Oeno 18 2003 en
Oeno 18 2003 en
Upon the proposal of the Sub-commission of the Methods of Analysis and Appraisal of Wine,
DECIDES:
To replace and complete Chapter II of the International Oenological Codex by the following
analytical and control techniques:
                                                                                       1
                  RESOLUTION OENO 18/2003
1. APPARATUS
        1.1 The apparatus used for separating NH3 is either a distillation apparatus with a
rectifying column or a distillation apparatus under a current of steam (diagram) made up
of:
        A 1 l flask A of borosilicate glass used as a boiler with a stopcock funnel for filling.
It can be heated by a gas or electric furnace.
        An adapter C which gathers the spent liquid from the bubbler B.
        A bubbler B of 500 ml with an inclined neck; the supply tube must reach the
lowest part of the flask. The out-going tube has an anti-entrainment ball that makes up
the top part of the bubbler. A stop-cock funnel E allows to introduce the liquid to be
treated and alkaline lye.
        A cooler 30 to 40 cm long, vertical, with a ball with fine dowel bush on the tip.
        A 250 ml conical flask for the distillate.
2. REAGENTS
       Concentrated sulphuric acid (R).
       Mineralisation catalyser (R).
       Sodium hydroxide solution at 30% (m/m) (R).
       Boric acid solution at 4% (R).
       Hydrochloric acid solution 0.1 M.
       Mixed-based indicator with methyl red (R) and methylene blue.
       The boiler must contain acidulated water by 1 per 1 000 of sulphuric acid. It is
advisable to boil this liquid before any operation, with the drain cock P open to let the
CO2 escape.
3. PROCEDURE
        Heat in an open flame under a hood. The neck of the flask is maintained inclined
until the solution becomes colourless and the walls of the flask are clear of carbonised
products.
        After cooling, dilute with 50 ml of water and cool; introduce this liquid in the
bubbler B with the funnel E, then add 40 to 50 ml of sodium hydroxide solution at 30%
(R) in order to obtain frank alkalinisation of the liquid. Entrain the ammoniac with the
vapour by gathering the distillate in 5 ml of boric acid solution (R) placed beforehand in a
receiving conical flask with 10 ml of water, with the tip of the ampoule plunged into the
liquid. Add 1 or 2 drops of mixed-based indicator and gather 70 to 100 ml of distillate.
                                                                                              2
                 RESOLUTION OENO 18/2003
       Titrate the distillate with the hydrochloric acid solution 0.1 M until the indicator
turns pink violet.
                                                                                         3
                 RESOLUTION OENO 18/2003
1 – FIELD OF APPLICATION
        This method is applied to the analysis of mercury in oenological products in the
    concentration range of 0 to 10 µg/l.
                                                                                         4
                RESOLUTION OENO 18/2003
                                                                                      5
                 RESOLUTION OENO 18/2003
4 – APPARATUS
    4.1. Glassware:
        4.1.1. graduated flasks 100, 200 and 1000 ml (class A)
        4.1.2. graduated pipettes 0.5; 1.0; 2.0; 5; 10 and 20 ml (class A)
        4.1.3. precautions: before use, the glassware must be washed with nitric acid at
10%, left in contact for 24 hours, then rinsed with demineralised water.
    4.2. Mineralisation apparatus (see Compendium of international methods of
analysis of wines and musts)
    4.3. Thermostatic heating mantle
    4.4. Peristaltic pump
    4.5. Cold vapour generator
        4.5.1. gas-liquid separator
   4.6. Dessicator (hygroscopic membrane) covered by an air current (supplied by a
compressor) and placed before the detector
    4.7. Spectrofluorimeter:
        4.7.1. mercury vapour lamp, adjusted to the wave length of 254 nm
        4.7.2. specific atomic fluorescence detector
    4.8. PC:
         4.8.1. software that adjusts the parameters of the vapour generator and atomic
fluorescence detector and allows calibration and the analysis of results.
        4.8.2. printer that archives results
    4.9. Bottle of neutral gas (argon)
                                                                                         6
                 RESOLUTION OENO 18/2003
the white smoke. This operation is repeated; after a third time, boil an instant, cool,
stabilise with several drops (about 10) of potassium permanganate (aqueous sol.) at 5%
(m/m) and add water to the liquid to reach 40 ml.
