Interface sheet - BTL 08 PC ECG CPET Treadmill -
Communication Card
Service documentation
Lode BV
Zernikepark 16
NL 9747 AN Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02
@: ask@lode.nl
www.lode.nl
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02   2/24
                          Table of contents
1 - List of symbols used                                        4
2 - Validity                                                    4
3 - BTL08-PC Setup instructions                                 5
   3.1 - Cable slection                                         5
   3.2 - Software setup                                         5
   3.3 - Ergometer setup                                        7
4 - Setting up the hardware                                     8
   4.1 - Connection overview                                    8
   4.2 - Connecting with the usb cable                          9
   4.3 - Connecting with the RS232 cable                        9
5 - Treadmill or ergometer with 3.5" touch screen unit         10
6 - Treadmill or ergometer with 7" touch screen control unit   15
7 - Nummerical Control Unit - Treadmill only                   19
8 - Treadmill or Bicycle without control unit                  21
   8.1 - Device without control unit                           21
1 - List of symbols used
         Next
         Previous
2 - Validity
This interface form covers
* All OEM treadmills manufactured by Lode B.V. with the part number 939900/01/02/07/08/09/10
starting from serial #939900-20090001;
* All Katana treadmills 9349XX-201X0001
* All Valiant treadmills starting with serial number 932900-20122XXX and 9329XX-20121XXX
* All Valiant 2 treadmill starting with serial number 9389xx-20140001
If your treadmill is not covered by this interface form please contact Lode service service@lode.nl
This interface form covers
* All OEM treadmills manufactured by Lode B.V. with the part number 939900/01/02/07/08/09/10
starting from serial #939900-20090001;
* All Katana treadmills 9349XX-201X0001
* All Valiant treadmills starting with serial number 932900-20122XXX and 9329XX-20121XXX
* All Valiant 2 treadmill starting with serial number 9389xx-20140001
If your treadmill is not covered by this interface form please contact Lode service service@lode.nl
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02   4/24
3 - BTL08-PC Setup instructions
                                    3.1 - Cable slection
                                    The required cable for the BTL-08 PC is an USB A to B cable (Lode art. nr.
                                    930916) as shown in the picture.
                                    Connect the cable to the ergometer as described in the 'hardware setup'
                                    chapter. Connect the cable to the computer to a free USB port.
                                                                                     3.2 - Software
                                                                                     setup
                                                                                     Start the BTL-08 software;
                                                                                     select'setting' from the right
                                                                                     upper corner of the mainscreen.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02            5/24
                                                                                     The following menu will appear.
                                                                                     Select 'ergometer configuration'.
                                                                                     Select 'Trackmaster' and select
                                                                                     the communication port to
                                                                                     which your ergometer is
                                                                                     connected.
                                                                                     Note that you may need to
                                                                                     connect the ergometer to the
                                                                                     computer and switch it on
                                                                                     before a communication port is
                                                                                     initialized in Windows.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02              6/24
3.3 - Ergometer setup
The following protocols need to be set in the Lode ergometer to communicate with the BTL08-PC.
Treadmill: Trackmaster
The next chapters describe the procedures to setup these protocols depending on the type of
ergometer.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02   7/24
4 - Setting up the hardware
Lode ergometers have two ways of connecting external devices. The cable selection depends on the
connection possibilities of the external device. The previous chapter tells you which cable you need. This
can be a RS232 cable or a USB cable.
4.1 - Connection overview
                                                                                     Connect the cable to the USB
                                                                                     port if you are using a USB
                                                                                     cable.
                                                                                     Please note that you may have
                                                                                     to install the FTDI driver on the
                                                                                     external equipment. This
                                                                                     procedure is only needed if
                                                                                     windows fails to instal the driver
                                                                                     automatically.
                                                                                     You can find the driver on the
                                                                                     Lode website under the Services
                                                                                     tab.
Locations:
- For the Treadmill the communication ports are Located on the front side near the power switch.
- For the bicycle ergometer the communication ports are located on the bottom of the bicycle. Tilt the
ergometer on it's side to be able to acces these ports.
Communication Module:
Use the single RS232 port to connect the communication module to the external equipment.
Legacy communication card (picture above -> 1 * USB, 2 * RS232)
In case you are using a RS232 cable you must use the RS232 port closest
to the USB connector, on the picture indicated with RS232 (1).
