Centermax Guide
Centermax Guide
Coordinate measuring
machine for series
production
Operating Instructions
Read this first! • Please read these operating instructions, before starting up the
measuring machine.
• For your own safety, always keep all relevant documents within
easy reach at all times.
Chapter 1 Introduction
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
CE marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Safety of the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Standards and regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Requirements for person responsible for CMM . . . . . . . . . . . . . . . 1-9
Definition of a specialist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Safety symbols on the CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Requirements for the safe use of the CMM . . . . . . . . . . . . . . . . 1-10
Safety devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Probe components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Design and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Probe elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Principles for assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Screwing technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Clamping technique. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Storage of probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Selecting the installation site . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Prerequisites for installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Connecting the power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Connecting the compressed-air supply . . . . . . . . . . . . . . . . . . . . . 5-3
Requirements for compressed-air supply . . . . . . . . . . . . . . . . . 5-4
Compressed-air connection to the CMM . . . . . . . . . . . . . . . . . 5-4
Pressure gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Adjusting the pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Filter units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Care measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Inspection measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Checking the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Cleaning/changing the filter mat . . . . . . . . . . . . . . . . . . . . . . . 9-9
Changing the compressed air filters . . . . . . . . . . . . . . . . . . . . 9-10
Index
Glossar
! Danger!
Special caution is advisable here. The warning triangle and the adja-
cent text indicate a possible danger of injury.
Failure to observe this warning may result in injury to the operator.
! Important!
This symbol warns you of situations which might cause measuring
errors, interfere with a measuring run or even lead to collisions result-
ing in damage to the CMM and/or the workpiece.
Example Description
not – Words are emphasized by placing them in
italics.
– Italics are often used to set apart an inter-
mediate heading, e.g. Type of
measurement:
Protective circuit Specially emphasized words are printed in
bold type.
no grease film Words specially emphasized in warnings are
also printed in bold type.
Blue bold type is sometimes used to mark a
listing, e.g.
ST-ATAC: ...
VAST: ...
➤"Measurement“ on Reference to a text passage where you can
page ... find more information.
➤ Page ...
1 Introduction
.................................................................................................................................
This chapter informs you about the applications of the CMM and the
prerequisites for safe use.
General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Additional documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Proper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General information
Scope of delivery
The standard version of the CenterMax coordiante measuring
machine is delivered with the following components:
– Coordinate measuring machine (CMM)
– Control
– Control panel – standard version
– Computer with peripheral units
– Software – e.g. UMESS-UX
– Documents
– Probe system
– Master probe
– Reference sphere – for calibration of probes
– Temperature sensor – for measuring the workpiece temperature
– Temperature probe.
CE marking
The CE marking confirms the system's conformity with the basic
requirements of the following CE directives:
– Machinery directive 98/37/EC
– Low voltage directive 73/23/EEC
– EMC directive 89/336/EEC.
The fact that the CMM meets the above requirements is indicated by
the CE marking on its type plate.
Additional documents
Separate documents
Components for which separate documents are available are referred
to in these operating instructions.
Separate documents are available for the following components:
Standard The following documents are delivered with each CMM:
– General safety instructions
– Control console – standard.
Option The following document is delivered if the CMM is equipped with the
corresponding option.
– Loading device.
Supplementary literature
"Einfach messen – und was Sie dazu wissen sollten". The primer of
measurement technology.
Carl Zeiss, Unternehmensbereich Industrielle Messtechnik, order no.:
612302-9002.
Warranty
Notes
– The specifications and statements in the German operating
instructions are prevalent and binding for translations into other
languages.
– All rights pertaining to changes in the coordinate measuring
machine and its options, the software packages and the pertaining
documents reserved.
– All rights reserved, especially in cases of granting a patent or regis-
tering a utility model.
The warranty does not cover the following:
– incidentals
– wearing parts.
Exclusion of warranty
The manufacturer cannot be held liable.....
– for actions contrary to the instructions in this manual,
– if the coordinate measuring machine version delivered by us was
changed,
– if maintenance work is not carried out by personnel specially
trained at Zeiss,
– if measures for care are not taken by the operator or user accord-
ing to the specified measures,
– if original spare parts are not used for maintenance and repair
work,
– if the necessary maintenance work and measures for care are not
carried out according to the Customer Documentation.
All information regarding maintenance work and measures for care,
incidentals and wearing parts is contained in the Customer
Documentation manual.
Proper use
Warranty
The coordinate measuring machine (CMM) may be used only for its
intended purpose. The user is liable for any damage incurred during
improper use of the CMM.
CMM
The CMM can be used to determine the geometric sizes of work-
pieces. The workpieces may be made of metal or plastic.
Movable axes The CMM can be moved in all axes and probes the workpiece in the x,
y and z coordinate directions. A rotary axis is available when using a
rotary table.
Dimensions, position Geometric sizes are for example width, length, height as well as the
and form diameter and depth of bores.
The position of bores is calculated from the measuring data. Further-
more, the form of workpieces, e.g. the form of a turbine bucket can
be determined by means of a special software.
! Danger!
The CMM must not be used for determining geometric sizes of work-
pieces. The CMM must not be used for any purposes other than those
listed below.
Only workpieces and objects required for measurement may be
placed on the measuring table.
Improper use
The CMM must not be used for purposes other than the proper use.
Examples for use contrary to the proper use:
– The machine axes must not be used to move objects.
– The z axis of the CMM must not be used to lift objects.
– The CMM must not be used to measure living objects or parts of
the body.
Probe system
The probe system is a hightech product which may only be used only
for its intended purpose.
Probing CMM probe systems are designed to determine the coordinates of a
workpiece. This is usually achieved via a probing, during which the
workpiece is probed by a probe element (stylus tip). In some cases
optical measuring techniques are used.
Linear scales The probe system comprises several components, such as the probe
head and linear scales. The linear scales are integrated in the machine
axes of the CMM and protected by covers. These covers must be
attached when the CMM is in operation to prevent it from being used
improperly.
Probe
The probe consists of several components: adapter plate, probe styli
and probe joints. A probe may be equipped with one or more styli.
The probe element is located at the end of the stylus.
Functions of the probe:
– The adapter plate holds and positions the probe exactly in the
probe head.
– Workpiece probing is carried out by the probe element.
Probe head
The probe head is inserted in the quill. The probe used for probing is
inserted below the probe head. Handle the probe head and probe
very carefully.
Functions of the probe head:
– Holding the probe.
– Exact positioning of the probe.
Improper use
The probe head and probe must not be used for purposes other than
the proper use.
Examples:
– The probe head must not be used as a support.
– The probe must not be used as lever arm, e.g. to loosen a ring bolt.
– The probe must not be used as striking tool.
Safety
Definition of a specialist
A specialist is a person who due to his
– training
– experience
– instruction
and knowledge of
– relevant standards
– legal regulations
– accident prevention regulations
– and operating conditions
has been authorized by the safety officer responsible for the CMM to
perform the currently required actions and therefore is in a position to
recognize and avoid potential dangers.
! Danger!
Any work on the control cabinet must be carried out only by an elec-
trical engineer.
! Danger!
Components in the control cabinet are still live, e.g. line filter, even
after switching off the main switch.
Danger symbol:
This symbol indicates a danger for persons and
the CMM and its components.
This symbol is attached to the CMM, the con-
trol panel or the control cabinet.
• Only use power cables and connectors which are in perfect condi-
tion.
• Insert the power connectors only in sockets equipped with a per-
fect protective ground connection.
• Do not exceed the maximum permissible workpiece weight.
