRG )                        ON
END
                                                        ( PRG )
                                                                                            Perform the selection
                                                                                            0/1 of the blinking digit
                                                                        6
Configuration Methods                                  Basic Configuration (cont’d)
The Precise 400 Series Digital / Analog               To set one of the following parameters, CAPAC,
Transmitters can be configured by using the            SENSIT, NET, DEAD L, DSPDIV, or ANALOG,
front panel keys (Model 400) to navigate through       press the “PRG” key to view the default or
a series of menus, or by sending configuration         previously selected value for that parameter.
and calibration data to the RS-232 port with           To change the value, press the “O” key to select
INOVATIONan MS Windows based program                 a digit, then use the “UP” or “DOWN” keys to
included with the unit. INOVATIONsimplifies          increment or decrement the value of the
the configuration and calibration procedure.           selected digit. When the desired value appears
                                                       on the display, press the “PRG” key to store the
Configuration Procedures                               new value. Press either the “UP” or “DOWN”
                                                       keys to move to the next parameter.
There are two procedures available for setup of
the Precise400 Series Transmitters, “Basic           The SIGNAL parameter enables you to view the
Configuration” or “Complete Configuration”.            actual input signal from the transducer/s. Press
                                                       the “O” key to move to the next parameter.
The “Basic Configuration” procedure allows you
to change selective parameters, whereas, the           The CALIBR parameter is used to calibrate the
“Complete Configuration” procedure allows you          transmitter. The unit can be calibrated using
to change all of the parameters.                       either the data sheet or dead load procedure.
For example, if the transmitter requires               The data sheet calibration procedure allows the
calibration only, the “Basic Configuration”            transmitter to be calibrated without a transducer
procedure is recommended.                              simulator or test weights. This procedure is
                                                       based on the capacity and certified, full scale
The following pages provide a detailed                 mV/V output (sensitivity) of the transducer/s.
explanation of both configuration procedures.
                                                       To perform a data sheet calibration, enter the
Basic Configuration via the Front Panel                total scale capacity, sensitivity, net weight, dead
                                                       load weight, and display resolution, then press
To perform a “Basic Configuration”, press and          the “O” key. The display will indicate STORE ?,
hold the “PRG” key until the display shows             press the “PRG” key to save the changes and
CAPAC. Use either the “UP” or “DOWN” keys to           return to the operating mode.
navigate through the eight basic configuration
sub-menus shown below.                                 See page 9 for a detailed explanation of the
                                                       calibration procedure using test weights.
 Display            Sub-menu Functions                 The ANALOG parameter is used to enter the
                                                       full scale weight value that you want the analog
 CAPAC      Total capacity of transducer/s             output to represent. The default is net weight
                                                       value.
 SENSIT     Rated output of the transducer/s           To exit from the setup procedure, press the “O”
                                                       key when any one of the following parameters
   NET      Net (live) weight value                    are displayed CAPAC, SENSIT, … ANALOG.
                                                       The display will indicate STORE ?, press the
 DEAD L     Dead load weight value                     “PRG” key to save the changes and return to
                                                       the operating mode.
 DSPDIV     Increment size (resolution)
                                                   7
Basic Configuration Menu
                                                               Enter
             (
    (   O            CAPAC           ( PRG )                total scale                       ( PRG )
                                                             capacity
                                                               Enter
             (
    (   O            SENSIT          ( PRG )               transducer/s                       ( PRG )
                                                             sensitivity
                                                             Enter
             (
    (   O             NET            ( PRG )                “LIVE”                            ( PRG )
                                                          weight value
                                                             Enter
             (
    (   O            DEAD L          ( PRG )              “Dead Load”                         ( PRG )
                                                          weight value
                                                          Select display
             (
    (   O            DSPDIV          ( PRG )                divisions                         ( PRG )
                                                           0.001 - 50
                                                            View input
             (                                                                                          (
    (   O            SIGNAL          ( PRG )               signal value                   (     O
                                                              mV/V
                                                            See calib.
             (
                                                                                  *
                                                                                      (
    (   O            CALIBR          ( PRG )              procedure on        (
                                                            next page
                                                            Enter FS
             (
    (   O           ANALOG           ( PRG )                value for                         ( PRG )
                                                           analog out
Place the unit in “Basic Configuration” mode by       then use the “UP” or “DOWN” keys to increment
pressing and holding the “PRG” key until the          or decrement the value of the selected digit.
display shows CAPAC. Use either the “UP” or           Press the “PRG” key to enter the new value,
“DOWN” keys to scroll through the eight basic         then press either the “UP” or “DOWN” keys to
configuration menus shown above. To set one           move to the next parameter. To exit the setup
of the following parameters, CAPAC, SENSIT,           procedure, press the “O” key when any one of
NET, DEAD L, DSPDIV, or ANALOG, press the             the sub-menu parameters are displayed. The
“PRG” key to view the default or previously           display will indicate STORE ?, press the “PRG”
selected value for that parameter. To change          key to save the changes and return to the
the value, press the “O” key to select a digit,       operating mode.
                                                  8
Basic Configuration Menu (cont’d)
NOTE: In order to ensure greater stability during the calibration procedure, the digital filter increases
to the maximum value automatically (value 9), therefore the weight indication updates very slowly at
this stage. Once the calibration procedure has been completed, the digital filter will return to the pre-
vious value automatically.
Cont’d from
                                *
                                    (
                            (           the previous
                                            page
                                                                      (            Sets ZERO
                        ( PRG )            CAL             (   O
                                                                                   calibration
                                                                     Enter the
                                        WT-VALUE                   actual weight             ( PRG )
                                                                   value (SPAN)
                                           CAL
                                                                          NOTE:
                                                                          Input signal from the transducer/s
                                         ( PRG )                          must be stable when entering the
              Back to                                                     zero and span values.
            “CALIBR”
            (Main menu)
The Precise400 Series Transmitters can be            4. Enter the actual weight value using the “O” key
calibrated via the front panel keys (Model 400),          to select each digit, then use the “UP” or
or with the INOVATIONconfiguration utility.             “DOWN” keys to increment or decrement the
                                                          value of the selected digit.
Both methods consist of the following steps.           5. Press the “PRG” key to enter the setting (CAL
                                                          will be blinking on the display).
   Zero calibration
   Apply a known load (test weight)                  6. Press the “PRG” key, (the display indicates
   Edit weight value                                    CALIBR), press either the “UP” or “DOWN”
   Span calibration                                     keys to move to the next parameter.
   Optional linearization (See page 15)              7. If no additional parameter changes are
                                                          required, press the “O” key (the display
2-point calibration via the front panel keys              indicates STORE ?), press the “PRG” key to
                                                          save the changes and return to the operating
To calibrate the transmitter with the front panel         mode.
keys, proceed as follows:
                                                       Should it be necessary to erase either of the
1. With the unit in “Basic Configuration” mode         previous calibration entries, proceed as follows:
   and the message “CALIBR” displayed, press
                                                       When the message CAL is blinking on the
   the “PRG” key (CAL will be blinking on the
                                                       display, to erase the zero calibration, press and
   display).
                                                       hold the “FUN” key, and then press the “O” key.
