Through-Wall Vital Signs Tracking
Through-Wall Vital Signs Tracking
Article
Through-Wall Multiple Targets Vital Signs Tracking
Based on VMD Algorithm
Jiaming Yan † , Hong Hong † , Heng Zhao † , Yusheng Li, Chen Gu and Xiaohua Zhu *
 School of Electronic and Optical Engineering, Nanjing University of Science and Technology,
 Nanjing 210094, China; jamlff2494@163.com (J.Y.); hongnju@njust.edu.cn (H.H.);
 soniczhao@live.com (H.Z.); leassun@126.com (Y.L.); gc_njust@163.com (C.G.)
 * Correspondence: zxh_njust@126.com; Tel.: +86-25-8431-5550
 † These authors contributed equally to this work.
 Abstract: Targets located at the same distance are easily neglected in most through-wall multiple
 targets detecting applications which use the single-input single-output (SISO) ultra-wideband (UWB)
 radar system. In this paper, a novel multiple targets vital signs tracking algorithm for through-wall
 detection using SISO UWB radar has been proposed. Taking advantage of the high-resolution
 decomposition of the Variational Mode Decomposition (VMD) based algorithm, the respiration
 signals of different targets can be decomposed into different sub-signals, and then, we can track
 the time-varying respiration signals accurately when human targets located in the same distance.
 Intensive evaluation has been conducted to show the effectiveness of our scheme with a 0.15 m thick
 concrete brick wall. Constant, piecewise-constant and time-varying vital signs could be separated
 and tracked successfully with the proposed VMD based algorithm for two targets, even up to three
 targets. For the multiple targets’ vital signs tracking issues like urban search and rescue missions, our
 algorithm has superior capability in most detection applications.
1. Introduction
      Through-wall detection has become a fundamental topic in homeland security, rescue missions
and obscured objects detecting applications in recent years [1–8]. The traditional optical image sensors
fail to provide visual information behind walls and shelters. The through-wall radar sensors, especially
the ultra-wideband (UWB) radar have therefore become the preferred choices. Different from other
radar systems, UWB radar system has strong penetrating power compare to the optical system, and
has high range resolution and good resolving ability compare to the continuous wave (CW) radar
system [9–14]. By using impulse signal or frequency modulated (FM) signal, UWB radar can not only
measure the micro-doppler movements but can also distinguish the closely positioned targets.
      Under this circumstance, through-wall human detection using UWB radar has emerged as one
of the keen research topic. The applications of UWB system include indoor target localization, target
tracking and multi-dimensional targets imaging [15–18]. In the meantime, with the development
of remote biomedical monitoring, Doppler radar detection of vital signs has become a promising
technique for health monitoring and life sensing applications [19–23]. This technique is based on the
detection of significant features of human periodic motions, and UWB radar is a great candidate to
receive signals containing periodic motions information from human targets related to the periodic
contraction of lungs and the beating of hearts in the process of breathing and heartbeat respectively.
      According to the degree of difficulty of the detection, the through-wall vital signs detection
can divide into two parts, single target detection and multiple targets detection. For single target
detection, a great deal of work has been achieved in the field of detecting human respiration, heartbeats
and gait [24–26]. Compared with the various work of detecting a single target, the multiple targets
detection faces difficulties like low resolution and overlapped harmonics of respiration waveform.
Only a few effective algorithms have been proposed to detect human respiration through a wall [27–29].
In [27], a self-injection-locked (SIL) radar was presented for ranging and vital signs detection of targets
through wooden wall when it worked in frequency modulated continuous wave (FMCW) mode.
However, the FMCW transmitter signal can easily leak and susceptible to near ground clutter. In [28],
the impulse UWB radar system was successfully used to monitor the micro-Doppler signatures like
breathing rates of two subjects through a cement wall. In [29], the group used UWB radar system to
detect and identify life characteristics of two targets at different distance. However, the number of
multiple targets in these cases is limited to two person and they cannot track the rapid change of the
respiration signals. Therefore, the through-wall detection of vital signs features from multiple targets
located in the same distance is easily to be neglected.