No warranty claims The manufacturer of the machine cannot be held liable for any dam-
age caused by unauthorized interventions in the measuring system. In
case of unauthorized manipulations, all warranty claims against the
manufacturer and supplier as well as the validity of the EC declaration
of conformity are null and void.
Documents Correct operation of the CMM is required for a safe measuring run. It
is assumed that you are familiar with the documents included in the
scope of delivery.
• Ensure that the operating instructions for the CMM are always
available within easy reach of the CMM.
Safety devices
The coordinate measuring machine (CMM) is equipped with safety
devices. Some devices are standard devices, others are optional.
Standard The standard safety devices include:
– The EMERGENCY STOP button on the control panel
– The quickstop button on the control cabinet
– Thrust force limiting (hardware) and drive monitoring (software)
– Drop protection of the quill
– Collision protection for the probe head and the probe; effective in
manual mode (vmax = 70mm/s).
Option Optional safety devices are:
– Light barrier
– Footswitch mats.
Identification of a CMM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Design of the CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Components and their functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Probe rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Identification of a CMM
The CenterMax is identified by the size of its working range. This
range is specified by its length in the x, y and z axes.
Identification of CenterMax:
Identification: 11/12/7
Ref. no. Working range
11 x: 1100 mm
12 y: 1200 mm
7 z: 700 mm
NOTE The working range does not correspond to the measuring range.
Please refer to the technical data for more information on the measur-
ing range.
Type plate The identification of your CMM is given on the type plate. The type
plate is located on the front side of the measuring table.
1 2 3 4
1 Serial number
2 Year of manufacture
3 Number of the CMM type
4 Series and identification of the CMM type.
2
6
3
1 Quill cover
2 X-bridge with concertina cover
3 Drive-side and guideway-side
4 Probe head
5 Measuring table
6 Guideway-side
Bridge
The CenterMax is a bridge coordinate measuring machine. The bridge
is composed of a crossbeam and two guideways. The quill is guided
on the crossbeam of the x-bridge.
3
4
1
1 Crossbeam
2 Drive side and guideway side
3 Quill
4 Guideway.
Measuring table
The workpieces to be measured are placed on the measuring table.
Threaded holes Threaded holes are provided in the measuring table. These are used to
fasten workpieces and calibration tools on the measuring table. The
threaded holes are distributed on the measuring table. For more infor-
mation, please refer to the document "Installation Instructions“.
Measuring volume The measuring volume is the space in which workpieces can be
probed.
1 Measuring volume
Coordinate axes
Travelling in the x, y and z axes is possible with the coordinate mea-
suring machine.
Travel directions The drawing shows the possible travel directions of the bridge.
x direction
Crossbeam The x-bridge crossbeam is mounted on the columns and supports the
quill. The quill is guided on the crossbeam and can be moved back
and forth. Thus, probing in the x direction is possible.
y direction
Columns The drive and guideway side columns support the cross beam and the
quill and the frictionless air bearings enable movement of the com-
plete bridge in the y direction.
z direction
Quill The quill is guided along the crossbeam. It can be moved in the
vertical direction and allows probing in the z direction. The probe sys-
tem is located at the bottom of the quill.
Control
Control cabinet
32 bit / C99 The control for the CenterMax is located in a separate control cabinet.
The CenterMax is equipped with a 32 bit control. The control has the
designation C99.
IP54 The housing of the control cabinet has protection class IP54.
Front side of the control cabinet:
2
3
1 Operating elements
2 Type plate for control cabinet
3 Door lock
4 Warning symbol for electric voltage: The control cabinet must be opened
only by an electrical engineer.
1
2
3
4
5
6
7
8
9
10
1 Fan
2 Air filter
3 Warning label: Replace or clean soiled air filter!
4 Door lock
5 Warning symbol for electric voltage: The control cabinet must be opened
only by an electrical engineer.
6 Plate with technical data
7 Attention symbol: Observe the information on the plate!
8 Connector for peripheral devices
9 Power supply of the control cabinet
10 Cables leading to the CMM.
8 7 6 4 2 1
5 3
Control Rotary switch (2) and indicator lamp (3): The rotary switch is used to
switch the control on. If it is in the ON position, the control is
switched on. The indicator lamp above the switch lights up.
Drives Rotary switch (4) and indicator lamp (5): The rotary switch is used to
switch the drives on.
Turn the switch clockwise. The switch does not remain in the new
position. It returns to its initial position.
When the drives are switched on, the indicator lamp above the switch
lights up.
Operating modes Key-operated switch (6): The key-operated switch is used to deter-
mine the operating mode in which the measurements are to be car-
ried out.
– MAN: The measurement is carried out manually. The workpiece
is probed by means of the joysticks on the control console. The
maximum speed is limited ➤“Travel speed” on page 3-3.
– OFF: No travel movement possible.
– AUTO: The measurement is carried out automatically, e.g. by
means of a CNC program. Before setting this operating mode, you
should test the CNC program in the MAN operating mode.
Working-hour meter The working-hour meter (7) counts the number of hours during
which the drives remain activated.
Quickstop button Push-and-turn switch (8): The quickstop button can be pressed to
stop all travel movements in cases of emergency. It engages when
pressed. All drives are switched off.
The quickstop button must be released before switching on the
drives. Release the button by turning it counterclockwise. Then, the
drives can be switched on again by means of the rotary switch
➤ page 2-8.
Control console
The CenterMax is equipped with the standard control console on
delivery. However, it can also be delivered with the Dynalog control
console on request. Each of these control consoles is described in a
separate manual.
1 3
4
2
EMERGENCY STOP but- The drives are switched off by pressing the EMERGENCY STOP but-
ton ton. Press the button to switch off the drives. The button engages. No
further travel movements are possible.
The button must be unlocked in order to switch the drives on again.
• Unlock the button by turning it.
Then the button is pushed up. The button is unlocked.
Right joystick
x and y direction The right joystick is used to move the probe in the x and y direction.
The bridge moves in the y direction, the quill in the x direction.
NOTE You may reverse the travel direction of the joysticks. This is advisable
when operating the control console and performing probing from
behind the CMM.
You can reverse the travel direction by pressing a button on the con-
trol console ➤ Operating instructions for the control console.
Left joystick
z direction and direc- The left joystick is used to move the quill in the z direction.
tion of rotation of the
rotary table (option)
NOTE If the joysticks are not activated within a certain period of time, they
will be blocked. The blocking must be cancelled before it is possible to
carry out travel movements by means of the joystick ➤ Operating
instructions for the control panel.
Calibration standard
A high-precision sphere is used as the calibration standard. The sphere
is made of a ceramic material and has a diameter of 30mm.
The calibration standard shown is included in the scope of delivery.
The calibration standard has a magnetic stand.
1 Calibration sphere
2 Rotary switch for activating the magnet
3 Magnetic stand
NOTE A metal plate is required for mounting the calibration standard on the
measuring table. This plate is delivered with the calibration standard.
For more information see ➤“Mounting the calibration standard on
the measuring table” on page 7-18.
Options
The following special equipment is available for CenterMax:
– Loading device
– Illuminating device.
Probe system
VAST
Application Universal
Measuring principle measuring
Further information ➤ page 2-15
VAST
Probe head ×
Adapter plate ×
Probe stylus ×
Registration + transmis- The probe head holds the adapter plate and the mounted probes.
sion Furthermore, the probe head registers the deflection of the adapter
plate and the probe during probing. A signal is sent to the computer
after each deflection.
