Phcond 3320
Phcond 3320
ba77054d03 03/2018
332
0
pH/Cond 3320
METER FOR 2 SENSORS (PH/ORP/COND)
pH/Cond 3320
2 ba77054d03 03/2018
pH/Cond 3320 Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 pH/Cond 3320 meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 Safety information in the operating manual . . . . . 9
2.1.2 Safety signs on the meter . . . . . . . . . . . . . . . . . . 9
2.1.3 Further documents providing safety information . 9
2.2 Safe operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2 Requirements for safe operation . . . . . . . . . . . . 10
2.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . . 12
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 General operating principles . . . . . . . . . . . . . . . . . . . . . . 13
4.1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.1.3 Status information (meter) . . . . . . . . . . . . . . . . . 14
4.1.4 Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 16
4.4.2 Measured value display . . . . . . . . . . . . . . . . . . . 17
4.4.3 Menus and dialogs . . . . . . . . . . . . . . . . . . . . . . . 17
4.4.4 Elements in menus and dialogs . . . . . . . . . . . . . 17
4.4.5 Navigation example 1:Setting the language. . . . 18
4.4.6 Example 2 on navigation: Setting the date
and time ............................ 20
4.5 Channel display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.5.1 Display of several sensors in the measuring
mode ............................ 22
ba77054e03 03/2018 3
Contents pH/Cond 3320
5 pH value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1.1 Measuring the pH value . . . . . . . . . . . . . . . . . . . 23
5.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 24
5.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.2 When do you have to calibrate? . . . . . . . . . . . . . 25
5.2.3 Automatic calibration (AutoCal) . . . . . . . . . . . . . 25
5.2.4 Manual calibration (ConCal) . . . . . . . . . . . . . . . . 28
5.2.5 Calibration points . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.6 Calibration data. . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.2.7 Continuous measurement control (CMC
function) ............................. 35
6 ORP voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.1.1 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . . 36
6.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 37
6.2 ORP calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7 Ion concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.1.1 Measuring the ion concentration . . . . . . . . . . . . 38
7.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 39
7.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.2 When to calibrate? . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.3 Calibration (ISE Cal) . . . . . . . . . . . . . . . . . . . . . . 40
7.2.4 Calibration standards . . . . . . . . . . . . . . . . . . . . . 43
7.2.5 Calibration data. . . . . . . . . . . . . . . . . . . . . . . . . . 43
8 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.1.1 Measuring the conductivity . . . . . . . . . . . . . . . . . 45
8.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 46
8.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.2.2 When to calibrate? . . . . . . . . . . . . . . . . . . . . . . . 47
8.2.3 Determining the cell constant (calibration in
control standard) ...................... 47
8.2.4 Calibration data. . . . . . . . . . . . . . . . . . . . . . . . . . 48
9 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
9.1 Measurement settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
9.1.1 Settings for pH measurements . . . . . . . . . . . . . . 50
9.1.2 Buffer sets for calibration . . . . . . . . . . . . . . . . . . 51
4 ba77054e03 03/2018
pH/Cond 3320 Contents
10 Data memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
10.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
10.2 Automatic data storage at intervals .............. 67
10.3 Measurement data memory . . . . . . . . . . . . . . . . . . . . . . 69
10.3.1 Editing the measurement data memory . . . . . . . 69
10.3.2 Erasing the measurement data memory . . . . . . 71
10.3.3 Measurement dataset . . . . . . . . . . . . . . . . . . . . 71
10.3.4 Memory locations . . . . . . . . . . . . . . . . . . . . . . . . 71
13 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
13.1 pH/ORP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
13.2 ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
13.3 Conductivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
13.4 General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
ba77054e03 03/2018 5
Contents pH/Cond 3320
14 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
14.1 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 82
14.1.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
14.1.2 ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
14.1.3 Cond . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
14.2 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
15 Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
16 Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
17 Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
6 ba77054e03 03/2018
pH/Cond 3320 Overview
1 Overview
1.1 pH/Cond 3320 meter
The pH/Cond 3320 meter enables you to perform measurements (pH, U, ISE,
conductivity) quickly and reliably.
