FT - BACGene L.mono
FT - BACGene L.mono
EGS 38/03-01/17
ALTERNATIVE ANALYTICAL
METHODS FOR AGRIBUSINESS
www.afnor-validation.com
BACGene
Table of Contents
1 INTRODUCTION ........................................................................................................................... 3
3 Agilent Aria™: Instructions for use with BACGene Listeria monocytogenes kit ......................... 22
4 Bio-Rad CFX96 Touch™ Instructions for use with BACGene Listeria monocytogenes Kit ........ 26
1 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
5 Data Interpretation....................................................................................................................... 32
7 TROUBLESHOOTING ................................................................................................................ 34
2 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
1 INTRODUCTION
The BACGene Listeria monocytogenes detection kit provides materials for rapid detection of Listeria
monocytogenes DNA from all human food products and environmental samples from food production
facilities.
The kit is intended to be used in analytical laboratories and may also be applied for other purposes in
food product research and e.g. microbial monitoring of production processes. The kit is not intended for
diagnostic use with medical specimens.
The BACGene Listeria monocytogenes kit is validated for use with the Agilent AriaMx™ and Bio-Rad
CFX96 Touch™ Deep Well PCR platforms.
1.1 Certifications
3 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
The BACGene Listeria monocytogenes kit is certified by AOAC-Research Institute under the
Performance Tested MethodsSM Program for detection of Listeria monocytogenes for:
25 g / 25 mL samples of
- Raw milk
- Ice cream with chocolate inclusions (with PREraser)
- Smoked salmon (with PREraser)
- Mayonnaise-based vegetable salad
- Frozen cantaloupe balls
- Frankfurters
- Process water
Stainless steel environmental surface
Ceramic environmental surface
Plastic environmental surfaces with sponge (with PREraser)
The end of the validity of the AOAC certification is indicated on the certificate 061703.
PREraser BACGene (see sections 1.3 and 2.5.1) has been performance tested for three relevant
matrices. Therefore, provided that additional matrices are verified, PREraser BACGene can be applied
to other matrices within the scope of the method.
4 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
DNA amplification and detection methods take advantage of the nucleotide sequence conservation
found in bacterial genomes that ensures the potential for high specificity and sensitivity in detection of
food-borne, pathogenic bacteria.
After enrichment, the microbial DNA is extracted by a simple thermal / enzymatic lysis step and rapidly
analysed by real-time PCR. In this way Listeria monocytogenes is detected from enrichment cultures
with extraordinary high sensitivity.
By means of specific primers (P) a nucleotide sequence of the species Listeria monocytogenes is
amplified during PCR. Primers do not react with DNA derived from closely related species from the
Bacillales order.
The amplified fragments for Listeria monocytogenes are detected with a R6G fluorescence-labelled
hybridisation probe quenched by non-fluorescent Tide Quencher™ 2 (TQ2).
An internal positive control (IPC) is included in the MasterMix. IPC DNA is amplified in parallel and
detected using a Cy5™ fluorescence-labelled hybridisation probe quenched by non-fluorescent Tide
Quencher™ 3 (TQ3). IPC detection indicates the proper functioning of the PCR.
Probe
Probe
5 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Important Notes
Store all reagents as indicated below, through to the expiration date printed on the label.
Never store components of the kit together with samples or post-PCR products. Shelf life is
indicated on the labels of the components.
Do not mix components from different kit lots.
*Note: additional domed caps for lysis can be purchased in a package of 250x 8 cap strips (cat. no.
5613900301).
6 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
7 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
*Note: Products pre-moistened with neutralising buffers containing aryl sulfonate (sodium
dodecylbenzenesulfonate) should not be used in conjunction with this kit. The use of Dey-Engley
neutralizing buffer, Letheen broth or HiCap broth is recommended.
8 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
1.4.2 Lysis
Equipment
2x Block heater (e.g. block heater, digital, VWR cat. no. 460-0350).
2x Insert for block heater, 0.2 mL, 96 wells (e.g. double block, 1×96-well PCR plate,
VWR cat. no. 460-3226).
1x 96 well cooling block (e.g. PCR-Cooler, VWR cat. no. 479-1162).
1x Lysis support grid, blue, (e.g. cat. no. 5617702209, pack of 8 grids).
1x Lysis rack (e.g. cat. no. 5613901401).
1x Strip centrifuge:
Capacity of 2x 8-well strips: (e.g. Carl Roth GmbH, Rotilabo® centrifuge with butterfly
rotor cat. no. T465.1).
