SBI-505 Indicator
USER MANUAL
Rev. A
Safety Instruction
For safety operation please follow the safety instruction.
WARNING
Settings, Calibration, Inspection and Maintenance should only be
done by a trained person or staff.
WARNING
Please make sure the indicator has sufficient grounding.
WARNING
The indicator is a static sensitive piece of
equipment. Cut off the power during electrical
connections and internal components touched by
hand are prohibited.
Please take all measure of anti-static protection.
Instruction
This indicator is designed for basic weighing function.
Main function
Weighing function:
Zero, tare, G.W, N.W, accumulation, printing, animal -weighing.
kg/lb conversion.
Print format: S.N. G.W N.W Tare. Date, Time
Options:
Pinter
RS232/RS485 serial interface or second display
Technical parameter
Accuracy class 6000 e
Resolution display: 30, 000 ADC: 2,000,000
Zero stability error TK0 < 0.1μV//K
Span stability error TKspn < ± 6 ppm//K
Sensitivity (internal) 0. 3 μV /d
Input voltage -30~30mV DC
Excitation circuit 5 VDC, 4 wire connection,
Maximum connect 6 load cell of 350Ω
AC power AC100~250V
Operation temperature - 10 °C ~ + 40 °C
Operation humidity ≤90%RH
Storage temperature - 40 °C ~ + 70 °C
1
Installation and calibration
Power supply connection
The indicator is powered by A/C adapter that plugs directly into the
“DC” pin at the bottom of the indicator.
Connection of load cell and indicator
The indicator can be connected up to 6 load cell of 350Ω, 4 wire or 6
wire load cell both ok.
Quick disconnect, as below:
PC Board connection
-EXC -IN GND +IN +EXC -EXC
2
Communication interface
RS232 : DB9 Pin
DB9 definition
5 4 3 2 1
9 8 7 6
Pin function and definition as bellows:
DB9 joint Definition Function
2 TXD Sending data
3 RXD Receiving data
5 GND Ground interface
Note: if RS485, The connection pin is 2 and 5 pin.
Basic operation
Power on & off
Press 2 seconds to power on or off the indicator. After self test
it will go to the weighing mode.
Zero operation
1. Initial zero setting
During power up if the weight on the scale is within the initial zero
tolerance, indicator show zero automatically.
2. Manually Zero setting
When the scale is stable and not the negative weight you can zero the
weight within tolerance by press key.
3
Tare operation
During “TARE” operation when gross weight is tared the indicator will
show the Net weight. The “Net” “tared” status light is on. In tare mode,
Press” TARE” key to clear the tare weight and the indicator will show the
gross weight.
Accumulation operation
With Zero on the Display load weight, Press goes to
accumulation mode, ”Total” light is on, display” n 001”, and then the display
goes back to the loaded weight; unload the weight , Display shows zero, load
the second weight again. Press display”n002” then display the
second loaded weight. Repeat it again and again for a maximum 999 times.
Check the accumulation
Press “TOTAL “key and hold it then press “ON/OFF” key, display ”n**”, (it is
the accumulating times) then show total weight. If the total weight is beyond
the display capability, It will show the first 4 digits then the last 4 digits. For
example, the first 4 digits is”0012”, the last 4 digits is”34,56” It means the
actual weight is “1234.56”
EXIT the accumulation function
When the indicator show the last 4 digits, Press hold it, the
indicator show “ clr n”, it means don’t clear the total Weight, Press “PRINT”
key to exit it; if you want to clear total weight Press “ZERO” or “TARE” key,
“clrn” change to “clry” it means clear total weight? Then Press “PRINT” to
clear the total weight and exit accumulating mode.
4
Print
If the weight is stable, press” PRINT” you can print the weight.
COUNT
1.In weighing mode, load the Pieces on the platform scales, Press”
Count” the indicator show” PCS 0” press” Zero” key in the quantity,
press” Print” to confirm it
2. Load the goods on the platform scales, the indicator will show the
quantity.