The pH/Cond 3320 provides the maximum degree of operating comfort, reli-
ability and measuring certainty for all applications.
332
0
1 Keypad
2 Display
3 Connectors
1.2 Sensors
A measuring system ready to measure consists of the pH/Cond 3320 meter
and a suitable sensor.
ba77054e03 03/2018 7
Overview pH/Cond 3320
8 ba77054e03 03/2018
pH/Cond 3320 Safety
2 Safety
2.1 Safety information
WARNING
indicates a possibly dangerous situation that can lead to seri-
ous (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not fol-
lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
ba77054e03 03/2018 9
Safety pH/Cond 3320
This meter is authorized exclusively for pH, ISE, ORP and conductivity
measurements in a field and laboratory environment.
This meter is authorized exclusively for pH, ISE and ORP measurements in a
field and laboratory environment.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 14 TECHNICAL DATA, page 82).
Any other use is considered unauthorized.
10 ba77054e03 03/2018
pH/Cond 3320 Commissioning
3 Commissioning
3.1 Scope of delivery
MeterpH/Cond 3320
4 batteries 1.5 V Mignon type AA
Short instructions
CD-ROM with
– USB drivers
– detailed operating manual
– Software MultiLab Importer
ba77054e03 03/2018 11
Commissioning pH/Cond 3320
1 Screws
2 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor-
rectly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
12 ba77054e03 03/2018
pH/Cond 3320 Operation
4 Operation
4.1 General operating principles
4.1.1 Keypad
ba77054e03 03/2018 13
Operation pH/Cond 3320
4.1.2 Display
4
3 5
pH
2
6.093
25.0 °C 6
1 AutoCal TEC 7
HOLD AR 8
15.02.2014
14:15
USB output 9
14 ba77054e03 03/2018
pH/Cond 3320 Operation
4.1.4 Connectors
5 1 2 4 3
1 pH/ISE electrode
2 Temperature sensor
3 Conductivity measuring cell
4 Mini USB-B interface
5 Service interface
CAUTION
Only connect sensors to the meter that cannot return any volt-
ages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors fulfill these conditions.
pH
6.949
25.0 °C
15.02.2014
Menu 14:15
ba77054e03 03/2018 15
Operation pH/Cond 3320
Automatic shut-off The instrument has an automatic shut-off function in order to save the batteries
function (see section 9.2.1 SYSTEM, page 61). The automatic shut-off function switches
off the meter if no key is pressed for an adjustable period.
The automatic shut-off function is not active
if the communication cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
Display illumination The meter automatically switches off the display illumination if no key is
pressed for 30 seconds. The illumination is switched on with the next keystroke
again.
You can also generally switch the display illumination on or off (see section
9.2.1 SYSTEM, page 61).
4.4 Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
Operating Explanation
mode
Measuring The measurement data of the connected sensor are shown
in the measured value display
Calibration The course of a calibration with calibration information, func-
tions and settings is displayed
Storing in The meter stores measuring data automatically or manually
memory
Transmit- The meter transmits measuring data and calibration records
ting data to a USB-B interface automatically or manually.
Setting The system menu or a sensor menu with submenus, set-
tings and functions is displayed
Only those displays and functions are available in the active operating mode
that are currently being required.
16 ba77054e03 03/2018
pH/Cond 3320 Operation
In the measured value display, open the setting menus with <ENTER>. The
current functions of the softkeys are shown on the display.
Use <ENTER> (short pressure) to open the menu for calibration and mea-
surement settings for the displayed measured parameter.
Use <ENTER _
> (long keystroke (approx. 2 s) to open the Storage & config
menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or cali-
bration (see section 4.1.1 KEYPAD, page 13). These functions are not available
in other operating situations.
The menus for settings and dialogs in procedures contain further subelements.
To select a subelement, use the <>< > keys. The current selection is dis-
played with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <M> to return to the measured value display.
