UM Diomede TA
UM Diomede TA
User manual
2022-10-30
010.027.012
Atlantamed.co.uk
Diomede user’s manual
AtlantaMed Ltd.
Unit 2, Kingsmill Business Park, Kingston upon Thames
KT1 3GZ, London, UK
Phone: +44 20 3417 9600
www.atlantamed.co.uk
Atlantamed English|010.027.012
Diomede user’s manual
Atlantamed English|010.027.012
Diomede user’s manual
TABLE OF CONTENT
1. INTRODUCTION 1
1.1. Brief device description 2
1.2. Intended use 2
1.3. Intended user 2
1.4. Intended environment 2
1.5. Warranty information 2
1.6. Disclaimer 2
1.7. Technical support 3
2. SAFETY GUIDELINES 4
2.1. Overview 5
2.2. Signal words 5
2.3. Electromagnetic Interference 5
2.4. Placement 5
2.5. Back up ventilation 6
2.6. Electrical hazards 6
2.7. Power supply and battery 6
2.8. Supply Gases 7
2.9. Detachable parts and accessories 7
2.10. Inadvisable modifications 7
2.11. General operating conditions 8
2.12. Cleaning and Maintenance 8
2.13. Disposal 8
2.14. Storage 9
2.15. Transport 9
2.16. Contraindications 9
3. SYSTEM OVERVIEW 10
3.1. Main components of Diomede 11
3.2. Standard and Optional Accessories 12
3.3. Front view 13
3.4. Rear view 14
3.5. Symbols and their definitions 15
Atlantamed i English|010.027.012
Diomede user’s manual
Atlantamed ii English|010.027.012
Diomede user’s manual
5. VENTILATION SETTINGS 39
5.1. Parameter Settings 40
5.1.1. Ventilation control parameters 41
5.1.2. Control parameters in each ventilation mode 43
5.2. Others menu 44
5.2.1. Capnography setting 45
5.2.2. O2 suction 45
5.2.3. Tube compensation 46
5.2.4. Leak compensation 48
5.2.5. Volume compensation 48
5.2.6. Humidifier 48
5.3. Adjustable Alarm Limits 49
5.4. Start Ventilation 49
5.5. Stop ventilation 50
5.6. Patient type readjustment 50
5.7. Emergency ventilation 50
5.7.1. Back to normal ventilation 51
6.2.11. Flow-cycled 56
6.2.12. Time-cycled 56
6.2.13. Expired minute volume 56
6.2.14. O2 concentration 56
6.2.15. Mandatory ventilation 56
6.2.16. Spontaneous breath 57
6.2.17. Assisted breath 57
6.2.18. Supported breath 57
6.3. Sigh 58
6.3.1. Pressure limit 58
6.3.2. Added VT 58
6.3.3. Per hour 58
6.3.4. Number 58
6.4. Apnea ventilation 59
6.4.1. Apnea time 60
6.4.2. Apnea Ventilation parameters 60
6.5. VCV (Assisted) 61
6.5.1. How it works 61
6.5.2. Ventilation parameters 61
6.5.3. Other functions available 61
6.5.4. Special considerations 62
6.6. PCV (Assisted) 62
6.6.1. How it works 62
6.6.2. Ventilation parameters 62
6.6.3. Other functions available 62
6.6.4. Special considerations 63
6.7. PRVC (Assisted) 63
6.7.1. How it works 63
6.7.1. Ventilation parameters 64
6.7.2. Other functions available 64
6.7.3. Special considerations 64
6.8. VSIMV+ PS 64
6.8.1. How it works 64
6.8.2. Ventilation parameters 65
6.8.3. Other functions available 65
6.9. PSIMV+PS 66
Atlantamed iv English|010.027.012
Diomede user’s manual
7. NEONATAL VENTILATION 73
7.1. General 74
7.2. Apnea ventilation 74
7.3. TCPL 74
7.3.1. How it works 74
Atlantamed v English|010.027.012
Diomede user’s manual
Atlantamed vi English|010.027.012
Diomede user’s manual
Atlantamed ix English|010.027.012
LIST OF TABLES
Atlantamed x English|010.027.012
Table 12-3 monitor specifications 138
Table 12-4 ambient conditions 138
Table 12-5 audible sound pressure emission 138
Table 12-6 Essential performance 139
Table 12-7 Pneumatic specifications 140
Table 12-8 electrical specifications 142
Table 12-9 control parameters, ranges and accuracy 143
Table 12-10 control parameters, ranges and accuracy 145
Table 12-11additional functions 147
Table 12-12 measurement accuracy 148
Table 12-13 monitored parameters 149
Table 12-14 adjustable alarm limits 151
Table 12-15 electromagnetic emission 152
Table 12-16 Electromagnetic Immunity 153
Table 12-17 Electromagnetic immunity 154
Atlantamed xi English|010.027.012
LIST OF FIGURES
1. INTRODUCTION
Atlantamed 1 English|010.027.012
Introduction
Diomede is a life supporting critical care Diomede ventilator comes with a warranty
ventilator that helps patients with on all parts excluding single use parts and
respiratory problems who need invasive or those accessories supplied by other
noninvasive ventilation. manufacturers, which starts on purchase
date.
Diomede can deliver a precisely controlled
mixture of oxygen and air at controlled The single use parts or parts with limited
pressure or volume into the patient’s lung. service life must be replaced on a regular
basis in accordance with the information
In addition to the conventional ventilation
provided in the user’s manual to secure
modes, the Diomede offers a wide range the proper functioning of the product.
of modes from which the operator can
select the one perfectly tailored to the This warranty covers manufacturing
needs of each patient. defects in materials and workmanship
within the aforementioned window of time.
The Diomede is a mains powered device
to provide patients with continuous This warranty takes effect only if the
ventilator support. It can, however, run on warranty certificate is dated, fully and
the internal backup battery in case of correctly filled out by an authorized
emergency or intrahospital transportation. AtlantaMed Ltd. dealer.
Atlantamed 2 English|010.027.012
Introduction
Atlantamed 3 English|010.027.012
2. SAFETY GUIDELINES
2.1. Overview 5
2.2. Electromagnetic Interference 5
2.3. Placement 5
2.4. Back up ventilation 6
2.5. Electrical hazards 6
2.6. Power supply and battery 6
2.7. Supply Gases 7
2.8. Detachable parts and accessories 7
2.9. Inadvisable modifications 7
2.10. General operating conditions 8
2.11. Cleaning and Maintenance 8
2.12. Transport 9
2.13. Contraindications 9
Atlantamed 4 English|010.027.012
Safety Guidelines
2.1. Overview
NOTICE
⚠
Manual.
Warning
The ventilator system as a restricted
medical device may only be operated by The Diomede ventilator can be vulnerable
trained personnel under direct medical to being undesirably affected in
supervision of a hospital physician. electromagnetic environment.
Bear in mind that only common safety Install the Diomede ventilator in
issues regarding the general operation of accordance with precautionary measures
the Diomede ventilator has been taken in introduced in technical specifications
to consideration in this chapter. regarding EMC (chapter11). Otherwise,
WARNINGS and CAUTIONS associated performance deterioration of the
with the subsystems or any other specific equipment may occur.
features will appear in the relevant
In order to avoid malfunctions during
chapters.
operation, increase the separation
2.2. Signal words between Diomede and high-frequency
devices including cell or wireless phones,
defibrillators, and electrocautery devices.
The choice of signal word depends on the
risk of harm and the severity of the harm In the vicinity of magnetic resonance
for those in contact with the Diomede or imaging (MRI) equipment, the proper
the Diomede ventilator itself. performance of the ventilator may be
ruined. This could damage the ventilator
permanently, and threaten the patient’s
⚠ Warning life.
Portable and mobile RF communications
A WARNING indicates that failure to pay equipment should not be used close to
attention to this information may lead to any part of ventilator, including cables.
injury, death or other severe
The recommended distance from such
consequences.
equipment can be calculated based on the
frequency of the transmitter (see 12.16)
⚠
A CAUTION indicates that failure to pay
attention to the information provided for Caution
the user may cause damage to the device
itself or other property. The Diomede is designed to be used at
the temperature range of 10-35°C and
altitude lower than 4000 meters. Using the
Atlantamed 5 English|010.027.012
Safety Guidelines
device outside these conditions can affect Do not remove the back panel under any
its performance. circumstance, while the ventilator is in use
it might cause an electrical shock. Doing
Do not place the Diomede in such a way
otherwise will be assumed as a breach of
that restricts the flow of cooling air, which
contract and the manufacturer will be
leads the device to overheat. Leave
under no obligation and no longer have
around 20 cm of space on either side of
any liabilities in connection with the
the ventilator.
product.
Make sure the loudspeaker is not covered
Antistatic or electrically conductive hoses
or positioned too close to the wall so that
or tubing shall not be used in the breathing
the alarm sound can be easily heard.
system and gas supply hoses.
To avoid performance loss and
degradation, prevent stacking or using any
2.7. Power supply and battery
devices on Diomede or closer than 30 cm.
Do not block the anti-suffocation inlet
located next to oxygen and air inlets in
⚠ Caution
order to avoid interference with patient The Diomede can run on internal backup
ventilation. battery for a limited duration of time. If the
backup battery is discharged, with no
2.5. Back up ventilation external supply power, the device will stop
working.
