Manual DuraScan G5 English
Manual DuraScan G5 English
DuraScan 50 G5 / 70 G5 / 80 G5
Contents
1 Introduction 1-1
1.1 Preface 1-1
1.2 Safety instructions 1-1
1.3 Manufacturer's notes 1-2
1.4 Icons and typographic conventions 1-4
1.5 Technical data 1-5
2 Initial startup 2-1
2.1 Unpacking the machine 2-1
2.2 Inspect the delivery 2-1
2.3 Weight and dimensions 2-4
2.4 Transporting the machine 2-4
2.5 Installing the machine 2-5
2.6 Starting up the machine 2-7
2.6.1 Mounting the necessary accessories 2-11
2.6.1.1 Indenter and lenses 2-11
3 Design and functions 3-1
3.1 Design of the tester 3-1
4 Basic operation 4-1
4.1 Switching the Machine ON and OFF 4-1
4.2 Graphical interface 4-2
4.3 Control elements 4-4
4.3.1 Positioning in the Z-axis 4-5
4.3.1.1 Slider 4-5
4.3.1.2 AF Camera 4-5
4.3.1.3 AF Indentation 4-5
4.3.2 Positioning in the X-Y-axis 4-6
4.3.2.1 Test anvil and specimen holder 4-6
4.3.2.2 Joystick 4-6
4.3.2.3 Movement in the live image 4-6
4.3.3 Turret 4-6
4.4 Buttons 4-6
4.5 T-Cursor 4-12
4.6 Editing tables 4-13
4.7 Selection window 4-14
4.8 Crosshairs 4-14
4.9 Colour-coding 4-14
4.10 Ring light 4-15
5 Examples of ecos Workflow 5-1
5.1 Single measurement 5-1
1-1
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
1-2
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
1-3
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
1-4
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
1 Introduction
1.1 Preface
Our hardness testing machines represent state-of-the-art technology and comply with
current norms and standards. Both the hardware and software in our machines can be
upgraded on an ongoing basis. As a result, you can ensure that the technology in your
machine is always up-to-date.
The machines are equipped with a measuring system and electronic power control.
They are also distinguished by their ergonomic design and user-friendly software, which
complies with DIN standards.
Take time to read this instruction manual closely. You will be amazed by the many
application options available.
Operating instructions Read all operating instructions before you start the machine.
Keep the operating instructions in the vicinity of the machine. Order a new copy if you
lose the manual.
Moving the machine To avoid injuries, never move the machine on your own.
Transport safety Remove all transport safety devices before you start the machine.
devices Never use the machine in a potentially explosive environment.
Fire and explosion ▪ Position the machine on a secure and stable working surface.
protection ▪ Ensure that its working height is in accordance with ergonomic principles and does
Installation require any anatomically questionable postures.
▪ Ensure that there is sufficient space available to access the machine and that an
adequate escape route is kept clear to provide for unforeseen incidents.
▪ Ensure that the work station has adequate lighting.
▪ The hardness testing machine must be protected against shocks and vibrations that
can influence the test result.
Working clothes and Ensure that working clothes are close-fitting and cannot be entangled in the moving
protective equipment parts of the machine. Make sure you use your personal protective equipment.
Electrical power Only trained and qualified persons are allowed to connect the electrical power supply.
connection
1-1
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
Work time Do not work with the machine for too long or when you are unable to concentrate
properly.
Protection from The machine must not come into contact with water. Protect the machine from
splashing splashing.
Startup Before each startup, ensure that the machine is in a faultless and maintained condition,
and that all safety devices are installed and are functional. Damaged cables must be
replaced immediately by authorized personnel.
Operation by Ensure that the machine is only operated by authorized and properly trained persons.
authorized personnel Ensure that the machine cannot be started or operated by unauthorized persons.
Single-person machine Ensure that the machine is only operated by one person at any one time.
Unauthorized Do not make any unauthorized modifications to the safety equipment of the machine.
modifications Bypassing control devices and any interference with the electrical/electronic
components of the machine are prohibited and will nullify the warranty in all cases.
Emergency OFF In an emergency, press the EMERGENCY OFF button to stop the unit at once.
Securing workpieces Before testing, check that the measuring tools are securely clamped.
Do NOT attempt to touch the internal parts of the machine while it is running.
Use appropriate fixings to secure any workpieces that do not stay in place without
support.
Never try to hold a workpiece with your hand!
Wear protective gloves when handling heavy or sharp workpieces or removing chips.
Resetting the machine Prior to resetting the machine, switch it off at the main switch.
Supervision Never leave the machine unattended while it is operating. Always switch the machine off
before you leave the work station.
Work station Keep your work station clean and tidy. A cluttered work station increases the risk of
injury.
Maintenance Use the main switch to turn off the machine. Remove the power plug to disconnect the
machine from the power supply before you carry out any maintenance or readjustment.
Any work on the electrical system should only be carried out by skilled electricians or
service technicians.
Make sure you comply with the inspection and maintenance intervals specified (see
standard EN ISO 6507).
Claims Contact your supplier in the event of collision or damage. Always specify the unit
number and software version of your machine in cases of complaint or damage or when
making inquiries or ordering replacement parts.
1-2
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
Ambient temperature The machine is adjusted for the ambient temperature specified in the technical data.
A stable autofocus function can only be guaranteed if there are no major temperature
fluctuations in the installation area.
Unauthorized uses of Do not use the machine to test workpieces of a low mechanical stability, that can be
the machine distorted or break when they are clamped.
Workpieces that are stable but do not have the required stability may only be clamped if
they are secured in an appropriate fixture.
Do not use the machine for workpieces which are toxic or harmful to health.
EC conformity The CE mark and the EC conformity declaration certify that the machine and this
manual comply with the stipulations of the directives that apply to the product.
Disposal The WEEE symbol on your unit shows that it is a WEEE-relevant machine that must
be disposed of separately. Further information about recycling this product can be
obtained from the local authorities.
The warranty period for new machines is 24 months from the date of delivery by the
Warranty conditions for manufacturer with unlimited operating hours.
new machines If a defect is detected, inform your sales partner or nearest service centre of the
manufacturer immediately, providing a detailed description of the defect in written form,
over the phone or in person.
Defects that are properly reported and are covered by the manufacturer's warranty will
be corrected free of charge either by repair or replacement delivery. If requested,
defective parts are to be returned to the manufacturer at the customer's expense and
risk.
The manufacturer's warranty does not apply to defects that are caused by one or more
of the following:
▪ Failure to fully comply with the operating instructions, safety and licensing
regulations or other instructions relating to the delivery, installation, startup or use of
the machine
▪ Incorrect assembly or startup of the machine
▪ Unauthorized tampering with with or modifications to the machine by the customer
or a third party
▪ Improper or inappropriate use of the machine for purposes other than those for
which it is intended
1-3
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
DANGER
The safety instruction DANGER stands for an extremely dangerous situation.
Nonobservance will lead to severe and even fatal injuries (burns, for example).
Observe the danger warning, and take greatest care (using protective equipment, for
example) to face the danger. Persons must work very carefully in this area.
