B&K 1014 Instruction Manual
B&K 1014 Instruction Manual
··~·
,. ~~
. '
; if<
.
~·. •
,._
•
:=:;.-~ ';:~- .·:e
.:r:. ·' • ~.-
.-
<J>-
• • w
··~.~~
<ill>~..... •
.•.
·. .. :r . . .
I
•
BHUEL&KJJEH
Nrerum, Denmark . Phone 80 05 00 . Telegrams: BRUKJA, Copenhagen
BB 1014
Beat Frequency Oscillator
Type 1014
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Group A. Electronic Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measurement of Frequency Response of Four-Terminal Networks .. 25
A.C. Bridge Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Measurements of Temperatures and Temperature Differences ...... 27
Measurement of Gain in A.F. Amplifiers .......................... 30
Group B. Acoustical Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Recording of the Frequency Response of Loudspeakers ............ 30
Recording of the Frequency Response of Microphones . . . . . . . . . . . . . 32
Recording the Frequency Characteristic of Hearing Aids and Ear-
phones ...................................................... 33
Checking of Hearing Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measurements on Air Filters, Carburettor Inlets etc. . . . . . . . . . . . . . . . . 36
Testing the Qualities of Airborne Sound Insulation . . . . . . . . . . . . . . . . 38
Measurement of Reverberation Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Absorption Qualities on Sound Insulation Material . . . . . . . . . . . . . . . . 44
Group C. Mechanical Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Strain Measurement on Vibrated Objects .......................... 45
The A.F. Response and Spectrum Recorder Type 3326 .................. 47
Description ...................................................... 47
The Audio Frequency Spectrometer .............................. 48
The Level Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Recording Paper ................................................ 50
Copying of Recorded Information ........ . .... .... ............... 51
Operation ...................................................... 52
General ......................................................... 52
Synchronization ................................................. 52
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Automatically Recording Harmonics ............................. . 54
Vibration Measurements ... .................................... . .. 57
Noise Measurements ............................................. 57
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Description
General.
The Beat Frequency Oscillator Type 1014 is primarily designed for electrical
and electro-acoustical measurements and consists of an oscillator- and mixer-
section , and an amplifier section.
It works on the heterodyne principle using two high frequency oscillators,
one uf which operates on a fixed frequency , while the frequency of the other
can be altered by a variable capacitor. The required audio frequency is then
obtained as the difference between the two high frequencies and can be read
off a large illuminated scale, the pointer of which is connected to the variable
eapacitor. The scale is logarithmic and graduated from 20 to 20ooo c/s. An
"Incremental Scale" is also provided , allowing exact frequency selection in
the range - 50 to +50 c/ s around any setting on the main scale.
The zero-adjustment is carried out by obtaining a beat between the ' frequency
of the mains voltage, and that of the oscillator voltage occurring'' when the
oscillator is tuned to the frequency of the mains and the pressbutton marked
" Power Frequency Beat" on the front panel of the oscillator is press~d.
The variable capacitor has two control knobs , one of which is in a fixed
position on the capacitor spindle and is used for quick setting to the de-
sired frequ en cy. The other will , when depressed , rotate the spindle with a
ratio of 1 to 5 giving greater accuracy and final selection of the frequency.
A worm gear permits the capacitor to be tuned automatically, for example,
with the aid of the motor of the Level Recorder Type 2305. The mechanical
connection to the Level Recorder is effected by means of a flexible shaft
which can be screwed onto the bushing on the side of the Oscillator's cabinet
during which the motor should be kept running. The worm gear can be
engaged and released with the aid of a built-in electromagnetic clutch,
operated from a switch on the front panel, or by a remote control arrange-
ment. The electromagnetic clutch is a friction-device allowing manual tuning
of the variable capacitor even when the clutch is engaged.
Being also designed for use in room-acoustical measurements, the Beat Fre-
quency Oscillator is equipped with frequency modulation, for which a re-
actance tube controlled by saw-tooth oscillations from a built-in oscillator is
switched into tbe circuit of the fixed oscillator. Both the frequency and the
amplitude of the saw-tooth oscillation are adjustable and may b~ read off
two printed dials. Provision is also inade for external modulation, whereby
very wide limits of frequency modulation can be obtained.
By means of a compressor circuit, which can be controlled from an externn.l
5
voltage, it is possible to keep the voltage, current, or sound pressure constant
during measurements when using the oscillator as a power source.
Impedance
Fixed Variable )J Low Pass Output Matching
Oscillator Amplifier Mixer Filter Amplifier Circuit
Load
Output
Regulating Variable
Amplifier Oscillator
100
°C(~ emote
L.~_Jfontrol
Ref. Sig.
Attenuator
Output
reactance tube circuit acts as a variable inductance and the modulation swing
can be continuously varied from 0 to ± 200 c/ s by means of a potentiometer
on the front panel of the apparatus, marked "Modulation Swing".
By means of the switch marked "Modulation Frequency" the frequency of
the built-in saw-tooth oscillator may be chosen. Frequencies of 1- 2- 4- 8- 16
and 32 c/ s are available. The oscillator is a blocking type, tuned to approx-
imately 7 Mc/s, and the frequency of the saw-tooth oscillations is set by
changing the grid resistor.
Provision is made for external modulation. The external generator should then
be connected to two terminals of the socket on the front plate marked
"Remote Control". For external modulation a voltage of approximately 7 volts
is necessary when a modulation swing of ± 200 c/s is required. The im-
pedance of the external generator must be low (approx. 1 kfl).
When external modulation is employed the switch marked "Modulation
Frequency" must not be in position " Mod. Off.", as in this position of the
switch the reactance tube is cut off.
A variable capacitor of 60 pF, inserted in the tuned circuit of the fixed
oscillator, and operated by the knob marked "Frequency Increment", permits
exact frequency selection in the range ± 50 c/ s for any setting on the main
scale.
6
By means of a noiseless switch on the front panel, marked "Oscillator Stop",
the voltage on the anode of the 120 kc/s oscillator can be disconnected. This
-301t---------
"OUTPUT VOLTAGE"
CONTROL
-1.0
- 501t--------
-60
-70
1
: -80
w
<.!)
~0
>
..,:9
:::>
a..
1-
:::>
0
-
100
o .1 .2 .3
----!0~~~~0
t. .5 .6 .7 B 9 1.0 1.1 1.2 1.3
VOLTAGE ON "COMPRESSOR" INPUT _ _ _ __.
1.1. 1.5 VOLTS
1605+7
("COMPRESSOR VOLTAGE" ON MAXIMUM)
7
innovation is specially provided for reverberation measurements. The same
method is used for remote control, the appropriate wiring of which can be
seen by referring to the circuit diagram of the Oscillator.
The output voltage from the fixed oscillator is fed to the grid circuit of a
pentode, tpe grid bias of which, is controlled, by means of a regulating
amplifier. To obtain a higher degree of control the working-point of the
pentode is chosen on the non-linear portion of the Ia-Eg characteristic, near
cut-off.