Capacity of 4x 8-well strips: (e.g. IKA, Mini Centrifuge Mini G cat. no. 0003958000
or VWR, MiniStar silverline cat. no. 521-2844P).
OR
1x Plate centrifuge:
Capacity of two times 12x 8-well strips: (e.g. Benchmark Scientific, PlateFuge™
microplate microcentrifuge cat. no. C2000).
1x Capping/Uncapping tool for 8-well strips (e.g. Thermo Fisher Scientific,
MicroAmp® Cap Installing Tool, cat. no. 4330015).
1x Stepper pipette (with 1 mL tip), (e.g. HandyStep® S Brand®, cat. no. 705110)
1x Single channel pipette, (e.g. Brand®, Transferpette® S 100 - 1000 µL,
cat. no. 705880).
1x Single channel pipette, (e.g. Brand®, Transferpette® S 10 - 100 µL,
cat. no. 705874)
1x Single channel pipette, (e.g. Brand®, Transferpette® S, 0.5 - 10 µL,
cat. no. 705870)
1x Pipette filler, (e.g. VWR, Pipetboy acu 2 for serological pipettes, cat. no. 612-0926)
9 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Equipment
1x PCR support grid, yellow (e.g. cat. no. 5617702208, pack of 8 grids).
Note: 1x PCR support grid is provided with each kit.
1x PCR rack (e.g. cat. no. 5613901301).
1x Capping/Uncapping tool for 8-well strips (e.g. Thermo Fisher Scientific,
MicroAmp® Cap Installing Tool, cat. no. 4330015).
1x Single channel pipette, (e.g. Brand®, Transferpette® S, 0.5 - 10 µL,
cat. no. 705870).
AND/OR
1x Multichannel pipette, (e.g. Brand®, Transferpette® S-8 Kanal / 0.5 – 10 µL,
cat. no. 705900)
1x Disposable transfer rack (e.g. an empty tip box) as mount for the yellow support grid
and the PCR plate
1x Real-time PCR Thermocycler:
- Agilent AriaMx™ (cat no.: G8830A) (up to Software v.1.5), with FAM™ and Cy5™
filter set.
- Bio-Rad CFX96 Touch™ (CFX Manager™ Software v. 3.1 / CFX MaestroTM Software
1.1).
- Bio-Rad CFX96 Touch™ Deep Well (CFX Manager™ Software v. 3.1 / CFX
MaestroTM Software 1.1).
Windows 7/10 PC with Microsoft Excel 2010 32 bit (Excel Version 14.7140.xxxx or later).
BACGene Evaluation Sheet version 2
OR
FastFinder automated PCR analysis software (UgenTec NV, Hasselt, Belgium) with
BACGene Listeria monocytogenes plugins with version 1
10 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
All samples should be handled with caution as they are potentially infectious.
In order to safeguard the health of laboratory personnel, it is essential that the whole of this
method be carried out only by skilled personnel using good laboratory practices.
A BSL-2 laminar flow biosafety cabinet is recommended for activities with potential for
producing aerosols of pathogens.
Relevant national Health and Safety Regulations relating to this organism must be adhered to.
Care must be taken in the disposal of all infectious materials.
Do not eat, drink or apply cosmetics in the work area in which the test is performed.
Do not pipette by mouth.
Avoid contact of kit components with injured skin.
Listeria monocytogenes should not be handled by pregnant women, children, the elderly and
immuno-compromised individuals due to the high infection risk and fatal health consequences
for this group, in particular for the unborn child in case of pregnant women.
The BACGene Listeria monocytogenes kit contains enzymes (proteinase K and lysozyme)
which may cause allergic reactions.
Lysis buffer L can cause serious eye irritation.
Always wear appropriate Personal Protective Equipment (PPE).
The remaining components of the BACGene Listeria monocytogenes kit are not classified
according to Regulation (EC) No. 1907/2006 (REACH). For more information, please refer to
the kit’s safety data sheet (SDS).
11 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
For use of this kit, two different types of contamination can impair results: Bacterial contamination is
prevented by regular disinfection, DNA contamination is prevented by regular decontamination.
Before commencing work and after completion of work, ensure workspaces to be clean:
Dispose of any waste which is potentially contaminated with pathogenic bacteria according to your
internal and local regulations.
12 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
2.4 Enrichment
Enrichment samples shall be prepared in pre-warmed (37 ± 1 °C) Actero™ Listeria Enrichment Media.
Please ensure that the prepared medium is stored at 2° – 8 °C under light protection.
For preparation of initial suspensions follow instructions of EN ISO 11290 and EN ISO 6887standards.
For sampling of environmental samples follow instructions EN ISO 18593 standard.
It is strongly recommended to prepare samples in stomacher bags containing a side filter. It is also
strongly recommended to prepare enrichment controls (positive and negative) in parallel with the
samples. Growth in the negative enrichment control indicates contamination of the medium used and
therefore enrichment should be repeated.
After the enrichment protocol, samples can be stored at 2-8 °C for up to 72 h before testing.
Raw milk,
Smoked salmon (incl.
PREraser),
Mayonnaise-based
vegetable salad,
Frozen cantaloupe 1:10
balls, Actero™ Listeria
Frankfurters, Enrichment Media
Ice-cream with (Pre-warmed to 37 °C
chocolate inclusions ± 1 °C)
(incl. PREraser)
25 g
Process water
25 mL
37 ± 1 °C 18 - 24 h 30 µL 5 µL
10 mL
Stainless steel
environmental surface Actero™ Listeria
1” x 1” area Enrichment Media
Tool: swab (Pre-warmed to 37 °C
± 1 °C)
Ceramic environmental
surface
4” x 4” area 100 mL
Tool: sponge Actero™ Listeria
Plastic environmental Enrichment Media
surface (incl. (Pre-warmed to 37 °C
PREraser) ± 1 °C)
4” x 4” area
Tool: sponge
13 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
10 mL
Swab Actero™ Listeria
Enrichment Media
1
(Pre-warmed to 37°C ±
1°C) 37 ± 1°C 18 - 24 h 30 µL 5 µL
225 mL
Wipe, cloth Actero™ Listeria
Enrichment Media
1
(Pre-warmed to 37°C ±
1°C)
100 mL
Sponge Actero™ Listeria
Enrichment Media
1
(Pre-warmed to 37°C ±
1°C)
14 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Note: The volume of Lysis buffer L of the kits with cat nos. 5123222001 and 5123222011 (3.5 mL) is
sufficient for 48 reactions. Lysis buffer L volume of the kit with cat no. 5123222010 (35 mL) is sufficient
for 480 reactions or five 96 well plates.
An optional elimination of free DNA from enrichments prior to the lysis step can be incorporated into the
workflow and has been AOAC-RI certified with the use of the kit.
PREraser BACGene can be applied for each sample in case of sample types with expected high levels
of free DNA (e.g. environmental samples) or after a first PCR positive prior to a second BACGene PCR
to exclude the presence of free DNA.
Please refer to the PREraser BACGene user’s manual to include PREraser BACGene to the workflow.
PREraser BACGene is not part of the AFNOR certified workflow for Listeria monocytogenes.
See section 1.4.1 for ordering information.
15 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
1. Label lysis buffer plate/strips to ensure correct orientation when pipetting samples.
2. Using a serological pipette, remove an aliquot of enrichment culture and dispense into a sterile
container. This step is highly recommended to avoid cross-contamination, as opposed to inserting
a micropipette directly into the enrichment bag.
3. Use an Un-/Capping Tool (with blue lysis label, dedicated to lysis area) to open the first lysis strip.
Note: Open each strip individually, add samples and close. Only then proceed to the next strip. The
correct method to use this tool is to place the “teeth” of the tool under the connection between the
caps and lever the caps open.
4. Transfer 30 µL of enrichment into appropriate wells.
Note: Do not add enrichment samples to wells A1 and B1 to ensure the same layout in the following
PCR (A1: positive control, B1: negative control).
5. Seal the strip using the Un-/Capping Tool.
6. Proceed to the next lysis strip until all samples have been included.
7. Place lysis plate/strips with blue frame onto the 37°C (± 2°C) heating block for 20 min.
8. Transfer lysis plate/strips with blue frame onto the 95°C (± 2°C) heating block for 10 min.
9. Finally transfer lysis plate/strips with blue frame onto the blue 96 well cooling block (stored at 4°C ±
2°C) for 5 min.
10. Ensure that no condensation remains in the lids by transferring the lysis plate/strips with the blue
frame to a 96-well plate centrifuge or a centrifuge with adapter for 0.2 mL in 8-strip format. Spin
down samples for 30 sec within the range of 400 x g to 2000 x g. If condensate remains in the lids,
please repeat this step.
11. If particles are visible in the lysis vessel, centrifuge the plate/strip following the procedure in the step
above.
12. Transfer lysis plate and blue frame into the working rack with transparent inlay.
16 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
2.7 PCR
PCR is an exponential reaction. The detection of single DNA targets is possible. The extreme sensitivity
requires special precautions for handling and equipment. After a successful amplification, several billion
amplicons are present in the reaction tube. Each of them might lead to a false positive result when
contaminating sample material, i.e. by spreading in aerosols.