WARNING
The safety instruction WARNING stands for a very dangerous situation.
Nonobservance can lead to severe, irreversible and even fatal injuries, and loss of
limbs.
Observe the danger warning, and take greatest care (using protective equipment, for
example) to face the danger.
CAUTION
The safety instruction CAUTION stands for a dangerous situation. Nonobservance
can lead to injuries or material damage.
Observe the danger warning, and take greatest care (using protective equipment, for
example) to face the danger.
NOTE
The safety instruction NOTE refers to situations that can lead to material damage.
They show how damage can be avoided.
1-4
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
Icons The following icons and typographic conventions are used in this instruction manual:
Crushing hazard
indicates a risk of crushing, threatening life and limb of persons.
Electric current
indicates a risk of electric current, threatening life and limb of persons.
Danger
indicates a risk that threatens life and limb of persons.
Note
indicates a risk of damage to property, or the need to proceed with special care.
Information
indicates additional information and tips
Ambient conditions
Room temperature (to ISO/ASTM) [°C] 23 (±5)°C
1-5
Hardness Tester Introduction
DuraScan 50 G5 / 70 G5 / 80 G5
1-6
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
2 Initial startup
2.1 Unpacking the machine
Unpacking the machine
Remove the plastic and paper packaging.
1 Machine stand
2 Test unit
3 Accessory case
4 Transportation safety device
5 Motorized cross slide
The serial number can be found on the rear of the tester.
2-1
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
1 Machine stand
2 Test unit
3 Monitor, keyboard and mouse
4 Motorized cross slide
5 Turret with indenter and lenses
(indenter and lenses not yet fitted on delivery.)
6 Sample holders (optional)
7 EMERGENCY OFF button
The DuraScan is controlled using a PC and operated with the ecos Workflow Software.
You can either order the PC along with the DuraScan or use a different PC.
Please contact your service partner as the PC must comply with the manufacturer's
specifications.
The following description only applies to the PC ordered along with the machine.
2-2
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Accessories
1 Accessory case
2 Patch cable
3 USB cable
4 USB cable (option)
5 Allen key 2.0 mm
6 Allen key 1.3 mm
7 Allen key 1.5 mm
8 Fork spanner
9 Two replacement fuses (T6,3 A slow-blowing)
10 Ring bolt
11 Cover
12 CD-ROM 01 with machine data, documentation, and conformity test
CD-ROM 02 with machine-specific data and certificate
13 Cable 3-wire US
14 Cable 3-wire EU
Standard accessories
▪ Indenter
▪ Objective lenses
If you ordered an indenter, it is included in the delivery and must be fitted as well as any
lenses supplied.
Optional accessories
▪ Calibrated test blocks
▪ Additional indenter
▪ Commercially available USB printer
▪ Sample holders
To order additional accessories and replacement parts, contact your supplier. Only
released products have been approved for use with this machine.
2-3
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Dimensions DS50
Dimensions DS50
2-4
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Secure the cover (2) over the hole for the ring bolt (1).
The cover is supplied with a Velcro fastener.
2-5
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
2-6
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Other settings that are possible in the ecos Workflow software can be found in Chapter
"Configuration and calibration", page 8-1.
2-7
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Plug an Ethernet cable into the Ethernet port (RJ45) and plug the opposite end into
the Ethernet port (RJ45) on the PC.
Plug the USB cable into the USB port on the machine and plug the opposite end
into one of the USB ports on the PC.
Repeat this step if the machine has two USB interfaces.
Starting up the PC If you wish to order your own PC, please contact your service partner as the PC must
comply with the manufacturer's specifications.
The PC is delivered in the manufacturer's original packaging together with an instruction
manual. The operating system is already installed. Use the operating instructions to
familiarize yourself with the operation of the PC.
Consult the instruction manual to find out how the PC needs to be set up.
Connect monitor, keyboard and mouse.
Follow the instructions in Chapter "Starting up the machine", page 2-7 to connect
the DuraScan to the PC.
Switch on the PC. Follow the instructions for your operating system in the
documentation.
Insert the CD-ROM 01 into the DVD drive and install the ecos Workflow software.
Insert CD-ROM 02 into the DVD drive: File Loader installs the machine-related
data.
After installation, enter the following IP address in the Windows Network
environment window: 10.10.150.2.
2-8
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
This completes installation, and you can now start the ecos Workflow.
Accept the preconfigured user name "Cal" and select Login without entering a
password.
2-9
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
If you are unable to log in without a password, contact your administrator or Service
team to ask for your login data.
The Configuration wizard screen opens. The first tab General is active. You can use
this during startup to first select the language and unit of measure.
2-10
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
2-11
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Hold the indenter in position (2) with the flat wrench until you have secured the
indenter with the Allen key (1).
Set the lens in the right position on the turret and secure.
Repeat the process for other lenses as necessary.
The lens adjusting procedure can be found in Chapter "Adjusting lenses (Assistant -
Objective Lenses)", page 8-17.
2-12
Hardness Tester Initial startup
DuraScan 50 G5 / 70 G5 / 80 G5
Note on calibration
The unit has already been calibrated and adjusted by the manufacturer or dealer. After
the transport and after indenter and lenses have been installed, some new settings have
to be made with the "Configuration wizard". Calibration should only be carried out by
trained technicians. Further information can be found at "Configuration and calibration",
page 8-1.
Select Exit to close the Configuration wizard and to start the ecos Workflow.
After a short time, the following screen appears:
2-13
Hardness Tester Design and functions
DuraScan 50 G5 / 70 G5 / 80 G5
1 Machine stand
2 Test unit
3 Monitor, keyboard and mouse
4 Motorized cross slide
5 Turret with indenter and lenses
6 Sample holders
7 EMERGENCY OFF button
The essential components of the hardness tester are the machine stand, the test unit
plus turret including the indenter/lenses, and the test anvil.
Operation is via the monitor, keyboard and mouse, which are connected to a standard
PC (with Windows operating system). The ecos Workflow software is installed on this
PC.
The machine stand is made of cast aluminium. The stand contains all electrical and
electronic components of the unit, and all drive units.
The test unit is powered by a DC motor. The test unit can move up and down at a speed
of 2 mm/s. It is capable of measuring workpieces with a maximum height of 260 mm.
The machine is equipped with a motorized 6-fold turret.
The test anvil is designed as a motorized cross slide.
Table of lenses The optical Vickers and Knoop test methods require lenses to show the indentation.
The following rule applies: The lower the test force, the greater the optical magnification.
A table with guide values for the range of application of the individual lenses can be
found at www.emcotest.com.
3-1
Hardness Tester Design and functions
DuraScan 50 G5 / 70 G5 / 80 G5
Safety equipment An EMERGENCY OFF button sits on the front of the machine as a safety device. When
you press this button, the current operation is cancelled and the machine is switched off.
The test unit stops moving and the screen is dimmed.
Rotate the EMERGENCY OFF button clockwise to release it.
WARNING
In an emergency, press the EMERGENCY OFF button to stop the unit immediately.
Never modify the safety equipment of the unit without authorization.
