Calibration Technology Starts Here
Martel
Precision Calibrators “10” Series Calibrators
Work better and get better work with Martel TEN series
multifunction calibrators. This family of 5 models scale up to do
any size job you need when calibrating process instrumentation.
Start at the top with the DMC-1410 documenting multifunction
calibrator. It’s versatile, providing access to a complete range of
calibration functions while performing automated on the fly calibration data collection
and storage. A simple easy-to-use software package is included that allows the user to
build a database of all assets that need calibration and download work orders to the
calibrator.
Next in line is the MC-1210 multifunction calibrator. It’s a rugged and reliable universal
calibrator. Like all the others in this series, the MC-1210 is based on the proven reliable, accurate and
stable MC-1200. It’s dual display and isolated readback allows it to power a transmitter under test while
reading its milliamp output. Truly an all-in-one calibrator. The MC1210 also has a wide range of
switch test features for both pressure and temperature switches.
The MC-1010 provides a high level of functions and features at an easy to swallow price for
the less demanding user who does not require the isolated read-back feature found on the
DMC-1410 or MC-1210.
For those who need specialty temperature calibration with high accuracy, the PTC-8010 is the
choice. Special display features show the cold junction temperature and milliVolt equivalents at
a glance for thermocouples. Ohms equivalents are shown when using the RTD functions.
The PSC-4010 is a superior loop calibrator with voltage, current and frequency functions. With the best
display in the business it makes the essentials of instrument calibration easier than ever. And, a
bonus feature not found in other loop calibrators is frequency in and out.
This innovative series features the introduction of a new, high contrast
ClearBrite™ graphic display. The display features a vivid white backlight that
makes the display easy to read in all light conditions.
All of these models have Martel’s easy to learn yet powerful 3 key menu
structure. It’s the same menu used in all of the company’s BetaGauge pressure
calibrators, too. Learn it once and you’ll know how to use every significant
calibrator Martel makes.
. New ClearBrite™ Graphic Display
. Auto Stepping/Auto Ramping
General Features
. 3 Key Martel Menu System
. Scroll/Step Output Control
. NIST Calibration Certificate
. Proven Rugged Reliable Design
. Fuseless Protection to 250 VAC
For more information:
- Call us toll free at 800-821-0023