VAST
Probe head VAST
Master probe 1
Adapter plate 1
Probe rack
one rail eightfoldb
Adapter platea for 2
probe rack
a The adapter plate is identical with the adapter plate for the master probe.
b
Number of holders
Components
1
2
limiting values
The limiting values for weight, length and torque of a probe must be
considered when assembling or mounting the probe.
Probe weight
600g The adapter plate receptacle may only be loaded with a maximum
weight of 600g.
0 450mm Probe length
The length of the probe head including extension must not exceed
450mm.
Torque
The maximum torque of the probe is 0.1Nm.
! Important!
The weight of all components and the length of the probe styli, exten-
sions and joining elements must be considered when assembling
probes.
• Observe also the weights indicated for the individual probes, join-
ing elements and the adapter plate when assembling probes.
• Observe the lengths indicated for the probes, extensions and other
joining elements when assembling probes.
Temperature-sensing probe
Application
The temperature-sensing probe is used to measure the the tempera-
ture of the workpiece surface and the surrounding air. Both measure-
ments can also be carried out via a CNC program. Thus, temperature
monitoring is possible throughout the entire measurement.
NOTE The VAST probe system is required for the operation of a tempera-
ture-sensing probe.
Design
A Adapter plate
B RST temperature-sensing probe
C Temperature sensor
1 Connector
2 Knurled ring with internal thread.
Probe
Probe components
The probe consists of
– adapter plate
– one or more styli
– joining elements.
Probe configuration A distinction is made between probe configuration and probe combi-
nation. The probe configuration may consist of one or more styli. Sev-
eral styli may form a probe combination.
1 Probe combinations
2 Probe configuration
The adapter plate is inserted in the probe head and held by magnetic
force. The groove in the adapter plate allows correct positioning. The
pin in the adapter plate receptacle of the probe head must engage in
this groove.
The VAST adapter plate is equipped with a probe mounting cube with
five threaded holes used for mounting probe components. The size of
the connection thread is M5.
A
1
1 Anchor plate
2 Groove for pin in the adapter plate receptacle of the probe head; the pin
allows correct positioning of the adapter plate.
3 M5 threaded hole for probe components, e.g. styli.
Joining elements
The probe kits for the probe systems contain probes or styli of differ-
ent sizes and components used to create probe configurations, e.g.
extensions.
Probes
Shaft + probe element A probe or stylus consists of a shaft and a probe element or stylus tip.
Shafts differ with regard to size and material. Furthermore, probe ele-
ments differ with regard to form ➤ page 2-21.
Probe elements
Probe elements may have various forms. Examples of probe elements:
1 2 3 4
1 Sphere
2 Cylinder
3 Cone
4 Disc
Cylinder Cylinders are preferably used for probing thin sheet-metal parts and
narrow workpiece edges.
Disc A disc is used for probing deep grooves and large holes. For large
holes, a disc offers the advantage of shorter travel paths as compared
with the sphere.
! Important!
The limiting values for probes must be observed when assembling
probes.
Screwing technique
The probe components have threads. In this way the components can
be screwed together. Depending on the probe system, different
thread sizes are used: M2, M3 and M5.
– M5: VAST
Probe kits
There are different probe kits, e.g. for small and large workpieces, for
each probe system.
Clamping technique
Laterally oriented styli are inserted in grooved discs and clamped by
the vertically oriented stylus. The end stylus is screwed into the probe
extension or directly into the adapter plate. In this way the laterally
oriented styli are clamped.
The probe components for the clamping technique and assembly
instructions are included in the probe kit.
Advantages and disad- Advantages:
vantages of the clamp-
– high flexibility
ing technique
– low weight.
Disadvantages:
– high measuring inaccuracy caused by poorly definable clamping
joints.
– with VAST: loss of rigidity at each joint.
Storage of probes
Once the probe has been assembled, it can be used as long as
required. It does not need to be disassembled. Careful handling is
necessary. This applies particularly to the adapter plate and probe ele-
ments.
Probe protection The probe must be protected against the following influences:
– Force of impact and other external mechanical infuences
– Dust and dirt.
A probe rack is recommended for storage if you measure every day
with the CMM. When no measurements are carried out with the
CMM over a longer period, the probes should be stored in a place
protected against environmental influences:
Probe rack
Order no.
1038-094 Included in the CenterMax standard
version
1038-093 Option
NOTE The probe rack with the order no. 1038-093 is mounted at the drive
end. However, the measuring range in the x direction is reduced.
3 Technical data
.................................................................................................................................
The following text provides technical data on the CMM and the probe
system.
CMM: Data referring to dimensions, measuring range, weight, noise
level, parameters, environmental conditions, requirements for the
electric power supply and compressed-air supply.
Probe system: Data referring to several probe systems, e.g. limiting
values and environmental conditions.
General data
CenterMax 11/12/7
Dimensions
Width [mm] 2090
Length [mm] 2126
Heighta [mm] 3000
Measuring range
x [mm] 1100
y [mm] 1200
z [mm] 700
Weight
CMM [kg] 6000
Workpiece [kg] 1000
Noise level of the CMM [dBA] <70
a An additional distance is required for mounting work: min. 200mm.
CMM parameters
Category Characteristic
value
Overvoltage category III
Pollution degree 2
Protection class 1
Connection data
Electric data
Compressed air
Certain requirements have to be met before connecting the com-
pressed air in order not to impair the function of the CMM.
Requirements
Pressure 6 - 10bar
Degree of purity 99%; without solid particles, gas-
eous proportions of oil and water
Air consumption max. 20l/min with 5bar
Travel speed
Environmental conditions
The following conditions must be fulfilled to guarantee perfect opera-
tion of the coordinate measuring machine.
Probe system
CMM dimensions
Length 91mm
Width 91mm
Height 200mm
Probing directions 6; ±x, ±y, ±z
Measuring force 0.05 to 1N / continuous
Diameter of adapter plate 69mm
Probe head deflection max. ±5mm
Probe weight, max. 600g / incl. adapter plate
Probe length, max. 450mm / stylus + extension
Repeatability following probe <1µm / with 200mm probe
change length
Environmental conditions
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Transport
Separate document
NOTE You received the document "Installation Instructions“ prior to the
delivery of the coordinate measuring machine. This document informs
you about all measures to be taken for transport and installation. In
the following, we will refer to some of the most important points con-
tained in this document.
To be observed on delivery
The coordinate measuring machine is delivered disassembled on
transport pallets. The individual parts are wrapped in insulated pack-
aging. For overseas transport, the individual parts are packed in ship-
ping containers.
Requirements regard- A fork-lift truck with sufficient carrying force has to be used for trans-
ing the fork-lift truck port. The fork-lift truck must be suitable for the weight of the individ-
ual transport containers. In order to avoid tilting of the load, the forks
must be completely underneath the transport pallets or shipping con-
tainers and the fork width must be variable. You will find the weight
and dimensions of the individual parts in the document "Installation
Instructions“.
! Important!
The packing material or the transport containers must not be dam-
aged. The packing material may only be removed at the installation
site by a ZEISS service engineer.
Ambient temperature The transport pallets and shipping containers must be stored in a cov-
+5°C to +40°C ered area until the machine is installed. The ambient temperature in
the storage area must be within +5°C to +40°C.
Transport conditions
The transport route from the storage area to the installation site must
be selected carefully. It might be necessary to prepare it in order for it
to meet the requirements.
Prerequisites for trans- Observe the following regarding transport:
port
– Weight: Is the bearing capacity of the floor along the routes suit-
able for the weight of the transport containers and the fork-lift
truck?