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02              8/24
4.2 - Connecting with the usb cable
                                    If there is no other connection available you can also use a USB cable to
                                    connect.
                                    1. You will need an USB A to B cable (Lode part no. 930916) to establish
                                    communication between the Lode ergometer and your stress system or
                                    ergospirometrysystem. See picture left.
                                    2. Connect the cable to the Lode ergometer.
                                    3. Connect the cable to the external device.
4.3 - Connecting with the RS232 cable
Connecting with the RS232 cable
                                    1. Depending on the external device you may need to use a straight
                                    RS232 cable (Lode part no. 930911) or a cross linked cable. The cable
                                    selection chapter specifies which cable is needed.
                                    2. Connect the cable to the Lode ergometer.
                                    3. Connect the cable to the external device.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02       9/24
5 - Treadmill or ergometer with 3.5" touch screen unit
                                                                                     Start the Lode ergometer.
                                                                                     During start up information
                                                                                     about the ergometer, like the
                                                                                     kind of device, the serial
                                                                                     number and the connected
                                                                                     modules is displayed.
                                                                                     After initialisation the terminal
                                                                                     mode is automatically
                                                                                     displayed. Swipe with your
                                                                                     finger over the screen as
                                                                                     indicated with the red arrow in
                                                                                     the picture to go to the main
                                                                                     menu.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02              10/24
                                                                                     Select "Settings" to change
                                                                                     settings of the device.
                                                                                     You can choose your Language
                                                                                     in the "Language" menu. The
                                                                                     menu "Sleep Mode" sets a
                                                                                     screensaver. The menu
                                                                                     "Terminal Settings" lets you set
                                                                                     the communication settings for
                                                                                     external communication. The
                                                                                     menu "Service" is only intended
                                                                                     for service purposes to be
                                                                                     executed by factory trainend
                                                                                     personel.
                                                                                     Select "Terminal Settings" to
                                                                                     change the communication
                                                                                     protocol.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02                11/24
                                                                                     Select the Communication
                                                                                     protocol as desribed in
                                                                                     "Ergometer setup" chapter.
                                                                                     Select "Default" for the
                                                                                     baudrate.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02             12/24
                                                                                     Select "RS232" or "USB"
                                                                                     depending in which cable you
                                                                                     have connected to the Lode
                                                                                     ergometer.
                                                                                     Select the button with the house
                                                                                     to go back to the main menu.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02            13/24
                                                                                     Select "Terminal" to set the
                                                                                     ergometer in terminal mode.
                                                                                     Your device is now ready for
                                                                                     external control.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02           14/24
6 - Treadmill or ergometer with 7" touch screen control unit
                                                                                     Start the device. Information
                                                                                     about the device is shown on
                                                                                     the display during boot mode.
                                                                                     After boot up the device shows
                                                                                     the Test window or main menu.
                                                                                     To access the menu from the
                                                                                     Test window you need to swipe
                                                                                     the graph area of the touch
                                                                                     screen as indicated in the
                                                                                     picture.
                                                                                     *Note that for this instruction
                                                                                     the graphic view for a Lode
                                                                                     Corival is used. The view for a
                                                                                     treadmill might difer a bit but
                                                                                     the procedure is the same.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02             15/24
                                                                                     Select "settings" to go to the
                                                                                     settings menu.
                                                                                     Scroll down using the red
                                                                                     marked scroll button and select
                                                                                     "Terminal".
                                                                                     Select the communication
                                                                                     protocol ad described in the
                                                                                     "Ergometer setup" chapter.
                                                                                     Use the scroll button to view the
                                                                                     complete list.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02                 16/24
                                                                                     Select "Default" for the
                                                                                     baudrate.
                                                                                     Select "RS232"or "USB"
                                                                                     depending on the type of cable
                                                                                     that you connected to the Lode
                                                                                     ergometer.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02           17/24
                                                                                     Press the home button on the
                                                                                     left bottom corner and select
                                                                                     "Terminal". Your ergometer is
                                                                                     now ready for external
                                                                                     communication.
                                                                                     *optional*
                                                                                     Your ergometer can be set to
                                                                                     boot up in terminal mode.