Filter on filters without cinders. Introduce 10 ml of filtrate into a 50 ml flask. Add
potassium permanganate (3.8.) until persistence of coloration. Solubilise the precipitate
(MnO2) with the reducing solution (3.10.). Complete to volume with demineralised water
(3.1.).
Do a blank test with demineralised water.
6 – PROCEDURE
    6.1. Analytical determination
    Turn on the fluorimeter; the apparatus is stabilised after 15 minutes.
    The peristaltic pump draws up the blank solution (3.3.), the tin chloride (II) solution
(3.13.) and the calibrations or samples (5.1.) or (5.2.).
    Check if there is a bubbling in the gas-liquid separator.
     Present successively the calibration solutions (5.1.); start the programming of the
vapour generator. The computer software sets up the calibration curve (percentage of
fluorescence depending on the concentration of mercury in µg/l).
    Then present the samples (5.2.).
    6.2. Self-check
     Every five determinations, an analytical blank solution and a calibration are analysed
in order to correct a possible drift of the spectrofluorimeter.
7 – EXPRESSION OF RESULTS
    The results are given by the computer software and are expressed in p.p.b. (or µg/l).
     The concentration of mercury in oenological products is calculated according to the
test sample and the dilution of the mineralisate. It is expressed in µg/kg.
8 – CONTROL OF RESULTS
      The quality control is performed by placing, after the set of calibration solutions and
all five samples, a reference material whose mercury content is known with certainty.
     A control card is set up for each reference material used. The control limits are set
at: +/- 2SR intra (SR intra: standard deviation for reproducibility).
                                                                                           7
                RESOLUTION OENO 18/2003
10 – BIBLIOGRAPHY
     10.1. CAYROL M., BRUN S., 1975. Dosage du mercure dans les vins. Feuillet Vert de
l’O.I.V. n°371.
     10.2. REVUELTA D., GOMEZ R., BARDON A., 1976. Dosage du mercure dans le vin
par la méthode des vapeurs froides et spectrométrie d’absorption atomic. Feuillet Vert de
l’O.I.V. n°494.
    10.3. CACHO J., CASTELLS J.E., 1989. Determination of mercury in wine by flameless
atomic absorption spectrophotometry. Atomic Spectroscopy, vol. 10, n°3.
    10.4. STOCKWELL P.B., CORNS W.T., 1993. The role of atomic fluorescence
spectrometry in the automatic environmental monitoring of trace element analysis.
Journal of Automatic Chemistry, vol. 15, n°3, p 79-84.
    10.5. SANJUAN J., COSSA D., 1993. Dosage automatique du mercure total dans les
organismes marins par fluorescence atomique. IFREMER, Rapport d’activité.
     10.6. AFNOR, 1997. Dosage du mercure total dans les eaux par spectrométrie de
fluorescence atomique. XPT 90-113-2.
    10.7. GAYE J., MEDINA B., 1998. Dosage du mercure dans le vin par analyse en flux
continu et spectrofluorimétrie. Feuillet Vert de l’O.I.V. n°1070.
                                                                                       8
                RESOLUTION OENO 18/2003
   Heavy metals react with the thiol function to form sulphurs. The coloration that
   results is compared to a standard.
2. Reagents
3. Procedure
   3.1 Test solution: pour 5 ml of buffer solution (2.9), 25.0 g of sample and about 15
        ml of water into a 50 ml graduated flask. Complete with water up to the
        reference mark.
   3.2.2. Comparative solution: mix 2.0 ml of test solution (3.1), 2.0 ml of buffer
           solution (2.9), 0.5 ml of standard lead solution (2.8.2), 4.5 ml of water and
           5.0 ml of methanol in a test tube.
   3.2.3. Control solution: mix 12.0 ml of test solution (3.1), 2.0 ml of buffer solution
           (2.9) and 0.5 ml of standard lead solution (2.8.2) in a test tube.
                                                                                          9
                RESOLUTION OENO 18/2003
- the sample solution must not be darker than the comparative solution.
