0 ratings 0% found this document useful (0 votes) 49 views 76 pages HP 3325B Installation
The installation manual for the HP 3325B Synthesizer/Function Generator provides essential information on installation, safety precautions, and operational verification procedures. It includes guidelines for power requirements, grounding, and environmental conditions necessary for optimal performance. Additionally, it outlines warranty details and limitations, as well as customer assistance options available through Hewlett-Packard.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here .
Available Formats
Download as PDF or read online on Scribd
Go to previous items Go to next items
Save HP 3325B Installation For Later INSTALLATION MANUAL
MODEL 3325B
SYNTHESIZER/FUNCTION GENERATOR
Serial Number: ALL
IMPORTANT NOTICE
This manual applies to all instruments.Documentation changes
required after the printing of this manual are shown on a manual
change supplement which accompanies this manual.
WARNING
To prevent potential fire or shock hazard,
do not expose instrument to rain or moisture.
Manual Part No. 03325-90006
Microfiche Part No. 03325-90206
©Copyright Hewlett-Packard Company 1978, 1981, 1984, 1988
8600 Soper Hill Road
Everett, Washington 98205-1298 U.S.A.
Printed: January 1988TA eackano
CERTIFICATION
Hewlett-Packard Company certifies that this product met its published specifications at the
time of shipment from the factory. Hewlett-Packard further certifies that its calibration
measurements are traceable to the United States National Bureau of Standards, to the extent
allowed by the Bureau's calibration facility, and the calibration facilities of other
International Standards Organization Members.
WARRANTY
‘This Hewlett-Packard product is warranted against defects in material and workmanship for
a period of one year from date of shipment. During the warranty period, Hewlett-Packard
Company will, at its option, either repair or replace products which prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated
by -hp-. Buyer shall prepay shipping charges to -hp- and -hp- shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and
taxes for products returned to -hp- from another country.
HP software and firmware produets which are designated by HP for use with a hardware
product, when properly installed on the hardware produet, are warranted not to fail to
execute their programing instructions due to defects in materials and workmanship. If HP
receives notice of such defects during their warranty period, HP shall repair or replace
software media and firmware which do not execute their programming instructions due to
such defects. HP does not warrant that the operation of the software, firmware or hardware
shall be uninterrupted or error free.
LIMITATION OF WARRANTY
‘The foregoing warranty shall not apply to defects resulting from improper or inadequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR,
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES.
HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT,
OR ANY OTHER LEGAL THEORY.
ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available for
Hewlett-Packard products.
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office. Addresses
are provided at the back of ihis manual.SAFETY SUMMARY
‘The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrament. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument.
Hewlett-Packard Company assumes no liability for the customer's failure to
comply with these requirements. This is a Safety Class 1 instrument.
GROUND THE INSTRUMENT
‘To minimize shock hazard, the instrument chassis and cabinet must be connected to an
electrical ground. The instrument is equipped with a three-conductor ae power cable. The
power cable must either be plugged into an approved three-contact electrical outlet or used
with a three-contact to two-contact adapter with the grounding wire (green) firmly
connected to an electrical ground (safety ground) at the power outlet. The power jack and
mating plug of the power cable meet International Electrotechnical Commission (IEC) safety
standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of
any electrical instrament in such an environment constitutes a definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS.
Operating personnel must not remove instrument covers. Component replacement and
internal adjustments must be made by qualified maintenance personnel. Do not replace
components with power cable connected. Under certain conditions, dangerous voltages may
exist even with the power cable removed. To avoid injuries, always disconnect power and
discharge circuits before touching them,
DO NOT SERVICE OR ADJUST ALONE
Do not attempt internal service or adjustment unless another person, capable of rendering
first aid and resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts or
perform any unauthorized modification to the instrument. Return the instrument to a
Hewlett-Packard Sales and Service Office for service and repair to ensure the safety features
are maintained.
DANGEROUS PROCEDURE WARNINGS
Warnings, such as the example below, precede potentially dangerous procedures throughout
this manual. Instructions contained in the warnings must be followed.
WARNING
Dangerous voltages, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and
adjusting.SAFETY SYMBOLS
General Definitions of Safety Symbols Used On Equipment or In Manuals.
A
Instruction manual symbol: the product will be marked with this,
symbol when it is necessary for the user to refer to the
instruction manual in order to protect against damage to the
instrument.
Indicates dangerous voltage (terminals fed from the interior by
voltage exceeding 1000 volts must be so marked.)
Protective conductor terminal. For protection against electrical
shock in case of a fault. Used with field wiring terminals to
indicate the terminal which must be connected to ground before
operating equipment.
Low-noise or noiseless, clean ground (earth) terminal. Used for
a signal common, as well as providing protection against
electrical shock in case of a fault. A terminal marked with this
symbol must be connected to ground in the manner described
in the installation (operating) manual, and before operating the
‘equipment.
Frame or chassis terminal. A connection to the frame (chassis)
of the equipment which normally includes all exposed metal
structures.
Alternating current (power ine.)
Direct (power line.)
Alternating or direct current (power line.)
WARNING The WARNING sign denotes a hazard. Itcalls attention to a procedure, practice,
condition or the like, which if not correctly performed or adhered fo, could
result in injury or death to personnel.
CAUTION
The CAUTION sign denotes a hazard. It calls attention to an operating procedure,
practice, condition or the like, which, itnot correctly performed or adhered to, could
result in damage to or destruction of part or all of the product.
‘The NOTE sign denotes important information, it calls attention to procedure,
practice, condition or the like, which is essential to highlight.HP 33258 Installation
General Installation Information
Initial Inspection
Power Requirements ........cceee00
Line Voltage Selection .
Over-Voltage Protect Circut Breaker
Power Cable and Grounding .
Requirements
Operating Environment
Temperature ores seceerene
Humiclty eae tess
Altitude
Instrument Cooling .
Instattation
HPAB System Interface Connections |.
‘Storage And Shipment ......
Operational Verification
Required Test Equipment
Soff Test
Sine Wave Veriiction
‘Square Wave Verification
Triangle and Ramp Verification
Amplitude Flatness Chack
Sync Output Check
Frequency Accuracy .
Culput Level and Attenuator Check |
Harmonic Distortion
Table of Contents
wi 46
we 46
148
46
we 6
46
48
2410
ant
ceed
2442
412
413
414
415
AIS
416
416
447
Close-tn Spurious Signal
Performance Tests ;
Required Test Equipment .....,
Harmonic Distortion
Spurious Signal .
Integrated Phase Noise
Amplitude Modulation Envelope
Distortion
Square Wave Riso Time .
and Aberrations
Ramp Retrace Time... 6...
‘Syne Output . . ae
‘Square Wave Symmetry ...
Frequency Accuracy
Phase Increment Accuracy
Phase Modulation Linearity
‘Amplitude Accuracy
DC Offset Acouracy (BC Oniy)
DC Offset Accuracy with AC Funetions
Triangle Linearity... see
X Drive Linearity oe.
Ramp Period Variation ...
Operational Verification Record
Performance Test Record
Specifications (Appendix A)
Index
419
4.20
420
423
425
427
428
. 429
4-30
430
43
4.02
4.33
434
4.38
4-05
446
447
450Chapter4
HP 3325B Installation
‘This section contains instructions for installing and interfacing the HP 3325B
Synthesizer/Function Generator as well as tests to verify performance. Included are initial
inspection procedures, power and grounding requirements, operating environment, available
acessories and options, installation instructions, interfacing procedures, and instructions for
repacking and shipping.
