Sae J1400 2017
Sae J1400 2017
SURFACE VEHICLE
                                                                                                                                                      RECOMMENDED PRACTICE                                                                 Issued                  1982-05
                                                                                                                                                                                                                                           Reaffirmed              1989-05
                                                                                                                                                                                                                                           Revised                 2017-07
RATIONALE
                                                                                                                This document defines an alternative method to ASTM E90 and ASTM E2249 for measuring sound transmission loss (STL)
                                                                                                                or noise reduction for flat sample materials.
FOREWORD
                                                                                                                This SAE Recommended Practice has been modified in response to user comments and to clarify some of the theoretical
                                                                                                                basis for the testing. The fundamental methodology of this procedure has not been changed from previous revisions.
INTRODUCTION
                                                                                                                This document is intended to provide a means of measuring the STL of materials. At each test frequency, the STL is
                                                                                                                determined based on the measured noise reduction (MNR) of the test specimen using a correlation factor (CF). The
                                                                                                                respective CF for the test condition is determined as the difference between the MNR of a homogeneous limp panel, such
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                as mass loaded vinyl, and its theoretically calculated field-incidence STL. Note that the calculated STL is for a theoretical
                                                                                                                panel of infinite dimensions, and the intent is to remove the effect of different size test windows found in different labs, with
                                                                                                                consideration of low frequency limitations imposed by smaller openings. This recommended practice then recognizes that
                                                                                                                many laboratories have measurement variances that can be corrected to a certain extent using this methodology.
                                                                                                                __________________________________________________________________________________________________________________________________________
                                                                                                                SAE Technical Standards Board Rules provide that: “This report is published by SAE to advance the state of technical and engineering sciences. The use of this report is entirely
                                                                                                                voluntary, and its applicability and suitability for any particular use, including any patent infringement arising therefrom, is the sole responsibility of the user.”
                                                                                                                SAE reviews each technical report at least every five years at which time it may be revised, reaffirmed, stabilized, or cancelled. SAE invites your written comments and
                                                                                                                suggestions.
                                                                                                                Copyright © 2017 SAE International
                                                                                                                All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted, in any form or by any means, electronic, mechanical, photocopying,
                                                                                                                recording, or otherwise, without the prior written permission of SAE.
                                                                                                                TO PLACE A DOCUMENT ORDER:               Tel:     877-606-7323 (inside USA and Canada)                SAE values your input. To provide feedback on this
                                                                                                                                                         Tel:     +1 724-776-4970 (outside USA)                       Technical Report, please visit
                                                                                                                                                         Fax:     724-776-0790                                        http://standards.sae.org/J1400_201707
                                                                                                                                                         Email: CustomerService@sae.org
                                                                                                                SAE WEB ADDRESS:                         http://www.sae.org
                                                                                                                SAE INTERNATIONAL                                           J1400™ JUL2017                                                                         Page 2 of 23
TABLE OF CONTENTS
1. SCOPE .......................................................................................................................................................... 4
                                                                                                                2.       REFERENCES.............................................................................................................................................. 4
                                                                                                                2.1      Applicable Documents .................................................................................................................................. 4
                                                                                                                2.1.1    SAE Publications ........................................................................................................................................... 4
                                                                                                                2.1.2    Accredited Publications ................................................................................................................................. 4
                                                                                                                2.1.3    INCE-USA Publications................................................................................................................................. 4
                                                                                                                2.1.4    ASTM Publications ........................................................................................................................................ 5
                                                                                                                2.1.5    AES Publications ........................................................................................................................................... 5
                                                                                                                2.1.6    Published Books ........................................................................................................................................... 5
                                                                                                                3.       INSTRUMENTATION.................................................................................................................................... 5
                                                                                                                3.1      Sound Level Meter ........................................................................................................................................ 5
                                                                                                                3.2      Filter Requirements ....................................................................................................................................... 5
                                                                                                                3.3      Microphone Calibrator ................................................................................................................................... 5
                                                                                                                3.4      Source Room Speakers ................................................................................................................................ 5
                                                                                                                3.5      Instrumentation ............................................................................................................................................. 5
                                                                                                                3.6      Ambient Sensors ........................................................................................................................................... 6
                                                                                                                4.       FACILITIES ................................................................................................................................................... 6
                                                                                                                4.1      Layout of Chambers ...................................................................................................................................... 7
                                                                                                                4.2      Receiving Chamber....................................................................................................................................... 7
                                                                                                                4.3      Chamber Sizes .............................................................................................................................................. 7
                                                                                                                4.4      Source Room Lower Cutoff Frequency and Diffusion .................................................................................. 7
                                                                                                                4.5      Ambient Conditions ....................................................................................................................................... 8
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                5.       PROCEDURE ............................................................................................................................................. 10
                                                                                                                5.1      Sample Mounting ........................................................................................................................................ 10
                                                                                                                5.2      Sample Conditioning ................................................................................................................................... 10
                                                                                                                5.3      Measurements ............................................................................................................................................ 10
                                                                                                                5.3.1    Background Noise ....................................................................................................................................... 10