- the control solution must not be lighter than the comparative solution.
4. Results:
The conditions described in 3.2.4 are obtained if the heavy metal content is less than 10
mg/l expressed in lead and with a precision of 1 mg/l.
                                                                                     10
                 RESOLUTION OENO 18/2003
       In a 160 × 16 mm test tube, place the volume prescribed of the solution obtained
by the means indicated in each monography; add 1 ml of diluted hydrochloric acid (R);
adjust to 20 ml with water and add 2 ml of barium chloride solution at 10% (R).
       Compare the opalescence or any cloudiness to the control sample prepared with 1
ml of solution at 0.100 g of sulphuric acid per litre (i.e. 0.10 mg of H2SO4,) with 1 ml of
diluted hydrochloric acid (R) and water until volume of 20 ml and 2 ml of barium chloride
solution (R). This tube contains 100 µg of H2SO4.
                                                                                        11
                 RESOLUTION OENO 18/2003
BROMINE INDEX
       The bromine index is the quantity of bromine expressed in grammes, that 100 g
of the substance can set.
1. APPARATUS
      A graduated flask of 300 to 400 ml with an interior tube welded at the bottom, an
emery stopper and a tube with a handle, compliant with the following diagram
2. SOLUTIONS
                                                                                       12
                 RESOLUTION OENO 18/2003
3. TECHNIQUE
         Using a tube with a handle, put about 0.50 g of potassium iodide in the recipient
inside the flask; (it is convenient to make a circular mark on the tube corresponding to
the salt’s weight so as not to have to weigh each dosage). Caution has to be taken so as
not to introduce iodide on the external part of the flask. Then introduce the measured
volume of the solution of the product to be measured, dissolved in neutral or alkaline
water, in the external part of the flask, then 25 ml of potassium bromate solution 0.016
M measured with a pipette, and 2 g of pure potassium bromide. Rinse the sides with
water to come to a total volume of about 100 ml, then add 5 ml of concentrated
hydrochloric acid (R); quickly close the flask with the stopper, the joint being humid with
distilled water; by a circular movement homogenise the content and allow to stand the
prescribed time. Shake the flask vigorously so as to put the potassium iodide in contact
with the liquid so as to enable the vapour bromine to react; open the flask while rinsing
the joint and the stopper with a spray of distilled water, and determine iodine using 25
ml of sodium thiosulphate solution 0.1 M; titrate the excess of sodium thiosulphate with
the iodine solution 0.05 M in the presence of starch paste;
                                                                                        13
                   RESOLUTION OENO 18/2003
        In a 160 × 16 mm test tube, place the volume prescribed of the solution obtained
 by the means indicated in each monography; add 5 ml of diluted nitric acid (R); complete
 to 20 ml and add 0.5 ml of silver nitrate solution at 5% (R).
         Compare the opalescence or any cloudiness to the control sample prepared with
0.5 ml of hydrochloric acid at 0.10 g per litre (0.05 mg of HCl) with 5 ml of diluted nitric
acid (R), and adjust to 20 ml with distilled water. Add 0.5 ml of silver nitrate solution at
5% (R). This tube contains 50 µg of HCl.
                                                                                               14
                 RESOLUTION OENO 18/2003
Method applicable for determining the following elements: calcium, magnesium, sodium,
iron, copper, zinc.
                     Gently burn the sample with the flame of a Bunsen burner under a
hood.
                     Put the capsule in a muffle oven at 525°C ± 25°C for 12 hours.
                     Take up the residue with a few ml of demineralised water.
                     Evaporate water over a water bath at 100°C.
                     Replace the capsule containing the sample in the oven.
                     The mineralisation is over when the cinders are white.
       Method applicable for determining the following elements: arsenic, cadmium, lead
in oenological products containing water.
                                                                                            15
                 RESOLUTION OENO 18/2003
Three grammes of tantalum powder are put in a 100 ml Teflon  cylindrical vase.
Add 10 ml of hydrofluoric acid diluted to a half, 3 g of dehydrated oxalic acid and 0.5 ml
of hydrogen peroxide at 30 vol.
Heat carefully to dissolve the metal.