      In this paper, we apply a SISO UWB radar system with high precision in through-wall multiple
targets vital signs detection. Based on the system, we addressed VMD algorithm [30] to decompose the
respiration characteristics from human targets in the same distance. Then we used Hilbert transform
to track the time-varying respiration signals. The UWB system has high resolution and could provide
great penetration. So it could provide high accuracy and separation in through-wall multiple targets
vital signs detecting applications. The VMD algorithm has not been applied to radar signal processing
before and it’s a novel time domain algorithm to separate different frequency signals in low noise
levels. The method has been experimentally validated in this paper and has the ability to detect
multiple targets respiration rates through the wall.
      This paper is organized as follows. In Section 2, we describe the theory of UWB radar vital signs
measurement. In Section 3, the multiple targets vital signs tracking algorithm is described. In Section 4,
UWB radar system and its configuration are proposed, and the experimental results are used to assess
the effectiveness of the proposed procedure. Finally, Section 5 contains the conclusions.
Figure 1. Block diagram of the single-input single-output (SISO) ultra-wideband (UWB) radar system.
     We assume that targets are located at a fixed distance d0 with multiple targets physiological
displacement x (t). Thus, the total distance between the transmitter and targets d(t) can be expressed as:
d ( t ) = d0 + x ( t ) (1)
                                                         m
                                              x (t) =   ∑ ai sin(2π fi t)                                         (2)
                                                        i =1
where i = [1, 2, · · · , m] denotes the number of the targets, ai and f i represent the respiration amplitude
and frequency of the i-th target, respectively.
     In this situation, the received signal can be represented as the sum of the responses of the channel,
So the impulse responses h(t, τ ) in the range profile reflected from multiple targets will be:
                                                                m
                                      h(t, τ ) = hc (t, τ ) + ∑ Ai ∂(τ − τiv (t))                                 (3)
                                                               i =1
     hc (t, τ ) denotes the response from environmental noise and other multipath component.
Where τiv (t) is the time delay and Ai is the amplitude of the pulse reflected on the i-th target.
This indicates that the respiration movements modulate the received signal.
     Due to stationarity of reflectors in the scene, each time delay τv associated with the respirations
are modeled as the sum of time of arrival τ0 plus some sinusoidal delays associated to the respiration
displacements of multiple targets:
                                                                m
                                                 2d(t)
                                      τv (t) =         = τ0 + ∑ τi sin(2π f i t)                                  (4)
                                                   C          i =1
where C denotes the propagation speed, and τi represents the respiration displacement of i-th target.
As the transmit signal T (t) is reflected by the targets, the received signal r (t, τ ) can be written as:
    Then discrete instants the received waveforms r (t, τ ) and the values are stored in a J × N matrix
R = R[ J, N ]:
                                    R[ J, N ] = r (t = jTs , τ = nT f )                             (6)
where j = [1, 2, · · · , J ] indicates the slow time index and n = [1, 2, · · · , N ] indicates the fast time index.
Ts and T f are the sampling period in slow time and fast time, respectively.
Figure 2. Block diagram of the multiple targets vital signs tracking algorithm.
     The received signal R[ J, N ] derived from the UWB radar system divide into J range bins and N
discrete-time sequences. We apply a moving window of 2048 points to track the candidate range bin
based on calculate the variance Vj , which can be described by:
                                             2                           2                   2
                          Vj = x j,1 − x j        + · · · + x j,n − x j        + x j,N − x j             (7)
where x j is the average value of J range bins and x j,n is the n-th slow time sample of the j-th range bin.
     After calculating the variance of each pulse in each range bin and find the range bin with the
maximum variance, the candidate range bin V can be picked to get the desired respiration signals.
The expression can be written as:
                                               V = max Vj                                                (8)
     Here we get the original signal f v (t) from range bin V, which can be used later on for respiration
signals separation. After f v (t) is obtained and then, is filtered by a low-pass filter from 0 Hz to 0.7 Hz
(0.7 Hz corresponding to 42 Beats/Minute) to reduce the magnitude of the high frequency component.
where µ̂k (ω ) is the Fourier transform of the sub-signal µk (t). Dual ascent for all ω ≥ 0. In VMD
algorithm, the Lagrangian multiplier λ can be used to enforce exact reconstruction of the input
signal [30], the updates of Lagrangian multipliers λ is as follows:
                                                                                          !