‘There are two sets of tests: the first, operational verification, is a subset of the second,
performance tests, an exhaustive test of the HP 3325B specifications. The operational
verification is typically used as an incoming inspection tool upon initial receipt. The
performance tests are used just before shipping from the factory, after any service work, and
when a full calibration is performed.
44General Installation Information
Initial Inspection
‘The HP 3825B was carofully inspected both mechanically and electrically before shipment.
It should be free of mars or scratches and in perfect olecirical order upon receipt. To assure
that this is the case, perform the following steps:
+ Inspect the HP 3325B for physical damage incurred in transit. If the HP 3325B
was damaged in transit, file a claim with the carrier.
Warning The Integrity of the protective earth ground may be interrupted if the
HP 3325B is mechanically damaged. Under no circumstances should the
HP 3325B be connected to power if itis damaged.
+ Check for supplied accessories (listed in Chapter 3 of the Operating Manual).
Inspection will be completed after testing the electrical performance using the Operational
Verification tests which appear later in this document. Also included in this document is the
Performance Test. This is a very detailed test procedure designed to verify that the
HP 3325B meets all the performance specifications.
Power Requirements
Caution _Be’ore applying ac line power to the HP 3325B, ensure the voltage selector on the
HP 33258 rear panel is set for the proper line voltage and the correct line fuse is
installed in the fuse holder. Procedures for changing the line voltage selector and fuse
are contained in the following section for “Line Voltage Selection.”
‘The HP 33255 can operate from any single phase ac power source supplying 100V, 120V,
220V or 240V in the frequency range from 47 to 66 Hz (see table 4-1). With all options
stalled, power consumption is less than 100 VA.
Table 4-1. Line Voltage Ranges
Selector ‘AC Voltage
Voltage Range
100 90-108
120 108-126
220 198-231
240 216-252V
42General Installation Information
Line Voltage Selection
‘The line voltage selector is set at the factory to correspond to the most commonly used line
voltage of the country of destination. The line voltage selected for the HP 3325B is indicated
on the line voltage selector (refer to figure 4-1). Refer to table 4-1 for the line voltage ranges
and table 4-2 to set the line voltage and seleet the appropriate fuse. To change the line
voltage and fuse:
Remove the power cord.
2. Pry open the power selector cover on the rear panel with a small screwdriver
(sce figure 4-1),
3. To cheek or replace the fuso, pull tho white fuse holder out of the power
selector and remove the fuse from the fuse holder.
4, ‘To reinstall the fuse, insert a fuse with the proper rating into the fuse holder.
‘The white arrow on the fuse holder handle should point toward the top of the
instrument. Push the fuse holder into the power selector.
5. Tochange the line voltage, remove the cylindrical line voltage selector.
Caution Remove line voltage selector to change voltage. Rotating the selector without
removing the cylinder could damage the module.
6. Reinstall the cylindrical line voltage selector and ensure the required voltage
label is facing out of the power selector. The eylinder is keyed so that it ean
not be installed backwards,
7. Close the power selector by pushing the side catches in (toward the center of
the cover) and then pressing down firmly on the cover.
8. Check that the correct line voltage appears through the window in the power
selector cover,
Table 4-2. Line Voltage and Fuse Selection
Line Setting Fuse Type | HP Part Number
s00v/120 1A 250V Quick-Acting F) 2110-0782
220v/240V 500 mA 250V Quick-Acting (F) | 2110-0733
Over-Voltage Protect Circuit Breaker
In addition to the current protection provided by the line fuse, the HP 3925B is protected by
an over-voltage circuit breaker. This device disconnects the power supply from the main
power connector when the line voltage exeeeds the upper limit. The reset switch, located on
the rear panel (figure 4-1), pops out when this occurs. If this occurs:
1, ‘Turn the power switch to STANDBY (6) and disconnect the power cord.
2. Check the setting of the line-voltage selector, as described earlier in this
chapter, to be sure that it matches the power connected to the HP 8325B,
43General Installation information
|p gee ae
Figure 4-1. Line Voltage Selection, Fuse Replacement, and Circuit Breaker
3. Reset the circuit breaker by pushing the reset switch on the rear panel.
4, Reconnect the power cord and turn the power switch on,
If the circuit breaker pops out when power is restored and the line voltage level is within the
limits described in table 4-1, send the HP 3325B to a qualified service facility for repair.
Warning Line voltages should be measured by a qualified service person who Is aware
of the hazards involved.
Ifthe circuit breaker does not open and the HP 3325B does not operate, remove power and
check line fuse.
Power Cable and Grounding Requirements
‘The HP 33258 is equipped with a three-conductor power cord which, when plugged into an
appropriate receptacle, grounds the HP 3325B cabinet. The type of power cable plug shipped
with each instrument doponds on the country of destination. Refer to figure 4-2 for the part
number of the power cable and plug configurations available.
44General installation Information
—
eff i
sso0- | a
cours swsonser ofS] — | couesomsen BEL
On eg . Sipe te tag
* a sais Eee aa Newtra
| “Se
Epp genie as
Plus: ceer-viy 2e0v 64 zeov- ea |
(GreLe™ S081 S808 OPERATION CABLE: Sost.sera OPERATION
prion 903 ves.scanasa cpticn #17 eapunnic of fovtn
ae cere
tester oe
Pe re aiay eee
See eee rude (at eee
ye [aaa aes a
SE a AA
a ae Sta
aa |
oroies ssins EAS | | cours Bh
“The nuinber shown forthe pf the industry icetitier for the plug only
The number shown for the cable san H1P part amber for a complete cable including the plug
“UL listed for use in the United States of America
Figure 4-2. Power Cables
WARNING The power cable plug must be inserted into a socket outlet provided with a
protective earth terminal. Defeating the protection of the grounded
instrument cabinet can subject the operator to lethal voltages.
45General Installation Information
Operating Environment
‘The following summarizes the HP 3325B operating environment ranges. In order for the
HP 83258 to mect specifications, the operating environment must be within these limits,
Warning ‘The HP 9325B is not designed for outdoor use. To prevent potential fire or
‘shock hazard, do not expose the HP 33258 to rain or other excessive
moisture.
‘Temperature
‘The HP 3325B may be operated in temperatures from 0°C to 55°C.
Humidity
‘The HP 3325B may be operated in environments with humidity up to 95% (0°C to +40°C),
However, the HP 3325B should be protected from temperatures or temperature changes
which cause condensation within the instrument.
Altitude
‘The HP 3325B may be operated at altitudes up to 4572 meters (15,000 feet).
Instrument Cooling
‘The HP 3325B is equipped with a cooling fan mounted on the rear panel. ‘The HP 3325B
should be mounted so that air can freely circulate through it. When operating the HP 3325B,
choose a location that provides at least 75 mm (3 inches) of clearance at the rear, and at least
25 mm (1 inch) of clearance at each side. Failure to provide adequate air clearance will result
in excessive internal temperature, reducing instrument reliability. The filter for the cooling
fan can be cleaned without removing it. The filter (HP part number 3150-0387) should be
cleaned with a vacuum cleaner every thirty days.
Installation
‘The HP 8325B is shipped with plastic feet in place, ready for use as a portable bench
instrument. The plastie fect are shaped to make full width modular instruments self-align
when they are stacked. The clearances provided by the plastic feet in bench stacking and the
filler strip in rack mounting allow air passage across the top and bottom cabinet surfaces.