Atlantamed 6 English|010.027.012
Safety Guidelines
⚠
parts.
Caution
The ventilator system must be serviced at
Connect Diomede only to the accessories regular intervals by personnel who have
recommended by the manufacturer. received authorization and specialized
training by the manufacturer.
Using cables, accessories, and
transducers other than ones provided by
the manufacturer could lead to expanded
Atlantamed 7 English|010.027.012
Safety Guidelines
Service, repair and installation must only The device shall be serviced by the
be performed by personnel authorized by authorized service personnel every 5000
the manufacturer. hour of operation.
Make sure these components match the Do not use acetone, chloroform, strong
associated intended use for the target acidic substances or chlorinated solvents
patient group. to clean the plastic parts and hoses.
The alarm system can only warn you of an Do not use sodium hypochlorite (bleach)
adverse condition. Therefore, the at1000 ppm to clean the patient circuit
ventilated patient must not be left hoses.
unattended. Some situations need Do not use Ethylene oxide to sterilize,
immediate corrective care. because it may damage the ventilator
The user must be a trained medical components.
professional such as a nurse, physician, or Do not use Iodophor and phenolic to
critical care expert who have a thorough disinfect the exterior surfaces.
knowledge of the device and its functions.
2.13. Disposal
While a patient is connected to the
ventilator, do not connect it to any external
equipment that fails to satisfy electrical Do not dispose of the electronic device or
safety requirements for medical devices.
its parts (O2 sensor, battery, etc.) with
2.12. Cleaning and Maintenance ordinary waste.
When replacement of any parts required,
disposal of them must be according to the
⚠ Caution hospital’s protocols.
Comply with all your country’s regulations
relevant to environmental protection.
See their instructions for use for those
accessories supplied by other
manufacturers.
Atlantamed 8 English|010.027.012
Safety Guidelines
2.14. Storage
2.15. Transport
⚠ Caution
The Diomede ventilator only allows
intrahospital transportation.
The device is not intended for
transportation outside the hospital or for
use in the home environment.
Check the battery charge level before
transporting the patient on ventilator within
the hospital.
2.16. Contraindications
⚠ Caution
Turning the ventilation system on without
any specialist physician assistance.
Connecting the ventilation system to
inadequate mains power.
Using the ventilator when it is not
protectively earthed.
Mobilization outside healthcare settings.
Driving an anesthesia machine with this
device.
Using the ventilator in the presence of
flammable anesthetic gases.
Locating the ventilation system nearby the
magnetic resonance equipment or
electromagnetic radiation.
Using the ventilator in a hyperbaric
chamber.
Using the ventilator in an oxygen rich
environment.
Atlantamed 9 English|010.027.012
3. SYSTEM OVERVIEW
Atlantamed 10 English|010.027.012
System Overview
③
⑥
Atlantamed 11 English|010.027.012
System Overview
⚠ Warning
Atlantamed 12 English|010.027.012
System Overview
③
①
④ ②
⑤
⑥
Atlantamed 13 English|010.027.012
System Overview
Ventilation unit
1. Cooling fan
2. Loud speaker
3. Air inlet
4. Oxygen inlet
②
5. Anti-suffocation inlet
6. LAN port
7. USB port ①
8. Hard Fault silent button
9. Fuse ⑥
10. AC Power inlet
11. On/Off button and indicator
⑦ ⑤
⑧ ④
⑪ ③
Compressor unit ⑩
12. Air inlet ⑨
13. Air outlet
⑮ ⑫
14. Power switch
15. Power cord connector ⑭ ⑬
Atlantamed 14 English|010.027.012
System Overview
Expiratory port
Table 3-2 Symbols and their definitions
Humidifier
Symbol Definition
Nebulizer
Authorized representative
O2 Sensor O2 sensor
Serial number
Capno Capnograph
Manufacturer
⚠ Caution
Standby
Fuse
USB Port
Ethernet connector
Operation instruction
inlet
Atlantamed 15 English|010.027.012
System Overview
①
③
Atlantamed 16 English|010.027.012
System Overview
Atlantamed 17 English|010.027.012
System Overview
3.7.2. Features in common in all The operator can have access to all
pages pages, menus and control buttons via the
rotary knob and the touch screen.
⑩ ⑨ ⑦ ⑧ ㉓ ⑤ ④ ③② ①
⑰ ⑯ ⑮ ⑭ ⑬ ㉒ ⑫ ⑪
㉑
⑱ ⑲ ⑳
Atlantamed 18 English|010.027.012
System Overview
• No alarm is activated.
Selected mode
8 pending for
confirmation
Atlantamed 19 English|010.027.012
System Overview
• Volume compensation is in
15 Volume Compensation
progress.
• Tube compensation is in
16 Tube Compensation
progress.
• Leak compensation is in
17 Leak Compensation
progress.
Atlantamed 20 English|010.027.012
System Overview
Atlantamed 21 English|010.027.012
System Overview
Atlantamed 22 English|010.027.012
System Overview
Touchscreen
• Touch the screen and swipe left or right The borders of the adjusted parameter
to move from one page to another. pending to be confirmed will turn yellow
(figure 3-9).
• Tap on a dot in navigation bar so you will
be taken to the corresponding page.
Rotary knob
Or
• Touch the parameter required to be
adjusted. The color of the selected item
• Use +/- to assign a new value to the
will change. An input field will open on the
parameter.
right. (figure 3-7)
• Touch apply to save the change
Atlantamed 23 English|010.027.012
System Overview
Atlantamed 24 English|010.027.012
System Overview
3.8. Device Settings Using the clock, you can define day time
and night time. The clock is divided into
two parts. The dark green part and the
You can adjust the display brightness and light green part represent the night time
alarm volume as well as time and date in and day time respectively.
this page.
To adjust the day time and night time
• Touch the screen and swipe left to reach
the last page. • Touch one of the radial lines and
slide your finger around to set one
3.8.1. Time and Date end of the period.
⚠ Caution
Make sure the auditory alarm signal sound
pressure levels are higher than ambient
levels. Otherwise, the operator may lose
the recognition of alarm signals activated.
Atlantamed 25 English|010.027.012
System Overview
To adjust sound level and brightness • Drag and drop the help icon on any item
during the night and day time or area of the touchscreen about which
you are looking for information.
• Touch and drag the slider up or down
• A window will open which contains
• Tap the apply button to save changes.
useful information.
The display setting will be changed
accordingly when night is turned to day or • Scroll down if necessary to access all
vice versa. the information.
3.8.3. Theme
3.9. Help
Atlantamed 26 English|010.027.012
System Overview
Atlantamed 27 English|010.027.012
4. PREPRATION FOR INSTALLATION
4.1. power supply 29
4.2. gas supplies 29
4.3. patient breathing circuits 32
4.4. set up a humidifier 32
4.5. set up a nebulizer 33
4.6. set up a proximal flow sensor 34
4.7. set up a capnograph 34
Atlantamed 28 English|010.027.012
Preparation for Installation
Atlantamed 29 English|010.027.012
Preparation for Installation
Atlantamed 30 English|010.027.012
Preparation for Installation
Atlantamed 31 English|010.027.012
Preparation for Installation
The size of breathing tube for adults, Figure 4-7 O2 sensor installation
pediatrics and neonates differs from one
to another according to the patient’s age
group. The diameters are 22, 15 and 10
mm respectively.
Atlantamed 32 English|010.027.012
Preparation for Installation
⚠ Caution ⚠ Caution
Atlantamed 33 English|010.027.012
Preparation for Installation
NOTICE
⚠ Caution Capnograph is not used in neonatal
Make sure the flow sensor tube is inclined ventilation.
at an appropriate angle at which moisture
accumulation will not occur. Otherwise, Capnograph uses infrared waves to
flow sensor measurements will be prone measure CO2. It is founded on the
to inaccuracies. principle that CO2 molecules absorb
infrared light energy at specific
NOTICE wavelengths.
adapter.
While the exhaled gas is passing through
the airway adapter, the CO2 sensor emits
a beam of infrared light. A fraction of the
Atlantamed 34 English|010.027.012
Preparation for Installation
Atlantamed 35 English|010.027.012
Preparation for Installation
4.9. Turn on the device 3. Slide right to turn the ventilator off as
shown in
⚠ Caution
Atlantamed 36 English|010.027.012
Preparation for Installation
Atlantamed 37 English|010.027.012
Preparation for Installation
4.11.5. Start tubing examination When the test has passed successfully,
the next arrow icon turns yellow and the
ventilator will go to standby mode.
• To start the tubing examination, Tapping on the next arrow icon will take
block the Y-piece by hand and then you to the Parameter Setting page.
select “Start”.
See chapter 5 for more details on when to
By passing each test, the message start ventilation.
“passed” will appear on the screen in
green text next to them.
During the tubing examination, if the ⚠ Warning
device detects a leak rate less than
10L/min, a message “LEAK WARNING” To prevent the possible failure in the
will appear on the screen. Expiratory Volume, PEEP, and Trigger,
AtlantaMed Ltd. recommends not to
By selecting “OK”, the message “Passed” connect the patient to the ventilation
will appear in the tubing examination system for the leak rate higher than
window. 10L/min.