The purpose of this circuit is to control automatically the output power of
the Beat-Frequency Oscillator by an A.F. control voltage. For example, the
voltage from a standard microphone placed in the sound field of a loud-
speaker which is fed from the Oscillator. In this case the output power of
the Oscillator will be so controlled that a constant sound pressure is main-
tained on the standard microphone.
The A.F. control voltage should be fed onto the screened socket marked
" Compressor Input" on the front panel of the Oscillator. A variable potentio- ·
meter, marked " Compresso'r Voltage" is inserted in the input circuit of the
regulating amplifier and can be used as volume control for the output power
from the Oscillator when automatic regulation is employed. The regulating
amplifier has a linear frequency characteristic from 20 to 20ooo c/s and
should have an input signal of approximately 1 volt on the grid of the A.F.
amplifier tube for full regulation. The input impedance, measured across the
terminals of the socket marked " Compressor Input" is approximately 100
k.Q, and the maximum obtainable range of regulation is 45 db.
The amplified A.F. control voltage is rectified in a full-wave double-diode
rectifier, designed to give a DC output voltage proportional to the average
value of the A.F.-control voltage. ·
By means of the switch marked " Compressor Speed" on the front panel of
the oscillator the regulation speed can be varied. Regulation speeds of 30-
100-300 or 1ooo db/ sec. may be chosen by changing the value of the
capacitor in the A-C filtering network for the rectified control voltage. When
the switch "Compressor Speed" is in position "Comp. Off." the output from
the rectifier is short-circuited thereby disconnecting the automatic regulation
circuit.
To make regulation of the output level possible, even when maximum output
power is required from the Oscillator, the level of the high frequency voltage
from the 120 kc/ s fixed Oscillator is raised 10- 15 db when the automatic
compression ' is switched in.
The anode-circuit of the pentode in the variable-It amplifier is tuned to
120 kc/s, forming a band-pass filter, the output of which is fed to the mixer.
In the mixer tube, which is one half of a twin triode, the 120 kc/s voltage is
mixed with the output voltage from the variable oscillator. The frequency
of the variable oscillator can be altered continuously from 120 to 100 kc/s
by means of a specially designed variable capacitor. This capacitor is made
8
with a high degree of accuracy and a maximum deviation of o. 7 degrees
from a logarithmic frequency curve is obtained. A worm gear, connected to
the capacitor spindle, permits automatic tuning with the aid of an external
motor, for example the motor in the Level Recorder Type 2305, and the
worm gear can be set and released by means of a magnetic clutch. This is
operated from a switch on the front panel of the oscillator, or it can be
operated from an external switch or relay. Connection must then be made
to the appropriate terminals of the socket marked "Remote Control" on the
front panel, and the control switch for the magnetic clutch must be in
position " Clutch On".
By means of a pushbutton marked "1ooo c/s Ref. Signal" an extra capacitor
is introduced in the tuning circuit of the variable oscillator. This changes
the frequency of the oscillator so as to increase the output frequency an
amount of 980 c/s. With the scale pointer set to 20 c/s the output frequency
is thus 1ooo c/ s and may be used for adjustment purposes as explained on
page 19, item 17.
As ,previously mentioned, the frequency scale is logarithmic and calibrated
20-20ooo c/ s. When the capacitor is set to frequencies above 20ooo cis
or below 20 c/ s the fixed Oscillator is blocked, and consequently no output
voltage will be obtained. For automatic recording of frequency characteristics,
i.e. when using the Level Recorder Type 2305, this is a great advantage as no
unwanted curves will then appear on the corresponding section of the fre-
quency calibrated paper.
The voltage developed across the grid circuit of the variable capacitor is
fed onto the mixer tube and mixed with the voltage from the variable-.u
amplifier. The mixer tube is of the triode type, whereby a low hum level
is obtained in spite of the AC-heating of the filament.
A low-pass filter having a cut-off frequency of 50 kc/ s is inserted in the
anode circuit of the mixer tube, thus passing only the lower frequency
obtained by the frequency conversion, onto the grid circuit of the first tube
in the output amplifier section.
Frequency Marking.
An arrangement, which can be operated by the control switch for the
magnetic clutch, is the marking of a certain frequency. At this frequency
the fixed Oscillator is short-circuited, while at all other frequencies the
normal output voltage is available. Basically, the blocking is effected by
closing a pair of relay contacts. These contacts are operated by a cam disc
which is mounted concentrically on the tuning capacitor spindle. Normally
the cam rotates with the spindle, being held in position by a spring-loaded
friction clutch. However, by means of a pawl the cam disc can be located at
the point of its cycle where the short circuiting relay is closed.
This is carried out by setting the control switch for the magnetic clutch to
position "Marking Adjustment", and turning the capacitor spindle until a
9
small "clickl' is heard. Because of the friction clutch, it is now possible to
turn the capacitor and the scale pointer relative to, and independent of the
cam. Release of the pawl, by means of the control switch, permits the cam
to rotate once again with the capacitor.
Frequency Marking
10
writing pen. This is a great asset in for example measurements where the
compressor circuit of the B.F.O. is used. In this instance the pen-lifting
arrangement of the Level Recorder can be con~rolled from the frequency
blocking circuit by making the appropriate connections to the "Remote
Control" jack of the B.F.O. In cases where the entire frequency range
(20-20ooo c/s) of the B.F.O. is utilized, the normal frequency blocking,
which functions outside the scale graduation, should be set out of pperation.
The writing pen of the Level Recorder can now be lifted from the paper
outside the frequency range of interest and a proper working of the som-
pressor also at the initial frequency (20 c/s) is ensured during the automatic
scan. If the described method is not utilized, the following would take place:
No signal will be present in the range 20ooo c/s to 20 c/s (outside the scale
graduation), i.e. the compressor of the B.F.O. will be in such a condition to
give full output signal of the B.F.O. Consequently, when the scale pointer
goes inside the scale graduation (20 c/ s) full output level will be transmitted
at 20 c/s, and fi~st after the chosen time delay (Compressor Speed) the signal
level will be compressed to the proper (preset) value. A deflection on the
recording paper which is not a response of the measured object would thus
be recorded.
11
10
Ofb
0.5
c 0.2
~
:0 0.1
i
D
z
0.05 - - - = Typical Distortion Curves
------ • Guaranteed Upper Limit
' 0.02
~ 1 ~1o------------~1o~o-------------1+oo-o----------~1o~o~oo~ f C/s
---Frequency ~6~072
Fig. 4. Distortion curves for different loads. The curve marked "Att. 10 volts"
is obtained from measurements taken on the "Attenuator" output terminals:
open circuit.
the output impedance being constant and approximately 50 ohms. The overall
accuracy of the attenuator is approximately 2 %.
