UT513A
UT513A
Model UT513
                                    OPERATING
                                    MANUAL
                                                Table of Contents
                            Title                           Page
                                                Model UT513: OPERATING MANUAL
    Introduction                                                          5
    Unpacking the Meter                                                   5
    Safety Information                                                    6
    International Electrical Symbols                                      8
    Battery Saver (Sleep Mode)                                            8
    Battery Indication                                                    8
    The Meter Structure                                                   9
    Display                                                               12
    Key Functions                                                         14
    Measurement Operation                                                 17
        A. Measuring Voltages                                             17
        B. Measuring Insulation Resistance                                19
           a) Continuous Measurement                                      20
           b) Timed Measurement                                           20
           c) Polarization Index (PI) Measurement                         21
           d) Compare Function                                            22
    The Use of Power Adaptor                                              23
    Title                                                               Page
1
USB Interface                                        24
Maintenance                                          25
   A. General Service                                25
   B. Replacing the Battery                          26
Specifications                                       27
Safety and Compliances                               27
Physical Specifications                              27
General Specifications                               28
Feature Summary                                      28
Detailed Accuracy Specifications                     29
   A. Voltage Measurement                            29
   B. Insulation Resistance Measurement              30
                                          List of Tables
1.              Unpacking                           5
                Inspection                          8
                                                    8
                                                    10
                                                    11
                                                    13
                                                       2
                                                    14
Model UT513: OPERATING MANUAL                                      Model UT513: OPERATING MANUAL
3
Model UT513: OPERATING MANUAL
                                                                                                                  4
                                                                       Model UT513: OPERATING MANUAL
5
                       Danger
  Use of instrument in a manual not specifed by
  the    manufactuer      may      impair   safety
  features/protection provided by the equipment.
  Read the following safety information carefully
  before using or servicing the instrument.
       Do not apply more than 600V.
       Do not use the Meter around explosive
       gas, vapor or dust.
         Do not use the Meter in a wet
       environment. When using the test leads,
  keep your figures away from the lead contacts.
  Keep your figures behind the finger guards on
  the leads. Do not use the Meter with any
  parts or cover removed.
                                                     6
    When carrying out insulation measurement,      When performing resistance tests, remove all
    do not contact the circuit under test.          power from the circuit to be measured and
                                                    discharge all the power.
                      Warning
                                                   When servicing the Meter, use only the same
  Do not use the Meter if it is damaged or metal    model number or identical electrical
   part is exposed. Look for cracks or missing      specifications of test leads and power
   plastic.                                         adaptor. Do not use the Meter if the battery
  Be careful when working above 33V rms,            indicator
   46.7V ac rms or 70V DC. Such voltages            ( ) shows a battery empty condition. Take
   pose a shock hazard.                             the battery out from the Meter if it is not
Discharge all loading of circuit under test         used for a long time.
after measuring high voltage.                      Do not use or store the Meter in an
Do not change battery when the Meter is in          environment of high temperature, humidity,
wet environment.                                    explosive, inflammable and strong magnetic
  Place test leads in proper input terminals.       field. The performance of the Meter may
Make sure all the test leads are firmly             deteriorate after dampened.
connected to the Meterís input terminals.          Soft cloth and mild detergent should be used
Make sure the Meter is turned off when              to clean the surface of the Meter when
opening the battery compartment.                    servicing. No abrasive and solvent should
                                                    be used to prevent the surface of the Meter
                      Caution                       from corrosion, damage and accident.
                                                   Dry the Meter before storing if it is wet.
7
                                                                           Model UT513: OPERATING MANUAL
                                                                                                              8
Model UT513: OPERATING MANUAL                                                        Model UT513: OPERATING MANUAL
10.6V~11.5V
11.6V or more
9
Model UT513: OPERATING MANUAL
 4      Data Clear the Display Backlight Button   19    High voltage line shielding input terminal
 5        Down Button                                  (two plugs red test lead to one alligator clip)
 6      On/Off Button                             20       GUARD: Grounding protection input
 7      Compare Button                                                     terminal
                                                         (one plug black test lead to one alligator
                                                                             clip)
 8      Insulation Resistance Button              21   EARTH: High resistance measurement
 9      DC Voltages measurement Button                 input terminal (one plug green test lead to
                                                       one alligator clip)
 10     Timer Button.
 11     AC Voltages measurement Button            22   Testing leads:
 12     Test Button                                    Two plugs red test lead to one alligator
 13     USB Button                                     clip. One plug black test lead to one
                                                       alligator clip. One plug green test lead to
 14     Data Store Button.
                                                       one alligator clip.