Atlantamed 38 English|010.027.012
5. VENTILATION SETTINGS
Atlantamed 39 English|010.027.012
Ventilation Settings
Atlantamed P a g e | 40
Ventilation Settings
Parameter Definition
VT Volume of gas that enters and leaves the lung during a breath
I:E Ratio of the inspiratory time to the expiratory time in a respiratory cycle
Number of assured inflations that are set to occur in a specified period of time,
F
expressed as breaths per minute
Quantity by which the baseline airway pressure is set to be positively offset from the
PEEP
ambient pressure
PEEPHigh higher baseline airway-pressure level in ventilation-modes labelled as bi-level AV
Indication of the time for the regulated parameter to rise to a set value following the
Rise Time
initiation of an inflation
Inflation-type that acts to generate a constant airway pressure, after a set rise time. Only
PS made available for selection with ventilation-patterns where it cannot be initiated other
than in response to a patient-trigger event
Atlantamed P a g e | 41
Ventilation Settings
Parameter Definition
A fraction of inspiratory peak flow (Exp. Trig) that determines the end of inspiration in
Exp. Trig. flow cycled ventilation modes. When the flow rate reaches this amount the ventilator
cycles from the inspiration phase into the expiratory phase.
Volume of gas either passing to or leaving the lung during inspiratory or inflation phases,
V̊E
or expiratory phases, respectively, expressed as a volume per minute
It determines the maximum pressure in the airway pressure. Changing PEEP pressure
Ptotal
does not change this amount. (NSIMV, TCPL)
The continuous flow (Cont.Flow) that the ventilator delivers to the breathing circuit In
Cont. Flow
CPAP+CF mode of ventilation.
Insp. Pause
Duration of an inspiratory pause
Time
Sigh* This function provides additional tidal volume in accordance with preset sigh parameters.
Apnea is a backup mode of ventilation to which the ventilator will switch in the absence
Apnea* of valid breathing effort over a time equal to Apnea Time.
Atlantamed P a g e | 42
Ventilation Settings
Flow Shape
Pinsp
Rise Time
PS
Exp. Trig.
V̊E
Ptotal
Cont.
Time Flow
Insp. Pause
Insp. Trig.
Sigh*
Apnea*
Others*
Parameter
VCV (Assisted) ✓ ✓ - - - - ✓ ✓ - - ✓ ✓ - - - - - - - - ✓ ✓ - ✓
Mode
PCV (Assisted) - ✓ - - - ✓ ✓ ✓ - - ✓ - - ✓ ✓ - - - - - - ✓ - - ✓
PSV - - - - - - - ✓ - - ✓ - - - ✓ ✓ ✓ - - - - ✓ - ✓ ✓
CPAP - - - - - - - ✓ - - ✓ - - - - - - - - - - ✓ - ✓ ✓
VSIMV + PS ✓ ✓ - - - - ✓ ✓ - - ✓ - ✓ - ✓ ✓ ✓ - - - ✓ ✓ - ✓ ✓
PSIMV + PS - ✓ - - - - ✓ ✓ - - ✓ - - ✓ ✓ ✓ ✓ - - - - ✓ - ✓ ✓
NIV - - ✓ - - ✓ ✓ ✓ - - ✓ - - - ✓ ✓ ✓ - - - - ✓ - - ✓
MMV+PSV - - - - - - - ✓ - - ✓ - - - ✓ ✓ ✓ ✓ - - - ✓ - ✓ ✓
PSV (VTGuaranteed) ✓ - - - - - ✓ - - ✓ - - - ✓ ✓ ✓ - - - - ✓ - ✓ ✓
APRV - - - ✓ ✓ - - - ✓ ✓ ✓ - - - ✓ ✓ ✓ - - - - ✓ - ✓ ✓
PRVC(assist) ✓ ✓ - - - - ✓ ✓ - - ✓ - - - ✓ - - - - - - ✓ - - ✓
TCPL - ✓ - - - ✓ ✓ ✓ - - ✓ - - - - - - - ✓ ✓ - ✓ - - ✓
NSIMV + PS - ✓ - - - - ✓ ✓ - - ✓ - - - - ✓ ✓ - ✓ ✓ - ✓ - ✓ ✓
CPAP + CF - - - - - - - ✓ - - ✓ - - - - - - - - ✓ - - - ✓ ✓
Atlantamed P a g e | 43
Ventilation Settings
None
NOTICE
Humidifier Active
The flow sensor configuration is only
Passive
available in neonatal ventilation modes.
Distal
Flow Sensor
Proximal
⚠ Warning ON/OFF
Bear in mind that changing humidifier type Tracheotomy/
requires an alteration to the patient Tube Endotracheal
Compensation
breathing circuit. In this case, tubing Diameter
examination needs to be repeated Compensate
Others
Volume
ON/OFF
• Others menu provides the user with a list Compensation
Leak
of items that can be set before or during ON/OFF
Compensation
ventilation. The list is as follows:
In Parameter Settings page, tap on others O2 Suction Period
button and then others menu will open. Settings O2 enrichment
Waveform
In neonatal ventilation, if the proximal flow (ON/OFF)
sensor is connected to the breathing circuit, Initialize
the option “proximal” is active. Otherwise, Capnography
Settings Save Settings
the option “distal” is active.
Zero Sensor
Reset Sensor
Atlantamed P a g e | 44
Ventilation Settings
5.2.2. O2 suction
Atlantamed P a g e | 45
Ventilation Settings
Atlantamed P a g e | 46
Ventilation Settings
The tracheal pressure peak value is • Use the rotary knob or drag the
displayed on the patient’s data window. range slider to set the diameter.
The maximum inspiratory pressure alarm
behaves in a particular way when the • Touch Compensation button. An
Tube compensations enabled. input field will open on the right.
Tube type, tube diameter and the
• Use the rotary knob or drag the
percentage of compensation can be
range slider to set the percentage of
configured in this page.
compensation.
Tube compensation settings
• Touch apply to save all changes.
Take the following steps to set the
parameters
Atlantamed P a g e | 47
Ventilation Settings
Atlantamed P a g e | 48
Ventilation Settings
5.3. Adjustable Alarm Limits • Drag the range slider or turn the rotary
knob to adjust the parameter
• Tap the Apply button
⚠ Caution
Operator should check that the preset
alarm limits are appropriate prior to use on
each patient.
Atlantamed P a g e | 49
Ventilation Settings
Atlantamed P a g e | 50
Ventilation Settings
In doing so, the emergency ventilation will As soon as the emergency condition is
start. Also, Emergency Ventilation alarm resolved or stabilized, the device should
will be activated. be configured again in order to function
correctly,
An emergency indicator on the status bar
It will be required to use an alternative
will also illuminate red.
means of ventilation in the meanwhile.
Take the following steps below to back to
normal ventilation
Control parameters
F Respiratory rate
Ti Inspiration time
Rise Time Rise Time
O2 Oxygen concentration
Atlantamed P a g e | 51
Explanation of Ventilation Modes
6.1. Introduction 53
6.2. Ventilation parameters 53
6.3. Sigh 58
6.4. Apnea ventilation 59
6.5. VCV (Assisted) 61
6.6. PCV (Assisted) 62
6.7. PRVC (Assisted) 63
6.9. PSIMV+PS 66
6.10. MMV + PSV 67
6.11. PSV (VT guaranteed) 68
6.12. PSV 69
6.13. CPAP 70
6.14. APRV 71
6.15. NIV 72
Atlantamed P a g e | 52
Explanation of Ventilation Modes
It is the airway pressure during an
inspiratory or inflation phase
Paw (cm H2O)
Time(S
)
Figure 6-1 inspiratory time
Time
Atlantamed P a g e | 53
Explanation of Ventilation Modes
Flow
(L/m)
Pi
PEE
P
Tim
e Tim
Figure 6-4 PEEP e
Time
Figure 6-7 rise time
Figure 6-5 tidal volume
Atlantamed P a g e | 54
Explanation of Ventilation Modes
Make sure that the value assigned to the Large enough respiratory
Paw (cm H2O) effort to trigger the
inspiratory trigger (triggering sensitivity) is inspiratory phase
Pressure
trigger
not so low that the ventilator triggers itself sensitivity
instead of triggering in response to
respiratory efforts of the patient. Time
Flow (L/m)
Inspiratory trigger can be either flow-
triggered or pressure-triggered.
Time
6.2.8. Flow trigger
Figure 6-9 pressure trigger sensitivity
When changes of pressure in the
Flow Large enough respiratory
ventilator breathing system reach a set
(L/m) effort to trigger the threshold level (trigger sensitivity) the
inspiratory phase
Flow trigger
ventilator will start inspiratory phase.
sensitivity The sensitivity is set as high as possible
to make sure that the ventilator will
Time initiate a breath in response to the
patient’s breathing efforts.
