pH51/pH54
Pocket-size Waterproof pH Meters
with replaceable electrode and manual calibration
Waterproof budget pH testers. Their waterproof casing and double junction replaceable
electrodes make them suitable also for heavy duty applications, such as wastewater
treatment and agriculture.
The modular design allows easy electrode and battery replacement.
Manual calibration prolongs the battery life up to 1500 hours.

Specifications

pH51 pH54
Range 0.0 to 14.0 pH 0.00 to 14.00 pH
Resolution 0.1 pH 0.01 pH
Accuracy (@25°C) ±0.1 pH ±0.1 pH
Typical EMC Deviation ±0.1 pH ±0.1 pH
Calibration manual at 2 points through trimmers manual at 2 points through trimmers
pH electrode MA73600 (replaceable) MA73600 (replaceable)
Environment 0 to 50°C; max RH 100% 0 to 50°C; max RH 100%
Battery Type 3 x 1.5V alkaline 3 x 1.5V alkaline
Battery Life more than 1500 hours of continuous use more than 1500 hours of continuous use
Packaging dimensions 254 x 67 x 47 mm 254 x 67 x 47 mm
Packaging weight 186 g 186 g

Accessories
MA73600 Replaceable electrode for M10007B pH 7.01 buffer solution 20 mL
pH51 and pH54 sachet (25 pcs)
M10000B Electrode rinse solution, 20 mL M10010B pH 10.01 buffer solution 20 mL
sachet (25 pcs) sachet (25 pcs)
M10004B pH 4.01 buffer solution 20 mL MA9015 Electrode storage solution, 230 mL
sachet (25 pcs) MA753 Hard carrying case for 2 testers

Ordering Information Packaging Information


All testers are supplied in a carton box complete with Optionally pH51 is also available in a kit (Mi5165,
calibration solution, batteries, instruction manual Mi5166, Mi5175, Mi5176) together with C65,
and screwdriver for calibration. C66, T75 or T76.

pH600
pH Economical Pocket Tester
Milwaukee’s budget pH tester with 1 point calibration is an
easy-to-use instrument for applications such as aquarium,
swimming pool and hydroponics.

Specifications pH600
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy ±0.1 pH
Calibration manual, 1 point
Environment 0 to 50°C / 32 to 122°F; max RH 95%
Battery Type / Battery Life 3 x 1.5V alkaline / 700 hours of use
Packaging dimensions 180 x 65 x 32 mm
Packaging weight 120 g

Ordering Information Accessories


pH600 is supplied in a plastic hard carrying case, M10004B pH 4.01 buffer solution 20 mL
complete with protective cap, calibration screwdriver, sachet (25 pcs)
batteries and instructions. M10007B pH 7.01 buffer solution 20 mL
sachet (25 pcs)
M10010B pH 10.01 buffer solution, 20 mL
sachet (25 pcs)
MA9015 Electrode storage solution, 230 mL
MA9016 Electrode cleaning solution, 230 mL

 21
 
EC59 PRO/EC60 PRO
Pocket-size EC/TDS/Temp Meters
Water-resistant pocket-size EC/TDS/Temp meters include features such as
replaceable probe, temperature in °C or °F, automatic temperature compensation with
adjustable β, battery level indicator, stability indicator, automatic shut-off and automatic
calibration all in a floating, water-resistant casing.

EC59 PRO shows on the dual-level LCD the EC (3999 µS/cm) or TDS (2000 ppm) value.
It also displays the temperature from 0.0 to 60.0°C (or 32.0 to 140.0°F) on the secondary
level at the same time.

EC60 PRO shows on the dual-level LCD the EC (20.00 mS/cm) or TDS (10.00 ppt) value.
It also displays the temperature from 0.0 to 60.0°C (or 32.0 to 140.0°F) on the secondary
level at the same time.

