Digital Force Gauge Guide
Digital Force Gauge Guide
Instruction manual
digital force gauge
SAUTER FL Version 2.0
01/2020
Congratulations on purchasing a digital force measuring device with an internal or external measuring cell from
SAUTER. We hope you will enjoy your quality measuring device
with a wide range of functions. Please do not hesitate to contact us if you have any questions, requests, or
suggestions.
Contents
1 Introduction
2 Scope of delivery
3 Additional information when using the AFH
software
4 Technical data
5 Battery indicator / power supply
6 Display indication
7 Control buttons
8 Functions
9 Computer control of the force gauge
10 RS232 output signal
11 Warnings
12 Attention:
13 Warranty
14 Adjustment instruction FL
15 Technical drawings
16 Documents / Resources
17 Related Posts
Introduction
Please read these operating instructions carefully before commissioning, even if you already have experience with
SAUTER measuring instruments.
After receipt of the force gauge, it should be checked in advance that no transport damage has occurred, that the
outer packaging, the plastic housing, other parts or even the gauge itself have not been damaged. If any damage
is evident, please notify SAUTER GmbH immediately.
The FL can measure tensile and compressive forces very accurately and it is easy to
operate. It can be held in the hand for measurements or mounted on top of a suitable test stand. SAUTER offers
optional software and accessories to make the measuring instrument more versatile in use. Please ask SAUTER
or the SAUTER supplier or visit our website www.sauter.eu.
Scope of delivery
If you want to connect an FL with the combination test bench and software (AFH-LD or AFH-FD) to a PC, you
must connect the FL to the equipment port of the test bench
via the FL-A04 cable (Sub-D 9-pin to PS2) and connect the test bench to the PC via the computer port.
Technical data
Measurement uncert
±0.2% of Max (measuring range)
ainty
Dimensions Display u
nit 160x75x30mm
(LxWxH)
Thread M6
2 kN 76x51x19mm M12x1.75
20 kN 76x51x28mm M12x1.75
4.4 Technical Data FL TM
5N; 10N; 25N; 50N; 100N; 250N; 5 1kN; 2,5kN; 5kN; 10kN; 20kN;
Adjustable measuring ranges
00N 50kN
4.4.1 Sensors
The sensor must be ordered separately. Service FL-C01 necessary.
4.4.2 Load cell connection
Solder the cable from the measuring cell to the supplied connector. The pin assignment at the connector is as
follows:
Pin 1: Supply ( – )
Pin 2: Signal ( + )
Pin 3: Signal ( – )
Pin 4: Supply ( + )
Pin 1: EXC ( – )
Pin 2: Signal ( + )
Pin 3: Signal ( – )
Pin 4: EXC ( + )
Battery indicator / power supply
If the battery voltage is less than 3.9V, a message appears that the battery is exhausted and the device
switches off automatically
The FL has equipped with four Nickel Metal Hybrid (NiMH) batteries type AAA. For safety reasons during
transport, the batteries are supplied uncharged. For maximum battery life, we recommend that you charge the
batteries only with the original charger (supplied) for a period of 14-16 hours before using the meter for the first
time.
Display indication
Position Description
1 Measurement result
8 Force diagram
Control buttons
MENU/ESC:
ZERO (zeroing):
ENTER:
Press the key briefly: Switch between N, gf, kgf, of, lbf, mN
Mode (mode change):
Assignment with three functions:
ON / OFF:
Functions
Colour Function
Green This indicates that the pass-fail criterion has been reached
An upper and a lower limit value can be programmed. The measuring instrument compares the measurement
result with the limit values and outputs the result as a signal in red or green light diodes.
The setting of the limit values see MENU →PASS-FAIL With the left arrow key the cursor is moved to the desired
value. With the UP or DOWN key, the value can be changed. By pressing and holding one of these keys, scrolling
is possible. With the right arrow key, the unit can be changed. With the ENTER-key the settings are saved and you
return to the main menu.
The display now shows a “PF”.
The PASS-FAIL option is automatically disabled if the upper and lower limits are set to 0 Newton. The lower value
must also always be smaller than the upper one.
8.2 Invert display
The display can be rotated by 180°. To do this, do the following: When the power is off, press and hold the
“MENU/ESC” key. In addition, the “On/Off” key is pressed and the unit switches on, thus rotating the display.
8.3 Store measured value in memory
Any measured value can be stored in the memory at any time by pressing the ENTER key. A maximum of 500
readings can be stored.
8.4 Memory / Storage
The memory is used to view the stored data. It can store 500 readings. Currently, individual entries or all entries
can be deleted, or the measurement data of the memory can be printed out.
To enter the MEMORY menu, the MENU/ESC button must be pressed. Use the UP or DOWN button to move the
cursor to MEMORY. Then confirm with the ENTER
button and you are on the memory page. Press the ESC button to return to the main menu page.
With the arrow keys UP and DOWN the page can be changed within the memory and by pressing and holding
one of these two keys it is possible to scroll up or down. By pressing the PRINT key, the memory contents can be
printed out via the serial port. By pressing the ZERO button, the DELETE menu is accessed
Use the UP and DOWN arrow keys to select the desired delete option. If NO is selected here and confirmed with
the ENTER key, the meter returns to the memory page.
When DELETE is selected and the ENTER button is pressed to confirm, the currently stored measurement is
deleted and the instrument returns to the memory page. When DELETE ALL is selected and the ENTER button is
pressed to confirm, all stored measurements are deleted and the instrument returns to the memory page.