The voltage on the output terminals is indicated by a vacuum-tube voltmeter
which measures the average value of the A.F. voltage. It is calibrated in
r.m.s. values of a sinusoidal voltage, and the accuracy in the frequency
range 20-20ooo c/s is 1.5 % of full scale deflection.
The sensitivity of the voltmeter is automatically changed when the position
of the switch marked "Matching Impedance" is altered. Full deflection of
the meter is indicated on the switch. When the "Matching Impedance" switch
is in position "Att." the output voltage available from the Oscillator will
depend on the position of the "Attenuator" switch, in this case full
deflection of the meter corresponds to the value indicated by the switch
position.
The signal-to-noise ratio of the Oscillator is greater than 70 db for maximum
output voltage. It is independent of the position of the attenuator, but some-
what dependent on the position of the potentiometer marked "Output
Voltage". The best result is obtained when this is positioned on or around
the point marked 8.
Harmonic distortion is dependent on the setting of the "Output Voltage
Potentiometer". The distortion increases as the output voltage is increased,
but as long as the output is kept within the meter range, the distortion will
be of the order indicated in Fig. 4.
12
Power Supply.
The Oscillator can be operated from a 240, 220, 150, 127, 115 or 100 volts AC
power line, the power consumption being about 70 watts .
The proper voltage is selectable by a switch-fuse combination situated at
the rear of the instrument. To select the voltage it is necessary to firstly
remove the fuse by turning the hexagonal disc head in the centre of the
switch anti-clockwise . Then with the aid of a screwdriver turn the head of
the voltage adjuster until the white mark is aligned with the required voltage.
The fuse is then replaced.
It should be noted that if the apparatus is to be operated from a DC power
line, or from an accumulator, a vibrator unit or a rotary converter is
required.
13
Operation
General.' ·
Firstly ascertain that the Beat Frequency Oscillator is set to the appropriate
power supply voltage by means of the selector at the rear of the instrument
and that the Remote Control plug on the front panel is firmly in its place.
After connection to the power supply, the instrument can be switched on
by the toggle switch marked " Power" on the front panel. The dial lights in
the meter and in the frequency scale should immediately come on.
Attenuator
Output
Magnetic Load
a~ch ~--------~~~...
Corrtrol
Compressor
Input
Output
Modulation
Frequency
1 000 r;s Ref. Signal
160590
Modulation Compressor Frequency
Control Swing Speed Scale
Adjustment
14
A. Calibration.
1. Turn the toggle switch marked "Power" to " On" and allow two minutes
warm up.
2. Set "Modulation Frequency" and "Compressor Speed" to their "Off"
position.
3. Set "Clutch, Marking" to "Marking Adjustment" position. Rotate the main
scale until a small click is heard and extra weight is felt on the drive.
Continue to rotate until the pointer is set within uncalibrated portion of
scale, i.e. between 20ooo c/s and 20 c/s. Then return clutch switch to
"Off".
4. Now turn main scale pointer until it is on the frequency of the line
voltage (e.g. 50 or 60 c/s), checking that the frequency incremental
scale is on zero. If not, set by "Frequency Increment" knob to this
point.
5. Set suitable deflection on the meter by tuning the knob marked "Output
Voltage" to higher than center scale reading.
6. Press "Pow~r Frequency Beat" button and hold to "in" position and at
the same time rotate "Frequency Scale Adjustment Fine" slowly, until a
large fluctuation registers, slows up, and practically ceases on the meter
dial. Two points may be found where this occurs, only one of which is
correct and therefore a check as outlined in the following paragraph
should be carried out, firstly releasing the "Power Frequency Beat"
button.
7. Turn the main scale pointer to 20 c/ s and with the "Frequency Increment"
knob reduce scale reading to -20 c/s or -30 cis depending upon supply
frequency in use. If the frequency is correct the meter needle will drop to
zero indicating that the B.F.O. is properly tuned. If it does not reach zero,
repeat procedure from item 4.
8. Finally return "Frequency Increment" to zero. The B.F.O. is ready
for use.
N.B. If a zero point cannot be found and is outside the range of the '"Fre-
quency Scale Adjustment Fine", re-align the variable capacitor marked
)<Coarse" with a screwdriver to give a suitable ~etting, which should occur at
some point between 4 and 6 on "Frequency Scale Adjustment Fine".
15
4. Turn the pointer on the main frequen 1y dial to the desired frequency,
finely adjusting the Frequency Incren1 ent if necessary. (For automatic
frequency sweep, see under E).
5. Select a suitable output voltage by h rning the knob marked "Output
Voltage".
D. Frequency Modulation.
When a frequency modulated output ~ :gnal is required, the following proce-
dure should be adopted:-
1. Turn the knob marked " Modula :ion Frequency" to the required fre-
quency.
2. Turn the knob marked "Modulati .m Swing" to zero.
3. Re-calibrate the Oscillator as described in A.
4. Set the "Modulation Swing" knob to the required frequency swing
(bandwidth).
5. Proceed as described in B items 2 to 5, or C items 2 to 5, dependent on
the requirement.
E. Automatic Recording.
By combining B.F.O. Type 1014 and Level Recorder Type 2305, or using
Automatic Frequency Response Recorder Type 3304, it is possible to auto-
matically record the frequency response of four terminal networks. When
using B.F.O. Type 1014 and Level Recorder 2305, it is necessary to connect
the two instruments mecha:Qically by a Flexible Shaft UB 0040 as in Fig. 6
and to make the electrical connec tions also shown. Fig. 7 depicts the use
of the Automatic Frequency Response Recorder Type 3304 with the required
external connections.
For setting-up, calibrating and synchronising the combination shown in
Fig. 6 the following procedure should be adopted : -
1. Ensure power supplies are correct and switch powe~ toggles to the
" On" position.
16
1014
Object
under !60519
test
3304
Object
under
!60SSO
test
Evant Marking
Potentiometer Range
db
Range Potentiometer
Drive Shaft n
Drive Shaft I
5. Load the Level Recorder with the desired recording paper. (Follow
instructions in Level Recorder Manual).
6. Select and insert required Range Potentiometer. (N.B. Place "Potentio-
meter Range db" switch to "Standby" when changing potentiometers).
7. Switch " Potentiometer Range db" until figure corresponds to the Range
Potentiometer being used, i.e. " 10", "25", "50" or 80".
8. By means of the switch " Rectifier Response", select R.M.S. or if
specially required one of the other three positions Average, Peak, or D.C.
9. Turn the "Lower Limiting Frequency" switch to the cut-off value (10,
20, 50 ox: 200 c/s) .
10. Set "Writing Speed" to required position.
(Full ex]>lanations of items 8, 9 and 10 can be obtained from the Level
Recorder · Manual).