A front handle kit can be installed for ease of handling the HP 3325B on the bench. The part
number for the front handle kit is listed in table 3-3 of the HP 3325B Operating Manual,
46General Installation Information
RACK MOUNT FLANGE KIT a
RACK MOUNT FLANGE/ERONT HANDLE KIT
Figure 4-3, Rack Mount and Handle Kits.
Option 908 (rack mount flange kit) and 909 (rack mount flange kit with handles) enable the
HP 33258 to be mounted in an equipment cabinet. The rack mount for the HP 3325B is EIA
standard width of 482.6 mm (19 inches). To install the HP 3325B in an equipment cabinet:
If installed, remove the plastic trim (seo figure 4-3) and front handles from the
HP 33258.
+ Remove the plastic fect from the bottom of the HP 83253.
Install the rack flange kit with or without handles according to instructions
included with the kit. (Kit part numbers are listed in figure 3-3 of the HP 3325B.
Operation Manual.)
Note “The rack mount flange kit of Option 908 will not provide the space requirement for
rack mounting when used withthe front handle kit of Option 907. If front handles
are not available, use the combination kit of Option 909 to rack mount with
handles. If Option 907 front handles are available, use Rack Mount Flange Kit, HP
part number 5061-2072 to add rack mounting.
+ Install an instrument support rail on each side of the instrument cabinet.
(The instrament support rails, used to support the weight of the instrument, are
included with HP instrument cabinets.)
* Lift the HP 3325B to its position in the cabinet on top of the instrument support
rails.
* Using the appropriate screws, fasten the HP 3325B rack mount flanges to the
front of the instrument cabinet.
47General Installation Information
ep Piggybeck
Connectors
Figure 4-4, Typical HP-IB System Interconnection
HP-IB System Interface Connections
‘The HP 93258 instrument is compatible with the Hewlett-Packard Interface Bus (HP-IB).
‘The HP-IB is Howlett-Packard’s implementation of IEEE Standard 4881978 and ANSI
Standard MC 1.1. The HP 3325B is connected to the HP-IB by connecting an HP-IB interface
cable to the connector located on the rear panel. Figure 4-4 illustrates a typical HP-IB system
interconnection.
With the HP-IB system, up to 15 HP-IB compatible instruments can be interconnected. The
HP 10833 HP-IB cables have identical piggy-back connectors on each end so that several
cables can be connected to a single source without special adapters or switch boxes. System
components and devices can be connected in virtually any configuration, There must, of
course, be a path from the controller to every device operating on the bus. As a practical
matter, avoid stacking more than three or four cables on any one connector. If the stack gets
too long, any force on the stack can damage the connector mounting. Be sure that each
connector is firmly screwed in place to keep it from working loose during use. The HP 3325B
uses all the available HP-IB lines, therefore, any damaged connector pins may adversely
affect HP-IB operation. Refer to figure 4-5 for a description of the HP-IB connector,
48General installation Information
he 33258 consin mai threaded HPD ceble mounting stds as
Ieee Sree rere meee i =
iri eine incorrect He
Sciortino naa erste ieee |e |
Sees 7B, |
See ghee i |e
ear | Bes ccvse cme
fewer come | | | Reamer Urea,
me (68) | 2 | eee ur
Figure 4-5. HP-IB Connector Information
‘To achieve design performance with the HP-IB, proper voltage levels and timing
relationships must be maintained. If the system cable is too long, the lines cannot be driven
properly and the system will fail to perform (aee figure 4-5 for HP-IB cable lengths).
‘Therefore, when interconnecting an HP-IB system, it is important to observe the
following rule:
‘Total cable length for the system must be less than or equal to 20 meters
(65 feet) or 2 meters (6 feet) times the total number of devices connected
to the bus, whichever is less,
49General Installation Information
Storage And Shipment
‘The HP 3325B should be stored in a clean, dry environment. The following are
environmental limitations that apply to both storage and shipment:
Temperature + 40°C to +75°C
Humidity Up to 95%
Altitude Up to 15,300 maters (50,000 feet)
‘The HP 33258 should also be protected from temperatures or temperature changes which
cause condensation within the instrument.
Containers and materials identical to those used in factory packaging are available through
Hewlett-Packard offices. If the instrument is being returned to Hewlett-Packard for service,
attach a tag indicating the type of service required, return address, model number, and full
serial number. Also, mark the container FRAGILE to ensure carefull handling. In any
correspondence, refer to the instrument by model number and fall serial number.
‘The following general instructions should be used for repacking with commercially available
materials:
‘+ Wrap the instrument in heavy paper or anti-static plastic. If shipping to a
Hewlett-Packard office or service center, attach a tag to the instrument
indicating type of service required, return address, model number, and full serial
number,
+ Use a strong shipping container. A double-wall carton made of 350-pound test
material is adequate.
+ Use a layer of shock absorbing material 70 to 100 mm (3 to 4 inch) thick around
all sides of the HP 3325B to provide firm cushioning and prevent movement
inside of the container. Protect the contro! panel with cardboard.
Caution Styrene pellets in any shape should not be used as packing material. The pellets do
not adequately cushion the instrument and do not prevent the instrument from shitting
in the carton. The pellets also create static electricity which can damage electronic
‘components,
+ Seal shipping container securely.
+ Mark shipping container FRAGILE to ensure careful handling,
+ In any correspondence, refer to the instrument by model number and full serial
number.
4-10Operational Verification
‘The following procedures are recommended for incoming inspection and for testing the
instrument after repair. Additional tests to be performed following repair of eertain circuits
are indicated in Section VIII of the HP 3325B Service Manual. An Operational Verification
Record is located at the end of this section. For ease of recording the test data at various
times, copies of the blank Operational Verification Record may be made without written
permission from Hewlett-Packard.
Operational Verification includes the following procedures:
Self Test
Sine Wave Vertication
‘Square Wave Verification
Triangle and Ramp Verification
Amplitude Flatness Check
‘Sync Output Check
Frequency Accuracy
Output Level and Attenuator Check
Harmonic Distortion
Close-in Spurious Signal
Required Test Equipment
‘The test equipment required for Operational Verification is listed in table 4-3, Any
equipment that satisfies the critical specifications given in the table may be substituted for
the recommended model.
Table 4-3, Test Equipment Required for Operational Verification
Instrument Critical Specifications Recommended Model
Analog Vertical: HP 1740A/TEK 2245
Oscilloscope Bandwidth: de to 100 MHz
Defiection: 0.01 to 5 V/div
| Horizontal:
Sweep: 0.05 ps to 0.5 s/civ |
x 10 Magnification |
Delayed Swoop
| Electronic Counter Frequency measurement to 20 MHz | HP 5328A,
| Accuracy: + 2 counts with Opt. 010,040, and
Resolution: 8 digits (041/53288 with Opt. 010
DC Digital Voltmeter Ranges: 0.1 to 100V HP 3455434784 |
Resolution: 5 1/2 digits
I Accuracy: * 0.1%
ant‘Operational Verification
Table 4-3. Test Equipment Required for Operational Verification (Cont'd)
Instrument Critical Specifications Recommended mode!
500 Feedthru ‘Accuracy: 0.2% HP 11048C.
Termination | Power Rating: 1W
High Frequency | Frequency Range: 1 to 80 MHz HP 1417/85528/85598/
Spectrum Analyzer | Amplitude Accuracy: + 0.5 oB 8565A/8568A |
Noise: >70 dB below reference
i
Low Frequency Frequency Range: 100 Hz to 50 kHz_| HP 358095854
‘Spectrum Analyzer Amplitude Range: 2 mV to 20V
Noise: >80 0B below input
|| reference or 140 d8v |
| |
Resistor | 4700 aw s% HP 0698-3604
Resistor 56.20 1/8W 1.0% HP 0757-0395
| adapter BNC female-to-dual banana plug _| HP 1251-2277
Self Test
‘This test uses the control, ROM, and control clock cireuits to verify operation of these and
other circuits, The following front panel indications result from this test.