Submenus
The name of the submenu is displayed at the upper edge of the frame. Sub-
menus are opened by confirming with <ENTER>. Example:
System
General
Interface
Clock
Service information
Reset
15.02.2014
Back
14:15
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <ENTER>. Subsequently,
the setting can be changed with <>< > and <ENTER>. Example:
ba77054e03 03/2018 17
Operation pH/Cond 3320
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with <ENTER>.
Example: Display the Calibration record function.
pH
Calibration record
Calibration data storage
Buffer: TEC
One point calibration: Yes
Calibration interval: 7d
Unit for slope: mV/pH
Messages
Information is marked by the i symbol. It cannot be selected. Example:
pH
Calibration record
Calibration data storage
Buffer: TEC
One point calibration: Yes
Calibration interval: 7d
Unit for slope: mV/pH
18 ba77054e03 03/2018
pH/Cond 3320 Operation
pH
6.949
25.0 °C
15.02.2014
Menu
14:15
15.02.2014
Back
14:15
System
General
Interface
Clock
Service information
Reset
Back 15.02.2014
14:15
ba77054e03 03/2018 19
Operation pH/Cond 3320
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
The meter has a clock with a date function. The date and time are shown in the
measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the fol-
lowing functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
After a fall of the supply voltage (empty batteries), the date and time
are reset.
20 ba77054e03 03/2018
pH/Cond 3320 Operation
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
Clock
Date format: dd.mm.yyyy
Date: 15.02.2014
Time: 14:15:25
15.02.2014
Back
14:15
4. Change and confirm the setting with <>< > and <ENTER>.
The minutes are highlighted.
5. Change and confirm the setting with <>< > and <ENTER>.
The seconds are highlighted.
6. Change and confirm the setting with <>< > and <ENTER>.
The time is set.
7. If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8. To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
ba77054e03 03/2018 21
Operation pH/Cond 3320
Channel display
Display of the plug position
pH for the respective parameter
The filled bar indicates for
6.949 each connected sensor to
which plug position of the
24.8 °C meter it is connected.
15.02.2014
14:15 USB output
The measured values of the connected sensors can be displayed in the follow-
ing ways:
Clear display of all connected sensors
Detailed display of one sensor
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
pH 6.966 pH
25.0 °C 6.966
ϰ 0.470 1/cm 25.0 °C
25.0 °C
15.02.2014 15.02.2014
USB output USB output
14:15 14:15
22 ba77054e03 03/2018
pH/Cond 3320 pH value
5 pH value
5.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
pH
6.949
24.8 °C
15.02.2014
Menu 14:15
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 9.2.3 AUTOMATIC STABILITY CONTROL,
page 62).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value
ba77054e03 03/2018 23
pH value pH/Cond 3320
- 1 °C Manual
24 ba77054e03 03/2018
pH/Cond 3320 pH value
5.2 Calibration
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see 9.1.1 SETTINGS FOR PH MEASUREMENTS, PAGE 50).
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with Technical buffers (TEC) is described. When other buffer
sets are used, other nominal buffer values are displayed. Apart from that, the
procedure is identical.
ba77054e03 03/2018 25
pH value pH/Cond 3320
pH Buffer 1
-180.0 mV
24.8 °C
AutoCal TEC
15.02.2014
14:15
pH Buffer 10.011
-180.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
9. Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
26 ba77054e03 03/2018
pH/Cond 3320 pH value
pH Buffer 7.000
0.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
15. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
16. If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
Continuing with
17. Thoroughly rinse the electrode with deionized water.
three- to five-point
calibration 18. Immerse the electrode in the next buffer solution.
19. When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
20. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
ba77054e03 03/2018 27
pH value pH/Cond 3320
pH Buffer 4.010
180.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
21. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).
22. If necessary, use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is
correctly selected (see section 9.1.1 SETTINGS FOR PH MEASUREMENTS, page
50).
Use any one to five buffer solutions in ascending or descending order.