Specifications

EC59 PRO EC60 PRO


Range EC 3999 μS/cm 20.00 mS/cm
TDS 2000 ppm 10.00 ppt
Temp 0.0 to 60.0°C / 32.0 to 140.0°F 0.0 to 60.0°C / / 32.0 to 140.0°F
Resolution EC 1 μS/cm 0.01 mS/cm
TDS 1 ppm 0.01 ppt
Temp 0.1°C / 0.1°F 0.1°C / 0.1°F
Accuracy EC ±2% Full Scale ±2% Full Scale
(@20°C) TDS ±2% Full Scale ±2% Full Scale
Temp ±0.5°C / ±1°F ±0.5°C / ±1°F
Typical EMC EC ±2% Full Scale ±2% Full Scale
Deviation TDS ±2% Full Scale ±2% Full Scale
Temp ±0.5°C / ±1°F ±0.5°C / ±1°F
Calibration automatic, 1 point with 1413 µS/cm automatic, 1 point with 12880 µS/cm
calibration solution calibration solution
Temperature Compensation automatic, with β=0.0 to 2.4%/°C automatic, with β=0.0 to 2.4%/°C
Probe Mi59P (replaceable) Mi59P (replaceable)
Environment 0 to 50°C / 32 to 122°F; max RH 100% 0 to 50°C / 32 to 122°F; max RH 100%
Battery Type 4 x 1.5V; IEC LR44, A76 (included) 4 x 1.5V; IEC LR44, A76 (included)
Battery Life approx. 100 hours of use approx. 100 hours of use
Auto-off after 8 minutes of non-use after 8 minutes of non-use
Packaging dimensions 254 x 67 x 47 mm 254 x 67 x 47 mm
Packaging weight 180 g 180 g

EC/TDS/Temp probe
Accessories Exposed temperature sensor
Mi59P Replaceable probe for EC59 & EC60 M10038B 6.44 ppt (g/L) calibration solution,
M10000B Rinse solution, 20 mL sachet, 25 pcs 20 mL sachet, (25 pcs)
M10030B 12880 µS/cm calibration solution, MA9060 12880 µS/cm calibration solution,
20 mL sachet, 25 pcs 230 mL bottle
M10031B 1413 µS/cm calibration solution, MA9061 1413 µS/cm calibration solution, Replaceable EC/TDS probe
20 mL sachet, 25 pcs 230 mL bottle
M10032B 1382 ppm (mg/L) calibration MA9016 Cleaning solution, 230 mL bottle
solution, 20 mL sachet, (25 pcs) MA753 Hard carrying case for 2 testers Replaceable probe
Replace the probe in a fast and simple way yourself!
Just unscrew the plastic ring on the top of the probe and
replace the probe with a new one.
Ordering Information
EC59 PRO is supplied in a carton box complete
with protective cap, 20 mL 1413 µS/cm sachet of
calibration solution, batteries and instructions.

EC60 PRO is supplied in a carton box complete


with protective cap, 20 mL 12880 µS/cm sachet of
calibration solution, batteries and instructions.

Optionally EC59 PRO and EC60 PRO is also


availablbe in a kit (Mi5559 or Mi5560) together with
pH55 PRO pH/Temp Meter.

28

MW10/MW11
Low cost digital photometers to
measure Free & Total Chlorine
Chlorine is the most commonly used water
disinfectant. Applications vary from treatment of
drinking water and wastewater to pool and spa
sanitization and food processing to sterilization.

Milwaukee offers 2 models:

MW10 for measuring free chlorine


(0.00 to 2.50 mg/L)

MW11 to measure total chlorine


(0.00 to 3.50 mg/L).

Key features include:

• User friendly;
• Small & Ergonomic case design;
• Inexpensive;
• Large and easy to read display;
• Good accuracy and immediate results;

Specifications

MW10 MW11
Free chlorine Total chlorine
Range 0.00 to 2.50 ppm 0.00 to 3.50 ppm
Resolution 0.01 ppm 0.01 ppm
Accuracy (@25°C) ±0.03 ppm ±3% of reading ±0.03 ppm ±3% of reading
Typical EMC Dev. ±0.01 ppm ±0.01 ppm
Light Source Light Emitting Diode @ 525 nm Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell Silicon Photocell
Method Adaptation of USEPA method Adaptation of USEPA method
330.5. The reaction between 330.5. The reaction between
free chlorine and the DPD free chlorine and the DPD
reagent causes a pink tint reagent causes a pink tint
in the sample. in the sample.
Environment 0 to 50°C (32 to 122°F) 0 to 50°C (32 to 122°F)
max. 95% RH non-condensing max. 95% RH non-condensing
Battery Type 1 x 1.5V AAA 1 x 1.5V AAA
Auto-off after 2 minutes of non use after 2 minutes of non use
Packaging dimensions 115 x 115 x 84 mm 115 x 115 x 84 mm
Packaging weight 180 g 180 g

Accessories The handy photometers are supplied in a


Mi526-25 Free Chlorine powder reagent, (25 pcs)
carton box including all accessories.
Mi0011 Glass cuvets (2 pcs)
Mi524-25 Total Chlorine powder reagent (25 pcs) Mi0013 Stoppers for cuvets (2 pcs)

Ordering information:
All handy photometers are supplied in a carton box including 2 cuvets, 6 powder reagents,
1 x 1.5 V AAA battery and instructions.

