General                                       EJA130E
Specifications                                Differential Pressure Transmitter
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                            GS 01C31B04-01EN     Feb. 10, 2017-00
<<Contents>> <<Index>>                                                                                                              3
Burst Pressure Limits                                                     Supply & Load Requirements
  132 MPa (19100 psi)                                                      (Output signal code D and J. Optional features
Self Diagnostics                                                           or approval codes may affect electrical
  CPU failure, hardware failure, configuration error,                      requirements.)
  and over-range error for differential pressure, static                   With 24 V DC supply, up to a 550Ω load can be
  pressure and capsule temperature.                                        used. See graph below.
  User-configurable process high/low alarm for
  differential pressure and static pressure is also
  available.
Signal Characterizer                                                                600
  (Output signal code D, J and Q)                                                               E-10.5
  User-configurable 10-segment signal characterizer                                        R=
                                                                            External            0.0244              Digital
  for 4 to 20 mA output.                                                    load                                 Communication
SIL Certification                                                           resistance                               range
  EJA-E series transmitters except Fieldbus,                                                                    BRAIN and HART
  PROFIBUS PA and 1-5V DC with HART (Low Power)                                  250
  communication types are certified in compliance with
  the following standards;
  IEC 61508: 2010; Part1 to Part 7                                          R (Ω)
  Functional Safety of Electrical/electronic/
  programmable electronic related systems; SIL 2
  capability for single transmitter use, SIL 3 capability                                    10.5 16.6         25.2                42
  for dual transmitter use.                                                                Power supply voltage E (V DC)
                                                                                                                               F02E.ai
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                               GS 01C31B04-01EN   Feb. 10, 2017-00
<<Contents>> <<Index>>                                                                                                          4
	 Load inductance                                                      Weight
  3.3 mH or less                                                        [Installation code 7, 8 and 9]
	 Input impedance of communicating device                               6.8 kg (14.3 lb) without integral indicator, mounting
  10 kΩ or more at 2.4 kHz.                                             bracket, and process connector.
                                                                        Add 1.5 kg (3.3 lb) for amplifier housing code 2.
EMC Conformity Standards
  EN 61326-1 Class A, Table2                                           Connections
  EN 61326-2-3                                                          Refer to “MODEL AND SUFFIX CODES.”
  EN 61326-2-5 (for fieldbus)                                           Process Connection of Cover Flange: IEC61518
European Pressure Equipment Directive
  2014/68/EU                                                           < Related Instruments>
  Sound Engineering Practice                                             FieldMate Versatile Device Management Wizard:
                                                                            Refer to GS 01R01A01-01E.
  With option code /PE3                                                  BRAIN TERMINAL: Refer to GS 01C00A11-00E
                                                                         Power Distributor: Refer to GS 01B04T01-02E or
                                                                            GS 01B04T02-02E
 Category III, Module H, Type of Equipment: Pressure
 Accessory-Vessel, Type of Fluid: Liquid and Gas,
 Group of Fluid: 1 and 2                                               < Reference >
                                                                         •	            ; Registered trademark of Yokogawa
EU RoHS Directive
                                                                             Electric Corporation.
 EN 50581
                                                                         •	 FieldMate; Registered trademark of Yokogawa
Safety Requirement Standards                                                 Electric Corporation.
 EN 61010-1, C22.2 No.61010-1                                            •	 Teflon; Trademark of E.I. DuPont de Nemours &
 •	 Installation category: I                                                 Co.
    (Anticipated transient overvoltage 330 V)                            •	 Hastelloy; Trademark of Haynes International Inc.
 •	 Pollution degree: 2                                                  •	HART®: Registered trademark of FieldComm
 •	 Indoor/Outdoor use                                                       Group.
                                                                         •	FOUNDATION Fieldbus; Trademark of FieldComm
□	 PHYSICAL SPECIFICATIONS                                                   Group.
