■ STANDARD SPECIFICATIONS
  Refer to GS 01C22T02-00E for FOUNDATION Fieldbus
  communication type and GS 01C22T03-00E for                                       Load capacitance; 0.22 µF or less.
  PROFIBUS PA communication type marked with “.”                                  Load inductance; 3.3 mH or less.
Measurement Ranges :                                                               Spacing from power line; 15 cm or more.
                                                                                   Input impedance of receiver connected receiving resistance;
                                                     Measurement                      10 k or more at 2.4 kHz
     Capsule            Measurement Span                Range                        (See Optional Specifications for Intrinsically safe type)
       A   *1               0.06 to 3 MPa             -0.1 to 3 MPa              Accuracy:
                         {0.6 to 30 kgf/cm2}        {-1 to 30 kgf/cm2}            See Table 2.
                            0.46 to 14 MPa           -0.1 to 14 MPa              Ambient Temperature Limits:
     EJA438W *1
                         {4.6 to 140 kgf/cm2}      {-1 to 140 kgf/cm2}            -40 to 60 ˚C (-40 to 140 ˚F) (general-use type)
 B
                            0.46 to 7 MPa             -0.1 to 7 MPa               -30 to 60 ˚C (-22 to 140 ˚F) (with integral indicator)
      EJA438N *1
                         {4.6 to 70 kgf/cm2}        {-1 to 70 kgf/cm2}               (Note: The ambient temperature limits must be within the fill
                                                                    T01E.EPS
                                                                                            fluid operating temperature range, see table 1.)
     *1:        Measurement range is within the flange rating.                     (See ‘Optional Specifications’ for Explosion-protected
                                                                                   types)
Output Signal “”:
 4 to 20 mA DC, 2-wire system with digital communication                         Ambient Temperature Effect:
                                                                                  See Table 2.
Failure Alarm:
  Output status at CPU failure and hardware error;                               Process Temperature Limits:
    Up-scale: 110%, 21.6 mA DC or more(standard)                                   See Table 1.
    Down-scale: -5%, 3.2 mA DC or less                                             (See ‘Optional Specifications’ for Explosion-protected
                 -2.5%, 3.6 mA DC or less (Optional code /F1)                      types)
     Note: Applicable for Output signal code D and E                             Ambient Humidity Limits:
Supply Voltage “”:                                                               5 to 100 % R.H.(at 40 ˚C)
  10.5 to 42 V DC for general use and flameproof type                            Working Pressure Limits:
  10.5 to 32 V DC for lightning protector (Optional code /A)                      2.7 kPa abs{20 mmHg abs}to max. measurement
  10.5 to 30 V DC for intrinsically safe, Type n,                                 rangevalue.
              nonincendive, or non-sparking type                                     (Note: The working pressure must be within flange rating
  10.5 to 28 V DC for TIIS intrinsically safe type                                          pressure.)
Conditions of Communication Line “”:                                              For atmospheric pressure or below, see Figure 2.
 Power supply voltage;                                                           Power Supply Effect “”:
    16.4 to 42 V DC                                                                0.005 %/V (21.6 to 32 V DC, 350 )
 Load resistance; See Figure 1.
     Note: In case of an intrinsically safe transmitter, external
                                                                                 Mounting:
           load resistance includes safety barrier resistance.                    Transmitter; 2-inch pipe mounting
  Communication distance;                                                         Diaphragm seal; flange mounting
    2 km, when CEV polyethylene-insulated                                        Mounting Flange Rating:
    PVC-sheathed control cables are used.                                         See ‘Model and Suffix Codes.’
     Note: Communication distance varies depends on kind of                       Flange conforming to ANSI are serrated on the gasket
           cable.                                                                 surfaces (ANSI B16.5)
                                                                                     Note: For model EJA438W wetted parts material code H, T,
                                                                                           or U, no serration is worked.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                         GS 01C22J03-00E    April 01, 2013-00
<<Contents>> <<Index>>                                                                                                                                      3
1 {7.5}
                                                                                          0.1 {0.75}
                                                                                                    -50 0 50 100 150 200 250 300
                                                                                                        Process Temperature (C)                       F02E.EPS
                                                         Pref                                                             Pref
                       Zero shift      ± (0.2 + 0.5      x ) % / 50 C                                ± (0.3 + 0.6       x ) % / 50 C
   Ambient
   temperature
   effect *2                           ± 1.4 % / 50 C                          For x ≥ Pref           ± 1.6 % / 50 C                        For x ≥ Pref
                       Total shift                    Pref                                                            Pref
                                       ± (0.7 + 0.7  x ) % / 50 C             For x < Pref           ± (0.8 + 0.8  x ) % / 50 C           For x < Pref
                                                                                                                                                        T07E.EPS
    *1:       ‘x’ is the highest value among the absolute value of the lower range value (LRV) and higher range value (HRV), and the span value in
              a calibration range.