  15     Data Recall Button
  16       Scroll Button
Below Figure 2 and Table 5 shows the Meter side             Table 5. Meter Side Description
structure and description
                                                                                                     10
                                                                       Model UT513: OPERATING MANUAL
                                             1   Safety Shutter
                                             2   Power adaptor Input Terminal
                                             3   USB Port
11
Model UT513: OPERATING MANUAL
Figure 3. Display
                                Table 6. Display Description
                                                               12
                                                                                  Model UT513: OPERATING MANUAL
13
          hold the button for 1 second to turn      LOAD       Press once to recall the first stored
          the Meter on. Press again to turn off                value.
          the Meter.                                           Press again to exit Load feature.
          The Meter default range is 500V                      Load feature can only be used
          insulation resistance continuous                 when there is no high voltage output.
          measurement when turning on.                      When the insulation resistance
                                                             measurement has no testing voltage
E-STOP    Emergency stop button. Press this                  output, press to select one voltage
          button when the Meter is hang and                  range up.
          cannot turn off the power.                        Under load mode: press to recall the
CLEAR /   Press to turn on or off the display                previous stored value.
          backlight
          Press and hold to clear the stored data            When the insulation resistance
                                                             measurement has no testing voltage
SAVE      Press      to   store   the   current              output, press to select one voltage
          measurement value. The maximum                     range down.
          number of stored reading is 18.                    Under load mode: press to recall the
          When the stored readings memory is                 next stored value.
          full, the Meter shows FULL and stop
          storing. Press and hold CLEAR /
          to clear the stored value in order to
          store the next measurement
                                                                                                   14
Model UT513: OPERATING MANUAL
15
                                   Model UT513: OPERATING MANUAL
           press to
        Table 7. Key Description
COMP   Set a pass / fail limit
       for insulation tests.
       The default value is
       10M
TIME   Pres to step through
       continuous
       measurement, timed
       measurement         and
       polarization      index
       measurement           in
       sequence.
TEST   Press to stop or start
       an insulation
       resistance test
IR     Press to initiate
       insulation resistance
       measurement
DCV    Press to initiate DC
       voltage
                                                             16
Model UT513: OPERATING MANUAL
            measurement
 ACV        Pres to initiate AC
            voltage
            measurement
 Measurement Operation               A.
                                     Measuring
                                     Voltages
Below section explains how to make measurements.
17
                                                                          Model UT513: OPERATING MANUAL
                                                                                                     18
Model UT513: OPERATING MANUAL
19
                                                           Model UT513: OPERATING MANUAL
5000V measure resistance lower than          5. Connect the red and black alligator clip to the
20M                                               circuit to be measured, negative voltage
To measure insulation resistance, set up          output from LINE terminal.
the Meter as Figure 5 and do the             6. Choose        below    insulation    resistance
following:                                        measurement mode.
                                             Press TIME button to select continuous
1. Press IR button to select insulation      measurement mode, there is no timer icon on
   resistance measurement.                   the LCD.
2. When there is no testing voltage          Press and hold TEST button for 1 second to
   output, press      and       button to    carry out continuous measurement.           Output
   select voltages of 500V, 1000V,           insulation resistance testing voltage, TEST
   2500V or 5000V.                           button light up, blinks on every 0.5 seconds.
3. When        performing       insulation   Press TEST button to close the insulation
   resistance tests, remove all power        resistance      measurement       voltage    when
   from the circuit to be measured and       measurement is completed. TEST button lights
     discharged all the power.               off,     disappears. The LCD shows the current
4. Insert the red test lead into the LINE    insulation resistance measurement value.
   input terminal and the black test lead
   into GUARD input terminal.
                                             Press TIME button to select timed measurement
                                             mode, the LCD displays TIME 1 and symbols.
                                                                                            20
Press and buttons to set the time          buttons to set the time (00:10~15:00). Within 1
(00:10~15:00). Within 1 minute, the time   minute, the time increment or decrement by
increment or decrement by every 5          every 5 seconds. Afterward, the time increment
seconds. Afterward, the time increment     or decrement by every 30 seconds. Press TIME
or decrement by every 30 seconds.          button again. TIME 2, PI and symbols appear
    Then press and hold TEST button        on the LCD.
    for 2 second to carry out timed
    measurement. TIME 1 and are
    displayed and blinked on the LCD on
    every 0.5 seconds.
   When the set time is reached, the
    insulation resistance measurement
    voltage will be closed and the
    measurement will be automatically
    stopped. The LCD displays the
    insulation resistance reading.
21
Model UT513: OPERATING MANUAL                            Model UT513: OPERATING MANUAL
Press and buttons to set the time               resistance reading and TIME 2 insulation
(00:15~15:30). Within 1 minute, the time        resistance reading.
increment or decrement by every 10
                                             Information:
seconds. Afterward, the time increment
                                             PI = 3 minutes ~10 minutes reading / 30 seconds
or decrement by every 30 seconds.
