Item
minimum technical requirements                                            Quantity
     No
1           Biosafety cabinet class III                                                                                          2 set
            Description
            Type: Class 3, ISO 3 air cleanliness
            Minimum Technical requirement
           Nominal workable size: 5 feet
           Usable chamber dimension (HxD), without bench: ≥ 60 x 50 cm
           Number of glove port: 4 (2 pair)
           Arm cuffs diameter: 300 ± 15% mm
            Glove type: Neoprene or Butyl
           Airlock: Automatic
           Front window glass thickness: ≥ 5 mm, anti-uv protected
            Exhausted air vented through a system of double (2) HEPA/ULPA filter
           Filter efficiency: 99.999% at 0.1- 0.3 μm or better
           Initial air flow volume: ≥ 750 m3/h
           Minimum negative pressure: ≥ -120 Pa
            Noise: ≤ 65 dBA
           Lamp type: Housing LED/fluorescent
           Interior illumination: ≥ 750 lux
            Work surface material: Stainless steel
            Main body construction: Corrosion and chemical resistant material
           Display: Digital
           Display parameter: Airflow velocity, temperature, filter life, UV-lamp life, usable time
            Audio/visual alarm for incorrect airflow velocity
            UV lamp: ≥ 20 W, 254 nm
           Accessories
           (i) Additional Tap (Fuel Gas / Non-Fuel Gas / Vacuum)
           (ii) Gas cylinders (tri gas), regulators and fittings
           (iii) Passbox/transfer hatch
           (iv) Pressure gauge
           (v) Vacuum airlock
           (vi) Stainless steel suport cart (trolly) with caster wheel with lock feature
           (vii) Hanging bar (SS)
           (viii) Thimble Duct Exhaust Transition
           Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
                                                                                                                                 Qty
2          Product name: Bioasafety cabinet Class 2-type 2                                                                       2 set
           Minimum Technical requirement
           Type: Class II-B2
           Nominal size: 4 feet
           External dimension (with table); DxH: (850 x 2400 mm) ± 10%
           The external body construction material should be chemical and corrosive resistant and with anti bacterial
           powder coating
           Interior should be made up of stainless steel (316 or 304) with side tempered glass to ensure corrosion resistant
           Cabinet interior shall process continuous, rounded corners
           Base stand: Manufacturer recommended support base stand shall be included
           Work surface height: ≥ 700 mm
           Maximum window opening: ≥ 425 mm
    Safety working height: ≥ 200 mm
    Front window moving mechanism: Motorized
    Front window glass thickness: ≥ 5 mm, anti-uv protected
    Airflow Volume (Inflow): 350 m³/h or better
    Airflow Volume (Down flow): 65%: 675 m³/h or better
    Airflow Volume (Exhaust): 5%: 360m³/h or better
    Should have 2 HEPA/ULPA filter with 0.12 µm particles ≥99.999%
    Single/dual fan blower
    Air speed should be adjustable to automatically to maintain constant air speed
    UV lamp: ≥ 20 W, 254 nm
    Lamp type: LED/fluorescent
    Interior illumination: ≥ 750 lux
    Display: Digital
    Display parameters: Panel should display all required parameters
    (i) Inflow and down flow velocities
    (ii) Filter life
    (iii) UV lamp life
    (iv) System work time
    Audio and or visual alarm facility should be available for
    (i) Abnormal airflow velocity
    (ii) Filter change
    (iii) Front window over height
    Noise: ≤ 65 dBA
    Should come with minimum
    (i) Cabinet with base stand
    (ii) LED/fluorescent and UV lamp
    (iii) SS water tap x 1 Nos
    (iv) Gas tap x 1 Nos
    (v) Drain ball valve
    (vi) 2 x 13 A multifunctional water proof socket
    (vii) Include a leak tight duck: Minimu 2 m ducting should be included
    (viii) Leak proof damper in the duct above the cabinet
    Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
                                                                                                                        Qty
3   Biosafety Cabinets Class I                                                                                           2
    Minimum Technical requirement
     Type: Class 1
    Nominal size: 3 feet
    External dimension (with table); DxH: (700 x 2400 mm)±10%
    The external body construction material should be chemical and corrosive resistant and with anti bacterial
    powder coating
    Interior should be made up of stainless steel (316 or 304) with side tempered glass to ensure corrosion resistant
    Cabinet interior shall process continuous, rounded corners
    Base stand: Manufacturer recommended support base stand shall be included
    Maximum window opening: ≥ 425 mm
    Safety working height: ≥ 200 mm
    Work surface height: ≥ 700 mm
    Display: Digital
    Should have 2 HEPA/ULPA filter with 0.3 µm particles ≥99.97%
    Airflow Velocity (Inflow): 0.75 m/s or better
    Air speed should be adjustable to automatically to maintain constant air speed
    UV lamp: ≥ 20 W, 254 nm
Noise: ≤ 65 dBA
Lamp type: LED/fluorescent
Interior illumination: ≥ 1000 lux
Audio and or visual alarm facility should be available for
(i) Abnormal airflow velocity
(ii) Filter change
(iii) Front window over height
Display: Digital
Display parameters: Panel should display all required parameters
(i) Inflow and down flow velocities
(ii) Filter life
(iii) UV lamp life
(iv) System work time
Front sliding sash interlock with fan and UV lamp
Should come with minimum
(i) Cabinet with base stand
(ii) LED/fluorescent and UV lamp
(iii) SS water tap x 1 Nos
(iv) Gas tap x 1 Nos
(v) Drain ball valve
(vi) 2 x 13 A multifunctional water proof socket
(vii) Include a leak tight duck: Minimum 2 m ducting should be included
(viii) Leak proof damper in the duct above the cabinet
Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
All standard accessories shall be supplied