Paw (cm
H2O)
6.2.10. Expiratory trigger
Atlantamed P a g e | 55
Explanation of Ventilation Modes
Tim
e 6.2.14. O2 concentration
Figure 6-10 flow cycled ventilation The percentage of Oxygen (O2) in the
gaseous mixture that is delivered to the
It is a method of inspiratory phase
patient.
termination. The ventilator cycles from
the inspiration phase into the expiratory 6.2.15. Mandatory ventilation
phase when the flow has decreased to
a preset value during inspiration. In
Diomede, the flow cycling variable Ti Ti
(Exp.Trig) is a percentage fraction of the T< Ti
Paw (cm H2O)
peak flow rate achieved during
inspiration.
6.2.12. Time-cycled
Flow (L/m)
Ti max
Flow
(L/m) Peak Flow
Figure 6-12 mandatory ventilation
Atlantamed P a g e | 56
Explanation of Ventilation Modes
Paw (cm
H2O)
Tim
e
Flow (L/m)
Tim
e
VT
Tim
Figure 6-13 spontaneous ventilation e
Atlantamed P a g e | 57
Explanation of Ventilation Modes
6.3.2. Added VT
Atlantamed P a g e | 58
Explanation of Ventilation Modes
⚠
ventilation modes where the normal
Warning function of ventilator is to automatically
deliver an inspiration to the patient in case
Apnea is a backup mode of ventilation to any valid respiratory effort has not been
which the ventilator will switch in the detected during a period of time as long
absence of valid breathing effort over a time as TApnea.
equal to TApnea . In case apnea is activated,
the ventilation parameters related to the The ventilator automatically switches into
preset mode will be disabled. apnea backup ventilation.
Atlantamed P a g e | 59
Explanation of Ventilation Modes
Atlantamed P a g e | 60
Explanation of Ventilation Modes
Atlantamed P a g e | 61
Explanation of Ventilation Modes
⚠ Warning
An increase in the resistance or decrease
Pi
in the compliance will result in the rise of
the airway pressure. Make sure to assign
Time(s)
a suitable value to the upper pressure limit
in order to prevent the excessive pressure Flow
in the patient’s lung.
Time(s)
VT
6.6. PCV (Assisted)
Time(s)
Assist Pressure Control Ventilation Figure 6-19 curves in PCV ventilation mode
• Pinsp
6.6.1. How it works • F
• Ti
• I:E
The Pressure Controlled mode can be
• Rise Time
adopted when the maximum airway
• PEEP
pressure must be controlled.
• Insp.Trig
In PCV mode, the ventilator maintains the • O2
airway pressure at a clinician set level and
this value remains unchanged regardless 6.6.3. Other functions available
of any changes in the respiratory system
mechanics. If the patient tries to exhale
Additional functions enabled while this
during the inspiration, the expiratory valve
mode is adopted are as follows:
will allow exhalation if the pressure
exceeds the set pressure level by 3 Table 6-4 other functions available in PCV
cmH2O.
In more menu In others menu
The delivered tidal volume depends on the
airway pressure, compliance and Nebulization Volume compensation
resistance in the respiratory system. The
%O2 suction Leak compensation
flow during inspiration is decreasing.
In the absence of the patient’s respiratory Manual inspiration Tube compensation
effort, breaths are delivered at the set
intervals. The patient can trigger extra Inspiratory/Expiratory
breaths, however. To do so, the ventilator pause
must recognize a breathing effort greater
than the set sensitivity.
The shorter the rise time is set at the
beginning, the sharper the slope of the
pressure-time graph and the sooner the
pressure will reach to its maximum level.
Atlantamed P a g e | 62
Explanation of Ventilation Modes
⚠
constant flow rate and pause time set to
Warning 0.5 second.
From then on, the ventilator begins to
The delivered tidal volume can vary deliver pressure-controlled breaths. The
according to changes in respiratory ventilator will carry on the ventilation with
mechanics. Special attention should be the estimated pressure to check if it is
given to monitoring the VT in this mode. precise enough to supply the desirable
Therefore, the alarm for VT minimum tidal volume. If not, the ventilator will make
and maximum should be carefully continuous adjustments to the pressure
adjusted. level breath-by-breath in order to achieve
the set tidal volume.
NOTICE The degree of automatic pressure
adjustment from one breath to another will
The inspiratory pressure (Pi) is not exceed 3 cmH2O.
considered above PEEP pressure.
Airway Pressure= Pinsp+ PEEP The pressure can fluctuate within a certain
range from a low of 5 cmH2O above
PEEP to a high of the upper pressure limit
6.7. PRVC (Assisted) minus 5 cmH2O.
If these maximum or minimum pressure is
Pressure Regulated Volume reached, and the target tidal volume is not
achieved, the alarm signal “Volume target
Controlled
not reached” will be activated.
The maneuver to estimate respiratory
6.7.1. How it works
mechanics will be repeated in the event of
one of the following conditions
This mode is similar to Synchronized
Intermittent Mandatory Ventilation (SIMV).
• Having put the ventilator on
It combines the advantages of Volume
standby mode9, the user wants to
Controlled and Pressure Controlled
carry on PRVC ventilation.
ventilation.
• The tidal volume delivered to the
The set tidal volume can be acquired by patient exceeds twice as high as
automatic regulation of pressure. the target tidal volume at the same
pressure.
The respiratory mechanics tends to vary in
• Assigning a new value to the target
response to the clinical condition of the
volume
patient, but the ventilator tries to adjust the
inspiratory pressure to the lowest possible
level to guarantee the set tidal volume.
Firstly, the ventilator carries out a
maneuver in order to calculate the
respiratory system compliance. It will then
estimate an approximate value for the
pressure required for delivering the set
tidal volume.
Atlantamed P a g e | 63
Explanation of Ventilation Modes
⚠ Warning
Paw (cm H2O)
The upper limit of the airway pressure is not
possible to be set lower than PEEP + 10
Time(s)
cmH2O. In order to reach the set target tidal
Flow (L/m) volume, there must be enough margin for the
ventilator to adjust the inspiratory pressure.
Otherwise, the alarm “target volume not
Time(s) reached” will be activated.
VT(mL)
Atlantamed P a g e | 64
Explanation of Ventilation Modes
The weaker the patient’s effort, the more 6.8.3. Other functions available
pressure support is needed to deliver an
adequate tidal volume.
Additional functions enabled while this mode is
The pressure support, therefore, should be adopted are as follows:
adjusted with respect to the patient’s total
respiratory rate and spontaneous tidal volume. Table 6-6 other functions available in VSIMV+PSV
This pressure support helps the patient to
overcome the respiratory system resistance. In more menu In others menu
Time(s)
Flow
(L/m)
Time(s)
VT(mL
)
Time(s)
• VT • Flow Shape
• PS • Rise Time
• F • PEEP
• Ti • Insp.Trig
• I:E • O2
• Exp.Trig
Atlantamed P a g e | 65
Explanation of Ventilation Modes
Inspiratory/Expiratory pause
Atlantamed P a g e | 66
Explanation of Ventilation Modes
Pressure Support
Time(s)
6.10.1. How it works
Flow (L/m)
Atlantamed P a g e | 67
Explanation of Ventilation Modes
Flow (L/m)
This mode delivers spontaneous breaths
with support pressure in which the patient
triggers every breath. It can, however, Time(s)
switch from PSV mode to VCV (with
square flow profile) to supply the clinician- VT(mL)
set tidal volume.
Time(s)
In response to adequate breathing effort Figure 6-24 PSV (VT guaranteed)
from the patient, the inspiratory phase will
be triggered and continue in PSV mode • VT
until inspiratory flow decreases to a preset • PS
percentage of peak inspiratory flow • Exp.Trig
(Exp.Trig). • Rise Time
• PEEP
At this moment, ventilation in PSV mode • Insp.Trig
will continue if the delivered tidal volume • O2
(VT) is equal to the target tidal volume.
Otherwise, the ventilation mode will switch
6.11.3. Other functions available
to VCV mode with square flow waveform.
The ventilator will extend the inspiratory
Additional functions enabled while this
phase with constant flow rate, aiming to
mode is adopted are as follows:
deliver the set tidal volume.
There is also a safety mechanism to Table 6-9 other functions available in PSV
(VT guaranteed)
prevent excessive inspiratory time.
If the inspiration time exceeds 3 seconds In more menu In others menu
for ADL or 1.5 seconds for PED, the
Nebulization Volume compensation
ventilator cycles from inspiration to
expiration. Manual breath Leak compensation
Atlantamed P a g e | 68
Explanation of Ventilation Modes
VT(mL)
The pressure is kept constant for the rest
of the inspiratory phase while the flow
shape is decreasing. Time(s)
The tidal volume (VT) delivered to the
Figure 6-25 PSV
patient depends on the set support
pressure (PS) and the respiratory system
resistance and compliance. • PS
The higher the set support pressure level, • Exp.Trig
the more gas delivers to the patient. • Rise Time
• PEEP
Peak pressure is the summation of set • Insp.Trig
PEEP and set Pressure Support (PS). • O2
PSV is flow-cycled ventilation. The
ventilator will cycle into the expiratory
phase once the flow rate has dropped to a
predetermined percentage of the peak
flow.
The expiratory trigger sensitivity (Exp.Trig)
can be set within a range from 5 to 80
percentage of the peak flow.
In the event of one of the following
conditions, expiratory phase will be
triggered in order to secure the patient’s
lung from hyperinflation.