                                                                                     Follow these steps to activate
                                                                                     this function:
                                                                                     - Go to the settings menu
                                                                                     - Select "Default Start Menu"
                                                                                     - Select "Terminal"
                                                                                     - Switch the ergometer off and
on. It will boot directly into terminal mode.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02            18/24
7 - Nummerical Control Unit - Treadmill only
                                                                                     1. Press and hold key #1 and #2
                                                                                     as shown on "Figure 1" for 10
                                                                                     seconds.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02           19/24
2. Press the enter key #4 until the display shows terminal 0
3. By using the "+" and "-" keys for the grade adjustment #3 you can change the "0" behind terminal to
"1" to activate the Terminal mode.
4. Press enter #4 and "prot id" will be shown on the display, for Lode protocol this should be set to "1".
Set the protocol as described in the "Ergometer setup" chapter.
The following protocols can be selected
Lode = Lode protocol
trAck = Trackmaster communication protocol
UUood = Woodway communication protocol
HPcoS = HP Cosmos communication protocol
gE = GE protocol
5. Press enter #4 and "baudrate" will be displayed to set the correct baudrate, select "0" for auto select.
The following baudrates can be set:
Baudrate = 0 (Autoselect dependent on protocol-id)
2400;
4800;
9600;
19200;
38400;
57600
6. Press enter #4 and "interface" will be displayed, you can choose the communication port which you
would like to select.
Interface = CIM interface choice. Select "RS232" or "USB" depending on the cable that you connected to
the treadmill.
0 = RS232
1 = RS232-AUX
2 = USB
7. Press enter #4; the next step "Neg Elev" is read only and is "1" when a negative elevation option has
been added to the treadmill, without a negative elevation option this value shows "0".
Neg Elev = Negetive elevation (read-only)
0 = not present
1 = present
Press the stop button #2 on the control unit or wait 10 seconds to leave the settings menu and let your
changes take effect.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02   20/24
8 - Treadmill or Bicycle without control unit
8.1 - Device without control unit
In case the device does not have a control unit the correct protocol can be set from a computer by using
the Lode Service Manager tool that can be downloaded from the Lode website if you are a registered
user. http://lode.nl/en/product/getDownload/469
Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cable from the
computer to the RS232 port
Start the Lode Service Manager and login with the same account details that you use to login on the
website.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02   21/24
                                                                                     1. Select Terminal
                                                                                     2. Choose Setup to select the
                                                                                     COM port on which the device is
                                                                                     connected.
                                                                                     3. Select preset Lode Bus
                                                                                     Service Protocol
                                                                                     4. Press OK
                                                                                     5. Press Break
                                                                                     6. Switch the device off, wait 10
                                                                                     seconds
                                                                                     7. Switch the device on
                                                                                     The screen as shown left should
                                                                                     show up within 5 seconds after
                                                                                     powering up.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02             22/24
                                                                                     A menu is showed. use the
                                                                                     nummerical keys to make your
                                                                                     selection. The next menu will
                                                                                     automatically come up when a
                                                                                     number is pushed.
                                                                                     Select settings (1) and choose
                                                                                     the protocol as described in the
                                                                                     "Ergometer set-up" chapter.
                                                                                     Choose which port you want to
                                                                                     use RS232 (0) or USB (1)
                                                                                     This setting refers to the type of
                                                                                     cable that you have connected
                                                                                     to the Lode ergometer.
                                                                                     Select default for the baudrate.
                                                                                     Note that Default should always
                                                                                     be selected unless different
                                                                                     specified by your dealer.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02              23/24
                                                                                     Select the kmh/kmh (1)
                                                                                     conversion for if you are using
                                                                                     the Trackmaster protocol.
                                                                                     If you are using a protocol other
                                                                                     than Trackmaster select "no
                                                                                     conversion" (0).
                                                                                     Select "Disable" (0) for Global
                                                                                     communication timeout.
                                                                                     Press "Y" to save the new
                                                                                     settings and restart the device.
                                                                                     Your device is now ready for use
                                                                                     with the external controller.
                                                                                     Note that the device is
                                                                                     automatically switched to
                                                                                     terminal mode when no manual
                                                                                     controller is present at the
                                                                                     device itself.
INTERFACE INSTRUCTION - BTL 08 PC ECG CPET TREADMILL - COMMUNICATION CARD / VERSION: 1.0 - 2015-09-02              24/24