                                                       Release both keys, the previous zero calibration
2. With the system unloaded, press the “O” key
                                                       entry has been erased. To erase the span
   to perform the ZERO calibration.
                                                       calibration, press and hold the “FUN” key, and
3. Apply a known load not less than 10% of the         then press the “SET” key. Release both keys, the
   net weight value and press the “UP” key.            previous span calibration entry has been erased.
                                                       End the procedure as described above (points 6
                                                       and 7).
                                                       9
Complete Configuration via the Front Panel               Configuration using INOVATION
(Model 400)
                                                         To configure the Precise400 Series Digital /
In order to perform a complete configuration, the        Analog Transmitters with the INOVATION
unit must be placed in setup mode. To enter the          program, proceed as follows:
setup mode, press and hold the “PRG” and SET”            1. Install INOVATIONon an IBM-compatible
keys on the front of the unit. The “SET” key                computer running Windows 95, 98, Me or XP.
must be pressed after the “PRG” key. Release
                                                         2. Minimum system requirements are 8MB of
both keys when the word CONFIG is shown on
                                                            extended memory and at least 5MB of
the display. The CONFIG menu is the first of six
                                                            available hard drive space.
main menus used to completely configure the
instrument. Use either the “UP” or “DOWN”                3. Remove power from both units, and connect
keys to navigate through the six main menus                 the PC’s serial port to the RS-232 terminals
shown below.                                                on the transmitter. Refer to page 27 for wiring.
                                                            Use a std. USB cable in case of using the
                                                            USB port. Refer to page 50 for USB drivers
                                                            installation on the PC.
                                                         4. Apply power to the PC and the transmitter,
 Display          Main Menu Functions                       then start the INOVATIONprogram.
CONFIG Parameters for total scale capacity,              5. Switch the transmitter into remote mode by
       transducer sensitivity, net and dead                 removing the upper left cover and moving
       load weight values, display divisions.               jumper (J1) to the remote control position as
                                                            shown below. The Model 400 can also be
CALIBR Zero and full scale calibration.                     switched into remote mode by going into
                                                            complete configuration and then selecting
                                                            ‘Remote’ from the serial menu. See page 19.
PARAM      Parameters for digital filtering,
           motion, auto zero and zero tracking                      Pins 1 - 2 = Remote Control
                                                                    Pins 2 - 3 = Default Setting
 IN-OUT    Operating mode selection for
           outputs, and input and output test.
                                                                                   3
                                                                                   2
SERIAL     Serial output configuration.                                            1
                                                    10
Configuration using INOVATION (cont’d)                      If the operator answers “Yes”, the “Dead load”
                                                             value inside the first file folder will change
On this screen you can switch the display                    automatically according to the new zero.
among net, gross and peak reading, set the                   If the operator answers “No”, the “Dead load”
serial address, enter values for the two                     value inside the first file folder will remain set to
setpoints, check the two logic inputs status and             the previous value.
the
                                                             The third file folder contains the parameter
software release of the instrument.                          selections for digital filtering, zero tracking, etc.
Click on the SETUP button, seven file folders will
                                                             The fourth file folder contains the parameter
appear on the lower half of the screen. The first            selections for configuring the inputs and outputs
file folder labeled “Parameters” will be open.
                                                             The fifth file folder contains the parameter
The top section in this folder is for entering the           selections for the serial output.
total scale capacity, sensitivity, net weight, and
dead load weight values (required parameters                 The sixth folder contains the parameter
for performing a data sheet calibration)                     selections for the analog output,
The data sheet calibration procedure allows the              The seventh folder provides a means for testing
transmitter to be calibrated without a transducer            the outputs and analog output
simulator or test weights. This procedure is                 The Main Setup Screen and file folders 1
based on the capacity and certified, full scale              through 6 all have SEND buttons on them for
mV/V output (sensitivity) of the transducer/s.
                                                             downloading the various parameter changes.
To perform a data sheet calibration, enter the
                                                             When configuration and calibration has been
above values, then click on the SEND button for              completed, the data can be saved on a PC or
that group of parameters.
                                                             floppy disk or downloaded (sent) to the
The next section in this file folder is used to              transmitter. This method of configuring and
select the display resolution. It consists of a              calibrating is very beneficial when several
drop down menu and a SEND button.                            transmitters with similar configurations are being
                                                             set up, or when a transmitter has to be replaced.
The Precise units allow to count by 1’s up to
60,000 counts, Yet, for values greater than
                                                             To exit INOVATION, click on the Exit Setup
10,000, the units automatically select a lower
                                                             icon. Now click on Option in the upper left corner
resolution to ensure greater stability. After
                                                             of the screen, from the drop down menu select
selecting the desired resolution, press the SEND
                                                             EXIT, a “Save Changes” window will appear,
button for this parameter. Do not press any of
                                                             click on OK. This will return you to the
the other SEND buttons in this folder or the unit
                                                             Configuration Utility Screen, click on QUIT.
will revert back to a default setting of x2, x5, etc.
Changes to the display resolution should                     Figure 2 shows an example of one of the
only be done after the unit has already been                 INOVATION configuration screens.
calibrated using the data sheet or dead load
procedure.
The bottom parameter selection in this file folder                                 Figure 2
is labeled “Operating Mode”. It enables the unit
to retain the mode of operation in the event of a
power loss. When the power is re-applied, the
transmitter automatically returns to the mode
that it was operating in prior to the power loss.
Changes to the Operating Mode cause an
automatic power off - power on sequence of the
instrument when the configuration will be saved.
This allows the new Operating Mode to take
effect.
The second file folder enables you to calibrate
the unit with a transducer simulator or actual
weight, (recommended method of calibration).