                            λ̂n+1 (ω ) ← λ̂n (ω ) + τ          fˆ (ω ) − ∑ µ̂nk +1 (ω )                (11)
                                                                               k
where τ is the update parameter of Lagrangian multiplier (pick 0 for denoising), and the cut-off
condition of µk is as follows:
                                    ∑k k µ̂nk +1 − µ̂nk k22
                                                            <e                              (12)
                                          k µ̂nk k22
Sensors 2016, 16, 1293                                                                                       5 of 11
where e is the tolerance of convergence criterion (typically around 10−6 ). The final step is reconstruction
each mode µk (ω ) converges after iterations. Then recognize the respiration signals and obtain K R
modes time domain respiration signals µKR (t) and the each center frequency ωKR .
where W (t) is a rectangular function of width T and height 1, with T set to be three times of the
dominant oscillatory period of the selected modes.
     Considering the impact of a real experimental environment, we carried out a comparative
simulation experiments to verify the performance of our algorithm. The simulated signal is a
composition of two simple components which simulated respiration rates were set at 12 Beats/Minute
(0.2 Hz) and 21 Beats/Minute (0.35 Hz) respectively. To verify the anti-noise performance of our
algorithm, we compared the simulated results between noiseless condition and one with SNR of
−10 dB. The comparison results are as follows.
     Figure 3a shows the simulated signal and the two respiration signals decompose by VMD
algorithm are shown in Figure 3b,c. Then the results from Hilbert transform after VMD algorithm
is presented in Figure 3d. From this picture, our algorithm not only gets the accurate frequency
information but also separates each respiratory signal dynamically. At the control group, we add white
Gaussian noise and change the SNR into −10 dB to simulate the experiment. Figure 3e–h correspond
to the former four figures respectively. For sufficiently low noise levels, we calculate the correlation of
each mode. The correlation of Figure 3b,f is 97.66%, the correlation of Figure 3c,g is 95.66%. The results
show the VMD based algorithm have a good anti-noise performance.
      Figure 3. Simulated respiration signals and separation results. (a) Simulated signal, (b) and (c) two
      modes decompose by Variational Mode Decomposition (VMD) based algorithm; (d) time-varying
      respiration signals; (e) Simulated signal with −10 dB SNR, (f) and (g) two modes decompose by VMD
      based algorithm; (h) time-varying respiration signals.
Sensors 2016, 16, 1293                                                                             6 of 11
     Based on the system, we designed four different types of experiments to track vital signs
features from multiple targets through the wall. In each experiment, we set each target almost
at the same distance from the antenna and make sure the environment variables virtually unchanged.
The experimental settings and results are as follows.
1 0.5
                           Normalized Amplitude
                                                  0.5
                                                                                        mode 1
                                                    0                                                       0
-0.5
                                                   -1                                                 -0.5
                                                         0   10   20    30    40   50                            0   10   20    30    40      50
                                                                   time (s)                                                time (s)
                                                                      (a)                                                     (b)
                                                  0.5                                                      0.7
                                                                                                           0.6                        mode1
                                                                                                                                      mode2
                                                                                          Frequency (Hz)
                                                                                                           0.5
                           mode 2
                                                                                                           0.4
                                                    0
                                                                                                           0.3
                                                                                                           0.2
                                                                                                           0.1
                                                  -0.5                                                      0
                                                         0   10   20    30    40   50                            0   10   20    30    40      50
                                                                   time (s)                                                time (s)
                                                                      (c)                                                     (d)
      Figure 5. Two targets with constant respiration rates (k = 2, α = 20, 000). (a) Original signal; (b) and
      (c) are two modes decompose by VMD based algorithm; (d) is the time-varying respiration signal.