                                                                                                                5.3.2    Reference Sample ...................................................................................................................................... 11
                                                                                                                5.3.3    Reference Sample Surface Density ............................................................................................................ 11
                                                                                                                5.3.4    Signal-to-Noise Ratios ................................................................................................................................ 11
                                                                                                                5.3.5    Measurement Conditions ............................................................................................................................ 11
                                                                                                                5.3.6    Test Samples .............................................................................................................................................. 12
                                                                                                                5.4      Data Analysis .............................................................................................................................................. 12
                                                                                                                5.4.1    Background Noise Correction ..................................................................................................................... 12
                                                                                                                5.4.2    Measured Noise Reduction ......................................................................................................................... 12
                                                                                                                5.4.3    Correlation Factors ...................................................................................................................................... 13
                                                                                                                5.4.4    Sound Transmission Loss ........................................................................................................................... 13
                                                                                                                5.4.5    Insertion Loss .............................................................................................................................................. 13
                                                                                                                5.5      Reporting ..................................................................................................................................................... 13
                                                                                                                5.5.1    Basic Information ........................................................................................................................................ 13
                                                                                                                5.5.2    Ambient Conditions ..................................................................................................................................... 13
                                                                                                                5.5.3    Sound Transmission Loss ........................................................................................................................... 14
                                                                                                                5.5.4    Confidence Limits........................................................................................................................................ 14
                                                                                                                5.5.5    Maximum Measurement Capability............................................................................................................. 14
                                                                                                                SAE INTERNATIONAL                                               J1400™ JUL2017                                                                        Page 3 of 23
                                                                                                                7.           NOTES ........................................................................................................................................................ 15
                                                                                                                7.1          Revision Indicator........................................................................................................................................ 15
                                                                                                                TABLE 1      SUGGESTED MINIMUM SOURCE ROOM VOLUME AT LOWEST MEASUREMENT FREQUENCY ...... 8
                                                                                                                TABLE 2      EFFECT OF DIAGONAL DIMENSION OF A SQUARE TEST WINDOW AT LOWEST
                                                                                                                             MEASUREMENT FREQUENCY AND COMPARISON WITH ¼-WAVELENGTH OF LOWEST
                                                                                                                             1/3-OCTAVE BAND FREQUENCY ............................................................................................................ 10
                                                                                                                TABLE 3      TARGET SOUND TRANSMISSION LOSS VALUES - CONTROL SAMPLE ............................................ 15
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                SAE INTERNATIONAL                                  J1400™ JUL2017                                           Page 4 of 23
1. SCOPE
                                                                                                                This SAE Recommended Practice presents a test procedure for determining the airborne sound insulation performance of
                                                                                                                materials and composite layers of materials commonly found in mobility, industrial and commercial products under
                                                                                                                conditions of representative size and sound incidence so as to allow better correlation with in-use sound insulator
                                                                                                                performance. The frequency range of interest is typically 100 to 8000 Hz 1/3 octave-band center frequencies.
                                                                                                                This test method is designed for testing flat samples with uniform cross section, although in some applications the
                                                                                                                methodology can be extended to evaluate formed parts, pass-throughs, or other assemblies to determine their acoustical
                                                                                                                properties. For non-flat parts or assemblies where transmitted sound varies strongly across the test sample surface, a more
                                                                                                                appropriate methodology would be ASTM E90 (with a reverberant receiving chamber) or ASTM E2249 (intensity method
                                                                                                                with an anechoic or hemi-anechoic receiving chamber).
2. REFERENCES
                                                                                                                The following publications form a part of this specification to the extent specified herein. Unless otherwise indicated, the
                                                                                                                latest issue of SAE publications shall apply.
                                                                                                                Available from SAE International, 400 Commonwealth Drive, Warrendale, PA 15096-0001, Tel: 877-606-7323 (inside USA
                                                                                                                and Canada) or +1 724-776-4970 (outside USA), www.sae.org.
                                                                                                                Ebbitt, G. and Hansen, M., "Mass Law - Calculations and Measurements," SAE Technical Paper 2007-01-2201, 2007,
                                                                                                                doi:10.4271/2007-01-2201
ANSI S1.11 Specification for Octave Band and Fractional Octave Band Filter Sets
ANSI S1.26 Method for Calculation of the Absorption of Sound by the Atmosphere
Available from INCE-USA, 12100 Sunset Hills Road, Suite 130, Reston, VA 20190. Tel: 703-234-4073, www.inceusa.org.
                                                                                                                Beranek, Leo L., Noise and Vibration Control - Revised Edition, Institute of Noise Control Engineering, Washington, DC,
                                                                                                                1988
                                                                                                                Moritz, C., Shaw, J. and Carrera, A., “The Influence of Test Fixture Damping on the Measurement of Sound Transmission
                                                                                                                Loss,” proceedings of NOISE-CON 2014
                                                                                                                SAE INTERNATIONAL                                    J1400™ JUL2017                                             Page 5 of 23
                                                                                                                Available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, Tel: 610-
                                                                                                                832-9585, www.astm.org.
                                                                                                                ASTM C423        Standard Test Method for Sound Absorption and Sound Absorption Coefficients by the Reverberation
                                                                                                                                 Room Method
ASTM E90 Laboratory Measurement of Airborne Sound Transmission Loss of Building Partitions
                                                                                                                ASTM E2249       Standard Test Method for Laboratory Measurement of Airborne Transmission Loss of Building Partitions
                                                                                                                                 and Elements Using Sound Intensity
                                                                                                                Available from Audio Engineering Society, 551 Fifth Ave., Suite 1225, New York, NY 10176, Tel: +1 212-661-8528,
                                                                                                                www.aes.org
Bohn, D., A., “Environmental Effects on the Speed of Sound” J. Audio Eng. Soc., Vol. 36, No. 4, April 1988
                                                                                                                Beranek, Leo L. and Vér, István L., Noise and Vibration Control Engineering: Principles and Applications, John Wiley &
                                                                                                                Sons, New York, NY, 1992 (2nd edition published 2005)
                                                                                                                3. INSTRUMENTATION
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                A sound level meter that meets the Type 1 requirements of ANSI S1.4a is required. As an alternative to making direct
                                                                                                                measurements using a qualified sound level meter, a microphone, measuring amplifier and a sound data acquisition system
                                                                                                                may be used, provided the system meets the requirements of SAE J184.