Installation
Separate document
NOTE You received the document "Installation Instructions“ prior to the
delivery of the coordinate measuring machine. This document informs
you about all measures to be taken for transport and installation. In
the following, we will refer to some of the most important points con-
tained in this document.
Connections
NOTE For more information on the electric power supply, please refer to the
document entitled "Installation Instructions“.
2
3
Pressure gages
There are five pressure gages for five devices. These devices and their
required pressures are listed below:
1 2 3 4 5 6 7 8
Indicated pressures:
Device No. Required pressurea
y drive 3 2bar
Weight balancer 4 5bar
Option: 5 5bar
Option: e.g. loading device 6 5bar
Air bearings in the CMM 7 5bar
a
The values set by the ZEISS service engineer may slightly deviate from the indicated
values.
! Important!
The valves of pressure gages which are not connected must be closed.
The connection shank should be covered by a cap.
Compressed-air valve
with pressure gage
1 2 1 3
1 Control knob for pressure adjustment – locked; only for service engineers
2 Air pressure gage
3 Connection of the compressed-air line.
! Important!
The pressure is adjusted individually by a ZEISS service engineer during
the initial start-up. After this, the pressure must not be readjusted.
If the required pressure is not indicated, proceed as follows:
• Identify the cause of the decrease or increase of pressure ➤“Faults
during the measuring run” on page 8-3.
• Call a ZEISS service engineer if you cannot find or eliminate the
cause.
Filter units
Several filter combinations are offered for the CMM. The CMM fea-
tures a combination of a fine filter and an extra fine filter as standard
equipment. If the required compressed air quality cannot be main-
tained, additional filters can be installed later. The possible versions
are specified below::
Combination
1 Fine and extra fine Standard
filter
Combination
2 Combination 1 + Optional
membrane dryer
for filtering out
moisture
3 Combination 2 + Optional
combination 1
Installation procedure
The holders are preassembled on the profile rail. The profile rail can be
attached directly to the guideway side.
! Important!
Exact probe rack alignment is necessary to enable trouble-free probe
change. The deviation must not exceed 0.1mm over the total length
of the profile rail.
• Align the probe rack so that the deviation of the profile rail does
not exceed 0.1mm.
1
2
The holders are fastened to the profile rail using Allen screws. Two
screws are required for VAST holders.
A Cap
B Holder for probe system VAST;
a: top plan view; b: front view
1 Mounting aid
2 Screw; M6×16
3 Cover of a holder
4 Brush for cleaning the adapter plate
• Distribute the holders evenly on the profile rail and tighten the
screws (2).
2
3
NOTE If you require more holders than can be mounted on two profile rails,
it is possible to mount a third profile rail on the drive side. However,
this will decrease the effective measuring range.
Visual check
NOTE The visual check must be part of the routine work to be carried out
during the daily measuring operation.
Housing covers
The CMM may only be operated if all CMM housing covers are
mounted.
• Mount all housing covers before taking further measures.
Compressed-air line
The line and connections must be in perfect condition. The line must
not be ...
– ...bent or
– ...damaged.
• Make sure that the line is in perfect condition.
• Make sure that ...
– no heavy objects are lying on the line.
– the line is connected correctly to the supply line and the CMM.
Cables
Cable duct The cables from the control cabinet to the CMM are installed in a
cable duct.
• Make sure that the cable duct is not damaged.
Power supply
– Is the power supply connected?
– Are the cables in perfect condition?
Compressed-air supply
– Is the compressed-air supply connected?
– Does the compressed-air supply have sufficient pressure?
– Is the required pressure indicated?
6 Start-up
.................................................................................................................................
This chapter describes the work to be carried out regularly.
You want to clamp a new workpiece on the measuring table or pre-
pare the probe system for the measuring run? Or you want to switch
the CMM on again after it has been put out of operation? The follow-
ing pages provide the corresponding information.
Furthermore, the procedure for checking the protective circuit for the
drives is described.
Safety measures
! Danger!
You are exposed to the danger area of the CMM when working on
the measuring table. Certain safety measures have to be observed to
avoid injuries.
• Switch off the drives to prevent unintentional travel movements.
• Activate the "crane interlock“ when using lifting equipment to
place the workpiece on the measuring table.
For more information see ➤ Operating instructions for the measur-
ing software.
! Danger!
There must not be any grease film on the measuring table. Work-
pieces may slide and fall off of the measuring table, thus endangering
personnel. Make sure that the measuring table is free of lubricants
before placing a workpiece on top of it.
• Remove any greasy or oily substances from the measuring table.
! Important!
The workpieces must have full contact with the measuring table. Any
hard particles lying underneath the workpiece may lead to inaccurate
measurements. Furthermore, the surface of the measuring table and
the workpiece may be damaged.
• Clean the measuring table before placing a workpiece on the mea-
suring table.
! Important!
Certain safety measures have to be observed to avoid damage to the
CMM and the measuring table.
• Observe the permissible workpiece weight ➤ page 3-2.
• Lower the workpiece onto the measuring table carefully. Lower
heavy workpieces at a maximum speed of 2mm/s.
The CMM may be damaged if the workpiece is set down hard.
• Make sure that the workpiece or lifting device does not strike the
bridge, quill, or drive side of the CMM.
Even light jolts may affect the functioning of the CMM.
• Place the workpiece upright on the measuring table.
• Do not slide the workpiece on the measuring table.
The surface of the table could be scratched when sliding the work-
piece, thus impairing the measuring accuracy of the CMM.
! Important!
The maximum permissible torque for the threaded holes in the mea-
suring table is 40Nm.
• Tighten the screws to a maximum torque of 40Nm.
Chucks
Chucks are screwed onto the measuring table using the threaded
(M12) holes.
Metal, plastic, wood Chucks are generally made of metallic materials. Chucks made of
plastic or wood may also be used if necessary.
Workpiece positioning
Position the workpiece so that all required measurements can be car-
ried out without changing the workpiece position.
Measuring operation When using a probe rack, position the workpiece and probe rack so
without collision that measurements and probe change can be carried out without col-
lision.
• Arrange the workpiece and probe rack so that no collisions can
occur during the measuring run.
For more information on installing a probe rack,, please refer to .
Heavy workpieces
Heavy workpieces have a hard impact if lowered into the measuring
table too quickly.
• Lower heavy workpieces at a maximum speed of 2mm/s.
Elastic supports • First place the workpiece on elastic supports.
• Then replace the elastic elements with rigid supports.
NOTE Metal blocks clamped by clamping devices can be used as rigid sup-
ports. Screw fasteners can be used alternatively. These fasteners can
be screwed into the threaded holes in the measuring table.
Safety measures
! Danger!
A visual check must be carried out before switching the CMM on. The
CMM may be switched on only if it is in perfect condition.
• Make sure that the CMM is in perfect condition ➤“Visual check”
on page 5-14.
Switch-on sequence
Proceed as follows when switching the control on:
1 Connect power supply; switch power supply on by means of the
main switch
2 Switch the control on.
3 Switch the drives on.
4 Select the operating mode
5 Switch the computer on and start the measuring software.
0125853 h
6 4 2 1
5 3
NOTE For more information on the operating elements, please refer to the
chapter ➤“Operating elements of the control cabinet” on page 2-8.
! Danger!
When starting a CNC program, make sure that no one is in the dan-
ger area of the CMM. The operating mode setting (MAN or AUTO) is
irrelevant.
• Check to make sure that no one is in the danger area of the CMM.
! Important!
The AUTO mode may only be set if the CNC program was tested in
the manual mode beforehand. This prevents collisions at high travel
speeds in the automatic mode.
