MMTCL Manual

The document provides specifications for a load indicator, detailing its input signal, range, display features, and power supply requirements. It also includes installation instructions, operational guidelines for configuration and calibration, and troubleshooting tips for common symptoms and faults. The load indicator is designed for measuring various parameters using strain gauge-based transducers and features a programmable interface for user customization.

Uploaded by

sales
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
10 views10 pages

MMTCL Manual

The document provides specifications for a load indicator, detailing its input signal, range, display features, and power supply requirements. It also includes installation instructions, operational guidelines for configuration and calibration, and troubleshooting tips for common symptoms and faults. The load indicator is designed for measuring various parameters using strain gauge-based transducers and features a programmable interface for user customization.

Uploaded by

sales
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 10

1.

SPECIFICATIONS

LOAD INDICATOR

1. Input signal : 0-15mV D.C

2. Range : 50KG

3. Linearity : +/-0.01% of FS

4. Input impedance : Greater than 10 Meg.ohms

5. Bridge excitation : +5V dc for a max. of 4 load cells.

350 ohms of input impedance.

6. Display : 5 digit,13 mm high intensity Red LED.


1. ACTUAL WEIGHT
2. SET WEIGHT
3. DEVIATION
Output

a. SET POINT : FOR TOWER LAMP


b. CURRENT : 4- 20 mA

7. Power supply : 230V ac +/-10%, 50Hz, 15VA

8. General

Operating temp. : 5 to 50 deg.C.

Physical : metal enclosure with enamel paint finish


2.INSTALLATION

The power supply should be made available with a proper earthing within

2 meters of the indicator. Connect the instrument to ac mains through the

connector provided at the rear of the indicator.

Mains voltage to be used is given in the specification

INSTALLATION:_

Connect the transducer to the equipment through the connector provided

at the back of the indicator. Connect the Tower lamp. Switch on the

Indicator.

Now the equipment is ready to be configured for measurement as given

in the following sections.


3. DESCRIPTION

INDICATOR :

The Load indicator is a field programmable indicator designed to be used with


strain gauge based transducers to measure parameters like load, pressure, torque
etc. It can take an input of 30 mV DC and can read up to a maximum of 60000
counts with selectable resolution of 0 / 5. This is a panel mounting type
equipment.

The main features of the indicator are :

1. Programmable range up to 60000


2. Calibration through keyboard.
3. Taring through keyboard.
4. Potential free contact

The indicator can be programmed through keyboard for range, resolution,


calibration, and
decimal point. The front panel has got a digital display and a keyboard. The
display can
indicate the following.

1. Tare facility - "TArE"


2 Range - " rAnGE "
3. Resolution - "rSn"
4. Decimal point - “dp”
5. Calibration - "CAL-U "
4.OPERATION

System configuration may be carried out using the keyboard.


CONFIGURING THE SYSTEM

RANGE SETTING:
1. Press SET (3)key, and switch on the indicator the display shows “rNG “.
Press  (4) Key.
2. The display shows `00000' with 1st digit blinking 0 may be set to 0 to 9
by using INC (1) increment key.
2. Use the (4) key to move the blinking to the right side digit. Change
the value by using INC (1) key. Continue this step till the required value
is entered.
3. Press (2) key to store the set value of range. Now the instrument
automatically goes to the next programming parameter i.e. resolution.

RESOLUTION SETTING:
The display shows `rSn' Press (4) Key with ‘0’ blinking press
INC (1) key to change the value. Press (2) key to store the set value
display will go to next programming parameter Decimal Point.

DECIMAL POINT SETTING:


Display shows dP Press Press (4) Key with ‘0’ blinking press
INC (1) key to change the value. Press (2) key to store the set value
display will go to next programming parameter Current.
S-20 (CURRENT SETTING)
The display shows S-20 Press Press (4) Key with ‘0’ blinking press
INC (1) key to change the value. Press (2) key to store the set value
display will go to next programming parameter
Calibration CAL-U switch OFF and ON the Indicator

CALIBRATION
1. Press SET (3) and switch ON the indicator.
2. Display will be showing “rNG” (range).
3. Keep pressing (2) till the display shows “CAL-U”
4. Press (4) then display will show pre-programmed “CAL value” with
one digit blinking.
5. Set the required Weight by using INC (1) for Increment (4) for
SHIFT Shifting the digit WITOUT KEEPING ANY WEIGHT ON THE
PLATFORM.
6. Press (2) to store the value. Then Display will show some reading,
TARE the value by pressing INC (1) . Then Display will show “00”.
7. Now keep the weight to be calibrated on the platform.
8. Display will be showing some reading.
(NOTE:- The display should show the reading more than that of the
CALIBRATED value, if not then weight will not be correct)
9. Press (2) now the Display will show the correct weight which we kept
on to the platform.
10.Remove the weight and see that the Display comes back to “00”.

NOTE: CALIBRATION WEIGHT VALUE TO BE ENTERED


WITHOUT KEEPING ANY WEIGHT ON THE PLAT FORM
After calibration Set the SET weight by Selecting the selector switch in SET
Mode in and press INC or DEC switch to set to required value and after
Select Switch to DEVATION mode and and press INC or DEC switch to set
the required value and keep back the Selector switch to SET mode.
Now the displays show the set value and deviation value according the
weight set. And corresponding lamps will glow according to the set weight.
TARE key is provided to make the value to Zero.
Current output is provided for Actual weight.

5.KEY BOARD LAYOUT


TARE

1 2

SET

4 3
SYMPTOMS & FAULTS

SYMPTOMS POSSIBLE FAULT

1.Blinking display or a. Low power supply


unsteady readings b. Bad earthing
c. Loadcell cable not connected in the
same sequence as in the drawing.

2.The indicator display other a. Calibration disturbed.(required


than actual reading, but steady re-calibration)

3.The display shows ‘HELP’ a. Over loaded. The indicated load is


greater than Range + 9 times the
resolution.

1. SPECIFICATION
2. INSTALLATION
3. DESCRIPTION
4. OPERATION
5. KEYBOARD LAYOUT
6. SYMPTOMS AND FAULTS