                                                                                                                A third-octave filter set covering the range of center frequencies of interest. The filters shall meet the Class III requirements
                                                                                                                of ANSI S1.11.
A sound level calibrator that meets the Class 1 requirements of ANSI S1.40.
                                                                                                                An acoustical sound generating system shall be selected to generate random noise containing a continuous distribution of
                                                                                                                frequencies over each test band.
3.5 Instrumentation
                                                                                                                                            Random
                                                                                                                                                                      Spectrum              Power
                                                                                                                                             Noise                                                                    Speaker(s)
                                                                                                                                                                      Shaper(s)           Amplifier(s)
                                                                                                                                           Generator(s)
                                                                                                                                        Source
                                                                                                                                        Room
                                                                                                                                                                       Preamp &
                                                                                                                                                                                          Spectrum
                                                                                                                                                                       Amplifier                                      Computer
                                                                                                                                                                                          Analyzer
                                                                                                                                                                       Channels
                                                                                                                                       Receiving
                                                                                                                                       Chamber
                                                                                                                                                                           Data Processing System
                                                                                                                Temperature and humidity sensors should be used to monitor and record ambient test conditions in the source and receiving
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
• ±1 °C for temperature
4. FACILITIES
                                                                                                                                                              Reverberant
                                                                                                                                                              Source Room                            Receiving
                                                                                                                                                                                                     Chamber
                                                                                                                Two adjacent chambers are located such that they share a common test window, into which test samples are mounted.
                                                                                                                The two chambers must be isolated from each other so that the only significant transmission path is through the sample
                                                                                                                window. Test windows are normally oriented in the vertical plane, but may also be oriented horizontally or even non-
                                                                                                                orthogonally to the vertical or horizontal planes.
                                                                                                                It is recommended that the receiving chamber be fully anechoic in order to minimize the influence of flanking paths and
                                                                                                                modal coupling between the chambers. However, good results can also be achieved with semi-anechoic or even reverberant
                                                                                                                receiving chamber designs if proper care is taken for isolation between the chambers, modal decoupling through the test
                                                                                                                window and spatial averaging of receiving chamber microphone(s).
                                                                                                                Source room volume is required to be at least 50 m3 (1765 ft3), with approximately 200 m3 (7100 ft3) being the recommended
                                                                                                                volume. Ideally, room/chamber sizes should be established based on the presence of at least twenty (20) natural frequency
                                                                                                                modes within the lowest 1/3 octave frequency band of interest for good diffusion of sound. For rectilinear design, room
                                                                                                                proportions of 1:21/3:41/3, or 1:1.26:1.59, are also recommended for good diffusion and modal separation. Table 1 shows
                                                                                                                low frequency cutoff for various room sizes using the recommended dimensional proportions. However, rooms smaller than
                                                                                                                those specified in Table 1 may be usable if measured diffusion criteria are met – this can be accomplished by use of low
                                                                                                                frequency absorbers. Appendix X1 of ASTM E90 suggests some alternate volume criteria based on a smaller number of
                                                                                                                resonant modes in each 1/3-octave band. Reverberant receiving chamber proportions should be the same; but sized at
                                                                                                                least 10% smaller or larger in interior volume to avoid natural frequencies which coincide with the source room. It is also
                                                                                                                advisable to keep reverberant receiving chamber boundaries and sound absorber elements at least ¼-wavelength from the
                                                                                                                sample test window. It is possible to do a test using the methodology of SAE J1400 using a small reverberation chamber
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
as a source room, but this will increase the low frequency cutoff for valid test results.
                                                                                                                To qualify a test chamber once it has been built or whenever changes are made to the chamber which may affect diffusion,
                                                                                                                it is recommended that the population standard deviation of twenty randomly located sound pressure measurements in the
                                                                                                                source room be no more than 2 dB at the 1/3-octave center frequency which is one octave higher than the low frequency
                                                                                                                cutoff and above (for example, 200 Hz and above for a room with 100 Hz low frequency cutoff), with a representative test
                                                                                                                sample mounted in the test window. Microphones should be spaced at least ¼-wavelength from each other, from diffusers
                                                                                                                and from any room boundaries at the lowest measurement frequency. Diffusion in the reverberation source room can be
                                                                                                                enhanced by the use of rotating or stationary diffusers. Note that excessively large reverberant source rooms may not exhibit
                                                                                                                diffuse field characteristics at high frequencies due to air absorption. See Section 7 of ASTM C423 for guidance on
                                                                                                                reverberation room design. See Annex A3 and Appendix X1 of ASTM C423 regarding room qualification and the use of
                                                                                                                diffusers to reduce the variability of sound pressure levels within the source room. See 4.9 regarding proper location and
                                                                                                                orientation of source room speaker(s). NOTE: ASTM C423 and ASTM E90 procedures specify 5000 Hz as the highest
                                                                                                                1/3-octave band. In order to measure accurately at higher frequencies, diffusion at these higher frequencies must be
                                                                                                                adequate. It is reasonable to extend the 5000 Hz values in Table A3.1 of ASTM C423 to higher frequencies for the purposes
                                                                                                                of room qualification. Ambient conditions (temperature and relative humidity) in the source room must be controlled to
                                                                                                                minimize air absorption – see Section 6 of ASTM C423 and ANSI S1.26.