                                                    1                                                      0.5
                           Normalized Amplitude
                                                  0.5
                                                                                        mode 1
0 0
-0.5
                                                   -1                                                 -0.5
                                                         0   10   20    30    40   50                            0   10   20    30    40      50
                                                                   time (s)                                                time (s)
                                                                      (a)                                                     (b)
                                                  0.5                                                      0.5
                                                                                                                                      mode1
                                                                                                           0.4                        mode2
                                                                                          Frequency (Hz)
                           mode 2
                                                                                                           0.3
                                                    0
                                                                                                           0.2
0.1
                                                  -0.5                                                      0
                                                         0   10   20    30    40   50                            0   10   20    30    40      50
                                                                   time (s)                                                time (s)
                                                                      (c)                                                     (d)
      Figure 6. Two targets with piecewise-constant respiration rates (k = 3, α = 20, 000). (a) Original signals;
      (b) and (c) two modes decompose by VMD based algorithm; (d) time-varying respiration signal.
Sensors 2016, 16, 1293                                                                                                                                                                                                             8 of 11
     The UWB echo data are shown in Figure 6a. From this figure, it is difficult to find a respiration
tendency from the two targets. Then the respiration signals separation results are shown in Figure 6b,c
and we can see the dynamic characteristic of the two respiration modes clearly. Because target B
held his breath first, he has a large period of deep breathing until the end of the experiment. Then
in Figure 6d, the time-frequency analysis results indicate a respiration rate of 20 Beats/Minute for
the seated target A, and a respiration rate of 12 Beats/Minute for the seated target B, respectively.
Finally, Figure 7a shows the FFT results without VMD based algorithm and Figure 7b shows the FFT
results with VMD based algorithm, which the comparison is very obvious. The VMD based algorithm
removes the interference of the wall and separates the respiration signal of each target very well.
                                                 1                                                                                                                       1
                                                                                                                                                                                                                       mode1
                                                                                                                                              Normalized Amplitude
                         Normalized Amplitude
                                                                                                                                                                                                                       mode2
                                                0.8                                                                                                                  0.8
0.6 0.6
0.4 0.4
0.2 0.2
                                                 0                                                                                                                       0
                                                  0                            0.2        0.4        0.6        0.8        1                                              0       0.2      0.4           0.6        0.8        1
                                                                                        Frequency (Hz)                                                                                   Frequency (Hz)
                                                                                              (a)                                                                                              (b)
      Figure 7. The comparison of the traditional FFT method and VMD based method. (a) FFT results
      without VMD based algorithm; (b) FFT results with VMD based algorithm.
                                                                               1
                                                                                                                                                          0.5
                                                      Normalized Amplitude
                                                                             0.5
                                                                                                                                    mode 1
0 0
                                                                             -0.5
                                                                                                                                                   -0.5
                                                                              -1
                                                                                    0    5      10     15       20    25       30                                        0    5     10     15       20         25      30
                                                                                                     time (s)                                                                            time (s)
                                                                                                        (a)                                                                                 (b)
                                                                                                                                                          0.7
                                                                             0.5                                                                                                                               mode1
                                                                                                                                                          0.6
                                                                                                                                                                                                               mode2
                                                                                                                                      Frequency (Hz)
                                                                                                                                                          0.5
                                                      mode 2
                                                                                                                                                          0.4
                                                                               0
                                                                                                                                                          0.3
                                                                                                                                                          0.2
                                                                                                                                                          0.1
                                                                             -0.5
                                                                                                                                                                     0
                                                                                    0    5      10     15       20    25       30                                        0    5     10     15       20         25      30
                                                                                                     time (s)                                                                            time (s)
                                                                                                        (c)                                                                                 (d)
      Figure 8. Two targets with time-varying respiration rates and VMD based algorithm results
      (k = 3, α = 20, 000). (a) Original signals; (b) and (c) two modes decompose by VMD based algorithm;
      (d) time-varying respiration signals.