Atlantamed P a g e | 69
Explanation of Ventilation Modes
6.12.3. Other functions available Any patient effort to open valves that
provide the flow is regulated by controlling
the Inspiratory Sensitivity.
Additional functions enabled while this
Properly set Insp.Trig will prevent
mode is adopted are as follows:
collapse of airways.
Table 6-10 other functions available in PSV When the ventilator does not detect huge
enough inspiratory efforts, the device
In more menu In others menu regards it as apneic condition and will
Nebulization Volume compensation automatically switch to backup ventilation.
Manual breath Leak compensation Always set the Apnea time appropriately
to fit the current clinical status of the
patient.
6.12.4. Special considerations
⚠
6.13.2. Ventilation parameters
Warning
If the lung mechanics changes, then
delivered tidal volume (VT) will be Paw (cm H2O)
changed as well. In this case, the
pressure support (PS) must be readjusted
Time(s)
to deliver the desired tidal volume. Flow (L/m)
Time(s)
VT(mL)
6.13. CPAP
Time(s)
Continuous Positive Airway
Pressure Figure 6-26 CPAP
• PEEP
6.13.1. How it works • Insp.Trig
• O2
The mode Continuous Positive Airway
Pressure (CPAP) is a noninvasive
operative mode and can be used when
the patient is able to breathe
spontaneously.
All breaths are spontaneous and starts
upon patient effort.
CPAP provides constant positive pressure
during inspiration and expiration in order
to open the upper airway and collapsed
alveoli.
When the patient starts to breath, the
ventilator delivers a flow proportional to
the patient respiratory demand in order to
keep the airway pressure constant.
Atlantamed P a g e | 70
Explanation of Ventilation Modes
Time(s)
6.13.4. Special considerations
VT(mL)
NOTICE Time(s)
Figure 6-27 APRV
CPAP is similar to PSV when the pressure
support is set to zero. (PS=0). In this case, • PEEP high
CPAP ventilation features are the same as • PEEP low
PSV. • PS
• T high
6.14. APRV
• T low
• Exp.Trig
Airway Pressure Release Ventilation • Rise Time
• Insp.Trig
• O2
6.14.1. How it works
6.14.3. Other functions available
APRV is a time-cycled, pressure-
controlled mode of ventilation. The Additional functions enabled while this
ventilator in this mode maintains airway mode is adopted are as follows:
pressure at higher pressure (High PEEP)
with intermittent pressure drop to Low Table 6-12 other functions available in APRV
PEEP, while allowing spontaneous
breaths at any time throughout ventilation. In more menu In others menu
Switching between two different airway Nebulization Volume compensation
pressures occurs at clinician specified
time. %O2 suction Leak compensation
Pressure support can be set larger than Manual inspiration Tube compensation
zero to boost spontaneous inspirations.
By design, any breathing efforts or
changes in respiratory mechanics do not
Atlantamed P a g e | 71
Explanation of Ventilation Modes
Non-Invasive Ventilation
Paw (cm H2O)
Flow (L/m)
This mode shares common characteristics
with PSV and PCV. Spontaneous patient-
Time(s)
triggered breaths are accompanied with
pressure support.
VT(mL)
If the rate of spontaneous breaths falls
below the target rate, the ventilator will Time(s)
send mandatory pressure limited breath at
specified interval in accordance with the • PS • Rise Time
set frequency. • F • PEEP
• Ti,max • Insp.Trig
All breaths are normally flow-cycled so the
• I:E • O2
ventilator cycles from inspiration to • Exp.Trig
expiration when the flow reaches a certain
fraction of peak flow. If the flow rate fails
6.15.3. Other functions available
to reach that amount in clinician-set
maximum inspiratory time (Ti, max), the
ventilator will begin the expiratory phase. Additional functions enabled while this
mode is adopted are as follows:
In this mode, leak compensation is
activated by default and cannot be Table 6-13 other functions available in NIV
disabled.
In more menu In others menu
The size of the compensatory leak
depends on the patient type. Nebulization Volume compensation
Compensation leak method allows the %O2 suction Leak compensation (activated)
stability of the regulated pressure,
prevention of auto triggering, and control Manual inspiration
of the respiratory synchronization.
The exhaled volume of the patient can
6.15.4. Special considerations
differ from the measured exhaled volume
due to leaks around the mask.
NOTICE
Atlantamed P a g e | 72
7. NEONATAL VENTILATION
7.1. General 74
7.2. Apnea ventilation 74
7.3. TCPL 74
7.4. NSIMV 75
7.5. CPAP + CF 76
Atlantamed P a g e | 73
Neonatal Ventilation
Compulsory Optional
PSV VSIMV+PSV Flow (L/m)
Apnea Ce PAP PSIMV+PSV
APRV
CPAP+ CF VT(mL)
TCPL
PRVC
Figure 7-1 TCPL
7.3. TCPL
• P Total
• F
Time-Cycled Pressure-Limited • Ti
ventilation • I:E
• Cont.Flow
7.3.1. How it works • PEEP
• Insp.Trig.
• O2
This mode is specific to the neonatal
ventilation. This mode is time-cycled, and 7.3.3. Other functions available
inspiratory pressure can be regulated.
Atlantamed P a g e | 74
Neonatal Ventilation
In more menu In others menu During the expiratory time, the patient can
trigger a spontaneous breath.
Nebulization Volume compensation Spontaneous patient-triggered breaths are
accompanied with pressure support.
%O2 suction Leak compensation
The spontaneous breaths are
Manual inspiration Distal/proximal flow sensor. synchronized with mandatory breaths
similar to what happens in SIMV modes.
Depending on the time the patient makes
7.3.4. Special considerations a valid inspiratory effort the ventilator will
deliver either pressure supported breath
or TCPL breath.
NOTICE
If inspiratory effort is not close to next
mandatory inspiration phase, the ventilator
In this mode, the inhaled VT is based on will deliver a pressure supported breath to
the continuous flow being delivered and the patient.
the patient’s expiration, and it may,
therefore, differ slightly from the results If the inspiratory effort is close enough to
obtained f other means. the next mandatory inspiration time, the
ventilator will shift backward. In this case,
the patient will receive a breath the same
The selection of high continuous flow can as TCPL ones.
lead to an increase in pressure baseline.
In this case the pressure baseline can be This will contribute to synchronization of
more than PEEP pressure. When the spontaneous pressure supported breaths
constant flow increases, the pressure and TCPL breaths, which result in a more
gradient across the entire respiratory comfortable and efficient ventilation
system will increase. the operator should therapy.
be experienced enough to assign
appropriate values to PEEP and
continuous flow.
7.4. NSIMV
Non-Invasive Synchronized
Intermittent Mandatory Ventilation
with Pressure Support
Atlantamed P a g e | 75
Neonatal Ventilation
• P Total
7.5.1. How it works
• PS
• F
• Ti This non-invasive mode is specific to
• I:E neonatal-infant age group. A nasal prong
• Exp.Trig. is used to connect the breathing circuit to
• Cont.Flow the patient. This mode can be used when
• PEEP the patient is able to breathe
• Insp.Trig. spontaneously.
• O2
Spontaneous breaths are not supported
by the ventilator. The ventilator provides a
7.4.3. Other functions available continuous flow during the entire
respiration. The amount of this flow that
can be delivered to the patient is propor-
tional to the intensity of the patient
Table 7-3 other functions available in NSIMV
respiratory effort.
In more menu In others menu Apnea ventilation is on by default in this
mode. It can be deactivated, however.
Nebulization Volume compensation
When the ventilator does not detect
%O2 suction Leak compensation
inspiratory efforts during the clinician set
Manual Distal/proximal flow Apnea time, the device regards it as an
inspiration sensor. apneic condition and will automatically
switch to backup ventilation. The apnea
alarm will then be activated.
Leakage compensation is activated in this
mode and cannot be turned off.
Assigning a large value to continuous flow
can raise the pressure baseline due to the
airway system resistance.
Atlantamed P a g e | 76
Neonatal Ventilation
Therefore, the operator should be While using this mode (CPAP+ CF), a well
experienced enough to assign appropriate experienced clinician must always be
values to the PEEP and continuous flow present to check the patient’s current
levels. situation. Because of the high resistance
of the nasal prongs for NEO-INF patients
7.5.2. Ventilation parameters and the large leaks that usually occur in
this ventilation mode, nasal prong
disconnection warning conditions may not
be detected.
NOTICE
In alarm setting page, the alarm limits of
tidal volume and minute volume are
Figure 7-3 CPAP+CF disabled in this mode.
Leak compensation
⚠ Warning
Atlantamed P a g e | 77
Monitoring and Record Keeping
Atlantamed P a g e | 78
Monitoring and Record Keeping
Inspiratory Time:
TI s
Duration of an inflation phase or inspiratory phase
Expiratory Time:
TE s
Duration of an expiratory phase
Mean Pressure
PMean CmH2O It indicates the mean pressure
of each breath
Peak Pressure:
PPeak CmH2O
Highest airway pressure reached during a previous respiratory cycle
PPlat CmH2O
Plateau Pressure:
Atlantamed P a g e | 79
Monitoring and Record Keeping
Airway pressure during an inspiratory pause when the flow at the patient-
connection port is zero
Tracheal pressure
PTrach CmH2O It indicates the pressure drop through an endotracheal tube when Tube
Compensation is activated.