                                                                                                                SAE INTERNATIONAL                                       J1400™ JUL2017                                                    Page 8 of 23
                                                                                                                         NOTE: Minimum room volume figures are based on a minimum of 20 resonant modes in the lowest 1/3-octave band of interest using
                                                                                                                               recommended dimensions of 1:1.26:1.59. The low frequency limit for a given size room may be extended even lower by good
                                                                                                                               design of low frequency absorption/diffusion.
                                                                                                                         NOTE: Calculations use sound speed of 343 m/s (1125 ft/s) corresponding with an air temperature of 20 °C (68 °F).
                                                                                                                Source and receiving chamber temperatures should be controlled to 22 °C ± 5 °C. Relative humidity should be controlled
                                                                                                                for both chambers at 40 to 70%. Temperature and humidity should not vary by more than ±3 °C and ±5% RH between
                                                                                                                measurements of the reference sample(s) and the test sample(s).
                                                                                                                The source and receiver chambers should have wall constructions of a sufficient STL to minimize flanking paths into the
                                                                                                                receiving chamber. If the existing isolation between test chambers is adequate, supplemental insulation added to potential
                                                                                                                flanking noise surfaces should have little or no effect on measured STL of maximum transmission loss samples. If
                                                                                                                supplemental insulation is beneficial, it should be left in place where feasible.
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                When the STL of the wall(s) separating the receiving chamber from the source room is not sufficiently greater than the
                                                                                                                sample under test, the measurements may be compromised by sound transmitted from source to receiving chamber via
                                                                                                                paths other than through the specimen under test. The maximum STL measurement capability of the system should be
                                                                                                                determined by measuring the STL of an insulator assembly composed of a 10 to 50 mm foam or fiber decoupler layer in
                                                                                                                combination with a 2 to 10 kg/m2 septum or barrier layer. Continue to add successive insulator assemblies and measure
                                                                                                                STL after adding each insulator assembly until no further changes in STL are noted (less than 1.0 dB change at all
                                                                                                                frequencies). Note that a second set of measurements may be required using an insulator assembly composed strictly of
                                                                                                                mass layers to determine the maximum STL at lower frequencies. The maximum STL capability is limited by flanking paths
                                                                                                                or residual noise in the overall measurement system. Measured sample STL values should fall at least 10 dB below
                                                                                                                maximum STL capability levels at all frequencies. Maximum STL levels can be improved by increasing the STL of the
                                                                                                                common wall between the rooms, by improving the sample sealing system, increasing the structure-borne vibration isolation
                                                                                                                between rooms or by decreasing the residual noise of the measurement system. Maximum STL measurement capability
                                                                                                                does not need to be run with every sample; but, it should be measured periodically (recommended annually) or whenever
                                                                                                                changes may have been made affecting the measurement facility and/or data acquisition system.
                                                                                                                A test sample fixture should hold the test sample securely between the source and receiving chambers. The fixture should
                                                                                                                be well sealed to prevent leakage between the source and receiving chambers through the fixture (see recommended
                                                                                                                sample mounting system in Appendix A). The fixture should provide means to maintain typical in-use contact between the
                                                                                                                various layers of the test assembly. Unless intended, care must be taken to assure that no air gaps are induced in a
                                                                                                                multilayer assembly or within sample layers during sample mounting or sealing. Although it is common practice to use a
                                                                                                                vertical test window orientation, horizontal test windows may provide a means of using gravity to naturally hold test layers
                                                                                                                properly together or to represent the in-use compression of layers.
                                                                                                                SAE INTERNATIONAL                                    J1400™ JUL2017                                              Page 9 of 23
                                                                                                                For materials with low inherent damping, the sample fixture may impart significant damping into the assembly, influencing
                                                                                                                the measured STL, especially near coincidence. This effect is more significant as the sample size decreases. The STL of
                                                                                                                a measured substrate should be compared with the theoretical or large sample value to insure that the test fixture does not
                                                                                                                adversely affect the results.
4.9 Loudspeakers
                                                                                                                One or more broadband loudspeakers of sufficient sound power capability should be used to produce sound pressure levels
                                                                                                                in the receiving chamber at least 10 dB above the noise floor of the measurement system and chamber at all frequencies
                                                                                                                with the test sample in place. It is recommended that sound spectrum shaping in the source room be utilized in order to
                                                                                                                reduce the span between the lowest and highest levels versus frequency of the measured sound spectrum in the receiving
                                                                                                                chamber to within the dynamic range capability of the measurement system. Typically, a source room sound pressure
                                                                                                                spectrum rising at 6 dB/octave will significantly reduce the dynamic range requirements in the receiving chamber. Selection,
                                                                                                                exact placement and orientation of the loudspeaker(s) within the source room are often trial-and-error processes to achieve
                                                                                                                desired source levels and diffusion. Should multiple loudspeakers be used, it is recommended that uncorrelated signals be
                                                                                                                fed to the broadband loudspeaker sources for best low frequency diffusion. To help define proper loudspeaker placement
                                                                                                                and orientation, the population standard deviation of the randomly located or spatially averaged sound pressure
                                                                                                                measurements in the source room should follow the population standard deviation guideline as previously defined in 4.4.