Inside this 2nd file folder, once the zero
calibration has been performed, the program
asks the operator if the new zero has to modify
                                                        11
                                                 Figure 2.1
                                                                            PRINT CONFIGURATION
The Inovation software also allows the
                                                 Date 12/01/2004     Test Procedure for DAT instruments
operator to create new configurations,
                                                              Status Instrument
save new files and open existing files           Device address                                                                     2
without having any instrument con-               Setpoint 1 value                                                                1500
nected to the computer.                          Setpoint 2 value                                                                3500
                                                 Input 1 status                                                                     0
                                                 Input 2 status                                                                     0
Thanks to this feature, the download of          Instrument software version                                                    0.4
a given configuration to the instrument                       Parameters
can be performed at any time.                    Total capacity of transducers                                                   10000
                                                 Full scale output value (mV/V)                                                 2.0000
To perform the operation mentioned               Net weight (value)                                                               8000
                                                 Dead load (value)                                                                1500
above click on the “Edit” menu at the left       Display resolution                                                                  1
end upper corner (See figure 2.1), then          Input signal                                                                    0.599
enable the function “Edit Configuration          Internal counts (Reference only)                                              155729
                                                 Operating mode                                                               Net
Offline”.                                                       Filters / z-track
The message “Edit” will appear on the            Digital filter value (0 to 9: 0 = no filter; 9 = max filter; default 5)            5
display, showing this particular condition.      Motion band value (0 to 4: 0 = minimum; 4 = maximum; default 2)                    1
                                                 Auto zero range setting at power up (0 to 10.0% of net weight: 0 = disble)        0.0
Another useful function concern the              Zero tracking (0 to 4: default 2)                                                  2
                                                      12
Complete Configuration Menu
( PRG )
          (                                                  Enter
(    O            CAPAC             ( PRG )               total scale                 ( PRG )
                                                           capacity
                                                             Enter
         (
(    O            SENSIT          ( PRG )                transducer/s                 ( PRG )
                                                           sensitivity
                                                            Enter
         (
(    O              NET           ( PRG )                  “LIVE”                     ( PRG )
                                                         weight value
                                                            Enter
(    O
         (
                  DEAD L            ( PRG )              “Dead Load”                  ( PRG )
                                                         weight value
                                                        Select display
         (
(    O            DSPDIV            ( PRG )               divisions                   ( PRG )
                                                         0.001 - 50
                                                          View input
         (                                                                                      (
(    O            SIGNAL            ( PRG )              signal value             (      O
                                                            mV/V
                                                           View the
         (                                                                                      (
(    O            COUNTS            ( PRG )                  Raw                  (      O
                                                          A/D counts
                                                                                                    Operating
         (
(    O           OPMODE                            GROSS                 NET           PEAK H       mode @
                                ( PRG )                                                             power up
                                                                        ( PRG )
         (
(    O            UPLOAD        ( PRG )           RECEIV
         (
(    O            DNLOAD        ( PRG )       2 secs.              DONE
Place the unit in complete configuration               a digit, then use the “UP” or “DOWN” keys to
mode by pressing and holding the “PRG”                 increment or decrement the value of the
and SET” keys on the front of the unit. The            selected digit. Press the “PRG” key to enter
“SET” key must be pressed after the “PRG”              the new value, then press either the “UP” or
key. Release both keys when CONFIG is                  “DOWN” keys to move to the next sub-menu
displayed.                                             parameter or press the “O” key to return to the
Press the “PRG” key to enter the sub-menu,             main menu.
then use either the “UP” or “DOWN” keys to             To exit the configuration procedure, return to the
select one of the sub-menu parameters. Press           main menu and press the “O” key, (the display
the “PRG” key to view the default or previously        will indicate STORE ?), press the “PRG” key to
selected value for that parameter.                     save the changes and return to the operating
                                                       mode.
To change the value, press the “O” key to select
                                                  13
Complete Configuration Menu (cont’d)
NOTE: In order to ensure greater stability during the calibration procedure, the digital filter increases
to the maximum value automatically (value 9), therefore the weight indication updates very slowly at
this stage. Once the calibration procedure has been completed, the digital filter will return to the pre-
vious value automatically.
( PRG )
                                              (            Sets ZERO
  ( PRG )          CAL             (   O
                                                           calibration
                                             Enter the
               WT-VALUE                    actual weight             ( PRG )
                                           value (SPAN)
  ( PRG )          CAL
                                                                          NOTE:
                                                                          Input signal from the transducer/s
                                                                          must be stable when entering the
                                                                          zero and span values.
NOTE: Prior to calibration, the transmitter must           5. Enter the actual weight value using the “O” key
be configured, thus giving it a set of operating              to select each digit, then use the “UP” or
parameters.                                                   “DOWN” keys to increment or decrement the
                                                              value of the digit. Press the “PRG” key to enter
The unit can be calibrated using the 2-point                  the setting (CAL will be blinking on the display).
method or the optional multi-point (linearity
compensation) method. See page 15 for a                    6. Press the “PRG” key, (the display indicates
detailed description of the multi-point calibration           CALIBR), press either the “UP” or “DOWN
procedure.                                                    keys to move to the next parameter.
2-Point calibration via the front panel keys               7. If no additional parameter changes are
                                                              required, press the “O” key (the display
To calibrate the transmitter using the front panel            indicates STORE ?), press the “PRG” key to
keys, proceed as follows:                                     save the changes and return to the operating
                                                              mode.
1. With the unit in “Complete Configuration”
   mode, press either the “UP” or “DOWN”                   Should it be necessary to erase either of the
   keys until CALIBR is shown on the display.              previous calibration entries, proceed as follows:
2. Press the “PRG” key (CAL will be blinking on            To erase the zero calibration, press and hold the
   the display).                                           “FUN” key, and then press the “O” key.
                                                           Release both keys, the previous zero calibration
3. With the system unloaded, press the “O” key.            entry has been erased. To erase the span
                                                           calibration, press and hold the “FUN” key, and
4. Apply a known weight not less than 10% of               then press the “SET” key. Release both keys, the
   the net weight value and press the “UP” key.            previous span calibration entry has been erased.
                                                      14
Complete Configuration Menu (cont’d)
                                              Apply a known
           Lin P1                             weight on the
                                                  scale
( PRG )
                                                  Display and
         WT VALUE                                   edit test        ( PRG )
                                                  weight value
The Model 400 features a 10-point linearization         5. Press the “PRG” key. The unit will increment
option (available only with the front panel keys).         to the next linearization point (Lin P2).
Any number of points up to 10 may be used,
however, the last point must always be the full         6. Repeat steps 3 through 5 for additional points.
scale value of the weighing system.
                                                        7. If all 10 points will be linearized, follow the 1st.
Example: A one point linearization correction              procedure. For linearization correction using
would require a total of two points, one point             less than 10 points, follow the 2nd. procedure.
within the range, and one point for full scale.
                                                        1st Procedure (10-point linearization)
Prior to performing the linearization correction,
at least the zero of the unit must have already
                                                        7a. Enter the weight value for linearization point
been calibrated, while the full scale value is not
                                                            Lin P9, and press the “PRG” key (CAL will be
mandatory. The unit must be in CALIBR mode,
                                                            blinking on the display). Press the “PRG” key
and you should have known weights available
                                                            again, (the display indicates CALIBR).
equal to the points being set.
                                                        7b. Press the “O” key (the display will indicate
Multi-point Calibration via Front Panel Keys
                                                            STORE ?), press the “PRG” to save the
1. With the unit in either “Basic or Complete               changes and return to the operating mode.
   Configuration” mode, press the “PRG” key
   (CAL will be blinking on the display).               2nd. Procedure (less than 10-point linearization)
2. Press and hold the “SET” key until Lin P1 is         7a. The last linearization point must be the full
   blinking on the display.                                 scale value. Press the “O” key, (CAL will be
                                                            blinking on the display).
3. Apply a known weight for the first linearization
   point, and press the “PRG” key.                      7b. Press the “PRG” key (the display will show
                                                            CALIBR). Press the “O” key (the display will
4. Enter the actual weight value using the “O”              indicate STORE ?).
   key to select each digit, then use the “UP”
   and “DOWN” keys to increment or decrement            7c. Press the “PRG” key to save the changes
   the value of the selected digit.                         and return to the operating mode.