18
11. Place "Reverse/Forward" switch to "Forward".
12. Select "Paper Speed" to a suitable speed, e.g. 10 mm/sec.
13. Pull gear-lever marked "x" to the outer position. (See Fig. 36).
The actual paper drive speed now corresponds to the small numQers
marked around the "Paper Speed" knob.
14. Two types of recording can be made:-
(a) Single chart recording (automatic recording over a length of 250 mm
paper only),
(b) Continuous recording over any length of paper.
19
taking a recording of frequency characteristics, where the lowest
attenuation is around 1ooo c/s, that the deflection of the stylus lies
within the scalar limits of the paper during the recording).
F. Marking Adjustment.
Should an identity mark at some particular frequency be desired the follow -
ing procedure should be used:-
1. Set the magnetic clutch control switch to the position "Marking Adjust-
ment".
2. Turn the pointer of the main frequency dial until a small "click" is
heard, the click being associated with an increased effort in turning the
pointer.
3. Set the pointer to the frequency at which marking is desired.
4. Release the magnetic clutch control from the Marking Adjustment posi-
tion, reverting to clutch "Off" or "On" as required.
The instrument is now operable as in B, C, D or E the output voltage
~isappearing at the "marked" frequency.
20
plate, pass it carefully between the contact spring at "A" and the
insulated bushing assembly "D".
6. Likewise remove the outer insulated cam ring "E", which will now
be loose, in the same manner as the metal retaining plate. The re-
maining insulated cam ring should not be removed.
7. Install the cam ring to be fitted, making sure that the white spot
on the surface near the slot on the perifery faces outwards towards
the operator.
8. Replace the metal retaining plate, and the three screws together with
the three spiral springs, making sure that the screws are properly
tightened .
9. Replace the steel ball.
10. Replace the perforated back plate to the cabinet.
21
3. Set pointer to:
a. The frequency at which the blocking should be cancelled, if
blocking of the initial scale part is wanted .
b. Two decades higher, approximately, (will be outside the scale
graduation), at which the blocking should commence, if blocking
of the last part of the scale is wanted.
Check for proper commence of blocking. If necessary readjust
as in item b.
4. Release "Clutch" control from position "Marking Adjustment", re -
verting to position "Off" or "On" as required.
The B.F.O. is now operable as in B, C, D or E.
22
Rep:~.ote
Control.
In the main description of the apparatus several forms of remote control are
mentioned. To carry out any one of these methods use must be made of the
"Remote Control" jack on the front panel, the appropriate connections being
made to the pins of the six-poled socket. Fig. 10 shows the different pins
on the socket.
Oscillator Stop
( +100 Volts d. c. approx.)
Internal
d c
Marking Contact
e b External Modulation
f a
Ground Magnetic Clutch
MIO•' .
Fig. 10. ''Remote Control'' jack, viewed looking towards front panel.
Trouble Shooting.
If th~ B. F. 0. is not ~orking properly when switched on, check the follow
• ing:-
1. · That 6-poled plug for the "Remote Control" jack is in position.
2. That scale-pointer is not situated in the uncalibrated section of the main
dial, i.e., between 20ooo c/s and 20 c/s.
3. That scale-pointer is not on the "marked" frequency or on a section
chosen for "partial blocking of frequency range", see pages 9 and 10.
Accessories
Output Transformer TU 0005.
This transformer is designed to allow symmetrical output from the altenu-
ator out'pqt of the B.F.O. 1014. (Symmetry better than o.1 %). The output
impedance' is 600 Q and the distortion o.5 % at 20 c/s with maximum output
voltage from the B.F.O. (12.6 V). The accuracy of the Transformer is ± o.2
db in the frequency range 10 c/s to 35 kc/s. In addition a core material has
been chosen for the transformer, which makes it possible to "preload" the
secondary winding with a current of 100 rnA without causing additional
distortion for frequencies above 300 c/s. The transformer ratio is ylo : 1.
24'
Applications
The field for use of the Beat Frequency Oscillator Type 1014 is so extensive
that only a few of the possible applications are illustrated in the following
pages, these being classified into three sections, showing the instrument being
used as a power source for:-
(A) Electronic Measurements
(B) Acoustical Measurements
(C) Mechanical Measurements.
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Brilol & Kl-
25
If an automatic recording of the frequency response is wanted, the Level
Recorder Type 2305 should be used. The mechanical coupling between the
motor in tqe Level Recorder and the tuning capacitor of the B.F.O. is
effected with a Flexible Shaft UB 0040 which is delivered with the B.F.O.
The measuring arrangement which is employed to obtain the frequency
characteristic of an A.F.-filter is shown in Fig. 11.
Should the compressor circuit be used to regulate the output signal from
the Oscillator it is advisable to verify that the voltage at the "Compressor
Input" is approximately the required 1 volt. When it is intended to use the
equipment for automatic recording of frequency characteristics, the input
of the Level Recorder may first be connected to the input of the compressor,
and a recording of the compressor input voltage made for the complete fre-
quency raf\ge in which measurements are to be taken. With the compressor
working correctly the resultant recording should be a straight line. If this
is the case the input to the L~vel Recorder can then be disconnected, and the
desired measurements carried out.
Fig. 12. Tile B.F.O. Type 1014 used as voltage source for AC Bridge Measure-
m ents . Th e Output Transformer TU 0005 provides a symmetrical output from
the B.F.O.
26
former "arms" is better than o.1 %. The voltage transmission loss of the
transformer when loaded by 600 [J is approximately 16 db and the distortion
less than 1 % for a transformer input voltage of approximately 12 V.
Due to the selectivity of the Frequency Analyzer it is well-suited as an
indicating instrument in a bridge circuit. The decibel scale on the instru-
ment meter will often prove useful when it is desired to measure the quality
of different components placed within the bridge.
27
2. The current has to be determined by measuring the voltage drop across
a fixed resistor which is connected in series with the respective termistor
(see sketch 14a). The obtained voltages are fed to the two input terminals
of the above mentioned selector in the Level Recorder section of the
Type 3304.
Heat
®
Source
3. Prior to the final measurements the set-up must be calibrated, this being
done by placing the thermistors in a bath of warm oil the temperature
of which has been taken with an ordinary thermometer. During the
cooling period the voltage drop across each resistor is taken at requisite
intervals and plotted on the side of the paper to be used.
As the ther~istors have a negative temperature co-efficient a decrease
in temperature will cause the voltage drop across the fixed resistor to
fall as can be seen on the calibrated curve in Fig. 14b.
Fig. 14c shows the recordings of a "simultaneous" measurement taken
f
with the arrangement as in 14a. The wall having firstly been warmed on
one side by a constant heat source to the desired temperature, then re-
28
~ D 0 D 0 D 0 0 D D D D D 0[
lllllllllll
~
~on 50 Ln 30 20
h
_l
'1
)
""n I
"L..
t;..,
I)...,
n-.
lh_
lll-'lr.
II-...,_
II th.