or 603-434-1433
- Fax us at 603-434-1653
- Email us at sales@martelcorp.com
and ask for the “10” Series Calibrators
- Or go to our website,
www.martelcalibrators.com
and click on Multifunction Calibrators
Martel “10” Series Calibrators
10 Series Function Table
†with optional BPPA-100 pressure module adapter
10 Series Feature Table
General Specifications (applies to all models)
Operating Temperature -10 to 50°C
Storage Temperature -20 to 70°C
Power (4) AA Alkaline or optional rechargeable batteries
Low Battery Warning Yes, on display
ClearBrite™ Display High contrast 128 x 64 pixel addressable graphic
LCD w/daylight backlight
2.4 x 1.8 in. • 63 x 44 mm
Serial Communications Yes, ASCII, RS-232, requires optional Martel
1919069 serial cable or 1919896 USB cable
CE – EMC EN50082-1 and EN55022: 1994 Class B
Safety CSA C22.2 No. 1010-1: 1992
Weight (with batteries) 1.8 lb (0.82 kg)
Size 8.5 x 4.8 x 2.1 in. (22 x 12 x 5.3 cm)
Other IP54
Protected against misconnection to 250 VAC/VDC
DMC 1410 MC 1210 MC 1010 PTC 8010 PSC 4010
DC Voltage and Current
DC Voltage DC Voltage DC Current DC Current
Upper Isolated Lower Non-Isolated Upper Isolated Lower Non-Isolated
Model Measurement Measurement Source Measurement Measurement Source
DMC-1410 Range 0.000V - 30.000V 0.000V - 20.000V 0.000V - 20.000V 0.000mA - 24.000mA 0.000mA - 24.000mA 0.000mA - 24.000mA
Accuracy 0.01% ± 2 mV 0.01% ± 2 mV 0.01% ± 2 mV 0.01% ± 2µA 0.01% ± 2µA 0.01% ± 2µA
MC-1210 Range 0.000V - 30.000V 0.000V - 20.000V 0.000V - 20.000V 0.000mA - 24.000mA 0.000mA - 24.000mA 0.000mA - 24.000mA
Accuracy 0.015% ± 2 mV 0.015% ± 2 mV 0.015% ± 2 mV 0.015% ± 2µA 0.015% ± 2µA 0.015% ± 2µA
MC-1010 Range N.A. 0.000V - 20.000V 0.000V - 20.000V N.A. 0.000mA - 24.000mA 0.000mA - 24.000mA
Accuracy N.A. 0.015% ± 2 mV 0.015% ± 2 mV N.A. 0.015% ± 2µA 0.015% ± 2µA
PSC-4010 Range N.A. 0.000V - 20.000V 0.000V - 20.000V N.A. 0.000mA - 24.000mA 0.000mA - 24.000mA
Accuracy N.A. 0.015% ± 2 mV 0.015% ± 2 mV N.A. 0.015% ± 2µA 0.015% ± 2µA
Note: optional 50 mA reange available
Frequency
Model Measurement Source Measurement Source Measurement Source
DMC-1410 Range 2.0 CPM - 600.0 CPM 2.0 CPM - 600.0 CPM 1.0 Hz - 1000.0 Hz 1.0 Hz - 1000.0 Hz 1.00 kHz - 10.00 kHz 1.00 kHz - 10.00 kHz
Accuracy 0.05% ± 0.1 CPM 0.05% 0.05% ± 0.1 Hz 0.05% 0.05% ± 0.01 kHz 0.05%
MC-1210 Range 2.0 CPM - 600.0 CPM 2.0 CPM - 600.0 CPM 1.0 Hz - 1000.0 Hz 1.0 Hz - 1000.0 Hz 1.00 kHz - 10.00 kHz 1.00 kHz - 10.00 kHz
Accuracy 0.05% ± 0.1 CPM 0.05% 0.05% ± 0.1 Hz 0.05% 0.05% ± 0.01 kHz 0.05%
MC-1010 Range 2.0 CPM - 600.0 CPM 2.0 CPM - 600.0 CPM 1.0 Hz - 1000.0 Hz 1.0 Hz - 1000.0 Hz 1.00 kHz - 10.00 kHz 1.00 kHz - 10.00 kHz