Wetted Parts Materials                                                   •	 PROFIBUS; Registered trademark of Profibus
                                                                             Nutzerorganisation e.v., Karlsruhe, Germany.
	 Diaphragm, cover flange, process connector,
                                                                         Other company names and product names used in
  capsule gasket, and vent/drain plug
                                                                         this material are registered trademarks or trademarks
  Refer to “MODEL AND SUFFIX CODES.”
                                                                         of their respective owners.
	 Process connector O-ring
  Fluorinated rubber
Non-wetted Parts Materials
	 Bolting
  B7 carbon steel, 316L SST or 660 SST
	 Housing
  •	 Low copper cast aluminum alloy
  •	 Low copper cast aluminum alloy with corrosion
      resistance properties (copper content ≤ 0.03%,
      iron content ≤ 0.15%) (optional)
  •	 ASTM CF-8M Stainless steel (optional)
	 Coating of housing
  [for aluminum housing]
  Urethane curing type polyester resin powder coating
  Mint-green paint (Munsell 5.6BG 3.3/2.9 or its
  equivalent)
  [for option code /P or /X2]
  Epoxy and polyurethane resin solvent coating
	 Degrees of protection
  IP66/IP67, Type 4X
	 Cover O-rings
  Buna-N, fluoro-rubber (optional)
	 Name plate and tag
  316 SST
	 Fill fluid
  Silicone, fluorinated oil (optional)
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                           GS 01C31B04-01EN   April 10, 2018-00
<<Contents>> <<Index>>                                                                                                                                                  5
	MODEL AND SUFFIX CODES
      Model                                   Suffix Codes                                                                      Description
    EJA130E               ...............................                                   Differential pressure transmitter
Output signal            -D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     4 to 20 mA DC with digital communication (BRAIN protocol)
                         -J . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     4 to 20 mA DC with digital communication (HART 5/HART 7 protocol)*1
                         -F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .     Digital communication (FOUNDATION Fieldbus protocol, refer to
                                                                                                 GS 01C31T02-01EN)
                         -G  . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      Digital communication (PROFIBUS PA protocol, refer to
                                                                                                 GS 01C31T04-01EN)
                   -Q  . . . . . . . . . . . . . . . . . . . . . . . . . . . . .            Low Power, 1 to 5 V DC with digital communication (HART 7 protocol)
Measurement           M . . . . . . . . . . . . . . . . . . . . . . . . . . .               1 to 100 kPa (4 to 400 inH2O)
span (capsule)        H . . . . . . . . . . . . . . . . . . . . . . . . . . .               5 to 500 kPa (20 to 2000 inH2O)
Wetted parts                S  . . . . . . . . . . . . . . . . . . . . . . . . .            Refer to “Wetted Parts Material” Table below.
material *2
Process connections              3  . . . . . . . . . . . . . . . . . . . . . .             with 1/4 NPT female process connector*3
                                 4  . . . . . . . . . . . . . . . . . . . . . .             with 1/2 NPT female process connector*3
                           ► 5  . . . . . . . . . . . . . . . . . . . . . .                 without process connector (1/4 NPT female on the cover flanges)
Bolts and nuts materia                 J  . . . . . . . . . . . . . . . . . . .             B7 carbon steel
                                       G . . . . . . . . . . . . . . . . . . .              316L SST
                                       C . . . . . . . . . . . . . . . . . . .              660 SST
Installation                                -7 . . . . . . . . . . . . . . . . .            Vertical piping, left side high pressure, and process connection downside
                                            -8 . . . . . . . . . . . . . . . . .            Horizontal piping and right side high pressure
                                      ► -9 . . . . . . . . . . . . . . . . .                Horizontal piping and left side high pressure
                                            -U . . . . . . . . . . . . . . . . .            Universal flange
Amplifier housing                                 1  . . . . . . . . . . . . . .            Cast aluminum alloy
                                                  3  . . . . . . . . . . . . . .            Cast aluminum alloy with corrosion resistance properties*4
                                                  2  . . . . . . . . . . . . . .            ASTM CF-8M stainless steel*5
Electrical connection                                  0  . . . . . . . . . . .             G1/2 female, one electrical connection without blind plugs
                                                 ► 2  . . . . . . . . . . .                 1/2 NPT female, two electrical connections without blind plugs