3-2
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
4 Basic operation
4.1 Switching the Machine ON and OFF
Switching the tester on Turn the main switch to the ON position.
The green indicator lamp in the main switch lights up.
If the machine does not start up:
Rotate the EMERGENCY OFF button clockwise to release it.
The machine is initialized and the nose cone released after approximately 30 seconds.
Switch on the PC.
You cannot start the ecos Workflow before the machine has been initialized and the
nose cone released.
4-1
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Switching the tester off Select File > Exit to close theecos Workflow.
Shut down the PC.
If you do not plan to use the unit for some time (> 1 week), switch off the machine too:
Turn the main switch to OFF.
The green indicator lamp in the main switch lights goes out.
The following figure shows the basic elements of the graphical interface.
4-2
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
1 Menu bar
2 Workflow bar
3 Work area
4 Controls
5 Tool bar
6 Specimen area
7 Tool bar (at the side)
8 Status panel
4-3
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Whenever making positioning movements, always check that the test unit is positioned
high enough above the specimen holder to prevent a collision between lenses/indenter
and specimen holder.
Should a lens interfere with the specimen holder, you may have to recalibrate the lens.
See Chapter "Adjusting lenses (Assistant - Objective Lenses)", page 8-17.
4-4
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
4.3.1.1 Slider
The slider can be used to move the test unit up and down. This allows you to:
▪ focus the image.
▪ raise or lower the indenter.
With the bar in the middle, the test unit is moved continuously up or down.
The two arrows at the top and bottom end of the slider move the test unit up or down in
steps.
When rapid traverse (rabbit icon) is active, the test unit moves at a 10 times higher
speed. Manual focussing is not possible when rapid traverse is active. Rapid traverse
can only be activated when the indenter is swivelled in. It is deactivated automatically
when the turret position is changed.
Use the slider to move the optical system to the specimen or surface area. Next, use the
AF Camera function to optimize the image.
4.3.1.2 AF Camera
The AF Camera function is used when the lens is already in the specimen or surface
area.
In other words, use the slider to move the lens to the specimen or the surface before
you select the AF Camera function.
The AF Camera function automatically creates an optimum image without any further
user activities:
▪ The image is focused.
▪ Brightness is optimized.
Wherever possible, select the AF Camera function to ensure measured values that are
objectively comparable.
The perception of the focus by the human eye varies widely, which can result in user-
related differences in the test evaluation.
4.3.1.3 AF Indentation
The AF Indentation function is used for automatically locating and identifying the
surface of a specimen.
The indenter is automatically moved to the workpiece, where it leaves a little indentation
with a test load of approx. 10 g and a diagonal of 10 - 15 µ. Next, the indenter is raised
slightly.
The lens with the lowest magnification is swivelled in and automatically adjusted for an
optimum image.
The optimum working height is then approximately 0.3 mm above tje workpiece.
4-5
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
4.3.2.2 Joystick
The joystick permits a manual movement on the X-Y-axis. Operation of the joystick (in 8
directions) results in a continuous movement of the image.
4.3.3 Turret
The turret is equipped with lenses and indenters. Click on the relevant symbol to swivel
in an indenter or lens. The tool which has currently been swivelled in is marked with a
red arrow.
4.4 Buttons
General symbols The following buttons can be found in the different screens.
Specimen area The following buttons can be found in the Specimen area.
4-6
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
4-7
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Select Test Points Selects the entire series or individual test points
independently of the function "Switch View T-Cursor
/ Test Points".
"T-Cursor" is selected: The entire row is selected
and can be moved and rotated.
"Test Points" is selected: Individual test points are
selected and can be moved.
Draw Guide Lines Draws a guide line to facilitate the positioning of test
points / series.
4-8
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Switch over View Switches the view to the T-Cursor or the individual
T-Cursor/Test test points
Points
Move Start Point Moves the start point of the row to the position
to this Position marked by the crosshairs
Set Core Hardness Adds a new core hardness point with Nht
Point measurement
The system switches to the "Test Points" view.
4-9
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
4-10
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Test Point Editor The following buttons can be found in the Test Point Editor.
Set Core Hardness Adds a new core hardness point with Nht
Point measurement
The system switches to the "Test Points" view.
Draw Guide Lines Draws a guide line to facilitate the positioning of test
points / series.
4-11
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Save Test Point Saves the test point pattern created in a file
Pattern
Load Test Point Loads a saved test point pattern from a file
Pattern
4.5 T-Cursor
The T-Cursor is used to position and align your row to the workpiece.
There are two options for positioning.
4-12
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
delete lines You can delete lines in tables in the Result screen and in the Test Point Editor onthe
Coordinates tab. This is no longer possible in the History screen.
In the table, click on the line you wish to delete and use the right mouse button to
open the context menu.
Select Delete or press the Del key.
The selected line is deleted immediately.
4-13
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Showing at series level In a table with CHD, Nht and Rht measurements, you can decide in the Result and
History screens whether you want to show all values or only the measurement series.
Click on the table and use the right mouse button to open the context menu.
Select Show CHD-, Nht-, Rht-Results at Series Level.
The table view changes.
In the work area, click on the button in the Test type box.
The selection list opens.
Select the required setting (here: CHD) from the selection list.
The selection list is closed. The selected setting is displayed in the box.
In the second document, selection via the selection window is described in the following
instruction:
In the work area select Test Type > CHD.
4.8 Crosshairs
In the ecos Workflow, you can only execute a command when the green crosshairs are
visible.
4.9 Colour-coding
Once all mandatory boxes have been completed for a specific work step, the specimen
in the specimen area takes on the colour of the related section in the Workflow. For
example: Pink for Method or blue for Position.
4-14
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Installing the ring light To install the ring light, first remove the lens ring and replace it with the ring light
adapter.
Remove the lens (if installed).
Attach the ring light lens adapter (1) and tighten it with the screws.
Install the required lens. See "Installing the lens".
If necessary: Use an adapter ring.
Slip the aluminium slide-on attachment (ring light) (2) on the lens such that the long
side face of the slide-on attachment is parallel to the engraved number on the turret.
Tighten the stud screw.
4-15
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Measuring with the ring To take a measurement with the ring light, you must additionally activate the ring light
light via an optional button on the Method screen in the software. Dark field illumination is
activated automatically. The evaluation algorithm is inverted and the image evaluation
program searches for a bright test indentation on a dark background. To return to using
the standard evaluation algorithm, simply use the same button to deactivate the ring
light. The ring light can remain attached to the lens in this case.
Adjusting the ring light If a test indentation does not have optimum illumination, the distance between the ring
light and the test indentation must be adjusted. To do this, slide the aluminium slide-on
attachment onto the lens in the correct position, and fasten it with the stud screw.
4-16
Hardness Tester Basic operation
DuraScan 50 G5 / 70 G5 / 80 G5
Loosen the stud screw on the ring light and change its position on the lens.
Fasten the attachment with the stud screw.
The position of the ring light does not need to be changed in this case.
Adjusting the Once the ring light is in the correct position, brightness must be optimized. You can
brightness adjust brightness on the Position screen.
Switch to the Position screen.
Select the Ring light menu item to display the brightness control setting for the ring
light.