17 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
PCR is performed in a volume of 25 µL in 0.1 mL reaction tubes/plates according to the real-time PCR
cycler instructions.
Test samples:
20 µL MasterMix + 5 µL lysate per reaction
Important
For every PCR, it is necessary to prepare a positive and a negative control reaction. These are required
for the evaluation of the PCR results! The layout should start with the positive control in A1, the negative
control in B1, samples in the following wells.
Required:
1 Positive control (C+):
5 µL Listeria monocytogenes positive control plasmid DNA
1 Negative control (C-):
5 µL heat-inactivated lysis buffer (or BACGene PCR Negative Control) added to MasterMix
Recommended:
1 Negative enrichment control (E-):
Listeria monocytogenes free enrichment medium control
1 Positive enrichment control (E+):
Lysed enrichment medium inoculated with Listeria monocytogenes
18 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
2.7.4 PCR
Plate Setup
The following plate document shows the recommended distribution of reactions:
1 2 3 4 5 6 7 8 9 10 11 12
A C+ 7 15 23 31 39 47 55 63 71 79 87
B C- 8 16 24 32 40 48 56 64 72 80 88
C 1 9 17 25 33 41 49 57 65 73 81 89
D 2 10 18 26 34 42 50 58 66 74 82 90
E 3 11 19 27 35 43 51 59 67 75 83 91
F 4 12 20 28 36 44 52 60 68 76 84 P-*
G 5 13 21 29 37 45 53 61 69 77 85 E-
H 6 14 22 30 38 46 54 62 70 78 86 E+
*The use of a PREraser negative control (C- or E- with PREraser) is recommended when applying a PREraser
treatment to the samples.
19 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
PCR Setup
Before you begin
Make sure you have switched on the computer, the PCR instrument and ensure the sample
layout for the PCR plate is suitably documented and programmed.
Only remove the number of PCR strips, containing pre-dispensed MasterMix, necessary for
testing the amount of samples to be analysed and store the remainder at -20 °C ± 2 °C.
Important: In case of cutting an 8-strip into smaller parts it is crucial to let the plastic warm up
at room temperature for around 1 minute prior to cutting. Do not place the strip into a cooling
block.
Note: Frequent freezing and thawing might cause inactivation of the reagents. Do not
freeze/thaw more than 3 times.
Spin down the strips with MasterMix for 30 sec between 400 2000 × g.
Visually check that all MasterMix is in the bottom of the wells.
1. Place PCR strips, together with the yellow PCR support grid, into an appropriate support rack. Avoid
exposing the PCR plate to light for long periods of time.
2. The lysis and PCR plate should be arranged in the same orientation in order to facilitate handling
with the multichannel pipette and in order to avoid cross contamination of wells (place the lysis plate
i.e. behind the PCR plate).
3. Open lysis plate as described in section 2.6.
4. Open the corresponding PCR strip by fixing the upper and lower end with one hand and removing
the foil starting with the loose end from top to bottom in 180 degree angle (see Figure 1A).
Note: Open each strip individually, add samples and close. Only then proceed to the next strip.
5. Transfer 5 µL of each lysate by using an 8-well multichannel pipette to the corresponding PCR strip.
Note: Make sure that the lysate is taken from the upper half of the lysis volume, avoid touching a
potential pellet of debris. Ensure the lysate is pipetted directly on top of the MasterMix.
6. Add 5 µL positive control DNA (vial with yellow cap) to well A1.
7. Close the PCR strip with an optical cap strip with clean, gloved hands (see Figure 1C).
8. Re-close the lysis strip with fresh domed caps (not supplied with the kit).
9. Repeat steps 3 - 8 with the remaining strips.
10. Transfer the PCR plate to the real-time instrument and start the run (refer to section 3 for Agilent
AriaMx™ + section 4 for Bio-Rad CFX96 Touch™).
11. Store the lysis plate at -20 ± 2 °C in case a re-run of the PCR is required.
20 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Figure 1 A. Removing foil from PCR strips. B. Incorrect way to close the lids in one direction. C. Correct
way to close lids by softly fixing the ends of the cap strip to the reaction tubes, then applying even
pressure to tightly close the PCR strips.
21 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
The BACGene Listeria monocytogenes kit is validated for the use on the Agilent AriaMx™ with software
versions up to 1.5.
The PCR platform should be prepared prior to commencing the practical work. For a more detailed
description of the instrument programming please refer to the user manual of the Agilent AriaMx™.
