SIZE MASTERS CALIBRATION SERVICES
Final Plot No. 123, Lane No. 10, Ramtekdi Industrial Estate, Hadapser,
411 019. (INDIA) Tel. : +91 9890644445, 8297644445,
E-mail : kiranzanwar@yahoo.com / smoscalibration@gmail.com
Punt
CERTIFICATE OF CALIBRATION
Certificate No. 2019 / 38461 /05 | Date of Issue — 16/11/2019 Page No. 1 of2 ]
ULR: CC246019400001551F Discipline: Mechanical Product Group: Force |
NAME OF CUSTOMER MACHINE DETAILS ]
Type: Universal (Uniaxial) 1S: 1828(Partl): 2076 |
‘Mis. Walchandnagar Testing Machine. cence 1807500-1: 2018 |
Industries Ltd. Make: FIE Procedure No. W214 |
Walchandnagar, Tal - Indapur, | Model: UTE - 60 Calibrated On’ 14/11/2019 |
Dist. Pune - 413 114, ieee Agreed Due Date | 13/11/2020 |
a
Maharashtra. Max. Cap. : 600 KN Calibrated By DG Nanaware |
Mode of cal": Gompression_| Location: LAB |
Other Details:
Make / Model SR.NO./IDNO. Capacity
Pressure Call NAINA. NAINA 00 kN
Digital Indicator FIE/ UTE - SERVO NAINA,
Details of Force Proving Instrument used for calibration:
Make/ModeVCapacity ADI Artech/0210/50 kN | ADI Artech/90310/500KN
Sr. No/Gauge Sr. No. NT-LIM05/NT-LE-105 | 365316 14687 / 14687
Calibration Cerificate No. | FRV09/16/10808 FRIMQT/OTAS 17080619/D1 0510-496
& Date. & ct 29/09/2018 & at 31/1272017 & at 25/112017 |
Next Calibration due 29/11/2020 02/03/2020 23/01/2020 |
Calibrated at Temp. 24°C 24°C 231°C |
UL IM, at 95.45% C.L. £0.08 % atK=2 0.080 % at K=2 £0.12 % at K= i
FRI, Ichalkaranji FRI, Ichalkaranji |
Traceability to NPL through | qasi-CC2029) (NABL-CO-2029) ee
Observations
Force Standard] Temp. Avg. oF
Proving | Test Fore® | instrument | Corrected | Observed | %Errorof | %Erorof | ULL.
Instrument | (gn) Read. | Std.Read | Read. Accuracy | Repeatability | % (2)
used ( (Div) (Div) (Div) |
Range: 0- 600 KN Gerlified Range: 10 600 KN Temp. 227°
L.C.: 0-120 KN: 0,03 KN, 120-240 KN: 0.06 KN & 240-600 KN: 0.15 kN
10 10004 : 10069.33 0.85 0.20 018 |
on 20 20007 : 2011167, 052 0.02 018 |
49985 : 50188.67 0.35 ona 018 |
ae. Ls
(ANAWARE
Calibrated By
Note : This certificate is issued subject to conditions stated overleat.SIZE MASTERS CALIBRATION SERVICES
Final Plot No. 123, Lane No. 10, Ramtokdi Industrial Estate, Hadapser,
Pune - 411 013. (INDIA) Tel. : +91 9890644445, 8237644445
E-mail : kiranzanwar@yahoo.com / smoscalibration@gmail.com
CERTIFICATE OF CALIBRATION
Continuation of Report No. : 2019/ 38461 /05._ | ULR: CC246019400001551F _| Page, 2 of2
Observations. -
Proving | Test Force tnarumert | canetea | ontenea %emerof | %emorot | ULM |
Instrument | (gi) Read. | Sti.Read | — Read, ‘Accuracy | Repeatability | 9% (2)
used (Dv) (Div) (div)
Range: 0- 600 EN Certified Range: 10-600 KN Temp. 227°C |
L.C.: 0-120 KN: 0.03 KN, 120-240 KN: 0.06 KN & 240-600 KN: 0.15 kN
UY 50 49.22 - 49.21 -0.03 ore | 018
se’ [100 99.47 - 99:37 0.10 00s | one |
150 149.69 : 149.87 0.42 0.08 a |
160 10020 - 10083.00 063 0.16 0.18
200 13356 : 13385.00 022 0.07 018
3o00KN [__300 20029 = 20133.67 052 010 | ote
te. 400 26702 : 26646.67 0.24 026 | o2t
500 33375 : 3e2e86.67 | -0.26 0.04 ote |
600 40048 : 39931.67 0.29 0.04 o18 |
Resolution: 0.03 KN ‘Max. Observed Reading on removal of force: O.00KN |
Max. Relative Resolution : 0.30 % Max. Zero Error: 0.00 % |
Note: Standard & Observed readings of 50 kN & 3000 kN Load Gell are In Div & 500 KN Load Cell are in kN.
\ _U.I.M. is calculated with the coverage factor 2 corresponding to 95.45% confidence level
Class of the machine: Class 1
Classification of the machine as per IS: 1828 (Part 1): 2015 & ISO 7500-1: 2018
Class of ‘Maximum Permissible value, %
Machine Relative Error of.
Range ‘Accuracy (a) [Repeatability (6) | Zero (fo) | Relative Resolution (a)
1 21.0 4.0 £01 05
2 £20 2.0 202 4.0
3 £3.0 3.0 £03 1.6
ES
SE ‘NANAWARE 1 OX te vw. R .
Calibrated By