28 ba77054e03 03/2018
pH/Cond 3320 pH value
pH Buffer 1
0.0 mV
24.8 °C
ConCal
15.02.2014
14:15
pH Buffer 7.000
0.0 mV
24.8 °C
ConCal
AR
15.02.2014
14:15
9. Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the setting of the nominal buffer value appears.
pH Buffer 1
7.000
24.8 °C
ConCal
15.02.2014
14:15
10. Set the nominal buffer value for the measured temperature with
< >< >.
ba77054e03 03/2018 29
pH value pH/Cond 3320
pH Buffer 2
4.035
24.8 °C
ConCal
15.02.2014
14:15
18. Set the nominal buffer value for the measured temperature with
< >< >.
19. Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
20. Finish the calibration procedure as a two-point calibration with <M>.
The calibration record is displayed.
30 ba77054e03 03/2018
pH/Cond 3320 pH value
Continuing with
21. Thoroughly rinse the electrode with deionized water.
three- to five-point
calibration 22. Immerse the electrode in the next buffer solution.
23. When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
24. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value appears.
pH Buffer 3
9.958
24.8 °C
ConCal
15.02.2014
14:15
26. Set the nominal buffer value for the measured temperature with
< >< >.
27. Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
28. Use <M> to finish calibration or
Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
ba77054e03 03/2018 31
pH value pH/Cond 3320
You can display the slope in the units, mV/pH or % (see section
9.1.1 SETTINGS FOR PH MEASUREMENTS, page 50).
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <ENTER> / Calibration / Calibration record. To open it, press the <CAL _ > key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>/[USB output] key.
Displaying the cali- The calibration records of the last calibrations are to be found in the menu,
bration data memory <ENTER> / Calibration / Calibration data storage.
32 ba77054e03 03/2018
pH/Cond 3320 pH value
Calibration evalua- After calibrating, the meter automatically evaluates the calibration. The zero
tion point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
ba77054e03 03/2018 33
pH value pH/Cond 3320
Calibration record
pH/Cond 3320
(example) Ser. no. 11292113
CALIBRATIONpH
15.02.2014 15:55
AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV
Voltage 2 3.0 mV
Voltage 3 -177.0 mV
Temperature 1 24.0 °C
Temperature 2 24.0 °C
Temperature 3 24.0 °C
Slope -60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++
etc...
34 ba77054e03 03/2018
pH/Cond 3320 pH value
1
2
pH 0 14
6.093
3
4
25.0 °C
15.02.2014
USB output
14:15
The limits of the calibrated range are determined by the buffers used for cali-
bration:
ba77054e03 03/2018 35
ORP voltage pH/Cond 3320
6 ORP voltage
6.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
U
157.0 mV
24.8 °C
15.02.2014
Menu 14:15
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 9.2.3 AUTOMATIC STABILITY CONTROL,
page 62).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value within the monitored time interval.
36 ba77054e03 03/2018
pH/Cond 3320 ORP voltage
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <AR> or <M>.
The HOLD function is switched off.
The [HOLD] status display disappears.
- 1 °C Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
ba77054e03 03/2018 37
Ion concentration pH/Cond 3320
7 Ion concentration
7.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
ISE
0.157 mg/l
25.0 °C
Ag
15.02.2014
Menu 12:12
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 9.2.3 AUTOMATIC STABILITY CONTROL,
38 ba77054e03 03/2018
pH/Cond 3320 Ion concentration
page 62).
Criteria The AutoRead criteria affect the reproducibility of the measured values. The
following criteria can be adjusted:
high: highest reproducibility
medium: medium reproducibility
low: lowest reproducibility
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
- 1 °C Manual
ba77054e03 03/2018 39
Ion concentration pH/Cond 3320
7.2 Calibration
40 ba77054e03 03/2018
pH/Cond 3320 Ion concentration
ISE Standard 1
0.0 mV
24.8 °C
Ag 0.010 mg/l
15.02.2014
12:12
0.0 mV
24.8 °C
AR
15.02.2014
12:12
11. Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The calibration display for the next standard solution appears.
ba77054e03 03/2018 41
Ion concentration pH/Cond 3320
ISE Standard 2
0.0 mV
24.8 °C
Ag 0.020 mg/l
15.02.2014
12:12
8.9 mV
24.8 °C
AR
15.02.2014
12:12
16. Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <ENTER>.
The calibration display for the next standard solution appears.