Once placed and qualified, loudspeaker position(s) and orientations must be maintained for all tests.
4.10 Microphones
                                                                                                                One or more random incidence microphones (reverberant chambers) or free field microphones (hemi-anechoic or anechoic
                                                                                                                chambers) shall be positioned within the source and receiving chambers. The number and spacing of microphone positions
                                                                                                                required in each room depends on the statistical precision desired in the time and space average band sound pressure
                                                                                                                levels.
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
4.10.1 Microphone Placement in Reverberant Source Room and Reverberant Receiving Room
                                                                                                                Randomly placed microphones or the traverse of a spatially averaged microphone should maintain at least ¼-wavelength
                                                                                                                distances from room boundaries, room diffusers, noise sources and the test sample window at the lowest test frequency
                                                                                                                consistent with the microphone placement for source room qualification specified in 4.4. For more specifics please refer to
                                                                                                                Section 10.1 of ASTM C423.
                                                                                                                The exact number and placement of receiving chamber microphones for best results is often a trial-and-error process. A
                                                                                                                recommended starting point is to place microphones 10 to 100 mm from the receiving room surface of the test sample. The
                                                                                                                population standard deviation of the randomly located or spatially averaged sound pressure measurements in the source
                                                                                                                room should still follow the population standard deviation guideline as previously defined in 4.4 – i.e., be no more than 2 dB
                                                                                                                at 200 Hz 1/3-octave center frequency and above, with a representative test sample mounted in the test window. The
                                                                                                                distance of the microphone from the surface of the sample facing the receiving room should remain constant. For example,
                                                                                                                if 20 mm from the face of the sample is selected for a test of an initial sample, and the next sample in the test is substantially
                                                                                                                thicker or thinner than the previous sample, the microphones should still be placed 20 mm from the face of the new sample.
                                                                                                                Note that the construction of a control sample is defined in 6.3, with target STL values that are defined in 6.4. These can
                                                                                                                be used to aid in the above trial-and-error process.
Once placed, microphone positions must be maintained throughout the test sequence and ideally for all tests.
                                                                                                                For finite size panels, random incidence or field-incidence transmission loss in the mass-controlled region will be higher for
                                                                                                                smaller panels, and this effect will be more pronounced at lower frequencies. The CF calculated using the reference limp
                                                                                                                barrier essentially converts the MNR to field-incidence STL for an infinite panel.
                                                                                                                SAE INTERNATIONAL                                           J1400™ JUL2017                                                     Page 10 of 23
                                                                                                                There is not a specific limit to window size. However, in order to meet the CF requirements in 5.4.3 (in terms of difference
                                                                                                                between MNR and calculated limp barrier STL) the user should be aware of the low frequency effect of finite window size.
                                                                                                                Appendix E contains a brief discussion of the theory behind this. Table 2 shows the minimum window size necessary for
                                                                                                                5 dB or 10 dB deviation of ideal MNR from calculated limp barrier STL. The actual CF for a given window may be somewhat
                                                                                                                higher or lower, depending on microphone placement and other factors. As long as CF requirements are met as specified
                                                                                                                in 5.4.3 the window opening sizes in Table 2 do not need to be strictly observed. However, the smaller the test window is,
                                                                                                                the more the CF will approach the specified limits. Although Table 2 is specific to square test samples, the theory is generally
                                                                                                                applicable to non-square openings and these figures are useful as a guideline. A useful empirical rule of thumb is to consider
                                                                                                                the low frequency limit to be that for which the diagonal opening of the test window is ¼ of the wavelength (in air).
                                                                                                                                    Table 2 - Effect of diagonal dimension of a square test window at lowest measurement frequency
                                                                                                                                        and comparison with ¼-wavelength of lowest 1/3-octave band frequency
5. PROCEDURE
                                                                                                                Test samples must be mounted and sealed completely within the test fixture so as to ensure a minimum of sound flanking
                                                                                                                the test sample. The recommended sample mounting system is shown in Appendix A.
                                                                                                                Test samples should be conditioned to the same temperature and humidity as the test chambers for at least 12 hours prior
                                                                                                                to testing.
5.3 Measurements
                                                                                                                Background noise levels within both the source and receiving chambers shall be measured and noted in all measurement
                                                                                                                bands and averaged over all measurement positions for each measurement series. Background noise levels in the receiving
                                                                                                                chamber must be measured at the same gain settings as during normal measurements in order to include the noise floor of
                                                                                                                the measurement system. Ultra-low noise microphones and preamps are available for laboratory use and can be effective
                                                                                                                in lowering the noise floor of the measurement system.
                                                                                                                SAE INTERNATIONAL                                  J1400™ JUL2017                                          Page 11 of 23
                                                                                                                Install and seal the reference sample, a homogeneous limp material such as lead, PVC sheet, EVA sheet or another
                                                                                                                monolithic limp material that does not show a critical frequency phenomenon in the frequency range of interest, into the test
                                                                                                                opening so that its field-incidence STL can be calculated from the relation:
                                                                                                                                                                                                 β 2 +1   
                                                                                                                                      STL(reference sample) = -0.192 + 10log(β 2 ) - 10log ln             
                                                                                                                                                                                             0.043227β +1  
                                                                                                                                                                                                          2
                                                                                                                                                                                                                                      (Eq. 5)
where:
β = ρsω/2ρ0c
ω=2πf
ρ0 = volumetric density of air (kg/m3) at measurement barometric pressure, temperature and humidity
                                                                                                                The procedure for calculating the speed of sound accounting for temperature, humidity and barometric pressure is found in
                                                                                                                Appendix D.