                                                       4  . . . . . . . . . . .             M20 female, two electrical connections without blind plugs
                                                       5  . . . . . . . . . . .             G1/2 female, two electrical connections and a blind plug*6
                                                       7  . . . . . . . . . . .             1/2 NPT female, two electrical connections and a blind plug*6
                                                       9  . . . . . . . . . . .             M20 female, two electrical connections and a blind plug*6
                                                       A . . . . . . . . . . .              G1/2 female, two electrical connections and a SUS316 blind plug
                                                       C . . . . . . . . . . .              1/2 NPT female, two electrical connections and a SUS316 blind plug
                                                       D . . . . . . . . . . .              M20 female, two electrical connections and a SUS316 blind plug
Integral indicator                                           D . . . . . . . .              Digital indicator*7
                                                             E  . . . . . . . .             Digital indicator with the range setting switch (push button)*8
                                                      ► N . . . . . . . .                   None
Mounting bracket	                                        ► B . . . . . .                    304 SST 2-inch pipe mounting, flat type (for horizontal piping)
                                                                  D . . . . . .             304 SST or SCS13A 2-inch pipe mounting, L type (for vertical piping)
                                                                  J  . . . . . .            316 SST 2-inch pipe mounting, flat type (for horizontal piping)
                                                                  K . . . . . .             316 SST or SCS14A 2-inch pipe mounting, L type (for vertical piping)
                                                                  N . . . . . .             None
Optional Codes                                                                              / Optional specification
    The “►” marks indicate the most typical selection for each specification.
    *1:	 HART 5 or HART 7 is selectable. Specify upon ordering.
    *2: 	Users must consider the characteristics of selected wetted parts material and influence of process fluids. Specifying
          inappropriate materials has the potential to cause serious damage to human body and plant facilities resulted from an
          unexpected leak of the corrosive process fluids.
    *3:	 Lower limit of ambient and process temperature is -15°C.
    *4:	 Not applicable for electrical connection code 0, 5, 7, 9 and A.
    *5:	 Not applicable for electrical connection code 0, 5, 7 and 9.
    *6:	 Material of a blind plug; aluminum alloy for code 5 and 9, and SUS304 for code 7.
    *7:	 Not applicable for output signal code G.
    *8:	 Not applicable for output signal code F.
Table. Wetted Parts Materials
Wetted parts
                       Cover flange               Process connector                                 Capsule                       Capsule gasket      Vent/Drain plug
material code
                                                                                          Hastelloy C-276 (Diaphragm)
                                                                                                          *2
         S#               F316 SST                   ASTM CF-8M *1                                                     Teflon-coated 316L SST             316 SST
                                                                                          F316L SST, 316L SST (Others)
    *1:	 Cast version of 316 SST. Equivalent to SCS14A.
    *2:	 Hastelloy C-276 or ASTM N10276.
    The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR0175/ISO15156.
    Please refer to the latest standards for details. Selected materials also conform to NACE MR0103.
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                                                  GS 01C31B04-01EN    June 1, 2017-00
<<Contents>> <<Index>>                                                                                                                       6
	OPTIONAL SPECIFICATIONS (For Explosion Protected type) “◊”
For other agency approvals and marine approvals, please refer to GS 01C25A20-01EN.
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                     GS 01C31B04-01EN       June 1, 2017-00
<<Contents>> <<Index>>                                                                                                                       7
      Item                                                                Description                                                  Code
Canadian             CSA Explosionproof Approval *1
Standards                 Certificate: 2014354
Association               Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.0.5, C22.2 No.25, C22.2 No.30,
(CSA)                     C22.2 No.94, C22.2 No.60079-0, C22.2 No.60079-1, C22.2 No.61010-1, C22.2 No.61010-2-030
                          Explosion-proof for Class I, Groups B, C and D.