Sensors 2016, 16, 1293                                                                                                                                            9 of 11
     The UWB echo data are shown in Figure 8a, from this figure, it is difficult to get the time-varying
respiration rate from each target by using the FFT algorithm directly. Then the VMD based algorithm
results are shown in Figure 8b,c, from which we can see the dynamic characteristics of the two
respiration signals clearly. Finally in Figure 8d, from the Hilbert transform we are able to see two
instantaneous respiration rates more directly. Target A varied from 12 Beats/Minute (0.2 Hz) to
18 Beats/Minute (0.3 Hz) and target B’s respiration rate varied from 30 Beats/Minute (0.5 Hz) to
24 Beats/Minute respectively. The results show our algorithm has the excellent capability to detect
multiple targets time-varying respiration signals.
                                                  1                                                          0.5
                                                                                                   mode 1
                         Normalized Amplitude
                                                                                                               0
                                                0.5                                                         -0.5
                                                                                                                   0   5        10     15       20   25      30
                                                                                                             0.5
                                                                                                   mode 2
                                                  0                                                            0
                                                                                                            -0.5
                                                                                                                   0   5        10     15       20   25      30
                                                -0.5                                                         0.5
                                                                                                   mode 3
                                                                                                               0
                                                 -1                                                         -0.5
                                                       0   5   10     15       20        25   30                   0   5        10     15       20   25      30
                                                                    time (s)                                                         time (s)
                                                                       (a)                                                              (b)
                                                0.7
                                                0.6                                                                    mode1           mode2         mode3
                               Frequency (Hz)
                                                0.5
                                                0.4
                                                0.3
                                                0.2
                                                0.1
                                                  0
                                                       0        5                   10               15                    20               25               30
                                                                                              time (s)
                                                                                                 (c)
      Figure 9. Three targets with constant respiration rates (k = 4, α = 20, 000). (a) Original signals; (b) three
      modes decompose by VMD based algorithm; (c) time-varying respiration signal.
     The original data are presented in Figure 9a and the three sub-signals are illustrated in Figure 9b.
From the picture, we can see the dynamic characteristic with different periods from three targets.
The respiration characteristics of three targets are presented in the Figure 9c, which shows that target
A has a respiration rate of 0.12 Hz (7 Beats/Minute), target B has a respiration rate of 0.34 Hz
(20 Beats/Minute) and target C has a respiration rate of 0.45 Hz (27 Beats/Minute) respectively.
Based on the results of this experiment, we still performed the experiment three times and the
separating results were all successful. This experiment demonstrates the VMD based algorithm
we proposed has high accuracy and strong decomposition to detect the constant vital signs signals of
three targets.
5. Conclusions
     In this paper, we consider the superimposition and interference issues of echo signals from
the same distance, which most through-wall multiple targets detecting applications are unable to
Sensors 2016, 16, 1293                                                                                        10 of 11
detect. The modes decomposition method, i.e., the VMD algorithm, has been elegantly extended
to SISO UWB radar system. A proof of concept is provided by using experimental data collected
in a laboratory environment with up to three targets at different respiratory status, i.e., constant,
piecewise-constant and time-varying respiration rate. The proposed technique has proven to provide
significant advantages in dynamic monitoring while maintaining desirable signal quality. The results
show that the VMD based algorithm is capable of detecting time-varying respiration rate of each target
in the same distance. In the near future, we will address more complex experiments, such as detecting
heartbeats and tracking through-wall multiple human respiration rates.
Acknowledgments: This work was supported by the Special Foundation of China Postdoctoral Science under
Grant 2013T6054, the National Natural Science Foundation of China under Grant 61301022, the National Key
Technology Support Program 2015BAI02B04, and by the the Natural Science Foundation of Jiangsu Province
under Grant BK20140801.
Author Contributions: Jiaming Yan, Hong Hong and Heng Zhao conceived and designed the experiments;
Yusheng Li developed the hardware system; Jiaming Yan, Heng Zhao and Chen Gu performed the experiments;
Jiaming Yan and Heng Zhao analyzed the data; Jiaming Yan, Hong Hong and Xiaohua Zhu participated in the
analysis of the results; Jiaming Yan wrote the paper.
Conflicts of Interest: The authors declare no conflict of interest.
References
1.    Gennarelli, G.; Vivone, G.; Braca, P.; Soldovieri, F.; Amin, M.G. Multiple extended target tracking for
      through-wall radars. IEEE Trans. Geosci. Remote Sens. 2015, 53, 6482–6494.