NOTE When starting a CNC program, the key should be removed and stored
in a safe place to prevent the setting from being changed accidentally.
! Danger!
In order to ensure the safety of the CMM during measurement, the
function of the protective circuit must be checked regularly.
• Check the protective circuit on a monthly basis.
Check
When performing the check, you must proceed as follows:
x axis
• Move the crossbeam forward just far enough so that you can
grasp the quill housing with one hand.
vx F
! Important!
You may press only against the quill housing.
• Do not press against the quill or against the probe head.
The probe head and quill could otherwise be damaged.
Check by a ZEISS service Under normal circumstances, the force of your hand should suffice to
engineer stop probe movement. If the movement can not be stopped in this
way, the protective circuit must be checked immediately by a ZEISS
service engineer.
• Call your service engineer and arrange for him to check the protec-
tive circuit.
! Danger!
If the protective circuit is not functioning properly, the CMM must be
taken out of operation until the protective circuit has been checked by
a ZEISS service engineer.
Reactivating the drives If you can stop the travel motions, you must reactivate the drives
➤ Operational controls.
y axis
• Move the control panel to the drive side for y-axis travel.
The drive side is located on the right side of the CMM.
vy F
! Important!
You may press against the crossbeam only on the drive side.
• Do not press against the probe head and quill or against the cross-
beam on the guideway side.
The probe head, quill and crossbeam could be damaged.
! Important!
With the CenterMax, the probe head must not be changed by the
operator. The probe head may be changed only by a ZEISS service
engineer.
Probe assembly
Safety measures
! Important!
Probes must not be assembled on the probe head. The probe head
could thus be damaged.
• Do not attach any probe elements to the probe while it is inserted
in the adapter plate receptacle.
! Important!
Assemble the probe configuration carefully. Do not exert any force
when screwing the components together.
• Screw the components together manually. Assemble them fully.
• Use the pin included in the probe kit to tighten the components.
! Important!
Observe the limiting values regarding weight and length when assem-
bling the probe. The limiting values depend on the probe system.
• Observe the limiting values ➤“Limiting values” on page 6-14.
NOTE Check the balance by holding the adapter plate between two fingers
and letting it swing back and forth until it comes to a stop. The probe
must point in a vertical downward direction.
– perfect condition
– clean
– grease-free
• Observe the limiting values; probe length and probe weight.
Limiting values
The maximum weight and length must not be exceeded in order to
ensure error-free measurement by the probe system.
The limiting values depend on the probe system used.
Example
Connecting parts are usually required to assemble the required probe
configuration. It is also possible to screw the probe directly into the
adapter plate ➤ page 6-15.
Screwing technique Example of probe assembly via the screwing technique using the
VAST adapter plate:
A B
! Important!
Probes must not be assembled on the probe head. The probe head
could be damaged.
• Insert only finished probes equipped with all necessary probe ele-
ments.
• Do not attach any probe elements to the probe as long as it is
inserted in the adapter plate receptacle.
! Important!
A VAST probe automatically falls off of the adapter plate receptacle
after a specified period of time.
• Hold the probe immediately after activating change in order to
avoid damage to the probe, the workpiece or the measuring table.
The period after which the probe falls off is set in the measuring soft-
ware.
Start-up checklist
Clamping the workpiece
– Is the workpiece in the correct position? ➤ page 6-3.
– Is it possible to measure all workpiece dimensions in one cycle?
– Is measurement possible without collision? E.g. collision with the
probe rack.
– Is the workpiece clamped? ➤ page 6-3.
Probe assembly
– Have the criteria for probe assembly been observed? ➤ page 6-13.
– Have the limiting values for the probe been observed?
➤ page 6-14.
7 Measuring operation
.................................................................................................................................
Connections carried out, workpiece clamped on measuring table,
probe system prepared. The preparations for the measuring run can
begin.
This chapter describes the procedure for correct measurement with
the CMM. Furthermore, the possible causes of unusual measured
results are also explained.
Coordinate systems
There are three coordinate axes corresponding to the travel move-
ments: x, y and z. Together these axes form a machine coordinate sys-
tem. Two additional coordinate systems are required for exact
calculation: the workpiece and control coordinate systems. The coor-
dinate axes are assigned the letters G, W and S to identify of the coor-
dinate systems:
– Machine coordinate system; xG, yG, zG
– Workpiece coordinate system; xW, yW, zW
– Control coordinate system; xS, yS, zS
For more information on the coordinate systems, please refer to the
➤ operating instructions for the measuring software.
Probing directions
Six (=6) probing direc- Workpieces can be probed from six directions, ±x, ±y, ±z. The illustra-
tions tion below shows five probing directions. The rectangular block
shown cannot be probed in the positive z direction. However, in case
the block has lateral recesses and bores, it is possible to probe in the
positive z direction by using a corresponding probe configuration.
3
4
1 Probe
2 Rectangular block (=workpiece)
3 Possible probing directions
4 Measuring table
Probing oblique work- It is also possible to probe oblique surfaces perpendicular to the prob-
piece surfaces ing surface. This is done by means of the measuring software. The
workpiece coordinate system is aligned corresponding to the work-
piece surface ➤ Operating instructions for the measuring software.
NOTE Clamp the workpiece on the measuring table before probing. Other-
wise, the workpiece may be shifted by the probing. Correct measure-
ment is not possible if the workpiece is shifted.
Types of measurement
Overview
Measurement in static The measuring data are not accepted immediately after probing.
status Instead, they are initially polled following a damping period of approx.
one second. I.e. the actual measurement occurs only after the
machine reaches a static state.
Multipoint measurement
Multipoint measurement is only possible with the VAST probe system.
During multipoint measurement, the probe is moved perpendicular to
the probing direction.
The probing axis runs parallel to the workpiece surface. During mea-
surement, the probe element constantly remains in contact with the
workpiece surface. The measuring force remains effective until prob-
ing is completed.
B
A C
A Lowering the probe and probing the workpiece; with the left joystick.
B - D Move the probe over the workpiece with the right joystick. The
position in the z direction remains unchanged.
D Lifting the probe.
There are two options You can influence the transfer of measured values individually. The
for transferring mea- measured value can be transferred either dynamically or statically.
sured values.
– Dynamic transfer of measured values:
The measured value is immediately transferred by pressing the
push-button of the right joystick. The probe movement is not
stopped.
– Static transfer of measured values:
Each time the CMM stops traveling a measured value is created,
but not transferred immediately. Measurement transfer occurs fol-
lowing a damping period of approx. one second.
Terminating the multi- To terminate the measurement, deflect the joystick counter to the
point measurement probing direction.
Scanning
Scanning is only possible with VAST probe system. During scanning,
the surface of a workpiece is continuously probed. The measuring
points make it possible to calculate surfaces or provide information on
the form of a surface.
No axis clamping The axes of the probe head are generally not clamped during scan-
ning. I.e. measurement takes place with freely movable axes.
The measuring software offers several scanning modes:
VAST circle scanning: for measuring shafts and bores whereby the
CMM automatically differentiates between inside and outside bores.
VAST surface scanning: for measuring surface-like workpiece areas.
VAST line scanning: for flatness measurements or acquisition of
curved shapes in predefined section planes. In this case, it is advisable
to clamp one axis.
Different measuring routines are available for each scanning mode:
Four VAST stages – VAST stage 1: exact acquisition of dimensions, position and
form.
Scanning with maximum precision for measuring dimensions, form
and position.
– VAST stage 2: rapid acquisition of dimensions, position and form.