8.5 Diagnostics
This function is used to check the loadcell. If it is suspected that the loadcell has been overloaded, the status can
be determined immediately. To do this, place the device in a horizontal position on a flat surface and call up the
main menu page. To enter the DIAGNOSTIC menu, the MENU/ESC button must be pressed. Use the UP or
DOWN button to move the cursor to DIAGNOSTIC and confirm with the ENTER button. You are now on the
DIAGNOSTIC page. Press the ESC key to return to the main menu page.
If the percentage drop is between 5% – 10%, please contact the supplier to have the loadcell replaced. These
values are given as a guide only. The actual need for calibration/replacement of the loadcell varies with its
individual characteristics.
With the RS 232 interface, a connection from the force gauge to the PC can be established by sending RS232
commands.
The reading of the FL shown on the display can be transferred to the PC by pressing the PRINT button or by
issuing a command instruction to the force gauge from the PC.
RS232 command Action
– Send the direct reading with the unit if the current mode is the track mode
“x” or
– Send peak value (on pull) with the unit if the current mode is peak mode
pressing
– Send peak value (on pressure) with the unit if the current mode is the peak value pri
PRINT KEY
nt mode
Send information about the device such as model, capacity, serial number, firmware c
““!”
heck, original offset, current offset, overload counter
– Data bit: 8
– Parity: none
– Stop bit: 1
-Startt bit 1
3 GND Ground
Warnings
Incorrectly performed force measurements can lead to serious injury to persons and damage to objects and must
therefore only be performed by trained and experienced personnel.
In particular, it must be avoided that forces act on the purchased measuring an instrument that exceeds the
maximum load (Max) of the instrument or which do not act axially via the external and internal load cell; or if high
impulse forces act on the measuring instrument.
Avoid twisting the loadcell, otherwise, it could be damaged and the measuring accuracy will decrease in any case.
Inappropriate use
Do not use the measuring instrument for medical weighing. If small quantities of the material to be measured are
removed or added, incorrect measurement results may be displayed due to the “stability compensation” in the
measuring instrument! (Example: Slow flow of liquids out of a container suspended from the measuring cell). Do
not apply a continuous load to measuring instruments with an external measuring cell.
Overloads
Please prevent the measuring instrument from being overloaded beyond the specified maximum load (Max),
minus any tare load that may already be present. This can damage the measuring instrument (danger of
breakage!)
Attention:
Make sure that never let people or objects are under the load, as it injures or could be damaged!
The measuring instrument is not suitable for weighing people, do not use it as an infant measuring instrument!
The measuring device does not comply with the German Medical Devices Act (MPG).
Never operate the measuring instrument in rooms where there is a risk of explosion. The standard version is
not explosion-proof.
The design of the measuring instrument must not be changed. This can lead to incorrect measurement results,
safety-related defects, and the destruction of the measuring device.
The measuring instrument may only be operated or maintained by trained personnel.
The measuring instrument may only be used in accordance with the described specifications.
SAUTER must give written approval for any other areas of use/applications.
Warranty
Note:
To view the CE declaration, please click on the following link: https://www.kern-
sohn.com/shop/de/DOWNLOADS/
Adjustment instruction FL
Introduction:
This manual describes the adjustment procedure, although some programming modes are password protected.
Important! During the adjustment procedure, all weights for loading the sensor should be calibrated.
Tempering:
Before the measuring instrument is handed over to the calibration laboratory, the respective instrument must be
checked for possible damage and the order specification. Before starting adjustment, allow the measuring
instrument to reach the working temperature of the calibration laboratory for at least 2 hours.
Preparation:
Insert a full set of charged batteries in the battery compartment or connect a standard power supply
Calibration:
CALIBRATION MENU
1. MAXIMUM LOAD
2. VOLTAGE INCREASE
3. COMPRESSION GAIN
4. SET NEW PASSWORD
3. Select the correct value of the weight for the respective measuring instrument. To set the load, use the UP and
DOWN arrow keys to select the MAX CAPACITY menu function and press ENTER. The menu window for setting
the maximum load appears on the screen.
4. Use the UP and DOWN arrow keys to select the maximum load, press the selected value and hold for a
moment. Press the ENTER key to save the selected value. Then return to the calibration menu.
5. To calibrate in the voltage direction: use the UP and DOWN arrow keys to select TENSION GAIN in the
calibration menu and then press ENTER. The display shows a menu window for voltage increase.
Set up the measuring device together with the necessary accessories for performing the voltage calibration
apply a force of 100% of the meter’s power to the weight
Remove the load, leave the device standing
as soon as the display becomes stable, press the ZERO key to save the zero display
6. Compression calibration: In the calibration menu, use the UP and DOWN arrow keys to select the
COMPRESSION GAIN menu function and then press ENTER. The display will show a menu window for
compression gain.
Set up the measuring device and all necessary accessories for performing the compression calibration
apply a force of 100% of the meter’s power to the weight
Remove the load, leave the device standing
as soon as the display becomes stable, press the ZERO key to save the zero display
7. After successful calibration, remove all accessories from the meter and place the meter on a flat surface with its
backside facing up. Select the voltage increase menu window, press the ZERO key to obtain the primary
calibration offset, date, and time, and reset the overload counter.
Technical drawings
Sauter GmbH
Ziegelei 1
D-72336 Balingen
e-mail: info@kern-sohn.com
Phone : +49- [0]7433- 9933-0
Fax: +49- [0]7433-9933-149
Internet: www.sauter.eu
FL-BA-e-2020.docx