Ark 1
Ark 1
Model
Auto refractometer
                                                                  ARK-1s/AR-1s                                            ARK-1a/AR-1a                            ARK-1/AR-1                 Auto Ref/Keratometer   ARK-1s/1a/1
                                                                                                                                                                                                                     AR-1s/1a/1
 Measurement range                         Sphere -30.00 to +25.00 D (VD = 12 mm)
                                                    (0.01/0.12/0.25 D increments)
                                           Cylinder 0 to ±12.00 D (0.01/0.12/0.25 D increments)                                                                                                   Auto Refractometer
                                           Axis      0 to 180° (1°/5° increments)
 Minimum measurable pupil diameter         ø2 mm
Auto keratometer*1
 Measurement range                         Curvature radius 5.00 to 13.00 mm
                                                              (0.01 mm increments)
                                           Refractive power 25.96 to 67.50 D (n = 1.3375)
                                                              (0.01/0.12/0.25 D increments)
                                           Cylindrical power 0 to ±12.00 D
                                                              (0.01/0.12/0.25 D increments)
                                           Axis               0 to 180º (1°/5° increments)
  Sagittal measurement                     25º each from the center
                                           (superior side, inferior side, temporal side, nasal side)
VA measurement
 Measurement mode                          Uncorrected VA, Corrected VA (distance, near)
 Measurement range                         Less than 0.1, 0.1, 0.25, 0.32, 0.4, 0.5, 0.63, 0.8, 1.0, 1.25 or
                                           Less than 20/200, 20/200, 20/80, 20/60, 20/50,
                                                      20/40, 20/30, 20/25, 20/20, 20/16                                    Not available                         Not available
  Correction range                         Sphere -20.00 to +20.00 D (VD = 12 mm)
                                                     (0.25 D increments)
                                           Cylinder 0 to ±8.00 D (0.25 D increments)
                                           Axis      0 to 180º (1º/5º increments)
Vision comparison                          Available with VA chart                                                Available with scenery chart        Available*2 / Not available*3
Retroillumination image                    Available                                                                                                          Not available
Accommodation measurement range            0 to 10.00 D (0.01/0.12/0.25 D increments)                                                                         Not available
Pupillary distance measurement range       30 to 85 mm (1 mm increments)
                                           (Near point PD: 28 to 80 mm at WD = 40 cm)
Corneal size measurement range*1           10.0 to 14.0 mm (0.1 mm increments)
Pupil size measurement range               1.0 to 10.0 mm (0.1 mm increments)
Auto tracking                              X-Y-Z directions                                                                                                       Y direction
Auto shot                                  Available
Display                                    Tiltable 6.5-inch color LCD
Printer                                    Thermal line printer with easy loading and auto cutter
Interface                                  RS-232C: 2 ports
                                           USB, LAN: 1 port each
Power supply                               100 to 240 V AC, 50/60 Hz
Power consumption                          100 VA
Dimensions/mass                            260 (W) x 495 (D) x 457 (H) mm / 20 kg
                                           10.2 (W) x 19.5 (D) x 18.0 (H)" / 44 lbs.
Standard accessories                       Printer paper, Power cord, Dust cover, Chinrest paper,
                                           Fixing pins for chinrest paper, Spherical model eye with
                                           integrated contact lens holder*4
Optional accessories                       Eye Care card,Communication cable (RS-232C)
                                                                                                                                                                        ARK-1/AR-1_B01E009
The Superior Auto Ref/Keratometer
        and Auto Refractometer
                          What is the superior auto ref/keratometer and auto refractometer?
   The ARK-1/AR-1 series speaks for itself, surpassing conventional auto ref/keratometer
                   and auto refractometer with technologically enhanced functions
                    generating greater accuracy and greater clinical information.
                                                                                                      Fog under
                             VA                              Vision     Accommodation     Opacity                       Auto
           Model                              Glare test                                             astigmatism
                         measurement                       comparison    measurement    assessment                    tracking
                                                                                                      correction
        ARK-1s/AR-1s                                                                                               X-Y-Z directions
       ARK-1a/AR-1a                                                                                                X-Y-Z directions
         ARK-1/AR-1                                            *                                                     Y direction
        : Available, : Not available
      *Available with scenery chart for the ARK-1
       Not available for the AR-1
The Superior Auto Ref/Keratometer
        and Auto Refractometer
                          What is the superior auto ref/keratometer and auto refractometer?