                          Dustignition-proof for Class II/III, Groups E, F and G.
                          When installed in Division 2, “SEAL NOT REQUIRED” Enclosure: Type 4X,
                          Temp. Code: T6...T4
                                                                                                                                       CF1
                          Ex d IIC T6...T4 Enclosure: IP66/IP67
                          Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
                          Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
                          –50 to 75°C(–58 to 167°F) for T6 *2
                     Process Sealing Certification
                          Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
                          No additional sealing required
                          Primary seal failure annunciation: at the zero adjustment screw
                     CSA Intrinsically safe Approval *1*3
                          Certificate: 1606623
                          [For CSA C22.2]
                             Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.25, C22.2 No.94, C22.2 No.157,
                             C22.2 No.213, C22.2 No.61010-1, C22.2 No.60079-0, C22.2 No.61010-2-030
                             Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, Groups E, F & G,
                             Class III, Division 1, Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2,
                             Groups F & G, Class III, Division 1
                             Enclosure: Type 4X, Temp. Code: T4 Amb. Temp.: –50 to 60°C(–58 to 140°F) *2
                             Electrical Parameters: [Intrinsically Safe] Vmax=30V, Imax=200mA, Pmax=0.9W, Ci=10nF, Li=0 µH
                             [Nonincendive] Vmax=30V, Ci=10nF, Li=0 µH                                                                 CS1
                          [For CSA E60079]
                             Applicable Standard: CAN/CSA E60079-11, CAN/CSA E60079-15, IEC 60529:2001
                             Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66/IP67
                             Amb. Temp.: –50 to 60°C(–58 to 140°F) *2, Max. Process Temp.: 120°C(248°F)
                             Electrical Parameters:	[Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0 µH
                             	                         [Ex nL] Ui=30V, Ci=10nF, Li=0 µH
                     Process Sealing Certification
                          Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01
                          No additional sealing required
                          Primary seal failure annunciation: at the zero adjustment screw
                     Combined CF1 and CS1 *1*3                                                                                         CU1
IECEx                IECEx Flameproof Approval *1
                          Applicable Standard: IEC 60079-0:2011, IEC60079-1:2007-4
                          Certificate: IECEx CSA 07.0008
                          Flameproof for Zone 1, Ex d IIC T6...T4 Gb Enclosure: IP66/IP67                                              SF2
                          Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
                          Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
                          –50 to 75°C(–58 to 167°F) for T6
                     IECEx Intrinsically safe and Flameproof Approval *1*3
                     Intrinsically safe Ex ia
                          Certificate: IECEx DEK 11.0081X
                          Applicable Standard: IEC 60079-0:2011, IEC 60079-11:2011
                          Ex ia IIC T4 Ga Enclosure: IP66/IP67
                          Amb. Temp.: –50 to 60 °C(–58 to 140 °F), Max. Process Temp.: 120 °C(248 °F)
                          Electrical Parameters: Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=27.6 nF, Li=0 μH
                     Intrinsically safe Ex ic
                          Certificate: IECEx DEK 13.0061X
                          Applicable Standard: IEC 60079-0:2011, IEC 60079-11:2011
                                                                                                                                       SU21
                          Ex ic IIC T4 Gc IP code: IP66
                          Amb. Temp.: –30 to 60°C(–22 to 140°F) *2, Max. Process Temp.: 120°C(248°F)
                          Electrical Parameters: Ui=30V,Ci=27.6 nF, Li=0 μH
                     Flameproof
                          Certificate: IECEx CSA 07.0008
                          Applicable Standard: IEC 60079-0:2011, IEC60079-1:2007-4
                          Flameproof for Zone 1, Ex d IIC T6...T4 Gb Enclosure: IP66/IP67
                          Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
                          Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
                          –50 to 75°C(–58 to 167°F) for T6
    *1:	   Applicable for Electrical connection code 2, 4, 7, 9, C and D.