              Percent of span = Percent of x 
                                                   x
                                                  span
    *2:       In case of 6 to 10 m. for capillary length, value of ambient temperature effect is twice that given in the table.
              The ambient temperature effect specification applies to the range of 0 to 60 C (32 to140 F).
              (Below 0 C (32 F), value is three times that given in the table.)
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                     GS 01C22J03-00E    Mar. 23, 2006-00
<<Contents>> <<Index>>                                                                                                                                                                4
■ MODEL AND SUFFIX CODES
● Model EJA438W [Flange size: 2-inch (50 mm) and 3-inch (80 mm)]
            Model                                     Suffix Codes                                                                  Description
         EJA438W                   ..................................                                  Diaphragm Sealed Gauge Pressure transmitter (Flush diaphragm type)
       Output Signal              -D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   4 to 20 mA DC with digital communication (BRAIN protocol)
                                  -E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   4 to 20 mA DC with digital communication (HART protocol) *1
                                  -F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   Digital communication (FOUNDATION Fieldbus protocol) *6
                                  -G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   Digital communication (PROFIBUS PA protocol) *10
       Measurement span               A.............................                                   0.06 to 3 MPa {0.6 to 30 kgf/cm2}{8.6 to 430 psi}{0.6 to 30 bar}
       (capsule)                      B.............................                                   0.46 to 14 MPa {4.6 to 140 kgf/cm2}{66 to 2000 psi}{4.6 to 140 bar}
       Wetted parts material *8                                                                        [Diaphragm]           [Others]
                                         S# . . . . . . . . . . . . . . . . . . . . . . . . . . . .    JIS SUS316L           JIS SUS316L *11
                                         H# . . . . . . . . . . . . . . . . . . . . . . . . . . . .    Hastelloy C-276 *9    Hastelloy C-276 *9
                                         T............................                                 Tantalum              Tantalum
                                         U............................                                 Titanium              Titanium
       Process flange rating              J1 . . . . . . . . . . . . . . . . . . . . . . . . . .       JIS 10K
                                          J2 . . . . . . . . . . . . . . . . . . . . . . . . . .       JIS 20K
                                          J4 . . . . . . . . . . . . . . . . . . . . . . . . . .       JIS 40K
                                          J6 . . . . . . . . . . . . . . . . . . . . . . . . . .       JIS 63K
                                          A1 . . . . . . . . . . . . . . . . . . . . . . . . . .       ANSI Class 150        P1 . . . . . . . . . JPI Class 150
                                          A2 . . . . . . . . . . . . . . . . . . . . . . . . . .       ANSI Class 300        P2 . . . . . . . . . JPI Class 300
                                          A4 . . . . . . . . . . . . . . . . . . . . . . . . . .       ANSI Class 600        P4 . . . . . . . . . JPI Class 600
                                          D2 . . . . . . . . . . . . . . . . . . . . . . . . . .       DIN PN10/16
                                          D4 . . . . . . . . . . . . . . . . . . . . . . . . . .       DIN PN25/40
                                          D5 . . . . . . . . . . . . . . . . . . . . . . . . . .       DIN PN64
       Process flange                     A.........................                                  2-inch (50 mm) / JIS S25C
       size / material                        B.........................                               2-inch (50 mm) / JIS SUS304 *12
                                              C.........................                               2-inch (50 mm) / JIS SUS316 *13
                                              D.........................                               3-inch (80 mm) / JIS S25C
                                              E.........................                               3-inch (80 mm) / JIS SUS304 *12
                                              F.........................                               3-inch (80 mm) / JIS SUS316 *13
       Cover flange bolts material          A.......................                                  JIS SCM435
                                                  B.......................                             JIS SUS630
       Fill fluid                                                                                                                                  [Process temp.] [Ambient temp.]