Add a few drops of hydrogen peroxide as soon as the reaction slows down; when the
dissolution is complete, add 4 g of oxalic acid and 30 ml of water.
The acid is dissolved and the solution is brought to 50 ml with ultra pure demineralised
water.
Store this solution in a plastic flask.
The platform is placed inside the graphite tube or used pyrolytic graphite tube. It is set to
the unit of atomisation of the spectrophotometer.
REFERENCE:
                                                                                          16
                 RESOLUTION OENO 18/2003
1 – FIELD OF APPLICATION
    This method applies to the analysis of arsenic in the concentration range of 0 to 200
µg/l with prior mineralisation for oenological products.
                                                                                      17
                 RESOLUTION OENO 18/2003
4 – APPARATUS
    4.1. Glassware:
        4.1.1. graduated flasks 50 and 100 ml (class A)
        4.1.2. graduated pipettes 1, 5, 10 and 25 ml (class A)
        4.1.3. cylindrical vases 100 ml
    4.2. Hot plate with thermostat
    4.3. Ashless filter paper
    4.4. Atomic absorption spectrophotometer:
        4.4.1. air-acetylene burner
        4.4.2. hollow-cathode lamp (arsenic)
        4.4.3. deuterium lamp
    4.5. Accessories:
        4.5.1. vapour generator (or gas-liquid separator)
        4.5.2. quartz absorption cell placed on the air-acetylene burner
        4.5.3. bottle of neutral gas (argon)
                                                                                        18
                 RESOLUTION OENO 18/2003
6. PROCEDURE
    6.1. Instrumental parameters of the atomic absorption spectrophotometer
         (given as an example)
        6.1.1. oxidant air-acetylene flame
        6.1.2. wave length: 193.7 nm
        6.1.3. width of the monochromator’s slit: 1.0 nm
        6.1.4. intensity of the hollow-cathode lamp: 7 mA
        6.1.5. correction of the non specific absorption with a deuterium lamp
    6.3. Self-check
     Every five determinations, an analytical blank solution and a calibration are analysed
in order to correct a possible deviation of the spectrometer.
7. EXPRESSION OF RESULTS
        The results are directly printed by the printer connected to the computer.
    The concentration of arsenic in oenological products is expressed in µg/kg while
    taking into account the test sample.
8. CONTROL OF RESULTS
    The quality control is performed by placing, after the set of calibration solutions and
every five samples, a reference material whose content in arsenic is known with certainty.
    A control card is set up for each reference material used. The control limits were set
at: +/- 2SR intra (SR intra : standard deviation of reproductibility).
10. BIBLIOGRAPHY
   10.1. PESQUE M., 1982. Dosage de l’arsenic dans le vin. Rapport de stage. Diplôme
d’œnologue. Institut d’œnologie de Bordeaux.
    10.2. GAYE J., MEDINA B., 1998. Dosage de l’arsenic dans le vin par spectrométrie
d’absorption atomique. Feuillet Vert de l’O.I.V. n°1069.
   10.3. GAYE J., MEDINA B., 1999. Arsenic dans les vins. Feuillet Vert de l’O.I.V.
n°1087.
                                                                                       19
                 RESOLUTION OENO 18/2003
                         DETERMINATION OF CADMIUM BY
                        ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
                                                      volumes injected in µl
                                solution of Cd              blank               matrix modifier
                                   at 8 µg/l
            blank                      0                         10                   2
 calibration N° 1 at 8 µg / l          1                          9                   2
calibration N° 2 at 16 µg / l          2                          8                   2
calibration N° 3 at 24 µg / l          3                          7                   2
calibration N° 4 at 32 µg / l          4                          6                   2
Sample to be dosed 5 5 2
                                                                                            20
                 RESOLUTION OENO 18/2003
3. REAGENTS
        Demineralised water
        Pure nitric acid for analysis at 65%
        Anhydrous palladous chloride (59% in Pd)
        Magnesium nitrate with 6 water molecules (ultra pure)
        Ammonium dihydrogenophosphate
        Matrix modifier: palladous chloride and magnesium nitrate mixture (dissolve 0.25
g of PdCl2 and 0.1 g of Mg(NO3)2.6H2O in 50 ml of demineralised water) or ammonium
dihydrogenophosphate at 6% (dissolve 3 g of NH4H2PO4 in 50 ml of demineralised water).