    *2:	   Lower limit of ambient temperature is –15°C (5°F) when /HE is specified.
    *3:	   Not applicable for output signal code Q.
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                    GS 01C31B04-01EN        June 1, 2017-00
<<Contents>> <<Index>>                                                                                                                        8
	OPTIONAL SPECIFICATIONS
                 Item                                                            Description                                            Code
Painting           Color change      Amplifier cover only*2                                                                              P
                                     Amplifier cover and terminal cover, Munsell 7.5 R4/14                                               PR
              Coating change         Anti-corrosion coating*1                                                                            X2
316 SST exterior parts               316 SST zero-adjustment screw and setscrews*3                                                       HC
Fluoro-rubber O-ring                 All O-rings of amplifier housing. Lower limit of ambient temperature: –15°C (5°F)                   HE
Lightning protector                  Transmitter power supply voltage: 10.5 to 32 V DC (10.5 to 30 V DC for intrinsically safe type.)
                                     Allowable current: Max. 6000 A (1×40 µs), Repeating 1000 A (1×40 µs) 100 times                       A
                                     Applicable Standards: IEC 61000-4-4, IEC 61000-4-5
Oil-prohibited use*4                 Degrease cleansing treatment                                                                        K1
                                     Degrease cleansing treatment with fluorinated oilfilled capsule.
                                                                                                                                         K2
                                     Operating temperature −20 to 80°C (−4 to 176°F)
Oil-prohibited use with              Degrease cleansing and dehydrating treatment                                                        K5
dehydrating treatment*4              Degrease cleansing and dehydrating treatment with fluorinated oilfilled capsule.
                                                                                                                                         K6
                                     Operating temperature −20 to 80°C ( −4 to 176°F)
Capsule fill fluid                   Fluorinated oil filled in capsule
                                                                                                                                         K3
                                     Operating temperature −20 to 80°C (−4 to 176°F)
Calibration units*5                  P calibration (psi unit)                                                                            D1
                                     bar calibration (bar unit)                       (See Table for Span and Range Limits.)             D3
                                     M calibration (kgf/cm2 unit)                                                                        D4
Long vent*6                          Total length: 119 mm (standard: 34 mm); Total length when combining with option code K1,
                                                                                                                                         U1
                                     K2, K5, and K6: 130 mm. Material: 316 SST
Gold-plated capsule gasket *7        Gold-plated 316L SST capsule gasket. Without drain and vent plugs.                                  GS
Gold-plated diaphragm *18            Surface of isolating diaphragms are gold plated, effective Gold plate thickness: 3 μm               A1
                                     for hydrogen permeation.                                       Gold plate thickness: 10 μm          A2
Output limits and failure            Failure alarm down-scale : Output status at CPU failure and hardware error is −5%, 3.2mA DC
                                                                                                                                         C1
operation*8                          or less for 4 to 20 mA output type, and −5%, 0.8V DC or less for 1 to 5 V output type.
                                                                        Failure alarm down-scale: Output status at CPU
                                     NAMUR NE43 Compliant                                                                                C2
                                                                        failure and hardware error is −5%, 3.2 mA DC or less.
                                     Output signal limits:
                                     3.8 mA to 20.5 mA     *17          Failure alarm up-scale: Output status at CPU
                                                                                                                                         C3
                                                                        failure and hardware error is 110%, 21.6 mA or more.
Body option*9                        Right side high pressure, without drain and vent plugs                                              N1
  Terminal                           N1 and Process connection, based on IEC61518 with female thread on both sides of cover
    Side                                                                                                                                 N2
                                     flange, with blind kidney flanges on back.