II I fTh- h.ltlln. on
..... &. .I.&. lllllllllh
.... .._...... l
·~··
~--
..
_
• AU.::IU m V I I I"----'
0 D D D 0 D D n 0 0 0 D 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Fig. 14 (c) Si m ultan eous rec ord ing of th e temperature on both sides of
th e wall .
29
moving the source whereby the temperature gradually falls on the heated
surface but still rises on the other surface due to the "inertion" of the
heat transmission through the structure, see Fig. 14c.
~ --:.- •
-o- •
i
·:· :·
""'"i= JL .;~-~ .:§_
Amplifier -"7i:=
1014 2804
t60~57
The output voltage from the amplifier under test should be measured with an
Audio Frequency Voltmeter Type 2409 (or 2410), or a Microphone Amplifier
Type 2603 (or Type 2604) an example of the arrangement being given in
Fig. 15.
30
Absorbing material
~0
.
I@-
: e::
:: :.
. '\;;P ~"""'!!!~-----__.
2803 3304
.... ........
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
.... ..... "'"'
31
The loudspeaker under test should be fed with a constant voltage or current,
the latter producing a mechanical force of constant amplitude which is
applied to the diaphragm.
Fig. 16a shows a set-up for recording the frequency characteristic of a
loudspeaker. The loudspeaker is fed from the B.F.O. section of the Auto-
matic Frequency Response Recorder Type 3304 via a series resistor the
voltage drop across which is led to the Compressor Input of the B.F.O. A
constant current will therefore be obtained in the circuit when the voltage
across the resistor is approximately 1 volt.
The output voltage from the Condenser Microphone Type 4131 is fed to the
Level Recorder in the Automatic Frequency Response Recorder via an
Amplifier. This amplifier can be a Microphone Amplifier Type 2603 (or 2604)
or a Frequency Analyzer Type 2107 or 2112. The Amplifier should be
switched to have a linear frequency characteristic.
An advantage gained by employing a Frequency Analyzer as an amplifier is
that distortion measurements can be carried out with the same measuring
set-up. Fig. 16b shows a recording obtained with the described set-up.
4131
i•
•.
.,
f6H60
A
Fig. 17a. Measuring set-up for automatic recording of the frequency response
of microphones.
32
In the set-up depicted, the microph~nc to be tested is connected to the Level
Recorder Type 2305, via a Microphone Amplifier Type 2604, the originating
sound source being a loudspeaker which is fed from the B. F. 0. Type 1014.
As the sound pressure in front of the microphone under test has to be kept
constant, it is necessary to place it relatively close to another microphone
(in this case a Condenser Microphone Type 4131) which is coupled to a
second Microphone Amplifier Type 2604, the output of which is fed to the
"Compressor Input" of the BFO ensuring a constant sound source. It is
essential that the two microphones are symmetrically placed in th ~ radiated
sound field and the correct compressor speed selected. If the acoustical delay
time required for the sound to travel from the loudspeaker to the micro-
phone is r, this period must be small in comparison to the time constant
determining the compressor speed. Under normal circumstances these con-
ditions are easily fulfilled. l
.........
DD D DDOIJDIJDIJDOOO OO OOODOODO DDO D DD DD D DDIJDDDODDD
..c
-~
To give reliable measurements the room to be used need not be fully anechoic
as the regulating effect of the compressor will compensate for any minor
reflections set-up . However, for correct operation of the regulation circuit,
the reverberation time of the room must not be too long and a low scanning
speed should be used for the frequency sweep.
In Fig. 17b will be seen a recording showing the frequency response of a
microphone recorded by employing the previously outlined system.
33
The B.F.O. 1014 feeds the earphone under test which is placed in the
Artificial Ear Type 4151. The ear consists basically of a base mounted on
a board and a replaceable acoustical coupler. Different types of couplers are
available. A DB 0138 2 cm3 which conforms to ASA Z 24. 9. 1949 and the new
IEC standards is suitable for measurements on insert type of earphones.
For headsets and similar external earphones a 6 cm3 can be supplied e.g.
DB 0160 (N . B. S. type) or DB 0161 (A. S. A. type).
(a)
+
-.- : ·e:
·;·~~
.•;,;-,·!- . -~~~~·:·
21.03 1014
(b)
-
DDDDDDDDDDDDDDDDDOODDDOODDDDDDOODOODOODDDDDD
Fig.18.
(a) Measuring arrangement.
(b) Recording of the frequency characteristic of an earphone.
34
Checking of Hearing Aids.
An arrangement for the checking of Hearing Aids is illustrated in Fig. 19a.
This set-up makes it possible to automatically record the frequency character-
istic of a complete hearing aid, under what are approximately free field
conditions.
l61t767
Fig.19.
(a) Arrangement for automatically checking the frequency characteristic of a
hearing aid.
The hearing aid earphone under examination is placed on the ear of the
Hearing Aid Test Box Type 4212, which consists of an external artificial ear,
a regulating microphone, a built-in loudspeaker, the latter two of which are
enclosed in a small anechoic chamber. The chamber is effectively insulated
against both airborne and impact noise, allowing measurements to be taken
down to 50 db re 2 X 10-4 .ubar approximately.
35
The hearing aid and the regulating microphone are placed symmetrically in
the sound field. The regulating microphone is connected to the Microphone
Amplifier Type 2603, which amplifies the signal and then applies it to the
Compressor· input of the iF.O. Type 1014. This combination enables the
sound pressure level on the hearing aid to be kept constant without in-
fluencing the practically free sound field conditions.
The B. F. 0. Type 1014 supplies the required power for the loudspeaker in
the chamber, while a B & K Condenser Microphone, which is placed in the
Artificial Ear, is used for tpe measurement of the acoustic output from the
hearing aid. The microphone is connected to a Microphone Amplifier Type
2603, and the amplified voltage is led to the input of a Level Recorder Type
2305.
IJ o
1
o o o o o o o o o : o o o o o o ~ o o. o o o o o o o o o o o o o o o o o o o o o o o o o o o o o o
BrD~I•r +.502:5
-----...
~h~~~cr;r,stjc
of hearmg ajd
.... max . amplification
reduced amplification
Fig. 19. (b) Recording by 19 (a). Taken for different settings of the hearing
aid volume control.
36
Recorder Type 3304) is shown in Fig. 20, the item under test being an air
filter. Fig. 21 shows the resultant curves which give an indication of the
attenuation of sound waves in the air filter. Curve A is measured with the
Sound Pressure Source 4211 open and Curve B with the filter mounted on
the Source opening. The sound pressure level at the air filter input in both
cases being 110 db with reference to 2 X 10-4 ,ubar, and to avoid directional
influences the measurements were carried out in a highly reflective hard-
walled chamber.
In many instances there will be a requirement to find the most suitable
position to mount absorbent material in order to silence the .object on test.