                                                         -A *2 . . . . . . . . . . . . . . . . .      For general use (silicone oil)               -10 to 250 C    -10 to 60 C
                                                          -B . . . . . . . . . . . . . . . . .         For general use (silicone oil)               -30 to 180 C    -15 to 60 C
                                                          -C *3. . . . . . . . . . . . . . . . . . .   For high temperature use (silicone oil)      10 to 300 C     10 to 60 C
                                                          -D *4 . . . . . . . . . . . . . . . . .      For oil-prohibited use (fluorinated oil)     -20 to 120 C    -10 to 60 C
                                                          -E . . . . . . . . . . . . . . . . . . .     For low temperature use (ethylene glycol)    -50 to 100 C    -40 to 60 C
                                                              A .................               Always A
       Capillary length (m)                                      *5 . . . . . . . . . . . .   Specify capillary length from 1 to 10 m in  . (Example for 2 m : 02)
       Installation                                               -9 . . . . . . . . . . . . . .Horizontal impulse piping type, left side high pressure
       Electrical connection                                        0.............              G1/2 female, one electrical connection
                                                                2.............                 1/2 NPT female, two electrical connections without blind plug
                                                                    3.............              Pg 13.5 female, two electrical connections without blind plug
                                                                    4.............              M20 female, two electrical connections without blind plug
                                                                    5.............              G1/2 female, two electrical connections and a blind plug
                                                                    7.............              1/2 NPT female, two electrical connections and a blind plug
                                                                    8.............              Pg 13.5 female, two electrical connections and a blind plug
                                                                    9.............              M20 female, two electrical connections and a blind plug
                                                                    A.............              G1/2 female, two electrical connections and a SUS316 blind plug
                                                                    C.............              1/2 NPT female, two electrical connections and a SUS316 blind plug
                                                                    D.............              M20 female, two electrical connections and a SUS316 blind plug
       Integral indicator                                               D..........             Digital indicator
                                                                        E..........             Digital indicator with the range setting switch *7
                                                                       N..........             (None)
       Mounting bracket                                              A.........                JIS SECC               2-inch pipe mounting (flat type)
                                                                          B.........            JIS SUS304             2-inch pipe mounting (flat type)
                                                                          J.........            JIS SUS316             2-inch pipe mounting (flat type)
                                                                          N.........            (None)
       Optional codes                                                              / Optional specification
                                                                                                                                                                           T09E.EPS
    The ‘’ marks indicate the most typical selection for each specification. Example: EJA438W-DASA1AA-AA02-92NA/
    The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use of
    SUS316 material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for details.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                        GS 01C22J03-00E          Jul. 08, 2009-00
<<Contents>> <<Index>>                                                                                                                                5
    *1:    Refer to GS 01C22T01-00E for HART Protocol version.
    *2:    In case of Wetted parts material code T (Tantalum), the process temperature limit is -10 to 200 C.
    *3:    Wetted parts material code T (Tantalum) cannot be applied.
    *4:    Even in case where Fill fluid code D (fluorinated oil) is selected, if degrease cleansing treatment or both degrease cleansing and
           dehydrating treatment for the wetted parts is required, specify Optional code K1 or K5.
    *5:    In case of Wetted parts material code H (Hastelloy C), T (Tantalum), and U (Titanium) or Fill fluid code C (for high temperature
           use), specify capillary length from 1 to 5 m.
    *6:    Refer to GS 01C22T02-00E for Fieldbus communication.
    *7:    Not applicable for Output signal code F and G.
    *8:          Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of
                 inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to
                 plant facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken diaphragm and the
                 fill fluid can contaminate the user's process fluids.
                 Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium
                 hypochlorite, and high-temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted
                 parts material.
    *9:    Hastelloy C-276 or ASTM N10276.
    *10:   Refer to GS 01C22T03-00E for PROFIBUS PA communication.
    *11:   JIS SUS316L or ASTM grade 316L.
    *12:   JIS SUS304 or ASTM grade 304. Forged version may be used.