           L          H              N2, and Material certificate for cover flange, diaphragm, capsule body, and blind kidney flange     N3
                          F03E.ai
Wired tag plate                  316 SST tag plate wired onto transmitter                                                                N4
Data configuration at factory*10                                                                  Software damping, Descriptor,
                                 Data configuration for HART communication type                                                          CA
                                                                                                  Message
                                     Data configuration for BRAIN communication type              Software damping                       CB
European Pressure                    PED 2014/68/EU
Equipment Directive*11               Category: III, Module: H, Type of Equipment: Pressure Accessory-Vessel,                            PE3
                                     Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2
Material certificate*12              Cover flange *13                                                                                   M01
                                     Cover flange, Process connector *14                                                                M11
Pressure test/                                                                                 Nitrogen(N2) Gas or Water*16
                                     Test Pressure: 32 MPa(4500 psi)                                                                    T09
Leak test certificate*15                                                                       Retention time: one minute
    *1:	       Not applicable with color change option. Not applicable for amplifier housing code 2.
    *2:	       Not applicable for amplifier housing code 2 and 3.
    *3:	       The specification is included in amplifier code 2.
    *4:	       Applicable for Wetted parts material code S.
    *5:	       The unit of MWP (Max. working pressure) on the name plate of a housing is the same unit as specified by option codes D1,
               D3, and D4.
    *6:	       Applicable for vertical impulse piping type (Installation code 7) and Wetted parts material code S.
    *7:	       Applicable for wetted parts material code S; process connection code 5; and installation code 8 and 9. Not applicable for
               option code U1, N2, N3 and M11. No PTFE is used for wetted parts.
    *8:	       Applicable for output signal codes D and J. The hardware error indicates faulty amplifier or capsule.
    *9:	       Applicable for wetted parts material code S; process connection codes 3, 4, and 5; installation code 9; and mounting
               bracket code N. Process connection faces on the other side of zero adjustment screw.
    *10:	      Also see ‘Ordering Information’.
    *11:	      If compliance with category III is needed, specify this option code.
    *12:	      Material traceability certification, per EN 10204 3.1B.
    *13:	      Applicable for process connections codes 5.
    *14:	      Applicable for process connections codes 3 and 4.
    *15:	      The unit on the certificate is always Pa unit regardless of selection of option code D1, D3 or D4.
    *16:	      Pure nitrogen gas or pure water is used for oil-prohibited use (option codes K1, K2, K5, and K6).
    *17:	      The 1 to 5 V voltage output corresponding to 4 to 20 mA current output is applied to output signal code Q which is non-
               compliant to NAMUR NE43.
    *18:	      /A2 is not applicable with FM approval.
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                     GS 01C31B04-01EN       Feb. 10, 2017-00
<<Contents>> <<Index>>                                                                                                                                                                             9
	DIMENSIONS
                                                                                                                                                                 Unit: mm (approx.inch)
                                                                                  (0.24)
                                             256(10.1)                                                         132(5.2)
                                                                                  6
                                                197(7.76)                                                      110(4.33)
                                                                                      (2.13)
                                                                                                               9      39              2-inch pipe
                                                                                        54
                                  97              143(5.63)
                                (3.82)                                                                      (0.35) (1.54)             (O.D. 60.5 mm)
                                              Mounting bracket                                                                    Vent/Drain plugs
                                              (L-type,optional)        External indicator
                                                                       Conduit connection
             (2.05)
               52
                                                                       (optional)
                                              Integral indicator
                                                  (optional)
             (4.88)
              124
277(10.9)
                                                                                                                                               192(7.56)*2
                                                                                                                              ø70(2.76)
                                                                                           ø78(3.07)
                                                                         (3.74)
                                                                           95
            (3.66)
              93
                                                                                  Process
                                                                   Conduit        connector
                                                                   connection     (optional)
             Process connection                         Zero adjustment                                High       54       Low
                                                                                                       pressure (2.13)     pressure
                                                    Ground terminal                                    side*1              side*1                              F04E.ai
                       54                                                                                                                                                    (2.76)
                     (2.13)
                                                                                                                                                                              ø70
        6
                                                                                    197(7.76)
                                 159(6.26)
                                                                                                                                                                                      143(5.63)
   (0.24)
                                                                                                       Vent plug,          Ground
                                                                                                        Drain plug*5       terminal                                                               Vent plug*5
                                                                       68
                                                                                                                                                                54(2.13)         Low
                                                                     (2.68)                                              High
                                                                                                                         pressure                                                pressure
                                                                                                       Process connector side*1                                                  side*1
                                                               169(6.65)
                                                                                                       (optional)
                       (1.63)
                        41.3
                                  (1.85)
                                    47
                                                               Mounting bracket
                                                               (Flat-type,optional)                                     2-inch pipe
                                                                                                                    (O.D. 60.5 mm)
                                                                                                                                                              154(6.06)*3
                                                                                                                                                                                                       F05E.ai
            *1:	    When installation code 8 is selected, high and low pressure side on above figure are reversed.
