TECHNICAL DATA SHEET ECG ELECTRODE“LESSA”
99.830.18
Description:
Pre-gelled electrode with solid gel and Foam support.
General Features:
• Foam support protects the sensor and gel from cleaning and surgical products.
• Solid gel offers good adhesion even in skins that have not been adequately prepared and allows high quality
records with minimal adverse reactions.
The product features enhanced, non-sensitizing, acrylic copolymer coated foam. It provides a comfortable, soft
conformable electrode. This non-irritating, disposable electrode meets and exceeds all parametric tests as per
ANSI / AAMI EC12: 2000 Standard.
The product is designed to be used in a rest-type procedure. For example: Resting ECG / monitoring and echocardiogram procedures. It complies with
the RoHs regulations.
EXPECTED USE DIMENSIONS
Expected Use Resting ECG Electrode shape Oval
Monitoring
ECHO Procedures
Characteristics disposable, Electrode size (mm) 40x36
Latex free, PVC free, Area [mm²] 1162
not sterile
Product shel life 36 meses Area Gel [mm²] 201
Storage (min/max) +5° C/ +30° C Adhesive Area [mm²] 893
CLASSIFICATION AND REGULATION
CLASSIFICATION (MDD ART. 9) Class I
CLASSIFICATION (CFR 21 870.2360.) Class II
MATERIALS PACKAGING MATERIALS
Connector Staple Stainless steel Bag, interior layer Polyethylene (PE)
Label PET layer Bag, central layer Aluminium (Al)
Support Foam Bag, external layer Paper
Adhesive Medical grade Acrylic Boxes Cardboard
Sensor (Ag/AgCl)
Sponge -
Gel Adhesive solid gel
Release Liner Silicone PET layer (transparent)
PACKAGING Códe: 99.830.18 BIOCOMPATIBILITY Test - ISO 10993-1
Blíster 3
Units / pouch 30 Support (adhesive) Gel
Units / box 600 Cytotoxity passed passed
Batch and expiration date (it is indicated in each bag and box) Skin irritation test passed passed
Skin Sensititation Test passed passed
ELECTRICAL VALUES Units Typical Values AAMI limits
DC-Offset [mV] 1,5 100
DC Offset (5 seg after capacitatos discharge) [mV] 15 100
Recovery slope [mV/sec] -0,3 1
AC-Impedance with 10 Hertz [Ω] 100 2000
ABGR7.2-01 Rev.01
Ficha técnica 9983654 eng