    *13:   JIS SUS316 or ASTM grade 316. Forged version may be used.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                            GS 01C22J03-00E        Jul. 08, 2009-00
<<Contents>> <<Index>>                                                                                                                                                                        6
● Model EJA438N [Flange size: 4-inch (100 mm)]
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                              GS 01C22J03-00E              Jul. 08, 2009-00
<<Contents>> <<Index>>                                                                                                                                                                                       7
● Model EJA438N [Flange size: 3-inch (80mm)]
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                                     GS 01C22J03-00E                Jul. 08, 2009-00
<<Contents>> <<Index>>                                                                                                                                8
■ OPTIONAL SPECIFICATIONS (For Explosion Protected Types “”)
    For FOUNDATION Fieldbus explosion protected type, see GS 01C22T02-00E.
    For PROFIBUS PA explosion protected type, see GS 01C22T03-00E.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                             GS 01C22J03-00E       April 01, 2013-00
<<Contents>> <<Index>>                                                                                                                                9
             Item                                                               Description                                                      Code
                                    CSA Explosionproof Approval *1 *3 *4
                                        Applicable standard: C22.2 No. 0, No. 0.4, No. 25, No. 30, No. 94, No. 142
                                        Certificate: 1089598
                                        Explosionproof for Class I, Division 1, Groups B, C and D
                                        Dustignitionproof for Class II/III, Division 1, Groups E, F and G
                                        Division2 ‘SEALS NOT REQUIRED’ , Temp. Class: T4, T5, T6 Encl Type 4x                                     CF1
                                        Max. Process Temp.: T4; 120C (248F), T5; 100C (212F), T6; 85C (185F)
                                        Amb. Temp.: –40 to 80C (–40 to 176F)
                                    Process Sealing Certification
                                        Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
  Canadian Standards                    No additional sealing required. Primary seal failure annunciation: at the zero adjustment screw
  Association (CSA)
                                    CSA Intrinsically safe Approval *1 *3 *4
                                        Applicable standard: C22.2 No. 0, No. 0.4, No. 25, No. 30, No. 94, No. 142, No. 157, No. 213
                                        Certificate: 1053843
                                        Class I, Groups A, B, C and D Class II and III, Groups E, F and G
                                        Encl Type 4x, Temp. Class: T4, Amb. Temp.: –40 to 60C (–40 to 140F)                                     CS1
                                        Vmax=30 V, Imax=165 mA, Pmax=0.9 W, Ci=22.5 nF, Li=730 H
                                    Process Sealing Certification
                                        Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
                                        No additional sealing required. Primary seal failure annunciation: at the zero adjustment screw
                                    Combined CF1 and CS1 *1 *3 *4                                                                                CU1
                                    IECEx Intrinsically safe, type n and Flameproof Approval *3 *4 *9
                                    Intrinsically safe and type n
                                         Applicable Standard: IEC 60079-0:2004, IEC 60079-11:1999, IEC 60079-15:2005,
                                         IEC 60079-26:2005
                                         Certificate: IECEx KEM 06.0007X
                                         Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP67
                                         Amb. Temp.: –40 to 60C (–40 to 140F), Max. Process Temp.: 120C (248F)
  IECEx                                  Electrical Parameters: [Ex ia] Ui=30 V, Ii=165 mA, Pi=0.9 W, Ci=22.5 nF, Li=730 H
                                                                                                                                                  SU2
                                                                [Ex nL] Ui=30 V, Ci=22.5 nF, Li=730 H
                                    Flameproof
                                         Applicable Standard: IEC 60079-0:2004, IEC60079-1:2003
                                         Certificate: IECEx KEM 06.0005
                                         Ex d IIC T6...T4 Enclosure: IP67
                                         Max.Process Temp.: T4;120C (248F), T5;100C (212F), T6; 85C (185F)
                                         Amb.Temp.: –40 to 75C (–40 to 167F) for T4, –40 to 80C (–40 to 176F) for T5,
                                                       –40 to 75C (–40 to 167F) for T6
                                    TIIS Flameproof Approval, Ex do IIC T4X *3 *5 *7 *8 *10
                                        Certificate: TC15296 (Without integral indicator)
                                                                                                                                                  JF3
                                                     TC15297 (With integral indicator)
  TIIS certification                    Amb. Temp.: –20 to 60C, Process Temp.: –20 to 120C
                                    TIIS Intrinsically safe Approval, Ex ia IIC T4 *6 *8
                                        Certificate: TC14632                                                                                      JS3
                                        Amb. Temp.: –20 to 60C, Process Temp.: –20 to 120C
  Attached flameproof               Electrical connection: G1/2 female                                           1 pc.                            G11
  packing adapter *5                Applicable cable: O. D. 8 to 12 mm                                           2 pcs.                           G12
                                                                                                                                                 T16E.EPS
    *1:    Applicable for Electrical connection code 2, 7 and C (1/2 NPT female).