ISE Standard 3
20.5 mV
24.8 °C
Ag 0.050 mg/l
15.02.2014
12:12
42 ba77054e03 03/2018
pH/Cond 3320 Ion concentration
Continuing with Repeat the steps 12 to 17 in the same way with the third and further standard
three- to seven-point solutions as necessary. The new calibration values are displayed after the last
calibration calibration step was completed.
Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <ENTER> / Calibration / Calibration record. To open it, press the <CAL _ > key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>/[USB output] key.
ba77054e03 03/2018 43
Ion concentration pH/Cond 3320
Displaying the cali- The calibration records of the last calibrations are available in the menu,
bration data memory <ENTER> / Calibration / Calibration data storage.
Calibration evalua- After calibrating, the meter automatically evaluates the calibration.
tion
Display Calibration Magnitude of the slope [mV]
record
+++ 50.0 ... 70.0 or 25.0 ... 35.0
CALIBRATIONISE
18.01.2013 08:09:10
44 ba77054e03 03/2018
pH/Cond 3320 Conductivity
8 Conductivity
8.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
The selection of the measuring cell and the setting of the cell con-
stant is done in the measurement settings menu for conductivity
(see section 9.1.6 SETTINGS FOR CONDUCTIVITY MEASURING CELLS,
page 57). The cell constant to be set must either be taken from the
operating manual of the measuring cell or is printed on the measur-
ing cell.
ϰ
1417 µS/cm
25.0 °C
[Tr25][nLF][ 0.470 1/cm]
15.02.2014
USB output
08:00
ba77054e03 03/2018 45
Conductivity pH/Cond 3320
The factor to calculate the total dissolved solids is set to 1.00 in the
factory. You can adjust this factor to meet your requirements in the
range 0.40 ... 1.00. The factor is set in the Measurement menu for
the parameter, TDS.
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 9.2.3 AUTOMATIC STABILITY CONTROL,
page 62).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
46 ba77054e03 03/2018
pH/Cond 3320 Conductivity
8.2 Calibration
Aging slightly changes the cell constant, e.g. due to coatings. As a result, an
inexact measured value is displayed. The original characteristics of the cell can
often be restored by cleaning the cell. Calibration determines the current value
of the cell constant and stores this value in the meter.
Thus, you should calibrate at regular intervals (we recommend: every 6
months).
You can determine the actual cell constant of the conductivity measuring cell
by calibrating with the control standard in the following ranges:
In the default condition, the calibrated cell constant of the meter is set to
0.475 cm-1 (conductivity sensor TetraCon 325).
For this calibration procedure, the Measuring cell setting must be set to cal in
the Measurement menu. Proceed as follows to determine the cell constant:
ba77054e03 03/2018 47
Conductivity pH/Cond 3320
ϰ
0.479 1 /cm
25.1 °C
15.02.2014
08:00
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <ENTER> / Calibration / Calibration record. To open it, press the <CAL _ > key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>[USB output] key.
Displaying the cali- The calibration records of the last calibration procedures are available in the
bration data memory <ENTER>/Calibration / Calibration data storage and <ENTER>/Storage &
config / Data storage / Calibration data storage menu.
48 ba77054e03 03/2018
pH/Cond 3320 Conductivity
or
Data storage /
Calibration data stor-
age / Output to USB
Calibration evalua- After calibration, the meter automatically evaluates the current status of the cal-
tion ibration. The evaluation appears on the display and in the calibration record.
Calibration record
pH/Cond 3320
(example) Ser. no. 11292113
CALIBRATION Cond
15.02.2014 16:13:33
ba77054e03 03/2018 49
Settings pH/Cond 3320
9 Settings
9.1 Measurement settings
The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <ENTER> key. After completing the settings, switch to the measured
value display with <M>.
In the following table, only those settings are listed that concern the pH mea-
surement.
Default settings are printed in bold.