Scanning with high dynamic for measuring dimensions, form and
position (2xV2).
– VAST stage 3: exact acquisition of dimensions and position.
Scanning with maximum precision for measuring dimensions and
position.
– VAST stage 4: rapid acquisition of the position.
Scanning with maximum dynamics for measuring the position.
Collision protection
All probe systems provide complete collision protection for the probe
head and probe in the setup mode.
After collision
If the quill, probe head or probe collides with an obstacle and the
drives are switched off, proceed as follows:
• Eliminate the cause of the collision.
• Reactivate the drives on the control cabinet.
• Turn the rotary switch clockwise approx. 90° and hold it in this
position until the indicator lamp above lights up.
Initializing the CMM If the drives cannot be switched on, the CMM has to be reinitialized.
For more information on the initialization procedure, please refer to
the ➤ Operating instructions for the measuring software.
NOTE If the quill or the probe head collides with an obstacle, the guideway
and other components of the CMM may be damaged. In this case,
perfect measuring operation cannot guaranteed.
• If measurements are not possible after a collision, call a ZEISS ser-
vice engineer or our hotline.
NOTE After a collision, the probe must be recalibrated and the workpiece
position (w position) must be determined again.
! Important!
In general, the measurement has to be carried out with the same axis
clamping as for calibration ➤“Probe calibration” on page 7-16.
Example: In case of unclamped axes, probing should be carried out
using a probe stylus. This stylus must be calibrated in the same way.
I.e.: All axes must be unclamped during the calibration of the stylus.
1
1
2
3
Clamped axes
You can cause the measuring force to act in a certain axis by clamping
specified axes.
Two axes are clamped when probing single points. The probe can be
moved in the axis of the probing direction.
Self-centering probing
If the probe can be moved freely in all axes, it will be centered auto-
matically, e.g. a conical probe in a bore. A combination of one
clamped and two unclamped axes is also possible.
NOTE All axes are unclamped via the measuring software. Axis clamping
must be carried out by means of the control console.
Measuring force
With self-centering probing it may be advisable to increase the mea-
suring force or to specify the direction in which the measuring force is
to act. The direction is specified by clamping the axes in which prob-
ing should not take place.
Measuring software or The measuring software makes it possible to alter the value of the
control console measuring force and to specify the direction in which the measuring
force is to act. The direction of the measuring force can also be set via
the control console.
Examples
One axis clamped Bottom of narrow V grooves.
z 1
y
2
x
3
z 1
y
2
x
2
Temperature-sensing probe
Notes on temperature measurement
– The temperature measurement should be carried out on thick-
walled workpieces.
– The workpiece surface must be clean and flat.
– The surface to be probed should be larger than the contact surface
of the temperature sensor.
– Measuring force: 1N.
– The workpiece should be probed in the axis direction of the tem-
perature-sensing probe.
The max. angle between the axis direction and the normals of the
workpiece surface must no exceed ±5°. Perfect measuring opera-
tion cannot be guaranteed with larger angles.
A
1
B 3
1 Temperature sensor
2 Workpiece
3 Deflection of the probe head
4 Maximum deflection of temperature sensor: ±5°.
Possibilities
A temperature sensing probe offers the following possibilities:
– Temperature compensation.
– Inclusion of the temperature-sensing probe in a CNC program.
– Temperature monitoring.
– Temperature check.
– Temperature recording.
Possibility Comment
Temperature com- When using the temperature-sensing probe for
pensation temperature compensation, the measuring
time after positioning the temperature-sensing
probe is approximately five seconds. During
this time, measured values are continuously
transferred, evaluated and saved.
Inclusion of the If the temperature-sensing probe is to be
temperature-sens- included in a measuring run, the position of
ing probe in a CNC the temperature sensor must be known before-
program hand.
• Determine the position of the temperature-
sensing probe.
Temperature moni- It is possible to define limiting values for the
toring temperature. Furthermore, it is possible to
define when the temperature measurement is
to be carried out.
There are two possibilities if the limiting value is
exceeded: A warning appears in the record or
the CNC program is cancelled.
Temperature check The ambient air temperature can be measured
at any time. The temperature is compared with
the limiting valuess.
There are two possibilities when the limiting
values is exceeded: A warning appears in the
record or the CNC program is cancelled.
Temperature If the data from the last temperature measure-
recording ment is required, the temperatures of the
workpiece and the air can be recorded.
Operating mode
The CMM allows measuring in two operating modes: manual and
automatic.
MAN operating mode In the manual mode the workpiece is probed by means of the joy-
sticks. The joysticks are located on the control console ➤ operating
instructions for the control console.
AUTO operating mode In the automatic mode the workpiece is measured by means of a CNC
program. The measuring software enables programming of CNC
runs.
! Important!
Before starting the CNC program in the AUTO operating mode, you
should test the program in the MAN operating mode. The test is a
safety measure to ensure that the measurement can be carried out
without collision.
Measuring software
In addition to the control console, the CMM is also operated by
means of the measuring software. There are different measuring soft-
ware versions for the UNIX and Windows NT operating systems. The
operation of the measuring software versions varies.
• Please read the corresponding chapters in the operating instruc-
tions for the measuring software before carrying out the working
steps.
Safety instructions
! Danger!
There is a risk of injuries during all movements of the CMM.
• Ensure that nobody is at risk during manual operation of the
machine.
• Keep a safe distance away from to the CMM during automatic
measuring runs.
! Danger!
The CMM is provided with a protective circuit which monitors the
driving forces during movements of the bridge in the x and y axes. If a
collision occurs, the protective circuit causes the drives to be cut out.
The function of the protective circuit must be checked regularly in
order to guarantee the safety of the CMM.
• Check the function of the protective circuit every month
➤ page 6-8.
! Important!
Radio signals may interfere with operation of the CMM.
• Therefore, do not use any mobile phones or walkie-talkies within a
3m radius of the CMM.
! Important!
For calibration, the same measuring force and clamping of the axes
must be set as later for workpiece measurement.
• Perform the calibration in the same way as you intend to probe the
workpiece later on, i.e. with the same measuring force and axis
clamping.
4 1
5 2
1 Reference point
2 Measuring volume
3 Safety position
4 Zero point
5 Coordinate system
! Danger!
The reference point travel is carried out automatically. Before starting
the reference point travel, make sure that nobody is at risk.
• Ensure that nobody is endangered by the travel motions.
! Important!
The reference point travel is carried out automatically. The probe head
first moves in the z direction until it reaches the mechanical end posi-
tion and then in the x and y direction.
Make sure that the reference point travel can be carried out without
any collision with components on the measuring table.
• Move the probe head to a position from where the reference point
travel can be carried out automatically.
Start by the measuring The reference point travel is started by the measuring software.
software
• Read the corresponding chapter in the ➤ operating instructions
for the measuring software.
Safety position after After the reference point travel, the probe head moves to the safety
reference point travel position (only with UMESS). The coordinates of this position are
defined in the measuring software. The coordinates must be within
the measuring volume.
Example for safety position: x= 30, y= -30, z= -30.
Probe calibration
NOTE The measuring software is required for the following procedure.
• Please read the corresponding chapters in the operating instruc-
tions for the measuring software.
Preparation
– Master probe
The master probe must be used for the reference measurement
➤“Carrying out a reference measurement” on page 7-19.
– Probes used for workpiece measurement.
4
3
5
1
Calibration procedure
After the reference measurement has been carried out, the probes
can be calibrated.
NOTE Semi-automatic calibration should be carried out for all probe sys-
hss
tems.