When this situation arises measurements should be carried out under "free
~.-
if ~ ~-~:.
,....... ~···
@
~ .!;
Fig. 20. Arrangement for determining the sound attenuation in an air filter.
-
OD D DDDD DD DDO OOODDDDDDDDDDDDDDDDDDOOOOOOOOOOOOOOOOOO
!rO~jatr .
5025
z-L..; _rop.Sp.: _
LIJ..ft; _Lc.No.: _
''"-
w•. s,..: _
-""'"- -"'
-
s.,: _ _
""' --~s.....• - oo~
37
receiver room transmitter room
4131 4131
L.
2812 2812
Fig.22.
(a) Measuring arrangement for automatic reading of the sound insulation
properties of a wall.
38
necessary sound is generated by the BFO 1014 section of the Type 3304
in conjunction with a loudspeaker. The Beat Frequency Oscillator should be
frequency modulated and the loudspeaker (or loudspeakers) placed so that a
sound field, as diffuse and isotropic as possible, is built up.
"'·--""" -_
w.s,..:_s..... _ -
...,, _ _ F,..Scol.o _
1000 XIOOO c!-«10004
("11/2'rii) .A. I
QP 0123 10 100 1000 10000
~6(0717
39
circuit and the possibility of frequency modulation , which makes it very
suitable for this type of measurement. The compressor circuit serves to
keep the sound radiated in the room at a constant value throughout the
frequency range of the measurements. Frequency modulation of the signal
mdiated in the room ensures that a great number of eigentones of the room
are excited in the frequency band covered by the frequency modulated
signal. The resuJtant recorded decay curves will in this manner appear with
a smooth slope. That would not be the case when a pure sine-wave signal
is radiated in the room, as distinct standing waves would arise. The fre-
quency modulation is easily adjusted in frequency swing and modulation
frequency by the controls "Modulation Swing" and "Modulation Frequency"
respectively.
Various measuring arrangements for reverberation measurements can be set
tip where the B.F.O. is an integrating part. Here will be discussed an arrange-
ment which wotks automatically and where the m easured decays are re-
corded by the B & K Level Recorder Type 2305. The set-up is illustrated in
Fig. 23. The B.F.O . Type 1014 and the loudspeaker constitute the trans -
mitting part, whereas one of the B & K Condenser Microphones, the Audio
Frequency Spectrometer Type 2112 and the Level Recorder Type 2305 make
Frontal Connections
2112 2305 1014
84 A2
Ground Ground
(b)
Fig. 24. Connection between instruments.
(a) Connections between remote control jacks.
(b) Electrical circuit of the remote arrangement.
40
up the recetvmg part. As the amplifier for the Microphone is chosen the
Spectrometer whioh makes a selective reception in 1h or Yt octave bands
possible, thereby reducing the influence of the room's background noise.
A sufficient dynamic range is in this manner obt.a ined when measurements
are carried out in rooms where the background noise cannot be removed.
The measuring arrangement shown allows decay curves of the reverberation
to be recorded automatically throughout the frequency range 25 c/s to 20ooo
cis with intervals of 1h octave. All the decays are registered on ' a frequency
calibrated part of recording paper being only 250 mm in length (refer Fig.
26). If greater spacing between the individual decay curves is required, the
recording has to be made on non-frequency calibrated paper. Below is given
a brief d·e scl'iption of the principal working of the two types of m:e asurements.
Frequency Calibrated Paper. For recording the decay of the sound in the
room the sound source has to be disconnected at definite intervals, this is
achieved by stopping the oscillator in the B.F.O. To ensure that only the
part of the measurement is recorded · which is of interest, the writing pen
should lift from the paper in the interval between two decays, and as
selective reception is utilized, the filters in the Spectrometer should be
switched in successively. The disconnecting of the sound source, the lifting
of the pen and the switching of the filters in the Spectrometer can all be
automatically controlled by a special switch in the Level Recorder. (The
Two-Channel Selector). The necessary connections between the different
instruments are shown in Fig. 24. The connections to the respective Remote
Control Jacks are shown in Fig. 24a, while F~g. 24b gives the electrical
circuits for the remote controlling arrangement. ,
Overlapping junction.
4:1
When placing a loop of 50 mm paper width (Fig. 25) in the Level Recorder
with a length of 49·5 mm (i.e. two chart lengths minus 5 mm; 5 mm being
the distance between two perforated holes) it is possible to have the curves
for the different frequencies placed with a spacing of 1h octave as shown
in Fig. 26. By cohesively synchronizing the paper movement with the fre-
-
OOODDDOODOOOODOODDOODDDDODDDDDDDDODDDOODODODOOOODDD
BrO~Jc:.r.
5025 ~ro.~•q_ .W6r., lro.lloq.. lS
f6f85-+
quency scanning of the B. F. 0., with the filter switching on the Spectrometer
and with the switching off moment of the sound, the starting points of the
decay curves will correspond to the center frequency of the respective filters,
represented bY'' small squares on the top of the preprint of the recording
paper QP 0123, QP 0223 and QP 0323, see Fig. 26. It is possible, lo a certain
degree, to keep the sound pr:e ssure level at the point of measurement
independent of loudspeaker and room response by utilizing the compressor
circuit of the B. F. 0. as indicated in Fig. 23. This method ensures that all
the decay curves commence at the same level on the recording paper.
42
10 mm spacing between the curves. In such instances the recoiding has to be
carried out on the lined recording paper, e.g. QP 0102, QP 0202 or QP 0203,
and it is necessary to " mark" one or more frequencies on the paper. The
marking can be readily done by means of the Level Recorder 's "Event-
Marker" arrangement.
If only a few reverberation curves are to be taken, the situation may not
warrant the use of automatic measuring, in the se circumstances use should
be made of the pressbutton marked "Oscillator Stop" on the B. F. 0. Also,
when it is desired to record the decay curves with a spacing less than 1/:{
octave, the described function of the automatically working arrangement
cannot be used immediately. The manually or remotely operated "Oscillator
Stop" may then be utilized.
16f8G6
43
2. The reverberation time in seconds is then read on the scale at the point
through which the horizontal line passes. Vide Fig. 28.
The decay curves should preferably be approximated into a straight line
making it easier to determine -the average slope.
If paper speeds other than 10 and 30 mm/ sec. have been used, the determined
reverberation times should be multiplied or divided by factors of 10.
Example.
50 db Range Potentiometer.
Paper Speed 100 mm/sec.: Use the section "50 db 10 mm/sec." and divide the measured
result by 10, see also Fig. 28.
161962
400'1
45
the Accelerometer's sensitivity curve. From this the driving force can be
calculated, using Newton's equation F = m X a, where F = the driving force,
m = the mass, and a = the acceleration.
To measure the mechanic~! strain in the object under test a Strain Gage
JJ 2612 2613
""i = ~ .;::·~· g
~ :.'C."!..
