Service-Manual for DC/AC Power Supply

PeakTech® 5995
PeakTech® 5995 ADJUSTMENT INSTRUCTION
DC

RP2 CV Max. Output Voltage trim no load, output voltage knob turn
clockwise all to the end. Adjust to 32.0V
(on DC Regulation board)

RP4 CC Max. Output Current trim with load enough to cc state, output
current knob turn clockwise all to the
(on DC Regulation board)
end, Adjust to 6.10A

RP3 Output voltage display Track the output

(on Display board)

RP4 Output current display Track the output

(on Display board)

AC

RP1 Max. Output current trim With 6.30A load, output current knob
turn clockwise all to the end, OC LED
(on AC Mainboard) should be lighted and output will be
interrupted. With 6.00A load, output
should not be interrupted.

RP1 Output voltage display Track the output

(on Display board)

RP2 Output current display Track the output

(on Display board)