Tensor calibration should be carried out for the VAST measuring
probe system. This is important since the measuring force causes a
1 Insert the probe into the adapter plate receptacle ➤ page 6-16.
2 Open the menu for calibration ➤ operating instructions for the
measuring software.
3 Select the clamping status of the axes, i.e. clamped or unclamped.
The same axis clamping status must be set as will be required later
on for workpiece measurement to prevent measuring errors.
Axis clamping must be performed via the control panel.
4 Probe system specific actions:
• The measuring force must be set for the VAST probe system.
5 Select the stylus to be calibrated on the control panel.
4
3
5
1
6 Call the menu for calibration in the measuring software and probe
the calibration sphere ➤ Operating instructions for the measuring
software.
The calibration sphere diameter is output as the result.
7 Check the result.
Standard deviation The result is evaluated based on the standard deviation. There is no
maximum standard deviation which applies to all applications. The
tolerable deviation depends on many factors: e.g. resolution and
precision of the CMM, length and rigidity of the probe shaft, qual-
ity of the probe tip, cleanness of the environment. Normally, the
standard deviation should be within a range of a few micrometers
(e.g. 0.002).
If the result is not satisfactory, repeat the calibration or look for
plausible causes. It might be necessary to decide whether or not to
accept measuring inaccuracies.
8 Repeat points 2 to 6 for all other styli of the probe.
Temperature compensation
What is temperature compensation?
The measuring software calculates the workpiece geometry for the
material properties at 20°C. 20°C is the reference temperature. Since
the actual temperature deviates from 20°C, the geometric data have
to be corrected by a correction value.
Expansion coefficient The correction value is calculated by the measuring software. How-
ever, the measuring software needs to know the temperature of the
CMM and the workpiece. Furthermore, the expansion coefficient
must be entered in the measuring software ➤ operating instructions
for the measuring software.
Procedure:
• Enter the expansion coefficient for the workpiece material in the
measuring software.
• Connect the temperature sensors ➤ page 7-22.
Calculating the correc- The measured values are automatically transferred to the measuring
tion value software. The measuring software calculates the correction value.
All of the geometric data measured on the workpiece are corrected by
the correction value with respect to the reference temperature of
20°C. Exact workpiece measurement is thus guaranteed.
1 Copper block
2 Connector
Application of the sen- • Place the copper blocks at two different locations in or on the
sors workpiece to be measured.
! Important!
Regular inspection of the probes is required to ensure correct work-
piece measurement.
• Check the styli, the probe components and the adapter plate regu-
larly.
• Remove any particles or grease film from the probe elements and
the adapter plate.
• Replace any damaged styli.
Probing procedure
Selecting the type of probing
The type of probing depends on the measuring job and the required
number of measuring points. Position measurement requires only one
or several single point probings. Multipoint probing or scanning can
be used to measure surfaces. Scanning must be used for form mea-
surements.
Probing sequence
Probing conditions
For probing, certain conditions must be fulfilled. The conditions
depend on the probe system. The following conditions must be ful-
filled:
– Constant probing speed.
Probing speed
Probing should be carried out at constant speed.
The speed depends on the deflection range of the joystick. At the
beginning and end of the deflection range, the speed is proportional
to the deflection range. The speed between the beginning and the
end is constant. This speed corresponds to the probing speed.
Probing speed depen-
dent on the deflection
range of the joystick
v Speed
v1 Probing speed;
optimum speed during probing.
Maximum possible The maximum possible speed is reached at full deflection range of the
speed joystick. In the setup mode, the max. speed is approx. 70mm/s. The
actual speed depends on the standard value defined in the measuring
software and the position of the control knob on the control panel.
Proceed as follows:
1 Enter a standard value of 100% in the measuring software.
2 Adjust the speed using the control knob on the control panel.
If the control knob is turned clockwise as far as possible, the maxi-
mum possible speed corresponds to the standard value, e.g.
100%, defined in the measuring software.
The control knob allows to reduce the speed to zero. If the speed
reaches a value of less than 10% of the standard value, this will be
indicated on the control panel. An LED on the standard control
panel lights up.
• If the speed is too high for certain probings, reduce the maxi-
mum possible speed.
Turn the control knob counterclockwise to reduce the speed.
NOTE When probing oblique workpiece surfaces with an angle between 30°
and 50°, the probing speed has to be reduced.
• Loosen the three screws in the adapter plate so that the probe
configuration can be turned. Do not completely unscrew the
screws!
• Insert the probe in the adapter plate receptacle.
• Turn the probe configuration until the required position is reached.
• Hold the probe configuration and remove the probe from the
adapter plate receptacle.
• Tighten the screws with a torque of 2Nm.
NOTE The measuring force must be set prior to calibrating the probe. Set
the measuring force for each probe individually if required.
Why does the measur- During probing, the probe element causes changes of the workpiece
ing force have to be form. The extent of these form changes influences the precision of
set? the calculated values. In order to avoid falsified measuring data, the
measuring force should be adapted to the characteristics of the work-
piece material.
A B
0,2N – normal The measuring force can be set continuously. Normally, a measuring
force of 0.2N has to be used for measurement. If larger form changes
are to be expected, the measuring force should be 0.1N.
• Set the measuring force according to your requirements.
Notes on scanning
! Important!
When measuring at high travel speed and with low measuring forces
(F< 0.1N), the probe may lift off the workpiece for a short time. In
unfavorable cases, this may lead to incorrect measured values.
• Reduce the travel speed in case of low measuring forces.
For this reason, probe elements should be checked regularly during all
scanning operations and replaced if necessary.
• Check the condition of the probe elements and clean them.
For information on how to remove aluminum deposits, please refer
to ➤ page 9-6.
• If the probe element is damaged, replace the stylus.
NOTE The measuring software is required in order to carry out this proce-
dure.
• Please read the corresponding chapters in the operating instruc-
tions for the measuring software.
Shutdown procedure
After the CMM is put out of operation, the probe head remains in its
last position. The space coordinates of its position are lost.
In order to avoid faults when the CMM is put back into operation, the
measuring run should be terminated correctly. For information on the
operator‘s controls, please refer to ➤ page 2-8.
• Proceed as follows:
CMM 1 Move the probe head to a safe position.
This position must be inside of the measuring volume. Prepare for
the next start-up by positioning the probe head close to the refer-
ence point.
Computer 2 Terminate the measuring software.
3 Power down the operating system.
4 Switch off the computer.
Control cabinet 5 Switch off the drives.
• Press the quickstop button on the control cabinet and unlock it
again.
6 Switch off the control.
• Set the rotary switch to the "OFF" position.
7 Switch off the power supply.
• Turn the main switch counterclockwise to the "0" position.
• If the above mentioned errors can be excluded, put the CMM out
of operation and repeat the start-up procedure.
• If it is then still not possible to carry out a measuring run, call a
ZEISS service engineer or our hotline ➤ page 8-9.
The following table helps you to identify and eliminate faults occur-
ring during the measuring run.
NOTE A fault may have different causes. One or more measures may be nec-
essary to eliminate a certain cause. If two measures are proposed, this
does not mean that both of them have to be taken to eliminate the
fault.
Special measures
After collision
• Eliminate the cause of the collision.
• Switch on the drives.
• Turn the rotary switch on of the control cabinet clockwise and
hold it until the indicator lamp located above lights up.
• Initialize the CMM if the cause cannot be eliminated by means of
the above measure.
➤ Operating instructions for the measuring software.
• Then switch on the drives.
• Push against the marked spot on the cover and move the quill
back into the measuring volume.
Depending on the quill position, you have to push against the
cover from the left- or right-hand side.