Accuracy 0.05% ± 0.1 CPM 0.05% 0.05% ± 0.1 Hz 0.05% 0.05% ± 0.01 kHz 0.05%
PSC-4010 Range 2.0 CPM - 600.0 CPM 2.0 CPM - 600.0 CPM 1.0 Hz - 1000.0 Hz 1.0 Hz - 1000.0 Hz 1.00 kHz - 10.00 kHz 1.00 kHz - 10.00 kHz
Accuracy 0.05% ± 0.1 CPM 0.05% 0.05% ± 0.1 Hz 0.05% 0.05% ± 0.01 kHz 0.05%
Resistance Measurement PSC-4010 MilliVolts Measurement/Source
Model Ohms Low Ohms High Function Range Accuracy
DMC-1410 Range 0.00Ω - 400.00Ω 5.00Ω - 400.0Ω Read 0000 mV – 90.000 mV 0.02% ± 10 µV
Accuracy 0.015% ± 0.05Ω 0.015% ± 0.5Ω Source 0.000 mV – 100.000 mV 0.02% ± 10 µV
MC-1210 Range 0.00Ω - 400.0Ω 5.00Ω - 400.0Ω
Accuracy 0.025% ± 0.05Ω 0.025% ± 0.5Ω
MC-1010 Range 0.00Ω - 400.0Ω 5.00Ω - 400.0Ω
Accuracy 0.025% ± 0.05Ω 0.025% ± 0.5Ω
PTC-8010 Range 0.00Ω - 400.0Ω 5.00Ω - 400.0Ω
Accuracy 0.025% ± 0.05Ω 0.025% ± 0.5Ω
Resistance Source
Ohms Source Low Ohms Source High Ohms Source High
Model Range Excitation Current Accuracy Range Excitation Current Accuracy Range Excitation Current Accuracy
DMC-1410 5.0Ω - 400.0Ω 0.1 – 0.5 mA 0.015% ± 0.1Ω 400Ω - 1500Ω 0.05 – 0.8 mA 0.015% ± 0.5Ω 1500Ω - 4000Ω 0.05 – 0.4 mA 0.015% ± 0.5Ω
5.0Ω - 400.0Ω 0.5 - 3 mA 0.015% ± 0.05Ω
MC-1210 5.0Ω - 400.0Ω 0.1 – 0.5 mA 0.025% ± 0.1Ω 400Ω - 1500Ω 0.05 – 0.8 mA 0.025% ± 0.5Ω 1500Ω - 4000Ω 0.05 – 0.4 mA 0.025% ± 0.5Ω
5.0Ω - 400.0Ω 0.5 - 3 mA 0.025% ± 0.05Ω
MC-1010 5.0Ω - 400.0Ω 0.1 – 0.5 mA 0.025% ± 0.1Ω 400Ω - 1500Ω 0.05 – 0.8 mA 0.025% ± 0.5Ω 1500Ω - 4000Ω 0.05 – 0.4 mA 0.025% ± 0.5Ω
5.0Ω - 400.0Ω 0.5 - 3 mA 0.025% ± 0.05Ω
PTC-8010 5.0Ω - 400.0Ω 0.1 – 0.5 mA 0.025% ± 0.1Ω 400Ω - 1500Ω 0.05 – 0.8 mA 0.025% ± 0.5Ω 1500Ω - 4000Ω 0.05 – 0.4 mA 0.025% ± 0.5Ω
5.0Ω - 400.0Ω 0.5 - 3 mA 0.025% ± 0.05Ω
RTD Measurement/Source mV/Thermocouples (DMC-1410,
(DMC-1410, MC-1210, MC-1010 and MC-1210, MC-1010 and PTC-8010 only)
PTC-8010 only)
MilliVolts (mV) Range Accuracy
RTD Type Range (°C) Accuracy (°C) Read -10.000 mV – 75.000 mV 0.015% ± 10 µV
Ni120 (672) -80.0 – 260.0 0.2 Source -10.000 mV – 75.000 mV 0.015% ± 10 µV
Cu10 -100.0 – 260.0 1.4 Maximum current output in voltage ranges is 1 mA with an output
Cu50 -180.0 – 200.0 0.4 impedance of ≤ 1 Ω
Cu100 -180.0 – 200.0 0.3
YSI400 15.00 – 50.00 0.1 TC Type Range (°C) Accuracy (°C)
Pt100 (385) -200.0 – 100.0 0.2 With CJ OFF With CJ ON
100.0 – 300.0 0.3 J -210.0 – 0.0 0.4 0.6
300.0 – 600.0 0.4 0.0 – 800.0 0.2 0.4
600.0 – 800.0 0.5 800.0 – 1200.0 0.3 0.5
Pt200 (385) -200.0 – 100.0 0.8 K -200.0 – 0.0 0.6 0.8
100.0 – 300.0 0.9 0.0 – 1000.0 0.3 0.5
300.0 – 630.0 1.0 1000.0 – 1372.0 0.5 0.7
Pt500 (385) -200.0 – 100.0 0.4 T -250.0 – 0.0 0.6 0.8
100.0 – 300.0 0.5 0.0 – 400.0 0.2 0.4
300.0 – 630.0 0.6 E -250.0 – -100.0 0.6 0.8