BACGene offers two methods for evaluating the final results. Users can choose between the automated
evaluation software FastFinder or the BACGene Evaluation Sheet. Evaluation specific comments are
marked in the protocol below in coloured boxes:
Evaluation sheet only
FastFinder only
The following instructions outline how to set up a PCR run on the Agilent AriaMx™ using an external
computer.
Note: If you are using the cycler via the touch screen without an external computer please contact us at
kits@eurofins.com for more information.
Note: The Aria software is able to interface with a Laboratory Information Management System “LIMS”.
Please contact kits@eurofins.com for more information.
Note: The necessary run template (.amxt) file is available via kits@eurofins.com. Alternatively, a run
template can be generated from LIMS, please contact us for more information.
1. Turn on Agilent AriaMx™ cycler and the remote computer.
2. Open the Agilent AriaMx™ software and continue via the software:
3. Go to “File”, then “Open…”
4. Browse to find the required run template file.
Evaluation sheet only
22 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
“Target Name”- in order to check target names (table below) click on the arrow next to “Target”
on the right side of the screen. If more than one system is analysed on one plate, please use
the following target name for respective wells of the BACGene Listeria monocytogenes kit and
verify the target names for the other assays with the respective assay manual:
Note: The target names have to be correct to ensure the correct evaluation using the BACGene
Evaluation Sheets or FastFinder software. The correct target names have already been assigned to the
wells in run templates provided. For more information please contact us via kits@eurofins.com.
Check sample type definitions
Evaluation sheet only
1 HOLD 42 CYCLES
enzyme activation denaturation annealing & extension
15 min at 95°C 15 sec at 95°C 60 sec at 60°C
no data collection no data collection data collection
Important: The run should always be initiated using a template containing the required information!
10. Select “Run” in the “Thermal Profile” tab to open the “Instrument Explorer”.
11. Select the correct AriaMX cycler and click on “Send Config”
12. If required, enter your login details for the selected instrument.
13. Save the experiment under the desired name and in the desired location.
23 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
FastFinder only
Note: For the use with FastFinder save run files in the folder defined as “To be processed” in the
software (see separate short guide “FastFinder for BACGene kits”).
14. The run information has now been sent to the selected instrument.
15. The following icon will appear on the bottom right of the screen of the cycler:
16. Click the icon and press “Open Primed Experiment” on the pop-up.
Note: In case “Sync Plate” message appears, please confirm and continue.
17. The experiment will open on the screen of the cycler.
18. Under “Thermal Profile” click “Run Experiment”.
After run completion, data should be evaluated using either the BACGene Evaluation Sheet or the
FastFinder software. The data export for the evaluation using the BACGene Evaluation Sheet is outlined
below.
The evaluation using FastFinder is done using the run files directly without requiring an additional data
export, for more details refer to separate short guide “FastFinder for BACGene kits”.
Note: Please do not use the “Disc Symbol” option on the touch screen for saving runs on a USB stick.
Only use the “Copy & paste” option as described in the instrument documentation.
Evaluation sheet only
Export of Data for the Evaluation Using the BACGene Evaluation Sheet
12. Open Agilent AriaMx™ software (up to v 1.5) on your PC and open the run file by going to “File”,
“Open…” and browsing to the correct location.
Note: If you are using the cycler via the touch screen without an external computer please contact us
at kits@eurofins.com for more information.
13. Ensure all relevant wells are selected in the “Plate Setup” screen.
14. As a final check before export go to “Graphical Displays” under the “Analysis” header and ensure
that the “Background Based Threshold” is set to Cycle Range “5 thru 9” and “Sigma Multiplier 10”.
This option can be found by clicking the arrow next to “Amplification Plots” on the right side of the
software interface.
15. To prepare a default configuration for result export, follow these instructions. This step only needs
to be carried out the first time you use the software. If the default configurations have already been
selected continue to step 5.
Go to “Export Data” tab.
24 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
For evaluation using FastFinder refer to the separate short guide “FastFinder for BACGene kits”.
25 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
The BACGene Listeria monocytogenes kit is validated for use with the Bio-Rad CFX96 Touch™ and
Bio-Rad CFX96 Touch™ Deep Well with CFX Manager™ software version 3.1 / CFX Maestro™
software version 1.1.
The PCR platform should be prepared prior to commencing the practical work. For a more detailed
description of the instrument programming please refer to the user manual of the Bio-Rad CFX96
Touch™ or CFX96 Touch™ Deep Well.