LED check: Turns on all LEDs for about two seconds.
‘The following messages are displayed for about one second:
PASS 0 or FAIL 02n - tests Amptd Cal of de offset.
PASS 1 or FAIL 02n - tests Amptd Cal of sine wave.
PASS 2 or FAIL 02n — tests Amptd Cal of square wave.
PASS 3 or FAIL 02n ~ tests Ampta Cal of triangle wave,
(nis a number from 0 to 9)
Press the blue [Shift] key, then press the [Amptd Cal] key. All LEDs should light, and the
display should not indicate any failures.
Sine Wave Verification
‘This procedure visually checks the sine wave output for the correct frequency and any visible
irregularities.
Equipment Required: Analog Oscilloscope
a, Connect the HP 3325B signal output to the oscilloscope vertical input. Set
the input switch to the 500 position. If your oscilloscope does not have a 500
input, use a 500 feedthru termination at the input.
412L
Operational Verification
Set the HP 3325B as follows:
High Votage Output (option 002) off
Funetion Sine
Frequency 20 MHz.
“Amplitude 10Vpp
Set the oscilloscope vertical control to 2 Vidiy, horizontal to 0.05 ys/div,
‘The oscilloscope should display one cycle per division, approximately five
divisions peak-to-peak.
Change the HP 33258 frequeney to 1 MHz,
Change oscilloscope horizontal control to 0.1 us/div.
‘The oscilloscope should display one sine wave having no visible irregularities,
High Voltage Output (option 002)
Sot the oscilloscope vertical control to 5 Vidiv.
Set the oscilloscope input switch to 1 MQ de coupled position (or disconnect
external 500 feedthru termination).
Select the high voltage output on the HP 3325B. A LED near the key
indicates that the high voltage output is on.
Change the amplitude to 40 Vpp. The oscilloscope should display one sine
wave approximately eight divisions peak-to-peak having no visible
irregularities,
‘Turn off the high voltage output.
Square Wave Verification
‘This procedure checks the square wave output for frequency, rise time, and aberrations,
Equipment Required: Analog Oscilloscope
a
Connect the HP 3325B signal output to the oscilloscope vertical input. Set
the input switch to the 50 position. If your oscilloscope does not have a 500
input, use a 500 feedthru termination at the input.
Set the HP 93258 as follows:
High Voltage Output (option 002) on
Function Square
Frequency 1 MHz
Amplitude 10 Vpp,
Set the oscilloscope vertical control to 2 Vidiv, horizontal to 0.2 ys/div. The
oscilloscope should display two square waves, approximately five divisions
peak-to-peak.
413Operational Verification
Switch the oscilloscope vertical control to 1 Vidiv, so that the aberrations
(overshoot and ringing) can be measured. Aberration excursion should be less,
than 500 mV (1/2 div).
Repeat step d at 2 kHz and 0.1 ms/div.
Adjust the oscilloscope vertical and horizontal controls so that the square
wave rise time between the 10% and 90% points can be measured. Rise time
should be less than 20 ns.
Triangle and Ramp Verification
‘This procedure checks the triangle and ramp output signals for frequency, shape, and ramp
retrace time.
ang
Equipment Required: Analog Oscilloscope
a.
&
Connect the HP 3325B signal output to the oscilloscope vertical input. Set
the input switch to the 502 position, If your oscilloscope does not have a
500 input, use a 500 feedthru termination at the input.
Set the HP 3325B as follows:
High Voltage Output (option 002) oft
Function Tangle
Frequency 40 kHz
Amplitude 10Vpp
Set the oscilloscope vertical control to 2 Vidiv, horizontal to 0.1 ms/div. The
oscilloscope should display one triangle wave per division, approximately five
divisions peak-to-peak.
Change the HP 3325B function to positive slope ramp. ‘The display should be
one ramp per division, approximately five divisions peak-to-peak.
Change the function to negative slope ramp. The display should be one ramp
per division, approximately five divisions peak-to-peak.
Change the oscilloscope horizontal and vertical controls so that the ramp
retrace time from the 90% to 10% points can be measured, Retrace time
should be less than 8 ys.
Change the HP 3325B funetion to positive slope ramp and repeat stop £
Change the function to triangle.
Set oscilloscope vertical control to 2 Vidi, horizontal to 10 us/div. The
oscilloscope should display one triangle wave with no visible irregularities in
either slope.Operational Verification
Amplitude Flatness Check
‘This procedure provides a visual check of the sine wave amplitude flatness,
Equipment Required: Analog Oscilloscope
a, Connect the HP 3325B signal output to the oseilloscope vertical input. Set
the input switch to the 502 position. If your oscilloscope does not have
2.500 input, use a 502 feedthru termination at the input.
b. Set the HP 3325B as follows:
High Vottage Output (option 002) on
Function Sine
Frequency 2kHz
“Amplitude 10Vpp
Sweep Start Freq oHz
Sweep Stop Freq 20 MHz
Sweep Marker Freq 5 MHz
Sweep Time 0.01 sec
©. Connect the HP 3325B X-Drive output to the oscilloscope channel B input.
Connect the signal output to the oscilloscope channel A input.
‘td. Set the oscilloscope as follows:
Display AvsB
Channel A Sensitivity 1Vidiv
(uncal — adjust for full vertical deflection)
Channel B Sensitivity 0.5 vidiv
(uncal — adjust for ful horizontal sweep)
‘Settings may vary from one oscilloscope to another. Note that whichever oscilloscope is used, it
should be operated in a XY made with the HP 3925B X-Drive output driving the horizontal (X)
‘channel and the signal output driving the vertical (Y) channel.
e. Pross the HP 3325B [Start Cont] key.
£ The oscilloscope display should show a sweep that is essentially flat, dropping
no more than 3.5%. Any de variations should be ignored, taking the
peak-to-peak reading for flatness comparison.
Sync Output Check
‘This test verifies the syne output signal levels.
Equipment Required: Analog Oscilloscope.
Connect the HP 3325B syne output to the oscilloscope vertical input. Set the
input switch to the 500 position, If your oscilloscope does not have a
500 input, use a 500 feedthru termination at the input.
‘b. Set the HP 3325B function to sine, frequency to 20 MHz.
¢. Adjust the oscilloscope controls to measure the high and low voltage levels of
the syne square wave. The high level should be greater than +1.2V and the
low level should be less than +0,2¥,
415Operational Verification
Frequency Accuracy
‘This test compares the accuracy of the HP 33258 output signal to the following specification:
45 x 10 of selected frequency (20°C to 30°C).
Equipment Required: Electronic Counter (calibrated within three months or with an
accurate 10 MHz external reference input)
Connect the HP 33258 signal output to the electronic counter channel A
input with a 500 feedthra termination, Allow HP 3325B to warm up for 20
minutes and the counter to warm up for its specified period.
Set the HP 33258 output as follows:
Function Sine
Frequency 20 MHz
Amplitude 0.99 Vpp
DC Offset ov
Set the counter to count the frequency of the A input with 0.1 Hz resolution,
and adjust for stable triggering. Electronic counter should indicate
20 000 000.0 Hz +100 Hz.