50 ba77054e03 03/2018
pH/Cond 3320 Settings
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during calibration.
ba77054e03 03/2018 51
Settings pH/Cond 3320
52 ba77054e03 03/2018
pH/Cond 3320 Settings
* Brand names or trade names are trademarks of their respective owners protected
by law.
ba77054e03 03/2018 53
Settings pH/Cond 3320
Setting the calibra- The calibration interval is set to 7 days in the factory.
tion interval You can change the interval (1 ... 999 days):
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <ENTER>key. After completing the settings, switch to the measured
value display with <M>.
In the following table, only those settings are listed that influence the ORP mea-
surement.
Default settings are printed in bold.
54 ba77054e03 03/2018
pH/Cond 3320 Settings
The settings for ISE measurements are made in the menu for calibration and
measurement settings of the ISE measurement. To open the settings, display
the required measured parameter in the measured value display and press the
<ENTER> key. After completing the settings, switch to the measured value dis-
play with <M>.
In the following table, only those settings are listed that influence the ISE mea-
surement:
ba77054e03 03/2018 55
Settings pH/Cond 3320
56 ba77054e03 03/2018
pH/Cond 3320 Settings
The settings are made in the Measurement menu for the measured parameter,
conductivity. To open the settings, display the required measured parameter in
the measured value display and press the <ENTER> key. After completing the
settings, switch to the measured value display with <M>.
Default settings are printed in bold.
ba77054e03 03/2018 57
Settings pH/Cond 3320
58 ba77054e03 03/2018
pH/Cond 3320 Settings
Setting the calibra- The calibration interval is set to 150 days in the factory.
tion interval You can change the interval (1 ... 999 days):
ba77054e03 03/2018 59
Settings pH/Cond 3320
Application tips Set the temperature compensation suitable for your test sample:
The factor to calculate the total dissolved solids is set to 1.00 in the delivery
condition.
You can adjust this factor to meet your requirements in the range 0.40 ... 1.00.
The factor is set in the menu for the parameter TDS (Measurement / TDS fac-
tor).
60 ba77054e03 03/2018
pH/Cond 3320 Settings
9.2.1 System
To open the Storage & config menu, press the <ENTER _ > key in the mea-
sured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
ba77054e03 03/2018 61
Settings pH/Cond 3320
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see section 10 DATA MEMORY, page 66).
You can activate or switch off the automatic Stability control function (see sec-
tion 9.2.1 SYSTEM, page 61).
Manually starting Irrespective of the setting for automatic Stability control (see section 9.2.1 SYS-
the stability control TEM, page 61) in the System menu, you can start the Stability control function
manually at any time.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
62 ba77054e03 03/2018
pH/Cond 3320 Settings
9.3 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings
with the Reset function:
ba77054e03 03/2018 63
Settings pH/Cond 3320
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <ENTER> (or <F1>/[Menu]) key.
Cond The following settings for conductivity measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <ENTER> (or <F1>/[Menu]) key.
64 ba77054e03 03/2018
pH/Cond 3320 Settings
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu in the measured value display, press the
<ENTER _> (or <F1 _ >/[Menu]) key.
ba77054e03 03/2018 65
Data memory pH/Cond 3320
10 Data memory
You can transmit measured values (datasets) to the data memory:
Manual data memory (see section 10.1 MANUAL STORAGE, page 66)
Automatic storage at intervals (see section 10.2 AUTOMATIC DATA STORAGE
AT INTERVALS, page 67)
With each data storage process, the current datasets of the sen-
sors indicated on the display are transmitted to the interface at the
same time.
15.02.2014 11:24:16
pH 7.000 24.8 °C AR +++
ID number: 1
Continue
15.02.2014
Back
14:15
66 ba77054e03 03/2018
pH/Cond 3320 Data memory
If the memory is full The following window appears if all 500 storage locations are occupied:
Warning
Yes
No
Back 15.02.2014
14:15
Configuring the
automatic memory
1. Press the <STO _
> key.
The menu for automatic data storing appears.
function
ID number 1
1
Interval 30 s
Duration 180 min
Continue
0d03h00min 2
0 1d17h33m
in 15.02.2014
Back 14:15
3
ba77054e03 03/2018 67
Data memory pH/Cond 3320
Settings You can configure the automatic data storing function with the following
settings:
Starting the auto- To start the automatic storing function, select Continue with <>< > and con-
matic storing func- firm with <ENTER>. The meter switches to the measured value display.
tion
pH 0 14
1
6.949
25.0 °C
0d03h00min
2
15.02.2014
Eco mode 14:15
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.