Use the brightness control setting to optimize brightness.
4-17
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Login After the unit has been switched on, the ecos Workflow starts with the login screen.
5-1
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The Method screen opens.
5-2
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The Position screen opens.
5-3
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
5-4
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In certain cases, the indentation left by the AF Indentation function is not desirable. In
such a case, manually lower the indenter with the slider until it is close to the workpiece.
5-5
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
WARNING
Crushing hazard
Keep your hands away from the area around the moving test unit.
5-6
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The History screen opens.
5-7
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Login After the unit has been switched on, the ecos Workflow starts with the login screen.
5-8
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
5-9
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The Method screen opens.
5.2.2 Creating measurement series, selecting measurement type, test method and lens
Perform the following operations in the Method screen:
▪ Select two measurement series.
▪ Select test procedure, test method and lens for both rows created.
5-10
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
5-11
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The Position screen opens.
5-12
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
▪ Align the indenter and workpiece in relation to each other, i.e. adjust to the optimum
working height.
▪ Specify two measurement series using the Test Point Editor.
▪ Position the start point for each measurement series on the workpiece.
▪ Start measurement.
The overview camera is used for rough positioning, the evaluation camera for fine
positioning.
Adjusting to the First, align the indenter and workpiece in relation to each other.
optimum working If necessary: In the work area, select the Evaluation Camera tab.
height Put the workpiece under the lens on the test anvil.
5-13
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Setting test points In the Specimen area, select the Row tab.
The Test Point Editor opens.
5-14
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the tool bar, select the button to set the test points.
Set three test points by clicking on the Coordinates box.
In the tool bar, select the button to set the test points.
Set three test points by clicking on the Coordinates box.
Coarse positioning In the work area, select the Overview Camera tab.
5-15
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Position the workpiece such that the required start point of the measurement series
is located in the zero point of the crosshairs.
5-16
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the tool bar, select the button to position the start point of the series at the
zero point of the crosshairs.
The T-Cursor moves to this position.
You can select the "T-Cursor" view by clicking on the button in the tool bar. This
allows you to switch between "T-Cursor" and "Test Points" view.
In the tool bar, select the button to specify the alignment of the measurement
series.
Click on the T-Cursor. Move over one of the axes and rotate the T-Cursor with the
arrow in the required direction of measurement.
5-17
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Saving the image for Use the following procedure to save an image for your test report:
the test report
In the tool bar, select the button and open a frame with the cursor.
The contents of the frame is saved for subsequent printing.
Fine positioning In the Specimen area, select Created Series > Series 1.
In the work area, select the Evaluation Camera tab.
The position of the selected series is moved to automatically.
Position the workpiece so that the required start point of your measurement series
is located exactly in the zero point of the crosshairs.
In the tool bar, select the button to position the start point of the series at the
point selected.
The T-Cursor moves to this position.
In the tool bar, select the button to correct the alignment of the measurement
series.
Click on the T-Cursor. Move over one of the axes and rotate the T-Cursor with the
arrow in the required direction of measurement.
5-18
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the Specimen area, specify your specimen by selecting Specimen > Created
Specimen.
In the Work area, select the button to start measurement of both series.
WARNING
Crushing hazard
Keep your hands away from the area around the moving test unit.
5-19
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
5-20
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The History screen opens.
Displaying In the work area in the Trend Line tab, you can have the measured data displayed as a
measurement series measurement series.
5-21
Hardness Tester Examples of ecos Workflow
DuraScan 50 G5 / 70 G5 / 80 G5
Printing a
measurement report Use the button to print the measurement report.
5-22
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
The Specimen area can be used to create the specimens to be measured and select the
test type.
You have the following options for creating specimens:
▪ You can create new specimens.
▪ You can load specimens that have already been measured and saved, and use
them as templates.
6-1
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
At Directory, you can see the default directory that is used to store the specimen data.
How to change the default directory is described in chapter "File Locations tab", page 8-
8.
At Test Type, you can select the type of measurement you wish to perform: Single
measurement, series measurement, CHD, Nht, Rht.
A basic distinction is made between measuring an individual test point and measuring a
series.
Single measurement Single measurement is carried out to measure an individual test point.
6-2
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen onwards in the
Workflow.
The Method screen opens.
6-3
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
The ecos Workflow automatically allocates the name <Specimen name> - Pattern for
the loaded template.
You can change the name:
Enter the required name in the Specimen Name box.
The name is changed.
A duplicate name is recognized by the system and rejected.
In the work area, select the button to move the specimen onwards in the
Workflow.
The Method screen opens.
6-4
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Test Here you can select the indenter, and thus the test procedure A
procedure maximum of two indenters (Vickers, Knoop) are available here.
Normally, only one indenter is installed in the turret. A second indenter
can optionally be installed.
Test method Here you can choose one of the test methods listed.
Objective Up to five lenses can be fitted in a 6-fold turret, which are available for
lenses selection here.
Zoom level By default, you can select one of two zoom levels.
6.2.1 Conversion
When you select Use Conversion, the following screen opens:
When dealing with different hardness test methods, it is often necessary to convert the
hardness value measured with one method into that of a different method. For this
reason, empirical values have been determined on the basis of numerous comparative
6-5
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
measurements. Conversion tables have been drawn up and standardized in the relevant
standard. Different tables are valid for different materials. Consult the related standard
for a conversion in accordance with the standard.
For example, the conversion tables also state the tensile strength of steel in N/mm².
Select Conversion Table lists the test methods offered by the machine as well as its
standard-based conversion options.
The following conversion tables are available for selection:
▪ DIN EN 50150
▪ ASTM E140-05
▪ DIN EN ISO 18265
If you selected conversion table ASTM or DIN EN ISO 18265, you can choose between
several materials.
After selecting a conversion table, choose material and test unit.
6-6
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
If you have selected series measurement, you can specify here whether you wish to
have the test points set automatically or want to set them later manually in the Test
Point Editor in the Position screen.
When you select Automatic Indentation Spacing, you can see the following default
settings:
▪ Distance to edge 0.1 mm
The first test point is set 0.1 mm away from the edge.
▪ Dx3
The second test point position is calculated from the diagonal of the first test point,
multiplied by 3.
▪ Number of test points 10
You can change the distance to the edge, the multiplier and the number of test points.
6.2.4 Limits
You can enter different limits depending on the test type selected.
If a measured result is within the limits, OK is displayed for the measurement.
At the evaluation of the results in the History screen, the limits are displayed as red
lines in the line diagram. Further information see Chapter "Information in the work area",
page 7-2.
Statistical analysis can only be performed when limit values are activated.
The hardness limits are specified in the same unit of hardness as for the test method
selected, e.g. HV for the Vickers test method.
6-7
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
This allows you to optimize the time taken for the measurement series, where only the
necessary number of test points is measured.
Select the number you require in the Number of Indentations after Reaching
Hardness Limit box.
Continue in the Workflow once you have made all necessary selections.
In the work area, select the button to move the specimen onwards in the
Workflow.
The Position screen opens.
▪ From the Result screen if you wish to replace a value and have selected the
button.