                                                      15
Complete Configuration Menu (cont’d)
( PRG )
                                                              Enter value
               (
    (     O            FILTER            ( PRG )                0 to 9              ( PRG )
                                                              Default = 5
                                                              Enter value
               (
    (     O            MOTION            ( PRG )                0 to 4              ( PRG )
                                                              Default = 2
                                                               Enter value
               (
    (     O            AUTO 0            ( PRG )              0.0 to 10.0%          ( PRG )
                                                              0 = Disabled
                                                              Enter value
               (
    (     O            0 TRAC            ( PRG )                0 to 4              ( PRG )
                                                              Default = 2
                                                              Enter value
               (
    (     O            0 BAND            ( PRG )               0 to 200             ( PRG )
                                                             Default = 100
1. With the unit in “Complete Configuration”            5. Press the “PRG” key to enter the new value,
   mode, press either the “UP” or “DOWN”                   then press either the “UP” or “DOWN” keys to
   keys until PARAM is shown on the display.               move to the next sub-menu parameter or
                                                           press the “O” key to return to the main menu.
2. Press the “PRG” key to enter the sub-menu,
   then use either the “UP” or “DOWN” keys to           6. To exit the configuration procedure, return to
   select one of the sub-menu parameters.                  the main menu and press the “O” key, (the
                                                           display indicates STORE ?), press the “PRG”
3. Press the “PRG” key to view the default or              key to save the changes and return to the
   previously selected value for that parameter.           operating mode.
See Table 2 on page 28 for a detailed explanation of the “PARAM” parameter selections
                                                   16
        Complete Configuration Menu (cont’d)
( PRG )
               (
    (     O             MODE 1         ( PRG )
                                                                                  ( PRG )    NET
                                                                                  ( PRG )   GROSS
                                                    See notes
               (                                                                  ( PRG )   PEAK H
    (     O             HYST 1        ( PRG )      on page 30      ( PRG )
                                                   Default = 2
                                                                                  ( PRG )    N.O.
               (                                   Enter value
    (     O             TIMER 1       ( PRG )      0.00 - 100.0    ( PRG )        ( PRG )    N.C.
                                                   Default = 0
                                                      Refer to
    (     O
               (
                        MODE 2        ( PRG )      procedure for                            NORMAL
                                                     MODE 1
                                                                                            STABLE
                                                    See notes
               (
    (     O             HYST 2        ( PRG )      on page 30      ( PRG )
                                                   Default = 2
                                                   Enter value
               (
    (     O             TIMER 2       ( PRG )      0.00 - 100.0    ( PRG )
                                                   Default = 0
               (                                   Enter value
    (     O             DELAY 2       ( PRG )      0.00 - 100.0    ( PRG )
                                                   Default = 0
                                                         17
        Complete Configuration Menu (cont’d)
                                                                                           Closing Input 1 = 10
                                                                                           Closing Input 2 = 02
                (                                                                 (
    (     O            TEST IN        ( PRG )        IN 0 0           (       O
                (
    (     O           TEST OUT          ( PRG )
                                                                          (
                                                              (   O               OUT 00
                                                                          (
                                                              (   O               OUT 10             See
                                                              (   O
                                                                          (
                                                                                  OUT 02            Table 1
                                                                          (
                                                              (   O               OUT 12
Table 1
OUT 10 ON OFF
OUT 02 OFF ON
OUT 12 ON ON
See Table 6 on page 30 for a detailed explanation of the “IN - OUT” parameter selections
                                                        18
                  SERIAL             ANALOG              CONFIG         CALIBR              PARAM             IN-OUT
                  ( PRG )
      (
              (
                            BAUD2R    BAUD1R
                                               *         ( PRG )
          O           BAUD R
                                                                                                               ( PRG )          2400
                                                                                                               ( PRG )          9600
              (
      (   O            PROT-1                 ( PRG )                                                          ( PRG )         19200
                                                         ( PRG )           CONTIN                              ( PRG )         38400
                                                         ( PRG )        DEMAND                                 ( PRG )        115200
                                                         ( PRG )           AUTOM
                                                         ( PRG )           SLAVE
                                                                        MODBUS          ( PRG )
                                                         ( PRG )            BCD
                                                         ( PRG )           PRINT                                ( PRG )          n 8 2
                                                         ( PRG )           NONE                                 ( PRG )         E 8 1
( PRG ) o 8 1
      (
              (                  **           ( PRG )
          O            PROT-2
      (   O
              (
                      ADDRES                   ( PRG )
                                                             Enter value
                                                               01 - 32         ( PRG )                   *“BAUD1R”, “BAUD2R” and “DC-ADD”
                                                              Max # 32
                                                                                                         come up on the display only if the
                                                                                                         “PROT-2” menu has been set as
                                                                                                         “DEVNET”.
                                       *
                                     DC-ADD
              (                                              Enter value
      (               PR-ADD                   ( PRG )                         ( PRG )
          O                                                   001 - 126
                                                                                                         **
                                                                                                         When the “PROT-2” menu is set as
                                                                                                         “PROFIB” or “DEVNET”, the “ANALOG”
                                                                                                         menu will be automatically disabled.
                                                                                                         Therefore the access to the “ANALOG”
                                                             Enter value                                 menu is possible only when the
              (
      (   O            DELAY                   ( PRG )       0.00 - 1.00       ( PRG )                   “PROT-2” has been set as “NONE”.
                                                             Max 1 sec.
              (                                                                              (                                OPERATING
      (   O           REM-CO                   ( PRG )       REMOTE               (     O                                       MODE
                                                                              ( PRG )            n 8 1
                                               ( PRG )
              (
      (   O            DATA F                                                 ( PRG )            n 8 2
                                                                              ( PRG )            E 8 1
                                                                              ( PRG )            o 8 1
                                                                              ( PRG )            n 7 2
                                                                              ( PRG )            E 7 1
                                                                              ( PRG )            E 7 2
See Tables 3 through 5 for                                                    ( PRG )            o 7 1
a detailed explanation of the                                                 ( PRG )            o 7 2
“SERIAL” parameter selections
                                                                   19
    Complete Configuration Menu (cont’d)
( PRG )
           (                                      Enter span
(     O            F-SCAL         ( PRG )           value              ( PRG )
           (
(     O             MODE            ( PRG )
                                                                                               ( PRG )               NET
                                                                                               ( PRG )           GROSS
           (                                     Enter weight
(     O           ANZERO          ( PRG )           value              ( PRG )
                                                                                               ( PRG )           PEAK H
           (
(     O             TEST            ( PRG )
                                                                                                         (
                                                                                           (     O               OUT       0
           (
(     O            RANGE          ( PRG )                                                                (
                                                                                           (     O               OUT 100
                                              ( PRG )        0-20 mA
                                              ( PRG )        4-20 mA
                                              ( PRG )           0-10 V
                                              ( PRG )           0-5 V
           (
(     O            OFFSET         ( PRG )
    1. With the unit in “Complete Configuration”             5. To edit numerical values, use the “O” key to
       mode, press either the “UP” or “DOWN”                    select the digit, then use either the “UP” or
       keys until ANALOG is shown on the display.               “DOWN” keys to increment or decrement the
                                                                value of the selected digit.
    2. Press the “PRG” key to enter the sub-menu,
       then use either the “UP” or “DOWN” keys to            6. Press the “PRG” key to enter the new value,
       select one of the sub-menu parameters.                   then press either the “UP” or “DOWN” keys to
                                                                move to the next sub-menu parameter or
    3. Press the “PRG” key to view the default or               press the “O” key to return to the main menu.
       previously selected value for that parameter.