@
2103
oooooooooooooooooooooooooo~
50
db
---~-+
20
10
0
50 100 200 C/s 300 ~0
46
is used. This is a pick-up device which is comprised of a looped resistance
(or resistances) sandwiched between insulating material which is cut in the
form of a strip, and which can be glued on to the test object. The object
when subjected to mechanical strain is distorted and this will alter the Gage
resistance, the variation being register ed by a sensitive measuring bridge
arrangement, e.g. the Strain Gage Apparatus Type 1516. (For further · informa-
tion refer to manual).
The output voltage from the Strain Gage Apparatus is then fed to the input
of the Level Recorder of the Type 3304, to give an automatic recording.
An example of such a recording, taken on a thin metal bar, showing the
mechanical strain and indicating its resonant frequency, is shown in Fig. 30b.
Fig. 31. Photograph of the A.F. Response and Spectrum Recorder Type 3326.
47
2305, all being housed in one unit. The complete assembly gives a compact
means of carrying out and automatically recording almost any desired
electrical, electro-acoustical, or acoustical measurement in the audio fre -
quency range.
Fig. 31 shows a photograph of the equipment. For a comprehensive technical
description of the Spectrometer _and the Level Recorder, reference should
be made to their respective manuals. However, a brief outline of their bask
principles is given in the following paragraphs.
48
supplied with a large amount of negative feedback. The metering circuit
connected to the output amplifier can be switched to measure half the peak
to peak, average or RMS value of the input signal.
The linear frequency range covered is 2 c/s to 45ooo c/s and the three
weighting networks conform to the IEC-proposed standards for precision
sound level measurem ents.
IS9J99
49
only necessary to release a snap-lock arrangement on the moving arm
which holds the stylus.
The writing speed is determined by the amount of damping applied to the
writing system which is selectable by the rotary switch "Writing Speed".
This is variable in 15 spot settings shown by the figures outside the switch.
The large figures denote the settings for 50 mm p'lper while the small
figures are for recordings on paper of 100 mm width.
Recording Paper.
Different types of pre-printed recording paper to be used on the Level
Recorder in conjunction with the Beat Frequency Oscillator Type 1014 and
Spectrometer Type 2112 combination are available, the paper being manu-
factured with a printed logarithmic frequency scale covering the range from
10 c/s to 40ooo c/s.
Three main types of paper, having various features, and which come in
two widths, can be supplied. Also included in the range is polar diagram
recording paper to be used when the Level Recorder is required for this
application (see Level Recorder Manual). Two types of writing are catered
for, either ink or stylus, the applicable surfaces being treated accordingly.
\Vhile paper for ink writing is available, with pre-printed line~ or outlines
of frequency diagrams and is obtainable in the widths of 100 mm or 50 mm.
The polar diagram type comes in packs of 100 sheets and has a 100 mm rarlius.
The waxed paper comes in two types but only in 50 mm widths, i.e. white
waxed, or red waxed, and are intended for use with a sapphire stylus. The
white waxed paper consists of coated black paper and exhibits printed lines
or frequency di~grams as the case may be. The red waxed type is comprised
of transparent pa:per with a thin covering of red wax, having printed lines
or frequency diagrams superimposed on it. Waxed paper is particular useful
50
when high writing speeds have to be used, giV~ng a very clear definition
of the recording. When using these papers the sapphire stylus will leave
a thin black line on the white waxed paper and a transpar(}nt line on the
red waxed paper, the latter being specially made to enable ,blue prints to
be taken. It is necessary to double-copy the latter, as the scales being
printed in black will not show up if a direct blue print is take~ from it.
51
OPERATION
General.
The instruments contained in the A. F. Respon se and Spectrum Recorder
can be used separately, or together, in various combinations. The logarithmic
frequency scale of the B. F. 0 . 1014 allows it to be completely synchronized
with the Spectrometer.
Synchronization.
To fully synchronize the three units the following sequence of operation
is recommended. Switch on the power of each instrument and connect the
" Recorder" terminal of the Spectrometer to the "Input" terminal of the
Level Recorder, then calibrate as follows:-
BFO 1014 Section.
Calibrate the BFO as in " A. Calibration" under Operation on page 15.
S pee trome ter.
1. "Input Switch" to " Direct".
2. "Meter Range" to "Ref.".
3. "Meter Switch" to " Fast" "RMS".
4. "Range Multiplier" to "x 1, 0 db" .
5. " Function Selector" to "Linear, 2- 40ooo c/ s".
6. "Automatic Switching" to " Off " .
Other knobs in any position.
Meter should show a deflection to the red mark on the scale. If
necessary any deviation can be recorded by means of the potentiometer
marked. " Sensitivity Amplifier Input" situated on the front panel.
Adjust with screwdriver.
Level Re~order.
(N.B. In this case a 50 db potentiometer is used. For other ranges refer
manual for Level Recorder Type 2305).
Set control knobs to the following positions:-
1. "Potentiometer Range db" to "50".
2. "Rectifier Response" to "RMS".
3. " Lower Limiting Frequency " to " 20".
4. " Writing Speed" :
50 mm paper: 500 mmj sec. (large figures ).
100 mm paper: 1000 mmjsec. (small figures ).
5. "Paper Drive" to "Stop" and " Forward" positions.
6. Motor to "On".
7. Set "Input Attenuator" to " 10" .
8. Using the " Input Potentiometer" adjust stylus to full deflection
- 4 db (e. g. using · 50 db Range it will be 50 db- 4 db = 46 db).
9. Insert the desired type of frequency calibrated paper. (If necessary
refer Level Recorder Manual).
52
10. Pull the Synchronizing Gear Lever (1 : 10) marked "X" in Fig. 36
to outer position.
11. With a screwdriver turn the screw " S" on Fig. 36 until marking
cut is in vertical position.
12. Set the "Paper Speed" to 10 mm/sec. (small figures). The spring
loaded knob is operated by pulling, turning and dropping to correct
position.
p1 p1
z 5
z
X 2305
Drive
Shafts
13. The toggle switch "Paper Drive" is set to "Start" whereby the
paper should start moving. If not, press the pushbutton "Single
Chart - Continuous Recording" and release it again, the paper will
move and after a chart length or less automatically stop.
14. Move the recording paper by means of the finger wheel "Z" shown
in Fig. 36 until stylus rests on 10 c/s line.
At this stage final synchronization of the units takes place by firstly syn-
chronizing the Spectrometer and Level Recorder sections as follows:-
53
4. As a means to see how far the paper has to be shifted, it is re-
commended to draw a line, by means of the "100 mV Ref." button
on the front plate of the Recorder, at the point where the paper has
stopped. By using this line as a reference the paper can be shifted
the appropriate distance to give correct synchronization.