In the Position screen, you work with the evaluation camera and the optional overview
camera.
▪ The evaluation camera is always included in the DuraScan.
▪ The overview camera is included in the DuraScan 70 and DuraScan 80.
6-8
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
The functions for creating measurement series are only available if series, CHD, Nht or
Rht measurement has been selected as the test type.
Both the overview and the evaluation camera allow you to move in the live image:
6-9
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
▪ Click on the required position in the camera image. You move immediately to the
selected position.
▪ Small movements can best be carried out using "Move with Line". Here, you press
and hold the left mouse button to move in the camera image.
The Position screen offers the following elements in the Evaluation Camera and
Overview Camera tabs:
▪ In the Control elements area, all elements required to move in the X-Y-Z-axes
▪ Buttons in the tool bar
▪ Additionally for measurement series: All tools required to define your measurement
series are in the tool bars.
Controls
The following controls can be found in the lower part of the screen. See also "Control
elements", page 4-4.
Controlling the Z-axis
With the slider you can
▪ focus the image
▪ raise or lower the indenter
If you move the slider up or down with the mouse button pressed and
held, the test unit moves along the Z-axis.
When rapid traverse (rabbit icon) is active, the test unit moves at a 10
times higher speed. Manual focussing is not possible when rapid
traverse is active. Rapid traverse can only be activated when the
indenter is swivelled in. It is deactivated automatically when the turret
position is changed.
6-10
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Turret control
The turret is equipped with lenses and indenters. Click on the relevant
symbol to swivel in an indenter or lens. The tool which is swivelled in
is marked with a red arrow in the turret.
As the machine normally detects the required function, this merely
acts as an additional control option.
During all positioning movements ensure that the test unit is high enough above the
specimen holder to prevent interference between lenses/indenter and specimen holder.
Any interference between lens and specimen holder requires the objective lens to be
calibrated. See Chapter "Adjusting lenses (Assistant - Objective Lenses)", page 8-17.
Tool bar
You can see the following symbols directly under the camera image in the tool bar.
Superimposes a grid over the image.
Brightness control
Further information about symbols can be found in Chapter "Buttons", page 4-6.
6-11
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
You can use the button to set test points manually, the and buttons to
distribute them along a defined line, the button to select test points individually and
move them or delete them using the Del key.
Once you have finished defining the test point template, you can save and reload
it.
Further information about symbols can be found in Chapter "Buttons", page 4-6.
In the lower part of the screen you can select the grid representation and define the grid.
6-12
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
With the millimetre specifications for the X-Y-axis, you can set the test points closer
together or further apart by adjusting the size of the grid.
You can use the slider to zoom into the grid and increase or reduce the size of the
display.
The coordinates specifications are used for orientation for setting test points.
If you have activated the Snap to Grid checkbox, inexactly positioned test points are
automatically aligned to a grid line.
If you have already selected the Automatic Indentation Spacing checkbox in the
Method screen, you only see a message that tells you that processing test points is not
possible.
6-13
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
If you wish to change a test point, use the X-distance between Test Points or Y-
distance between Test Points box to change the related value, and select
Change.
If you wish to change a method or lens, use the Objective Lenses or Method box
to specify another lens or method, and select Change.
If you wish to sort all test points, select a series and select Sort.
The Sort function ensures that the test points are moved to one by one in succession
without unnecessary traverse paths, thus optimizing the time taken by the measurement
series.
If you have already selected the menu item Automatic Indentation Spacing in the
Method screen, you see only a message that tells you that processing test points is not
possible.
There are two control elements in the lower part of the screen. Further information about
these control elements can be found in the Chapter "Evaluation camera and overview
camera", page 6-8.
During all positioning movements ensure that the test unit is high enough above the
specimen holder to prevent interference between lenses/indenter and specimen holder.
Any interference between lens and specimen holder requires the objective lens to be
calibrated. See Chapter "Adjusting lenses (Assistant - Objective Lenses)", page 8-17.
You can use the button to set test points manually, the and buttons to
distribute them along a defined line, the button to select test points individually and
move them or delete them using the Del key.
6-14
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
You can also use the button to position the start point of the series in the centre of
the crosshairs and the button to switch between T-Cursor and test points.
A detailed description of this function can be found in the Chapter "Specifying the
position", page 5-12.
Further information about symbols can be found in the Chapter "Buttons", page 4-6.
There are control elements in the lower part of the screen. Further information about
these control elements can be found in the Chapter "Evaluation camera and overview
camera", page 6-8.
During all positioning movements ensure that the test unit is high enough above the
specimen holder to prevent interference between lenses/indenter and specimen holder.
Any interference between lens and specimen holder requires the objective lens to be
calibrated. See Chapter "Adjusting lenses (Assistant - Objective Lenses)", page 8-17.
You can use the button to set test points manually, the button to select test
points individually and move them or delete them using the Del key.
You can use the button to approach individual test points and perform fine
positioning, for example if an indentation is already available.
You can also use the button to position the start point of the series in the centre of
the crosshairs and the button to switch between T-Cursor and test points.
A detailed description of this function can be found in the Chapter "Specifying the
position", page 5-12.
Further information about symbols can be found in the Chapter "Buttons", page 4-6.
6-15
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Camera used:
Evaluation Camera tab
6-16
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Camera used:
Evaluation or overview camera
Camera used:
Overview camera
In the tool bar, select the button and open a frame with the cursor.
The content of the frame is saved for a subsequent printout.
Camera used:
Evaluation or overview camera
With measurement series you can select the series and specimens you wish to measure
in the Specimen area.
Specimen tab
An individual specimen is selected by clicking on it. To select two or more specimens,
press and hold the Ctrl key during the selection.
6-17
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Series tab
An individual series is selected by clicking on it. To select two or more series, press and
hold the Ctrl key during the selection.
Select the specimen(s) or series you wish to measure.
In the work area, select the related tab, either Evaluation Camera or Overview
Camera.
WARNING
Crushing hazard
Keep your hands away from the area around the moving test unit.
Here you have two options for replacing a previously measured value:
▪ You can automatically remeasure a previously measured indentation.
▪ You can perform a replacement measurement, i.e. a new indentation is set and
measured. The old indentation is discarded and replaced with the new one.
6-18
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Remeasurement
WARNING
Crushing hazard
Keep your hands away from the area around the moving test unit.
The new indentation will be measured and the value displayed in the Result screen.
The old indentation is discarded.
6-19
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
In the Specimen area, you can select the specimens and series for which you require
information. Select an individual value in the table in the status panel.
Specimen tab
An individual specimen is selected by clicking on it. To select two or more specimens,
press and hold the Ctrl key during the selection.
Series tab
An individual series is selected by clicking on it. To select two or more series, press and
hold the Ctrl key during the selection.
Status panel
In the status panel, select the value for which you require information.
6-20
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
6-21
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Four measurement lines appear. The tool bar shows another button , which can be
used to rotate the measurement lines.
6-22
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
To rotate
To move
Move the cursor over a measurement line.
An arrow will appear.
Move the measurement line to the required position.
Proceed in the same manner for the other lines.