! Important!
You may only push against the cover of the quill.
• Do not push against the quill or the probe head.
The probe head and quill may be damaged.
• Push against the marked spot on the cover and move the cross-
beam back into the measuring volume.
Depending on the crossbeam position, you have to push against
the crossbeam from the left- or right-hand side.
! Important!
You may only push against the crossbeam at the drive end.
• Do not push against the probe head and the crossbeam at the
drive side.
The probe head, quill or crossbeam may be damaged.
z direction: You must release the quill fall brake before moving the
probe head back into the measuring volume. Proceed as follows:
• Move the crossbeam in the y direction to the front area of the
measuring volume. Always observe the following: ➤ page 8-6.
Move the crossbeam forward far enough so that you can grasp the
cover with your hand.
The right-hand side of the cover (1) is provided with an opening. You
can access the switch for deactivating the fall brake through this
opening.
• Press the switch with a thin object while pulling the quill down-
wards.
! Important!
Do not pull the probe head.
• Hold the quill above the probe head (2) in order to pull the quill
downwards.
A Front view
B View from the right
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Maintenance
Purpose of mainte- Proper maintenance ensures …
nance work
– safe measuring operation,
– that there are no downtimes,
– that the CMM always measures with maximum accuracy.
In order to guarantee these features in the long run, the CMM
requires regular maintenance.
! Danger!
Maintenance work must be carried out only by skilled personnel.
These persons must have received special training on the correspond-
ing CMM qualifying them to carry out all necessary maintenance
work.
We recommend:
Components Examples
Safety devices (standard For example drive monitoring; colli-
equipment) sion protection in the Series measure-
ment mode
Compressed-air supply For example air bearings, pressure
controller, air filters, valves
Measuring system For example functioning, precision
Control For example operator’s controls on
the control cabinet; control panel
Care
Although the CMM is robust and easy to care for, a certain degree of
care is still required. The sensitive measuring system of the CMM
requires a certain degree of cleanliness. Even the smallest dust parti-
cles on the probe element or the calibration sphere may cause inaccu-
rate measurements.
Cleaning and checking Measures for care also include the regular cleaning and checking of
certain components.
For cleaning, you require:
– Mild detergent
– A fluff-free cloth, e.g. made of linen
– A vacuum cleaner
– If necessary, special solvent for removing aluminum deposits from
the probe element, e.g. KOH (10%).
! Danger!
When carrying out care measures, make sure that no travel move-
ments are performed.
• Switch off the drives on the control cabinet in order to enable safe
execution of the care measures.
hss
! Important!
• Do not use compressed air for cleaning.
NOTE Preventive care also includes making sure that all workpieces to be
measured are clean. The workpieces must be free from machining res-
idues (e.g. metal chips, oil) and dust.
• Clean the workpieces before placing them on the measuring table.
Overview
The components requiring care are listed in the following. The time
intervals apply only if all installation site requirements have been met.
Care measures
Measuring table
The workpieces must have full contact with the measuring table. Any
hard particle below the workpiece may lead to inaccurate measure-
ments. Furthermore, the surface of the measuring table and the
workpiece may be damaged.
! Danger!
There must not be any grease film on the measuring table. Work-
pieces may slide and fall off the measuring table, thus presenting a
risk to the personnel. Before placing a workpiece on the measuring
table make sure that the measuring table is free from lubricants.
• Remove any greasy and oily substances from the measuring table.
Threaded holes
Dust may accumulate in the threaded holes and threads. In order to
ensure perfect condition of the threads, remove any dirt deposit from
the threaded holes with a vacuum cleaner.
Probe elements
Probe elements may be soiled e.g. by dust particles. Furthermore,
material from the workpiece being probed may settle on the probe
element, especially during scanning.
• Clean the probe elements with a fluff-free cloth.
• Use a cleaning agent if required.
Make sure that the probe element is free from cleaning agent resi-
dues.
NOTE The styli must be handled carefully. If force is exerted, the glued bond
between the probe element and the probe shank may be separated
and the probe shank may be bent or broken.
• Do not exert any force when cleaning.
• Rinse the probe elements immediately with water and dry them.
hss
! Danger!
Caustic solvents may represent a health hazard. When using caustic
solvents, always take protective measures, e.g. wear rubber gloves
and safety eyeglasses.
• Observe the general protective measures when handling harmful
solvents.
NOTE Whenever the styli are not in use, store them in a safe place, ideally in
the probe kit.
Calibration sphere
The calibration sphere must be clean and in perfect condition to
ensure correct calibration.
• Clean the calibration sphere with a fluff-free cloth.
• Use a cleaning agent if required.
Make sure that the probe element is free from cleaning agent resi-
dues.
Adapter plate
The adapter plate must be protected against soiling.
• Store the adapter plate in a dust-free and clean place, e.g. in a
cabinet or drawer, whenever it is not in use.
Cleaning:
Wipe off the adapter plate with a dry or damp cloth. Dry the adapter
plate in order to avoid oxidation of its contacts.
If the adapter plate is dirty:
• Clean it with a mild cleaning agent.
• Then dry it.
Make sure that the adapter plate is free from cleaning agent residues.
The function of the adapter plate could otherwise be adversely
affected.
Probe head
The adapter plate and the adapter plate receptacle must be protected
against soiling.
• Put a protective cap on the adapter plate receptacle of the probe
head whenever no probe is inserted in the adapter plate receptacle
or the probe head is not in use.
• Store the probe head in a dust-free and clean place, e.g. in a cabi-
net or drawer.
Cleaning:
Clean the probe head with a dry or damp cloth.
If the adapter plate receptacle is dirty:
• Clean the adapter plate receptacle with a mild cleaning agent.
• Then dry the adapter plate receptacle.
Make sure that the adapter plate receptacle is free from residues of
the cleaning agent. The function of the adapter plate receptacle may
be affected.
Probe rack
Vacuum the probe rack and the storage positions with a vacuum
cleaner and clean them with a mild cleaning agent.
• Remove all cleaning agent residues.
Concertina cover
Vacuum the concertina cover with a vacuum cleaner and clean it with
a damp cloth.
! Important!
The cover must not be removed.
Housing covers
Vacuum the housing covers and clean them with a damp cloth.
! Important!
The housing covers must not be removed.
Inspection measures
Checking the probe
Intensive care of the CMM also includes inspection of the compo-
nents used.
Regular inspection of the probes is required to ensure correct work-
piece measurement.
• Check the styli, other probe components and the adapter plate
regularly.
• Remove any particles or grease film from the probe elements and
the adapter plate.
• Replace any damaged styli.
For this reason, probe elements should be checked regularly during all
scanning operations and replaced if necessary.
• Check the condition of the probe elements and clean them.
For information on how to remove aluminum deposits, please refer
to ➤ page 9-6.
• Replace the stylus if the probe element is damaged.
1
2
3
1 Fan
2 Air filter
3 Warning label: "Verschmutztes Luftfilter austauschen oder reinigen. /
Clean or replace soiled air filter.“
! Danger!
Switch off the compressed air before changing the filters.
A B
A Fine filter
B Extra fine filter
A B C
1 4
2
6
3
With the extra fine filter, the viewing glass (C) indicates when the fil-
ter has to be changed. When changing the extra fine filter, you should
also change the fine filter.
U
UMESS 7-12
UNIX measuring software 7-12
Unlock fall brake 8-7
V
VAST probe system 2-15
adapter plate 2-19
application 2-15
components 2-16
limiting values 2-16
Components
Probe systems in
operation
Calibration: RDS/RST probe system: