CERTIFICATE OF PERFORMANCE FOR PATIENT MONITOR
Customer : Aster MEDCITY
Kerala, COCHIN
Equipment : Umec 100 - Multipara-Monitor
Serial No. : FT4-41000799
Manufacturer : Shenzhen Mindray Biomedical Equipment Co. Ltd
Date of Calibration : 08/07/2025
Valid Up to : 07/07/2026
CALIBRATION STANDARD USED:
SL. CAL. CAL DUE
ASSET NO TEST EQUIPMENT MANUFACTURER MODEL NO SERIAL NO
NO. DATE DATE
PATIENT SAFETY FLUKE ESA615 4975032 12/09/24 12/09/25
1
ANALYZER
FLUKE PROSIM8 5387027 20/09/24 20/09/25
2
PATIENT SIMULATOR
ECG & RESPIRATION:
Test Item Test Method Standard/Requirement Test Result
ECG Display/Waveform HR:80BPM, RES: 20 80±1BPM / 20±1RPM 80BPM / 21 RPM
ECG Wave Gain Change Gain The ECG wave changes OK
as per to gain.
Lead Selection Change lead status Normal. No interference OK
ECG Monitoring Type Monitoring Type: Full Normal. No interference OK
lead
ECG Filter Change Filter The ECG Waves changes OK
as per Filter
Pace Analysis Stimulate PACE Respond correctly. OK
Signal Output. Set
pace analysis of the
main unit to ON.
Lead off Stimulate Lead Off Lead Off is prompted and OK
Status alarmed correctly
Notch Notch Normal OK
SpO2:
Test Item Test Method Standard / Requirement Test Result
Tone Modulation Increase and Tone modulation changes OK
decrease SP02 when SP02 value
Value by stimulator. increase and decrease
SP02 Cal Test Stimulates different SP02 Value & PR OK.
SP02 & PR Value. displays with1% Accuracy.
NIBP
Test Item Standard Specs Test Result
Accuracy 120/80 ±8mmhg 123/86 -OK
Accuracy 100/60 ±8mmhg 105/68 -OK
Accuracy 60/30 ±8mmhg 63/35 -OK
Leakage Test 1mmhg/min OK
General:
Test Item Standard Specs Test Result
Operates the Menu Respond correctly OK
Date & Time / Language Realtime clock running & Defalut OK
Language “English”
Equipment version Check the latest SW ver & OK
configuration
Lable Check the lable & Serial No. OK
Panel buttons Respond correctly. OK
Storage function Change the equipment settings,
turn off and turn on. The
OK
changed setting should be
restored
Power Supply:
Test Item Standard Specs Test Result
AC input (230Vac/50HZ) 230V / 50HZ with Proper OK
Grounding.
Battery power supply Battery should power the OK
equipment normally.
Fan Should Run Freely OK
The equipment is met with the standard specification recommended by us and found working with its optimal
performance.
For Mindray Medical India Private Limited,
REMITH
Authorized Signatory