The measurement lines are correctly positioned when the inside of the lines is touching
the corners of the indentation (the "clear distance" is measured here).
If measurement lines are outside the visible range, they can be found by rotating to
move the line into the visible range.
6-23
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Prerequisite
Measurement has started.
6-24
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
In the work area, select the button to move the specimen back to
Position.
In the Specimen area, select the Series tab.
If necessary: Now make your changes.
6-25
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Prerequisite
The measurement series is completed.
In the work area, select the button to move the specimen back to
Position.
In the Specimen area, select the Series tab.
Add other test points now.
If necessary: In the work area, select the Coordinates tab.
Select the Sort button to sort the test points.
The test points are sorted by "X-Distance to Start Point", and an optimized
measurement series is available.
6-26
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
Right click on one of the names of the columns in the result list (e.g. hardness
value).
Select the Edit Column Info sub-menu and switch to the Additional Test Point
Info button.
The following screen opens:
6-27
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
If required, enter the desired column name for each additional information.
If you want to use the additional information for your measurement, press the
related Use button to activate the additional information.
If required, activate the Enter additional info before each measurement button. If
this button is activated, additional information can be entered before each hardness
testing. When you press the Start Testing button, a screen appears for each
activated additional information that permits the additional data to be entered. This
data is automatically entered in the result list after the hardness testing.
If you only want to fill the additional information into the result list, deactivate the
Enter additional info before each measurement button.
To confirm the settings, press OK.
6-28
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
To fill the additional boxes of the result list with data, double-click on the related
additional box.
The following screen opens:
After you have filled in the boxes and pressed Next, the entered data is
automatically entered into the result list.
6-29
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
6-30
Hardness Tester Enhanced settings and
functions
DuraScan 50 G5 / 70 G5 / 80 G5
In the Specimen area, you can select the specimens and series for which you require
information. Select an individual value in the table in the status box.
Specimen tab
To select an individual specimen, click on it. To select two or more specimens, press
and hold the Ctrl key during the selection.
Series tab
To select an individual series, click on it. To select two or more series, press and hold
the Ctrl key during the selection.
Status panel
In the status panel, select the value for which you require information.
If changes are necessary, you can move a stored specimen from the History screen
back to the Result screen.
6-31
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
In the Specimen area, select the specimens and/or series for which you require
information. Select an individual value in the table in the status box.
Specimen tab
To select an individual specimen, click on it. To select two or more specimens, press
and hold the Ctrl key during the selection.
Series tab
To select an individual series, click on it. To select two or more series, press and hold
the Ctrl key during the selection.
Status box
In the status box, select the value for which you require information.
7-1
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
Thumbnails tab
7-2
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
Trend Line tab The Trend Line tab shows the progression of measurement series.
One or more specimens can be selected in the Specimen area. The specified limits
must be identical for the selected specimens.
Limit values are shown as red lines.
7-3
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
Histogram tab The Histogram tab shows the distribution of the values in percent over the hardness
range as a histogram.
One or more specimens can be selected in the Specimen area. The specified limits
must be identical for the selected specimens.
The Statistical Values tab is used for the statistical analysis of values.
Statistical vlaues tab One or more specimens can be selected for analysis in the Specimen area. The
specified limits must be identical for the selected specimens.
Meaningful analysis can only be carried out if you have specified hardness limits.
7-4
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
User boxes tab You can use the User boxes tab to view entries in the individual boxes. These boxes
are also output in the test report.
You can additionally amend user boxes.
Further information about the tables see "User boxes tab", page 8-6.
7-5
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
A minimum of ten user boxes has to be created for a test report to be generated. If you
have defined more than ten user boxes, the test report will only include the first ten
boxes, due to lack of space.
If you wish to edit templates or create your own templates, you must know how to use
the List&Label tool. The utilization of this tool is not described in this manual.
With the DuraScan 70/80 you can save an image for your test report in the Position
screen (see "Saving image for test report", page 6-17). In this case select a template
with image for the test report. The saved photo will be included in the report.
Types of test reports You specify the form to be used to output your test report.
You have the following options:
▪ You can generate a print preview.
▪ You can generate a PDF file.
▪ You can start printing immediately.
7-6
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
You can save this preview (as a PDF file, for example).
7-7
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
Select Print.
The Ecos WorkFlow – List&Label Report window opens.
7-8
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
In this window you can select the output format (Adobe PDF or a printer, for example).
If you select a report with an image, set the graphics resolution as required.
Once you have made all necessary selections you can save the PDF file.
7-9
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
Select Print.
The Ecos WorkFlow – List&Label Report window opens.
7-10
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
7-11
Hardness Tester Evaluation and documentation
DuraScan 50 G5 / 70 G5 / 80 G5
7-12
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the tool bar, you'll find the additional button which you can use to switch display
of the length measurement on and off.
In the tool bar, select the button to enable measurement lines and display of
the length measurement.
The tool bar also shows an additional button which you can use to rotate the
measurement lines.
You have three lines which you can rotate and move.
If necessary: Select the button in the tool bar and rotate the measurement
lines in the required direction.
To measure the indentation, move the lines to the required position.
The measured length is displayed.
8-1
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Select the button in the tool bar and save the image with length measurement
at the required storage location.
Condition
You are in the Position screen and have performed the following operations:
8-2
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Reference points can be displayed under the evaluation camera or the overview
camera.
Sample reference points can only be displayed in conjunction with the overview camera.
Select the camera you require.
In the Specimen area, select a series with Created Series or press the Ctrl key to
select all series you wish to display.
If necessary: In the tool bar, select the button to display the individual test
points.
The individual test points are displayed.
Select View > Show Reference Points.
The series reference points or specimen reference points are also shown and mark the
start points of the series or specimens.
You can now save an image that shows the reference points for the test report.
In the tool bar, select the button and open a frame with the cursor.
The contents of the frame will later be used for the printout.
8-3
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The General settings menu includes functions that are mainly used for the
administration of the unit.
The Configuration wizard menu includes all functions you require to calibrate the
machine.
8-4
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The measurement counter reading is for information only. The measurement counter
automatically counts the number of measurements performed on the machine.
Selecting unit of In the Unit box, you can select the required unit (mm or inch). All length specifications –
measure such as the coordinates in series measurement – appear immediately in the selected
unit of measurement.
Selecting a language On delivery, the unit is set to English. Select the language you require in the Language
box.
8-5
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Dwell time Setting the dwell time allows you to specify the exposure time for the test force. The
values selected here are global parameters for all test methods.
Please observe the applicable standards when you modify the dwell time.
Configuration wizard In a newly delivered machine, the query window opens to start the Configuration wizard
query window when you call up the ecos Workflow. This default setting is expedient, since the unit
must be calibrated after the installation of the lenses and indenters. Later, you can call
up the Configuration wizard as required, and do without the query window.
For further information about test reports see "Generating a test report", page 7-5.
Configuring user boxes To adjust the prefabricated test report to your requirements, you can allocate any names
to ten user boxes with default settings. Use the User Box button to create additional
user boxes. Use the Delete User Box button to delete a selected user box.