Energy saving mode If the automatic storing function is active, the meter provides an energy saving
([Eco mode]) mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
68 ba77054e03 03/2018
pH/Cond 3320 Data memory
saving mode switches off functions of the meter that are not required for the
automatic storage of measurement data (such as the display). By pressing any
key the energy saving mode is switched off again.
Terminating the Proceed as follows to switch off the automatic data storing function before the
automatic memory adjusted storing duration has expired:
function prema-
turely
1. Press the <STO _ > key.
The following window appears.
Warning
Yes
No
Back 15.02.2014
14:15
The settings are explained here using the manual data memory as
an example. The same settings and functions are available for the
automatic data memory.
ba77054e03 03/2018 69
Data memory pH/Cond 3320
15.02.2014
Back
14:15
15.02.2014 10:56:20
pH/Cond 3320
Ser. no. 08502113
ID number 2
________________________________________
etc...
Quitting the display To quit the display of stored measurement datasets, you have the following
70 ba77054e03 03/2018
pH/Cond 3320 Data memory
options:
Switch directly to the measured value display with <M>.
Quit the display and move to the next higher menu level with <F1>/[Back].
Erasing the measurement data memory (see section 10.3.1 EDITING THE
MEASUREMENT DATA MEMORY, page 69).
The pH/Cond 3320 meter has two measurement data memories. The
measured values recorded either manually or automatic are stored separately
in individual measurement data memories.
ba77054e03 03/2018 71
Transmitting data (USB interface) pH/Cond 3320
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measure-
ment datasets, calibration records).
72 ba77054e03 03/2018
pH/Cond 3320 Transmitting data (USB interface)
11.2 Connecting a PC
Connect the pH/Cond 3320 to the PC via the USB interface.
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
Installation of the System requirements of the PC for installation of the USB driver:
USB driver on the
PC
1. Insert the supplied installation CD in the CD drive of your PC.
2. Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3. Connect the pH/Cond 3320 to the PC via the USB interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
ba77054e03 03/2018 73
Maintenance, cleaning, disposal pH/Cond 3320
1 Screws
2 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor-
rectly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
74 ba77054e03 03/2018
pH/Cond 3320 Maintenance, cleaning, disposal
12.2 Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lint-
free cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain sol-
vents. Remove any splashes immediately.
12.3 Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
12.4 Disposal
At the end of its operational lifetime, the meter must be returned to the
disposal or return system statutory in your country. If you have any questions,
please contact your supplier.
ba77054e03 03/2018 75
Maintenance, cleaning, disposal pH/Cond 3320
12.4.1 ISE
12.4.2
76 ba77054e03 03/2018
pH/Cond 3320 What to do if...
13 What to do if...
13.1 pH/ORP
ba77054e03 03/2018 77
What to do if... pH/Cond 3320
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Temper if necessary
78 ba77054e03 03/2018
pH/Cond 3320 What to do if...
13.2 ISE
ba77054e03 03/2018 79
What to do if... pH/Cond 3320
13.3 Conductivity
80 ba77054e03 03/2018
pH/Cond 3320 What to do if...
ba77054e03 03/2018 81
Technical data pH/Cond 3320
14 Technical data
14.1 Measuring ranges, resolution, accuracy
14.1.1 pH/ORP
82 ba77054e03 03/2018
pH/Cond 3320 Technical data
14.1.2 ISE
ba77054e03 03/2018 83
Technical data pH/Cond 3320
14.1.3 Cond
84 ba77054e03 03/2018
pH/Cond 3320 Technical data
ba77054e03 03/2018 85
Technical data pH/Cond 3320
86 ba77054e03 03/2018
pH/Cond 3320 Firmware update
15 Firmware update
Available firmware updates are provided on the Internet.
With the firmware update program and a PC you can update the firmware of the
pH/Cond 3320 to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the pH/Cond 3320).