5. To check the synchronization, run the recording until the pen is
stopped on for example the 2ooo cjs line. When correctly syn-
chronized, the switching of the 800 cjs to the 1ooo cjs filter should
now take place at the 900 cjs line. If this is not the. case repeat
from item 2.
6. Finally reset the Writing Speed on 250 mm or lower if the large
figures are being used, and to a figure of 500 mm or less if using
the small figures.
To complete the synchronization sequence the BFO must be synchronized
with the frequency calibration on the paper.
1. Rotate main scale of BFO manually until it corresponds to the
frequency denoted by the stylus, firstly moving it to a higher fre-
qen'cy and then rotating it back until it arrives on the desired
frequency. This will take up any possible backlash with the gears
of the BFO.
2. Move "Clutch Switching" of the BFO to the "On" position.
All units should now be in complete synchronization and on
switching the toggle switch of the "Paper Drive" on the Recorder
to "Start", the combination will operate in complete unison.
APPLICATIONS
Applications for the combined B. F. 0. 1014 and Level Recorder 2305 have
been described, and illustrated, under this heading in the earlier phase of
this manual. Now one important use for the treble combination, B. F. 0. +
Level Recorder + Spectrometer (i.e. the A.F. Response and Spectrum Re-
corder Type 3326) will be described, followed by an other example using
the Spectrometer and Level Recorder with the exclusion of the B. F. 0. 1014
section.
54
Fig. 37 (a) Measuring arrangement to measure harmonic distortion in
hearing aids.
55
In the case where the harmohics of a four-terminal network are to be
automatically recorded, the synchronization of the units should have been
completed and the Single Chart recording method should be utilized .
The harmonics are measured by setting the filter switch on the Spectro-
meter so that it runs ahead of the frequency scanning of the B. F. 0 ., the
selected frequency difference being in accordance with the harmonic which
is going to be measured.
By means of the Reverse/Forward switch the recording is returned by the
si:Rgle chart length already run ·o ff; and the process repeated until all the
required harmonics are recorded on the same chart as the fundamental.
The dips seen on the curve are due to the shape of the Spectrometer filter,
and the manner of scanning the frequency range . When properly syn-
chronized the "open " dips will occur with a depth that corresponds to
around 3 db.
During measurements the sensitivity of the Spectrometer or the Level
Recorder should not be altered. If, however, circumstances require an
alteration to be made e. g. to obtain a clear recording of some of the
0 -
~ D D DODDDDOOODODDDDDD~DDDOODDDDODD DO DDDDDDDDODDDODOO
BrU:!!,!Iatr . . . . . !10 too.~•q. Joa.I6Q. M.l~ll- \0 1S
25
~~' --
Measuring of
....
fundamental .. "'
and harmonics
1n heormg a1ds ,. ,
Fundamental
2nd harm
z-L...'.~IOS
~....u.. F.__so- - 3rd harm . ·.
w•. .,._ 2o_o__
.... .,._ 10_ _
.........,.Fotq.Seol.loy•_l_o o
Vibration Measurements.
By using one of the Accelerometer Sets Type 4308/09/10/11 with a Vibration
Pick-up Preamplifier Type 1606, or one of the Microphone Cathode
56
Followers, vibrations in buildings, machinery, ships, etc., can be measured.
In Fig. 38a a layout is shown, where the Type 3326 is being used to auto-
matically record vibration measurements taken on the base of an electrical
motor, the Spectrometer allowing an analysis of the vibrations to be made .
The output from the Accelerometer is fed via the Preamplifier 1606 to the
"Condenser Microphone" input of the Spectrometer. Information on the
calibration of the arrangement is given in the manual for the Type 1606.
@•
• :i::
·i·~!
.x:;;p@
~·:;;'i:!- · ·~~~··
Noise Measurements.
Noise within factories , offices, cities, airports , etc. varies with the time of
day. Therefore it is a firm requirement that recordings of the noise should
be carried out over a considerable period of time as a single recording
taken at a particular instant would provide incorrect inform,ation.
Fig. 39 shows a recording taken of the noise level in a m~~hanical work-
shop over a selected period of time by using one of the Weighting Networks
57
~ oooooooooo:ooooooooooooooooooooo o ooooooooooooooooo
lrD:!!!Iaor • .50~ ~.o.~•r;. , ~.G.~•KI- ~oo.~•K~-
...
10 75
- " ' -- -
~e~~~~f~~gtn .. .,
on a electncg[
motor frame
~~~
.,~ BE!;l_
'-" RM5
z-w...~ 1;; S
~: ·"~
:~~ 2go
Mu~~~pt,F,.Scoleby~_j_ o 010 30000 c/1 .0000 A I C lill.
(1611/ 2111lA I C IJA. ·
UP 1123 10 100 1000 10000
Fig. 38b . Recording obtained with measuring set-up as shown in Fig. 38a.
in the Spectrometer. As can be seen the noise level varies considerably with
time and a reliable anai.ysis cannot be obtained by only taking a few
measurements. It is therefore necessary, when doing such a test, to record
continuously and then to statistically assess the final result.
To measure, analyse and record such noise the Spectrometer + Level
Recorder (i.e. the receiving part of the A. F. Response and Spectrum
Recorder Type 3326) in combination with a B & K Condenser Microphone
can be used.
o o o o o o o o o o o o o o o o o o o o o o o o octo o o o o o o o o o o o o o o o o o o o o o o o o
8rD:!!!,Ic.t . . s.o.~•r.., .w•IQw ~.o.~•r..,.
Fig. 39. Recording of the noise level in a mechanical workshop over a period
of time.
58
Specification
Frequency Range: 20-20ooo c/s.
Frequency Scales:
Main Scale: Logarithmic from 20-20ooo c/s.
Tolerance ± o. 7 degrees of theoretical logarithmic curve. Vernie1
driven.
Increment Scale: Range -50 to +50 c/s of main scale reading. Both
scales illuminated.
Frequency Accuracy:
Main scale: 1 % ± 1 c/s.
Increment scale: ± o.5 c/s.
Outputs:
Matching: Switchable matching impedance for 6, 60, 600 or 6ooo ohms
load.
Max. power output 2.5 watts approx.
Attenuator: Variable in steps of 10 db (within ± o.2 db} from 125 !f-V to
12.5 V. Continuously variable by potentiometer within each step.
Distortion:
Frequency in c/s ....... 20 200 2ooo 20ooo
"Attenuator" terminal.
No load with 10 V
output approx . . .. ..
.. . 1.0% o.1 ·% o.1 o/o 0.7%
"Load" terminal.
(Loaded 1 watt) ....... . 2.0 •% 0.3% 0.3% 1.2 •%
59
frequency range 50- 20ooo c/s. Input impedance 100 kohms. Regulation
speed variable in steps: 30- 100-300 and 1ooo db/sec.
Centimetres 50 40 20
I I
Inches 20 16 8
I I
Weight
I 22 kg
I
49lbs.
60
B K