Use the Empty circulating memory user fields button to delete the saved data of the
user fields that was entered in the Workflow step Specimen Workflow step.
If the Export Editor option is available in your machine, activate the related button to
add the user boxes to the manual export.
8-6
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Completing user boxes The created user boxes can be completed in the following screens:
▪ in the Specimen screen
▪ in the Result screen
▪ in the History screen.
User levels The ecos Workflow is characterized by a role-based utilization concept with three
different user levels. The table shows the main user privileges per level:
Level User privileges
1 - View Viewing results in the History screen
2 - User The "View" privileges, plus carrying out measurements, changing
general settings (such as user administration), and performing simple
calibrations (such as adjusting lenses)
3 - Cal Same privileges as "User", plus performing load calibration
To avoid operating errors due to incorrect machine settings, be very careful when you
assign user privileges. Users with "Cal" privilege MUST be trained adequately.
8-7
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Editing a user Use the Created Users box to select the user whose name, user level or password
you wish to modify.
Select the Change User button.
Modify the relevant data.
Select the Apply button.
The data is modified.
Deleting a user On each level, one user with the related name (View, User, Cal) is preconfigured. These
preconfigured users and the user who is currently logged in cannot be deleted.
Caution
The deletion is executed at once and without any query!
Use the Created Users box to select the user you want to delete.
Select the Delete button.
The user is deleted at once.
Select the file type whose file location you wish to change.
Select the related Change Path button.
Select the file location you want to use.
Select the OK or Apply button to confirm the settings.
8.3.1.5 Advanced
Use the Advanced tab for additional settings for single and series measurements
8-8
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
8-9
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
At General, set the required settings for Port, Baud Rate, Stop Bits, Decimal
Separator and Value Separator.
8-10
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Select the Export List button to select the data you want to export.
Values for the serial export can only be included after they have been selected. This
requires the fields that are displayed in the Non-exported values list to be marked
and, with the arrow key "right" moved to the Exported values list. Values that shall
not be in included can be moved back with the "left" arrow key.
If required, use the up and down arrow keys to arrange the values in the Exported
values list.
Press Apply to confirm the settings.
8.3.1.8 Colours
You can individually adjust the colour settings for all graphic objects that are used in the
evaluation or overview camera.
Select the Colour tab.
8-11
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Displaying Info
Here you can find information about machine configuration and program versions that
can be of interest to the Service team.
8-12
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Enter the product key you received from your dealer on purchase.
Select the Apply button.
The login screen opens. Log in again and try out the new functions.
8-13
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The number of tabs displayed depends on the machine configuration. The tabs 2nd
Indenter and Overview Camera are only displayed if these options are available.
You need "Cal" privileges to make all selections in the Configuration wizard screen
which are necessary for a proper and accurate functioning of the machine. If users have
not yet been configured (initial startup), you can log in as "Cal" user (without password).
If you have "User" rights, you can adjust position and focus of the lenses and define the
turret configuration.
8-14
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Select the equipped positions of the turret one by one in succession and use the list
to specify which indenter / lens is mechanically inserted in the position.
Check that the representation on the screen corresponds to the actual allocation of
the equipment.
Ensure that indenters and lenses are correctly allocated. The machine will not function
properly is this is not the case.
8-15
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The first subtab, Manual, is active. The control elements are available at the bottom of
the screen (see also "Control elements", page 4-4).
In the boxes Test procedure and Test method, select the test procedure and the
test method for which you want to check load calibration.
Use the slider (control element) to move the test unit upwards.
Place the load cell of the calibration unit on the test anvil.
In load calibration, the tip of the indenter touches the load cell of the calibration unit. You
should therefore place a workpiece on the load cell as an intermediate layer, and set the
display of the calibration unit to zero.
8-16
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Read the measured value on the calibration equipment, and enter it (in "g") in the
box provided.
Select the Apply button.
The machine performs a measurement with the corrected value. If no error message is
displayed, the value is confirmed.
Repeat the process if an error message is displayed.
Repeat these steps for the other test procedures and test methods for which you
want to check load calibration.
The difference of the measured values is calculated automatically, and load correction is
carried out. All load corrections are listed in a table.
Table
Select the Table subtab.
The table lists all possible loads for the machine and the applied load corrections.
8-17
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
You can see three subtabs: Position, Focus and Magnification. The first subtab
Position is active. The control elements are available at the bottom of the screen (see
also "Control elements", page 4-4.
Adjusting the position Use the set screws to adjust the lenses so that they are positioned right above the
indentation of the indenter.
Select the Position subtab (if not already active).
In the boxes Measurement Type and Test Method, select the type of
measurement and test method for which you want to adjust the setting.
Place a workpiece that is suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement starts with the indenter that is fitted at position 1 of the turret. The image
is shown by the lens with the lowest magnification.
8-18
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
When prompted, use the grub screws on the lens to adjust the position so that the
crosshairs are in the middle of the indentation.
The indentation must be positioned on the workpiece so that the crosshairs intersect
with the corners. Otherwise you must turn the indenter slightly with the fork spanner.
See "Indenter and lenses", page 2-11.
Correct position (example of Vickers): Crosshairs centred in the indentation, intersecting with the corners
Repeat these steps for the other lenses (Next Lens button).
Adjusting the focus For all lenses, adjust the position on the Z-axis at which a focussed image is displayed.
Perform this calibration by the eye.
Select the Focus subtab.
8-19
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the boxes Measurement Type and Test Method, select the type of
measurement and test method for which you want to adjust the setting.
Put a workpiece that suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement starts with the indenter that is fitted at position 1 of the turret. The image
is shown by the lens with the greatest magnification.
Use the slider (control element) to focus the image.
Select the Save focus button.
Select the Next Lens button if you wish to adjust other lenses.
Adjusting the To adjust the magnification factor, you require a calibrated test block (micrometre
magnification factor objective).
Select the Magnification subtab.
Place the test block on the test anvil and align so that the scale is visible.
In the Required value box, enter the required value of the test block for the scale
distance you are remeasuring.
8-20
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Use the mouse to align the green lines to the scale and to measure the scale
distance you entered.
Ensure that the green lines are precisely aligned, for example to each left-hand edge of
a scale marking.
Select the Save magnification factor button when the green lines are positioned
precisely.
The actual dimension is displayed and the magnification factor is calculated
automatically.
Select the Next Lens button if you wish to adjust other lenses.
Adjusting the position The position of the second indenter must also be adjusted so that with all lenses the
(2nd indenter) indentation is centred in the image.
Select the Position subtab.
8-21
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the boxes Measurement Type and Test Method, select the type of
measurement and test method for which you want to adjust the setting.
Place a workpiece that is suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement is performed with the second indenter. The lens with the lowest
magnification is swivelled in.
Move the X-Y-axis until you have found the indentation and positioned it exactly in
the centre of the image.
Select the Calculate Offset button.
The Configuration wizard calculates the offset between the first and the second
indenter. This is compensated for automatically by the slide system during the
measurements. Adjusting the position mechanically is not necessary.
Adjusting the focus The offset between the first and second indenter has also to be calculated and saved for
(2nd indenter) the focus of the lenses.