Cadmium reference solution at 1 g/l, commercial or prepared as follows: dissolve 2.7444
g Cd(NO3)2.4H2O in a solution of HNO3 0.5 M, adjust to 1 l with HNO3 0.5 M.
Cadmium solution at 10 mg/l: place 1 ml of the reference solution in a 100 ml graduated
flask, add 5 ml of pure nitric acid and complete to volume with demineralised water.
Cadmium solution at 0.8 g/l: place 4 ml of the diluted solution in a 50 ml graduated
flask, add 2.5 ml of pure nitric acid and complete to volume with demineralised water.
Calibration range at 0, 8, 16, 24 and 32 µg/l of cadmium.
4. PREPARATION OF SAMPLES
5. PROCEDURE
      Each calibration solution is passed right after the blank solution. Perform 2
successive absorbance readings and establish the calibration curve.
      Calculate the cadmium content of the samples while taking into account the test
sample of different dilutions.
                                                                                           21
                 RESOLUTION OENO 18/2003
                         DETERMINATION OF CALCIUM BY
                       ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
     The calcium is directly determined in the liquid oenological product (or in the
mineralisation solution) suitably diluted by atomic absorption spectrometry by air-
acetylene flame after the addition of spectral buffer.
2. APPARATUS
          Instrumental parameters (given as an example)
          Atomic absorption spectrophotometer
          Reducing air-acetylene flame
          Hollow-cathode lamp (calcium)
          wave length: 422.7 nm
          width of slit: 0.2 nm
          intensity of the lamp: 5 mA
          No correction of non specific absorption.
3. REAGENTS
4. PREPARATION OF SAMPLES
                                                                                       22
                 RESOLUTION OENO 18/2003
5. PROCEDURE
        Pass each solution of the set in ascending order of the concentration of calcium.
        For each solution, perform 2 absorbance readings when they are perfectly
stabilised (integration time of signal: 10 seconds).
        Pass each sample twice and calculate the calcium content.
                                                                                            23
                RESOLUTION OENO 18/2003
                        DETERMINATION OF CHROME BY
                      ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
      The chrome is determined by atomic absorption spectrophotometer without flame.
2. APPARATUS
      2.1 Experimental parameters (given as an example)
            Atomic absorption spectrophotometer
            wave length: 357.9 nm
            hollow-cathode lamp (Chrome)
            width of slit: 0.2 nm
            intensity of the lamp: 7 mA
            correction of continuum by the Zeeman effect
            introduction in hot conditions of the samples in the graphite oven
            measurement of the signal: peak height
            time of measurement: 1 second
            number of measurements per sample: 2
            pyrolytic graphite tube:
            pyrolytic graphite oven containing a platform L’Vov tantalised
            tantalisation of platform (see above)
            inert gas: argon - hydrogen mixture (95%; 5%)
            parameters for oven:
                                                                                        24
                 RESOLUTION OENO 18/2003
                                                 volumes injected in µl
                               chrome solution         blank            matrix modifier
                                  at 50 µg/l
            blank                     0                  17                   3
 calibration N° 1 at 50 µg/l          5                  12                   3
calibration N° 2 at 100 µg/l          10                  7                   3
calibration N° 3 at 150 µg/l          15                  2                   3
sample to be measured 5 12 3
3. REAGENTS
       3.1 pure demineralised water for analysis
       3.2 pure nitric acid for analysis at 65%
       3.3 anhydrous palladous chloride (59% in Pd)
       3.4 pure hexahydrated magnesium nitrate for analysis
       3.5 ammonium dihydrogenophosphate
       3.6 matrix modifier: mixture of palladium chloride and magnesium nitrate
(dissolve 0.25 g of PdCl2 and 0.1 g of Mg(NO3)2.6H2O in 50 ml of demineralised water)
ammonium dihydrogenophosphate at 6% (dissolve 3 g of NH4H2PO4 in 50 ml of
demineralised water).
       3.7 reducing agent: L-ascorbic acid in solution at 1% m/v.