    *2:    Applicable for Electrical connection code 2, 4, 7, 9, C and D (1/2 NPT and M20 female).
    *3:    Applicable for Output signal code D and E. For intrinsically safe approval, use the safety barrier certified by the testing laboratories
           (BARD-400 is not applicable).
    *4:    Lower limit of ambient temperature is –15˚C (5˚F) when /HE is specified.
    *5:    If cable wiring is to be used to a TIIS flameproof type transmitter, add the YOKOGAWA-assured flameproof packing adapter.
    *6:    Applicable for Output signal code D. See <Safety Barrier for TIIS Intrinsically Safe Type>.
    *7:    In case that the ambient temperature exceeds 50°C or that the ambient temperature exceeds 45°C with the process temperature of
           90°C or above, use heat-resistant cables with maximum allowable temperature of 75°C or above.
    *8:    TIIS (The Technology Institution of Industrial Safety) Certification is a new notation for the explosionproof approval in Japan
           instead of JIS.
    *9:    Applicable for Electrical connection code 2, 4, 7, C and D (1/2 NPT and M20 female).
    *10:   Not applicable for Electrical connection code A, C and D.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                             GS 01C22J03-00E       April 23, 2012-00
<<Contents>> <<Index>>                                                                                                                                               10
■ OPTIONAL SPECIFICATIONS
                           Item                                                              Description                                                  Code
                                                  Amplifier cover only                                                                                     P
                               Color change
           Painting *8                            Amplifier cover and terminal cover, Munsell 7.5 R4/14                                                    PR
                               Coating change     Epoxy resin-baked coating *9                                                                             X1
                                                  Exterior parts on the amplifier housing(name plates, tag plate, zero-adjustment screw,
           316 SST exterior parts                                                                                                                          HC
                                                  stopper screw) will become 316 SST *10
           Fluoro-rubber O-ring                   All O-rings of amplifier housing. Lower limit of ambient temperature: –15°C (5°F)    *11                 HE
                                                  Transmitter power supply voltage : 10.5 to 32 V DC ( 10.5 to 28 V DC for TIIS intrinsically safe
                                                  type, 10.5 to 30 V DC for intrinsically safe type other than TIIS, or 9 to 32 V DC for Fieldbus
           Lightning protector                    communication type.)                                                                                      A
                                                  Allowable current : Max. 6000 A (140 s), Repeating 1000 A (140 s) 100 times
           Oil-prohibited use                     Degrease cleansing treatment                                                                             K1
           Oil-prohibited use
                                                  Degrease cleansing and dehydrating treatment                                                             K5
           with dehydrating treatment
           No serration   *1                      No serration work on the flange gasket surface ( for ANSI flange only )                                   Q
           Teflon film                            Diaphragm protection from sticky process fluid by FEP Teflon film attached with fluorinated oil.
                                                                                                                                                            T
                                                  Operation range: 20 to 150°C, 0 to 2 MPa(Not applicable for vacuum service).
           Operating temperature correction *2    Adjusting range : 80°C to Maximum temperature of specified fill fluid                                     R
                                                  Update time: 0.125 sec
           Fast response *6                       Amplifier assembly damping time constant: 0.1 to 64 sec in 9 increments.                                 F1
                                                  Response time (with min. damping time constant): max. 0.3 sec (excluding diaphragm seal units)
           Failure alarm down-scale *3            Output status at CPU failure and hardware error is –5%, 3.2 mA or less.                                  C1
                                                                                 Failure alarm down-scale: output status at CPU failure and
                                                                                                                                                           C2
                                                  Output signal limits:          hardware error is –5%, 3.2 mA or less.
           NAMUR NE43 compliant *3 *7             3.8 mA to 20.5 mA
                                                                                 Failure alarm up-scale: output status at CPU failure and
                                                                                                                                                           C3
                                                                                 hardware error is 110%, 21.6 mA or more.