1. Press” Count” to return to weighing mode.
2. If you want to weigh different goods, at weighing mode, put the
sample on the platform scales, press” Count” the indicator show”0”
Press “Count” hold it and then press “ON/OFF” the indicator show”
PCS 0”, press “Zero” input the sample quantity, press “Print” to
confirm it. Then repeat the step 2 and 3.
Calibration and Parameter setting
Enter setting
There have two methods to enter the setting menu:
1. when the switch “ CAL” is off, press the “PRINT” hold it and
then press” Total” this enters the C08-39 settings.
2. Take out the sealing screw on the back of indicator, move the
sealing switch to second position. Press “PRINT” hold it and then
press “Total” key this enters C01-C39 settings.
Inside View, Calibration Switch
5
The key functions in setting:
Enter
Up
Down
Left
Power switch, exit setting without saving, “Count” exit with
saving changes.
Calibration operation:
According to the second method which can enter setting menu,
C01-C39
step Method of operation display Remark
1 [C01 ] After you enter calibration
mode, it display [C01 ]
2 press [C01 1] Calibration Weight
option:1=kg
2=lb
3 press [C02 ] Set decimal position
press [C02 0]
option:0/1/2/3/4
press or [C02 2]
Select decimal digit
example : two decimal point
over:[C02 2]
6
4 press [C03 ] Set graduation
press [C03 1]
option:1/2/5/10/20/50
press or [C03 5]
Select required graduation
example:graduation 5:[C03 5]
5 press [C04 ] Max capacity
press [0100.00]
press or / [0100.00]
example:max weighing 100kg:
[0100.00]
press [C05 ] Zero calibration
press [C05 0] Option
press [C05 1] 0=no need zero calibration
press [CAL 9] 1=need zero calibration
calibration zero please choose 1 and
。。
。。。。
6 ensure scale is empty and “stable”
[0000.00] light is on
Ensure zero calibration, countdown.
Till show[0.00](example for two
decimal point)。
7 press [C06 ] calibration
press [C06 0] option:
0=No need calibration
press or [C06 1] 1= need calibration
Load weights on scales
press [SPAN ] according to max. capacity.
[0100.00] Suggest close to the max
capacity, at least 10% of max.
capacity.
press or [0080.00]
press [CAL 9] For example: the weights is
…… 80kg
7
[0080.00] As bellows:
[CAL End] Input the 0080.00, count down ,
then indicator shows 0080.00 ,
calibration is over.
If you want to set application
function parameter. Press
“PRINT” if you want to exit
press “Count”
press [C07 ] Default parameters setting
8 press [07 0] option:0=non-restore default
press or [07 1] parameters
1=restore default parameters
Note: after the above
parameters setting finish, please
do not select 1 to avoid losing
new setting parameters.
Application function parameters setting chart
Setting
Function parameters setting and instruction
Item
C08 Options: 0 = no key tone
key tone warning 1 = key tone
tone
option:0=close auto power off
10= power off automatically if no
C09
Automatic change within 10 minute.
Automatic
power off 30= power off automatically if no
power off
change within 30 minute.
60= power off automatically if no
change within 60 minute.
C10 LED Version ONLY:
Power saving
Power saving
setting option:0= close power saving setting
setting
8
3= close display if no change within 3min.
5= close display if no change within 5 min.
LCD Version:
0=Close he backlight
1= backlight when the weight change or
press the keyboard
2=constant backlight
option:0=close hold function
1=Peak hold /2=Data Hold
Hold
C11 Instruction:
function
Hold mode Peak-hold: it shows the max. data,
N.A
mainly application for materials testing,
SBI-505
such as tension and pulling force.
Date-hold: it shows current weight value.
Mainly application for animal weighing.