                                                             The “ANZERO” parameter is a negative weight
    4. To change a parameter, press either the “UP”          value to which the zero offset of the analog
       or “DOWN” keys to view the choices for that           output can be referred.
       parameter.
       When the desired selection appears on the             Example: If the selected range is 4/20 mA and
       display, press the “PRG” key to enter the             the “ANZERO” value is “10”, the analog output
       selection and move to the next parameter.             value will be 4 mA when the weight will be “–10”
                                                        20
Serial Communication
Baud Rates
2400, 9600, 19200, 38400, and 115200
Data Formats
All serial characters are in ASCII format, and consist of the following.
1 Start Bit, 7 or 8 Data Bits, 1 or 2 Stop Bit, Parity No, Even or Odd.
The above data depend on the selection made in the “SERIAL” menu (see Table 3 on page 28)
Transmission of the serial data can be initiated in several ways as described below.
Automatic mode
The “Automatic” mode is used to interface with printers or other data acquisition devices. The serial
data is transmitted automatically, upon stabilization whenever the weight value has increased or
decreased more than 20 divisions.
This function is not performable when the “MOTION” parameter is set to 0 (weight stability check
disabled).
Continuous mode
The “Continuous” mode is used to interface with computers, remote displays, or other types of
devices that require constant data updating.
Demand mode
The “Demand” mode is used to interface with printers and requires a manual print command from the
front panel keys (Model 400) or a remote contact closure to initiate the output.
In the above modes, the transmitter sends serial data using the format shown below:
STX <status> <net weight> <gross weight> <peak value> ETX <chksum> EOT
STX (start of text) = 02h ETX (end of text) = 03h EOT (end of transmission) = 04h
NOTE: The brackets “<“ “>“ are not sent, they are only shown for clarification.
The result of the above calculation equals 1 character, which can be a numerical value of 0 to 9 or
A to F (Hexadecimal)
                                                   21
Serial Communication (cont’d)
Slave mode
The slave mode is used for interfacing with distributed control systems (DCS) or programmable logic
controllers (PLC). This mode requires a data request from the master to initiate the output.
In this mode the host sends serial data to the transmitter using the formats shown below:
After receiving the request, the transmitter responds with the following data string.
<addr> “N” <status> <net weight> <gross weight> <peak value> ETX <chksum> EOT
In case of an error, the transmitter will respond with the following data string.
Character strings
NOTE: The brackets “<“ ”>“ are not sent, they are only shown for clarification.
The checksum control value is obtained by performing the XOR operation.
It is expressed with 2 Hexadecimal digits. Example: 25 = 1D
The result of the above calculation equals 1 character, which can be a numerical value of 0 to 9 or
A to F (Hexadecimal)
In case of an error, the transmitter will respond with the following data string.
The programmed values are stored in RAM, however, they are not retained if the power is removed.
There is no limit on the number of times that these commands can be performed.
                                                   22
Store set-point values in Eeprom
In case of an error, the transmitter will respond with the following data string.
In case of an error, the transmitter will respond with the following data string.
In case of an error, the transmitter will respond with the following data string.
                                                   23
                                         Switch to Net weight mode
In case of an error, the transmitter will respond with the following data string.
In case of an error, the transmitter will respond with the following data string.
Modbus mode
The Modbus mode is used to communicate with Modicon Programmable Controllers using a unique
message structure that the controllers can understand regardless of the type of networks being used.
Please refer to the tables on pages 31 through 41 for information regarding the MODBUS RTU
Protocol Specifications.
Modbus® and Modicon® are registered trademarks of the Schneider Automation Group.
                                                        24
Print mode
This protocol enables the communication with a printer. The data on the printed report are the follo-
wing: NET, GROSS and TARE (+ PEAK, if enabled).
The print command takes place by pressing the PRG key or by activating the remote input #2. (*)
The printed report is issued only if the following conditions are met:
       Gross weight positive
       Net weight positive
       Stable weight (**)
       BLIND function disabled (***)
(*) In case the PRG key is locked (see page 6) the print command can take place through the
    remote input #2 only.
(**) The weight stability condition is controlled only if the “MOTION” parameter (see page 16) is set
     to a value  0.
     If MOTION = 0 the printed report is issued even if the weight is unstable..
(***)In case the BLIND function is enabled (see page 6) the print command can take place through
     the remote input #2 only, not via the PRG key.
Between one printout and the next one the weight must change for at least 20 counts.
                                                                        Rx
                                                                        Tx
                                                                        Gnd
Tx Rx Gnd
27 24 19
                                                     25
                            Serial Output Wiring Connections
RS-232 serial connections to a computer, remote serial display or other data acquisition device.
See Figure 3.
Figure 3
                                                                       Rx
                                                                       Tx
                                                                       Gnd
Tx Rx Gnd
27 24 19
RS-422 and 485 serial configurations allow connection to a distributed control system (DCS) or
programmable logic controller (PLC) over long distances (up to 3200 ft.)
RS-422 allows a point-to-point connection, while RS-485 allows a multi-drop connection of up to 32
units. See Figure 4 for the RS422 connection and Figure 5 for the RS485 connection.
+ Rxd
- Rxd
+ Txd
- Txd
22 19
                                                        26
                                       Figure 5          RS485 Connection
                                                                                                              Gnd
                                                                                                             RS485   +
                                                                                                             RS485
  24 23 22 21 20 19                                               24 23 22 21 20 19
                          Transmitter # 32                                                 Transmitter # 1
Rx (Pin 2) Rx (Pin 3)
Tx (Pin 3) Tx (Pin 2)
Tx Rx Gnd
  27            24          19
                                                                Std. USB cable
                                                                                              PC (USB port)
                                                                  3
                                                        USB       2
                                                                  1
                                                        Precise
                                                          DAT
                                                          400
Not only the RS232 or USB, but also the RS422 or RS485 can be used for the communication
with the Inovation software. Refer to Figures 3, 4, 5 depending on the desired connection.
                                                         27
 Sub-menu parameter selections
Digital filtering affects the update rate and speed of the unit.        0 BAND:
Filter setting = 0         Update rate = 25 Hz                          this parameter defines the operating
Filter setting = 1         Update rate = 10 Hz
                                                                        range of the “Zero” key (and Input 1),
Filter setting = 2         Update rate = 5 Hz
Filter setting = 3         Update rate = 2.5 Hz                         used to zero the unit in Gross mode.
Filter setting = 4         Update rate = 1.5 Hz
Filter setting = 5         Update rate = 1 Hz
Filter setting = 6         Update rate = 0.7 Hz
Filter setting = 7         Update rate = 0.4 Hz
Filter setting = 8         Update rate = 0.2 Hz
Filter setting = 9         Update rate = 0.1 Hz
                                                        28
Sub-menu parameter selections (cont’d)
Continuous transmission:
The transmission speed depends on the digital filter setting and the baud rate value selected.
See Table 4 below for additional information.