Pt1000 (385) -200.0 – 100.0 0.2 -100.0 – 1000.0 0.2 0.4
100.0 – 300.0 0.3 R 0.0 – 1767.0 1.2 1.4
300.0 – 630.0 0.4 S 0.0 – 1767.0 1.2 1.4
Pt10 (385) -200.0 – 100.0 1.4 B 600.0 – 800.0 1.2 1.0
100.0 – 300.0 1.6 800.0 – 1000.0 1.3 1.5
300.0 – 600.0 1.8 1000.0 – 1820.0 1.5 1.7
600.0 – 800.0 2.0 C 0.0 – 1000.0 0.6 0.8
Pt50 (385) -200.0 – 100.0 0.4 1000.0 – 2316.0 2.3 2.5
100.0 – 300.0 0.5 XK -200.0 – 800.0 0.2 0.4
300.0 – 600.0 0.6 BP 0.0 – 800.0 0.9 1.1
600.0 – 800.0 0.7 800.0 – 2500.0 2.3 2.5
Pt100 (3926) -200.0 – 100.0 0.2 L -200.0 – 0.0 0.3 0.5
100.0 – 300.0 0.3 0.0 – 900.0 0.2 0.4
300.0 – 630.0 0.4 U -200.0 – 0.0 0.5 0.7
Pt100 (3916) -200.0 – 100.0 0.2 0.0 – 600.0 0.3 0.5
100.0 – 300.0 0.3 N -200.0 – 0.0 0.8 1.0
300.0 – 630.0 0.4 0.0 – 1300.0 0.4 0.6
Notes
All specifications apply at 23°C ± 5°C unless otherwise stated. Outside of this range add ±0.005% of reading/°C.
Specifications are for battery operation only.
Accuracy is % of reading ± floor spec.
Maximum current output in voltage ranges is 1 mA with an output impedance of ≤ 1Ω.
Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5 – 30 V.
Frequency input voltage amplitude range is 1V to 20V zero based square wave only. Output amplitude is adjustable from 1V to 20V, and is a square wave
with 50% duty cycle. For output frequency, a negative offset of approximately -0.1V is present to assure zero crossing.
In Ohms source and RTD source modes, units are compatible with smart transmitters and PLCs that use a strobing excitation current. Frequency
response is 5 msec.
BetaGauge Pressure Modules
Martel Electronics offers 27 standard pressure
AVAILABLE MODULES modules, covering gauge, vacuum, absolute,
PARAMETER/ FULLSCALE VACUUM OVER compound, and differential measurements. All
RANGE ACCURACY ACCURACY PRESSURE modules are directly compatible with the Beta-
Isolated Gauge (PSIG): Gauge II. With the Model BPPA-100 Pressure
0 to15 (0 to 1 Bar) ±0.025 % 300 %
0 to 30 (0 to 2 Bar) ±0.025 % 300 %
Module Adapter, all modules (with the exception
0 to 500 (0 to 35 Bar) ±0.025 % 200 % of the DC measurement model) are fully compatible
0 to1000 (0 to70 Bar) ±0.025 % 200 % with the Martel MC-1210 and MC-1010 Multi-
0 to1500 (0 to 100 Bar) ±0.035 % 200 % Function Calibrators, the BetaGauge 330, 321A,
0 to3000 (0 to200 Bar) ±0.05 % 200 % 311A and 301 Pressure Calibrators, the DMC-1410
0 to5000 (0 to340 Bar) ±0.05 % 200 %
0 to10000 (0 to 700 Bar) ±0.01 % 120 %
Documenting Multi-Function Calibrator, and the
Non Isolated Compound (PSIG): Martel Electronics M2001 and 3001 Laboratory/Bench
-0.4 to 0.4 (-20 to 20 mBar) ±0.1 % ±0.15 % 400 % Standards.