Only mandatory breaths are taken into account in calculation of this parameter.
The breaths that Can be triggered by the ventilator or the patient except for
PSV or CPAP.
V̇E,Mand L/min
In VCV or PCV modes all breaths are mandatory.
In APRV, any switch from low PEEP level to high PEEP level is considered as
mandatory breath.
V̇E,Spon L/min minute volume that is additional to the assured minute volume
It is the minute volume calculated from spontaneous breaths generated from
PSV or CPAP.
Inspiratory-Time Fraction
Ti/Ttot -
Ratio of the inspiratory time to the respiratory cycle time
Atlantamed P a g e | 80
Monitoring and Record Keeping
Leak percentage
Leak % Percentage of loss of gas from the ventilator breathing system
Airway Leak:
Leak L/min loss of respiratory gas from its pathway between the patient-connection port
interface and the lungs
The slope of flow-volume curve using two points corresponding to 50% and
TCEXP s
75% of the expiration phase.
The percentage of Oxygen (O2) in the gaseous mixture that is delivered to the
O2% %
patient.
Atlantamed P a g e | 81
Monitoring and Record Keeping
Tidal Volume
VTE
volume of gas that enters and leaves the lung during a breath
Atlantamed P a g e | 82
Monitoring and Record Keeping
Auto-PEEP
portion of the stabilized airway pressure above the set
PEEP, at the end of an expiratory-hold procedure that
Auto PEEP CmH2O
temporarily occludes the airway and in the absence of
any respiratory activity
Ri cmH2O/L/s
Re cmH2O/L/s
Cstat mL/cmH2O
Ri cmH2O/L/s
Cdyn mL/cmH2O
Tidal Volume L
Trapped volume L
Current vital
L
capacity
PImax cmH2O
PV Flex
Upper IP cmH2O
Lower IP cmH2O
Cmax mL/cmH2O
Atlantamed P a g e | 83
Monitoring and Record Keeping
VD/VT Physiologic
PaCO2
PeCO2
Atlantamed P a g e | 84
Monitoring and Record Keeping
Atlantamed P a g e | 85
Monitoring and Record Keeping
8.3.3. Loops
Atlantamed P a g e | 86
Monitoring and Record Keeping
Numeric display
• Select the third option from the graphics
menu. Parameter Ventilation Mode
In doing so, time-graphs will be shown in
the monitoring page. the variables on the FSpon All modes
vertical axes are as follows:
All modes except for
• Airway pressure(cmH2O) FMand PSV, PSVVT, MMV AND
• Tidal volume(mL) CPAP
• Flow (L/min)
All modes except for
V̇E,Mand PSV, PSVVT, MMV AND
CPAP
• Airway pressure
• Tidal volume
• Flow
Airway pressure, tidal volume and flow- Figure 8-6 time graphs
time graphs will be displayed onscreen.
Atlantamed P a g e | 87
Monitoring and Record Keeping
8.3.6. Capnography
⚠ Warning
BA210 CO2 sensor and nebulizer should
not be used simultaneously. Otherwise,
the optical transmission of the airway
Figure 8-7 CO2 sensor warm up
adapter may be affected.
To monitor CO2 concentration
Condensation of the humid gas exhaled
from the patient’s lung may leave drops of • In the monitoring page, tap on the
water on optical windows of the airway graphics icon
adapter. This may result in overcounting • Select the option on which CO2 is
of CO2 concentration. displayed
⚠
In doing so, time-graphs will be shown in
Caution the monitoring page. The variables on the
vertical axes are as follows:
Make sure the capnograph is installed • Airway pressure(cmH2O)
properly; otherwise, the option in graphics • Tidal volume(mL)
menu provided for capnography will be • Flow (L/m)
disabled. (The item marked as number 5 Plus, two other graphs illustrate the end-
in figure 8-2) tidal CO2 versus time and the end-tidal
As soon as the capnograph is connected CO2 versus the volume of CO2 expired per
to the device, the message “Sensor Warm breath. Apart from graphical
Up” will appear onscreen. representation of the patient’s data, some
essential parameters are displayed in
numbers. These parameters are gathered
in table 8-4.
Atlantamed P a g e | 88
Monitoring and Record Keeping
In parameter setting window, follow the To display the capnograph data onscreen
steps below to open capnography menu
• Tap on the “Waveform”
• Parameter setting>>Others>> • Turn it on
Capnography
8.4.2. Initialize
The followings are required to take care of
if necessary
If the sensor is connected to the ventilator
• Waveform
and a message “ERROR” appears
• Initialize onscreen, the sensor must be activated
• Save settings manually. Take the following steps to fix it.
• Zero sensor
• Reset sensor Parameter setting>> Others >>
Capnography >> Initialize.
Capnography parameters
ETCO2 End-tidal carbon dioxide. The percentage of carbon dioxide released at the end of an exhaled breath
VAlveolar The amount of gas per unit of time that reaches the alveoli and becomes involved in gas exchange
VCO2
Amount of CO2 expired per minute
(STPD)
Partial pressure of CO2 of the exhaled gas mixture (alveolar gas + gases of the anatomical dead
PECO2
space)
Atlantamed P a g e | 89
Monitoring and Record Keeping
Atlantamed P a g e | 90
Monitoring and Record Keeping
When the CO2 value is detected more than 150 If this error persists, the sensor value
CO2 OUT OF RANGE
mmHg. must be set to zero.
(during adapter changing or when there has been When the sensor is removed from the
CO2 inside it at the time of reset to zero). adapter check the adapter for any
CHECK PATIENT
AIRWAY ADAPTER It is recommended to clean the adapter occlusion or blockage.
Atlantamed P a g e | 91
Monitoring and Record Keeping
Atlantamed P a g e | 92
Monitoring and Record Keeping
mode
Atlantamed P a g e | 93
Monitoring and Record Keeping
Respiration pause can be applied to either • Tap the “O2 suction” button
inspiration or expiration phases.
• Press and hold the respiration pause
in inspiration phase.
The inspiration phase will continue for 7
seconds and the INS PAUSE indicator Figure 8-16 O2 suction
below the status bar will illuminate green.
(Figure 3-5)
The ventilator increases the oxygen
• Press and hold the respiration pause concentration delivered to the patient to
in expiration phase 100% and maintain this level throughout
the pre-processing stage which takes 60
The expiration phase will continue for 20
seconds.
seconds and the EXP PAUSE below the
status bar will illuminate green. (Figure 3-
5)
Atlantamed P a g e | 94
Monitoring and Record Keeping
Over the suction stage, pulmonary The ventilator ventilates temporarily with
secretions are removed from the patient’s air only. The low O2 Gas alarm will be
airway. This stage takes 20 seconds. activated in this instance.
Having finished the suctioning stage, the • Connect an oxygen cylinder to the
ventilator will rise the oxygen ventilator
concentration delivered to the patient to • Shut down the wall air supply and
100% once again and maintain this level remove the air hose from the
during the post processing stage which ventilator.
takes 60 seconds.
This will trigger the Low air pressure
8.7.6. Calibrate O₂ alarm.
The Diomede ventilator allows The “Low Air Gas” and “Power Loss”
intrahospital transportation. This function alarms will be deactivated when
can only be activated when the air supply “Transportation” is on.
is missing.
8.8. Event log
To transport the patient, wall air and
oxygen supply are required to be
disconnected from the ventilator.
Event logs records all necessary data
• Shut down the wall O2 supply and about ventilator activities, including
remove the O2 hose from the activation and deactivation of alarms,
ventilator. readjustments of target parameters, alarm
Atlantamed P a g e | 95
Monitoring and Record Keeping
8.9. Trend
Atlantamed P a g e | 96
Monitoring and Record Keeping
Atlantamed P a g e | 97
Monitoring and Record Keeping
8.9.3. Read parameter values • Touch and hold your finger on one of the
Y-axis labels (. Flow, VT or PCO2). A list
menu will open.
The value of every parameter is displayed • Select the parameter you want to be
in yellow on the vertical axis of each substituted for the one that is already
graph. being shown.
• Drag the zoom slider to find the time 8.9.5. Change time-scale
about which you are looking for
patient’s data
• Tap on the set of graphs below .A To monitor the patient record for the last 6
vertical line will appear which months
intersects with all three graphs.
• Tap on 6 Months button
• The corresponding value of the
parameter at the point of intersection
To monitor the patient’s data for a shorter
will appear on the vertical axis.
period of time
• Touch and drag the vertical line
across the x-axis to read the value of
• Tap on + button on the right
parameters at the specific time you
want. On both sides of each set of graphs, there
are two buttons labeled with the time
8.9.4. Change vertical axis variables
length during which the patient information
is being displayed. Take the following
The Diomede ventilator keeps record of figure for example. The time length for
important time-varying parameters. Flow, graphs above (5 minutes) is ten times
VT and PCO2 are plotted on Y axis by more than the time length for the set of
default. graphs below (30 seconds).