Select the Focus subtab.
8-22
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the Indenter box, select the indenter for which you want to adjust the settings.
Place a workpiece that is suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement is performed with the second indenter. The image is shown by the lens
with the greatest magnification.
Use the slider (control element) to focus the image.
Select the Save focus button.
If your machine is fitted with a second indenter, an additional tab 2nd indenter appears.
In the 2nd indenter you can adjust position and focus of the lenses for the second
indenter.
Such a calibration of the second indenter requires quite some experience and should
preferably only be carried out by the Service team.
Adjusting the position The position of the second indenter must also be adjusted so that with all lenses the
(2nd indenter) indentation is centred in the image.
Select the Position subtab.
8-23
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the boxes Measurement Type and Test Method, select the type of
measurement and test method for which you want to adjust the setting.
Place a workpiece that is suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement is performed with the second indenter. The lens with the lowest
magnification is swivelled in.
Move the X-Y-axis until you have found the indentation and positioned it exactly in
the centre of the image.
Select the Calculate Offset button.
The Configuration wizard calculates the offset between the first and the second
indenter. This is compensated for automatically by the slide system during the
measurements. Adjusting the position mechanically is not necessary.
Adjusting the focus The offset between the first and second indenter has also to be calculated and saved for
(2nd indenter) the focus of the lenses.
Select the Focus subtab.
8-24
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the Indenter box, select the indenter for which you want to adjust the settings.
Place a workpiece that is suitable for the test method on the test anvil.
Select the Start Measurement button.
Measurement is performed with the second indenter. The image is shown by the lens
with the greatest magnification.
Use the slider (control element) to focus the image.
Select the Save focus button.
Adjusting the To adjust the magnification factor, you need a rule with a suitable scale (mm, 1/10 mm),
magnification factor for example a sliding caliper.
(overview camera) Select the Magnification subtab.
Ensure that the height of the overview camera is correct. To do so, switch to the
Objective Lenses tab and select the AF Indentation function in the Controls area, for
example.
Put the rule on the test anvil and align it such that the scale is visible.
8-25
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
In the Required Value box, enter the required value for the scale distance you are
remeasuring.
Use the mouse to align the green lines to the scale so you can measure the scale
distance you entered.
Ensure that the green lines are precisely aligned, for example to each left-hand edge of
a scale marking.
Select the Save magnification factor button when the green lines are positioned
precisely.
The actual dimension is displayed and the magnification factor is calculated
automatically.
Put a workpiece with a sharp edge on the test anvil and align it such that the edge
touches the crosshairs in the centre of the image.
Select the Save button.
8-26
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The current position is saved. A lens of the evaluation camera is swivelled in and the
evaluation camera view is displayed.
Move the edge manually to the centre of the image.
Select the Save button.
The offset from the centre of the image to the centre of the image of the two cameras is
calculated.
Repeat the procedure to check the result.
If your machine (DuraScan 70 / DuraScan 80) is equipped with an overview camera, an
Overview Camera tab appears additionally.
Use the Overview Camera tab to adjust magnification factor and offset between
evaluation camera and overview camera.
Calibrating the overview camera requires quite some experience and should preferably
only be carried out by the Service team.
Adjusting the To adjust the magnification factor, you need a rule with a suitable scale (mm, 1/10 mm),
magnification factor for example a sliding caliper.
(overview camera) Select the Magnification subtab.
Ensure that the height of the overview camera is correct. To do so, switch to the
Objective Lenses tab and select the AF Indentation function in the Controls area, for
example.
Put the rule on the test anvil and align it such that the scale is visible.
In the Required Value box, enter the required value for the scale distance you are
remeasuring.
8-27
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
Use the mouse to align the green lines to the scale so you can measure the scale
distance you entered.
Ensure that the green lines are precisely aligned, for example to each left-hand edge of
a scale marking.
Select the Save magnification factor button when the green lines are positioned
precisely.
The actual dimension is displayed and the magnification factor is calculated
automatically.
Put a workpiece with a sharp edge on the test anvil and align it such that the edge
touches the crosshairs in the centre of the image.
Select the Save button.
8-28
Hardness Tester Configuration and calibration
DuraScan 50 G5 / 70 G5 / 80 G5
The current position is saved. A lens of the evaluation camera is swivelled in and the
evaluation camera view is displayed.
Move the edge manually to the centre of the image.
Select the Save button.
The offset from the centre of the image to the centre of the image of the two cameras is
calculated.
Repeat the procedure to check the result.
8-29
Hardness Tester Maintenance and care
DuraScan 50 G5 / 70 G5 / 80 G5
NOTE
Risk of damage
Only trained personnel are permitted to replace fuses.
The fuse cover (1) sits between the main switch (with the green indicator lamp) and the
mains connection.
NOTE
Risk of damage
Switch off the machine at the main switch before you replace a fuse.
9-1
Hardness Tester Maintenance and care
DuraScan 50 G5 / 70 G5 / 80 G5
DANGER
Electric current
Connect the machine only to a grounded electrical outlet with a protective earth
conductor contact.
Any intervention in the electrical/electronic part of the unit is prohibited.
9-2
Hardness Tester Startup and Fitting with
Optional Accessories
DuraScan 50 G5 / 70 G5 / 80 G5
10-1
Hardness Tester Startup and Fitting with
Optional Accessories
DuraScan 50 G5 / 70 G5 / 80 G5
NOTE
Risk of damage
Indenters should always be stored in their plastic case when not in use in order to avoid
damage.
NOTE
Risk of damage
The indenter adapter comes with a transport safety device. Never remove the safety
device before the adapter has been installed on the turret.
Fasten the adapter with 4 Allen head screws in any desired position of the turret.
After installation of the indenter adapter, remove the transport lock with an Allen
wrench.
10-2
Hardness Tester Startup and Fitting with
Optional Accessories
DuraScan 50 G5 / 70 G5 / 80 G5
If required, select the 'Create file after every testing' button to automatically create
a separate file for each measured value. These are saved in the selected file
location for autoexport files.
The file name is composed of date/time/file type.
If required, select the Save values in common file button to automatically save all
values in a common file.
To save the values in a common file, the file name box must additionally be filled
in. This file is saved in the file path selected for autoexport files.
The file name consists of file name/file type.
Automatic data export is not performed if you select none of the two buttons.
Manual data export, however, can be performed as before.
Select the Set export button to individually adjust the export.
10-3
Hardness Tester Startup and Fitting with
Optional Accessories
DuraScan 50 G5 / 70 G5 / 80 G5
At General, select the required settings for output format, decimal marks and value
separators.
Select the 'Export List' button to specify the data that shall be included in the export.
The settings are used for manual and for automatic export.
A value can only be included in the export after it has been selected. This requires
the boxes that are displayed in the Non-exported values list to be marked and,
with the arrow key "right" moved to the Exported values list. If values shall not be
included, use the arrow key "left" to move them back.
If required, use the up and down arrow keys to arrange the values in the Exported
values list.
To confirm the settings, press Apply.
10-4
Hardness Tester Messages and problems
DuraScan 50 G5 / 70 G5 / 80 G5
11-1