                                                                                                                The surface density of the reference sample should be selected to be within 50 to 200% of the surface density of each test
                                                                                                                panel or multi-layer test assembly as long as the requirements of 5.3.1 are met and the reference sample surface density
                                                                                                                does not exceed 10.0 kg/m2. Reference samples of different surface densities may be required to cover various test sample
                                                                                                                materials
                                                                                                                The source signals may be amplified or filtered versus frequency so that, with the test sample sealed in place, the source
                                                                                                                room and receiving chamber signal levels are each at least 10 dB, and preferably more than 15 dB, higher than the
                                                                                                                background noise levels within the respective chambers and within the dynamic range capability of the measurement system
                                                                                                                at all frequencies of interest.
                                                                                                                The time and spatially-averaged third-octave band levels in both the source and receiving chambers shall be measured and
                                                                                                                recorded over the desired measurement bands with the reference sample sealed into the fixture in the test opening.
                                                                                                                Optionally, time averaged, single point measurements may be used on the receiving side of the sample mounting window
                                                                                                                if they can be shown to give STL results within ±2.0 dB at all measurement frequencies to time and spatially-averaged
                                                                                                                measurements. Averaging times shall be long enough to provide an estimate of the time-averaged level to within ±0.5 dB
                                                                                                                for 95% confidence limits at all measurement frequencies. See 5.5.4 and Appendix B. Measurement distance to the sample
                                                                                                                mounting plane and number of measurement positions in the receiving chamber which give best results will vary from lab
                                                                                                                to lab and are usually determined through trial and error. See 4.10 for microphone placement guidelines.
                                                                                                                Note that the construction of a control sample is defined in 6.3, with target STL values defined in 6.4. These can be used
                                                                                                                to aid in the above trial-and-error process.
                                                                                                                SAE INTERNATIONAL                                   J1400™ JUL2017                                       Page 12 of 23
                                                                                                                After removing the reference sample, the test sample is installed and sealed into the same opening and in the same manner
                                                                                                                as the reference sample. The measurements are then repeated as in 5.3.5. The test sample may be a homogeneous single
                                                                                                                layer material, a multi-layer material, a combination of multilayer materials with a sheet metal backing, a porous material
                                                                                                                without an impervious layer, or any of the previous materials with a pass-through, opening or intentional leakage path. The
                                                                                                                results for each of these systems are compared to the results for the reference sample tested in 5.3.5. If microphones are
                                                                                                                moved during test sample mounting, they must be accurately replaced (within 3 mm) to the same positions as in prior
                                                                                                                measurements, including reference sample measurements. In a hemi-anechoic or anechoic receiving room, distance from
                                                                                                                the surface of the sample facing the receiving room shall remain the same.
                                                                                                                Some studies have found that the method of sealing the edges of a lightly damped test sample (e.g., a steel or aluminum
                                                                                                                panel) can affect the STL measured according to SAE J1400, particularly in the vicinity of the coincidence frequency of the
                                                                                                                test panel. To minimize this artifact it is strongly recommended that the minimum necessary amount of material be used to
                                                                                                                seal the edges of the sample to the test fixture. A convenient tool to use to check for sealing gaps is an ultrasonic leak
                                                                                                                detector.
The following procedures are used to calculate the field incidence STL of the test sample.
                                                                                                                If the difference between background noise level and signal plus background noise level is between 10 and 15 dB, correct
                                                                                                                for background noise levels at each measured third octave frequency band of interest and at each microphone for both the
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
where:
                                                                                                                For both the reference sample and the test sample, compute the measured noise reduction (MNR) at each one-third octave
                                                                                                                band of interest. Using the corrected band pressure levels, if required, for each measurement band, subtract the receiving
                                                                                                                chamber band pressure level from the source room band pressure level to obtain the MNR for both samples. Where
                                                                                                                applicable, use spatially-averaged values for source and/or receiving room/chamber.
                                                                                                                Determine the correlation factor applicable to the test opening and source/receiving chamber pair at each test frequency
                                                                                                                (CFf) as the difference between the measured noise reduction of the reference sample, MNRf (reference) and its calculated
                                                                                                                STLf (reference).
                                                                                                                NOTE: Correlation factors at all frequencies should fall within +10/-0 dB for a well-implemented test system, +15/-0 dB for
                                                                                                                      a typical system and should not exceed the range of +15/-5 dB. Methods to reduce the correlation factor without
                                                                                                                      major facility changes would include changing the position of the receiving microphone(s), increasing the absorption
                                                                                                                      in the receiving chamber, improving the sealing and/or sample mounting system, increasing the number of receiving
                                                                                                                      room microphones, averages or spatial averaging in the receiving room.
                                                                                                                NOTE: The correlation factors (CFf) determined using this methodology are subsequently used for computing the STL of
                                                                                                                      multi-wall samples as well as single wall samples.
                                                                                                                Compute the sound transmission loss (STLf) of the test sample at each test frequency by subtracting the CFf from the MNRf
                                                                                                                of the test sample:
                                                                                                                STLf (sample) at each frequency band of interest should be rounded to the nearest whole dB.