Change the HP 3325B funetion to square wave. Frequency automatically
changes to 10 MHz. Electronic counter should indieate 10 000 000.0 Hz
+50 Fz.
Change the function to triangle. Frequency automatically changes to 10 kHz,
Move the counter input to the syne output of the HP 3325B. Set the counter
to average 1000 periods. Electronic counter should indicate 100 000.00 ns
0.5 ne.
Change the function to positive slope ramp. Electronic counter should
indicate 100 000.00 ns +0.5 ns.
Output Level and Attenuator Check
‘Dhis procedure checks the output level and the attenuator by using the “de only” function,
46
Equipment Required: DC Digital Votmeter
500 Feedthru Termination
Connect the HP 3325B signal output directly to a 500 feedthru termination
and then with a cable to the voltmeter input.
If the instrument has high voltage output (option 002), make sure the high
voltage output is off (high voltage indicator light in the lower right corner of
the front panel is off).
Press whichever function key is presently active, indicated by a lighted
indieator beside the key. This removes the ae output. The indicator beside the
[DC Offset] key should light.
Set the HP 3325B de offset to ~BV, then press the [Amptd Call] key.e
Operational Vertication
‘The voltmeter reading should be —4.980 to -5.020 V.
Change the HP 3325B de offset to +5V. Voltmeter reading should be
+4,980 to +5.020 V.
‘Change the HP 39258 de offset to the following voltages. The voltmeter
reading should be within the tolerances shown.
DC offset Tolerances
1.4990 +£1.49300 to 1.50499 V
2499.9 mV +£0.49790 to 0.50190 V
2149.9 mv +£0.14990 t 0 0.15050 V
£49.99 mV £0.04979 to 0.05019 V
£14.99 mV £0.01499 t0 0.01505 V
24.999 mV +£0,004979 10 0,005019 V
4.499 mV.
High Voltage Output (option 002) DC Offset
Remove the 502 feedthru termination and connect the HP 8326B output
directly to the voltmeter input.
Select high voltage output on the HP 3325B. A LED near the key
indicates that high voltage output is on.
Set the HP 3325B de offset to 20V. Voltmeter reading should be
+19.775 to +20.225 V.
Set the HP 3325B de offfset to —20V, Voltmeter reading should be
= 19.775 to ~20.225 V.
Harmonic Distortion
‘This procedure tests the harmonie
following specifications:
istortion of the HP 3325B sine wave output to the
Harmonic Distortion (relative to fundamental)
Fundamental No Harmonie
Greater Than
0.1 Hz to 50 kHz. 65 dB
50 10 200 kHz. 60.08
200 kHz 10 2 MH2 40.08
210 15 MHz 30.43
15t0.20 MHz =25.03_
Equipment Required: High Frequency Spectrum Analyzer
Low Frequency Spectrum Analyzer
500 Feedthru Termination
Resistor 4700 2W 5%
Resistor 56.20 1/6W 1%
447‘Operational Verification
418
a.
db,
Set the HP 33258 output as follows:
High Vottage Output (option 002) off
Function Sine
Frequency 20 MHz
Amplitude 989 Mvp
Connect the HP 3325B signal output to the high frequency spectrum
analyzer 600 input,
Set the spectrum analyzer controls to display the fundamental and at least
four harmonics. Verify that all harmonies are 25 dB below the fundamental.
Set the HP 3325B to 15 MHz and verify that all harmonics are at least
30 dB below the fundamental.
Disconnect the HP 3325B from the high frequency spectrum analyzer and
connect it to the low frequency spectrum analyzer 500 input. Set the
HP 3326B to the following frequencies and verify the specified levels, relative
to the fundamental.
2MHz -40.0B
200 kHz = 60 0B
Set the HP 8825B frequency to 50 kHz and the amplitude to 9.99 mVpp.
Set the spectrum analyzer controls to display the fundamental and at least
three harmonics. (It may be necessary to decrease the video bandwidth to
separate the harmonies from the noise floor.) Verify that all harmonics are at
least 65 dB below the fundamental.
Set the HP 3325B to the following frequencies and verify that all harmonics
are 65 dB below the fundamental.
40 kHz
1 kHz
100 Hz
High Voltage Output (option 002)
Connect the HP 3325B signal output to the low frequency spectrum analyzer
high impedance input (see figure 4-6).
Select the high voltage output on the HP 3325B. Set the amplitude to
40 Vpp and the frequency to 100 Hz,
Set the spectrum analyzer controls to display the fundamental and at least
three harmonics. Verify that all harmonics are 65 dB below the fundamental.
Set the HP 3325B to the following frequencies and verify that their
harmonies are below the specified levels, relative to the fundamental,
10 KHz 65 dB
200 KHz 60 dB
1 MHz 4008
‘Turn off the high voltage output,Operational Verification
Low Frequency
Spectrum Anoiyzer
aoa
aw
NA
S620
2 lew
|
v
Figure 4-6. Harmonic Distortion Verification (High Voltage Output).
Close-In Spurious Signal
‘This procedure tests the sine wave output for spurious signals which may be generated by
the HP 3325B frequency synthesis circuits. The spurious signals must be more than
70 dB lower than the fundamental signal.
Equipment Required: Spectrum Analyzer
a. Set the HP 3325B as follows:
High Voltage Output (option 002) of
Function Sine
Frequency 20.001 MHz
Amplitude 2.99 dBm.
DC Offset ov
b, Connect the HP 3325B signal output to the spectrum analyzer 500 input.
¢. Set the spectrum analyzer controls for a center frequency of 20.001 MHz,
a resolution bandwidth of 30 Hz, a frequency span of 100 Hz/div, and the
fundamental referenced to the top of the display graticule.
d. Set the spectrum analyzer center frequency to 20.002, 20.003, and
20.004 MHr, verifying in each case that all spurious signals are more than
70 dB below the fundamental.
419Performance Tests
‘The following procedures compare the instrument operation to its specifications listed in
Appendix A. Performance Test Records are located at: the end of this section. These test
records lists all of the tested specifications and acceptable limits. For ease of recording data
at various times, copies of the blank Performance Test Records may be made without written
permission from Hewlett-Packard
‘The Performance Tests include the following:
Harmonie Distortion
Spurious Signal
Integrated Phase Noise
‘Ampitude Modulation Envelope Distortion
‘Square Wave Rise Time and Aberrations
Ramp Retrace Time
Sync Output
‘Square Wave Symmetry
Frequency Accuracy
Phase Increment Accuracy
Phase Modulation Linearity
‘Amplitude Accuracy
DC Offset Accuracy (OC Oniy)
DC Offset Accuracy with AC Functions
Tiiangle Linearty
X Drive Linearity
Ramp Period Variation
Required Test Equipment
‘The test equipment required for the Performance Tests is listed in table 4-4. Any equipment
that satisfies the critical specifications given in the table may be substituted for the
recommended model.
4-20Performance Tests
Table 4-4, Test Equipment Required for Performance Tests.
Instrument Critical Specifications | Recommended Mode!