           Data configuration at factory          Description into “Descriptor” parameter of HART protocol                                                 CA
                                                  Amplifier housing material : JIS SCS14A stainless steel
           Stainless steel amplifier housing *4                                                                                                            E1
                                                  (equivalent to JIS SUS316 cast stainless steel or ASTM CF-8M)
           Gold-plate                             Inside of isolating diaphragms(fill fluid side) are gold plated, effective for hydrogen permeation.
                                                                                                                                                           A1
                                                  ( applicable for Wetted parts material code S and H )
           Wired tag plate                        Stainless steel tag plate wired onto transmitter                                                         N4
           Capillary without PVC sheaths          When ambient temperature exceeds 100°C, or use of PVC is prohibited                                       V
                                                  P calibration ( psi unit )                                                                               D1
           Calibration units *5                   bar calibration ( bar unit )                                                 (See Table 4.)              D3
                                                  M calibration (kgf/cm2 unit )                                                                            D4
                                                  Sealant(liquid silicone rubber) is coated on JIS SUS630 cover flange mounting nuts against stress
           Sealing treatment to JIS SUS630 nuts                                                                                                             Y
                                                  corrosion cracking.
                                                                                                                                                          T13E.EPS
    *1:      This item cannot be applied to model EJA438W Wetted part material code H, T, or U, whose gasket contact surface are not serrated
             as standard specifications.
    *2:      Specify the process operating temperature for zero correction. Example: Zero correction by process temperature 90C.
    *3:      Applicable for Output signal code D and E. The hardware error indicates faulty amplifier or capsule.
             When combining with Option code F1, output status for down-scale is –2.5%, 3.6 mA DC or less.
    *4:      Applicable for Electrical connection code 2, 3, 4, A, C and D. Not applicable for Option code P, X1, and JF3.
    *5:      The unit of MWP (Max. working pressure) on name plate of a housing is the same unit as specified by Option code D1, D3, and D4.
    *6:      Applicable for Output signal code D and E. Write protection switch is attached for Output code E. Not applicable for Integral
             indicator code E.
    *7:      Not applicable for Option code C1.
    *8:      Standard polyurethan painting can be used in acid atmosphere, whereas the epoxy resin-baked coating (Option code X1) can be used
             in alkaline atmosphere. Anti-corrosion coating, the combination of polyurethan and epoxy resin-baked coating, is available by
             special order as sea water, alkaline, and acid resistant.
    *9:      Not applicable for color change option.
    *10:     316 or 316L SST. The specification is included in option code /E1. Not applicable with option code /JF3.
    *11:     Not applicable with option code /JF3.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                          GS 01C22J03-00E           July 01, 2011-00
<<Contents>> <<Index>>                                                                                                                                                                11
                                                For A-capsule
                                                                JIS 40K                   3 MPa {30    kgf/cm2}                                                                     T43
                                                                JIS 63K                   3 MPa {30 kgf/cm2}                 EJA438W                                                T45
                                                                ANSI/JPI Class 150        3 MPa {29.8          kgf/cm2}                                                             T46
                                                                                                                             EJA438W/EJA438N
                                                                ANSI/JPI Class 300        3 MPa {30 kgf/cm2}                                                                        T47
                                                                ANSI/JPI Class 600        3 MPa {30 kgf/cm2}                 EJA438W                       Nitrogen (N2) Gas *2     T49
                                                                JIS 10K                   2 MPa {20 kgf/cm2}                                                                        T31
                                                                                                                              EJA438W/EJA438N              Retention time:
                                                                JIS 20K                   5 MPa {50 kgf/cm2}                                                        10 minutes      T32
                                                                JIS 40K                   10 MPa {100 kgf/cm2}               EJA438W                                                T33
                                                For B-capsule
                Span       8.6 to 430 psi             0.6 to 30 mbar         0.6 to 30 kgf/cm2                       Span         8.6 to 430 psi    0.6 to 30 mbar   0.6 to 30 kgf/cm2
            A                                                                                                  A
  EJA438W
EJA438N
Range -15 to 430 psi -1 to 30 mbar -1 to 30 kgf/cm2 Range -15 to 430 psi -1 to 30 mbar -1 to 30 kgf/cm2
                Span       66 to 2000 psi          4.6 to 140 mbar          4.6 to 140 kgf/cm2                       Span         66 to 1000 psi    4.6 to 70 mbar   4.6 to 70 kgf/cm2
            B                                                                                                   B
                Range      -15 to 2000 psi           -1 to 140 mbar         -1 to 140 kgf/cm2                        Range        -15 to 1000 psi    -1 to 70 mbar   -1 to 70 kgf/cm2
                                                                                                                                                                                  T12E.EPS
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                                GS 01C22J03-00E      Jul. 08, 2009-00
<<Contents>> <<Index>>                                                                                                                                                                                 12
■ DIMENSIONS
● Model EJA438W
                                                                                                                                                                Unit : mm (approx. inch)
                                                                                        25 (0.98)
                                                                                                                                t