            	       (i.e. High pressure side is on the right side.)
            *2:	    When option code K1, K2, K5 or K6 is selected, add 15mm(0.59 inch) to the value in the figure.
            *3:	    When option code K1, K2, K5 or K6 is selected, add 30mm(1.18 inch) to the value in the figure.
            *4:	    15mm (0.59 inch) for right side high pressure.
            *5:	    Not available when option code GS is selected.
            *6:	    When electrical connection code 7 or C is selected, a blind plug is protruded upto 8 mm from the conduit connection.
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                                                          GS 01C31B04-01EN               June 1, 2017-00
<<Contents>> <<Index>>                                                                                                                                              10
                                                                                                                                        Unit: mm (approx.inch)
                                                                                                                ø78(3.07)
            (2.13)
                                                                                                                                               (2.76)
       6
                                                                                                                                                ø70
                            159(6.26)
(0.24)
                                                                                                                                                        143(5.63)
                                                                  197(7.76)
                                                                                              Ground terminal
  Electrical connection
  for code 5, 9, A, and D.
                                                                                                                                                             Vent plug
               Drain plug
                                                     68                                                                                                 Drain plug
                                                                          Process connection                                         54
               Vent plug                           (2.68)                                                    High                  (2.13)          Low
                                               169(6.65)                      Process connector              pressure            154(6.06)*1       pressure
                 Drain plug                                                                                  side                                  side
                                                                              (optional)
                                                                                                                                                                    F06E.ai
    *1:	   When Option code K1, K2, K5, or K6 is selected, add 30 mm (1.18 inch) to the value.
    *2:	   When electrical connection code 7 or C is selected, a blind plug is protruded upto 8 mm from the conduit connection.
                                                                              CHECK       +         3
                                                                                          –         2
                                                                                                           External indicator (ammeter) terminals*1*2
Ground terminal
                                                                       Three or four wire connection. For four wire connection, both supply
                                                                       and signal lines use SUPPLY - terminal.
                                                                                                                                                               F07E.ai
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                                                        GS 01C31B04-01EN     June 1, 2017-00
<<Contents>> <<Index>>                                                                                                                11
< Ordering Information > “◊”                                           < Factory Setting > “◊”
Specify the following when ordering
                                                                        Tag number          As specified in order
1.	 Model, suffix codes, and option codes
2.	 Calibration range and units                                         Software
                                                                                            ‘2.00 s’ or as specified in order
    1)	Calibration range can be specified with range                    damping *1
       value specifications up to 5 digits (excluding                   Output mode         ‘Linear’ unless otherwise specified in order
       any decimal point) for low or high range limits                  Calibration range
                                                                                            As specified in order
       within the range of -32000 to 32000. When                        lower range value
       reverse range is designated, specify Lower                       Calibration range
       Range Value(LRV) as greater than Upper Range                     upper range         As specified in order
       Value(URV). When square root output mode is                      value
       specified, LRV must be “0 (zero) ”.                                                  Selected from mmH2O, mmH2O(68°F),
    2)	Specify only one unit from the table, ‘Factory                                       mmAq*2, mmWG*2, mmHg, Pa, hPa*2,
       setting.’                                                        Calibration range   kPa, MPa, mbar, bar, gf/cm2,
3.	 Select linear or square root for output mode and                    unit                kgf/cm2, inH2O, inH2O(68°F), inHg, ftH2O,
                                                                                            ftH2O(68°F) or psi. (Only one unit can be
    display mode.                                                                           specified.)