The variables plotted on Y axis can be • Tap on time length buttons on either
changed to illustrate other target side to shift the time axis
parameters.
To substitute the variable of the Y-axis
Atlantamed P a g e | 98
Monitoring and Record Keeping
Atlantamed P a g e | 99
9. ASSESSMENT OF RESPIRATORY MECHANICS
Atlantamed P a g e | 100
Assessment of Respiratory Mechanics
Atlantamed P a g e | 101
Assessment of Respiratory Mechanics
Atlantamed P a g e | 102
Assessment of Respiratory Mechanics
9.2.7. Modes
Atlantamed P a g e | 103
Assessment of Respiratory Mechanics
Atlantamed P a g e | 104
Assessment of Respiratory Mechanics
Atlantamed P a g e | 105
Assessment of Respiratory Mechanics
Atlantamed P a g e | 106
Assessment of Respiratory Mechanics
This maneuver can be performed using Take the following steps to measure PImax
two different instructions.
• Go to the Respiratory Mechanics
The patient can be encouraged to make page
respiratory effort so the maximum • Tap on the PImax tab
inspiratory effort will be achieved. • Tap on the Start button
Atlantamed P a g e | 107
Assessment of Respiratory Mechanics
Atlantamed P a g e | 108
Assessment of Respiratory Mechanics
9.7.3. Modes
Atlantamed P a g e | 109
Assessment of Respiratory Mechanics
Atlantamed P a g e | 110
Assessment of Respiratory Mechanics
PCV (Assist)
Guaranteed)
VSIMV+PSV
PSIMV+PSV
PSV(VT
APRV
CPAP
PSV
NIV
Auto-PEEP ✓ ✓ ✓ - - - - - - - -
RC Measure ✓ ✓ ✓ - - - - - - - -
Trapped Volume ✓ ✓ ✓ - - - - - - - -
Slow Vital Capacity ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓
P0.1 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ - ✓ ✓
PI max - - - ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓
PV Flex ✓ ✓ ✓ - - - - - - - -
VD/VT Physiologic ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓
Atlantamed P a g e | 111
Alarms Handling
PRVC (Assist)
VCV (Assist)
PCV (Assist)
NSIMV+PSV
VSIMV+PSV
PSIMV+PSV
NEO
APRV
CPAP
TCPL
PSV
NIV
Auto-PEEP ✓ - ✓ ✓ - - - - - - -
RC Measure ✓ - ✓ ✓ - - - - - - -
Trapped Volume ✓ - ✓ ✓ - - - - - - -
P0.1 - - - - - - - - - - -
PI max - - - - - - - - - - -
PV Flex - - - - - - - - - - -
VD/VT Physiologic - - - - - - - - - - -
Atlantamed P a g e | 112
Alarms Handling
Atlantamed P a g e | 113
Alarms Handling
High 10
Information about then activated and
deactivated alarm is stored in the Event Medium 3
log. Low 1
Atlantamed P a g e | 114
Alarms Handling
• While the condition that triggered the When the highest priority alarm is cleared,
alarm exists, tap on the alarm icon to the second highest priority alarm displays
view the name of the alarms. Alarms in status bar.
are listed in order of priority.
• Tap on the alarm icon in the status bar
to find more details about other alarms
with lower importance.
Atlantamed P a g e | 115
Alarms Handling
Low O2 Concentration %
O2 Concent. O2%
High O2 Concentration %
– –
PEEP PEEP
High PEEP cmH2O
Atlantamed P a g e | 116
Alarms Handling
Priority
Alarm Adjustability Condition Action needed
Which modes
Technical/Physiological
Atlantamed P a g e | 117
Alarms Handling
Atlantamed P a g e | 118
Alarms Handling
Medium The ventilator has not • Check the value of the configured
Non-adjustable detected any valid respiratory inspiratory trigger, and correct it if
MMV+PSV effort on the side of the necessary.
Apnea PSV+VT/PSV/CPAP patient for as long as preset • Check the presence of the patient's
VSIMV+PSV/PSIMV+PSV apnea time. inspiratory efforts.
APRV • Check that there are no leaks in the
Physiological patient circuit.
Atlantamed P a g e | 119
Alarms Handling
Low
Nebulization is interrupted
Nebulization Non-adjustable
due to the absence of flow or
Stopped Nebulized treatment
gas deficiency.
Physiological
Atlantamed P a g e | 120
Alarms Handling
Atlantamed P a g e | 121
Alarms Handling
High The alarm will be activated • Disconnect the patient from the
Non-adjustable when no pulse is received ventilator.
Tech.Error 01 All modes from the power module • Use alternative ventilation
Technical processor. immediately.
• Contact the customer service
Atlantamed P a g e | 122
Alarms Handling
Atlantamed P a g e | 123
Alarms Handling
Device Status
Alarm
Startup Ventilation Suction Standby Transport
Emergency Ventilation --- High High --- High
High Pressure --- High High --- High
Persistent High Pressure --- High High --- High
Low Gas (Air) High High High Low ---
Low Gas (O2) High High High Low High
O2 Concentration Too Low --- High High --- High
Patient Circuit Open --- High --- --- High
Low Pressure --- High --- --- High
High Tidal Volume --- Medium --- --- Medium
Low Tidal Volume --- Medium --- --- Medium
Low O2 Concentration --- Medium --- --- Medium
High O2 Concentration --- Medium --- --- Medium
Apnea --- Medium Medium --- Medium
High Rate --- Low --- --- Low
Low Rate --- Low --- --- Low
High Minute Volume --- Medium --- --- Medium
Low Minute Volume --- Medium --- --- Medium
Nebulization Stopped --- Low --- --- Low
Low Gas (Air and O2) High High High Low High
Standby --- Low --- Low Low
High ETCO2 --- High --- --- High
Low ETCO2 --- High --- --- High
Target Volume Not Reached --- Medium --- --- Medium
Transporting --- Low --- --- Low
High PEEP --- Medium --- --- Medium
Power Loss High Medium Medium Medium ---
Low Battery Medium Medium Medium Medium Medium
Critical Battery High High High High High
Fan failure Medium Medium Medium Medium Medium
Tech.Error 01 Medium Medium Medium Medium Medium
Tech.Error 02 High High High High High
Tech.Error 03 High High High High High
Tech.Error 04 High High High High High
Tech.Error 05 High High High High High
Tech.Error 06 High High High High High
Atlantamed P a g e | 124
Alarms Handling
Low Gas (Air) 7. Disconnect the air hose from the air inlet connection
Atlantamed P a g e | 125
Alarms Handling
Atlantamed P a g e | 126
Alarms Handling
Target Volume Not 7. In PRVC mode, set the upper limit and lower limit of airway
Reached pressure very close to one another.
Atlantamed P a g e | 127
11. MAINTENANCE
Atlantamed P a g e | 128
Maintenance
Annually
Annually
Preventive maintenance is necessary to
ensure continued safe use of the Diomede.
Air Fully serviced by
Every 5000 hours
compressor authorized
Negligence of preventative maintenance
service
with accordance to the time schedule may
personnel
deteriorate the device performance and
even put the patient’s life at risk. Expiratory
valve Between patients
The Diomede shall be fully serviced by diaphragm
authorized service personnel annually or
every 5000 hours of operation.
Every 3 month
Performance
Battery
inspection
Replacement if
required
Daily
Water traps
Empty the water
High O2
Annually concentration and
Replacement the ambient
O2 sensor
temperature may
narrow down this
period
Patient Manufacturer’s
Between patients
circuit instructions
Compressor 6 months
air filter Replacement
Air inlet 6 months
filter Replacement
Oxygen 6 months
inlet filter Replacement
patient
Manufacturer’s Daily
circuit water
instructions for Between patients
traps
use
Atlantamed P a g e | 129
Maintenance
11.2.1. Battery
Performance inspection
Every three months the Diomede should be
unplugged from the supply mains and runs
on the internal battery to check the battery
whether it maintains a satisfactory
performance. When it is fully charged, the
Diomede must run on the battery for an hour
and thirty minutes.
Replacement
The battery will need to be replaced by the
authorized service personnel if its
performance is lower than it should be. In
this case contact AtlantaMed’s technical
support team.
Proper battery usage
The proper way to use the battery to extend
its lifespan is to keep the Diomede
connected to the mains supply.
If kept in storage, the battery must be
recharged every three months.
11.2.2. Fuse
Atlantamed P a g e | 130
Maintenance
Atlantamed P a g e | 131
Maintenance
Atlantamed P a g e | 132
Maintenance
Atlantamed P a g e | 133
Maintenance
11.5.2. Cleaning of the expiratory valve stages if there is any residual visible on
the components.
Atlantamed P a g e | 134
11.5.6. Reassembly
×
Figure11-14 diaphragm
installation
✓ Line up the
three
notches with
the locking
nibs and
push firmly.