                                               ød *1                                                                        f                                34
                                                                                                                                                           (1.34)
øg
                                                                                                                                         øD
                                                                                                                                øC
                                             n-øh
                                              140(5.51)
                                                                                                                                                               Wetted parts material
                                                                                                                                         110 (4.33)             code U (Titanium)
               External indicator
                                           94(3.70)
                                                                        Conduit
               conduit connection                                       connection                                                    12
               (Optional)                                                                                                           (0.47)
                                                                                                                                                          Terminal
                                                                          Internal                                                                        side
                                                                                                                       (3.07)
                                                                          indicator
                                                                                                                        ø78
                                                                          (Optional)
                                                                                                                                                                     146(5.75)
                                                                                        197(7.76)
                                                                                                                                                          Ground
                                                                                                                                                          terminal
                 333(13.11)
                                                                                                                                                               Open to
                                                                                                         Zero                                                  atmosphere
                              (4.88)
                                                                                                         adjustment                                            (ø5mm)
                               124
                                                       92
                                                     (3.62)
                                                                                 Shrouding bolt *2
                              (1.85)
                                47
                                                        2-inch pipe
                                                       (O.D. 60.5mm)                                Mounting bracket
                                                                                                    (Flat-type, Optional)
              *1: Indicates inside diameter of gasket contact surface.
              *2: Applicable only for ATEX, IECEx, and TIIS Flameproof type.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                                             GS 01C22J03-00E            Mar. 23, 2006-00
<<Contents>> <<Index>>                                                                                                                                                                                13
● Model EJA438N
                                                                                                                                               14 (0.55)
                                                                                 X2
                                                                                                                                                           f
                                                                                                                                                                t
                                                                          120                                               ø30
                                               140(5.51)                (4.72)                                              (1.18)          n-øh
          External indicator
                                             94(3.70)                                                                                     110 (4.33)
                                                                           Conduit
          conduit connection                                               connection                                                   12
          (Optional)                                                                                                                  (0.47)
                                                                                                                                                                    Terminal
                                                                            Internal                                                                                side
                                                                                                                          (3.07)
                                                                            indicator
                                                                                                                           ø78
                                                                            (Optional)
                                                                                                                                                                               146(5.75)
                                                                                         197(7.76)
                                                                                                                                                                Ground
                                                                                                                                                                terminal
            333(13.11)
                                                                                                                                                                      Open to
                                                                                                        Zero                                                          atmosphere
                          (4.88)
                                                                                                        adjustment                                                    (ø5mm)
                           124
                                                         92