Table 4
                                             BAUD RATE
                         2400             9600              19,200           38,400   115,200
        D      0         6 Hz             25 Hz             50 Hz            50 Hz    50 Hz
        I      1         6 Hz             25 Hz             50 Hz            50 Hz    50 Hz
        G
        I      2         6 Hz             25 Hz             50 Hz            50 Hz    50 Hz
        T      3         6 Hz             25 Hz             25 Hz            25 Hz    25 Hz
        A
        L      4         6 Hz             25 Hz             25 Hz            25 Hz    25 Hz
        F      5         6 Hz             12 Hz             12 Hz            12 Hz    12 Hz
        I      6         6 Hz             12 Hz             12 Hz            12 Hz    12 Hz
        L
        T      7         6 Hz             12 Hz             12 Hz            12 Hz    12 Hz
        E
        R      8         6 Hz             6 Hz              6 Hz             6 Hz      6 Hz
               9         6 Hz             6 Hz              6 Hz             6 Hz      6 Hz
Table 5
                                             BAUD RATE
                                                  2400               9600
                                D     0           12 Hz              50 Hz
                                I     1           12 Hz              50 Hz
                                G
                                I     2           12 Hz              50 Hz
                                T     3           12 Hz              25 Hz
                                A
                                L     4           12 Hz              25 Hz
                                F     5           12 Hz              12 Hz
                                I     6           12 Hz              12 Hz
                                L
                                T     7           12 Hz              12 Hz
                                E
                                R     8           6 Hz               6 Hz
                                      9           6 Hz               6 Hz
                                                    29
Sub-menu parameter selections (cont’d)
Table 6
NOTES:
       1. The hysteresis setting must be less than it’s respective setpoint value.
          Entering a hysteresis setting greater than the setpoint value will automatically
          be replaced with a default setting of (2) counts.
       Setpoint 1 is used to fill a vessel with 5,000 lbs. of product, and setpoint 2 will be
       used as a high weight alarm. When the weight in the vessel reaches 5,000 lbs.,
       relay # 1 de-energizes stopping the flow, feed, etc. The product is drawn from the
       vessel as required, when the weight in the vessel reaches 10 lbs., setpoint # 1’s
       relay re-energizes and automatically refills the vessel to 5,000 lbs.
                                                  30
                           MODBUS RTU PROTOCOL SPECIFICATIONS
1. PRECAUTIONS
The Modbus protocol writes directly into the memory of the instrument.
Use caution when sending data to the instrument:
The data being sent must be within the specified ranges given in the tables on the following pages.
Some of the data is written into the E2prom’s memory (refer to the column in the tables “Stored in
E2prom”) for additional information. This memory can only be written to 100,000 times, therefore,
you should avoid writing continuously into this memory.
To confirm the storage of a new value in the E2prom, perform the MAKE - BACKUP function.
If this function is not performed, all of the newly stored data will be lost when power is removed from
the instrument.
The “Set point” setting does not require the MAKE - BACKUP function to be performed.
The set point values are expressed as number of divisions
This means that any set point value (temporary or permanent) has to be multiplied by the “Display
division” data, in order to obtain the actual value of the set point.
E.G. Assuming that the “display division” value is 0.2; if the PLC asks for a set point value and get
the data “3949” from the instrument, this means that the actual value of the set point is 789.8
(3949 x 0.2 = 789.8)
Even the other weight values are given as number of divisions, except for the “Net weight” and
“Gross weight”.
“Net weight” and “Gross weight” (Modbus addresses 40006 and 40007) are given as absolute
values
Note:
The numeric values on the following pages for (addresses, codes, and data) are represented as
decimal values.
The addresses given in the following tables refer to the standards provided in the Modicon Modbus
Protocol Reference Guide PI-MBUS-300 Rev J.
2. DATA FORMAT
   Start Bit : 1
   Data Bits : 8, the Least Significant Bit (LSB) is sent first.
   Parity : No, Even, Odd
   Stop Bits : 1, 2
                                           Possible Data Format Selection: N-8-2     E-8-1   O-8-1
        Function                                            Description
         01 (01)         READ COIL STATUS
         03 (03)         READ HOLDING REGISTERS
         04 (04)         READ INPUT REGISTERS
         05 (05)         FORCE SINGLE COIL
         06 (06)         PRESET SINGLE REGISTER
         15 (0F)         FORCE MULTIPLE COILS
         16 (10)         PRESET MULTIPLE REGISTERS
                                                  31
Each single function is explained in detail on pages 34, 35 and 36 of the operator’s manual.
Each function is composed by a Query (request master             instrument) and a Response
(answer instrument         master). Queries and responses are composed by a sequence of data that
are briefly explained here below (***), please note that the code “0x” before any value indicates that
the value is represented as hexadecimal value.
Besides, all the Modbus addresses (the ones mentioned in the left end column on pages 38 and
39) have to be represented as hexadecimal values, but, only for these addresses, there is an
important rule to be taken into consideration:
 Before converting the address into an hexadecimal value, the 1st digit on the left has not to
         be considered and the remaining 4 digits have to be decremented by “1”.
                                                    1st Example:
To represent the address 40150 (“display division” on page 38) as hexadecimal value, exclude the
digit “4” (remains “0150”), then decrement by 1 the number “0150”. The result is “0149”, now perform
the conversion into hexadecimal. The result is “00 95”. This is the value to be used in the query.
                                                    2nd Example:
To represent the address 40102 (“load cells sensitivity” on page 38) as hexadecimal value, exclude
the digit “4” (remains “0102”), then decrement by 1 the number “0102”. The result is “0101”, now
perform the conversion into hexadecimal. The result is “00 65”. This is the value to be used in the
query.
(***)
Address: is the number of the instrument’s address, represented as hexadecimal value
           (in this case do not decrement by 1 the number).
Function: is the number of the function to be performed (see the table above)
1st Output address: is the address number where the reading of the outputs status has to start
Number of outputs: is the number of the outputs to be read
1st Input address: is the address number where the reading of the inputs has to start
Number of inputs: is the number of the inputs to be read
Number of bytes: represents the number of bytes which compose a data
2 byte CRC: “Cyclical Redundancy Check”. It’s an algorithm used as checksum.
Exceptions:
Query Master to slave 01 Response time slave 01 Response slave 01 20 mSec delay Query Master to slave 02
Query Master to slave 01 Response time slave 01 Response slave 01 delay Query Master to slave 02
                                                        32
Scheme of the communication time in a Master-Slave RS-485 system, using a double twisted pair.
Tx PC
Tx Slave 01
Tx Slave 01
Tx PC
Tx Slave 01
Tx Slave 01
15 mSec
                                                 33
                                           5. LIST OF TRANSMISSION STRINGS
QUERY
  Address           Function          1st. Output address       Number of outputs        2 byte      Total byte
      A              0 x 01                 0 x 0000                  0 x 0008            CRC            8
RESPONSE
QUERY
Address Function 1st. Register address Number of register 2 byte Total byte
RESPONSE
Address Function Number of bytes 1st. Register 2nd. Register 2 byte Total byte
QUERY
Address Function 1st. Register address Number of register 2 byte Total byte
RESPONSE
                                                                 34
5.5. FUNCTION 5: FORCE SINGLE COIL
QUERY
RESPONSE
QUERY
RESPONSE
QUERY
Address Function 1st. Output address Nº of outputs Nº of bytes Status of outputs 2 byte Total byte
RESPONSE
Number of outputs: Number of the output to be written, starting from the address.