                                             ~1 minute reading
Then press and hold TEST button for 2
seconds       to    carry    out    timed     PI          4 or more 4~2    2.0~1.0 1.0 or less
measurement. TIME 1 and                are    Standard The best Good Warning Bad
displayed and blinked on the LCD on          Press COMP button to select compare feature.
every 0.5 seconds before TIME 1 set          COMP symbol displays on the LCD..
time is reached.                             Press and buttons to set the compare value
TIME 2 and are displayed and blinked         Below is the list in sequence of the compare
on the LCD on every 0.5 seconds before       value:
TIME 2 set time is reached.                  10M , 20M , 30M , 40M , 50M , 60M , 70M ,
When the two set time are reached, the       80M    , 90M ,100M , 200M , 300M , 400M ,
insulation resistance measurement            500M , 600M , 700M , 800M , 900M ,1G         ,
voltage will be closed and the
                                             2G , 3G , 4G , 5G , 6G , 7G , 8G , 9G ,
measurement will be automatically
                                             10G ,20G , 330G , 40G , 50G , 60G , 70G ,
stopped. The LCD displays the
                                             80G    , 90G , 100G ,200G , 300G , 400G ,
polarization index reading.
                                             500G , 600G , 700G , 800G , 900G
    Press , to set through the
    polarization index, TIME 2 insulation
                                                                                           22
Model UT513: OPERATING MANUAL
Press and hold TEST button for 2              The NG symbol will display if the insulation
seconds to carry out the measurement.         resistance value is smaller than resistance
                                              value.
                                              Otherwise GOOD symbol will be displayed.
The Use of Power Adaptor                 1.
The use of power adaptor, see figure 6   2.
23
                                                                               Model UT513: OPERATING MANUAL
Figure 6. The Use of Power Adaptor                              Connecting the USB interface, see figure 7
Open the side safey shutter, then you will see there is a power
adaptor input terminal. Make sure the Meter is power off and    Figure 7. USB Interface Connection
insert the UT513 power adaptor to the input terminal. It is     1. Install the included software, the installation
highly recommeded to take out all the batteries when you are       guide can be seen from the CD.
using the power adaptor. Make sure the Meter is power off
                                                                2. Open the side safety shutter, then you will see
when you disconnect the UT513 power adaptor from the
                                                                   there is a USB port.
Meter. It is highly recommeded to use Uni-Trend supplied
                                                                3. Insert the included USB cable to the Meterís
UT513 power adaptor to avoid dangerous.
                                                                   USB port and the other end to the computer.
USB Interface
                                                                                                               24
Maintenance
This section provides basic maintenance information
including battery replacement instruction.
Warning
Model UT513: OPERATING MANUAL
B. Replacing the Battery                               Donít mix to use old and new batteries. Be
                                                       careful the polarity is correct when installing
                                                       batteries.
                                                       Do not use the Meter if the battery indicator (
                                                            ) shows a battery empty condition. Do
                                                       you carry out measuring during the battery
                                                       compartment is open.
1
Model UT513: OPERATING MANUAL
    Icon Display                     Equips with function and battery indicator icons.
    Current Consumption              Maximum: around 1.0A
                                     Average: around 20mA
Feature Summary
    Display Backlight                Bright backlight for clear readings in poorly lighted areas.
    Computer connection              Via USB interface.
    Data Logging and Recall          18
    Autorange                        The Meter automatically selects best range
    Warning                             and red light will on.
    Voltage                          Auto release voltage
    COMP Measurement                 Use the Compare function to set a pass/fail compare level for the insulation
                                     measurements.
    PI Measurement                   Polarization Index is the ratio of insulation resistance. You can pre-set two
                                     point of times and automatically carry out the measurement.
    TIME                             To carry out measurement by setting a specified time within 15 minutes.
Detailed Accuracy Specifications
Accuracy: ([% of reading] + [number of least significant digits), guarantee for 1 year.
Operating temperature: 18     ~28 Relative
humidity: 45~75%RH
A. Voltage Measurement
                                                                                                                     2
                                                                                       Model UT513: OPERATING MANUAL
                                               DC Voltage                                 AC Voltage
Operating Caution
3
Model UT513: OPERATING MANUAL                                                   Model UT513: OPERATING MANUAL
At any output voltage, when the tested resistance is les than 10M , the testing time cannot exceed 10
seconds continuously.
                                                     *END*
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China Postal Code: 523
925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
                                                                                                           4
Fax: (852) 2950 9303 Email:
info@uni-trend.com
http://www.uni-trend.com