C12
Kg/lb Kg/lb C12=0 stop kg/lb conversion
conversion conversion C12=1 kg/lb conversion is ok
C13
Upper/lower Upper limit You can set it within the max. capacity limit
limit alarm alarm value
C14
Lower limit
alarm value
Inner Code C15
display Check inner enter C15 to check the inner code
code
C16 Enter C16, you can set the date,
Date from left to right: year/month/day
Date and time
C17 Enter C17, you can set the time from
Time left to right: hour/min./sec.
9
option:0= Close serial interface data
output
1=Continuous sending, connect big
display
C18 2=Print method, connect printer.
Serial interface
3=Command request method ,
data output method
Communication
connect computer.
setting
4=PC continues sending format,
connect computer.
5=PC/ big display continuous
sending format.
option:
C19
Baud rate 0=1200/1=2400/2=4800/3=9600
Fixed 8,none & 1 stop bit
Option:
0= close manually zero setting
1=±1% max capacity
C20
2=±2% max capacity
Manually zero
4=±4% max capacity
range
10=±10% max capacity
20=±20% max capacity
Zero range 100=±100% max capacity
option:0= no initial zero setting
1=±1% max capacity
C21
2=±1% max capacity
Initial zero range
5=±1% max capacity
10=±1% max capacity
20=±1% max capacity
10
Options:
0= close zero tracking
0.5=±0.5d
1.0=±1.0d
C22 2.0=±2.0d
Automatically zero 3.0=±3.0d
tracking range 4.0=±4.0d
5.0=±5.0d
Zero tracking Note: 1. d = division
2. the zero tracking range can
not bigger than manual zero range.
C23 Options:
0= close zero tracking time
Automatically zero 1= 1 second
tracking time 2= 2 seconds
3= 3 seconds
Overload range C24
option:00= close overload range
Overload range
01d~99d
remark:d =division
Negative display C25
Option:0=-9d
Negative display
range 10=10% max. capacity
20=20% max. capacity
50=50% max. capacity
100=100% max. capacity
Standstill time C26 Option:
Standstill time 0= quick 1= medium 2= slow
11
C27 Option:
Standstill range 1= 1d 2=2d 5=5d 10=10d
D= division
Digital filter C28
option:0= close dynamic filter
Dynamic filter
1=1 digital filter strength
Instruction :
2=2 digital filter strength
Dynamic filter is 3=3 digital filter strength
collecting the data 4=4 digital filter strength
filter before loaded 5=5 digital filter strength
weight stable. 6=6 digital filter strength
When loaded
Note : Pls setting dynamic filter
weight easily
shaking (for strength carefully, the No. is
example animal) , bigger, more stable. if the
you can set this loaded weight shake not too
filter to make much. The setting is less than 3
weight display
more stable
C29
option:0=close noise filter
Noise filter
1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength
C30 C30=0 yy.mm.dd
Print time and date C30=1 mm.dd.yy
C30=2 dd.mm.yy
C30=3 yy.mm.dd
Analog output C31 output type C31=0 0~5Vouput
setting C31=1 4~20mA output
NA SBI-505
12
4~20mA current C32 calibrate Refer to 2.5
calibrate current
NA SBI-505
Relay output C33 Relay output C33=0 close relay output
setting C33=1 Open relay output function 1
NA SBI-505 C3=2 Open relay output function2
C33=3 Preserved menu
Muti C34 C34= 0~99 Add. Code
communication Communication
add. NA SBI-505 add.