After switching the meter off and on you can check whether the meter has taken
over the new software version (see YOU WANT TO KNOW WHICH SOFTWARE
VERSION IS IN THE METER, PAGE 81).
ba77054e03 03/2018 87
Glossary pH/Cond 3320
16 Glossary
pH/ORP/ISE
Conductivity
88 ba77054e03 03/2018
pH/Cond 3320 Glossary
Temperature compen- Name of a function that considers the temperature influence on the
sation measurement and converts it accordingly. Depending on the measured
parameter to be determined, the temperature compensation functions
in different ways. For conductimetric measurements, the measured
value is converted to a defined reference temperature. For potentiomet-
ric measurements, the slope value is adjusted to the temperature of the
test sample but the measured value is not converted.
General topics
Adjusting To manipulate a measuring system so that the relevant value (e.g. the
displayed value) differs as little as possible from the correct value or a
value that is regarded as correct, or that the difference remains within
the tolerance.
AutoRange Name of the automatic selection of the measuring range.
Calibration Comparing the value from a measuring system (e.g. the displayed
value) to the correct value or a value that is regarded as correct. Often,
this expression is also used when the measuring system is adjusted at
the same time (see adjusting).
Channel A channel is a display indication that corresponds to a physical connec-
tion on the meter.
Measured parameter The measured parameter is the physical dimension determined by
measuring, e.g. pH, conductivity or D.O. concentration.
Measured value The measured value is the special value of a measured parameter to be
determined. It is given as a combination of the numerical value and unit
(e.g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
Reset Restoring the original condition of all settings of a measuring system.
Resolution Smallest difference between two measured values that can be dis-
played by a meter.
Stability control (Au- Function to control the measured value stability.
toRead )
ba77054e03 03/2018 89
Glossary pH/Cond 3320
Standard solution The standard solution is a solution where the measured value is known
by definition. It is used to calibrate a measuring system.
Temperature function Name of a mathematical function expressing the temperature behavior
of a test sample, a sensor or part of a sensor.
Test sample Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and
original sample are identical if the test sample was not processed.
90 ba77054e03 03/2018
pH/Cond 3320 Index
17 Index
E T
Energy saving mode . . . . . . . . . . . . . . . . . 68 TDS factor . . . . . . . . . . . . . . . . . . . . . . . . . 60
Temperature compensation . . . . . . . . . . . 59
Temperature measurement
F
ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Firmware update . . . . . . . . . . . . . . . . . . . . 87 pH . . . . . . . . . . . . . . . . . . . . . . . . . 24, 37
Three-point calibration
I ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Initial commissioning . . . . . . . . . . . . . . . . . 11 pH . . . . . . . . . . . . . . . . . . . . . . . . . 27, 31
Initialize . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Transmitting data . . . . . . . . . . . . . . . . . . . 72
Interval for calibration . . . . . . . . . . . . . . 53, 59 Automatic . . . . . . . . . . . . . . . . . . . . . . 72
Manual . . . . . . . . . . . . . . . . . . . . . . . . 72
K Transmitting measured values . . . . . . . . . 72
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Two-point calibration
ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
M pH . . . . . . . . . . . . . . . . . . . . . . . . . 26, 30
Measured value display . . . . . . . . . . . . . . . 17
Measurement accuracy . . . . . . . . . . . . 54, 59
ba77054e03 03/2018 91
Index pH/Cond 3320
Z
Zero point of pH electrode . . . . . . . . . . . . . 25
92 ba77054e03 03/2018
What can Xylem do for you?
We're a global team unified in a common purpose: creating innovative solutions to
meet our world's water needs. Developing new technologies that will improve the way
water is used, conserved, and re-used in the future is central to our work. We move,
treat, analyze, and return water to the environment, and we help people use water
efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we
have strong, long-standing relationships with customers who know us for our powerful
combination of leading product brands and applications expertise, backed by a legacy
of innovation.
Service address:
Xylem Analytics Germany
Sales GmbH & Co. KG
WTW
Dr.-Karl-Slevogt-Str. 1
82362 Weilheim
Germany