44
 „
MA871/MA872/MA873/MA881
Digital Refractometers for Brix, Fructose, Glucose
and Invert Sugar Measurements
The digital refractometers
are optical instruments that
employ the measurement of
refractive index to determine
the % Brix of sugar (MA871), % Fruc-
tose (MA872), % Glucose (MA873) and
% Invert Sugar (MA881) in aqueous
solutions.

The method is both simple and quick.


Samples are measured after a simple
user calibration with deionized or
distilled water. Within seconds the
instruments measure the refractive index of the
sample and convert it to % Brix or % by weight concentra-
tion units.

The digital refractometers eliminate the uncertainity


associated with mechanical refractometers and are easily
portable for measurements in the field.
The measurement technique and temperature
Key features include: compensation employ methodology recommended in
• Dual-level LCD the ICUMSA Methods Book (Internationally recognized
• Automatic Temperature Compensation (ATC) body for Sugar Analysis). Temperature (in °C or °F) is
• Easy setup and storage displayed simultaneously with the measurement on the
• Battery operation with Low Power indicator (BEPS) large dual level display along with icons for Low Power and
• Automatically turns off after 3 minutes of non-use other helpful message codes.

Specifications

MA871 MA872 MA873 MA881


Brix Fructose Glucose Invert Sugar
Range 0 to 85% Brix 0 to 85% mass 0 to 85% mass 0 to 85% mass
0 to 80°C / 32 to 176°F 0 to 80°C / 32 to 176°F 0 to 80°C / 32 to 176°F 0 to 80°C / 32 to 176°F
Resolution 0.1% Brix 0.1% 0.1% 0.1%
0.1°C / 0.1°F 0.1°C / 0.1°F 0.1°C / 0.1°F 0.1°C / 0.1°F
Accuracy ±0.2% Brix ±0.2% ±0.2% ±0.2%
±0.3°C / ±0.5°F ±0.3°C / ±0.5°F ±0.3°C / ±0.5°F ±0.3°C / ±0.5°F
Light source yellow LED yellow LED yellow LED yellow LED
Measurement Time approximately 1.5 seconds approximately 1.5 seconds approximately 1.5 seconds approximately 1.5 seconds
Minimum Sample Volume 100 μL (cover prism totally) 100 μL (cover prism totally) 100 μL (cover prism totally) 100 μL (cover prism totally)
Sample Cell SS ring and flint glass prism SS ring and flint glass prism SS ring and flint glass prism SS ring and flint glass prism
Temperature automatic between automatic between automatic between automatic between
Compensation 10 and 40°C / 50 to 104°F 10 and 40°C / 50 to 104°F 10 and 40°C / 50 to 104°F 10 and 40°C / 50 to 104°F
Case Material ABS ABS ABS ABS
Battery Type 1 x 9V AA (included) 1 x 9V AA (included) 1 x 9V AA (included) 1 x 9V AA (included)
Battery Life 5000 reading 5000 reading 5000 reading 5000 reading
Auto-shut off after 3 minutes of non-use after 3 minutes of non-use after 3 minutes of non-use after 3 minutes of non-use
Packaging dimensions 268 x 122 x 118 mm 268 x 122 x 118 mm 268 x 122 x 118 mm 268 x 122 x 118 mm
Packaging weight 660 g 663 g 666 g 644 g

Ordering Information Stainless Steel Sample


MA871, MA872, MA873 and MA881 are supplied in a carton box, Well and Prism
complete with 9V battery, pipette and instruction manual. Optionally you
Place a few drops of the sample in
can also order the refractometers in a hard carrying case (MA752).
the well and press the READ key.

48
­  
Measuring pH in meat
Using MW102 PRO+ pH portable meter
with a MA920B/1 pH electrode
The pH changes occurring in a carcass
during the first 24 hours after slaughter are
important for the quality of the final meat
or meat products. Protein denaturation will
occur if pH falls to too low a level or if a
relatively low pH sets in at a time after
slaughter where the carcass temperature is
still high. This will result in meat with poor
water holding capacity and in extreme cases
in meat that is PSE.

Calibrate the pH meter using pH 7 and pH 4


standardization buffers.

1. Cut meat sample into small pieces.

2. Weight approximatley 10 grams into a blender cup.


Run duplicates on each sample.

3. Add 100 ml of distilled deionized water and blend


for 30 seconds on high speed.

4. Transfer sample to a beaker. Read the pH as soon


as possible.

5. By pressing the HOLD key you can activate the


hold function. The measured value is frozen on the
display and the “HOLD” tag lights up. Release “HOLD”
by pressing HOLD key again.

6. Blender cups, beakers and stir bars can be rinsed


in distilled water between samples. The pH electrode
should be rinsed with distilled water between each
sample and periodically rinsed with acetone from a
squeeze bottle to remove fat buildup.

56