Wireless C35 C35=0~99 signal
communication
NA SBI-505
Gravity of C36 C36=9.7000~9.9999
calibration
location
Gravity of C37 C37=9.7000~9.9999
destination
Version No. C38
Preserved menu C39
13
Output format
Big display continuous sending format
Output continuous format
S S S S C
C
T W W W X X X X X X X X X X X X K
R
X A B C S
1 2 3 4 5 6
State A
Bits0,1,2
0 1 2 Decimal point position
1 0 0 XXXXXX0
0 1 0 XXXXXXX
1 1 0 XXXXX.X
0 0 1 XXXX.XX
1 0 1 XXX.XXX
Bits3,4 Division
0 1 X1
1 0 X2
State B
BitsS function
Bits0 gross=0, net=1
Bits1 Symbol: positive =0,negative =1
Bits2 Overload(or under zero)=1
Bits3 dynamic=1
14
Bits4 unit:lb=0, kg=1
Bits5 Constant 1
Bits6 Constant 0
State C
Bit2 Bit1 Bit0 unit
0 0 0 Kg or lb
0 0 1 g
0 1 0 t
Bit 3 printing=1
Extend
Bit 4
display=1
Bit 5 Constant 1
Bit 6 Constant 0
Computer continuous sending format
, , CR LF
S1 S2 S3 Data S4
S1: weight status, ST= standstill, US= not standstill, OL= overload
S2: weight mode, GS=gross mode, NT=net mode
S3: weight of positive and negative, “+” or ” –“
S4: “kg” or “lb”
Data: weight value, including decimal point
CR: carriage return
LF: line feed
Serial interface reception command:
15
RS232COM serial interface can receive simple ASCII command.
Command word and role as follows:
Command NAME Function
T TARE Save and clear tare
Z ZERO Zero gross weight
P PRINT Print the weight
R G.W/N.W Read gross weight or net weight
C Kg/lb Kg/lb conversion
G G.W Check gross weight at net weight mode
R command receive data format
Print format
ID.NO. 004 (Serial No.)
Date: XX.XX. XX (yy.mm.dd)
Time: XX.XX.XX (hh.mm.ss)
GROSS 8.88kg (gross weight)
TARE 2.88kg (tare)
NET 6.00kg (net weight)
PC or Big display continuous sending format
16
17
6. Maintenance
6.1 Regular error and solution
ERROR REASON SOLUTION
1. Overload 1. reduce the weight
2. wrong connection 2. check load cell connection
with load cell 3. Inspect load cell. Check the
UUUUUU
3. Load cell has input and output voltage 0R
problem. input and output resistance to
judge it is good or not.
1. Under load 1. Check platform if it is level or
2. LC connection not.
3. load cell has 2. Check load cell connection.
problem
nnnnnnn 3. Check load cell:Check the
input and output voltage 0R
input and output resistance to
judge it is good or not.
Input the correct weights
During calibration, no
ERR1
input of weights or the
weight is overload
The calibration weights
During calibration , the Minimum is 10% of Max. cap.
ERR2
weights is below than Recommend 60%-80% of Max.
Min. required weights Cap.
18
1. check the connection is
During calibration, the correct
ERR3 input signal is 2. check load cell
negative
During calibration, the After the platform is stable, start
ERR4
signal is unstable calibration
ERR5 Change PCB
Default parameters
Parameter instruction Default
C01 Calibration 1
C02 Decimal digits 0
C03 Resolution 1
C04 Max. capacity 10000
C05 Empty calibration 0
C06 Capacity calibration 0
C07 Restore default 0
C08 Warning tone 1
C09 Power-off automatically 0
C10 Power saving mode 0
C11 Hold function 0
C12 Prohibit kg/lb conversion
1
C13 Upper limit alarm 000000
C14 Under limit alarm 000000
C15 Inner code
19
C16 Date setting
C17 Time setting
C18 Serial interface data output 0
C19 Serial interface Baud rate
3(9600)
C20 Zero manually 10
C21 Initial zero 10
C22 Zero tracking range
0.5
C23 Zero tracking time 1
C24 Overload range 9
C25 Negative range 10
C26 Standstill time 1
C27 Standstill range 2
C28 Dynamic filter 0
C29 Noisy filter 2
C30 Print format 0
C31 Analog signal options 1
C32 4~20mA testing 4
C33 Relay output setting 1
C34 Muti PC communication add. 0
C35 Wireless communication channel 6
C36 Calibration location gravity 9.7936
C37 Destination gravity 9.7936
C38 Version No. check
C39 Reserved menu
20