Analog Vertical HP 1740A/TEK 2245,
Oscilloscope Bandwidth: de to 100 MHz
Deflection: 0.01 10 § Vialv
Sweep: 0.05 ys o 0.5 s/div
|
Horizontal |
10 Magnification |
Delayed Sweep
Sampling Vertical TEK 7603* with 7711/
Oscilloscope Defiection: 2 mVidiv | 7811 andS-1
Horizontal |
Sweep: 10 ps to 50 jis/div
Transient response
| Aberrations:
|< +0.5%, ~ 3% Vp
| first Sins following step transition
< 21% Vpp after 5 ms |
Electronic Counter Frequency measurement HP 5328 with
Frequency Range: to 20 MHz Opt. 010, 040, and
Resolution: 8 digs 041/53288 with Opt. 010
Accuracy: + 2 counts
Time interval Average A to B
Resolution: 0.01 ns i
ACIDE Digital ‘AC Function (True RMS) | HP 9455A\34780
Vottmeter Ranges: 1 to 100
Accuracy : + 0.2%
Resolution: § 1/2 digits
Crest Factor: 4:1
DC Functions
Ranges: 0.1 to 100
Accuracy: * 0.05%
Resolution: 5 1/2 digits
500 Feedthru Accuracy: + 0.2% HP 11048C
Termination Power Rating: 1W
High Frequency Frequency Range: 1 kHz to 60 MHz | HP 1417/85525/85538/
‘Spectrum Analyzer | Amplitude Accuracy: = 0.5 dB B566A/8568A
Noise: > 70 d8 below reference
Low Frequency Frequency Range: 20 Hz to 50 kHz | HP S580A/3585A
‘Spectrum Analyzer | Amplitude Accuracy: = 0.5 dB
‘Spurious Responses: 80 dB
below reference |
(9) This equipment is only necessary to perform the Square Wave Rise Time and Aberrations test,
421Performance Tests
Table 4-4, Test Equipment Required for Performance Tests. (Cont'd)
Instrument
Critical Specifications
Recommended Model
Frequency Synthesizer
Double Balanced Mixer
1 MHz Low Pass Fitter
15 kHz Filter
AC Voltmeter
Sine Wave Signal
Source
DC Power Supply
‘Thermal Converter
Resistive Divider
Resistive Divider
Frequency Range: 100 kHz to 21 MHz
Amplitude Range: to +13.01 dBm
Output Impedance: 500.
Phase Noise (Integrated):
9.9 MHz: < ~ 63.08
20 MHz: < ~ 70 dB
Spurious: > 75 dB below fundamental
Impedance: 500
Frequency Range: 1-20 MHz
Cut-off Frequency: 1 MHz
Stopband Atten: 50 dB by 4 MHz
Stopband Freq: 4— 80 MHz
Consisting of:
Resistor: 10k 1%
Capacitor: 1600 pF 5%
Ranges: 0.1 to 1 V
Frequency Range: 20 Hz~1 MHz
Input impedance: 21 MQ
Meter: Log scale
‘Ace (100 Hz to 10 kHz): 1%
Frequency: 10 kHz
‘Amplitude: 1 Vrms Into 20 kQ
Distortion: —60 dB
Volts: 010 +5
‘Amps: 10 mA
Floating Output
Input impedance: 500
Input Voltage: 1 Vrms
Frequency: 2 kHz to 20 MHz
Frequency Response: #0.05 dB
2 kHz 10 20 MHz
Consisting of:
2 Resistors: 61.110 0.1% 1/4W
2 Resistors: 36.550 0.1% 1/8W
Consisting of:
Capacitor: 300 pF 5%
‘3 Resistors: 13300 0.1% 1/4W
Resistor: 430 0.1% 1/8W
HP 33354,
HP 105348,
Model J903
TTE Inc.
22148. Benny Ave.
Los Angeles, CA 80066
HP 0757-0340
HP 0160-2223
HP 400FL/34008
HP 2040/3325/3336
HP 62140/6214B
HP 11050A/Ballantine
Model 13954-1 with
cable 125774 Opt. 10
Ballantine Labs, nc.
P.O. Box 97
Boonton, NJ 07005
HP 0699-0090
HP 0698-7169,
HP 01602207
HP 0698-7453,
HP 0698.8264
422Performance Tests.
Table 4-4, Test Equipment Required for Performance Tests. (Cont'd)
Instrument | Critical Specifications Recommended Model
HighSpeed DC | DC Voltage: Oto +10 HP 34378
DigtalVotmeter | External Tigger Low True
TTL Edge Trigger
Trigger Delay: Selectable 10 to 1408
BNC:to-Triax Adapter | Female BNC to Male Triax HP 1250.0595
Resistive Divider Consisting of:
+25 Resistor: 300 1% 1/4W HP 0698-7533
Resistor: 200 1% 1/4W HP 068.6296
Resistive Divider Consisting of: |
+26 Resistor: 100 kO 1% 1/8W HP 0757-0465
Resistor: 162 KO 13% 1/8W HP 0757-0470
Resistor 4700 2W 5% HP 0698-9634
Resistor 56.20 1/BW 1.0% | Hp 0757-0395
| Adapter BNC female to dual banana plug | HP 1251-2277
BNC Tee HP 1250-0781
Stop Attenuator 0-12.08; 1 dB stops HP 3556
| 0-40.08 HP 3550"
(*) This equipment is only necessary to perform the Square Wave Rise Time and Aberrations test.
Harmonic Distortion
‘This procedure tests the harmonic distortion of the HP 3825B sine wave output to the
following specifications:
Harmonic Distortion (relative to fundamental)
Fundamental No Harmonic
Frequency Greater than
10.1 Hz 10 50 kHz 65 dB
50 10 200 KHz 60.68
200 kHz to 2 Miz 40.08
21015 MHz 90 6B
1510.20 MHz 25 0B
Equipment Required: High Frequency Spectrum Analyzer
Low Frequency Spectrum Analyzer
500 Feedtthru Termination
Resistor 4700 2w 5%
Resistor 56.20 1/8W 1%
423Performance Tests
4.24
a
Set the HP 3325B output as follows:
High Voltage Output (option 002) of
Function Sine
Frequency 20 MHz
Amplitude 999 MVpp,
Connect the signal output to the high frequency spectrum analyzer 500
input,
Set the spectrum analyzer controls to display the fundamental and at least,
four harmonics. Verify that all harmonies are 25 dB below the fundamental.
Set the HP 3325B to 15 MHz and verify that all harmonies are at least
30 dB below the fundamental.
Disconnect the HP 3325B from the high frequency spectrum analyzer and
connect it to the low frequency spectrum analyzer 500 input. Set the
HP 3325B to the following frequencies and verify the specified levels, relative
to the fundamental.
2MHz 40 dB
200 kHz ~60 08
Set the HP 3925B frequency to 50 kHz and the amplitude to 9.99 mVpp.
Set the spectrum analyzer controls to display the fundamental and at least
three harmonies, (It may be necessary to decrease the video bandwidth to
separate the harmonics from the noise floor.) Verify that all harmonics are at
least 65 dB below the fundamental.
Set the HP 3325B to the following frequencies and verify that all harmonics
are 65 dB below the fundamental.
10 kHz
kHz
400 Hz
High Voltage Output (option 002)
Connect the HP 3325B signal output to the low frequency spectrum analyzer
high impedance input (see figure 4-6).
Select the high voltage output on the HP 3325B. Set the amplitude to
40 Vpp and the frequency to 100 Hz.
Set the spectrum analyzer controls to display the fundamental and at least
three harmonies. Verify that all harmonies are 65 dB helow the fundamental.
Set the HP 3825B to the following frequencies and verify that their
harmonies are below the specified level, relative to the fundamental
10kHz -65. cB
200 kHz 60 08,
MHz 40 08
‘Turn off the high voltage output.Performance Tests
Spurious Signal
‘This procedure tests the HP 3325B sine wave output for spurious signals. Circuits within the
HP 3325B may generate repetitive frequencies that are not harmonically related to the
fundamental output frequency. All spurious signals must be more than 70 dB below the
fundamental signal or less than ~90 Bm, whichever is greater.
