Specifications
Data Acquisition Unit
Description
The Data Acquisition Unit (DAU) is part of the S8000 Plus System. It
samples the signal inputs such as rotating speed, vibration, process
variables, converts them from analog to digital and transfers the data to
WEB8000 Plus.
It supports the TCP/IP protocol based on the Ethernet (10M/100M)
standard, and transmits data via the enterprise network. In addition it is
characterized by high speed, high accuracy, high reliability, easy
Figure: PWR8000 Plus
installation and connection, maintenance-free.
Meanwhile, the DAU is the first field acquisition station in the world with
a WEB function. From a computer on the same network with the DAU, a
user can view spectrum and other data from the DAU with IE browser.
According to the application and capacity, there are 6 types of DAU. They
are NET8000 Plus, NET8000 Dual, PWR8000 Plus, PWR8000 Dual and
HYD8000 Plus, HYD8000 Dual.
Figure: PWR8000 Plus Rear View
Type of the DAU
Application
NET8000 Plus
PWR8000 Plus
HYD8000 Plus
NET8000 Dual
PWR8000 Dual
HYD8000 Dual
Rotating
>4000 RPM
>1000RPM
12RPM
Speed
32000 RPM
4000RPM
1000RPM
Petroleum
Industry
Coal Fired Power,
&chemical,
Nuclear Power
Steel & Aluminum
Hydro Power
Number of Inputs of the DAU
Signal Type
NET8000 Plus
NET8000 Dual
PWR8000 Plus
PWR8000 Dual
HYD8000 Plus
HYD8000 Daul
Rotating Speed
Vibration
24
48
Hard-Wired Process
12
24
Page 1 of 5
Specifications of DAU
Turn green if the corresponding vibration signal is
normal
Inputs
STATIC
(1-12/24) LEDs:
Power Supply:
AC220V10% 50Hz
Turn green if the corresponding process
AC110V10% 50Hz
signal is normal
Power
Consumption:
Max. 80 Watts
Communications
Protocol:
Data Port
(Backboard):
Max. 19.2kbps Serial Communication (2
MODBUS:
Communication with DCS system to
RS232 ports)
obtain process signals
10Base-T/100Base-TEthernet (1 RJ45
HTTP:
Communication with WEB8000 Plus
port)
server to transmit data
Outputs
Backboard
Communications:
Front Panel LEDs:
RS232
Protocol
PWR LED:
Turn green when the DAU is powered on
Supported:
MODBUS
OK LED:
Turn green when the DAU is operating
Baud Rate:
Max. 19.2kbps
properly
Cable Length:
LAN LED:
Max.15meters
Turn green if the communication with
WEB8000 Plus is fine via Ethernet
Connector:
DB9 serial cable
COM1 LED:
Turn green if the communication is fine
Backboard
on COM1 port
Communications:
Ethernet, 10Base-T/100Base-TX
COM2 LED:
Turn green if the communication is fine
Protocol
on COM2 port
Supported:
HTTP
Cable Length:
K1-K4/K8 LEDs:
Max. 100 meters
Turn green if the corresponding rotating
speed signal is normal.
DYNAMIC
(1-24/48) LEDs:
Connector:
RJ45 (telephone jack style)
Data Collection
Page 2 of 5
Filter Performance: 2-step RC filter,
Rotating Speed Inputs:
cut-off frequency: 2Hz
Supports max. 8 rotating speed
signals, 1 rotating speed processing
cards, 8 A/D cards. The speed range
Data Processing and Analysis
support is 12~32000rpm.
Characteristics of Data Processing:
Input Type: Differential DC Input
Max. Sampling Speed:
Input Resistance: 400K
4096 samples/revolution
Input Range: 100V
A/D Resolution: 14bits
Linear Range: 18V
Max. Spectrum Lines: 3200
Filter Performance: Filters AC
User-configurable Synchronous
signals whose peak-peak value is
waveform sampling rates:
less than 20% of reference threshold
-1024 samples/rev for 8 revolutions
Volts.
- 512 samples/rev for 16/8
Phase Delay:
revolutions
<1 degree when
- 256 samples/rev for 32/16/8
rotating speed24000 RPM
revolutions
Rotating Speed Error <1.5rpm
-128 samples/rev for 64/32/16/8
Reference Threshold Voltage:
revolutions
Strong Rotating Speed Signal:
-64 samples/rev for
1V/2V/3V/4V/5V/6V/7V/8V;
128/64/32/16/8 revolutions
Weak Rotating Speed Signal:
-32 samples/rev for
0.2V/0.4V/0.6V/0.8V/1.0V/1.2V/1.
256/128/64/32/16/8 revolutions
4V/1.6V
Characteristics of Data analysis:
Vibration Inputs:
Supports max. 48 vibration signals,
Vibration waveform
12 vibration condition cards
Tacho phase signal waveform
Input Type: Displacement
Harmonic component vibration
/velocity/acceleration
Vibration harmonic component:
Input Resistance: 400K
1X,
Input Range: 100V
frequency bands
DC Linear Range: 20V
Different vibration type, ture p-p,
AC Linear Range: 4V/2V (Max.)
p-p, Rms, P;
DC Error:
<0.01V
Parameter value.
AC Error:
<0.008V
1P, 2X, 2P, 0.5X, two
Filter Performance: 1-step RC filter,
Data Storage
cut-off frequency: 0.1Hz, 1.6K Hz
Sensitive Monitoring Technology:
Distinguish frequency component
Process Inputs:
Support Max. 24 process signals, 2
of vibration: overall value, 1X, 1P,
process condition cards
2X, 2P, frequency band X, residual
Input Type: Differential DC Input
Input Resistance: 400K
Thresholds for event identification:
Input Range: 100V
-Overall value amplitude
Linear Range: 20V
-1X amplitude and phase
DC Error:
-2X amplitude and phase
<0.01V
Page 3 of 5
-Frequency band 1 amplitude
Max. relative humidity 80%
-Frequency band 2 amplitude
Main supply voltage fluctuations up
-Residual amplitude
to 10% of the nominal voltage
Automatic statistic learning to get
Rated pollution degree II
thresholds
Degree of protection IP 20
Characteristics of Data Storage:
Time-driven
CE Mark Directives
Event-driven
EMC Compliance
Data Type:
Management Group Issued
Real time data
Certificate No.:
EC10147
Trend data
Previous hour data
EC61326:
2006, Emissions
8, previous day data
9, previous week data
EN55011:
10, previous month data
1998+A1: 1999, Radiated and
11, previous year data
Conducted Emissions
13, startup/shutdown data
EN61000-3-2:
2000, Harmonic Current Emissions
14, sensitive monitoring data
3, 4, 5, black box data
EN61000-3-3:
1995+A1:2001,Voltage Fluctuations
6, before sensitive monitoring data
7, load throw-off data
EC61326:
2006, Immunity
Data Upload
EN61000-4-2:
Edition 1.2(2001), Electrostatic
Under good network condition;
Discharger Immunity Test
- Data upload frequency: up to
1s/time
EN61000-4-3:
2002, Radiated, Radio-Frequency,
- With high ratio data compress
Electromagnetic Field Immunity Test
technology, narrow net band
needed: 2kbps/channel.
EN61000-4-4:
2004, EFT/Burst Immunity
When network interrupted:
- Data be temporarily saved in
EN61000-4-5:
Edition 1.1:2001, Surge Immunity Test
the memory
- Memory space can last for 15
EN61000-4-6:
Edition 2.0(2003), Conducted Immunity
days.
- Upload be continued
EN61000-4-8:
automatically when network
Edition 1.1(2001), Power Frequency
recovered
Magnetic Field Immunity Test
Environmental Requirement
EN61000-4-11:
2004, Voltage Dips and Short
Interruptions Immunity Test
Indoor use
Altitude up to 2000m
Temperature 5 to 40
EC61010:
Safety requirement for electrical
Page 4 of 5
equipment for measurement, control,
and laboratory use
Physical Parameters
Dimensions
(Height Width Depth):
134mm452mm451mm
Weight:
17kg
Rack Installation
Requirement:
3U