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
                                                                                                                Sometimes an insertion loss (IL) is desired. In the context of SAE J1400, IL is defined as the difference between the STL
                                                                                                                of a substrate (e.g., a metal panel) and the STL of the substrate plus insulator. This can be easily calculated at each
                                                                                                                frequency band as:
Similarly, if the CF meets the requirements of 5.4.3, then the IL may be calculated directly from the MNR as:
5.5 Reporting
The following shall be included when reporting results of these test procedures.
                                                                                                                Measurement date, test location, person performing tests, sample description(s) and reference sample(s) used. It is also
                                                                                                                required to specifically state any deviations to the SAE J1400 procedure requirements, if any. Sample description shall
                                                                                                                include dimensions, weight, overall composition/number of layers, which surface faces the source room and sample
                                                                                                                orientation (horizontal, vertical or other). Photos of the each sample as installed should be included. Microphone positions
                                                                                                                within source/receiving rooms, number of microphones used (or a single microphone on a rotating boom) should all be
                                                                                                                identified. Microphone/preamp model numbers should be recorded.
                                                                                                                Ambient temperature, local barometric pressure and humidity conditions in each test chamber at time of measurements.
                                                                                                                SAE INTERNATIONAL                                  J1400™ JUL2017                                          Page 14 of 23
                                                                                                                Sound transmission loss rounded to the nearest whole dB versus 1/3-octave center frequencies in Hz. Measurements
                                                                                                                which have been corrected for background noise should be marked with an asterisk and a note explaining such. See 5.4.1.
                                                                                                                If the user does not have specific data presentation requirements (e.g., internal company standards) the SAE J2629
                                                                                                                document and spreadsheets provide data presentation formats for SAE J1400 and other acoustical tests.
                                                                                                                95% Confidence Limits in dB versus 1/3-octave center frequencies in Hz are defined in Appendix B. Calculations for
                                                                                                                Confidence Limits do not have to be made for each test, but should be done annually or following any changes made to the
                                                                                                                chamber and/or measurement system which may affect Confidence Limits. Confidence Limits may be shown graphically
                                                                                                                as upper and lower bounds for STL at each frequency (i.e., “error bars”).
Maximum STL capability rounded to the nearest whole dB versus 1/3-octave center frequencies in Hz. See 4.7.
6. GENERAL COMMENTS
                                                                                                                It is essential that technically qualified personnel trained in the current techniques of sound measurements select equipment
                                                                                                                and perform the tests.
Instrumentation manufacturers' or quality standard recommended calibration practices should be followed before each test.
                                                                                                                A control sample has been developed to aid in the assessment and minimization of intra-laboratory variability and/or bias.
                                                                                                                See Appendix C for construction details. Specific sources of construction materials are mentioned as guidelines. Alternative
                                                                                                                sources may work as well.
                                                                                                                The following table is based on a weighted average of multiple control samples produced independently and tested in a
                                                                                                                number of independent test laboratories. All qualified laboratories should be able to reproduce these results within 3 dB at
                                                                                                                all frequencies. Laboratories having trouble with reproducibility particularly below 1000 Hz should seek improvements as
                                                                                                                suggested in 4.3 and 4.8. If measured STL is too low relative to target levels above 2500 Hz, improvements as suggested
                                                                                                                in 4.6 and 4.9 are recommended. Note that stiffness effects of finite size panels may affect low frequency STL. Every lab
                                                                                                                may not be able to reproduce the table shown below in its full frequency and STL ranges. A good understanding of lab
                                                                                                                limitations (flanking paths, source room diffuse field cutoff, effects of window size, etc.) should provide understanding of
                                                                                                                how much of Table 3 is valid.
                                                                                                                SAE INTERNATIONAL                                   J1400™ JUL2017                                           Page 15 of 23
7. NOTES
                                                                                                                A change bar (l) located in the left margin is for the convenience of the user in locating areas where technical revisions, not
                                                                                                                editorial changes, have been made to the previous issue of this document. An (R) symbol to the left of the document title
                                                                                                                indicates a complete revision of the document, including technical revisions. Change bars and (R) are not used in original
                                                                                                                publications, nor in documents that contain editorial changes only.
Holding Frame
                                                                                                                                                                                                     Sealing Putty
                                                                                                                                                                                                                           Test Sample
                                                                                                                                                                                                                            Sheetmetal
                                                                                                                                                                                                                             (optional)
                                                                                                                                                                                                            Sealing Putty
Insulation as required
                                                                                                                The test sample shown here is installed from the source room, which is recommended for convenience of sample mounting.
                                                                                                                In some cases, test samples may not have a sheet metal base panel or it may be a different material/composite. Test
                                                                                                                window orientation may be horizontal or non-orthogonal in order to duplicate end-use orientation, particularly if gravity holds
                                                                                                                test samples in place. The sheet metal-to-sample orientation may be reversed (sheet metal facing source room) in order
                                                                                                                to have the sound field incident on the same face as the intended application.
                                                                                                                SAE INTERNATIONAL                                   J1400™ JUL2017                                        Page 17 of 23
B.1 ACCURACY
                                                                                                                Accuracy determines how well the test data measures the STL properties of the test specimen. Since this test protocol is
                                                                                                                defined, the accuracy of a test conducted by this method is governed by the design of the test facility and by the sample
                                                                                                                mounting and microphone placement procedures.
B.2 PRECISION
                                                                                                                Precision determines the variation of the average measurement for a certain confidence level. The precision of this test is
                                                                                                                determined by the variation in each measured data set and is calculated from the confidence intervals for each data set.
