Sor 6547
Sor 6547
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-02
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-03
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-04
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-05
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-06
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-07
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-08
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                         CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-09
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
             Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                         CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-10
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
             Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                         CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-11
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
             Comments:
             The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
             instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                         CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-12
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
             Comments:
             The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
             instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-13
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-14
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
50 50.40 0.40 1
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-15
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-16
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-17
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-18
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-19
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-20
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-21
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-22
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-23
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-24
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.
                                          CALIBRATION CERTIFICATE
Date:       30/October/2024                                                                                Certificate No:   LIF241030-25
Calibration Results
Applied Value (A) Measured Value (M) Deviation (M-A) Uncertainty (U)
  Based on ANSI/NCSL Z540-1, The reported uncertainty is the expanded uncertainty with k=2 according to the ISO/IEC "Guide expansion of
  uncertainty in measurement" providing a level of confidence of approximately 95%.
Note:
1.    Yes     instrument is "used" and in good condition on receiving.
2.    No      Instrument was new on receiving.
3.    No      Instrument is adjusted and post adjustment values are reported.
4.    Yes     No adjustment is carried out and measurements in this certificate are as received figures.
Comments:
              The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the long-term stability of the
              instrument.
Calibrated By Approved By
This certificate is issued in accordance with the requirements of ISO 17025. It provides traceability of measurements to recognized
national/international Standards and to units of measurements realized to recognized national/international standard laboratories. This certificate may
not be reproduced, except in full, without the prior written approval of the laboratory.