Number of bytes: Number of bytes transmitted as status of outputs (8 outputs each byte)
Status of outputs: 1 output each bit (1 = On, 0 = Off); the 1st. Output coincides with the least significant bit (LSB) of the 1st.
Byte. Non-significant bits are set to zero.
The response includes the identification of the modified outputs.
                                                                35
5.16 FUNCTION 16: PRESET MULTIPLE REGISTERS
QUERY
Address      Function            1st.              Number of           Number of       1st.           2nd.      2 byte    Total byte
                            Register output         register             bytes       Register        Register
                                                                                      value           value
    A         0 x 10            0 x 0000            0 x 0002            0 x 04       0 x 0000        0 x 0000   CRC      7 + 2 x Nº of
                                                                                                                         registers + 2
RESPONSE
Number of registers: Number of the registers to be written, starting from the address.
Number of bytes: Number of bytes transmitted as registers value (2 bytes each register)
Registers value: Content of the registers, starting from the 1st.
The response includes the identification of the modified registers.
6 BROADCAST MODE
The communication strings are controlled by the CRC (Cyclical Redundancy Check). If a communication error occurs, the
slave unit doesn’t answer. The master unit controls a timeout when waiting for the response from the slave. If the slave
doesn’t answer during this timeout, it means that a communication error has occurred.
If the string is received correctly but is not executable, the slave answers the master with an EXCEPTION RESPONSE.
The field “function” is transmitted having the most significant digit (MSD) set to 1.
EXCEPTION RESPONSE
            Code                                                          Description
             1                  ILLEGAL FUNCTION (The function is not valid or not supported)
             2                  ILLEGAL DATA ADDRESS (The specified data address is not available)
             3                  ILLEGAL DATA VALUE (The data value received is not valid)
                                                                   36
6 BROADCAST MODE
The communication strings are controlled by the CRC (Cyclical Redundancy Check). If a communication error occurs, the
slave unit doesn’t answer. The master unit controls a timeout when waiting for the response from the slave. If the slave
doesn’t answer during this timeout, it means that a communication error has occurred.
If the string is received correctly but is not executable, the slave answers the master with an EXCEPTION RESPONSE.
The field “function” is transmitted having the most significant digit (MSD) set to 1.
EXCEPTION RESPONSE
          Code                                                    Description
            1                ILLEGAL FUNCTION (The function is not valid or not supported)
            2                ILLEGAL DATA ADDRESS (The specified data address is not available)
            3                ILLEGAL DATA VALUE (The data value received is not valid)
                                                            37
                                       List of the Holding Registers (4x)
 40100    Load cell/s total capacity (kg) H                CAPAC               0-500000 (1)              YES
 40101    Load cell/s total capacity (kg) L
 40102    Sensitivity of load cell/s                       SENSIT           1.0000 - 4.0000 (5)          YES
 40103    Net weight of the weighing system (H)             NET              0 to full scale (1)         YES
 40104    Net weight of the weighing system (L)
 40105    Dead load of the weighing system (H)             DEAD L            0 to full scale (1)         YES
 40106    Dead load of the weighing system (L)
 40110    Operating mode                                   OPMODE                 0-2 (3)                YES
 40150    Display divisions                                DSPDIV                0 - 14 (2)              YES
                                                                               See Table “A”
                      Weighing Parameters
 40180    Digital filter value                             FILTER                 0-9 (3)                YES
 40181    Motion band value                                MOTION                 0-4 (3)                YES
 40182    Auto zero value                                  AUTO 0            0.1-10.00 (3) (5)           YES
 40183    Zero tracking value                              0 TRAC                 0-4 (3)                YES
                    Set Points Configuration
 40200    Operating mode Set point 1                       MODE 1                                        YES
 40201    Hysteresis Set point 1                           HYST 1              0 to full scale           YES
 40202    Timer Set point 1                                TIMER1              0.1-100.0 (5)             YES
 40203    Delay Set point 1                                DELAY1              0.1-100.0 (5)             YES
 40204    Operating mode Set point 2                       MODE 2                                        YES
 40205    Hysteresis Set point 2                           HYST 2              0 to full scale           YES
 40206    Timer Set point 2                                TIMER2              0.1-100.0 (5)             YES
 40207    Delay Set point 2                                DELAY2              0.1-100.0 (5)             YES
(1)   The combined theoretical net and tare weight values must not exceed the total capacity value.
(2)   Refer to the 15 values from 0.001 to 50
(3)   For the list of values please refer to the user manual.
(4)   They are stored in the E2prom if writing the function 0000 in the status register.
(5)   When setting these values the decimal point has not to be included.
    Once the “serial address” and “response delay” have been modified, they take effect immediately.
    The serial port mode can’t be modified via Modbus.
    When setting a new baud rate value, the power to the instrument must be removed and re-applied in order for the
      changes to take effect.
                                                           TABLE “A”
                                                     IDENTIFICATION CODES
                                                             39
                Table of the Input Status bytes
(*)        The scale is considered unloaded (empty) when the weight value is within the “0 Band”
           range (see “Table 2”, page 28)
          Modbus                                                           Description
          Address
           30003              A/D converter internal counts H
           30004              A/D converter internal counts L
           30005              Millivolt value
           30006              Instrument software release
           30007              Instrument “On-line” (*)
Modbus address 30007 =           00 01: the weight value is shown on the display. The instrument is “On-line”
Modbus address 30007 =           00 00: one of the error conditions (underload, overload, offrange) occurs, or any
                                 programming procedure is underway. The instrument is “Off-line”.
                                                          Bit’s meaning
      Modbus            Description                                                            Range value             Stored in
      Address                                                                                                          E2prom
                                                      0                      1
       00001            Logic output 1          De-activated              Activated                 1 bit                 NO
       00002            Logic output 2          De-activated              Activated                 1 bit                 NO
                                                                     40
                                       List of the functions “Command register”
For additional information regarding the Modbus protocol, please refer to the Modbus Protocol Manual PI-MBUS-300 Rev J.
                                                              41
                 Transfer of a complete set-up configuration (Download - Upload)
With regard to the use of the InovationTM software, the Upload and Download functions represent an
alternative for configuring the instrument, and offers remarkable differences:
     The InovationTM software must be installed on a PC, while the Upload/Download procedures
       do not require any software installation.
     The InovationTM software allows to configure the instrument (or to transfer an existing
       configuration) from a PC, and parameters can be edited at any time prior to starting the
       transfer process
     The Upload and Download functions allow to transfer the complete set-up configuration
       from one instrument to another, passing through a PC, without enabling parameters
       modification.
       In this case, the instrument can send to a PC (download) or receive from a PC a text file (*.txt)
       containing the configuration. Upload and/or Download can be managed through the use of any
       PC application able to transfer text files: MS WindowsTM Hyper Terminal is an example.
Practically:
With the InovationTM software the PC is used to create a set-up configuration to be transferred
to an instrument.
