HIGH VOLTAGE TANK FOR X-RAY GENERATORS E-40G HFD
FOR RADIOGRAPHY AND FLUOROSCOPY
40 kHz - 125 kV - 3,5÷15 kW
This High Voltage X-Ray Generator is suited for the supplying of the X-Ray tube’s housing of a
cardiological fluoroscopy system, where high power, 15kW pulsed fluoroscopy has to be reached, at up
to 30 frames per second.
X-RAY GENERATOR ASSEMBLY Use in Radiography
(2) Maximum Power: . . . . . . . . . . . . .3,5 kW (Inverter E-9040-5)
Operating Frequency (H.V.) 40 kHz 5 kW (Inverter E-9040-15P)
Maximum Voltage to X-Ray Tube . . . . . . . . . . . . . . . . . . . . .125 kV
Use in Fluoroscopy Receptacle (-) to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62,5 kV
(1) Maximum Power: Receptacle (+) to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62,5 kV
Maximum Current to X-Ray Tube: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Continuous Fluoroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360 W
in accordance with (2) and the
Pulsed Fluoroscopy Duty Cycle 25%- 50% . . . . . . . . . . . . . . . . . . . .
X-Ray Tube Specifications
3,5 kW (Inverter E-9040-5)
Ripple at Maximum Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 1,5 %
15 kW (Inverter E-9040-15P)
kV Rise Time to Maximum Power . . . . . . . . . . . . . . . . . . . . .< 1 ms
Maximum Voltage to X-Ray Tube . . . . . . . . . . . . . . . . . . . . .120 kV
Receptacle (-) to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 kV
Output Signals
Receptacle (+) to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 kV
kV+ (Analog Signal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0÷6,25 V
Maximum Current to X-Ray Tube: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
kV- (Analog Signal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0÷6,25 V
in accordance with (1) and the
mA+ & mA- (Analog Signal) . . . . . . . . . . . . anodic mA readings
X-Ray Tube Specifications
Ripple at Maximum Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .< 1,5 %
Dimensions and Weight
kV Rise Time to Maximum Power . . . . . . . . . . . . . . . . . . . . .< 1 ms
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (see picture)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17,5 kg
Thermal Features
Compensation Bellow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0,2 dm3
Thermal Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57°C ACCESSORIES
400 Hz Filament Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .S15
The Generator can be delivered with Big or Small HV-Receptacles.
Doc.: 9921
Specifications subject to change without notice.
I.M.D. Generators s.r.l.
Viale Matteotti, 28/a - 24050 Grassobbio - BG - ITALY
generators Ph.:+39.035.526344 • Fax.:+39.035.526086 • E-mail: info@imdxray.com • web: www.imdxray.com