   The ARK-1/AR-1 series speaks for itself, surpassing conventional auto ref/keratometer
                   and auto refractometer with technologically enhanced functions
                    generating greater accuracy and greater clinical information.
                                                                                                      Fog under
                              VA                             Vision     Accommodation     Opacity                       Auto
           Model                              Glare test                                             astigmatism
                          measurement                      comparison    measurement    assessment                    tracking
                                                                                                      correction
        ARK-1s/AR-1s                                                                                               X-Y-Z directions
       ARK-1a/AR-1a                                                                                                X-Y-Z directions
         ARK-1/AR-1                                            *                                                     Y direction
        : Available, : Not available
      * Available with scenery chart for the ARK-1
        Not available for the AR-1
   Accurate Refraction Measurement                                                                                                                     Easy VA Measurement with Glare Test
Large pupil zone imaging method                                                                                                                     VA measurement with built-in charts and lenses (available for the ARK-1s and AR-1s)
The large pupil zone imaging method enables the measurement of wider area refraction up to 6 mm diameter and can                                    The ARK-1s and AR-1s provides visual acuity (VA) measurement. This unique function enables a quick check of the
indicate the difference between the wide area refraction and central area refraction of up to 3.5 mm diameter. These                                patient's refractive error by comparing subjective measurement with objective measurement. Even corrected near visual
pupil diameters are measured simultaneously. The difference of the measurement allows assessment of the effect of                                   acuity is measurable with easy operability to determine the need for a progressive lens.
pupil size such as a vision in dim light.
                                                                                                                Central area: ø1 to 3.5 mm area
                               Central area: ø1 to 4 mm area                                                                                        Recall function for instant vision comparisons (available for the ARK-1s/1a/1 and AR-1s/1a)
                                                                                                                 Large area: ø1 to 6 mm area
                                                                                                                                                    The recall function provides the instant comparison between vision corrected with AR
                                                                                                                                                    data and uncorrected vision or vision corrected with the data of patient’s own glasses.
                                                                                                                                                    For the patient, this function demonstrates the difference in vision and the necessity
                                                                                                                                                    for more appropriate vision correction.
                                                                                                                                                    Contrast and glare test (available for the ARK-1s and AR-1s)
Super luminescent diode and highly sensitive CCD                                                                                                    After a low contrast VA chart is
                                                                                                                                                    presented, a glare source is illuminated
Incorporation of the super luminescent diode (SLD) provides a sharper and clearer image compared to a conventional
                                                                                                                                                    beside the chart and the glare VA can
LED. The highly sensitive CCD detects the ring image even if the fundus reflection is weak. The system, which combines
                                                                                                                                                    be measured. With this test the effect
the SLD and highly sensitive CCD, significantly improves measurement capability even in dense cataractous eyes.
                                                                                                                                                    of glare and halo to visual performance
                                                                                                                                                    can be measured. This utility can be
                                                                                                                                                    used for cataract and refractive surgery      Simulated patient’s vision of         Vision with glare and halo          Normal vision
                                                                                                                                                    patients.                                        low contrast VA chart          Simulated patient’s vision of VA chart and glare source
                                   Half
   Eye                             mirror
                                                                                                  SLD (normal eye)          LED (normal eye)
                                                Ring lens
    Reflective                                                                                                                                         Keratometry Measurement with Mire Ring (available for the ARK-1s/1a/1)
    mirror
                                              Highly                                                                                                   The mire ring is used to measure keratometry. It reduces eyelid artifacts.
                             SLD            sensitive           Hyperopia Emmetropia   Myopia
                                                CCD
Optimal fogging to minimize accommodation (available for the ARK-1s/1a and AR-1s/1a)                                                                                          Measurement with mire ring
Fogging is performed after correcting the patient’s astigmatism with built-in cylinder lenses. This allows the patient
to view the target clearly and minimizes the interference with accommodation even in high astigmatism.