With the Upload/Download procedures the PC is only used as a “bridge” (a tool) to transfer
the set-up configuration from an instrument to another one.
The advantages related to Upload and Download functions can be summed up as follows:
(*)
While InovationTM software saves the theoretical calibration (Data-sheet) only, the Upload and
Download functions also save the actual Zero and FS Calibration previously performed as Dead
weight calibration.
In other words, a new instrument that must replace a faulty one will retain an identical behaviour if it
has received the configuration of the previously installed unit .
This function is very handy on those weighing equipments where both a new Zero and a new FS
calibration would be difficult or need the use of large volume sample masses, or worst, in case of a
tank filled with a product that cannot be emptied for some time, so that the Zero calibration cannot be
done.
                                                   42
DOWNLOAD (transfer of configuration from instrument to PC)
1.    From the CONFIG menu, select “DNLOAD” and leave the instrument in this condition (the
      instrument is now ready to transfer its own configuration to the PC).
Instrument’s display
DNLOAD
2.    Open the Hyper Terminal application on your PC (Start - All Programs - Accessories -
      Communications)
      Assign any name to the “Connection description” and Click on OK.
3.    At “Connect using”, select the serial port to be used for communicating with the DAT
      (the example here below selects the PC’s COM1 serial port), then Click on OK.
                                              43
4.        Set the serial port communication parameters.
          Baud rate (Bits per second) and Data format (Data Bits, Parity, Stop Bits) must
            correspond to those programmed in the DAT (check SERIAL menu on page 19)
          Select “Flow Control” = None).
          Click on “Apply”, then “OK”.
                                                  44
6.        On the upper part of the screen (“ASCII Sending”):
         Enable “Send line ends with line feeds”
         Set “Line Delay” at 200 milliseconds
Then click on OK two times (*) = 1st time - (**) = 2nd time
(*) (**)
8.        Press PRG key on the DAT to start data transfer to the PC.
          When transfer is complete (lasting a few seconds), the instrument displays “DONE”.
Instrument’s display
                                                      45
9.     DAT configuration data received on PC will have this format:
10.    Close the connection with a click on           button (Disconnect) then close the Hyper
      Terminal application using the X button on the upper right corner of the screen
12.    In the destination directory previously decided at Section 7, a text file was created (and a
       name assigned) containing the instrument’s configuration. The file can be open with the
       Notepad application although the related data cannot be interpreted and most of all must
not    be modified.
                                                 46
UPLOAD (transfer of configuration from PC to instrument)
13.    From the CONFIG menu, select “UPLOAD” and leave the instrument in this condition (the
       instrument is now ready to receive from the PC the text file containing the configuration ).
                                       Instrument’s display
UPLOAD
14.    Open the Hyper Terminal session on your PC, saved as in Section 11 (Start - All Programs -
       Accessories - Communications). The session name is ABC in the following example.
16.    Get the instrument ready to receive the text file from the PC by pressing the PRG key
       (message on the display: “Receiv”)
Instrument’s display
17.    Browse the directory previously decided at Section 7 and locate the text file to be transferred
       to the instrument, a double-click on it will start the data tranfer process.
       Upon receipt of file, the instrument automatically returns to CONFIG menu.
Instrument’s display
18.    Close the Hyper Terminal application using       X button on the upper right corner of the screen
      then, click on “Yes” when the following message is prompted:
19.    Press the “>0<” key on the instrument (message “STORE?”), then click on the PRG key.
       The instrument will now operate according to the new configuration just received.
                                                   47
     Intrinsic Safety Barriers
                                          - Signal                               - Signal
                                                        8     Z961H      1
          - Exc.      13                  + Signal                               + Signal      1     - Sig     + Exc      1
                                                        5                4                           + Sig      - Exc
          + Exc.      14                                                                       2                          2
          + Sens      15                  - Exc.                                 - Exc.        3     - Exc      + Sig     3
          - Sens      16                                6                3                     4     - Sens      - Sig    4
                                                              Z966H
          - Sig.      17                  + Exc.                                 + Exc.        5     + Sens    Shield     5
          + Sig.      18                                5                4                     6     + Exc
                                          - Sense                                - Sense       7     Shield              TB1
                                                        8     Z961H      1
                                          + Sense                                + Sense      TB5              + Exc      1
                                                        5                4                                      - Exc     2
                                                                                                                + Sig     3
48
                                                                                                                 - Sig    4
                                                                                                               Shield     5
TB2
- Exc. Black
- Sense Yellow
+ Signal Green
- Signal White
       Shield           Bare
Options
                                                Precise™
                                         24 Volt Power Supply
Installation
   Make sure the installation complies with local regulations and electrical codes.
   Connect AC voltage to the terminals marked “L” and “N”.
   The DC voltage is available on the terminals marked “+” and “-”. The second set of terminals
     are used when powering two transmitters from a single power supply (See notes).
   A red LED is illuminated when the power supply is “ON”.
Notes: The power supply should be used to power two transmitters only when a single load cell is
       connected to each transmitter. For multiple load cell applications it is recommended that an
       individual power supply be used for each transmitter. This enables the power supply to
       compensate for any fluctuations in the incoming line voltage. Recent field experiences have
       shown that voltage drops of greater than 10% can cause the transmitters to drift and become
       unstable.
Fuse
                                     L      N    LED        + -
                                     Mains in              24 Vdc
Fuse Replacement
The following procedures require work inside the power supply enclosure and should be performed
by qualified service personnel.
         In the event of a malfunction, please contact the nearest distributor for assistance.
        Any attempt to modify or repair the power supply will void the manufacturers warranty.
                                                  49
                          The USB connection between the DAT-400 and the PC
Introduction
The hardware module installed inside the DAT-400 and allowing the instrument to be interfaced with a PC
through the USB port, is named CP210x, manufactured by Silicon Laboratories.
The CP210x module needs the drivers to be installed in the PC when the connection to the DAT-400 is
established for the first time.
Virtual COM Port driver installation for Windows 2000 / XP / Server 2003 / Vista
Download the drivers from the Manufacturer’s web site using the following link:
http://www.silabs.com/tgwWebApp/public/web_content/products/Microcontrollers/USB/en/mcu_vcp.htm
        Mouse click on “VCP Driver Kit”, to start the download of the file named:
          “CP210x_VCP_Win2K_XP_S2K3.exe” will start.
 Double mouse click on the icon of the file just downloaded to start the set-up wizard.
In some cases the system needs to be re-started in order to complete the installation process.
Switch-on the DAT-400 and connect it to the PC using a std. USB cable, the message “Found new hardware”
appears on the PC screen, then a message advising that the new hardware is ready to be used will appear.
If the COM port # automatically assigned by the system to the USB adapter is < 6 the Inovation software can
be used immediately, otherwise it will be necessary to change the COM port # manually.
Perform the following steps to see (or to change) the COM port # assigned to the USB adapter:
    4.    Double mouse click on “CP210x USB to UART Bridge Controller (COM X)”.
    5.    Select “Port settings” then “Advanced”.
    6.    Open the “COM port number” popup menu and assign a COM port number 1 to 6.
    7.    Click on OK and come back to the “Ports (COM & LPT)” list.
50