       3.8 chrome reference solution at 1 g/l, commercial or prepared as follows:
dissolve 7.6952 g of Cr(NO3)3.9H2O in a solution of HNO3 0.5 M, adjust at 1 l with HNO3
0.5 M
       3.9 chrome solution at 10 mg/l: place 1 ml of the reference solution in a 100
ml graduated flask, add 5 ml of nitric acid at 65% and complete to volume with
demineralised water.
       3.10 set of calibration solutions: 0, 50, 100 and 150 µg/l of chrome (see
table: adjustments of the automatic sampler).
4. PREPARATION OF SAMPLES
5. PROCEDURE
      Pass each solution of the set in ascending order of the concentration of chrome;
      Pass each sample twice and calculate the chrome content while taking into
account the test sample.
                                                                                          25
                 RESOLUTION OENO 18/2003
                         DETERMINATION OF COPPER BY
                       ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
3. APPARATUS
3. REAGENTS
                                                                                         26
               RESOLUTION OENO 18/2003
5. PROCEDURE
                                                                  27
                 RESOLUTION OENO 18/2003
                          DETERMINATION OF IRON BY
                       ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
      2.1 Instrumental parameters: (given as an example)
            atomic absorption spectrophotometry
            flame: oxidant air-acetylene
            hollow-cathode lamp (iron)
            wave length: 248.3 nm
            width of slit: 0.2 nm
            intensity of the lamp: 5 mA
            no correction of non specific absorption.
3. REAGENTS
4. PREPARATION OF SAMPLES
5. PROCEDURE
       Pass successively the calibration solutions and the blank which will be
demineralised water or a water-acid solution with concentrations used for samples of
solid oenological products mineralised by dry process and perhaps diluted.
                                                                                       28
               RESOLUTION OENO 18/2003
                        DETERMINATION OF NICKEL BY
                     ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
                                                                                    29
                   RESOLUTION OENO 18/2003
3. REAGENTS
        3.1 Pure demineralised water for analysis
        3.2 Pure nitric acid for analysis at 65%
        3.3 Anhydrous palladium chloride (59% in Pd)
        3.4 Pure hexahydrated magnesium nitrate for analysis
        3.5 Ammonium dihydrogenophosphate
        3.6 Matrix modifier: mixture of palladium chloride and magnesium nitrate
(dissolve 0.25 g of PdCl2 and 0.1 g of Mg(NO3)2.6H2O (3.4) in 50 ml of demineralised
water) ammonium dihydrogenophosphate at 6% (dissolve 3 g de NH4H2PO4 in 50 ml of
demineralised water), (3.1).
        3.7 L-ascorbic acid
        3.8 Analytical blank solution: L-ascorbic acid solution at 1% (m/v).
        3.9 Nickel reference solution at 1 g/l (1000 µg/ml) off the shelf or prepared
as follows: dissolve 4.9533 of Ni(NO3)2.6H2O in a solution of HNO3 0.5 M, adjust at 1 l
with HNO3 0.5 M.
4. PROCEDURE
       Nickel solution at 10 mg/l: place 1 ml of the reference solution (3.8) in a 100 ml
graduated flask, add 5 ml of nitric acid (3.2); complete to volume with demineralised
water.
       Nickel solution at 50 µg/l: place 1 ml of the nickel solution at 10 mg/l in a 200 ml
graduated flask, 10 ml of nitric acid (3.2) and complete with demineralised water.
       Set of calibration solution: 0, 50, 100 and 150 µg/l of nickel.
       The automatic distributor cycle enables to perform this calibration on the platform
from a nickel solution at 50 µg/l.
5. PREPARATION OF SAMPLES
6. DETERMINATIONS
                                                                                        30
                 RESOLUTION OENO 18/2003
                          DETERMINATION OF POTASSIUM
                      BY ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
       The potassium is determined by mineralisation by dry process by atomic
absorption spectrometry.
       The addition of a spectral buffer (cesium chloride) to avoid the ionisation of the
potassium is necessary.