2.    Lubecke, V.M.; Boric-Lubecke, O.; Host-Madsen, A.; Fathy, A.E. Through-the-wall radar life detection and
      monitoring. In Proceedings of the 2007 IEEE/MTT-S International Microwave Symposium, Honolulu, HI,
      USA, 3–8 June 2007; pp. 769–772.
3.    Setlur, P.; Alli, G.; Nuzzo, L. Multipath exploitation in through-wall radar imaging via point spread functions.
      IEEE Trans. Image Process. 2013, 22, 4571–4586.
4.    Dehmollaian, M.; Sarabandi, K. Refocusing through building walls using synthetic aperture radar. IEEE Trans.
      Geosci. Remote Sens. 2008, 46, 1589–1599.
5.    Li, G.; Burkholder, R.J. Hybrid matching pursuit for distributed through-wall radar imaging. IEEE Trans.
      Antennas Propag. 2015, 63, 1701–1711.
6.    Browne, K.E.; Burkholder, R.J. Nonlinear optimization of radar images from a through-wall sensing system
      via the Lagrange multiplier method. IEEE Geosci. Remote Sens. Lett. 2012, 9, 803–807.
7.    Yoon, Y.-S.; Amin, M.G. High-resolution through-the-wall radar imaging using beamspace MUSIC. IEEE Trans.
      Antennas Propag. 2008, 56, 1763–1774.
8.    Browne, K.E.; Burkholder, R.J.; Volakis, J.L. Fast optimization of through-wall radar images via the method of
      Lagrange multipliers. IEEE Trans. Antennas Propag. 2013, 61, 320–328.
9.    Nguyen, V.; Pyun, J. Location detection and tracking of moving targets by a 2D IR-UWB radar system. Sensors
      2015, 15, 6740–6762.
10.   Yarovoy, A.G.; Ligthart, L.P.; Matuzas, J.; Levitas, B. UWB radar for human being detection. IEEE Aerosp.
      Electron. Syst. Mag. 2006, 21, 10–14.
11.   Zhuge, X.; Yarovoy, A.G. A sparse aperture MIMO-SAR-Based UWB imaging system for concealed weapon
      detection. IEEE Trans. Geosci. Remote Sens. 2011, 49, 509–518.
12.   Yang, Y.; Fathy, A.E. Development and implementation of a real-time see-through-wall radar system based on
      FPGA. IEEE Trans. Geosci. Remote Sens. 2009, 47, 1270–1280.
13.   Charvat, G.L.; Kempel, L.C.; Rothwell, E.J.; Coleman, C.M.; Mokole, E.L. A through-dielectric radar imaging
      system. IEEE Trans. Antennas Propag. 2010, 58, 2594–2603.
14.   Fioranelli, F.; Salous, S.; Raimundo, X. Frequency-modulated interrupted continuous wave as wall removal
      technique in through-the-wall imaging. IEEE Trans. Geosci. Remote Sens. 2014, 52, 6272–6283.
15.   Kocur, D.; Svecova, M.; Rovnakova, J. Through-the-wall localization of a moving target by two independent
      Ultra Wideband (UWB) radar systems. Sensors 2013, 13, 11969–11997.
16.   Huang, Q.; Qu, L.; Wu, B.; Fang, G. UWB Through-wall imaging based on compressive sensing. IEEE Trans.
      Geosci. Remote Sens. 2010, 48, 1408–1415.
Sensors 2016, 16, 1293                                                                                      11 of 11
17. Wang, F.; Horng, T.; Peng, K.; Jau, J.; Li, J.; Chen, C. Seeing through walls with a self-injection-locked radar
    to detect hidden people. In Proceedings of the 2012 IEEE MTT-S International Microwave Symposium
    Digest (MTT), Montreal, QC, Canada, 17–22 June 2012; pp. 1–3.
18. Zhang, Z.; Zhang, X.; Lv, H.; Lu, G.; Jing, X.; Wang, J. Human-target detection and surrounding structure
    estimation under a simulated rubble via UWB radar. IEEE Geosci. Remote Sens. Lett. 2013, 10, 328–331.