Figure11-16
Figure11-13 expiratory cap
silicon diaphragm installation
Atlantamed P a g e | 135
Figure 11-19 inappropriate installation
Figure 11-17 expiratory valve installation
11.5.8. Disposal
Atlantamed P a g e | 136
12. SPECIFICATIONS
12.1.Classification 138
12.2.Physical characteristics 138
12.3.Touch screen 138
12.4.Ambient conditions 138
12.5.Audible energy emission 138
12.6.Essential performances 139
12.7.Pneumatic specification 140
12.8.Breathing circuit characteristics 141
12.9.Electrical specifications 142
12.10.Control settings,ranges and accuracy 143
12.11.Additional functions 147
12.12.Measurement accuracy 148
12.13.Monitored parameters 149
12.14.Measurement accuracy 150
12.15.Alarm 151
12.16.Electromagnetic compatibility 152
12.17.Pneumatic diagram 156
Atlantamed P a g e | 137
Specifications
12.1. Classification
Class C
Risk Medical Device Directive
93/42/EEC
IP protection IP21 H
Altitude
12.3. Touch screen
12.5. Audible energy emission
Table 12-3 monitor specifications
Table 12-5 audible sound pressure emission
Sensitive capacitive touch screen/
Type Sound pressure level
color display
Atlantamed P a g e | 138
Specifications
Component Requirements
Gas supply failure The failure of gas supply should be recognized, and warned the operator.
Atlantamed P a g e | 139
Specifications
Maximum supply pressure 7.0 kg/cm2 or bar (686.4 kPa - 100 psi)
Maximum working pressure determining Using safety valve and setting alarm limits based on it.
method
Gas source inlet flow 180 L/min (120 L/min for air compressor)
The average of gas flow required by the 55 L/min (at pressure of 280 kPa, measured at the gas input port)
ventilator in 10 seconds
The maximum average of gas flow is 180 L/min (at pressure of 280 kPa, measured at the gas input
required by the VENTILATOR in 3 seconds port)
The maximum required time to change the 20 Seconds
oxygen concentration from 21 % to 90 %
ADL:130 L/min
Maximum minute volume resulting PED: 40 L/min
NEO-INF: 17 L/min
Connector hoses from the gas supplier to Air: DISS 3/4" _ 16 female connector (both ends).
the ventilator Oxygen: DISS 9/16" – 18 female connector (both ends).
O2 or Air hose length is 3m
Atlantamed P a g e | 140
Specifications
Atlantamed P a g e | 141
Specifications
Atlantamed P a g e | 142
Specifications
Minimum Maximum
Patient Type
ADL PED NEO ADL PED NEO ADL PED NEO ADL PED NEO
Parameter Unit
0.050 0.020 0.100a 0.030 0.005 0.001
VT L 0.400 0.050 0.010 0.002 0.150 0.001
0.100a 0.030 2.500 0.300 0.010 0.005
1.00-2.00 : 0.1
THigh s 5.0 1.00 30.00 2.00-10.00 : 0.5
10.00-30.00 : 1.0
0.2 2.0 0.20-2.00 : 0.1
TLow s 1.5 2.0 10.0 2.00-10.00 : 0.5
10.0 30.0 10.00-30.00 : 1.0
O2 % 50 21 100 1
PS cmH2O 5 0 100 1
Exp.Trig. % 25 5 50 5
Ptotal cmH2O - - 10 - - 2 - - 70 - - 1
Cont.
L/min - - 8 - - 2 - - 40 - - 1
Flow
Atlantamed P a g e | 143
Specifications
Insp.
Pause s Off Off Off 0.25 0.25 - 2.00 2.00 - 0.25 0.25 -
Time
Insp.Trig 0.2-1.0 : 0.1
L/min 3.0 3.0 1.0 0.2 15.0
1.0-15.0 : 1.0
(flow)
Insp.Trig 0.2-0.5 : 0.1
cmH2O 1.5 1.5 0.3 0.2 20.0
0.5-20.0 : 0.5
(pressure)
a.
Atlantamed P a g e | 144
Specifications
Minimum Maximum
Patient Type
NE NE AD
ADL PED ADL PED PED NEO ADL/PED/NEO
Parameter Unit O O L
ADL:
0.050-0.100 : 0.005
0.100-2.500 : 0.010
PED:
VT L 0.400 0.050 0.010 0.050 0.020 0.002 2.500 0.300 0.150
0.020-0.030 : 0.001
0.030-0.300 : 0.005
NEO:
0.002-0.150 : 0.001
1.00-2.00 : 0.1
THigh s 5.0 1.00 30.00 2.00-10.00 : 0.5
10.00-30.00 : 1.0
0.20-2.00 : 0.1
TLow s 1.5 0.2 30.0 2.00-10.00 : 0.5
10.00-30.00 : 1.0
I:E - 1:4.0 1:3.0 1:5.6 1:199 1:599 1:599 4.0:1 4:0.1 4:0.1 ----
O2 % 50 21 100 1
100-PEEPd
Pinsp cmH2O 15 8 8 2 35-PEEPe 1
100-PEEPfHigh
PS cmH2O 5 0 100 1
Exp.Trig. % 25 5 50 5
NEO
Ptotal cmH2O - - 10 - - 2 - - 70
1
NEO
Cont. Flow L/min - - 8 - - 2 - - 40
1
Atlantamed P a g e | 145
Specifications
Atlantamed P a g e | 146
Specifications
TApnea s 5 5 60 1
Number - 1 1 - 1 1 - 3 3 - 1 1 -
Per Hour - 5 5 - 5 5 - 20 20 - 5 5 -
Sigh
Added VT % 30 30 - 10 10 - 100 100 - 10 10 -
Pressure
cmH2O 40 40 5 5 - 120 120 - 5 5 -
Limit
Period s 10 20 20 10 40 5
O2
Suction 5 20 5
O2
% max max 10
Enrichment 100-
20 80
max
Atlantamed P a g e | 147
12.12. Measurement accuracy
.
PARAMETER ACCURACY
0 – 40 mmHg ±2 mmHg
41 – 70 mmHg ±5% of the reading
CO2
71 – 100 mmHg ±8% of the reading
101 – 150 mmHg ±10% of the reading
End-expiratory pressure
±(2 cmH2O + 4% of the programmed pressure)
(PEEP)
Atlantamed P a g e | 148
12.13. Monitored parameters
It is derived by the
Controlled, monitored
Volume* corresponding flow ±10%
and plotted volumes
measurements
Inspiratory and
Expiratory times ±0,06 s
Time Quartz crystal
respiratory rate I:E ratio. ± 1 rpm
Internal clock.
Inspiratory Pressure
Controlled Pressure Pressure transducers ±10%
Pathway
Expiratory Pressure
Monitored Pressure Pressure transducers ±10%
Pathway
Atlantamed P a g e | 149
12.14. Measurement accuracy
Tolerance deviation of
Parameter
measurement
≤ 50 ml: ± 1%
Volume
> 50 ml: ± 1.75%
±1%
O2
Atlantamed P a g e | 150
12.15. Adjustable Alarm limits
0.200a
Low VTE L 0.025 0.005 0.010 0.001 0.001 High VTE - Step 0.010 – 0.030: 0.001
0.150b
0.030 – 0.100: 0.005
0.015a 0.100 – 3.000: 0.010
High VTE L 0.600 0.075 Low VTE + Step 3.000 0.500 0.250
0.050b
High
cmH2O 20 1 50 1
PEEP
Low
mmHg 30 1 High ETCO2 - 1 1
ETCO2
High
mmHg 50 Low ETCO2 + 1 150 1
ETCO2
a. Modes that have VT as a control parameter
b. Modes in which VT is not a control parameter
c. MMV+PSV
d. All modes except for MMV+PSV
Atlantamed P a g e | 151
12.16. Electromagnetic compatibility
Atlantamed P a g e | 152
12.16.2. Electromagnetic immunity
Atlantamed P a g e | 153
Table 12-17 Electromagnetic immunity
Guidance and manufacturer’s declaration – Electromagnetic immunity
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the
ventilator should ensure that it is used in such an electromagnetic environment..
Electromagnetic environment –
IMMUNITY test IEC 60601 test level Compliance level
guidance
6 Vrms 10 Vrms
Recommended separation distance
150 kHz to 80 MHz in 150 kHz to 80 MHz in
Conducted RF
ISM ISM
IEC 61000-4-6
bands Bands
d = 0.35 √𝑃
d = 1.2 √𝑃
3 V/m 10 V/m
Radiated RF d = 1,2 √𝑃 80 MHz to 800 MHz
80 MHz - 2.7 GHz 80 MHz - 3.0 GHz
IEC 61000-4-3 d = 2,3 √𝑃 800 MHz to 2,5 GHz
80 % AM at 1 kHz 80 % AM at 1 kHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from the structures, objects and people.
a) The ISM (Industrial, Scientific and Medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;
13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Atlantamed P a g e | 154
b) The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz
to 2,5 GHz are intended to decrease the probability of mobile/portable communications equipment that could cause
interference if it is brought into patient areas inadvertently. For this reason, an additional factor of 10/3 has been
incorporated into the formulae used in calculating the recommended separation distance from transmitters in these
frequency ranges.
c) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically accurately. To
assess the electromagnetic environment due to the fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ventilator is used exceeds the applicable RF
compliance level, the ventilator should be observed to verify the normal operation. If any abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating ventilator.
d) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Atlantamed P a g e | 155
12.17. Pneumatic diagram
Atlantamed P a g e | 156
Atlantamed P a g e | 157