2. APPARATUS
       2.1 Glassware
             100 and 200 ml graduated flasks (class A)
             1, 2, 4 and 10 ml graduated pipettes (class A)
             100 ml cylindrical vase
3. REAGENTS
                                                                                            31
                 RESOLUTION OENO 18/2003
4. PREPARATION OF SAMPLES
5. DETERMINATIONS
                                                                                         32
                   RESOLUTION OENO 18/2003
                          DETERMINATION OF LEAD BY
                       ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
                                                                                         33
                  RESOLUTION OENO 18/2003
                                                  volumes injected in µl
                            lead solution at 50            blank            matrix modifier
                                   µg / l
          blank                      0                      10                     2
    calibration N°   1               1                       9                     2
    calibration N°   2               2                       8                     2
    calibration N°   3               3                       7                     2
    calibration N°   4               4                       6                     2
    calibration N°   5               6                       4                     2
Sample to be measured 10 0 2
3. REAGENTS
      3.1 Pure demineralised water for analysis
      3.2 Pure nitric acid for analysis at 65%
      3.3 Ammonium dihydrogenophosphate
      3.4 Matrix modifier: ammonium dihydrogenophosphate at 6%.
Lead solution at 0.1 mg/l: place 1 ml of the lead solution at 10 mg/l in a 100 ml
graduated flask,
      add 1 ml of nitric acid at 65%; complete to volume with pure demineralised water
      for analysis.
Set of calibration solutions: 0, 50, 100, 150, 200, 300 µg/l of lead.
               The automatic distributor cycle allows to directly inject these quantities of
lead on the platform from the lead solution at 0.050 mg/l.
4. PREPARATION OF SAMPLES
       The liquid or solution samples must have concentrations between 0 and 300 µg/l
of lead.
       The solid samples will be mineralised by wet process (attack by nitric acid).
       The blank is made up of pure water for analysis containing 1% of nitric acid at
65%.
5. PROCEDURE
                                                                                           34
                  RESOLUTION OENO 18/2003
                         DETERMINATION OF SELENIUM BY
                        ATOMIC ABSORPTION SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
        2.1 Glassware
              Graduated flasks 50, 100 ml (class A)
              Graduated pipettes 1, 5 and 10 ml (class A)
              Polypropylene tubes 50 ml with screw top.
                                                                                       35
                 RESOLUTION OENO 18/2003
3. REAGENTS
4. PREPARATION OF SAMPLES
5. DETERMINATIONS
                                                                                       36
                 RESOLUTION OENO 18/2003
                         DETERMINATION OF SODIUM BY
                       ABSORPTION ATOMIC SPECTROMETRY
1. PRINCIPLE
2. APPARATUS
       2.1 Glassware
             Graduated flasks 50 and 100 ml (class A)
             Graduated pipettes 2.0; 5.0; 10.0 ml (class A)
             Automatic pipette 1000 µl
             Cylindrical vase 100 ml.
3. REAGENTS
              The calibration solutions prepared contain 1 g of cesium per litre; they are
              stored in polyethylene flasks.
                                                                                        37
               RESOLUTION OENO 18/2003
5. PREPARATION OF SAMPLES
5. DETERMINATIONS
                                                                                     38
                 RESOLUTION OENO 18/2003
                          DETERMINATION OF ZINC BY
                       ATOMIC ABSORPTION SPECTROMETRY
1 .PRINCIPLE
      The zinc is determined directly by atomic absorption spectrometry by flame.
2. APPARATUS ()
      Instrumental parameters: (given as an example)
            atomic absorption spectrometer
            oxidant air-acetylene flame
            wave length: 213.9 nm
            hollow-cathode lamp (zinc)
            width of slit: 0.5 nm
            intensity of the lamp: 3.5 mA
            correction of the non specific absorption with a deuterium lamp.
3. REAGENTS
        3.1 Pure demineralised water for analysis
        3.2 Pure nitric acid for analysis at 65%
        3.3 Zinc reference solution at 1 g/l commercial or prepared as follows:
dissolve 4.5497 g of Zn(NO3)2. 6H2O in a solution of HNO3 0.5 M, adjust at 1 l with HNO3
0.5 M.
        3.4 Zinc solution at 10 mg/l:
         place 1 ml of the zinc reference solution in a 100 ml graduated flask, 1 ml of
nitric acid (3.2) and complete to volume with pure demineralised water for analysis.