                                                       (3.62)
                                                                                   Shrouding bolt *1
                          (1.85)
                            47
                                                          2-inch pipe
                                                         (O.D. 60.5mm)                               Mounting bracket
                                                                                                     (Flat-type, Optional)
                     *1: Applicable only for ATEX, IECEx, and TIIS Flameproof type.                                       Diaphramg extension length code
                                                                                                                             2 : X2 = 50 mm (2 inch)
                                                                                                                             4 : X2 = 100 mm (4 inch)
                                                                                                                             6 : X2 = 150 mm (6 inch)
                               Process flange size : 4 inch (100 mm)
                           Flange Rating            øD            øC             øg                  øA               t               f*       n               øh
                            JIS 10K             210(8.72)     175(6.89)      155(6.10) 96±0.5(3.78)              18(0.71)             0        8           19(0.75)
                            JIS 20K             225(8.86)     185(7.28)      155(6.10) 96±0.5(3.78)              24(0.94)             0        8           23(0.91)
                            JIS 40K             250(9.84)     205(8.07)      155(6.10) 96±0.5(3.78)              36(1.42)             0        8           25(0.98)
                            ANSI Class 150 228.6(9.00) 190.5(7.50) 155(6.10) 96±0.5(3.78) 23.9(0.94) 1.6(0.06)                                 8           19.1(0.75)
                            ANSI Class 300 254(10.00) 200.2(7.88) 155(6.10) 96±0.5(3.78) 31.8(1.25) 1.6(0.06)                                  8           22.4(0.88)
                            JPI Class 150       229(9.02) 190.5(7.50) 155(6.10) 96±0.5(3.78)                     24(0.94)          1.6(0.06)   8           19(0.75)
                            JPI Class 300      254(10.00) 200.2(7.88) 155(6.10) 96±0.5(3.78)                     32(1.26)          1.6(0.06)   8           22(0.87)
                            DIN PN 10/16        220(8.66)     180(7.09)      155(6.10) 96±0.5(3.78)              20(0.79)             0        8           18(0.71)
                            DIN PN 25/40        235(9.25)     190(7.48)      155(6.10) 96±0.5(3.78)              24(0.94)             0        8           22(0.87)
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                                                                             GS 01C22J03-00E           Mar. 23, 2006-00
<<Contents>> <<Index>>                                                                                                                14
  ● Terminal Configuration                                             ● Terminal Wiring
                                                                                       +
                                                                          SUPPLY               Power supply and output terminal
                                                                                       –
< Ordering Information > “”                                           < Safety Barrier for TIIS Intrinsically Safe Type >
Specify the following when ordering                                         Supplier        Type               Model
1. Model, suffix codes, and optional codes                                                            MTL3046B
2. Calibration range and units:                                                MTL
                                                                                                      MTL4041B
  1) Calibration range can be specified with range value                                   Isolator
  specifications up to 5 digits (excluding any decimal point)                                         KFD2-STC3-Ex 1
                                                                               P+F
  for low or high range limits within the range of -32000 to                                          KFD2-STV3-Ex 1-1, 2, 3
                                                                                                                               T001E.EPS
  32000.
  2) Specify only one unit from the table ‘Settings when                   Note: Requirements of capacitance and inductance for cable
  shipped.’                                                                      wiring.
3. Select normal or reverse for operation mode                                      Cw ≤ Co – 11[nF]
    Note: If not specified, the instrument is shipped in normal                     Lw ≤ Lo – 730[µH]
          operation mode.                                                           (Co: Max. external capacitance)
4. Display scale and units (for transmitters equipped with                          (Lo: Max. external inductance)
  integral indicator only)
  Specify either 0 to 100 % or engineering unit scale and              < Reference >
  ‘Range and Unit’ for engineering units scale :                        1. JIS SUS316L stainless steel; Equivalent to AISI 316L.
  Scale range can be specified with range limit specifications          2. JIS SUS316 stainless steel; Equivalent to AISI 316.
  up to 5 digits (excluding any decimal point) for low or               3. JIS SUS304 stainless steel; Equivalent to AISI 304.
  high range limits within the range of -19999 to 19999.                4. JIS S25C carbon steel; Equivalent to AISI 1025.
5. Tag Number (if required)                                             5. JIS SECC; Carbon steel.
6. Process fluid temperature for zero compensation (if                  6. Teflon; Trademark of E.I. duPont de Nemours & Co.
  required)                                                             7. JIS SCM435 chrome molybdenum steel; Equivalent to
                                                                          AISI 4137.
< Related Instruments > “”                                             8. JIS SUS630 stainless steel; Equivalent to ASTM 630.
  Power Distributor: Refer to GS 01B04T01-02E or                        9. Hastelloy; Trademark of Haynes International Inc.
                     GS 01B04T02-02E                                    10. JIS SCS14A stainless steel; Equivalent to JIS SUS316
  BRAIN TERMINAL: Refer to GS 01C00A11-00E                                cast stainless steel or ASTM CF-8M.
                                                                        11. HART; Trademark of the HART Communication
                                                                          Foundation.
                                                                        12. FOUNDATION; Trademark of Fieldbus Foundation.
                                                                        13. PROFIBUS; Registered trademark of Profibus
                                                                          Nutzerorganisation e.v., Karlsruhe, Germany.
                                                                        14. Other company names and product names used in this
                                                                          material are registered trademarks or trademarks of their
                                                                          respective owners.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation                                  GS 01C22J03-00E   June 01, 2007-00
Subject to change without notice.