-1 to 1 (-70 to 70 mBar) ±0.05 % ±0.1 % 400 %
-5 to 5 (-350 to 350 mBar) ±0.075 % ±0.1 % 400 % Pressure ranges may be displayed in any of 13 user
-7.2 to 7.2 (-500 to 500 mBar) ±0.07 % ±0.1 % 300 % selectable units. Water density correction factors of 4 °C,
-10 to 10 (-700 to 700 mBar) ±0.03 % ±0.05 % 300 % 20 °C, or 60 °F can be selected for either water column
-15 to 15 (-1 to 1 Bar) ±0.04 % ±0.04 % 300 % unit. The choice of pressure unit may be restricted by
-15 to 30 (-1 to 2 Bar) ±0.025 % ±0.025 % 300 %
Isolated Compound (PSIG):
limitations on resolution of the instrument display of the
-12 to 50 (-0.8 to 3.5 Bar) ±0.03 % ±0.03 % 200 % particular calibrator the module is used with. For optimum
-12 to 100 (-0.8 to 7 Bar) ±0.025 % ±0.025 % 200 % mechanical strength, external pressure connection is made
-12 to 150 (-0.8 to 10 Bar) ±0.03 % ±0.03 % 200 % by a 1/8” FNPT 316SS connector welded to a stainless steel
-12 to 300 (-0.8 to 20 Bar) ±0.025 % ±0.025 % 150 % metal plate.
..
Isolated Absolute (PSIA):
0 to 15 (0 to 1 Bar) ±0.04 % 300 %
0 to 30 (0 to 2 Bar) ±0.025 % 300 %
0 to 50 (0 to 3.5 Bar) ±0.03 % 300 %
..
General Features
27 standard ranges
0 to 100 (0 to 7 Bar) ±0.025 % 300 %
0 to 300 (0 to 20 Bar) ±0.025 % 200 % Gauge, vacuum, absolute, compound, and
Differential (PSID): differential measurements
0 to 5 (0 to 350 mBar) ±0.075% 400 % Accuracy specified over 15 °C to 35 °C range
0 to 30 (0 to 2 Bar) ±0.025% 300 %
0 to 50 (0 to 3.5 Bar) ±0.03% 300 % Isolated and non-isolated measurements,
range dependant
BetaGauge Model BPPA-100
Pressure Modules Pressure Module Adapter
..
MARTEL 3001
Lab The 3001
.
General Features
standard The Martel 3001 Bench Calibrator combines the power and Superior calibration accuracy to 0.0025% of reading
Direct keyboard entry or cursor entry with decade
features of the M2000 voltage/current calibrator,
multifunction with the addition of thermocouple, RTD, and pressure
..
control
calibrators calibration, for a single laboratory calibration instrument Source/Read thermocouple (13), RTD (9), Voltage,
.
unmatched in versatility, performance, and value. As
Current, Pressure (read only)
Custom RTD and SPRT profiles
..
with every Martel calibrator, the 3001's world-class
Nine (9) setpoints for each output range and type
performance and features are accessed through a very
Beryllium-Copper binding posts reduce thermal
.
simple-to-use, intuitive user interface. The Martel 3001 EMFs
is truly a "process calibration laboratory in a box." RS232, USB and IEEE-488 remote control
Compatible with Fluke Met/Cal® software
-
Isolated measurement channel
-
Need similar performance for
Two (2) voltage ranges: 10V and 100 V DC
-
MilliAmp range 0 to 52 mA
voltage and current only?
If your calibration needs are limited to voltage and current,
MilliAmp range with simultaneous
-
the Martel M2000 Bench Calibrator provides all of the
24 VDC power
-
performance, functionality, and ease of use of the 3001.
Selectable 250 Ohm HART™ resistor
Accuracy of 0.005% of reading on
voltage ranges
Need similar performance
for voltage and current only?
If your calibration needs are limited to voltage and current, the
Martel M2000 Bench Calibrator provides all of the performance,
functionality, and ease of use of the 3001.