                                                                                                                The overall confidence interval for the resulting data is derived from the confidence intervals for the individual measured
                                                                                                                quantities. The following paragraphs describe the steps to be taken to collect and calculate confidence intervals.
                                                                                                                Calculate the standard deviation, sf, for the mean of each time and space averaged one-third octave band sound pressure
                                                                                                                level for LSRf, LRRf, LSTf, and LRTf
where:
                                                                                                                      LSRf = source room sound pressure level for the reference sample
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
LRRf = receiving chamber sound pressure level for the reference sample
LSTf = source room sound pressure level for the test sample
LRTf = receiving chamber sound pressure level for the test sample
where:
Xif = individual measurement of sound pressure level at each frequency band of interest
XMEANf = arithmetic mean of the set of sound pressure levels at each frequency band of interest
                                                                                                                      n = number of measurements of sound pressure level in each set; a minimum of six (6) is recommended
                                                                                                                SAE INTERNATIONAL                                     J1400™ JUL2017                                         Page 18 of 23
Calculate the 95% confidence interval for the individual measurement frequency bands from:
where:
                                                                                                                                                       13                                        0.604
                                                                                                                                                       14                                        0.577
                                                                                                                                                       15                                        0.544
                                                                                                                                                       16                                        0.533
                                                                                                                                                       17                                        0.514
                                                                                                                                                       18                                        0.497
                                                                                                                                                       19                                        0.482
                                                                                                                                                       20                                        0.468
                                                                                                                                                       21                                        0.455
                                                                                                                                                       22                                        0.443
                                                                                                                                                       23                                        0.432
                                                                                                                                                       24                                        0.422
                                                                                                                                                       25                                        0.413
                                                                                                                                                n greater than 25                       a = n0.5/(0.512n-0.71)
Calculate the confidence interval for each frequency band of interest from:
delta STLf = SQRT { (delta LSRf)2 + (delta LRRf)2 +(delta LSTf)2 +(delta LRTf)2} (Eq. B3)
B.3 REPEATABILITY
                                                                                                                Repeatability determines the success in obtaining identical test results on the same sample. Within the limits for precision
                                                                                                                defined in B.2.4, the repeatability of a test on any sample is influenced by the mounting and microphone
                                                                                                                placement/replacement procedures. If mounting and microphone placement practices are closely replicated, then
                                                                                                                repeatable results should be expected within the limits of precision.
                                                                                                                SAE INTERNATIONAL                          J1400™ JUL2017                                       Page 19 of 23
                                                                                                                For accurate calculation of reference sample STL, and ultimately the STL of the test sample, the speed of sound should be
                                                                                                                corrected to account for ambient conditions temperature, barometric pressure and humidity.
                                                                                                                           Step by step calculations to compute speed of sound (c) with temperature, humidity and
                                                                                                                           barometric pressure:
                                                                                                                               •   Speed of sound in dry air (cd) with temperature:
                                                                                                                                                        𝑇𝑇
                                                                                                                                   𝑐𝑐𝑑𝑑 = 331.45 �1 +   273
                                                                                                                                                                    (m/s)                                               (A-1)
                                                                                                                                                                      𝑐𝑐            𝛾𝛾𝑤𝑤
                                                                                                                                                                          = 4.5513�
                                                                                                                                                                     𝑐𝑐𝑑𝑑          𝑀𝑀𝑤𝑤
                                                                                                                                                                                      𝛾𝛾
                                                                                                                                                                 𝑐𝑐 = 4.5513 𝑐𝑐𝑑𝑑 �𝑀𝑀𝑤𝑤          (m/s)
                                                                                                                                                                                           𝑤𝑤
                                                                                                                SAE INTERNATIONAL                                J1400™ JUL2017                                  Page 21 of 23
For example:
As a convenient reference, Table D1 shows water vapor pressure values at various temperatures.
                                                                                                                The discussion following is based on chapter 9 of the first edition of Noise and Vibration Control Engineering: Principles and
                                                                                                                Applications. Field-incidence includes angles of incident sound from 0 (normal to the panel) to 78°, as opposed to random-
                                                                                                                incidence, ranging from 0 to 90°. Field-incidence prediction for infinite panels has been found to correlate better to measured
                                                                                                                results for large panels than random-incidence prediction. A generally accepted rule is that for normal-incident STL greater
                                                                                                                than 15 dB, field-incident STL at frequencies well under coincidence frequency and in the mass-controlled region for an
                                                                                                                infinite panel can be found as:
                                                                                                                For a finite panel in the mass-controlled region and well below coincidence frequency, a random-incidence prediction which
                                                                                                                agrees well with measured results is:
where
and
                                                                                                                σF is the forced-wave radiation efficiency for random-incident sound, k0 is the wavenumber calculated for the 1/3-octave
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51
band frequency of interest and S is the panel surface area. σF increases with frequency and with panel dimensions.
                                                                                                                To illustrate the effect of panel size see Figure E1, which shows nominal field incidence transmission loss in the mass
                                                                                                                controlled region for an infinite plate as compared with finite size plates of 0.5 x 0.5 m and 1.4 x 1.4 m.
                                                                                                                SAE INTERNATIONAL                            J1400™ JUL2017                                      Page 23 of 23
                                                                                                                 Figure E1 - Comparison of calculated field incidence transmission loss for finite size square plates versus an
                                                                                                                                                                  infinite plate
Normen-Download-Beuth-Shlomo Ronen Advanced Building Consulting-KdNr.7916586-LfNr.8597737001-2018-10-14 13:51