BACGene offers two methods for evaluating the final results. Users can choose between the automated
evaluation software FastFinder or the BACGene Evaluation Sheet. Evaluation specific comments are
marked in the protocol below in coloured boxes:
Evaluation sheet only
FastFinder only
The following instructions outline how to set up a PCR run on the Bio-Rad CFX96 Touch™ and Bio-Rad
CFX96 Touch™ Deep Well with an external computer.
Note: If you are using the cycler via the touch screen without an external computer please contact us
at kits@eurofins.com for more information.
26 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Note: The required protocol and plate layout templates are available via kits@eurofins.com.
Furthermore, the CFX Manager™/ CFX Maestro™ 1.1 software can be configured for use with a
Laboratory Information Management System “LIMS”. Please contact kits@eurofins.com for more
information.
Note: This procedure only needs to be carried out the first time you are running the assay.
27 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Make sure that the correct protocol and plate layout templates have been imported into the Express
Load folder according to section 4.1.1. Alternatively, a LIMS template import can be generated directly
from LIMS, for more information please contact us at kits@eurofins.com.
1. Open the CFX Manager™ 3.1 or CFX Maestro™ 1.1 software.
2. In the “Startup wizard” select the correct instrument in the “Select instrument” drop-down menu.
3. Under “Select run type” click “User defined”
4. In the “Run setup” window select the BACGene protocol template from the “Express load” drop-
down menu.
Important: The run should always be initiated using the correct protocol and layout templates.
5. Prior to starting the run check that the following parameters are correct:
In the “Protocol” tab check:
- Sample volume: 25 µL
- Thermal cycler times and temperatures
- Thermal Profile:
1 HOLD 41 REPETITIONS
enzyme activation denaturation annealing & extension
15 min at 95°C 15 sec at 95°C 60 sec at 60°C
no data collection no data collection data collection
6. Click on “Next” and open the plate layout template from the “Express load” drop-down menu:
Evaluation sheet only
28 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
9. Sample names can be added either manually, via copy and paste or importing a “*.csv” file (click on
“Import”).
10. After adding sample names, close the spreadsheet view by clicking “OK”.
11. In the “Plate editor” check:
Plate Type: BR White
Scan Mode: All Channels
Fluorophores: VIC™ and Cy5™ are selected for all relevant 96 wells
Target Names: To check, click “Edit Selected” and the target names are shown on the right
side of the screen, compare with target names in table below. If more than one system is
analysed on one plate, please use the following target names for respective wells of the
BACGene Listeria monocytogenes kit and verify the target names for the other assays with
the respective assay manual:
Note: The target names have to be correct to ensure the correct evaluation using the BACGene
Evaluation Sheet or FastFinder software. The correct target names are assigned to the wells in
templates provided by us. For more information please contact us via kits@eurofins.com
Check sample type definitions:
Evaluation sheet only
29 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
12. Click “OK” to close the “Plate Editor” and select “Next” in the “Run setup” tab to get to the “Start
Run” tab.
13. Select the cycler from the list.
14. Click “Open Lid”, load the plate into the cycler and then press “Close Lid”
15. Click “Start Run”.
16. In the directory, choose the location to save the run file and then click “Save”
30 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
After run completion, data should be evaluated using the BACGene Evaluation Sheet or FastFinder
software. The data exports from the Bio-Rad CFX96 Touch™ or Bio-Rad CFX96 Touch™ Deep Well
for the BACGene Evaluation Sheet are described below.
The evaluation using FastFinder is described in the separate short guide “FastFinder for BACGene kit”.
Evaluation sheet only
Export of Data for the Evaluation Using the BACGene Evaluation Sheet
1. Open Bio-Rad CFX Manager™ software version 3.1 / CFX Maestro™ software 1.1 and select the
saved run. This can be done using one of the following options:
a. Double click on the saved run file. This will also automatically open the Bio-Rad CFX Manager™
3.1 / CFX Maestro™ 1.1 software.
a. Files can be dragged and dropped into the open Bio-Rad CFX Manager™ 3.1 / CFX Maestro™
1.1 software.
b. In the open software, go to “File”, then click “Open” and select the run file via the directory.
Note: If you are using the cycler via the touch screen without an external computer please contact us
at kits@eurofins.com for more information.
2. Once the run file is opened, go to “Settings” tab and ensure the following options are selected:
Cq Determination Mode = Single Threshold.
Baseline Setting = Baseline Subtracted Curve Fit.
Ensure that “Apply Fluorescence Drift Correction” is selected under “Baseline Settings”.
3. To export results, follow these instructions:
Go to “Export” tab.
Select “All Data Sheets” and export as text files (*.txt). Choose a location for the files.
Got to “Export” and select “Custom Export” choose text format (*.txt) with the following columns
and save in chosen location:
- Well
- Fluorophore
- Target Name
- Content
- Sample Name
- Cq
- End RFU
Important: The order of the exported columns is mandatory.
Note: Save all export files in the same folder.
4. Click “Export”
5. In the “Export Complete” pop-up, click “Ok”
6. Proceed to BACGene Evaluation Sheet for data interpretation.
31 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
FastFinder only
For evaluation using FastFinder refer to the separate short guide “FastFinder for BACGene kits”.
5 DATA INTERPRETATION
Data is interpreted using the FastFinder evaluation software or ‘BACGene Evaluation Sheet’ provided.
For FastFinder evaluation please refer to the “FastFinder for BACGene kits” manual.
For an evaluation using the evaluation sheet upload data as described in ‘BACGene Evaluation Sheet’
file.
Interpretation of final sample results is summarized in the following table:
In the context of NF VALIDATION and AOAC-PTM certification, all samples identified as positive by the
BACGene Listeria monocytogenes kit must be confirmed by (one of) the following test(s):
Option 1:
Using the conventional tests described in the methods standardized by CEN or ISO from colonies
(including the purification step). The confirmation step must start from the enrichment broth or from
typical colonies isolated on selective media.
Option 2:
Direct streaking of 10 µL from enrichment in Actero™ Listeria Enrichment Media (by FoodChek™) onto
O&A agar.
Incubation of the O&A plates at 37°C ±1°C for 24 ± 3h and if necessary (negative) further 24h ± 3h.
Check for the presence of characteristic colonies which are presumed to be Listeria monocytogenes.
The presence of characteristic colonies is sufficient to confirm the presence of Listeria monocytogenes.
32 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
33 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
7 TROUBLESHOOTING
BACGene offers two methods for evaluating the final results. Users can choose between the automated
evaluation software FastFinder or the BACGene Evaluation Sheet. Evaluation specific troubleshooting
comments are marked below in colored boxes:
Evaluation sheet only
FastFinder only
Observation:
MasterMix plate is not frozen upon arrival and without appropriate cooling (dry ice)
Observation
MasterMix was thawed.
34 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
Frozen MasterMix film covering the well
Observation
Liquid MasterMix film covering the well
Observation
MasterMix is spilling during removing the seal
Observation
Optical cap strips do not align with wells and plastic is damaged
35 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
Data import into the BACGene Evaluation Sheet was not successful/possible.
Agilent AriaMx™:
Ensure that all relevant fluorophores/targets are selected in the
“Graphical Displays” section for all relevant wells.
Ensure that all columns are selected for export in “Tabular results”
file (see section 3.2).
36 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
FastFinder only
Observation
The samples are not automatically labelled (sample definition and/or assay type) on the plate
layout in the FastFinder analysis
Observation
No positive/negative controls set. Evaluation is not possible.
Alternatively: Define the respective Tasks in the original run file and
repeat the export and evaluation.
FastFinder only
Restart analysis and identify/select correct location of C+ and C-.
37 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
The negative control (C-) is positive for at least one target assay.
Observation
C- not in range (in the IPC system) - evaluation not possible.
Evaluation of the IPC is dependent on the C- control reaction. If this reaction is not in range, the IPC
cannot be evaluated in the sample reactions.
38 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
C+ is negative/too late in the required target system - evaluation is not possible.
Note: The warning message „Late Cq value of C+“ is an indication reduced performance of the C+.
However, evaluation of the results is still valid. Also review troubleshooting recommendations below.
The positive control was If incorrect storage is suspected, use a fresh tube of positive control
not stored correctly. material. When starting a new kit, always use the positive control
material from that kit.
The positive control Use a fresh tube of positive control material. Do not freeze/thaw
material has undergone too more than 6 times. If C+ continues to show a negative signal,
many freeze-thaw cycles. contact Eurofins GeneScan Technologies.
The frozen MasterMix was Always visually check the level of the MasterMix before use. If the
not in the bottom of the MasterMix is not at the bottom of the tubes, allow to thaw, spin
PCR tube, but in the lid. down and re-freeze before use.
C+ has not been assigned Make sure that the correct well for C+ is selected.
correctly.
Observation
Cq value of the C+ is too early (in the required target system) – evaluation is not possible.
Observation
Linear, straight lines instead of amplification curves or amplification curves that lay completely
below the threshold
39 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
A “Q” or invalid result is observed for sample
40 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Important: Before manually adjusting any settings, please make sure that all steps described in the
data export section 3.2 (for AriaMx™) and 4.2 (for Bio-Rad CFX96 Touch™ Standard and Deep Well)
were followed.