    Note:	 If not specified, the instrument is shipped set for
           linear mode.                                                                     Designated differential pressure value
4.	 Display scale and units (for transmitters equipped                                      specified in order. (% or user scaled
                                                                        Display setting
                                                                                            value.) Display mode ‘Linear’ or ‘Square
    with the integral indicator only)
                                                                                            root’ is also as specified in order.
	 Specify either 0 to 100 % or ‘Range and Unit’ for
    engineering units scale:                                                                ‘0 to 32 MPa’ for M and H capsule,
                                                                        Static pressure
                                                                                            absolute value.
	 Scale range can be specified with range limit                         display range
                                                                                            Measuring high pressure side.
    specifications up to 5 digits (excluding any decimal
    point) for low or high range limits within the range                  *1:	   To specify these items at factory, option code CA
    of -32000 to 32000. Unit display consists of 6-digit,                        or CB is required.
    therefore, if the specified scaling unit excluding ‘/’ is             *2:	   Not available for HART protocol type.
    longer than 6-characters , the first 6 characters will
    be displayed on the unit display.                                  < Material Cross Reference >
5.	 HART PROTOCOL
	 When output signal code is “J”, specify the HART                      ASTM                 JIS
    protocol revision “5” or “7”.                                       316                  SUS316
6.	 TAG NO (if required)                                                F316                 SUSF316
	 Specified characters (up to 16 characters for                         316L                 SUS316L
    BRAIN, 22 characters for HART, or 16 characters
                                                                        F316L                SUSF316L
    for /N4 tag) are engraved on the stainless steel tag
    plate fixed on the housing.                                         304                  SUS304
7.	 SOFTWARE TAG (for HART only. if required)                           F304                 SUSF304
	 Specified characters (up to 32 characters) are set                    660                  SUH660
    as “Tag” (the first 8 characters) and “Long tag”*1                  B7                   SNB7
    (32 characters) in the amplifier memory. Use                        CF-8M                SCS14A
    alphanumeric capital letters.
	 When the “SOFTWARE TAG” is not specified,
    specified “TAG NO” is set as “Tag” (the first 8                    <Information on EU WEEE Directive>
    characters) and “Long tag”*1 (22 characters) in the                EU WEEE (Waste Electrical and Electronic Equipment)
    amplifier memory.                                                  Directive is only valid in the EU.
    *1: applicable only when HART 7 is selected.                       This instrument is intended to be sold and used only
8.	 Other factory configurations (if required)                         as a part of equipment which is excluded from WEEE
    Specifying option code CA or CB will allow further                 Directive, such as large-scale stationary industrial
    configuration at factory. Following are configurable               tools, a large-scale fixed installation and so on, and,
    items and setting range.                                           therefore, subjected to the exclusion from the scope
	 [/CA : For HART communication type]                                  of the WEEE Directive. The instrument should be
    1)	Descriptor (up to 16 characters)                                disposed of in accordance with local and national
    2)	Message (up to 30 characters)                                   legislation/regulations.
    3)	Software damping in second (0.00 to 100.00)
	 [/CB : For BRAIN communication type]
    1)	Software damping in second (0.00 to 100.00)
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation                              GS 01C31B04-01EN        April 10, 2018-00
Subject to change without notice.