Equipment Required: Spectrum Analyzer
Mixer Spurious:
a, Connect the HP 3325B signal output to the spectrum analyzer 500 (RF)
input and the HP 3325B EXT REF input to the analyzer 10 MHz reference
output (see figure 4-7).
b. Set the HP 33258 as follows:
Function Sine
Amplitude =20 08m
Frequency 2.001 MHz
©. Set the analyzer controls as follows:
Center Frequency 2.001 MHz
Frequency Span kHz
Video 87 400 Hz
Resolution BW 30 Hz
d. Adjust the spectrum analyzer to referenee the fundamental to the top display
graticule,
e. Without changing the reference level, change the spectrum analyzer center
frequency to 27.999 MHz to display the 2:1 mixer spur. Verify that this spur
is at least 70 dB below the fundamental.
£ Change the spectrum analyzer center frequency to 25.998 MHz to display the
8:2 mixer spur. Verify that this spur is at least 70 dB below the fundamental.
g — Inasimilar manner, change the HP 3325B frequency and the spectrum
analyzer center frequency to the following frequencies. For each setting,
verify that all spurious signals are 70 dB below the fundamental.
Spectrum Analyzer
HP 39258 renter Frequency
2:1 Spur 8:2 Spur
4.100 MHz 25.9 MHz 21.8 MHz
6.100 MHz. 23.9 MHz 178M
8.100 MHz 21.9 MHz 13.8 MHz
40.100 MHz 49.9MHz 9.8 MHz
12.100 MHz. 17.9 MHz 5.8 MHz
14.100 MHz. 18.9 MHz 1.8 MHz
36.100 MHz. 18.9 MHz 2.2MHz
18.100 MHz 14.9 MHz 6.2MHz
20.100 MH2 9.9 MHz 10.2 Miz,
425Performance Tests
‘SPECTRUM ANALYZER
Figure 4-7. Mixer Spurious.
Close-in Spurious (Fractional N Spurs)
h. Set the HP 33258 frequency to 5.001 MHz and the amplitude to ~2.99 dBm.
i. Sot the spectrum analyzer controls as follows:
Center Frequency 5.001 MHz
Frequency Span 1 kHe
Video BW 100 Hz
Resolution BW 30Hz
j. Adjust the spectrum analyzer to reference the fundamental to the top display
graticule,
k. Without changing the reference level, change the spectrum analyzer center
frequency to 5,002 MHz. to display the API 1 spur. It may be necessary to
decrease the video bandwidth to optimize the display resolution.
1. All spurious (non-harmonie) signals should be at least 70 dB below the
fundamental.
m. Without changing the reference level, set the HP 3325B frequency and the
spectrum analyzer center frequency to the frequencies listed below. For each
sotting, verify that all spurious signals are at least 70 dB below the
fundamental.
Spectrum Analyzer
Hp 33258 Genter Frequency
5.0001 Miz ‘5.0071 MH
5.00001 M2 ‘5.00101 M2
5.000001 MHz 5.003001 MHz
20001 MH 20.002 MHz
20.001 wiz 20.003 Miz
20.001 MHz 20.004 MHz
20,008 Mie 20,005 Mie
426Performance Tests
Integrated Phase Noise
‘This test compares the HP 3325B integrated phase noise to the following specification:
60 dB for a 30 kHz band centered on a 20 MHz cartier (exclu
about the cartier).
get Hz
Equipment Required: Frequency Synthesizer
Double Balanced Mixer
‘500 Feedthru Termination
AC/DC Digital Voltmeter
‘AC Voltmeter
15 kHz noise equivalent fiter consisting of:
Resistor: 10 KO + 1%
‘Capacitor: 1600 pF + 5% (soe figure 4-8)
4 MHz Low Pass Filter
Connect the equipment as shown in figure 4-8, connecting the 15 kHz noise
equivalent filter output to the ac voltmeter. Phase lock the HP 3325B and the
signal generator together.
b. Set the HP 3325B as follows:
Function Sine
Frequency 19.901 MHz
Amplitude o.dBm
c. Sot the synthesizer (reference) as follows
Frequency 19.9 MHz
Amplitude +7.00 dam
4. Record the ac voltmeter reading (dB seale).
e. Change the HP 33258 frequency to 19.9 MHz,
f. Connect the 15 kHz filter output to the digital voltmeter,
g Pros the HP 8325B [Phase] key. Using the modify keys, adjust the output
phase for a minimum reading on the digital voltmeter.
h. Disconnect the 15 kHz filter output from the digital voltmeter and connect it
to the ac voltmeter.
i. Record the ac voltmeter reading (4B scale) and subtract it from the reading
recorded in step d. The difference should be ~64 dB or greater. Add —6 dB to
this number and enter on the Performance Test Record. The 6 dB is a
correction factor compensating for the folding action of the mixer.
NOTE Frequencies used minimize the phase noise contribution of the frequency
‘synthesizer.
427Performance Tests
AC.
VOLTMETER
vn 1D.
La Fase
FLIER son
c1
=| +—
Figure 4-8. Integrated Phase Noise.
Amplitude Modulation Envelope Distortion
‘This procedure tests the HP 8325B amplitude modulation envelope distortion to the
following specification:
—30 dB to 80% modulation at 1 kHz, OV de offset
Equipment Required: Sine Wave Signal Source
‘Spectrum Analyzer
a, Connect the equipment as shown in figure 4-9.
b. Set the HP 33258 output as follows:
Function Sine
Frequency 1 MHz
Amplitude 3V ep
DC Offset ov
High Voltage Output (option 002) off
AN on
428Performance Tests
SPECTRUM ANALYZER
Figure 4-9. AM Envelope Distortion.
¢. Set the modulating signal source frequency to 1 kHz. and adjust the level to
produce 80% modulation of the HP 3325B output. This is indicated by
modulation sidebands being 8.0 dB down from the carrier, as viewed on the
2 dB/div display of the spectrum analyzer.
4. Adjust the spectrum analyzer to display the fundamental frequency, the
1 kHz sideband frequency, and at least 4 harmonics of the sidebands. All
harmonies should be at least 30 dB lower than the modulation sidebands.
Square Wave Rise Time and Aberrations
‘This procedure compares the HP 3525B square wave output to its rise/fall time and
overshoot specifications,
Rise and Fall Time: <20 ns, 10% to 90% at full output
Overshoot: <5% of peak-to-peak amplitude at full output
Equipment Required: Sampling Oscilloscope
40 dB Attenuator
a, Connect the HP 33258 signal output to the attenuator input and the
attenuator output to the oscilloscope input. Set the attenuator for 40 dB
attenuation,
b. Set the HP 33258 as follows:
High Voltage Output (option 002) on
Funetion Square
Frequency MHz
Amplitude 40 pp
429Performance Tests.
Adjust the oscilloscope vertical and horizontal controls so that the square
‘wave rise time between the 10% and 90% points can be measured. Rise time
should be loss than 20 ns.
Adjust the oscilloscope vertical and horizontal controls so that the square
wave fall time between the 10% and 90% points can be measured. Fall time
should be less than 20 ns.