19. Gu, C.; Li, C. Assessment of human respiration patterns via noncontact sensing using doppler multi-radar
    system. Sensors 2015, 15, 6383–6398.
20. Lazaro, A.; Girbau, D.; Villarino, R. Techniques for clutter suppression in the presence of body movements
    during the detection of respiratory activity through UWB radars. Sensors 2014, 14, 2595–2618.
21. Li, C.; Chen, F.; Jin, J.; Lv, H.; Li, S.; Lu, G.; Wang, J. A method for remotely sensing vital signs of human
    subjects outdoors. Sensors 2015, 15, 14830–14844.
22. Ren, L.; Koo, Y.S.; Wang, H.; Wang, Y.; Liu, Q.; Fathy, A.E. Noncontact multiple heartbeats detection and
    subject localization using UWB impulse doppler radar. IEEE Microw. Wirel. Compon. Lett. 2015, 25, 690–692.
23. Yan, J.; Zhao, H.; Li, Y.; Sun, L.; Hong, H.; Zhu, X. Through-the-wall human respiration detection using
    impulse ultra-wide-band radar. In Proceedings of the 2016 IEEE Topical Conference on Biomedical Wireless
    Technologies, Networks, and Sensing Systems (BioWireleSS), Austin, TX, USA, 24–27 January 2016; pp. 94–96.
24. Bugaev, A.S.; Chapursky, V.V.; Ivashov, S.I.; Razevig, V.V.; Sheyko, A.P.; Vasilyev, I.A. Through wall sensing
    of human breathing and heart beating by monochromatic radar. In Proceedings of the Tenth International
    Conference on Ground Penetrating Radar, GPR 2004, Delft, The Netherlands, 21–24 June 2004; pp. 291–294.
25. Liu, L.; Liu, Z.; Barrowes, B.E. Through-wall bio-radiolocation with UWB impulse radar: Observation,
    simulation and signal extraction. IEEE J. Sel. Top. Appl. Earth Observ. Remote Sens. 2011, 4, 791–798.
26. Li, W.; Jing, X.; Li, Z.; Wang, J. A new algorithm for through wall human respiration monioring using GPR.
    In Proceedings of the 2012 14th International Conference on Ground Penetrating Radar (GPR), Shanghai,
    China, 4–8 June 2012; pp. 947–952.
27. Wang, F.; Horng, T.; Peng, K.; Jau, J.; Li, J.; Chen, C. Detection of concealed individuals based on their vital
    signs by using a see-through-wall imaging system with a self-injection-locked radar. Microw. IEEE Trans.
    Theory Tech. 2013, 61, 696–704.
28. Wang, Y.; Liu, Q.; Fathy, A.E. CW and pulse-doppler radar processing based on FPGA for human sensing
    applications. IEEE Trans. Geosci. Remote Sens. 2013, 51, 3097–3107.
29. Li, J.; Zeng, Z.; Sun, J.; Liu, F. Through-wall detection of human being’s movement by UWB radar. IEEE Geosci.
    Remote Sens. Lett. 2012, 9, 1079–1083.
30. Dragomiretskiy, K.; Zosso, D. Variational mode decomposition. IEEE Trans. Signal Process. 2014, 62, 531–544.
31. Huang, N.E.; Shen, Z.; Long, S.R.; Wu, M.C.; Shih, H.H.; Zheng, Q.; Yen, N.C.; Tung, C.C.; Liu, H.H.
    The empirical mode decomposition and the Hilbert spectrum for nonlinear non-stationary time series analysis.
    Proc. R. Soc. Lond. A 1998, 454, 903–995.
32. Hjortland, H.A.; Lande, T.S. CTBV integrated impulse radio design for biomedical applications. IEEE Trans.
    Biomed. Circuits Syst. 2009, 3, 79–88.
                          c 2016 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access
                         article distributed under the terms and conditions of the Creative Commons Attribution
                         (CC-BY) license (http://creativecommons.org/licenses/by/4.0/).