        3.5 Set of calibration solution: 0.2; 0.4; 0.6; 0.8; 1.0 mg/l: place successively
1, 2, 3, 4, 5 ml of the zinc solution at 10 mg/l in 5, 50 ml graduated flasks, complete to
volume with pure demineralised water for analysis.
4. PREPARATION OF SAMPLES
                The liquid or solution samples must have concentrations between 0 and 1
                mg/l of zinc.
                The solid samples are mineralised by dry process.
                The blank solution is made up of pure water for analysis containing 1% of
nitric acid at 65%.
5. PROCEDURE
      Pass successively the blank, the calibration solutions and the samples of
oenological products.
      The absorbency readings are performed for 10 seconds and the measurements
are duplicated.
      The concentrations of zinc in the samples are obtained from absorbency values.
                                                                                        39
                 RESOLUTION OENO 18/2003
1. PRINCIPLE
              The gases are controlled by chromatography in gaseous phase using a
"molecular sieve" type column and detection by catharometer or flame ionisation.
2. SAMPLING
      Either use
             - a stainless steel flask for sampling gas
             - a Teflon sampling bag for gas.
3. INJECTION METHOD
Wide bore
                                                                                    40
                  RESOLUTION OENO 18/2003
1. PRINCIPLE
            Mobile phase:
            solvent A: Deionised and microfiltered water
solvent B: acetonitrile
                                                                                    41
                 RESOLUTION OENO 18/2003
3. RESULTS
                                                                                      42
                 RESOLUTION OENO 18/2003
DETERMINATION OF 5-(HYDROXYMETHYL)FURFURAL
1. PRINCIPLE
3. PREPARATION OF SAMPLES
     The samples and the calibration solution HMF are injected after filtration on a 0.45
µm membrane.
4. PROCEDURE
       The chromatographic column is stabilised with the mobile phase for about 30 min.
       Calculate the concentration of HMF of the sample from the peak surfaces.
                                                                                          43
               RESOLUTION OENO 18/2003
GRAPE SUGAR:
1. PRINCIPLE
2.1 Chromatograph
      The chemicals used for the reference solution preparation are of "pure for
      analysis" quality.
      The composition of this solution is about 10 g/l for each sugar (fructose, glucose
         and saccharose).
      The reference solution is prepared every two weeks (maximum) and stored in
      the refrigerator in the 100 ml graduated flask used for the preparation.
                                                                                     44
                  RESOLUTION OENO 18/2003
SULPHURIC CINDERS
       The sulphuric cinders result from the calcination after being in contact with air
after being attacked by sulphuric acid.
        Heat a silica or platinum crucible of low form for 30 min until red; allow to cool in
a vacuum dessicator and tare the crucible. Place the exactly weighed test sample in the
crucible and wet it with a sufficient quantity of concentrated sulphuric acid (R) diluted
beforehand by an equal volume of water. Heat until dry evaporation, then in a muffle
oven, first carefully until red without exceeding the temperature of 600°C ± 25°C.
Maintain calcination until the black particles disappear, allow to cool, add 5 drops of
sulphuric acid diluted to half to the residue, then evaporate and calcinate as previously
until constant weight; weigh after cooling in the desiccator.
TOTAL CINDERS
The total cinders result from the calcination of the product after contact with air.
        Heat a silica or platinum crucible of low form for 30 min until red. Allow to cool in
a vacuum dessicator and tare the crucible. Dispose homogenously the exactly weighed
test sample in the crucible. Desiccate for an hour in the incubator at 100°C-105°C.
Incinerate in the muffle oven, first carefully to avoid that the sample catches fire, then
until red at a temperature of 600°C ± 25°C. Maintain the calcination until the black
particles disappear. For 30 min allow to cool in a vacuum desiccator. Weigh. Continue the
calcination until constant mass.
        If the black particles persist, take up the cinders in hot distilled water. Filter these
cinders on an ashless filter paper (porosity 10 µm). Incinerate the filter and residue until
constant mass. Group the new cinders with the filtrate. Evaporate the water. Incinerate
the residue until constant mass.
45