Adjust the oseilloseope vertical and horizontal controls so that the square
wave overshoot can be measured. Overshoot should be less than 500 mV at
and negative peaks.
posi
Ramp Retrace Time
‘This test compares the HP 3325B retrace time of the positive and negative slope ramps to
the following specification:
$3 128 90% to 10%
Equipment Required: Analog Oscilloscope
a
a
Connect the HP 3325B signal output to the oscilloscope vertical input. Set
the input switch to the 502 position, If your oscilloscope does not have a
‘500 input, use a 500 feedthru termination at the input.
Set the HP 3325B as follows:
High Voltage Output (option 002) oft
Function Posttive Slope Ramp
Frequency 40 kHz
Amplitude 10 Vpp
‘Adjust the oscilloscope vertical and horizontal controls so that the ramp
retrace time from the 90% to 10% points can be measured. Retrace time
should be less than 3 us.
Change function to negative slope ramp and repeat step e.
Sync Output
‘This procedure checks the voltage levels of the square wave on the HP 8325B front and rear
panel syne outputs to the following specifications:
Vhigh > +1.2V; Viow < +0.2V into SOQ
Equipment Required: Analog Oscilloscope
a
Connect the HP 3325B front syne output to the oscilloscope vertical input,
Sot the input switch to the 500 position, If your oscilloscope does not have a
500 input, use a 500 feedthru termination at the input.
Sot the HP 3325B function to sine, frequency to 20 MHz.Performance Tests
3325, ELECTRONIC
Bee) = hE
it}
Figure 4-10. Square Wave Symmetry.
Adjust the oscilloscope controls to measure the high and low levels of the
syne square wave. The high level should be greater than +1.2V and the low
level should be less than +0.2V.
Repeat the measurement for the rear panel FAST “syne output, The high
level should be greater than +1.5V and the low lovel less than +0.5V.
Square Wave Symmetry
‘This procedure checks the symmetry of the HP 3325B square wave signal output to the
following specification:
50.02% of period +3 nanoseconds
Equipment Required: Electronic Counter
@ — Connect the HP 8325B signal output to both inputs of the electronic counter,
using a BNC tee (see figure 4-10).
b. Set the HP 3325B output as follows:
Function ‘Square
Frequency MHz
“Amplitude ims
DC Offset ov
© Adjust the electronic counter to measure time interval average A to B, with
Slope A +, Slope B -. Note the reading.
a
Change Slope A to ~, Slope B to +. Reading should be equal to the reading in
stepe+ <8.2ns,
FAST ™ is a trademark of Fairchild Semiconduetor Corporation,Performance Tests
ELECTRONIC
COUNTER
Figure 4-11. Frequency Accuracy.
Frequency Accuracy
‘This test compares the accuracy of the HP 3325B output signal to the following
specifications:
£5 x 10 of selected frequency (20°C to 30°C).
Equipment Required: Electronic Counter (calibrated within three months or with an
ce.
482
accurate 10 MHz extemal reference input)
Connect the HP 3325B signal output to the electronic counter channel A
input with a 500 feedthru termination, Allow the HP 3325B to warm up for
20 minutes and the counter’s frequency reference to warm up for its specified
period.
Sot the HP 3325B output as follows:
Function sine
Frequency 20 MHz
Amplitude 0.99 Vp
DC Offset ov
Sot the counter to count the frequency of the A input with 0.1 Hz resolution,
and adjust for stable triggering. Electronic counter should indicate
20 000 000.00 Hz +100 Hz.
Change the HP 3325B function to square wave. Frequency automatically
changes to 10 MHz. Electronic counter should indicate 10 000 000.0 Hz
=50 Hz,
Change the HP 3325B function to triangle. Frequency automatically changes
to 10 kHz, Move the counter input to the syne output of the HP 3325B, Set
the counter to average 1000 periods, Electronic counter should indicate
100 000.00 ns +0.5 ns.
Change the HP 3325B funetion to positive slope ramp. Electronic eounter
should indicate 100 000.00 ns +0.5 ns.Performance Tests
To EXT
REF IN ELECTRONIC COUNTER
lowe| SYNTHESIZER
REF.
7 REFERENCE,
Figure 4-12. Phase Increment Accuracy.
Phase Increment Accuracy
‘This test compares the HP 3325B phase increment accuracy to the following specification:
10.2"
Equipment Required: Frequency Synthesizer
Electronic Counter
‘Connect the equipment as shown in figure 4-12.
b. Set the HP 89258 as follows:
High Voltage Output (option 002) on
Function Sine
Frequency 100 kHz
Amplitude 13.d8m
¢. Set the synthesizer as follows:
Frequency 0.1 MHz,
Amplitude 13.dBm,
4. Set the counter as follows:
Function rear e
Frequency Resolution, N 1
Inputs: 500, Separate
Slope A and B Positive
‘Sample Rate Maximum
e. Press the HP 3325B [Phase] key to display phase. Using the modify keys,
adjust the phase until the counter reads approximately 200 ns. Press the
blue [Shift] key, then the [Asgn Zero ©] key.
433Performance Tests
Set the counter sample rate to hold, then reset the counter. Record the
counter reading (to 2 decimal places) on the Performance Test Record in the
space for Zero Phase Time Interval. This is the phase difference (in
nanoseconds) between the HP 33258 output and the reference signal.
Set the HP 3325B phase to ~1°.
Reset the counter, Record the counter reading (to 2 decimal places) in the
space for 1° Increment Time Interval.
Determine the time difference between the counter readings in steps h and f,
and record in the Time Difference column. The difference should be from
22.22 to 33.34 ns,
Set the HP 3325B phase to —10°,
Reset the counter. Record the counter reading in the space for 10° Increment
Time Interval.
Enter the time difference between the Zero Phase Time Interval and the
reading in step k in the Time Difference column. This should be from
272.22 to 283.34 ns,
Set the HP 3325B phase to —100°,
Reset the counter. Record the counter reading in the space for 100°
Incremental Time Interval.
Enter the time difference between the Zero Phase Time Interval and the
reading in step n in the Time Difference column, It should be from
2722.22 to 2783.34 ns,
Phase Modulation Linearity
‘This procedure compares the HP 3325B phase modulation linearity to the following
specification:
434
a.
+0.5%, best ft straight line
Equipment Required: Frequency Synthesizer
Electronic Counter
DC Power Supply
Digital Voltmeter
Connect the equipment as shown in figure 4-13,Performance Tests
POWER SUPPLY MoD iN’
0 PHASE 35
ELECTRONC COUNTER
=
ea T
ro anc be
DIGITAL
yee To ExT
REF IN
VOLTNETER
loves
REF OUT SYNTHESIZER
REFERENCE
a,
e
Figure 4-13, Phase Modulation Linearity.
Set the HP 3325B as follows:
of
Sine
4100 kHz
43.dBm
Phase Modulation On
Set the synthesizer as follows:
Frequency 400 kHz,
Ampiitude 13.d8m
Sot the electronic counter as follows:
Function Tinge interval Avg A to B
Frequency Resolution, N 10
Inputs 500, Separate
Slope A and B Positive
‘Sample Rate Maximum
Using the voltmeter to monitor the de power supply output, sot the de voltage
as near ~5.0000V as possible.
Press the HP 3325B [Phase] key to display phase, Using the modify keys,
adjust the phase until the counter reads approximately 200 ns. Record the
counter reading as a reference for the following steps.
‘As soon as possible after recording the counter reading, note the voltmeter
reading and record on the Performance Test Record in the DVM Reading,
xy space.
Press the HP 3325B blue [Shift] key, then the [Asgn Zero 9] key.
Change the de power supply output to —4.0000V.
Using the modify keys, adjust the HP 3325B phase to return the counter
reading to the value recorded in step f