   Accurate Refraction Measurement                                                                                                                     Easy VA Measurement with Glare Test
Large pupil zone imaging method                                                                                                                     VA measurement with built-in charts and lenses (available for the ARK-1s and AR-1s)
The large pupil zone imaging method enables the measurement of wider area refraction up to 6 mm diameter and can                                    The ARK-1s and AR-1s provide visual acuity (VA) measurement. This unique function enables a quick check of the
indicate the difference between the wide area refraction and central area refraction of up to 3.5 mm diameter. These                                patient's refractive error by comparing subjective measurement with objective measurement. Even corrected near visual
pupil diameters are measured simultaneously. The difference of the measurement allows assessment of the effect of                                   acuity is measurable with easy operability to determine the need for a progressive lens.
pupil size such as a vision in dim light.
                                                                                                                Central area: ø1 to 3.5 mm area
                               Central area: ø1 to 4 mm area                                                                                        Recall function for instant vision comparisons (available for the ARK-1s/1a/1 and AR-1s/1a)
                                                                                                                 Large area: ø1 to 6 mm area
                                                                                                                                                    The recall function provides the instant comparison between vision corrected with AR
                                                                                                                                                    data and uncorrected vision or vision corrected with the data of patient’s own glasses.
                                                                                                                                                    For the patient, this function demonstrates the difference in vision and the necessity
                                                                                                                                                    for more appropriate vision correction.
                                                                                                                                                    Contrast and glare test (available for the ARK-1s and AR-1s)
Super luminescent diode and highly sensitive CCD                                                                                                    After a low contrast VA chart is
                                                                                                                                                    presented, a glare source is illuminated
Incorporation of the super luminescent diode (SLD) provides a sharper and clearer image compared to a conventional
                                                                                                                                                    beside the chart and the glare VA can
LED. The highly sensitive CCD detects the ring image even if the fundus reflection is weak. The system, which combines
                                                                                                                                                    be measured. With this test the effect
the SLD and highly sensitive CCD, significantly improves measurement capability even in dense cataractous eyes.
                                                                                                                                                    of glare and halo to visual performance
                                                                                                                                                    can be measured. This utility can be
                                                                                                                                                    used for cataract and refractive surgery      Simulated patient’s vision of         Vision with glare and halo          Normal vision
                                                                                                                                                    patients.                                        low contrast VA chart          Simulated patient’s vision of VA chart and glare source
                                   Half
   Eye                             mirror
                                                                                                  SLD (normal eye)          LED (normal eye)
                                                Ring lens
    Reflective                                                                                                                                         Keratometry Measurement with Mire Ring (available for the ARK-1s/1a/1)
    mirror
                                              Highly                                                                                                   The mire ring is used to measure keratometry. It reduces eyelid artifacts.
                             SLD            sensitive           Hyperopia Emmetropia   Myopia
                                                CCD
Optimal fogging to minimize accommodation (available for the ARK-1s/1a and AR-1s/1a)                                                                                          Measurement with mire ring
Fogging is performed after correcting the patient’s astigmatism with built-in cylinder lenses. This allows the patient
to view the target clearly and minimizes the interference with accommodation even in high astigmatism.
    Patient-friendly Accommodation Measurement                                                                                                                           Practical and User-friendly Features
    (available for the ARK-1s/1a and AR-1s/1a)
The retroillumination image enables the observation of opacity of the optical media
of the eye. NIDEK cataract indices indicate the severity of the opacity and help to
assess the progression of pathology.
                                                                                                                                                                                                                                            Intelligent                               Auto
  COI.H        Opacity size within a diameter of 3 mm of the center (vertical diameter): mm                                                                                                                                                  Refractor                              Lensmeter
COI.A Opacity proportion within a diameter of 3 mm of the center: % Eye with dense opacity
The retroillumination image enables the observation of opacity of the optical media
of the eye. NIDEK cataract indices indicate the severity of the opacity and help to
assess the progression of pathology.
                                                                                                                                                                                                                                            Intelligent                               Auto
  COI.H        Opacity size within a diameter of 3 mm of the center (vertical diameter): mm                                                                                                                                                  Refractor                              Lensmeter
COI.A Opacity proportion within a diameter of 3 mm of the center: % Eye with dense opacity
ARK-1/AR-1_B01E009