The interpretation of real-time PCR runs depends on automated algorithms that analyse and interpret
the measured fluorescence data.
These algorithms are thoroughly validated by the manufacturers of the PCR cyclers and the respective
evaluation software and provide reliable results for the vast majority of analyses.
In rare cases, when the automated threshold calculation of the cycler software fails, manual adjustments
are reasonable and justified in order to get a correct interpretation of the measured data (e.g. automatic
threshold settings might calculate a threshold which is not set in the exponential phase of the
amplification curves).
Figure 2. Exemplary amplification plot of a real -time PCR in linear scale (Graph Type “Linear”). 1,
background; 2, exponential amplification phase; 3, linear amplification phase; 4, threshold line.
not part of the AOAC validation study.
41 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Figure 3: Exemplary amplification plot of a real-time PCR in linear scale (Box “Log Scale” is unchecked).
1, background; 2, exponential amplification phase; 3, linear amplification phase; 4, threshold line.
42 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Observation
The positive result in the Listeria monocytogenes PCR assay(s) cannot be confirmed by culture
(no isolated typical colonies).
Quality of selective agar Make sure agar plates are quality tested (e.g. according to ISO
plates is not sufficient. 11133).
The detection level of the Subculture 100 µL of the pre-enrichment in 9.9 mL Full Fraser and
confirmation method is not incubate up to 48h at 37°C ± 1°C or/and spread 100 µL pre-
reached (e.g. for slow- enrichment onto O&A plates (for samples with low levels of bacterial
growing Listeria background).
monocytogenes serovars).
Target organism cannot be Make sure the selective agar plates are prepared and stored
detected on plates as they according to the manufacturer’s instructions and quality-controlled
are overgrown with (e.g. according to ISO 11133).
background flora.
Contamination with Make sure all recommendations regarding DNA contamination
genomic (free DNA from prevention are followed and take measures to identify the source of
dead bacteria) in the a possible contamination (i.e. environmental swabbing for DNA
environment causes contamination and/or block contamination test for Agilent AriaMx).
positive PCR results.
Free DNA from dead target General:
organism cell in the sample Use of PREraser BACGene is recommended for samples in which a
produces a positive PCR high level of free DNA from dead cells is expected. PREraser is
result. applied before lysis of the living cells in order to eliminate signals
from free DNA. Refer to the PREraser BACGene manual and
relevant certifications for details about the scope and applicability of
the method for the matrix in question.
Environmental samples:
Environmental samples should be taken in a way that reduces
sampling of dead bacteria, e.g. by rinsing of surfaces after
decontamination to remove dead bacteria.
Make sure that swabs are enriched according to certified protocol.
43 / 44
BACGene Listeria monocytogenes
Cat. No. 5123222001, 5123222011 and
5123222010
Eurofins GeneScan Technologies GmbH (“GeneScan”) warrants the products manufactured by it will
be free of defects in materials and workmanship when used in accordance with the applicable
instructions before the expiration date marked on the product packaging and when stored under the
storage conditions recommended in the instructions and/or on the package. GeneScan makes no other
warranty, expressed or implied. There is no warranty of merchantability or fitness for a particular
purpose.
GeneScan’s sole obligation with the respect to the foregoing warranties shall be, at its option, to either
replace or to refund the purchase price of the product(s) or part thereof that proves defective in materials
or workmanship within the warranty period, provided the customer notifies GeneScan promptly of any
such defect. GeneScan shall not be liable for any direct, indirect or consequential damages resulting
from economic loss or property damages sustained by buyer or any customer from the use of the
product(s). A copy of Eurofins GeneScan Technologies GmbH terms and conditions can be obtained
on request, and is also provided in our price lists.
This kit is developed, designed, and sold for research purposes only. It is not to be used for diagnostic
purposes or analysis of food and feed unless expressly cleared for that purpose by the competent
regulatory authorities in the country of use. All due care and attention should be exercised in the handling
of the materials described in this text.
10 IMPORTANT NOTES
Registered names, trademarks, etc. used in this document, even when not specifically marked as such,
are not to be considered unprotected by law.
Eurofins GeneScan Technologies GmbH is only responsible for the content of this manual in its original
language (English) and not liable for any translations.
44 / 44
TECHNICAL SUPPORT SERVICE
For technical assistance and more information please reach out to your local sales
support as stated on the Eurofins Technologies website or reach out to your local
distributor.
Eurofins Technologies
Fóti út 56.
1047 Budapest
Hungary
https://www.eurofins-technologies.com/contacts
© Eurofins GeneScan Technologies GmbH, all rights reserved.